Fetal Dose Calculations and Impact on Patient Care

Size: px
Start display at page:

Download "Fetal Dose Calculations and Impact on Patient Care"

Transcription

1 Fetal Dose Calculations and Impact on Patient Care Matt Hough, MS, DABR, DABMP Florida Hospital Diagnostic Medical Physics and Radiation Safety

2 Resource ACR-SPR Practice Parameter for Imaging Pregnant or Potentially Pregnant Adolescents and Women with Ionizing Radiation (Revised 2013) Includes Overall guidance Appendices with detailed Risks Sample Policy Sample Patient History Form Sample Consent

3 Estimated Fetal Dose Ranges for Diagnostic Procedures All non-abdominal imaging Negligible scatter dose to fetus Abdominal Primary Exposure General X-ray and Fluoro Well below 20 mgy CT (Single Phase) Usually less than 35 mgy (15-20 mgy) CT (Multi Phase) Multiple of single phase doses Fluoroscopically Guided Interventional procedure Can be well above 100 mgy

4 ACR-SPR Practice Parameter for Imaging Pregnant 2013

5 Prospective Scenario Screening and Pregnancy Testing Only for high risk procedures? What if you have a negative test? What if you have a positive test? Avoid multi-phase CT and FGI if possible Is there an alternate procedure? ACR Appropriateness Criteria If procedure must be done Radiologist should Consult a Qualified Medical Physicist for dose estimate Counsel and Consent Patient

6 Retrospective Scenario Pregnancy discovered during/after procedure Radiologist should Consult a Qualified Medical Physicist for dose estimate Counsel Patient s ordering physician? Counsel Patient

7 Qualified Medical Physicist Role QMP Dose Estimation Where do we start? What options do we have? Communicating risk associated with that Dose estimate How do we communicate our estimate to the Radiologist? Do we ever speak directly to patients?

8 CTDI is Computed Tomography Dose Index A standard parameter used to measure Scanner Radiation Output in a reproducible manner There are different types CTDI 100, CTDI W, CTDI VOL There are different sizes of CTDI phantoms Adult Body: 32 cm diameter Adult Head, Ped Body, Ped Head: 16 cm diameter CTDI VOL is the one dose metric required to be reported by scanner

9 What affects Scanner Radiation Output? CTDI VOL Avg = mgy CTDI VOL Min = mgy CTDI VOL Max = mgy

10 CTDI is not Actual Patient Dose!

11 How do we estimate Organ and Fetal Doses in CT? Different approaches to the same problem Direct Measurement Ion Chamber in Rando Phantom (Felmlee Method) Monte Carlo Simulation Voxelized Patient Phantom (Angel Method) Stylized Mathematical Phantom (Radimetrics one vendor) Other methods available

12 CT Fetal Dose Estimation - Felmlee Methodology Felmlee, et Al Estimated Fetal Radiation Dose AJR 154: , January 1990

13 CT Fetal Dose Estimation Important Notes Scanners from 1980s - Felmlee Methodology Data is averaged over multiple scan parameters (e.g. kvp) Phantom based Overestimate for large patient, underestimate for small patient What CTDI? Not scanner reported CTDI vol CTDI 100 for 16 cm phantom Have to back-calculate based on scanner specific measurements Can be TCM corrected manually Can be pitch corrected

14 CT Fetal Dose Estimation - Angel Methodology Angel, et Al Radiation Dose to Fetus Radiology Vol 249 Number 1 October 2008

15 CT Fetal Dose Estimation Important Notes - Angel Methodology Monte Carlo based calculation GE LightSpeed 16 Source Model 120 kvp only Patient based but not patient specific? Dose estimate based on Mother s Perimeter and Fetal Depth as modeled from study s sample Provides Fetal Dose per 100 mas Can be TCM corrected manually Can be Pitch corrected manually

16 CT Fetal Dose Estimation - Radimetrics Methodology

17 CT Fetal Dose Estimation - Radimetrics Methodology Important Notes Monte Carlo based calculation Stylized mathematical model of patient anatomy Scan based but not patient specific Provides Uterus and Fetal Dose for 1 st, 2 nd, and 3 rd trimester patient/fetal models Dose is scaled based on slice specific CTDI VOL Automatically TCM corrected Automatically Pitch corrected Automatically kvp corrected

18 First look data

19 Correcting for model limitations

20 What does this show? QMP must understand model limitations and how to correct for them When accounting for model limitations, there seems to be good agreement Radimetrics can be used as a surrogate for other methods to quickly ascertain a fetal dose estimate

21 How do we communicate risk Know your audience Radiologist (knows something about radiation and dose) Ordering Physician (probably knows very little) Patient (probably knows very little and is very scared) Remain in a positive tone Compare estimated dose to threshold for deterministic effects based on fetal age Note potential for stochastic effects

22 Conclusions Institutions should have detailed Policies ACR-SPR is a great resource Benefit versus Risk should always be taken into consideration The majority of diagnostic imaging procedures should not pose deterministic risks If there is a high risk procedure, QMP involvement is important Account for any limitations in Dose estimations Risk should be communicated with a positive tone

National Cancer Institute

National Cancer Institute National Cancer Institute A Dosimetry Summary of CT Participants in the National Lung Screening Trial (NLST) U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES National Institutes of Health AAPM 2015 Anaheim,

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

Estimating Patient Radiation Dose from Computed Tomography

Estimating Patient Radiation Dose from Computed Tomography Estimating Patient Radiation Dose from Computed Tomography C. Cagnon, J. DeMarco, E. Angel, M. McNitt-Gray UCLA David Geffen School of Medicine 1 Patient Dose from CT Advances in Technology... Helical

More information

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates Patient Dose Estimates from CT Examinations Patient Dose Estimates from CT Examinations John M. Boone, Ph.D. Professor and Vice Chairman of Radiology Professor of Biomedical Engineering University of California,

More information

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Learning Objectives: Limitations for estimating patient dose for CT Methods

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

CT Head Dose Reduction Using Spiral Scanning Protocol

CT Head Dose Reduction Using Spiral Scanning Protocol CT Head Dose Reduction Using Spiral Scanning Protocol Reed, William J MD; Broderick, Daniel F, MD; Weindling, Steven M, MD; Czervionke, Leo F MD; and Morin, Richard L; PhD; Mayo Clinic, Department of Radiology;

More information

State of the art and future development for standardized estimation of organ doses in CT

State of the art and future development for standardized estimation of organ doses in CT State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES A THESIS SUBMITTED TO THE GRADUATE FACULTY in partial fulfillment of the requirements

More information

Radiation Units and Dosimetry 15 August Kalpana M. Kanal, Ph.D., DABR 1

Radiation Units and Dosimetry 15 August Kalpana M. Kanal, Ph.D., DABR 1 Introduction Radiation Units and Dosimetry Radiation dose quantities are used as indicators of the risk of biologic damage to patients from x-rays and thus a good knowledge of the different dose parameters

More information

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial Medical Physics and Informatics Original Research Larke et al. Estimated Radiation Dose for Low-Dose Chest CT Medical Physics and Informatics Original Research Frederick J. Larke 1 Randell L. Kruger 2

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Re: PSM of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report

Re: PSM of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report Re: PSM-044-10 of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report Dear members of the National Quality Forum Board: As Professors of Radiologic Physics (at Mayo Clinic

More information

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations Aiping Ding, X. George Xu Rensselaer Polytechnic Institute Troy, NY USA http://rrmdg.rpi.edu 1 Acknowledgements

More information

IMAGE GENTLY HOW CAN YOU HELP?

IMAGE GENTLY HOW CAN YOU HELP? IMAGE GENTLY HOW CAN YOU HELP? Keith J. Strauss, MSc, FAAPM, FACR Director, Radiology Physics & Engineering Children s s Hospital Boston Harvard Medical School Acknowledgment Marilyn J. Goske,, MD Robert

More information

Doses from CT examinations to children suffering from hydrocephalus

Doses from CT examinations to children suffering from hydrocephalus Doses from CT examinations to children suffering from hydrocephalus E. G. Friberg 1, L. J. Holmedal 2, I. Børretzen 1, K. Rosendahl 3 and H. M. Olerud 1 1 Norwegian Radiation Protection Authority, P.O.

More information

CT Dose Reduction in Pediatric Patients

CT Dose Reduction in Pediatric Patients CT Dose Reduction in Pediatric Patients By Kelly Firestine, RT(R)(CT)(M) Executive Summary CT is an incredibly valuable imaging tool, but there are unique concerns with pediatric patients, including the

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

RADIATION SAFETY. Junior Radiology Course

RADIATION SAFETY. Junior Radiology Course RADIATION SAFETY Junior Radiology Course Expectations for the Junior Radiology Course Medical School wants students to learn basic principles, factual knowledge, safety info, etc. Medical Students want

More information

Patient / Organ Dose in CT

Patient / Organ Dose in CT Patient / Organ Dose in CT Patient specific and organ dose estimation H.D. Nagel Dr. HD Nagel, Science & Technology for Radiology Buchholz / Germany www.sascrad.com 1 Topics CTDI & patient dose SSDE Organ

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

AAPM Annual Meeting. ACR Accreditation Update in CT

AAPM Annual Meeting. ACR Accreditation Update in CT AAPM Annual Meeting ACR Accreditation Updates in CT, Ultrasound, Mammography and MRI: ACR Accreditation Update in CT Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiological Sciences

More information

CT Dosimetry in the Clinical Environment: Methods and Analysis

CT Dosimetry in the Clinical Environment: Methods and Analysis CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida

More information

Patient Radiation Doses from Adult and Pediatric CT

Patient Radiation Doses from Adult and Pediatric CT Radiation Doses from CT CT Imaging Original Research Walter Huda 1 Awais Vance 1,2 Huda W, Vance A Keywords: MDCT, pediatric CT, physics of radiology, radiation dose DOI:.2214/AJR.06.01 Received January

More information

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could

More information

How to Develop CT Protocols for Children

How to Develop CT Protocols for Children How to Develop CT Protocols for Children Introduction Prior to 2001 the vast majority of CT imaging of children was conducted using the same or similar techniques used for adult imaging. In 2001, several

More information

Radiation Safety in the Catheterization Lab

Radiation Safety in the Catheterization Lab SCAI FALL FELLOWS COURSE - 2015 Radiation Safety in the Catheterization Lab V. Vivian Dimas, MD, FSCAI Associate Professor Pediatrics, Cardiology UT Southwestern Medical Center Dallas TX None Disclosures

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center Mannudeep K. Kalra, MD Webster Center for Quality and Safety Massachusetts General Hospital Harvard Medical School

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them) Rebecca Milman Marsh, Ph.D. University of Colorado Department of Radiology Who in the Facility Works

More information

Justification, Optimization and Communication in Pediatric CT Imaging: Recent Improvements and Persistent Challenges Designated Emphasis in Nuclear

Justification, Optimization and Communication in Pediatric CT Imaging: Recent Improvements and Persistent Challenges Designated Emphasis in Nuclear Justification, Optimization and Communication in Pediatric CT Imaging: Recent Improvements and Persistent Challenges Designated Emphasis in Nuclear Sciences 2016 Seminar Jerrold Bushberg Ph.D., DABMP,

More information

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

Multislice CT versus cone beam CT: dosimetric and image quality comparision. Multislice CT versus cone beam CT: dosimetric and image quality comparision. Poster No.: C-2589 Congress: ECR 2015 Type: Authors: Scientific Exhibit C. Dionisi 1, A. Crispo 2, E. Stefani 3, A. Lovaglio

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Minimizing CT Radiation Dose CT examinations improve health care and are an essential part

More information

NCRP PAC 4: Radiation Protection in Medicine. Overview. NCRP and PAC 4. Donald L. Miller, MD FSIR FACR Center for Devices and Radiological Health FDA

NCRP PAC 4: Radiation Protection in Medicine. Overview. NCRP and PAC 4. Donald L. Miller, MD FSIR FACR Center for Devices and Radiological Health FDA NCRP PAC 4: Radiation Protection in Medicine Donald L. Miller, MD FSIR FACR Center for Devices and Radiological Health FDA 1 Overview NCRP and PAC 4 Recent PAC 4 publications Reports in preparation 2 NCRP

More information

Cardiovascular Imaging in Pregnancy. Diana Litmanovich, MD

Cardiovascular Imaging in Pregnancy. Diana Litmanovich, MD Cardiovascular Imaging in Pregnancy Diana Litmanovich, MD Cardiovascular disorders 4% of all pregnancies PE DVT Aortic dissection Peripartum cardiomyopathy Acute coronary syndrome Hypertension Pneumonia

More information

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD) Journal of Physics: Conference Series PAPER OPEN ACCESS Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD) To cite this

More information

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific

More information

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Radiation Dose To Pediatric Patients in Computed Tomography in Sudan Omer Osman,Saeed Medical Physics Department, ALNeelain University, Sudan Presentation outlines Introduction Objectives Materials and

More information

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4 Patient Safety Concerns in Diagnostic Radiology? Lawrence T. Dauer, PhD, CHP Assistant Attending Health Physicist Department of Medical Physics RAMPS/GNYCHPS Spring Symposium April 30, 2010 Benefits?

More information

Implementation & optimization of a lung cancer screening CT program. Presented by Izabella Barreto at the 2016 Florida AAPM Chapter Meeting

Implementation & optimization of a lung cancer screening CT program. Presented by Izabella Barreto at the 2016 Florida AAPM Chapter Meeting Implementation & optimization of a lung cancer screening CT program Presented by Izabella Barreto at the 2016 Florida AAPM Chapter Meeting Izabella Barreto, Nathan Quails, Catherine Carranza, Nathalie

More information

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information

HI-Res Extremity Sensation 16

HI-Res Extremity Sensation 16 Page 1 Routine Extremity - (2/14/2013) CTDI: ~20 mgy per acquisition Used for evaluation of: Humerus Forearm Femur Knee Tib/Fib Billing: 1. CT Upper/Lower Extremity of concern without contrast, with contrast,

More information

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT R Aamir, C Lowe, J Damet, P Carbonez, A P H Butler, N Schleich, N G Anderson MARFO, Emmanuel Geant4 User Workshop 2017, Wollongong Overview MARS

More information

Radiation Dosimetry for CT Protocols

Radiation Dosimetry for CT Protocols Radiation Dosimetry for CT Protocols This document contains radiation dosimetry information from CT scans and can be used by investigators to estimate the dosimetry information required by the JRSC or

More information

Dose Metrics in the Patient Report: Myopic or Foresighted?

Dose Metrics in the Patient Report: Myopic or Foresighted? Dose Metrics in the Patient Report: Myopic or Foresighted? BEVERLEY NEWMAN M.B.B.CH FACR Professor Of Radiology, Stanford University, Lucile Packard Children s Hospital, Palo Alto, CA Background: California

More information

Impact of ICRP-89 Based Models on Dose Estimates for Radiopharmaceuticals and CT Exams. Stabin MG, Kost SD, Clark JH, Pickens DR, Price RR, Carver DE

Impact of ICRP-89 Based Models on Dose Estimates for Radiopharmaceuticals and CT Exams. Stabin MG, Kost SD, Clark JH, Pickens DR, Price RR, Carver DE Impact of ICRP-89 Based Models on Dose Estimates for Radiopharmaceuticals and CT Exams Stabin MG, Kost SD, Clark JH, Pickens DR, Price RR, Carver DE Vanderbilt University, Nashville, TN, USA Abstract New

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Journal of Radiology and Imaging

Journal of Radiology and Imaging Journal of Radiology and Imaging An Open Access Publisher http://dx.doi.org/10.14312/2399-8172.2017-2 Original research Open Access Reduction of radiation dose in adult CT-guided musculoskeletal procedures

More information

Review of TG-186 recommendations

Review of TG-186 recommendations Review of TG-186 recommendations Implementation of advanced brachytherapy dose calculation algorithms beyond TG-43 Rowan M. Thomson Carleton Laboratory for Radiotherapy Physics Carleton University Ottawa

More information

Computer based delineation and follow-up multisite abdominal tumors in longitudinal CT studies

Computer based delineation and follow-up multisite abdominal tumors in longitudinal CT studies Research plan submitted for approval as a PhD thesis Submitted by: Refael Vivanti Supervisor: Professor Leo Joskowicz School of Engineering and Computer Science, The Hebrew University of Jerusalem Computer

More information

Using body mass index to estimate individualised patient radiation dose in abdominal computed tomography

Using body mass index to estimate individualised patient radiation dose in abdominal computed tomography O Neill et al. European Radiology Experimental (2018) 2:38 https://doi.org/10.1186/s41747-018-0070-5 European Radiology Experimental ORIGINAL ARTICLE Using body mass index to estimate individualised patient

More information

Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography

Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography The British Journal of Radiology, 81 (2008), 499 503 SHORT COMMUNICATION Modelling the effect of lead and other materials for shielding of the fetus in CT pulmonary angiography 1 G R IBALL, MSc, BSc, 2

More information

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan *M.T.Bahreyni Toossi, **S.Mohandes Dastgherdi Medical Physics Dep., Faculty

More information

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Automated CT Protocol Design Advantages and Pitfalls of Algorithm-Based Technique Selection in Pediatrics. Disclosures 7/22/2014. Learning Objectives

Automated CT Protocol Design Advantages and Pitfalls of Algorithm-Based Technique Selection in Pediatrics. Disclosures 7/22/2014. Learning Objectives Automated CT Protocol Design Advantages and Pitfalls of Algorithm-Based Technique Selection in Pediatrics Robert MacDougall, M.Sc. Department of Radiology Boston Children s Hospital Disclosures 2 Learning

More information

Regional diagnostic reference levels and collective effective doses from CT scanners in India

Regional diagnostic reference levels and collective effective doses from CT scanners in India Regional diagnostic reference levels and collective effective doses from CT scanners in India Roshan S Livingstone and Paul M Dinakaran Department of Radiology Christian Medical College, Vellore, S India

More information

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015 MR QA/QC for MRgRT Rick Layman, PhD, DABR Department of Radiology July 13, 2015 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research Institute

More information

STANDARD OF PRACTICE. Dental CT Scanners CONTENTS. April 2011

STANDARD OF PRACTICE. Dental CT Scanners CONTENTS. April 2011 April 2011 STANDARD OF PRACTICE Approved by Council April 18, 2011 Dental CT Scanners This document is the standard of practice in relation to the use of dental computed tomography (CT) scanners with respect

More information

Tracking Doses in the Pediatric Population

Tracking Doses in the Pediatric Population Tracking Doses in the Pediatric Population Frederic H. Fahey DSc Boston Children s Hospital Harvard Medical School frederic.fahey@childrens.harvard.edu Disclosures Sadly, none that pay me any money! SNMMI

More information

Hi RES Extremity - (04/18/2011) CTDI: ~13 mgy per acquisition Used for evaluation of: Ankle Elbow Hand Wrist Foot /Calcaneous Toes Fingers

Hi RES Extremity - (04/18/2011) CTDI: ~13 mgy per acquisition Used for evaluation of: Ankle Elbow Hand Wrist Foot /Calcaneous Toes Fingers P a g e 1 Hi RES Extremity - (04/18/2011) CTDI: ~13 mgy per acquisition Used for evaluation of: Ankle Elbow Hand Wrist Foot /Calcaneous Toes Fingers Billing: 1. CT Upper/Lower Extremity of concern without

More information

Radiation Exposure 1980 to 2006

Radiation Exposure 1980 to 2006 Radiation Exposure 1980 to 2006 Background 3-6 msv/yr Natural (85% 45%) Radon Cosmic Rays Air travel Living at Altitude Man-made (15% 55%) Medical Imaging** mgy Radiation Therapy cgy Radiation Whole Body

More information

An audit of radiation dose of 4D CT in a radiotherapy department

An audit of radiation dose of 4D CT in a radiotherapy department An audit of radiation dose of 4D CT in a radiotherapy department Poster No.: R-0097 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit T. Hubbard, J. Callahan, J. Cramb, R. Budd, T. Kron;

More information

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling Investigations and research Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling J. Wang Z. Ying V. Yao L. Ciancibello S. Premraj S. Pohlman

More information

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program CT Low Dose Lung Cancer Screening Part I Journey to LDCT LCS Program Paul Johnson, M.S., DABHP, DABR Cleveland Clinic September 26, 2015 Lung Caner is No. 1 In Cancer Related Death In The United States

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

What Have We Learned from Children? Donald P. Frush, MD

What Have We Learned from Children? Donald P. Frush, MD What Have We Learned from Children? Donald P. Frush, MD Medical Radiation and Children: Accomplishments We can lower doses (read noisier exams) We should lower doses Multiphase exams should be justified

More information

Modern Computational Phantoms and Their Applications

Modern Computational Phantoms and Their Applications Modern Computational Phantoms and Their Applications Joint ICRP-RERF-JHPS Workshop Recent Progress on Radiation Dosimetry for Epidemiology and Radiological Protection Sunday, December 2, 2017 Wesley Bolch,

More information

DANIEL J. LONG UNIVERSITY OF FLORIDA

DANIEL J. LONG UNIVERSITY OF FLORIDA MONTE CARLO CALCULATIONS OF PATIENT ORGAN DOSES IN TOSHIBA COMPUTED TOMOGRAPHY EXAMINATIONS WITH AUTOMATIC TUBE CURRENT MODULATION: A FEASIBILITY STUDY By DANIEL J. LONG A DISSERTATION PRESENTED TO THE

More information

Patient Dose in the Diagnosis of PE

Patient Dose in the Diagnosis of PE Patient Dose in the Diagnosis of PE IZAAZ BADSHAH 2018 CANM-CAMRT Joint Annual Meeting March 22-24, 2018 Disclosure I do not have a financial interest, arrangement or affiliation including receipt of honoraria

More information

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems Dental ConeBeam Computed Tomography (CBCT) X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL CONEBEAM COMPUTED TOMOGRAPHY (CBCT) X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability.

More information

Original Article Thoracic Imaging

Original Article Thoracic Imaging Original Article Thoracic Imaging https://doi.org/10.3348/kjr.2018.19.6.1179 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2018;19(6):1179-1186 Size-Specific Dose Estimation In the Korean Lung Cancer

More information

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research

More information

CT Radiation Dosimetry Study Using Monte Carlo Simulation and. Computational Anthropomorphic Phantoms

CT Radiation Dosimetry Study Using Monte Carlo Simulation and. Computational Anthropomorphic Phantoms CT Radiation Dosimetry Study Using Monte Carlo Simulation and Computational Anthropomorphic Phantoms by Yakun Zhang Graduate Program in Medical Physics Duke University Date: Approved: Ehsan Samei, Supervisor

More information

Breast CT and Dosimetry

Breast CT and Dosimetry 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of

More information

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Conventional and spiral CT dose indices in Yazd general hospitals, Iran Iran. J. Radiat. Res., 2006; 3 (4): 183-189 Conventional and spiral CT dose indices in Yazd general hospitals, Iran F. Bouzarjomehri 1*,M.H.Zare 2, D. Shahbazi 2 1 Department of Medical Physics, Shahid

More information

Patient dose assessment of CT perfusion scanning at the RSCH

Patient dose assessment of CT perfusion scanning at the RSCH Patient dose assessment of CT perfusion scanning at the RSCH Lesley Leavesley, Emma Whitehead, Matthew Pryor, Debbie Peet Regional Radiation Protection Service Royal Surrey County Hospital, Guildford Overview

More information

Radiation Safety for New Medical Physics Graduate Students

Radiation Safety for New Medical Physics Graduate Students Radiation Safety for New Medical Physics Graduate Students John Vetter, PhD Medical Physics Department UW School of Medicine & Public Health Background and Purpose of This Training This is intended as

More information