Field-size correction factors of a radiophotoluminescent glass dosimeter for small-field and intensity-modulated radiation therapy beams

Size: px
Start display at page:

Download "Field-size correction factors of a radiophotoluminescent glass dosimeter for small-field and intensity-modulated radiation therapy beams"

Transcription

1 Field-size correction factors of a radiophotoluminescent glass dosimeter for small-field and intensity-modulated radiation therapy beams Shimpei Hashimoto a) Department of Radiation Oncology, Tokyo Metropolitan Cancer and Infectious Diseases Center Komagome Hospital, Honkomagome, Bunkyo-ku, Tokyo, Japan Yukio Fujita Department of Radiation Oncology, Tokai University School of Medicine, 143 Shimokasuya, Isehara-shi, Kanagawa, Japan Tetsurou Katayose Department of Radiation Oncology, Chiba Cancer Center, Nitona-cho Chuo-ku, Chiba-shi, Chiba, Japan Hideyuki Mizuno Department of Radiation Measurement and Dose Assessment, National Institute of Radiological Sciences, National Institutes for Quantum and Radiological Science and Technology, Anagawa, Inage-ku, Chiba-shi, Chiba, Japan Hidetoshi Saitoh Graduate School of Human Health Sciences, Tokyo Metropolitan University, Higashiogu, Arakawa-ku, Tokyo, Japan Katsuyuki Karasawa Department of Radiation Oncology, Tokyo Metropolitan Cancer and Infectious Diseases Center Komagome Hospital, Honkomagome, Bunkyo-ku, Tokyo, Japan (Received 27 March 2017; revised 19 October 2017; accepted for publication 24 October 2017; published 5 December 2017) Purpose: We evaluated the energy responses of a radiophotoluminescent glass dosimeter (RPLD) to variations in small-field and intensity-modulated radiation therapy (IMRT) conditions using experimental measurements and Monte Carlo simulation. Methods: Several sizes of the jaw and multileaf collimator fields and various plan-class IMRT-beam measurements were performed using the RPLD and an ionization chamber. The field-size correction factor for the RPLD was determined for 6- and 10-MV x rays. This correction factor, together with the perturbation factor, was also calculated using Monte Carlo simulation with the EGSnrc/ egs_chamber user code. In addition, to evaluate the response of the RPLD to clinical-class-specific reference fields, the field-size correction factor for the clinical IMRT plan was measured. Results: The calculated field-size correction factor ranged from to (for 6-MV x rays) and from to (for 10-MV x rays) as the jaw-field size ranged from 1 9 1cm 2 to cm 2. The atomic composition perturbation factor for these jaw fields decreased by 3.2% and 1.9% for the 6- and 10-MV fields, respectively. The density perturbation factor was unity for field sizes ranging from 3 9 3cm 2 to cm 2, whereas that for field sizes ranging from 3 9 3cm 2 to 1 9 1cm 2 decreased by 3.2% (for 6-MV x rays) and 4.3% (for 10-MV x rays). The volume-averaging factor rapidly increased for field sizes below cm 2. The results for the MLC fields were similar to those for the jaw fields. For plan-class IMRT beams, the field-size correction and perturbation factors were almost unity. The difference between the doses measured using the RPLD and ionization chamber was within 1.2% for the clinical IMRT plan at the planning-target volume (PTV) region. Conclusions: For small fields of size cm 2 or less, it was clarified that the volume averaging and density perturbation were the dominant effects responsible for the variation in the RPLD response. Moreover, perturbation correction is required when measuring a field size cm 2 or less. Under the IMRT conditions, the difference in the responses of the RPLD between the reference conditions and the PTV region calculated by Monte Carlo simulation did not exceed 0.8%. These results indicate that it is feasible to measure IMRT dosage using an RPLD at the PTV region American Association of Physicists in Medicine [ Key words: energy response, intensity-modulated radiation therap, radiophotoluminescent glass dosimeter, small field 1. INTRODUCTION The accuracy of the output dose is one of the most important factors in radiotherapy. Recently, several groups have performed dosimetry audits for intensity-modulated radiation therapy (IMRT) using a thermoluminescent dosimeter (TLD) and an alanine dosimeter. 1 4 Radiophotoluminescent glass dosimeters (RPLDs) have been used 382 Med. Phys. 45 (1), January /2018/45(1)/382/ American Association of Physicists in Medicine 382

2 383 Hashimoto et al.: Field-size correction factors of RGD for small-field 383 for the in vivo dosimetry 5,6 and postal dosimetry audits of megavoltage radiotherapy photon beams 7,8 because of their small fading effects, precise readings owing to repeatable readout, and ease of handling. 9 Therefore, because of the above properties, RPLDs can be used in patient-specific measurements, in vivo dosimetry, and postal dosimetry audits for IMRT. The absorbed dose to water measured in the nonreference field D f nr was defined as the following equation, which is the modified equation of Alfonso et al.: 10 D f nr ¼ M f nr N D;w;Q0 k Q;Q0 ;Q ; (1) where M f nr is the reading of the dosimeter in the nonreference field f nr. N D;w;Q0 is the calibration coefficient in terms of the absorbed dose to water for an ionization chamber at a reference beam quality Q 0 (usually 60 Co). N D;w;Q0 is measured at the standards laboratory for a reference field of size cm 2. k Q;Q0 is the beam-quality correction factor, which corrects the differences between the reference beam quality Q 0 at the standards laboratory and the beam quality Q of the reference field f ref. ;Q is a field-size correction factor that corrects the differences between the conditions of field size of f ref and f nr. Several studies reported the energy and field-size dependence for RPLD because of changes in the photonenergy spectrum. 7,9,11 13 Mizuno et al. reported a correction method for the energy and field-size dependence of postaldose audits for reference and nonreference conditions (e.g., several field-size and wedge-beam conditions) of conventional linac photon beams. 13 IMRT comprises a smaller field, together with on- and off-axis multileaf collimator (MLC) fields. Therefore, the responses of the RPLD to the IMRT beams may vary. Moreover, in small-photon-field measurements, detector response is largely determined by the mass densities of detector materials, whose influence on the small-to-large-field ratio of detector response is greater than that of atomic number. 14,15 In addition, a nonuniform dose distribution over the detection volume leads to a volume-averaging effect in a small field. The mass density of the RPLD is 2.61 g/cm 3, and the active volume is 1.5 mm in diameter and 6.0 mm in length. Therefore, the RPLD may be affected by density perturbation and volume averaging in the IMRT measurements. Although Azangwe et al. evaluated the RPLD response for field sizes ranging from cm 2 to cm 2 field, 16 a considerably smaller field and an IMRT beam have not been previously evaluated. Therefore, to evaluate the reliability of the RPLD for the small-field and IMRT-beam measurements, we evaluated the energy response and perturbation factor of the RPLD for the small-field, MLC field, and IMRT conditions using experimental measurements and Monte Carlo simulation. 2. MATERIAL AND METHODS 2.A. Measurement of field-size correction factor To evaluate the RPLD response to variations in the small field and IMRT beam, a field-size correction factor was determined. An RPLD (GD-302M, Asahi Techno Glass, Shizuoka, Japan) and an ionization chamber (A1SL, Standard Imaging, Middleton, WI, USA) were positioned at the center of a cm 3 water-equivalent phantom (WE211, Kyotokagaku, Kyoto, Japan), and the RPLD was oriented perpendicular to the beam axis (Fig. 1). The source-to-detector distance was 100 cm. Irradiated-field conditions included jaw fields, MLC fields, off-axis MLC fields (Fig. 2), simple- IMRT fields, and two plan-class-specific reference fields. The jaw fields were defined using the X Y collimator jaws alone. The collimator jaws defined square fields with side lengths of 1, 1.6, 2, 3, 4, 5, 10, and 20 cm. In the clinical FIG. 1. Diagram of the measurement setup and detector orientation. [Color figure can be viewed at wileyonlinelibrary.com]

3 384 Hashimoto et al.: Field-size correction factors of RGD for small-field 384 FIG. 2. Diagram of the multileaf collimator (MLC) position of the cm 2 field. The cross indicates the isocenter, whereas the circle indicates the field center. 2-cm off-axis and 4-cm off-axis indicate that the distance between the isocenter and the field center is 2 and 4 cm, respectively. [Color figure can be viewed at wileyonlinelibrary.com] situation, an IMRT beamlet is formed using the MLC, and the slit shape is often used for the IMRT beamlet. Therefore, the MLC field was investigated. The MLC fields used MLC alone, with the collimator jaws set to cm 2. The MLC was used to define 1, 2, 4, 6, and cm 2 fields. A simple-imrt field (s-imrt) was created to duplicate the simulated field-five uniform beamlets of a cm 2 field. The two plan-class-specific reference fields used the AAPM TG119 prostate (PR) and head-and-neck (HN) plans 17 with seven and nine fields, respectively. In these plans, the measurement points (isocenter) and the center of the planning-target volume (PTV) were aligned. Under each condition, irradiation was performed five times and one RPLD was used with each irradiation. In addition, a specially made plug which is made of WE211 was employed to hold the RPLD. Irradiation was performed using a Clinac 21EX for 6-MV x ray and TrueBeam (Varian Medical Systems, Palo Alto, CA, USA) for 10-MV x ray. The RPL of the RPLD was read five times using an RPL reader (FDG-1000, Asahi Techno Glass), and the outputs were averaged. The field-size correction factor ;Q can be expressed as follows: ;Q ¼ Dfnr D f ref =M f nr w;q ref =M f ref Q ref ¼ D f nr =RPL f nr D f ref w;q ref =RPL f ; (2) ref Q ref where RPL is a radiophotoluminescence reading. The absorbed dose to water was determined using the ionization chamber according to the Japanese standard dosimetry code, 18 which is similar to IAEA TRS The reference condition was 10 cm depth in water and cm 2 jaw field. While measuring a small field, the response of the A1SL chamber varies from the reference condition. Therefore, the field factor ;Q for the A1SL chamber was calculated for a small field using Monte Carlo simulation. This factor was used to correct the measured absorbed dose to water for the A1SL chamber. ;Q for the A1SL chamber is expressed as follows: ;Q ¼ Dfnr = D f nr ch; D f ref w;q = Df ref ch;q ; (3) where D ch is the absorbed dose to air in an ion chamber in the reference and nonreference fields. Simulations were performed using the egs_chamber user code for the EGSnrc Monte Carlo code system. 20 The A1SL chamber was modeled using blueprint specifications obtained from the manufacturer (Standard Imaging, Inc., Middleton, WI, USA). The calculation parameters are describedinsection2.b. To validate the accuracy of A1SL modeling, the determined correction factor was compared with that obtained by Kamio et al. 21 In their investigation, the collimator jaws defined square fields with side lengths of 0.5, 0.75, 1, 1.5, and 4 cm, and the volume of water in which the dose is absorbed is the sensitive volume of the detector filled with water. In this validation, the calculation setting was the same as their setting. Figure 3 shows the field-size correction factor for the A1SL chamber as a function of field size. The average difference between our result and that of Kamio et al. is 0.4% (i.e., good agreement was obtained). For the cm 2 field, the difference is 1.4%. The electron-collecting region is not defined in the blueprint of an A1SL chamber. The field-size correction factor calculated via the Monte Carlo method depends on how the collecting region is defined in a considerably small field. Therefore, in a cm 2 field, the difference in the field-size correction factor is greater than other fields. 2.B. Monte Carlo simulation To evaluate the cause of response variation of the RPLD to the small field and IMRT beam, the field-size correction FIG. 3. Variation in the calculated field-size correction factors as a function of the jaw-field size. [Color figure can be viewed at wileyonlinelibrary.com]

4 385 Hashimoto et al.: Field-size correction factors of RGD for small-field 385 factor ;Q, density perturbation factor (P q), atomic composition perturbation factor (P med ), and volume-averaging factor (P vol ) were calculated using the EGSnrc code system Each perturbation factor was determined by a ratio of doses scored in two different simulation geometries. Each scored dose was calculated using the egs_chamber code. 20,24 As with previous studies, 14,26,27 a series of cavity doses is defined as follows: (1) D RPLD : absorbed dose in the sensitive volume of the RPLD. (2) D WERPLD : absorbed dose in the sensitive volume of the WERPLD, which is an artificial material that has all the dosimetric (and atomic) properties of normal RPLD, except that its density is equal to that of water. (3) D w,vol : absorbed dose in the sensitive volume of the RPLD, replaced with water. (4) D w,point : absorbed dose in a mm 3 voxel of water placed at the centroid of the active volume of the RPLD, representing the absorbed dose at a point in water at the location of measurement. The series of scoring volumes and each perturbation factor are illustrated in Fig. 4. ;Q was determined as follows: ;Q ¼ Dfnr = D f nr RPLD; D f ref w;q = Df ref RPLD;Q ¼ P q P med P vol P q P med P vol fnr fref Q : (4) The calculation-field conditions were the same as the measurement condition. These factors were calculated with a statistical uncertainty of 0.1%. The energy threshold and cutoff energy were set to AE = ECUT = MeV and AP = PCUT = 0.01 MeV, respectively. For the Monte Carlo calculations using a photon-beam source, radiation transport in the accelerator was modeled using the BEAMnrc Monte Carlo code. 28,29 The accelerator geometry and materials were obtained from the manufacturer s data for the Clinac 21EX and TrueBeam. Simulation parameters were optimized to reproduce the actual dose distribution, and reproducibility was confirmed by the good agreement between the measured and simulated dose distributions at the first step of the Monte Carlo procedure C. Dose measurement using an RPLD and an ionization chamber for clinical-class-specific reference fields To evaluate the response of the RPLD to clinical-classspecific reference fields, the field-size correction factor for the clinical IMRT plan was determined. To study the representative clinical IMRT plan, 24 cases were selected from the clinical case databases of the Tokyo Metropolitan Komagome Hospital (TMKH) and the Chiba Cancer Center (CCC). The selected cases comprised 12 PR and 12 HN cancer patients. All plans used 2.0 Gy per fraction. The RPLD and A1SL ionization chamber were alternately positioned at the center of the cm 3 WE211 phantom and irradiated at the same coordinates and under the same conditions. The measurement point was set to high-dose and low-dose gradient regions. Measurements at the TMKH were performed using Siemens ONCOR and a Varian Clinac 21EX linear accelerator and those at the CCC were performed using a Varian TrueBeam linear accelerator. Five RPLDs were used for each planned measurement. For reference, five RPLDs were irradiated to a dose of 2.0 Gy at each machine and beam energy, each under the same experimental conditions using a reference field of size cm 2. The field-size correction factor was obtained using Eq. (2). 3. RESULTS AND DISCUSSION 3.A. Field-size correction factors for jaw fields Figure 5 shows the measured and calculated ;Q for several jaw fields. The differences between the measured and calculated ;Q values were within 0.8%. This result is in good agreement with the calculated ;Q and the data from other studies. 7,12,13 To analyze the field-size dependence of various correction factors, we evaluated the field response in three subgroups: large fields ( cm cm 2 ), small fields (3 9 3cm cm 2 ), and very small fields ( cm cm 2 ). For large fields, the calculated ;Q values increased by 2.6% and 1.2% with decreasing field sizes for the 6- and 10-MV x rays, respectively. Figure 6 shows P med, P q,andp vol, as calculated using Monte Carlo simulations for several jaw-field conditions. The FIG. 4. Illustration of the series of cavity doses simulated to calculate the perturbation factors. [Color figure can be viewed at wileyonlinelibrary.com]

5 386 Hashimoto et al.: Field-size correction factors of RGD for small-field 386 FIG. 5. Variation in the measured and Monte Carlo-calculated field-size correction factors as a function of the jaw-field size for (a) the 6- and (b) 10-MV x rays. [Color figure can be viewed at wileyonlinelibrary.com] P med value in the large fields increased by 2.9% and 1.6% with decreasing field sizes for the 6- and 10-MV x rays, respectively. These results show that the variation in P med was the largest contributor to the variation in ;Q for the large fields. For small fields, the calculated ;Q decreased by 0.9% and 0.4% with decreasing field size for the 6- and 10-MV x rays, respectively. In this field range, no variation in P med was observed, P q decreased by 1.0% and 2.0%, and P vol was unity and increased by 1.4% for the 6- and 10-MV x rays, respectively, when the field size decreased. In addition, P q had the dominant influence on the variation in ;Q. Scott et al. reported that in the cm 2 6-MV photon field, the P q value was 0.90 and 0.96 for diamond density water (q = 3.5 g/cm 3 ) and silicon density water (q = 2.3 g/cm 3 ), respectively. 14 For jaw fields of size 2 9 2cm 2 or less, the lateral charged particle equilibrium (LCPE) was not established for the 6- and 10-MV x rays. Under the condition without LCPE, the contribution of the absorbed dose by the primary beam directed at the cavity in the dense material FIG. 6. Variation in the Monte Carlo-calculated perturbation factor as a function of the jaw-field size for (a) the 6- and (b) 10-MV x rays. [Color figure can be viewed at wileyonlinelibrary.com] increased; therefore, the dense-detector response increased and P q for this detector decreased. 15 The density of RPLD was 2.61 g/cm 3, and RPLD was affected by the density perturbations in the smaller field. For jaw fields of size 2 9 2cm 2 or less, the P q value for the RPLD decreased. The calculated ;Q values for very small fields increased by 0.7% and 1.4% with decreasing field sizes for the 6- and 10-MV x rays, respectively. The variation in P med was unity, the P q values were decreased by 2.2% and 2.3%, and the P vol values were increased by 2.7% and 3.7% for the 6- and 10- MV x rays for field size cm 2. Azangwe et al. reported that the P vol value for the RPLD in the cm 2 field was 1.03 for the 6-MV x ray. 16 In this field range, P vol had the dominant influence on the variation in ;Q. However, P vol increased and P q decreased with decreasing field size. Therefore, P vol and P q canceled each other. Overall, these results show that the volume-averaging and density perturbation effects were dominant in terms of the variation in the RPLD response in a small-photon field. Moreover, for fields of size 1 9 1cm 2 or less, perturbation corrections are required for measurements using an RPLD.

6 387 Hashimoto et al.: Field-size correction factors of RGD for small-field B. Field-size correction factor for an MLC field and an IMRT beam Figure 7 shows the measured and calculated ;Q values for MLC fields. The ;Q value was within 1.0% of unity for all MLC field sizes. The differences between the measured and calculated ;Q values were within 0.5%. For MLC fields with sizes ranging from cm 2 to cm 2, the calculated ;Q value was increased by 0.4% and 0.2% for the 6- and 10-MV x rays, respectively. For cm 2 MLC field, ;Q decreased by 0.5% and 0.9% for the 6- and 10-MV x rays, respectively, from its value for a cm 2 MLC field. Figure 8 shows P med and P q for the MLC fields. P med in the MLC field with size ranging from cm 2 to cm 2 increased by 1.1% and 0.8% for the 6- and 10- MV x rays, respectively. P q was unity over the cm 2 MLC field and decreased by 0.8% and 1.6% for the 6- and 10-MV x rays, respectively, from its value for a cm 2 MLC field. These results were similar to those of the jaw field. For MLC fields with sizes ranging from cm 2 to cm 2, P med had the dominant influence on the variation in k f clin;f ref Q clin ;Q and P q had an influence on the variation in ;Q for MLC fields with size cm2 or less. However, in these MLC fields, the variation in ;Q was within 1.0% for all field sizes, which can be ignored. Figure 9 shows the measured and calculated ;Q values, and Fig. 10 shows P med and P q for the off-axis MLC, simple- IMRT, and plan-class-specific reference fields. The differences between the measured and calculated ;Q values were within 1.5%. The calculated ;Q values in the 2- and 4-cm off-axis MLC field were and for 6 MV and and for 10 MV, respectively. These values are considerably large, and correction is required. The P med values for the 2- and 4-cm off-axis MLC fields were and for 6 MV and and for 10 MV, respectively. Scarboro et al. reported that for cm 2 6-MV photon field, the mean photon energy was 0.31 and 0.28 MeV, and the fieldsize correction factor for the optically stimulated luminescent dosimeter was 0.85 and 0.81 at 15- and 20-cm off-axis distance, respectively. 31 For out-of-field measurement, the number of low-energy scattered photons increased and photon FIG. 7. Variation in the measured and Monte Carlo-calculated field-size correction factor as a function of the X side of the MLC field size for (a) the 6- and (b) 10-MV x rays. The Y-collimator jaws were set to 10 cm for all MLC fields. [Color figure can be viewed at wileyonlinelibrary.com] FIG. 8. Variation in the Monte Carlo-calculated perturbation factor as a function of the X side of the MLC field size for (a) the 6- and (b) 10-MV x rays. [Color figure can be viewed at wileyonlinelibrary.com]

7 388 Hashimoto et al.: Field-size correction factors of RGD for small-field 388 FIG. 9. Variation in the measured and calculated field-size correction factors for several field conditions for (a) the 6- and (b) 10-MV x rays. [Color figure can be viewed at wileyonlinelibrary.com] energy decreased with increasing distance from the field edge to the measurement position. 31 Therefore, P med decreased for off-axis MLC field. The P q values for 2- and 4-cm off-axis MLC fields were and for 6 MV and and for 10 MV, respectively. Moreover, the average path length (and consequently, the fluence) of electrons was lower at high densities; hence, the fluence decreased with increasing cavity-mass density. Consequently, the dense-detector response decreased and P q for the dense detector increased. These results show that out-of-field variation in ;Q is caused by P med and P q. In the measurement of the IMRT plan for the organ-at-risk region (low-dose region), response correction may be required for accurate measurement because there are many contributions to the off-axis field and perturbation may increase in this region. The IMRT plan differs for each patient, and the contribution of the off-axis field is dependent on each irradiation plan; thus, in the measurement of the IMRT plan for the organ-at-risk region, a correction factor is required for each plan. The respective ;Q, P med, andp q values were within 0.6% for simple-imrt and plan-class-specific reference fields. The PTV region for the IMRT plan had high- and lowdose gradient regions. This region was constructed by many small on- and off-axis fields. When a high- and low-dose gradient region comprises small on- and off-axis fields, the lack FIG. 10. Variation in the Monte Carlo-calculated perturbation factor for several field conditions for (a) the 6- and (b) 10-MV x rays. [Color figure can be viewed at wileyonlinelibrary.com] of LCPE is compensated and the effect of density perturbation becomes small. Moreover, in the low-dose gradient region, the detector volume averaging is negligible. Therefore, perturbation was very small for measurement of the IMRT plan at a PTV region using RPLD. Scarboro et al. investigated the energy response of an OSLD and a TLD for the IMRT conditions. 31,32 They reported that the calculated correction factor of the OSLD and TLD for the simple-imrt condition, which was created to duplicate the simulated uniform field-five beamlets of the cm 2 field, was 1.00 and that a correction factor for the IMRT beam was unnecessary. In this study, the calculated correction factors were and for the 6- and 10-MV simple-imrt conditions, respectively. These results were almost the same as the TLD and OSLD results. 3.C. Clinical-class-specific reference field Figure 11 shows a histogram of ;Q for 24 IMRT plans. The ;Q value was less than the 1.2% difference for all plans. The mean ;Q value was with a standard deviation of This result shows that the perturbation effects of the RPLD for the IMRT plan were insignificant and that an RPLD could be used for IMRT-plan measurement.

8 389 Hashimoto et al.: Field-size correction factors of RGD for small-field 389 FIG. 11. Histogram of the k fnr;fref Qnr;Q values for 24 intensity-modulated radiation therapy (IMRT) plans. 4. CONCLUSIONS In this study, we evaluated the field-size corrections and perturbation of the RPLD under the small-field and IMRT conditions using experimental measurements and Monte Carlo simulation. We showed that atomic composition perturbation was the dominant effect for the variation in the RPLD response over a 2 9 2cm 2 field. Moreover, it clarifies that the volumeaveraging effect caused by high-dose gradient and the density perturbation caused by lack of LCPE were the dominant effects causing the variation in the RPLD response for a 2 9 2cm 2 field or less and that a correction factor is required for dosimetry in smaller fields. Although the high- and low-dose gradient PTV regions comprised multiple small on- and off-axis fields, the differences in the RPLD responses between the reference and IMRT conditions were close to unity. This is because a lack of LCPE was compensated for and the values of P q and P vol were close to unity under the IMRT condition. These results indicate that the RPLD is a suitable detector for IMRT-dose measurements in the PTV region. ACKNOWLEDGMENTS This study was supported by research grants from the Association for Nuclear Technology in Medicine. The authors would like to thank Enago ( for the English language review. a) Author to whom correspondence should be addressed. Electronic mail: s.hashimoto@cick.jp. REFERENCES 1. Molineu A, Followill DS, Balter PA, et al. Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the Radiation Therapy Oncology Group. Int J Radiat Oncol Biol Phys. 2005;63: Followill DS, Evans DR, Cherry C, et al. Design, development, and implementation of the radiological physics center s pelvis and thorax anthropomorphic quality assurance phantoms. Med Phys. 2007;34: Molineu A, Hernandez N, Nguyen T, Ibbott G, Followill D. Credentialing results from IMRT irradiations of an anthropomorphic head and neck phantom. Med Phys. 2013;40: Budgell G, Berresford J, Trainer M, Bradshaw E, Sharpe P, Williams P. A national dosimetric audit of IMRT. Radiother Oncol. 2011;99: Nose T, Koizumi M, Yoshida K, et al. In vivo dosimetry of high-doserate brachytherapy: study on 61 head-and-neck cancer patients using radiophotoluminescence glass dosimeter. Int J Radiat Oncol Biol Phys. 2005;61: Nose T, Koizumi M, Yoshida K, et al. In vivo dosimetry of high-doserate interstitial brachytherapy in the pelvic region: use of a radiophotoluminescence glass dosimeter for measurement of 1004 points in 66 patients with pelvic malignancy. Int J Radiat Oncol Biol Phys. 2008;70: Mizuno H, Kanai T, Kusano Y, et al. Feasibility study of glass dosimeter postal dosimetry audit of high-energy radiotherapy photon beams. Radiother Oncol. 2008;86: Rah JE, Kim S, Cheong KH, et al. Feasibility study of radiophotoluminescent glass rod dosimeter postal dose intercomparison for high energy photon beam. Appl Radiat Isot. 2009;67: Tsuda M. A few remarks on photoluminescence dosimetry with high energy X-rays. Jpn J Med Phys. 2000;20: Alfonso R, Andreo P, Capote R, et al. A new formalism for reference dosimetry of small and nonstandard fields. Med Phys. 2008;35: Araki F, Moribe N, Shimonobou T, Yamashita Y. Dosimetric properties of radiophotoluminescent glass rod detector in high-energy photon beams from a linear accelerator and cyber-knife. Med Phys. 2004;31: Araki F, Ohno T. The response of a radiophotoluminescent glass dosimeter in megavoltage photon and electron beams. Med Phys. 2014;41: Mizuno H, Fukumura A, Fukahori M, et al. Application of a radiophotoluminescent glass dosimeter to nonreference condition dosimetry in the postal dose audit system. Med Phys. 2014;41: Scott AJ, Kumar S, Nahum AE, Fenwick JD. Characterizing the influence of detector density on dosimeter response in non-equilibrium small photon fields. Phys Med Biol. 2012;57: Bouchard H, Seuntjens J, Duane S, Kamio Y, Palmans H. Detector dose response in megavoltage small photon beams. I. Theoretical concepts. Med Phys. 2015;42: Azangwe G, Grochowska P, Georg D, et al. Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams. Med Phys. 2014;41: Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36: Japan Society of Medical Physics (JSMP). Standard Dosimetry of Absorbed Dose in External Beam Radiotherapy. Tokyo: Tsusho-Sangyo- Kenkyu-Sha; Andreo P, Burns DT, Hohlfeld K, et al. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. IAEA TRS; 2000; Wulff J, Zink K, Kawrakow I. Efficiency improvements for ion chamber calculations in high energy photon beams. Med Phys. 2008;35: Kamio Y, Bouchard H. Correction-less dosimetry of nonstandard photon fields: a new criterion to determine the usability of radiation detectors. Phys Med Biol. 2014;59: Kawrakow I. Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version. Med Phys. 2000;27:

9 390 Hashimoto et al.: Field-size correction factors of RGD for small-field Kawrakow I, Rogers DW. The EGSnrc Code System: Monte Carlo Simulation of Electron and Photon Transport Technical Report PIRS-701. Ottawa: NRCC; Kawrakow I, Mainegra-Hing E, Tessier F, Walters B. The EGSnrc C++ Class Library, NRC Report PIRS-898 (rev A). Ottawa: NRCC; Rogers D, Kawrakow I, Seuntjens J, Walters B. NRC User Codes for EGSnrc NRC Technical Report PIRS-702. Ottawa: NRCC; Bouchard H, Seuntjens J, Carrier JF, Kawrakow I. Ionization chamber gradient effects in nonstandard beam configurations. Med Phys. 2009;36: Francescon P, Beddar S, Satariano N, Das IJ. Variation of kqclin, Qmsr (fclin, fmsr) for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio. Med Phys. 2014;41: Rogers D, Faddegon B, Ding G, Ma CM, We J, Mackie T. BEAM: a Monte Carlo code to simulate radiotherapy treatment units. Med Phys. 1995;22: Rogers D, Walters B, Kawrakow I. BEAMnrc Users Manual. Ottawa: NRCC; Fujita Y, Saitoh H, Myojoyama A. Bremsstrahlung and photoneutron leakage from steel shielding board impinged by MeV electrons beams. J Radiat Res. 2009;50: Scarboro SB, Followill DS, Kerns JR, White RA, Kry SF. Energy response of optically stimulated luminescent dosimeters for non-reference measurement locations in a 6 MV photon beam. Phys Med Biol. 2012;57: Scarboro SB, Followill DS, Howell RM, Kry SF. Variations in photon energy spectra of a 6 MV beam and their impact on TLD response. Med Phys. 2011;38:

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams Ernesto Mainegra-Hing a) and Bryan R. Muir Measurement Science and Standards, National Research Council of Canada,

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS PAPER EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS Isra Israngkul Na Ayuthaya *, Sivalee Suriyapee, Phongpheath Pengvanich

More information

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Megan A. Hyun, M.S. Under the supervision of Larry A. DeWerd, Ph.D. NCCAAPM Fall Meeting October 30, 2015 Medical Radiation Research Center

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)

More information

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Ion recombination corrections of ionization chambers in flattening filter-free photon radiation Yuenan Wang, a Stephen B. Easterling,

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

Applications of Monte Carlo simulations to radiation dosimetry

Applications of Monte Carlo simulations to radiation dosimetry Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

Dosimetry of Small Static Fields Used in External Beam Radiotherapy

Dosimetry of Small Static Fields Used in External Beam Radiotherapy Technical Reports SeriEs No.483 Dosimetry of Small Static Fields Used in External Beam Radiotherapy An International Code of Practice for Reference and Relative Dose Determination Sponsored by the IAEA

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

Neutron dose evaluation in radiotherapy

Neutron dose evaluation in radiotherapy Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head

More information

The influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainLAB m3 micro-multileaf collimator

The influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainLAB m3 micro-multileaf collimator JOURNA OF APPIED CINICA MEDICA PHYSICS, VOUME 13, NUMBER 6, 2012 The influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainAB m3 micro-multileaf collimator

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the

More information

The clinical use of OSLD

The clinical use of OSLD The clinical use of OSLD Stephen F. Kry PhD Radiological Physics Center, Imaging and Radiation Oncology Core - Houston IROC Houston The University of Texas MD Anderson Cancer Center, Houston, TX Spring

More information

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS} High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement Accurate clinical dosimetry Part I: Fundamentals and New developments in reference dosimetry Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement

More information

Small Field Dosimetry

Small Field Dosimetry Small Field Dosimetry Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin Madison WI Conflict

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Much of what we will say

Much of what we will say New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

Neutron Measurements for Intensity Modulated Radiation Therapy

Neutron Measurements for Intensity Modulated Radiation Therapy SLAC-PUB-8443 April 2 Neutron Measurements for Intensity Modulated Radiation Therapy N. E. Ipe et al. Presented at Chicago 2 World Congress on Medical Physics and Biomedical Engineering, 7/23/2 7/28/2,

More information

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA M. Pimpinella, A. S. Guerra, S. La Civita and R. F. Laitano Istituto

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.

More information

PRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR.

PRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR. IX Latin American IRPA Regional Congress on Radiation Protection and Safety - IRPA 2013 Rio de Janeiro, RJ, Brazil, April 15-19, 2013 SOCIEDADE BRASILEIRA DE PROTEÇÃO RADIOLÓGICA - SBPR PRECISE DOSIMETRY

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators

Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators Madelaine K. Tyler,

More information

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article. Original Article PMP Progress in Medical Physics 29(1), March 218 https://doi.org/1.14316/pmp.218.29.1.16 pissn 258-4445, eissn 258-4453 Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers

More information

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields Dosimetric Uncertainties in Reerence and Relative Dosimetry o Small Fields Jan Seuntjens, Ph.D., FAAPM, FCCPM McGill University Health Centre Canada Outline o Presentation Uncertainty concepts and requirements

More information

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD Proceedings of the Tenth EGS4 Users' Meeting in Japan, KEK Proceedings 2002-18, p.65-73 DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD N. Tohyama, H. Saitoh, T. Fujisaki 1, S. Abe 1

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

Detector comparison for small field output factor measurements with flattening filter free photon beams

Detector comparison for small field output factor measurements with flattening filter free photon beams Detector comparison for small field output factor measurements with flattening filter free photon beams 1 Wolfgang Lechner, 1 Lukas Sölkner, 2 Paulina Grochowska, 1 Dietmar Georg 1 Christian Doppler Laboratory

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada IAEA/AAPM Code o Practice or the Dosimetry o Static Small Photon Fields Jan Seuntjens McGill University Montréal, Canada Acknowledgements IAEA/AAPM small and composite ield working group: Hugo Palmans

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1) Indra J. Das/ Small Fields (Page-1) Small Field Dosimetry: Overview of AAP TG-155 and the IAEA-AAP Code of Practice (Therapy) Indra J. Das, PhD, FAAP, FACR, FASTRO Department of Radiation Oncology Indiana

More information

How accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues

How accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues Accurate Clinical dosimetry Part I: Fundamentals and Issues Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement effect of dose uncertainty

More information

*Corresponding author. Tel: (Ext: 91074); Fax: ;

*Corresponding author. Tel: (Ext: 91074); Fax: ; Journal of Radiation Research, 2014, 55, 600 607 doi: 10.1093/jrr/rrt135 Advance Access Publication 1 January 2014 Design, development of water tank-type lung phantom and dosimetric verification in institutions

More information

AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy

AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy Stephen F. Kry a) Department of Radiation Physics, MD Anderson Cancer Center, Houston,

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Patient dosimetry for total body irradiation using single-use MOSFET detectors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Patient dosimetry for total body irradiation using single-use MOSFET detectors Tina Marie Briere, 1 Ramesh Tailor, 1 Naresh Tolani,

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP Iran. J. Radiat. Res., 2006; 4 (1): 15-19 Dose distribution and dosimetry parameters calculation of MED3633 palladium- source in water phantom using MCNP A.A. Mowlavi 1*,A. Binesh 2, H. Moslehitabar 3

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment Oncology and Translational Medicine DOI 10.1007/s10330-016-0183-3 December 2017, Vol. 3, No. 6, P260 P266 REVIEW ARTICLE A comparative study between flattening filter-free beams and flattening filter beams

More information

IMRT point dose measurements with a diamond detector

IMRT point dose measurements with a diamond detector IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,

More information

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method Iran. J. Radiat. Res., 2007; 5 (3): 147-152 Dosimetric characteristics of Cs sources used in after loading Selectron system by Monte Carlo method M.B.Tavakoli, D. Shahbazi-Gahrouei *, M. Hosseinpour Department

More information

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System James D Rijken Thesis submitted for the degree of Master of

More information

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning C-M M Charlie Ma, Ph.D. Department of Radiation Oncology Fox Chase Cancer

More information

RPC s Credentialing Programs for Clinical Trials

RPC s Credentialing Programs for Clinical Trials RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups

More information

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data J. A. Bencomo, N. Hernandez, and W.F. Hanson University of Texas M.D. Anderson Cancer Center Medical physics,

More information

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Professor, Department of Medical Physics

More information

Protection of the contralateral breast during radiation therapy for breast cancer

Protection of the contralateral breast during radiation therapy for breast cancer Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,

More information

Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT

Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT Rajesh

More information

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National

More information

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an

More information

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe 3D printing of air slab caps for correction free small field dosimetry Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe Small fields? Defined as fields sizes too small to achieve

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West

More information

Quality Assurance of Helical Tomotherapy Machines

Quality Assurance of Helical Tomotherapy Machines Quality Assurance of Helical Tomotherapy Machines Wolfgang Tomé 1,2, Hazim Jaradat 2, David Westerly 1, John Fenwick 3, Nigel Orton 2, Rock Mackie 1,2, and Bhudatt Paliwal 1,2 1 Human Oncology and 2 Medical

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

Radiation qualities in carbon-ion radiotherapy at NIRS/HIMAC

Radiation qualities in carbon-ion radiotherapy at NIRS/HIMAC Radiation qualities in carbon-ion radiotherapy at NIRS/ Shunsuke YONAI Radiological Protection Section Research Center for Charged Particle Therapy National Institute of Radiological Sciences (NIRS) E-mail:

More information

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction

More information

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output Presented by: Rebecca M. Howell PhD, DABR, FAAPM Results from Voluntary Independent External Peer Review of Beam Output What do we hope to achieve with independent peer review? Independent peer review

More information