Featured Article. Evaluation of Four Qualitative Third-Generation HIV Antibody Assays and the Fourth-Generation Abbott HIV Ag/Ab Combo Test

Size: px
Start display at page:

Download "Featured Article. Evaluation of Four Qualitative Third-Generation HIV Antibody Assays and the Fourth-Generation Abbott HIV Ag/Ab Combo Test"

Transcription

1 Submitted Accepted Evaluation of Four Qualitative Third-Generation HIV Antibody Assays and the Fourth-Generation Abbott HIV Ag/Ab Combo Test Frank H. Wians Jr., PhD, MT(ASCP), DABCC, FACB, 1 Holly A. Moore, AS, MT(AMT), 1 Donna Briscoe, BS, MT(ASCP), 1 Kathy M. Anderson, BA, MT(ASCP), 1 Pamela S. Hicks, BS, MLT(ASCP), MT(HEW), 2 Debbie L. Smith, AD, MLT(ASCP), 2 Tammy A. Clark, BS, MT(ASCP)SH, 3 Maggie M. Preston, BS, MT(HHS), MLT (ASCP), 4 Beckie Gammons, BS, MT(ASCP), 4 Cynthia S. Ray, BS, MT(ASCP), 5 Christine Bond, BS, MT(ASCP), 6 Joshua T. West, BS, MLS(ASCP) 7 (Departments of Pathology, 1 Baylor University Medical Center, Dallas, 2 Texas Health Arlington Memorial Hospital, Arlington, 3 Baylor All Saints, Fort Worth, 4 Texas Health Presbyterian Hospital, Dallas, 5 Texas Health Harris Methodist Hospital, Fort Worth, 6 Texas Health Presbyterian Hospital, Plano, 7 Texas Health Harris Methodist Hospital Southwest, Fort Worth, TX) DOI: /LM38NCGQ0GWJIMZV labmedicine.com September 2011 Volume 42 Number 9 LABMEDICINE 523

2 Abstract Objective: Compare the agreement between four third-generation rapid or automated human immunodeficiency virus (HIV) tests vs the fourth-generation Abbott ARCHITECT HIV Ag/Ab Combo assay (Combo Test) (Abbott Laboratories, Abbott Park, IL) in 7 clinical laboratories located throughout the Dallas-Fort Worth, TX, metroplex area. Methods: We tested a total of 220 specimens from the general population of patients being screened for HIV antibody status, including Western blot (WB) and HIV-1 RNA viral load by NAAT in patients with initially positive HIV screening tests. We performed a method precision study of the Combo Test and evaluated the agreement between the 4 comparative methods (CMs) against the Combo Test method (TM) using a 2 2 contingency table to determine values for positive percent agreement (PPA), negative percent agreement (NPA), and overall percent agreement (OPA) according to the guideline (EP12-A2) for the evaluation of qualitative assays published by the Clinical Laboratory Standards Institute (CLSI). Results: The Combo Test yielded consistent (reproducible) results when the HIV antibody (or p24 antigen) concentration was 20% away or more from C 50. In addition, values for PPA, NPA, and OPA for CM vs TM results were all 100%, or not significantly different from 100%. Conclusion: The high detection rate of the Combo Test, when coupled with its high diagnostic sensitivity and specificity, rapid TAT (<30 minutes) of test results using a fully automated instrument (ARCHITECT), and excellent agreement with HIV-1/2 antibody results by several commercially available HIV-1/2 screening tests and with WB and NAAT results, indicate that the Abbott HIV-1/2 Ab/Ag Test is a useful screening test in the identification of HIV-infected individuals. Keywords: human immunodeficiency virus, fourth-generation assay, comparative method, test method, agreement Corresponding Author Frank H. Wians, Jr., PhD, MT(ASCP), DABCC, FACB frank.wians@baylorhealth.edu Abbreviations HIV, human immunodeficiency virus; CM, comparative method; TM, test method; PPA, positive percent agreement; NPA, negative percent agreement; OPA, overall percent agreement; CLSI, Clinical Laboratory Standards Institute; AIDS, acquired immunodeficiency syndrome; UNAIDS, United Nations Programme on HIV/acquired immunodeficiency syndrome; CDC, Centers for Disease Control; gag, group-specific antigen; EIAs, enzyme immunoassays; CAP, College of American Pathologists; WB, Western blot, NAAT, nucleic acid amplification test; AMH, Arlington Memorial Hospital; BAS, Baylor All-Saints; BUMC, Baylor University Medical Center; THD, Texas Health Presbyterian Hospital, Dallas; THFW, Texas Health Harris Methodist Hospital; THP, Texas Health Presbyterian Hospital; THSW, Texas Health Harris Methodist Hospital Southwest; R, reactive; NR, nonreactive; EDTA, ethylenediaminetetraacetic acid; LL, lower limits; UP, upper limits; S/CO, signal-to-cutoff; %pos, percentage of positive; %neg, percentage of negative; FP, false positive; FN, false negative; POC, point-of-care; TAT, turnaround time The genetic variability of the human immunodeficiency virus (HIV) is illustrated by the number of groups, genetic subtypes, and the geographic distribution of these groups and subtypes throughout the world (Figure 1). Moreover, HIV-1 group M accounts for the majority of HIV infections in the United States, while HIV infection with a closely related virus, HIV-2, is rare in the United States but prevalent in West Central and West Africa (Figure 1). In addition, the prevalence of HIV-1 group O infection was thought to be low in the United States the first case of this infection was reported in the United States in There are marked differences in the global distribution of HIV infection (Figure 2). 2 The 2010 Global Report of the Joint United Nations Programme on HIV/acquired immunodeficiency syndrome (AIDS) (UNAIDS) indicated that in the 10-year period between 1999 and 2009, there was a 27% increase in the number of people living with HIV, with Sub-Saharan Africa still bearing the largest burden of HIV-infected individuals (Figure 2). 2 In addition, during this same time period, the number of new HIV infections and risk of HIV transmission to infants declined by ~20% with interventions (eg, use of highly active antiretroviral therapy) available in the United States and other developed countries. 2,3 The factors contributing to these encouraging findings include the cost-effective availability of antiretroviral drug therapy, more widespread use of condoms, and the introduction of significant educational programs and materials, especially in high-risk populations, related to the prevention of HIV infection and AIDS. Despite these successes, the Centers for Disease Control (CDC) has estimated that as many as 25% of newly infected individuals in the United States are unaware of their infection; these individuals account for more than 50% of new HIV infections; and, they remain concentrated among men who have sex with men and African Americans. 2,4 Fortunately, the recent availability of fourth-generation HIV assays, capable of identifying individuals with HIV infection earlier than prior generation assays (Figure 3), is likely to improve this statistic if the individuals who are unaware of their HIV status are screened using a fourth-generation HIV assay. Several studies have demonstrated that fourth-generation HIV assays can reduce the window (or seroconversion) period (ie, period during which a pathogen can be detected in a patient specimen by nucleic acid testing but not by serology, as the patient s immune system has not yet responded to the pathogen by producing measurable levels of antibodies) such that newly infected individuals can be identified, treated, and counseled. 5-8 The reduction in the window period afforded by fourth-generation HIV assays is due to the ability of these assays to detect not only HIV-1/2 antibodies but also the 24 kilodalton HIV group-specific (capsid or core) antigen (gag), p24, which can be detected as early as, and in some cases, even sooner than, 17 days post-hiv infection and up to 5-20 days sooner than third-generation HIV antibody enzyme immunoassays (EIAs) (Figure 3). This is not surprising when one considers that HIV antigen presentation must, obligatorily, precede antibody production to these antigens at levels detectable by HIV-1/2 antibody assays. Currently, there are no fourth-generation HIV rapid tests available and, because of the only recent availability of fully automated fourth-generation HIV assays, the most popular HIV screening assays are third-generation EIAs, 9-11 especially qualitative rapid tests such as the OraSure OraQuick ADVANCE (OraSure Technologies, Bethlehem, PA) and Clearview 524 LABMEDICINE Volume 42 Number 9 September 2011 labmedicine.com

3 Genetic Subtypes Geographic Distribution A Groups Group M (major) B Predominant Subtype Geographic Distribution Lower Frequency than HIV-1 HIV-1 C D F-H J K Group N (Non-M, Non-O) Group O (Outlier) Group P (Proposed) HIV-2 CRFs CRF01_AE Figure 1_Phylogeny of HIV groups and subtypes, and their geographic distribution. Color maps (top right): North America, Europe, South America, Japan, Australia, and Asia; (bottom center): West-Central (Congo Republic, Gabon, Cameroon, Central African Republic, Equatorial Guinea; Congo [Democratic Republic of Congo]; and, Angola]), and West Africa (ie, all countries shown except those considered West-Central Africa). CRFs, circulating recombinant forms. labmedicine.com September 2011 Volume 42 Number 9 LABMEDICINE 525

4 No Data <.1%.1%- <.5%.5%-<1% 1%-<5% 5%-<15% >15%-28% Figure 2_Global distribution of HIV disease (percentage of population) in Below the thick black dotted line on the continent of Africa is sub-saharan Africa. STAT-PAK HIV-1/2 (Inverness Medical, Princeton, NJ) assays. Among all laboratories participating in the most recent College of American Pathologists (CAP) external proficiency survey, AHIV-C, the percentage of all labs (n=~825) reporting results for the HIV-1/2 antibody methods shown in Table 1 decreased in the order: 49.8% (OraQuick ADVANCE) >43.4% (STAT-PAK) >2.5% (Other) >1.9% (Abbott, but not the Combo Test) >1.5% (Ortho) >0.9% (Multispot). Thus, these 2 rapid tests accounted for greater than 90% of the HIV tests used by all laboratories reporting HIV results for this CAP Survey. The performance characteristics, including diagnostic sensitivity and specificity, of the OraQuick and STAT-PAK HIV-1/2 assays, have been reported previously. 12 When evaluating qualitative assays, most laboratorians use a similar strategy to that used when evaluating quantitative assays select, without obvious bias, a minimum of patient samples for testing by the test method (TM) being evaluated and hope that there are no discordances (ie, 100% agreement) between results for the current or comparative method (CM) and the TM so the issue of deciding how many discordances can occur without invalidating the TM as an acceptable replacement for their CM is moot. It is important to note, however, that the discordances provide valuable information, especially if patient-specific information and/or the results of additional testing (eg, WB, quantitative HIV RNA level, and/ or NAAT) are combined to determine where truth lies (ie, was it the CM or the TM that provided the correct result?). Using this strategy, typically, the assumption is made that the precision, diagnostic sensitivity, and diagnostic specificity of the TM have been validated previously by other studies as consistent with the information for these analytical and clinical performance characteristics indicated in the manufacturer s product insert. Fortunately, the Clinical Laboratory Standards Institute (CLSI) has provided an evaluation protocol, EP12-A2, for evaluating qualitative test performance that provides statistical validity to the results of such an evaluation compared to the aforementioned strategy. 13 The CLSI protocol EP12-A2 recognizes 2 possible approaches when evaluating qualitative methods: 1 in which the comparator is diagnostic accuracy criteria and the other in which the comparator is not diagnostic accuracy criteria. There are marked differences in the information on TM performance provided by these 2 different approaches. Using the approach involving diagnostic accuracy criteria, applied to the evaluation of a qualitative HIV TM, samples from patients with known HIV disease, based on clinical findings and on results from a test (eg, HIV-1 RNA viral load) different from the TM being evaluated and with high diagnostic accuracy, are tested by the TM, while in the absence of diagnostic accuracy criteria, the performance of the qualitative TM is evaluated by testing these samples by both the CM and the TM. In both cases, a 2 2 contingency table (Figure 4) of the test results by the CM and TM is constructed; however, when diagnostic accuracy (ie, diagnostic sensitivity and specificity) criteria are used, this table allows the user to determine estimates for the following parameters of TM performance: Diagnostic sensitivity Diagnostic specificity Predictive value of a positive test result (PPV) Predictive value of a negative test result (NPV) Detection rate ([number of individuals with known disease positive by a CM/total number of individuals tested] 100) Moreover, a statistical test (eg, McNemar test) can be used to conclude whether or not there is a statistically significant difference between the (sensitivity, specificity) pairs of the 2 tests. An excellent example of the evaluation of qualitative HIV methods using diagnostic accuracy criteria is provided by the publication of Pandori and colleagues LABMEDICINE Volume 42 Number 9 September 2011 labmedicine.com

5 HIV-RNA p24 Ag EIA 3rd Gen. HIV Ab EIAs 1st Gen. HIV Ab EIAs 2nd Gen. Ab Assay HIV RNA (plasma) Anti-HIV Ab HIV p24 Ag Days Post-Infection 2nd Gen. Ab Assay 1st Gen. Ab Assay 3rd Gen. Ab Assay p24 Ag Assay 4th Gen. Ag/Ab Assay Using HIV Ag/Ab Combo assays closes the window to detect HIV infection up to 5 days. Figure 3_The effect of assay improvements with each generation of HIV assays on the window period for identifying individuals with acute or recent infection (adapted from a figure available from Abbott Diagnostics compiled from information provided in References 5-8). Table 1_All Participant Summary Data for Anti-HIV-1/2 From CAP Survey AHIV-C 2010 AHIV-11 AHIV-12 AHIV-13* AHIV-14 AHIV-15 Manufacturer Pos Neg Pos Neg Pos Neg Pos Neg Pos Neg Abbott** Bio-Rad Multispot Clearview STAT-PAK OraSure OQ ADVANCE Ortho Other Total Referee labs n.p. n.p. n.p. n.p n.p. n.p. n.p. n.p. *Due to lack of consensus among referee labs, participant labs were not graded on this challenge. **Not the Combo Test, but, rather, the OraQuick assay licensed previously by OraSure Technologies for distribution by Abbott. This license is no longer in force. CAP, College of American Pathologists; AHIV, anti-human immunodeficiency virus; Pos, positive; Neg, negative; n.p., not provided (by CAP). labmedicine.com September 2011 Volume 42 Number 9 LABMEDICINE 527

6 In the absence of diagnostic accuracy criteria, the same 2 2 contingency table can be used to obtain estimates for the following parameters of agreement between the CM and TM results: Positive percent agreement (PPA) Negative percent agreement (NPA) Overall percent agreement (OPA) The disadvantages of the evaluation protocol not using diagnostic accuracy criteria compared to the protocol that uses diagnostic accuracy criteria are: agreement is not a measure of correctness agreement measures are not universally applicable to all populations because of possible differences in the prevalence of HIV infection/disease between 2 different populations TM Pos Neg Total P, % Pos (a) 7 (c) 5 (a + c) PPA (TM) = [7/12] x 100 = 58.3% NPA (TM) = [39/39] x 100 = 100.0% OPA = [46/51] x 100 = 90.2% CM Neg (b) 0 (d) 39 (b + d) % Score Confidence Interval for Agreement: PPA Q 1,ppa = 2a = 2(7) = Q 2,ppa = [4(7)(5)/12] = 7.72 Q 3,ppa = 2(a+c) = 2(7 + 5) = Total (a + b) 7 (c + d) 44 LL = [(Q 1,ppa Q 2,ppa )/Q 3,ppa ] x 100 = [( ) / 31.68] x 100 = 31.68% UL = [(Q 1,ppa + Q 2,ppa )/Q 3,ppa ] x 100 = [( ) / 31.68] x 100 = 80.68% NPA Q 1,npa = 2d = 2(39) = Q 2,npa = [4(0)(39)/(0 +39)] = 3.84 Q 3,npa = 2(b+d) = 2(0+39) = LL = [(Q 1,npa Q 2,npa )/Q 3,npa ] x 100 = [( ) / 85.68] = 91.04% UL = [(Q 1,npa + Q 2,npa )/Q 3,npa ] x 100 = [( ) / 85.68] = % 51 P, % OPA Q 1,opa = 2(a+d) = 2(7 + 39) = Q 2,opa = [4(7+39)(0+5) / 51] = 9.17 Q 3,opa = 2n = 2(51) = LL = [(Q 1,opa Q 2,opa )/Q 3,opa ] x 100 = [( ) / ] x 100 = 79.02% UL = [(Q 1,opa + Q 2,opa )/Q 3,opa ] x 100 = [( ) / ] x 100 = 95.74% McNemar Test On Difference Between Paired Proportions (P, %): Difference between paired proportions = (P pos,cm - P pos,tm ) = ( ) = 9.8 = (P neg,tm P neg,cm ) = ( ) = 9.8 = n McNemar test (pair proportions) Data Pos Neg Pos % Neg % 23.5% 76.5% Results Difference 95% CI Exact probability Comment CM = OraQuick TM = Combo Test 9.80% 0.43% to 9.8% P = Pos, positive; Neg, negative; CM, comparative method; TM, test (or candidate) method; PPA, percent positive agreement; NPA, negative percent agreement; OPA, overall percent agreement; LL, lower limit; UL, upper limit. Figure 4_2 2 Contingency Table for BUMC Data. 528 LABMEDICINE Volume 42 Number 9 September 2011 labmedicine.com

7 the usual measures of test performance (ie, diagnostic sensitivity, specificity, PPV, and NPV) cannot be determined Despite these disadvantages, the CLSI protocol, EP12-A2, indicates: reporting how often the candidate method and comparative method (CM) agree could be useful. 13 Because of the potential advantages of the decreased window period of fourth-generation over earlier generation HIV screening assays, 15 we undertook an evaluation of the Abbott ARCHITECT (Image 1) fourth-generation HIV Ag/Ab Combo (or Combo Test) assay against currently used HIV screening assays in an unselected (ie, no information was available on the HIV status of the individuals tested or the results of any Table 2_Comparison of Qualitative HIV Assay Characteristics Characteristic 1 OraQuick ADVANCE HIV 1/O/2 Enhanced HIV 1/2 STAT-PAK Anti-HIV 1+2 HIV Ag/Ab Combo Test Manufacturer Orasure Siemens Chembio Ortho Abbott Assay generation 3 rd 3 rd 3 rd 3 rd 4 th HIV Ab subtype detected 1 and 2 1 (inc Gp O) and 2 1 and 2 1 and 2 1 (Gps O and M) and 2 HIV-1 Ag detected n.d. n.d. n.d. n.d. p24 Device/instrument SU-POCT ADVIA Centaur SU-POCT VITROS ARCHITECT Sample type OF, P 2, WB 3 S, P 4 S, P 5, WB 3 S, P 4 (S, P) 6 Sample volume, ml n.s Immunoassay type LFIC-PAG ABMCI LFIC-CGD IBA CMIA Solid phase Nitrocellulose membrane PMPs Nitrocellulose membrane Plastic well PMPs Capture Ag s/ab s/ Synthetic HIV env Ag s SB-([HIV 1 and 2 rag s] + CG-(ABP+HIV-1&2 Ag s) rag s 7 HIV-1&2 Ag s + HIV-1 Substance Gp O peptide Ag) p24 mouse Mcl Ab Signal molecule/antigen Protein-A gold AE-([HIV 1 and 2 rag s] + CGD particles Luminol derivative AE-conjugates 8 Gp O peptide Ag) Number of calibrators n.a. 2 n.a. 1 1 Number of controls Reproducibility, %CV range 9 Between run Between testing site Read time or TTFR, min Diagnostic sensitivity, % 99.6 ( ) ( ) 99.7 ( ) ( ) Diagnostic specificity, % 99.9 ( ) 99.9 ( ) 99.9 ( ) 99.6 ( ) ( ) 1 All data from manufacturer s product insert. 2 EDTA plasma. 3 EDTA, sodium heparin, sodium citrate, or ACD Solution A whole blood (WB) from venipuncture or fingerstick. 4 EDTA, sodium heparin, sodium citrate, or ACD Solution A plasma. 5 EDTA, sodium heparin, or sodium citrate plasma; WB from venipuncture or fingerstick. 6 Serum in glass or plastic tube with or without serum separator; tripotassium (K 3 )- or disodium (Na 2 )-EDTA plasma in glass; Na 2 -EDTA plasma with or without gel separator, sodium heparin, or lithium heparin with gel separator in plastic tube. 7 HIV-1 Env 13, HIV-2 Env AL, HIV-1 ENV 10, and HIV-1 p24 antigens. 8 5 Conjugate types: 1) AE-(HIV-1 Ag); 2) AE-(HIV-2 Ag); 3) AE-(HIV-1 Group M Ag); 4) AE-(HIV-1 Group O Ag); 5) AE-(HIV-1 p24 Ab). Ab, antibody; SU-POCT, single use-point-of-care test; OF, oral fluid; P, plasma; WB, whole blood; n.s., not stated; LFIC, lateral flow immunochromatography; PAG, protein-a gold; CGD, colloidal gold dye; HIVenv Ag s, antigens representing HIV envelope epitopes; Ab s, antibodies; n.d., not detected; n.a., not applicable; inc, including; Gp(s), group(s); ABMCI, antigen bridging microparticle chemiluminescent immunoassay; PMPs, paramagnetic particles; SB, streptavidin-biotinylated; rag s, recombinant antigens; CG-ABP+HIV-1/2 Ag s, colloidal gold dye particles-(antigen binding protein+hiv-1&2 antigens) conjugate; IBA, immunometric bridging assay; CMIA, chemiluminescent microparticle immunoassay; Mcl, monoclonal; AE, acridinium ester; TTFR, time to first result. 9 Variability of assay performance for all samples/replicates per sample/testing sites/reagent lots/days/techs: 5/27/3/3/3/9 (OraQuick); 12/5/3/3/6/3 (HIV 1/O/2 Enhanced); 16/5/3/3/3/3 (HIV1/2 STAT-PAK); 4/2/3/3/20/1 (Anti-HIV 1+2); 13/4/3/3/5/3 (HIV Ag/Ab Combo). Table 3_Summary of Testing Sites and Number of Patients Specimens Tested by Comparative Method and Test Method, Western Blot, and Quantitative or Qualitative NAAT for HIV RNA No. of Patients Specimens Tested Testing Site HIV Assay (CM) Type of Specimen Tested CM and TM WB NAAT AMH OraQuick ADVANCE K 3 -EDTA-P 21 1 BAS Centaur HIV 1/O/2 Enhanced Serum BUMC OraQuick ADVANCE K 3 -EDTA-P THD Centaur HIV 1/O/2 Enhanced K 3 -EDTA-P THFW Vitros Anti-HIV 1+2 Serum 13 THP OraQuick ADVANCE Serum 25 1 THSW OraQuick ADVANCE K 3 -EDTA-P 21 Total labmedicine.com September 2011 Volume 42 Number 9 LABMEDICINE 529

8 HIV testing, eg, Western blot [WB], HIV-1 viral load, and/or nucleic acid amplification testing [NAAT] performed prior to the evaluation) population of individuals at 7 sites participating in this evaluation. The Combo Test was the first fourth-generation HIV diagnostic assay approved by the U.S. Food & Drug Administration (June 18, 2010). We determined the PPA, NPA, and OPA for each CM vs the Combo TM. Materials and Methods Sites The laboratories that participated in this study are located in 1 of the following 7 sites: 1) Texas Health Arlington Memorial Hospital (AMH), Arlington; 2) Baylor All Saints (BAS), Fort Worth; 3) Baylor University Medical Center (BUMC), Dallas; 4) Texas Health Presbyterian Hospital (THD), Dallas; 5) Texas Health Harris Methodist Hospital (THFW), Fort Worth; 6) Texas Health Presbyterian Hospital (THP), Plano; and, 7) Texas Health Harris Methodist Hospital Southwest (THSW), Fort Worth, TX. HIV Assays and Type/Number of Specimens Tested The characteristics of the qualitative HIV assays used in this study are summarized in Table 2. The HIV assay used at each participating site, and the type and number of patients specimens tested using these assays are shown in Table 3. These assays included: OraQuick ADVANCE Rapid HIV-1/2 Antibody Test (OraSure Technologies), HIV 1/O/2 Enhanced (Siemens, Tarrytown, NY), Anti-HIV 1+2 (Ortho-Clinical Diagnostics, High Wycombe, U.K.), and HIV Ag/Ab Combo (Abbott, Wiesbaden, Germany). All patients specimens were selected with no obvious bias and obtained by fingerstick or venipuncture. Specimens obtained by venipuncture were collected using 1 of the collection tube types indicated in the manufacturer s product insert for each HIV assay. All specimens were tested according to the manufacturer s instructions using the assays shown in Table 3. At each testing site, all samples were assayed by both the CM specific to each site and the candidate or TM (ie, the Abbott HIV Ag/Ab Combo Test) within 2 hours of specimen collection, or after thawing (THD), in a single run by a single operator. According to the manufacturer s product insert for the ADVIA Centaur EHIV assay, when 10 specimens were subject to 6 freeze/thaw cycles, no clinically significant differences were observed. Test results were reported as reactive (R) or nonreactive (NR) based on the information provided in the manufacturer s product insert. A total of 220 test results were generated across 7 testing sites and all assays. In addition, 27 specimens across all sites were assayed by WB, and 14 specimens were assayed for HIV-1 ribonucleic acid (RNA) by quantitative or qualitative NAAT (Table 3). WB Testing Western blot confirmatory testing of positive immunoassay screening tests was performed either in-house or by an external, CAP-accredited referral lab (Med Fusion, Lewisville, TX) using the Bio-Rad (Redmond, WA) Genetic Systems (GS) HIV-1 WB assay. Interpretation of patients serum or (ethylenediaminetetraacetic acid [EDTA]/heparin/sodium citrate/cpda-1/ ACD)-plasma was based on the pattern of bands observed on the immunoblot (Table 4). Quantitative and Qualitative HIV-1 RNA Testing by NAAT Quantitative or qualitative HIV-1 RNA testing was performed using EDTA-plasma obtained from whole blood within 24 hours of collection by centrifugation at g for 20 minutes at room temperature. Quantitative HIV-1 RNA testing was performed at Med Fusion, Lewisville, TX, using the AmpliPrep/COBAS TaqMan HIV-1 Test (Roche Diagnostics, Indianapolis, IN). According to the manufacturer s product Table 4_Interpretation of Western Blot Patterns All Bands Pattern Interpretation 2 of the following 3 major bands present and at least as intense as the low positive control gp 120 band: 1. (gp160 and/or gp120)=1 major band 2. Broad and diffuse gp41 band 3. p24 Positive 1 band present but band(s) present do not meet the criteria for positive Indeterminate No bands present Negative gp, glycoprotein; env, envelope; pol, polymerase; gag, group-specific antigen. 530 LABMEDICINE Volume 42 Number 9 September 2011 labmedicine.com

9 insert, the linear range of this test is: 20 cp/ml to cp/ ml. Qualitative HIV-1RNA testing was performed using the Aptima HIV-1 RNA Qualitative Assay (Gen-Probe, San Diego, CA). According to the manufacturer s product insert, the analytical sensitivity of this assay is 30 cp/ml (ie, 98.5% of samples containing 30 cp/ml of HIV-1 RNA yield positive results). Method Precision Using patients serum samples and the protocol described in EP12-A2, 13 we performed a method precision study on the TM by preparing 3 samples: 1 with concentration at C 50 ; 1 with concentration 20% below C 50 (C -20 ); and, 1 with concentration 20% above C 50 (C +20 ). The C 50 sample contains an analyte concentration near the assay cutoff value such that this sample yields 50% positive results and 50% negative results on testing of many replicates of this sample. Each sample was assayed by the TM 30 times in the same run. Method precision was considered acceptable if the C -20 to C +20 concentration range bounded (contained) the C 5 -C 95 interval. The C 5 -C 95 interval corresponds to the concentrations around the cutoff value such that observed results at concentrations outside this interval are consistently negative (ie, concentrations <C 5 ) or consistently positive (ie, concentrations >C 95 ). The true percentage of positive results lying within an estimated C 5 -C 95 confidence interval varies with the number of samples (n) tested and approaches 95% as n increases. 13 Statistical Analysis of Agreement Between CMs and TM All data were analyzed according to the guidelines for the evaluation of qualitative test performance provided in EP12-A2. 13 For each comparison of CM results vs TM results, a 2 2 contingency table was constructed and values for PPA, NPA, OPA, and the lower (LL) and upper (UL) limits of the 95% score confidence interval were calculated as shown in Figure 4. When results by the CM are based on diagnostic accuracy criteria, PPA and NPA are the diagnostic sensitivity and specificity, respectively, of the TM. 13 The use of score confidence limits, in lieu of exact confidence limits, is recommended for reasons beyond the scope of this article. 13 In addition, the McNemar test was performed using MedCalc software (MedCalc Software, Mariakerke, Belgium) to determine if there was a significant difference between paired proportions at a P value of Results TM Precision Signal-to-CutOff (S/CO) values (mean ± SD) for the 30 replicates of the C -20, C 50, and C +20 samples using the TM were: 0.79 ± 0.04, 0.99 ± 0.04, and 1.19 ± 0.03, respectively. The percentage of positive and negative (%pos, %neg) results among the 30 replicates by the TM for samples C -20, C 50, and C +20 were: (0.0, 100.0), (40.0, 60.0), and (100.0, 0.0), respectively. Therefore, the largest confidence that can be achieved where the lower end of the one-sided confidence interval for both the C -20 and C +20 samples is at or above 95% is: % = 78.68%, where is obtained from a table comparing confidence (probability scale) as a function of N (number of replicate measurements) and the fraction (as a %) of positive (or negative) results (see Appendix, p39 in Reference 13). A confidence value of 79% indicates that the Combo Test yields Image 1_Abbott Architect i2000sr instrument (operated by Holly Moore, MT(ASCP), BUMC). consistent (reproducible) results when the HIV antibody (or p24 antigen) concentration is 20% away or more from C 50. Agreement Between CMs and TM The PPA, NPA, and OPA for OraQuick results vs results by the Combo Test HIV assay were all 100%, or not significantly different from 100%, at 3 of 4 sites (AMH, THSW, THP) using the OraQuick HIV assay, while at 1 site (BUMC), PPA and OPA were less than 100% (Figure 5). However, the difference between paired proportions was not statistically significant (Figure 4). The PPA for the results of the OraQuick vs results by the Combo Test HIV assays at BUMC was the lowest (58.3%) for all HIV assays at all sites because there were 5 samples (Nos. 7 to 11, Table 5) that were initially R by OraQuick, weakly or NR on repeat testing, NR by the Combo Test HIV assay, and WB negative (ie, no bands detected). In addition, Combo Test results for all 16 discordant (ie, CM and TM results disagreed) samples across all sites were consistent with WB and/or NAAT results (Table 5). Among the 2 sites (BAS and THD) using the Centaur HIV assay, values for the 3 measures of agreement (PPA, NPA, and OPA) between this assay vs the Combo Test HIV assay were: BAS: 84.2%, 93.9%, and 90.4%; THD: 80.0%, 100.0%, and 89.2% (Figure 5). The Vitros HIV assay at site THFW provided 100% agreement for PPA, NPA, and OPA vs results by the Combo Test HIV assay (Figure 5). In Figure 5, a table is included indicating the number of potential false positive (FP) and false negative (FN) results that cause values for PPA, NPA, and OPA shown in the figure to vary from 100%. The data in the table are referred to as potential FP or FN results because a diagnostic accuracy criterion was not used to definitively identify the individuals in this study as with or without HIV infection prior to testing with the comparative and TMs. labmedicine.com September 2011 Volume 42 Number 9 LABMEDICINE 531

10 Potential Agreement With TM (%) AMH- OQ BAS- Centaur BUMC- OQ Site AMH BAS BUMC THD- Centaur Site-HIV Method THFW- Vitros Discussion Differences between values for PPA, NPA, or OPA for comparative and TMs between sites (eg, OraQuick ADVANCE HIV-1/2 vs Combo Test Results at BUMC compared to the 3 other sites [AMH, THP, THSW]; Figure 5) are most likely due to sample size and differences in the prevalence of HIV disease THP- OQ THSW- OQ PPA NPA OPA THD THFW THP THSW FP FN Figure 5_Percentage agreement between individual site-specific HIV assays vs the test method (TM; Abbott HIV Ag/Ab Combo Test) (error bars represent the lower and upper limits [UL] of the 95% score confidence interval). Abbreviations are same as those shown in Tables 2 and 3. FP, false positive; FN, false negative; PPA, positive percent agreement; NPA, negative percent agreement; OPA, overall percent agreement. in the unselected (ie, not based on known HIV status by a gold standard method [eg, HIV-1 RNA viral load]) population of patients being screened for HIV antibodies at each individual site. Among the 7 sites in this study, BUMC is the largest (1025 licensed beds; 39,000 admissions/y; 67,000 ED visits/y; and, 241,000 outpatient visits/y), with an active epidemiology and infection control department, and it is likely that this site screens patients with a higher prevalence of HIV disease and a greater heterogeneity of HIV antibody and antigen levels in their blood than the patients screened at the other 6 sites. Estimates for diagnoses of HIV infection (n=1246) in 2008 for the Dallas (the location of BUMC) and Fort Worth metroplex area, and adjacent cities (the location of the other 6 sites participating in this study), were 974 (78.2%) and 272 (21.8%), respectively. 16 This heterogeneity resulted in 5/51 samples tested at BUMC demonstrating initially R OraQuick results that were all Combo Test NR and WB negative (Table 5). Fortunately, these patients would have been correctly classified as screen negative because all initially R specimens are repeat tested, and if the initial and repeat results disagree, the specimens are sent for WB testing. The Combo Test would have correctly classified these 5 patients as screen negative without the need for repeat screening or confirmatory WB testing (Table 5). Table 5_Results for Samples With Discordant Comparative and Test Method Results on Which WB and/or HIV-1 RNA Testing Was Performed Interpretation Quantitative (cp/ml) or Comp HIV Result by Repeat Avg S/CO Ratio of Combo WB Qualitative (Pos/Neg) No. Site ID Method Comp Method Result by CM by Combo Test Test Results WB Band Pattern HIV-1 RNA by NAAT 1 AMH OraQuick NR 4.64 R 1.7 x BAS Centaur R 0.09 NR Neg No bands present 3 BAS Centaur R 0.27 NR Neg No bands present 4 BAS Centaur R 0.08 NR Neg No bands present 5 BAS Centaur NR 6.04 R Neg No bands present Pos 6 BAS Centaur NR 9.65 R Neg No bands present Pos 7 BUMC OraQuick R NR 0.18 NR Neg No bands present 8 BUMC OraQuick R WR 0.12 NR Neg No bands present 9 BUMC OraQuick R WR 0.17 NR Neg No bands present 10 BUMC OraQuick R NR 0.16 NR Neg No bands present 11 BUMC OraQuick R WR 0.15 NR Neg No bands present 12 THD Centaur R 0.22 NR Neg No bands present 13 THD Centaur R 0.20 NR Neg No bands present 14 THD Centaur R 0.16 NR Neg No bands present 15 THD Centaur R 0.30 NR Neg No bands present 16 THP OraQuick NR R Pos WB, Western Blot (Pos, Neg, or Ind); cp/ml, copies/milliliter; NAAT, nucleic acid amplification test; Comp, comparative; S/CO, signal-to-cutoff ratio; AMH, Texas Health Arlington Memorial Hospital; BAS, Baylor All-Saints; BUMC, Baylor University Medical Center ; THD, Texas Health Presbyterian Hospital; THP, Texas Health Presbyterian Hospital ; R, reactive; NR, nonreactive; WR, weakly reactive; Ind, indeterminate; Pos, positive; Neg, negative. 532 LABMEDICINE Volume 42 Number 9 September 2011 labmedicine.com

11 The Vitros HIV antibody test provided 100% agreement with Combo Test results on the relatively small number of samples tested at site THFW (n=13), while OPA for the Centaur method was 89.2% (95% Score CI: 75.3% to 95.7%) at site THD (n=37) and 90.4% (95% Score CI: 79.4% to 95.8%) at site BAS (Figure 5). The Centaur method would be expected to yield similar results as the Combo Test because among the HIV screening tests we evaluated, the principle of these methods was most similar between the Centaur and Architect methods, including the ability to detect Group O HIV-1 antibodies (Table 2 and Figure 1). The weakness of the WB confirmatory test in correctly classifying patients, especially those with acute or recent HIV infection, as HIV positive or negative, was illustrated in the study by Pandori and colleagues. 14 In this study, among 5 patients with acute or recent HIV infection, based on positive HIV-1 RNA findings (range of quantitative HIV-1 RNA levels: 335 cp/ ml to cp/ml), 4 of these patients had NR OraQuick Advance results, indeterminate WB results, and positive HIV-1 viral loads of 30,734 cp/ml to cp/ml. In addition, the remaining patient had a R OraQuick result, an indeterminate WB result, and a positive HIV-1 RNA viral load (4,571,787 cp/ml). The Combo Test was R on samples from all 5 of these patients (Table 6). The ability of the Combo Test to yield R results in samples that are NR by a point-of-care (POC) rapid test but contain high levels of HIV-1 RNA highlights the enhanced clinical utility of the Combo Test in correctly classifying HIV status on the basis of this single test (Table 6; samples 1 to 4). The data in Table 6 underscore the heterogeneity of the time course of antibody production in individuals with significant HIV viremia that can cause third-generation HIV assays, which detect only HIV antibodies, to be NR and fourth-generation assays, which detect both antibodies and HIV p24 antigen, to be R. It is likely that this advantage of fourth-generation HIV screening assays, when coupled with the reduced window period provided by Table 6_Summary of Quantitative HIV-1 Results * No. OraQuick-FS Avg S/CO Ratio by Combo Test Combo Test Result WB Result HIV-1 RNA Viral Load, cp/ml 1 NR 2.05 R Ind 30,734 2 NR R Ind 500,000 3 NR R Ind 413,186 4 NR R Ind 10,000,000 5 R R Ind 4,571,787 *Data from Pandori et al. 14 OraQuick-FS, OraQuick ADVANCE HIV 1/2 Rapid Test-Fingerstick whole blood sample tested; Avg, average; S/CO, signal-to-cutoff ratio; WB, Western blot; cp, copies; R, reactive; NR, nonreactive; Ind, indeterminate. A B A1: 4th Generation HIV-1/2 IA A1: 1st Generation HIV-1 IA A1+ A1( ) A1+ 2 Tests A1( ) A2 HIV-1/HIV-2 discriminatory assay Result: ( ) for HIV-1, HIV-2 antibodies, and p24 antigen A2 WB Confirmatory Test Result ( ) HIV-1+ HIV-2+ HIV-1&2( ) Positive Negative HIV-1 antibodies detected: initiate care viral load testing HIV-2 antibodies detected: initiate care A3: NAAT Result (+) Result ( ) NAAT+ NAAT ( ) Acute HIV infection: initiate care HIV-1 ( ) Figure 6_HIV screening algorithms: A) algorithm recommended by U.S. Public Health Service in 1989; B) proposed algorithm by CDC APHL HIV Diagnostics Conference, A1, test-1; A2, test-2; A3, test 3; HIV, human immunodeficiency virus; IA, immunoassay; WB, Western blot; +, reactive or positive; (-), nonreactive or negative; NAAT, nucleic acid amplification test. labmedicine.com September 2011 Volume 42 Number 9 LABMEDICINE 533

12 Table 7_Two-Test Comparisons of OraQuick, STAT-PAK, and Combo Test HIV Results* 2-Test Comparison PPA (%) NPA (%) OPA (%) OraQuick vs Combo Test ( ) 13.2 ( ) 28.1 ( ) STAT-PAK vs Combo Test ( ) 13.2 ( ) 28.1 ( ) OraQuick vs STAT PAK ( ) ( ) ( ) *Data adapted from Pandori and colleagues 14 (values in parentheses are the 95% score confidence interval). PPA, positive percent agreement; NPA, negative percent agreement; OPA, overall percent agreement. these assays, 5-8,14,15 underlies the central role of these assays in the proposed CDC HIV screening algorithm (Figure 6B). The difficulty in achieving 100% agreement for HIV-1/2 antibody results for currently available POC and non-poc TMs is illustrated further by the all-participant summary results from a recent CAP Survey (AHIV-C, 2010). Among laboratories participating in this Survey that reported results for the Clearview STAT-PAK (n=358 labs) and OraQuick ADVANCE (n=411 labs) rapid TMs on survey challenge AHIV-13, 306/358 labs (85.5%) reported negative results by the STAT-PAK method, while 372/411 labs (90.5%) reported positive results by the OraQuick method (Table 1). However, referee results (n=18), by an unspecified assay/method on survey sample AHIV-13, yielded 13/18 (72.2%) positive results and 5/18 (27.8%) negative results. As expected, values for PPA, NPA, and OPA are different for any 2-test comparison between HIV antibody results obtained on samples from a selected vs an unselected population of individuals. This point is illustrated using data from the study by Pandori and colleagues 14 on selected samples (n=64) from patients with acute or recent HIV infection, as indicated by positive (>20 cp/ml) quantitative HIV-1 RNA results (range of values: 335 cp/ml to cp/ml). Using these data, we determined values for PPA, NPA, and OPA for each 2-test comparison: OraQuick vs Combo Test; STAT-PAK vs Combo Test; and, OraQuick vs STAT-PAK (Table 7; Note: Pandori and collagues 14 did not evaluate the other HIV antibody methods used in our study). As shown in Table 7, values for PPA, NPA, and OPA were all 100% for the OraQuick vs STAT-PAK method; however, when results for both of these methods were compared to those of the Combo Test, PPA was 100% for both tests, but NPA and OPA were considerably smaller: 13.2% and 28.1% for both tests, respectively (Table 7). In addition, in the study by Pandori and colleagues, 14 the detection rate (ie, [no. of positive results by TM/total No. of known patients with acute or recent HIV infection tested] 100) for the OraQuick, STAT-PAK, and Combo TMs was: 17.2%, 17.2%, and 89.1%, respectively. When using a laboratory test to screen an unselected population for a disease (eg, HIV) for which good to excellent treatment options are available that have an optimum impact on disease course and long-term outcome, the earlier the patient with disease is detected, the test should have high diagnostic sensitivity (ie, Table 8_Potential Causes of False-Negative and False-Positive Results in HIV Assays * False Positive False Negative Any of the following conditions, infections, findings, or lifestyle choices Pre-analytical or analytical causes Hemodilution Mislabeling of samples or wells Variability in test kits, including the inability to detect HIV-1 subtype O individuals using an HIV kit that does not detect subtype O antibodies Masking of HIV antigenic determinants or insufficient antigenic determinants for detection By the HIV assay Pipetting error Assay inhibition due to: Powder from powdered gloves Storing samples in serum separator tubes Biological, pathologic, or pharmacologic determinants Window (preseroconversion) period status Delayed antibody synthesis in infants Diminished immune response in patients With HIV 1-related immune dysfunction On immunosuppressive therapy With concurrent EBV or CMV infection Congenital or drug-induced hypogammaglobulinemia Formation of antigen-antibody complexes Renal failure Alpha-interferon therapy in hemodialysis patients Flu or flu vaccination HSV I or II URI Recent viral infection or vaccination against HBV or with tetanus toxoid Pregnancy in multiparous women Malaria RA Blood transfusions Myeloma Hemophilia Stevens-Johnson syndrome Organ transplantation ALD or AH Autoimmune disease Receptive anal sex Anti-lymphocyte, -collagen, -smooth muscle, -nuclear, -mitochondrial, -parietal cell, -HAV IgM, -HBc IgM, or -polystyrene antibodies Heat-treated, lipemic, hemolyzed, or icteric specimens Normal human ribonucleoproteins Visceral leishmaniasis Fungal infections, including TB and Mycobacterium avium *Adapted from Reference 25. HIV, human immunodeficiency virus; EBV, Epstein-Barr virus; CMV, cytomegalovirus; HSV, herpes simples virus; URI, upper respiratory tract infection; HBV, hepatitis B virus; RA, rheumatoid arthritis; ALD, alcoholic liver disease; AH, alcoholic hepatitis; HAV, hepatitis A virus; HBc, hepatitis B core; TB, tuberculosis. 534 LABMEDICINE Volume 42 Number 9 September 2011 labmedicine.com

13 few FNs). However, when FPs are also undesirable because of the emotional effect on the patient, there is a premium on high diagnostic specificity (ie, few FPs) as well. The study by Pandori and colleagues 14 demonstrated that the Combo Test had the highest detection rate for acute and recent HIV disease among all CMs tested, including the OraQuick and STAT-PAK tests; other studies have shown high diagnostic sensitivity and specificity for the Combo test; and, the turnaround time (TAT) of test results by all 3 of these tests is approximately the same (~20-28 minutes; Table 2). It is important to note that immunoassays reported to have 100% diagnostic sensitivity and specificity or 100% agreement between a CM and a TM are not perfect (ie, there are never any falsepositive or false-negative results) because all immunoassays, regardless of the generation of the assay, are prone to falsepositive or false-negative results for a variety of reasons summarized in Table 8. There has been at least 1 report indicating that fourthgeneration HIV assays, albeit not the Combo Test, may still be negative in patients with HIV seroconversion illness (ie, evolving WB reactivity and a positive, often markedly elevated, viral load in a patient with symptoms of HIV infection). 22 Clearly, such reports validate the difficulty of any fourth-generation HIV assay in reducing the window period to the time of HIV RNA detectability. To resolve this problem, Stekler and colleagues 23 have suggested that, NAAT be integrated into HIV testing programs that serve populations that undergo frequent testing and that have higher rates of HIV acquisition, particularly if rapid HIV antibody testing is employed. It remains to be proven if this same statement applies to fourth-generation HIV assays, despite the suggestion in some studies, including the study by Pandori and colleagues, 14 based on negative or indeterminate WB results on samples with positive fourth-generation HIV assay and NAAT results, that this may be the case. Recently, a new algorithm has been proposed to replace the algorithm that has changed little since it was recommended by the U.S. Public Health Service in 1989 (Figure 6A). 24 This new algorithm takes advantage of the higher diagnostic accuracy (ie, sensitivity and specificity) of a fourth-generation HIV screening assay and NAAT confirmatory testing (Figure 6B). It should be noted, however, that the WB is not being abandoned completely, as it is well recognized that WB testing will continue to be useful in: Confirming long-standing HIV infection Evaluating diagnostic dilemmas Testing vaccine recipients for antibody response Monitoring patients with end-stage AIDS The high detection rate of the Combo Test for individuals with acute or recent HIV infection, when coupled with its high diagnostic sensitivity and specificity, rapid TAT (<30 minutes) of test results using a fully automated instrument (ARCHITECT), excellent agreement with HIV-1/2 antibody results by several commercially available HIV-1/2 screening tests (Figure 5), including 1 (OraQuick ADVANCE) of the 2 most popular tests among laboratories participating in the CAP AHIV Survey (Table 1), and with WB and/or NAAT results among discordant samples (Table 5) indicate that the Abbott HIV-1/2 Ab/Ag Test is a useful screening test in the identification of HIV-infected individuals. LM Acknowledgements: The authors thank Abbott Diagnostics for providing HIV Ag/Ab Combo Test kits for evaluation. 1. Marquez C, Zetola NM, Klausner JD. HIV testing: An update. MLO Med Lab Obs. 2008;40:12-14,16,18 passim. 2. Global report: UNAIDS report on the global AIDS epidemic ISBN (NLM classification: WC 503.4). 3. Anderson BL, Cu-Uvin S. Pregnancy and optimal care of HIV-infected patients. Clin Infect Dis. 2009;48: Hall HI, Song R, Rhodes P, et al. Estimation of HIV incidence in the United States. JAMA. 2008;300: Fiebig EW, Wright DJ, Rawal BD, et al. Dynamics of HIV viremia and antibody seroconversion in plasma donors: Implications for diagnosis and staging of primary HIV infection. AIDS. 2003;17: Owen SM, Yang C, Spira T, et al. Alternative algorithms for human immunodeficiency virus infection diagnosis using tests that are licensed in the United States. J Clin Microbiol. 2008;46: Kleinman SH, Lelie N, Busch MP. Infectivity of human immunodeficiency virus-1, hepatitis C virus, and hepatitis B virus and risk of transmission by transfusion. Transfusion. 2009;49: Patel P, Mackellar D, Simmons P, et al. Detecting acute human immunodeficiency virus infection using 3 different screening immunoassays and nucleic acid amplification testing for human immunodeficiency virus RNA, Arch Intern Med. 2010;170: Pumarola T, Freeman J, Saxton E, et al. Performance evaluation of the ADVIA Centaur HIV Ag/Ab Combo assay. J Virol Methods. 2010;170: Schappert J, Wians FH Jr, Schiff E, et al. Multicenter evaluation of the Bayer ADVIA Centaur HIV 1/O/2 enhanced (EHIV) assay. Clin Chim Acta. 2006;372: Thompson MP, Hidalgo D, Kim A, et al. The VITROS immunodiagnostics products anti-hiv 1+2 assay: A rapid and reliable method for HIV screening. LabMed. 2011;42: Delaney KP, Branson BM, Uniyal A, et al. Evaluation of the performance characteristics of 6 rapid HIV antibody tests. Clin Infect Dis. 2011;52: CLSI. User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline. 2nd ed. EP12-A2. Wayne, PA: CLSI; 2008;28: Pandori MW, Hackett J Jr, Louie B, et al. Assessment of the ability of a fourthgeneration immunoassay for human immunodeficiency virus (HIV) antibody and p24 antigen to detect both acute and recent HIV infections in a high-risk setting. J Clin Microbiol. 2009;47: Speers D, Phillips P, Dyer J. Combination assay detecting both human immunodeficiency virus (HIV) p24 antigen and anti-hiv antibodies opens a second diagnostic window. J Clin Microbiol. 2005;43: Centers for Disease Control. Diagnoses of HIV infection and AIDS in the United States and dependent areas, HIV Surveillance Report. 2008;20: Stekler JD, Swenson PD, Coombs RW, et al. HIV testing in a highincidence population: Is antibody testing alone good enough? Clin Infect Dis. 2009;49: Eshleman SH, Khaki L, Laeyendecker O, et al. Detection of individuals with acute HIV-1 infection using the ARCHITECT HIV Ag/Ab Combo assay. J Acquir Immune Defic Syndr. 2009;52: Ly TD, Laperche S, Brennan C, et al. Evaluation of the sensitivity and specificity of six HIV combined p24 antigen and antibody assays. J Virol Methods. 2004;122: Kang HJ, Yoo KH, Kim HS, et al. [Evaluation of Abbott Fourth Generation HIV Antigen and Antibody Assays.]. [Korean] Korean J Lab Med. 2006;26: Weber B. A new automated fourth-generation HIV screening assay with sensitive antigen detection module and high specificity. In: Zhu T, ed. Methods in Molecular Biology vol Human Retrovirus Protocols: Virology and Molecular Biology. Totowa, NJ: Humana Press. 22. Shields M. HIV seroconversion illness Latest HIV assays may still be negative. Aust Fam Physician. 2006;35: Stekler JD, Swenson PD, Coombs RW, et al. HIV testing in a highincidence population: Is antibody testing alone good enough? Clin Infect Dis. 2009;49: CDC American Public Health Laboratories (APHL) HIV Diagnostics Conference, Technical Series of the Tulip Group: HIV testing: Evolution and practical issues. Goa, India: Qualpro Diagnostics, labmedicine.com September 2011 Volume 42 Number 9 LABMEDICINE 535

ARCHITECT HIV Ag/Ab Combo: Moving HIV Diagnostics Forward in the U.S.

ARCHITECT HIV Ag/Ab Combo: Moving HIV Diagnostics Forward in the U.S. ARCHITECT HIV Ag/Ab Combo: Moving HIV Diagnostics Forward in the U.S. Catherine Brennan, Ph.D. Research Fellow Infectious Diseases Research Abbott Diagnostics 1 Agenda ARCHITECT HIV Ag/Ab Combo Assay What

More information

New HIV Tests and Algorithm: A change we can believe in

New HIV Tests and Algorithm: A change we can believe in New HIV Tests and Algorithm: A change we can believe in Esther Babady, PhD, D (ABMM) Memorial Sloan-Kettering Cancer Center New York, New York Learning Objectives After this presentation you should be

More information

HIV Update in Laboratory Testing. Patricia Slev, PhD, D(ABCC)

HIV Update in Laboratory Testing. Patricia Slev, PhD, D(ABCC) HIV Update in Laboratory Testing Patricia Slev, PhD, D(ABCC) Objectives Explain the advances in HIV diagnostics, including fourth generation Ag/Ab combination HIV screening assays Describe the new CDC

More information

Learning Objectives. New HIV Testing Algorithm from CDC. Overview of HIV infection and disease 3/15/2016

Learning Objectives. New HIV Testing Algorithm from CDC. Overview of HIV infection and disease 3/15/2016 New HIV Testing Algorithm from CDC ASCLS-Michigan March 31, 2016 Dr. Kathleen Hoag Learning Objectives Following attendance and review of material provided, attendees will be able to: 1. Describe the new

More information

HIV Testing Technology and the Latest Algorithm

HIV Testing Technology and the Latest Algorithm HIV Testing Technology and the Latest Algorithm David Warshauer, PhD, D(ABMM) Deputy Director, Communicable Diseases Wisconsin State Laboratory of Hygiene HIV Testing has changed over time Patients with

More information

HIV Test Technologies, Best Practices, and New Algorithm. Jenny R. McFarlane DSHS HIV Prevention and Care Branch

HIV Test Technologies, Best Practices, and New Algorithm. Jenny R. McFarlane DSHS HIV Prevention and Care Branch HIV Test Technologies, Best Practices, and New Algorithm Jenny R. McFarlane DSHS HIV Prevention and Care Branch Testing History 1985 1 st Gen HIV-1 IA 1987 HIV-1 WB 1990 HIV-2 IA HIV-1 IA DBS 1991 2 nd

More information

1 st and 2 nd Generation EIA

1 st and 2 nd Generation EIA HIV Diagnostic Tests Bernard M. Branson, M.D. Associate D irector for Laboratory D iagnostics Division of HIV/AIDS Prevention Centers for D isease Control & Prevention The views expressed in this presentation

More information

A Summary of Clinical Evidence

A Summary of Clinical Evidence A Summary of Clinical Evidence Supporting the use of the Alere Determine HIV-1/2 Ag/Ab Combo Rapid Test to assist in the diagnosis of Human Immunodeficiency Virus (HIV) TAP HERE TO SEE THE PRODUCTS Table

More information

Diagnostic Tests for HIV

Diagnostic Tests for HIV Mountain West AIDS Education and Training Center Diagnostic Tests for HIV David Spach, MD Principal Investigator, Mountain West AETC Professor of Medicine, University of Washington Last Updated: June 22,

More information

Evolution of HIV Diagnostics. Goals for the 2010 Conference. Bernard M. Branson, M.D. CDC Division of HIV/AIDS Prevention

Evolution of HIV Diagnostics. Goals for the 2010 Conference. Bernard M. Branson, M.D. CDC Division of HIV/AIDS Prevention Evolution of HIV Diagnostics and Goals for the 2010 Conference Bernard M. Branson, M.D. Associate Director for Laboratory Diagnostics CDC Division of HIV/AIDS Prevention Evolution or other influences like

More information

Advantages and disadvantages of different types of FDA-approved HIV immunoassays used for screening by generation and platform*

Advantages and disadvantages of different types of FDA-approved HIV immunoassays used for screening by generation and platform* Advantages and disadvantages of different types of FDA-approved HIV immunoassays used for screening by generation and platform* HIV immunoassays grouped by generation, platform, and CLIA complexity + Advantages

More information

Point-of-care HIV testing

Point-of-care HIV testing Point-of-care HIV testing Joanne Stekler, MD MPH Associate Professor University of Washington December 5, 2016 Disclaimer: The findings and conclusions in this presentation are those of the author(s) and

More information

HIV Serology Quality Assessment Program Summary for Panel HIVS Oct22

HIV Serology Quality Assessment Program Summary for Panel HIVS Oct22 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Serology Quality Assessment Program Summary

More information

2018 HIV and HCV Diagnostic Testing Survey

2018 HIV and HCV Diagnostic Testing Survey 2018 HIV and HCV Diagnostic Testing Survey This survey is designed to capture the 2017 HIV and HCV testing practices in state and local public health laboratories (PHL). The results of the survey will

More information

The Latest in HIV Tests: What Do the Results Mean?

The Latest in HIV Tests: What Do the Results Mean? The Latest in HIV Tests: What Do the Results Mean? Bernard M. Branson MD Principal Consultant, Scientific Affairs LLC Atlanta, Georgia Objectives At the end of this workshop, participants will be able

More information

HIV Serology Quality Assessment Program Summary for Panel HIVSER 2017Apr19

HIV Serology Quality Assessment Program Summary for Panel HIVSER 2017Apr19 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Serology Quality Assessment Program Summary

More information

Using all the data: Immunoassay signal-to-cutoff values provide useful information that should be considered in HIV diagnostic

Using all the data: Immunoassay signal-to-cutoff values provide useful information that should be considered in HIV diagnostic Using all the data: Immunoassay signal-to-cutoff values provide useful information that should be considered in HIV diagnostic algorithms KP Delaney, M Pentella, B Bennett, and K Landgraf for the CDC/APHL

More information

Complicated viral infections

Complicated viral infections Complicated viral infections Clinical case discussion Diagnostic dilemmas NSW State Reference Laboratory for HIV St Vincent s Hospital Sydney Diagnostic dilemmas Indeterminate or discordant serology (western

More information

HIV: 2009 DIAGNOSTIC SURVEY FACING NEW CHALLENGES

HIV: 2009 DIAGNOSTIC SURVEY FACING NEW CHALLENGES ISSUES IN BRIEF: MEMBER SURVEY OF HIV ASSAY USE ASSOCIATION OF PUBLIC HEALTH LABORATORIES MARCH 2010 HIV: 2009 DIAGNOSTIC SURVEY FACING NEW CHALLENGES Twenty years ago, the Association of Public Health

More information

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ www.micropathology.com info@micropathology.com Micropathology Ltd Tel 24hrs: +44 (0) 24-76 323222 Fax / Ans: +44 (0) 24-76 - 323333 University of Warwick Science Park, Venture Centre, Sir William Lyons

More information

The Evolving Landscape of HIV Prevention and Diagnosis

The Evolving Landscape of HIV Prevention and Diagnosis The Evolving Landscape of HIV Prevention and Diagnosis Jenny R. McFarlane Texas Department of State Health Services HIV/STD/TB/Viral Hepatitis Unit Is it okay if we just talk? NHAS: A Call to ACT Reduce

More information

A Case of False-Positive Test Results in a Pregnant Woman of Unknown HIV Status at Delivery

A Case of False-Positive Test Results in a Pregnant Woman of Unknown HIV Status at Delivery A Case of False-Positive Test Results in a Pregnant Woman of Unknown HIV Status at Delivery Pascale Akl, MD, 1 Kenneth E. Blick, PhD 1* Lab Med Spring 2014;45:259-263 DOI: 10.1309/LMAAGVXK05LUWOQN ABSTRACT

More information

Reduction of the HIV Seroconversion Window Period and False Positive Rate by Using ADVIA Centaur HIV Antigen/Antibody Combo Assay

Reduction of the HIV Seroconversion Window Period and False Positive Rate by Using ADVIA Centaur HIV Antigen/Antibody Combo Assay Original Article Diagnostic Immunology Ann Lab Med 2013;33:420-425 http://dx.doi.org/10.3343/alm.2013.33.6.420 ISSN 2234-3806 eissn 2234-3814 Reduction of the HIV Seroconversion Window Period and False

More information

SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World

SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World MIT OpenCourseWare http://ocw.mit.edu SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World Spring 2009 For information about citing these materials or our

More information

New Generation of Nucleic Acid Testing. Michele Owen, Ph.D Division of HIV/AIDS Prevention Centers for Disease Control & Prevention

New Generation of Nucleic Acid Testing. Michele Owen, Ph.D Division of HIV/AIDS Prevention Centers for Disease Control & Prevention New Generation of Nucleic Acid Testing Michele Owen, Ph.D Division of HIV/AIDS Prevention Centers for Disease Control & Prevention Percentage (%) Persons Living with Diagnosed or Undiagnosed HIV Infection

More information

List the steps in the fourth generation HIV screening algorithm Describe the relationship between rapid HIV antibody tests and the fourth generation

List the steps in the fourth generation HIV screening algorithm Describe the relationship between rapid HIV antibody tests and the fourth generation Alere (honoraria) List the steps in the fourth generation HIV screening algorithm Describe the relationship between rapid HIV antibody tests and the fourth generation testing algorithm Interpret challenging

More information

HIV Guideline Sakchai Dettrairat

HIV Guideline Sakchai Dettrairat HIV Guideline 2016 Sakchai Dettrairat Division of Clinical Immunology Department of Medical Technology Faculty of Associated Medical Sciences Chiang Mai University Appearance of HIV markers in early HIV

More information

Routine HIV Testing Community of Practice Session #2

Routine HIV Testing Community of Practice Session #2 Routine HIV Testing Community of Practice Session #2 Presenters: Denver Prevention Training Center Broward Community & Family Health Centers, Inc. 20 March 2016 Communities of Practice Routine HIV Testing

More information

Technical Bulletin No. 104b

Technical Bulletin No. 104b CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 104b June 22, 2016 Update to Guidelines for Diagnosing HIV Infection CPAL s Testing Algorithm for Diagnostic HIV Testing HIV Geenius

More information

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx 0244-055-00 OraQuick HCV Rapid Antibody Test Kit with product codes 1001-0270

More information

AccuSet Anti-HIV-1 Mixed Titer Performance Panel

AccuSet Anti-HIV-1 Mixed Titer Performance Panel PACKAGE INSERT PRB205(M3) / (0800-0379) INTENDED USE The Performance Panel PRB205(M3) / (0800-0379) is intended for use by diagnostics manufacturers and clinical laboratorians to evaluate their HIV-1 test

More information

TESTS, REF TESTS)

TESTS, REF TESTS) ENGLISH Read the Package Insert completely before using the product. Follow the instructions carefully when performing testing. Failure to do so may result in inaccurate test results. For In Vitro Diagnostic

More information

Research Article Decision on conducting HCV Immunoblot and HCV Viral Load Tests Dependent upon the Result of the Screening Tests

Research Article Decision on conducting HCV Immunoblot and HCV Viral Load Tests Dependent upon the Result of the Screening Tests IBIMA Publishing Journal of Virology & Microbiology http://www.ibimapublishing.com/journals/jvm/jvm.html Vol. 2013 (2013), Article ID 332501, 7 pages DOI: 10.5171/2013.332501 Research Article Decision

More information

The Alphabet Soup of Viral Hepatitis Testing

The Alphabet Soup of Viral Hepatitis Testing The Alphabet Soup of Viral Hepatitis Testing August 18, 2011 Patricia Slev, PhD, DABCC Medical Director, Serologic Hepatitis and Retrovirus Laboratory, ARUP Laboratories Assistant Professor of Pathology,

More information

READ HIGHLIGHTED CHANGES

READ HIGHLIGHTED CHANGES BIO-FLASH anti-hbs 3000-8581 100 tests The BIO-FLASH anti-hbs is a fully automated chemiluminescent simultaneous immunoassay for quantitative measurement of antibodies to Hepatitis B surface antigen (anti-hbs)

More information

Overview of HIV Testing Practices and Technology

Overview of HIV Testing Practices and Technology Overview of HIV Testing Practices and Technology 2019 HIV Diagnostics Conference March 25 th, 2019 Michele Owen, Ph.D National Center for HIV/AIDS, Viral Hepatitis, STD and TB Prevention Centers for Disease

More information

Draft Report WHO evaluation of HIV 1/2 STAT-PAK (Chembio Diagnostic Systems Inc)

Draft Report WHO evaluation of HIV 1/2 STAT-PAK (Chembio Diagnostic Systems Inc) Date: 6 January 2012 Draft Report WHO evaluation of HIV 1/2 STAT-PAK (Chembio Diagnostic Systems Inc) WHO Collaborating Centre for HIV/AIDS Diagnostic and Laboratory Support Department of Clinical Sciences

More information

100 Abbott Park Road Abbott Park, IL Trade Name: ARCHITECT HIV Ag/Ab Combo Assay Approved Date: 18-JUN-2010

100 Abbott Park Road Abbott Park, IL Trade Name: ARCHITECT HIV Ag/Ab Combo Assay Approved Date: 18-JUN-2010 Currently Approved CBER Device Premarket Applications (PMAs) As of Nov 10, 2014 Sorted by Applicant Name BP090080-0 BP060002-0 BP120037-0 BP090022-0 BP090032-0 BP040046-0 BP100064-0 BP140120-0 Abbott Laboratories

More information

Viral Hepatitis Diagnosis and Management

Viral Hepatitis Diagnosis and Management Viral Hepatitis Diagnosis and Management CLINICAL BACKGROUND Viral hepatitis is a relatively common disease (25 per 100,000 individuals in the United States) caused by a diverse group of hepatotropic agents

More information

HEALTH. Re: Interim Guidelines for Laboratories on the use of a new Diagnostic Testing Algorithm for Human Immunodeficiency Virus (HIV) Infection

HEALTH. Re: Interim Guidelines for Laboratories on the use of a new Diagnostic Testing Algorithm for Human Immunodeficiency Virus (HIV) Infection ===========ir ~~!!~~fk li=:='========= Nirav R. Shah, M.D., M.P.H. Commissioner May 16, 2013 HEALTH Sue Kelly Executive Deputy Commissioner Re: Interim Guidelines for Laboratories on the use of a new Diagnostic

More information

AccuSet Anti-HIV-1 Mixed Titer Performance Panel

AccuSet Anti-HIV-1 Mixed Titer Performance Panel signal to cut-off (s/co) DATA SHEET OVERVIEW PRB205(M3) is a modified 16-member panel originating from Anti- HIV-1 Mixed Titer PRB205(M2). Panel members 3, 5, 6, 17, 21, 22, 23, 24, and 25 from the original

More information

Module 3: Overview of HIV Testing Technologies

Module 3: Overview of HIV Testing Technologies Module 3: Overview of HIV Testing Technologies Purpose Pre-requisite Modules Module Time Learning Objectives To provide the participants with a basic knowledge of HIV testing and how HIV rapid test results

More information

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere HIV Combo WHO reference number: PQDx

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere HIV Combo WHO reference number: PQDx WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: Alere HIV Combo WHO reference number: PQDx 0243-013-00 Alere HIV Combo with product codes 7D2842, 7D2843, 7D2843SET manufactured by Alere

More information

CEPHIA Consortium for the Evaluation and Performance of HIV Incidence Assays STANDARD OPERATING PROCEDURE

CEPHIA Consortium for the Evaluation and Performance of HIV Incidence Assays STANDARD OPERATING PROCEDURE CEPHIA Consortium for the Evaluation and Performance of HIV Incidence Assays STANDARD OPERATING PROCEDURE TITLE : SOP for (off board dilution) Less Sensitive Modified VITROS Enzyme Immunoassay CEPHIA DOCUMENT

More information

HIV Serology Quality Assessment Program Revised Summary for Panel HIVSER 2017Oct27

HIV Serology Quality Assessment Program Revised Summary for Panel HIVSER 2017Oct27 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Serology Quality Assessment Program Revised

More information

Trends in molecular diagnostics

Trends in molecular diagnostics Trends in molecular diagnostics Detection of target genes of interest Quantification Infectious diseases HIV Hepatitis C & B TB / MAC Cytomegalovirus Herpes simplex Varicella zoster CT/GC HPV Profiling

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Peters PJ, Westheimer E, Cohen S, et al. Screening yield of HIV antigen/antibody combination and pooled HIV RNA testing for acute HIV infection in a high-prevalence population.

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: HIV 1/2 STAT-PAK Number: PQDx Abstract

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: HIV 1/2 STAT-PAK Number: PQDx Abstract WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: HIV 1/2 STAT-PAK Number: PQDx 0007-006-00 Abstract The HIV 1/2 STAT-PAK with product code HIV101, manufactured by Chembio Diagnostic

More information

Expression of HIV-1 Markers During Progression of Infection E E E Days Since 1st Bleed

Expression of HIV-1 Markers During Progression of Infection E E E Days Since 1st Bleed Seroconversion OVERVIEW HIV-1 AccuVert TM Seroconversion PRB970 is a 4 member, 1.0 ml per vial panel of undiluted, naturally-occurring plasma samples. members represent serial bleeds collected from a single

More information

INTEGRATING HIV AND HCV TESTING

INTEGRATING HIV AND HCV TESTING INTEGRATING AND HCV TESTING OVERVIEW Increasingly health departments (HDs) are seeking new strategies to improve the productivity and yield of testing and linkage programs, expand availability of and access

More information

AccuSet HIV-1/2 Performance Panel

AccuSet HIV-1/2 Performance Panel OVERVIEW AccuSet HIV-1/2 (0800-0380) is a 13-member validation panel of undiluted, naturally occurring plasma samples (1 vial per member, 1.0 ml per vial). Panel members represent bleeds from multiple

More information

HIV testing has significantly improved

HIV testing has significantly improved HIV testing: an update By Nicola Zetola, MD, and Jeffrey D. Klausner, MD, MPH HIV testing has significantly improved since the first diagnostic test in 1985. Increased understanding of HIV transmission

More information

Adventures in Discordance- HIV Testing

Adventures in Discordance- HIV Testing Anne Gaynor, PhD On behalf of Monica M. Parker, PhD Director, Bloodborne Viruses Laboratory Wadsworth Center, NYSDOH Adventures in Discordance- HIV Testing HIV Laboratory Testing Algorithm HIV-1/2 antigen/antibody

More information

HUMAN IMMUNODEFICIENCY VIRUS (HIV) NON-IMMEDIATE NOTIFICATION STD PROGRAM. Version

HUMAN IMMUNODEFICIENCY VIRUS (HIV) NON-IMMEDIATE NOTIFICATION STD PROGRAM. Version 1 HUMAN IMMUNODEFICIENCY VIRUS (HIV) NON-IMMEDIATE NOTIFICATION STD PROGRAM Event Name: Event Time Period: ADULT HIV 900 (AIDS.gov 12/31/2015) HIV Lifelong HIV (human immunodeficiency virus) is a retrovirus

More information

Objectives. Seven days of separation. Clinical situation a transplant. The Donor COOMBS MD PHD 1. Laboratory Diagnosis & Monitoring of HIV Infection

Objectives. Seven days of separation. Clinical situation a transplant. The Donor COOMBS MD PHD 1. Laboratory Diagnosis & Monitoring of HIV Infection Laboratory Diagnosis & Monitoring of HIV Infection An Introduction to Diagnostic HIV Testing Robert W Coombs MD, PhD, FRCPC Professor, Departments of Laboratory Medicine & Medicine University of Washington,

More information

Primary Sample Manual Infectious Serology Issue No Effective Date: 20/09/17 Page 1 of 15 EUROFINS BIOMNIS

Primary Sample Manual Infectious Serology Issue No Effective Date: 20/09/17 Page 1 of 15 EUROFINS BIOMNIS Issue No. 2.02 Effective Date: 20/09/17 Page 1 of 15 Written / Revised By: Dr. Mike Louw, Medical Director Date: Reviewed By: Date: Dr. Sinead Kelly, Infectious Serology Consultant Authorised By: Jean-Sébastien

More information

HIV Serology Quality Assessment Program Summary for Panel HIVSER 2018Apr19

HIV Serology Quality Assessment Program Summary for Panel HIVSER 2018Apr19 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Serology Quality Assessment Program Summary

More information

NEW CHALLENGES FOR ACCURATE HIV-1 DIAGNOSIS ALL BELGIAN AIDS REFERENCE LABORATORIES KAROLIEN STOFFELS (ARL VUB STP) & FIEN VANROYE (ARL ITM)

NEW CHALLENGES FOR ACCURATE HIV-1 DIAGNOSIS ALL BELGIAN AIDS REFERENCE LABORATORIES KAROLIEN STOFFELS (ARL VUB STP) & FIEN VANROYE (ARL ITM) NEW CHALLENGES FOR ACCURATE HIV-1 DIAGNOSIS ALL BELGIAN AIDS REFERENCE LABORATORIES KAROLIEN STOFFELS (ARL VUB STP) & FIEN VANROYE (ARL ITM) 30 NOVEMBER 2018, BRUGGE 0 NEW CHALLENGES FOR ACCURATE HIV-1

More information

5. Over the last ten years, the proportion of HIV-infected persons who are women has: a. Increased b. Decreased c. Remained about the same 1

5. Over the last ten years, the proportion of HIV-infected persons who are women has: a. Increased b. Decreased c. Remained about the same 1 Epidemiology 227 April 24, 2009 MID-TERM EXAMINATION Select the best answer for the multiple choice questions. There are 60 questions and 9 pages on the examination. Each question will count one point.

More information

AccuVert HIV-1 Seroconversion Panel PRB974 ( )

AccuVert HIV-1 Seroconversion Panel PRB974 ( ) PACKAGE INSERT PRB974 (0600-0258) INTENDED USE PRB974 (0600-0258) is a group of serial bleeds from an individual plasma donor during seroconversion. This panel is intended for use by diagnostics manufacturers

More information

The Challenges in Developing and Commercializing HIV Tests that are Useful in Differentiating VISP/R VISP/R Workshop Bethesda, MD March 2013

The Challenges in Developing and Commercializing HIV Tests that are Useful in Differentiating VISP/R VISP/R Workshop Bethesda, MD March 2013 The Challenges in Developing and Commercializing HIV Tests that are Useful in Differentiating VISP/R VISP/R Workshop Bethesda, MD March 2013 Christopher Bentsen, M.S.,RAC, FRAPS Bio-Rad Laboratories, Redmond,

More information

ISBT WP-TTID Annual Report for Subgroup on Virology Drs. Michael Busch, Kurt Roth and Susan Stramer

ISBT WP-TTID Annual Report for Subgroup on Virology Drs. Michael Busch, Kurt Roth and Susan Stramer ISBT WP-TTID Annual Report for Subgroup on Virology Drs. Michael Busch, Kurt Roth and Susan Stramer Questionnaire on NAT Screening of Blood Donations for an International Forum on 10 years of NAT Screening

More information

AccuVert HBV Seroconversion Panel PHM941(M) ( )

AccuVert HBV Seroconversion Panel PHM941(M) ( ) PACKAGE INSERT PHM941(M) (0605-0061) INTENDED USE PHM941(M) (0605-0061) is a group of serial bleeds from an individual plasma donor during HBV seroconversion. This panel is intended for use by diagnostics

More information

TAP HERE TO SEE THE PRODUCT

TAP HERE TO SEE THE PRODUCT TAP HERE TO SEE THE PRODUCT Make every testing opportunity count. Acute infection accounts for 5-20% of all cases of HIV infection among persons seeking testing.1 This acute phase of infection is associated

More information

Screening and Diagnosis of Hepatitis Virus Infections

Screening and Diagnosis of Hepatitis Virus Infections Screening and Diagnosis of Hepatitis Virus Infections Prof. Jean-Michel Pawlotsky, MD, PhD National Reference Center for Viral Hepatitis B, C and delta Department of Virology & INSERM U955 Henri Mondor

More information

HIV-1 Seroconversion Panel PRB973

HIV-1 Seroconversion Panel PRB973 A SERACARE PANEL PRODUCT INTENDED USE The is a group of serial bleeds from an individual plasma donor during seroconversion. This panel is intended for use by diagnostics manufacturers and clinical laboratorians

More information

HIV-1 Seroconversion Panel

HIV-1 Seroconversion Panel A SERACARE PANEL PRODUCT Seroconversion INTENDED USE The Seroconversion is a group of serial bleeds from an individual plasma donor during seroconversion. This is intended for use by diagnostics manufacturers

More information

HIV Diagnostic Testing

HIV Diagnostic Testing In The name of God HIV Diagnostic Testing By : Dr. Shahzamani PhD of Medical virology Purpose of HIV Testing To identify asymptomatic individuals To diagnose HIV infection in those who practice high risk

More information

Sensitivity of the Procleix HIV-1/HCV Assay for Detection of Human Immunodeficiency Virus Type 1 and Hepatitis C Virus RNA in a High-Risk Population

Sensitivity of the Procleix HIV-1/HCV Assay for Detection of Human Immunodeficiency Virus Type 1 and Hepatitis C Virus RNA in a High-Risk Population JOURNAL OF CLINICAL MICROBIOLOGY, July 2002, p. 2387 2391 Vol. 40, No. 7 0095-1137/02/$04.00 0 DOI: 10.1128/JCM.40.7.2387 2391.2002 Copyright 2002, American Society for Microbiology. All Rights Reserved.

More information

2006 HIV Diagnostics Survey

2006 HIV Diagnostics Survey Public Health Laboratory Issues in Brief: 2006 HIV Diagnostics Survey Association of Public Health Laboratories November 2007 Background Nearly 20 years ago, the Association of Public Health Laboratories

More information

Spectrum of false positivity for the fourth generation human immunodeficiency virus diagnostic tests

Spectrum of false positivity for the fourth generation human immunodeficiency virus diagnostic tests DOI 10.1186/s12981-015-0086-3 AIDS Research and Therapy CASE REPORT Spectrum of false positivity for the fourth generation human immunodeficiency virus diagnostic tests Peter Liu 1*, Patrick Jackson 2,

More information

detection of HIV-1/2 antibodies

detection of HIV-1/2 antibodies CVI Accepts, published online ahead of print on 18 June 214 Clin. Vaccine Immunol. doi:1.1128/cvi.153-14 Copyright 214, American Society for Microbiology. All Rights Reserved. 1 2 Evaluation of Bio-Rad

More information

FINAL REPORT HBV Serology External Quality Assessment Scheme

FINAL REPORT HBV Serology External Quality Assessment Scheme FINAL REPORT HBV Serology External Quality Assessment Scheme Programme codes: HBVC4310, HBVC435 Panel ID: 2008/Mar/11 HBV Serology EQAS Panel ID 2008/Mar/11 Page 1 of 9 The NRL is a: NATA-accredited proficiency

More information

HIV-1 Seroconversion Panel PRB964

HIV-1 Seroconversion Panel PRB964 A SERACARE PANEL PRODUCT Seroconversion INTENDED USE The is a group of serial bleeds from an individual plasma donor during seroconversion. This is intended for use by diagnostics manufacturers and clinical

More information

AccuSet HCV Performance Panel

AccuSet HCV Performance Panel Signal to cutoff (s/co) DATA SHEET OVERVIEW AccuSet HCV (0810-0204) is a 25-member validation panel of undiluted, naturally occurring plasma samples (1 vial per member, 1.0 ml per vial). Panel members

More information

Diagnosing HIV Infection AMNERIS E. LUQUE, M.D. PROFESSOR OF MEDICINE/INFECTIOUS DISEASES MEDICAL DIRECTOR, PARKLAND HIV SERVICES

Diagnosing HIV Infection AMNERIS E. LUQUE, M.D. PROFESSOR OF MEDICINE/INFECTIOUS DISEASES MEDICAL DIRECTOR, PARKLAND HIV SERVICES Diagnosing HIV Infection AMNERIS E. LUQUE, M.D. PROFESSOR OF MEDICINE/INFECTIOUS DISEASES MEDICAL DIRECTOR, PARKLAND HIV SERVICES Disclosures None Objectives 1. Outline the CDC recommendations for HIV

More information

Comparison between ELISA and chemiluminescence immunoassay for the detection of Hepatitis C virus antibody

Comparison between ELISA and chemiluminescence immunoassay for the detection of Hepatitis C virus antibody Original Research Article DOI: 10.18231/2394-5478.2017.0078 Comparison between ELISA and chemiluminescence immunoassay for the detection of Hepatitis C virus antibody Pampi Majumder 1, Anup Kumar Shetty

More information

HIV-1 Seroconversion Panel

HIV-1 Seroconversion Panel OVERVIEW This Data Sheet contains test results specific for HIV-1. s are undiluted aliquots from plasma units collected from a single donor. No preservatives were added. CAUTION: Potentially infectious

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Lookback for Hepatitis C Virus (HCV): Product Quarantine, Consignee Notification, Further Testing, Product Disposition, and Notification of Transfusion Recipients Based on Donor Test

More information

SOLE-SOURCE DETERMINATION

SOLE-SOURCE DETERMINATION SOLE-SOURCE DETERMINATION The Purchasing Division has been requested to approve a sole source purchase for the commodity or service described below. Pursuant to West Virginia Code 5A-3-10c, the Purchasing

More information

List of HIV diagnostics eligible to tender for procurement by WHO in 2011

List of HIV diagnostics eligible to tender for procurement by WHO in 2011 Since 1988, WHO has performed assessments of commercially available diagnostics for HIV, hepatitis B, hepatitis C and Chagas disease. Up to 2006, this process was called the WHO Test Kit Evaluation programme,

More information

Opportunities Created by Diagnostic HCV and HIV Nucleic Acid Tests

Opportunities Created by Diagnostic HCV and HIV Nucleic Acid Tests Opportunities Created by Diagnostic HCV and HIV Nucleic Acid Tests Ann Winters, MD Medical Director, Viral Hepatitis Program, Bureau of Communicable Disease, New York City Department of Health and Mental

More information

BIO-FLASH. BIOKIT, S.A. - Can Malé s/n Lliçà d Amunt - Barcelona - SPAIN

BIO-FLASH. BIOKIT, S.A. - Can Malé s/n Lliçà d Amunt - Barcelona - SPAIN READ HIGHLIGHTED CHANGES BIO-FLASH anti-hbc 3000-8578 100 tests The BIO-FLASH anti-hbc is a fully automated chemiluminescent two-step immunoassay for qualitative measurement of total antibodies to hepatitis

More information

Technical Bulletin No. 104a

Technical Bulletin No. 104a CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 104a April 22, 2014 Update to Guidelines for Diagnosing HIV Infection - CPAL s Testing Algorithm for Diagnostic HIV Testing Contact:

More information

READ HIGHLIGHTED CHANGES

READ HIGHLIGHTED CHANGES BIO-FLASH Toxo IgM 3000-8554 50 tests The BIO-FLASH Toxo IgM is a fully automated chemiluminescent two-step immunoassay for qualitative measurement of IgM antibodies to Toxoplasma gondii in human serum

More information

Joanne Stekler, MD, MPH Assistant Professor of Medicine Deputy Director, Public Health Seattle & King County HIV/STD Program

Joanne Stekler, MD, MPH Assistant Professor of Medicine Deputy Director, Public Health Seattle & King County HIV/STD Program NORTHWEST AIDS EDUCATION AND TRAINING CENTER HIV Testing Joanne Stekler, MD, MPH Assistant Professor of Medicine Deputy Director, Public Health Seattle & King County HIV/STD Program Presentation prepared

More information

Serology and International units

Serology and International units Serology and International units L. Grangeot-Keros, National Reference Laboratory for Rubella, Virology Department, A. Béclère Hospital, Clamart, France Detection of rubella-specific IgG antibody Assays

More information

Bringing excellence to life

Bringing excellence to life Who believes in oral HIV point-of-care tests? A summary of 3 years experience and more than 2000 OraQuick ADVANCE tests N Garrett, J Saunders, K Moir, J Zelin, C Estcourt On behalf of North East London

More information

The cost-effectiveness of NAT for HIV, HCV, and HBV in whole-blood donations Jackson B R, Busch M P, Stramer S L, AuBuchon J P

The cost-effectiveness of NAT for HIV, HCV, and HBV in whole-blood donations Jackson B R, Busch M P, Stramer S L, AuBuchon J P The cost-effectiveness of NAT for HIV, HCV, and HBV in whole-blood donations Jackson B R, Busch M P, Stramer S L, AuBuchon J P Record Status This is a critical abstract of an economic evaluation that meets

More information

HIV and PEP. LTC Rose Ressner WRNMMC ID staff Oct 2014 UNCLASSIFIED

HIV and PEP. LTC Rose Ressner WRNMMC ID staff Oct 2014 UNCLASSIFIED HIV and PEP LTC Rose Ressner WRNMMC ID staff Oct 2014 UNCLASSIFIED Disclaimer The views expressed in this presentation are those of the speaker and do not reflect the official policy of the Department

More information

Hiv diagnosis By: k. baesi

Hiv diagnosis By: k. baesi Hiv diagnosis By: k. baesi In contrast to the epast, HIV testing has now gained immense therapeutic relevance: starting HAART on time may improve the quality of life and indeed dprolong lf life considerably.

More information

Treponema-Specific Tests for Serodiagnosis of Syphilis: Comparative Evaluation of Seven Assays

Treponema-Specific Tests for Serodiagnosis of Syphilis: Comparative Evaluation of Seven Assays JOURNAL OF CLINICAL MICROBIOLOGY, Apr. 2011, p. 1313 1317 Vol. 49, No. 4 0095-1137/11/$12.00 doi:10.1128/jcm.02555-10 Copyright 2011, American Society for Microbiology. All Rights Reserved. Treponema-Specific

More information

Technical Bulletin No. 161

Technical Bulletin No. 161 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 161 cobas 6800 HIV-1 Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: ABON HIV 1/2/O Tri-Line Human Immunodeficiency Virus Rapid Test Device Number: PQDx 0141-051-00 Abstract ABON HIV 1/2/O Tri-Line Human

More information

Notice. Guidance for Manufacturers of Human Immunodeficiency Virus (HIV) Test Kits intended to be used in the Laboratory

Notice. Guidance for Manufacturers of Human Immunodeficiency Virus (HIV) Test Kits intended to be used in the Laboratory December 7, 2011 Notice Our file number: 11-121739-524 Re: Test Kits intended to be used in the Laboratory is pleased to announce the release of the final version of the Guidance for Manufacturers of Test

More information

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO)

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO) Diagnostic Methods of HBV infection Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO) Hepatitis B-laboratory diagnosis Detection of HBV infection involves

More information

Model Performance Evaluation Program (MPEP) HIV Rapid Testing Survey: Report of Sample Shipment Results, September 2009

Model Performance Evaluation Program (MPEP) HIV Rapid Testing Survey: Report of Sample Shipment Results, September 2009 Use of trade names and commercial sources is for identification only and does not constitute endorsement by the Public Health Service or the U.S. Department of Model Performance Evaluation Program (MPEP)

More information

Diagnostic Methods of HBV and HDV infections

Diagnostic Methods of HBV and HDV infections Diagnostic Methods of HBV and HDV infections Zohreh Sharifi,ph.D Blood Transfusion Research Center, High Institute for Research and Education in Transfusion Medicine Hepatitis B-laboratory diagnosis Detection

More information