To Shield or Not to Shield? Lincoln L. Berland, M.D.

Size: px
Start display at page:

Download "To Shield or Not to Shield? Lincoln L. Berland, M.D."

Transcription

1 To Shield or Not to Shield? Lincoln L. Berland, M.D.

2 Disclosures Consultant to: Nuance, Inc. Page 2

3 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation. Radiation to the breast on CT pulmonary angiography may be times higher than 2-view screening mammogram ( mgy). Bismuth shield is designed to, and is proven to decrease breast exposure on CT. Page 3

4 Concepts However, is there an alternative that can achieve an equivalent reduced radiation exposure with at least equivalent image quality on chest CT? What are the advantages and disadvantages of various strategies? Page 4

5 Our Phantom Study Lincoln L. Berland, M.D., F.A.C.R. Michael Yester, Ph.D.

6 Methods Compared strategies of using Bismuth breast shield with standard radiation exposure factors vs. no shield with reduced ma. Chest region of an anthropomorphic phantom, without and with Bismuth breast shield Phantom alone Phantom with 1 additional 3.5 cm simulated fat layer Phantom with 2 additional 3.5 cm simulated fat layers Page 6

7 Methods mas adjusted to achieve nearly equivalent standard deviations in the mediastinum without and with the shield. CT attenuation numbers recorded from: Mediastinum Superficial soft tissues Page 7

8

9

10 Results: Without and With Shield Effect on radiation exposure of using shield without modifying technique - constant ma. No Shield With Shield ma Mediastinal SD Dose Reduction- Surface Dose Reduction - Mediastinum % 20-25% Page 10

11 Equivalent Radiation Without Shield What reduced ma without shield is calculated to result in noise (based on SD) in mediastinum similar to routine ma with shield? What reduced ma without shield actually achieves such noise? No Shield - Calculated No Shield - Actual ma Mediastinal SD Dose Reduction- Surface and Mediastinum % % Page 11

12 Radiation at Surface In each case, using noise in mediastinum with shield to calculate what lower ma should be used without a shield suggests the need for a higher exposure than is necessary to achieve the same noise level. Therefore, noise is higher with a shield than one would expect based on the actual dose reduction. These higher noise levels with the shield are not visible as streak artifacts and may be the result of the shield attenuation in both AP and PA orientations. Page 12

13 Radiation at Mediastinum For equivalent noise within mediastinum, radiation dose reduction within mediastinum: No fat rings One added ring Two added rings No Shield, lower ma 42% 41% 40% Shield, routine ma 20-25% 18% 15% Page 13

14 Hounsfield Unit Variation Attenuation in mediastinum varied 3 HU between shield, no shield Surface attenuation 120 kvp No fat rings One added ring Two added rings No Shield -5 HU -69 HU -69 HU Shield 27 HU -39 HU -44 HU Page 14

15 Streak Artifacts Shield creates streak artifacts, particularly marked at upper and lower edges. Page 15

16 Why Do We Expect the Shield to Work? Why is the use of shielding an attractive concept? Use of lead shields ingrained in our training and seems intuitive However, fundamental difference: Lead shields are intended to attenuate radiation for areas in patients or for people who are not being examined, but Bismuth shields intended to limit radiation to areas that are being examined Page 16

17 Why Doesn t the Shield Work? Shield attenuates photons both entering (from the front), but also exiting (through the front), wasting radiation Causes beam-hardening, streak artifacts, HU changes Page 17

18 Why Doesn t the Shield Work? Does the shield selectively attenuate low-energy photons? No. Preinstalled filters attenuate low-energy photons before reaching the patient or shield. Spectrum from one manufacturer: mean photon energy of 69.1 kev for a 120-kV, but shields only increase the mean energy to 70.5 kev kev. McCollough CH, Wang J, Berland LL. Bismuth Shields for CT Dose Reduction: Do They Help or Hurt? Journal of the American College of Radiology. Volume 8, Issue 12, Pages , December 2011 Page 18

19 May confuse AEC programs Why Doesn t the Shield Work? If place shield before radiograph, AEC compensates with increased exposure If place shield after radiograph, exposure is too low, decreasing image quality If real-time AEC, then will partially compensate with increased exposure when shield in field, whether you put the shield on before or after scanogram Do you know if your system has real-time AEC? Page 19

20 Why Doesn t the Shield Work? Four other phantom studies all led to same conclusion: Geleijns J, Wang J, McCollough. The use of breast shielding for dose reduction in pediatric CT: arguments against the proposition. Pediatr Radiol 40: , 2010 Vollmar SV, Kalender WA. Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations. Eur Radiol 18: , 2008 Wang J, Duan X, Christner JA, Leng S, Yu L, McCollough CH. Radiation dose reduction to the breast in thoracic CT: Comparison of bismuth shielding, organ-based tube current modulation, and use of a globally decreased tube current. Med Phys 38: , 2011 Wang J, Duan X, Christner JA, Leng S, Grant KL, McCollough CH. Bismuth Shielding, Organ-based Tube Current Modulation, and Global Reduction of Tube Current for Dose Reduction to the Eye at Head CT. Radiology 262: , 2012 Page 20

21 Why Not Reduce Dose and Add Shield? Dose reduction is linear proportional to mas. Theoretical example: Standard, no shield: 180 ma 120 ma, no shield decreases breast dose 33% 180 ma, with shield decreases breast dose 33% Theoretical reduced dose plus shield: 120 ma, with shield, reduces dose to 44% of original 80 ma, no shield also reduces dose to 44% of original, better noise, better sectional dose profile, no artifacts, etc. Page 21

22 Disadvantages of Shield Same or lower noise in mediastinum can be obtained with the same or lower exposure to breasts without shield using a reduced ma as with shield using a baseline ma Similar results with added simulated fat rings Substantially lower radiation to mediastinum and remainder of region with lower ma without the shield Shield alters attenuation values Shield creates streak artifacts In field: deep to shield At upper and lower edges Worse if any motion Page 22

23 Other Disadvantages of Shield Proper use depends on knowing how your manufacturer, model of scanner and version of software perform AEC Have to stop after scout and place the shield Costs money Need to use infection control techniques because comes in contact with patient Page 23

24 Recommendations Chest CT Identify population of patients for whom decreased breast dose desired: E.g. women under 50 yo, without terminal condition Reduce ma by 30%, do not use a shield Page 24

25 Breast Exposure on Abdominal CT Breasts sometimes in field, depending on: Technologist upper range Diaphragm position Breast size Concerns with using breast shield on abd CT: Causes streaks through upper abdomen, particularly at edge of shield Alters attenuation Page 25

26 Use of shield depends on: Breast, diaphragm location Conclusions Abdominal CT Patient comorbidities and age Need for high-quality images through upper liver Other measures to limit breast radiation: Lead apron or shield can protect from overranging if breasts not in field Can keep bra on or retract breasts above field Page 26

27 Recommendations Abdominal CT Train and encourage technologists to limit overranging Keep bra on if not also scanning chest, attempt to retract breasts If possible, reduce ma 30% for upper component of abdominal scan Page 27

28 Challenges Although several studies suggest that lower dose chest CT is diagnostically acceptable, there are no specific standards for appropriate ma for chest CT. Challenge is to design low-dose protocols and procedures that assure use of appropriately lower ma for chest CT in vulnerable women. Use of Bismuth or lead shield for abdominal CT depends on multiple factors Page 28

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

Ask EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT

Ask EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Shielding in pediatric CT Claudio Granata (IRCCS Istituto Giannina Gaslini, IT) Joana Santos (ESTeSC-Coimbra Health School, PT) Elina

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Ask EuroSafe Imaging Tips & Tricks. CT Working Group

Ask EuroSafe Imaging Tips & Tricks. CT Working Group Ask EuroSafe Imaging Tips & Tricks CT Working Group Organ Based Tube Current Modulation to Reduce Radiation Dose to Superficial Radiosensitive Organs Eileen Kelly (Galway University Hospitals, IE) Matthias

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Automated CT Protocol Design Advantages and Pitfalls of Algorithm-Based Technique Selection in Pediatrics. Disclosures 7/22/2014. Learning Objectives

Automated CT Protocol Design Advantages and Pitfalls of Algorithm-Based Technique Selection in Pediatrics. Disclosures 7/22/2014. Learning Objectives Automated CT Protocol Design Advantages and Pitfalls of Algorithm-Based Technique Selection in Pediatrics Robert MacDougall, M.Sc. Department of Radiology Boston Children s Hospital Disclosures 2 Learning

More information

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Learning Objectives: Limitations for estimating patient dose for CT Methods

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

CT angiography techniques. Boot camp

CT angiography techniques. Boot camp CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

Paediatric Dose Reduction and Image Quality

Paediatric Dose Reduction and Image Quality Paediatric Dose Reduction and Image Quality Alan Whiteside The majority of this work was undertaken as part of MSc Thesis of Helen Dixon. Introduction Paediatric CT protocols result in a higher effective

More information

Estimating Iodine Concentration from CT Number Enhancement

Estimating Iodine Concentration from CT Number Enhancement Estimating Iodine Concentration from CT Number Enhancement Rosemary Eaton, Andrew Shah, Jane Shekhdar Medical Physics, Mount Vernon Hospital CT Users Group 4 th October 212, Edinburgh Summary Background

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Dual-Energy CT Applications in Radiation Therapy

Dual-Energy CT Applications in Radiation Therapy THE UNIVERSITY OF WISCONSIN MADISON Dual-Energy CT Applications in Radiation Therapy - Jessica Miller 1 Disclosures Funding provided by Siemens Medical 2 Learning objectives General principles of dual

More information

Dose Reduction of the Female Breast in Chest CT

Dose Reduction of the Female Breast in Chest CT Cardiopulmonary Imaging Original Research Seidenfuss et al. Reduction of CT Radiation Dose to Female Breast Cardiopulmonary Imaging Original Research Downloaded from www.ajronline.org by 46.3.197.9 on

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

To Shield or Not to Shield: Application of Bismuth Breast Shields

To Shield or Not to Shield: Application of Bismuth Breast Shields Cardiopulmonary Imaging Opinion Colletti et al. Bismuth Breast Shields Cardiopulmonary Imaging Opinion FOCUS ON: Patrick M. Colletti 1 Orlando A. Micheli Kai H. Lee Colletti PM, Micheli OA, Lee KH Keywords:

More information

Breast Dose Reduction Options During Thoracic CT: Influence of Breast Thickness

Breast Dose Reduction Options During Thoracic CT: Influence of Breast Thickness Cardiopulmonary Imaging Original Research Revel et al. Breast Dose Reduction in Thoracic CT Cardiopulmonary Imaging Original Research Marie-Pierre Revel 1 Isabelle Fitton 2 Etienne Audureau 3 Joseph Benzakoun

More information

Dual Energy CT Aortography: Can We Reduce Iodine Dose??

Dual Energy CT Aortography: Can We Reduce Iodine Dose?? Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest

More information

Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography

Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography The British Journal of Radiology, 81 (2008), 499 503 SHORT COMMUNICATION Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography 1 G R IBALL, MSc, BSc, 2

More information

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems Dental ConeBeam Computed Tomography (CBCT) X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL CONEBEAM COMPUTED TOMOGRAPHY (CBCT) X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability.

More information

Dual Energy CT of the Liver

Dual Energy CT of the Liver 34th Annual Course October 2011 Washington, DC Dual Energy CT of the Liver Vassilios Raptopoulos, MD Beth Israel Deaconess Medical Center Harvard Medical School Dual Energy CT (DECT) Different materials

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

Radiation Dose Estimations to the Thorax Using Organ-Based Dose Modulation

Radiation Dose Estimations to the Thorax Using Organ-Based Dose Modulation Cardiopulmonary Imaging Original Research Lungren et al. Organ-Based Dose Modulation Cardiopulmonary Imaging Original Research Matthew P. Lungren 1 Terry T. Yoshizumi 1 Samuel M. Brady 1 Greta Toncheva

More information

How to Develop CT Protocols for Children

How to Develop CT Protocols for Children How to Develop CT Protocols for Children Introduction Prior to 2001 the vast majority of CT imaging of children was conducted using the same or similar techniques used for adult imaging. In 2001, several

More information

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Rebecca Milman Marsh, Ph.D. University of Colorado Department of Radiology Who in the Facility Works

More information

Preparing for Medical Physics Components of the ABR Core Examination

Preparing for Medical Physics Components of the ABR Core Examination Preparing for Medical Physics Components of the ABR Core Examination The ABR core examination for radiologists contains material on medical physics. This content is based on the medical physics that is

More information

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT protocols Tips Leukaemia / lymphoma: ~ 35% acute lymphoblastic

More information

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens DEfinitely Siemens True Dual Energy Dr. Stefan Ulzheimer, Siemens Healthcare GmbH International version. Not for distribution in the US. Unrestricted Siemens AG 2015 All rights reserved. The products/features

More information

Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT

Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Hoang et al. Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Jenny K. Hoang 1,2 Terry T. Yoshizumi 1 Kingshuk Roy

More information

Dual-Energy 101: Principles, Methods and Dose

Dual-Energy 101: Principles, Methods and Dose Dual-Energy 101: Principles, Methods and Dose Juan Carlos Ramirez-Giraldo, Ph.D Staff Scien2st, Collabora2ons Manager SE Region ISCT San Francisco, 2017 Siemens Medical Solu2ons USA, Inc., 2017 Page 1

More information

Effects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography

Effects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography Original Article Effects of shielding the radiosensitive superficial organs of ORNL pediatric phantoms on dose reduction in computed tomography Parisa Akhlaghi, Hashem Miri Hakimabad, Laleh Rafat Motavalli

More information

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Debra Pennington, MD Director of Imaging Dell Children s Medical Center Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates Patient Dose Estimates from CT Examinations Patient Dose Estimates from CT Examinations John M. Boone, Ph.D. Professor and Vice Chairman of Radiology Professor of Biomedical Engineering University of California,

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016 Computed Tomography The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge that, when coupled with selected clinical experiences,

More information

The Computed Tomography Examination

The Computed Tomography Examination CONTENT SPECIFICATIONS The Computed Tomography Examination The purpose of The American Registry of Radiologic Technologists (ARRT ) Computed Tomography Examination is to assess the knowledge and cognitive

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Young Hen Lee 1 Eun-tae Park 1 Pyong Kon Cho 1 Hyung Suk Seo 1 Bo-Kyung Je 1 Sang-il Suh 2 Kyung-Sook

More information

Metal Artifact Reduction by Dual Energy CT

Metal Artifact Reduction by Dual Energy CT Metal Artifact Reduction by Dual Energy CT Poster No.: C-0108 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper T. Johnson, F. Bamberg, A. Dierks, H.-C. Becker, M. F. Reiser; Munich/DE

More information

Medical Diagnostic Imaging

Medical Diagnostic Imaging Medical Diagnostic Imaging Laboratories Medical Diagnostic Imaging Lab Name Location Person in Charge Programs Served Courses Served Patient Care and Management (2) Introduction to MDI Radiographic Technique

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

What s New in Breast Imaging. Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia

What s New in Breast Imaging. Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia What s New in Breast Imaging Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia Disclosure Hologic, Inc. Shareholder and research agreement. Volpara Solutions, Ltd. Shareholder

More information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation

Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation Medical Physics and Informatics Original Research Angel et al. Radiation Dose During Chest CT Medical Physics and Informatics Original Research Erin Angel 1,2 Nazanin Yaghmai 1 Cecilia Matilda Jude 1 John

More information

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center Mannudeep K. Kalra, MD Webster Center for Quality and Safety Massachusetts General Hospital Harvard Medical School

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

An Introduction to Dual Energy Computed Tomography

An Introduction to Dual Energy Computed Tomography An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the

More information

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School OUTLINE 1 st Presentation: Dose optimization strategies

More information

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters Medical Physics and Informatics Review Maldjian and Goldman Reducing Radiation Dose in Body CT Medical Physics and Informatics Review FOCUS ON: Pierre D. Maldjian 1 Alice R. Goldman Maldjian PD, Goldman

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

1 Course Syllabus + Study Guide for Lecture and Laboratory

1 Course Syllabus + Study Guide for Lecture and Laboratory 1 Course Syllabus + Study Guide for Lecture and Laboratory /BERGEN COMMUNITY COLLEGE Division of Health Professions/Radiography Program Fall 2014 A. General Course Information Title: Radiography I Credits:

More information

Lens Dose in Routine Head CT: Comparison of Different Optimization Methods With Anthropomorphic Phantoms

Lens Dose in Routine Head CT: Comparison of Different Optimization Methods With Anthropomorphic Phantoms Medical Physics and Informatics Original Research Nikupaavo et al. Lens Dose in Head CT Medical Physics and Informatics Original Research Ulla Nikupaavo 1,2 Touko Kaasalainen 1,3 Vappu Reijonen 1 Sanna-Mari

More information

Breast CT and Dosimetry

Breast CT and Dosimetry 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of

More information

Documentation, Codebook, and Frequencies

Documentation, Codebook, and Frequencies Documentation, Codebook, and Frequencies Dual-Energy X-ray Absorptiometry Femur Bone Measurements Examination Survey Years: 2005 to 2006 SAS Transport File: DXXFEM_D.XPT January 2009 NHANES 2005 2006 Data

More information

A Snapshot on Nuclear Cardiac Imaging

A Snapshot on Nuclear Cardiac Imaging Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the

More information

CT Radiation Safety in Adults: Where are we now? What can be done?

CT Radiation Safety in Adults: Where are we now? What can be done? CT Radiation Safety in Adults: Where are we now? What can be done? David A Leswick MD FRCPC University of Saskatchewan Canadian Association of Radiologists Annual General Meeting Friday April 26, 2013

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

Radiation dose reduction in computed tomography: techniques and future perspective

Radiation dose reduction in computed tomography: techniques and future perspective REVIEW Radiation dose reduction in computed tomography: techniques and future perspective Despite universal consensus that computed tomography (CT) overwhelmingly benefits patients when used for appropriate

More information

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert University of Groningen Quantitative CT myocardial perfusion Pelgrim, Gert IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check

More information

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Dose Reduction Options in Cardiac CT

Dose Reduction Options in Cardiac CT Dose Reduction Options in Cardiac CT Doyle P, Ball P*, Donnelly P # Radiological Sciences & Imaging, Forster Green Hospital *Department of Radiology, Ulster Hospital # Department of Cardiology, Ulster

More information

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017 EXAMINATION CONTENT SPECIFICATIONS Computed Tomography The purpose of the computed tomography examination is to assess the knowledge and cognitive skills underlying the intelligent performance of the tasks

More information

Cardiopulmonary Imaging Original Research

Cardiopulmonary Imaging Original Research Cardiopulmonary Imaging Original Research Vardhanabhuti et al. Three Algorithms to Assess Standard- and Low-Dose Chest CT Cardiopulmonary Imaging Original Research FOCUS ON: Varut Vardhanabhuti 1,2 Robert

More information

354 Korean J Radiol 9(4), August 2008

354 Korean J Radiol 9(4), August 2008 Review of Failed CT Phantom Image Evaluations in 2005 and 2006 by the CT Accreditation Program of the Korean Institute for Accreditation of Medical Image Hye Jung Park, MD 1 Seung Eun Jung, MD 1, 2 Young

More information

The effect of breast shielding during lumbar spine radiography

The effect of breast shielding during lumbar spine radiography 26 research article The effect of breast shielding during lumbar spine radiography Nejc Mekis 1, Dejan Zontar 2, Damijan Skrk 2 1 University of Ljubljana, Faculty of Health Sciences, Medical Imaging and

More information

Liver, Liver Fat, Fibrosis and HCC Screening with DECT

Liver, Liver Fat, Fibrosis and HCC Screening with DECT Liver, Liver Fat, Fibrosis and HCC Screening with DECT Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org

More information

Outline. Lifetime Attributable Risk 10 mgy in 100,000 exposed persons (BEIR VII 2006) SPECT/CT and PET/CT Dosimetry

Outline. Lifetime Attributable Risk 10 mgy in 100,000 exposed persons (BEIR VII 2006) SPECT/CT and PET/CT Dosimetry SPECT/CT and PET/CT Dosimetry Lifetime Attributable Risk 10 mgy in 100,000 exposed persons (BEIR VII 2006) All Solid Tumors Leukemia Frederic H. Fahey DSc Male Female Male Female Children s Hospital Boston

More information

Radiation Exposure 1980 to 2006

Radiation Exposure 1980 to 2006 Radiation Exposure 1980 to 2006 Background 3-6 msv/yr Natural (85% 45%) Radon Cosmic Rays Air travel Living at Altitude Man-made (15% 55%) Medical Imaging** mgy Radiation Therapy cgy Radiation Whole Body

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski White Paper October 2012 SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine

More information

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Original Article Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Shun-Yu Gao, Xiao-Peng Zhang, Yong Cui, Ying-Shi

More information

Contrast-Enhanced Digital Mammography

Contrast-Enhanced Digital Mammography 2015 ARRS Breast Symposium Contrast-Enhanced Digital Mammography John Lewin, M.D. Diversified Radiology of Colorado CEDM - Outline History Technique Literature Review / Cases Clinical Status Inexpensive,

More information

Collapsed Cone Convolution 2D illustration

Collapsed Cone Convolution 2D illustration Collapsed Cone Convolution 2D illustration 8 cones Energy desposition decreases very quickly with distance Energy is absorber in blue pixels only. IGRT1 technologies Paweł Kukołowicz Warsaw, Poland IGRT

More information

Breast Cancer PET/CT Imaging Protocol

Breast Cancer PET/CT Imaging Protocol Breast Cancer PET/CT Imaging Protocol Scanning Protocol: Patients are scanned from the top of the neck through the pelvis. Arms-up position is used to avoid beam-hardening artifact in the chest and abdomen.

More information

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3 Coronary Artery Bypass Graft Imaging with CT Angiography and Iterative Reconstruction: Quantitave Evaluation of Radiation Dose Reduction and Image Quality Simon Nepveu 1, Irina Boldeanu 1, Yves Provost

More information

Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety

Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February 2008. Radiation Safety Sub -headings Background Radiation Effects of Radiation Safe Levels Effective Doses ALARA Principle Radiation

More information

Gender differences in CT calcium scoring: A phantom study

Gender differences in CT calcium scoring: A phantom study Gender differences in CT calcium scoring: A phantom study Nicholas Petrick, Qin Li, Benjamin Berman, Marios A Gavrielides, Rongping Zeng, Berkman Sahiner CDRH/OSEL/DIDSR U.S. Food and Drug Administration

More information

Patient / Organ Dose in CT

Patient / Organ Dose in CT Patient / Organ Dose in CT Patient specific and organ dose estimation H.D. Nagel Dr. HD Nagel, Science & Technology for Radiology Buchholz / Germany www.sascrad.com 1 Topics CTDI & patient dose SSDE Organ

More information

CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings

CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings Poster No.: C-3289 Congress: ECR 2010 Type: Scientific Exhibit Topic: Chest - Your latest results Authors: M. K. Gill,

More information

USING TEAM AUDITS TO IMPROVE IMAGE QUALITY: Our Institution s Experience

USING TEAM AUDITS TO IMPROVE IMAGE QUALITY: Our Institution s Experience USING TEAM AUDITS TO IMPROVE IMAGE QUALITY: Our Institution s Experience Lincoln L. Berland, MD Deborah D. Flint, PE, BIE, MBA Beth S. Winningham, BSRT, MBA UAB Health System 900-bed Facility and Level

More information

Whole Body CT Protocol Update 2018

Whole Body CT Protocol Update 2018 Whole Body CT Protocol Update 2018 10 th Nordic Course in Trauma Radiology Gothenburg, Sweden K.SHANMUGANATHAN M.D. Disclosure of Commercial Interest Neither I nor my immediate family members have a financial

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Christner et al. Estimating Effective Dose for CT Medical Physics and Informatics Original Research FOCUS ON: Jodie A. Christner 1 James M. Kofler Cynthia

More information

Effective Utilization of Imaging. John V. Roberts, M.D. Premier Radiology Abdominal Imaging

Effective Utilization of Imaging. John V. Roberts, M.D. Premier Radiology Abdominal Imaging Effective Utilization of Imaging John V. Roberts, M.D. Premier Radiology Abdominal Imaging Safety Contrast and Radiation What to order Abdomen/Pelvis Brain/Spine Chest Musculoskeletal Ob/Gyn Head and Neck

More information

Pitfalls of the Pediatric Chest and Abdomen SPR 2017

Pitfalls of the Pediatric Chest and Abdomen SPR 2017 Pitfalls of the Pediatric Chest and Abdomen SPR 2017 Richard I. Markowitz, MD, FACR Children s Hospital of Philadelphia Perelman School of Medicine University of Pennsylvania No Disclosures Cognitive Perceptual

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information