Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Size: px
Start display at page:

Download "Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR."

Transcription

1 Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR.

2 Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of out-of-field doses Second cancers, fetal dose, pacemakers Techniques to assess out-of-field doses Limitations

3 Outline Magnitude of out-of-field doses Impact of out-of-field doses Assessment of out-of-field doses

4 Outline Magnitude of out-of-field doses Impact of out-of-field doses Assessment of out-of-field doses

5 Basics conventional X-rays Sources of out-of-field radiation: Patient scatter, collimator scatter, head leakage. Fraction of Dose Leakage Patient scatter Collimator scatter 0.2 Leakage and Scatter from machine Distance from field edge (cm) Kry et al, 2009, AAPM Scatter from within patient

6 Basics conventional X-rays Doses studied extensively Trends in the data generalizations

7 Basics conventional X-rays Dose decreases ~ exponentially away from edge of field 1% at 10cm with field size Const. with energy Const. with depth Stovall et al. 1995, Medical Physics

8 Basics conventional X-rays Wedges Beam modifiers Physical wedges increase out of field dose by 2-4 times (Sherazi et al, 1985, Int J Radiat Oncol Biol Phys) Dynamic or universal wedges no increase (Li et al, 1997, Int J Radiat Oncol Biol Phys) MLC Secondary MLC no impact on out-of-field dose (Mutic et al, 2002, J Appl Clin Med Phys) Tertiary MLC is extra shielding decrease out of field dose by 30-50% (Stern, 1999, Med Phys)

9 IMRT Near target, dose reduced Treat smaller volume Farther from target, higher doses Modulation increases head leakage Leakage Patient scatter Collimator scatter Fraction of Dose Fraction of Dose Leakage Patient scatter Collimator scatter Distance from field edge (cm) Kry et al, 2009, AAPM Distance from field edge (cm)

10 IMRT Overall: generally higher doses with IMRT Extent of modulation changing Modulation factor decreasing Early 2000s: 3-5 (Followill, 1997, Int J Radiat Oncol Biol Phys. Kry, 2005, Int J Radiat Oncol Biol Phys) Clinically (2010): often <2 Dose (msv) Kry 18 MV IMRT Kry 18 MV Conv Howell 6 MV IMRT Howell 6 MV Conv Distance from central axis (cm) Kry, 2005, Int J Radiat Oncol Biol Phys Howell, 2006, Med Phys

11 High-energy photon therapy Above ~10 MV neutrons Dicentric chromosome induction in human lymphocytes Nolte, 2005 Mutagenesis in mouse fibroblast cell line Hall, 1995 Other measurements have found RBE = 1 NCRP 104 Large uncertainties in neutron RBE

12 High-energy photon therapy This RBE is usually described with: Radiation weighting factor 1.E H = w R D Neutron Fluence (A.U.) 8.E-10 6.E-10 4.E-10 Spectrum wr wr 2.E E+00 1.E-08 1.E-07 1.E-06 1.E-05 1.E-04 1.E-03 1.E-02 1.E-01 1.E+00 1.E+01 E (MeV) ICRP 92 0

13 High-energy photon therapy Neutron dose equivalent with beam energy Howell 2009 Med Phys ~Constant with field size, distance off-axis Dosimetry is difficult varied results in literature Neutrons: 25% of out-offield dose equivalent (18 MV IMRT) 10 MV may be optimal energy for deep tumors (Kry 2005, Int J Radiat Oncol Biol Phys) % of total dose Testes Colon Stomach Lung Thyroid Prostate Liver Esophagus Heart Distance from Central Axis (cm) Kry et al., 2009, Radiother Oncol

14 Cyberknife Other photon modalities Older version 4X higher doses than linac (Petti 2006, Med Phys) Newer versions added shielding Doses more consistent with linacs (Chuang, 2008, Med Phys) Be aware of beams along patient axis Gamma Knife Doses higher near field, lower farther away Zytkovicz, 2007, Phys Med Biol

15 Other photon modalities Tomotherapy (Hi-Art) Many more MUs but extra shielding, no flattening filter Mixed/comparable results relative to IMRT (Ramsey, 2006, J Appl Clin Med Phys) Overall: Some variation between different accelerators All in the same ballpark (factor of 2)

16 Proton therapy Sources of out-of-field dose equivalent: Mostly external neutrons Some internal neutrons A few photons Fontenot, 2008, Phys Med Biol

17 Proton therapy Neutron dose eq. varies with: Treatment energy SOBP and snout size Air gap Small change with distance or field size Zheng, 2007, Phys Med Biol Zheng, 2007, Phys Med Biol Mesoloras, 2006, Med Phys

18 Proton therapy How much dose equivalent is there? Variations in beam parameters Beam energy, SOBP, aperture, air gap Variations in experimental design Size and material of phantom, manufacturer of accelerator Challenges in Dosimetry Lack of high energy response Xu, 2008, Phys Med Biol Unique machines

19 Protons vs. Photons Conventional photon therapy Photons: More dose near treatment field Comparable dose beyond cm from field edge Xu, 2008, Phys Med Biol Stovall, 1995, Med Phys

20 Protons vs. Photons Depending on what proton data you use, you can get very different results: Hall, 2006, Int J Radiat Oncol Biol Phys Zytkovicz, 2007, Phys Med Biol

21 Protons vs. Photons Near to field, dose equivalent much lower with p+ Less lateral scatter Less exit dose Less total out of field dose Effect more pronounced at lower p+ energy Fontenot, 2008, Phys Med Biol. HT/D as a function of lateral distance (along the patient axis) from the isocenter from this work compared to IMRT values collected from Kry et al (2005) and Howell et al (2006).

22 Scanning Proton Beams Much interest in scanning beams No external neutrons Still internal neutrons, gammas Up to half of dose equivalent to near organs Negligible dose to distant organs Scanning beam is an improvement, but is not free from out-offield dose

23 Outline Magnitude of out-of-field doses Impact of out-of-field doses Assessment of out-of-field doses

24 How much dose is there? Out of field doses vary with treatment type/parameters. Typical numbers, (20 cm from field edge, 60 Gy treatment): msv, scanned proton beam msv, passive scatter protons 250 msv, Conventional photons msv, IMRT What do these doses mean?

25 How much is this dose? Put these doses in perspective msv, scanned proton beam msv, passive scatter protons 250 msv, Conventional photons msv, IMRT Courtesy of David Brenner msv

26 Late Effects Radiation has been shown to increase the risk of Cardiovascular disease Diabetes Stroke Hereditary effects Second cancers Potentially relevant for all patients

27 Hereditary effects Historically a major concern Mutant babies Gonad dose considered highest risk Not supported with data Neel Am J Hum Genet 1990, Little Br Med Bull 2003 Gonad dose no longer considered high risk

28 Heart Disease Increased risk with radiation A-bomb survivors Left-sided breast patients Clinically: Minimize dose to the heart

29 Second Cancers Female second cancer incidence. Lifetime cases/100k exposures to 0.1 Gy Cases Stomach Colon Liver Lung Breast Bladder Other Thyroid Leukemia Uterus Ovaries BEIR VII Age at exposure Much higher risk for younger populations

30 Second cancers Cancer takes at least 5 years, typically 10+ years to develop after exposure Risk continues to be elevated 50 years later Absolute Risk for 10+ year survivors: 1 in 70 Brenner Cancer 2000

31 Second Cancers Where do second cancers occur? 12% within geometric field 66% beam-bordering region 22% out-of-field (>5 cm away) Diallo IJROBP 2009 Get most second cancers in high-dose regions This is also the area we re trying to treat, so hard to reduce this dose Want to reduce the dose outside PTV

32 Therapy Symposium Late Effects from Radiation Therapy and Diagnostic Imaging Thursday, 12:30-2:20 Ballroom A

33 Fetal dose Fetus is collection of rapidly dividing cells Very sensitive to radiation damage Sensitivity and damage depend on stage of development

34 Stovall et al, 1995 Fetal dose

35 Fetal dose Important to put these risks into perspective Risk of radiation induced birth defects vs. normal birth defects Stovall et al, 1995

36 Fetal dose Treatment decision will be made by physician (treat, recommend abortion, chemo only, etc.) Physicists must carefully assess dose Impact of this dose may require discussion as well.

37 Fetal dose - shielding Shielding should always be used Even if not necessary Pb is not an effective shield against neutrons Use BPE Shielding is more effective the farther from the treatment field you go ~50% reduction Fraction of Dose 1.0 Leakage Patient scatter 0.8 Collimator scatter Distance from field edge (cm)

38 Pacemakers/Defibrillators Electronics are susceptible to radiation damage Irritation for cameras, CTs etc. Patient safety issue with pacemakers/defibrillators Can be transient effects while beam is on, or can have permanent effects

39 Pacemakers/Defibrillators Inside treatment field is clearly a high dose May be sensitive to relatively low doses: High intensity imaging procedures High LET radiation high energy X-rays/proton therapy Observed failure after 0.9 Gy of neutrons [Raitt Chest 1994]

40 Pacemakers/Defibrillators Ensure there are policies (and follow them) Determine dose to device and compare with device tolerances Monitor functionality of device during/after patient treatment Assess need to move device before treatment TG-34: Management of radiation oncology patients with implanted cardiac pacemakers TG-203: Management of radiotherapy patients with implanted cardiac pacemakers and defibrillators

41 Outline Magnitude of out-of-field doses Impact of out-of-field doses Assessment of out-of-field doses

42 Calculations: Treatment Planning Systems Distance from Field Edge (cm) Dose calc (cgy) Dose meas (cgy) Percent Difference % Howell et al, in preparation % % % This is for a simple, conventional field. As treatment complexity increases, more scatter contributing to dose, less accuracy See this with Mesothelioma cases even within the treatment field we are often off by several percent due to the limitations of the treatment planning system to handle scatter.

43 Calculations - simple TG-36 Peridose software (van der Giessen Ratiother Oncol 2001) Accuracy Pretty good (±30%) Double check with measurement Limitations of treatments Neutrons, IMRT, electrons, brachytherapy

44 Additional source Fetal dose from a large number of procedures, including radiotherapy, brachytherapy, nuclear medicine, imaging

45 Calculations - Detailed Monte Carlo studies Benchmark simulations against measurements Use detailed model Bednarz Med Phys 2008

46 Measurements Ion chamber measurements

47 Measurements In vivo measurements Taken directly on/in patient Caution: surface dose is elevated (4x)!! Cover detector with bolus TG-36 Kry Med Phys 2006

48 Additional TLD comments TLD very nice dosimeter TLD-100 over-responds to neutrons Factor of 10 in reported dose for 18 MV 2.3 mgy reported as 27.0 mgy Kry JACMP 2007 Do not use TLD-100 for out-of-field measurements for E>10 MV (OK inside treatment field) Use TLD-700, ion chamber

49 Other considerations Energy spectrum is much different outside the treatment field Energy dependent dosimeters will have a different response TLD, diodes. Even ion chambers

50 Perspective and final thoughts Out of field doses are not negligible Risk is justified Must be cognizant of, and work to minimize, stray radiation Be aware of dose assessment pitfalls Approach to assess out-of-field dose should be based on the required accuracy of the information.

51 Thank you!

Outline. Outline 3/30/12. Second Cancers from. Radiotherapy Procedures. Stephen F. Kry, Ph.D., D.ABR.

Outline. Outline 3/30/12. Second Cancers from. Radiotherapy Procedures. Stephen F. Kry, Ph.D., D.ABR. Second Cancers from Radiotherapy Procedures Stephen F. Kry, Ph.D., D.ABR. Outline Radiation and cancer induction Medically exposed people Estimating risk of second cancers Minimizing the risk Outline Radiation

More information

Radiation Related Second Cancers. Stephen F. Kry, Ph.D., D.ABR.

Radiation Related Second Cancers. Stephen F. Kry, Ph.D., D.ABR. Radiation Related Second Cancers Stephen F. Kry, Ph.D., D.ABR. Objectives Radiation is a well known carcinogen Atomic bomb survivors Accidental exposure Occupational exposure Medically exposed Radiotherapy

More information

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS 6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy

AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy Stephen F. Kry a) Department of Radiation Physics, MD Anderson Cancer Center, Houston,

More information

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS PAPER EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS Isra Israngkul Na Ayuthaya *, Sivalee Suriyapee, Phongpheath Pengvanich

More information

Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy

Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy Susannah Lazar David Followill, Ph.D. John Gibbons, Ph.D. * Anita Mahajan, M.D. Mohammad Salehpour, Ph.D. Marilyn Stovall, Ph.D. R. Allen White,

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

8/2/2017. Acknowledgement. Disclaimer. How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations

8/2/2017. Acknowledgement. Disclaimer. How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations Moyed Miften, Ph.D. Department of Radiation Oncology University of Colorado School of Medicine Disclaimer

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Dosimetry for Epidemiology Cohorts Who Receive Radiation Therapy

Dosimetry for Epidemiology Cohorts Who Receive Radiation Therapy Dosimetry for Epidemiology Cohorts Who Receive Radiation Therapy Wayne Newhauser, PhD Eurados Winter School, Milan, 2016 Introduction About 1 in 2 men and women born today will be diagnosed with some form

More information

Acknowledgments. Introduction. Managing the Care of Radiotherapy Patients Implanted with Cardiac Devices. Outline

Acknowledgments. Introduction. Managing the Care of Radiotherapy Patients Implanted with Cardiac Devices. Outline Managing the Care of Radiotherapy Patients Implanted with Cardiac Devices Acknowledgments Members of AAPM TG203 Joann Prisciandaro Moyed Miften, PhD Dimitris Mihailidis, PhD Professor and Chief Physicist

More information

Radiation qualities in carbon-ion radiotherapy at NIRS/HIMAC

Radiation qualities in carbon-ion radiotherapy at NIRS/HIMAC Radiation qualities in carbon-ion radiotherapy at NIRS/ Shunsuke YONAI Radiological Protection Section Research Center for Charged Particle Therapy National Institute of Radiological Sciences (NIRS) E-mail:

More information

Radiation Protection: Are We Doing Enough To Protect Our Patients and Staff?

Radiation Protection: Are We Doing Enough To Protect Our Patients and Staff? Radiation Protection: Are We Doing Enough To Protect Our Patients and Staff? Wayne Newhauser, Ph. D. Univiversity Texas M. D. Anderson Cancer Center Proton Therapy Course, 2008-05-21, Jacksonville Acknowledgments

More information

Neutron dose evaluation in radiotherapy

Neutron dose evaluation in radiotherapy Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

Risk of a second cancer after radiotherapy

Risk of a second cancer after radiotherapy Risk of a second cancer after radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Medical radiological procedures worldwide 2.5 billion diagnostic radiological examinations 78%

More information

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Poster No.: C-0096 Congress: ECR 2012 Type: Scientific Paper Authors: M. Mazonakis,

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Out-of-field Radiation Risks in Paediatric Proton Therapy

Out-of-field Radiation Risks in Paediatric Proton Therapy Out-of-field Radiation Risks in Paediatric Proton Therapy Charlot Vandevoorde NRF ithemba LABS Contact: cvandevoorde@tlabs.ac.za Seventh NCS Lustrum Proton Therapy Amsterdam Charlot Vandevoorde 27 October

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

Radiation treatment planning and delivery strategies for a pregnant brain tumor patient

Radiation treatment planning and delivery strategies for a pregnant brain tumor patient Received: 19 October 2017 Revised: 7 December 2017 Accepted: 18 December 2017 DOI: 10.1002/acm2.12262 RADIATION ONCOLOGY PHYSICS Radiation treatment planning and delivery strategies for a pregnant brain

More information

On the use of bolus for pacemaker dose measurement and reduction in radiation therapy

On the use of bolus for pacemaker dose measurement and reduction in radiation therapy Received: 8 February 2017 Revised: 16 October 2017 Accepted: 23 October 2017 DOI: 10.1002/acm2.12229 RADIATION ONCOLOGY PHYSICS On the use of bolus for pacemaker dose measurement and reduction in radiation

More information

Spot scanning proton therapy minimizes neutron dose in the setting of radiation therapy administered during pregnancy

Spot scanning proton therapy minimizes neutron dose in the setting of radiation therapy administered during pregnancy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Spot scanning proton therapy minimizes neutron dose in the setting of radiation therapy administered during pregnancy Xin Wang, 1a

More information

Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield

Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield Amir M. Owrangi, Donald A. Roberts,

More information

Predicting Maximum Pacemaker/ICD Dose in SAVI HDR Brachytherapy

Predicting Maximum Pacemaker/ICD Dose in SAVI HDR Brachytherapy Predicting Maximum Pacemaker/ICD Dose in SAVI HDR Brachytherapy Chaitanya Kalavagunta Ph.D. University of Maryland School of Medicine, Baltimore, Maryland. Background Aging population - Number of patients

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

Neutron Measurements for Intensity Modulated Radiation Therapy

Neutron Measurements for Intensity Modulated Radiation Therapy SLAC-PUB-8443 April 2 Neutron Measurements for Intensity Modulated Radiation Therapy N. E. Ipe et al. Presented at Chicago 2 World Congress on Medical Physics and Biomedical Engineering, 7/23/2 7/28/2,

More information

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction

More information

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring

More information

A Monte Carlo Study for Photoneutron Dose Estimations around the High-Energy Linacs

A Monte Carlo Study for Photoneutron Dose Estimations around the High-Energy Linacs www.jbpe.org A Monte Carlo Study for Photoneutron Dose Estimations around the High-Energy Linacs Original 1 Physics Department, Faculty of Sciences, Ferdowsi University of Mashhad, Iran Mohammadi N 1,

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

Intensity Modulated RadioTherapy

Intensity Modulated RadioTherapy Intensity Modulated RadioTherapy A clinical application of a small accelerator University Medical Center Groningen A.A. van t Veld PhD I. Hoveijn PhD part 1 part2 CERN/KVI accelerator school, Zeegse, June

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

RESEARCH ARTICLE. Vasanthan Sakthivel 1,3 *, Ganesh Kadirampatti Mani 1,2, Sunil Mani 3, Raghavendiran Boopathy 3. Abstract.

RESEARCH ARTICLE. Vasanthan Sakthivel 1,3 *, Ganesh Kadirampatti Mani 1,2, Sunil Mani 3, Raghavendiran Boopathy 3. Abstract. DOI:10.22034/APJCP.2017.18.7.1897 RESEARCH ARTICLE Radiation-Induced Second Cancer Risk from External Beam Photon Radiotherapy for Head and Neck Cancer: Impact on in-field and Out-of-Field Organs Vasanthan

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the

Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the PMMA ring holder with 7 containers are also shown. Relative

More information

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics Neutron Interactions Part 2 George Starkschall, Ph.D. Department of Radiation Physics Neutron shielding Fast neutrons Slow down rapidly by scatter in hydrogenous materials, e.g., polyethylene, paraffin,

More information

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

Radiation Safety for New Medical Physics Graduate Students

Radiation Safety for New Medical Physics Graduate Students Radiation Safety for New Medical Physics Graduate Students John Vetter, PhD Medical Physics Department UW School of Medicine & Public Health Background and Purpose of This Training This is intended as

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

[Setawati et. al., Vol.4 (Iss10): October, 2017] ISSN: DOI: /zenodo

[Setawati et. al., Vol.4 (Iss10): October, 2017] ISSN: DOI: /zenodo EXTERNAL RADIATION SIMULATION OF LINAC TO DETERMINE EFFECTIVE DOSE IN ORGANS USING MONTE CARLO METHOD Evi Setawati *1, Muchammad Azam 1, Ngurah Ayu Ketut Umiati 1, Hammam Oktajianto 1 *1 Physics Department,

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR Therapeutic Medical Physics Stephen J. Amadon Jr., Ph.D., DABR Outline 1. Why physicists are needed in medicine 2. Branches of medical physics 3. Physics in Radiation Oncology 4. Treatment types and Treatment

More information

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser. Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

Outlines. An Institutional Experience Managing the Care of Patients with CIEDs

Outlines. An Institutional Experience Managing the Care of Patients with CIEDs An Institutional Experience Managing the Care of Patients with CIEDs Joann I. Prisciandaro, Ph.D. The Department of Radiation Oncology University of Michigan Outlines Discuss functionality and components

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Proton Treatment. Keith Brown, Ph.D., CHP. Associate Director, Radiation Safety University of Pennsylvania

Proton Treatment. Keith Brown, Ph.D., CHP. Associate Director, Radiation Safety University of Pennsylvania Proton Treatment Keith Brown, Ph.D., CHP Associate Director, Radiation Safety University of Pennsylvania Proton Dose vs. Depth Wilson,. R.R. Radiological use of fast protons. Radiology 47:487-491, 1946.

More information

Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning

Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca

More information

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER Bro. Dr. Collie Miller IARC/WHO Based on trends in the incidence of cancer, the International Agency for Research on Cancer (IARC) and WHO

More information

Treatment Planning (Protons vs. Photons)

Treatment Planning (Protons vs. Photons) Treatment Planning Treatment Planning (Protons vs. Photons) Acquisition of imaging data Delineation of regions of interest Selection of beam directions Dose calculation Optimization of the plan Hounsfield

More information

Today, I will present the second of the two lectures on neutron interactions.

Today, I will present the second of the two lectures on neutron interactions. Today, I will present the second of the two lectures on neutron interactions. 1 The main goal of this lecture is to tell you a little about clinical neutron therapy, first with fast neutron beams, and

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters 133 Clinical Radiation Oncology Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters Bhaskar Mukherjee 1,2, Carolina Lina 1 Vladimir Mares 3 1Westdeutsches

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

Estimating Risks from CT Scans - in the Context of CT Scan Benefits

Estimating Risks from CT Scans - in the Context of CT Scan Benefits Estimating Risks from CT Scans - in the Context of CT Scan Benefits David J. Brenner Center for Radiological Research Columbia University Medical Center djb3@cumc.columbia.edu There is no question that

More information

Additional Questions for Review 2D & 3D

Additional Questions for Review 2D & 3D Additional Questions for Review 2D & 3D 1. For a 4-field box technique, which of the following will deliver the lowest dose to the femoral heads? a. 100 SSD, equal dmax dose to all fields b. 100 SSD, equal

More information

Future upcoming technologies and what audit needs to address

Future upcoming technologies and what audit needs to address Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

Normal tissue dose in pediatric VMAT Piotr Zygmanski

Normal tissue dose in pediatric VMAT Piotr Zygmanski Normal tissue dose in pediatric VMAT Piotr Zygmanski Department of Radiation Oncology Brigham and Women s Hospital Radiation induced effects in children Risk of inducing 2 nd cancers higher than for adults

More information

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 IROC Prostate Phantom Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 The study groups are requesting that each institution keep the phantom for a period of time

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence SCIENCE & TECHNOLOGY To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence Alburuj R. Rahman*, Jian Z. Wang, Dr. Z. Huang, Dr. J. Montebello Department of

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output Presented by: Rebecca M. Howell PhD, DABR, FAAPM Results from Voluntary Independent External Peer Review of Beam Output What do we hope to achieve with independent peer review? Independent peer review

More information

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST Time : p.m p.m.

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST Time : p.m p.m. POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST 2015 Date :- 24 th August 2015 PAPER I Time :- 2.00 p.m. - 4.15 p.m. If the examiners cannot

More information

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,

More information

HEALTH RISKS FROM EXPOSURE TO FAST NEUTRONS. Prof. Marco Durante

HEALTH RISKS FROM EXPOSURE TO FAST NEUTRONS. Prof. Marco Durante HEALTH RISKS FROM EXPOSURE TO FAST NEUTRONS Prof. Marco Durante Risk from neutrons Risk from exposure to fission spectrum neutrons has been extensively studied in the 60 s at nuclear reactors using animal

More information

Hampton University Proton Therapy Institute

Hampton University Proton Therapy Institute Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI

More information

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad

More information

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS} High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

Managing the imaging dose during image-guided radiation therapy

Managing the imaging dose during image-guided radiation therapy Managing the imaging dose during image-guided radiation therapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Richmond VA Imaging during radiotherapy Radiographic

More information

Proton Stereotactic Radiotherapy: Clinical Overview. Brian Winey, Ph.D. Physicist, MGH Assistant Professor, HMS

Proton Stereotactic Radiotherapy: Clinical Overview. Brian Winey, Ph.D. Physicist, MGH Assistant Professor, HMS Proton Stereotactic Radiotherapy: Clinical Overview Brian Winey, Ph.D. Physicist, MGH Assistant Professor, HMS Acknowledgements Radiation Oncologists and Physicists at various institutions (MGH, MDACC,

More information

Evaluation of The Second Cancer s Risk In Conformal Therapy And Intensity Modulated Radiotherapy For The Organs Inside The Primary Radiation Fields

Evaluation of The Second Cancer s Risk In Conformal Therapy And Intensity Modulated Radiotherapy For The Organs Inside The Primary Radiation Fields IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 8-4861.Volume 8, Issue Ver. I (Mar. - Apr. 016), PP 44-5 www.iosrjournals.org Evaluation of The Second Cancer s Risk In Conformal Therapy And Intensity

More information

Estimation of Effective Doses for Radiation Cancer Risks on ISS, Lunar and Mars Missions with Space Radiation Measurements

Estimation of Effective Doses for Radiation Cancer Risks on ISS, Lunar and Mars Missions with Space Radiation Measurements Estimation of Effective Doses for Radiation Cancer Risks on ISS, Lunar and Mars Missions with Space Radiation Measurements Myung-Hee Y. Kim Wyle Laboratories, Houston, Texas, 77058 and Francis A. Cucinotta

More information