IMRT QA: Point Dose Measurements or 2D Array?
|
|
- Hilary Henderson
- 5 years ago
- Views:
Transcription
1 QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify dose delivery Andrew Hwang, PhD Necessity of IMRT QA Potential error sources Planning R&V Delivery accuracy of dose delivery should be documented by irradiating a phantom containing a calibrated dosimetry to verify that the dose delivered is the dose planned. ACR-ASTRO practice guideline 1
2 Objective Describe our experience with two different IMRT QA methods Compare the two methods Discuss ideas about interpreting results Phantom Measurements Two plastic phantoms Cylindrical phantom Slab phantom Phantom size chosen to approximate treatment site Dose measurements Central ion chamber Peripheral MOSFETs Procedure Copy treatment plan to CT scan of phantom Adjust isocenter position as needed Determine planned dose to ion chamber / MOSFET s Deliver plan Compare measurement w/ expected dose Results N=77 Mean deviation -0.1±.0% (ion chamber) 1.3 ±.3% (MOSFET) Consistently able to achieve our goal of 5% tolerance.
3 Phantom measurement pros and cons Advantages End-to-end check Easy to interpret results Disadvantages Checks limited portion of field Harder to set-up May be difficult to obtain good results in high gradient regions D Arrays Popular tool for performing IMRT QA Available from multiple vendors Sun Nuclear MapCheck 157 diodes, 7 mm spacing 3x6 cm active area PTW seven9 79 ion chambers 7x7 cm area IBA Matrixx 100 ion chambers cm area MapCheck Array Questions Pass / Fail criteria? Best way to use array? Rotating with gantry vs. stationary Field by field vs. composite What information can be determined from data? array Mounting frame 3
4 Gamma Analysis Method used for comparing two distributions Combines dose difference and distance to agreement (DTA) γ Γ ( r ) = min{ Γ( r, r )} [ r] m ( r, r ) m c = r m c ( r, r ) δ ( r, r ) + m c m c d M D M c Pass/Fail criteria 3%/3mm is the most common criteria chosen for gamma analysis (Nelms & Simon, 007) AAPM TG-119 (IMRT commissioning) recommends a 90% pass rate for 3%/3mm for per field analysis Stricter criteria may be more sensitive to dosimetric / MLC errors DTA criteria Dose criteria Value less than 1 = satisfactory result See Low, Harms, Mutic, and Purdy, Med Phys, vol. 5, Results Pass rates dependent on resolution of planar dose calculation Switching fluence (and dose grid) resolution from 4 mm to mm improved average passing rate Average gamma (3%/3mm) pass rates 97.1±.7% Composite pass rates did not correlate with field by field measurements. Average pass rate (3%/3mm) was 96.8±3.9% Gamma pass rates did not vary by beam angle Composite vs. Individual Fields Individual Field Pass Rate (%) Mean R = Minimum R = Composite Pass Rate (%) Pass rate for composite plotted vs. individual field pass rates Composite 3mm/3% pass rate: 96.8±3.8% 4
5 1 5 9 Effect of Gantry Angle Data from one linac plotted here ANOVA shows differences not significant Control Plots Plot of process vs. time Have control limits ( specifications) Based on short-term variation x± 3σ short n Proper design needed for maximum utility Help to identify special cause variation but does not identify cause Control Plots KD Potential error sources Plotted data by machine Subgroup by patient Red lines show control limits Pass Rate 3%/3mm (%) Patient # 9 8 KD Planning R&V Delivery Measurement & Analysis Standard Deviation (%) Patient # 5
6 Interpreting Results Unclear relationship between gamma index pass rates and 3D dose distribution MLC errors may be detected by D array measurements (Yan et al, 009). Systematic MLC errors (1- mm) result in significant dosimetric changes (Mu et al, 008) Gamma analysis of individual field measurements is insensitive to inaccuracies of overall plan (Kruse, 010) Gamma pass rates do not predict clinically relevant errors (Nelms et al, 011) Array Pros and Cons Advantages Check larger portion of the field Potentially able to detect atic MLC position errors Simple to use, less user dependent Quicker to set-up Disadvantages Difficult to interpret results Time Comparison Phantom Array Plan Preparation 5 minutes 5 minutes Set-up / take down time 0-5 minutes minutes Delivery Time minutes minutes Analysis / Documentation 5 minutes 10 minutes Total (1 plan) minutes minutes Conclusions Compared and contrasted two methods to perform patient-specific IMRT QA Most (~75%) IMRT QA currently done with D array Results demonstrated that our IMRT delivery is accurate Methods have different strengths and weaknesses and provide different information Tests are screening and not diagnostic Total (5 plans) minutes minutes 6
7 Acknowledgements Jeff Bellerose, Angelica Perez-Andujar, Devan Krishnamurthy, Josephine Chen, Ping Xia. References G Mu, E Ludlum, and P Xia, Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancer, Phys Med Biol, vol. 53, pp , 008. JJ Kruse, On the insensitivity of single field planar dosimetry to IMRT inaccuracies, Med Phys, vol 37, pp , 010. BE Nelms, JA Simon, A Survey of planar IMRT QA analysis, J Appl Clin Med Phys, vol. 8, pp. 1-15, 007. BE Nelms, H Zhen, and W Tome, Per-beam IMRT QA passing rates do not predict clinically relevant patient dose errors, Med Phys, vol 38, pp , 011. G Yan, C Liu, TA Simon, L Peng, C Fox, JG Li, On the sensitivity of patient-specific IMRT QA to MLC positioning errors, J Appl Clin Med Phys, vol. 10, pp , 009. GA Ezzell, JW Burmeister, et al, IMRT commissioning: multiple institution planning and dosimetry comparisons: a report from AAPM task group 119, Med Phys, vol 36, pp ,
GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119
GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119 M. DUMITRACHE 1,3, Alina TĂNASE 2,3 1 Institute of Oncology Prof. Dr. Al. Trestioreanu, Bucharest,
More informationGAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119
Romanian Reports in Physics, Vol. 68, No. 3, P. 1130 1139, 2016 GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119 M. DUMITRACHE 1,3, Alina TĂNASE
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationEvaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell
Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating Characteristic (ROC) Analysis by Allison Lorraine Mitchell Graduate Program in Medical Physics Duke University Date: Approved:
More informationA comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy
Radiology and Oncology Ljubljana Slovenia www.radioloncol.com research article 307 A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy Jaeman Son 1,2,
More informationThe MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates Daniel W. Bailey, 1,2 a Jason D. Spaans, 2
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationCHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA
CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy
More informationVolumetric Modulated Arc Therapy - Patient Specific QA
Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck
More informationQuality assurance and credentialing requirements for sites using inverse planned IMRT Techniques
TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery
More informationEvaluation of Dosimetry Check software for IMRT patient-specific quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,
More informationConfidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea Jung-in Kim, 1 Jin-Beom Chung, 2 Ju-Young
More informationA model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms
1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,
More information8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient
Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,
More informationSunCHECK Patient Comprehensive Patient QA
The SunCHECK Platform SunCHECK Patient Comprehensive Patient QA SunCHECK Patient // Comprehensive Patient QA Automate Your Patient QA SunCHECK Patient automates all your Patient QA needs from Secondary
More informationMeasurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry
Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Conflict of interest None Acknowledgement: Jim
More informationAre Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?
Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or
More informationDebate Motion. Theoretical benefits of VMAT: Real Costs of VMAT. "Arc Based Techniques Will Make Conventional IMRT Obsolete
Austin, TX July 21, 2014 Debate Motion "Arc Based Techniques Will Make Conventional IMRT Obsolete AGAINST the Motion: Peter Balter, Ph.D. Potential Conflicts of Interest: Dr. Balter has research support/collaborations
More informationSRS/SBRT Errors and Causes
SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research
More informationVerification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,
More informationDosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar
More informationEric E. Klein, Ph.D. Chair of TG-142
Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing
More informationDosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy Francis Kar-ho Lee, a Simon Kar-yiu
More informationCan we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationLearning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy
3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy Sofie Ceberg, PhD Medical Physicist and Lund University Lund, Sweden Learning objectives 1: 3D Dosimetry in the Clinic: Background
More informationIntroduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?
More informationAn Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center
An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions
More informationPMP. Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy.
Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.61 pissn 2508-4445, eissn 2508-4453 Gamma Evaluation with Portal Dosimetry for Volumetric Modulated
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationDosisverifikation mit Delta 4 Discover während der Behandlung
Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover
More informationEPID dosimetry confi guration and pre-treatment IMRT verifi cation
Rep Pract Oncol Radiother, 2007; 12(6): 307-312 Original Paper Received: 2007.04.26 Accepted: 2007.08.29 Published: 2007.12.27 Authors Contribution: A Study Design B Data Collection C Statistical Analysis
More informationPre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system
Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system E. Mezzenga, * E. Cagni, A. Botti, M. Orlandi and M. Iori Medical Physics Unit, IRCCS Arcispedale
More informationIMRT pre-treatment verification basic approaches
School of Medical Physics for Radiation Therapy: Dosimetry and Treatment Planning for Basic and Advanced Applications 27-march 7-april 2017 IMRT pre-treatment verification basic approaches Eugenia Moretti
More informationIBA Dosimetry Company Presentation. Patient Safety & Quality Control
IBA Dosimetry Company Presentation Patient Safety & Quality Control IBA Mission Protect, enhance and save lives Protect, Enhance, and Save Lives - 2 - The History of IBA Dosimetry A story of innovation
More informationVerification of Advanced Radiotherapy Techniques
Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December
More informationTraceability and absorbed dose standards for small fields, IMRT and helical tomotherapy
Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationIMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors
IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors S. Sathiyan, M. Ravikumar, C. Varatharaj, S.S. Supe, S.L. Keshava Department of Radiation Physics, Kidwai
More informationNormal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT
Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using
More informationS. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY
More informationEPID dosimetry for pretreatment quality assurance with two commercial systems
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 EPID dosimetry for pretreatment quality assurance with two commercial systems Daniel W. Bailey, 1,2a Lalith Kumaraswamy, 1 Mohammad
More information7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.
UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems
More informationEffective and efficient radiotherapy dosimetry audit: Where to next?
Effective and efficient radiotherapy dosimetry audit: Where to next? Catharine Clark, Eduard Gershkevitsh, Milan Tomsej and Nuria Jornet National Physical Laboratory Royal Surrey County Hospital Radiotherapy
More informationABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007
ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory
More informationStatistical process control analysis for patient-specific IMRT and VMAT QA
Journal of Radiation Research, 2013, 54, 546 552 doi: 10.1093/jrr/rrs112 Advance Access Publication 7 December 2012 Statistical process control analysis for patient-specific IMRT and VMAT QA Taweap SANGHANGTHUM
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the
More informationEPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 EPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams Narges Miri, 1a Peter Keller, 2 Benjamin
More informationVMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT
VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT Poster No.: C-0520 Congress: ECR 2011 Type: Scientific Exhibit Authors: Y. Kawasaki, S. Tadokoro,
More informationAgenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry
Controlli di qualità preclinici in IMRT statica e dinamica Michele Stasi S.C. Fisica Sanitaria, A.O. Ordine Mauriziano D.O. Fisica Sanitaria, IRCCS Candiolo TORINO mstasi@mauriziano.it Agenda 1. Introduction
More informationLearning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan
Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationEVALUATION OF INTENSITY MODULATED RADIATION THERAPY (IMRT) DELIVERY ERROR DUE TO IMRT TREATMENT PLAN COMPLEXITY AND IMPROPERLY MATCHED DOSIMETRY DATA
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2011 EVALUATION OF INTENSITY MODULATED RADIATION THERAPY (IMRT) DELIVERY
More informationPatient QA Recommended Publications
Patient QA Recommended Publications Your Most Valuable QA and Dosimetry Tools 3DVH Accuracy Studies Cross-Validation of Two Commercial Methods for Volumetric High-Resolution Dose Reconstruction on a Phantom
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationTreatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)
Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize
More informationUnrivaled, End-to-End
PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by
More informationbiij Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies Krishna Murthy K *
Available online at http://www.biij.org/2012/4/e28 doi: 10.2349/biij.8.4.e28 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Patient-specific quality assurance of RapidArc treatments:
More informationQuality Assurance of Helical Tomotherapy Machines
Quality Assurance of Helical Tomotherapy Machines Wolfgang Tomé 1,2, Hazim Jaradat 2, David Westerly 1, John Fenwick 3, Nigel Orton 2, Rock Mackie 1,2, and Bhudatt Paliwal 1,2 1 Human Oncology and 2 Medical
More informationImplementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy
Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out
More informationAbstract: A National Project for the in-vivo dosimetry in radiotherapy: first results
Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results A.Piermattei 1, L. Azario 1, S. Cilla 2, A.Fidanzio 1, F.Greco 1, M.T.Russo 3, S. Zucca 4 1 Istituto di Fisica e Unità
More informationDose rate response of Digital Megavolt Imager detector for flattening filter-free beams
Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free
More informationPatient-specific quality assurance for intracranial cases in robotic radiosurgery system
JBUON 2018; 23(1): 179-184 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Patient-specific quality assurance for intracranial cases in robotic
More informationFollow this and additional works at:
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2010 Commissioning and implementation of an EPID based IMRT system Dosimetry Check for 3D absolute dose measurements
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationComparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma
Lin et al. Radiation Oncology 2014, 9:71 RESEARCH Open Access Comparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma Hailei
More informationEvaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc
Original Article PMP Progress in Medical Physics 28(4), December 217 https://doi.org/1.14316/pmp.217.28.4.181 pissn 258-4445, eissn 258-4453 Evaluation of Dynamic Delivery Quality Assurance Process for
More informationph fax
This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY
More information8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation
Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research
More informationAssessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine
2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe
More informationAmendment No. 2. Item No. 2 (Rfx/ Event number )
Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have
More informationQuality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.
Quality Assurance of Ultrasound Imaging in Radiation Therapy Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Louis, Missouri Typical Applications of Ultrasound Imaging in Radiation Therapy
More informationDosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.87 Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Mee
More information6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS
6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationDevelopment and evaluation of an end-to-end test for head and neck IMRT with a novel multiple-dosimetric modality phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Development and evaluation of an end-to-end test for head and neck IMRT with a novel multiple-dosimetric modality phantom Viatcheslav
More informationPGY-1. Resident Review Session Schedule
1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT
More informationA VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 401 410, 2014 A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM * M. D. SUDITU 1,2, D. ADAM 1,2, R. POPA 1,2, V. CIOCALTEI
More informationDosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter
Applications of Nuclear Techniques (CRETE15) International Journal of Modern Physics: Conference Series Vol. 44 (2016) 1660221 (9 pages) The Author(s) DOI: 10.1142/S2010194516602210 Dosimetric characteristics
More informationWork partially supported by VisionRT
Work partially supported by VisionRT Background of frameless intracranial stereotactic radiosurgery UCSD SRS/SRT procedure Clinical Results Summary Total prescribed doses : order of 10 50 Gy Planning targets
More informationA review of RTTQA audit. Karen Venables
A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines
More informationSBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session
Stereotactic Body Radiation Therapy Quality Assurance Educational Session J Perks PhD, UC Davis Medical Center, Sacramento CA SBRT fundamentals Extra-cranial treatments Single or small number (2-5) of
More information8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit
MR-Linac is a research programme. It is not available for sale and its future availability cannot be guaranteed 8/2/2017 Clinical Implementation of an MR-Guided Treatment Unit Geoffrey S. Ibbott, Ph.D.
More informationSpatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.
Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation
More informationAn anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery
An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,
More informationMEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO
HOW TO QUALITY CONTROL AND PRE-TREATMENT QUALITY ASSURANCE APPLICATION OF EPID (as1000) FOR FF AND FFF BEAM VMAT PLANS Y. Mekuria 1, M. Bjorkqvist 1, J. Kulmala 1 1 Turku University Hospital/Radiotherapy
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationUse of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments
Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Jean M. Moran, Ph.D. Department of Radiation Therapy, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor
More informationEvaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf
More informationPre-treatment verification for IMRT / IGRT and Rotational therapy
In OmniPro-I mrt Vivo treatment software verification Pre-treatment verification for IMRT / IGRT and Rotational therapy Introduction Introduction Introduction Introtro Intro Intro The rapid development
More informationM. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course
Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics
More informationSpecial Procedures Rotation I/II SBRT, SRS, TBI, and TSET
University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,
More informationQuality assurance of volumetric modulated arc therapy using Elekta Synergy
Acta Oncologica, 2009; 48: 11931197 ORIGINAL ARTICLE Quality assurance of volumetric modulated arc therapy using Elekta Synergy AKIHIRO HAGA 1, KEIICHI NAKAGAWA 1, KENSHIRO SHIRAISHI 1, SAORI ITOH 1, ATSURO
More informationReliability of the gamma index analysis as a verification method of volumetric modulated arc therapy plans
Park et al. Radiation Oncology (2018) 13:175 https://doi.org/10.1186/s13014-018-1123-x RESEARCH Open Access Reliability of the gamma index analysis as a verification method of volumetric modulated arc
More informationOnline Care, Online Control, Online Confidence
Online Care, Online Control, Online Confidence * Online Treatment Monitoring and much more! Designed for Efficiency provides you with online control and online confidence of treating your patients exactly
More informationARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.
ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael
More informationDELIVERED PATIENT DOSE
SOFTWARE DELIVERED PATIENT DOSE The only software that provides a true picture of patient and dose at each stage of treatment. Adaptivo patient dosimetry software does something other patient dosimetry
More information