Electron Beam ET - Reloaded Therapy - Reloaded
|
|
- Lesley Mathews
- 5 years ago
- Views:
Transcription
1 8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles Suitable for superficial tumors Requires collimation close to skin Clinical e unchanged for decades and underutilized in cancer radiotherapy ~ 15% utilization EAT Electron Arc Therapy Electron beam delivered in partial arc Modifications to linac/cone and TPS Shortened cone Different PDDs Cutout shape Patient-specific cast/shield and bolus Primarily for chest wall irradiation Univ. of Utah Leavitt 1985 Electron ARC therapy: physical measurement and treatment planning techniques IJROBP 11 McNeely 1988 Electron arc therapy: chest wall irradiation of breast cancer patients IJROBP 14(6) 1
2 8/4/216 ECT Electron Conformal Therapy One or a few electron beams Bolus electron conformal therapy Tissue equivalent Modulate electron energy Keep PTV within 9% isodose volume Minimize dose to distal/underlying critical structures and normal tissue Perkins 21 A custom three-dimensional electron bolus technique for optimization of postmastectomy irradiation IJROBP51 MERT Modulated Electron Radiotherapy Multiple electron beams Energy modulation Bolus or energy selection Intensity modulation Cutout emlc pmlc Scanning beam Standard MERT : 6 MV e : 6 / 9 / 12 MeV 12 segments Henzen et al 214 Beamlet based direct aperture optimization for MERT using a photon MLC. Med Phys, 41 Henzen et al 214 Monte Carlo based beam model using photon MLC for modulated electron radiotherapy. Med Phys, 41 DEAR Dynamic Electron Arc Radiotherapy Electron radiation is delivered in ARC mode Electron applicator and cut-out are kept to provide lateral beam constriction Treatment couch is in simultaneous motion with gantry rotation to prevent collision. Beam always normal to skin and SSD = 1 cm Couch motion, gantry rotation, and dose rate are modulated to produce desirable dose distributions Rodrigues, Yin, and Wu. "Dynamic Electron Arc Radiotherapy (DEAR): a feasibility study", Phys Med Biol, (2) 2
3 8/4/216 DEAR design DEAR design Gantry Couch Plan comparison: chest wall phantom DEAR Photons DEAR: 6, 9, 12 MeV. Photons: 6X, tangent, no wedge 3
4 Gantry Angle (deg) Couch LAT pos (cm) Couch VRT pos (cm) 8/4/216 DEAR delivery DEAR delivery demo Room View Patient View Gantry and couch motion Position 5 4 Speed -5 1 Gantry Angle(deg) Expected Actual Gantry Rot (deg/s) Expected Actual -1 Couch VRT (cm) Couch VRT (cm/s) Couch LAT (cm) Couch LAT (cm/s) Dose (MU) Time (s) 4
5 Dose Rate (MU/min) Dose (MU) Couch VRT pos (cm) Couch LAT pos (cm) Dose Rate (MU/min) 8/4/216 Beam hold Dose (MU) Beam Hold Beam Hold Expected Actual Couch VRT (cm) Couch LAT (cm) Time (s) Dose rate modulation 7 6e 7 6x 6 Actual 6 5 Expected Time (s) stdev = 1.6 MU/min Time (s) stdev =.2 MU/min Dosimetry: 6e on cylindrical phantom depth 1.5 cm Static (16x1 cm 2,G=⁰) DEAR (3x1 cm 2,G=315-45⁰)
6 8/4/216 Virtual scanning mode 6 MeV electron D=1cm cutout CR film 148 CP (7MU/CP) Beam hold (D->e) Potential DEAR case Need for fine energy resolution Photons DEAR 6
7 8/4/216 Small field electron beam dosimetry Small field electron beam dosimetry Composite dose map 6 MeV 1x1 cm 2 kernel Small field dosimetry 6 MeV 1x1 cm 2 PDD 7
8 Technology Complexity 8/4/216 Small field dosimetry 16 MeV 1x1 cm 2 PDD DEAR Summary Advantages Uniform dose distributions over large and curved targets while maintaining narrow penumbra Treated area > cone size DEAR delivery Fixed cone/cutout DEAR delivery has high accuracy Expected and actual plans agree very well Trajectory log file can be used as a QC tool Dosimetry Challenges Small field dosimetry Not ready in clinical operation Couch motion Coarse energy selection Bolus Planning tools Forward and inverse optimization Dynamic delivery collimators Motorized electron collimator (MEC) Electron MLC (emlc) QA Tools Summary: e vs. γ FFF VMAT IMRT ARC 3DCRT Photons SFF DEAR MERT EAT ECT Electrons VMATe 3D Printing 8
9 8/4/216 Prototype emlc Eley, J. G., K. R. Hogstrom, et al. (211). "Potential of discrete Gaussian edge feathering method for improving abutment dosimetry in emlc-delivered segmented-field electron conformal therapy." Medical Physics 38(12): Looking to the Future Antolak Commercial emlc Gauer, T., D. Albers, et al. (26). "Design of a computer-controlled multileaf collimator for advanced electron radiotherapy." Physics in medicine and biology 51(23): Antolak 9
10 8/4/216 Hogstrom, 23 Slide 31 Hogstrom, 23 Slide 32 1
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?
More informationA model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms
1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,
More informationElectron therapy Class 3: Clinical procedures
Electron therapy Class 3: Clinical procedures Laurence Court lecourt@mdanderson.org Reference: Faiz M. Khan, The Physics of Radiation Therapy Slide acknowledgements: Karl Prado, Rebecca Howell, Kent Gifford,
More informationA TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,
More informationOutline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams
Chapter 12 Treatment Planning Combination of Beams Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther Outline Combination
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationVariable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10
Variable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10 Simon Goodall Radiation Oncology Physicist Genesis Care Western Australia Introduction
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationLung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014
Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this
More informationRADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident
Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate
More informationProduction and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study
Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad
More informationAdditional Questions for Review 2D & 3D
Additional Questions for Review 2D & 3D 1. For a 4-field box technique, which of the following will deliver the lowest dose to the femoral heads? a. 100 SSD, equal dmax dose to all fields b. 100 SSD, equal
More informationIROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014
IROC Prostate Phantom Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 The study groups are requesting that each institution keep the phantom for a period of time
More informationPGY-1. Resident Review Session Schedule
1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT
More informationARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.
ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael
More informationEric E. Klein, Ph.D. Chair of TG-142
Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing
More informationIROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015
IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During
More informationIROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014
IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 The study groups are requesting that each institution keep the phantom for a period of time
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationIROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015
IROC Lung Phantom 3D CRT / IMRT Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 The IROC requests that each institution keep the phantom for no more than 2 weeks. During
More informationEfficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO
Investigations and research Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO M. Kunze-Busch P. van Kollenburg Department of Radiation Oncology, Radboud University Nijmegen Medical
More informationQuality assurance of volumetric modulated arc therapy using Elekta Synergy
Acta Oncologica, 2009; 48: 11931197 ORIGINAL ARTICLE Quality assurance of volumetric modulated arc therapy using Elekta Synergy AKIHIRO HAGA 1, KEIICHI NAKAGAWA 1, KENSHIRO SHIRAISHI 1, SAORI ITOH 1, ATSURO
More informationBorges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1
Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationIMRT Planning Basics AAMD Student Webinar
IMRT Planning Basics AAMD Student Webinar March 12, 2014 Karen Chin Snyder, MS Senior Associate Physicist Department of Radiation Oncology Disclosures The presenter has received speaker honoraria from
More informationIndependent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA
Independent Dose Verification for IMRT Using Monte Carlo C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Outline Why Monte Carlo for IMRT QA? Experimental verification
More informationImplementation of advanced RT Techniques
Implementation of advanced RT Techniques Tibor Major, PhD National Institute of Oncology Budapest, Hungary 2. Kongres radiološke tehnologije, Vukovar, 23-25. September 2016. Current RT equipments at NIO,
More informationNIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed
More informationA dosimetric comparison between volumetric-modulated arc therapy and dynamic conformal arc therapy in SBRT
JBUON 2019; 24(2): 838-843 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE A dosimetric comparison between volumetric-modulated arc therapy and
More informationDosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar
More informationA Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population
A Dosimetric Comparison of Whole-Lung Treatment Techniques in the Pediatric Population Corresponding Author: Christina L. Bosarge, B.S., R.T. (R) (T) Indiana University School of Medicine Department of
More informationRadiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery
Therapy SAM Symposium: WE-A-BRCD-1 Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery Kamil M. Yenice, PhD Associate Professor Chief of Clinical Physics
More informationA VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI
More informationTrajectory Modulated Arc Therapy: Application to Partial Breast Irradiation. Research and development to advance radiotherapy
Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation Dimitre Hristov Radiation Oncology Stanford University Research and development to advance radiotherapy New imaging platforms:
More informationTREATMENT PLANNING: CLINICAL ELECTRONS John A. Antolak, Ph.D. Prepared for the AAPM Therapy Physics Review Course Austin, TX July 19 20, 2014 PURPOSE
TREATMENT PLANNING: CLINICAL ELECTRONS John A. Antolak, Ph.D. Prepared for the AAPM Therapy Physics Review Course Austin, TX July 19 20, 2014 PURPOSE The purpose of this lecture is to provide the medical
More informationTreatment Planning for Lung. Kristi Hendrickson, PhD, DABR University of Washington Dept. of Radiation Oncology
Treatment Planning for Lung Kristi Hendrickson, PhD, DABR University of Washington Dept. of Radiation Oncology Outline of Presentation Dosimetric planning strategies for SBRT lung Delivery techniques Examples
More informationTreatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)
Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize
More information3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients
3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients G Asuni *1, T vanbeek 1, E VanUtyven 1, P McCowan 1,2, and B.M.C. McCurdy 1,2,3 1 Medical Physics Department, CancerCare
More informationIMAT: intensity-modulated arc therapy
: intensity-modulated arc therapy M. Iori S. Maria Nuova Hospital, Medical Physics Department Reggio Emilia, Italy 1 Topics of the talk Rotational IMRT techniques: modalities & dedicated inverse-planning
More informationAdvanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*
Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark* Purpose: To evaluate an institution s 3D treatment planning process and the institution
More informationMeasurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry
Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Conflict of interest None Acknowledgement: Jim
More informationTHE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS
THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More informationVerification of Advanced Radiotherapy Techniques
Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December
More informationWHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY
doi:10.1016/j.ijrobp.2009.03.029 Int. J. Radiation Oncology Biol. Phys., Vol. 75, No. 1, pp. 253 259, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front
More informationProstate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014
Prostate Phantom Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014 GENERAL INFORMATION: Each institution may keep the phantom for a period of time no more than 2 weeks.
More information7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis
Acknowledgments Thanks to Saiful Huq for an illuminating conversation about the application of TG-100 Jennifer O Daniel, Ph.D. & Fang-Fang Yin, Ph.D. Duke University Medical Center Annual AAPM Meeting,
More informationClinical experience with planning, quality assurance, and delivery of burst-mode modulated arc therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Clinical experience with planning, quality assurance, and delivery of burst-mode modulated arc therapy Kristofer Kainz, a Douglas
More informationCollimator rotation in volumetric modulated arc therapy for craniospinal irradiation and the dose distribution in the beam junction region
Li et al. Radiation Oncology (2015) 10:235 DOI 10.1186/s13014-015-0544-z METHODOLOGY Open Access Collimator rotation in volumetric modulated arc therapy for craniospinal irradiation and the dose distribution
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationTo Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence
SCIENCE & TECHNOLOGY To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence Alburuj R. Rahman*, Jian Z. Wang, Dr. Z. Huang, Dr. J. Montebello Department of
More informationAAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction
AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall
More informationNormal tissue dose in pediatric VMAT Piotr Zygmanski
Normal tissue dose in pediatric VMAT Piotr Zygmanski Department of Radiation Oncology Brigham and Women s Hospital Radiation induced effects in children Risk of inducing 2 nd cancers higher than for adults
More informationEvaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas
1 Carol Boyd Comprehensive Case Study July 11, 2013 Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas Abstract: Introduction:
More informationIROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014
IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014 The IROC Houston is requesting that each institution keep the phantom for two weeks. During
More information8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient
Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,
More informationQUALITY ASSURANCE IN RADIOTHERAPY THERAPY EQUIPMENT
QUALITY ASSURANCE IN RADIOTHERAPY THERAPY EQUIPMENT G. Guidi Az. Ospedaliero-Universitaria di Modena - Policlinico, Modena Guidi.Gabriele@policlinico.mo.it Phone: +39 059 422 5699 AGENDA Overview of the
More informationVarian Treatment. Streamlined Treatment Delivery Management Application. Specifications
Varian Treatment Streamlined Treatment Delivery Management Application Specifications Specifications Varian Treatment 1 Introduction Streamlined Treatment Delivery Management Varian Treatment verifies
More informationEvaluation of Dosimetry Check software for IMRT patient-specific quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,
More informationA Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer
A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer Tony Kin Ming Lam Radiation Planner Dr Patricia Lindsay, Radiation Physicist Dr John Kim, Radiation Oncologist Dr Kim Ann Ung,
More informationDebate Motion. Theoretical benefits of VMAT: Real Costs of VMAT. "Arc Based Techniques Will Make Conventional IMRT Obsolete
Austin, TX July 21, 2014 Debate Motion "Arc Based Techniques Will Make Conventional IMRT Obsolete AGAINST the Motion: Peter Balter, Ph.D. Potential Conflicts of Interest: Dr. Balter has research support/collaborations
More informationClinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations
Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.111 pissn 2508-4445, eissn 2508-4453 Clinical Implications of High Definition Multileaf Collimator
More informationActivity report from JCOG physics group
2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer
More informationGuidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy
Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...
More informationImplementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy
Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationIMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor
IMRT/IGRT Patient Treatment: A Community Hospital Experience Charles M. Able, Assistant Professor Disclosures I have no research support or financial interest to disclose. Learning Objectives 1. Review
More informationSpatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.
Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation
More informationOriginal Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem
Original Article Treatment Planning Evaluation of Sliding Window and Multiple Static Segments Technique in Intensity Modulated Radiotherapy for Different Beam Directions Teyyiba Kanwal, Muhammad Khalid,
More informationMedical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center
Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center All students accepted into the Medical Dosimetry Graduate Certificate Program
More informationA comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment
Oncology and Translational Medicine DOI 10.1007/s10330-016-0183-3 December 2017, Vol. 3, No. 6, P260 P266 REVIEW ARTICLE A comparative study between flattening filter-free beams and flattening filter beams
More informationKenny Guida, DMP, DABR March 21 st, 2015
Kenny Guida, DMP, DABR March 21 st, 2015 Breast Cancer Treatment planning and delivery Hypofractionation Trials Hybrid Planning History Techniques RTOG 1005 Trial Hybrid-VMAT Research project 3D Tangents
More informationTraceability and absorbed dose standards for small fields, IMRT and helical tomotherapy
Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,
More informationA review of RTTQA audit. Karen Venables
A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines
More informationDosisverifikation mit Delta 4 Discover während der Behandlung
Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover
More informationQuality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose
Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel
More informationAn Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements
An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements Jean M. Moran, a Donald A. Roberts, Teamour S. Nurushev, Larry E. Antonuk, Youcef El-Mohri, and Benedick A. Fraass Department
More informationIMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors
IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors S. Sathiyan, M. Ravikumar, C. Varatharaj, S.S. Supe, S.L. Keshava Department of Radiation Physics, Kidwai
More informationProtons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.
Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic
More informationOptimization of RapidArc for Head-and-Neck Radiotherapy. Jessica Emily Salazar. Department of Medical Physics Duke University.
Optimization of RapidArc for Head-and-Neck Radiotherapy by Jessica Emily Salazar Department of Medical Physics Duke University Date: Approved: Shiva Das, Supervisor Ryan McMahon Robert Reiman Thesis submitted
More information4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest
SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific
More information8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation
Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research
More information3D-CRT Breast Cancer Planning
3D-CRT Breast Cancer Planning Tips and Tricks Bednář, V. 3D-CRT Obsolete or not? There are more advanced techniques than 3D-CRT, but 3D-CRT has some advantages: Availability and price Forward planning
More informationVOLUMETRIC-MODULATED ARC THERAPY VS. 3D-CONFORMAL RADIOTHERAPY FOR BREAST CANCER
Romanian Reports in Physics, Vol. 67, No. 3, P. 978 986, 2015 VOLUMETRIC-MODULATED ARC THERAPY VS. 3D-CONFORMAL RADIOTHERAPY FOR BREAST CANCER D. ADAM 1, 2, M.D. SUDITU 1, 2, R. POPA 1, 2, V. CIOCALTEI
More information6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS
6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types
More informationImaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT
Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationRotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies
Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies J. Y. C. Cheung and K. N. Yu a Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong,
More informationTomoTherapy. Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA. May 2016
TomoTherapy Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA May 2016 TomoTherapy The Facts Greek Tomo = slice Advanced form of IMRT 3D computerised tomography (CT) imaging immediately prior
More informationLearning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy
3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy Sofie Ceberg, PhD Medical Physicist and Lund University Lund, Sweden Learning objectives 1: 3D Dosimetry in the Clinic: Background
More informationCan we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationImage Guided Proton Therapy and Treatment Adaptation
Image Guided Proton Therapy and Treatment Adaptation www.hollandptc.nl d.r.schaart@tudelft.nl Cancer in The Netherlands About 1 in 3 people get cancer in some stage of their life 86.800 new cancer patients
More informationEstablishing the impact of temporary tissue expanders on
Establishing the impact of temporary tissue expanders on electron and photon beam dose distributions A. Asena 1,*, T. Kairn 1,2, S.B. Crowe 1 and J.V Trapp 1 1 School of Chemistry, Physics and Mechanical
More informationCODING GUIDELINES. Radiation Therapy. Effective January 1, 2019
CODING GUIDELINES Radiation Therapy Effective January 1, 2019 Coding guidelines for medical necessity review of radiation therapy services. Please note the following: CPT Copyright 2018 American Medical
More informationDosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.87 Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Mee
More information