Monte Carlo Dose Calculation for Radiotherapy Treatment Planning
|
|
- Berenice Rogers
- 6 years ago
- Views:
Transcription
1 Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning C-M M Charlie Ma, Ph.D. Department of Radiation Oncology Fox Chase Cancer Center Philadelphia, PA 19111, USA Chetty I, Curran B, Cygler J, DeMarco J, Ezzell G, Faddegon B, Kawrakow I, Keall P, Liu H, Ma C-M, Sheikh-Baghery D, Rogers D, Seuntjens S, Siebers S. The AAPM Task Group 105 Report Med Phys (2007) 34: What is Monte Carlo? (a mathematical method using random sampling) The originators : Von Neumann and Ulam 1949 The method : Random sampling from pdf s to construct solutions to problems. Why call it Monte Carlo? John Von Neumann ( ) Stanislow M. Ulam ( ) After the city in the Monaco principality... 1
2 What is Monte Carlo Radiation Transport? Photons Random sampling of particle interactions a good supply of random numbers probability distributions governing the physics processes Information obtained by simulating large number of histories e- e+ CPU time required 1 Gy => 1 billion electrons 1% uncertainty for 0.3 cm cubes requires a few million electrons (hours on a PC) 1 Gy => 1000 billion photons 1% uncertainty for 0.3 cm cubes requires up to a few billion photons (days on a PC) Applications of M-C in radiotherapy Fluence and spectrum calculations Dosimetric parameters (stopping powers, etc.) Correction factors (BSF, HS, PS, P/S ratio...) Dosimeter response simulations Treatment head simulations Treatment planning dose calculations 2
3 Correction factor for bone for 6 MV photons water bone Correction factor for bone Cl2300 C/D 6MV calculated by FOCUS measured with ion chamber calculated by EGS4/DOSXYZ water Comparisons of Pencil Beam and Monte Carlo Mohan et al (1997) Conventional Method (Pencil Beam) PEREGRINE 70 Gy Tumor 70 Gy cm Depth (cm) Dose Verification for IMRT (Pawlicki et al 2000) 100 DVH Comparison 80 MC Corvus GTV Monte Carlo Gy Corvus Volume (%) RT Lung Cord Dose (Gy) 3
4 Implementation of MCTP Simulation of Clinical Accelerators Accelerator simulation Source modeling Beam commissioning CT data conversion and phantom setup Dose calculation algorithms Data processing and plan evaluation Plan optimization Tissue Type/Mass Density Determination Frequency (Pawlicki 1998) 512 x x x CT Number CT Number to Medium Conversion 2000~2.088g/cm3 125~1.101g/cm3-700~0.302g/cm3-950~0.044g/cm3-1000~0.001g/cm3 Mass density (g/cm^3) air CT Number teflon Skeleton - Ribs polycarbonate Vertebral body water polyethylene Alderson Muscle Alderson Lung 4
5 M-C Simulation Geometry Source plane contours Density map Medium map Beam modifier CT phantom Monte Carlo Algorithms Electron beams reported VMC DOSXYZ MCDOSE/MCSIM PENELOPE MCRTP DPM Photon beams reported XVMC PEREGRINE MCNP DOSXYZ MCDOSE/MCSIM/MCRS Current Status of MCTP BrainLab NOMOS Varian Accuray CMS Nucletron AAPM TG157 Survey 5
6 Measured vs MC Reconstructed Dose Distributions Beam Commissioning for MC 6 MV 40cmx40cm 18 MV 40cmx40cm Yang et al, Phys Med Biol (2004) 49: Combined Dose Distribution: Film vs Monte Carlo Patient plan (Corvus) Delivery in film phantom Monte Carlo dose calculation Calculations vs. Measurements superior inferior cm from isocenter Energy M-C Meas CORVUS 4 MV Gy Gy Gy 15 MV Gy Gy Gy Lee et al (2001) 6
7 Treatment Plan Comparison Prostate Treatment Plan Comparison Monte Carlo Gy Corvus Prostate T11-L1 Vertebra Volume (%) Rectum MC Corvus Bladder Prostate Monte Carlo Gy Corvus Dose (Gy) 90 7
8 100 T11-L1 Vertebra CyberKnife Case Study Volume (%) Cord MC Corvus GTV Dose (Gy) Comparison of Dose Distribution Variation of Dose Kernels with Density 12.5 cm depth 5 cm depth High-density medium Low-density medium 8
9 d=3.3cm calculated d=3.3cm measured d=10cm calculated d=10cm measured Variation of Dose Kernels with Density Beamlet Dose Distributions High-density medium Low-density medium FSPB Monte Carlo High-density medium The accuracy depends on how it is implemented. Pawlicki and Ma (2001) Effect of Couch Bar on IMRT Dose Distribution Attenuation for 18 MV photons Other Applications for Advanced RT Relative Dose (%) 20 Without bar Off-axis Distance (cm) MC vs film measurement Yang et al 2005 With bar 9
10 100 Effect of Couch Bar on IMRT Dose Distribution IMRT using 18 MV photons 18MV DVH 100 IMRT using 6 MV photons DVH Data TPS Monte Carlo Volume (%) bladder-no bar rectum-no bar CTV-no bar bladder-with bar rectum-with bar CTV-with bar Volume (%) bladder - no bar rectum - no bar CTV - no bar bladder - with bar rectum - with bar CTV - with bar Dose (cgy) Dose Yang et al 2005 Dose Volume Histogram (H&N) 100 Pencil beam Volume (%) LT Eye Target Monte Carlo MC for Modulated Electron RT Optic chiasm Dose (cgy) 10
11 MU real patient vs.water tank (MC / Water tank= 292 / 256=1.14) MU real patient vs.water tank Target 1,2 MC based MU Target 1,2 water tank based MU Lt eye water tank based MU Rt eye water tank based MU Rt eye MC based MU Lt eye MC based MU Courtesy of Joanna Cygler Internal mammary nodes (MC / Water tank= 210 / 206=1.019) Isodose Distributions of MBRT for Hypofractionated Breast Treatment Courtesy of Joanna Cygler A. Photon IMRT 58.0Gy 56.0Gy 47.3Gy 45.0Gy 42.8Gy 40.5Gy 36.0Gy 31.5Gy 22.5Gy 13.5Gy 4.5Gy B. Electron beam C. Combined dose 11
12 Mixed Beam Breast Plan Statistics and Denoising Optimize IMRT based on a fixed e- field Optimize IMRT and the e- field weight Xiong et al (PMB 2004) Optimize IMRT and MERT Statistics and Smoothing Techniques Dose Prescription Smoothing or denoising is safe at 3-4% statistical level 12
13 CTV or PTV Dose Prescription Dose to what medium? 4% stats 2mm σ 4mm σ 9% 5% 3% 1% Dose prescription based on DVH and isodose distributions Dose to Water or Dose to Medium? Summary MC dose calculation is becoming a practical tool for advanced radiotherapy treatments The accuracy of MC dose calculation depends on the implementation The potential of MC dose calculation remains to be explored 13
14 Acknowledgments The FCCC/Stanford Monte Carlo Team Charlie Ma Bob Price Lili Chen Eugene Fourkal Jinsheng Li Lu Wang Yan Chen James Fan Teh Lin Max Jin Steve Jiang Todd Pawlicki Jun Deng Bilal Shahine Ajay Kapur Michael Lee Qianyi Xu Iavor Veltchev Alain Tafo Ahmed ElDib Francis Tang Sotirios Stathakis Jay Chen Wei Luo Jie Yang Lihong Qin Meisong Ding Omar Chibani Grisel Mora Thai Bing Nguyen William Xiong Antonio Leal Freek Du Plessis Thank You FCCC Monte Carlo course, April
Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA
Independent Dose Verification for IMRT Using Monte Carlo C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Outline Why Monte Carlo for IMRT QA? Experimental verification
More informationOutline. Clinical Implementation of Monte Carlo Methods for External Photon Therapy. Outline. Indrin J. Chetty Henry Ford Hospital, Detroit MI
Outline Clinical Implementation of Monte Carlo Methods for External Photon Therapy Indrin J. Chetty Henry Ford Hospital, Detroit MI A. Introduction to the Monte Carlo Method as applied to radiation transport
More informationIMRT for Prostate Cancer
IMRT for Cancer All patients are simulated in the supine position. Reproducibility is achieved using a custom alpha cradle cast that extends from the mid-back to mid-thigh. The feet are positioned in a
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationAssessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom
Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson
More informationIntroduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy
X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction
More informationAAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction
AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall
More informationBenchmarking treatment planning dose calculations using a Monte Carlo dose engine as a potential activity of a primary standard lab
Benchmarking treatment planning dose calculations using a Monte Carlo dose engine as a potential activity of a primary standard lab Nick Reynaert (LSDG, Belgium/Erasmus MC, The Netherlands) Hugo Palmans
More informationMONTE CARLO BASED DOSIMETRY OF HEAD-AND-NECK PATIENTS TREATED WITH SIB-IMRT
doi:10.1016/j.ijrobp.2005.09.049 Int. J. Radiation Oncology Biol. Phys., Vol. 64, No. 3, pp. 968 977, 2006 Copyright 2006 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/06/$ see front
More informationLinac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR
Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,
More informationEvaluation of Dosimetry Check software for IMRT patient-specific quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,
More informationFilm-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom
Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationCommissioning and quality assurance of a commercial stereotactic treatment-planning system for extracranial IMRT
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Commissioning and quality assurance of a commercial stereotactic treatment-planning system for extracranial IMRT Lu Wang, Jinsheng
More informationIROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015
IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During
More informationPre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system
Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system E. Mezzenga, * E. Cagni, A. Botti, M. Orlandi and M. Iori Medical Physics Unit, IRCCS Arcispedale
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationThe Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom
The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom Gary Fisher, B.S. David Followill, Ph.D. Geoffrey Ibbott, Ph.D. This investigation
More informationTransition to Heterogeneity Corrections. Why have accurate dose algorithms?
Transition to Heterogeneity Corrections Eric E. Klein, M.S., Washington University, St. Louis, MO Craig Stevens, M.D., Ph.D., MD Anderson Cancer Center, Houston, TX Nikos Papinikolou, Ph.D., University
More informationQuality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose
Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationIndependent absorbed-dose calculation using the Monte Carlo algorithm in volumetric modulated arc therapy
Haga et al. Radiation Oncology 2014, 9:75 RESEARCH Open Access Independent absorbed-dose calculation using the Monte Carlo algorithm in volumetric modulated arc therapy Akihiro Haga *, Taiki Magome, Shigeharu
More informationMonte Carlo for CyberKnife Incise TM MLC
Division of Medical Radiation Physics Monte Carlo for CyberKnife Incise TM MLC Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital, Bern University Hospital, and University
More informationFeasibility of 4D IMRT Delivery for Hypofractionated High Dose Partial Prostate Treatments
Feasibility of 4D IMRT Delivery for Hypofractionated High Dose Partial Prostate Treatments R.A. Price Jr., Ph.D., J. Li, Ph.D., A. Pollack, M.D., Ph.D.*, L. Jin, Ph.D., E. Horwitz, M.D., M. Buyyounouski,
More informationApplications of Monte Carlo simulations to radiation dosimetry
Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationTHE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS
THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationDosimetric advantage of using 6 MV over 15 MV photons in conformal therapy of lung cancer: Monte Carlo studies in patient geometries
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 1, WINTER 2002 Dosimetric advantage of using 6 MV over 15 MV photons in conformal therapy of lung cancer: Monte Carlo studies in patient geometries
More informationFrom position verification and correction to adaptive RT Adaptive RT and dose accumulation
From position verification and correction to adaptive RT Adaptive RT and dose accumulation Hans de Boer Move away from Single pre-treatment scan Single treatment plan Treatment corrections by couch shifts
More informationStereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy?
Establishing SBRT Program: Physics & Dosimetry Lu Wang, Ph.D. Radiation Oncology Department Fox Chase Cancer Center Outlines Illustrate the difference between SBRT vs. CRT Introduce the major procedures
More informationIntroduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.
Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy
More informationFeasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment
Chinese Journal of Cancer Original Article Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment Suresh Rana 1 and ChihYao Cheng 2 Abstract The volumetric modulated
More informationMeasurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional
Measurement of Dose to Critical Structures Surrounding the Prostate from Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional Conformal Radiation Therapy (3D-CRT); A Comparative Study Erik
More informationHeterogeneity Corrections in Clinical Trials
Heterogeneity Corrections in Clinical Trials Niko Papanikolaou Departments of Radiology and Radiation Oncology U of TX Health Science Center, San Antonio, TX Eric Klein Department of Radiation Oncology
More informationEvaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases
Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases CASE STUDY Institution: Odette Cancer Centre Location: Sunnybrook
More informationActivity report from JCOG physics group
2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer
More informationTechnical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.
Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution
More informationIMAT: intensity-modulated arc therapy
: intensity-modulated arc therapy M. Iori S. Maria Nuova Hospital, Medical Physics Department Reggio Emilia, Italy 1 Topics of the talk Rotational IMRT techniques: modalities & dedicated inverse-planning
More informationAn anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery
An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,
More information8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}
High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics
More informationA comparative dosimetric analysis of the effect of heterogeneity corrections used in three treatment planning algorithms
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2010 A comparative dosimetric analysis of the effect of heterogeneity corrections used in three treatment planning
More informationIROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015
IROC Lung Phantom 3D CRT / IMRT Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 The IROC requests that each institution keep the phantom for no more than 2 weeks. During
More informationEvaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc
Original Article PMP Progress in Medical Physics 28(4), December 217 https://doi.org/1.14316/pmp.217.28.4.181 pissn 258-4445, eissn 258-4453 Evaluation of Dynamic Delivery Quality Assurance Process for
More informationTITLE: MR Imaging Based Treatment Planning for Radiotherapy of Prostate Cancer
AD Award Number: W81XWH-04-1-0023 TITLE: MR Imaging Based Treatment Planning for Radiotherapy of Prostate Cancer PRINCIPAL INVESTIGATOR: Lili Chen, Ph.D. CONTRACTING ORGANIZATION: FOX Chase Cancer Center
More informationBorges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1
Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior
More informationHampton University Proton Therapy Institute
Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI
More informationImplementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy
Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out
More informationA Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population
A Dosimetric Comparison of Whole-Lung Treatment Techniques in the Pediatric Population Corresponding Author: Christina L. Bosarge, B.S., R.T. (R) (T) Indiana University School of Medicine Department of
More informationA VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI
More information8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient
Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,
More informationMR Imaging Based Treatment Planning for Radiotherapy. Distribution Unlimited
AD Award Number: W81XWH-04-1-0023 TITLE: MR Imaging Based Treatment Planning for Radiotherapy of Prostate Cancer PRINCIPAL INVESTIGATOR: Lili Chen, Ph.D. CONTRACTING ORGANIZATION: Fox Chase Cancer Center
More informationRange Uncertainties in Proton Therapy
Range Uncertainties in Proton Therapy Harald Paganetti PhD Professor of Radiation Oncology, Harvard Medical School Director of Physics Research, Massachusetts General Hospital, Department of Radiation
More informationCan we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?
Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas Dept. of Medical Physics & Engineering, Strahlenklinik, Klinikum Offenbach GmbH 63069 Offenbach, Germany
More informationA model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms
1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,
More informationA treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse
Peters et al. Radiation Oncology 2014, 9:153 RESEARCH Open Access A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse Samuel Peters
More informationContributors to this Talk
Tomotherapy Thomas Rockwell Mackie Tomotherapy Research Group Depts. Of Medical Physics, Human Oncology and Biomedical Engineering University of Wisconsin Madison WI 53706 Phone: (608) 262-7358 Email:
More informationDosimetric Analysis Report
RT-safe 48, Artotinis str 116 33, Athens Greece RT-safe +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ------ Irradiation system: ------
More informationIMRT Planning Basics AAMD Student Webinar
IMRT Planning Basics AAMD Student Webinar March 12, 2014 Karen Chin Snyder, MS Senior Associate Physicist Department of Radiation Oncology Disclosures The presenter has received speaker honoraria from
More informationCommissioning of brachytherapy Treatment Planning Systems
Commissioning of brachytherapy Treatment Planning Systems Jacco L.G. Steenhuijsen, PhD Catharina Hospital Eindhoven Landelijke brachydag, 17-11-17 Brachytherapy Treatment Planning Systems 2 Brachytherapy
More information8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation
Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research
More informationProton and helium beams: the present and the future of light ion beam therapy
Proton and helium beams: the present and the future of light ion beam therapy Dr. Andrea Mairani Group Leader Biophysics in Particle Therapy Heidelberg Ion Beam Therapy Center HIT Department of Radiation
More informationCan we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?
Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas 1,3, Natasa Milickovic 1, Nikolaos Zamboglou 2 1 Dept. of Medical Physics & Engineering, 2 Strahlenklinik,
More informationNIH/NCI Varian Medical Systemss Philips HealthCare
Treatment Planning of Lung Cancer: Dosimetric Considerations Indrin J. Ch hetty, PhD Henry Ford Health System Disclo osure My department receives research support from: NIH/NCI Varian Medical Systemss
More informationCalibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System
Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System James D Rijken Thesis submitted for the degree of Master of
More informationAccounting for center-of-mass target motion using convolution methods in Monte Carlo-based dose calculations of the lung
Accounting for center-of-mass target motion using convolution methods in Monte Carlo-based dose calculations of the lung Indrin J. Chetty, a) Mihaela Rosu, Daniel L. McShan, Benedick A. Fraass, James M.
More informationDosimetric Comparison: Acuros XB vs AAA
Dosimetric Comparison: Acuros XB vs AAA Leo El Hage, Chris James, Jim Frantzis, Roger Li, Venkata Sheshadri Epworth Radiation Oncology Epworth HealthCare Excellence. Everywhere. Everyday Background Epworth
More informationAutomated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist
Automated Plan Quality Check with Scripting Rajesh Gutti, Ph.D. Clinical Medical Physicist Veera.Gutti@BSWHealth.org Outline Introduction - BSW Automation in Treatment planning Eclipse Scripting API Script
More informationOrgan Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy
Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy Rasha Makkia Biomedical Physics Ph.D. student East Carolina University March 3 rd, 2016 Outlines The Purpose Wilms Tumor,
More informationA Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer
A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer Tony Kin Ming Lam Radiation Planner Dr Patricia Lindsay, Radiation Physicist Dr John Kim, Radiation Oncologist Dr Kim Ann Ung,
More informationClinical Impact of Couch Top and Rails on IMRT and Arc Therapy
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2011 Clinical Impact of Couch Top and Rails on IMRT and Arc Therapy
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationEQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS
PAPER EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS Isra Israngkul Na Ayuthaya *, Sivalee Suriyapee, Phongpheath Pengvanich
More informationEvaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer
1 Charles Poole April Case Study April 30, 2012 Evaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer Abstract: Introduction: This study
More informationVerification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System
Original Article Research in Oncology 2018; Vol. 14, No. 1: 34-38. DOI: 10.21608/resoncol.2018.3026.1047 Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System
More informationQA for Clinical Dosimetry with Emphasis on Clinical Trials
QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions
More informationCommission, Ganakbari, Savar, Dhaka, Bangladesh
International Letters of Chemistry, Physics and Astronomy Submitted: 2016-05-17 ISSN: 2299-3843, Vol. 68, pp 54-60 Revised: 2016-07-08 doi:10.18052/www.scipress.com/ilcpa.68.54 Accepted: 2016-07-12 2016
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationThe Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist
The Radixact System Experience at Miami Cancer Institute Tino Romaguera, D.Sc. Senior Physicist Disclosure An honorarium is provided by Accuray for this presentation The views expressed in this presentation
More informationApplications of Modern Radiotherapy Systems
Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie Professor University of Wisconsin Co-Founder and Chairman of the Board TomoTherapy Inc. Financial Disclosure I am a founder and Chairman
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More informationHot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis
Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline
More informationFeasibility of improving cone-beam CT number consistency using a scatter correction algorithm.
Thomas Jefferson University Jefferson Digital Commons Department of Medicine Faculty Papers Department of Medicine 11-4-2013 Feasibility of improving cone-beam CT number consistency using a scatter correction
More informationThis article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationIGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose
IGRT Protocol Design and Informed Margins DJ Vile, PhD Conflict of Interest I have no conflict of interest to disclose Outline Overview and definitions Quantification of motion Influences on margin selection
More informationAdditional Questions for Review 2D & 3D
Additional Questions for Review 2D & 3D 1. For a 4-field box technique, which of the following will deliver the lowest dose to the femoral heads? a. 100 SSD, equal dmax dose to all fields b. 100 SSD, equal
More informationDOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD
Proceedings of the Tenth EGS4 Users' Meeting in Japan, KEK Proceedings 2002-18, p.65-73 DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD N. Tohyama, H. Saitoh, T. Fujisaki 1, S. Abe 1
More informationTESTICULAR DOSES IN IMAGE-GUIDED RADIOTHERAPY OF PROSTATE CANCER
doi:10.1016/j.ijrobp.2011.01.071 Int. J. Radiation Oncology Biol. Phys., Vol. 82, No. 1, pp. e39 e47, 2012 Copyright Ó 2012 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/$ - see front
More information4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest
SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific
More informationMeasure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)
IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.Volume 10, Issue 2 Ver. I (Mar. Apr. 2018), PP 55-59 www.iosrjournals.org Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient
More informationOnline in vivo dosimetry in conformal radiotherapies with MOSkin detectors
Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors G.Gambarini 1, C.Tenconi 1, N.Mantaut 1 M.Carrara 2, M.Borroni 2, E.Pignoli 2 D.Cutajar 3, M.Petasecca 3, I.Fuduli 3, M.Lerch
More information8/2/2012. Transitioning from 3D IMRT to 4D IMRT and the Role of Image Guidance. Part II: Thoracic. Peter Balter, Ph.D.
8/2/2012 Transitioning from 3D IMRT to 4D IMRT and the Role of Image Guidance Part II: Thoracic Peter Balter, Ph.D. Disclosure Dr. Balter is Physics PI on a trial comparing Cyberknife based SBRT with surgery,
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationSpecifics of treatment planning for active scanning and IMPT
Specifics of treatment planning for active scanning and IMPT SFUD IMPT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Treatment planning for scanning 1. Single Field,
More informationReena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery
Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery Reena Phurailatpam Tejpal Gupta, Rakesh Jalali, Zubin Master, Bhooshan Zade,
More informationMAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:
SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,
More informationCan we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationThe effect of body contouring on the dose distribution delivered with volumetric-modulated arc therapy technique
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 The effect of body contouring on the dose distribution delivered with volumetric-modulated arc therapy technique Jaegi Lee, 1,2 Jong
More informationTreatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)
Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize
More information