1 Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa ICTP, Trieste, Nov 14,

2 Papers in PMB and Med Phys with Monte Carlo in title or abstract 2/64

3 Radiation dosimetry in radiotherapy primary standards air kerma, absorbed dose electron & photon beams beta-ray fields clinical dosimetry protocols dose in a water tank TG51, TG61, TG43, TRS-398 radiotherapy treatment planning dose in a (CT) patient 3/64

4 radiation dosimeters many types of radiation dosimeters for radiotherapy ion chambers- the work horse for clinical reference dosimetry and air kerma primary standards calorimeters for absorbed dose primary standards free air chambers for x-ray air kerma standards TLDs LiF diodes, MOSFETS radiographic and radiochromic films chemical (Fricke) dosimeters Monte Carlo calculations have been used to elucidate all of these. 4/64

5 Ion chambers Farmer ion chamber from John McCaffrey, NRC 5/64

6 Cavity theory: stopping-power ratios Relates dose in cavity to dose in medium. sprs are fundamental to med gas -dosimetry protocols -primary standards 6/64

7 What is (L/ρ)? A Spencer-Attix spr - stopping-power ratio 7/64

8 Dosimetry in a phantom P wall, P gr, P fl, P cel all 1% or less effects -major variation comes from spr for complete definitions of P wall etc see 8/64

9 Electron beam depth-dose curve 12 MeV 9/64

10 sprs in electron beams Ding et al, Med Phys 22(1995) /64

11 Realistic electron beam sprs BEAM code used to simulate realistic accelerator beams Ding et al Med Phys 22 (1995)489 11/64

12 Effects of realistic sprs Ding et al MP 22(1995)489 12/64

13 How to use realistic sprs David Burns noted: changing d ref simplifies everything. d ref = 0.6 R (cm) The basis of electron beam dosimetry in IAEA TRS-398 and AAPM TG-51 clinical protocols Burns et al MP 23(1996)383 13/64

14 Realistic sprs: d ref =0.6R Burns et al MP 23(1996)383 14/64

15 Photon beams: specifying beam quality NAP -nominal accelerating potential %dd(10) -percentage depth dose at 10 cm depth in a 10x10 cm 2 field on surface at SSD 100 cm %dd(10) X -the photon component of %dd(10) TG-51 (i.e., ignoring electron contamination) TPR ratio of absorbed doses at depths 20 TRS-398 and 10 cm in a water phantom, measured with a constant source-chamber distance of 100 cm and a field size of 10x10 cm 2 at the plane of the chamber 15/64

16 sprs for photon beams filled: heavily filtered open: lightly filtered Kalach and Rogers 30 (2003) /64

17 sprs for photon beams filled: heavily filtered open: lightly filtered Kalach and Rogers 30 (2003) /64

18 What happens without a flattening filter? For IMRT, flattening filter is not needed (Titt et al, Med Phys 33(2006) 3270). A single fit handles both sets of beams using %dd(10) x. Major effect is on %dd(10) x due to non-flat beams Xiong and Rogers, in prep, 2007 Based on full BEAM simulations. 18/64

19 Flattening filter free: TPR Two sets of k Q values will be needed, one for with flattening filters, one for machines without them. 19/64

20 Summary: protocol dosimetry the major quantity which varies in protocol dosimetry is the stopping power ratio hence the discussion of it but other aspects of protocols such as TG-51 and TRS-398 which are based on MC calculated values P wall for plane parallel chambers in Co-60 beams P cel for aluminium electrodes relationship between I 50 and R 50 in e- beams plus on-going research on other aspects P wall for all beams, P repl, effective point of measurement 20/64

21 Primary standards of air-kerma in Co-60 Primary standards in Co-60 beams are based on cavity ion chambers and S-A cavity theory D gas D wall /D gas D air /D wall for complete definitions see 21/64

22 How accurately can we calculate ion chamber response? Fano cavity chamber, - walls and gas the same material (assume graphite) with a density ratio of about establish kerma to graphite in a parallel 60 Co beam. Fano s theorem => no fluence correction (traditionally ignored, but in principle needed). All other K = 1.00 ie we can check our D gas calculation 22/64

23 How accurately can we calculate ion -cover of EGSnrc manual -against own cross sections -ESTEPE is max fractional step size chamber response? (cont) This is the toughest test I know for any electron-photon Monte Carlo code 23/64

24 How accurately can we calculate ion chamber response? (cont) against measured data Kawrakow & Rogers, MC2000, p135 based on data of Nilsson et al, IAEA Proceedings, /64

25 K wall : attenuation and scatter K air eqn ignores attenuation and scatter in chamber walls D gas or Monte Carlo K wall scores without / D gas with scatter and attenuation Or regenerate interacting photons & ignore scattered photons 25/64

26 K wall : non-linear extrapolation Rogers & Bielajew, PMB 35 (1990) /64

27 Some measured confirmations of MC K wall rotate the chamber in Co-60 response*k wall =response/a wall should be constant graphite walled chamber at NRC 27/64

28 Response vs angle of Mark IV If A wall is correct, R/A wall should be constant. It is, within 0.3% despite 8% variation. (residual 0.3% is a K an effect) McCaffrey et al PMB 49(2004) /64

29 PTB/OMH: cylindrical chamber radial axial axis of rotation 45 measured response vs wall thickness. Should all extrapolate to same value. Only the calculated K wall correction gave a constant response Büermann et al PMB 48 (2003) /64

30 K an : axial non-uniformity Bielajew developed an analytic theory to account for point sources not parallel beams (PMB 35(1990)501 & 517) A brute force MC calculation with a parallel beam or a point source, confirms the analytic theory. The corrections are all very small for Co-60 sources at 1 m from typical chambers 30/64

31 Revision of air-kerma standards Using EGSnrc calculated K wall and K an values, revise the reported values Rogers and Treurniet, 1999 (NRC Report PIRS-663) extending work of Bielajew and Rogers, PMB 37(1992) /64

32 Revision of air-kerma standards (cont) Note: the BIPM baseline moved up by 0.3% Monte Carlo => world s air kerma standards increased 0.8% (double stated uncertainty) Rogers & Treurniet 1999 NRC Report 32/64

33 How accurate are calculations? If we are going to use Monte Carlo calculated factors, we need to know their uncertainty How sensitive are they to: -algorithm/computer code used -cross sections -spectrum used -size of source Rogers & Kawrakow Med Phys 30 (2003)521 33/64

34 Calculated response of NRC 3C chamber 34/64

35 K wall for NRC 3C 35/64

36 (L/ρ) for different algorithms 36/64

37 K wall vs incident spectrum 37/64

38 K an vs incident spectrum 38/64

39 spr vs incident spectrum 39/64

40 K an for 3C vs source radius 40/64

41 Uncertainty estimates (%) spr K wall K K an comp Stats <0.01 < Algorithm Spectrum 0.01 < e- X-sec γ X-sec Rogers & Kawrakow Med Phys 30 (2003)521 41/64

42 Verification of cavity theory? Can Monte Carlo verify the accuracy of cavity theory? EGSnrc can calculate D gas to 0.1% (proof: Fano cavity calculations) Cavity theory assumes that photon interactions in the cavity do not occur But Ma and Nahum showed they did. PMB 36(1991)413 So does cavity theory hold for Ir-192 or lower energy photon beams? 42/64

43 Accuracy of Spencer-Attix cavity theory Another thought/computational experiment For a parallel beam incident on a stemless chamber filled with dry air spectrum SPRRZnrc DOSRZnrc CAVRZnrc CAVRZnrc EGSnrc 43/64

44 Accuracy of Spencer-Attix cavity theory Only this good because graphite and air so similar. Calculations used = 10 kev for spr. Using larger values brings value within 0.1% of unity Borg et al, Med Phys 27(2000) /64

45 The use of silicon diode detectors a common assumption is that diode detectors measure dose directly ie no spr correction etc but sprs actually change quite a bit as the beam quality changes Why don t we need to correct for this? 45/64

46 water/silicon stopping powers are not constant calculate ratio of dose in small active region of diode detector isolated from rest of detector to dose to water at same location. Use CSnrc which uses correlated sampling Wang Med Phys 34 (2007) /64

47 model of diode detector (Scanditronix EFD) McKerracher and Thwaites Radioth Oncol 79(06) /64

48 dose water/dose silicon active region Wang Med Phys 34 (2007) /64

49 effect of backscatter from rest of chip Wang Med Phys 34 (2007) /64

50 diode response at d max vs field size mostly a change in spr effect as d max changes Wang Med Phys 34 (2007) /64

51 Summary re diode detectors diodes measure dose directly within +-1% as a function of depth and beam quality in electron beams one exception - small radius electron beams the silicon backing of the active region and the epoxy play an important role in the flat response 51/64

52 P TP : the pressure-temperature correction for ion chambers PTP is constructed so t e- ion chamber independent of ρ So E dep (ρ) is proportional to the density ρ. E dep (ρ ο ) is independent of the density ρ. 52/64

53 P TP (cont) e- What happens if the electron does not cross the cavity? ion chamber independent of ρ E dep (ρ) is no longer proportional to the density ρ. Hence the standard P TP correction factor may no longer work. 53/64

54 Pressure vs. altitude 54/64

55 A4 EGSnrc Monte Carlo code NE2571 A12 NRC x- ray monitor cross-sections for DRY air of different densities calculate D cav (dose to air) standard P TP correction inherent in results PTB catalogued spectra 55/64

56 Thimble chamber calculations 56/64

57 Conclusions of P TP paper I there is a significant breakdown of the standard P TP correction for low energy photon beams basic cause: e- stopping in the cavity, not crossing magnitude of the effect depends on: mismatch of wall to air cross sections fraction of dose due to photon interactions in the cavity air a similar effect was reported in 2005 by the UW ADCL for well ion chambers for I /64

58 Experiments at NRC to demonstrate the effect with Malcolm McEwen complete BEAMnrc model to give x-ray spectrum 58/64

59 A variety of chambers studied A2 C552 aluminium A12 C552 NE2571 NE2505 A19 graphite dural, C552 Kawrakow s egs_view Calculations with cavity.cpp, using Kawrakow s C++ geometry package & interface to EGSnrc 59/64

60 Farmer-like chambers: 60 kv Closed symbols: P TP corrected measured responses open symbols: calculated responses 60/64

61 Effects of geometry details CAVRZnrc uses a cylindrical model cavity.cpp includes the conical end. These geometry differences have no effect in a Co-60 beam 61/64

62 Summary re P PT corrections measurements confirm the calculated breakdown of the P TP correction factor for low-energy x-rays EGSnrc is capable of reproducing air-kerma calibration coefficients well within 1% N K vs beam quality curves allow quantification of the size of impurity effects geometry details have some effects at these low energies although not at Co-60 impurities are important at low photon energies 62/64

63 Summary MC techniques play a fundamental role in radiation dosimetry sprs and other corrections for ion chambers used in clinical dosimetry correction factors for primary standards verification of cavity theory accuracy elucidation of detector response (eg was diode) investigation of pressure-temperature effects and much, much more TLDs, OSL, alanine,fricke, well chambers, brachytherapy dosimetry etc 63/64

64 Acknowledgements The work described here has been done in conjunction with many colleagues, grad students and research associates, without whom it wouldn t get done. the various works described involved: Iwan Kawrakow, David Burns, George Ding, Guoming Xiong, Nina Kalach, Jette Borg, Alex Bielajew, John McCaffrey, Joanne Truerniet, Lilie Wang and Dan La Russa, but many more were involved in the overall project of Monte Carlo in radiation dosimetry Support from the Canada Research Chairs program and 64/64

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams Ernesto Mainegra-Hing a) and Bryan R. Muir Measurement Science and Standards, National Research Council of Canada,

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement

How accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement Accurate clinical dosimetry Part I: Fundamentals and New developments in reference dosimetry Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA M. Pimpinella, A. S. Guerra, S. La Civita and R. F. Laitano Istituto

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

NRC Activities and Publications, Report to CCRI(I) Meeting, BIPM May

NRC Activities and Publications, Report to CCRI(I) Meeting, BIPM May CCRI(I)/03 47 NRC Activities and Publications, 2001 2003 Report to CCRI(I) Meeting, BIPM May 21 23 2003 D.W.O. Rogers, I. Kawrakow, N.V. Klassen, J. P. McCaffrey, M. R. McEwen, P. R. B. Saull, C.K. Ross

More information

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams Peter R. Almond, Brown Cancer Center,Louisville, KY 40202 Peter J. Biggs, Department of Radiation Oncology,

More information

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System James D Rijken Thesis submitted for the degree of Master of

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Ion recombination corrections of ionization chambers in flattening filter-free photon radiation Yuenan Wang, a Stephen B. Easterling,

More information

How accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues

How accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues Accurate Clinical dosimetry Part I: Fundamentals and Issues Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement effect of dose uncertainty

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Dosimetry in digital mammography

Dosimetry in digital mammography Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol

More information

Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS,

Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS, Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at, 2005-2007 1. Introduction Gerhard Stucki, Solange Gagnebin, Damian Twerenbold, Sándor Vörös

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom Iran. J. Radiat. Res., 2003; 1(3): 143-149 Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom A. Mostaar 1, M. Allahverdi 1,2, M. Shahriari 3 1 Medical Physics Department,

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Dosimetric characterization of surface applicators for use with the Xoft ebx system Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP Iran. J. Radiat. Res., 2006; 4 (1): 15-19 Dose distribution and dosimetry parameters calculation of MED3633 palladium- source in water phantom using MCNP A.A. Mowlavi 1*,A. Binesh 2, H. Moslehitabar 3

More information

The influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainLAB m3 micro-multileaf collimator

The influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainLAB m3 micro-multileaf collimator JOURNA OF APPIED CINICA MEDICA PHYSICS, VOUME 13, NUMBER 6, 2012 The influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainAB m3 micro-multileaf collimator

More information

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

EDITORIAL NOTE. factor, which has been used instead of the chamber factor N D

EDITORIAL NOTE. factor, which has been used instead of the chamber factor N D IAEA/WHO NETWORK OF SECONDARY STANDARD DOSIMETRY LABORATORIES Prepared by the Joint IAEA/WHO Secretariat of the SSDL Network Published semi-annually by the International Atomic Energy Agency Vienna No.

More information

Much of what we will say

Much of what we will say New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

Dosimetry and QA of proton and heavier ion beams

Dosimetry and QA of proton and heavier ion beams Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference

More information

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article. Original Article PMP Progress in Medical Physics 29(1), March 218 https://doi.org/1.14316/pmp.218.29.1.16 pissn 258-4445, eissn 258-4453 Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations

Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,

More information

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation. Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation Samuel Chiquita 1 1 University of Porto, Porto, Portugal Abstract In

More information

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics Neutron Interactions Part 2 George Starkschall, Ph.D. Department of Radiation Physics Neutron shielding Fast neutrons Slow down rapidly by scatter in hydrogenous materials, e.g., polyethylene, paraffin,

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

Recent advances in dosimetry in reference conditions for proton and light-ion beams

Recent advances in dosimetry in reference conditions for proton and light-ion beams Recent advances in dosimetry in reference conditions for proton and light-ion beams S. Vatnitskiy a), P. Andreo b) and D.T.L. Jones c) a) MedAustron, Wiener Neustadt, Austria b) Medical Radiation Physics,

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

Review of TG-186 recommendations

Review of TG-186 recommendations Review of TG-186 recommendations Implementation of advanced brachytherapy dose calculation algorithms beyond TG-43 Rowan M. Thomson Carleton Laboratory for Radiotherapy Physics Carleton University Ottawa

More information

Implantable MOSFET dosimeter response to 192 Ir HDR radiation

Implantable MOSFET dosimeter response to 192 Ir HDR radiation Implantable MOSFET dosimeter response to 192 Ir HDR radiation Jessica Fagerstrom University of Wisconsin Madison, Department of Medical Physics Medical Radiation Research Center North Central Chapter AAPM

More information

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Manik Aima, Larry A. DeWerd, Wesley S. Culberson Manik Aima, Larry A. DeWerd, Wesley S. Culberson University of Wisconsin Medical Radiation Research Center, Madison, WI 25 th Annual Meeting of the Council of Ionizing Radiation Measurements and Standards,

More information

Staff Exposure Monitoring in Interventional Radiology

Staff Exposure Monitoring in Interventional Radiology Conference on Physics in Medicine: From Diagnosis to Treatment "Enhancing safety and quality in radiation medicine". KFMC, Riyadh 7-9 November 2017 Staff Exposure Monitoring in Interventional Radiology

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY E. BADITA 1, C.VANCEA 1,3, I. CALINA 1,3, D. STROE 2, M. DUMITRACHE 2,3,

More information

Neutron dose evaluation in radiotherapy

Neutron dose evaluation in radiotherapy Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry C. Candela-Juan, J. Vijande, T. García-Martínez, Y. Niatsetski,

More information

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms 1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,

More information

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD Proceedings of the Tenth EGS4 Users' Meeting in Japan, KEK Proceedings 2002-18, p.65-73 DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD N. Tohyama, H. Saitoh, T. Fujisaki 1, S. Abe 1

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS} High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning C-M M Charlie Ma, Ph.D. Department of Radiation Oncology Fox Chase Cancer

More information

Metrology and quality of radiation therapy dosimetry of electron, photon and epithermal neutron beams

Metrology and quality of radiation therapy dosimetry of electron, photon and epithermal neutron beams FI0000014 STUK-A 164 August 1999 Metrology and quality of radiation therapy dosimetry of electron, photon and epithermal neutron beams Antti Kosunen Radiation and Nuclear Safety Authority, STUK Department

More information

Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations

Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations Iran. J. Radiat. Res., 21; 7 (4): 217-222 Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations Downloaded from ijrr.com at 6:52 +33

More information

HDR Brachytherapy Dosimetry at the NPL

HDR Brachytherapy Dosimetry at the NPL 1 Introduct HDR Brachytherapy Dosimetry at the NPL Brachytherapy is a special procedure in radiotherapy that utilises the irradiat of a target volume (e.g. malignant cells) with radioactive sources placed

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

Dosimetry. Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology

Dosimetry. Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology Kantonsspital Luzern Dosimetry Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology 3.9.2007 Introduction Metrological traceability; the dosimetry chain Absolute / relative

More information

Current Status of Electronic Brachytherapy Dosimetry

Current Status of Electronic Brachytherapy Dosimetry Current Status of Electronic Brachytherapy Dosimetry 2014 NCCAAPM Fall Meeting La Crosse, WI Wes Culberson, PhD, DABR University of Wisconsin Madison University of Wisconsin Medical Radiation Research

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor

Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor BNL-66807 1 1 1 1 2 Holden, N.E., Hu, J-P., Greenberg, D.D., Reciniello, R.N., Farrell, K. and Greenwood, L.R. Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor Reference:Holden,N.E.,

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Megan A. Hyun, M.S. Under the supervision of Larry A. DeWerd, Ph.D. NCCAAPM Fall Meeting October 30, 2015 Medical Radiation Research Center

More information

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)

More information

Patient Dosimetry in Mammography and Tomosynthesis:

Patient Dosimetry in Mammography and Tomosynthesis: 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

Dosimetry of Small Static Fields Used in External Beam Radiotherapy

Dosimetry of Small Static Fields Used in External Beam Radiotherapy Technical Reports SeriEs No.483 Dosimetry of Small Static Fields Used in External Beam Radiotherapy An International Code of Practice for Reference and Relative Dose Determination Sponsored by the IAEA

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Validation of the MC platform GATE for radiation therapy applications Evaluation of dose deposited by electrons. Yann Perrot, LPC, CNRS/IN2P3

Validation of the MC platform GATE for radiation therapy applications Evaluation of dose deposited by electrons. Yann Perrot, LPC, CNRS/IN2P3 Validation of the MC platform GATE for radiation therapy applications Evaluation of dose deposited by electrons Yann Perrot, LPC, CNRS/IN2P3 1 Projects @ PCSV 2 Projects @ PCSV Computation for life science

More information

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

Recent Activities in Measurement Standards and Dosimetry at ARPANSA,

Recent Activities in Measurement Standards and Dosimetry at ARPANSA, CCRI(I)/13-17 Report to the 21 st Meeting of the CCRI(I), March 2013 Recent Activities in Measurement Standards and Dosimetry at ARPANSA, 2011-2013 Nkem Anele, Duncan Butler, Andrew Cole, Ramanathan Ganesan,

More information

Recent proceedings in Brachytherapy Physics

Recent proceedings in Brachytherapy Physics Recent proceedings in Brachytherapy Physics Frank-André Siebert UKSH, Campus Kiel, Germany Clinic of Radiotherapy Dept. Medical Physics Physical characteristics of brachytherapy (Courtesy Luc Beaulieu,

More information

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe 3D printing of air slab caps for correction free small field dosimetry Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe Small fields? Defined as fields sizes too small to achieve

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,

More information

Physics. Copyright 2010 Xoft, Inc.

Physics. Copyright 2010 Xoft, Inc. Physics Outline The XOFT System Dosimetry of X-Ray Source: TG43 Balloon Applicators and APBI Endo Rectal Applicators Cervical Applicator TG61 Surface Applicators Radioprotection: Shielding Room exposure

More information

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative

More information

NRC Activities and Publications, Report to the CCRI(I) Meeting, BIPM March 2013

NRC Activities and Publications, Report to the CCRI(I) Meeting, BIPM March 2013 NRC Activities and Publications, 2011-2013 Report to the CCRI(I) Meeting, BIPM March 2013 Compiled by Malcolm McEwen Ionizing Radiation Standards National Research Council Ottawa, Canada K1A 0R6 Tel: 613-993-2197,

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information