Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Size: px
Start display at page:

Download "Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams"

Transcription

1 Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health Sciences, Health Campus USM, Universiti Sains Malaysia, Kubang Kerian, Kelantan, Malaysia b Department of Nuclear Medicine, Radiotherapy and Oncology, School of Medical Sciences, Health Campus, Universiti Sains Malaysia,16150 Kubang Kerian, Kelantan, Malaysia *Corresponding author: reduan@usm.my ABSTRACT: Particular attention may be paid to reassess for treatment planning systems (TPSs) that deal with specialized techniques such as stereotactic radiosurgery. It is important for clinical medical physicist to be able to quantify errors involved in the dosimetric parameter data before it can be downloaded into the TPS. The purpose of this study was to verify the measurement of relative output factor (ROF), which is one of the dosimetric parameter data required for Stereotactic Radiosurgery (SRS) treatment planning. ROF of 5 mm to 45 mm diameter Radionic circular cone collimators were measured using three types of ionization chambers, which were A12 Extradin Farmer-type Chamber (0.65 cc), A14 Extradin Microchamber (0.016 cc) and PTW Pinpoint Ionization Chamber (0.016 cc). The ROF measurements were verified using a Monte Carlo (EGSnrc) validated model and other previous studies. In conclusion, pinpoint ionization chamber gave the most reliable ROF result for field size from 5 mm to 45 mm compared to other ionization chambers used in this study. The ROF measurement using pinpoint ionization chamber was in good agreement with the Monte Carlo calculation (± 4.9% percentage deviation). Therefore for measurement of ROF for SRS circular field size, pinpoint ionization chamber was suggested due to its accuracy. This ROF validation test can be carried out during the commissioning and installation of the new TPS and also for annually quality assurance (QA). 20

2 Keywords: Relative Output Factor (ROF), Stereotactic Radiosurgery (SRS), Monte Carlo, Small field dosimetry Introduction For installing and commissioning a new stereotactic radiosurgery (SRS) treatment planning computer, a comprehensive set of beam data of the linear accelerator (LINAC) machine must be measured and then downloaded into the treatment planning software (Abdullah et al., 2015). Three basic beam parameters to be measured for dose calculation are tissue phantom ratios (TMR) or percentage depth dose (PDD), relative output factors (ROF) and single beam profiles. There are two crucial components in SRS in order to achieve the aims of a treatment planning. The first component is to deliver a precise and uniform dose to the target lesion with multiple non-coplanar beams. The second factor is the accuracy in patient positioning, which can be achieved by introducing the stereotactic apparatus. With the combination of these two factors, SRS can provide uniform dose and sharp dose falloff for treating small lesion and at the same time minimizing the damage to critical organs surrounding the lesion (Khan et al., 2011). Although dose calculation can be carried out using dosimetric parameters as beam data, the method of collecting beam data presents challenges to the physicists. In particular, narrow stereotactic radiosurgery beams with sharp dose gradient require detectors and chambers with high spatial resolution. For this reason, the smallest detector available should be used when performing small field dosimetry. For central axis measurement such as depth dose, tissue phantom ratio and ROF, the detector dimensions should be significantly smaller than the field sizes. ROF is the ratio of output for a stereotactic radiosurgery cone diameter size to the reference condition (field size 10 x 10 cm 2 at the reference depth inside the water phantom and at 100 cm of source to surface distance (SSD)). Particular attention may be paid to reassess for treatment planning systems (TPSs) that deal with specialized techniques such as stereotactic radiosurgery. It is important for clinical medical physicist to be able to quantify errors involved in the input data (measured beam data) before it can be downloaded into the TPS. All input data should be correct and accurate before it can be used clinically. In this study, ROF verification is carried out to make sure the 21

3 input data is accurate by comparing the measured data of ROF and ROF calculated using MC. This ROF validation test can be carried out during the commissioning and installation of new TPS and also for annually quality assurance (QA). Materials and Methods Measurement of ROF In this study, the main focus was to measure the relative output factor (ROF) for Radionic circular cones used for SRS treatment from 5 mm to 45 mm. The measurement of ROF was conducted inside a mini water phantom with the dimension of 35 cm x 35 cm x 50 cm. Firstly, the gantry and collimator position was positioned at zero degree and SSD was at 100 cm. Then, the pinpoint ionization chamber was placed 5 cm from the water surface (Figure 1). The 6 MV photon used was energized by Primus (Siemens Medical, USA) linear accelerator. The irradiations were repeated three times and the average values were calculated. The circular cone of 45 mm diameter was then replaced with other circular cones available in the department and the procedure was repeated with smaller circular cone size. The ROF for each circular cone was calculated by finding the ratio of charge obtained at 50 mm depth to the charge (C w,ic,a ), obtained at similar depth for reference field size (100 x 100 mm 2 ) (C w,ic,ref ). ROFs were calculated using the formula from Equation 1: ROF (A) = C w,ic,a / C w,ic,ref (Equation 1) The procedures were repeated by using different types of ionization chamber: A14 microchamber and A12 Farmer type ionization chamber. 22

4 Figure 1: The experimental setup for relative output factor measurement Modelling and Simulation Recently, Monte Carlo (MC) calculation method has reported as most accurate method of predicting dose distribution in radiotherapy (Abdul Aziz et al., 2011). In this present work, MC simulation and calculation is a useful tool in evaluating the ROF measurement results. Calculation of ROF for Radionic circular cones used for SRS treatment from 5 mm to 45 mm was performed using BEAMnrc and DOSXYZnrc (National Research Council Canada, NRCC) source code (EGSnrc package) on personal desktop. BEAMnrc is a general purpose MC simulation system for modelling radiotherapy source while DOSXYZnrc is used to calculate dose in three dimensional (3D) medium (Abdul Aziz et al., 2011). The simulations were performed in three stages. BEAMnrc was used to simulate the Siemens Primus linac head to obtain phase space data for the first 20 cm after the target, right after the projection mirror. This phase space file was used in the subsequent stereotactic cones and 100 x 100 mm 2 field sizes. In this stage, 50 million particle histories were used. In the second stage, all thirteen Radionic stereotactic cones diameter starting from 5 mm to 45 mm were simulated. The cone was simulated as stainless steel materials with a density of 8.03 g/cm 3. The geometry of each cone was measured manually since the technical drawing of the component was unavailable. The Radionic assembly holder and the cones were simulated using BEAMnrc s flattening filter (FLATFILT) component module. The 50 x 50 mm 2 (phase 23

5 space file was used for the second stage) and space phase files were generated for all cones at SSD = 80 cm. The values for energy cutoff for photon (PCUT) and electron (ECUT) transport were set to 0.01 and 0.7 MeV, respectively. The phase space files generated during the second stage of simulations were used in the final stage to calculate the three dimensional (3D) dose distributions in water phantom using DOSXYZ software. Results and Discussion The measurement of ROF with 6 cm x 6 cm MLC size in comparison with Sharma et al. (2007) is presented in Figure 2. ROFs were measured for both studies using the same type of pinpoint ionization chamber. In general, both studies showed good agreement of ROF values for field size from 12.5 mm to 45 mm. The highest difference was 2.1% for field size 22.5 mm. X jaws or multi-leaves collimator (MLC) and Y jaws were shaped according to the radiation field size. It was important to verify by inspection that the area outside the central part of the conical collimators was completely shut by the jaws. The X and Y jaws were also ensured to not overlap with the circular collimator opening. This situation could happen when the MLC is stuck and not calibrated properly. The advantage of using smaller jaws size is to avoid radiation leakage at the Linac s head. Contributions to the output factor or scatter factor may arise from two groups of photon beams. Primary photons arise from the Linac s head, while scattered photons arise from collimation devices including Y jaws, MLC, collimator housing and circular cone. ROF are strongly dependent on the field size and the energy of the photon beam. If more photons pass through the collimator opening and reach the point of interest inside the water medium, then the lateral scatter equilibrium effect is said to exist, which arises from the collimate field size. As the diameter of field size increases, the primary photons will remain the same, but the contribution from the scattered photon from the cone size differs. In small field dosimetry, less contribution of scattering radiation from the medium (water), called the phantom scatter, is observed. The effect of phantom scatter on increased field size was also explained in Khan (1994). The explanation is general and agreed with what happens in the small field size dosimetry. The effect of field size on the scatter factor due to phantom scatter alone is significant as long as 24

6 ROF the distance between the point of measurement and the edge is shorter than the range of the lateral scatter electron produced. When the certain distance is reached, there is no further increased in the scatter factor caused by the phantom scatter. When the field size is reduced below the level required for lateral scatter equilibrium, the scatter factor decreases rapidly Cone Diameter (mm) Measurement Sharma et al Figure 2: ROF vs. Radionic cone diameter measured using pinpoint ionization chamber. Figure 3 illustrates our ROF results compared to that of the study done by Shepard (2009) which measured ROF using BrainLab assembly holding the cones and circular cones. The graph shows a small variation difference between both studies for field size from 17.5 mm to 30 mm. Stereotactic radiosurgery beams exhibit a sharp decrease in the output with decreasing field size (Shepard, 2009). It is obviously observed in Figure 3 that ROF of small fields showed strong field size dependence with rapidly decreasing ROF as the field diameter decreased (Sharma et al., 2007). This is due to the rapid decrease in the primary photon for field size smaller than lateral electron range, where the lateral electron equilibrium no longer exists. Figure 3 also reveals that ROF of BrainLab (BrainLab, 2012) and Radionic circular cones were close to each other (within 2%) for every cone diameters. The slight variation in the ROF of the SRS circular cones from these two manufacturers may be due to the difference in the physical dimension and material of the cone (Sharma et al., 2007). Radionic s collimator housing geometry was longer compared to the collimator housing of BrainLab, which caused the collimation of field to be closer to the isocenter. The distance from the isocenter to the 25

7 ROF lower end of the mounted collimator (with gantry in 0 degree angle) was 230 mm. It allows for more primary radiation and scattering radiation coming from the circular cone collimator to reach the detector and reduce the scattering radiation produced by the interaction with air because the air gap between the surface of the circular cones and the isocenter was reduced. Khelashvili et al. (2012) reported that for small fields, the x-ray target is partially occluded by the field boundaries, due to the divergence of the small fields. In effect, the extrapolations of the openings toward the target converge slightly above the target position. Therefore, the projection of the field opening at the position of the target is smaller than the focal spot size and a partial occlusion of x-ray source occurs. If part of the target is occluded, the geometrical penumbras from opposite edges of the fields are overlapped, resulting in less ROF value in comparison to the bigger field size, in which the entire focal spot on the target is viewed by the detector. Hence, this occlusion effect is strongly dependent on the design of the collimator housing with the circular cones and also the divergence of the small field to the target Cone Diameter (mm) Measurement Shepard 2009 Figure 3: ROF vs. cone diameter of different SRS cones manufacturers Figure 4 shows the measurement of ROF with different sizes of multi-leaves collimator (MLC) opening. The graph shows the variation of ROF with circular cone field sizes, from 5 mm to 45 mm, measured using pinpoint ionization chamber with different MLC sizes. From Figure 4, it is shown that the ROF values for 5 cm x 5 cm MLC size was consistently low compared to ROF values of 6 cm x 6 cm MLC size except for the smallest cone, 5 mm. The 26

8 ROF ROF values of 6 cm x 6 cm MLC size were 1.06% higher than ROF of 5 cm x 5 cm MLC size. Smaller MLC size caused larger amount primary beams to pass through. It also reduced the production of scattered photon by limiting the collision between the primary beam to MLC and Y jaws inside the Linac s head before passing through the circular cone and reaching the target area cm x 6 cm 5 cm x 5 cm Cone Diameter (mm) Figure 4: ROF for circular cone diameters with diffrent MLC size Further comparison of ROF for circular cone field sizes measured using three different types of ionization chambers was plotted in Figure 5. Generally, ROF values increased with field sizes. From Figure 5, it is observed that ROF measured using pinpoint ionization chambers showed small deviation of difference compared to ROF calculated using the MC method. ROF values measured using the pinpoint chamber were consistently lower than those measured using A14 microchamber for all Radionic radiosurgery cones sizes, except for the 5 mm cone. However, ROF measured using A12 Farmer type ionization chamber showed large differences compared to those measured using both pinpoint and A14 microchamber for all field sizes. The differences range from 35% to 80%, except for the largest circular cone field size, 45 mm (3.9% difference). This was because in smaller field, less scattering photons were produced due to fewer lateral scattering from the primary radiation. As a result, fewer ion pair produced and small amount of charge was collected by the ionization chamber. For 27

9 ROF ionization chambers that have larger active volume such as A12 Farmer, the charge collected by the ionization chamber was very small and thus leading to the volume averaging effect Cone Diameter (mm) MC PP A12 A14 Figure 5: ROF measured using different types of ionization chambers Table 1: Results of ROF measured using different types of ionization chambers and ROF calculated with MC A14 A12 PP MC Cone (mm) ROF ROF ROF ROF

10 Table 2: The percentage deviation values of ROF measured using A14, A12 and pinpoint ionization chambers compare with MC-calculated A14/ MC A12/ MC PP/ MC Cone (mm) % % % A comparison of ROF for circular fields measured using different types of ionization chamber detectors and MC-calculation is summarized in Table 1 and Table 2. The results of ROF measured using A12 was too small for 5 mm field size compared to both A14 and pinpoint measurement. ROF measured using A14 microchamber compared with MCcalculation, with less than 7 % deviation for all measurements except for 5mm cone size. ROF measured using pinpoint ionization chamber showed a good agreement with MCcalculation, with less than 5 % deviation for all measurements. Conclusion ROF measurement must be done carefully than the usual care for small field sizes dosimetry. Smaller field can cause greater error. This study indicates that the absence of lateral electronic equilibrium due to small size circular cone affect the result of ROF measured. Besides that, other factors such as the volume averaging effect due to the detector itself, energy dependence and misalignment of positioning the detector rightly at the central axis of the beam during ROF measurement can contribute to ROF errors. Even though the detector was considered as small, the size was probably large enough for small beam dosimetry. 29

11 From our study, pinpoint ionization chamber gave the most reliable result for ROF for field size from 5 mm to 45 mm compared to other ionization chambers which were A14 microchamber and A12 Farmer type chamber. The results were also supported by the agreement with MC-calculation and previous measurements conducted by Shepard (2009) and Sharma et al. (2007). Therefore for measurement of ROF for SRS circular field size, pinpoint ionization chamber was suggested due to its accuracy. Acknowledgements The authors would like to thank all oncologists, physicists and therapy radiographers in the Department of Nuclear medicine, radiotherapy and oncology of the Hospital Universiti Sains Malaysia and also all supporting staff in Medical Radiation Programme of School of Health Sciences for their cooperation and help. In addition, we would like to express our gratitude to Dr. Ahmad Lutfi Yusof for the MC simulations. References Abdul Aziz, M.Z., Salikin, M.S., Yusoff, A.L. and Abdullah, R. (2011). Monte Carlo simulation of electron beam in 3D water phantom. World Academy of Science and Technology, 60: Abdullah, R., Nik Idris, N.R., Zakaria, A., Yusoff, A.L., Mohamed, M., and Mohsin, N.I.(2015). Measurement of dosimetric parameters and dose verification in stereotactic radiosurgery (SRS). Jurnal Sains Kesihatan Malaysia, 13(1): BrainLab Physic. (2012). Technical Reference Guide Rev USA: BrainLab. Khan, F.M. (1994). The Physic s of Radiation Therapy. Second Edition. USA: William and Wilkins Khan, F.M., Gibbons, J., Mihailidis, D., and Alkhatib, H. ( 2011). Khan 's Lectures: Handbook of the Physics of Radiation Therapy. USA: Lippincot Williams and Wilkins. 30

12 Khelashvili, G., Chu, J., Diaz, A. and Turian, J. (2012). Dosimetric characteristics of small diameter BrainLab cones used for stereotactic radiosurgery. Journal of Applied Clinical Medical Physics, 13(1): Sharma, S.D., Kumar, S., Dagaonkar, S.S., Bisht, G., Dayanand, S., Devi, R., Deshpande, S.S., Chaudry, S., Bhatt, B.C. and Karman, S. (2007). Dosimetric comparision of linear accelerator-based stereotactic radiosurgery system. Journal of Medical Physics, 32(1): Shepard, D. ( 2009). Quality assurance in stereotactic radiosurgery and fractionated stereotactic radiotherapy. AAPM 51st Annual Meeting. California, July. 31

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams.

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams. IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams. To cite this

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

SRS/SBRT Errors and Causes

SRS/SBRT Errors and Causes SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS SB Crowe, E Whittle, C Jones, T Kairn SMALL FIELD DOSIMETRY The measurement of small field tissue-maximum ratio (TMR) and percentage depth dose (PDD)

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC) Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf

More information

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Basic dosimetric data for treatment planning system

Basic dosimetric data for treatment planning system Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny

More information

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS} High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate

More information

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Sang Hyun Cho and Geoffrey S. Ibbott Department of Radiation Physics The

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad

More information

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

More information

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams Ernesto Mainegra-Hing a) and Bryan R. Muir Measurement Science and Standards, National Research Council of Canada,

More information

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms 1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,

More information

ph fax

ph fax This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY

More information

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.

More information

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the

More information

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland IGRT1 technologies Paweł Kukołowicz Warsaw, Poland Minimal prerequisite for good, efficient radiotherapy ICTP 2015 Paweł Kukołowicz 2/29 Minimal prerequisite for good, efficient radiotherapy Well trained

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

A new geometric and mechanical verification device for medical LINACs

A new geometric and mechanical verification device for medical LINACs JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein

More information

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Rhys Carrington, Ysbyty Glan Clwyd, Bodelwyddan. MSc Medical Radiation Physics. Small field dosimetry

More information

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery Therapy SAM Symposium: WE-A-BRCD-1 Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery Kamil M. Yenice, PhD Associate Professor Chief of Clinical Physics

More information

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system JBUON 2018; 23(1): 179-184 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Patient-specific quality assurance for intracranial cases in robotic

More information

Expectations of Physics Knowledge for Certification

Expectations of Physics Knowledge for Certification Expectations of Physics Knowledge for Certification Bhudatt Paliwal, Ph.D. University of Wisconsin Medical School Department of Human Oncology Madison, Wisconsin 53792 USA Guiding Principles Test that

More information

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel

More information

Dosimetric characteristics of the small diameter BrainLab TM cones used for stereotactic radiosurgery

Dosimetric characteristics of the small diameter BrainLab TM cones used for stereotactic radiosurgery JOURNL OF PPLIED CLINICL MEDICL PHYSICS, VOLUME 13, NUMBER 1, 2012 Dosimetric characteristics of the small diameter BrainLab TM cones used for stereotactic radiosurgery Gocha Khelashvili, 1a James Chu,

More information

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National

More information

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM A Thesis Presented to The Academic Faculty by Ryan J. Underwood In Partial Fulfillment of the Requirements for the Degree Master of Science in the School

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West

More information

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)

More information

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD Proceedings of the Tenth EGS4 Users' Meeting in Japan, KEK Proceedings 2002-18, p.65-73 DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD N. Tohyama, H. Saitoh, T. Fujisaki 1, S. Abe 1

More information

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D. ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael

More information

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Dosimetric characterization of surface applicators for use with the Xoft ebx system Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal

More information

Online Care, Online Control, Online Confidence

Online Care, Online Control, Online Confidence Online Care, Online Control, Online Confidence * Online Treatment Monitoring and much more! Designed for Efficiency provides you with online control and online confidence of treating your patients exactly

More information

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 37-013 GARD Users Manual March 2005 Manual No. 37-013-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D. Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation

More information

IMRT point dose measurements with a diamond detector

IMRT point dose measurements with a diamond detector IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,

More information

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,

More information

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,

More information

IMRT QA: Point Dose Measurements or 2D Array?

IMRT QA: Point Dose Measurements or 2D Array? QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify

More information

SRS Uncertainty: Linac and CyberKnife Uncertainties

SRS Uncertainty: Linac and CyberKnife Uncertainties SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis

More information

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry Journal of Physics: Conference Series PAPER OPEN ACCESS Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry To cite this article: N A Costa et al 2016 J. Phys.: Conf. Ser. 733 012098 View

More information

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

I. Equipments for external beam radiotherapy

I. Equipments for external beam radiotherapy I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,

More information

Electron Beam ET - Reloaded Therapy - Reloaded

Electron Beam ET - Reloaded Therapy - Reloaded 8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY E. BADITA 1, C.VANCEA 1,3, I. CALINA 1,3, D. STROE 2, M. DUMITRACHE 2,3,

More information

Much of what we will say

Much of what we will say New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer

More information

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis Acknowledgments Thanks to Saiful Huq for an illuminating conversation about the application of TG-100 Jennifer O Daniel, Ph.D. & Fang-Fang Yin, Ph.D. Duke University Medical Center Annual AAPM Meeting,

More information