Monte Carlo for CyberKnife Incise TM MLC
|
|
- Lesley Atkinson
- 6 years ago
- Views:
Transcription
1 Division of Medical Radiation Physics Monte Carlo for CyberKnife Incise TM MLC Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital, Bern University Hospital, and University of Bern, Switzerland
2 Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation The views expressed in this presentation are those of the presenters and do not necessarily reflect the views or policies of Accuray Incorporated or its subsidiaries. No official endorsement by Accuray Incorporated or any of its subsidiaries of any vendor, products or services contained in this presentation is intended or should be inferred.
3 Outline CyberKnife M6 System at the Inselspital Bern Commissioning Evaluation In-house independent MC dose calculation (IDC) for InCise MLC Evaluation Benchmarking of MC dose calculation for InCise MLC Summary
4 CyberKnife M6 System at the Inselspital Bern 2014: Installation of CyberKnife M6 System. May 2014: Start of patient treatments. July 2015: Installation of InCise MLC. August 2017: CyberKnife -11 Upgrade (requirement for the use of Monte Carlo (MC) for the InCise MLC).
5 CyberKnife M6 System at the Inselspital Bern
6 Commissioning Finite Size Pencil Beam (FSPB) algorithm must be commissioned before starting with MC commissioning. No additional data/measurements required for the MC MLC algorithm. Estimated time needed for commissioning: 3-5 days.
7 Commissioning - Workflow FSPB MC START Define Source Model (FWHM, Energy) Calculate TPR Calculate OCR Calculate OF ok ok Inselspital Bern: FWHM = 2.0 mm Energy = 6.3 MeV Compare calc. vs meas. TPR Compare calc. vs meas. OCR Compare calc. vs meas. OF ok FINISH nok nok nok
8 Evaluation In-house independent MC dose calculation (IDC) for InCise MLC Geometry model of the CyberKnife System equipped with the InCise MLC Ref: P.-H. Mackeprang et al., "Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy", Phys. Med. Biol. 63, 2018.
9 Evaluation In-house independent MC dose calculation (IDC) for InCise MLC Model is validated with measurements. Ref: P.-H. Mackeprang et al., "Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy", Phys. Med. Biol. 63, 2018.
10 Evaluation In-house independent MC dose calculation (IDC) for InCise MLC An MC-based modular IDC-Framework was successfully developed, implemented and validated against measurements and is now available to perform patient-specific QA. Ref: P.-H. Mackeprang et al., "Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy", Phys. Med. Biol. 63, 2018.
11 Evaluation Benchmarking of MC dose calculation for InCise MLC IDC - Framework COMPARISON Precision Dose distributions in water 11 rectangular MLC field sizes TPR, OCR Dose distributions in lung phantom (CIRS Thorax Phantom) Single MLC field Dose distributions in patients 2 prostate cases 7 lung cases Calculation time
12 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in water -> OCR x mm 2 Framework and Precision depth = 1.5 cm Agreement between IDC 7.6 x 7.7 mm 2 within 3.1% / 1 mm x 46.2 mm 2 Measurement xxx IDC Framework Precision
13 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in water -> TPR Measurement xxx IDC Framework Precision Agreement between IDC Framework and Precision within 2.3% / 1 mm x mm x 46.2 mm x 7.7 mm 2
14 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in lung phantom (CIRS Thorax Phantom) Precision IDC Framework Dose difference
15 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in patients (lung case) Dose difference Gamma (2% / 1 mm)
16 Evaluation Benchmarking of MC dose calculation for InCise MLC Dose distributions in patients
17 Evaluation Benchmarking of MC dose calculation for InCise MLC Calculation time
18 Evaluation Benchmarking of MC dose calculation for InCise MLC
19 Summary Once the FSPB algorithm is commissioned, no additional beam data measurements are needed in order to commission the MC algorithm for the InCise MLC. An in-house independent dose calculation framework (IDC- Framework) was developed and validated against measurements. The accuracy of the MC algorithm within Precision was evaluated by comparing it with the IDC-Framework. The calculation time for a clinically reasonable dose distribution with Monte Carlo is typically below one hour.
20 Thank you for your attention!
The Journey of Cyberknife Commissioning
The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources
More informationIt s All About Margins. Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018
It s All About Margins Maaike Milder, Ph.D. Accuray Symposium April 21 st 2018 Why margins? The smaller the better! Short Introduction Erasmus MC has been using the CyberKnife Robotic Radiosurgery System
More informationIndependent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA
Independent Dose Verification for IMRT Using Monte Carlo C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Outline Why Monte Carlo for IMRT QA? Experimental verification
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationReview of TG-186 recommendations
Review of TG-186 recommendations Implementation of advanced brachytherapy dose calculation algorithms beyond TG-43 Rowan M. Thomson Carleton Laboratory for Radiotherapy Physics Carleton University Ottawa
More informationFilm-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom
Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with
More informationThe Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist
The Radixact System Experience at Miami Cancer Institute Tino Romaguera, D.Sc. Senior Physicist Disclosure An honorarium is provided by Accuray for this presentation The views expressed in this presentation
More informationClinical experience with TomoDirect System Tangential Mode
Breast Cancer Clinical experience with TomoDirect System Tangential Mode European Institute of Oncology Milan, Italy Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation The
More informationAssessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom
Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson
More informationMonte Carlo Dose Calculation for Radiotherapy Treatment Planning
Monte Carlo Dose Calculation for Radiotherapy Treatment Planning Monte Carlo Dose Calculation for Radiotherapy Treatment Planning C-M M Charlie Ma, Ph.D. Department of Radiation Oncology Fox Chase Cancer
More informationThe TomoTherapy System as a Tool of Differentiation in Quality and Marketability
The TomoTherapy System as a Tool of Differentiation in Quality and Marketability John J. Kresl, MD, PhD, FACRO, FACR Medical Director Phoenix CyberKnife & Radiation Oncology Center Managing Partner Phoenix,
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationQuality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose
Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationA. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.
Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1
More informationAlbert Lisbona Medical Physics Department CLCC Nantes Atlantique a
The Equipment : safety, commissioning, re commissioning calibration : Equipements : sûreté, recette, modification calibration A medical physicist point of view. Albert Lisbona Medical Physics Department
More informationBasic dosimetric data for treatment planning system
Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny
More informationThe RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff
The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an
More information8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}
High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics
More informationA model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms
1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,
More informationBorges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1
Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior
More informationOut-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients
EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More informationPrecisely Maximize Dose, side effects, Patient
Precisely Maximize Dose, Minimize side effects, Maximize Patient Comfort Key Differentiators: The CyberKnife M6 Series Industry-leading precision and patient comfort Enables full-body robotic radiosurgery,
More informationCYBERKNIFE M6 SERIES: Unmatched Precision and Patient Comfort
CYBERKNIFE M6 SERIES: Unmatched Precision and Patient Comfort Precisely maximize dose, minimize side effects and maximize patient comfort The new CyberKnife M6 Series has the capabilities and efficiency
More informationImaging of Radiation Dose Using Cherenkov Light
Imaging of Radiation Dose Using Cherenkov Light Eric Brost 1, Yoichi Watanabe 1, Fadil Santosa 2, Adam Green 3 1 Department of Radiation Oncology, University of Minnesota 2 Institute for Mathematics and
More informationDosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy Francis Kar-ho Lee, a Simon Kar-yiu
More informationDosimetry of the Gamma Knife TM
Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife
More informationHot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis
Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationOne stop shop. Natalie D. Klass, MD
One stop shop Natalie D. Klass, MD AERO Academy Conference Innovation & Safety Lisbon 26th and 27th January 2018 Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation The views
More informationLevel of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013
Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific
More informationVertebral Body Compression Fracture Following Spine SBRT
RADIATION ONCOLOGY & MOLECULAR RADIATION SCIENCES Vertebral Body Compression Fracture Following Spine SBRT Kristin J. Redmond, MD, MPH Disclosure & Disclaimer An honorarium is provided by Accuray for this
More informationStrengthening the weakest link: improving tumour definition
Strengthening the weakest link: improving tumour definition Marianne C Aznar marianne.aznar@manchester.ac.uk University of Manchester / The Christie NHS Clinical Trial Service Unit, University of Oxford
More informationThe PreciseART Approach to Adaptive Radiotherapy with the RADIXACT System. Prof. Anne Laprie Radiation Oncologist
The PreciseART Approach to Adaptive Radiotherapy with the RADIXACT System Prof. Anne Laprie Radiation Oncologist Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation The views
More informationImplementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy
Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out
More informationQuality assurance in external radiotherapy
Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance
More informationAAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction
AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall
More informationSBRT Credentialing: Understanding the Process from Inquiry to Approval
SBRT Credentialing: Understanding the Process from Inquiry to Approval David Followill, Ph.D. (presented by Andrea Molineu, M.S.) IROC Houston QA Center July 15, 2015 What is credentialing? Verification
More informationFiducial-Free Lung Tracking and Treatment with the CyberKnife System: A Non-Invasive Approach
Fiducial-Free Lung Tracking and Treatment with the CyberKnife System: A Non-Invasive Approach Jesse McKay, MS, DABR Erlanger Health System Chattanooga, TN JASTRO 2014, Yokohama Japan Disclosure I have
More informationIndependent corroboration of monitor unit calculations performed by a 3D computerized planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and
More informationTreating Multiple. Brain Metastases (BM)
ESTRO 36 5-9 May 2017, Vienna Austria, Accuray Symposium Treating Multiple Brain Metastases (BM) with CyberKnife System Frederic Dhermain MD PhD, Radiation Oncologist Gustave Roussy University Hospital,
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the
More informationSTEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS
PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic
More informationAn Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center
An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions
More informationReport to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS,
Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at, 2005-2007 1. Introduction Gerhard Stucki, Solange Gagnebin, Damian Twerenbold, Sándor Vörös
More informationGuidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy
Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...
More informationBenchmarking treatment planning dose calculations using a Monte Carlo dose engine as a potential activity of a primary standard lab
Benchmarking treatment planning dose calculations using a Monte Carlo dose engine as a potential activity of a primary standard lab Nick Reynaert (LSDG, Belgium/Erasmus MC, The Netherlands) Hugo Palmans
More informationQuality assurance and credentialing requirements for sites using inverse planned IMRT Techniques
TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery
More informationtypical IMRT fraction time and expand high definition radiotherapy anywhere in the body with the widest range of motion of the
Precisely maximize dose, Versatile, efficient and effective for the BENEFITS OFandTHE CYBERKNIFE TREATMENT minimize side effects range of radiation oncology patients maximize patient comfort DELIVERY SYSTEM
More informationPRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT
PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT BENEFITS OF THE CYBERKNIFE TREATMENT DELIVERY SYSTEM True robotic precision: Enable high definition radiotherapy anywhere in the body with the
More informationCALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE
CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of
More informationNovel techniques for normal tissue toxicity modelling. Laura Cella Institute of Biostructures and Bioimaging National Research Council of Italy
Novel techniques for normal tissue toxicity modelling Laura Cella Institute of Biostructures and Bioimaging National Research Council of Italy Disclosure: I have no conflict of interest An honorarium is
More informationHeterogeneity Corrections in Clinical Trials
Heterogeneity Corrections in Clinical Trials Niko Papanikolaou Departments of Radiology and Radiation Oncology U of TX Health Science Center, San Antonio, TX Eric Klein Department of Radiation Oncology
More informationImmunotherapy & radiotherapy
Immunotherapy & radiotherapy Eric Deutsch MD PhD Gustave Roussy Cancer Campus Image courtesy of Gustave Roussy Institute Disclosure & Disclaimer An honorarium is provided by Accuray for this presentation
More informationPlanning for the Installation of a Compact Superconducting Proton Therapy Unit in a Busy Medical Center Environment
Planning for the Installation of a Compact Superconducting Proton Therapy Unit in a Busy Medical Center Environment Barry W. Wessels, Ph.D. Director of Medical Physics and Dosimetry Case Western Reserve
More informationQA for Clinical Dosimetry with Emphasis on Clinical Trials
QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions
More informationTransition to Heterogeneity Corrections. Why have accurate dose algorithms?
Transition to Heterogeneity Corrections Eric E. Klein, M.S., Washington University, St. Louis, MO Craig Stevens, M.D., Ph.D., MD Anderson Cancer Center, Houston, TX Nikos Papinikolou, Ph.D., University
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationWhat Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff
What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG
More informationApplication of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom
Iran. J. Radiat. Res., 2003; 1(3): 143-149 Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom A. Mostaar 1, M. Allahverdi 1,2, M. Shahriari 3 1 Medical Physics Department,
More informationNew Thinking on Fractionation in Radiotherapy
New Thinking on Fractionation in Radiotherapy Alan E. Nahum Visiting Professor, Physics dept., Liverpool university, UK alan_e_nahum@yahoo.co.uk 1 An honorarium is provided by Accuray for this presentation
More informationProduction and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study
Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad
More informationEffective and efficient radiotherapy dosimetry audit: Where to next?
Effective and efficient radiotherapy dosimetry audit: Where to next? Catharine Clark, Eduard Gershkevitsh, Milan Tomsej and Nuria Jornet National Physical Laboratory Royal Surrey County Hospital Radiotherapy
More informationS. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY
More information4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest
SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific
More informationDosimetric Comparison: Acuros XB vs AAA
Dosimetric Comparison: Acuros XB vs AAA Leo El Hage, Chris James, Jim Frantzis, Roger Li, Venkata Sheshadri Epworth Radiation Oncology Epworth HealthCare Excellence. Everywhere. Everyday Background Epworth
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationAssessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit
More informationVerification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams
Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health
More informationIntegrating the Radixact into a Highly Comprehensive Center
Integrating the Radixact into a Highly Comprehensive Center Alonso N. Gutiérrez, PhD, MBA ASTRO Meeting 2018 Disclosures An honorarium is provided by Accuray for this presentation (donated) Accuray and
More informationInhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system
Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.
More information7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.
UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems
More informationSRS Uncertainty: Linac and CyberKnife Uncertainties
SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis
More informationClinical Implementation of SRS/SBRT
Clinical Implementation of SRS/SBRT Anil Sethi, PhD, FAAPM Loyola University Medical Center November 4, 2017 Disclosures Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch 2 Learning
More informationElectron Beam ET - Reloaded Therapy - Reloaded
8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles
More informationAccuracy of the Small Field Dosimetry Using the Acuros XB Dose Calculation Algorithm within and beyond Heterogeneous Media for 6 MV Photon Beams *
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2012, 1, 78-87 Published Online November 2012 (http://www.scirp.org/journal/ijmpcero) http://dx.doi.org/10.4236/ijmpcero.2012.13011
More informationS Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M
s y n c h r o n y r e s p i r a t o r y t r a c k i n g s y s t e m S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M The Synchrony System tracks respiration in real time and automatically
More information02 CyberKnife: Treatment Delivery
TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options
More informationCyberKnife SBRT for Prostate Cancer
CyberKnife SBRT for Prostate Cancer Robert Meier, MD Swedish Radiosurgery Center Swedish Cancer Institute Seattle, WA 2017 ESTRO Meeting, Vienna Austria 5-year safety, efficacy & quality of life outcomes
More informationIntroduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?
More informationAdvanced Tumor Treatment
Advanced Tumor Treatment What is CyberKnife? The CyberKnife Stereotactic Radiosurgery System, offered by Southwest Washington Medical Center, is a noninvasive outpatient treatment without the risks and
More informationProtons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.
Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic
More informationCommissioning of brachytherapy Treatment Planning Systems
Commissioning of brachytherapy Treatment Planning Systems Jacco L.G. Steenhuijsen, PhD Catharina Hospital Eindhoven Landelijke brachydag, 17-11-17 Brachytherapy Treatment Planning Systems 2 Brachytherapy
More informationSpecial Procedures Rotation I/II SBRT, SRS, TBI, and TSET
University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,
More informationTechnical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.
Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution
More informationHampton University Proton Therapy Institute
Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI
More informationA review of RTTQA audit. Karen Venables
A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines
More informationOrgan Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy
Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy Rasha Makkia Biomedical Physics Ph.D. student East Carolina University March 3 rd, 2016 Outlines The Purpose Wilms Tumor,
More informationVerification of Advanced Radiotherapy Techniques
Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December
More informationAnalytical Methods for Normal- Tissue Dose Reconstructions. Acknowledgments. Disclosures. Learning Objectives 8/2/2017. Wayne Newhauser, PhD
Analytical Methods for Normal- Tissue Dose Reconstructions Wayne Newhauser, PhD newhauser@lsu.edu AAPM Annual Meeting, August 2, 2017 Acknowledgments Bella Bowman Foundation Numerous current and former
More informationPSMMP PUBLICATIONS 2016
PSMMP PUBLICATIONS 2016 Dosimetric and radiobiological comparison of CyberKnife M6 InCise multileaf collimator over IRIS variable collimator in prostate stereotactic body radiation therapy, Vindu Kathriarachchi,
More informationStereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations
Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife State-of of-the-art and New Innovations Chad Lee, PhD CK Solutions, Inc. and CyberKnife Centers of San Diego Outline Basic overview
More informationNew CyberKnife M6 Service at HKSH for Cancer Patients Shorter Treatment Time, Fewer Side Effects and Improved Patient Outcomes
For Immediate Release New CyberKnife M6 Service at HKSH for Cancer Patients Shorter Treatment Time, Fewer Side Effects and Improved Patient Outcomes (13 August 2017, Hong Kong) Hong Kong Sanatorium & Hospital
More informationJournal extended Abstracts FL-AAPM Spring Meeting April 28-30, 2017, Orlando FL
PSMMP PUBLICATIONS 2017 A GPU accelerated simulation annealing algorithm for IMRT optimization, P. Galanakou, T. Leventouri, A. Georgakilas, G. Kalantzis, (IEEE proc. SNPD 2017). A Novel Test to Verify
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More information