Neutron dose evaluation in radiotherapy
|
|
- Gordon Williamson
- 5 years ago
- Views:
Transcription
1 Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA
2 Radiation therapy with a linear accelerator (LINAC)
3 Photoneutron production in accelerator head Photoneutrons produced by interaction of photon beam with accelerator components Produced mainly in the target, primary collimator, flattener and jaws/collimators Typical materials are copper,tungsten, gold, lead and iron Neutron production in electron mode is lower than in photon mode Direct production of neutrons by electrons is at least 2 orders of magnitude lower Lower electron current Nisy E. Ipe, 2007 AAPM Summer School
4 Photoneutron production in accelerator head Photon must have energy greater than binding energy of nucleus in atom S n =Separation Energy Neutron production in primary laminated barrier Lead has lower S n than Iron Iron Pb-207 (22.1%): S n =6.74 MeV (NCRP 79) Pb-208 (52.4%) : S n = 7.37 MeV Fe-57 (2.1%): S n = 7.65 MeV Fe-56 (91.7%): S n =11.19 MeV Lead has a higher neutron yield than iron Steel is a better choice for reducing neutron production Nisy E. Ipe, 2007 AAPM Summer School
5 Schematic of Varian accelerator head
6 Photoneutron production in accelerator head
7 Photoneutron production Photoneutron spectrum from accelerator head resembles a fission spectrum Spectrum changes after penetration through head shielding Concrete room scattered neutrons will further soften the spectrum Spectrum outside the concrete shielding resembles that of a heavily shielded fission spectrum Nisy E. Ipe, 2007 AAPM Summer School
8 Photoneutron production Two Processes Direct Emission Average energy of direct neutrons is ~ few MeV Spectra peak at energies > 2 MeV Have a sin 2 J angular distribution, therefore forward peaked Contributes about 10 to 20% of neutron yield for bremsstrahlung with upper energies of 15 to 30 MV Nisy E. Ipe, 2007 AAPM Summer School
9 Photoneutron production Two Processes Evaporation Neutrons Dominant process in heavy nuclei Emitted isotropically Spectral distribution is independent of photon energy for energies that are a few MeV above neutron production threshold Average energy is ~ 1-2 MeV Spectra peak at ~ kev Nisy E. Ipe, 2007 AAPM Summer School
10 Recent treatment modalities IMRT SmartBeam IMRT "paints" a dose to the tumor with pinpoint precision, while sparing healthy normal tissue. Minimizes hot spots Improves target inhomogeneity Provides detailed dose painting to the target Sculpts dose around critical structures more effectively Allows treatments to occur in conventional minute time slots Dose resolution, with up to 500 segments per field Spatial resolution of 2.5-5mm
11 Recent treatment modalities VMAT RapidArc RapidArc uses a unique algorithm that provides excellent treatment delivery control. Its single gantry rotation speeds treatment delivery so clinicians can develop treatments that take one-half to one-eighth the time of conventional IMRT treatments just two minutes in many cases. A RapidArc treatment may also result in less radiation leakage and scatter, so peripheral tissues receive a lower overall dose.
12 Recent treatment modalities - Tomotherapy
13 Photoneutrons: NCI-WG 2001 recommendations Total-body dose for IMRT patients is higher, generally increasing with the number of monitor units used for treatment. The potential for complications related to this increased dose should be recognized and considered. Increased neutron production for high-energy machines used for IMRT should be considered. Increased workload values for IMRT (may be 2 5 times larger than in conventional therapy) should be considered for the leakage/transmission part of the treatment room shielding. Because IMRT is inherently less efficient (per MU) than conventional RT, vendors should consider the use of more internal shielding in the design of future IMRT machines.
14 Neutron dosimetry Conversion coefficients Neutrons particle fluence to ambient or personal dose equivalent influence of phantom (multiple scattering) and quality factor
15 Neutron radiation protection dosimetry Utilization of physical phenomena resembling dose (equivalent) deposition in tissue. Measurements of LET spectra and convolution over Q(L). Design of systems mimicking the fluence to dose equivalent conversion coefficient
16 Neutron detection by thermoluminescence Neutrons are particles without charge, difficult to detect TL induced by secondary particles from nuclear reactions: (n, ) (n,p) (n,d) etc TLD for neutrons have high concentration of isotopes with high cross section to neutrons in LiF ~ 7,4% of Li is 6 Li n Radiation LiF 6 Li(n, ) 3 H
17 Some neutron capture cross sections 3 He(n,p) 3 H 10 B(n, ) 7 Li 6 Li(n, ) 3 H
18 Albedo-based neutron dosimeters
19 TL albedo neutron dosimeter response
20 Plastic Nuclear Track Detectors Particles of ionizing radiation cause molecular size damage in solid material The damage can be enlarged to microscopic range by chemical etching
21 How CR-39 PNTDs Work
22 How CR-39 PNTDs Work
23 How CR-39 PNTDs Work
24 Atomic force microscopy of latent tracks Latent tracks nm and must be enlarged to 5-30 µm to be visible with optical microscopy Tracks are enlarged by chemical or electrochemical etching
25 Atomic force microscopy of latent tracks
26 How CR-39 PNTDs Work
27 Result of Etching
28 Typical microscopic view-field with developed tracks Overlapping Tracks Identified Tracks Track density is proportional to the exposure value [dose] With calibration the [equivalent] dose can be deducted
29 Neutron response mechanism In practice, converter layers are utilized: Neutrons are non ionizing radiation, but CR-39 (C 12 H 18 O 7 ) is sensitivity to fast neutron recois Neutrons interact with H producing recoil protons Recoil protons create latent tracks in the CR-39 material - Converters modify (improve) the energy dependence of the reponse - Converters also protect the CR-39 chip against alpha particles from environmental radon
30 Neutron energy dependent response Good energy dependence of response But, relatively low sensitivity
31 Sensitization by CO 2 pretreatment
32 Superheated emulsions Fluorocarbon droplets kept in a steady superheated state by emulsification in compliant gels. Bubble nucleation triggered by neutrons above selectable threshold energies. Can be totally insensitive to photons.
33 Some current technologies
34 Neutron energy dependence of emulsions Composition closely tissue equivalent Insensitive to photons Response resembling kerma equivalent coefficient The emulsions can measure dose equivalent in phantom without disrupting charged particle equilibrium
35 Isocenter neutron doses
36 Neutron doses from various x-ray beams 18 MV x-rays 15 MV x-rays 10 MV x-rays
37 In vivo photoneutron measurements
38 SDD fluence response vs quality-factor weighted kerma coefficient and photoneutron spectrum
39 Bubble counting by scattered light Photo diodes LEDs Instant read out Rate insentive Position sensitive
40 Study of a prostate treatment
41 Characteristics of various phantoms for out-of-field measurements
42 Clinical simulations of prostate radiotherapy BOMAB = BOttle MAnnikin ABsorber phantom, industry standard (ANSI 1995) for calibrating whole body counting systems
43 Irradiation channels ( pipes )
44 BOMAB CT scan and simulated organs Bladder Rectum Transverse plane section view Prostate PTV Boost volume Sagittal Sagittal plane section view plane section view 5 cm 5 cm 5 cm 5 cm
45 Organs of interest: bladder, prostate, rectum
46 Organs of interest: bladder, prostate, rectum
47 Organs of interest: bladder, prostate, rectum
48 Treatment modalities in Pisa and Krakow Pisa Varian Clinac 2100 C MU/ 2 Gy Modalities Krakow Varian Clinac 2300 CD 6 MV single 10x10 cm 2 field (ref) MV single 10x10 cm 2 field (ref) 15 MV single 10x10 cm 2 Used for reference field (ref) MV single 10x10 cm 2 field (ref) 6 MV 4-field MLC 15 MV 5-field MLC 6 MV IMRT MV 4-field MLC 6 MV IMRT 18 MV IMRT 6 MV VMAT (RapidArc) MV Tomotherapy MU/ 2 Gy
49 Treatment modalities 5 fields with MCL
50 Treatment modalities IMRT
51 Treatment modalities - VMAT
52 Dose profiles from different treatment plans
53 LEG S Dosimeter in-phantom placement BTI sensitive detector length: 7.0 cm SDD sensitive detector length: 2.5 cm Measurement points center of sensitive length D = 6.8 D = 12 D = 16.8 D = 31.8 D = 46.8 D = 56.8 D = 12 D = 27 D = 42 D distance from outer flange legs side. Dimensions in cm
54 μsv/gy IMRT neutron dose as a function of nominal energy 1,E+03 1,E+02 1,E MV - Krakow 6 MV - Pisa 1,E Penetration distance along the sagittal plane (cm) 5 cm is bladder, 10 cm is prostate and 15 cm is rectum
55 μsv/gy MLC multiple field treatments 1,E+04 PTV Krakow 18 MV SDD PTV Pisa 15 MV SDD 1,E+03 PTV Krakow 18 MV PADC PTV Pisa 15 MV PADC 1,E+02 1,E+01 1,E Off-axis distance (cm)
56 μsv/gy Reference phantom literature comparison - SDD 6,E+03 5,E+03 4,E+03 3,E+03 This work (SDD, 20 MV) This work (BDPND, 20 MV) d'errico 2000 (SDD, 18 MV) d'errico 2000 (SDD, 15 MV) Awotwi-Pratt 2007 (SDD, 15 MV) Lin 2007 (BDPND, 15 MV) This work (SDD, 12 MV) 2,E+03 1,E+03 0,E Off-axis distance (mm)
57 TrueBeam 10 cm x 10 cm field results
58 Conclusions and Prospects The Pisa BOMAB proved a viable approach The selected neutron dosimetry methods appear reliable and reproducible But, measurement times are long and additional campaigns are needed for a systematic analysis Tomotherapy appears to deliver the smallest neutron doses A small neutron contamination is present already at 6 MV and warrants further investigation
Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments
Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331
More informationToday, I will present the second of the two lectures on neutron interactions.
Today, I will present the second of the two lectures on neutron interactions. 1 The main goal of this lecture is to tell you a little about clinical neutron therapy, first with fast neutron beams, and
More informationNeutron Measurements for Intensity Modulated Radiation Therapy
SLAC-PUB-8443 April 2 Neutron Measurements for Intensity Modulated Radiation Therapy N. E. Ipe et al. Presented at Chicago 2 World Congress on Medical Physics and Biomedical Engineering, 7/23/2 7/28/2,
More informationSHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES
SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring
More informationIntroduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.
Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy
More informationNeutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics
Neutron Interactions Part 2 George Starkschall, Ph.D. Department of Radiation Physics Neutron shielding Fast neutrons Slow down rapidly by scatter in hydrogenous materials, e.g., polyethylene, paraffin,
More informationEQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS
PAPER EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS Isra Israngkul Na Ayuthaya *, Sivalee Suriyapee, Phongpheath Pengvanich
More informationAN ABSTRACT OF THE THESIS OF. Leanna R. Eller for the degree of Master of Science in Radiation Health Physics
AN ABSTRACT OF THE THESIS OF Leanna R. Eller for the degree of Master of Science in Radiation Health Physics presented on November 30, 2010. Title: An Investigation on Photoneutron Production from Medical
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER
ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER Bro. Dr. Collie Miller IARC/WHO Based on trends in the incidence of cancer, the International Agency for Research on Cancer (IARC) and WHO
More informationRisk of a second cancer after radiotherapy
Risk of a second cancer after radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Medical radiological procedures worldwide 2.5 billion diagnostic radiological examinations 78%
More informationAAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy
AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy Stephen F. Kry a) Department of Radiation Physics, MD Anderson Cancer Center, Houston,
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationMonte Carlo Modelling: a reliable and efficient tool in radiation dosimetry
Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative
More informationShielding Design Basis and its Calculation of High Energy Medical Linac Installed in Bangladesh Atomic Energy Commission, Bangladesh
Universal Journal of Medical Science 5(2): 27-31, 2017 DOI: 10.13189/ujmsj.2017.050202 http://www.hrpub.org Shielding Design Basis and its Calculation of High Energy Medical Linac Installed in Bangladesh
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationCamille J. Lodwick. Photoneutron contamination produced during high-energy, external beam
AN ABSTRACT OF THE THESIS OF Mary Elizabeth Catchpole for the degree of Master of Science in Medical Physics presented on December 1, 2010. Title: Comparison of Secondary Neutron Dose from 10 MV Intensity
More informationNeutrons. ρ σ. where. Neutrons act like photons in the sense that they are attenuated as. Unlike photons, neutrons interact via the strong interaction
Neutrons Neutrons act like photons in the sense that they are attenuated as I = I 0 e μx where Unlike photons, neutrons interact via the strong interaction μ = The cross sections are much smaller than
More informationTHERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL
THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL G. Gambarini Dipartimento di Fisica dell Università, Milano, Italy e-mail grazia.gambarini http://users.unimi.it/~frixy/
More informationOut-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison
MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationPeak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam
Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch
More informationIntensity Modulated RadioTherapy
Intensity Modulated RadioTherapy A clinical application of a small accelerator University Medical Center Groningen A.A. van t Veld PhD I. Hoveijn PhD part 1 part2 CERN/KVI accelerator school, Zeegse, June
More informationGDR in Radiotherapy Treatment Fields with 18 MV Accelerators
GDR in Radiotherapy Treatment Fields with 18 MV Accelerators R. R. Martín-Landrove 1, J. Dávila 1, H. R. Vega-Carrillo 2, M. T. Barrera 3, A.J. Kreiner 4, F. Pino 3, H. Barros 3, E. D. Greaves 3 and L.
More informationRadiologic Units: What You Need to Know
Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report
More informationPractical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4
Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability
More informationProduction and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study
Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad
More informationImaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT
Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationVerification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,
More informationRadiation Related Second Cancers. Stephen F. Kry, Ph.D., D.ABR.
Radiation Related Second Cancers Stephen F. Kry, Ph.D., D.ABR. Objectives Radiation is a well known carcinogen Atomic bomb survivors Accidental exposure Occupational exposure Medically exposed Radiotherapy
More informationABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007
ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory
More informationIntroduction. Chapter 15 Radiation Protection. Advisory bodies. Regulatory bodies. Main Principles of Radiation Protection
Introduction Chapter 15 Radiation Protection Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. F.M. Khan, The Physics of Radiation Therapy, 4th ed., Chapter
More informationThis article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing
More informationVenue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry
Venue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry M. Caresana a, A. Sashala Naik a,c, S. Rollet b, M. Ferrarini a,d a Polytechnic
More informationA comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment
Oncology and Translational Medicine DOI 10.1007/s10330-016-0183-3 December 2017, Vol. 3, No. 6, P260 P266 REVIEW ARTICLE A comparative study between flattening filter-free beams and flattening filter beams
More informationNuclear Data for Radiation Therapy
Symposium on Nuclear Data 2004 Nov. 12, 2004 @ JAERI, Tokai Nuclear Data for Radiation Therapy ~from macroscopic to microscopic~ Naruhiro Matsufuji, Yuki Kase and Tatsuaki Kanai National Institute of Radiological
More informationA Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer
A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer Tony Kin Ming Lam Radiation Planner Dr Patricia Lindsay, Radiation Physicist Dr John Kim, Radiation Oncologist Dr Kim Ann Ung,
More informationY FILMS DOSIMETR Nederland België / Belgique
DOSIMETRY FILMS GAFCHROMIC Dosimetry Films The Self-developing Dosimetry Films that Allow You to Go Filmless Do away with the cost and headache of calibrating and maintaining a processor Do away with hazardous
More informationPRINCIPLES and PRACTICE of RADIATION ONCOLOGY. Matthew B. Podgorsak, PhD, FAAPM Department of Radiation Oncology
PRINCIPLES and PRACTICE of RADIATION ONCOLOGY Matthew B. Podgorsak, PhD, FAAPM Department of Radiation Oncology OUTLINE Physical basis Biological basis History of radiation therapy Treatment planning Technology
More informationExpectations of Physics Knowledge for Certification
Expectations of Physics Knowledge for Certification Bhudatt Paliwal, Ph.D. University of Wisconsin Medical School Department of Human Oncology Madison, Wisconsin 53792 USA Guiding Principles Test that
More informationTwo-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance
Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing
More informationRadiation Monitoring Instruments
Radiation Monitoring Instruments This set of slides is based on Chapter 4 authored byg. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics: A Handbook for Teachers
More informationDosisverifikation mit Delta 4 Discover während der Behandlung
Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover
More informationSarcoma and Radiation Therapy. Gabrielle M Kane MB BCh EdD FRCPC Muir Professorship in Radiation Oncology University of Washington
Sarcoma and Radiation Therapy Gabrielle M Kane MB BCh EdD FRCPC Muir Professorship in Radiation Oncology University of Washington Objective: Helping you make informed decisions Introduction Process Radiation
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationFETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS
FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção
More information6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS
6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types
More informationIMAT: intensity-modulated arc therapy
: intensity-modulated arc therapy M. Iori S. Maria Nuova Hospital, Medical Physics Department Reggio Emilia, Italy 1 Topics of the talk Rotational IMRT techniques: modalities & dedicated inverse-planning
More informationRadiation qualities in carbon-ion radiotherapy at NIRS/HIMAC
Radiation qualities in carbon-ion radiotherapy at NIRS/ Shunsuke YONAI Radiological Protection Section Research Center for Charged Particle Therapy National Institute of Radiological Sciences (NIRS) E-mail:
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationM. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course
Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics
More informationS. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY
More informationManaging the imaging dose during image-guided radiation therapy
Managing the imaging dose during image-guided radiation therapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Richmond VA Imaging during radiotherapy Radiographic
More informationnuclear science and technology
EUROPEAN COMMISSION nuclear science and technology Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields (EVIDOS) Editor: Helmut Schuhmacher, Physikalisch-Technische Bundesanstalt
More informationIntroduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?
More informationPHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)
PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce
More informationBELdART-2: Moving towards safer radiotherapy. Steven Lelie 1,2, Wouter Schroeyers 1, Brigitte Reniers 1, Sonja Schreurs 1
BELdART-2: Moving towards safer radiotherapy Steven Lelie 1,2, Wouter Schroeyers 1, Brigitte Reniers 1, Sonja Schreurs 1 1 Nucleair Technologisch Centrum (XIOS/Uhasselt), Wetenschapspark 27, 3590 Diepenbeek
More informationHelical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System
The Challenges Associated with Differential Dose Delivery using IMRT Chester Ramsey, Ph.D. Director of Medical Physics Thompson Cancer Center Knoxville, Tennessee, U.S.A Collaborators Chester Ramsey, Ph.D.
More informationCalibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System
Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System James D Rijken Thesis submitted for the degree of Master of
More informationDosimetry for Epidemiology Cohorts Who Receive Radiation Therapy
Dosimetry for Epidemiology Cohorts Who Receive Radiation Therapy Wayne Newhauser, PhD Eurados Winter School, Milan, 2016 Introduction About 1 in 2 men and women born today will be diagnosed with some form
More informationThe title of the presentation is: Neutron production with clinical LINACs for BNCT studies in physical, medical and biological fields.
The title of the presentation is: Neutron production with clinical LINACs for BNCT studies in physical, medical and biological fields. 1 Neutron production with clinical e-linac s is acheived by in-hospital
More informationIMRT Planning Basics AAMD Student Webinar
IMRT Planning Basics AAMD Student Webinar March 12, 2014 Karen Chin Snyder, MS Senior Associate Physicist Department of Radiation Oncology Disclosures The presenter has received speaker honoraria from
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationOutline. Outline 3/30/12. Second Cancers from. Radiotherapy Procedures. Stephen F. Kry, Ph.D., D.ABR.
Second Cancers from Radiotherapy Procedures Stephen F. Kry, Ph.D., D.ABR. Outline Radiation and cancer induction Medically exposed people Estimating risk of second cancers Minimizing the risk Outline Radiation
More informationAn anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery
An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,
More informationI. Equipments for external beam radiotherapy
I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),
More informationAssessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom
Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson
More informationV. 4. Design and Benchmark Experiment for Cyclotron-based Neutron Source for BNCT
CYRIC Annual Report 2002 V. 4. Design and Benchmark Experiment for Cyclotron-based Neutron Source for BNCT Yonai S., ItogaT., Nakamura T., Baba M., Yashima, H. Yokobori H. *, and Tahara Y. ** Cyclotron
More informationA TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,
More informationBasic radiation protection & radiobiology
Basic radiation protection & radiobiology By Dr. Mohsen Dashti Patient care & management 202 Wednesday, October 13, 2010 Ionizing radiation. Discussion issues Protecting the patient. Protecting the radiographer.
More informationReal Time Spectrometer for thermal neutrons from Radiotherapic Accelerators
Real Time Spectrometer for thermal neutrons from Radiotherapic Accelerators Aldo Mozzanica, Università degli Studi di Brescia, INFN sezione di Pavia mozzanica@bs.infn.it 2 Outlines The Boron Neutron Capture
More informationINTRODUCTION. Material and Methods
Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University
More informationCharacterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning
Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca
More informationPre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system
Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system E. Mezzenga, * E. Cagni, A. Botti, M. Orlandi and M. Iori Medical Physics Unit, IRCCS Arcispedale
More informationPresented by Dr. Amir H. Mohagheghi. Annual RMCC Workshop Middle East Scientific Institute for Security (MESIS) Amman, Jordan June 17-19, 2013
Neutron Activation Analysis and Dosimetry based on reports by R. Burrows, A. Mohagheghi, and D. Ward: Personnel Nuclear Accident Dosimetry at Sandia National Laboratories SAND96-2204, SAND2011-6416 Presented
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationMuch of what we will say
New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer
More informationCalibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia
Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationIPPE Iron shell transmission experiment with 14 MeV neutron source and its analysis by the Monte-Carlo method
IPPE Iron shell transmission experiment with 14 MeV neutron source and its analysis by the Monte-Carlo method S.P. Simakov 1,2, M.G. Kobozev 1, A.A. Lychagin 1, V.A. Talalaev 1, U. Fischer 2, U. von Möllendorff
More informationNIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed
More informationMany vendors are beginning to allow couch motion during radiation delivery.
Dynamic Couch Motion Many vendors are beginning to allow couch motion during radiation delivery. Varian developer mode allows institutions to perform research using these types of treatments. Tomotherapy
More informationIROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015
IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During
More informationCone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood
Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex
More informationBiological Effects of Ionizing Radiation & Commonly Used Radiation Units
INAYA MEDICAL COLLEGE (IMC) RAD 232 - LECTURE 2 & 3 Biological Effects of Ionizing Radiation & Commonly Used Radiation Units DR. MOHAMMED MOSTAFA EMAM How does radiation injure people? - High energy radiation
More information8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}
High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics
More informationImplementation of advanced RT Techniques
Implementation of advanced RT Techniques Tibor Major, PhD National Institute of Oncology Budapest, Hungary 2. Kongres radiološke tehnologije, Vukovar, 23-25. September 2016. Current RT equipments at NIO,
More informationOut-of-field Radiation Risks in Paediatric Proton Therapy
Out-of-field Radiation Risks in Paediatric Proton Therapy Charlot Vandevoorde NRF ithemba LABS Contact: cvandevoorde@tlabs.ac.za Seventh NCS Lustrum Proton Therapy Amsterdam Charlot Vandevoorde 27 October
More informationTreatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)
Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize
More informationHEALTH RISKS FROM EXPOSURE TO FAST NEUTRONS. Prof. Marco Durante
HEALTH RISKS FROM EXPOSURE TO FAST NEUTRONS Prof. Marco Durante Risk from neutrons Risk from exposure to fission spectrum neutrons has been extensively studied in the 60 s at nuclear reactors using animal
More informationRadiochromic film dosimetry in water phantoms
INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms
More informationThe ANDANTE project: a multidisciplinary approach to neutron RBE
The ANDANTE project: a multidisciplinary approach to neutron RBE Andrea Ottolenghi, Klaus Trott, Giorgio Baiocco, Vere Smyth Università degli Studi di Pavia, Italy On behalf of the ANDANTE project MELODI
More informationDollars and Sense: Are We Overshielding Imaging Facilities? Part 2
Disclosure Dollars and Sense: Are We Overshielding Imaging Facilities? Part 2 Bryon M. Murray, M.S., DABR Paid consultant to NELCO Worldwide Owner, CEO ZapIT! Medical Objectives Understand methods for
More informationMeasurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry
Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Conflict of interest None Acknowledgement: Jim
More informationProstate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014
Prostate Phantom Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014 GENERAL INFORMATION: Each institution may keep the phantom for a period of time no more than 2 weeks.
More informationIMRT QA: Point Dose Measurements or 2D Array?
QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify
More information