CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA
|
|
- Roderick Scott
- 5 years ago
- Views:
Transcription
1 CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy (IMRT), radiotherapy treatment precision requires evolving quality assurance. 2-D arrays with ionization chambers and portal dosimetry system have been widely used and characterized for dose verification of IMRT. Amerio et al. (2004) described the design and construction of ImatriXX 2-D array, whereas Herzen et al. (2007) extensively evaluated it s dosimetric properties. Dosimetric properties of asi1000 electronic portal imaging device (EPID) and it s validation and clinical use were mentioned in a literature survey by Elmpt et al. (2008). In that survey, they reported about the sag of portal imager while using for QA other than zero degree gantry angle. 2-D array detectors have also shown angular dependence while positioned on the treatment couch. Li et al. (2010) and Shimohigashi et al. (2012) reported the angular dependence of central and off-axis detectors in a 2-D ionization chamber array, MatriXX, and suggested the correction methods to improve the accuracy in patient specific QA results. Studies are limited about the dosimetric evaluation of 2-D arrays while positioned in gantry holder as well as the angular dependence of portal dosimetry system for IMRT patient specific QA in true gantry angles. With this background 68
2 this study was performed, to assess the angular response of asi1000 EPID of Varian medical system and the ImatriXX 2-D array system of IBA dosimetry and to validate the detectors for the IMRT patient specific QA measurements in true gantry angles. 5.2 Materials and Methods In this study all the measurements were performed on a 6 MV beam Varian Clinac ix linear accelerator (LINAC). The asi 1000 EPID of Varian medical system and ImatriXX 2-D array system of IBA dosimetry were subjected to study the angular response and to validate both the systems for IMRT patient specific QA with true gantry angles. To study the angular response of the detectors, profiles and output were measured at gantry increments of 10 degrees for a 10x10 cm 2 field. Flatness, symmetry and output values were compared with those for the reference zero degree gantry angle measurements. For ten dynamic IMRT plans (total of 65 fields), patient specific QA tests were performed using both detectors. Two sets of measurements were done (i) with all gantry angles reset to Zero and (ii) with true gantry angles as in the treatment plan. The results of the QA tests were compared using gamma criterion of 3%-3mm. Student s t-test was used to calculate the variation in gamma value due to angular changes in both the detectors. 5.3 Results Flatness, symmetry and output values were compared with those for the reference 0 degree gantry angle measurements. The output stability of the detectors were <0.5% (Figure 5.1 and Figure 5.2). Flatness and symmetry were well within acceptable limits of 2% for all gantry angles in both detectors (Figure 5.3 to 5.6). 69
3 Absolute dose(cgy) Absolute dose(cgy) Figure 5.1 Absolute dose (output) vs. gantry angle- asi1000 EPID Figure 5.2 Absolute dose (output) vs. gantry angle- ImatriXX 70
4 Symmetry(%) Flatness(%) Figure 5.3 Flatness vs. gantry angle- asi1000 EPID Figure 5.4 Symmetry vs. gantry angle- asi1000 EPID 71
5 Symmetry(%) Flatness(%) Figure 5.5 Flatness vs. gantry angle- ImatriXX Figure 5.6 Symmetry vs. gantry angle- ImatriXX 72
6 In case of portal dosimetry with all gantry angle reset to zero, three parameters were studied for each IMRT cases-namely (i) area gamma >1% (ii) average gamma and (iii) maximum gamma. The average values of these parameters for the 10 cases were as follows: 0.89, 0.3 and 1.89 respectively with the standard deviation of 0.46, 0.05 and 0.28 for 3%- 3mm criteria. With true gantry angles, the average value of the studied parameters for the 10 cases were 1.17, 0.34, and 1.66 respectively with the standard deviation of 0.48, 0.06 and 0.23 for the 3%-3 mm criteria (Table 5.1). In the patient specific QA with ImatriXX 2-D array system, on an average 98.9 % pixels passed the criterion of 3%-3 mm with a standard deviation of 0.5 for true gantry angle setup. With all gantry angles reset to zero, 99.2% of the pixels passed the said criterion with standard deviation of 0.3 (Table 5.1). P-value in student s t-test for true gantry angle vs. zero gantry angle was more than 0.05, indicating no significant variation in gamma value due to angular changes in both the detectors (Table 5.2). 73
7 Table 5.1 Patient specific QA results using portal dosimetry and ImatriXX 2-D array system: zero gantry angle vs. true gantry angle IMRT case Portal dosimetry with γ 3% & 3 mm Zero Gantry angle Area γ 1% Avera ge γ Max γ Area γ 1% True Gantry angle Average γ Max γ ImatriXX 2D array system γ = 3% & 3 mm Zero True Gantry Gantry angle angle %of pixels passed %of pixels passed Avg SD Table 5.2 Test of significance (t-test) results: Gamma results of zero gantry angle vs. true gantry angle Parameters measured p-value Area Gamma 1% (in portal dosimetry) 0.16 Average Gamma (in portal dosimetry) 0.10 Maximum gamma (in portal dosimetry) 0.09 % pixels passed Gamma 3%-3 mm (in ImatriXX 2-D array)
8 5.4 Discussion The sag of asi1000 detector and the angular dependence of the 2-D array detectors limited the IMRT patient specific QA measurements in zero gantry angles. The VMAT patient specific QA with these detectors are also a subject of current research. While using the 2-D arrays on treatment couch, the angular dependence that must be compensated by applying correction factor [Anup et al. (2009) and Shimohigashi et al. (2012)]. Currently, available correction factor tables have several underlying assumptions: first, these correction factors assume that the response of all ion chambers is identical for each angle; second, that the ion chamber array response from gantry angles zero degree to 180 degree are equivalent to the response from 180 degree to zero degree; and the third, that the response is independent of the direction of rotation. In this scenario, the use of 2-D arrays in gantry holder or use of portal dosimetry system are the alternate options for the IMRT patient specific QA in true gantry angle. While using the 2-D array detector system in gantry holder as well as the EPID for the measurements in true gantry angles, the QA results can be influenced by the gravity effect, MLC positional errors as well as sag in gantry and detectors. So before performing IMRT QA with true gantry angles, one must make sure that there is no significant variation in profile characteristics and output due to angular changes. In this study, the test results indicate no angular dependence on gamma value while using for measurements in true gantry angle with both detectors. Flatness and symmetry values for profiles did not exhibit any gantry angle dependence and so was the output. Both the detector system can be use for the patient specific QA of IMRT with fields in true gantry angle. Compared to 75
9 ImatriXX 2-D array system, the portal dosimetry system is easy to use for the measurements in true gantry angle and the set up and measurements are also much easier. 76
Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?
Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or
More informationDosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar
More informationEvaluation of Dosimetry Check software for IMRT patient-specific quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,
More informationEvaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell
Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating Characteristic (ROC) Analysis by Allison Lorraine Mitchell Graduate Program in Medical Physics Duke University Date: Approved:
More informationVerification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,
More informationQuality assurance in external radiotherapy
Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance
More informationbiij Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies Krishna Murthy K *
Available online at http://www.biij.org/2012/4/e28 doi: 10.2349/biij.8.4.e28 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Patient-specific quality assurance of RapidArc treatments:
More informationIMRT QA: Point Dose Measurements or 2D Array?
QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationEric E. Klein, Ph.D. Chair of TG-142
Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing
More informationVolumetric Modulated Arc Therapy - Patient Specific QA
Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck
More informationA comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy
Radiology and Oncology Ljubljana Slovenia www.radioloncol.com research article 307 A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy Jaeman Son 1,2,
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More information8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient
Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,
More informationDosisverifikation mit Delta 4 Discover während der Behandlung
Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover
More informationVerification of Advanced Radiotherapy Techniques
Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December
More informationIMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors
IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors S. Sathiyan, M. Ravikumar, C. Varatharaj, S.S. Supe, S.L. Keshava Department of Radiation Physics, Kidwai
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationPatient and Linac QA - present and the future
Patient and Linac QA - present and the future IBA Dosimetry Sandra Kos Service Department Customer Support Specialist sandra.kos@iba-group.com 1 IBA Dosimetry Part of IBA Group OUR MISSION At IBA we dare
More informationARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.
ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael
More informationLearning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy
3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy Sofie Ceberg, PhD Medical Physicist and Lund University Lund, Sweden Learning objectives 1: 3D Dosimetry in the Clinic: Background
More informationDose rate response of Digital Megavolt Imager detector for flattening filter-free beams
Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free
More informationOnline Care, Online Control, Online Confidence
Online Care, Online Control, Online Confidence * Online Treatment Monitoring and much more! Designed for Efficiency provides you with online control and online confidence of treating your patients exactly
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationDELIVERED PATIENT DOSE
SOFTWARE DELIVERED PATIENT DOSE The only software that provides a true picture of patient and dose at each stage of treatment. Adaptivo patient dosimetry software does something other patient dosimetry
More informationEPID dosimetry for pretreatment quality assurance with two commercial systems
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 EPID dosimetry for pretreatment quality assurance with two commercial systems Daniel W. Bailey, 1,2a Lalith Kumaraswamy, 1 Mohammad
More informationMEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO
HOW TO QUALITY CONTROL AND PRE-TREATMENT QUALITY ASSURANCE APPLICATION OF EPID (as1000) FOR FF AND FFF BEAM VMAT PLANS Y. Mekuria 1, M. Bjorkqvist 1, J. Kulmala 1 1 Turku University Hospital/Radiotherapy
More informationQuality assurance and credentialing requirements for sites using inverse planned IMRT Techniques
TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery
More informationUse of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments
Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331
More informationAgenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry
Controlli di qualità preclinici in IMRT statica e dinamica Michele Stasi S.C. Fisica Sanitaria, A.O. Ordine Mauriziano D.O. Fisica Sanitaria, IRCCS Candiolo TORINO mstasi@mauriziano.it Agenda 1. Introduction
More informationActivity report from JCOG physics group
2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer
More informationCyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017
CyberKnife Technology in Ablative Radiation Therapy Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 Objectives Components and work flow of CyberKnife Motion management of CyberKnife
More informationPMP. Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy.
Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.61 pissn 2508-4445, eissn 2508-4453 Gamma Evaluation with Portal Dosimetry for Volumetric Modulated
More informationA new geometric and mechanical verification device for medical LINACs
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein
More informationUse of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments
Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Jean M. Moran, Ph.D. Department of Radiation Therapy, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor
More informationEPID dosimetry confi guration and pre-treatment IMRT verifi cation
Rep Pract Oncol Radiother, 2007; 12(6): 307-312 Original Paper Received: 2007.04.26 Accepted: 2007.08.29 Published: 2007.12.27 Authors Contribution: A Study Design B Data Collection C Statistical Analysis
More informationIBA Dosimetry Company Presentation. Patient Safety & Quality Control
IBA Dosimetry Company Presentation Patient Safety & Quality Control IBA Mission Protect, enhance and save lives Protect, Enhance, and Save Lives - 2 - The History of IBA Dosimetry A story of innovation
More informationUnrivaled, End-to-End
PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by
More informationTwo-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance
Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing
More information3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients
3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients G Asuni *1, T vanbeek 1, E VanUtyven 1, P McCowan 1,2, and B.M.C. McCurdy 1,2,3 1 Medical Physics Department, CancerCare
More informationNormal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT
Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using
More informationIntroduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology
Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology Philip Mayles Clatterbridge Centre for Oncology, Wirral, UK 1 What is RapidArc (or VMAT)?
More informationiviewdose Confidence and assurance in dose delivery
iviewdose Confidence and assurance in dose delivery As radiotherapy techniques advance and treatment plans become more complex, involving escalated doses and increasingly conformal deliveries, the need
More informationDosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.87 Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Mee
More informationLearning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan
Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams
More informationImplementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy
Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out
More informationA review of RTTQA audit. Karen Venables
A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines
More informationPre-treatment verification for IMRT / IGRT and Rotational therapy
In OmniPro-I mrt Vivo treatment software verification Pre-treatment verification for IMRT / IGRT and Rotational therapy Introduction Introduction Introduction Introtro Intro Intro The rapid development
More informationTrajectory Modulated Arc Therapy: Application to Partial Breast Irradiation. Research and development to advance radiotherapy
Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation Dimitre Hristov Radiation Oncology Stanford University Research and development to advance radiotherapy New imaging platforms:
More informationOptimized Varian asi portal dosimetry: development of datasets for collective use
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 6, 2013 Optimized Varian asi portal dosimetry: development of datasets for collective use Ann Van Esch, 1,2a Dominique P. Huyskens, 1,2 Lukas
More informationThe impact of continuously-variable dose rate VMAT on beam stability, MLC positioning, and overall plan dosimetry
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 The impact of continuously-variable dose rate VMAT on beam stability, MLC positioning, and overall plan dosimetry Christopher Boylan,
More informationClinical Impact of Couch Top and Rails on IMRT and Arc Therapy
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2011 Clinical Impact of Couch Top and Rails on IMRT and Arc Therapy
More informationA model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms
1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,
More informationAn Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements
An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements Jean M. Moran, a Donald A. Roberts, Teamour S. Nurushev, Larry E. Antonuk, Youcef El-Mohri, and Benedick A. Fraass Department
More informationGuidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy
Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...
More informationVarian Treatment. Streamlined Treatment Delivery Management Application. Specifications
Varian Treatment Streamlined Treatment Delivery Management Application Specifications Specifications Varian Treatment 1 Introduction Streamlined Treatment Delivery Management Varian Treatment verifies
More informationDebate Motion. Theoretical benefits of VMAT: Real Costs of VMAT. "Arc Based Techniques Will Make Conventional IMRT Obsolete
Austin, TX July 21, 2014 Debate Motion "Arc Based Techniques Will Make Conventional IMRT Obsolete AGAINST the Motion: Peter Balter, Ph.D. Potential Conflicts of Interest: Dr. Balter has research support/collaborations
More informationMany vendors are beginning to allow couch motion during radiation delivery.
Dynamic Couch Motion Many vendors are beginning to allow couch motion during radiation delivery. Varian developer mode allows institutions to perform research using these types of treatments. Tomotherapy
More informationS. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY
More informationREPLACING PRETREATMENT VERIFICATION WITH IN VIVO EPID DOSIMETRY FOR PROSTATE IMRT
doi:10.1016/j.ijrobp.2006.11.047 Int. J. Radiation Oncology Biol. Phys., Vol. 67, No. 5, pp. 1568 1577, 2007 Copyright 2007 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/07/$ see front
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More information7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis
Acknowledgments Thanks to Saiful Huq for an illuminating conversation about the application of TG-100 Jennifer O Daniel, Ph.D. & Fang-Fang Yin, Ph.D. Duke University Medical Center Annual AAPM Meeting,
More informationFeasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry
Baek et al. Radiation Oncology 2014, 9:273 RESEARCH Open Access Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry Tae Seong Baek 1,2, Eun Ji
More informationAmendment No. 2. Item No. 2 (Rfx/ Event number )
Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have
More informationEPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 EPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams Narges Miri, 1a Peter Keller, 2 Benjamin
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationTechnical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.
Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution
More informationA comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment
Oncology and Translational Medicine DOI 10.1007/s10330-016-0183-3 December 2017, Vol. 3, No. 6, P260 P266 REVIEW ARTICLE A comparative study between flattening filter-free beams and flattening filter beams
More informationComparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma
Lin et al. Radiation Oncology 2014, 9:71 RESEARCH Open Access Comparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma Hailei
More informationAn Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center
An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions
More informationA new transmission methodology for quality assurance in radiotherapy based on radiochromic film measurements
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 A new transmission methodology for quality assurance in radiotherapy based on radiochromic film measurements Leonardo L. do Amaral,
More information7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.
UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems
More informationThe MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates Daniel W. Bailey, 1,2 a Jason D. Spaans, 2
More informationI. Equipments for external beam radiotherapy
I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),
More information5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany
Evaluation of 3D diamond detectors for application in medical radiation dosimetry K. Kanxheri (1,2), L. Servoli (2), C. Zucchetti (5), A. C. Dipilato (5), M. Iacco (5), S. Lagomarsino (3,4), A. Morozzi
More informationiba-dosimetry.com INTEGRATED QUALITY ASSURANCE Solution Overview
iba-dosimetry.com Solution Overview 2 3 IBA DOSIMETRY WHY? Protect, Enhance and Save Lives At IBA we share ideas and know-how with our customers and partners to provide innovative solutions for the diagnosis
More informationD DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology
Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran
More informationThe Journey of Cyberknife Commissioning
The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationLinac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR
Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,
More informationPre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system
Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system E. Mezzenga, * E. Cagni, A. Botti, M. Orlandi and M. Iori Medical Physics Unit, IRCCS Arcispedale
More informationTraceability and absorbed dose standards for small fields, IMRT and helical tomotherapy
Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationIMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor
IMRT/IGRT Patient Treatment: A Community Hospital Experience Charles M. Able, Assistant Professor Disclosures I have no research support or financial interest to disclose. Learning Objectives 1. Review
More informationSpatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.
Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation
More informationFollow this and additional works at:
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2010 Commissioning and implementation of an EPID based IMRT system Dosimetry Check for 3D absolute dose measurements
More informationYour Global Quality Assurance Platform All-in-One. All Connected. All Secure.
Your Global Quality Assurance Platform All-in-One. All Connected. All Secure. www.iba-dosimetry.com Dealing with complexity in your QA? All-in-One. Linacs OK Patients OK Reports APPROVED Satellite Center
More informationReal-time Volumetric Scintillation Dosimetry for Radiation Therapy
Real-time Volumetric Scintillation Dosimetry for Radiation Therapy Sam Beddar, Ph.D., FCCPM, FAAPM Professor & Chief of Research Department of Radiation Physics The University of Texas MD Anderson Cancer
More informationBeam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate Ziyad A. Alrowaili, 1,2 Michael L.F. Lerch,
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationEVALUATION OF INTENSITY MODULATED RADIATION THERAPY (IMRT) DELIVERY ERROR DUE TO IMRT TREATMENT PLAN COMPLEXITY AND IMPROPERLY MATCHED DOSIMETRY DATA
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2011 EVALUATION OF INTENSITY MODULATED RADIATION THERAPY (IMRT) DELIVERY
More informationRADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident
Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate
More information8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation
Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research
More informationStatistical process control analysis for patient-specific IMRT and VMAT QA
Journal of Radiation Research, 2013, 54, 546 552 doi: 10.1093/jrr/rrs112 Advance Access Publication 7 December 2012 Statistical process control analysis for patient-specific IMRT and VMAT QA Taweap SANGHANGTHUM
More informationLarge-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance
Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance J. Gajewski1, M. Kłosowski1, S. Greilich2, P. Olko1 Institute of Nuclear Physics, Kraków, Poland German Cancer
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationIn-Room Radiographic Imaging for Localization
In-Room Radiographic Imaging for Localization Fang-Fang Yin, Zhiheng Wang, Sua Yoo, Devon Godfrey, Q.-R. Jackie Wu Department of Radiation Oncology Duke University Medical Center Durham, North Carolina
More information