PRACTICES OF COATING COLLECTION SURFACES OF CASCADE IMPACTORS: A SURVEY OF MEMBERS OF THE EUROPEAN PHARMACEUTICAL AEROSOL GROUP (EPAG)

Size: px
Start display at page:

Download "PRACTICES OF COATING COLLECTION SURFACES OF CASCADE IMPACTORS: A SURVEY OF MEMBERS OF THE EUROPEAN PHARMACEUTICAL AEROSOL GROUP (EPAG)"

Transcription

1 PRACTICES OF COATING COLLECTION SURFACES OF CASCADE IMPACTORS: A SURVEY OF MEMBERS OF THE EUROPEAN PHARMACEUTICAL AEROSOL GROUP () J.P. Mitchell on behalf of members of the European Pharmaceutical Aerosol Group Summary members evaluate many different aerosols from both pressurized metered dose inhalers (pmdis) and dry powder inhalers (s) by the use of a cascade impactor. It is widely recognized that pre-coating of particle collection surfaces is a prerequisite in testing to avoid bias caused by particle bounce. A summary of a recent survey is presented, in which members were asked to provide information about their impactor type and coating methodology, including the composition of the coating material applied to the collection surfaces. There is no standard approach in terms of either materials used or method of applying the coating. However, standardization is unnecessary, since each organization has independently selected and validated their procedure with their own formulations. The newcomer can therefore select from a variety of methods to eliminate particle bounce, and at the same time ensure quantitative recovery and assay of the active pharmaceutical ingredient(s). Introduction Coating of the collection surfaces of the cascade impactor with a media that provides a tacky surface is recognized as an appropriate precaution to minimize particle bounce and reentrainment, especially when testing dry powder inhalers (s) [1,2]. Studies with both the Andersen 8-stage impactor (ACI) [3] and Next Generation Pharmaceutical Impactor () [4] indicate that collection surface coating may also be required for some formulations delivered by pressurized metered dose inhaler (pmdi), especially when measurements are being made with a low number of actuations from the inhaler. The table summarizes the coating methods that are currently in use by members of. Impactor Collection Surface Coating Methods ACI Andersen 8-stage impactor, MMI Marple-Miller impactor, Next Generation Pharmaceutical impactor, VHC valved holding chamber

2 Table 1. Coating Methods Used by Organizations 1 ACI pmdi 2 ACI 3 ACI, 4 ACI pmdi & 5 ACI pmdi, or Aluminium alloy 6 ACI, 7 ACI, 8 ACI, pmdi, Ethanolic solution of Brij35 (polyoxyethylenelaurylether) is mixed with glycerol 11% polypropylene glycol in Hexane A solution of Brij, glycerol and ethanol (ethanol evaporates after applying, before analysis) Glycerol or Brij 35. Applied only if required 1 ml of Brij 35 in 20 ml ethanol dissolved in 5 g glycerol n-hexane + 1% of SPAN 85 (sorbitan trioleate) 1% w/v silicone oil in hexane Dip plate into coating solution, leave in a fume hood to allow solvent to evaporate before assembly of impactor Submerge plate, remove plate allowing the bulk to drain off, then air dry Apply standard amount of coating material (thin) evenly (in practise done with a sponge). Apply standard quantity of solution to give a thin film of coating substance on each plate. Distribute 50 µl per plate For, distribute 100 µl in pre-separator Immerse each plate in a crystalizer containing the coating material. Allow 0.5 hr for solvent evaporation in a fume hood before assembly of impactor Dip plates in solution and allow hexane to evaporate prior to assembly of impactor Inox Glycerine Coat with a thin layer of glycerine

3 9 ACI 10 ACI 11 ACI Liquid systems pmdi, Method 1: Brij 35 (2% w/v) in glycerol Method 2: Span-85 (1% w/v) in hexane 1% soln. Tween20 in methanol. The pre-separator is not coated Non-coated or Brij 35/ethanol 12 ACI Apiezon spray (Dekati Ltd) 13 ACI, 14 ACI, 15 ACI,, pmdi 1% w/v silicone oil in hexane 1% Tween in EtOH alternatively 1% span 85/hexane 3 g Brij 35 P dissolved in 20 ml ethanol, 1 ml of the Brij solution mixed with 5 g glycerol; alternatively: 1 % silicone oil in hexane; alternatively: % Tween80 in glycerol; Dissolve 2.5 g Brij in 25 ml ethanol. Add 4-ml to 20 g glycerol and mix well. Apply 50 µl as coating and spread to coat entire plate surface Apply coating via a pipette, allowing excess to run of plate. Allow solvent to evaporate. Recovery of active by immersion of plate in sample diluent, gently swirling for 10-min Immerse plates, remove and air dry Pipette coating solution onto plate, wipe-off excess and allow to air dry Apply spray to plates once a day (max 6 collections) Dip plates in solution, allowing hexane to evaporate prior to assembly of impactor Dip plates in solution, allow to air dry in clean conditions. Check for homogeneous coating visually before use Distribute 50 µl solution per plate. Distribute 100 µl in pre-separator (ACI) for testing

4 16 ACI,, MMI pmdi/ VHC ACI; metal cups and MMI 17 ACI Aluminium or stainless A check is made for particle bounce at IQ for new impactors or formulations. Coat with a solution of Brij 35, glycerol and ethanol Releasil-B silicone spray (Dow Corning) Apply standard amount of coating material (thin) and evenly. Spray directly on to plate Conclusions The table demonstrates the wide variety of methods that are in current use by members to coat the collection surfaces of cascade impactors for the particle size assessment of a variety of inhaler types. The decision whether or not to coat the collection surfaces of the cascade impactor will depend on both the formulation and type of impactor, and should always be considered as part of the method development process [5]. References 1. USP 26-NF Chapter 601 Physical tests and determinations: Aerosols. United States Pharmacopeia, Rockville, MD, USA European Pharmacopeia Section Preparations for inhalation: aerodynamic assessment of fine particles. European Pharmacopeia: 3 rd Edn., Suppl Council of Europe, Strasbourg, France Nasr, M.M., D.L. Ross and N.C. Miller Effect of drug load and plate coating on the particle size distribution of a commercial albuterol metered dose inhaler (MDI) determined using the Andersen and Marple-Miller impactors. Pharm. Res. 14(10): Kamiya, A., Sakagami, M., Hindle, M. and Byron, P.R Particle sizing with the next generation impactor: A study of Vanceril metered dose inhaler. J. Aerosol Med., 16(2): Christopher D., Curry, P., Doub, B., Furnkranz, K., Lavery, M., Lin, K., Lyapustina, S., Mitchell, J., Rogers, B., Strickland, H., Tougas, T., Tsong, Y. and Wyka, B Considerations for the development and practice of cascade impaction testing, including a mass balance failure investigation tree. J. Aerosol Med. 16: member organizations are: 3M, Astra Zeneca, Aventis, Bespak, Boehringer Ingelheim, Chiesi, Clinical Designs, GlaxoSmithKline, Hovione, Innovata Biomed, Ivax Pharmaceticals, Novartis, Novo Nordisk, Pharmachenie, Pfizer, SkyePharma, Sofotec, Trudell Medical International and Vectura

5 Drug Delivery to the Lungs 14, The Aerosol Society, London, UK, 2003:75-78 PRACTICES OF COATING COLLECTION SURFACES OF CASCADE IMPACTORS: A SURVEY OF MEMBERS OF THE EUROPEAN PHARMACEUTICAL AEROSOL GROUP () J.P. Mitchell on behalf of members of the European Pharmaceutical Aerosol Group members evaluate many different aerosols from both pressurized metered dose inhalers (pmdis) and dry powder inhalers (s) by the use of a cascade impactor. It is widely recognized that pre-coating of particle collection surfaces is a prerequisite in testing to avoid bias caused by particle bounce. A summary of a recent survey is presented, in which members were asked to provide information about their impactor type and coating methodology, including the composition of the coating material applied to the collection surfaces. There is no standard approach in terms of either materials used or method of applying the coating. However, standardization is unnecessary, since each organization has independently selected and validated their procedure with their own formulations. The newcomer can therefore select from a variety of methods to eliminate particle bounce, and at the same time ensure quantitative recovery and assay of the active pharmaceutical ingredient(s).

The Role of EPAG in Standards and Regulatory Guidance Development

The Role of EPAG in Standards and Regulatory Guidance Development The Role of EPAG in Standards and Regulatory Guidance Development Jolyon P. Mitchell on behalf of EPAG Session 4: The Remit, Activities and Achievements of The European Pharmaceutical Aerosols Group 1

More information

Last updated: July 2011

Last updated: July 2011 EFFICIENT DATA ANALYSIS, ABBREVIATED IMPACTOR MEASUREMENTS, AERODYNAMIC PARTICLE SIZE DISTRIBUTIONS: PUBLICATIONS AND PRESENTATIONS BY MEMBERS OF THE IPAC-RS CASCADE IMPACTION WORKING GROUP AND EPAG 2000-2011

More information

The abbreviated impactor measurement concept

The abbreviated impactor measurement concept AS APPEARED IN Inhalation JUNE 2009 www.inhalationmag.com The abbreviated impactor measurement concept A potentially faster and more precise way to assess quality of inhaled drug products Jolyon P. Mitchell,

More information

EPAG Perspective - Regulatory Advances Related to Nasal Spray Pumps. Dr G.Williams Nasal Drug Delivery Management Forum London, 15 Apr 2010

EPAG Perspective - Regulatory Advances Related to Nasal Spray Pumps. Dr G.Williams Nasal Drug Delivery Management Forum London, 15 Apr 2010 1 EPAG Perspective - Regulatory Advances Related to Nasal Spray Pumps Dr G.Williams Nasal Drug Delivery Management Forum London, 15 Apr 2010 Overview 2 EPAG, what is it?, background, this project Regulatory

More information

Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pmdi)

Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pmdi) Respiratory Drug Delivery Europe 2017 Nagel and Suggett Equivalence Evaluation of Valved Holding Chambers (VHCs) with Albuterol Pressurized Metered Dose Inhaler (pmdi) Mark W. Nagel and Jason A. Suggett

More information

The effect of different coating materials on the prevention of powder bounce in the next generation impactor

The effect of different coating materials on the prevention of powder bounce in the next generation impactor Research in Pharmaceutical Sciences, June 2018; 13(3): 283-287 Received: June 2017 Accepted: December 2017 School of Pharmacy & Pharmaceutical Sciences Isfahan University of Medical Sciences Short Communication

More information

Caption: The equipment required for testing Fluticasone Propionate (FP) Inhalation Powder in line with a new product-specific monograph (USP36-NF31).

Caption: The equipment required for testing Fluticasone Propionate (FP) Inhalation Powder in line with a new product-specific monograph (USP36-NF31). Product-specific FDA guidance, and product-specific pharmacopeial monographs, point to the use of test equipment, some of which isn t included in the general USP/Ph. Eur. chapters for orally inhaled products

More information

EPAG Sponsored Workshop on Abbreviated Impactor Measurement (AIM) and Efficient Data Analysis (EDA) Concepts in Inhaler Testing. Overview of AIM EDA

EPAG Sponsored Workshop on Abbreviated Impactor Measurement (AIM) and Efficient Data Analysis (EDA) Concepts in Inhaler Testing. Overview of AIM EDA EPAG Sponsored Workshop on Abbreviated Impactor Measurement (AIM) and Efficient Data Analysis (EDA) Concepts in Inhaler Testing Overview of AIM EDA Jolyon Mitchell 1 and Mark Copley 2 1 Trudell Medical

More information

INTRODUCTION. size and total nozzle area decrease with stage number. Volumetric air flow rate through

INTRODUCTION. size and total nozzle area decrease with stage number. Volumetric air flow rate through CASCADE I M P A C T I O N Optimizing Cascade Impactor Testing for Characterizing Orally Inhaled & Nasal Drug Products By: Mark Copley INTRODUCTION Cascade impaction is a core analytical technique for characterizing

More information

Understanding cascade impaction and its importance for inhaler testing

Understanding cascade impaction and its importance for inhaler testing Understanding cascade impaction and its importance for inhaler testing Mark Copley, Technical Sales Manager Inhalation product development is an important area of activity for the pharmaceutical sector.

More information

Characterizing the Performance of Metered Dose Inhalers with Add-On Devices: New Methods For Clinically Relevant Testing

Characterizing the Performance of Metered Dose Inhalers with Add-On Devices: New Methods For Clinically Relevant Testing INSTRUMENTATION» Characterizing the Performance of Metered Dose Inhalers with Add-On Devices: New Methods For Clinically Relevant Testing Mark Copley Director Copley Scientific The recent introduction

More information

OVERVIEW OF EFFECTIVE DATA ANALYSIS (EDA)

OVERVIEW OF EFFECTIVE DATA ANALYSIS (EDA) OVERVIEW OF EFFECTIVE DATA ANALYSIS (EDA) Terrence P. Tougas on behalf of the IPAC-RS Cascade Impactor Working Group IPAC-RS 2011 Conference 1 CI WG Members 1. Steve Stein 3M 2. Mårten Svenson AstraZeneca

More information

Inhalation Product Research at FDA

Inhalation Product Research at FDA Inhalation Product Research at FDA Changning Guo Ph. D., Chemist Division of Pharmaceutical Analysis FDA/CDER/OPS/OTR 2016 GPhA CMC workshop, May 17, 2016 Disclaimer: This presentation reflects the views

More information

905 UNIFORMITY OF DOSAGE UNITS

905 UNIFORMITY OF DOSAGE UNITS Change to read: 905 UNIFORMITY OF DOSAGE UNITS [ NOTE In this chapter, unit and dosage unit are synonymous. ] To ensure the consistency of dosage units, each unit in a batch should have a drug substance

More information

Metered Dose Inhalers (MDIs) In Vitro Measures to Confirm Patient Perceptions: HFA vs. CFC

Metered Dose Inhalers (MDIs) In Vitro Measures to Confirm Patient Perceptions: HFA vs. CFC Metered Dose Inhalers (MDIs) In Vitro Measures to Confirm Patient Perceptions: HFA vs. CFC William H. Doub, Ph.D. Division of Pharmaceutical Analysis (DPA) US FDA/CDER/OPS/OTR The information presented

More information

DEBATING THE OPERATING CURVES FOR DDU TESTS ON MARKETED INHALERS

DEBATING THE OPERATING CURVES FOR DDU TESTS ON MARKETED INHALERS DEBATING THE OPERATING CURVES FOR DDU TESTS ON MARKETED INHALERS Industry s problem with the current FDA DDU requirements Presented by Bo Olsson, AstraZeneca R&D Lund on behalf of IPAC-RS DDU Working Group

More information

Adapting Pharmacopeial Testing for Pharmaceutical Equivalence Studies. Julie D. Suman, Ph.D. March 29, 2016

Adapting Pharmacopeial Testing for Pharmaceutical Equivalence Studies. Julie D. Suman, Ph.D. March 29, 2016 Adapting Pharmacopeial Testing for Pharmaceutical Equivalence Studies Julie D. Suman, Ph.D. March 29, 2016 AGENDA Regulatory Approaches for Bioequivalence (BE) Compendial Testing Vs Pharmaceutical Equivalence

More information

Assessing the role of breathing simulators in OIP testing

Assessing the role of breathing simulators in OIP testing As first AppeAred in Inhalation April 2014 www.inhalationmag.com Assessing the role of breathing simulators in OIP testing Exploring how the application of patient-representative inhalation profiles can

More information

REFERENCES OVERVIEW ADVANTAGES AEROSOLS. Aerosols. and Foams

REFERENCES OVERVIEW ADVANTAGES AEROSOLS. Aerosols. and Foams COMMONWEALTH OF AUSTRALIA Copyright Regulations 1969 WARNING This material has been reproduced and communicated to you by or on behalf of Monash University pursuant to Part VB of the Copyright Act 1968

More information

Testing Inhaled Generics

Testing Inhaled Generics Generic Bioequivalence Testing Inhaled Generics By Mark Copley at Copley Scientific New product-specific FDA guidance and USP monographs support the development of popular inhaled products. This article

More information

Go With the Flow REGULATORY LANDSCAPE. Mark Copley at Copley Scientific

Go With the Flow REGULATORY LANDSCAPE. Mark Copley at Copley Scientific Go With the Flow Image: Guzel Studio shutterstock.com Increasing global requirements for efficacious, inexpensive products to treat respiratory illnesses are driving the development of inhaled generics.

More information

OF COATING MATERIAL ON THE AERODYNAMIC PARTICLE SIZE DISTRIBUTION (PSD) OF OXIS TURBOHALER USING MIXING INLET WITH AN ANDERSEN CASCADE IMPACTOR (ACI)

OF COATING MATERIAL ON THE AERODYNAMIC PARTICLE SIZE DISTRIBUTION (PSD) OF OXIS TURBOHALER USING MIXING INLET WITH AN ANDERSEN CASCADE IMPACTOR (ACI) 165 J App Pharm 2(3): 165-178 (2011) Khan et al., 2011 EFFECT OF COATING MATERIAL ON THE AERODYNAMIC PARTICLE SIZE DISTRIBUTION (PSD) OF OXIS TURBOHALER USING MIXING INLET WITH AN ANDERSEN CASCADE IMPACTOR

More information

What Do Cascade Impaction Measurements Tell Us: In Vitro Aspects. Context of Presentation

What Do Cascade Impaction Measurements Tell Us: In Vitro Aspects. Context of Presentation What Do Cascade Impaction Measurements Tell Us: In Vitro Aspects Jolyon P. Mitchell, Ph.D. Trudell Medical International London, Canada IPAC-RS Conference November 2006 1 Context of Presentation This talk

More information

Best Practices for OINDP Pharmaceutical Development Programs Leachables and Extractables

Best Practices for OINDP Pharmaceutical Development Programs Leachables and Extractables Best Practices for OINDP Pharmaceutical Development Programs Leachables and Extractables PQRI Leachables & Extractables Working Group PQRI Training Course 12-13 April 2007 Chicago, IL What are leachables

More information

IVIVC in Pediatric OIPs

IVIVC in Pediatric OIPs IPAC-RS/UF Orlando Conference 2014 March 20, 2014 IVIVC in Pediatric OIPs Herbert Wachtel Declaration of Conflicts of Interest H. Wachtel is employee of Boehringer Ingelheim Pharma GmbH & Co. KG, Germany.

More information

AEROSOL THERAPY: THE PRACTICALITIES

AEROSOL THERAPY: THE PRACTICALITIES AEROSOL THERAPY: THE PRACTICALITIES Lester I. Harrison, PhD Section Head, Clinical Pharmacokinetics, 3M Pharmaceuticals, 3M Center 270-3S-05, St. Paul, MN, USA 55144 liharrison@mmm.com Introduction: Horses,

More information

Formulation Considerations for Inhaled Products

Formulation Considerations for Inhaled Products Formulation Considerations for Inhaled Products Formulation Considerations of Inhaled Products Inhalation Therapy Nebulizers and Formulations Dry Powder Inhalers and Formulations Metered Dose Inhalers

More information

Teerapol SRICHANA,*,a Anthony BRAIN, b Christopher MARRIOTT, c and Gary Peter MARTIN c

Teerapol SRICHANA,*,a Anthony BRAIN, b Christopher MARRIOTT, c and Gary Peter MARTIN c February 2000 Chem. Pharm. Bull. 48(2) 167 174 (2000) 167 A Study of Drug Carrier Interactions in Dry Powder Inhaler Formulations Using the Andersen Cascade Impactor, X-Ray Microanalysis and Time of Flight

More information

The Collapsible Chamber

The Collapsible Chamber NEW The Collapsible Chamber DESIGNED FOR PERFORMANCE AND CONVENIENCE INDICATIONS FOR USE Flexichamber is an anti-static valved collapsible holding chamber intended to be used by patients under the care

More information

Device Change Management for Inhaled Products. Loy Britto, Ph.D. GlaxoSmithKline ISAM Congress Munich 2015

Device Change Management for Inhaled Products. Loy Britto, Ph.D. GlaxoSmithKline ISAM Congress Munich 2015 Device Change Management for Inhaled Products Loy Britto, Ph.D. GlaxoSmithKline ISAM Congress Munich 2015 Topics to be covered Update on ISO/TC 084/WG 15: Guidelines for development of drug products- Quality

More information

International Pharmaceutical Aerosol Consortium on Regulation and Science

International Pharmaceutical Aerosol Consortium on Regulation and Science International Pharmaceutical Aerosol Consortium on Regulation and Science 1500 K Street NW Washington DC 20005 Telephone +1 202 230 5607 Fax +1 202 842 8465 Email info@ipacrs.org Web www.ipacrs.org Submitted

More information

Device Design Similarity

Device Design Similarity Device Design Similarity Dave Parkins Director DPI Product Development PQRI Workshop on Demonstrating Bioequivalence of Locally Acting Orally Inhaled Drug Products. Bethesda March 9-10, 2009 Device Similarity

More information

University of Groningen. Technology in practice Lexmond, Anne

University of Groningen. Technology in practice Lexmond, Anne University of Groningen Technology in practice Lexmond, Anne IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document

More information

ADDITIONAL TECHNOLOGIES FOR PRESSURIZED METERED DOSE INHALERS. Steve Newman Scientific Consultant Nottingham, UK

ADDITIONAL TECHNOLOGIES FOR PRESSURIZED METERED DOSE INHALERS. Steve Newman Scientific Consultant Nottingham, UK THE PRESS-AND-BREATHE pmdi ADDITIONAL TECHNOLOGIES FOR PRESSURIZED METERED DOSE INHALERS Steve Newman Scientific Consultant Nottingham, UK steve.newman@physics.org Compact, portable, convenient Asthma

More information

SELECTION AND VALIDATION OF CASCADE IMPACTOR TEST METHODS

SELECTION AND VALIDATION OF CASCADE IMPACTOR TEST METHODS Respiratory Drug Delivery IX, 2004 Byron et al. SELECTION AND VALIDATION OF CASCADE IMPACTOR TEST METHODS Peter R. Byron, 1 R. Harris Cummings, 2 Steven C. Nichols, 3 Guirag Poochikian, 4 Michael J. Smurthwaite,

More information

Metered Dose Inhaler Technology

Metered Dose Inhaler Technology Metered Dose Inhaler Technology Edited by Tol S. Purewal and David J. W. Grant Interpharm Press, Inc. Buffalo Grove, Illinois Contents 1. INTRODUCTION 1 T. S. Purewal Definition of the Metered Dose Inhaler

More information

Next Generation Pharmaceutical Impactor (A New Impactor for Pharmaceutical Inhaler Testing). Part I: Design ABSTRACT

Next Generation Pharmaceutical Impactor (A New Impactor for Pharmaceutical Inhaler Testing). Part I: Design ABSTRACT JOURNAL OF AEROSOL MEDICINE Volume 16, Number 3, 2003 Mary Ann Liebert, Inc. Pp. 283 299 Next Generation Pharmaceutical Impactor (A New Impactor for Pharmaceutical Inhaler Testing). Part I: Design VIRGIL

More information

Ingredient Listing Qty. Unit NDC # Supplier ml

Ingredient Listing Qty. Unit NDC # Supplier ml 8/28/2016; Page 1 SUGGESTED FORMULATION Ingredient Listing Qty. Unit NDC # Supplier Progesterone 0.25%/Spironolactone 0.5% Stock Solution 5.00 ml Medisca Foamil Base 45.00 g Alcohol (95%), USP q.s. to

More information

Implications of the new CHMP Guideline on the Pharmaceutical Quality of Inhalation and Nasal Products

Implications of the new CHMP Guideline on the Pharmaceutical Quality of Inhalation and Nasal Products Implications of the new CHMP Guideline on the Pharmaceutical Quality of Inhalation and Nasal Products Wissenschaftliche Prüfungsarbeit zur Erlangung des Titels Master of Drug Regulatory Affairs der Mathematisch-Naturwissenschaftlichen

More information

COMMENTS. Submitted by The International Pharmaceutical Aerosol Consortium

COMMENTS. Submitted by The International Pharmaceutical Aerosol Consortium COMMENTS on a draft Guidance for Industry Nasal Spray and Inhalation Solution, Suspension, and Spray Drug Products Chemistry, Manufacturing, and Controls Documentation (Docket No. 99D-1454) Submitted by

More information

IMPROVING THE REALISM AND RELEVANCE OF MOUTH-THROAT MODELS FOR INHALED PRODUCT TESTING

IMPROVING THE REALISM AND RELEVANCE OF MOUTH-THROAT MODELS FOR INHALED PRODUCT TESTING IMPROVING THE REALISM AND RELEVANCE OF MOUTH-THROAT MODELS FOR INHALED PRODUCT TESTING In this piece, Mark Copley, Sales Director of Copley Scientific, provides some background on mouth-throat models for

More information

COPLEY SCIENTIFIC. Quality Solutions for Inhaler Testing Edition. metered-dose inhalers dry powder inhalers nebulisers nasal sprays

COPLEY SCIENTIFIC. Quality Solutions for Inhaler Testing Edition. metered-dose inhalers dry powder inhalers nebulisers nasal sprays COPLEY SCIENTIFIC Quality Solutions for Inhaler Testing 2012 Edition metered-dose inhalers dry powder inhalers nebulisers nasal sprays Who are Copley Based in Nottingham, UK, Copley Scientific was founded

More information

Assessing Quality of Inhaled Products And Links to Efficacy and Safety

Assessing Quality of Inhaled Products And Links to Efficacy and Safety Assessing Quality of Inhaled Products And Links to Efficacy and Safety Prasad Peri, PhD ONDQA 2011 IPAC-RS Conference Bringing Value To The Patient In A Changing World March 30, 2011 1 Outline of the Presentation

More information

RITONAVIRI COMPRESSI RITONAVIR TABLETS. Final text for addition to The International Pharmacopoeia (July 2012)

RITONAVIRI COMPRESSI RITONAVIR TABLETS. Final text for addition to The International Pharmacopoeia (July 2012) July 2012 RITONAVIRI COMPRESSI RITONAVIR TABLETS Final text for addition to The International Pharmacopoeia (July 2012) This monograph was adopted at the Forty-sixth WHO Expert Committee on Specifications

More information

SUGGESTED FORMULATION. Lot Number. Expiry Date. Ingredient Listing Qty. Unit NDC # Supplier

SUGGESTED FORMULATION. Lot Number. Expiry Date. Ingredient Listing Qty. Unit NDC # Supplier 9/17/2016; Page 1 SUGGESTED FORMULATION Ingredient Listing Qty. Unit NDC # Supplier Aloe Vera (Powder) (Freeze Dried) 0.200 g Tacrolimus 5% Stock Powder Blend 0.600 g Ethoxy Diglycol 1.0 ml Medisca U-Mild

More information

21 Virginiamycin OH O. For chickens (except for broilers) broilers. Added amount 5~15 5~15 10~20 10~20

21 Virginiamycin OH O. For chickens (except for broilers) broilers. Added amount 5~15 5~15 10~20 10~20 21 Virginiamycin H H H H H H Virginiamycin M 1 C 28 H 35 3 7 MW: 525.6 CAS o.: 21411-53-0 Virginiamycin S 1 C 43 H 49 7 10 MW: 823.9 CAS o.: 23152-29-6 [Summary of virginiamycin] Virginiamycin (VM) is

More information

DRAFT PROPOSAL FOR THE INTERNATIONAL PHARMACOPOEIA: CARBAMAZEPINI COMPRESSI - CARBAMAZEPINE TABLETS

DRAFT PROPOSAL FOR THE INTERNATIONAL PHARMACOPOEIA: CARBAMAZEPINI COMPRESSI - CARBAMAZEPINE TABLETS December 2015 Draft document for comment 1 2 3 4 5 6 DRAFT PROPOSAL FOR THE INTERNATIONAL PHARMACOPOEIA: CARBAMAZEPINI COMPRESSI - CARBAMAZEPINE TABLETS (December 2015) REVISED DRAFT FOR COMMENT Should

More information

TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010)

TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010) June 2010 TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010) This monograph was adopted at the Forty-fourth WHO Expert Committee on Specifications for Pharmaceutical

More information

ARTESUNATE TABLETS: Final text for revision of The International Pharmacopoeia (December 2009) ARTESUNATI COMPRESSI ARTESUNATE TABLETS

ARTESUNATE TABLETS: Final text for revision of The International Pharmacopoeia (December 2009) ARTESUNATI COMPRESSI ARTESUNATE TABLETS December 2009 ARTESUNATE TABLETS: Final text for revision of The International Pharmacopoeia (December 2009) This monograph was adopted at the Forty-fourth WHO Expert Committee on Specifications for Pharmaceutical

More information

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) GUIDELINE ON THE PHARMACEUTICAL QUALITY OF INHALATION AND NASAL PRODUCTS

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) GUIDELINE ON THE PHARMACEUTICAL QUALITY OF INHALATION AND NASAL PRODUCTS European Medicines Agency Inspections London, 16 February 2005 Doc Ref.: EMEA/CHMP/QWP/49313/2005 corr. COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) GUIDELINE ON THE PHARMACEUTICAL QUALITY OF

More information

Smart asthma therapy. Patient information Spacer inhaling aid for metered-dose aerosol inhalers

Smart asthma therapy. Patient information Spacer inhaling aid for metered-dose aerosol inhalers FF Smart asthma therapy Patient information Spacer inhaling aid for metered-dose aerosol inhalers :: Metered-dose inhalers coordination technique In the treatment of asthma and COPD the medications are

More information

Process Drift and it s Resolution in the Manufacture of Drug Products. MDI s and DPI s Metered Dose Inhalations and Dry Powder Inhalations

Process Drift and it s Resolution in the Manufacture of Drug Products. MDI s and DPI s Metered Dose Inhalations and Dry Powder Inhalations Process Drift and it s Resolution in the Manufacture of Drug Products MDI s and DPI s Metered Dose Inhalations and Dry Powder Inhalations Ed Warner, Merck MMD December 2, 2010 PQRI-FDA Workshop on Process

More information

COPLEY SCIENTIFIC. Quality Solutions for Inhaler Testing. metered-dose inhalers dry powder inhalers nebulisers nasal sprays

COPLEY SCIENTIFIC. Quality Solutions for Inhaler Testing. metered-dose inhalers dry powder inhalers nebulisers nasal sprays COPLEY SCIENTIFIC Quality Solutions for Inhaler Testing 2015 Edition metered-dose inhalers dry powder inhalers nebulisers nasal sprays Who are Copley Scientific? Copley Scientific was founded in 1946 in

More information

DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA PAEDIATRIC RETINOL ORAL SOLUTION (August 2010)

DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA PAEDIATRIC RETINOL ORAL SOLUTION (August 2010) August 2010 RESTRICTED DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA PAEDIATRIC RETINOL ORAL SOLUTION (August 2010) DRAFT FOR COMMENT This document was provided by a quality control expert and was

More information

Product Guide. MSP Corporation Rice Creek Parkway, Suite 300. Shoreview, Minnesota 55126, U.S.A. Phone: Fax:

Product Guide. MSP Corporation Rice Creek Parkway, Suite 300. Shoreview, Minnesota 55126, U.S.A. Phone: Fax: Product Guide inhalertesting MSP Corporation 5910 Rice Creek Parkway, Suite 300 Shoreview, Minnesota 55126, U.S.A. Phone: 651.287.8100 Fax: 651.287.8140 sales@mspcorp.com www.mspcorp.com About This Guide

More information

Metering Valve Design and Material Selection; How can this be chosen to maximize product performance?

Metering Valve Design and Material Selection; How can this be chosen to maximize product performance? Metering Valve Design and Material Selection; How can this be chosen to maximize product performance? B. Grosjean Medical Plastics 2007 Inhalation Devices for Medicine Delivery 2007, 1-2 October 2007,

More information

Current Challenges and Opportunities in Demonstrating Bioequivalence

Current Challenges and Opportunities in Demonstrating Bioequivalence Current Challenges and Opportunities in Demonstrating Bioequivalence Gur Jai Pal Singh, Ph.D. Watson Laboratories, Inc. Corona, California, USA Demonstrating Bioequivalence of Locally Acting Orally Inhaled

More information

Quality of Nasal and Inhalation Drug Products

Quality of Nasal and Inhalation Drug Products Quality of Nasal and Inhalation Drug Products Good Development Practices, Specifications, and Analytical Methods for Inhalation Drug Products (Metered Dose Inhaler (MDI), Dry Powder Inhaler (DPI)) and

More information

Ensuring the Quality of Some Drug-Device Combination Products - FDA Perspective

Ensuring the Quality of Some Drug-Device Combination Products - FDA Perspective Ensuring the Quality of Some Drug-Device Combination Products - FDA Perspective PQRI Meeting, Rockville, MD - 2017 Ramesh Raghavachari, Ph.D. Chief, Branch I, DPMA I, OLDP OPQ 03/23/2017 OFFICE OF PHARMACEUTICAL

More information

The Montreal Protocol and the Phase-out of CFC-based Metered Dose Inhalers. Paul Krajnik UNIDO, Montreal Protocol Branch

The Montreal Protocol and the Phase-out of CFC-based Metered Dose Inhalers. Paul Krajnik UNIDO, Montreal Protocol Branch The Montreal Protocol and the Phase-out of CFC-based Metered Dose Inhalers Paul Krajnik UNIDO, Montreal Protocol Branch 1 The Montreal Protocol Binding international agreement for the preservation and

More information

Respiratory Therapy. Medical/Scientific/General Background

Respiratory Therapy. Medical/Scientific/General Background Respiratory Therapy Medical/Scientific/General Background Marketing Europe Dr. Rainer Jakobs PMM Europe 1 Dr. Rainer Jakobs, PMM Europe RT Medical/Scientific/General Background 2 Dr. Rainer Jakobs, PMM

More information

1. Module Details PENETRANT TESTING. 2. Module Purpose To enable students to develop a detailed understanding of the

1. Module Details PENETRANT TESTING. 2. Module Purpose To enable students to develop a detailed understanding of the 1. Module Details Module name Nominal duration PENETRANT TESTING One and one half modules It is anticipated that a learner holding the prescribed entry level skills will achieve the module purpose in 55

More information

LEVONORGESTREL AND ETHINYLESTRADIOL TABLETS. (January 2012) DRAFT FOR COMMENT

LEVONORGESTREL AND ETHINYLESTRADIOL TABLETS. (January 2012) DRAFT FOR COMMENT January 2012 RESTRICTED DRAFT PROPOSAL FOR The International Pharmacopoeia LEVONORGESTREL AND ETHINYLESTRADIOL TABLETS (January 2012) DRAFT FOR COMMENT This document was provided by a quality control expert.

More information

MDI. Recommended Equipment

MDI. Recommended Equipment 1 x DUSA for MDIs (Silicone Seals) 1 x Base Plate MDI 1 x DUSA for MDIs (Silicone Seals) 9 x Spare Collection Tubes for MDIs 18 x Spare Rinsing Caps for MDIs 1 x Stand for 10 Collection Tubes 1 x Waste

More information

Respiratory. Martin Jetzer DDL27 Edinburgh December 2016

Respiratory. Martin Jetzer DDL27 Edinburgh December 2016 Respiratory Investigating the Effect of the Force Control Agent Magnesium Stearate in Fluticasone Propionate Dry Powder Inhaled Formulations with Single Particle Aerosol Mass Spectrometry Martin Jetzer

More information

REVISED DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA RETINOL CONCENTRATE, OILY FORM. (August 2010)

REVISED DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA RETINOL CONCENTRATE, OILY FORM. (August 2010) August 2010 RESTRICTED REVISED DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA RETINOL CONCENTRATE, OILY FORM (August 2010) DRAFT FOR COMMENT This document was provided by a quality control expert

More information

8/13/11. RSPT 1410 Humidity & Aerosol Therapy Part 3. Humidification Equipment. Aerosol Therapy

8/13/11. RSPT 1410 Humidity & Aerosol Therapy Part 3. Humidification Equipment. Aerosol Therapy 1 RSPT 1410 Humidity & Aerosol Therapy Part 3 Wilkins: Chapter 35, p. 775-799 Cairo: Chapter 4, p. 88-143 2 Humidification Equipment A humidifier is a device that adds molecular liquid (e.g. water vapor)

More information

Effect of Interval Between Actuations of Albuterol Hydrofluoroalkane Pressurized Metered-Dose Inhalers on Their Aerosol Characteristics

Effect of Interval Between Actuations of Albuterol Hydrofluoroalkane Pressurized Metered-Dose Inhalers on Their Aerosol Characteristics Effect of Interval Between Actuations of Albuterol Hydrofluoroalkane Pressurized Metered-Dose Inhalers on Their Aerosol Characteristics Ariel Berlinski MD and David Pennington BACKGROUND: Albuterol hydrofluoroalkane

More information

Ingredient Listing Qty. Unit NDC # Supplier. q.s. to ml

Ingredient Listing Qty. Unit NDC # Supplier. q.s. to ml 12/20/2018; Page 1 SUGGESTED FORMULATION Ingredient Listing Qty. Unit NDC # Supplier, USP 10 MU Sucrose, NF 23.00 g Propylene Glycol, USP 7.0 ml Medisca Oral Mix (Flavored Suspending Vehicle) 50.0 ml Medisca

More information

A NEW, ADVANCED HIGH- THROUGHPUT SYSTEM FOR AUTOMATED INHALER TESTING

A NEW, ADVANCED HIGH- THROUGHPUT SYSTEM FOR AUTOMATED INHALER TESTING A NEW, ADVANCED HIGH- THROUGHPUT SYSTEM FOR AUTOMATED INHALER TESTING Two years ago, Novi Systems Ltd set out to shake up the inhaler automation market. On December 8th, 2015, at the Drug Delivery to the

More information

ZIDOVUDINE, LAMIVUDINE AND ABACAVIR TABLETS Draft proposal for The International Pharmacopoeia (September 2006)

ZIDOVUDINE, LAMIVUDINE AND ABACAVIR TABLETS Draft proposal for The International Pharmacopoeia (September 2006) September 2006 RESTRICTED ZIDOVUDINE, LAMIVUDINE AND ABACAVIR TABLETS Draft proposal for The International Pharmacopoeia (September 2006) This document was provided by a contracted quality control laboratory.

More information

Drug/Device Combination Products: Bioequivalence

Drug/Device Combination Products: Bioequivalence Drug/Device Combination Products: Bioequivalence Three stories:. The story of Nasal and Inhalation Product BE 2. The story of the Generic Auto-Injector 3. The story of User Interface Considerations Bioequivalence

More information

NEBULIZERS, METERED DOSE INHALERS, AND DRY POWDER INHALERS

NEBULIZERS, METERED DOSE INHALERS, AND DRY POWDER INHALERS NEBULIZERS, METERED DOSE INHALERS, AND DRY POWDER INHALERS Douglas S. Gardenhire, Ed.D, RRT-NPS MODULE 1 Manipulate Small Volume Nebulizers by Order or Protocol 1 Objectives for Module 1 At the end of

More information

21/03/2011 AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY. Fundamentals of aerosols

21/03/2011 AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY. Fundamentals of aerosols AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY AEROSOL DEPOSITION AND THE ASSESSMENT OF PULMONARY DRUG DELIVERY Steve Newman Scientific Consultant Norfolk, UK steve.newman@physics.org

More information

Ingredient Listing Qty. Unit NDC # Supplier TBD. Sterile Preparation

Ingredient Listing Qty. Unit NDC # Supplier TBD. Sterile Preparation 8/16/2018; Page 1 SUGGESTED FORMULATION Acyclovir, USP Ingredient Listing Qty. Unit NDC # Supplier Mineral Oil (Light), NF 3.0 ml Polyox TM WSR-301 Oral Adhesive Paste* *Formula # F 006 397 available on

More information

Cascade Impactor Study Comparing the Amsino OneMask Oxygen Mask and Predicate Mask from Hudson RCI on a Human Model

Cascade Impactor Study Comparing the Amsino OneMask Oxygen Mask and Predicate Mask from Hudson RCI on a Human Model (p) 93-850-6630 (f) 93-850-6635 Cascade Impactor Study Comparing the Amsino OneMask Oxygen Mask and Predicate Mask from Hudson RCI on a Human Model Abstract This in vitro study compares drug delivery efficiencies

More information

Preparation and Administration

Preparation and Administration Preparation and Administration A guide for healthcare professionals XOLAIR IS INDICATED FOR: Adults and adolescents (aged 12 years) with moderate-to-severe persistent asthma who have a positive skin test

More information

Senior Scientist / Principal Scientist. Compounding Manager (Sterile and non-sterile specials) Senior Pharmaceutical Assessor

Senior Scientist / Principal Scientist. Compounding Manager (Sterile and non-sterile specials) Senior Pharmaceutical Assessor Curriculum Vitae PERSONAL INFORMATION Andrew Twitchell WORK EXPERIENCE May 1987 September 1991 Senior Scientist / Principal Scientist 3M Health Care Ltd, (United Kingdom) Formulation, scale-up and manufacture

More information

Understanding Regulatory Global Requirements for Nasal Drug Products. Julie D. Suman, Ph.D. April 8, 2016

Understanding Regulatory Global Requirements for Nasal Drug Products. Julie D. Suman, Ph.D. April 8, 2016 Understanding Regulatory Global Requirements for Nasal Drug Products Julie D. Suman, Ph.D. April 8, 2016 AGENDA NDA vs ANDA Regulatory Approaches for Bioequivalence (BE) FDA Drug Specific Guidances FDA,

More information

The Use of Physics-Based Modeling to Better Design Drug- Device Interface. Yoen-Ju Son, PhD Merck Research Laboratory, Summit, NJ

The Use of Physics-Based Modeling to Better Design Drug- Device Interface. Yoen-Ju Son, PhD Merck Research Laboratory, Summit, NJ The Use of Physics-Based Modeling to Better Design Drug- Device Interface Yoen-Ju Son, PhD Merck Research Laboratory, Summit, NJ Presentation Outline Physics-based modeling in pharmaceutical industry Pulmonary

More information

Varroa Mite Detection Methods

Varroa Mite Detection Methods Varroa Mite Detection Methods Apiculture Bulletin # 222 Updated: 10/15 Effective mite control is dependent on frequent and reliable mite detection. Varroa mites spread rapidly between hives and apiaries

More information

Formulation and Development of Metered Dose Inhalations of Salbutamol in Solution Form

Formulation and Development of Metered Dose Inhalations of Salbutamol in Solution Form Research Paper Formulation and Development of Metered Dose Inhalations of Salbutamol in Solution Form ANUBHA KHALE* AND AMRITA BAJAJ 1 M. E. S. Instititute s, H. K. College of Pharmacy, Pratikshanagar,

More information

Rebaudioside a From Multiple Gene Donors Expressed in Yarrowia Lipolytica

Rebaudioside a From Multiple Gene Donors Expressed in Yarrowia Lipolytica Residue Monograph prepared by the meeting of the Joint FAO/WHO Expert Committee on Food Additives (JECFA), 82 nd meeting 2016 Rebaudioside a From Multiple Gene Donors Expressed in Yarrowia Lipolytica This

More information

Issued by: LABORATORY MANAGER Original Date: March 20, 2000 Approved by: Laboratory Director Revision Date: July 26, 2000 CRYPTOCOCCAL ANTIGEN

Issued by: LABORATORY MANAGER Original Date: March 20, 2000 Approved by: Laboratory Director Revision Date: July 26, 2000 CRYPTOCOCCAL ANTIGEN Policy # MI\TECH\11\v01 Page 1 of 6 Issued by: LABORATORY MANAGER Original Date: March 20, 2000 Approved by: Laboratory Director Revision Date: July 26, 2000 CRYPTOCOCCAL ANTIGEN Latex particles coated

More information

EUDRAGIT L 100 and EUDRAGIT S 100

EUDRAGIT L 100 and EUDRAGIT S 100 Technical Information EUDRAGIT L 100 and EUDRAGIT S 100 Specification and Test Methods Ph. Eur. Methacrylic Acid - Methyl Methacrylate Copolymer (1:1) Methacrylic Acid - Methyl Methacrylate Copolymer (1:2)

More information

Metered Dose Inhalation Formulations of Salbutamol Sulphate Using Non-CFC Propellant Tetrafluoroethane

Metered Dose Inhalation Formulations of Salbutamol Sulphate Using Non-CFC Propellant Tetrafluoroethane Available online at www.ijpsdr.com International Journal of Pharmaceutical Sciences and Drug Research 2011; 3(4): 292-296 Research Article ISSN 0975-248X Metered Dose Inhalation Formulations of Salbutamol

More information

SELECTING A DOSAGE FORM FOR DRUG DELIVERY TO THE LUNGS. Dr. Jay T. Holt, Ph.D., Sr. Director, Inhalation & Analytical

SELECTING A DOSAGE FORM FOR DRUG DELIVERY TO THE LUNGS. Dr. Jay T. Holt, Ph.D., Sr. Director, Inhalation & Analytical SELECTING A DOSAGE FORM FOR DRUG DELIVERY TO THE LUNGS Dr. Jay T. Holt, Ph.D., Sr. Director, Inhalation & Analytical www.cirruspharm.com February 2014 INTRODUCTION Cirrus, a Kemwell company, is frequently

More information

Hepatitis A virus IgM ELISA Kit

Hepatitis A virus IgM ELISA Kit Hepatitis A virus IgM ELISA Kit Catalog Number KA0285 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle

More information

Misty Max 10 nebulizer

Misty Max 10 nebulizer AirLife brand Misty Max 10 nebulizer Purpose Introduction Delivery of nebulized medication to the lungs is a complex process dependant upon a variety of clinical and device-related variables. Patient breathing

More information

Study of the British Pharmacopeia method on methimazole (thiamazole) content in carbimazole tablets

Study of the British Pharmacopeia method on methimazole (thiamazole) content in carbimazole tablets Journal of Pharmaceutical and Biomedical Analysis 16 (1998) 785 792 Study of the British Pharmacopeia method on methimazole (thiamazole) content in carbimazole tablets Maria Aletrari *, Popi Kanari, Dora

More information

The pulmonary route is gaining increasing

The pulmonary route is gaining increasing ELECTRONICALLY REPRINTED FROM FEBRUARY 2014 Pulmonary Drug Delivery Particle Engineering for Inhaled Therapeutics Moderated by Adeline Siew, PhD Industry experts discuss the various factors affecting drug

More information

Methimazole 5 mg Oral Chewable Treats (Solid Suspension, 60 x 1 ml Treats) FIN F v2. Ingredient Listing Qty. Unit NDC # Supplier

Methimazole 5 mg Oral Chewable Treats (Solid Suspension, 60 x 1 ml Treats) FIN F v2. Ingredient Listing Qty. Unit NDC # Supplier 11/23/2015; Page 1 SUGGESTED FORMULATION Ingredient Listing Qty. Unit NDC # Supplier Methimazole, USP 0.300 g Tuna Flavor 1.50 g Glycerin, USP** 6.0 ml Chew-A-Treat Compound A 22.80 g Chew-A-Treat Compound

More information

Ingredient Listing Qty. Unit NDC # Supplier. Sterile Preparation

Ingredient Listing Qty. Unit NDC # Supplier. Sterile Preparation 2/8/2014; Page 1 SUGGESTED FORMULATION Ingredient Listing Qty. Unit NDC # Supplier Baclofen, USP 1.000 g Cyclobenzaprine Hydrochloride, USP 1.000 g Diclofenac Sodium, USP 1.500 g Lidocaine Hydrochloride,

More information

WHAT IS NSP? ( ANTIFINGERPRINTS NANOCOATING by The Inox in Color )

WHAT IS NSP? ( ANTIFINGERPRINTS NANOCOATING by The Inox in Color ) WHAT IS NSP? ( ANTIFINGERPRINTS NANOCOATING by The Inox in Color ) NSP (Non Stain Processing) No fingerprint mark process Fingerprints can be a major problem on some stainless steel applications with public

More information

Sub-components in Nicotine cessation products?

Sub-components in Nicotine cessation products? Sub-components in Nicotine cessation products? Mr Arun Sarda, Director of Quality and Regulatory Affairs Member of the HEITKAMP & THUMANN GROUP Regulatory approval of sub-components in nicotine cessation

More information

VARIABILITY IN CASCADE IMPACTION: SOURCES, IMPACT AND STRATEGIES FOR REDUCTION

VARIABILITY IN CASCADE IMPACTION: SOURCES, IMPACT AND STRATEGIES FOR REDUCTION H&T Copley Presspart Scientific VARIABILITY IN CASCADE IMPACTION: SOURCES, IMPACT AND STRATEGIES FOR REDUCTION In this article, Mark Copley, Chief Executive Officer, Copley Scientific, considers cascade

More information

Emulsions. Purpose of emulsions and of emulsification:

Emulsions. Purpose of emulsions and of emulsification: Pharmacist Ghada Hamid Emulsions Emulsion is a dispersion in which the dispersed phase is composed of small globules of a liquid distributed throughout a vehicle in which it is immiscible. The dispersed

More information

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON LAMIVUDINE TABLETS

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON LAMIVUDINE TABLETS RESTRICTED INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON LAMIVUDINE TABLETS DRAFT FOR COMMENT Please address any comments you may have on this document, by 12 July 2006, to Dr S. Kopp, Quality Assurance and

More information

Aerosol Delivery Devices

Aerosol Delivery Devices Aerosol Delivery Devices Judy Bielenda LRRT Respiratory Therapy Clinical Specialist Pediatric Asthma Education Coordinator University of Michigan C.S. Mott Children s and Von Voigtlander Women s Hospital

More information

Mouse C-Peptide ELISA Kit

Mouse C-Peptide ELISA Kit Mouse C-Peptide ELISA Kit Cat.No: DEIA4507 Lot. No. (See product label) Size 96T Intended Use The Mouse C-Peptide ELISA kit is for the quantitative determination of c-peptide in mouse serum, plasma, and

More information