Monte Carlo for CyberKnife Incise TM MLC

Size: px
Start display at page:

Download "Monte Carlo for CyberKnife Incise TM MLC"

Transcription

1 Division of Medical Radiation Physics Monte Carlo for CyberKnife Incise TM MLC Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital, Bern University Hospital, and University of Bern, Switzerland

2 Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation The views expressed in this presentation are those of the presenters and do not necessarily reflect the views or policies of Accuray Incorporated or its subsidiaries. No official endorsement by Accuray Incorporated or any of its subsidiaries of any vendor, products or services contained in this presentation is intended or should be inferred.

3 Outline CyberKnife M6 System at the Inselspital Bern Commissioning Evaluation In-house independent MC dose calculation (IDC) for InCise MLC Evaluation Benchmarking of MC dose calculation for InCise MLC Summary

4 CyberKnife M6 System at the Inselspital Bern 2014: Installation of CyberKnife M6 System. May 2014: Start of patient treatments. July 2015: Installation of InCise MLC. August 2017: CyberKnife -11 Upgrade (requirement for the use of Monte Carlo (MC) for the InCise MLC).

5 CyberKnife M6 System at the Inselspital Bern

6 Commissioning Finite Size Pencil Beam (FSPB) algorithm must be commissioned before starting with MC commissioning. No additional data/measurements required for the MC MLC algorithm. Estimated time needed for commissioning: 3-5 days.

7 Commissioning - Workflow FSPB MC START Define Source Model (FWHM, Energy) Calculate TPR Calculate OCR Calculate OF ok ok Inselspital Bern: FWHM = 2.0 mm Energy = 6.3 MeV Compare calc. vs meas. TPR Compare calc. vs meas. OCR Compare calc. vs meas. OF ok FINISH nok nok nok

8 Evaluation In-house independent MC dose calculation (IDC) for InCise MLC Geometry model of the CyberKnife System equipped with the InCise MLC Ref: P.-H. Mackeprang et al., "Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy", Phys. Med. Biol. 63, 2018.

9 Evaluation In-house independent MC dose calculation (IDC) for InCise MLC Model is validated with measurements. Ref: P.-H. Mackeprang et al., "Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy", Phys. Med. Biol. 63, 2018.

10 Evaluation In-house independent MC dose calculation (IDC) for InCise MLC An MC-based modular IDC-Framework was successfully developed, implemented and validated against measurements and is now available to perform patient-specific QA. Ref: P.-H. Mackeprang et al., "Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy", Phys. Med. Biol. 63, 2018.

11 Evaluation Benchmarking of MC dose calculation for InCise MLC IDC - Framework COMPARISON Precision Dose distributions in water 11 rectangular MLC field sizes TPR, OCR Dose distributions in lung phantom (CIRS Thorax Phantom) Single MLC field Dose distributions in patients 2 prostate cases 7 lung cases Calculation time

12 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in water -> OCR x mm 2 Framework and Precision depth = 1.5 cm Agreement between IDC 7.6 x 7.7 mm 2 within 3.1% / 1 mm x 46.2 mm 2 Measurement xxx IDC Framework Precision

13 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in water -> TPR Measurement xxx IDC Framework Precision Agreement between IDC Framework and Precision within 2.3% / 1 mm x mm x 46.2 mm x 7.7 mm 2

14 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in lung phantom (CIRS Thorax Phantom) Precision IDC Framework Dose difference

15 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in patients (lung case) Dose difference Gamma (2% / 1 mm)

16 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in patients

17 Evaluation Benchmarking of MC dose calculation for InCise MLC Calculation time

18 Evaluation Benchmarking of MC dose calculation for InCise MLC

19 Summary Once the FSPB algorithm is commissioned, no additional beam data measurements are needed in order to commission the MC algorithm for the InCise MLC. An in-house independent dose calculation framework (IDC- Framework) was developed and validated against measurements. The accuracy of the MC algorithm within Precision was evaluated by comparing it with the IDC-Framework. The calculation time for a clinically reasonable dose distribution with Monte Carlo is typically below one hour.

20 Thank you for your attention!

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

It s All About Margins. Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018

It s All About Margins. Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018 It s All About Margins Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018 Why margins? The smaller the better! Short Introduction Erasmus MC has been using the CyberKnife Robotic Radiosurgery System

More information

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Independent Dose Verification for IMRT Using Monte Carlo C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Outline Why Monte Carlo for IMRT QA? Experimental verification

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

Review of TG-186 recommendations

Review of TG-186 recommendations Review of TG-186 recommendations Implementation of advanced brachytherapy dose calculation algorithms beyond TG-43 Rowan M. Thomson Carleton Laboratory for Radiotherapy Physics Carleton University Ottawa

More information

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with

More information

The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist

The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist The Radixact System Experience at Miami Cancer Institute Tino Romaguera, D.Sc. Senior Physicist Disclosure An honorarium is provided by Accuray for this presentation The views expressed in this presentation

More information

Clinical experience with TomoDirect System Tangential Mode

Clinical experience with TomoDirect System Tangential Mode Breast Cancer Clinical experience with TomoDirect System Tangential Mode European Institute of Oncology Milan, Italy Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation The

More information

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson

More information

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning

Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning C-M M Charlie Ma, Ph.D. Department of Radiation Oncology Fox Chase Cancer

More information

The TomoTherapy System as a Tool of Differentiation in Quality and Marketability

The TomoTherapy System as a Tool of Differentiation in Quality and Marketability The TomoTherapy System as a Tool of Differentiation in Quality and Marketability John J. Kresl, MD, PhD, FACRO, FACR Medical Director Phoenix CyberKnife & Radiation Oncology Center Managing Partner Phoenix,

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a The Equipment : safety, commissioning, re commissioning calibration : Equipements : sûreté, recette, modification calibration A medical physicist point of view. Albert Lisbona Medical Physics Department

More information

Basic dosimetric data for treatment planning system

Basic dosimetric data for treatment planning system Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny

More information

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an

More information

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS} High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics

More information

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms 1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,

More information

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Precisely Maximize Dose, side effects, Patient

Precisely Maximize Dose, side effects, Patient Precisely Maximize Dose, Minimize side effects, Maximize Patient Comfort Key Differentiators: The CyberKnife M6 Series Industry-leading precision and patient comfort Enables full-body robotic radiosurgery,

More information

CYBERKNIFE M6 SERIES: Unmatched Precision and Patient Comfort

CYBERKNIFE M6 SERIES: Unmatched Precision and Patient Comfort CYBERKNIFE M6 SERIES: Unmatched Precision and Patient Comfort Precisely maximize dose, minimize side effects and maximize patient comfort The new CyberKnife M6 Series has the capabilities and efficiency

More information

Imaging of Radiation Dose Using Cherenkov Light

Imaging of Radiation Dose Using Cherenkov Light Imaging of Radiation Dose Using Cherenkov Light Eric Brost 1, Yoichi Watanabe 1, Fadil Santosa 2, Adam Green 3 1 Department of Radiation Oncology, University of Minnesota 2 Institute for Mathematics and

More information

Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy

Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy Francis Kar-ho Lee, a Simon Kar-yiu

More information

Dosimetry of the Gamma Knife TM

Dosimetry of the Gamma Knife TM Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

One stop shop. Natalie D. Klass, MD

One stop shop. Natalie D. Klass, MD One stop shop Natalie D. Klass, MD AERO Academy Conference Innovation & Safety Lisbon 26th and 27th January 2018 Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation The views

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

Vertebral Body Compression Fracture Following Spine SBRT

Vertebral Body Compression Fracture Following Spine SBRT RADIATION ONCOLOGY & MOLECULAR RADIATION SCIENCES Vertebral Body Compression Fracture Following Spine SBRT Kristin J. Redmond, MD, MPH Disclosure & Disclaimer An honorarium is provided by Accuray for this

More information

Strengthening the weakest link: improving tumour definition

Strengthening the weakest link: improving tumour definition Strengthening the weakest link: improving tumour definition Marianne C Aznar marianne.aznar@manchester.ac.uk University of Manchester / The Christie NHS Clinical Trial Service Unit, University of Oxford

More information

The PreciseART Approach to Adaptive Radiotherapy with the RADIXACT System. Prof. Anne Laprie Radiation Oncologist

The PreciseART Approach to Adaptive Radiotherapy with the RADIXACT System. Prof. Anne Laprie Radiation Oncologist The PreciseART Approach to Adaptive Radiotherapy with the RADIXACT System Prof. Anne Laprie Radiation Oncologist Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation The views

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

SBRT Credentialing: Understanding the Process from Inquiry to Approval

SBRT Credentialing: Understanding the Process from Inquiry to Approval SBRT Credentialing: Understanding the Process from Inquiry to Approval David Followill, Ph.D. (presented by Andrea Molineu, M.S.) IROC Houston QA Center July 15, 2015 What is credentialing? Verification

More information

Fiducial-Free Lung Tracking and Treatment with the CyberKnife System: A Non-Invasive Approach

Fiducial-Free Lung Tracking and Treatment with the CyberKnife System: A Non-Invasive Approach Fiducial-Free Lung Tracking and Treatment with the CyberKnife System: A Non-Invasive Approach Jesse McKay, MS, DABR Erlanger Health System Chattanooga, TN JASTRO 2014, Yokohama Japan Disclosure I have

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

Treating Multiple. Brain Metastases (BM)

Treating Multiple. Brain Metastases (BM) ESTRO 36 5-9 May 2017, Vienna Austria, Accuray Symposium Treating Multiple Brain Metastases (BM) with CyberKnife System Frederic Dhermain MD PhD, Radiation Oncologist Gustave Roussy University Hospital,

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS,

Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS, Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at, 2005-2007 1. Introduction Gerhard Stucki, Solange Gagnebin, Damian Twerenbold, Sándor Vörös

More information

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...

More information

Benchmarking treatment planning dose calculations using a Monte Carlo dose engine as a potential activity of a primary standard lab

Benchmarking treatment planning dose calculations using a Monte Carlo dose engine as a potential activity of a primary standard lab Benchmarking treatment planning dose calculations using a Monte Carlo dose engine as a potential activity of a primary standard lab Nick Reynaert (LSDG, Belgium/Erasmus MC, The Netherlands) Hugo Palmans

More information

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery

More information

typical IMRT fraction time and expand high definition radiotherapy anywhere in the body with the widest range of motion of the

typical IMRT fraction time and expand high definition radiotherapy anywhere in the body with the widest range of motion of the Precisely maximize dose, Versatile, efficient and effective for the BENEFITS OFandTHE CYBERKNIFE TREATMENT minimize side effects range of radiation oncology patients maximize patient comfort DELIVERY SYSTEM

More information

PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT

PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT BENEFITS OF THE CYBERKNIFE TREATMENT DELIVERY SYSTEM True robotic precision: Enable high definition radiotherapy anywhere in the body with the

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

Novel techniques for normal tissue toxicity modelling. Laura Cella Institute of Biostructures and Bioimaging National Research Council of Italy

Novel techniques for normal tissue toxicity modelling. Laura Cella Institute of Biostructures and Bioimaging National Research Council of Italy Novel techniques for normal tissue toxicity modelling Laura Cella Institute of Biostructures and Bioimaging National Research Council of Italy Disclosure: I have no conflict of interest An honorarium is

More information

Heterogeneity Corrections in Clinical Trials

Heterogeneity Corrections in Clinical Trials Heterogeneity Corrections in Clinical Trials Niko Papanikolaou Departments of Radiology and Radiation Oncology U of TX Health Science Center, San Antonio, TX Eric Klein Department of Radiation Oncology

More information

Immunotherapy & radiotherapy

Immunotherapy & radiotherapy Immunotherapy & radiotherapy Eric Deutsch MD PhD Gustave Roussy Cancer Campus Image courtesy of Gustave Roussy Institute Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation

More information

Planning for the Installation of a Compact Superconducting Proton Therapy Unit in a Busy Medical Center Environment

Planning for the Installation of a Compact Superconducting Proton Therapy Unit in a Busy Medical Center Environment Planning for the Installation of a Compact Superconducting Proton Therapy Unit in a Busy Medical Center Environment Barry W. Wessels, Ph.D. Director of Medical Physics and Dosimetry Case Western Reserve

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Transition to Heterogeneity Corrections. Why have accurate dose algorithms? Transition to Heterogeneity Corrections Eric E. Klein, M.S., Washington University, St. Louis, MO Craig Stevens, M.D., Ph.D., MD Anderson Cancer Center, Houston, TX Nikos Papinikolou, Ph.D., University

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom Iran. J. Radiat. Res., 2003; 1(3): 143-149 Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom A. Mostaar 1, M. Allahverdi 1,2, M. Shahriari 3 1 Medical Physics Department,

More information

New Thinking on Fractionation in Radiotherapy

New Thinking on Fractionation in Radiotherapy New Thinking on Fractionation in Radiotherapy Alan E. Nahum Visiting Professor, Physics dept., Liverpool university, UK alan_e_nahum@yahoo.co.uk 1 An honorarium is provided by Accuray for this presentation

More information

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad

More information

Effective and efficient radiotherapy dosimetry audit: Where to next?

Effective and efficient radiotherapy dosimetry audit: Where to next? Effective and efficient radiotherapy dosimetry audit: Where to next? Catharine Clark, Eduard Gershkevitsh, Milan Tomsej and Nuria Jornet National Physical Laboratory Royal Surrey County Hospital Radiotherapy

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

Dosimetric Comparison: Acuros XB vs AAA

Dosimetric Comparison: Acuros XB vs AAA Dosimetric Comparison: Acuros XB vs AAA Leo El Hage, Chris James, Jim Frantzis, Roger Li, Venkata Sheshadri Epworth Radiation Oncology Epworth HealthCare Excellence. Everywhere. Everyday Background Epworth

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

Integrating the Radixact into a Highly Comprehensive Center

Integrating the Radixact into a Highly Comprehensive Center Integrating the Radixact into a Highly Comprehensive Center Alonso N. Gutiérrez, PhD, MBA ASTRO Meeting 2018 Disclosures An honorarium is provided by Accuray for this presentation (donated) Accuray and

More information

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.

More information

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D. UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems

More information

SRS Uncertainty: Linac and CyberKnife Uncertainties

SRS Uncertainty: Linac and CyberKnife Uncertainties SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis

More information

Clinical Implementation of SRS/SBRT

Clinical Implementation of SRS/SBRT Clinical Implementation of SRS/SBRT Anil Sethi, PhD, FAAPM Loyola University Medical Center November 4, 2017 Disclosures Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch 2 Learning

More information

Electron Beam ET - Reloaded Therapy - Reloaded

Electron Beam ET - Reloaded Therapy - Reloaded 8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles

More information

Accuracy of the Small Field Dosimetry Using the Acuros XB Dose Calculation Algorithm within and beyond Heterogeneous Media for 6 MV Photon Beams *

Accuracy of the Small Field Dosimetry Using the Acuros XB Dose Calculation Algorithm within and beyond Heterogeneous Media for 6 MV Photon Beams * International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2012, 1, 78-87 Published Online November 2012 (http://www.scirp.org/journal/ijmpcero) http://dx.doi.org/10.4236/ijmpcero.2012.13011

More information

S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M

S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M s y n c h r o n y r e s p i r a t o r y t r a c k i n g s y s t e m S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M The Synchrony System tracks respiration in real time and automatically

More information

02 CyberKnife: Treatment Delivery

02 CyberKnife: Treatment Delivery TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options

More information

CyberKnife SBRT for Prostate Cancer

CyberKnife SBRT for Prostate Cancer CyberKnife SBRT for Prostate Cancer Robert Meier, MD Swedish Radiosurgery Center Swedish Cancer Institute Seattle, WA 2017 ESTRO Meeting, Vienna Austria 5-year safety, efficacy & quality of life outcomes

More information

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?

More information

Advanced Tumor Treatment

Advanced Tumor Treatment Advanced Tumor Treatment What is CyberKnife? The CyberKnife Stereotactic Radiosurgery System, offered by Southwest Washington Medical Center, is a noninvasive outpatient treatment without the risks and

More information

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser. Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic

More information

Commissioning of brachytherapy Treatment Planning Systems

Commissioning of brachytherapy Treatment Planning Systems Commissioning of brachytherapy Treatment Planning Systems Jacco L.G. Steenhuijsen, PhD Catharina Hospital Eindhoven Landelijke brachydag, 17-11-17 Brachytherapy Treatment Planning Systems 2 Brachytherapy

More information

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,

More information

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou. Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution

More information

Hampton University Proton Therapy Institute

Hampton University Proton Therapy Institute Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI

More information

A review of RTTQA audit. Karen Venables

A review of RTTQA audit. Karen Venables A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines

More information

Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy

Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy Rasha Makkia Biomedical Physics Ph.D. student East Carolina University March 3 rd, 2016 Outlines The Purpose Wilms Tumor,

More information

Verification of Advanced Radiotherapy Techniques

Verification of Advanced Radiotherapy Techniques Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December

More information

Analytical Methods for Normal- Tissue Dose Reconstructions. Acknowledgments. Disclosures. Learning Objectives 8/2/2017. Wayne Newhauser, PhD

Analytical Methods for Normal- Tissue Dose Reconstructions. Acknowledgments. Disclosures. Learning Objectives 8/2/2017. Wayne Newhauser, PhD Analytical Methods for Normal- Tissue Dose Reconstructions Wayne Newhauser, PhD newhauser@lsu.edu AAPM Annual Meeting, August 2, 2017 Acknowledgments Bella Bowman Foundation Numerous current and former

More information

PSMMP PUBLICATIONS 2016

PSMMP PUBLICATIONS 2016 PSMMP PUBLICATIONS 2016 Dosimetric and radiobiological comparison of CyberKnife M6 InCise multileaf collimator over IRIS variable collimator in prostate stereotactic body radiation therapy, Vindu Kathriarachchi,

More information

Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations

Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife State-of of-the-art and New Innovations Chad Lee, PhD CK Solutions, Inc. and CyberKnife Centers of San Diego Outline Basic overview

More information

New CyberKnife M6 Service at HKSH for Cancer Patients Shorter Treatment Time, Fewer Side Effects and Improved Patient Outcomes

New CyberKnife M6 Service at HKSH for Cancer Patients Shorter Treatment Time, Fewer Side Effects and Improved Patient Outcomes For Immediate Release New CyberKnife M6 Service at HKSH for Cancer Patients Shorter Treatment Time, Fewer Side Effects and Improved Patient Outcomes (13 August 2017, Hong Kong) Hong Kong Sanatorium & Hospital

More information

Journal extended Abstracts FL-AAPM Spring Meeting April 28-30, 2017, Orlando FL

Journal extended Abstracts FL-AAPM Spring Meeting April 28-30, 2017, Orlando FL PSMMP PUBLICATIONS 2017 A GPU accelerated simulation annealing algorithm for IMRT optimization, P. Galanakou, T. Leventouri, A. Georgakilas, G. Kalantzis, (IEEE proc. SNPD 2017). A Novel Test to Verify

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information