Clinical application of the Lens Opacities Classification System III in the performance of phacoemulsification
|
|
- Abigail Beasley
- 5 years ago
- Views:
Transcription
1 Clinical application of the Lens Opacities Classification System III in the performance of phacoemulsification James A. Davison, MD, Leo T. Chylack Jr., MD Purpose: To report the correlation of features of cataracts graded by the Lens Opacities Classification System, version III (LOCS III), with phacoemulsification energy expenditure and the balanced salt solution (BSS ) volume used during cataract surgery. Setting: Wolfe Clinic, Marshalltown, Iowa, USA. Methods: This was a retrospective review of 2364 cases operated on by a single surgeon from January 1998 to July 2000 in which the cataract had been graded at the slitlamp using the 4 grading scales of the LOCS III: nuclear opalescence (NO), nuclear color (NC), cortical cataract (C), and posterior subcapsular cataract (P). Polynomial best-fit lines were derived using regression analysis correlating the 4 preoperative LOCS III characteristics with 3 intraoperative observations: machine-measured phacoemulsification time, mean power expenditure, and BSS volume. Results: As determined by best-fit lines and their coefficient of determination (R 2 ), there were exponential relationships between machine-measured phacoemulsification time and the degree of NC (R ) and NO (R ). Trends existed between NC and NO and the amount of BSS used (R and R , respectively). No relationships were observed between the LOCS III classes of cataract, C and P, at any intraoperative observation. Conclusions: Exponentially, greater phacoemulsification energy was required as NC and NO increased. The LOCS III cataract grading system enhanced the ability to estimate ultrasonic energy expenditure and BSS volume use during phacoemulsification. Preoperative LOCS III cataract classification can help to create a more formally organized, integrated, customized operative plan. J Cataract Refract Surg 2003; 29: ASCRS and ESCRS The Lens Opacities Classification System, version III (LOCS III), is a widely used, scientifically valid, standardized photographic comparison system for grading the features of the human age-related cataract. 1 5 It has been used to grade the type and severity of cataract in Accepted for publication September 12, From the Wolfe Clinic, Marshalltown, Iowa, USA. Presented at the Symposium on Cataract, IOL and Refractive Surgery, San Diego, California, USA, April 2001 Reprint requests to James A. Davison, MD Wolfe Clinic, 309 East Church Street, Marshalltown, Iowa, USA. cross-sectional studies and the progression of cataract in longitudinal studies. It also has been used to grade cataracts at the slitlamp. 3 Grading using LOCS III involves the assessment of 4 features shown on 3 sets of photographs on an 8.5-inch 11-inch color transparency (Figure 1). The features of nuclear opacification and brunescence are graded according to 1 set of 6 photographs. The brightness of scatter from the nuclear region has been designated nuclear opalescence (NO) and the intensity of brunescence, nuclear color (NC). The amount of cortical cataract (C) is determined by comparing the estimated aggregate of cortical spoking to that seen in 2003 ASCRS and ESCRS /03/$ see front matter Published by Elsevier Science Inc. doi: /s (02)
2 Figure 1. (Davison) The LOCS III standard images in an 8.5-inch 11-inch color transparency as used in the office at the slitlamp. The top row contains the standards for NO and NC. The second row contains the standards for grading C and the bottom row, for grading P. 5 separate photographs. Similarly, the estimated amount of posterior subcapsular cataract (P) is determined by comparing it to another 5 photographs depicting increasing amounts of posterior subcapsular cataract. The grade for each feature is derived by locating the image of the patient s lens on the scale of severity for each feature represented in the color transparency. The NC and NO are graded on a decimal scale of 0.1 to 6.9. The severity of C and P are graded on a decimal scale of 0.1 to 5.9. The final LOCS III grade comprises 4 decimal values, 1 each for NO, NC, C, and P. In the original publication describing the LOCS III system, 1 an assessment of within-grader and betweengrader reproducibility was done and expressed as a 95% confidence interval for each scale. On average, these 95% confidence intervals were approximately 0.7 units. The intervals were used to define a significant change in the severity of the cataract. For example, if the NO grade was 2.3 at baseline and 1 year later it was 3.5, the increase in severity ( ) was greater than the 95% confidence interval (0.7) and it was therefore reasonable to conclude that one was 95% confident that the change was statistically significant. If the increase had been only 0.5 units, the increase could more likely be the result of the intrinsic methodologic noise of the LOCS III system and not a true increase in the severity of the cataract. Since 1998, LOCS III has been used systematically in the clinical practice of one author (J.A.D.). Preoperative LOCS III grading of all surgical patients was done at the slitlamp to improve clinical documentation in the clinical record. Parenthetical descriptions were rarely added but usually included the critical nature of the centrality within the visual axis of subtle localized interfaced zones of NO and NC, cortical spokes, or subcapsular opacities. There was consistency in grading over time, and patients could be told with greater confidence whether their cataracts were stable or progressing. It was also helpful to have accurate measures of cataract severity when discussing cases with third-party payer representatives; surgical interventional decisions to offer surgery to some patients could be better defended. Over time, LOCS III slitlamp grading was empirically incorporated into the decision-making process and planning of upcoming surgeries. Ultimately, this resulted in a more organized, integrated, customized operative plan for each patient. Patients and Methods In this retrospective study, 2364 consecutive cases of phacoemulsification were analyzed with respect to the 4 LOCS III characteristics and machine-measured phacoemulsification time, average machine measured power (%), and the balanced salt solution (BSS ) volume used. All data points had been recorded in 2186 cases. Surgery was performed from January 1998 to July 2000 by a single surgeon (J.A.D), who also performed LOCS III grading of all cataracts at the slitlamp. A standardized nuclear-fracture quadrant-aspiration phacoemulsification technique was used with a straight 45-degree 0.9 mm ABS MicroTip with the Alcon Legacy 20,000 machine. 6 After topical anesthesia was administered, a temporal clear corneal incision was made. Grooving was accomplished with the machine in memory I (phacoemulsifica- J CATARACT REFRACT SURG VOL 29, JANUARY
3 Figure 2. (Davison) After initial grooving is completed, a second rotation of the nucleus allows for improved visualization while deeper grooving progresses in hard nuclei. Figure 3. (Davison) Nuclear cracking is created from the periphery toward the center with a cross action of the phacoemulsification tip and the cyclodialysis spatula. Figure 4. (Davison) The 45-degree phacoemulsification tip is turned sideways with its aperture facing the interior wall of the first nuclear quadrant. Low levels of emulsification energy are used to engage and impale the central portion of the quadrant. Figure 5. (Davison) A low level of ultrasonic energy and a higher level of vacuum allow the tip to adhere to the softer edge of the quadrant and drag it centrally. tion power maximum, 90%; vacuum maximum, 50 mm Hg; aspiration flow rate, 14 cc/minute) (Figure 2). Wider, deeper grooves were made as nuclear hardness increased. The posterior nuclear plate was torn from the periphery toward the center by cross action with the phacoemulsification tip and a custom, modified, 0.37 mm cyclodialysis spatula (Figure 3). Quadrant aspiration was accomplished in memory III (phacoemulsification power maximum, 70%; vacuum maximum, 500 mm Hg; aspiration flow rate, 35 cc/minute). The aspiration process was kept posterior to the iris plane within the capsular bag as much as possible to protect the corneal endothelium from damage from fragment impact and abrasion (Figures 4 to 6). Sodium hyaluronate 3.0% chondroitin sulfate 4.0% (Viscoat ) was usually injected into the anterior chamber before quadrant aspiration when the NC was greater than 4.5. In some cases, Viscoat was reinjected before aspiration of the fourth quadrant because of the tendency of the capsular bag to flatten after the third quadrant had been removed, which forces the remaining process anterior to the iris plane. In some eyes, soft quadrants or the last fragment of the fourth firm quadrant were removed using reduced vacuum (maximum 300 mm Hg), maximum power 140 J CATARACT REFRACT SURG VOL 29, JANUARY 2003
4 Figure 6. (Davison) The bulk of the quadrant is central while moderate ultrasonic energy, vacuum, and aspiration flow are used to remove the firm portions of the quadrant. Figure 7. (Davison) Aspiration can begin after an edge of the last quadrant is lifted free of the posterior capsule. (50%), and a reduced aspiration flow rate (25 cc/minute) (Figures 7 and 8). Polynomial best-fit lines were derived by regression analysis among the 4 LOCS III characteristics and the 3 phacoemulsification observations. Figure 8. (Davison) The cyclodialysis spatula protects the posterior capsule as phacoemulsification power is applied during removal of the last quadrant. Lower levels of vacuum and aspiration flow rate help protect the posterior capsule from inadvertent aspiration. Table 1. Scale Range Mean machine-measured ultrasound time (minutes). LOCS III NO Scale LOCS III NC Scale Total NO nuclear opalescence; NC nuclear color Results The mean machine-measured phacoemulsification time increased as the NO and NC grades increased (Table 1). As determined by the best-fit lines and their coefficients of determination (R 2 ), there was an exponential relationship between NO and machine-measured phacoemulsification time (R ). A trend existed between NO and the mean power (R ) and the amount of BSS used (R ) (Figure 9). A slightly stronger exponential relationship existed between NC and machine-measured phacoemulsification time (R ). A trend also existed between NC and mean power (R ) and the amount of BSS used (R ) (Figure 10). No relationships were observed between the LOCS III classes of cataract, C and P, and any intraoperative observation (Figures 11 and 12). Discussion In the original LOCS III publication, the NO and NC scales were highly correlated with each other and the relationships between objective measures of NO, NC, and the LOCS III grade of the standard images were J CATARACT REFRACT SURG VOL 29, JANUARY
5 Figure 9. (Davison) A significant relationship is seen between LOCS III NO and ultrasound time but not mean ultrasound power and BSS volume used. Figure 10. (Davison) A significant relationship is seen between LOCS III NC and ultrasound time but not mean ultrasound power and BSS volume used. 142 J CATARACT REFRACT SURG VOL 29, JANUARY 2003
6 Figure 11. (Davison) No relationship is seen between LOCS III C and ultrasound time, mean ultrasound power, and BSS volume used. Figure 12. (Davison) No relationship is seen between LOCS III P and ultrasound time, mean ultrasound power, and BSS volume used. J CATARACT REFRACT SURG VOL 29, JANUARY
7 nearly linear. The NO was assessed independently with image analysis of optical density, and NC was assessed independently with fast spectral scanning colorimetry of each LOCS III standardized image. Highly different grades of NO and NC in the same cataract were unlikely. Similarly, the relationship between the expenditure of phacoemulsification energy and the LOCS III grades of NO and NC appears to be linear for LOCS III grades below approximately 3.7. Above that, the relationship between LOCS III grade and phacoemulsification energy expenditure seems to become exponential. Although in LOCS III NO and NC are graded separately, most clinicians are accustomed to describing nuclear change by a single term, nuclear sclerosis. Both color change and increased light scattering are part of the process of nuclear sclerosis, and it is therefore not surprising to have found that both NO and NC are similarly positively correlated with ultrasonic energy exposure. When viewing a cataractous lens with LOCS III NO and NC grades of 4.0 to 5.0, most surgeons intuitively anticipate increased phacoemulsification times. The nucleus is simply harder, so it will take more energy to emulsify. The LOCS III NC grade has been correlated with hardness of the central nucleus as measured by the resistance to a fine conical probe measured by a dynamometer. 7 Phacoemulsification energy recorded by cumulative delivered energy as well as by postoperative anterior chamber flare has been shown to be correlated with LOCS III NC and NO. 8 The surgeon in this study finds it useful to be able to accurately grade opalescence and brunescence (and the severity of C and P) to better estimate when prolonged ultrasonic time or increased ultrasonic energy is likely. It has proved advantageous to integrate LOCS III grading into routine operative planning. By doing so, the surgeon will be better prepared to manage the more sclerotic nucleus and his or her preoperative planning can be more patient specific. The surgeon in this study empirically formulated several recommendations for customizing a surgical plan using preoperative LOCS III grading as a part of a complete preoperative ophthalmic examination. These are listed below. 1. A retentive viscoelastic material is recommended in cases in which the LOCS III NO or NC grade is 4.0 or greater. 2. If the LOCS III NO or NC grade is 5.0, the capsulorhexis will be more easily visualized if indocyanine green dye is used to stain the anterior capsule. 3. If the pupil is smaller than 4.5 mm and the LOCS III NO or NC grade is greater than 4.5, use of a pupil expansion/maintenance device should be considered. 4. If zonular integrity is compromised and the LOCS III NO or NC grade is 4.5 or greater, a backup ciliary-sulcus-supported intraocular lens (IOL) or anterior chamber IOL should be available. 5. Surgeons in training should learn and use LOCS III grading to enable them to select normal cases with nuclei of enough substance to easily manipulate yet readily emulsify. Very soft or very hard nuclei may present unnecessary challenges. 6. The LOCS III is helpful and should be used in the trials of new technologies for cataract removal. For example, in one author s experience (J.A.D.) with early AquaLase technology (Alcon Surgical), efficient removal of a nucleus with an NC grade greater than 3.7 was not possible. Of course, the relationships between LOCS III characteristics and phacoemulsification energy expenditure vary with the individual technique and equipment used. Despite this individual-specific variation, it is likely that the trends observed will be seen with other surgeons as long as a mechanical removal technique is used. Incorporating LOCS III grading into a preoperative regimen should not only improve surgical planning, it should also improve clinical documentation, communication among providers and with third-party payers, and the accuracy of longitudinal follow-up. Most important, preoperative LOCS III classification should contribute to more predictable and effective cataract surgery. References 1. Chylack LT Jr, Wolfe JK, Singer DM, et al. The Lens Opacities Classification System III; the Longitudinal Study of Cataract Study Group. Arch Ophthalmol 1993; 111: Maraini G, Pasquini P, Tomba MC, et al. An independent evaluation of the Lens Opacities Classification System II (LOCS II); the Italian-American Cataract Study Group Ophthalmology 1989; 96: Karbassi M, Khu PM, Singer DM, Chylack LT Jr. Evalu- 144 J CATARACT REFRACT SURG VOL 29, JANUARY 2003
8 ation of Lens Opacities Classification System III applied at the slitlamp. Optom Vis Sci 1993; 70: Milton RD, Chylack LT Jr, Köpcke W. Letters to the editor. Ophthalmic Epidemiol 1996; 3: van den Berg TJTP, Coppens JC. Conversion of lens slit lamp photographs into physical light-scattering units. Invest Ophthalmol Vis Sci 1999; 40: Davison JA. Performance comparison of the Alcon Legacy mm TurboSonics and 0.9 mm Aspiration Bypass System tips. J Cataract Refract Surg 1999; 25: Hu C, Zhang X, Hui Y, Chung H. [The nuclear hardness and associated factors of age-related cataract]. [Chinese] Yen Ko Tsa Chih 2000; 36: Ursell PG, Spalton DJ, Tilling K. Relation between postoperative blood-aqueous barrier damage and LOCS III cataract gradings following routine phacoemulsification surgery. Br J Ophthalmol 1997; 81: J CATARACT REFRACT SURG VOL 29, JANUARY
Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population
Original article: Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population Dr Nikhil Mahajan*, Prof. Swati Tomar** **Professor,*Resident Department of Ophthalmology,
More informationCOURSE DESCRIPTION BASIC FUNDAMENTALS
TACKLING POSTERIOR CAPSULE RUPTURE AND IOL IMPLANTATION: A VIDEO BASED COURSE TUESDAY - 29 th APRIL, 2014: 1.00 PM-2.30 PM, BCEC, ROOM 258 A ; SESSION 29-308 COURSE DESCRIPTION BASIC FUNDAMENTALS Early
More informationSECONDARY CAPSULOTOMY USING THE FEMTOSECOND LASER. Surendra Basti, MD Northwestern University Feinberg School of Medicine, Chicago, IL
SECONDARY CAPSULOTOMY USING THE FEMTOSECOND LASER Surendra Basti, MD Northwestern University Feinberg School of Medicine, Chicago, IL SMALL PUPIL MANAGEMENT DURING FEMTO CATARACT SURGERY Surendra Basti,
More informationDespite recent advances in techniques and technology,
Common Phaco Techniques for Rock-Hard Nuclei Tips to ensure a successful outcome. By Arup Chakrabarti, MS Despite recent advances in techniques and technology, phacoemulsification in a rock-hard nucleus
More informationInadvertent trypan blue staining of posterior capsule during cataract surgery associated with "Argentinian flag" event
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2016 Inadvertent trypan blue staining of posterior capsule during cataract surgery associated with "Argentinian
More informationMicroincisional cataract surgery (MICS) with pulse and burst modes
European Journal of Ophthalmology / Vol. 16 no. 6, 2006 / pp. 804-808 Microincisional cataract surgery (MICS) with pulse and burst modes M. BAYKARA 1, I.. ERCAN 2, H. OZCETIN 1 1 Department of Ophthalmology
More informationFate of anterior capsule tears during cataract surgery
J CATARACT REFRACT SURG - VOL 32, OCTOBER 2006 Fate of anterior capsule tears during cataract surgery Frederico F. Marques, MD, Daniela M.V. Marques, MD, Robert H. Osher, MD, James M. Osher, MS PURPOSE:
More informationCATARACT TRANSECTION WITH A NEW DEVICE
CATARACT SURGERY CATARACT TRANSECTION WITH A NEW DEVICE The miloop has the potential to change the landscape of cataract surgery. BY WILLIAM F. WILEY, MD Tremendous changes in cataract surgery have occurred
More informationPhacoemulsification with hydrodelineation and OVD-assisted hydrodissection in posterior polar cataract
Hua et al. BMC Ophthalmology (2018) 18:165 https://doi.org/10.1186/s12886-018-0845-8 RESEARCH ARTICLE Open Access Phacoemulsification with hydrodelineation and OVD-assisted hydrodissection in posterior
More informationDisorders of the. blood-aqueous barrier after. phacoemulsification in diabetic patients CLINICAL STUDY. Y Liu, L Luo, M He and X Liu
(2004) 18, 900 904 & 2004 Nature Publishing Group All rights reserved 0950-222X/04 $30.00 www.nature.com/eye CLINICAL STUDY Disorders of the blood-aqueous barrier after phacoemulsification in diabetic
More informationRuba Alobaidy Jia Y Ng Sathish Srinivasan
Ruba Alobaidy Jia Y Ng Sathish Srinivasan Department of Ophthalmology, University Hospital Ayr, Ayr, Scotland The authors have no financial interests to declare. Continuous curvilinear capsulorhexis (CCC)
More informationIntroduction. We are finally using a laser!!! The Use of a Femtosecond Laser for Complex Cataract Procedures. Financial Disclosure
ASCRS 2014 The use of the femtosecond laser for complex cataract surgery Cases I could not have performed without the femtosecond laser Course Director: Richard S. Davidson, M.D. Associate Professor and
More informationComplex cataract cases Managing catarocks : Better surgery on dense lenses, intumescent cataracts
Complex cataract cases Managing catarocks : Better surgery on dense lenses, intumescent cataracts by Vanessa Caceres EyeWorld Contributing Writer Hypermature white cataract. According to Dr. Donaldson,
More informationPosterior capsule rupture is a complication that is
techniques Posterior assisted levitation for nucleus retrieval using Viscoat after posterior capsule rupture David F. Chang, MD, Richard B. Packard, MD A pars plana injection of Viscoat (sodium hyaluronate
More informationIntroduction. We are finally using a laser!!! The Use of a Femtosecond Laser for Complex Cataract Procedures. Financial Disclosure
ASCRS 2014 The use of the femtosecond laser for complex cataract surgery Cases I could not have performed without the femtosecond laser Course Director: Richard S. Davidson, M.D. Associate Professor and
More informationIntraoperative biometry for intraocular lens (IOL) power calculation at silicone oil removal
European Journal of Ophthalmology / Vol. 13 no. 7, 2003 / pp. 622-626 Intraoperative biometry for intraocular lens (IOL) power calculation at silicone oil removal S.M. EL-BAHA, A. EI-SAMADONI, H.F. IDRIS,
More informationLearn Connect Succeed. JCAHPO Regional Meetings 2017
Learn Connect Succeed JCAHPO Regional Meetings 2017 Cataract Surgery in 2017 DARBY D. MILLER, MD MPH CORNEA, CATARACT AND REFRACTIVE SURGERY ASSISTANT PROFESSOR OF OPHTHALMOLOGY MAYO CLINIC FLORIDA Natural
More informationInternational Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR)
International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR) The International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubrics
More informationIntraoperative Floppy Iris Syndrome
Intraoperative Floppy Iris Syndrome Dr. Saurabh Sawhney, Dr. Aashima Aggarwal Insight Eye Clinic, Rajouri Garden, New Delhi 110027, India Introduction Phacoemulsification entails highly precise manipulations
More informationCapsule fixation device for cataract surgery
European Journal of Ophthalmology / Vol. 19 no. 1, 2009 / pp. 143-146 SHORT COMMUNICATIONS & CASE REPORTS Capsule fixation device for cataract surgery N.M. SERGIENKO 1, Y.N. KONDRATENKO 1, A.K. YAKIMOV
More informationASCRS 2016 Instructional Course Mastering Femtosecond Laser Assisted Phacoemulsification
ASCRS 2016 Instructional Course 07-410 Mastering Femtosecond Laser Assisted Phacoemulsification LASER CAPSULOTOMY TIMOTHY V ROBERTS MBBS (NSW), MMed (Syd), FRANZCO, FRACS, GAICD Vision Eye Institute, Sydney
More informationMorphological Study of Corneal Endothelium and Corneal Thickness in Pseudoexfoliation Syndrome
LABORATORY INVESTIGATIONS Morphological Study of Corneal Endothelium and Corneal Thickness in Pseudoexfoliation Syndrome Kenji Inoue*,, Kazuko Okugawa*,, Tetsuro Oshika and Shiro Amano *Department of Ophthalmology,
More informationComparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification
Received: 5.5.2007 Accepted: 22.1.2008 Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification Heidarali Moeini*, Fateme Eslami**, Akram Rismanchian*,
More informationStart with ME. LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL
LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. Start with ME. TECNIS TORIC 1-PIECE IOL INDICATIONS: The TECNIS Toric 1-Piece Posterior Chamber Lens is indicated for the visual correction
More informationORIGINAL ARTICLE. HIGH VOLUME CAMP SURGERIES A CLINICAL STUDY D. N. Prakash, K, Sathish, Sankalp Singh Sharma, Soujanya. K, Savitha Patil.
HIGH VOLUME CAMP SURGERIES A CLINICAL STUDY D. N. Prakash, K, Sathish, Sankalp Singh Sharma, Soujanya. K, Savitha Patil. 1. Assistant Professor. Department of Ophthalmology, MMC & RI, Mysore, 2. Associate
More informationDo you need to perform hydrodissection?
Hydrodissection in Femtoassisted cataract surgery The Why and the How no financial interest DrIan YEO Singapore National Eye Centre Do you need to perform hydrodissection? 1 Hydrodissectionin traditional
More informationOriginal Article Capsular tension ring implantation after lens extraction for management of subluxated cataracts
Int J Clin Exp Pathol 2014;7(7):3733-3738 www.ijcep.com /ISSN:1936-2625/IJCEP0000754 Original Article Capsular tension ring implantation after lens extraction for management of subluxated cataracts Xiao
More informationVisual acuity versus letter contrast sensitivity in early cataract
Vision Research 38 (1998) 2047 2052 Visual acuity versus letter contrast sensitivity in early cataract David B. Elliott a, *, Ping Situ b a Department of Optometry, Uni ersity of Bradford, Bradford BD7
More informationTITLE: Femtosecond Laser Cataract Surgery : How to Avoid and Manage Complications
ASCRS 2016 Course Handouts TITLE: Femtosecond Laser Cataract Surgery : How to Avoid and Manage Complications Course Director: Dr Seng-Ei TI Faculty: Prof Soon-Phaik CHEE, Dr Ron YEOH This course discusses
More informationIntraindividual comparison of surgical trauma after bimanual microincision and conventional small-incision coaxial phacoemulsification
ARTICLE Intraindividual comparison of surgical trauma after bimanual microincision and conventional small-incision coaxial phacoemulsification Günal Kahraman, MD, Michael Amon, MD, Carmen Franz, MD, Anna
More informationEndo Optiks. Clinical Publication Summaries
Endo Optiks Clinical Publication Summaries Effective. Safe. Simple. Four scientific studies demonstrating the proven clinical benefits of combined ECP and cataract surgery. ECP is an Effective, Safe, and
More informationManagement of Radial Tears During Capsulorhexis
4(2):56 60, 2006 C A T A R A C T S U R G E R Y C O M P L I C A T I O N S Management of Radial Tears During Capsulorhexis Mehrdad Mohammadpour, MD Ophthalmic Research Center Shaheed Beheshti University
More informationIFIS. OZil Torsional ultrasound and INTREPID fluidics give surgeons greater confidence with these challenging cases.
CRST0409Euro_IFIS_supp.qxd Supplement to 3/26/09 10:24 AM Page 1 Produced under an educational grant from Alcon Laboratories, Inc. April 2009 Managing IFIS With OZil Torsional Ultrasound OZil Torsional
More informationEUROPE. TORSIONAL AND MICRO-COAXIAL TECHNOLOGY: New Efficiencies in Challenging Cases
EUROPE TORSIONAL AND MICRO-COAXIAL TECHNOLOGY: New Efficiencies in Challenging Cases Highlights from a CME symposium held during the 2007 XXV ESCRS Congress in Stockholm, Sweden This supplement has been
More informationGives you the chamber stability you demand
Gives you the chamber stability you demand See better. Live better. 02 / Fluidics Fluidics / 03 Stellaris Activate gives you the chamber stability you demand Only Stellaris Activate gives you vacuum based
More informationEvery week, the John A. Moran Eye Center at the
Analyzing Explanted IOLs A review of the pathological processes that affect pseudophakic patients postoperative vision. BY LILIANA WERNER, MD, PHD Every week, the John A. Moran Eye Center at the University
More informationgo the distance NEW AcrySof IQ ReSTOR +2.5 D It s a wide world. Help your patients and everywhere in between.
NEW AcrySof IQ ReSTOR +2.5 D It s a wide world. Help your patients go the distance and everywhere in between. Now with ACTIVEFOCUS optical design for active-lifestyle patients. Recommend AcrySof IQ ReSTOR
More informationSurgical Notes. January 2017 Clear Cornea Knife. Capsulorhexis Forceps. Advancut Single Use Clear Corneal Knife
Surgical Notes January 2017 Clear Cornea Knife Advancut Single Use Clear Corneal Knife Make a perfect square and true to size incision Very sharp tip easy for penetration The uniquely designed edge of
More informationYou can see vivid colours again after cataract management at Sankar Foundation Eye Hospital
The Department of Cataract in our Sankar Foundation is equipped with state-of-the-art operation theatres, surgical microscope phacoemulsification machine and microsurgical instruments. And also the department
More informationIntegrating the LenSx Laser Laser into your Clinic: Our advice for those considering it
Integrating the LenSx Laser Laser into your Clinic: Our advice for those considering it (DAY) Facilitator: (Name) LenSx Laser Important Safety Information Caution United States Federal Law restricts this
More informationClinical and Experimental Ophthalmology 2017; 2018; : 46: doi: doi: /ceo.12998
Clinical and Experimental Ophthalmology 2017; 2018; : 46: 35 45 doi: doi: 10.1111/ceo.12998 Original Article Can we improve the efficacy of modern cataract surgery by using different tip designs? A comparison
More informationVisual perception during phacoemulsification cataract surgery under subtenons anaesthesia
(23) 17, 51 55 & 23 Nature Publishing Group All rights reserved 95-222X/3 $25. www.nature.com/eye Visual perception during phacoemulsification cataract surgery under subtenons anaesthesia SS Wickremasinghe,
More informationAnterior segment imaging
Article Date: 11/1/2016 Anterior segment imaging AS OCT vs. UBM vs. endoscope; case based approaches BY BENJAMIN BERT, MD, FACS AND BRIAN FRANCIS, MD, MS Currently, numerous imaging modalities are available
More informationIntroduction. Donor tissue preparation for Descemet Membrane Endothelial ASCRS Aim of dissection. DMEK graft preparation
Introduction Donor tissue preparation for Descemet Membrane Endothelial keratoplasty (DMEK) ASCRS - 2015 Endothelial Keratoplasty DSAEK / DSEK DMEK Donor lamellae stroma + DM + endothelium DM + endothelium
More informationComparison of the Long-term Clinical Results of Hydrophilic and Hydrophobic Acrylic Intraocular Lenses
Korean J Ophthalmol Vol. 19:29-33, 2005 Comparison of the Long-term Clinical Results of Hydrophilic and Hydrophobic Acrylic Intraocular Lenses Youngwoo Suh, MD 1, Chunghoon Oh, MD 2, Hyo-Myung Kim, MD,
More informationAssociation between Low Plasma Vitamin E Concentration and Progression of Early Cortical Lens Opacities
American Journal of Epidemiology Copyright 996 by The Johns Hopkins University School of Hygiene and Public Health All rights reserved Vol. 44, No. 5 Printed in U.SA. Association between Low Plasma Vitamin
More informationHalf-moon supracapsular nucleofractis phacoemulsification: Safety, efficacy, and functionality
ARTICLE Half-moon supracapsular nucleofractis phacoemulsification: Safety, efficacy, and functionality _Izzet Can, MD, Tamer Takmaz, MD, _ Ipek Genç, MD PURPOSE: To compare the safety, efficacy, and functionality
More informationStructural changes of the anterior chamber following cataract surgery during infancy
Structural changes of the anterior chamber following cataract surgery during infancy Matthew Nguyen, Emory University Marla Shainberg, Emory University Allen Beck, Emory University Scott Lambert, Emory
More informationLASER CATARACT SURGERY FOR COMPLEX CASES?
LASER CATARACT SURGERY FOR COMPLEX CASES? Five circumstances in which this technology is particularly beneficial. BY SAMUEL MASKET, MD The technology creates more barriers than benefits in these cases.
More information2 years experience with LenSx,what we learned?
2 years experience with LenSx,what we learned? Saleh Saif AL Messabi FRCS( CANADA) Medical Director,Samaya Eye Hospital and Centers,UAE President,Emirates Ophthalmic Society Financial Interest Speaker
More informationVanderbilt Eye Institute
Vanderbilt Eye Institute Joshua Ki Hu Vanderbilt Eye Institute Ophthalmology, PGY-4 DATE 05.30.08 Introduction Cataract surgery is one of the most commonly performed surgeries in the world, with over 1
More informationMegalocornea is a non-progressive, uniformly
Case Report 191 Anterior Megalophthalmos Chien-Kuang Tsai, MD; Ing-Chou Lai, MD; Hsi-Kung Kuo, MD; Mei-Chung Teng, MD; Po-Chiung Fang, MD We describe a 36-year-old female who suffered from presenile cataract
More informationFinancial Disclosures. Corneal Problems for the Cataract Surgeon. Four Common Problems. Dry Eye syndrome. Rose-Bengal 3/27/16
Corneal Problems for the Cataract Surgeon Financial Disclosures Consultant: AMO/VISX Consultant: Angiotech/Sharppoint Michael J Taravella, MD Director: Cornea and Refractive Surgery University of Colorado
More informationSurgical Applications and Complications
Ophthalmic Viscosurgical Devices and Anterior Segment Surgery: Surgical Applications and Complications Suresh K. Pandey M.D., Jaya Thakur M.D., Liliana Werner MD Ph.D., David J. Apple M.D. Background Viscoelastic
More informationInfluence of cataract on image quality and macular thickness measured using spectral domain optical coherence tomography: a prospective cohort study
International Journal of Advances in Medicine Lathika VK et al. Int J Adv Med. 2017Apr;4(2):546-550 http://www.ijmedicine.com pissn2349-3925 eissn 2349-3933 Original Research Article DOI: http://dx.doi.org/10.18203/2349-3933.ijam20171058
More informationComparison of Phaco Energy and Complications for Conventional Phacoemulsification and 25-gauge Phacovitrectomy Surgery
Ophthalmology Research: An International Journal 6(3): XX-XX, 2016; Article no.or.29870 ISSN: 2321-7227 SCIENCEDOMAIN international www.sciencedomain.org Comparison of Phaco Energy and Complications for
More informationOptometric Postoperative Cataract Surgery Management
Financial Disclosures Optometric Postoperative Cataract Surgery Management David Dinh, OD Oak Cliff Eye Clinic Dallas Eye Consultants March 10, 2015 Comanagement Joint cooperation between two or more specialists
More informationTransient Intraocular Pressure Elevation after Trabeculotomy and its Occurrence with Phacoemulsification and Intraocular Lens Implantation
Transient Intraocular Pressure Elevation after Trabeculotomy and its Occurrence with Phacoemulsification and Intraocular Lens Implantation Masaru Inatani*, Hidenobu Tanihara, Takahito Muto*, Megumi Honjo*,
More informationVISUAL OUTCOME AFTER MICROCOAXIAL PHACOEMULSIFICATION WITH MICRIOL PLUS LENS IMPLANTATION. of Corresponding Author:
VISUAL OUTCOME AFTER MICROCOAXIAL PHACOEMULSIFICATION WITH MICRIOL PLUS IJCRR Vol 04 issue 23 Section: Healthcare Category: Research Received on: 12/10/12 Revised on: 06/11/12 Accepted on: 27/11/12 Mohd
More informationCataract Surgery Management in Eyes with Extensive Iridoschisis
Cataract Surgery Management in Eyes with Extensive Iridoschisis Hassan Hashemi, MD 1,2 Golshan Latifi, MD 3 Sasan Moghimi, MD 4 S-Farzad Mohammadi, MD 5 Abstract Purpose: To demonstrate an approach to
More informationClinical study of traumatic cataract and its management
Clinical study of traumatic cataract and its management Original article Manjula Mangane, M.R. Pujari, Chethan N. Murthy Department of Ophthalmology, Basaweshwar Teaching and General Hospital, Gulbarga,
More informationIs There an Association of Topical Ocular Hypotensive Medication with Lens Opacification and Decreased Visual Function?
Is There an Association of Topical Ocular Hypotensive Medication with Lens Opacification and Decreased Visual Function? Supported by the National Eye Institute (EY 09307, EY 09341) National Center on Minority
More informationCataract. A cataract is a clouding of the lens in your eye. It
Cataract A cataract is a clouding of the lens in your eye. It affects your vision. Cataracts are very common in older people. By age 80, more than half of all Americans either have a cataract or have had
More informationCigdem Altan, MD, Sukru Bayraktar, MD, Tugrul Altan, MD, Hakan Eren, MD,
Anterior chamber depth, iridocorneal angle width, and intraocular pressure changes after uneventful phacoemulsification in eyes without glaucoma and with open iridocorneal angles Cigdem Altan, MD, Sukru
More informationClinical Evaluation of the BunnyLens IOL
Clinical Evaluation of the BunnyLens IOL Introduction: BunnyLens is a foldable Hydrophlic Acrylic IOL with four ear shaped haptic design. The lens design offers many advantages in terms of: 1. Centration
More informationOphthalmic viscosurgical devices (OVD) are used in eye
Int J Ophthalmol, Vol. 10, No. 5, May 18, 2017 www.ijo.cn Clinical Research Safety of hydroimplantation in cataract surgery in patients with pseudoexfoliation syndrome Tevfik Oğurel, Reyhan Oğurel, Zafer
More informationPros & Cons of Introducing Femtosecond. Training Programs. James P. McCulley M.D., FACS,FRCOph (UK) University of Texas Southwestern Medical School
Pros & Cons of Introducing Femtosecond Laser AssistedCataract Surgery in Residency Training Programs James P. McCulley M.D., FACS,FRCOph (UK) Department of Ophthalmology University of Texas Southwestern
More informationObjective Scatter Index: Working Toward a New Quantification of Cataract?
ORIGINAL ARTICLE Objective Scatter Index: Working Toward a New Quantification of Cataract? Florence Galliot, MD, PhD; Sunni R. Patel, PhD; Béatrice Cochener, MD, PhD ABSTRACT PURPOSE: To investigate the
More informationLABORATORY SCIENCES. Influence of Elevated Intraocular Pressure on the Posterior Chamber Anterior Hyaloid Membrane Barrier During Cataract Operations
LABORATORY SCIENCES Influence of Elevated Intraocular Pressure on the Posterior Chamber Anterior Hyaloid Membrane Barrier During Cataract Operations Shiro Kawasaki, MD; Yoshitaka Tasaka, MD; Takashi Suzuki,
More informationLaser Refractive Cataract Surgery with the LenSx Laser
Laser Refractive Cataract Surgery with the LenSx Laser a Novartis company 1 LenSx Laser Important Safety Information Caution: United States Federal Law restricts this device to sale and use by or on the
More informationQuantified Measurement of Intraocular Inflammation after Phacoemulsification Cataract Surgery
Med. J. Cairo Univ., Vol. 82, No. 2, December: 237-241, 14 www.medicaljournalofcairouniversity.net Quantified Measurement of Intraocular Inflammation after Phacoemulsification Cataract Surgery AYMAN M.
More informationComparison of Corneal Endothelial Cell Loss between Torsional and Linear Phacoemulsification
Comparison of Corneal Endothelial Cell Loss between Torsional and Linear Phacoemulsification Delhi J Ophthalmol 2014; 25 (1): 23-27 DOI: http://dx.doi.org/10.7869/djo.68 Amrita Sawhney, Uma Sridhar, Charu
More informationIOL Power Calculation for Children
1 IOL Power Calculation for Children Rupal H. Trivedi, MD MSCR M. Edward Wilson, MD The authors have no financial interest in the subject matter of this presentation. Intraocular lens (IOL) implantation
More informationComparison of OVD and BSS for Maintaining the Anterior Chamber during IOL Implantation
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2011;25(1):15-21 DOI: 10.3341/kjo.2011.25.1.15 Original Article Comparison of OVD and BSS for Maintaining the Anterior Chamber during IOL Implantation
More informationCURRICULUM VITAE JAMES ARTHUR DAVISON, M.D., F.A.C.S.
CURRICULUM VITAE JAMES ARTHUR DAVISON, M.D., F.A.C.S. PRIVATE PRACTICE: Wolfe Eye Clinic, P.C. 309 East Church Street Marshalltown, IA 50158 Cataract, Refractive Surgery & Comprehensive Ophthalmology August
More informationComparing Femtosecond Laser Assisted Cataract Surgery Before and After Phakic Intraocular Lens Removal
SURGICAL TECHNIQUE Comparing Femtosecond Laser Assisted Cataract Surgery Before and After Phakic Intraocular Lens Removal Tim Schultz, MD, FEBO; Luca Schwarzenbacher, MD; H. Burkhard Dick, MD, PhD ABSTRACT
More informationThe use of an ultrasonically
The News Magazine of the American Society of Cataract & Refractive Surgery EYEWORLD SUPPLEMENT MARCH 2006 Exploring and Understanding the Benefits of Torsional Phacoemulsification Understanding the Physics
More informationPREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland
AD Award Number: W81XWH-08-1-0531 TITLE: Virtual Mentor Cataract Surgery Trainer PRINCIPAL INVESTIGATOR: Principal Investigator: John I. Loewenstein MD Co-Investigator: Bonnie A. Henderson MD CONTRACTING
More informationA clinical study of corneal complications of manual small incision cataract surgery
Original Research Article Anitha S. Maiya 1,*, Akshatha M. Dharmesh 2, R. Jayaram 3 1 Associate Professor, 2 Senior Resident, 3 Professor, Dept. of Ophthalmology, Adichunchanagiri Institute of Medical
More informationIncidence, clinical findings and management of intraoperative floppy iris syndrome associated with tamsulosin
Incidence, clinical findings and management of intraoperative floppy iris syndrome associated with tamsulosin Ugur Keklikci, 1 Kenan Isen, 2 Kaan Unlu, 1 Yusuf Celik 3 and Mine Karahan 1 1 Department of
More informationThe Effect of Polymethylmethacrylate and AcrySof Intraocular Lenses on the Posterior Capsule in Patients with a Large Capsulorrhexis
The Effect of Polymethylmethacrylate and AcrySof Intraocular Lenses on the Posterior Capsule in Patients with a Large Capsulorrhexis W. R. Meacock*, D. J. Spalton*, E. J. Hollick*, S. Barman and J. F.
More informationINTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany
European Journal of Ophthalmology / Vol. 17 no. 3, 2007 / pp. 363-367 Anterior segment optical coherence tomography for evaluation of changes in anterior chamber angle and depth after intraocular lens
More informationSynchrony AIOL Key Features
Synchrony AIOL Key Features Spacers Provide consistent separation distance at emmetropia Prevent lens adhesion Synchrony AIOL Key Features Spring Haptics Bias the system open Provide consistent separation
More informationCLINICAL SCIENCES. The Effect of Texturing the Intraocular Lens Edge on Postoperative Glare Symptoms
CLINICAL SCIENCES The Effect of Texturing the Intraocular Lens Edge on Postoperative Glare Symptoms A Randomized, Prospective, Double-Masked Study Will R. Meacock, FRCOphth; David J. Spalton, FRCS, FRCOphth;
More informationCorneal Endothelial Change after Phacoemulsification
Original Article Abstract Background: Phacoemulsification is the modern method of sight restoring surgery for cataract and it is becoming popular day by day. We planned this study to find out any significant
More informationSTANDARD AUTOMATED PERIMETRY IS A GENERALLY
Comparison of Long-term Variability for Standard and Short-wavelength Automated Perimetry in Stable Glaucoma Patients EYTAN Z. BLUMENTHAL, MD, PAMELA A. SAMPLE, PHD, LINDA ZANGWILL, PHD, ALEXANDER C. LEE,
More informationFROM PRE-OP TO POST-OP, OPTIMIZE YOUR WORKFLOW WITH THE CATALYS SYSTEM MOBILE PATIENT BED.
FROM PRE-OP TO POST-OP, OPTIMIZE YOUR WORKFLOW WITH THE CATALYS SYSTEM MOBILE PATIENT BED. PUSH THE LIMITS Imagine if your CATALYS System patient bed could: Optimize your productivity throughout the full
More informationHooks & Rings In The Management Of Subluxated Cataracts. Introduction. Introduction- Capsular Support 3/31/2015
Hooks & Rings In The Management Of Subluxated Cataracts Financial Disclosure Ashu Agarwal Perfect Sight Centre New Delhi India I have no financial interests or relationships to disclose. 1 Introduction
More informationFROM PRE-OP TO POST-OP, HELP OPTIMIZE YOUR WORKFLOW WITH THE CATALYS SYSTEM MOBILE PATIENT BED.
FROM PRE-OP TO POST-OP, HELP OPTIMIZE YOUR WORKFLOW WITH THE CATALYS SYSTEM MOBILE PATIENT BED. PUSH THE LIMITS Imagine if your CATALYS System patient bed could: Optimize patient workflow throughout the
More informationThe Effect of Phacoemulsification with Posterior Chamber Intraocular Lens Implantation on Intraocular Pressure and Anterior Chamber Depth
The Effect of Phacoemulsification with Posterior Chamber Intraocular Lens Implantation on Intraocular Pressure and Anterior Chamber Depth Mohammad Javad Mohamadi, MD 1 Reza Soltani Moghadam, MD 1 Hasan
More informationInclusion Criteria Ages eligible for study: 40 Years to 70 Years Genders eligible for study: Both
IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 16, Issue 5 Ver. II (May. 2017), PP 49-53 www.iosrjournals.org A Prospective Comparative Study of Central
More informationInternational Journal of Research and Review E-ISSN: ; P-ISSN:
International Journal of Research and Review www.gkpublication.in E-ISSN: 2349-9788; P-ISSN: 2454-2237 Original Research Article Study of Efficacy and Safety of Low and High Fluidic Settings in Phacoemulsification
More informationNature and Science 2016;14(9)
Comparison between Implantable Contact Lens (ICL) versus Acry Sof Cachet Phakic Intraocular Lenses in Correction of Moderate to High Myopia Khaled Nada, M. Sc.; Mohammed Attia, M. D and Ashraf El Habbak,
More informationOriginal Article. Dharmik Sharma MBBS, DNB, Dipan Desai MS, Kaushik Solanki MS, DO. Delhi Journal of Ophthalmology
Original Article Delhi Journal of Ophthalmology To Compare Incisional Fluid Loss During Different Stages of Phacoemulsification and its effect on Anterior Chamber Stability in Micro Coaxial and Bimanual
More informationOriginal article. Amon M Pediatric micro-incision cataract surgery Nepal J Ophthalmol 2011;3(5):3-8
Original article Surgical management challenges and clinical results of bimanual micro-incision phacoemulsification cataract surgery in children with congenital cataract Department of Ophthalmology and
More informationRESEARCH ARTICLE. Madan Ashok *, Deshmukh Mona, Sute Supriya,Tekade Pradeep. Government medical college and hospital, Nagpur
RESEARCH ARTICLE TO COMPARE THE INTRAOPERATIVE EFFICIENCY, SAFETY AND VISUAL OUTCOMES BETWEEN COAXIAL CONVENTIONAL AND MICROINCISIONAL PHACOEMULSIFICATION Madan Ashok *, Deshmukh Mona, Sute Supriya,Tekade
More informationCataract Surgery in the Patient with a History of LASIK or PRK
Cataract Surgery in the Patient with a History of LASIK or PRK #56996-RS April 2018 Sebastian Lesniak, MD Matossian Eye Associates None Disclosures Bio Matossian Eye Associates, Hopewell NJ, 7/2015 Present
More informationOphthalmology. Cataract Surgery and IOL Implants
Ophthalmology Cataract Surgery and IOL Implants APOLLO_Krames_Cataract surgery and IOL implants_print.indd 1 5/9/2016 5:38:55 PM What is a cataract? Sclera A cataract is an eye disease in which Ciliary
More informationGives you the chamber stability you demand
Gives you the chamber stability you demand See better. Live better. STELLARIS ACTIVATE The platform developed for optimum phaco experience for patients, surgeons and theatre staff SAFETY Solid chamber
More information