An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Size: px
Start display at page:

Download "An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center"

Transcription

1 An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center

2 Outline of Presentation Definition of the problem Results of audits Possible solutions Resources to help

3 Learning Objectives Recognize that small field dosimetry is not easy and that there are tools to help you be as accurate as possible.

4 Sometimes we all like to believe that all across the kingdom it s bright and merry!

5 Ladies and Lords There is a scourge across the Kingdom Small field dosimetry

6 What is the truth? From Das et al 2000

7 Recommendation: TG-155 Small Field Dosimetry The unshielded stereotactic diode and microchambers are the detectors of choice for measurements in radiosurgery beams. However, we recommend comparing the measurements from at least two different detectors listed in Table 2a, and after applying corrections taking their average.

8 TomoTherapy results (since Aug 2010) Average: / irradiations 80 # results TLD/Institution

9 But the manufacturer told me it would work, the manufacturer set the value or I sent my own measured data to the TPS company and they modeled my beams for me. 1. Accuray calibrates all of the tomotherapy units using the Exradin A1SL chamber and the user monitors constancy. Correction factor of for A1SL based on Gago-Arias et al, Med Phys, 2012

10 Manufacturer modeling of small field data You make the measurements You send the measurements They create the beam model BUT They assume you used the correct dosimeter They assume you made measurements correctly They do not know to apply additional corr. factors RESULT Incorrect Small Beam Model

11 Small Field Dosimetry Corrections Situation is even worse if you consider using field sizes less then 0.5 x 0.5 cm 2 Francescon et al 2011 data

12 f, f k clin 10x10 Q clin,q10x10 calculated as the mean over all the values obtained by changing the linac model, (Primus TM Siemens 6 MV and Synergy Elekta 6 MV), the radial FWHM and energy of the electron source as presented by Francescon et al. 97 These values can also be used with caution for Varian 6 MV Linacs. Field Size (cm x cm) Detectors 0.5x x x x x x3.0 PTW ± ± ± ± ± ±0.001 Sun Nuclear Dedge 0.932± ± ± ± ± ±0.002 IBA SFD 0.972± ± ± ± ± ±0.001 Exradin D1V 0.980± ± ± ± ± ±0.001 Exradin A ± ± ± ± ± ±0.001 PTW PinPoint ± ± ± ± ± ± PTW MicroLion 1.023± ± ± ± ± ±0.001 Exradin W1 PSD 0.998± ± ± ± ± ±0.002 Unpublished data from P. Francescon

13 Gamma Knife results (since Sep 2007) 60 Average: / results # results TLD/Inst.

14 Depending on the dosimeter used, the magnitude of the additional correction, down to a 0.5 x 0.5 cm 2 field size can be as much as: 1% 1. 5% 2% 2. 2% 1% 3. No correction needed 91% 4. 13% 6% 5. 8%

15 The magnitude of the additional correction, down to a 0.5 x 0.5 cm 2 field size can be as much as: The answer is: 13% Francescon et al. Medical Physics, Vol. 38, No. 12, December 2011

16 TG-155 Small Field Dosimetry Recommendation: It is strongly recommended to independently verify dosimetric measurements in small fields, either through measurements carried out by a different person and/or though an independent external audit, such as that carried out by the RPC.

17 RPC Measurements during onsite visits Measurement configuration: depth = 10 cm, 100 cm SSD, Exradin A16 Secondary jaws kept at a 10 x x x

18 The Problem is that our Dragon is very small!

19 Tables of standard small field factors We even make mistakes Followill et al 2012 and also will be in TG-155

20 Corrected Varian Data Varian 6 MV Varian 10 MV Varian 15 MV Varian 18 MV Field Size (cm x cm) RPC Institution RPC Institution RPC Institution RPC Institution (cm x cm) 10 x 10 RPC Institution RPC Institution RPC Institution RPC Institution x x 6 (0.937)* (0.965) x (0.005) (0.008) (0.002) (0.003) (0.005) (0.007) (0.003) (0.006) (0.005)[0.8%] (0.008) (0.002) [0.5%] (0.003) (0.005) [0.4%] (0.007) (0.003) [0.5%] (0.006) [0.8%] (n=127) [0.5%] (n=22) [0.4%] (n=32) [0.5%] (n=41) (n=127) (n=22) (n=32) (n=41) x 4 (0.885) (0.924) (0.006) (0.013) (0.006) (0.010) (0.006) (0.009) (0.002) (0.009) (0.006) (0.013) (0.006) (0.010) (0.006) (0.009) (0.002) (0.009) [1.8%] [1.2%] [0.7%] [0.9%] [1.8%] [1.2%] [0.7%] [0.9%] (n=122) (n=25) (n=32) (n=37) (n=122) (n=25) (n=32) (n=37) 3 x (0.849) (0.884) (0.007) (0.012) (0.006) (0.008) (0.006) (0.012) (0.005) (0.019) [2.4%] [1.6%] [0.9%] [1.4%] (n=123) (n=25) (n=31) (n=41) 2 x (0.802) (0.814) (0.008) (0.008) (0.020) (0.020) (0.005) (0.005) (0.015) (0.015) (0.011) (0.011) (0.040) (0.040) (0.007) (0.007) (0.020) (0.020) [2.9%] [2.0%] [2.2%] [1.6%] (n=136) (n=23) (n=33) (n=40)

21 On-Site Dosimetry Review Audit Discrepancies Discovered (Jan. 05 April 13) Discrepancies Regarding: Number of Institutions Receiving rec. (n = 206) Review QA Program 152 (74%) Photon Field Size Dependence 138 (67%) Wedge Factor (WF) 66 (32%) Off-axis Factors (OAF)/Beam symmetry 60 (29%) Electron Calibration 35 (17%) Photon Depth Dose 33 (16%) Electron Depth Dose 25 (12%) Photon Calibration 16 (8%) This is a beam measurement issue and TPS beam modeling challenge.

22 There is guidance or possibly a fair maiden after the dragon is slain. The bigger challenge is with Varian machines due to the tertiary collimation (MLC) Varian/Pinnacle combination: Followill et al, JACMP, vol. 13, No. 5, 2012 Varian/Eclipse combination: Kron et al, Med Phys, vol 39 (2), 2012

23 TG-155 Small Field Dosimetry Huge Resource of Published data: 239 references 5 Tables 15 Figures 13 Conclusions and recommendations

24 New IROC Houston audit OSLD

25 Commissioning and Testing Three dosimeters were used to estimate the dose for each field size: Exradin A16 Exradin D1V diode Exradin D1V diode PTW CC04 Exradin W1

26 First Test Irradiation away from RPC

27 All of the following are current methods to provide a verification of your small field dosimetry data except: 1% 1. Use of published data 8% 86% 4% 2% 2. Your 2 nd set of measurements 3. Rely on manufacturer s data 4. Remote audit 5. Indep. meas. from another physicist

28 All of the following are current methods to provide a verification of your small field dosimetry data except: The answer is: Reliance on the manufacturers data Ref: Das et al, AAPM TG 155 report, submitted to Med Phys

29 The challenge continues for the <2 x 2 cm 2 field sizes The correct dosimeter is essential. There are independent audits and standard data to help. We will defeat this dragon! Summary Conclusion

30 Thank You Questions?

31 References Francescon et al. Medical Physics, Vol. 38, No. 12, December 2011 Das et al, AAPM TG 155 report, submitted to Med Phys Followill et al. J Appl Clin Med Phys 13(5):282-9, Followill DS: Erratum, J Appl Clin Med Phys 15 (2): /2014.

32 Acknowledgements Work supported by PHS grant CA10953 awarded by NCI, DHHS

SRS/SBRT Errors and Causes

SRS/SBRT Errors and Causes SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Sang Hyun Cho and Geoffrey S. Ibbott Department of Radiation Physics The

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

Basic dosimetric data for treatment planning system

Basic dosimetric data for treatment planning system Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1) Indra J. Das/ Small Fields (Page-1) Small Field Dosimetry: Overview of AAP TG-155 and the IAEA-AAP Code of Practice (Therapy) Indra J. Das, PhD, FAAP, FACR, FASTRO Department of Radiation Oncology Indiana

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Small Field Dosimetry

Small Field Dosimetry Small Field Dosimetry Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin Madison WI Conflict

More information

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data J. A. Bencomo, N. Hernandez, and W.F. Hanson University of Texas M.D. Anderson Cancer Center Medical physics,

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output Presented by: Rebecca M. Howell PhD, DABR, FAAPM Results from Voluntary Independent External Peer Review of Beam Output What do we hope to achieve with independent peer review? Independent peer review

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

IMRT QA: Point Dose Measurements or 2D Array?

IMRT QA: Point Dose Measurements or 2D Array? QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment Journal of Physics: Conference Series PAPER OPEN ACCESS Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment To cite this article: R Marzuki et al 2018 J. Phys.:

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Megan A. Hyun, M.S. Under the supervision of Larry A. DeWerd, Ph.D. NCCAAPM Fall Meeting October 30, 2015 Medical Radiation Research Center

More information

Clinical Implementation of SRS/SBRT

Clinical Implementation of SRS/SBRT Clinical Implementation of SRS/SBRT Anil Sethi, PhD, FAAPM Loyola University Medical Center November 4, 2017 Disclosures Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch 2 Learning

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

ph fax

ph fax This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY

More information

Small-Field Dosimetry: Its Importance In Clinic

Small-Field Dosimetry: Its Importance In Clinic Indra J. Das/ Small Fields (Page-1) Small-Field Dosimetry: Its Importance In Clinic isadministration edia Coverage IJDas (1) Indra J. Das, PhD, FAAP, FACR, FASTRO Department of Radiation Oncology Indiana

More information

Detector comparison for small field output factor measurements with flattening filter free photon beams

Detector comparison for small field output factor measurements with flattening filter free photon beams Detector comparison for small field output factor measurements with flattening filter free photon beams 1 Wolfgang Lechner, 1 Lukas Sölkner, 2 Paulina Grochowska, 1 Dietmar Georg 1 Christian Doppler Laboratory

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Dosimetry of the Gamma Knife TM

Dosimetry of the Gamma Knife TM Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada IAEA/AAPM Code o Practice or the Dosimetry o Static Small Photon Fields Jan Seuntjens McGill University Montréal, Canada Acknowledgements IAEA/AAPM small and composite ield working group: Hugo Palmans

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit MR-Linac is a research programme. It is not available for sale and its future availability cannot be guaranteed 8/2/2017 Clinical Implementation of an MR-Guided Treatment Unit Geoffrey S. Ibbott, Ph.D.

More information

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of: Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

The clinical use of OSLD

The clinical use of OSLD The clinical use of OSLD Stephen F. Kry PhD Radiological Physics Center, Imaging and Radiation Oncology Core - Houston IROC Houston The University of Texas MD Anderson Cancer Center, Houston, TX Spring

More information

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Professor, Department of Medical Physics

More information

Major accidents in radiotherapy

Major accidents in radiotherapy Major accidents in radiotherapy related to treatment units (a) IAEA International Atomic Energy Agency Incorrect decay data (USA) IAEA International Atomic Energy Agency Background A cobalt unit was used

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Patient and Linac QA - present and the future

Patient and Linac QA - present and the future Patient and Linac QA - present and the future IBA Dosimetry Sandra Kos Service Department Customer Support Specialist sandra.kos@iba-group.com 1 IBA Dosimetry Part of IBA Group OUR MISSION At IBA we dare

More information

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

More information

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy

More information

DEPT OF RADIATION ONCOLOGY DEPT OF RADIATION ONCOLOGY

DEPT OF RADIATION ONCOLOGY DEPT OF RADIATION ONCOLOGY Disclosures Lessons Learned from Misadministrations and Accidents in Radiation Therapy: Teletherapy Ryan Foster, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX I receive

More information

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields Dosimetric Uncertainties in Reerence and Relative Dosimetry o Small Fields Jan Seuntjens, Ph.D., FAAPM, FCCPM McGill University Health Centre Canada Outline o Presentation Uncertainty concepts and requirements

More information

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)

More information

IMRT point dose measurements with a diamond detector

IMRT point dose measurements with a diamond detector IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,

More information

Dosimetry of Small Static Fields Used in External Beam Radiotherapy

Dosimetry of Small Static Fields Used in External Beam Radiotherapy Technical Reports SeriEs No.483 Dosimetry of Small Static Fields Used in External Beam Radiotherapy An International Code of Practice for Reference and Relative Dose Determination Sponsored by the IAEA

More information

RPC s Credentialing Programs for Clinical Trials

RPC s Credentialing Programs for Clinical Trials RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups

More information

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with

More information

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D. UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe 3D printing of air slab caps for correction free small field dosimetry Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe Small fields? Defined as fields sizes too small to achieve

More information

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams

More information

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results A.Piermattei 1, L. Azario 1, S. Cilla 2, A.Fidanzio 1, F.Greco 1, M.T.Russo 3, S. Zucca 4 1 Istituto di Fisica e Unità

More information

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011 MVCT Image Guidance and QA Robert Staton, PhD DABR MD Anderson Cancer Center Orlando ACMP Annual Meeting 2011 Disclosures MDACCO has received grant funding from TomoTherapy, Inc. Overview TomoTherapy MVCT

More information

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications Varian Treatment Streamlined Treatment Delivery Management Application Specifications Specifications Varian Treatment 1 Introduction Streamlined Treatment Delivery Management Varian Treatment verifies

More information

Disclosures. Clinical Implementation of SRS/SBRT. Overview. Anil Sethi, PhD. Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch

Disclosures. Clinical Implementation of SRS/SBRT. Overview. Anil Sethi, PhD. Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch Clinical Implementation of SRS/SBRT Anil Sethi, PhD Loyola University Medical Center August 3, 2017 Disclosures Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch 2 Overview Physics Considerations

More information

SunCHECK Patient Comprehensive Patient QA

SunCHECK Patient Comprehensive Patient QA The SunCHECK Platform SunCHECK Patient Comprehensive Patient QA SunCHECK Patient // Comprehensive Patient QA Automate Your Patient QA SunCHECK Patient automates all your Patient QA needs from Secondary

More information

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS SB Crowe, E Whittle, C Jones, T Kairn SMALL FIELD DOSIMETRY The measurement of small field tissue-maximum ratio (TMR) and percentage depth dose (PDD)

More information

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a The Equipment : safety, commissioning, re commissioning calibration : Equipements : sûreté, recette, modification calibration A medical physicist point of view. Albert Lisbona Medical Physics Department

More information

Quality Assurance of Helical Tomotherapy Machines

Quality Assurance of Helical Tomotherapy Machines Quality Assurance of Helical Tomotherapy Machines Wolfgang Tomé 1,2, Hazim Jaradat 2, David Westerly 1, John Fenwick 3, Nigel Orton 2, Rock Mackie 1,2, and Bhudatt Paliwal 1,2 1 Human Oncology and 2 Medical

More information

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,

More information

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS Romanian Reports in Physics, Vol. 64, No. 3, P. 728 736, 2012 MEDICAL PHYSICS IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS G. VASILE 1,3, MADALINA VASILE 2,3, O.G. DULIU 3 1 Center of

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free

More information

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Rhys Carrington, Ysbyty Glan Clwyd, Bodelwyddan. MSc Medical Radiation Physics. Small field dosimetry

More information

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D. ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

Much of what we will say

Much of what we will say New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer

More information

Quality Assurance in Radiation Therapy

Quality Assurance in Radiation Therapy Quality Assurance in Radiation Therapy Purpose (from George) To make aware of the needs for quality assurance in radiation therapy To identify components of a quality assurance program in radiation therapy

More information

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system JBUON 2018; 23(1): 179-184 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Patient-specific quality assurance for intracranial cases in robotic

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry Journal of Physics: Conference Series PAPER OPEN ACCESS Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry To cite this article: N A Costa et al 2016 J. Phys.: Conf. Ser. 733 012098 View

More information

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC) Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf

More information

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK

More information

On the use of bolus for pacemaker dose measurement and reduction in radiation therapy

On the use of bolus for pacemaker dose measurement and reduction in radiation therapy Received: 8 February 2017 Revised: 16 October 2017 Accepted: 23 October 2017 DOI: 10.1002/acm2.12229 RADIATION ONCOLOGY PHYSICS On the use of bolus for pacemaker dose measurement and reduction in radiation

More information

Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea

Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea Jung-in Kim, 1 Jin-Beom Chung, 2 Ju-Young

More information

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies Abdul Kazi, Ph.D., DABR Assistant Professor, Department of Radiation Oncology University of Maryland

More information

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom Gary Fisher, B.S. David Followill, Ph.D. Geoffrey Ibbott, Ph.D. This investigation

More information

The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist

The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist The Radixact System Experience at Miami Cancer Institute Tino Romaguera, D.Sc. Senior Physicist Disclosure An honorarium is provided by Accuray for this presentation The views expressed in this presentation

More information