Radiation Physics Principles of DXA Basic Statistics for DXA Essential Anatomy for DXA

Size: px
Start display at page:

Download "Radiation Physics Principles of DXA Basic Statistics for DXA Essential Anatomy for DXA"

Transcription

1 Strong Bone Asia 2013 Osteoporosis in ASEAN Radiation Physics Principles of DXA Basic Statistics for DXA Essential Anatomy for DXA Chris Schultz Scientist-in-Charge (Bone Densitometry) Royal Adelaide Hospital, Adelaide, Australia Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

2 Strong Bone Asia 2013 Osteoporosis in ASEAN Radiation Physics Ionising Radiation in Bone Densitometry Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

3 Electromagnetic Spectrum Wavelength Frequency (Energy) Ionising ev Radio waves Powers of ten represent orders of magnitude difference, with respect to the visible spectrum, of the photon energy (or frequency) Also represent the approx. photon energy in electron volts (ev; 1eV = 1.6 x J) Infra-red Ultra violet Visible DXA X-rays Therapy X & rays ANZBMS Slide 3

4 Key properties of Diagnostic X-rays Diagnostic X-rays have photon energies in the range keV Absorption & scattering of X-rays occurs in matter - including human tissue. X-rays interact with atoms (ie electron clouds) not molecules or cells X-rays produce biological effects that are seen as direct or indirect actions on living cells Johns & Cunningham (1983) ANZBMS Slide 4

5 Penetration of Matter by Diagnostic X-rays aluminium lead photon energy X-ray beam Basis of X-ray attenuation, imaging, & shielding Aluminium selectively attenuates low energy photons (beam hardening) removing rays that don t contribute to the image ( filtration ) Body tissue allows some of the beam through without interaction, creating the image. This is the primary beam. The rest of the beam causes the radiation dose Lead blocks the beam and is suitable for radiation shielding ANZBMS Slide 5

6 Interactions between X-rays & Matter Absorption with production of secondary electrons &/or new X-rays Scattering resultant photon has lower energy and new direction Heggie et al (2001) Attenuation is the sum of these - the reduction in the forward intensity of the X-ray beam as it traverses matter - depending on: The physical density & atomic composition of the material The energy spectrum of the X-ray beam ANZBMS Slide 6

7 Inverse Square Law of radiation intensity or dose When the distance from a point radiation source is doubled, the intensity is reduced by one quarter. This is the basis of radiation protection by distance I 2 I 2 I 1 d d I 1 source d 2 d 1 ANZBMS Slide 7

8 Biological Effects of Ionising Radiation CELL DAMAGE ALTERED METABOLISM & FUNCTION TRANSFORMATION REPAIR CELL DEATH Radiosensitivity of organelles and cells varies DNA region > 100 times more sensitive than cytoplasm Ion pairs created in or near DNA by radiation (and by chemical environmental insults) Therefore rapidly dividing cells more susceptible to damage (eg ovaries, testes, bone marrow, intestinal epithelium) The foetus and young children are therefore more sensitive to the effects of radiation Hallenbeck (1994) ANZBMS Slide 8

9 Types of Radiobiological Effects (A): Deterministic Directly caused by the radiation (eg destruction of bone marrow cells following radiotherapy level dose) Cellular repair mechanisms overwhelmed by the severity of the insult and most cells die Occur only above a specific threshold - typically as high as 200 times yearly background dose Above this threshold severity of effect dosedependent eg; ICRP60 (1990) ANZBMS Slide 9

10 Types of Radiobiological Effects (B): Stochastic Stochastic effects are assumed to arise from damage to a single cell, from low radiation doses The majority of damaged cells die, but a few abnormal cells survive The effect cannot be causally related to a particular radiation incident, because the consequences usually arise years later The probability of occurrence, but not the severity of a stochastic effect is proportional to the dose received It is possible that there is no safe dose, though at low doses this is an assumption based on mathematical modelling and not evidence based Most radiation protection environments and calculations (including bone densitometry) are concerned with assessing and predicting stochastic effects eg ICRP60 (1990) ANZBMS Slide 10

11 Strong Bone Asia 2013 Osteoporosis in ASEAN Radiation Physics Quantitative Radiation Protection Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

12 Long-term Biological Effect of Exposure to Ionising Radiation The biological consequences of ionising radiation are among the most predictable of all environmental hazards This is because: the magnitude of radiation exposure can often be recorded (or estimated) reasonably accurately deep follow-up studies have been carried out on thousands of victims for up to 65 years The key to predicting long term detriment is to devise a quantitative measure of the radiobiological effect ANZBMS Slide 12

13 Calculating Effective Dose (E = H T. T ) Definition: Effective Dose (E) sum of the equivalent doses (H T ) for each organ or discrete exposed tissue multiplied by tissue weighting factor ( T ) for that organ sum for all tissues and organs within the exposed region of the body SI Unit : sievert (Sv) E in bone densitometry often quoted in microsievert ( Sv) Effective Dose is the most biologically direct, quantitative measure of the long term risk of severe detriment (cancer, inherited effects) from the stochastic effects of radiation Tissue or Organ Tissue Weighting Factor ( T ) * Gonads 0.08 Bone marrow (red), Colon, Lung, Stomach, Breast (each) Bladder, Liver, Oesophagus, Thyroid (each) Skin, Bone surface, Brain (each) Other organs/tissues (entire) Total 1.0 relative radio-sensitivities of organs/tissues ICRP60 (1990), ICRP103 (2007) ANZBMS Slide 13

14 Calculation of stochastic risks for Adults Risk of long term effects Multiply the effective dose in Sv by the coefficient below for the relevant population & detriment Exposed population Detriment (Sv -1 ) All Cancer Severe hereditary effects Total Adult Workers Whole Population Note that workers are more resistant to radiation ICRP103 (2007) ANZBMS Slide 14

15 Calculation of Stochastic Risks for Adults Equivalent examples Ionising Radiation Person receiving a man-made dose equal to background radiation per year (~2.5mSv) for 10 years has (2.5 x10-3 ) x 10 x 0.05 = ~0.1% extra subsequent lifetime risk of cancer that could be fatal (ie above normal risk ~25%) Roads Person has about the same (0.1%) risk of dying on Australian roads over a 10-year period ICRP60 (1990) Australian Bureau of Stats (2002) ANZBMS Slide 15

16 Standard DXA Effective Doses ( Sv) Scan Mode PA Spine (L1-L4) Prox. Femur (inc. ovaries) Total Body (ex. ovaries) Total body (inc. ovaries) QDR 4500/Delphi Fan Beam a Expert Fan Beam b Norland (Pencil Beam) Systems c <1 <0.5 3 <1 d <1 Notes: * Lateral spine can deliver up to 2-6 x higher dose than PA, depending on low vs high definition * Quoted values are effective doses, not the more frequently-quoted entrance surface doses (ESD), which are easier to determine but are 2 steps removed from biological significance. * In postmenopausal women, the ovaries no longer carry a significant organ weighting (w T ). Refs: a: Patel et al. (1996). b, c; Established from manufacturer quoted entrance exposures or doses, & Bezakova et al. (1997), Huda & Moran (1996). ANZBMS Slide 16

17 Comparative doses in medical diagnosis & daily life Comparison Dose ( Sv) BERT * Natural Background (all sources) 5-8/day a 1 day Trans-Pacific Return Flight (cosmogenic burden) ~50 b ~4-6 days Standard chest X-ray c 3-7 days Mammography ~450 d ~56 days Thoraco-lumbar Lateral & AP X-ray set e days Thoracic CT or Whole-body PET >7000 f 1000 days Two PA DXA scans (excluding LVA) <12 g < 1 day * Background Equivalent Radiation Time (days), based on average background dose Refs: a: UNSCEAR (2000). b:adapted from Hallenbeck (1994). c,d,e: Standard values from variety of sources, inc. Johns & Cunningham (1983) & NRPB Vol. 10 #1 (1999); f: Price et al calc.mtp, SCGH, Perth, based on 16-slice CT & GSO crystal PET camera (2005). g: See previous slide ANZBMS Slide 17

18 Strong Bone Asia 2013 Osteoporosis in ASEAN Radiation Physics Minimising Radiation Doses to Patient & Technologist Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

19 Explaining the Risk-Benefit of Radiation to Patients Yearly natural background ~ >100 times most DXA Radiation dose in intercontinental flights >5-10 times DXA Other commonly accepted risk factors (eg car accidents) outweigh radiation risks (be prudent though!) Benefits of the DXA investigation - likely impact on quality & length of life (~30% untreated lifetime fracture risk, reduced by >half by proper management, in the spine) The deep knowledge of risks of radiation (eg ICRP, IAEA), compared with other hazards ANZBMS Slide 19

20 Patients Scanning children Extra duty of care must be applied to children Reference ranges acquired using children should be subjected to the strictest of justifications Care should be taken by Technologist to ensure that no scan is wasted Radiosensitive organs (eg gonads) should be shielded, wherever possible Children would not be scanned using QCT, except as pqct (eminently suited to children) Dose limits for children recognise their developmental status eg; ICRP85 (2000); ARPANSA, #8 (2004); Hall, Paediatric Radiology (2002) ANZBMS Slide 20

21 Patients Pregnant or Possibly pregnant Exposures to the axial skeleton incur the highest burden Possibly pregnant patients require particular duty-of-care Pregnancy status of reproductive women should always be established ask! The relative risk depends on the stage of pregnancy. The most sensitive period is 6-10 weeks Bone densitometry is never so urgent that it cannot be postponed However, in all bone densitometry, in the event of any inadvertent dose to the foetus, stochastic risk is small, compared to the natural occurrence of 4-6% congenital risks Summary In all cases of the inadvertent use of DXA, the pregnant patient may be reassured of negligible risks eg; Given-Wilson (1993); Lloyd et al (1998); NHMRC (1986) ANZBMS Slide 21

22 Strategies for Minimising Technologist Doses (I) General ALARA Strategies Time Minimise contact with X-ray beam. Dose is proportional to time spent Distance Maximise distance from source. Protection afforded by inverse square law Shielding Screen (if necessary) between technologist and X-ray source, particularly if work station in same room as DXA. A lead glass screen of 0.25 mm lead-equiv. thickness represents >4HVL for DXA scattered radiation >93% of radiation blocked ANZBMS Slide 22

23 Personal Radiation Monitoring: Judging the Need Some Radiation Regulators require personal radiation monitoring of DXA Technologists, at least initially Application for exemption can be made on basis of dose records that show no significant readings (eg one year) If staff feel safer with personal radiation monitoring, then it might be prudent to offer it, until they acknowledge the lack of need Exceptions, suggestive of continuous monitoring, are: When work practices habitually require more physical interaction between technologist and patient during scanning (eg paediatrics) In the case of QCT using a whole body scanner, the technologist will be a licensed radiographer ANZBMS Slide 23

24 Strategies for Minimising Technologist Doses Pregnant staff ICRP 60 recommends that the dose to the surface of mothers abdomen during pregnancy should be <2 msv For bone densitometry operators this is easily avoided without taking extra precautions Pregnancy in the workplace should be handled with sensitivity. In DXA-only workplaces, the exposure to the foetus will always be negligible, but perception of risk creates anxiety Concern can be dealt with by education and negotiation. A leaded apron costs <$1000 and absorbs the vast majority of dose (0.25 mm lead equiv. Gown ~ >4 HVL of scattered DXA radiation) ICRP60 (1990) ANZBMS Slide 24

25 Strategies for Minimising Doses to Technologists Specific Strategies Strictly minimise exposure to direct beam (relevant for when restricting patient movement) Check that Safety signs (for both ionising radiation & lasers) & lights (if relevant) are logical and obeyed by personnel & patients ionising laser Conduct systematic, regular Best Practice Quality Assurance (reduces need for repeat scanning) Seek licensed maintenance & repair personnel ANZBMS Slide 25

26 Strong Bone Asia 2013 Osteoporosis in ASEAN Principles of Dual Energy X-ray Absorptiometry Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

27 Photon Absorption Mass attenuation coefficient Photon attenuation: Depth = attenuation Atomic number = attenuation photon energy = attenuation Mass attenuation coefficient ( ) for a given material: decreases with photon energy rises with atomic number Sharp changes associated with energy levels (shells) of atom ANZBMS Slide 27

28 Mass Attenuation coefficient (cm 2 /g) Mass attenuation coefficient & ideal energies 2 distinct energies can differentiate between tissue and bone Photon absorption water, tissue, muscle are similar in fat in bone Soft Tissue 1.00 Adipose Tissue Muscle, Skeletal Cortical Bone 0.10 Water Calcium Photon Energy (KeV) ANZBMS Slide 28

29 Theory of mass calculation (1) Solve for two masses bone and soft tissue Beam attenuation soft tissue low energy : high energy Equivalent to ratio of mass attenuation coefficients at each energy M M bone tissue I ln( I μ 0high high bonelow I ln( I μ R R 0high high tissuelow bone tissue ).R μ μ μ ).R μ μ μ tissue bonehigh bone I ln( I.R tissuehigh bonelow bonehigh tissuelow tissuehigh 0low low tissue I ln( I.R 0low low bone ) ) ANZBMS Slide 29

30 Mass Attenuation coefficient (cm 2 /g) Theory of mass calculation (2) Scan spine, femur, forearm and total body (any site) Fat correction possible Ideal energies calculated to be between ~20Kev and ~100KeV KeV 100KeV Soft Tissue 1.00 Adipose Tissue Muscle, Skeletal Cortical Bone 0.10 Water Calcium Photon Energy (KeV) ANZBMS Slide 30

31 Strong Bone Asia 2013 Osteoporosis in ASEAN Principles of Dual Energy X-ray Absorptiometry DXA Scanner structure, X-ray generation Beam collimation Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

32 Scanner structure Diagram adapted from Wahner HW et al J Nucl Med. (1984) ANZBMS Slide 32

33 X-ray generation Hologic Switch tube kv between 100 & 140kV (was 70 & 140kV) Pulse tube at power line frequency Calibrate & correct for beam hardening using filter wheel R tissue constant (corrected calibration empirically) CdWO 4 detector initially single, followed by array ANZBMS Slide 33

34 X-ray generation LUNAR Factory calibrated to ashed bone Rare earth filter (Ce or Sm) uses K shell absorption of photon energy as filter Tube ma & speed changes alter dynamic range R tissue corrected pixel by pixel using tissue adjacent to bone CdZnTe in narrow fan beam configuration, without need for dead time correction Phantom block for Quality Assurance checks engineering specification only ANZBMS Slide 34

35 Pencil beam vs Fan Beam Pencil beam 1:1 relationship of source & detector Relatively slow scan time Fan Beam Faster scan Inverse square law applies at edges (reduced photon flux) Magnification effect (source location proximity to patient bone) Narrow fan beam Rectilinear type scans Some beam overlap Potentially less artefact Slower than full fan beam ANZBMS Slide 35

36 Strong Bone Asia 2013 Osteoporosis in ASEAN Principles of Dual Energy X-ray Absorptiometry 3 tissue component approximation Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

37 3 compartments from 2 DXA is 2 compartment model bone and soft tissue Possible to derive fat and lean mass Relies on documented fat and lean R values Limited by masking due to bone areas estimate ANZBMS Slide 37

38 3 compartment derivation Given R material μ μ materiallow materialhigh and Fraction fat + Fraction lean = 1 (ie Tissue Total ) then so F F fat lean M M fat lean (R (R fat (R (R M R R fat M ) ) R) R fat tissue tissue lean lean.f lean fat.f ) lean ANZBMS Slide 38

39 Strong Bone Asia 2013 Osteoporosis in ASEAN Principles of Dual Energy X-ray Absorptiometry Limitations of DXA Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

40 Density DXA limitations Measurement is 2 dimensional DXA density is area density Normal Osteoporotic ANZBMS Slide 41

41 DXA limitations Measurement integrates all bone Changes in projection change results (up to 0.5SD for >15 ) Normal Rotated 22 ANZBMS Slide 42

42 DXA limitations Photon flux must be adequate to distinguish between tissue types Low enough to not affect R tissue or R bone ANZBMS Slide 43 ANZBMS Slide 43

43 DXA limitations Artefacts can affect calculations Oral contrast Intravenous contrast ANZBMS Slide 44

44 Strong Bone Asia 2013 Osteoporosis in ASEAN Basic anatomy for bone densitometry Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

45 Cancellous Bone Bone Microstructure NORMAL BONE OSTEOPOROTIC BONE ANZBMS Slide 47

46 Cortical Bone Bone Microstructure NORMAL BONE OSTEOPOROTIC BONE ANZBMS Slide 48

47 Bone Microstructure Periosteum membrane covering the outside of the bone. 2. Bone marrow cavity Endosteum membrane lining the inside of the cortical bone and the trabeculae ANZBMS Slide 49

48 Cortical bone Shafts of long bones Cortical rim of vertebrae ANZBMS Slide 50

49 Distribution of cortical and trabecular bone ANZBMS Slide 51

50 Lumbar Spine Essential Anatomy Proximal Femur Forearm ANZBMS Slide 52

51 Lumbar spine anatomy with relevance to bone densitometry ANZBMS Slide 53

52 Vertebral Anatomy Cortical rim Body Transverse Process Cancellous bone Pedicle Superior articular process Lamina Vertebral foramen Spinous process ANZBMS Slide 54

53 Projectional anatomy of the lumbar spine Lateral view Posterior view Anterior view ANZBMS Slide 55

54 Vertebral Appearance in Densitometry ANZBMS Slide 56

55 Anatomical variations of lumbar Spine Number of vertebrae Approximately 7% of people have missing ribs on T12 Only 2% had 6 lumbar vertebrae ANZBMS Slide 57

56 Anatomical variations of lumbar vertebrae 7.5% of subjects have 4 lumbar Vertebrae. Ribs on L1 are very rare. ANZBMS Slide 58

57 Sacralization L5 sacralization is a common anatomical anomaly, where the L5 vertebral body becomes incorporated into the S1 vertebral body. ANZBMS Slide 59

58 Basic anatomy of the proximal femur with relevance to bone densitometry ANZBMS Slide 60

59 Anatomy of the Hip Femoral neck Femoral head Greater trochanter Lesser trochanter ANZBMS Slide 61

60 Anatomy of Lesser Trochanter ANZBMS Slide 62

61 X ANZBMS Slide 63

62 Effect of rotation of the Lesser Trochanter ANZBMS Slide 64

63 Anatomy of the Forearm Ulna Ulna styloid Radius 33% Shaft 5 mm Ultra -distal Trabecular 5% 35% 75% ANZBMS Slide 65

64 Strong Bone Asia 2013 Osteoporosis in ASEAN Introduction to Statistics for Bone Densitometry Copyright This material is Copyright by the Council of the Australian & New Zealand Bone & Mineral Society (ANZBMS) and may not be copied or transmitted by any means without written permission from the Council of the ANZBMS.

65 Why Make Measurements? Places the observation in universally recognised context To compares one observation with another over time, in a single individual or group, or between individuals or groups To compare an observation with a reference range made up of many similar measurements of normal individuals, in order to determine how unusual it is To examine cause and effect relationships between different measured variables ANZBMS Slide 67

66 Frequency Description of a Measured Variable Magnitude of distribution Where does a measurement usually lie? Three common measures of the predominant value of a measurement Mode the most frequently occurring value Median the value where half the measurements lie above it and half lie below Mean an average that gives greater weighting to larger values ( values / N) = ( )/30 = 5.27 ANZBMS Slide 68

67 Frequency Description of a Measured Variable Measures of the spread of the distribution Within what limits does a measurement usually lie? Range The difference between the lowest and highest values Inter-quartile Range Difference between the top of the first (lowest) quartile and the bottom of the fourth (highest) quartile. 50% of values lie between these two values* Standard Deviation (SD) Estimate of the spread that gives greater weighting to values distant from the Mean * Deciles could also be used in replacement of quartiles n 1 SD = SQRT[(5.27-1) 2 + (5.27-2) 2 + (5.27-3) (5.27-9) 2 + ( ) 2 /(30-1)] = 2.12 x i x Lower quartile Range Inter-quartile range Upper quartile ANZBMS Slide 69

68 Symmetric Distributions the Bell-shaped Curve Most statistical analysis assume measured variable is distributed symmetrically (a bell-shaped curve) Three reasons: Multiple observations of many biological variables follow a bellshape or a mathematical function of it (eg log-normal) Variables not following a bell-shape often have the distribution of means of subgroups of its distribution tending to bell shaped. This powerful result places this type of curve at the centre of statistical analysis! The curve has a mathematical formula so mathematics of the statistical analysis based on the curve is greatly simplified Formally known as the Normal (Gaussian) Distribution ANZBMS Slide 70

69 mean Characteristics of Normally Distributed Data Variables following a Normal Distribution: Describe the predominant value as the Mean = median = mode Describe the spread of data under curve as Standard Deviation 68% of data is within ± 1 SD of the mean 95% of data is within ± 2 SD of the mean 99.89% of data is within ± 3 SD of the mean 50% data 50% data 16.7% data 16.7% data 2.5% data 2.5% data ANZBMS Slide 71

70 Mean Does a Measurement belong to a Particular Data Set? The Z-statistic Does belong to the distribution? ie What is the probability that it is so different from the other points that it is not a member of the group? To calculate the Z-statistic Choose data distribution to compare single data point with Need distribution Mean & SD (must be approximately normal) Z = (X-Mean)/SD X-Mean 1 SD Standard Deviations X ANZBMS Slide 72

71 Measurement Accuracy Systematic Errors Occur when repeated measurements made on the same subject generally agree, but in comparison with a more physically valid technique, are found to differ Difficult to estimate, unless a Gold Standard technique able give correct answer for comparison eg ashing & weighing cadaveric bones following DXA scanning Therefore repeat measurements may agree but may all be wrong Used to calculate the accuracy of the technique (eg DXA), usually expressed as a % variation from the real value, determined using the gold standard ANZBMS Slide 73

72 Measurement Precision Random Errors Occur when repeated measurements made on the same subject do not agree, even though subject s physiological state is unchanged Caused by: fluctuations in the operating characteristics of the machine random operator variations in patient positioning random variations in scan acquisition and scan analysis Easiest errors to estimate, as distribution usually Normal The distribution of random errors is used to calculate the precision of the measurement ANZBMS Slide 74

73 Accuracy vs Precision in clinical densitometry poor precision good accuracy (larger random error) Dual-energy QCT good precision poor accuracy (larger systematic error) DXA QUS Photogrammetry good precision good accuracy (small random & systematic errors) The ideal ANZBMS Slide 75

74 Determining Precision of Bone Densitometry Precision measurements should be made on the population they will be applied to (ie patients), rather than healthy volunteers A good precision estimate requires >28 subjects measured twice Each subject should be removed from scanning couch following first scan, and relocated for the second Such data can be used for short term precision calculation Ethics a source of duplicate scans Certain subjects likely to be reviewed regularly (and therapy varied) may benefit from duplicate measurements at baseline because a bad baseline corrupts future comparisons Duplicate measurements in a clinical practice may also be justified if transferring patients to a new DXA machine Bonnick & Lewis (2002) ANZBMS Slide 76

75 x Calculating Precision of Bone Densitometry SD (s) of the distribution of precision errors (ie random errors of duplicate measurements) is calculated: Where: d i = x i1 - x i2 is the difference in the i th duplicate data set n = the number of subjects Precision error is often expressed as coefficient of variation (CV%) s CV where is the Mean of the 2n measurements The closer to 0% the CV% the greater the reproducibility di 2n 2 s % 100 x Bonnick & Lewis (2002) ANZBMS Slide 77

76 Precision & Accuracy in Bone Densitometry Precision Accuracy Central DXA PA spine 1-2% 4-10% Lateral spine 2-4% 5-15% Femur 1.5-3% 6% Forearm 1% 5% Total body 1% 3% pdxa Forearm 1-2% 4-6% Calcaneus 1-2% 4-6% QCT Spine QCT 1.5-4% 5-15% Peripheral QCT 1-2% 2-8% Genant HK, et al. J Bone Miner Res 1996; 11(6):707 ANZBMS Slide 78

77 The Importance of Good Precision Precision is probably the most important factor to consider in routine bone densitometry High precisional error usually indicates poor work practices or machine compliance Better precision allows smaller changes in BMD to be detected at any selected confidence level and permits any change in bone status to be detected sooner Precision errors are minimised by: regular quality assurance scans to detect machine drift attention to technique - patient positioning, scan acquisition & analysis Significance of small changes in BMD can be assessed by a knowledge of the Least Significant Change (LSC) ANZBMS Slide 79

78 Least Significant Change (1) Serial measurements of BMD are important to manage the patient (eg input into decision as to whether to vary Rx) Important to know the smallest magnitude of change in BMD which is significant ie due to true change rather than random error The Least Significant Change (LSC) for two BMD measurements is the minimum change necessary between serial DXA scans to be certain of a real change in BMD LSC depends on: precision of BMD technique degree of statistical confidence demanded (ie p<0.05, p<0.01) ANZBMS Slide 80

79 Least Significant Change (2) LSC Z * 2 * s where: Z a is the Z-statistic for chosen significance level (eg Z = 1.96 if p<0.05) s is the precision error expressed as the SD of the distribution of differences in duplicate measurements Usually, we are interested in a significance of p<0.05 so: LSC =1.96 x 1.4 x s ~ 2.8 x s If s = 0.01 g/cm 2 then LSC = g/cm 2 If s = 0.15 g/cm 2 then LSC = 0.42 g/cm 2 ANZBMS Slide 81

80 LSC (as %) Importance of good precision in follow-up studies Difference in serial measurements to guarantee a real change 8 4 good poor LSC ~2.8 CV DXA precision (CV%) ANZBMS Slide 82

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017 STRUCTURED EDUCATION REQUIREMENTS Bone Densitometry The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge that, when coupled

More information

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7 STRUCTURED SELF ASSESSMENT CONTENT SPECIFICATIONS SSA LAUNCH DATE: JANUARY 1, 2018 Bone Densitometry The purpose of continuing qualifications requirements (CQR) is to assist registered technologists in

More information

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p.

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p. Dedication Preface Acknowledgments Continuing Education An Introduction to Conventions in Densitometry p. 1 Densitometry as a Quantitative Measurement Technique p. 2 Accuracy and Precision p. 2 The Skeleton

More information

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017 EXAMINATION CONTENT SPECIFICATIONS Bone Densitometry The purpose of the bone densitometry examination is to assess the knowledge and cognitive skills underlying the intelligent performance of the tasks

More information

The Bone Densitometry Examination

The Bone Densitometry Examination The Bone Densitometry Examination The purpose of The American Registry of Radiologic Technologist (ARRT ) Bone Densitometry Examination is to assess the knowledge and cognitive skills underlying the intelligent

More information

Norland Densitometry A Tradition of Excellence

Norland Densitometry A Tradition of Excellence Norland Densitometry A Tradition of Excellence Norland DXA Bone Density Measurement Osteoporosis is a disease marked by reduced bone strength leading to an increased risk of fractures. About 54 million

More information

Bone Densitometry Equipment Operator

Bone Densitometry Equipment Operator Bone Densitometry Equipment Operator The purpose of the Bone Densitometry Equipment Operator Examination, which is made available to state licensing agencies, is to assess the knowledge and cognitive skills

More information

Accuracy of DEXA scanning & other methods for determining BMD.

Accuracy of DEXA scanning & other methods for determining BMD. BMD- Measurement Site Accuracy of DEXA scanning & other methods for determining BMD. Ann Larkin In general, densitometry techniques can be performed in either the axial or the appendicular skeleton, depending

More information

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi Clinical Utility of Bone Densitometry Diagnosis (DXA)

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

CT Imaging of skeleton in small animals. Massimo Marenzana

CT Imaging of skeleton in small animals. Massimo Marenzana CT Imaging of skeleton in small animals Massimo Marenzana Introduction Osteoporosis is a disease in which bones become fragile and more likely to break. It can be defined as a systemic skeletal disease

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Radiation Safety for New Medical Physics Graduate Students

Radiation Safety for New Medical Physics Graduate Students Radiation Safety for New Medical Physics Graduate Students John Vetter, PhD Medical Physics Department UW School of Medicine & Public Health Background and Purpose of This Training This is intended as

More information

Radiation Safety - Things You Need to Know

Radiation Safety - Things You Need to Know Radiation Safety - Things You Need to Know Michael Casey Ph.D. Phlebotomy Autumn Seminar 13 th October 2012 Radiation is a form of energy transport What is Radiation? It is caused by electrical disturbances

More information

Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety

Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February 2008. Radiation Safety Sub -headings Background Radiation Effects of Radiation Safe Levels Effective Doses ALARA Principle Radiation

More information

9 Quality Assurance in Bone Densitometry section

9 Quality Assurance in Bone Densitometry section 9 Quality Assurance in Bone Densitometry section Introduction Bone densitometry is frequently used to determine an individual's fracture risk at a particular point in time but may also be used to assess

More information

2013 ISCD Official Positions Adult

2013 ISCD Official Positions Adult 2013 ISCD Official Positions Adult These are the Official Positions of the ISCD as updated in 2013. The Official Positions that are new or revised since 2007 are in bold type. Indications for Bone Mineral

More information

PHY138Y Nuclear and Radiation

PHY138Y Nuclear and Radiation PHY38Y Nuclear and Radiation Professor Tony Key MP40 key@physics.utoronto.ca Announcements MP problems set #4 due Sunday at midnight PS#5 WRITTEN now posted! - do in teams, no Lone Wolves!! NB correction

More information

Bone Mineral Densitometry with Dual Energy X-Ray Absorptiometry

Bone Mineral Densitometry with Dual Energy X-Ray Absorptiometry Bone Mineral Densitometry with Dual Energy X-Ray Absorptiometry R Gilles, Laurentius Ziekenhuis Roermond 1. Introduction Osteoporosis is characterised by low bone mass, disruption of the micro-architecture

More information

Lateral Vertebral Analysis DXA Body Composition Quality Assurance in DXA

Lateral Vertebral Analysis DXA Body Composition Quality Assurance in DXA Strong Bone Asia 2013 Osteoporosis in ASEAN Lateral Vertebral Analysis DXA Body Composition Quality Assurance in DXA Chris Schultz Scientist-in-Charge (Bone Densitometry) Royal Adelaide Hospital, Adelaide,

More information

X-rays How safe are they?

X-rays How safe are they? X-rays How safe are they? Patient information Thirty years ago, X-rays were the only way to see what was going on inside your body. Now other methods of medical imaging are available, some using different

More information

Basic radiation protection & radiobiology

Basic radiation protection & radiobiology Basic radiation protection & radiobiology By Dr. Mohsen Dashti Patient care & management 202 Wednesday, October 13, 2010 Ionizing radiation. Discussion issues Protecting the patient. Protecting the radiographer.

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Densitometry Techniques

Densitometry Techniques 2 Densitometry Techniques CONTENTS PLAIN RADIOGRAPHY IN THE ASSESSMENT OF BONE DENSITY QUALITATIVE MORPHOMETRY QUALITATIVE SPINAL MORPHOMETRY THE SINGH INDEX QUANTITATIVE MORPHOMETRIC TECHNIQUES CALCAR

More information

Documentation, Codebook, and Frequencies

Documentation, Codebook, and Frequencies Documentation, Codebook, and Frequencies Dual-Energy X-ray Absorptiometry Femur Bone Measurements Examination Survey Years: 2005 to 2006 SAS Transport File: DXXFEM_D.XPT January 2009 NHANES 2005 2006 Data

More information

Certified Clinical Densitometrist (CCD ) DXA Resource Materials

Certified Clinical Densitometrist (CCD ) DXA Resource Materials Certified Clinical Densitometrist (CCD ) DXA Resource Materials International Society for Densitometry DXA Resource Materials Editors Steven Petak, MD, JD, CCD Neil Binkley, MD, CCD Sue Broy, MD, CCD

More information

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy What is radiation? Radiation is energy emitted from a source, that travels through space and can penetrate matter. Listed below are two types that we are exposed to and contribute to our overall radiation

More information

Bone Densitometry. What is a Bone Density Scan (DXA)? What are some common uses of the procedure?

Bone Densitometry. What is a Bone Density Scan (DXA)? What are some common uses of the procedure? Scan for mobile link. Bone Densitometry What is a Bone Density Scan (DXA)? Bone density scanning, also called dual-energy x-ray absorptiometry (DXA) or bone densitometry, is an enhanced form of x-ray technology

More information

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine Radiation physics and radiation protection University of Szeged Department of Nuclear Medicine Radiation doses to the population 1 Radiation doses to the population 2 Sources of radiation 1 Radiation we

More information

UQ X-ray Safety Training Module

UQ X-ray Safety Training Module UQ X-ray Safety Training Module 23 January 2018, v2 1 UQ X-ray Safety Training Module Course Overview: This training module has been developed for workers at the University of Queensland, and forms part

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Module Rhodes

Module Rhodes Module 6 10-526-197 Rhodes Health Physicist Concerned with providing occupation radiation protection and minimizing radiation dose to the public. Diagnostic Imaging has changed our world Live longer Work

More information

ICRP Recommendations Evolution or Revolution? John R Cooper Main Commission

ICRP Recommendations Evolution or Revolution? John R Cooper Main Commission ICRP Recommendations Evolution or Revolution? John R Cooper Main Commission 3 September 2009 ICRP Recommendations 1. Reasons for new Recommendations 2. Summary of health risks 3. Summary of changes to

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT O S T E O P O R O S I S A S S E S S M E N T Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT Identify Patients at Risk The evaluation of bone health has become an essential

More information

2 Skeletal Anatomy in Densitometry

2 Skeletal Anatomy in Densitometry 2 Skeletal Anatomy in Densitometry CONTENTS CHARACTERIZING THE SKELETON IN DENSITOMETRY THE SPINE IN DENSITOMETRY THE PROXIMAL FEMUR IN DENSITOMETRY THE FOREARM IN DENSITOMETRY THE METACARPALS, PHALANGES,

More information

GUIDELINES ON IONISING RADIATION DOSE LIMITS AND ANNUAL LIMITS ON INTAKE OF RADIOACTIVE MATERIAL

GUIDELINES ON IONISING RADIATION DOSE LIMITS AND ANNUAL LIMITS ON INTAKE OF RADIOACTIVE MATERIAL RADIATION PROTECTION AUTHORITY OF ZIMBABWE (RPAZ) RADIATION PROTECTION ACT [CHAPTER 15:15] GUIDELINES ON IONISING RADIATION DOSE LIMITS AND ANNUAL LIMITS ON INTAKE OF RADIOACTIVE MATERIAL Compiled by Radiation

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Children s Healthcare of Atlanta

Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Children s Healthcare of Atlanta Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Technical: Patient positioning Performance of the scan Analysis of the data Theoretical: Identification

More information

Assessment of effective dose in paediatric CT examinations

Assessment of effective dose in paediatric CT examinations Assessment of effective dose in paediatric CT examinations E. Dougeni 1,2 CL. Chapple 1, J. Willis 1, G. Panayiotakis 2 1 Regional Medical Physics Department, Freeman Hospital, Freeman Road, Newcastle

More information

The activity. Suggested Answers

The activity. Suggested Answers Teacher notes Introduction This activity encourages students to consider the principles of radiation protection and how they might apply in the nuclear industry and in medicine. They discuss and answer

More information

Quality Control of DXA System and Precision Test of Radio-technologists

Quality Control of DXA System and Precision Test of Radio-technologists J Bone Metab 2014;21:2-7 http://dx.doi.org/10.11005/jbm.2014.21.1.2 pissn 2287-6375 eissn 2287-7029 Review Article Quality Control of DXA System and Precision Test of Radio-technologists Ho-Sung Kim 1,

More information

PRINCIPLES AND METHODS OF RADIATION PROTECTION

PRINCIPLES AND METHODS OF RADIATION PROTECTION PRINCIPLES AND METHODS OF RADIATION PROTECTION Lesson Outcomes At the end of the lesson, student should be able to: Define what is radiation protection (RP) Describe basic principles of RP Explain methods

More information

Radiation Safety Information for Students in Courses given by the Nuclear Physics Group at KTH, Stockholm, Sweden

Radiation Safety Information for Students in Courses given by the Nuclear Physics Group at KTH, Stockholm, Sweden Radiation Safety Information for Students in Courses given by the Nuclear Physics Group at KTH, Stockholm, Sweden September 2006 The aim of this text is to explain some of the basic quantities and units

More information

Lunar idxa. The intelligent DXA. gehealthcare.com

Lunar idxa. The intelligent DXA. gehealthcare.com Lunar idxa The intelligent DXA gehealthcare.com The best of DXA technology for bone and metabolic health assessment With a state-of-the-art design, Lunar idxa offers research-grade image resolution and

More information

Survey of Radiation Dose Levels in Patients in X-Ray Units of Some Selected Hospitals in Jos Metropolis

Survey of Radiation Dose Levels in Patients in X-Ray Units of Some Selected Hospitals in Jos Metropolis International Journal of Innovative Scientific & Engineering Technologies Research 6(4):1-9, Oct.-.Dec., 2018 SEAHI PUBLICATIONS, 2018 www.seahipaj.org ISSN: 2360-896X Survey of Radiation Dose Levels in

More information

RADIATION SAFETY. Junior Radiology Course

RADIATION SAFETY. Junior Radiology Course RADIATION SAFETY Junior Radiology Course Expectations for the Junior Radiology Course Medical School wants students to learn basic principles, factual knowledge, safety info, etc. Medical Students want

More information

Standard Operating Procedure TCRC Dual-Energy X-ray Absorptiometry (DXA)

Standard Operating Procedure TCRC Dual-Energy X-ray Absorptiometry (DXA) 1. RELEVANCE a. This SOP outlines the instructions to completing Duel Energy X-Ray Absorptiometry (DXA) including: scanning, analysis, review and filing. 2. SCOPE a. This SOP applies to all TCRC RDs. 3.

More information

2013 ISCD Combined Official Positions

2013 ISCD Combined Official Positions 2013 ISCD Combined Oicial Positions Oicial Positions of the International Society for Clinical Densitometry The International Society for Clinical Densitometry (ISCD) is a not-for-profit multidisciplinary

More information

Radiation Protection- Cath lab

Radiation Protection- Cath lab Radiation Protection- Cath lab Dr. Mawya A Khafaji Associate Prof. Medical Physics, Faculty of Medicine, KAU Head of Medical Physics Unit Dept. of Radiology -KAUH Head, Volunteer Office -KAUH Outline:

More information

Technical Principles of Dual Energy X-Ray Absorptiometry

Technical Principles of Dual Energy X-Ray Absorptiometry Technical Principles of Dual Energy X-Ray Absorptiometry Glen M. Blake and Ignac Fogelman Since its introduction nearly ten years ago, dualenergy x-ray absorptiometry (DXA) has become the single most widely

More information

BONE MINERAL DENSITY OF BANGLADESHI PEOPLE

BONE MINERAL DENSITY OF BANGLADESHI PEOPLE 6 DAFFODIL INTERNATIONAL UNIVERSITY JOURNAL OF SCIENCE BANU: BONE AND MINERAL TECHNOLOGY, DENSITY VOLUME OF BANGLADESHI 4, ISSUE 2, JULY PEOPLE 2009 BONE MINERAL DENSITY OF BANGLADESHI PEOPLE Dilruba Akhter

More information

L.W. Sun 1,2, G. Beller 1, D. Felsenberg 1. Introduction. Original Article. Abstract

L.W. Sun 1,2, G. Beller 1, D. Felsenberg 1. Introduction. Original Article. Abstract J Musculoskelet Neuronal Interact 2009; 9(1):18-24 Original Article Hylonome Quantification of bone mineral density precision according to repositioning errors in peripheral quantitative computed tomography

More information

The x-rays produced penetrate the body which absorbs, refracts, or reflects the x-ray beam energy depending on the tissue. Bone

The x-rays produced penetrate the body which absorbs, refracts, or reflects the x-ray beam energy depending on the tissue. Bone Authors Sari Cohen, Poh Yan Lim, Merng Koon Wong, Siew Hong Lau, Donna Russell-Larson 1.6.2 Image intensifier Poh Yan Lim, Merng Koon Wong The discovery of x-rays had a profound impact on the diagnosis

More information

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result. Minimizing Dose Understand the importance and methods of pre-procedure patient assessment including a review of previous radiologic exams, disease processes and anatomical considerations that may increase

More information

Introducing the future of DXA. Powerful images. Clear answers. Horizon DXA System

Introducing the future of DXA. Powerful images. Clear answers. Horizon DXA System Introducing the future of DXA Powerful images. Clear answers. Horizon DXA System Hologic turns ideas into innovation. Again. Hologic cares about you and your patients about keeping their bones healthy,

More information

Interpreting DEXA Scan and. the New Fracture Risk. Assessment. Algorithm

Interpreting DEXA Scan and. the New Fracture Risk. Assessment. Algorithm Interpreting DEXA Scan and the New Fracture Risk Assessment Algorithm Prof. Samir Elbadawy *Osteoporosis affect 30%-40% of women in western countries and almost 15% of men after the age of 50 years. Osteoporosis

More information

A Standardised Approach to Optimisation

A Standardised Approach to Optimisation The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation exposure of the Yazd population from medical conventional X-ray examinations Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status Journal of Clinical Densitometry, vol. 10, no. 3, 239e243, 2007 Ó Copyright 2007 by The International Society for Clinical Densitometry 1094-6950/07/10:239e243/$32.00 DOI: 10.1016/j.jocd.2007.03.002 Original

More information

Quality Assurance and Control in Dual energy X ray Absorptiometry

Quality Assurance and Control in Dual energy X ray Absorptiometry Mihail A. Boyanov Quality Assurance and Control in Dual energy X ray Absorptiometry Sofia, 2013 Central Medical Library Quality Assurance and Control in Dual energy X ray Absorptiometry Mihail A. Boyanov,

More information

Use of DXA / Bone Density in the Care of Your Patients. Brenda Lee Holbert, M.D. Associate Professor Senior Staff Radiologist

Use of DXA / Bone Density in the Care of Your Patients. Brenda Lee Holbert, M.D. Associate Professor Senior Staff Radiologist Use of DXA / Bone Density in the Care of Your Patients Brenda Lee Holbert, M.D. Associate Professor Senior Staff Radiologist Important Websites Resources for Clinicians and Patients www.nof.org www.iofbonehealth.org

More information

X-Ray Board Reviewing Rules

X-Ray Board Reviewing Rules Volume 24, Issue 3 July-Sept 2018 Tennessee X-Ray X-Ray Board Reviewing Rules The Tennessee Board of Radiologic Imaging and Radiation Therapy had their first meeting on January 25, 2018. The minutes from

More information

Nutritional Aspects of Osteoporosis Care and Treatment Cynthia Smith, FNP-BC, RN, MSN, CCD Pars Osteoporosis Clinic, Belpre, Ohio

Nutritional Aspects of Osteoporosis Care and Treatment Cynthia Smith, FNP-BC, RN, MSN, CCD Pars Osteoporosis Clinic, Belpre, Ohio Osteoporosis 1 Nutritional Aspects of Osteoporosis Care and Treatment Cynthia Smith, FNP-BC, RN, MSN, CCD Pars Osteoporosis Clinic, Belpre, Ohio 1) Objectives: a) To understand bone growth and development

More information

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry Code of Practice for Radiation Protection in Dentistry Code of Practice For Radiation Protection in Dentistry 10 OCTOBER 2017 CONTENTS 1. INTRODUCTION... 3 1.0 CITATION... 3 1.1 BACKGROUND... 3 1.2 PURPOSE

More information

Estimating the Absorbed Dose to Critical Organs During Dual X-ray Absorptiometry

Estimating the Absorbed Dose to Critical Organs During Dual X-ray Absorptiometry Estimating the Absorbed Dose to Critical Organs During Dual X-ray Absorptiometry M Mokhtari-Dizaji, PhD 1 A A Sharafi, PhD 2 B Larijani, MD 3 N Mokhlesian, MSc 4 H Hasanzadeh, MS 1 Index terms: Dual X-ray

More information

Radiation Safety Manual

Radiation Safety Manual King Abdulaziz University Faculty of Dentistry Radiation Safety Manual FOR X-RAY EQUIPMENT OPERATORS October 2009 Radioactivity and Radiation All matter in our environment is made of atoms. Most atoms

More information

Radiation Safety. Bethany Gillett 14th Feb After this lecture, you should be able to:

Radiation Safety. Bethany Gillett 14th Feb After this lecture, you should be able to: Radiation Safety Bethany Gillett bethany.gillett@addenbrookes.nhs.uk 14th Feb 2018 Learning Outcomes After this lecture, you should be able to: Understand different radiation protection quantities Explain

More information

ICRP = International Commission on. recommendations and guidance on. Functioning since 1928.

ICRP = International Commission on. recommendations and guidance on. Functioning since 1928. ICRP = International Commission on Radiological Protection; An advisory body providing recommendations and guidance on radiation protection; Functioning since 1928. While the use of ionising radiation

More information

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua

Radiation Safety For Anesthesiologists. R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua Radiation Safety For Anesthesiologists R2 Pinyada Pisutchareonpong R2 Nawaporn Sateantantikul Supervised by Aj Chaowanan Khamtuicrua Modern World Non Ionizing VS Ionizing Non Ionizing Harmless Ex. visible

More information

Page 1 of 5 Patient Safety: Radiation Dose in X-Ray and CT Exams What are x-rays and what do they do? X-rays are forms of radiant energy, like light or radio waves. Unlike light, x-rays can penetrate the

More information

The DXL Calscan heel densitometer: evaluation and diagnostic thresholds

The DXL Calscan heel densitometer: evaluation and diagnostic thresholds The British Journal of Radiology, 79 (2006), 336 341 The DXL Calscan heel densitometer: evaluation and diagnostic thresholds J A THORPE, MSc, BSc and S A STEEL, MSc, BSc Centre for Metabolic Bone Disease,

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

A basic study for automatic recognition of osteoporosis using abdominal X-ray CT images

A basic study for automatic recognition of osteoporosis using abdominal X-ray CT images A basic study for automatic recognition of osteoporosis using abdominal X-ray CT images Sadamitsu Nishihara* a, Hiroshi Fujita b, Tadayuki Iida a, Atsushi Takigawa a, Takeshi Hara b and Xiangrong Zhou

More information

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Silvia Capuani*, Giulia Di Pietro*, Guglielmo Manenti, Umberto Tarantino^

Silvia Capuani*, Giulia Di Pietro*, Guglielmo Manenti, Umberto Tarantino^ Silvia Capuani*, Giulia Di Pietro*, Guglielmo Manenti, Umberto Tarantino^ *CNR ISC, Physics Department Sapienza University, Rome, Italy Department of Diagnostic Imaging and Interventional Radiology, Molecular

More information

chapter Bone Density (Densitometry) RADIOPHARMACY INDICATIONS Radionuclide Localization Quality Control Adult Dose Range Method of Administration

chapter Bone Density (Densitometry) RADIOPHARMACY INDICATIONS Radionuclide Localization Quality Control Adult Dose Range Method of Administration 10766-04_CH04_redo.qxd 12/3/07 3:47 PM Page 17 chapter 4 Bone Density (Densitometry) RADIOPHARMACY Radionuclide Single radionuclide: 125 I t 1/2 : 60.1 days Energies: 23 31 kev Type: EC, x, γ, accelerator

More information

Preliminary technical data

Preliminary technical data Preliminary technical data 2D-FanBeam Whole Body Densitometer PROGRESS THROUGH INNOVATION 2D-Fan Beam Whole Body Densitometer Acquisition chain parameters Dual Emission X-ray Absorptiometry (DEXA) 2D-FanBeam

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection. John Harrison ICRP Committee 2

Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection. John Harrison ICRP Committee 2 Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection John Harrison ICRP Committee 2 Joint ICRP-RERF-JHPS Workshop: Tokyo, December 2017 Task Group 79 : Use of Effective Dose

More information

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Rebecca Milman Marsh, Ph.D. University of Colorado Department of Radiology Who in the Facility Works

More information

IONISING RADIATION REGULATIONS 99

IONISING RADIATION REGULATIONS 99 IONISING RADIATION REGULATIONS 99 & IRMER IONISING RADIATION MEDICAL EXPOSURE REGULATIONS BARBARA LAMB Specialist Radiographer Dental and maxillofacial radiography BarbaraHLamb@googlemail.com 07775994424

More information

This is a repository copy of Microarchitecture of bone predicts fractures in older women.

This is a repository copy of Microarchitecture of bone predicts fractures in older women. This is a repository copy of Microarchitecture of bone predicts fractures in older women. White Rose Research Online URL for this paper: http://eprints.whiterose.ac.uk/130351/ Version: Accepted Version

More information

Quantitative Computed Tomography 4 Introduction

Quantitative Computed Tomography 4 Introduction Quantitative Computed Tomography 4 Introduction Quantitative Computed Tomography (QCT) is a well recognised technique for the measurement of bone mineral density (BMD) in the lumbar spine1 and forearm2.

More information

Omnisense: At Least As Good As DXA

Omnisense: At Least As Good As DXA Omnisense: At Least As Good As DXA The following document summarizes a series of clinical studies that have been conducted to compare between different qualities of the Sunlight support the claim that

More information

ICRP RECOMMENDATIONS AND IAEA SAFETY STANDARDS: THEN AND NOW

ICRP RECOMMENDATIONS AND IAEA SAFETY STANDARDS: THEN AND NOW ICRP RECOMMENDATIONS AND IAEA SAFETY STANDARDS: THEN AND NOW Abdalla N. Al-Haj, PhD, FIPEM, CRadP, CSci, MSRP Chief Health Physicist King Faisal Specialist Hospital & Research Centre Riyadh, Saudi Arabia

More information

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER Bro. Dr. Collie Miller IARC/WHO Based on trends in the incidence of cancer, the International Agency for Research on Cancer (IARC) and WHO

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

2005 RECOMMENDATIONS OF ICRP

2005 RECOMMENDATIONS OF ICRP IRPA 11 11 th International Congress of the International Radiation Protection Association 23 28 May 2004, Madrid, Spain 2005 RECOMMENDATIONS OF ICRP ROGER H CLARKE CHAIRMAN FEATURES OF RECOMMENDATIONS

More information

Building Bone Density-Research Issues

Building Bone Density-Research Issues Building Bone Density-Research Issues Helping to Regain Bone Density QUESTION 1 What are the symptoms of Osteoporosis? Who is at risk? Symptoms Bone Fractures Osteoporosis 1,500,000 fractures a year Kyphosis

More information

Prodigy. from GE Healthcare. Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world. gehealthcare.

Prodigy. from GE Healthcare. Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world. gehealthcare. Prodigy from GE Healthcare Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world gehealthcare.com Prodigy High performance, efficient and reliable DXA system

More information

Live Educational Programs

Live Educational Programs Live Educational Programs 2014 Osteoporosis: Essentials of Densitometry, Diagnosis and Management - for Clinicians Osteoporosis: Essentials of Densitometry, Diagnosis and Management - for Technologists

More information

Thickness Computation Under In-Vivo Trabecular Bone CT Imaging

Thickness Computation Under In-Vivo Trabecular Bone CT Imaging Thickness Computation Under In-Vivo Trabecular Bone CT Imaging Gokul. S PG Scholar,M.E - II year Department of Computer Science and Engineering Jansons Institute of Technology Coimbatore, Tamilnadu Aarthi.K

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

DEXA Scores and Bone Density Measured on Routine CT Scans

DEXA Scores and Bone Density Measured on Routine CT Scans DEXA Scores and Bone Density Measured on Routine CT Scans Poster No.: R-0072 Congress: 2015 ASM Type: Scientific Exhibit Authors: C. Annabattula, P. Phadke; Sydney/AU Keywords: Screening, Nuclear medicine

More information

The effect of vertebral rotation of the lumbar spine on dual energy X-ray absorptiometry measurements: observational study

The effect of vertebral rotation of the lumbar spine on dual energy X-ray absorptiometry measurements: observational study JCY Cheng HL Sher X Guo VWY Hung AYK Cheung Key words: Absorptiometry, Bone density; Densitometry, X-ray; Lumbar vertebrae; Scoliosis "#$%& "# X HKMJ 2001;7:241-5 The Chinese University of Hong Kong, Prince

More information