IAEA TECDOC 1517 Quality Control in Mammography Software

Size: px
Start display at page:

Download "IAEA TECDOC 1517 Quality Control in Mammography Software"

Transcription

1 IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua. b Centro de Investigación en Ciencias Atómicas, Nucleares y Moleculares, Universidad de Costa Rica, Costa Rica. c Facultad de Medicina, Universidad Complutense, Madrid, España. d Radiological Protection of Patients Unit, Department of Nuclear Safety and Security, IAEA. Abstract. In October 2006, the IAEA published the TECDOC 1517 Quality Control in Mammography, whose main purpose was to give Latin American countries a protocol in Spanish with all necessary QC. This protocol harmonizes the tests and evaluation criteria of mammography equipment and its complementary equipment; it states the different responsibilities on all personnel and gives guidance on radiographic techniques. It was the joint effort of two ARCAL projects: RLA/6/043 and RLA/9/035. QC programs are needed to assure the final quality of mammography images and to optimize the radiation dose to the patients. Countries where national campaigns are used to improve the early detection of breast cancer among asymptomatic women require the establishment of QC programs. Specific software has been developed based on the TECDOC 1517 to facilitate the technologist, medical physicist and physician its implementation. It has a main menu bar tool and icons for rapid access to the different tests. The help option in each test pops a window with the same procedure written on the TECDOC for the user s convenience. The tests are divided on the same sections as in the document: visual inspection, storage of films, dark room, image system, radiological equipment, automatic exposure control, geometry, collimation, image visualization, film rejection analysis, image quality and dosimetry. On each test, data is introduced on specific color cells and when the user activates the calculation button, the results are compared against its tolerance levels, and indication of pass/fail is finally displayed. This software, available to all Member States, adds extra value to the TECDOC 1517 since errors in calculations will be reduced by its use. It will harmonize the way results are presented, it will facilitate comparisons, it reduces the time to evaluate the results of the test and finally it becomes a teaching tool for the TECDOC. KEYWORDS: mammography, quality control, software, TECDOC Introduction Breast cancer is the most common cancer among women worldwide and is a leading cause of cancer mortality in women. Breast cancer incidence increased per cent from the 1970s to the 1990s in most countries, with the most marked increases among women aged fifty years and older, although the incidence for women under fifty is increasing. Mammography is an X ray examination of the breast. Its principal purpose is to facilitate the detection of breast cancer at a point earlier in its natural history than possible by clinical examination. It has been demonstrated that routine screening with high quality mammography is effective in reducing mortality from breast cancer in women aged [1-2] One of the key elements to obtain a high quality mammogram is the establishment of a Quality Control (QC) program.[3-4] In 2006, the International Atomic Energy Agency (IAEA) published the IAEA TECDOC-1517 Quality Control in Mammography.[5] This document in Spanish was the joint effort of two ARCAL projects: RLA/6/043 and RLA/9/035; with the participation of the following countries: Argentina, Bolivia, Brazil, Chile, Colombia, Costa Rica, Cuba, Guatemala, El Salvador, Mexico, Nicaragua, Panamá, Paraguay, Perú, República Dominicana and Venezuela. This protocol harmonizes the tests and evaluation criteria of mammography equipment and its complementary equipment such as the processor; it states the different responsibilities on all personnel on a mammography facility and gives guidance on radiographic techniques. It is aimed to help optimize radiation doses in mammography [6] and provide a teaching material for the different personnel. patricia.mora@ucr.ac.cr

2 The software was developed for the Latin American mammography community to reduce calculation errors, harmonize the way results are presented, facilitate comparison among national professionals and with colleagues of other countries, reduce the time necessary to evaluate the test results and finally to become a teaching tool for the TECDOC General overview of the program The Principal Screen has a standard menu bar with a list of the entire QC tests to perform; see Figure 1. It also has a second bar with icons for easy access to each test. There are a total of 29 windows for each of the tests described in the IAEA TECDOC In each window all mathematical formulas have been introduced and are automatically calculated for easy comparison of the result with the tolerance value of the test. Each window can be saved or printed. Figure 1: Principal Screen Four different colors for easy identification of the purpose of the cells were used in all windows; see Figure 2. Light gray is used for fixed cells or cells displaying tolerance values. White cells are used by the user to introduce data to the software (either alphabetic or numerical values). Dark gray cells are used to perform calculations by the software and their results are showed on yellow cells. Figure 2: Use of different colors for cells in the program 2

3 During the execution of a window, the program can show 4 sub windows for: additional information, confirmation of an action, error message or warning message. The user moves through a window with the TAB button or positioning the mouse on a specific cell. Menu Files are managed as in other Windows based programs with icons for new, open, save, print and exit options. Once the Institution name has been added on the first window it will appear on all pages of the report. A general report can be printed with all the windows or individual prints can be done for each test. On Figure 3 an example of one of the reports generated is presented. Figure 3: Visual inspection of the x-ray mammography unit report. On the main tool bar a Help option will give the user a short manual with instructions on installation, programs designed and a general description on all QC tests. Also available for convenience of the user, on each window a Help icon will bring the text of the specific test as presented on the IAEA TECDOC Figure 4 shows an example of the help window. 3

4 Figure 4: Help window 3. Description of the windows Institution data: on this first window the user is asked to introduce the information on the x-ray mammography equipment, processor and film used at the facility being evaluated, also, information on the person doing the tests. Visual inspection: a simple table with YES/NO entries must be completed; it can be seen on Figure 3. Storage of films: an evaluation on the storage conditions of the films is done. Dark room evaluation: this window evaluates the general conditions of the darkroom, (such as temperature and humidity), light leakage and fog due to safety lights. Processor: ph of developer and fixer, processing time, replenishment rate and artifacts are evaluated with this window. Sensitometry: three different windows are used for sensitometric purposes. The first one establishes the initial operation levels (IOL) for the base + fog, low density, mid density and high density. The second window is the daily sensitometric test. As data is added everyday, a graphic representation of base + fog, mid density and density difference is displayed; as shown on Figure 5. Finally on a separate window a calculation is performed when a transition between boxes is done. 4

5 Figure 5: Sensitometric data window with associated graphics Image system: it contains the test for screen cleanness and screen-film contact. Screen tightness and uniformity: for all used cassettes spots are looked for and if found, recorded on the window. Also values of mas and OD are recorded for each screen to evaluate their uniformity among them. The following tests due to their recommended frequency and required equipment will be performed by the medical physicist (except those related to the mammographic phantom and film rejection rate). When an ionization chamber has to be used all windows will ask if the chamber corrects for temperature and pressure; if not, introduction of temperature and pressure is necessary to have the software calculate the correction factor. It is important to be aware of the units for different magnitudes (for example, µgy or mgy depending on the test). Also many cells have a choice of values to choose; for example, anode/filter combination, compression (yes/no), etc. Careful understanding of the following tests is necessary in order to use the software correctly. Radiation leakage: Using the measured air kerma rates at the points where leakage has been found, the software will extrapolate a value in msv/hr to compare to the regulatory 1 msv/hr at 1 meter value. kvp: this window will evaluate the accuracy of 7 nominal values of kvp and its repeatability at 28 kvp and Mo/Mo combination. Compliance with the tolerance values will be done when the button Conform is pressed. An example of this test is shown in Figure 6. 5

6 Figure 6: Kilovoltage evaluation test window Half value layer: establishment of the HVL will be done for reference conditions (e.g. 28 kvp and Mo/Mo) and for 3 clinical conditions set by the user based on the institution preference. For each filter condition 3 readings have to be entered. Output: again the output will be calculated for the reference conditions (e.g. 28 kvp and Mo/Mo) and for 2 user conditions. For 4 values of mas, 5 readings are needed to calculate the output, its linearity and repeatability. Time: for those units where the time is displayed after the exposure, the accuracy and repeatability will be calculated. Compression force: in this window the user has to indicate the units of the force (N, Lb or kg) and record the compression force for manual and automatic modes. Automatic Exposure Control (AEC): this window evaluates the repeatability of the AEC and its compensation for different thickness of the breast and beam quality. Density control setting evaluation: the user will obtain images for all the density control settings and comparison of the OD and mas values with the tolerance values will be established. Collimation system: this window contains several tests, first it verifies that the SID is > 60 cm, then a general overview of the collimation system is done by evaluation of coincidence between the radiation field and the image receptor; the coincidence between the light field and the radiation field and finally the alignment of the compression paddle with the margin of the breast support. View boxes: a general visual inspection of view boxes is done, followed by luminance measurements at 9 points in each view box. Then if more than one view box is used a comparison between them is carried out based on the central luminance of each panel. Room ambient illumination is also evaluated. In Figure 7 an example of the evaluation of view boxes and room ambient illumination window is shown. 6

7 Figure 7: View Box and Room ambient illumination window Film rejection rate: a simple evaluation chart for film rejection analysis is presented to be filled out. Common factors for film rejection are sorted by technique factors (e.g. movement, wrong marking, etc.) and darkroom and processor factors (e.g. fog). Image: this window contains the phantom image evaluation (using the known scoring system of fibers, masses and microcalcification groups) and the spatial resolution of the system. Incident air kerma and Mean Glandular Dose: there are 2 equal windows for this test, one for the reference condition and one for clinical conditions. The user has to pick if the x-ray equipment shows the mas after the exposure in order for the software to use the appropriate formulas. The formulas take into account the c DG50,Ki,PMMA factor that converts the entrance air kerma to the mean glandular dose for a 45 mm thick standard phantom and the corresponding s factor which gives a correction that depends on the target/filter combination. [7-8] Figure 8 shows an example of this window. 7

8 Figure 8: Window for Incident Kerma and Mean glandular dose calculation 4. Conclusions Most of the quality control protocols include data form sheets to illustrate how data must be taken by the technologist, medical physicist or radiologist. However, the lack of a complete definition or a glossary about the magnitudes linked to their tolerance values give rise to a wide spread of results. [9-10] This fact causes misunderstandings when the results on the applicability of the protocols are compared. This software has the main objective of harmonizing results which is of special importance for several reasons: First, mammography is a very compromising radiological examination since it is used to detect early breast cancer. As it is well known, the diagnostic signs linked to its early detection rely on fine details whose visualization imposes important requirements (for example high contrast and resolution, and low noise) to the whole mammographic imaging chain. On the other hand, early breast cancer programs are directed to asymptomatic women, where the benefit could be overcome by the risk of delivering high doses to the women. Therefore, image quality and dose values have to be strictly verified and optimized by applying the adequate quality control procedures. [11] Second, it will facilitate comparison among national professionals and with colleagues of other countries. These comparisons have been demonstrated to be an important element in the optimization process of the mammographic image acquisition technique. Third, in those countries where there is no tradition in the implementation of the quality control programs, the comparison of results can help to detect failure on the application of the protocol. In this way, the software can also be understood as a helpful tool to learn about quality control. 8

9 Finally, it reduces the time to evaluate the results of the numerous tests on the TECDOC This is very important since the number of medical physicists or technologists in Latin America in diagnostic medical physics involved in quality control tasks are usually very low or even inexistent. In consequence, when these professionals are engaged in QC programs, they are in charge of a high number of different radiological equipment. REFERENCES [1] WORLD HEALTH ORGANIZATION, IARC Handbooks of Cancer Prevention, Vol 7: Breast Cancer Screening, (INTERNATIONAL AGENCY FOR RESEARCH ON CANCER ed.), IARC Press, Lyons (2002). [2] TABAR, L., et al., Mammography service screening and mortality in breast cancer patients: 20 year follow up before and after introduction of screening, Lancet 361 (2003) [3] AMERICAN COLLEGE OF RADIOLOGY, Mammography Quality Control Manual, ACR, New York (1999). [4] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Equipment requirements and quality control for mammography, Report No. 29, American Institute of Physics, New York (1990). [5] INTERNATIONAL ATOMIC ENERGY AGENCY, TEC DOC 1517 Control de Calidad en Mamografía, Viena (2006). [6] INTERNATIONAL ATOMIC ENERGY AGENCY, International Basic Safety Standards for Protection Against Ionizing Radiation and for the Safety of Radiation Sources, Safety Series No. 115, Vienna (1996). [7] DANCE, D.R., Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose, Phys. Med. Biol., 35 (1990) [8] EUROPEAN COMMISSION, European Protocol on Dosimetry in Mammography, CEC- Report EUR 16263, Office for Official Publications of the European Communities, Luxembourg (1996). [9] EUROPEAN COMMISSION, European Guidelines for Quality Assurance in Mammography Screening, 3 rd Edition, Europe Against Cancer, Office for Official Publications of the European Communities, Luxembourg (2001). [10] EUROPEAN COMMISSION, European Guidelines for Quality Assurance in Breast Cancer screening and diagnosis. 4 th Edition, European Communities, Belgium (2006). [11] SOCIEDAD ESPAÑOLA DE FÍSICA MÉDICA-SOCIEDAD ESPAÑOLA DE PROTECCIÓN RADIOLÓGICA, Protocolo Español de Control de Calidad en Radiodiagnóstico (Aspectos Técnicos), Edicomplet, Madrid (1996). 9

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information

Hong Kong College of Radiologists Mammography Statement

Hong Kong College of Radiologists Mammography Statement Hong Kong College of Radiologists Mammography Statement The Hong Kong College of Radiologists would like to give the following comments concerning mammography. Mammography screening: Breast cancer is the

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420, Mail Code 25-01

More information

Mammography Quality Control: A Refresher

Mammography Quality Control: A Refresher Mammography Quality Control: A Refresher MATT WAIT, MS, DABR Objectives Attendees will re-familiarize themselves with the purpose of quality assurance and quality control Attendees will re-familiarize

More information

Dosimetry in digital mammography

Dosimetry in digital mammography Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol

More information

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis 60 Journal of Radiology and Diagnostic Imaging, 2013, 1, 60-64 Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

More information

Patient Dosimetry in Mammography and Tomosynthesis:

Patient Dosimetry in Mammography and Tomosynthesis: 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor

More information

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi

More information

Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE

Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE ACR Update on FFDM Accreditation Overview New ACR Activities Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO QC Today QC Tomorrow *No financial disclosures to report

More information

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program An De Hauwere, Hubert Thierens Ghent University, Department of Medical Physics,

More information

Survey of Mammography Practice: Initial Results from Serbia

Survey of Mammography Practice: Initial Results from Serbia Survey of Mammography Practice: Initial Results from Serbia Olivera Ciraj Bjelac *, Dusko Kosutic, Danijela Arandjic, Milojko Kovacevic Vinca Institute of Nuclear Sciences, Radiation and Environmental

More information

Update on the ACR FFDM QC Manual

Update on the ACR FFDM QC Manual Update on the ACR FFDM QC Manual Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO AAPM 2011 Vancouver, Canada 9000 8500 8000 7500 7000 6500 6000 5500 5000 4500 4000

More information

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2

More information

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics

More information

1. Introduction. 2. Technical procedures

1. Introduction. 2. Technical procedures MAMMOGRAPHY QUALITY ASSURANCE IN MOROCCO i inn mill IT -7 v T i * ^ T i XA0101632 F. Zaoui, K. Talsmat, K. Lalaoui Departement des Applications Medicales et Biologiques, Centre National del'energie des

More information

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY

More information

Implementation of a New Tomosynthesis Program: A Physicists Perspective

Implementation of a New Tomosynthesis Program: A Physicists Perspective Implementation of a New Tomosynthesis Program: A Physicists Perspective Bill Geiser, MS DABR Senior Medical Physicist wgeiser@mdanderson.org 1 Conflict of Interest None of the authors nor their immediate

More information

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation?

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation? The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Revised: September 7, 2017; updated questions in red) Table of Contents Application - General...

More information

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements Standards Revisions for Organizations Providing Fluoroscopy Services The Joint Commission has approved the following revisions for prepublication. While revised requirements

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

Direct half value layer measurements in mammography - is near enough good enough?

Direct half value layer measurements in mammography - is near enough good enough? Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview New ACR Activities Requirements

More information

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,

More information

IAEA activities in Medical Physics

IAEA activities in Medical Physics IAEA activities in Medical Physics Giorgia Loreti Training Officer (Medical Physics) G.Loreti@iaea.org IAEA - International Atomic Energy Agency Department of Nuclear Sciences and Applications Division

More information

Breast phantom with silicone implant for evaluation in conventional mammography

Breast phantom with silicone implant for evaluation in conventional mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, winter 2011 Breast phantom with silicone implant for evaluation in conventional mammography Fábio A. R. Silva, 1a Luíza F Souza, 1 Carlos

More information

Senior Management Team Meeting Target setting in Latin America

Senior Management Team Meeting Target setting in Latin America Senior Management Team Meeting Target setting in Latin America October 2006 Nancy Andrade-Castro RST Latin America Geneva/2006 For all enquiries, please contact Nancy Andrade-Castro, Acting Director for

More information

The New Jersey Radiographic Quality Assurance Program at 5 Years

The New Jersey Radiographic Quality Assurance Program at 5 Years The New Jersey Radiographic Quality Assurance Program at 5 Years Julie Timins, MD a, Paul Orlando, BS b, Jill Lipoti, PhD b Purpose: Five years ago, the New Jersey Bureau of Radiological Health decided

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

AAPM Annual Meeting. ACR Accreditation Update in CT

AAPM Annual Meeting. ACR Accreditation Update in CT AAPM Annual Meeting ACR Accreditation Updates in CT, Ultrasound, Mammography and MRI: ACR Accreditation Update in CT Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiological Sciences

More information

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose. Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives

More information

The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red)

The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red) The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red) Table of Contents General... 1 Applicability... 3 Transitioning

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system Original article NOWOTWORY Journal of Oncology 2017, volume 67, number 2, 115 120 DOI: 10.5603/NJO.2017.0018 Polskie Towarzystwo Onkologiczne ISSN 0029 540X www.nowotwory.edu.pl Mean glandular dose values

More information

A tissue-equivalent phantom series for mammography dosimetry

A tissue-equivalent phantom series for mammography dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOL. 5, NO. 4, FALL 2004 A tissue-equivalent phantom series for mammography dosimetry William P. Argo, 1 Kathleen Hintenlang, 2 and David E. Hintenlang 3 U.S.

More information

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro 3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:

More information

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA Facility Certification Extension Requirements, Quality Assurance and Medical Physicists role for Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT) System Kish chakrabarti, Ph.D. Senior Physicist

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

Dental Extraoral X-ray Systems

Dental Extraoral X-ray Systems Dental Extraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL EXTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability. A registrant s x-ray system used for dental

More information

Update on MQSA and Mammography Accreditation

Update on MQSA and Mammography Accreditation MQSA - Who s Who Update on MQSA and Mammography Accreditation The Law: Mammography Quality Standards Act (MQSA) The Regulator: US Food and Drug Administration (FDA) Priscilla F. Butler, M.S. Senior Director,

More information

Quality control in mammography: an initiative in France

Quality control in mammography: an initiative in France f 1994, The British Journal of Radiology, 67, 371-383 Quality control in mammography: an initiative in France 1 C MACCIA, PhD, 2 R RENAUD, MD, 1 S CASTELLANO, PhD, 2 P SCHAFFER, MD, PhD, 2 R WAHL, MD,

More information

Mammography calibration qualities establishment in a Mo- Mo clinical system

Mammography calibration qualities establishment in a Mo- Mo clinical system Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733

More information

Current Status in Latin America

Current Status in Latin America WHO International Consultation 15 to 17 October 2014 Munich, Germany JUSTIFICATION OF THE USE OF CT FOR INDIVIDUAL HEALTH ASSESSMENT OF ASYMPTOMATIC PEOPLE Current Status in Latin America Gloria Soto Giordani

More information

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems Dental ConeBeam Computed Tomography (CBCT) X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL CONEBEAM COMPUTED TOMOGRAPHY (CBCT) X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability.

More information

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Radiation Dosimetry in Digital Breast Tomosynthesis March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Syllabus 1. Introduction 2. Dosimetry in Mammography 3. Dosimetry

More information

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Caspian Journal of Health Research 2016; 2(2):18-27 Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Hossein Rahanjam 1, Hossein

More information

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.

More information

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in

More information

-Ray Producing Equipment

-Ray Producing Equipment Temple University R adiation Safety Guide For Diagnostic Imaging X -Ray Producing Equipment Environmental Health And Safety Radiation Safety Department Table of Content Page 1. Radiation 2 A. Radiation

More information

INTER-AMERICAN UNIVERSITY OF PUERTO RICO

INTER-AMERICAN UNIVERSITY OF PUERTO RICO Introduction INTER-AMERICAN UNIVERSITY OF PUERTO RICO POLICY FOR DIRECT AND INDIRECT SUPERVISION The Radiologic Protection is an essential measure to be followed in a Radiology Department, especially during

More information

IAEA-TECDOC-796. Radiation doses in diagnostic radiology

IAEA-TECDOC-796. Radiation doses in diagnostic radiology IAEA-TECDOC-796 Radiation doses in diagnostic radiology The IAEA does The originating Section FOREWORD It is well recognized that diagnostic radiology is the largest contributor to the collective dose

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Effects of Quality Assurance Regulatory Enforcement on Performance of Mammography Systems: Evidence From Large- Scale Surveys in Taiwan

Effects of Quality Assurance Regulatory Enforcement on Performance of Mammography Systems: Evidence From Large- Scale Surveys in Taiwan Quality Assurance in Mammography Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Yi-Shuan Hwang 1 Hui-Yu Tsai 2 Chien-Chuan Chen 1 Pei-Kwei Tsay 3 Huay-Ben

More information

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY PNRA-RG-904.05 (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY For Further Details Directorate of Policies & Procedures PAKISTAN

More information

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY March 2010 EAST AFRICAN MEDICAL JOURNAL 127 East African Medical Journal Vol. 87 No. 3 March 2010 OPTIMIsATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY G. K. Korir, BSc, MSc, PhD

More information

Analysis of the results from the implementation of quality control program in mammography in Bulgaria

Analysis of the results from the implementation of quality control program in mammography in Bulgaria Analysis of the results from the implementation of quality control program in mammography in Bulgaria *, J. Vassileva, A. Dimov National Centre of Radiobiology and Radiation Protection, Sofia, Bulgaria

More information

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,

More information

Dental Intraoral X-ray Systems

Dental Intraoral X-ray Systems Dental Intraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 2.0 4732.#### DENTAL INTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Commented [JC(1]: Based on part 4732.0880. Subpart 1. Applicability.

More information

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences 1313 C. Kimme-Smith1 J. Wang2 N. DeBruhl1 M. Basic2 L. W. Bassett1 Received October 27, 1 993; accepted after revision January 25, 1994. Presented at the annual meeting of the American Roentgen Ray Society,

More information

Mdiagnosing breast by using x-ray to both breasts. It

Mdiagnosing breast by using x-ray to both breasts. It Asian Journal of Medical and Clinical Sciences Estimation radiation risk during mammography in Sudan 1 1 1, Hifaa.Mohamed Khair, Hiba.Osman, Abdelmoneim Sulieman* 1 Al Neelain University, Faculty of Science

More information

Medical Diagnostic Imaging

Medical Diagnostic Imaging Medical Diagnostic Imaging Laboratories Medical Diagnostic Imaging Lab Name Location Person in Charge Programs Served Courses Served Patient Care and Management (2) Introduction to MDI Radiographic Technique

More information

Introduction 1. Executive Summary 5

Introduction 1. Executive Summary 5 Roman_pages 20-09-2005 21:01 Pagina IX Table of contents Introduction 1 Executive Summary 5 1. Epidemiological guidelines for quality assurance in breast cancer screening 15 1.10 Introduction 17 1.20 Local

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Bureau of Laboratory Quality Standards

Bureau of Laboratory Quality Standards 1. Condom -Lubricant ISO 4074:2002/Cor.1:2003 Annex C -Thickness ISO 4074:2002/Cor.1:2003 Annex F - Length ISO 4074:2002 / Cor1 : 2003 - Width ISO 4074 : 2002/Cor 1 : 2003 - Bursting pressure and Volume

More information

Mammography device use in Turkey, and quantity and quality analysis of mammography education

Mammography device use in Turkey, and quantity and quality analysis of mammography education Diagn Interv Radiol 2007; 13:129 133 Turkish Society of Radiology 2007 BREAST IMAGING ORIGINAL ARTICLE Mammography device use in Turkey, and quantity and quality analysis of mammography education Nuray

More information

Digital film dosimetry in radiotherapy and the development of analytical applications software

Digital film dosimetry in radiotherapy and the development of analytical applications software University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2005 Digital film dosimetry in radiotherapy and the development

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

IMPROVING THE RADIOLOGICAL OCCUPATIONAL PROTECTION IN PARAGUAY

IMPROVING THE RADIOLOGICAL OCCUPATIONAL PROTECTION IN PARAGUAY IMPROVING THE RADIOLOGICAL OCCUPATIONAL PROTECTION IN PARAGUAY Virginia Romero de González; Elsa Siris Duré Romero; Luis Edilberto Moré Torres Comisión Nacional de Energía Atómica, Dirección General de

More information

Challenges for Latin America s. Information on Drug Consumption

Challenges for Latin America s. Information on Drug Consumption Challenges for Latin America s Information on Drug Consumption Fatima Marinho, MD, MPH, PHD Coordinator of Information and Health Analysis - PAHO International Day against Drug Abuse & Illicit Trafficking

More information

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria International Journal of Medical Imaging 2015; 3(4): 69-74 Published online June 6, 2015 (http://www.sciencepublishinggroup.com/j/ijmi) doi: 10.11648/j.ijmi.20150304.11 ISSN: 2330-8303 (Print); ISSN: 2330-832X

More information

Skin Model and its impact on Digital Mammography

Skin Model and its impact on Digital Mammography Skin Model and its impact on Digital Mammography Rodrigo T. Massera; Alessandra Tomal Institute of Physics "Gleb Wataghin University of Campinas Campinas, Brazil 1 Outline Motivation Methodology Results

More information

Mammography. What is Mammography? What are some common uses of the procedure?

Mammography. What is Mammography? What are some common uses of the procedure? Mammography What is Mammography? Mammography is a specific type of imaging that uses a low-dose x-ray system to examine breasts. A mammography exam, called a mammogram, is used to aid in the early detection

More information

Y FILMS DOSIMETR Nederland België / Belgique

Y FILMS DOSIMETR Nederland België / Belgique DOSIMETRY FILMS GAFCHROMIC Dosimetry Films The Self-developing Dosimetry Films that Allow You to Go Filmless Do away with the cost and headache of calibrating and maintaining a processor Do away with hazardous

More information

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program ACR MRI Accreditation Program Update Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives Present requirements of the new

More information

MAMMOGRAPHY QC: MORE THAN PHANTOM IMAGES. Stephanie Schofield RTR QC Technologist Nova Scotia Health Authority

MAMMOGRAPHY QC: MORE THAN PHANTOM IMAGES. Stephanie Schofield RTR QC Technologist Nova Scotia Health Authority 1 MAMMOGRAPHY QC: MORE THAN PHANTOM IMAGES Stephanie Schofield RTR QC Technologist Nova Scotia Health Authority Disclaimer 2 I have nothing to disclose. Outline 3 QC tests for the mammography technologist

More information

Closing of Nucletron acquisition Building a stronger position in the cancer care market. September 15, 2011

Closing of Nucletron acquisition Building a stronger position in the cancer care market. September 15, 2011 Closing of Nucletron acquisition Building a stronger position in the cancer care market September 15, 2011 Important step in Elekta s growth strategy Nucletron is the world leader in Brachytherapy and

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

Radiation Protection in Dental Intraoral Radiography Edition

Radiation Protection in Dental Intraoral Radiography Edition Radiation Protection In Intraoral Dental Radiography is one of a series of training CDs designed for staff that use or are exposed to radiation in their daily work. It would be useful to take the Basic

More information

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY Journal of Bangladesh Academy of Sciences, Vol. 3, No. 1, 7-17, 211 A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY MAHFUZA BEGUM 1, A. S. MOLLAH

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,

More information

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT The use of dental CBCT equipment must comply with all the regulations (IRR99 and IR(ME)R2000)

More information

HISTORY OF MQSA AND ACR

HISTORY OF MQSA AND ACR HISTORY OF MQSA AND ACR DEBORAH THAMES R.T. (R)(M)(QM) WHY MQSA? In the United States, there was a lack of standards in mammography imaging. Reporting Imaging Type of imaging screening/diagnostic Equipment

More information

The relationship of the mean glandular dose with compressed breast thickness in mammography

The relationship of the mean glandular dose with compressed breast thickness in mammography Review Article Page 1 of 8 The relationship of the mean glandular dose with compressed breast thickness in mammography Xiang Du, Ningle Yu, Yimei Zhang, Jin Wang Department of Radiation Protection, Jiangsu

More information

Biodosimetry: Latin American Biological Dosimetry Network (LBDNet)

Biodosimetry: Latin American Biological Dosimetry Network (LBDNet) Biodosimetry: Latin American Biological Dosimetry Network (LBDNet) 23rd WiN Global Annual Conference- Women in Nuclear Meet Atomos for Peace Lic. Marina Di Giorgio mdigiorgio@arn.gob.ar 24 28 August 2015

More information

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

More information

1 Course Syllabus + Study Guide for Lecture and Laboratory

1 Course Syllabus + Study Guide for Lecture and Laboratory 1 Course Syllabus + Study Guide for Lecture and Laboratory /BERGEN COMMUNITY COLLEGE Division of Health Professions/Radiography Program Fall 2014 A. General Course Information Title: Radiography I Credits:

More information

ABSTRACT I. INTRODUCTION

ABSTRACT I. INTRODUCTION 2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography

More information

Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography

Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography Diagn Interv Radiol 2007; 13:134 139 Turkish Society of Radiology 2007 BREAST IMAGING ORIGINAL ARTICLE Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography

More information

Radiographic techniques in screen-film mammography

Radiographic techniques in screen-film mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 3, SUMMER 2002 Radiographic techniques in screen-film mammography Thomas R. LaVoy* 4314 Kelsey Drive, Syracuse, New York 13215 Walter Huda

More information

Film processing procedures have a direct effect on the quality of a radiograph.

Film processing procedures have a direct effect on the quality of a radiograph. . Film processing procedures have a direct effect on the quality of a radiograph. Processing is a series of steps that changes the latent image on the exposed film into a radiograph by producing a visible

More information

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions (Revised: July 31, 2017)

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions (Revised: July 31, 2017) The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions (Revised: July 31, 2017) Table of Contents APPLICATION - GENERAL... 1 MOVED FACILITIES AND UNITS...

More information