Dosimetry in digital mammography

Size: px
Start display at page:

Download "Dosimetry in digital mammography"

Transcription

1 Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom

2 Outline Why do dosimetry? History Essentials of European breast dosimetry protocol Practical breast dosimetry Uncertainties and limitations Key points

3 Why do dosimetry? To control the dose no higher than necessary BUT high enough to ensure that adequate image quality is obtained essential part of quality control optimise equipment selection, design and use To estimate risk comparison of risk and benefit

4 History

5 UK DOSE SURVEY cm phantom Entrance surface dose TLD Dose range 0.9 to 45 mgy (!) kv range kv HVL range 0.2 mm Al to 1.7 mm Al! Entrance dose is a poor measure of risk!

6 Transmission Transmission through 5 cm breast Adipose Gland Energy kev

7 Alternative dose measures Mid-breast dose Mean breast dose Average glandular dose (Mean glandular dose) Karlsson in 1976 Recommended by ICRP in 1987

8 Essentials of European breast dosimetry protocol

9 Original dosimetry requirements in UK AGD: risk related dose measure Easy to implement Applicable for patient and phantom dosimetry Standard phantom for QC Equivalent to a typical compressed breast

10 Average glandular dose Cannot measure AGD on patients Incident air kerma K (without backscatter) can be easily measured or estimated Need conversion coefficients which relate K to AGD AGD = K g Used in: IPSM Report European Protocol 1996

11 Focal spot Calculation of g using Monte Carlo model (Dance 1990) Calculate energy deposited in breast tissues Compression plate Simple breast phantom Breast support etc

12 Simple breast model Hammerstein mm adipose shield region Central region with mixture of glandular and adipose tissues Fraction by weight of glandular tissue in central region is known as the glandularity Hammerstein et al suggested the use of 50% glandularity for dosimetry

13 g-factors g-factor is for 50% glandularity In principle g-factors depend upon target/filtration and kv Simplification Tables of g-factor just based on HVL & breast thickness For spectra in use at that time worked within ± 5%

14 g-factor mgy/mgy g-factors HVL 0.3 mm HVL 0.6 mm Breast thickness cm

15 Improved formulation in 2000 Dance et al 2000, 2009, 2010, 2014 Real breasts don t all have 50% glandularity factors calculated for % Wider range of spectra used AGD = Kgcs

16 AGD = Kgcs c corrects for glandularity =1 for 50% glandularity s corrects for X-ray spectrum =1 for Mo/Mo Used in: UK (IPEM report ) European Guidelines (latest supplement is 2013) IAEA Code of Practice (TRS457)

17 c-factors Tabulated against HVL, breast thickness and glandularity c-factor cm breast 0.1% 25% 50% 75% % HVL (mm Al) But what glandularity should be used?

18 Breast glandularity (UK)

19 s-factors S Max Error Mo/Mo % Mo/Rh % Rh/Rh % Rh/Al % W/Rh % W/Ag % W/Al s-table needed

20 Standard breasts For quality control and inter-system comparison need a simple phantom or phantoms Cheap and easily reproducible (PMMA) Should be approximately equivalent to typical compressed breast(s)

21 Breast/PMMA equivalence Same incident air kerma at PMMA surface and breast surface Obtained by Monte Carlo simulation Validated by measurements with breast tissue equivalent phantoms

22 PMMA for standard breast dosimetry PMMA mm Breast mm Gland y %

23 PMMA equivalence age Dance et al, 2000

24 Practical breast dosimetry

25 European Guidelines Fourth Edition Supplements 2013 Method: Section 2b2.3 g,c and s factors: Appendix 5

26 Practical dosimetry using blocks of PMMA (6 monthly) Objective is to simulate exposure of standard breasts Stage 1: determine exposure settings using blocks of PMMA Stage 2: determine incident air kerma & HVL Stage 3: calculate AGD Stage 4: review results

27 Stage 1 - Exposure parameters for blocks of PMMA Use usual clinical exposure settings Record mas, kv and target/filter Repeat for 20, 30, 40, 45, 50, 60, 70 mm blocks

28 Exposure parameters for blocks of PMMA Notes I PMMA blocks Thickness correct within ± 0.5 mm PMMA and breast equiv thicknesses are different If AEC works on thickness, add expanded polystyrene spacer to adjust total thickness (do not cover chamber with spacer) If AEC works on attenuation, spacer should not be necessary

29 Exposure parameters for blocks of PMMA Notes II Apply standard compression AEC on mid-line and closest to chest wall If digital AEC looks at densest region of breast, exposures may not be a good match to those obtained clinically

30 Stage 2 - Determination of incident air kerma and HVL AGD = Kgcs g, c and s depend upon the HVL K and HVL are determined with the compression paddle in place

31 Determination of incident air kerma - I Incident air kerma at upper surface of PMMA Measurement of tube output at a known distance from tube focus Apply inverse square law to calculate air kerma at top surface of PMMA Measurement is made with paddle in contact with the dosimeter

32 Determination of incident air kerma - II Use a calibrated dosimeter suitable for mammography Position dosimeter on line joining focal spot to a point on the mid-line of the breast support 6 cm from chest wall edge.

33 Determination of incident air kerma - II For dosimeters with backscatter rejection position dosimeter on breast support For dosimeters without backscatter rejection position the dosimeter higher up to avoid or reduce the effect of backscatter

34 Stage 3 AGD = Kgcs Tables A5.1 & A5.2 give g and c vs PMMA cm & HVL Table A5.4 gives s

35 AGD mgy Stage 4: review results acceptable achievable PMMA thickness cm

36 AGD: 53 mm breast (45 mm PMMA) Number of systems MGD (mgy) 98% < 2.5 mgy

37 Patient dosimetry Measure tube output and HVL with paddle in place Record exposure parameters for patient series (e.g.50 or more) kv & target/filter mas compressed breast thickness Calculate AGD

38 Patient dosimetry Note Accuracy of displayed thickness should be checked by applying a typical force (e.g. 100 N) to rigid material of known thickness Accuracy of ±2 mm or better required

39 AGD = Kgcs Tables A5.5 A5.7 give g and c vs breast cm & HVL c for age or Table A5.4 gives s

40 Average patient AGD for different digital systems: 2010 UK data Oduko et al IWDM 2010 AGD (mgy) DR CR FS Image receptor type

41 Uncertainties & Limitations

42 Uncertainties in AGD using PMMA (typical) Dosimeter calibration 5% AEC and dosimeter reproducibility 2% Uncertainty in g due to 3% uncertainty in HVL 2% Chamber positioning (2mm) 1% Phantom thickness (0.5mm) 2% s-factor approximation 4% Overall 7%

43 Errors in using tabulated equivalent PMMA thickness Exact PMMA thickness required for equivalence varies with beam quality. Error is largest for highest kvs and thickest breasts mm breast: up to 1.0 mm PMMA 80 mm breast: up to 1.7 mm PMMA 100 mm breast: up to 2.4 mm PMMA

44 Errors in AGD associated with nonequivalence of tabulated PMMA thickness (W/Al spectrum) kv 35 kv 40 kv Error % Breast thickness (mm)

45 Air kerma measurement Measurements should be made with the dosimeter in contact with paddle (to mirror MC model) Scatter contributes e.g. ~7% to air kerma for 2.4 mm polycarbonate paddle Some semi-conductor dosimeters are collimated and will not record scatter Correction is then needed (Hemdal, Rad Prot Dosim. 2011)

46 Uncertainties in patient AGD typical and assuming breast model is correct Dosimeter calibration 5% Dosimeter etc. reproducibility 2% Uncertainty in g due to 3% uncertainty in HVL 2% Patient thickness (2 mm) impact on g and ISL 3% s-factor approximation 4% Overall, for one exposure 7% PLUS Standard error on mean from sample of 50 patients about 7%

47 Some limitations of the model Heel effect ignored Simple breast geometry Breast composition data (Hammerstein) based on 4 samples glandular tissue 8 samples adipose tissue 6 samples skin tissue results show considerable variability in composition

48 Effect of adipose shield thickness on g g / g(standard) mm adipose shield Thickness (cm)

49 g / g(standard) Effect of glandular tissue composition on g High O Low O Thickness (cm)

50 Real distributions of glandular tissue The distribution of glandular tissue in patients will not be uniform and large differences in conversion factors can be expected. Modelled using breast phantoms from Predrag Bakic (Dance et al, RPD 2005) Differences in AGD of up to ~40% found

51 European vs USA breast model (Wu) Adipose shield replaced by 4mm skin Different initial attenuation Different volume containing glandular tissue to share energy absorbed Unclear if dose from photons scattered in paddle included Different cross section compilation Differences of up to 16% in conversion coefficients

52 Key points

53 Key points The conversion factors are influenced by the model parameters. Whatever breast model is chosen for QC dosimetry, it need only be an approximate representation of the true situation. The present European and American models give standard procedures and are fine for QC. Large errors can occur if either model is used to estimate AGD for individual patients. But the models are fine for QC of patient doses.

54 Acknowledgements Recent work has been funded under the EU 6th Framework Programme the HighRex project and the UK funded OPTIMAM and OPTIMAM2 projects I am grateful to my colleagues at NCCPM for provision of data, and for discussions

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Direct half value layer measurements in mammography - is near enough good enough?

Direct half value layer measurements in mammography - is near enough good enough? Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.

More information

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi

More information

Patient Dosimetry in Mammography and Tomosynthesis:

Patient Dosimetry in Mammography and Tomosynthesis: 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor

More information

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY

More information

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2

More information

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in

More information

Skin Model and its impact on Digital Mammography

Skin Model and its impact on Digital Mammography Skin Model and its impact on Digital Mammography Rodrigo T. Massera; Alessandra Tomal Institute of Physics "Gleb Wataghin University of Campinas Campinas, Brazil 1 Outline Motivation Methodology Results

More information

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro 3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical

More information

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography

More information

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose. Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives

More information

Calibration coefficients of dosimeters used in mammography for various target/filter combinations

Calibration coefficients of dosimeters used in mammography for various target/filter combinations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Calibration coefficients of dosimeters used in mammography for various target/filter combinations Michiharu Sekimoto, 1a Yoh Katoh,

More information

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz

More information

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Radiation Dosimetry in Digital Breast Tomosynthesis March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Syllabus 1. Introduction 2. Dosimetry in Mammography 3. Dosimetry

More information

The relationship of the mean glandular dose with compressed breast thickness in mammography

The relationship of the mean glandular dose with compressed breast thickness in mammography Review Article Page 1 of 8 The relationship of the mean glandular dose with compressed breast thickness in mammography Xiang Du, Ningle Yu, Yimei Zhang, Jin Wang Department of Radiation Protection, Jiangsu

More information

Breast CT and Dosimetry

Breast CT and Dosimetry 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

Breast dosimetry system in screen/film mammography

Breast dosimetry system in screen/film mammography Bull Fae Health Sei, Okayama Univ Med Seh, 10: 99-106, 2000 (Original article) Breast dosimetry system in screen/film mammography Sachiko GOTO, Yoshiharu AZUMA, Toshinori MARUYAMA, Yoshitada NAKAGIRI,

More information

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system Original article NOWOTWORY Journal of Oncology 2017, volume 67, number 2, 115 120 DOI: 10.5603/NJO.2017.0018 Polskie Towarzystwo Onkologiczne ISSN 0029 540X www.nowotwory.edu.pl Mean glandular dose values

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program An De Hauwere, Hubert Thierens Ghent University, Department of Medical Physics,

More information

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,

More information

A tissue-equivalent phantom series for mammography dosimetry

A tissue-equivalent phantom series for mammography dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOL. 5, NO. 4, FALL 2004 A tissue-equivalent phantom series for mammography dosimetry William P. Argo, 1 Kathleen Hintenlang, 2 and David E. Hintenlang 3 U.S.

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

More information

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.

More information

A Standardised Approach to Optimisation

A Standardised Approach to Optimisation The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained

More information

IAEA TECDOC 1517 Quality Control in Mammography Software

IAEA TECDOC 1517 Quality Control in Mammography Software IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua.

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Development of a program for estimation of patient exposure in mammography

Development of a program for estimation of patient exposure in mammography Developmt of a program for estimation of patit exposure in mammography Y. Asada 1, S. Suzuki, M. Kobayashi, S. Koga 1 Faculty of radiological technology, Fujita Health University, School of Health Scices,

More information

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review Receptor dose and Patient Dose in Radiographic exposures - a 15 year review D J Peet*, N Tyler, M Pryor, P Hollaway, C Strudley and L Leavesley Regional Radiation Protection Service, Royal Surrey County

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

Glandularity Estimation in Japanese Women by Using a Breast Model Made from Mammographic Findings of European Women

Glandularity Estimation in Japanese Women by Using a Breast Model Made from Mammographic Findings of European Women Jpn. J. Health Phys. 51 (4) 238 244 (2016) DOI: 10.5453/jhps.51.238 Original Paper Glandularity Estimation in Japanese Women by Using a Breast Model Made from Mammographic Findings of European Women Ai

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

IAEA-TECDOC-796. Radiation doses in diagnostic radiology

IAEA-TECDOC-796. Radiation doses in diagnostic radiology IAEA-TECDOC-796 Radiation doses in diagnostic radiology The IAEA does The originating Section FOREWORD It is well recognized that diagnostic radiology is the largest contributor to the collective dose

More information

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,

More information

UNIFORMED SERVICES UNIVERSITY OF THE HEALTH SCIENCES 4301 JONES BRIDGE ROAD BETHESDA, MARYLAND JUNE 2003 APPROVAL SHEET

UNIFORMED SERVICES UNIVERSITY OF THE HEALTH SCIENCES 4301 JONES BRIDGE ROAD BETHESDA, MARYLAND JUNE 2003 APPROVAL SHEET UNIFORMED SERVICES UNIVERSITY OF THE HEALTH SCIENCES 4301 JONES BRIDGE ROAD BETHESDA, MARYLAND 20814-4799 18 JUNE 2003 APPROVAL SHEET Title of Thesis: " Digital Mammography Breast Dosimetry Using Copper-Doped

More information

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3 Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,

More information

The need for standardization of dosimetry in experimental radiation biology

The need for standardization of dosimetry in experimental radiation biology The need for standardization of dosimetry in experimental radiation biology Kurt Pedersen University of Wisconsin Medial Radiation Research Center Under the direction of Larry DeWerd, PhD. NCCAAPM Fall

More information

Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE

Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE ACR Update on FFDM Accreditation Overview New ACR Activities Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO QC Today QC Tomorrow *No financial disclosures to report

More information

Simulation of spiculated breast lesions

Simulation of spiculated breast lesions Simulation of spiculated breast lesions Premkumar Elangovan* a, Faisal Alrehily b, R Ferrari Pinto a, Alaleh Rashidnasab a,c, David R Dance b,d, Kenneth C Young b,d, Kevin Wells a a Centre for Vision,

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Staff Exposure Monitoring in Interventional Radiology

Staff Exposure Monitoring in Interventional Radiology Conference on Physics in Medicine: From Diagnosis to Treatment "Enhancing safety and quality in radiation medicine". KFMC, Riyadh 7-9 November 2017 Staff Exposure Monitoring in Interventional Radiology

More information

Applications of Monte Carlo simulations to radiation dosimetry

Applications of Monte Carlo simulations to radiation dosimetry Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

Survey of Mammography Practice: Initial Results from Serbia

Survey of Mammography Practice: Initial Results from Serbia Survey of Mammography Practice: Initial Results from Serbia Olivera Ciraj Bjelac *, Dusko Kosutic, Danijela Arandjic, Milojko Kovacevic Vinca Institute of Nuclear Sciences, Radiation and Environmental

More information

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose Academic Journal of Cancer Research 7 (2): 162-167, 2014 ISSN 1995-8943 IDOSI Publications, 2014 DOI: 10.5829/idosi.ajcr.2014.7.2.83313 Comparison Between Film-Screen and Digital Mammography for Woman

More information

ALARA in mammography screening. Hilde Bosmans et al.

ALARA in mammography screening. Hilde Bosmans et al. ALARA in mammography screening Hilde Bosmans et al. ALARA in the mammography Overview 1. Screening 2. The balance: 1. Image quality 2. Radiation dose 3. In practice Screening leeftijd Mammography Screening

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

nuclear science and technology

nuclear science and technology EUROPEAN COMMISSION nuclear science and technology Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields (EVIDOS) Editor: Helmut Schuhmacher, Physikalisch-Technische Bundesanstalt

More information

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS I. Villanueva (1), I. Amor (1), A. Brosed (), R. Villanueva () M. Ginjaume (3), X. Ortega (3), M. Roig (3) (1)

More information

Mammography calibration qualities establishment in a Mo- Mo clinical system

Mammography calibration qualities establishment in a Mo- Mo clinical system Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733

More information

Contrast Enhanced Spectral Mammography (CESM) Updates

Contrast Enhanced Spectral Mammography (CESM) Updates Contrast Enhanced Spectral Mammography (CESM) Updates Georgeta Mihai, PhD, DABR Medical Physicist, BIDMC, Boston Assistant Professor, Harvard Medical School, Boston Disclosures None Acknowledgments: Da

More information

ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΜΕΛΕΤΗ ΑΠΟΡΡΟΦΩΜΕΝΗΣ ΟΣΗΣ ΣΕ ΜΕΓΕΘΥΝΤΙΚΕΣ ΛΗΨΕΙΣ ΣΤΗ ΜΑΣΤΟΓΡΑΦΙΑ ΜΕ ΤΕΧΝΙΚΕΣ ΠΡΟΣΟΜΟΙΩΣΗΣ MONTE CARLO

ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΜΕΛΕΤΗ ΑΠΟΡΡΟΦΩΜΕΝΗΣ ΟΣΗΣ ΣΕ ΜΕΓΕΘΥΝΤΙΚΕΣ ΛΗΨΕΙΣ ΣΤΗ ΜΑΣΤΟΓΡΑΦΙΑ ΜΕ ΤΕΧΝΙΚΕΣ ΠΡΟΣΟΜΟΙΩΣΗΣ MONTE CARLO ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΤΜΗΜΑ ΙΑΤΡΙΚΗΣ ΤΜΗΜΑ ΦΥΣΙΚΗΣ ΙΑΤΜΗΜΑΤΙΚΟ ΠΡΟΓΡΑΜΜΑ ΜΕΤΑΠΤΥΧΙΑΚΩΝ ΣΠΟΥ ΩΝ ΣΤΗΝ ΙΑΤΡΙΚΗ ΦΥΣΙΚΗ ΜΕΛΕΤΗ ΑΠΟΡΡΟΦΩΜΕΝΗΣ ΟΣΗΣ ΣΕ ΜΕΓΕΘΥΝΤΙΚΕΣ ΛΗΨΕΙΣ ΣΤΗ ΜΑΣΤΟΓΡΑΦΙΑ ΜΕ ΤΕΧΝΙΚΕΣ ΠΡΟΣΟΜΟΙΩΣΗΣ

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

Dose standards for safe and secure breast cancer screening

Dose standards for safe and secure breast cancer screening Research paper Dose standards for safe and secure breast cancer screening - Establishment of X-ray dose standards for mammography and construction of a calibration service system- Takahiro Tanaka *, Tadahiro

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

The impact of SID and collimation on backscatter in radiography

The impact of SID and collimation on backscatter in radiography The impact of SID and collimation on backscatter in radiography Poster No.: C-2931 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: M. Joyce 1, J. T. Ryan 2, M. F. Mc Entee

More information

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT R Aamir, C Lowe, J Damet, P Carbonez, A P H Butler, N Schleich, N G Anderson MARFO, Emmanuel Geant4 User Workshop 2017, Wollongong Overview MARS

More information

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy

More information

Patient Specific Radiation Doses From Projection Radiography Images: Improving the Accuracy of Low Dose Radiation Calculations

Patient Specific Radiation Doses From Projection Radiography Images: Improving the Accuracy of Low Dose Radiation Calculations University of New Mexico UNM Digital Repository Nuclear Engineering ETDs Engineering ETDs Spring 4-9-2018 Patient Specific Radiation Doses From Projection Radiography Images: Improving the Accuracy of

More information

Evaluation of absorbed dose and image quality in mammography

Evaluation of absorbed dose and image quality in mammography Evaluation of absorbed dose and image quality in mammography Hemdal, Bengt 2009 Link to publication Citation for published version (APA): Hemdal, B. (2009). Evaluation of absorbed dose and image quality

More information

Technical quality control in the Norwegian Breast Cancer Screening Program with focus on radiation doses

Technical quality control in the Norwegian Breast Cancer Screening Program with focus on radiation doses Technical quality control in the Norwegian Breast Cancer Screening Program with focus on radiation doses Ingrid Helen Ryste Hauge Department of Physics Faculty of Mathematics and Natural Sciences University

More information

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System James D Rijken Thesis submitted for the degree of Master of

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

2001 AAPM Summer School Seattle, Washington ACR R/F Phantom

2001 AAPM Summer School Seattle, Washington ACR R/F Phantom 2001 AAPM Summer School Seattle, Washington ACR R/F Phantom Charles R. Wilson, Ph.D., FACR Medical College of Wisconsin Milwaukee, Wisconsin MCMC Circa 1975 ACR R/F Phantom This lecture has been approved

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

IRRADIATORS. September 11, :00 am-5:00 pm Hock Plaza Auditorium Duke University Medical Center, Erwin Road

IRRADIATORS. September 11, :00 am-5:00 pm Hock Plaza Auditorium Duke University Medical Center, Erwin Road IRRADIATORS Terry Yoshizumi,, PhD September 11, 2006 8:00 am-5:00 pm Hock Plaza Auditorium Duke University Medical Center, Erwin Road Radiation Countermeasures Center of Research Excellence 1 Acknowledgements

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME

EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME S. Ferrer (), M Ramos (), J.I. Villaescusa (), G Verdú (), M.D. Salas (3), M.D. Cuevas (4) () Departamento

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

Update on the ACR FFDM QC Manual

Update on the ACR FFDM QC Manual Update on the ACR FFDM QC Manual Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO AAPM 2011 Vancouver, Canada 9000 8500 8000 7500 7000 6500 6000 5500 5000 4500 4000

More information

Image Quality Requirements for Digital Mammography in Breast Cancer Screening

Image Quality Requirements for Digital Mammography in Breast Cancer Screening 5 Image Quality Requirements for Digital Mammography in Breast Cancer Screening Margarita Chevalier 1, Fernando Leyton 2,3, Maria Nogueira Tavares 3, Marcio Oliveira 3, Teogenes A. da Silva 3 and João

More information

Trends in Patients Exposure Doses during Radiographic Examination in Japan

Trends in Patients Exposure Doses during Radiographic Examination in Japan Trends in Patients Exposure Doses during Radiographic Examination in Japan S. Suzuki 1*, Y. Asada 1, M. Kobayashi 2, K. Kobayashi 2, Y. Matsunaga 3, T. Otsuka 4 1 School of Health Sciences, Fujita Health

More information

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural

More information

ABSTRACT I. INTRODUCTION

ABSTRACT I. INTRODUCTION 2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography

More information

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski White Paper October 2012 SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

Implementation of a New Tomosynthesis Program: A Physicists Perspective

Implementation of a New Tomosynthesis Program: A Physicists Perspective Implementation of a New Tomosynthesis Program: A Physicists Perspective Bill Geiser, MS DABR Senior Medical Physicist wgeiser@mdanderson.org 1 Conflict of Interest None of the authors nor their immediate

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

PRINCIPLES AND METHODS OF RADIATION PROTECTION

PRINCIPLES AND METHODS OF RADIATION PROTECTION PRINCIPLES AND METHODS OF RADIATION PROTECTION Lesson Outcomes At the end of the lesson, student should be able to: Define what is radiation protection (RP) Describe basic principles of RP Explain methods

More information

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method Iran. J. Radiat. Res., 2007; 5 (3): 147-152 Dosimetric characteristics of Cs sources used in after loading Selectron system by Monte Carlo method M.B.Tavakoli, D. Shahbazi-Gahrouei *, M. Hosseinpour Department

More information

Effect of thyroid shielding during mammography: measurements on phantom and patient as well as estimation with Monte Carlo simulation

Effect of thyroid shielding during mammography: measurements on phantom and patient as well as estimation with Monte Carlo simulation Pyka et al. European Radiology Experimental (2018) 2:14 https://doi.org/10.1186/s41747-018-0042-9 European Radiology Experimental ORIGINAL ARTICLE Effect of thyroid shielding during mammography: measurements

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,800 116,000 120M Open access books available International authors and editors Downloads Our

More information

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI john.sandrik@med.ge.com with many thanks to Vince Polkus, Advanced Applications Product Mgr. 1 Content

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information