Delivery Quality Assurance with a High Resolution Liquid Filled Ion Chamber Array for Robotic Radiosurgery

Size: px
Start display at page:

Download "Delivery Quality Assurance with a High Resolution Liquid Filled Ion Chamber Array for Robotic Radiosurgery"

Transcription

1 Adamczyk S, Chan M, Stelljes T, Schmitt D, Skrobala A, Poppinga D, Adamczyk M, Pawalowski B, Hildebrandt G, Poppe B, Blanck O Dr. Oliver Blanck University Clinic Lübeck, Germany Phone: Oliver@Blanck.de Delivery Quality Assurance with a High Resolution Liquid Filled Ion Chamber Array for Robotic Radiosurgery

2 Introduction CyberKnife radiosurgery has a high demand for precise (delivery) quality assurance Film has a great spatial resolution, but its dosimetric accuracy is only around 3% Diode Arrays are generally sparse in resolution and are prone to directional beam orientation We evaluated a 2D liquid filled ion chamber array (Octavius 1000 SRS, PTW, Freiburg, Germany) and present our first experiences regarding sensitivity and usability for the CyberKnife in clinical practice

3 Material and Methods Octavius 1000 SRS (PTV, Freiburg) 977 liquid filled ion chambers MicroLion (PTW) technology Chamber size: 2.3 x 2.3 x 0.5 mm 10 cm x 10 cm field 2.5 mm in central 5 cm x 5 cm field

4 Material and Methods Test 1: Resolution check 1000SRS versus Diamond Test 2: Detection of geometric beam profile positioning Single vertical CyberKnife beam pointed to the central chamber Robot was moved in 0.1 mm steps in all directions Registration of reference to shifted measurements using auto-alignment function in VeriSoft Test 3: Repetition With fiducial plate on 1000SRS delivery AQA beam multiple times Remove fiducial plate, change 1000SRS, separate multiple weeks

5 Material and Methods Test 4: Response changes from beam orientation Isocentrical beam delivery and comparing measurement versus plan calculation beam by beam Test 5: Response to system delivery errors Complex C-shaped dose distribution Deleting beams in a specific region (1%, 1.5%, 2.5% and 3% MU) Comparing results to calculated dose distribution of original non-modified treatment plan Test 6: Usage in clinical routine Initial validation on 20 extracranial and 20 intracranial clinical treatment plans

6 Test 1: Resolution check

7 Test 2: Detection of geometric beam profile positioning Almost perfect agreement Works with collimators >= 10mm Robot offset Autoalignment Difference baseline mm mm mm mm mm mm mm mm mm mm Test 3: Repetition Max displacement: 0.34mm Max dose difference : 0.48% (1 degree warmer in room) Auto align Dose Time X (mm) Y (mm) Dose (Gy) Diff (%) Baseline min min min min Next day min min min min Next day min min min min Screw Screw Screw Random Shift Random Shift Random Shift Mean Min Max

8 3 months later: No change that is a different array by the way 8

9 Test 4: Response changes from beam orientation

10 Test 4: Response changes from beam orientation Calibrated dose difference per CyberKnife Body Path node

11 Test 5: Response to system delivery errors (original, 1mm/2%)

12 Test 5: Response to system delivery errors (1% less MU)

13 Test 5: Response to system delivery errors (1.5% less MU)

14 Test 5: Response to system delivery errors (2.5% less MU)

15 Test 5: Response to system delivery errors (3% less MU)

16 Test 5: Response to system delivery errors (3% less MU, 2mm/2%!!)

17 Test 6: Usage in clinical routine Gamma (2% local pixel dose difference, 1mm distance to agreement) Intracranial passing rates: 96.3% (min 84.5%, max 100%, SD 4.0%) Extracranial passing rates: 89.1% (min 75.9%, max 99.4%, SD 6.3%)

18 Conclusion and Outlook We demonstrated that a high-resolution array of ion chambers (Octavius 1000 SRS) can be a precise dosimetric tool for routine and delivery quality assurance for robotic radiosurgery with beam angles below 85 degree Work in progress Vertical 1000 SRS holder and modifying the fiducial plate -> AQA Detailed analysis of treatment plan beam directions -> DQA Plan to 1000SRS registration with 3D max Gamma search -> E2E Synchrony treatment verification (log files) -> Patient delivery QA Small lesions

19 Patient delivery QA Static vs. Patient motion with additional 1cm drift!! Gamma 0mm/0.5%!!

20 Adamczyk S, Chan M, Stelljes T, Schmitt D, Skrobala A, Poppinga D, Adamczyk M, Pawalowski B, Hildebrandt G, Poppe B, Blanck O Dr. Oliver Blanck University Clinic Lübeck, Germany Phone: Oliver@Blanck.de Thanks to PTW

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

SRS Uncertainty: Linac and CyberKnife Uncertainties

SRS Uncertainty: Linac and CyberKnife Uncertainties SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis

More information

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 CyberKnife Technology in Ablative Radiation Therapy Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 Objectives Components and work flow of CyberKnife Motion management of CyberKnife

More information

Verification of Advanced Radiotherapy Techniques

Verification of Advanced Radiotherapy Techniques Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system JBUON 2018; 23(1): 179-184 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Patient-specific quality assurance for intracranial cases in robotic

More information

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery

More information

Precisely Maximize Dose, side effects, Patient

Precisely Maximize Dose, side effects, Patient Precisely Maximize Dose, Minimize side effects, Maximize Patient Comfort Key Differentiators: The CyberKnife M6 Series Industry-leading precision and patient comfort Enables full-body robotic radiosurgery,

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Quality Assurance in Stereotactic. Radiotherapy. Swedish Cancer Institute Seattle, WA

Quality Assurance in Stereotactic. Radiotherapy. Swedish Cancer Institute Seattle, WA Quality Assurance in Stereotactic iosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Volumetric Modulated Arc Therapy - Patient Specific QA

Volumetric Modulated Arc Therapy - Patient Specific QA Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck

More information

Work partially supported by VisionRT

Work partially supported by VisionRT Work partially supported by VisionRT Background of frameless intracranial stereotactic radiosurgery UCSD SRS/SRT procedure Clinical Results Summary Total prescribed doses : order of 10 50 Gy Planning targets

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 37-013 GARD Users Manual March 2005 Manual No. 37-013-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective

More information

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI

More information

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119 Romanian Reports in Physics, Vol. 68, No. 3, P. 1130 1139, 2016 GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119 M. DUMITRACHE 1,3, Alina TĂNASE

More information

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119 GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119 M. DUMITRACHE 1,3, Alina TĂNASE 2,3 1 Institute of Oncology Prof. Dr. Al. Trestioreanu, Bucharest,

More information

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients 3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients G Asuni *1, T vanbeek 1, E VanUtyven 1, P McCowan 1,2, and B.M.C. McCurdy 1,2,3 1 Medical Physics Department, CancerCare

More information

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS Stereotactic Radiosurgery Extracranial Stereotactic Radiosurgery Annette Quinn, MSN, RN Program Manager, University of Pittsburgh Medical Center Using stereotactic techniques, give a lethal dose of ionizing

More information

X-Ray Guided Robotic Radiosurgery for Solid Tumors

X-Ray Guided Robotic Radiosurgery for Solid Tumors X-Ray Guided Robotic Radiosurgery for Solid Tumors Mohan Bodduluri Accuray Incorporated 570 Del Rey Avenue Sunnyvale, CA 94085 USA and J. M. McCarthy Department of Mechanical and Aerospace Engineering

More information

Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations

Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife State-of of-the-art and New Innovations Chad Lee, PhD CK Solutions, Inc. and CyberKnife Centers of San Diego Outline Basic overview

More information

ph fax

ph fax This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY

More information

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011 MVCT Image Guidance and QA Robert Staton, PhD DABR MD Anderson Cancer Center Orlando ACMP Annual Meeting 2011 Disclosures MDACCO has received grant funding from TomoTherapy, Inc. Overview TomoTherapy MVCT

More information

IMRT QA: Point Dose Measurements or 2D Array?

IMRT QA: Point Dose Measurements or 2D Array? QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,

More information

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system E. Mezzenga, * E. Cagni, A. Botti, M. Orlandi and M. Iori Medical Physics Unit, IRCCS Arcispedale

More information

Margins in SBRT. Mischa Hoogeman

Margins in SBRT. Mischa Hoogeman Margins in SBRT Mischa Hoogeman MARGIN CONCEPTS Why do we use margins? Target / tumor To a-priori compensate for (unknown) deviations between the intended target position and the real target position during

More information

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas Dept. of Medical Physics & Engineering, Strahlenklinik, Klinikum Offenbach GmbH 63069 Offenbach, Germany

More information

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During

More information

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients IDEAL DOSIMETRY FOR IMRT/SRS Absolute dose measurement Precise

More information

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Rhys Carrington, Ysbyty Glan Clwyd, Bodelwyddan. MSc Medical Radiation Physics. Small field dosimetry

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

The CyberKnife Robotic Radiosurgery System in 2010

The CyberKnife Robotic Radiosurgery System in 2010 Technology in Cancer Research and Treatment ISSN 1533-0346 Volume 9, Number 5, October 2010 Adenine Press (2010) The CyberKnife Robotic Radiosurgery System in 2010 www.tcrt.org This review provides a complete

More information

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Quality Assurance of Ultrasound Imaging in Radiation Therapy Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Louis, Missouri Typical Applications of Ultrasound Imaging in Radiation Therapy

More information

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?

More information

Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac

Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac Iran. J. Radiat. Res., 2005; 3 (2): 73-78 Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac A.R. Khoshbin Khoshnazar 1*, M.T. Bahreyni Toossi 2, A.R. Hashemiyan

More information

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Amendment No. 2. Item No. 2 (Rfx/ Event number ) Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have

More information

A national dosimetric audit of VMAT and Tomotherapy in the UK

A national dosimetric audit of VMAT and Tomotherapy in the UK A national dosimetric audit of VMAT and Tomotherapy in the UK Catharine Clark 1,2,3, M Hussein 1,4, Y Tsang 3,5, D Wilkinson 3,5, R Thomas 2, J Snaith 2, C Gouldstone 2, G Bass 2, S Bolton 6,7, A Nisbet

More information

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

Quality assurance of volumetric modulated arc therapy using Elekta Synergy Acta Oncologica, 2009; 48: 11931197 ORIGINAL ARTICLE Quality assurance of volumetric modulated arc therapy using Elekta Synergy AKIHIRO HAGA 1, KEIICHI NAKAGAWA 1, KENSHIRO SHIRAISHI 1, SAORI ITOH 1, ATSURO

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy Sofie Ceberg, PhD Medical Physicist and Lund University Lund, Sweden Learning objectives 1: 3D Dosimetry in the Clinic: Background

More information

Agenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry

Agenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry Controlli di qualità preclinici in IMRT statica e dinamica Michele Stasi S.C. Fisica Sanitaria, A.O. Ordine Mauriziano D.O. Fisica Sanitaria, IRCCS Candiolo TORINO mstasi@mauriziano.it Agenda 1. Introduction

More information

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating Characteristic (ROC) Analysis by Allison Lorraine Mitchell Graduate Program in Medical Physics Duke University Date: Approved:

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

Brain Tumor Radiosurgery. Gabor Fichtinger, PhD

Brain Tumor Radiosurgery. Gabor Fichtinger, PhD Brain Tumor Radiosurgery Gabor Fichtinger, PhD Surgical CAD/CAM Preoperative Computerassisted planning Surgical Intraoperative Update Model Update Plan Surgical CAD CAM Patient-specific Model Computer-

More information

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms 1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,

More information

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 The study groups are requesting that each institution keep the phantom for a period of time

More information

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by: SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D. UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems

More information

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson

More information

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 IROC Lung Phantom 3D CRT / IMRT Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 The IROC requests that each institution keep the phantom for no more than 2 weeks. During

More information

A new geometric and mechanical verification device for medical LINACs

A new geometric and mechanical verification device for medical LINACs JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein

More information

Concepts of Uncertainty and their Application in Gamma Knife SRS. David J. Schlesinger, Ph.D. University of Virginia

Concepts of Uncertainty and their Application in Gamma Knife SRS. David J. Schlesinger, Ph.D. University of Virginia Concepts of Uncertainty and their Application in Gamma Knife SRS David J. Schlesinger, Ph.D. University of Virginia (djs9c@virginia.edu) Disclosures Research support: Elekta, AB Learning Objectives Review

More information

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas 1,3, Natasa Milickovic 1, Nikolaos Zamboglou 2 1 Dept. of Medical Physics & Engineering, 2 Strahlenklinik,

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

A review of RTTQA audit. Karen Venables

A review of RTTQA audit. Karen Venables A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines

More information

S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M

S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M s y n c h r o n y r e s p i r a t o r y t r a c k i n g s y s t e m S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M The Synchrony System tracks respiration in real time and automatically

More information

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry Journal of Physics: Conference Series PAPER OPEN ACCESS Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry To cite this article: N A Costa et al 2016 J. Phys.: Conf. Ser. 733 012098 View

More information

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with

More information

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize

More information

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of: Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization

More information

Dosimetry of the Gamma Knife TM

Dosimetry of the Gamma Knife TM Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife

More information

CyberKnife: Treatment Planning and Delivery Capabilities

CyberKnife: Treatment Planning and Delivery Capabilities CyberKnife: Treatment Planning and Delivery Capabilities Donald A. McRae, Ph.D. Department of Radiation Medicine Georgetown University Hospital CyberKnife System gun deck X-ray source magnetron Accelerating

More information

It s All About Margins. Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018

It s All About Margins. Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018 It s All About Margins Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018 Why margins? The smaller the better! Short Introduction Erasmus MC has been using the CyberKnife Robotic Radiosurgery System

More information

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou. Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution

More information

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System Varian Edge Experience Jinkoo Kim, Ph.D Henry Ford Health System Disclosures I participate in research funded by Varian Medical Systems. Outline of Presentation Review advanced imaging in Varian Edge Linear

More information

Dosimetric Analysis Report

Dosimetric Analysis Report RT-safe 48, Artotinis str 116 33, Athens Greece RT-safe +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ------ Irradiation system: ------

More information

Overview of MLC-based Linac Radiosurgery

Overview of MLC-based Linac Radiosurgery SRT I: Comparison of SRT Techniques 1 Overview of MLC-based Linac Radiosurgery Grace Gwe-Ya Kim, Ph.D. DABR 2 MLC based Linac SRS Better conformity for irregular target Improved dose homogeneity inside

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 IROC Prostate Phantom Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 The study groups are requesting that each institution keep the phantom for a period of time

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

Presentation Outline. Patient Setup Imaging in RT: Getting the Most Bang for your Buck. Main Errors in RT. Hypothetical Patient Examples

Presentation Outline. Patient Setup Imaging in RT: Getting the Most Bang for your Buck. Main Errors in RT. Hypothetical Patient Examples Patient Setup Imaging in RT: Getting the Most Bang for your Buck Olivier Morin, PhD UC SF Comprehensive Cancer Center San Francisco, CA Presentation Outline Errors in RT. Radiographic films in RT. On-board

More information

Intensity modulation techniques for spinal treatments and on-line VolumeView TM guidance

Intensity modulation techniques for spinal treatments and on-line VolumeView TM guidance IMAGE GUIDED RADIATION THERAPY Elekta Synergy Intensity modulation techniques for spinal treatments and on-line VolumeView TM guidance Institution: U.O. Radiobiologia Clinica, Universita Di Firenze Casa

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

University, Vellore, Tamilnadu, India, ABSTRACT

University, Vellore, Tamilnadu, India, ABSTRACT Volume 118 No. 18 2018, 4569-4578 ISSN: 1311-8080 (printed version); ISSN: 1314-3395 (on-line version) url: http://www.ijpam.eu ijpam.eu Validation of Patient-Specific Quality Assurance of Flattening Filter

More information

Intensity Modulated RadioTherapy

Intensity Modulated RadioTherapy Intensity Modulated RadioTherapy A clinical application of a small accelerator University Medical Center Groningen A.A. van t Veld PhD I. Hoveijn PhD part 1 part2 CERN/KVI accelerator school, Zeegse, June

More information

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014 IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014 The IROC Houston is requesting that each institution keep the phantom for two weeks. During

More information

IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS. A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University

IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS. A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University Content 1.Introduction 2.Methods and materials 3.Results and discussion 4.Conclusion

More information

Quality Assurance of Helical Tomotherapy Machines

Quality Assurance of Helical Tomotherapy Machines Quality Assurance of Helical Tomotherapy Machines Wolfgang Tomé 1,2, Hazim Jaradat 2, David Westerly 1, John Fenwick 3, Nigel Orton 2, Rock Mackie 1,2, and Bhudatt Paliwal 1,2 1 Human Oncology and 2 Medical

More information

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this

More information

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis Acknowledgments Thanks to Saiful Huq for an illuminating conversation about the application of TG-100 Jennifer O Daniel, Ph.D. & Fang-Fang Yin, Ph.D. Duke University Medical Center Annual AAPM Meeting,

More information

Leksell Gamma Knife Icon A New User s Perspective

Leksell Gamma Knife Icon A New User s Perspective Leksell Gamma Knife Icon A New User s Perspective Steve Herchko, DMP 2017 MFMER slide-1 Disclosures None 2017 MFMER slide-2 Outline Icon Overview Mayo Clinic Experience Frame-Based System Mask-Based System

More information

Cyberknife Stereotactic Treatment

Cyberknife Stereotactic Treatment Cyberknife Stereotactic Treatment Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA DISCLAIMER: I am not affiliated with any vendor and did not receive any financial support from any vendor.

More information

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel

More information