Quality assurance in external radiotherapy

Size: px
Start display at page:

Download "Quality assurance in external radiotherapy"

Transcription

1 Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas,

2 Acceptance and Commissioning

3 Acceptance Test Testing if machine and radiation parameters is within specified limits Used Customer Acceptance Procedures or International protocols Tests for all equipment: machines, planning systems, verification systems and et. 3

4 Acceptance Test 4 Customer Acceptance Procedures, Varian

5 Commissioning Acquiring all radiation beam data required for treatment Entering these data into a computerized treatment planning system Developing all dosimetry, treatment planning and treatment procedures Verifying the accuracy of these procedures Establishing quality control tests and procedures Training all personnel 5

6 Commissioning 6 Faiz M. Khan, The Physics of Radiation Therapy, 2003

7 Quality Assurance 7

8 Daily QA (Linac) 8 Klein et al., Task Group 142 Report: QA of Medical Accelerators, Med. Phys.,36 (9), 2009, p

9 Monthly QA (Linac) 9 Klein et al., Task Group 142 Report: QA of Medical Accelerators, Med. Phys.,36 (9), 2009, p

10 Annual QA (Linac) 10 Klein et al., Task Group 142 Report: QA of Medical Accelerators, Med. Phys.,36 (9), 2009, p

11 Annual QA (Linac) 11 Klein et al., Task Group 142 Report: QA of Medical Accelerators, Med. Phys.,36 (9), 2009, p

12 Wedges, MLC, Imaging QA (Linac) 12 Klein et al., Task Group 142 Report: QA of Medical Accelerators, Med. Phys.,36 (9), 2009, p

13 QA of Cobalt-60 Kutchar et al., Report of AAPM Radiation Therapy Committee Task Group 40, Med. Phys., 21(4), 1994, p

14 QA of Simulator Kutchar et al., Report of AAPM Radiation Therapy Committee Task Group 40, Med. Phys., 21(4), 1994, p

15 QA of Kilavoltage Equipment J.R. Williams et al., Radiotherapy Physics in Practice,

16 QA after Service (Linac) Beam Quality Assurance Guidelines High Energy Clinacs, Customer Technical Bulletin, Varian 16

17 Real Situation 17

18 Roles and Responsibilities of Medical Physicists Faiz M. Khan, The Physics of Radiation Therapy,

19 Radiotherapy Clinics 5 Clinics: Institute of Oncology, Vilnius University Hospital of Kaunas University of Medicine Oncological Hospital of Kaunas Medical University Hospital Klaipėda University Hospital Šiauliai Hospital 19

20 Radiotherapy (external) Equipment Equipment Linac Simulator Institute of Oncology, Vilnius University 3 Varian (2) 6, 15 MV Varian (1) 6 MV IMRT Varian Hospital of Kaunas University of Medicine 3 Varian (2) 6, 15 MV Siemens 6, 18 MV Stereotactic CT simulator GE X-ray therapy In future plans - Oncological Hospital of Kaunas Medical University Hosp. 1 Varian 6, 15 MV Varian Gulmay kv Klaipėda University Hospital 1 Siemens 6, 18 MV CT simulator Siemens Gulmay kv Šiauliai Hospital 1 Elekta 6, 15 MV CT Gulmay kv Treatment planning Eclipse Eclipse Eclipse Xio Precise Plan system Co-60 Russia - - Russia - Verification system Aria Aria Aria Lantis Mosaiq 20

21 Medical Physicists Institute of Oncology, Vilnius University: 8 medical physicists Hospital of Kaunas University of Medicine: 6 medical physicists Oncological Hospital of Kaunas Medical University Hospital: 4 medical physicists Klaipėda University Hospital: 4+1 (arriving) medical physicists Šiauliai County Hospital: 2+3 (arriving) medical physicists 21

22 Equipment for QA Lithuanian Hygiene Standard HN 95:2005 Radiation Protection and Quality Control in Radiotherapy 22

23 QA in Oncological Hospital (Linac) Daily (Rad. Technologist): Door interlock Beam on indicator Audiovisual monitor Laser localization Optical distance indicator X-ray output constancy X-ray beam symmetry X-ray beam flatness Photon beam energy constancy Radiation field size Monthly: X-ray output constancy Photon beam energy constancy Light/ radiation field coincidence Emergency beam off Quarterly: PDD constancy X-ray symmetry constancy X-ray flatness constancy Mechanical tests Annually: Customer Acceptance procedures Commissioning procedures (part) 23

24 QA in Oncological Hospital (Kilavoltage Unit) Weekly: Door interlock Beam on indicator Calibration constancy Annually: Absolute calibration Beam Quality (HVL) 24

25 QA in Oncological Hospital (Simulator, CT) Simulator (Quarterly, Annually): Lithuanian Hygiene Standard HN 78:2009 Quality Control in Radiology CT (Daily, Weekly, Annually): Lithuanian Hygiene Standard HN 78:2009 Quality Control in Radiology 25

26 QA Procedure QA in Lithuania (Linac) Institute of Oncology, Vilnius University Oncological Hospital of Kaunas Medical University Hospital Klaipėda University Hospital Šiauliai Hospital Daily Dosimetry (quick check) Dosimetrist Rad. Technologist Med. Physicist Rad. Technologist Mechanical Dosimetrist Rad. Technologist Rad. Technologist Rad. Technologist Safety Dosimetrist Rad. Technologist Rad. Technologist Rad. Technologist Weekly Mechanical Med. Physicist Med. Physicist Safety Med. Physicist Monthly Dosimetry (water phantom) Med. Physicist Med. Physicist Med. Physicist Med. Physicist Mechanical Med. Physicist Med. Physicist Med. Physicist Med. Physicist Safety Med. Physicist 3 Months Dosimetry (big water phantom) Mechanical Dosimetry (big water phantom) Med. Physicist Med. Physicist 6 Months Med. Physicist Med. Physicist Med. Physicist Annually Dosimetry (big water phantom) Med. Physicist Med. Physicist Med. Physicist Med. Physicist Mechanical Med. Physicist Med. Physicist Med. Physicist Med. Physicist Safety Med. Physicist Med. Physicist 26

27 QA in Lithuania (Kilavoltage Unit) QA Procedure Oncological Hospital of Kaunas Medical University Hospital Klaipėda University Hospital Šiauliai Hospital Daily Safety Rad. Technologist Rad. Technologist Weekly Dosimetry Med. Physicist Safety Med. Physicist Monthly Dosimetry Med. Physicist Med. Physicist 3 Months Dosimetry Annually Dosimetry Med. Physicist Med. Physicist Med. Physicist 27

28 Thank you for your attention

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Sang Hyun Cho and Geoffrey S. Ibbott Department of Radiation Physics The

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Patient and Linac QA - present and the future

Patient and Linac QA - present and the future Patient and Linac QA - present and the future IBA Dosimetry Sandra Kos Service Department Customer Support Specialist sandra.kos@iba-group.com 1 IBA Dosimetry Part of IBA Group OUR MISSION At IBA we dare

More information

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis Acknowledgments Thanks to Saiful Huq for an illuminating conversation about the application of TG-100 Jennifer O Daniel, Ph.D. & Fang-Fang Yin, Ph.D. Duke University Medical Center Annual AAPM Meeting,

More information

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy

More information

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Indiana University Health Proton Therapy Center. Chee-Wai Cheng, Ph.D.

Indiana University Health Proton Therapy Center. Chee-Wai Cheng, Ph.D. Indiana University Health Proton Therapy Center Chee-Wai Cheng, Ph.D. Machine configuration and layout Fixed beam line, double scattering and propeller Uniform scanning nozzle and snout Gantry room Range

More information

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,

More information

I. Equipments for external beam radiotherapy

I. Equipments for external beam radiotherapy I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Quality Assurance in Radiation Therapy

Quality Assurance in Radiation Therapy Quality Assurance in Radiation Therapy Purpose (from George) To make aware of the needs for quality assurance in radiation therapy To identify components of a quality assurance program in radiation therapy

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Implementation of advanced RT Techniques

Implementation of advanced RT Techniques Implementation of advanced RT Techniques Tibor Major, PhD National Institute of Oncology Budapest, Hungary 2. Kongres radiološke tehnologije, Vukovar, 23-25. September 2016. Current RT equipments at NIO,

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

DEPT OF RADIATION ONCOLOGY DEPT OF RADIATION ONCOLOGY

DEPT OF RADIATION ONCOLOGY DEPT OF RADIATION ONCOLOGY Disclosures Lessons Learned from Misadministrations and Accidents in Radiation Therapy: Teletherapy Ryan Foster, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX I receive

More information

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of: Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization

More information

Who Should Know Radiation Oncology Coding?

Who Should Know Radiation Oncology Coding? Why Should We Learn Radiation Oncology Coding? Terry Wu, Ph.D. Chief Physicist Radiation Oncology Department Willis-Knighton Cancer Center Who Should Know Radiation Oncology Coding? Radiation Oncologist

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

RADIOTHERAPY- CURRENT SITUATION AND FUTURE TRENDS

RADIOTHERAPY- CURRENT SITUATION AND FUTURE TRENDS HOSPITAL OF LITHUANIAN UNIVERSITY OF HEALTH SCIENCES KAUNO KLINIKOS RADIOTHERAPY- CURRENT SITUATION AND FUTURE TRENDS Prof. Elona Juozaitytė Perspectives of Czech- Lithuanian Research Partnerships About

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output Presented by: Rebecca M. Howell PhD, DABR, FAAPM Results from Voluntary Independent External Peer Review of Beam Output What do we hope to achieve with independent peer review? Independent peer review

More information

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data J. A. Bencomo, N. Hernandez, and W.F. Hanson University of Texas M.D. Anderson Cancer Center Medical physics,

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

SRS/SBRT Errors and Causes

SRS/SBRT Errors and Causes SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research

More information

A new geometric and mechanical verification device for medical LINACs

A new geometric and mechanical verification device for medical LINACs JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein

More information

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a The Equipment : safety, commissioning, re commissioning calibration : Equipements : sûreté, recette, modification calibration A medical physicist point of view. Albert Lisbona Medical Physics Department

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or

More information

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS 6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types

More information

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,

More information

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?

More information

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D. UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor IMRT/IGRT Patient Treatment: A Community Hospital Experience Charles M. Able, Assistant Professor Disclosures I have no research support or financial interest to disclose. Learning Objectives 1. Review

More information

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free

More information

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications Varian Treatment Streamlined Treatment Delivery Management Application Specifications Specifications Varian Treatment 1 Introduction Streamlined Treatment Delivery Management Varian Treatment verifies

More information

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D. ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael

More information

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR Therapeutic Medical Physics Stephen J. Amadon Jr., Ph.D., DABR Outline 1. Why physicists are needed in medicine 2. Branches of medical physics 3. Physics in Radiation Oncology 4. Treatment types and Treatment

More information

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT Purpose The purpose of this curriculum outline is to provide a framework for multidisciplinary training for radiation oncologists, medical

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research

More information

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies Abdul Kazi, Ph.D., DABR Assistant Professor, Department of Radiation Oncology University of Maryland

More information

Overview of TG225 - Medical Physics Practice Guideline #1

Overview of TG225 - Medical Physics Practice Guideline #1 Overview of TG225 - Medical Physics Practice Guideline #1 Evaluation and quality assurance of x-ray based image guided radiotherapy systems Jonas Fontenot, Ph.D. MPPG #1 Evaluation and Quality Assurance

More information

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa International Journal of Sciences: Basic and Applied Research (IJSBAR) ISSN 237-4531 (Print & Online) http://gssrr.org/index.php?journal=journalofbasicandapplied ---------------------------------------------------------------------------------------------------------------------------

More information

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams

More information

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital EBT Process Diagnosis Simulation Tx Planning Tx Verification Tx Delivery X-ray CT MRI NM Patho Process

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

RTOG DOSIMETRY DATA SUBMISSION

RTOG DOSIMETRY DATA SUBMISSION Radiation Therapy Oncology Group American College of Radiology 1818 Market Street, Suite 1600 Philadelphia, PA 19103-3604 (215) 574-3189 (800) 227-5463 Ext. 4189 (215) 928-0153 Fax Phoenix, Arizona RTOG

More information

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery

More information

Office: Cell Ph:

Office: Cell Ph: Rajeev K. Badkul, M.S., DABR Lead Medical Physicist & Assistant Professor Dept. of Radiation Oncology University Of Kansas Medical Center 3901 Rainbow Blvd., Kansas City, KS,66160 rbadkul@kumc.edu Office:

More information

iviewdose Confidence and assurance in dose delivery

iviewdose Confidence and assurance in dose delivery iviewdose Confidence and assurance in dose delivery As radiotherapy techniques advance and treatment plans become more complex, involving escalated doses and increasingly conformal deliveries, the need

More information

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,

More information

Understanding Radiation Therapy. For Patients and the Public

Understanding Radiation Therapy. For Patients and the Public Understanding Radiation Therapy For Patients and the Public Introduction to Radiation Oncology Radiation has been an effective tool for treating cancer for more than 100 years. Radiation oncologists are

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Amendment No. 2. Item No. 2 (Rfx/ Event number ) Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have

More information

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 CyberKnife Technology in Ablative Radiation Therapy Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 Objectives Components and work flow of CyberKnife Motion management of CyberKnife

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session Stereotactic Body Radiation Therapy Quality Assurance Educational Session J Perks PhD, UC Davis Medical Center, Sacramento CA SBRT fundamentals Extra-cranial treatments Single or small number (2-5) of

More information

ph fax

ph fax This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY

More information

SunCHECK Patient Comprehensive Patient QA

SunCHECK Patient Comprehensive Patient QA The SunCHECK Platform SunCHECK Patient Comprehensive Patient QA SunCHECK Patient // Comprehensive Patient QA Automate Your Patient QA SunCHECK Patient automates all your Patient QA needs from Secondary

More information

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA Catharine Clark NPL, Royal Surrey County Hospital and RTTQA Background Historically UK has been relatively independent of the rest of the world in audit NPL PSDL International intercomparison for reference

More information

Objective. Evaluation of Linear Accelerator Performance Standards using an Outcome Oriented Approach. Linac performance standards*

Objective. Evaluation of Linear Accelerator Performance Standards using an Outcome Oriented Approach. Linac performance standards* Quality in Radiation Therapy: what is it and how do you achieve it?. Overview of definitions and approaches to Quality Pawlicki. ROSIS Knöö öös. Peer Review Quality Audits Halvorsen. The Regulator s s

More information

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...

More information

Standards for Radiation Oncology

Standards for Radiation Oncology Standards for Radiation Oncology Radiation Oncology is the independent field of medicine which deals with the therapeutic applications of radiant energy and its modifiers as well as the study and management

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

In-Room Radiographic Imaging for Localization

In-Room Radiographic Imaging for Localization In-Room Radiographic Imaging for Localization Fang-Fang Yin, Zhiheng Wang, Sua Yoo, Devon Godfrey, Q.-R. Jackie Wu Department of Radiation Oncology Duke University Medical Center Durham, North Carolina

More information

Work partially supported by VisionRT

Work partially supported by VisionRT Work partially supported by VisionRT Background of frameless intracranial stereotactic radiosurgery UCSD SRS/SRT procedure Clinical Results Summary Total prescribed doses : order of 10 50 Gy Planning targets

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

McGill University Medical Physics Unit Radiotherapy Lab: MDPH 603 (2 credits) Winter 2017

McGill University Medical Physics Unit Radiotherapy Lab: MDPH 603 (2 credits) Winter 2017 McGill University Medical Physics Unit Radiotherapy Lab: MDPH 603 (2 credits) Winter 2017 Coordinators: Marija Popovic, Steve Davis Office hours: Monday 15:00-16:00, DS1.5116, McGill University Health

More information

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West

More information

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results A.Piermattei 1, L. Azario 1, S. Cilla 2, A.Fidanzio 1, F.Greco 1, M.T.Russo 3, S. Zucca 4 1 Istituto di Fisica e Unità

More information

Protura Robotic Patient Positioning System. for efficiency + performance

Protura Robotic Patient Positioning System. for efficiency + performance Protura Robotic Patient Positioning System for efficiency + performance Protura Robotic Patient Positioning System The Protura Robotic Patient Positioning System is the ultimate in robotic patient motion

More information

Curriculum Vitae. Iavor Veltchev, Ph.D., DABR

Curriculum Vitae. Iavor Veltchev, Ph.D., DABR Curriculum Vitae Iavor Veltchev, Ph.D., DABR CONTACT INFORMATION Iavor Veltchev E-mail: iavor.veltchev@fccc.edu Fox Chase Cancer Center office: +1 215 728 2854 Department of Radiation Oncology fax: +1

More information

Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST

Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST What is a Medical Physicist? Medical physics application of physics concepts

More information

Image Registration: The Challenge for QA Centers

Image Registration: The Challenge for QA Centers Image Registration: The Challenge for QA Centers Kenneth Ulin, Ph.D. Marcia M. Urie, Ph.D. Quality Assurance Review Center Imaging Studies Used for: Staging Protocol Eligibility Target Volume Definition

More information

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors S. Sathiyan, M. Ravikumar, C. Varatharaj, S.S. Supe, S.L. Keshava Department of Radiation Physics, Kidwai

More information

Commissioning and Radiobiology of the INTRABEAM System

Commissioning and Radiobiology of the INTRABEAM System Commissioning and Radiobiology of the INTRABEAM System Susha Pillai and Junan Zhang Scheme INTRABEAM System. Physics Commissioning, QA, and Radiation Protection Radiobiology 1 Disclosure OHSU is an INTRABEAM

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring

More information

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Major accidents in radiotherapy

Major accidents in radiotherapy Major accidents in radiotherapy related to treatment units (a) IAEA International Atomic Energy Agency Incorrect decay data (USA) IAEA International Atomic Energy Agency Background A cobalt unit was used

More information

Treatment Efficiency and Optimization of Patient Care with IBA ProteusOne

Treatment Efficiency and Optimization of Patient Care with IBA ProteusOne Treatment Efficiency and Optimization of Patient Care with IBA ProteusOne Terry Wu, Ph.D. Chief Physicist, Radiation Oncology Department Willis-Knighton Cancer Center/Proton Therapy Center Shreveport,

More information

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment Oncology and Translational Medicine DOI 10.1007/s10330-016-0183-3 December 2017, Vol. 3, No. 6, P260 P266 REVIEW ARTICLE A comparative study between flattening filter-free beams and flattening filter beams

More information