BELdART-2: Moving towards safer radiotherapy. Steven Lelie 1,2, Wouter Schroeyers 1, Brigitte Reniers 1, Sonja Schreurs 1
|
|
- Amelia Hodge
- 6 years ago
- Views:
Transcription
1 BELdART-2: Moving towards safer radiotherapy Steven Lelie 1,2, Wouter Schroeyers 1, Brigitte Reniers 1, Sonja Schreurs 1 1 Nucleair Technologisch Centrum (XIOS/Uhasselt), Wetenschapspark 27, 3590 Diepenbeek 2 BEFY (Vrije Universiteit Brussel), Laarbeeklaan 101, 1090 Jette International Symposium 50 years BVS-ABR, April 10 th, Brussels
2 TOC Introduction BELdART-1 BELdART-2 The future Conclusions
3 Introduction Succes/Failure absorbed doses Centers do QA What if: Wrong calibration Faulty equipment Need for an external audit
4 Introduction (2) What was/is available? (not free of charge) IAEA & WHO: TLD audit (since 1969) World wide TLD audits Postal audit for photon beams Focus on developing countries ESTRO-QUALity assurance network (since 1998) European postal audit with TLD Photon & Electron beams RPC-MD Anderson (1968) World wide TLD audits Moving towards OSL
5 Introduction (3) Starting from 2009 FANC asked for a national audit program for reference and non reference conditions via public tender NuTeC was selected to perform a national audit BELdART was born (feb 2009) No cost for center A total of 34 centers 91 clinical machines
6 BELdART-1 Visited audit of all centers (+ satellites) Auditing of Basic mechanical parameters Photon beams Electron beams Equipment independent of hospital
7 BELdART-1 (2) BELdART was unique: Visited national audit including all centers All centers participated with at least 1 accelerator Used the L-α-alanine dosimetry system
8 The alanine system ) ( 1 0 T T c k T T t c f t e k b sleeve sleeve b f f b T T Q Q M j j M j j b b m D k k k k k k k k with k D m m A 0, ) (7.0 d c t ) (1.8 K c T D w expressed as Dose Normalized Amplitude D w calculated with calibration Curve b ad w b D Q Co b D c w k D D 60 0 Base function using reference dosimeters of 0 Gy and 20 Gy Calibration using dosimeters in the range 2 Gy 25 Gy
9 The alanine system Measurement uncertainties Part of continuous internal QA procedure 5 measurements of calibration set over several months
10 The alanine system Uncertainty budget (4 Harwell detectors; 5 rotations, 4 Gy): Base function detectors (20 Gy): Dose (primary standard) 0.20% Amplitude: (A b D) 0.15% Mass: ( 50mg) 0.04% Field detector (4 Gy): Amplitude: (A D ; 30 mgy = worst case) 0.75% Mass:( 50mg) 0.04% Experimental conditions: Fading: 0.02% Irr temp: 0.03% Beam quality: 0.27% Positioning 0.05% Combined standard uncertainty 0.84%
11 Audit protocol Mechanical parameters Position of the isocenter Optical distance indicator Position of the laser lines Correspondence light-radiation field
12 Audit protocol (2) Dosimetrical tests (reference and non-reference) Exp Experiment Depth (cm) Irradiation distance Field size accessory Det. Dose ref 1 Ref. field d ref ssd or sad 10cm x 10cm no 1 4 Gy Tray factor D ref ssd or sad 10cm x 10cm tray 2 4 Gy Energy open beam 4 Energy beam with wedge 10&20 ssd or sad 10cm x 10cm no 3,4 4 Gy &20 ssd or sad 10cm x 10cm wedge 5,6 4 Gy Output factor1 8 ssd or sad 6cm x 6cm no 7 4 Gy Output factor2 8 ssd or sad 8cm x 20cm no 8 4 Gy Output factor3 8 ssd or sad 20cm x 8cm no 9 4 Gy Output factor4 8 sad or ssd 20cm x 20cm no 10 4 Gy MLC 1 Circular 8 ssd or sad 5.6cm circular no 11 4 Gy MLC 2 inverted Y 11 MLC 3 irreg.+ wedge 8 ssd or sad 15cm x 12cm no 12 4 Gy ssd or sad 12cm x 8 cm wedge 13 4 Gy 3.11
13 Audit protocol (3) Exp Experiment Depth (cm) Irradiation distance Field size accessory Nr det Delivered dose ref 12 Ref. field_ MeV1 d ref ssd or clinical practice 10cm x 10cm no 15 4 Gy Ref. field_ MeV2 (R 50 > 7 cm) d ref ssd or clinical practice 10cm x 10cm no 16 4 Gy 4.14 The relative deviation is classified in 4 levels: The relative deviation d º D measured - D center D center is classified into 4 levels: "within optimal level": d 3% "out of optimal level but within tolerance": 3% < d 5% "out of tolerance level": 5% < d 10% "alarm level": d > 10%
14 BELdART-1 results All centers were audited for at least 1 unit A total of 61 out of 91 machines Total number of 212 beams Not audited: Tomotherapy Cyberknife Gammeknife Mobetron
15 BELdART-1 results Mechanical parameters
16 optimal, within tolerance, out of tolerance and alarm for all high-energy photon beams as BELdART-1 results registered in the 1 st runs measurement campaign. The deviation in emergency level was connected to test D2 but appeared to be non significant, as the second run check of this beam was within Dosimetric parameters Fig. 3.7: tolerance level. No explanation could be found a posteriori. Relative and absolute frequencies for the four levels in deviation for the dosimetrical tests as a result of 2 nd run measurements in high-energy photon beams. 0% 0% 6% δ > 10% Eighty second run measurements (6%) were completed versus 1375 first run measurements. After 10% δ > 5% completing the 2nd run measuring campaign, 5% all test δ > 3% measurements δ are within # tolerance % level or 94% δ 3% δ > 10% 1 0,1 better (Fig. 3.7) that is: 96.7% of all measurements in high-energy photon beams are within optimal 10% δ > 5% 4 0,3 level and only 3.3% of the measurements remain in the tolerance 5% δ > 3% level. For 80 the latter, 5,9the local physicist was asked to take a close look at these measurements. δ 3% ,7 Fig. 3.6: Relative and absolute frequencies for the four levels in deviation for the dosimetrical tests as a result of 1 st run measurements in high-energy photon beams. 1% 4% 13% 82% δ > 10% 10% δ > 5% 5% δ > 3% δ 3% δ # % δ > 10% 1 1,0 10% δ > 5% 4 4,1 5% δ > 3% 13 13,3 δ 3% 80 81,6 Fig. 3.8: Percentages and absolute values of 1 st run measurements in electron beams regarding to their deviations.
17 BELdART-1 results Belgian Dosimetry Audits in Radiotherapy (BELdART) : final report. 50 Dosimetric parameters 0% 0% 3% Fig. 3.7: δ > 10% 10% δ > 5% 5% δ > 3% δ # % 97% δ 3% δ > 10% 0 0,0 10% δ > 5% 0 0,0 5% δ > 3% 44 3,3 δ 3% ,7 Belgian Dosimetry Audits in Radiotherapy (BELdART) : final report. 51 Relative and absolute frequencies for the four levels in deviation for the dosimetrical tests as a result of 2 nd run measurements in high-energy photon beams. 0% 0% 13% Eighty second run measurements (6%) were completed versus 1375 first run measurements. After δ > 10% completing the 2nd run measuring campaign, all test measurements are within tolerance level or 10% δ > 5% better (Fig. 3.7) that is: 96.7% of all measurements in high-energy photon beams are within optimal 5% δ > 3% δ # % level and 87% only 3.3% of the measurements δ remain 3% in the tolerance level. For the latter, the local δ > 10% 0 0,0 physicist was asked to take a close look at these measurements. 10% δ > 5% 0 0,0 5% δ > 3% 14 12,7 δ 3% 96 87,3 Fig. 3.9: Percentage and absolute values after the 2 nd run measurements in electron beams regarding to their deviations. Regarding measurements in high-energy electron beams, we can make an analogous analysis. Fig. 3.8
18 Audited centers
19 BELdART-2 Courtesy of Varian Medical Systems
20 BELdART-2 Advanced and dynamic radiotherapy IMRT RapidARC Tomotherapy Cyberknife
21 Protocol A combination of L-α-alanine dosimetry 2 pellets per point instead of 4 Both in high and low dose points Used for absolute dosimetry EBT-3 film dosimetry 3-channel algorithm 2D dose distributions In plane of alanine pellets (relating to absolute dose)
22 3-channel
23 BELdART-2 planning Devided in 2 phases Phase 1 (6 months) Getting some feeling with EBT3 What could we expect in Belgium Homogeneous phantom with selected centers Phase 2 Open for all radiotherapy centers Heterogeneous phantom
24 Phase 1: what did we check? Delivery of the modulation Beam configuration
25 BELdART-2: Phase 1 Use of easy-cube phantom Because of the shape: Prostate case with Prostate Rectum Bladder Film through all 3 structures (sagittal plane)
26 BELdART-2: Phase 1
27 Phase 2: Delivery of the modulation heterogeneities
28 Phase 2: Heterogeneous phantom: Lung or head & neck phantom Including Heterogeneities Imaging Tumor definition (using a description provided) Radiotherapy planning Image guided positioning (if applicable) Treatment
29 Dosimeters For both phases: Dosimeters are inserted by the BELdART-2 team Phantom delivered including dosimeters No visit of the auditor in the center during treatment/planning Phantom has to be treated as if a patient Audit is free of charge for all Belgian radiotherapy centers!
30 What is checked by the end?
31 BELdART-3? New technologies are being implemented
32 Conclusion Radiotherapy is in constant movement A rigid, independent auditing system is necessary to ensure safe radiotherapy BELdART-1 showed that all radiotherapy center work within tolerance levels BELdART-2 is the first national audit investigating the implementation of advanced radiotherapy techniques Within 3 years we will know how good we work in Belgium
33 A word of thanks to The scientific IMRT committee Michael Duchateau, UZBrussel Wouter Crijns, UZLeuven Yassine Boucours, Centre Hospitalier Peltzer-La Tourelle Antoine Delor, UCL Michel Mathot, University of Liege The steering committee François Sergent; Clinique et Maternité Ste-Elisabeth Stefaan Vynckier; UCL Dirk Verellen; UZBrussel Alex Rijnders; Cliniques de l Europe Karen Feyen; AZ St. Maarten
34 Thank you Symposium: Radiation protection for medical doctors Date: Saterday April 27 th 2013 Location: Klein Auditorium, Uhasselt campus Oude gevangenis, Hasselt Target audience: Surgeons, intervention cardiologists, reumatologists, dentists, More info?
International Practice Accreditation
International Practice Accreditation Program endorsed and supported by Jean-Xavier Hallet Punta Cana, Nov 2017 Short History Created in 2004 by ESTRO, independent since 2009 Dosimetry Laboratory (IGR):
More informationTraceability and absorbed dose standards for small fields, IMRT and helical tomotherapy
Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationIAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*
THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service
More informationWhat Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff
What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG
More informationLevel of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013
Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific
More informationRPC s Credentialing Programs for Clinical Trials
RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups
More informationThe RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff
The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an
More informationINTRODUCTION. Material and Methods
Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University
More informationS. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationQA for Clinical Dosimetry with Emphasis on Clinical Trials
QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions
More informationQuality Assurance of Helical Tomotherapy Machines
Quality Assurance of Helical Tomotherapy Machines Wolfgang Tomé 1,2, Hazim Jaradat 2, David Westerly 1, John Fenwick 3, Nigel Orton 2, Rock Mackie 1,2, and Bhudatt Paliwal 1,2 1 Human Oncology and 2 Medical
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationVerification of Advanced Radiotherapy Techniques
Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationNeutron dose evaluation in radiotherapy
Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationQuality assurance in external radiotherapy
Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance
More informationThe Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom
The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom Gary Fisher, B.S. David Followill, Ph.D. Geoffrey Ibbott, Ph.D. This investigation
More informationAssessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom
Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson
More informationUnrivaled, End-to-End
PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationStandard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationGuidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy
Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...
More informationAlbert Lisbona Medical Physics Department CLCC Nantes Atlantique a
The Equipment : safety, commissioning, re commissioning calibration : Equipements : sûreté, recette, modification calibration A medical physicist point of view. Albert Lisbona Medical Physics Department
More informationD DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology
Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran
More informationLearning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy
3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy Sofie Ceberg, PhD Medical Physicist and Lund University Lund, Sweden Learning objectives 1: 3D Dosimetry in the Clinic: Background
More informationProstate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014
Prostate Phantom Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014 GENERAL INFORMATION: Each institution may keep the phantom for a period of time no more than 2 weeks.
More informationBasic dosimetric data for treatment planning system
Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny
More informationActivity report from JCOG physics group
2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer
More informationAn Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center
An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions
More informationVerification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,
More informationMuch of what we will say
New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer
More informationIROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015
IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During
More informationRADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident
Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate
More informationAn anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery
An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,
More informationIROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014
IROC Prostate Phantom Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 The study groups are requesting that each institution keep the phantom for a period of time
More informationDOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS
Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationCatharine Clark NPL, Royal Surrey County Hospital and RTTQA
Catharine Clark NPL, Royal Surrey County Hospital and RTTQA Background Historically UK has been relatively independent of the rest of the world in audit NPL PSDL International intercomparison for reference
More informationIN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS
Romanian Reports in Physics, Vol. 64, No. 3, P. 728 736, 2012 MEDICAL PHYSICS IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS G. VASILE 1,3, MADALINA VASILE 2,3, O.G. DULIU 3 1 Center of
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationCan we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationReference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor
Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Sang Hyun Cho and Geoffrey S. Ibbott Department of Radiation Physics The
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationResponse evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry
Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Luciana C. Matsushima 1, Glauco R. Veneziani 1, Roberto K. Sakuraba
More informationImproving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations
Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationLung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014
Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this
More informationA review of RTTQA audit. Karen Venables
A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines
More informationImplementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy
Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationIntroduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?
More informationOut-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients
EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical
More information8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}
High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the
More informationOrigin, funding & collaboration
UK Brachytherapy Audit: Origin, funding & collaboration Tony Palmer OVERVIEW Purpose From idea to implementation Methodology, results & reflection Proposal for a Brachytherapy Audit Modern brachytherapy
More informationIntroduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.
Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy
More informationAdvanced film dosimetry for a UK brachytherapy audit
Advanced film dosimetry for a UK brachytherapy audit Tony Palmer Lead Scientist Medical Physics, Portsmouth Hospitals NHS Trust, UK University of Surrey, UK Institute of Physics and Engineering in Medicine
More informationReport GA 04/02/2011. Points :
Report GA 04/02/2011 Effective members present : Bacher, Van der Plaetsen, Bosmans, Weber, Del Re, Scheelen, Gysbrechts, Van Duyse, Tomsej, Greffe, Forthomme, Bleeser, Delval, Merlo, Infantino, Bogaert,
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationph fax
This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY
More informationLarge-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance
Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance J. Gajewski1, M. Kłosowski1, S. Greilich2, P. Olko1 Institute of Nuclear Physics, Kraków, Poland German Cancer
More informationDosisverifikation mit Delta 4 Discover während der Behandlung
Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover
More informationEffective and efficient radiotherapy dosimetry audit: Where to next?
Effective and efficient radiotherapy dosimetry audit: Where to next? Catharine Clark, Eduard Gershkevitsh, Milan Tomsej and Nuria Jornet National Physical Laboratory Royal Surrey County Hospital Radiotherapy
More information8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation
Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research
More informationA new geometric and mechanical verification device for medical LINACs
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein
More information3D DOSIMETRY IN END-TO-END DOSIMETRY QA
3D Dosimetry in the Clinic and Research 3D DOSIMETRY IN END-TO-END DOSIMETRY QA Geoffrey S. Ibbott, PhD Department of Radiation Physics MD Anderson Cancer Center Disclosures My institution holds Master
More informationAssessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital
Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri
More informationCredentialing for Clinical Trials -IGRT. Evidence Based Radiation Oncology. Levels of Clinical Evidence. Why Credentialing?
Acknowledgements Credentialing for Clinical Trials -IGRT Invitation from Organizers RTOG Headquarter QA Team ATC Collaborators Evidence Based Radiation Oncology Clinical Trials From Collaborative Groups
More informationUse of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments
Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331
More informationThe Journey of Cyberknife Commissioning
The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources
More informationTLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience
IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationHow accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement
Accurate clinical dosimetry Part I: Fundamentals and New developments in reference dosimetry Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement
More informationA VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI
More informationPresented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output
Presented by: Rebecca M. Howell PhD, DABR, FAAPM Results from Voluntary Independent External Peer Review of Beam Output What do we hope to achieve with independent peer review? Independent peer review
More informationIndependent corroboration of monitor unit calculations performed by a 3D computerized planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and
More informationZakithi Msimang SAAPMB/SARPA 2011
THE IAEA/WHO TLD POSTAL DOSE AUDIT SERVICE Zakithi Msimang SAAPMB/SARPA 2011 Introduction The idea of organizing a dosimetry audit for radiotherapy centres by the IAEA, was suggested for the first time
More informationEric E. Klein, Ph.D. Chair of TG-142
Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing
More informationIROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014
IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 The study groups are requesting that each institution keep the phantom for a period of time
More informationAccreditation of Dosimetry Services
Accreditation of Dosimetry Services J. Van Cauteren, E. De Geest, M. Bricoult AV-Controlatom Content History Regulatory requirements Discussion of reference documents/standards Practical implementation
More informationActive Scanning Beam 3 Checking Delivery/Dosimetry
Active Scanning Beam 3 Checking Delivery/Dosimetry C Algranati, J Salk, A Coray, A Lomax, E Pedroni, T Boeringer, S Lin, E Hug Center for Proton Radiation Therapy Checking Delivery/Dosimetry Contents Absolute
More informationCan we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?
Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas Dept. of Medical Physics & Engineering, Strahlenklinik, Klinikum Offenbach GmbH 63069 Offenbach, Germany
More informationTwo-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance
Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing
More informationProduction and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study
Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad
More informationSHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES
SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring
More informationLONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY
LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY E. BADITA 1, C.VANCEA 1,3, I. CALINA 1,3, D. STROE 2, M. DUMITRACHE 2,3,
More informationDosimetric Consideration in Diagnostic Radiology
Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure
More informationAnthropomorphic breast phantoms for quality assurance and dose verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 5, NUMBER 1, WINTER 2004 Anthropomorphic breast phantoms for quality assurance and dose verification 1 2 3 4 José A. Bencomo, 5 Connel Chu, Victor M.
More informationSmall field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST
Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST What is a Medical Physicist? Medical physics application of physics concepts
More informationMania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane
Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors
More informationStefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.
Experiences of the Secondary Standard Dosimetry Laboratory of the CPHR in the implementation and validation of the Postal Quality Audit in Radiotherapy, using TLDs Stefan Gutiérrez Lores 1, Gonzalo Walwyn
More informationQuality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study
Radiotherapy Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study Technology in Cancer Research & Treatment 2016, Vol. 15(1) 163 170 ª The Author(s) 2015 Reprints
More informationWHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY
doi:10.1016/j.ijrobp.2009.03.029 Int. J. Radiation Oncology Biol. Phys., Vol. 75, No. 1, pp. 253 259, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front
More information