Diagnostic reference levels as a quality assurance tool

Size: px
Start display at page:

Download "Diagnostic reference levels as a quality assurance tool"

Transcription

1 Australian Institute of Radiography The Radiographer 2009; 56 (3): Literature review Diagnostic reference levels as a quality assurance tool KD Edmonds Medical Physics Section, Medical Radiation Branch, Australian Radiation Protection and Nuclear Safety Agency, Yallambie, Victoria 3085, Australia. Correspondence keith.edmonds@arpansa.gov.au Abstract The objective of diagnostic reference levels (DRL) in radiology is to assist in the optimisation of radiation dose to patients, while maintaining diagnostic image quality, and to detect unusually high doses that do not contribute significantly to the clinical outcome of a medical imaging examination. DRL have been in existence overseas for more than a decade and its influence has contributed to a steady decline in dose for general radiography and fluoroscopic procedures. High dose modalities such as CT and interventional procedures are increasing dramatically both locally and internationally resulting in the unwanted outcome of a significant increase in population cumulative effective dose. This calls for urgent dose reduction and dose constraint measures. Utilising DRL is one method of optimising patient dose. Some local and international DRL dose levels for some common radiographic, interventional and CT examinations are presented as a suggestion for the application of this methodology in Australian radiology practice. Keywords: diagnostic reference levels, guidance levels, radiation dose, reference values, quality assurance. Introduction The objective of establishing diagnostic reference levels (DRL) in diagnostic imaging (also previously known as Guidance Levels or Reference Values 1,2 ) is to provide radiology and other departments that use x-ray imaging with a convenient DRL dose comparison to ensure that radiation doses to patients are kept within reasonable limits. The main task of radiation protection is not only to minimise the stochastic risks but also to avoid deterministic injuries. Stochastic refers to effects whose probability increases with increasing dose and for which there is no threshold dose. Any dose, no matter how small, has the potential to cause harm and this becomes apparent years after the exposure. Examples are leukaemia and hereditary effects. Deterministic effects are those in which the severity of the effect, rather than the probability, increase with increasing dose and for which there is a threshold dose. Examples are epilation, erythema and hematologic damage and are known as early effects. 3 A DRL, as defined by the International Commission on Radiological Protection (ICRP), is a form of investigation level, applied to an easily measured quantity, usually the absorbed dose in air, or tissue-equivalent material at the surface of a simple phantom or a representative patient. 4 The ICRP recommends the establishment of reference levels as a method of optimising the radiation exposure to patients. This is accomplished by comparison between the numerical value of the diagnostic reference level (derived from relevant national, regional or local data) and the mean or other appropriate value observed in practice for a suitable reference group of patients or a suitable reference phantom. Another definition by the Council of the European Union in its Council Directive 97/43 defines DRL as dose levels in medical radiodiagnostic practices or, in the case of radio-pharmaceuticals, levels of activity, for typical examinations for groups of standardsized patients or standard phantoms for broadly defined types of equipment. These levels are expected not to be exceeded for standard procedures when good and normal practice regarding diagnostic and technical performance is applied. 36 It reinforces the concept of references doses applying only to standard or representative patients. DRL therefore are not dose limits but a guide of good practice. It is not a dose constraint and the DRL values are not used for regulatory or commercial purposes. DRL act as an investigation trigger if the numerical values are consistently exceeded. Background The need for DRL Patient exposures in diagnostic radiology are increasing at a disquieting rate for certain radiographic, fluoroscopic and CT examinations. Regulla and Elder 5 pointed out that data obtained from the United Nations Scientific Committee on the Effects of Atomic Radiation (UNCEAR) show that there are significant differences in national radiation exposures and a very uneven distribution of patient doses among world population for the same or similar procedures. Mean annual x-ray effective dose of the population can vary by up to a factor of 60. In the United States, studies such as the Nationwide Exposure X-ray Trends (NEXT) surveys also showed that patient doses in radiology vary considerably from one facility to the next. 6 Gray, et al. 1 posed the question, why one radiology facility should use an exposure that is 10, 20 or 126 times greater than another facility to produce a radiographic image? Johnston and Brennan 7 and Carroll and Brennan 8 also reported wide variations in patient doses for the same radiographic examinations among hospitals in the UK and Europe. These patient doses are attributed to a wide range of factors such as type of image receptor, exposure factors, fluoroscopic times, number of images, type of anti-scatter grid and level of quality control as reviewed by Seeram, 9 Bushong 10 and Parry, et al. 11 In Australia, the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) Code of Practice, Section 3.1.8

2 Diagnostic reference levels as a quality assurance tool The Radiographer 33 (Radiation Protection Series No.14) states that the Responsible Person must establish a program to ensure that radiation doses administered to a patient for diagnostic purposes are: 1 Periodically compared with DRL for diagnostic procedures for which DRL have been established in Australia 2 If DRL are consistently exceeded, reviewed to determine whether radiation has been optimised. 12 In addition, the ARPANSA Safety Guide, Section 7.8 (Radiation Protection Series No.14.1), suggests that as part of the QA program, patient dose surveys are undertaken periodically to establish that the doses are acceptable when compared with published DRL. It also recommends that accrediting bodies such as RANZCR and the Australian Council on Healthcare Standards consider including compliance with DRL for a core set of examinations. If the radiology department observes dose values consistently exceeding the DRL, then this warrants further investigation however some flexibility should be allowed if higher doses are indicated by sound clinical judgement. 13,14 However, at this point in time, there are no DRL published in the Code of Practice or the Safety Guides. It would seem logical therefore to use published values from the literature from extensive surveys in countries with similar healthcare settings e.g. similar levels of education and training for imaging technologists, radiologists and similar provision of imaging equipment. 15 History of DRL National surveys of patient doses from x-ray examinations in Europe and the USA since the 1950s have demonstrated wide variations in doses between radiology departments and illustrated the need for quantitative guidance on patient exposure. It was only at this stage that dose measurements to patients began in earnest. National surveys in the USA and UK concentrated on measuring entrance surface doses with or without backscatter for common radiographic projections. The Nationwide Evaluation of X-ray Trends in the USA in the 1970s measured entrance skin exposure free-in-air for average exposure technique factors or used a standard phantom. The NRPB national patient dose survey in the UK in the 1980s measured entrance surface dose directly on the surface of the patient (including backscatter) using thermoluminescence dosemeters. A European trial supporting the Quality Criteria for Diagnostic Radiographic Images in 1991 used the same technique. 37 Dose guidelines began to appear in late 1980s. First was the USA, promoted by the Centre for Devices and Radiological Health (CDRH) in conjunction with the Conference of Radiation Control Program Directors Inc. Then in the UK it was conducted by the National Radiation Protection Board in collaboration with relevant professional bodies. Europe then followed with reference doses incorporated into Working Documents by EC Study Groups. 38 International recommendations then appeared on how to measure and set reference dose levels based on the initiatives led by the USA and the UK. The ICRP Publication 60 first made mention of the concept of investigation levels in 1990 followed by the current definition of DRL in ICRP Publication 73 in and the EC Medical Exposure Directive in The United Kingdom introduced DRL in 1990 for common diagnostic examinations based on a national patient dose survey in the mid-1980s conducted by the NRPB, now known as the Health Protection Agency (HPA). They are now based on the five-yearly reviews of the National Patient Dose Database and are currently in their third review. 16 The International Atomic Energy Agency (IAEA) in 1996 and 2002 also issued advice on the use of DRL (or guidance levels) in their safety standard series and included guidance levels for typical adult patient doses for general radiography, CT, fluoroscopy and mammography. 2,14, Many countries worldwide have now incorporated the European Community Directive in national legislative documents. 1 Several organisations currently providing guideline documents for establishing DRL include the ICRP, 4,20 the Health Protection Agency in the UK, 16,21 the Commission of European Communities 22 and the American College of Radiology. 23 Aim The overall aim of DRL are to better manage patient dose in diagnostic radiology using the principle of optimisation which is defined as exposure to radiation from justified activities should be kept as low as reasonably achievable, social and economic factors being taken into account. The European Commission and the ICRP provide a range of tools to achieve this. 15,24 Data obtained from patient dose surveys show that typical patient doses for the same type of x-ray examinations can vary considerably from one radiology practice to another. The establishment of DRL therefore is to give an indication of unusually high values. The DRL are usually set at the third quartile value of the distribution of typical doses derived from dose surveys both nationally and internationally. Using the third quartile or 75th percentile is a compromise between being overly stringent and overly complacent. 3 Essentially, if mean doses exceed a reference level dose an investigation should take place to establish the cause and take corrective action, unless the dose was clinically justified. Reference doses were also used to provide a trigger for practices in need of investigation and hopefully lead to dose optimisation. ICRP 73 4 recommended that DRL values be selected by professional bodies, be reviewed at regular intervals and be specific to a country or region. Wide variations in patient doses are to be expected and it is only sensible to compare mean or median values, which is less influenced by extreme outliers, on representative groups of patients to monitor trends with time, equipment or technique. Method From a practical perspective, the DRL should be expressed as a readily measurable patient-related quantity for the specified procedure. For example, 1 General radiographic examinations either entrance skin dose (ESD) or the dose area product (DAP) 2 Fluoroscopic examinations dose area product (DAP) 3 CT examinations computed tomography dose index ( CTDI w or CTDI vol ) and the dose length product (DLP). New CT scanners in accordance with Australian Standards, AS/NZS , 25 should display the volume CTDI vol and/or the DLP on the operator s console after the selection of technique factors and prior to the initiation of x-rays. 13 DRL used for film-screen technology should not necessarily be used for new digital radiography without prior adjustment. 26 Dosimetry methods International guidance on patient dosimetry techniques for x-rays used in medical imaging is published by the International Commission on Radiation Units and Measurements in ICRU Report This report contains advice on the relevant dosimetric quantities and how to measure or calculate them in a clinical setting which is directly applicable to the patient dose surveys needed to estimate population exposure.

3 34 The Radiographer KD Edmonds There are various methods of recording dose levels. Thermoluminescent dosimeters (TLD) are often used for plain film examinations and include dose contribution from backscatter if placed on the patient or phantom surface. The small TLD sachets are usually placed in the centre of the irradiated field on the entrance surface of the patient or phantom. The TLD can be stuck directly and unobtrusively to the patient s skin with very little interference in patient mobility or comfort. They do not interfere with the examination or obscure important diagnostic information on the radiographic image. They need to be calibrated with respect to radiation qualities used in diagnostic radiology. TLD are also prone to some inaccuracy due to signal fade, nonlinear response and dependency on beam energy. 3 Ionisation chambers are bulky and more difficult to attach to patients. The parallel plate ionisation chambers measure back scatter but the lead backed solid state detectors do not. They are not recommended for direct measurement of entrance surface dose on the skin of the patient. They can, however, be used to make measurements of the absorbed dose to air, in free air, without a patient or phantom present. The measurements can then be corrected using appropriate backscatter factors and the inverse square law to estimate the entrance surface dose. Newer technology such as optically stimulated luminescence dosimeters (OSLs) and radiochromic film may replace TLD. Radiochromic film is currently being evaluated by ARPANSA. Alternatively, the entrance skin dose may also be calculated from x-ray tube output measurements (mgy/mas) and the exposure parameters, kvp, filtration and mas. The incident air kerma is calculated from the tube output using the inverse square law and then multiplied by the backscatter factor to obtain the entrance skin dose. A dose area product (DAP) meter consists of an ionisation meter that is usually attached to the x-ray tube collimator and measures the dose in Gy. square centimetre (Gy cm 2 ) which is proportional to the beam area and incident air kerma. The unit unfortunately does not measure backscatter which is important in higher dose examinations such as cardiac and vascular interventional procedures. However DAP meters can be used for radiographic and fluoroscopic procedures such as barium meals, angiography and on mobile image intensifiers. 2,3,28 For CT machines, the CTDI w and /or CTDI vol (mgy) and the DLP values(mgy cm) are conveniently provided at the operator console before or after the examination. The CTDI w is the weighted sum of the CT dose (or air kerma ) index measured in the centre and periphery (1 cm under the surface) of a 16 cm diameter (head) or a 32 cm diameter (trunk) standard polymethylmethacrylate (PMMA) CT dosimetry phantom. The CT dose index is measured with a 100mm long pencil ionization chamber inside a standard PMMA CT dosimetry phantom. CTDI w = CTDI c + CTDI p where c is the centre position 3 3 and p is the peripheral position of the phantom. Units: mgy CTDI w corrected for pitch is the CTDI vol. CTDI vol = 1 CTDI w pitch 2 Units: mgy The DLP is the product of CTDI vol and the scan length of the examination. Thus DLP = CTDI vol x length irradiated. Units: mgy cm 15 As a starting point it is suggested that the UK 2000 survey review of DRL for general radiography for adults (Table 1) and fluoroscopy for adults (Table 2), Paediatric procedures (Table 3), the UK 2003 CT survey (Table 4), and mammography (Table 5), be adopted and /or adapted in the Australian context as there are currently no established national DRL. 29 Some state regulators though have provided local DRL guidance on radiography, fluoroscopy and CT. 34 Dose values should be reviewed as computed/digital radiography becomes more widespread in order to minimise the detrimental influence of exposure creep. 30 This phenomenon occurs after the change over from film-screen radiography to digital radiography where exposure factors may actually increase in order to reduce image noise. Uncoupling of display from acquisition in digital radiography introduces the potential for systematic overexposure without necessarily compromising image quality. 31 The wide exposure latitude and linear response to x-ray energy provides an image appearance that remains consistent throughout the exposure range and this in turn provides little feedback to the technologist. Underexposed images typically have a grainy, mottled appearance that causes radiologists to reject images. Over-exposed images, on the other hand, have a crisp, sharp appearance. In order to prevent repeating the image, the technologist may increase exposure factors especially for manual and mobile radiography. Exposure indices or exposure indicators provided by the various CR/DR manufacturers also have a wide range of acceptable values and are currently not standardised throughout the industry. In the case of CT examinations, care should be taken when following overseas DRL because of the wide variety of CT scanners and local examination protocols employed. In addition, the rapid advances in CT technology have also resulted in constantly changing scanning protocols. Nevertheless, the DRL provided in Tables 1 5, serve as a rough guide until new DRL emerge from current surveys in Australia. In future, a web based interactive dose survey software program will be provided by ARPANSA where each radiology department can access it to calculate their dose levels and compare them with DRL. Discussion The development of DRL practice in diagnostic radiology within Australia is still at an early stage as no national surveys have been carried out for any radiological examinations for the express purpose of establishing national DRL. At a local level, various organisations, regulatory authorities and individual practices have carried out limited general radiography, fluoroscopy and CT surveys. 34 There is a clear need to manage (optimise) the radiation doses from diagnostic radiology in order to minimise the risks from radiation induced cancers. The establishment and use of DRL is recommended by international radiation protection organisations as an important component of the management of these doses and many countries have incorporated them into their radiation protection regulations 12,36 Data from European countries shows a wide variation in common DRL which may be due to differences in socio-economic conditions, regulatory regime, activeness of professional bodies and health care implementation (private/public mix etc). 7,32 International radiation protection bodies such as the IAEA and ICRP therefore recommend that each country carry out its own national wide scale DRL survey. It is for this reason that Australia must develop its own set of common national DRL. The introduction of computed and digital radiography in recent years has had a significant impact on the potential for higher dose

4 Diagnostic reference levels as a quality assurance tool The Radiographer 35 Table 1: Recommended diagnostic reference doses for individual radiographs on adult patients. Radiograph ESD per radiograph (mgy) DAP per radiograph (Gy cm 2 ) Skull AP/PA 3 - Skull LAT Chest PA Chest LAT 1 - Thoracic spine AP Thoracic spine LAT 10 - Lumbar spine AP Lumbar spine LAT 14 3 Lumbar spine LSJ 26 3 Abdomen AP 6 3 Pelvis AP 4 3 Adopted from the UK 2000 DRL survey review. 29 Note: Adult is defined as a person of average size (70 80 kg) ESD = Entrance Skin Dose, DAP = Dose Area Product. Table 2: Recommended diagnostic reference doses for fluoroscopic/interventional examinations on adult patients. Examination DAP per exam (Gy cm 2 ) Fluoroscopy time per exam (mins) Barium (or water soluble) swallow Barium meal Barium follow through Barium (or water soluble) enema Small bowel enema Biliary drainage/intervention Femoral angiogram 33 5 Hickman line Hysterosalpingogram 4 1 IVU 16 - MCU Nephrostogram Nephrostomy Retrograde pyelogram 13 3 Sialogram T-tube cholangiogram 10 2 Venogram (leg) Coronary angiogram Oesophageal dilation Pacemaker implant Adopted from the UK 2000 survey review. 29 DAP = Dose Area Product. delivery. 30,31 In addition, the exponential increase in CT examinations has lead to the unwanted outcome of a significant increase in population cumulative effective dose. 35 Other causal agents that are linked to high doses include type of image receptor, exposure factors, fluoroscopic time, number of images, type of antiscatter grid and level of quality control. 9,10,11 DRL depend significantly on local practice and equipment. They may also change with time as optimisation strategies become successful. The UK experience over the past 20 years has shown that the implementation of DRL together with a dose optimisation program has resulted in a gradual reduction of doses. 29,32 For DRL to succeed, acceptance and application of the concept is required across as many radiology departments as possible. Complex calculations will only discourage participation. Furthermore, setting DRL is a resource intensive activity and requires a national response. Priority should be given to procedures with greatest dose implications, i.e. CT and Interventional procedures. DRL should be owned by the professions such as the Australian Institute of Radiography and the Royal Australian New Zealand College of Radiology. ARPANSA will assist in facilitating their development. Conclusion Overseas experience has shown that the use of DRL have proven be a useful quality assurance tool in optimising patient dose in

5 36 The Radiographer KD Edmonds Table 3: Recommended diagnostic reference doses for complete examinations on paediatric patients. Examination Standard age (y) DAP per exam (Gy cm 2 ) MCU Barium meal Barium swallow Adopted from the UK 2000 survey review. 29 Table 4: Recommended diagnostic reference levels for CT examinations (CTDI vol and DLP). Patient group Scan region CTDI vol (mgy) single slice/multi slice DLP (mgy cm) Single slice/multi slice Adults Post fossa Cerebrum 65/100 55/65 760/930 Abdomen (liver metastases ) 13/14 460/470 Abdomen and pelvis (abscess) 13/14 510/560 Chest, abdomen and pelvis (lymphoma staging or follow up). 22/26 760/940 Chest (lung cancer) 10/13 430/580 Chest Hi-res 3/7 80/170 Children 0 1 year-old Head (post fossa) Head (cerebrum) Thorax (whole exam) year-old Head (post fossa) Head (cerebrum) Thorax (whole exam) year-old Head (post fossa) Head (cerebrum) Thorax (whole exam) 370 Adopted from UK 2003 CT dose survey. 33 Dose values for adults relate to the 16 cm diameter CT dosimetry phantom for examinations of the head and the 32 cm diameter CT dosimetry phantom for examinations of the trunk.all dose values for children relate to the 16 cm diameter CT dosimetry phantom. CTDI vol = Computed Tomography Dose Index Volume, DLP = Dose Length Product.

6 Diagnostic reference levels as a quality assurance tool The Radiographer 37 Table 5: Recommended diagnostic reference level for mammography for a typical adult patient. For film screen examinations using a grid, the mean glandular dose (MGD) is 2 mgy based on the 4.2 cm acrylic American College of Radiologists phantom. 17,18,19 For a 50% adipose, 50% glandular 5 cm thick phantom the MGD is 3 mgy 17 Note: For digital mammography, the values quoted above represent an upper limit diagnostic radiology. It is recommended that local dose surveys be performed annually while national surveys every five years. 32,33 The imaging technologist is the main person who decides on the exposure factors and the visual image quality for the radiologist to make a diagnosis. The technologist should therefore be aware of the exposure options that minimise radiation doses while still maintaining good image quality and monitoring dose levels. DRL would therefore serve as an important means of minimising radiation doses as well as dose variations at minimal cost to radiology departments. They also increase staff awareness and imaging technologists will be better equipped to deal with patient enquiries. 3 ARPANSA is responsible for carrying out national DRL surveys in consultation with relevant stakeholders such as the Royal Australian & New Zealand College of Radiology, The Australian Institute of Radiography, Australasian College of Physical Scientists & Engineers in Medicine, Australian & New Zealand Society of Nuclear Medicine, Department of Health and Aging and the various State/Territory Regulators. Acknowledgements The author thanks Anthony Wallace for his advice in preparing this article. The author KD Edmonds DCR BHA Grad Dip Pub Health References 1 Gray JE, Archer BR, Butler PF, Hobbs BB, Mettler FA, Pizzutiello RJ, et al. Reference values for diagnostic radiology: application and impact. Radiology 2005; 235: IAEA 2002.Radiological Protection for Medical Exposure to Ionizing Radiation. IAEA Safety Standards Series. Safety Guide No RS-G-1.5. Vienna, Austria: International Atomic Energy Agency. 3 Seeram,E, Brennan P C, Diagnostic reference levels in radiology. Radiologic Technology 2006: 77: ICRP. Radiological Protection and Safety in Medicine. ICRP Publication 73, Annals of the ICRP 1996; 26 (2). 5 Regulla DF, Eder H. Patient exposures in medical x-ray imaging in Europe. Radiat Prot Dosimet 2005; 14: Conference of Radiation Control Program Directors. Nationwide Evaluation of X-ray Trends (NEXT): Tabulation and Graphical Summary of 1995 Abdomen and Lumbosacral Spine Survey CRCPD Publication Johnston DA, Brennan PC. Reference dose levels for patients undergoing common diagnostic x-ray examinations in Irish hospitals. Br J Radiol 2000: 73: Carroll EM, Brennan PC. Radiation doses for barium enema and barium meal examinations in Ireland: potential diagnostic reference levels. Br J Radiol 2003; 76: Seeram E. Rad Tech s Guide to Radiographic Protection. Boston: Blackwell Science; Bushong S. Radiologic Science for Technologists 8th ed. St Louis: Mosby; Parry RA, Glaze SA, Archer BJ. Typical patient radiation doses in diagnostic radiology. Radiographics 1999; 19: ARPANSA. Code of Practice for Radiation Protection in the Medical Applications of Ionizing Radiation, Radiation Protection Series 2008 RPS ARPANSA. Safety Guide for Radiation Protection in Diagnostic and Interventional Radiology Radiation Protection Series 2008 RPS IAEA International Atomic Energy Agency, International Basic safety standards for protection against ionizing radiation and for the safety of radiation sources. Safety Series No. 115, IAEA, Vienna. 15 European Commission. European guidance on estimating population doses from medical x-ray procedures Radiation Protection Publication No 154. TREN/04/NUCL/S National Radiological Protection Board. Guidelines on patient dose to promote the optimisation for diagnostic medical exposures. Documents of the NRPB 1999; 10 (1). 17 Craig AR, Heggie JCP, McLean ID, Coakley KS, Nicoll JJ. Recommendations for a mammography quality assurance program ACPSEM Position paper. Australas Phys Eng Sci Med 2001; 24 (107): Mammographic Quality Control Manual. Royal Australian and New Zealand College of Radiologists National Accreditation Standards. Breastscreen Australia 2002; pp International Commission on Radiological Protection recommendations of the International Commission on Radiological Protection. ICRP Publication No. 60. Oxford: Pergamon; HPA National protocol for patient dose measurements in diagnostic radiology. Dosimetry Working Party of the Institute of Physical Sciences in Medicine. Health Protection Agency.Chilton.Didcot. 22 Commission of European Communities. European Guidelines on Quality Criteria For Diagnostic Radiographic Images. The European Commission ACR Technical Standard for Diagnostic Medical Physics Performance Monitoring of Radiographic and Fluoroscopic Equipment. Revised. Reston: American College of Radiology; 2001: International Commission on Radiological Protection. Diagnostic reference levels in medical imaging: review and additional advice. Available online at: [verified 21st March]. 25 AS/NZS Standards Australia and Standards New Zealand. Medical electrical equipment Particular requirements for safety-x-ray equipment for computed tomography, AS/NZS Ed ICRP. Managing patient dose in digital radiography. ICRP Publication 93 Annals of the ICRP 2004; 34 (1). 27 ICRU. Patient Dosimetry for x-rays used in medical imaging. ICRU Report 74. Journal of the ICRU 2005; 5 (2). 28 IPSM (1992). Dosimetry Working Party of the Institute of Physical Sciences in Medicine. National protocol for patient dose measurements in diagnostic radiology. Chilton, UK: NRPB. 29 Hart D, Hillier MC, Wall BF. Doses to patients from medical x-ray examinations in the UK-2000 review. Chilton: NRPB 2002; NRPB-W Warren-Forward H, Arthur L, Hobson L, et al. An assessment of exposure indices in computed radiography for the posterior-anterior chest and the lateral lumbar spine. Br J Radiol 2007; 80: Willis C. Strategies for dose reduction in ordinary radiographic examinations using CR and DR. Pediatr Radiol 2004; 34 (Suppl 3): S196 S Hart D, Hillier MC,Wall BF. National reference doses for common radiographic, fluoroscopic and dental x-ray examinations in the UK. Br J Radiol 2009: 82: Shrimpton PC,Hillier MC,Lewis MA,Dunn M. National survey of doses from CT in the UK: Br J Radiol 2006: 79: Boal T, Hedt J, Einsiedel P. A survey of patient dose and image quality factors for CT scanners in Victoria. Aust Phys Engineer Sci Med 1999; 22 (3): Brenner DJ, Hall EJ. Computed tomography an increasing source of radiation exposure. New Engl J Med 2007; 357 (22): Office for Official Publications of European Communities, Luxemburg, European Commission, Council Directive 97/43/Euroatom of June 1997 on the health protection of individuals against the dangers of ionizing radiation in relation to medical exposure. Official Journal of the European Communities L180/ Wall BF, Shrimpton PC. The Historical development of reference doses in diagnostic radiology. Radiat Prot Dosimet 1998; 80 (1 3): EC Study Group. Quality Criteria for Diagnostic Radiographic Images. Working Document, 2nd ed, CEC XII/173/90 (Brussels: EC) 1990.

A Standardised Approach to Optimisation

A Standardised Approach to Optimisation The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained

More information

Radiation Protection Dosimetry Advance Access published April 22, 2008

Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 5 doi:10.1093/rpd/ncn144 EVALUATION OF EXPOSURE PARAMETERS IN PLAIN RADIOGRAPHY: A COMPARATIVE

More information

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,

More information

JMSCR Vol 06 Issue 12 Page December 2018

JMSCR Vol 06 Issue 12 Page December 2018 www.jmscr.igmpublication.org Impact Factor (SJIF): 6.379 Index Copernicus Value: 79.54 ISSN (e)-2347-176x ISSN (p) 2455-0450 DOI: https://dx.doi.org/10.18535/jmscr/v6i12.80 Diagnostic Reference Levels:

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Diagnostic Reference Levels (DRLs)

Diagnostic Reference Levels (DRLs) Section 7 Diagnostic Reference Levels (DRLs) Establishment and Use of Diagnostic Reference Levels (DRLs) The National Radiation Safety Committee (NRSC), established in November 2007, is a statutory committee

More information

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran Soraya Emamgholizadeh Minaei 1, Fariba Firouzi 1 2, 3,*, Hamid Reza Khosravi 1 Department of radiology, Shahid

More information

Patient dose in routine X-ray examinations in Yazd state

Patient dose in routine X-ray examinations in Yazd state Iran. J. Radiat. Res., 2004; 1(4): 199-204 Patient dose in routine X-ray examinations in Yazd state F. Bouzarjomehri Health Physics Dept. Shahid Sadoghi Univ. of Medical Sciences, Yazd, Iran ABSTRACT Background:

More information

Assessing the standard dose to standard patients for x-ray investigations

Assessing the standard dose to standard patients for x-ray investigations Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review Receptor dose and Patient Dose in Radiographic exposures - a 15 year review D J Peet*, N Tyler, M Pryor, P Hollaway, C Strudley and L Leavesley Regional Radiation Protection Service, Royal Surrey County

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation exposure of the Yazd population from medical conventional X-ray examinations Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics

More information

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural

More information

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES IJRRAS 9 (1) October 11 www.arpapress.com/volumes/vol9issue1/ijrras_9_1_16.pdf THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES Bamidele

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

European Guidelines on DRLs for Paediatric Imaging

European Guidelines on DRLs for Paediatric Imaging European Guidelines on DRLs for Paediatric Imaging Final complete draft 1 1 1 1 1 1 1 1 0 1 0 1 Contents PREFACE... EXECUTIVE SUMMARY... 1. Background.... Introduction.... Purpose and scope.... Definitions...

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Debra Pennington, MD Director of Imaging Dell Children s Medical Center Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)

More information

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3 Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,

More information

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran Global Journal of Health Science; Vol. 7, No. 5; 2015 ISSN 1916-9736 E-ISSN 1916-9744 Published by Canadian Center of Science and Education Measurement of Entrance Skin Dose and Calculation of Effective

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

Trends in examination frequency and population doses in Norway,

Trends in examination frequency and population doses in Norway, Workshop on European Populaltion Doses from Medical Exposure, 24-26 April 2012, Athens, Greece Trends in examination frequency and population doses in Norway, 2002-2008 OLERUD H.M. 1,2, FRIBERG E.G. 1,

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Doses Received by Patients during Thorax X-Ray Examinations

Doses Received by Patients during Thorax X-Ray Examinations Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 01, 45-51 Received: September 16, 01; Accepted: February 0, 013 Original Article Doses Received by Patients during Thorax X-Ray Examinations Nsikan

More information

ICRP Symposium on the International System of Radiological Protection

ICRP Symposium on the International System of Radiological Protection ICRP Symposium on the International System of Radiological Protection October 24-26, 2011 Bethesda, MD, USA M. Balonov, ICRP Committee 2, IRH, Russia P. Shrimpton, HPA, UK To be used just for regulatory

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

Alfred Health's CT quality control program

Alfred Health's CT quality control program Alfred Health's CT quality control program Poster No.: R-0037 Congress: Type: Authors: 2014 CSM Scientific Exhibit A. Perdomo, Z. Brady, N. Tran, L. Hudson, K. Provis; PRAHRAN/ AU Keywords: Quality assurance,

More information

EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2

EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2 Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 7 doi:10.1093/rpd/ncn143 EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY

More information

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY PNRA-RG-904.05 (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY For Further Details Directorate of Policies & Procedures PAKISTAN

More information

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY March 2010 EAST AFRICAN MEDICAL JOURNAL 127 East African Medical Journal Vol. 87 No. 3 March 2010 OPTIMIsATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY G. K. Korir, BSc, MSc, PhD

More information

The impact of SID and collimation on backscatter in radiography

The impact of SID and collimation on backscatter in radiography The impact of SID and collimation on backscatter in radiography Poster No.: C-2931 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: M. Joyce 1, J. T. Ryan 2, M. F. Mc Entee

More information

Trends in Patients Exposure Doses during Radiographic Examination in Japan

Trends in Patients Exposure Doses during Radiographic Examination in Japan Trends in Patients Exposure Doses during Radiographic Examination in Japan S. Suzuki 1*, Y. Asada 1, M. Kobayashi 2, K. Kobayashi 2, Y. Matsunaga 3, T. Otsuka 4 1 School of Health Sciences, Fujita Health

More information

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Caspian Journal of Health Research 2016; 2(2):18-27 Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Hossein Rahanjam 1, Hossein

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT Ask EuroSafe Imaging Tips & Tricks CT Working Group On the use of Diagnostic Reference Levels in CT Robert Bujila (Karolinska University Hospital, SE) Eileen Kelly (Galway University Hospitals, IE) Matthias

More information

National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations

National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations Eva Godske Friberg *, Anders Widmark and Ingrid Helen Ryste Hauge Norwegian Radiation

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

Shielding Calculation: Radiographic Room. Jerry Williams

Shielding Calculation: Radiographic Room. Jerry Williams Shielding Calculation: Radiographic Room Jerry Williams What you need to know Room use and layout DAP workload DAP averaged kv Distance to barrier Construction details Walls Ceilings/ floors Surrounding

More information

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Conventional and spiral CT dose indices in Yazd general hospitals, Iran Iran. J. Radiat. Res., 2006; 3 (4): 183-189 Conventional and spiral CT dose indices in Yazd general hospitals, Iran F. Bouzarjomehri 1*,M.H.Zare 2, D. Shahbazi 2 1 Department of Medical Physics, Shahid

More information

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Omer Osman,Saeed Medical Physics Department, ALNeelain University, Sudan Presentation outlines Introduction Objectives Materials and

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

Establishing the Greek National Reference Levels for Interventional Cardiology procedures

Establishing the Greek National Reference Levels for Interventional Cardiology procedures Establishing the Greek National Reference Levels for Interventional Cardiology procedures George Simantirakis*, Christina Koukorava, Maria Kalathaki, Christos Pa

More information

Quality Control and Patient Dosimetry on line for Computed Tomography

Quality Control and Patient Dosimetry on line for Computed Tomography Quality Control and Patient Dosimetry on line for Computed Tomography Jose I. Ten 1,2, Eliseo Vano 2,3, Jose M. Fernandez-Soto 2,3, Roberto Sanchez 3, Juan Arrazola 1,2 1 Diagnostic Radiology Service and

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements Standards Revisions for Organizations Providing Fluoroscopy Services The Joint Commission has approved the following revisions for prepublication. While revised requirements

More information

Dosimetry in digital mammography

Dosimetry in digital mammography Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol

More information

Australian Per Caput Doses from Diagnostic Imaging and Nuclear Medicine

Australian Per Caput Doses from Diagnostic Imaging and Nuclear Medicine Australian Per Caput Doses from Diagnostic Imaging and Nuclear Medicine A.J.M Hayton, P.N Johnston, J Baldas, P.A Marks, K Edmonds, A.B Wallace Australian Radiation Protection and Nuclear Safety Agency,

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

Doses to Patients arising from Dental X-ray Examinations in the UK, A Review of Dental X-ray Protection Service Data

Doses to Patients arising from Dental X-ray Examinations in the UK, A Review of Dental X-ray Protection Service Data HPA-RPD-22 Doses to Patients arising from Dental X-ray Examinations in the UK, 22 24 A Review of Dental X-ray Protection Service Data A D Gulson, T A Knapp and P G Ramsden ABSTRACT The Dental X-ray Protection

More information

International Journal of Radiation Research, January 2017

International Journal of Radiation Research, January 2017 Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing

More information

Why radiation protection matters?

Why radiation protection matters? Why radiation protection matters? Elias Brountzos Professor of Radiology 2 nd Department of Radiology Medical School, University of Athens Athens, Greece A definition for radiation protection Radiation

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

IMAGE GENTLY HOW CAN YOU HELP?

IMAGE GENTLY HOW CAN YOU HELP? IMAGE GENTLY HOW CAN YOU HELP? Keith J. Strauss, MSc, FAAPM, FACR Director, Radiology Physics & Engineering Children s s Hospital Boston Harvard Medical School Acknowledgment Marilyn J. Goske,, MD Robert

More information

The use of gonadal shielding in singular common diagnostic radiographic procedures

The use of gonadal shielding in singular common diagnostic radiographic procedures Australian Institute of Radiography Student paper The Radiographer 2007; 54 (2): 13 17 The use of gonadal shielding in singular common diagnostic radiographic procedures Sean A. Bowen, Ryan Clarke, Samuel

More information

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY Journal of Bangladesh Academy of Sciences, Vol. 3, No. 1, 7-17, 211 A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY MAHFUZA BEGUM 1, A. S. MOLLAH

More information

ABSTRACT I. INTRODUCTION

ABSTRACT I. INTRODUCTION 2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography

More information

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Establishment of local diagnostic reference levels for quality control in intraoral radiography Oral Radiol (27) 33:38 44 DOI.7/s282-6-24-9 ORIGINAL ARTICLE Establishment of local diagnostic reference levels for quality control in intraoral radiography Maki Izawa Yasuo Harata Noriyoshi Shiba Nobuhide

More information

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Assessment of Patient Dose from CT Examinations in Khorasan, Iran Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 2012, 233-238 Received: July 18, 2012; Accepted: November 17, 2012 Original Article Assessment of Patient Dose from CT Examinations in Khorasan,

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria

Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria Bidemi I.

More information

THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS

THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS Larysa Stadnyk, Pylypenko M., Shalepa O., Nosyk O. Grigorev Institute for Medical Radiology,

More information

Research Article. Corresponding author H. Osman

Research Article. Corresponding author H. Osman Scholars Journal of Applied Medical Sciences (SJAMS) ISSN 2320-6691 Sch. J. App. Med. Sci., 2013; 1(5):511-515 Scholars Academic and Scientific Publisher (An International Publisher for Academic and Scientific

More information

Radiation Protection Dosimetry Advance Access published September 13, 2006

Radiation Protection Dosimetry Advance Access published September 13, 2006 Radiation Protection Dosimetry Advance Access published September 13, 26 Radiation Protection Dosimetry (26), 1 of 6 doi:1.193/rpd/ncl113 TECHNICAL NOTE RADIATION EXPOSURE AND DOSE EVALUATION IN INTRAORAL

More information

Local reference dose evaluation in conventional radiography examinations in Iran

Local reference dose evaluation in conventional radiography examinations in Iran JOURNAL OF PLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 Local reference dose evaluation in conventional radiography examinations in Iran M. Shirin Shandiz, 1,2 M.T. Bahreyni Toossi, 3a S.

More information

People Exposed to More Radiation from Medical Exams

People Exposed to More Radiation from Medical Exams People Exposed to More Radiation from Medical Exams With its release of a new report, titled Ionizing Radiation Exposure of the Population of the United States (Report No. 160, 2009), the National Council

More information

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Rebecca Milman Marsh, Ph.D. University of Colorado Department of Radiology Who in the Facility Works

More information

ROMANIA MEDICAL EXPOSURE IN 2016 IN ROMANIAN RADIOLOGICAL DEPARTMENTS. Olga GIRJOABA and Alexandra CUCU

ROMANIA MEDICAL EXPOSURE IN 2016 IN ROMANIAN RADIOLOGICAL DEPARTMENTS. Olga GIRJOABA and Alexandra CUCU ROMANIA MEDICAL EXPOSURE IN 2016 IN ROMANIAN RADIOLOGICAL DEPARTMENTS Olga GIRJOABA and Alexandra CUCU National Institute of Public Health, Bucharest, Romania olga.garjoaba@insp.gov.ro PURPOSE Improvement

More information

Managing Patient Dose in Computed Tomography (CT)

Managing Patient Dose in Computed Tomography (CT) Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result. Minimizing Dose Understand the importance and methods of pre-procedure patient assessment including a review of previous radiologic exams, disease processes and anatomical considerations that may increase

More information

ACR AAPM PRACTICE GUIDELINE FOR DIAGNOSTIC REFERENCE LEVELS AND ACHIEVABLE DOSES IN MEDICAL X-RAY IMAGING

ACR AAPM PRACTICE GUIDELINE FOR DIAGNOSTIC REFERENCE LEVELS AND ACHIEVABLE DOSES IN MEDICAL X-RAY IMAGING BE IT RESOLVED, Sponsored By: RESOLUTION NO. 47 that the American College of Radiology adopt the ACR AAPM Practice Guideline for Diagnostic and Achievable Doses in Medical X-Ray Imaging ACR Council Steering

More information

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information

Revised: 8/05; 9/08; 9/09; 8/11; 8/12; 1/13 Reviewed: 3/10

Revised: 8/05; 9/08; 9/09; 8/11; 8/12; 1/13 Reviewed: 3/10 DCH RADIOGRAPHY PROGRAM CURRICULUM 17 Hours of Pre-requisites are required before entering into the professional phase of the Radiography Program. Pre-requisites English Comp I Intermediate College Algebra

More information

Radiation Safety in the Catheterization Lab

Radiation Safety in the Catheterization Lab SCAI FALL FELLOWS COURSE - 2015 Radiation Safety in the Catheterization Lab V. Vivian Dimas, MD, FSCAI Associate Professor Pediatrics, Cardiology UT Southwestern Medical Center Dallas TX None Disclosures

More information

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES A THESIS SUBMITTED TO THE GRADUATE FACULTY in partial fulfillment of the requirements

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

3 rd International Symposium on the System of Radiological Protection Seoul, October John Harrison

3 rd International Symposium on the System of Radiological Protection Seoul, October John Harrison 3 rd International Symposium on the System of Radiological Protection Seoul, October 2015 John Harrison UK Task Group 79 : Use of Effective Dose as a Risk-related Radiological Protection Quantity John

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

Patient doses in general radiography examinations

Patient doses in general radiography examinations peer reviewed ORIGINAL ARTICLE Patient doses in general radiography examinations Wambani JS 1 MBChB, M. Med (Rad), Paediatric (Rad) Onditi EG 2 MBChB, M.Med (Rad) Korir GK 3 B.Sc, MSc (Nuclear Science),

More information

A multicentric study on patient dose in multislice CT

A multicentric study on patient dose in multislice CT A multicentric study on patient dose in multislice CT A.Stratis 1, M.Molfetas 1, S.Kottou 2, A.Louizi 2 1. Medical Physics department, Evangelismos General hospital of Athens, Athens, Greece 2. Medical

More information

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua Radiation Safety For Anesthesiologists R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua Modern World Non Ionizing VS Ionizing Non Ionizing Harmless Ex. visible

More information

Page 1 of 5 Patient Safety: Radiation Dose in X-Ray and CT Exams What are x-rays and what do they do? X-rays are forms of radiant energy, like light or radio waves. Unlike light, x-rays can penetrate the

More information

Estimation of the number of total and pediatric CT procedures based. on a nationwide survey in Japan

Estimation of the number of total and pediatric CT procedures based. on a nationwide survey in Japan Estimation of the number of total and pediatric CT procedures based on a nationwide survey in Japan Koji Ono 1, Nobuhiko Ban 2, Kai Michiaki 1. 1 Oita University of Nursing and Health Sciences 2944-9,

More information

Facing the challenges: IAEA activities to protect the patients

Facing the challenges: IAEA activities to protect the patients Facing the challenges: IAEA activities to protect the patients Debbie Gilley Radiation Protection of Patients Radiation Safety and Monitoring Section Division of Radiation, Transport and Waste Safety 18

More information

Radiation Doses in Radiology: Influence of Standards and Regulations

Radiation Doses in Radiology: Influence of Standards and Regulations Radiation Doses in Radiology: Influence of Standards and Regulations Beebe Symposium National Academy of Sciences December 9, 2009 Washington D.C. Orhan H Suleiman MS PhD, FAAPM Senior Science Policy Adviser

More information

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit References References The accuracy, quality and relevance of these works are not guaranteed or uniform and more recent information may have superseded these works. This list is not exhaustive. It does

More information

Diagnostic reference levels in intraoral dental radiography in Korea

Diagnostic reference levels in intraoral dental radiography in Korea Imaging Science in Dentistry 2012; 42 : 237-42 http://dx.doi.org/10.5624/isd.2012.42.4.237 Diagnostic reference levels in intraoral dental radiography in Korea Eun-Kyung Kim, Won-Jeong Han, Jin-Woo Choi,

More information

Practice and Risk at Medical Facilities in Agency Operations

Practice and Risk at Medical Facilities in Agency Operations Practice and Risk at Medical Facilities in Agency Operations Igor Gusev Radiation Protection Unit IAEA International Atomic Energy Agency Outline What is medical radiation exposure? Radiation sources and

More information

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4 Patient Safety Concerns in Diagnostic Radiology? Lawrence T. Dauer, PhD, CHP Assistant Attending Health Physicist Department of Medical Physics RAMPS/GNYCHPS Spring Symposium April 30, 2010 Benefits?

More information

The Fluoroscopic Technique for Monitoring Distraction of a Non-Invasive Lengthening Device in Early Onset Scoliosis

The Fluoroscopic Technique for Monitoring Distraction of a Non-Invasive Lengthening Device in Early Onset Scoliosis Send Orders for Reprints to reprints@benthamscience.net Open Medicine Journal, 2014, 1, 23-28 23 Open Access The Fluoroscopic Technique for Monitoring Distraction of a Non-Invasive Lengthening Device in

More information

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy What is radiation? Radiation is energy emitted from a source, that travels through space and can penetrate matter. Listed below are two types that we are exposed to and contribute to our overall radiation

More information