Patient Radiation Doses from Adult and Pediatric CT

Size: px
Start display at page:

Download "Patient Radiation Doses from Adult and Pediatric CT"

Transcription

1 Radiation Doses from CT CT Imaging Original Research Walter Huda 1 Awais Vance 1,2 Huda W, Vance A Keywords: MDCT, pediatric CT, physics of radiology, radiation dose DOI:.2214/AJR Received January 19, 2006; accepted after revision June 7, Supported in part by National Institutes of Health grant R01 EB Department of Radiology, SUNY Upstate Medical University, 750 E Adams St., Syracuse, NY Address correspondence to W. Huda (hudaw@upstate.edu). 2 Present address: Department of Diagnostic Imaging, Rhode Island Hospital, Providence, RI CME This article is available for CME credit. See for more information. AJR 2007; 188: X/07/ American Roentgen Ray Society Patient Radiation Doses from Adult and Pediatric CT OBJECTIVE. The purpose of our study was to determine typical organ doses, and the corresponding effective doses, to adult and pediatric patients undergoing a single CT examination. MATERIALS AND METHODS. Heads, chests, and abdomens of patients ranging from neonates to oversized adults (120 kg) were modeled as uniform cylinders of water. Monte Carlo dosimetry data were used to obtain average doses in the directly irradiated region. Dosimetry data were used to compute the total energy imparted, which was converted into the corresponding effective dose using patient-size-dependent effective-dose-per-unit-energy-imparted coefficients. Representative patient doses were obtained for scanning protocols that take into account the size of the patient being scanned by typical MDCT scanners. RESULTS. Relative to CT scanners from the early 1990s, present-day MDCT scanners result in doses that are 1.5 and 1.7 higher per unit mas in head and body phantoms, respectively. Organ absorbed doses in head CT scans increase from 30 mgy in newborns to 40 mgy in adults. Patients weighing less than 20 kg receive body organ absorbed doses of 7 mgy, which is a factor of 2 less than for normal-sized (70-kg) adults. Adult head CT effective doses are 0.9 msv, four times less than those for the neonate. Effective doses for neonates undergoing body CT are 2.5 msv, whereas those for normal-sized adults are 3.5 msv. CONCLUSION. Representative organ absorbed doses in CT are substantially lower than threshold doses for the induction of deterministic effects, and effective doses are comparable to annual doses from natural background radiation. t is of interest to quantify the radiation dose received by a patient un- I dergoing a CT examination, and national surveys generally show that this imaging technique is the dominant contributor to medical radiation exposure [1 3]. Knowledge of an individual organ absorbed dose permits the determination of the probability of inducing deterministic effects such as skin burns or epilation and any corresponding stochastic risks of carcinogenesis, and genetic effects [2, 4]. In addition, the dose to the conceptus of a pregnant patient quantifies any possible detrimental effect in the irradiated embryo or fetus [5]. At the relatively low radiation doses normally encountered in diagnostic radiology, deterministic effects are extremely unlikely [6]. The effective dose is currently deemed to be the best available dose descriptor for quantifying stochastic risks in diagnostic radiology [7]. The effective dose takes into account the dose and the relative radiosensitivity of all irradiated organs and may be converted into a corresponding estimate of detriment (i.e., carcinogenesis and genetic effects) if proper account is taken of the age and sex of an exposed individual [4]. The dose characteristics of new MDCT scanners merit investigation to give practitioners a better understanding of how radiation doses from these newer systems compare with the single-detector systems prevalent in the 1990s. A major effort has recently been undertaken to optimize CT [8, 9] using imaging protocols that explicitly take into account the characteristics of the patient being examined [ 12]. It is therefore of interest to investigate how the introduction of MDCT scanners and patient-size-dependent imaging protocols have affected patient doses. Patients undergoing CT examinations can range from neonates to oversized adults. Radiation doses in CT, however, are generally measured in cylindric acrylic phantoms designed to simulate a head (16-cm diameter) or body (32 cm) [13]. It is difficult to obtain reliable quantitative values of patient doses from any measurements performed in these 540 AJR:188, February 2007

2 Radiation Doses from CT TABLE 1: Radii of Water Equivalent Cylinders Used to Model Three Body Regions as a Function of Patient Age and Weight Equivalent Radius of a Patient Water Cylinder (mm) for Age Weight (kg) Head Chest Abdomen Newborn mo y y y y y Adult Oversized adult standard dose phantoms because patients have sizes and body compositions that can differ markedly from the phantoms (e.g., pediatric patients). In this article, we modeled patients as uniform cylinders of water to estimate representative CT doses for current generation MDCT scanners. We investigated patients ranging from the neonate to the oversized adult who are scanned using protocols with radiographic techniques that explicitly account for patient size. Materials and Methods Patient Size For radiation dosimetry purposes, a specified body region may be modeled as a uniform cylinder of water, for which the cylinder radius is selected so that the cylinder mass is equal to the corresponding mass of the patient [14 16]. At diagnostic photon energies in CT examinations, the pattern of energy deposition in the water cylinder will be similar to that in the corresponding patient when irradiated using similar techniques (kv, mas, beam width, pitch ratio, scan length, and so forth) with an accuracy that is adequate for most patient CT dosimetry purposes. Data on head characteristics as a function of patient age have been published for patients ranging from the neonate to the adult [17]. The third column in Table 1 summarizes the radii of water cylinders that can be used to simulate heads of patients ranging from the neonate to the adult. Patient body dimensions and the corresponding average Hounsfield values have been published by Ogden et al. [18]. Column 4 of Table 1 summarizes the resultant water equivalent radii for the chest region used in this study, and column 5 summarizes the corresponding data for the abdomen. Mean Section Dose (mgy/mas) Relative Mean Section Dose Water Phantom Radius (mm) X-Ray Tube Voltage (kv) Dosimetry Organ absorbed doses When a water cylinder (radius r) located at a CT scanner isocenter undergoes a single rotation of the X-ray tube (i.e., 360 ) at a fixed technique value (kv/mas), approximately half the energy will be deposited in the directly irradiated region and the remaining half in scatter tails on either side of the directly irradiated region [19]. The mean section dose, D m, is defined as the total energy deposited in the water cylinder divided by the mass of the directly irradiated volume, or π r 2 T ρ, where T is the section thickness and ρ is the water density [20]. The mean section dose approximates the average (i.e., representative) dose in the directly irradiated region of the water cylinder for contiguous scanning in the axial scanning mode, or using a pitch ratio of 1 in the helical scanning mode. Figure 1 shows how the mean section dose, D m, varies with water cylinder radius obtained using Monte Carlo modeling techniques for a HiSpeed Fig. 1 Mean section dose (D m ) versus cylindric water phantom radius for a HiSpeed Advantage CT scanner (GE Healthcare) operated at 120 kv. Fig. 2 Values of relative mean section dose (R kv ) for HiSpeed Advantage scanner (GE Healthcare) normalized to unity at 120 kv, versus X-ray tube voltage for water cylinders with radius of 20 (dashed line), 60 (dotted line), and 200 (solid line) mm. Advantage CT scanner (GE Healthcare) operated at 120 kv [20]. Increasing the cylinder radius from 20 to 200 mm reduces the mean section dose from to mgy/mas, or by a factor of about 6. Figure 2 shows the relative mean section dose, R kv, as a function of X-ray tube voltage (constant mas), and normalized to unity at 120 kv for each phantom size. Increasing the X-ray tube voltage from 80 to 140 kv at a fixed mas increases D m by a factor of about 5.5. For the HiSpeed Advantage scanner at given kv and mas values, representative organ doses, D organ, may be obtained using the equation D organ =(D m mas R kv ) / P, (1) where P is the pitch ratio and R kv is taken from the data shown in Figure 2. The HiSpeed Advantage was introduced into clinical practice in the early 1990s and is not representative of current MDCT scanners. Table 2 summarizes data for head- and body-weighted AJR:188, February

3 TABLE 2: CTDI w for Head and Body Dose Phantoms for CT Scanners (120 kv) from Five Commercial Vendors CTDI w (mgy/mas) Vendor Model Detectors (mm) Head a Body b GE Healthcare LightSpeed Plus Philips Medical Systems Mx8000 IDT Siemens Medical Solutions Sensation Shimadzu Corporation 7000 TS Toshiba Medical Systems Aquilion Multi Average ± SD of all five systems ± ± Note CTDI w = Weighted CT dose index. a HiSpeed Advantage (GE Healthcare), mgy/mas (head/120 kv). b HiSpeed Advantage (GE Healthcare), mgy/mas (body/120 kv). Fig. 3 Effective dose per unit energy imparted, E / ε, versus patient weight for head (solid line) and body (dashed line) CT examinations. Effective Dose per Unit Energy Imparted (msv/j) 0 CT dose index (CTDI w ) data for recent CT systems from five major vendors [13]. The data presented in Table 2 indicate that head and body doses on presentday scanners (mgy/mas), when operated at 120 kv, would be 1.45 and 1.71 times higher, respectively, than those shown in Figure 1. Energy imparted and effective dose Dose is the energy per unit mass; accordingly, the total energy imparted to a cylindric water phantom (ε), with a mean section dose D m per 360 rotation of the X-ray tube, is ε =(D m ) (π r 2 T ρ) N, (2) where T is the nominal X-ray beam width and N is the total number of rotations of the X-ray tube around the phantom. For a CT scanner operated in the helical mode, N may be estimated by dividing the full scan length by the product of the X-ray beam width and the pitch ratio P. For a given patient size and scanned body region, the effective dose is directly proportional to 0 the energy imparted to the patient. Doubling the energy imparted to a patient undergoing head CT, for example, is generally expected to double the corresponding effective dose. Published Monte Carlo calculations of radiation doses at CT examinations provide the complete pattern of energy deposition in the patient and enable the mean dose to each organ and tissue to be determined. These organ doses enable the computation of the effective dose (E) and the energy imparted (ε) [20], where the ratio of these two parameters, E / ε, permits energy imparted to be converted into a corresponding effective dose using E = ε (E / ε) R,M, (3) where (E / ε) R,M is the effective dose per unit energy imparted coefficient specific for a body region R (head, chest, abdomen, and so forth.) and for a patient of weight M. Figure 3 shows (E / ε) R,M used in this study to convert energy imparted obtained from equation 2 into the corresponding value of ef- fective dose; these values were taken using published data for head [17] and body [21, 22] CT examinations. Differences between the (E / ε) for the chest and abdomen are only about 2%, and the body curves shown in Figure 3 are the mean chest and abdomen values. Protocols Patient doses in CT examinations depend on the choice of radiographic technique factors used to perform the scanning. Key parameters that affect patient dose, and which need to be defined in CT protocols, are the following: X-ray tube voltage (kv) Figure 2 shows how the choice of X-ray tube voltage affects patient dose in CT. X-ray tube current and scanning rotation time (mas) Patient doses will be directly proportional to the choice of mas. Axial or helical scanning In helical scanning, patient doses are inversely proportional to the pitch ratio, assuming a constant scan length. In axial scanning, it is possible to estimate doses by taking the axial pitch to be numerically equal to the ratio of the X- ray beam width to the table increment distance, and using this value in equation 1 for determining D organ. Scan length The total energy put into a patient is directly proportional to the total scan length. For helical scans, one additional rotation (20-mm beam width) was added when computing equation 2 to determine energy imparted to account for the overscanning required for interpolation purposes. Table 3 presents scanning protocols that explicitly take into account patient size. These protocols are based on those used at our own institution and are generally similar to those adopted in radiology departments that specialize in pediatric imaging (Frush DP, personal communication). Results Figure 4 shows mean section doses, D m, as a function of patient weight for current scanners at 120 kv, the X-ray tube voltage most commonly used clinically. The highest doses occur when scanning neonate chests (0.24 mgy/mas), and the lowest doses when scanning the abdomen of an oversized (120-kg) adult (0.067 mgy/mas). As expected, patient weight is more important in determining mean section doses in body imaging than in head imaging. Data in Figure 4 illustrate the intrinsic dose characteristics of CT scanners per unit mas and show the importance of taking the patient weight into account when performing patient dose estimates. Pediatric doses will be much larger than those of adults if radiographic techniques (mas) are kept constant for all CT examinations. 542 AJR:188, February 2007

4 Radiation Doses from CT TABLE 3: Patient Age- and Weight-Specific CT Protocols (120 kv) Used to Determine Representative Patient Organ and Effective Doses Patient Head (Axial or Contiguous) Chest (Pitch = 1.5) Abdomen (Pitch = 1.5) Age Weight (kg) mas Scan Length (cm) mas Scan Length (cm) mas Scan Length (cm) Newborn mo y y y y y Adult Oversized adult Fig. 4 Mean section dose (D m ) versus patient weight for an average of five MDCT systems listed in Table 2 operated at 120 kv for CT examinations of head (solid line), chest (dotted line), and abdomen (dashed line), where curves are least square fits to a secondorder polynomial. Mean Section Dose (mgy/mas) Figure 5 shows patient doses computed using the head protocol scanning parameters listed in Table 3, and Figure 6 shows the corresponding results obtained for body scans. The coefficients of determinations (r 2 ) for the six fitted curves shown in Figures 5 and 6 were all between 0.97 and Figure 5A shows organ absorbed doses in head CT increasing from 30 mgy in newborns to 40 mgy in adults. Figure 6A shows chest and abdominal CT result in similar doses; it also shows similar trends with changes in patient weight. In body CT, patients weighing less than 20 kg have organ doses of 7 mgy, which is a factor of 2 lower than that of a normal-sized (70-kg) adult. Figure 5B shows head CT effective doses for adults are 0.9 msv, but that neonate effective doses are approximately four times 0 higher. Figure 6B shows the effective dose for a neonate undergoing chest CT is 2.2 msv; the effective dose falls to a minimum of 1.5 msv for a -kg infant and increases to 4 msv for a normal-sized adult. A neonate undergoing abdominal CT has an effective dose of 3 msv; the effective dose falls to a minimum of 2 msv for a -kg infant and then increases to 3 msv for a normal-sized adult. Discussion The data presented in Table 2 show that modern MDCT scanners have normalized doses (i.e., CTDI w per unit mas) that are higher than those of single-detector CT scanners used clinically in the early 1990s. MDCT detector arrays have short scanning times of 0.4 seconds or less, and to maintain a reasonable exposure at the CT detector has required a reduction in the X-ray tube to isocenter distance. MDCT technology uses X- ray beams that are wider than the detector array (overbeaming), which also increases patient doses. Further changes that may increase patient doses include changes in beam-shaping filters (bow tie) and the use of X-ray tubes with metal frames that have replaced glass. After the acquisition of new CT equipment, protocols should be reviewed to help ensure that patient doses are kept as low as reasonably achievable (ALARA), and that any increase in patient dose is justified by a corresponding improvement in diagnostic information [4, 23]. The mean section dose (D m ) shown in Figure 1 relates to an old scanner, whereas those in Figure 4 pertain to an average of major CT scanners in use today. To obtain doses for any current (or future) scanner, it is possible to use the data shown in Figure 1 multiplied by a scale factor equal to the head (or body) CTDI value for the scanner of interest divided by the corresponding HiSpeed Advantage CTDI value given in the footnotes of Table 2. The data shown in Figure 2 show the importance of the choice of X-ray tube voltage (kv) on patient dose. Increasing the X-ray tube voltage from 80 to 140 kv at a constant mas results in a fivefold increase in patient dose, which is approximately independent of phantom size. Although most CT examinations to date have been performed at 120 kv, there may be specific diagnostic tasks for which the use of lower kv values could offer substantial dose reductions with no loss of diagnostic information. For example, visualization of iodinated contrast medium administered to patients could be performed at 80 kv to maximize iodine contrast resulting from increased photoelectric absorption by iodine at lower photon energies, which could substantially reduce patient doses [24]. For a novel scanner operating at an X-ray tube voltage that differs from 120 kv, two methods are possible to estimate patient doses at the new kv value [1]: use approximate kv correction factors from Figure 2; or use R kv values measured in head or body dosimetry phantoms. For head and pediatric CT examinations, the pattern of dose distribution in phantoms and patients is relatively uniform, and peripheral (skin) and central doses are similar. Accordingly, D m and organ doses for head and pediatric CT examinations may be taken to represent the dose to directly irradiated organs. In larger patients, however, the central AJR:188, February

5 50 4 Organ Dose (mgy) Patient Age (y) A Fig. 5 Radiation doses versus patient age for head CT examinations based on protocol in Table 3 for average of five MDCT systems listed in Table 2 operated at 120 kv. Curves are least square fits to a second-order polynomial. A and B, Graphs show organ (absorbed) (A) and effective (B) doses. Organ Dose (mgy) A Fig. 6 Radiation doses versus patient weight for body CT based on protocol in Table 3 for average of five MDCT systems listed in Table 2 operated at 120 kv for chest (dotted line) and abdomen (dashed line), where curves are least square fits to a second-order polynomial. A and B, Graphs show organ (absorbed) (A) and effective (B) doses. dose is expected to be lower than the peripheral dose. Central CTDIs in body phantoms, for example, are generally a factor of 2 lower than those obtained at the periphery [13]. Peripheral-to-central dose ratios as a function of phantom size have been published [25], which enable the average organ doses (Figs. 1 and 4) to be used to estimate the expected regional variations in dose for larger patients. CT scanners that use ma modulation will result in a more uniform pattern of dose deposition in patients [26], and mean section doses obtained by use of an average ma are unlikely to be in serious error. Effective Dose (mgy) Effective Dose (mgy) The standard CTDI dosimetry phantoms are made of acrylic and have a relatively high density of 1.19 g cm 3. The head phantom has a 16-cm diameter, which is equivalent to a water cylinder with a radius of 87 mm, and is therefore similar to water cylinder radii for older children and heads of adults (Table 1). The body phantom has a 32-cm diameter, which corresponds to a water phantom with a radius of 175 mm, and is much larger than typical adult body dimensions (Table 1). Accordingly, body CTDI will generally underestimate true patient organ dose because these CTDI measurements are made in a phantom that is much Patient Age (y) 0 larger than a normal-sized adult. Data in Figure 1 show that correcting for patient size would increase body CTDI by 50% for abdominal scans and double CTDIs for chest scans. CTDI doses are specified in acrylic or air, which are typically 30% lower than tissue doses [27]. In addition, CTDI uses integration distances of either ± seven slice-thickness values or 0 mm, which can underestimate true tissue doses by as much as 70% [28]. For all these reasons, CTDIs are poor predictors of patient organ doses; organ doses presented in Figures 1 and 4 will be superior because they are specified as soft-tissue doses, take into ac- B B 544 AJR:188, February 2007

6 Radiation Doses from CT TABLE 4: Typical Effective Dose Values Encountered in Everyday Life Affected Group Effective Dose Comments Radiation workers Current dose limit 50 msv/y Current regulatory limit for most radiation workers in U.S. Recommended dose limit 20 msv/y Limit recommended by ICRP and NCRP Occupational doses 5 msv/y Average doses of the most highly exposed radiation workers a Fetus of a radiation worker b 5 msv b Dose limit introduced to permit women of reproductive capacity to be radiation workers Pilots and air crew (~1,000 flight h) 5 msv/y At 30,000 feet, one receives ~5 µsv/h radiation exposure due to additional cosmic background General public Natural background radiation (U.S.) 3 msv/y 0.3 msv from cosmic radiation, 0.3 msv from terrestrial radioactivity, 0.4 msv from internal radionuclides (e.g., 40 K), 2 msv from radon Dose limit 1 msv/y Excludes natural background and medical exposures Note ICRP = International Commission on Radiological Protection, NCRP = National Council on Radiation Protection and Measurements. a For example, interventional radiologists and nuclear plant maintenance staff. b From declaration of pregnancy to birth, with additional restriction on dose rate (0.5 msv/mo). count the patient size, and have been computed for scan lengths representative of clinical practice (Table 3). Organ doses may be used to predict the probability of a deterministic effect such as skin erythema, epilation, and the induction of eye cataracts; a threshold dose of approximately 2 Gy is normally adopted for use in clinical practice [6]. Typical organ doses in head (Fig. 5) and body (Fig. 6) CT examinations are low (< 0.05 Gy), and deterministic effects are therefore most unlikely for protocols similar to those listed in Table 3. Deterministic effects are nonetheless possible in CT and have been reported for CT perfusion studies, when high techniques are combined with repeated scanning over the same head region [29]. Organ doses presented in this study may be used to estimate embryo or fetal doses in a pregnant patient. Our data suggest that embryo and fetal doses would generally be well below the action threshold of 0 mgy currently recommended by the International Commission on Radiological Protection (ICRP) [30]. Doses presented in Figures 4 and 6A are only approximate, and accurate embryo and fetal dose estimates may need to take into account the pattern of dose distribution in the patient and the location of the embryo or fetus [31]. Note that body organ absorbed doses (7 14 mgy) are considerably lower than those for head CT, whereas body effective doses are generally higher than effective doses for most head CT studies. The effective dose combines organ doses with their relative radiosensitivity (i.e., ICRP weighting factor [4]), and the low head effective dose reflects the absence of radiosensitive organs in the head. Although the chest and abdomen have lower organ doses, they contain the most radiosensitive organs in the body (i.e., lung, red bone marrow, colon, stomach), which results in effective doses that are much higher than those for head CT. Data in Figure 5B show that infants absorb less energy than adults but have effective doses that are about four times higher than those of adults. This increase in effective dose for smaller patients, when scanned at the same techniques as adults, has also been reported by other investigators [32, 33]; high effective doses in infant head CT occur because adjacent organs such as the thyroid and lung are small and their proximity causes them to receive proportionally more scattered radiation [15]. Adult effective doses reported here are somewhat lower than the values of 1.3 msv recently reported using a similar methodology for head examinations [15], 5.4 msv for chest examinations [16], and 3.9 msv for abdominal examinations [14]. Effective doses generally reflect the choice of scanning protocol used to perform a given CT examination, and our computed effective dose values are low because we used low mas values, minimized the scan length, and used a pitch of 1.5 for helical scanning (Table 3). It is noteworthy that current pediatric doses are now much lower than reported previously; for a -kg patient, the effective dose for a chest CT examination has been reduced from 9.6 to 1.5 msv [16], and the corresponding reduction for an abdominal CT examination is from 7 to 2 msv [14]. Because of the increased radiosensitivity of children and the increasing utilization of CT in this age group, it is most welcome that pediatric doses can be significantly reduced with no apparent loss of diagnostic information [1, 34]. Our methodology is based on modeling patients as uniform cylinders of water for the purposes of estimating the pattern of energy deposition and the total energy imparted to the patient. This simplification should not result in large dose errors because the attenuation properties of water for CT spectra, where the Compton effect is dominant, would be similar to those of patients, who are composed of bone and tissue. Our approach should approximate absolute patient dose values, but is expected to provide reliable relative dose estimates when technique factors (kv, mas, pitch) or the CT scanner type are changed. One major limitation of our study is that we have not explicitly included the effect of X- ray tube ma modulation as the X-ray tube rotates around the patient or along the patient s long axis. For a given body region, patient doses are expected to be directly proportional to the average mas value. If the average mas values were lower for a given CT scanner than those used in this study (Table 3), then the patient doses would be correspondingly lower. Patient radiation risks may be estimated from the effective doses given in this study. For example, the ICRP specifies that the nominal risk coefficient for the induction of fatal cancer is 5%/Sv, and the total detriment (i.e., induction of all cancers and genetic effects) is 7.3%/Sv when risk factors are averaged over a whole population [4]. However, note that any risk estimates need to take into account patient demographics, with large differences in the risks per unit effective dose between infants and older adults [8]. Of greater importance is the fact that there are large uncertainties in risk AJR:188, February

7 estimates at dose levels normally encountered in CT [2]. One helpful way to better understand the significance of a given CT effective dose value is to compare it with other types of radiation exposure encountered in society at large (Table 4). Patient doses in CT are broadly comparable to annual doses from natural background, and any risks from a single CT scan should be comparable to those from background doses delivered to every inhabitant of our planet each year. Acknowledgments We gratefully acknowledge the technical support provided by M. L. Roskopf and R. L. Lavallee and valuable discussions with D. P. Frush and T. Toth. References 1. Linton OW, Mettler FA Jr, National Council on Radiation Protection and Measurements. National conference on dose reduction in CT, with an emphasis on pediatric patients. AJR 2003; 181: UNSCEAR Sources and effects of ionizing radiation. New York, NY: United Nations, Shrimpton PC, Hillier MC, Lewis MA, Dunn M. Doses from computed tomography (CT) examinations in the UK: 2003 review. NRPB-W67. Chilcot, UK: National Radiological Protection Board, International Commission on Radiological Protection (ICRP) publication Recommendations of the International Commission on Radiological Protection. Ann ICRP 1991; 21: Wagner LK, Lester RG, Saldana LR. Exposure of the pregnant patient to diagnostic radiations: a guide to medical management, 2nd ed. Madison, WI: Medical Physics Publishing, Wagner LK, Eifel PJ, Geise RA. Potential biological effects following high X-ray dose interventional procedures. J Vasc Interv Radiol 1994; 5: McCollough CH, Schueler BA. Calculation of effective dose. Med Phys 2000; 27: Brenner D, Elliston C, Hall E, Berdon W. Estimated FOR YOUR INFORMATION risks of radiation-induced fatal cancer from pediatric CT. AJR 2001; 176: Paterson A, Frush DP, Donnelly LF. Helical CT of the body: are settings adjusted for pediatric patients? AJR 2001; 176: Huda W, Scalzetti EM, Levin G. Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 2000; 217: Donnelly LF, Emery KH, Brody AS, et al. Minimizing radiation dose for pediatric body applications of single-detector helical CT: strategies at a large Children s Hospital. AJR 2001; 176: Judy PF. Comment on In x-ray computed tomography, technique factors should be selected appropriate to patient size. Med Phys 2001; 28: Nagel HD. Radiation exposure in CT, 4th ed. Hamburg, Germany: CTB Publications, 2002: Ware DE, Huda W, Mergo PJ, Litwiller AL. Radiation effective doses to patients undergoing abdominal CT examinations. Radiology 1999; 2: Huda W, Chamberlain CC, Rosenbaum AE, Garrisi W. Radiation doses to infants and adults undergoing head CT examinations. Med Phys 2001; 28: Huda W, Scalzetti EM, Roskopf M. Effective doses to patients undergoing thoracic computed tomography examinations. Med Phys 2000; 27: Huda W, Lieberman KA, Chang J, Roskopf ML. Patient size and x-ray technique factors in head computed tomography examinations. I. Radiation doses. Med Phys 2004; 31: Ogden K, Huda W, Scalzetti EM, Roskopf ML. Patient size and x-ray transmission in body CT. Health Phys 2004; 86: Beck JW, Dunn WL, O Foghludha F. A Monte Carlo model for absorbed dose calculations in computed tomography. Med Phys 1983; : Huda W, Atherton JV, Ware DE, Cumming WA. An approach for the estimation of effective radiation dose at CT in pediatric patients. Radiology 1997; 203: Atherton JV, Huda W. Energy imparted and effective doses in computed tomography. Med Phys 1996; 23: This article is available for CME credit. See for more information. 22. Huda W, Atherton JA. Erratum: Energy imparted and effective doses in computed tomography. Med Phys 2003; 30: International Commission on Radiological Protection (ICRP) publication 87. Managing patient dose in computed tomography. Ann ICRP 2000; 30(3): Huda W, Lieberman KA, Chang J, Roskopf ML. Patient size and x-ray technique factors in head computed tomography examinations. II. Image quality. Med Phys 2004; 31: Nickoloff EL, Dutta AK, Lu ZF. Influence of phantom diameter, kvp and scan mode upon computed tomography dose index. Med Phys 2003; 30: Kalender WA, Wolf H, Suess C. Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements. Med Phys 1999; 26: Johns HE, Cunningham JR. The physics of radiology, 4th ed. Springfield, IL: Charles C Thomas, Boone J, Cooper VN 3rd, Nemzek W, McGahan J, Seibert J. Monte Carlo assessment of computed tomography dose to tissue adjacent to the scanned volume. Med Phys 2000; 27: Imanishi Y, Fukui A, Niimi H, et al. Radiation-induced temporary hair loss as a radiation damage only occurring in patients who had the combination of MDCT and DSA. Eur Radiol 2005; 15: International Commission on Radiological Protection (ICRP) publication 84. Pregnancy and medical radiation. Ann ICRP 2000; 30(3): Felmlee JP, Gray JE, Leetzow ML, Price JC. Estimated fetal radiation dose from multislice CT studies. AJR 1990; 154: Khursheed A, Hillier MC, Shrimpton PC, Wall BF. Influence of patient age on normalized effective doses calculated for CT examinations. Br J Radiol 2002; 75: Chapple CL, Willis S, Frame J. Effective dose in paediatric computed tomography. Phys Med Biol 2002; 47: Slovis TL. The ALARA concept in pediatric CT: myth or reality? Radiology 2002; 223: AJR:188, February 2007

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

How to Develop CT Protocols for Children

How to Develop CT Protocols for Children How to Develop CT Protocols for Children Introduction Prior to 2001 the vast majority of CT imaging of children was conducted using the same or similar techniques used for adult imaging. In 2001, several

More information

Doses from CT examinations to children suffering from hydrocephalus

Doses from CT examinations to children suffering from hydrocephalus Doses from CT examinations to children suffering from hydrocephalus E. G. Friberg 1, L. J. Holmedal 2, I. Børretzen 1, K. Rosendahl 3 and H. M. Olerud 1 1 Norwegian Radiation Protection Authority, P.O.

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

Radiation Dose in Pediatric Imaging

Radiation Dose in Pediatric Imaging Radiation Dose in Pediatric Imaging A Brief History of Radiology Dose: Why Does It Matter? Measuring Exposure and Dose Deterministic Effects Stochastic Effects Common Exams: What is the Risk? Reducing

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial Medical Physics and Informatics Original Research Larke et al. Estimated Radiation Dose for Low-Dose Chest CT Medical Physics and Informatics Original Research Frederick J. Larke 1 Randell L. Kruger 2

More information

Estimating Patient Radiation Dose from Computed Tomography

Estimating Patient Radiation Dose from Computed Tomography Estimating Patient Radiation Dose from Computed Tomography C. Cagnon, J. DeMarco, E. Angel, M. McNitt-Gray UCLA David Geffen School of Medicine 1 Patient Dose from CT Advances in Technology... Helical

More information

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation exposure of the Yazd population from medical conventional X-ray examinations Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

National Cancer Institute

National Cancer Institute National Cancer Institute A Dosimetry Summary of CT Participants in the National Lung Screening Trial (NLST) U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES National Institutes of Health AAPM 2015 Anaheim,

More information

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Conventional and spiral CT dose indices in Yazd general hospitals, Iran Iran. J. Radiat. Res., 2006; 3 (4): 183-189 Conventional and spiral CT dose indices in Yazd general hospitals, Iran F. Bouzarjomehri 1*,M.H.Zare 2, D. Shahbazi 2 1 Department of Medical Physics, Shahid

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

State of the art and future development for standardized estimation of organ doses in CT

State of the art and future development for standardized estimation of organ doses in CT State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida

More information

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Omer Osman,Saeed Medical Physics Department, ALNeelain University, Sudan Presentation outlines Introduction Objectives Materials and

More information

Fetal Dose Calculations and Impact on Patient Care

Fetal Dose Calculations and Impact on Patient Care Fetal Dose Calculations and Impact on Patient Care Matt Hough, MS, DABR, DABMP Florida Hospital Diagnostic Medical Physics and Radiation Safety Resource ACR-SPR Practice Parameter for Imaging Pregnant

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates Patient Dose Estimates from CT Examinations Patient Dose Estimates from CT Examinations John M. Boone, Ph.D. Professor and Vice Chairman of Radiology Professor of Biomedical Engineering University of California,

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Christner et al. Estimating Effective Dose for CT Medical Physics and Informatics Original Research FOCUS ON: Jodie A. Christner 1 James M. Kofler Cynthia

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua Radiation Safety For Anesthesiologists R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua Modern World Non Ionizing VS Ionizing Non Ionizing Harmless Ex. visible

More information

Radiographic techniques in screen-film mammography

Radiographic techniques in screen-film mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 3, SUMMER 2002 Radiographic techniques in screen-film mammography Thomas R. LaVoy* 4314 Kelsey Drive, Syracuse, New York 13215 Walter Huda

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference?

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference? Pediatric Imaging Original Research Reid et al. Parameter Selection for Pediatric Abdominal CT Downloaded from www.ajronline.org by 46.3.3.24 on 2/3/18 from IP address 46.3.3.24. Copyright ARRS. For personal

More information

Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection. John Harrison ICRP Committee 2

Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection. John Harrison ICRP Committee 2 Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection John Harrison ICRP Committee 2 Joint ICRP-RERF-JHPS Workshop: Tokyo, December 2017 Task Group 79 : Use of Effective Dose

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan *M.T.Bahreyni Toossi, **S.Mohandes Dastgherdi Medical Physics Dep., Faculty

More information

Radiation Exposure 1980 to 2006

Radiation Exposure 1980 to 2006 Radiation Exposure 1980 to 2006 Background 3-6 msv/yr Natural (85% 45%) Radon Cosmic Rays Air travel Living at Altitude Man-made (15% 55%) Medical Imaging** mgy Radiation Therapy cgy Radiation Whole Body

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation Medical Physics and Informatics Original Research Angel et al. Radiation Dose During Chest CT Medical Physics and Informatics Original Research Erin Angel 1,2 Nazanin Yaghmai 1 Cecilia Matilda Jude 1 John

More information

Radiation Safety for New Medical Physics Graduate Students

Radiation Safety for New Medical Physics Graduate Students Radiation Safety for New Medical Physics Graduate Students John Vetter, PhD Medical Physics Department UW School of Medicine & Public Health Background and Purpose of This Training This is intended as

More information

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations Aiping Ding, X. George Xu Rensselaer Polytechnic Institute Troy, NY USA http://rrmdg.rpi.edu 1 Acknowledgements

More information

Biological Effects of Ionizing Radiation Module 8 - AAPM/RSNA Curriculum. Basic Radiation Biology

Biological Effects of Ionizing Radiation Module 8 - AAPM/RSNA Curriculum. Basic Radiation Biology Biological Effects of Ionizing Radiation Module 8 - AAPM/RSNA Curriculum Basic Radiation Biology Kalpana M. Kanal, PhD, DABR Associate Professor, Radiology Director, Resident Physics Education a copy of

More information

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result. Minimizing Dose Understand the importance and methods of pre-procedure patient assessment including a review of previous radiologic exams, disease processes and anatomical considerations that may increase

More information

CT Dosimetry in the Clinical Environment: Methods and Analysis

CT Dosimetry in the Clinical Environment: Methods and Analysis CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida

More information

CT Dose Reduction in Pediatric Patients

CT Dose Reduction in Pediatric Patients CT Dose Reduction in Pediatric Patients By Kelly Firestine, RT(R)(CT)(M) Executive Summary CT is an incredibly valuable imaging tool, but there are unique concerns with pediatric patients, including the

More information

Current status of diagnostic imaging in dental university hospitals in Japan

Current status of diagnostic imaging in dental university hospitals in Japan Oral Radiol (2004) 20:15 21 Japanese Society for Oral and Maxillofacial Radiology and Springer-Verlag Tokyo 2004 DOI 10.1007/s11282-004-0010-3 ORIGINAL ARTICLE Takehito Sasaki Minoru Fujita Tsuguhisa Katoh

More information

Patient / Organ Dose in CT

Patient / Organ Dose in CT Patient / Organ Dose in CT Patient specific and organ dose estimation H.D. Nagel Dr. HD Nagel, Science & Technology for Radiology Buchholz / Germany www.sascrad.com 1 Topics CTDI & patient dose SSDE Organ

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Optimizing radiation dose by varying age at pediatric temporal bone CT

Optimizing radiation dose by varying age at pediatric temporal bone CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Optimizing radiation dose by varying age at pediatric temporal bone CT Daichi Noto, 1 Yoshinori Funama, 2a Mika Kitajima, 3 Daisuke

More information

GUIDELINES ON IONISING RADIATION DOSE LIMITS AND ANNUAL LIMITS ON INTAKE OF RADIOACTIVE MATERIAL

GUIDELINES ON IONISING RADIATION DOSE LIMITS AND ANNUAL LIMITS ON INTAKE OF RADIOACTIVE MATERIAL RADIATION PROTECTION AUTHORITY OF ZIMBABWE (RPAZ) RADIATION PROTECTION ACT [CHAPTER 15:15] GUIDELINES ON IONISING RADIATION DOSE LIMITS AND ANNUAL LIMITS ON INTAKE OF RADIOACTIVE MATERIAL Compiled by Radiation

More information

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Assessment of Patient Dose from CT Examinations in Khorasan, Iran Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 2012, 233-238 Received: July 18, 2012; Accepted: November 17, 2012 Original Article Assessment of Patient Dose from CT Examinations in Khorasan,

More information

Application of the Commission's Recommendations for the Protection of People in

Application of the Commission's Recommendations for the Protection of People in ICRP Publication 127 ICRP Publication 126 ICRP Publication 125 ICRP Publication 124 ICRP Publication 123 ICRP Publication 122 ICRP Publication 121 ICRP Publication 120 ICRP 2011 Proceedings Radiological

More information

Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose)

Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose) Eur Radiol (2002) 12:1571 1576 DOI 10.1007/s00330-001-1255-4 PHYSICS Holger Greess Anton Nömayr Heiko Wolf Ulrich Baum Michael Lell Bernhard Böwing Willi Kalender Werner A. Bautz Dose reduction in CT examination

More information

ICRP 128 ICRP ICRP ICRP 1928

ICRP 128 ICRP ICRP ICRP 1928 ICRP 1928 129 ICRP 1928 ICRP ICRP ICRP 1928 129 ICRP 129 ICRP 128 Radiological Protection in Cone Beam Computed Tomography (CBCT) Radiation Dose to Patients from Radiopharmaceuticals: A Compendium of Current

More information

Patient dose assessment of CT perfusion scanning at the RSCH

Patient dose assessment of CT perfusion scanning at the RSCH Patient dose assessment of CT perfusion scanning at the RSCH Lesley Leavesley, Emma Whitehead, Matthew Pryor, Debbie Peet Regional Radiation Protection Service Royal Surrey County Hospital, Guildford Overview

More information

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Rebecca Milman Marsh, Ph.D. University of Colorado Department of Radiology Who in the Facility Works

More information

Paediatric Dose Reduction and Image Quality

Paediatric Dose Reduction and Image Quality Paediatric Dose Reduction and Image Quality Alan Whiteside The majority of this work was undertaken as part of MSc Thesis of Helen Dixon. Introduction Paediatric CT protocols result in a higher effective

More information

Estimating the Radiation-Induced Cancer Risks in Pediatric Computed Tomography

Estimating the Radiation-Induced Cancer Risks in Pediatric Computed Tomography Iranian Journal of Medical Physics Vol. 13, No. 4, December 2016, 218-227 Received: August 20, 2016; Accepted: November 04, 2016 Original Article Estimating the Radiation-Induced Cancer Risks in Pediatric

More information

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information

Estimation of the Risk of Cancer Associated with Pediatric Cranial Computed Tomography

Estimation of the Risk of Cancer Associated with Pediatric Cranial Computed Tomography British Journal of Medicine & Medical Research 9(10): 1-7, 2015, Article no.bjmmr.18079 ISSN: 2231-0614 SCIENCEDOMAIN international www.sciencedomain.org Estimation of the Risk of Cancer Associated with

More information

IMAGE GENTLY HOW CAN YOU HELP?

IMAGE GENTLY HOW CAN YOU HELP? IMAGE GENTLY HOW CAN YOU HELP? Keith J. Strauss, MSc, FAAPM, FACR Director, Radiology Physics & Engineering Children s s Hospital Boston Harvard Medical School Acknowledgment Marilyn J. Goske,, MD Robert

More information

Quality Control and Patient Dosimetry on line for Computed Tomography

Quality Control and Patient Dosimetry on line for Computed Tomography Quality Control and Patient Dosimetry on line for Computed Tomography Jose I. Ten 1,2, Eliseo Vano 2,3, Jose M. Fernandez-Soto 2,3, Roberto Sanchez 3, Juan Arrazola 1,2 1 Diagnostic Radiology Service and

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

Estimates of Risks LONG-TERM LOW DOSE EFFECTS OF IONIZING RADIATION

Estimates of Risks LONG-TERM LOW DOSE EFFECTS OF IONIZING RADIATION Estimates of Risks LONG-TERM LOW DOSE EFFECTS OF IONIZING RADIATION Low Level Radiation Exposure Single exposure of 10 rad or less Larger exposures delivered over periods of days or longer (low dose

More information

Chief Radiographer TEI Clinical Associate 2016

Chief Radiographer TEI Clinical Associate 2016 MDCT Principles i and Applications Ε ΑGADAKOS MSc Ε. ΑGADAKOS MSc Chief Radiographer TEI Clinical Associate 2016 Aim To understand d recent technological advances in MSCT and how they can be effectively

More information

Estimated cumulative radiation dose from PET/CT in children with malignancies: a 5-year retrospective review

Estimated cumulative radiation dose from PET/CT in children with malignancies: a 5-year retrospective review Pediatr Radiol (1) :681 686 DOI 1.17/s247-9-1434-z ORIGINAL ARTICLE Estimated cumulative radiation dose from PET/CT in children with malignancies: a 5-year retrospective review Soni C. Chawla & Noah Federman

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Minimizing CT Radiation Dose CT examinations improve health care and are an essential part

More information

3 rd International Symposium on the System of Radiological Protection Seoul, October John Harrison

3 rd International Symposium on the System of Radiological Protection Seoul, October John Harrison 3 rd International Symposium on the System of Radiological Protection Seoul, October 2015 John Harrison UK Task Group 79 : Use of Effective Dose as a Risk-related Radiological Protection Quantity John

More information

Table of Contents. Introduction 3. Background 4

Table of Contents. Introduction 3. Background 4 Training manual Table of Contents Introduction 3 Background 4 What are X-rays? 4 How are X-rays Generated? 5 Primary and Scatter Radiation 6 Interactions with Matter 6 Biological Effects of Radiation 7

More information

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine Radiation physics and radiation protection University of Szeged Department of Nuclear Medicine Radiation doses to the population 1 Radiation doses to the population 2 Sources of radiation 1 Radiation we

More information

Managing Patient Dose in Computed Tomography (CT)

Managing Patient Dose in Computed Tomography (CT) Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES A THESIS SUBMITTED TO THE GRADUATE FACULTY in partial fulfillment of the requirements

More information

PRINCIPLES AND METHODS OF RADIATION PROTECTION

PRINCIPLES AND METHODS OF RADIATION PROTECTION PRINCIPLES AND METHODS OF RADIATION PROTECTION Lesson Outcomes At the end of the lesson, student should be able to: Define what is radiation protection (RP) Describe basic principles of RP Explain methods

More information

EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC)

EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC) EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC) BY KOFI OKYERE AKYEA-LARBI 10442894 THIS THESIS IS SUBMITTED TO THE MEDICAL

More information

ICRP RECOMMENDATIONS AND IAEA SAFETY STANDARDS: THEN AND NOW

ICRP RECOMMENDATIONS AND IAEA SAFETY STANDARDS: THEN AND NOW ICRP RECOMMENDATIONS AND IAEA SAFETY STANDARDS: THEN AND NOW Abdalla N. Al-Haj, PhD, FIPEM, CRadP, CSci, MSRP Chief Health Physicist King Faisal Specialist Hospital & Research Centre Riyadh, Saudi Arabia

More information

Radiation Units and Dosimetry 15 August Kalpana M. Kanal, Ph.D., DABR 1

Radiation Units and Dosimetry 15 August Kalpana M. Kanal, Ph.D., DABR 1 Introduction Radiation Units and Dosimetry Radiation dose quantities are used as indicators of the risk of biologic damage to patients from x-rays and thus a good knowledge of the different dose parameters

More information

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT

Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Correlation of Patient Weight and Cross-Sectional Dimensions with Subjective Image Quality at Standard Dose Abdominal CT Mannudeep K. Kalra, MD, DNB Michael M. Maher, MD, FRCR, FFR (RCSI) Srinivasa R.

More information

Radiation Dose Reduction Strategies in Coronary CT Angiography

Radiation Dose Reduction Strategies in Coronary CT Angiography Radiation Dose Reduction Strategies in Coronary CT Angiography Noor Diyana Osman, PhD noordiyana@usm.my Contents: Introduction Radiation dosimetry in CT Radiation risk associated with coronary CT angiography

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4 Patient Safety Concerns in Diagnostic Radiology? Lawrence T. Dauer, PhD, CHP Assistant Attending Health Physicist Department of Medical Physics RAMPS/GNYCHPS Spring Symposium April 30, 2010 Benefits?

More information

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Debra Pennington, MD Director of Imaging Dell Children s Medical Center Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)

More information

CT Radiation Dosimetry Study Using Monte Carlo Simulation and. Computational Anthropomorphic Phantoms

CT Radiation Dosimetry Study Using Monte Carlo Simulation and. Computational Anthropomorphic Phantoms CT Radiation Dosimetry Study Using Monte Carlo Simulation and Computational Anthropomorphic Phantoms by Yakun Zhang Graduate Program in Medical Physics Duke University Date: Approved: Ehsan Samei, Supervisor

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Patient Doses in Chest CT Examinations: Comparison of Various CT Scanners

Patient Doses in Chest CT Examinations: Comparison of Various CT Scanners SERBIAN JOURNAL OF ELECTRICAL ENGINEERING Vol. 10, No. 1, February 2013, 31-36 UDK: 621.386.8:616.24-073 DOI: 10.2298/SJEE1301031B Patient Doses in Chest CT Examinations: Comparison of Various CT Scanners

More information

ICRP Symposium on the International System of Radiological Protection

ICRP Symposium on the International System of Radiological Protection ICRP Symposium on the International System of Radiological Protection October 24-26, 2011 Bethesda, MD, USA M. Balonov, ICRP Committee 2, IRH, Russia P. Shrimpton, HPA, UK To be used just for regulatory

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Calculation of Normalised Organ and Effective Doses to Adult Reference Computational Phantoms from Contemporary Computed Tomography Scanners

Calculation of Normalised Organ and Effective Doses to Adult Reference Computational Phantoms from Contemporary Computed Tomography Scanners Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.165-171 (2011) ARTICLE Calculation of Normalised Organ and Effective Doses to Adult Reference Computational Phantoms from Contemporary Computed Tomography

More information

Estimation of the number of total and pediatric CT procedures based. on a nationwide survey in Japan

Estimation of the number of total and pediatric CT procedures based. on a nationwide survey in Japan Estimation of the number of total and pediatric CT procedures based on a nationwide survey in Japan Koji Ono 1, Nobuhiko Ban 2, Kai Michiaki 1. 1 Oita University of Nursing and Health Sciences 2944-9,

More information

Radiation Safety in the Catheterization Lab

Radiation Safety in the Catheterization Lab SCAI FALL FELLOWS COURSE - 2015 Radiation Safety in the Catheterization Lab V. Vivian Dimas, MD, FSCAI Associate Professor Pediatrics, Cardiology UT Southwestern Medical Center Dallas TX None Disclosures

More information

Re: PSM of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report

Re: PSM of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report Re: PSM-044-10 of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report Dear members of the National Quality Forum Board: As Professors of Radiologic Physics (at Mayo Clinic

More information