Patient Dosimetry in Mammography and Tomosynthesis:

Size: px
Start display at page:

Download "Patient Dosimetry in Mammography and Tomosynthesis:"

Transcription

1 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of Biomedical Engineering University of California Davis Medical Center Sacramento, California 1

2 Mammography and Tomosynthesis Dosimetry Mammography Why measure breast dose? Basic Concepts of Breast Dosimetry (how) Mean Glandular Dose (MGD) (what) DgN coefficients Skin Thickness Issues Breast Density Issues Tomosynthesis Differences between tomo and mammo Summary 2

3 U.S. Breast Cancer Statistics (2006) 212,290 new cases 40,970 deaths 1 / 8 women will get breast cancer (12.5%) Breast Cancer Lung Cancer Incidence * 31% 12% Mortality * 15% 26% Ravdin, et al., NEJM * of all cancers

4 death rate per 100,000 U.S. Cancer Mortality ( ) Jemal A, et al., Cancer Statistics

5 Mammography: Standard of Care CC MLO CC MLO 5

6 Annual screening dose accumulation average size breast 6

7 Mammography and Tomosynthesis Dosimetry Mammography Why measure breast dose? Basic Concepts of Breast Dosimetry (how) Mean Glandular Dose (MGD) (what) DgN coefficients Skin Thickness Issues Breast Density Issues Tomosynthesis Differences between tomo and mammo Summary 7

8 Digital mammography unit at UC Davis 8

9 Breast Dose is highly dependent upon the x-ray spectrum used. Different Anode/Filter combinations are used. ANODE Mo Mo Rh W W W FILTER Mo Rh Rh Rh Ag Al S/F Digital Tomo 9

10 In addition to the Anode / Filter combo, the kv and HVL need to be well characterized 10 compression paddle 10

11 X-ray Tube Voltage Evaluation 11 non-invasive kv meter 11

12 In addition to the Anode / Filter combo, the kv and HVL need to be well characterized mm thick Al filters (99.999%) ion chamber calibrated to mammo energies air gap 12

13 HVL as a function of tube voltage 13

14 14 Establishing x-ray tube output mgy air kerma per 100 mas at 50 cm ion chamber 14

15 X-ray output cm) versus tube voltage 15

16 anode / filter combo glandular fraction kv breast thickness HVL s 16

17 Validation of Breast Thickness Accuracy 4.1 cm breast phantoms (slabs) 17

18 Estimating Entrance Air Kerma From Bloomquist, et.al. Med Phys 33: ,

19 Estimating Entrance Air Kerma Recorded in patient s record (digital) anode filter kv ma exposure time compressed breast thickness 19

20 Estimating Entrance Air Kerma dgn values normalized glandular dose coefficients air kerma 20

21 Estimating Entrance Air Anode/Filter kv: air kerma per cm = k ESK = k (mas/100) ISL ISL 50 SID T 2 50 cm SID - T usually ~65 cm breast thickness 21

22 Mammography and Tomosynthesis Dosimetry Mammography Why measure breast dose? Basic Concepts of Breast Dosimetry (how) Mean Glandular Dose (MGD) (what) DgN coefficients Skin Thickness Issues Breast Density Issues Tomosynthesis Differences between tomo and mammo Summary 22

23 Mean Glandular Dose 23

24 Mean Glandular Dose skin glandular tissue adipose tissue 24

25 Mean Glandular Dose (Monte Carlo Calculations) skin homogeneous mixture glandular tissue? glandular fraction adipose tissue? 25

26 Mammography and Tomosynthesis Dosimetry Mammography Why measure breast dose? Basic Concepts of Breast Dosimetry (how) Mean Glandular Dose (MGD) (what) DgN coefficients Skin Thickness Issues Breast Density Issues Tomosynthesis Differences between tomo and mammo Summary 26

27 DgN values are based on MC Studies assumed 4 mm skin thickness 27

28 DgN values are based on MC Studies assumed 4 mm skin thickness 28

29 Monoenergetic MC eval of E deposition 29

30 Converting Energy imparted to dose. MGD = m en r m en r glandular breast comp Energy Imparted Mass f-factor (SI) 30

31 Mean Glandular Dose (Monte Carlo Calculations) ESK DgN = MGD ESK MGD 31

32 Dose Calc with Normalization by ESE 32

33 To convert the monoenergetic DgN values to realistic polyenergetic values, spectral models are used to weight the monoenergetic values 33

34 DgN Tables (poly) The units used in DgN Tables have varied over the years and by the country of origin. These units are mrad / R, but mgy/mgy is in more common usage today 34

35 35

36 DgN versus monoenergetic x-ray energy 2 cm 4 cm 6 cm 8 cm 36

37 Mammography and Tomosynthesis Dosimetry Mammography Why measure breast dose? Basic Concepts of Breast Dosimetry (how) Mean Glandular Dose (MGD) (what) DgN coefficients Skin Thickness Issues Breast Density Issues Tomosynthesis Differences between tomo and mammo Summary 37

38 t = 4 mm? 38

39 Developed algorithms to segment skin from breast CT images and measure thickness t 39

40 Skin Thickness measurements were relatively precise for each women (18%) Measurements from the same women 40

41 Skin Thickness measurements ranged from 0.9 mm to 2.3 mm (m = 1.45 mm) Measurements from 51 different women 41

42 Left / Right Comparison provided a consistency check, with good results 42

43 43

44 DgN values versus skin thickness 44

45 Change (in %) in DgN values versus skin thickness, Relative to the assumption of t = 4 mm 45

46 Mammography and Tomosynthesis Dosimetry Mammography Why measure breast dose? Basic Concepts of Breast Dosimetry (how) Mean Glandular Dose (MGD) (what) DgN coefficients Skin Thickness Issues Breast Density Issues Tomosynthesis Differences between tomo and mammo Summary 46

47 Mean Glandular Fraction high medium low 100% 50% 0% 47

48 48

49 Breast Density Analysis segmentation glandular % adipose risk assessment & dosimetry validation of 2D approaches (M. Yaffe) 49

50 including skin excluding skin UC Davis breast CT N =

51 Density analysis done on mammograms at University of Toronto Data from UC Davis Breast CT 51

52 UCD & UT data combined (N = 2831) Median 16 % 52

53 UCD & UT data combined (N = 2831) Cumulative Distribution 95% 90% 75% 50% 53

54 Breast density decreases about 5 % per decade 54

55 volume fraction (%) Areal Fraction versus Volume Fraction 100% 0% 0% 100% areal fraction (%) 55

56 DgN for 0% Glandular Breast 56

57 DgN for 100% Glandular Breast 57

58 12% 16% 50% increase in average DgN coefficient of: ( )/237 = 12% 5 cm 2 cm 4 cm 6 cm 8 cm 58

59 Mammography and Tomosynthesis Dosimetry Mammography Why measure breast dose? Basic Concepts of Breast Dosimetry (how) Mean Glandular Dose (MGD) (what) DgN coefficients Skin Thickness Issues Breast Density Issues Tomosynthesis Differences between tomo and mammo Summary 59

60 Tomosynthesis (limited angle tomography) 60

61 61

62 Conventional versus Tomosynthesis Conventional Tomosynthesis Conventional Tomosynthesis Similar display characteristics 62

63 Tomosynthesis (limited angle tomography) ±7.5 63

64 Combo Tomo / Mammo mode output mammography mr/mas or mgy per 100 mas tomosynthesis 64

65 65

66 Physical measurements of dose in Tomo/Mammo Polyethylene is an excellent surrogate for adipose 66

67 Output of real time exposure meter 15 tomo exposures mammo exposure filter changes & grid is inserted 67

68 Point dose measures: Tissue Air Ratios 68

69 Mammography and Tomosynthesis Dosimetry Mammography Why measure breast dose? Basic Concepts of Breast Dosimetry (how) Mean Glandular Dose (MGD) (what) DgN coefficients Skin Thickness Issues Breast Density Issues Tomosynthesis Differences between tomo and mammo Summary 69

70 Mammography and Tomosynthesis Dosimetry Summary Complete characterization of the x-ray system is necessary (kv accuracy, HVL, and air kerma / 100 anode/filter) Ionization chambers must be mammo beam compatible thin windowed and calibrated Practical Assessment requires table of DgN values specific to the conditions of the actual exam Current DgN tables may slightly under-estimate dose due to % glandular fraction and skin thickness issues Tomosynthesis is similar enough to the geometry of mammo anode/filter combo may differ; different DgN tables needed 70

71 71

Breast CT and Dosimetry

Breast CT and Dosimetry 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of

More information

Skin Model and its impact on Digital Mammography

Skin Model and its impact on Digital Mammography Skin Model and its impact on Digital Mammography Rodrigo T. Massera; Alessandra Tomal Institute of Physics "Gleb Wataghin University of Campinas Campinas, Brazil 1 Outline Motivation Methodology Results

More information

Dosimetry in digital mammography

Dosimetry in digital mammography Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol

More information

for Mammography Dosimetry

for Mammography Dosimetry Changing Perceptions and Updated Methods for Mammography Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR, FAIMBE Professor of Radiology & Biomedical Engineering University of California Davis Sacramento,

More information

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Radiation Dosimetry in Digital Breast Tomosynthesis March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Syllabus 1. Introduction 2. Dosimetry in Mammography 3. Dosimetry

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose Academic Journal of Cancer Research 7 (2): 162-167, 2014 ISSN 1995-8943 IDOSI Publications, 2014 DOI: 10.5829/idosi.ajcr.2014.7.2.83313 Comparison Between Film-Screen and Digital Mammography for Woman

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography

More information

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,

More information

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2

More information

Direct half value layer measurements in mammography - is near enough good enough?

Direct half value layer measurements in mammography - is near enough good enough? Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.

More information

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro 3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:

More information

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in

More information

Contrast Enhanced Spectral Mammography (CESM) Updates

Contrast Enhanced Spectral Mammography (CESM) Updates Contrast Enhanced Spectral Mammography (CESM) Updates Georgeta Mihai, PhD, DABR Medical Physicist, BIDMC, Boston Assistant Professor, Harvard Medical School, Boston Disclosures None Acknowledgments: Da

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree

8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree DBT Physics Basic to Advanced: Primer On Tomosynthesis Andrew D. A. Maidment, Ph.D. University of Pennsylvania Department of Radiology Acknowledgements of Support Research support from the Komen Foundation,

More information

A tissue-equivalent phantom series for mammography dosimetry

A tissue-equivalent phantom series for mammography dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOL. 5, NO. 4, FALL 2004 A tissue-equivalent phantom series for mammography dosimetry William P. Argo, 1 Kathleen Hintenlang, 2 and David E. Hintenlang 3 U.S.

More information

Implementation of a New Tomosynthesis Program: A Physicists Perspective

Implementation of a New Tomosynthesis Program: A Physicists Perspective Implementation of a New Tomosynthesis Program: A Physicists Perspective Bill Geiser, MS DABR Senior Medical Physicist wgeiser@mdanderson.org 1 Conflict of Interest None of the authors nor their immediate

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi

More information

Breast dosimetry system in screen/film mammography

Breast dosimetry system in screen/film mammography Bull Fae Health Sei, Okayama Univ Med Seh, 10: 99-106, 2000 (Original article) Breast dosimetry system in screen/film mammography Sachiko GOTO, Yoshiharu AZUMA, Toshinori MARUYAMA, Yoshitada NAKAGIRI,

More information

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY

More information

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography

Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography Diagn Interv Radiol 2007; 13:134 139 Turkish Society of Radiology 2007 BREAST IMAGING ORIGINAL ARTICLE Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography

More information

Digital Breast Tomosynthesis from a first idea to clinical routine

Digital Breast Tomosynthesis from a first idea to clinical routine International Master Programm Biomedical Engineering Digital Breast Tomosynthesis from a first idea to clinical routine Historical background 2D imaging of 3D objects has important limitations Jörg Barkhausen

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

Mammography calibration qualities establishment in a Mo- Mo clinical system

Mammography calibration qualities establishment in a Mo- Mo clinical system Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733

More information

Mdiagnosing breast by using x-ray to both breasts. It

Mdiagnosing breast by using x-ray to both breasts. It Asian Journal of Medical and Clinical Sciences Estimation radiation risk during mammography in Sudan 1 1 1, Hifaa.Mohamed Khair, Hiba.Osman, Abdelmoneim Sulieman* 1 Al Neelain University, Faculty of Science

More information

Contrast-Enhanced Digital Mammography

Contrast-Enhanced Digital Mammography 2015 ARRS Breast Symposium Contrast-Enhanced Digital Mammography John Lewin, M.D. Diversified Radiology of Colorado CEDM - Outline History Technique Literature Review / Cases Clinical Status Inexpensive,

More information

IAEA TECDOC 1517 Quality Control in Mammography Software

IAEA TECDOC 1517 Quality Control in Mammography Software IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua.

More information

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Dosimetric characterization of surface applicators for use with the Xoft ebx system Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal

More information

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA Facility Certification Extension Requirements, Quality Assurance and Medical Physicists role for Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT) System Kish chakrabarti, Ph.D. Senior Physicist

More information

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose. Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives

More information

Since its introduction in 2000, digital mammography has become

Since its introduction in 2000, digital mammography has become Review Article Smith A, PhD email : Andrew.smith@hologic.com Since its introduction in 2000, digital mammography has become an accepted standard of care in breast cancer screening and has paved the way

More information

The relationship of the mean glandular dose with compressed breast thickness in mammography

The relationship of the mean glandular dose with compressed breast thickness in mammography Review Article Page 1 of 8 The relationship of the mean glandular dose with compressed breast thickness in mammography Xiang Du, Ningle Yu, Yimei Zhang, Jin Wang Department of Radiation Protection, Jiangsu

More information

Breast Tomosynthesis An additional screening tool in the fight against breast cancer

Breast Tomosynthesis An additional screening tool in the fight against breast cancer What to Expect Breast Tomosynthesis An additional screening tool in the fight against breast cancer Every woman over 40 should be examined for breast cancer once a year. American Cancer Society What to

More information

WHAT TO EXPECT. Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC. The Women's Health Company

WHAT TO EXPECT. Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC. The Women's Health Company WHAT TO EXPECT Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC The Women's Health Company ...,. Screening for breast cancer Doctors and scientists agree that

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

TOMOSYNTHESIS: WORTH ALL THE HYPE?

TOMOSYNTHESIS: WORTH ALL THE HYPE? X-Ray Associates of New Mexico, P.C. TOMOSYNTHESIS: WORTH ALL THE HYPE? MICHAEL N. LINVER, MD, FACR MAMMOGRAPHY: THE GOOD, THE PRETTY GOOD, & THE NOT SO GOOD MAMMOGRAPHY: THE GOOD, THE PRETTY GOOD, & THE

More information

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD Update of Digital Breast Tomosynthesis Susan Orel Roth, MD NCI estimates that : Why DBT? Approximately 20% of breast cancers are missed at mammography screening Average recall rates approximately 10%

More information

Effect of thyroid shielding during mammography: measurements on phantom and patient as well as estimation with Monte Carlo simulation

Effect of thyroid shielding during mammography: measurements on phantom and patient as well as estimation with Monte Carlo simulation Pyka et al. European Radiology Experimental (2018) 2:14 https://doi.org/10.1186/s41747-018-0042-9 European Radiology Experimental ORIGINAL ARTICLE Effect of thyroid shielding during mammography: measurements

More information

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz

More information

CASE STUDIES. Population. Compression

CASE STUDIES. Population. Compression CASE STUDIES > Critical Insight for Breast Imaging Quality and Workflow VolparaAnalytics provides key mammography metrics to support breast imaging centers in delivering high quality breast screening services

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

State of the art and future development for standardized estimation of organ doses in CT

State of the art and future development for standardized estimation of organ doses in CT State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida

More information

Outline. Digital Breast Tomosynthesis: Update and Pearls for Implementation. Tomosynthesis Dataset: 2D/3D (Hologic Combo Acquisition)

Outline. Digital Breast Tomosynthesis: Update and Pearls for Implementation. Tomosynthesis Dataset: 2D/3D (Hologic Combo Acquisition) Outline Digital Breast Tomosynthesis (DBT) the new standard of care Digital Breast Tomosynthesis: Update and Pearls for Implementation Emily F. Conant, M.D. Professor, Chief of Breast Imaging Department

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Do women with dense breasts have higher radiation dose during screening mammography?

Do women with dense breasts have higher radiation dose during screening mammography? Received: 8 June 6 Revised: 6 September 6 Accepted: 9 September 6 DOI:./tbj.8 ORIGINAL ARTICLE Do women with dense breasts have higher radiation dose during screening mammography? Jonathan V. Nguyen MD

More information

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates Patient Dose Estimates from CT Examinations Patient Dose Estimates from CT Examinations John M. Boone, Ph.D. Professor and Vice Chairman of Radiology Professor of Biomedical Engineering University of California,

More information

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative

More information

#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? TERMS (AKA) WHAT IS TOMOSYNTHESIS? 3/3/2014. Digital breast tomosynthesis =

#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? TERMS (AKA) WHAT IS TOMOSYNTHESIS? 3/3/2014. Digital breast tomosynthesis = #46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? January K. Lopez, MD Hoag Breast Care Center Newport Beach, CA Disclosures: None TERMS (AKA) Digital breast tomosynthesis = DBT Tomo 3D Full

More information

Better View. Getting a. Plus. enters new phase of clinical testing. Conducting Employee Surveys Radiology Master s Program Ground Transients

Better View. Getting a. Plus. enters new phase of clinical testing. Conducting Employee Surveys Radiology Master s Program Ground Transients RT awardwinning magazine the weekly source for radiology professionals Getting a Better View Dedicated breast CT enters new phase of clinical testing Plus Conducting Employee Surveys Radiology Master s

More information

Patient Specific Radiation Doses From Projection Radiography Images: Improving the Accuracy of Low Dose Radiation Calculations

Patient Specific Radiation Doses From Projection Radiography Images: Improving the Accuracy of Low Dose Radiation Calculations University of New Mexico UNM Digital Repository Nuclear Engineering ETDs Engineering ETDs Spring 4-9-2018 Patient Specific Radiation Doses From Projection Radiography Images: Improving the Accuracy of

More information

Calibration coefficients of dosimeters used in mammography for various target/filter combinations

Calibration coefficients of dosimeters used in mammography for various target/filter combinations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Calibration coefficients of dosimeters used in mammography for various target/filter combinations Michiharu Sekimoto, 1a Yoh Katoh,

More information

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS ABBY DUROJAYE,M.D CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS BOTH HAVE BEEN SHOWN TO BE EFFECTIVE TOOLS EARLY DETECTION OF BREAST

More information

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical

More information

CT Dosimetry in the Clinical Environment: Methods and Analysis

CT Dosimetry in the Clinical Environment: Methods and Analysis CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida

More information

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Learning Objectives: Limitations for estimating patient dose for CT Methods

More information

Breast Tomosynthesis. What is breast tomosynthesis?

Breast Tomosynthesis. What is breast tomosynthesis? Scan for mobile link. Breast Tomosynthesis Breast tomosynthesis is an advanced form of mammography, a specific type of breast imaging that uses low-dose x-rays to detect cancer early when it is most treatable.

More information

Innovations and Applications of Tomosynthesis. Andrew D. A. Maidment, Ph.D. University of Pennsylvania Department of Radiology

Innovations and Applications of Tomosynthesis. Andrew D. A. Maidment, Ph.D. University of Pennsylvania Department of Radiology Innovations and Applications of Tomosynthesis Andrew D. A. Maidment, Ph.D. University of Pennsylvania Department of Radiology Acknowledgements of Support Grant support from the Komen Foundation, DOD, NIH,

More information

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology Iran. J. Radiat. Res., 2006; 4 (1): 21-27 Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology M.T. Bahreyni

More information

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system Original article NOWOTWORY Journal of Oncology 2017, volume 67, number 2, 115 120 DOI: 10.5603/NJO.2017.0018 Polskie Towarzystwo Onkologiczne ISSN 0029 540X www.nowotwory.edu.pl Mean glandular dose values

More information

Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis

Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis J. Antunes 1,2, J. Machado 1,2, L. Peralta 1,2, N. Matela 1,3 1 Departamento de Física da Faculdade de Ciências da Universidade

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

Module Rhodes

Module Rhodes Module 6 10-526-197 Rhodes Health Physicist Concerned with providing occupation radiation protection and minimizing radiation dose to the public. Diagnostic Imaging has changed our world Live longer Work

More information

Financial Disclosures

Financial Disclosures Financial Disclosures 3D Mammography: The Latest Developments in the Breast Imaging Arena I have no financial disclosures Dr. Katharine Lampen-Sachar Breast and Body Radiologist Radiology Associates of

More information

What s New in Breast Imaging. Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia

What s New in Breast Imaging. Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia What s New in Breast Imaging Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia Disclosure Hologic, Inc. Shareholder and research agreement. Volpara Solutions, Ltd. Shareholder

More information

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences 1313 C. Kimme-Smith1 J. Wang2 N. DeBruhl1 M. Basic2 L. W. Bassett1 Received October 27, 1 993; accepted after revision January 25, 1994. Presented at the annual meeting of the American Roentgen Ray Society,

More information

Breast-density measurement using photon-counting spectral mammography

Breast-density measurement using photon-counting spectral mammography Breast-density measurement using photon-counting spectral mammography Henrik Johansson and Miriam von Tiedemann Philips Health Systems, Mammography Solutions, Torshamnsgatan 30A, 164 40 Kista, Sweden Klaus

More information

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations Nadia Alatta 1, Abdelrahaman Elnour 2, Ikhlas Abdelaziz 3, Tugua Tageldeen 4, Duha Abdu 5, Asma Elamin

More information

UNIFORMED SERVICES UNIVERSITY OF THE HEALTH SCIENCES 4301 JONES BRIDGE ROAD BETHESDA, MARYLAND JUNE 2003 APPROVAL SHEET

UNIFORMED SERVICES UNIVERSITY OF THE HEALTH SCIENCES 4301 JONES BRIDGE ROAD BETHESDA, MARYLAND JUNE 2003 APPROVAL SHEET UNIFORMED SERVICES UNIVERSITY OF THE HEALTH SCIENCES 4301 JONES BRIDGE ROAD BETHESDA, MARYLAND 20814-4799 18 JUNE 2003 APPROVAL SHEET Title of Thesis: " Digital Mammography Breast Dosimetry Using Copper-Doped

More information

EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME

EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME S. Ferrer (), M Ramos (), J.I. Villaescusa (), G Verdú (), M.D. Salas (3), M.D. Cuevas (4) () Departamento

More information

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI john.sandrik@med.ge.com with many thanks to Vince Polkus, Advanced Applications Product Mgr. 1 Content

More information

Accurate Accurate boost or Simply Accuboost

Accurate Accurate boost or Simply Accuboost Accurate Accurate boost or Simply Accuboost Zoubir Ouhib MS DABR Lynn Cancer Institute of Boca Raton Community Hospital Disclosure: Advisory board Items to be discussed Big picture on management of Breast

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Radiographic techniques in screen-film mammography

Radiographic techniques in screen-film mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 3, SUMMER 2002 Radiographic techniques in screen-film mammography Thomas R. LaVoy* 4314 Kelsey Drive, Syracuse, New York 13215 Walter Huda

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

2001 AAPM Summer School Seattle, Washington ACR R/F Phantom

2001 AAPM Summer School Seattle, Washington ACR R/F Phantom 2001 AAPM Summer School Seattle, Washington ACR R/F Phantom Charles R. Wilson, Ph.D., FACR Medical College of Wisconsin Milwaukee, Wisconsin MCMC Circa 1975 ACR R/F Phantom This lecture has been approved

More information

ALARA in mammography screening. Hilde Bosmans et al.

ALARA in mammography screening. Hilde Bosmans et al. ALARA in mammography screening Hilde Bosmans et al. ALARA in the mammography Overview 1. Screening 2. The balance: 1. Image quality 2. Radiation dose 3. In practice Screening leeftijd Mammography Screening

More information

THE DIAGNOSTIC WORKUP: THE TEAM APPROACH

THE DIAGNOSTIC WORKUP: THE TEAM APPROACH X-Ray Associates of New Mexico, P.C. THE DIAGNOSTIC WORKUP: THE TEAM APPROACH MICHAEL N. LINVER, MD, FACR DAWN DERENBURGER, RTRM Disclosure There are no conflicts of interest or relevant financial interests

More information

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations Aiping Ding, X. George Xu Rensselaer Polytechnic Institute Troy, NY USA http://rrmdg.rpi.edu 1 Acknowledgements

More information

Lifetime risk of radiation-induced cancer from screening mammography

Lifetime risk of radiation-induced cancer from screening mammography Lifetime risk of radiation-induced cancer from screening mammography Poster No.: C-0558 Congress: ECR 2015 Type: Scientific Exhibit Authors: R. M. K. M.Ali, A. England, P. Hogg; Manchester/UK Keywords:

More information

Managing the imaging dose during image-guided radiation therapy

Managing the imaging dose during image-guided radiation therapy Managing the imaging dose during image-guided radiation therapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Richmond VA Imaging during radiotherapy Radiographic

More information

Development of a program for estimation of patient exposure in mammography

Development of a program for estimation of patient exposure in mammography Developmt of a program for estimation of patit exposure in mammography Y. Asada 1, S. Suzuki, M. Kobayashi, S. Koga 1 Faculty of radiological technology, Fujita Health University, School of Health Scices,

More information

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview New ACR Activities Requirements

More information

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality.

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality. Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality. Poster No.: C-1636 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit

More information

ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΜΕΛΕΤΗ ΑΠΟΡΡΟΦΩΜΕΝΗΣ ΟΣΗΣ ΣΕ ΜΕΓΕΘΥΝΤΙΚΕΣ ΛΗΨΕΙΣ ΣΤΗ ΜΑΣΤΟΓΡΑΦΙΑ ΜΕ ΤΕΧΝΙΚΕΣ ΠΡΟΣΟΜΟΙΩΣΗΣ MONTE CARLO

ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΜΕΛΕΤΗ ΑΠΟΡΡΟΦΩΜΕΝΗΣ ΟΣΗΣ ΣΕ ΜΕΓΕΘΥΝΤΙΚΕΣ ΛΗΨΕΙΣ ΣΤΗ ΜΑΣΤΟΓΡΑΦΙΑ ΜΕ ΤΕΧΝΙΚΕΣ ΠΡΟΣΟΜΟΙΩΣΗΣ MONTE CARLO ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΤΜΗΜΑ ΙΑΤΡΙΚΗΣ ΤΜΗΜΑ ΦΥΣΙΚΗΣ ΙΑΤΜΗΜΑΤΙΚΟ ΠΡΟΓΡΑΜΜΑ ΜΕΤΑΠΤΥΧΙΑΚΩΝ ΣΠΟΥ ΩΝ ΣΤΗΝ ΙΑΤΡΙΚΗ ΦΥΣΙΚΗ ΜΕΛΕΤΗ ΑΠΟΡΡΟΦΩΜΕΝΗΣ ΟΣΗΣ ΣΕ ΜΕΓΕΘΥΝΤΙΚΕΣ ΛΗΨΕΙΣ ΣΤΗ ΜΑΣΤΟΓΡΑΦΙΑ ΜΕ ΤΕΧΝΙΚΕΣ ΠΡΟΣΟΜΟΙΩΣΗΣ

More information