[ Care and Use Manual ]

Size: px
Start display at page:

Download "[ Care and Use Manual ]"

Transcription

1 MALDI Calibration Kit I. Introduction The MALDI Calibration Kit is a conveniently packaged selection of MALDI matrices and calibration standards (includes high-purity Neg Ion Mode Calibrant for accurate negative ion mass calibrations) qualified for MALDI-MS applications, and is applicable for calibration and testing of MALDI- MS instrumentation. The MALDI matrices are ultra purified matrices that aid in optimizing desorption/ionization of analytes from a target surface for demanding MALDI applications. Each matrix is prepared under strict quality control. The prepackaged calibration standards eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample preparation for TOF mass spectrometry. CONTENTS I. INTRODUCTION II. MALDI MATRICES b. Recommended Usage of MALDI Matrices III. CALIBRATION STANDARDS b. Preparation of Calibration Standards c. Calibration Mixture DIOS Low Mass d. Waters Calibration Mix MALDI Reflectron e. ACTH Standard f. Neg Ion Mode Calibrant IV. ORDERING INFORMATION

2 II. MALDI MATRICES The matrices are conveniently packaged in 1.5 ml tubes (10 mg of matrix pre-weighed in each tube) for ease of use. Tubes are vacuumsealed in foil pouches for convenient storage and protection from light and air. It is recommended that the matrices be stored at 4 C. After opening, the matrix may be used for 1 month if stored in the dark at 4 C. Once reconstituted in appropriate solvents, the matrix solution is stable for 3 days when stored in the dark at 4 C. b. Recommended Usage of MALDI Matrices The following methods are provided as a general guideline for matrix preparation on standard stainless steel targets. The described method only serves as a starting point. Depending on the application, other solvents and/or methods may give improved results. 1. For optimum performance, use solvents of the highest purity commercially available. CH 3 : Acetonitrile (J. T. Baker Cat. No.: ) C 2 H 5 OH: Ethanol (Aldrich Cat. No.: 45,984-4) TFA: Trifluoroacetic acid (Fisher, Cat. No: O ) O: Deionized (18.2 MW) water To prepare 0.1% TFA aqueous solution, add 30 µl of TFA to 30 ml of deionized water. 2. Each vial contains 10 mg of matrix material. Depending on the application (peptides, proteins, oligonucleotides, oligosaccharides, etc.), prepare matrix solution by adding appropriate solvent(s) into the matrix vial. 3. Vortex the vial for several seconds to ensure that the solid matrix is completely dissolved. 4. Mix equal volumes of sample solution with the prepared matrix solution. 5. Apply 1 µl onto the MALDI target. 6. Dry at room temperature before submitting target for MALDI analysis. Table 1: Recommended Matrix Solvents and Volumes Matrix Suggested Analytes Solvents and Volumes Concentration Alpha-Cyano-4-Hydroxycinnamic Acid (CHCA) Peptides 495 ml: 495 ml: 10 ml C 2 H 5 OH:CH 3 :0.1% TFA 10 mg/ml Sinapinic Acid Proteins 600 ml: 400 ml 0.1% TFA: CH 3 10 mg/ml 2,5-Dihydroxybenzoic Acid (DHB) Oligosaccharides 100 ml: 400 ml O: CH 3 20 mg/ml 3-Hydroxypicolinic Acid (HPA) Oligonucleotides 200 µl: 200 ml O : CH 3 25 mg/ml 2,4,6-Trihydroxyaceto-phenone (THAP) Oligonucleotides 200 µl: 200 ml O: CH 3 25 mg/ml 2

3 III. CALIBRATION STANDARDS The calibration mixtures are packaged as lyophilized solids in glass vials. The vials are vacuum-sealed in foil pouches to ensure protection from light and air. It is recommended that all products remain in the foil pouch (in the dark) at room temperature until use. Once reconstituted in appropriate solvents, the solutions are stable when stored at 4 C for the length of time indicated in table below. Stability can be extended by freezing samples at -20 C. b. Preparation of Calibration Standards The following methods are provided as a general guideline for preparation of the standards. The described method serves only as a starting point. Depending on your application, one may consider using other solvents and/or methods. Additionally, the selection of matrix should match that of samples to be analyzed. Optimization for specific applications is recommended. Other relative ratios and solvents may be used for reconstitution. In some cases further dilutions of this solution may yield better spectra depending on the instrument and substrate used. Upon applying 1µl of sample mixture to the target, allow the target to completely dry before submitting it to MALDI analysis. Table 2: Calibration Standards Intended Use and Mass Range PN Product Description Calibration Mix - MALDI Reflectron Calibration Mix - DIOS Low Mass ACTH Standard Neg Ion Mode Calibrant Intended Use Calibration of MALDI TOF for positive ion reflectron mode detection Calibration of MALDI TOF for small molecule detection in positive reflectron mode External calibration of MALDI TOF in reflectron or linear mode Calibration of MALDI TOF for negative ion reflectron mode detection m/z range Table 3: Calibration Standards Sample Preparation Methods PN Product Description Calibration Mix - MALDI Reflectron Calibration Mix - DIOS Low Mass ACTH Standard Neg Ion Mode Calibrant Volume Diluent added to lyophilized calibrant (µl)* Mixed 1:1 with matrix 250 Yes Target Plate Stainless Steel target 250 No DIOS-target Yes Yes Stainless Steel target Stainless Steel target * It is recommended as a starting point to use a 1:1 mixture of CH 3 : O as the sample diluent. For optimum performance, solvents of the highest purity commercially available are recommended. The following solvents were found to be satisfactory. CH 3 : Acetonitrile (J. T. Baker Cat. No.: ) O: High purity (18 MW) water c. Calibration Mixture DIOS Low Mass Waters Calibration Mix DIOS Low Mass contains polyethylene glycol (PEG) 200, 600, and 1000 with sodium iodide, rubidium iodide and cesium iodide (NaI, RbI and CsI). This mixture has not been optimized for use in electrospray systems. The following spectrum of the Calibration Mix DIOS Low Mass was processed by baseline subtraction, smoothing, and centroiding in preparation for calibration. Data was obtained on a Waters MALDI micro MX. Note that the Calibration Mix is relatively concentrated and tends to leave a visible, shiny residue on the surface. Adjust your instrument laser power so that the signal does not saturate the detector. 3

4 Table 4: MassPREP Calibration Standards Storage Table 5: Peak List for Calibration Mix DIOS Low Mass PN Product Description Solution Storage Calibration Mix - MALDI Reflectron Calibration Mix - DIOS Low Mass 1 year 1 year ACTH Standard 1 week Neg Ion Mode Calibrant 1 year at -20 C Figure 1: MALDI MS spectrum of the Calibration Mix DIOS Low Mass Peak List for Calibration Mix DIOS Low Mass The following reference file contains the masses of selected low mass clusters, which are formed by the mixture of sodium, rubidium, and cesium iodide as well as the sodium (and some rubidium) adducts of polyethylene glycol oligomers with chemical formula HO[C 2 H 4 O]nH + Na (where n = 1,2,3,...) between 200 and 1500 Da. Masses were calculated using the following values (where M (x) indicates the mass, M, of the chemical species, x, within the parentheses): M(C) = M(H) = M(N) = M(O) = M(Na) = M(K) = M(Rb) = M(I) = M(Cs) = Below m/z 150 the sodium, rubidium, and cesium ions provide the more dominant s. Above m/z 200 the sodiated PEG s provide the more dominant s. 4 Mass Identity Comment Mass Identity Comment Na Rb - Rb - isotope Cs sometimes weak PEG + Na PEG + Rb PEG + Na PEG + Rb PEG + Na very weak PEG + Na PEG + Na weak PEG + Rb PEG + Rb weak PEG + Na PEG + Na main PEG + Rb PEG + Rb main PEG + Na PEG + Na PEG + Rb PEG + Rb PEG + Na PEG + Na PEG + Rb PEG + Rb PEG + Na PEG + Na PEG + Rb PEG + Rb Mass Identity Comment Mass Identity Comment PEG + Na PEG + Rb PEG + Rb PEG + Na PEG + Na PEG + Rb PEG + Rb PEG + Na PEG + Na PEG + Rb PEG + Rb PEG + Na PEG + Na PEG + Rb PEG + Rb PEG + Na PEG + Na PEG + Rb PEG + Rb PEG + Na PEG + Na PEG + Rb

5 d. Waters Calibration Mix MALDI Reflectron Waters Calibration Mix MALDI Reflectron contains polyethylene glycol (PEG)1000, 2000, and 3000 with sodium iodide (NaI) and provides many uniformly-spaced s. The following spectrum of the Calibration Mix MALDI Reflectron mixed with MALDI Matrix - CHCA was processed by baseline subtraction, smoothing, and centroiding in preparation for calibration. Data was obtained on a Waters MALDI micro MX. Note that the Calibration Mix is relatively concentrated and tends to leave a visible, shiny residue on the surface. Adjust your instrument laser power so that the signal does not saturate the detector. Figure 2: MALDI MS Spectrum of the Calibration Mix MALDI Reflectron Peak List for Calibration Mix MALDI Reflectron The following reference file contains the masses of the sodium adducts of polyethylene glycol oligomers with chemical formula HO[C 2 H 4 O]nH + Na (where n = 1,2,3,...) between 565 and 4005 Da. Masses were calculated using the following values (where M (x) indicates the mass, M, of the chemical species, x, within the parentheses): M(H) = M(O) = M(C) = M(Na) = Table 6: Peak List for Calibration Mix MALDI Reflectron Sodiated PEG = M(HO[C 2 H 4 O]nH + Na) n Sodiated PEG =

6 e. ACTH Standard Waters ACTH standard is Adrenocorticotropic Hormone Fragment human, and provides a monoisotopic calibration at (M+H) +. Each vial contains 1nmol of lyophilized ACTH. The spectrum of the ACTH Standard seen below was processed by baseline subtraction and smoothing. The matrix used was CHCA (10mg/mL in 0.1% TFA/90% A/9.9% O). Data was obtained on a Waters MALDI micro MX. Peak List for Neg Ion Mode Calibrant The following reference file contains the masses of the 2-AA Dextran between 789m/z and 4027m/z. Table 7: Peak List for Neg Ion Mode Calibrant Number of hexose, n= M-H (m/z) Number of hexose, n= M-H (m/z) Figure 3: MALDI MS Spectrum of the ACTH Standard IV. ORDERING INFORMATION f. Neg Ion Mode Calibrant Waters Neg Ion Mode Calibrant provides polymeric s of 2-AA Dextran. The following spectrum was acquired on a Waters MALDI micro MX using the recommended procedure, including DHB matrix (typically mg/ml of DHB in solution with high organic content and 5-10 mm ammonium citrate). Adjust the laser power by aiming the laser at a variety of positions until adequate ionization is achieved. Table 8: MALDI Calibration Kit Product Description Part Number m/z range Units/kit Amount/Unit MALDI Calibration Kit MALDI Matrix CHCA MALDI Matrix DHB ACTH Neg Ion Mode Calibrant Calibration Mix R Calibration Mix Low Mass mg mg nmol 1 50 µg Figure 4: MALDI MS Spectrum of the Neg Ion Mode Calibrant 6

7 Sales Offices: Austria and European Export (Central South Eastern Europe, CIS and Middle East) Australia Belgium Brazil Canada x2205 China CIS/Russia Czech Republic Denmark Finland France Germany Hong Kong Hungary India and India Subcontinent Ireland Italy Japan Korea Mexico The Netherlands Norway Poland Puerto Rico Singapore Spain Sweden Switzerland Taiwan United Kingdom Waters, MALDI micro MX and MassPREP are trademarks of Waters Corporation Waters Corporation Rev B, 11/07 VW-PDF Waters Corporation 34 Maple Street Milford, MA U.S.A. T: F:

PosterREPRINT INTRODUCTION. 2-D PAGE of Mouse Liver Samples. 2-D PAGE of E.coli Samples. Digestion / Cleanup. EXPERIMENTAL 1-D PAGE of BSA Samples

PosterREPRINT INTRODUCTION. 2-D PAGE of Mouse Liver Samples. 2-D PAGE of E.coli Samples. Digestion / Cleanup. EXPERIMENTAL 1-D PAGE of BSA Samples INTRODUCTION Identification and characterization of low abundance proteins separated by 2D gel electrophoresis is complicated by two important factors; the use of suitable staining techniques for the visualization

More information

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis Providing a Universal, ne-step Alternative to Liquid-Liquid Extraction in Bioanalysis Jessalynn P. Wheaton, Erin E. Chambers, and Kenneth J. Fountain APPLICATIN BENEFITS n Simple, one-step sample preparation

More information

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT. Objective The aim of the study presented is to determine the presence of target Gingkolides in Gingko biloba leaf extract. The data presented shows Gingkolides determined to be present in the extract analyzed.

More information

Improving Food Process Control Using the

Improving Food Process Control Using the Improving Food Process Control Using the UPLC Amino Acid Analysis Solution Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION Food products have different desirability and value based in

More information

ApplicationNOTE A MULTI-RESIDUE LC/MS/MS METHOD FOR THE DETERMINATION OF 81 PESTICIDE RESIDUES IN FRUIT AND VEGETABLES: PART 1, METHOD OVERVIEW

ApplicationNOTE A MULTI-RESIDUE LC/MS/MS METHOD FOR THE DETERMINATION OF 81 PESTICIDE RESIDUES IN FRUIT AND VEGETABLES: PART 1, METHOD OVERVIEW ApplicationNOTE A MULTI-RESIDUE LC/MS/MS METHOD FOR THE DETERMINATION OF 81 PESTICIDE RESIDUES IN FRUIT AND VEGETABLES: PART 1, METHOD OVERVIEW Gordon Kearney1, Lutz Alder2, Anthony Newton1, Jeannette

More information

A New Era in SFC Column Technology

A New Era in SFC Column Technology B TM A ew Era in SFC Column Technology Combining state-of-the-art media manufacturing with the industry leading column technology, Viridis Supercritical Fluid Chromatography (SFC) columns bring a new level

More information

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link. Fully-Automated HbA 1c Testing

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link. Fully-Automated HbA 1c Testing Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c VARIANT II TURBO Link Fully-Automated HbA 1c Testing Bio-Rad Laboratories HEMOGLOBIN Testing VARIANT II TURBO Link Efficient streamlined process Powerful

More information

The first and only fully-automated, multiplexed solution for Measles, Mumps, Rubella and Varicella-zoster virus antibody testing

The first and only fully-automated, multiplexed solution for Measles, Mumps, Rubella and Varicella-zoster virus antibody testing Bio-Rad Laboratories BioPlex 2200 System BioPlex 2200 MMRV IgG Kit The first and only fully-automated, multiplexed solution for Measles, Mumps, Rubella and Varicella-zoster virus antibody testing Bio-Rad

More information

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis The Investigation of Factors Contributing to Immunosuppressant Drugs Variability in LC-MS/MS Analysis Joseph Herman, Dayana Argoti, and Sarah Fair Thermo Fisher Scientific, Franklin, MA, USA Overview Purpose:

More information

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link System. Fully-Automated HbA 1c Testing

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link System. Fully-Automated HbA 1c Testing Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c VARIANT II TURBO Link System Fully-Automated HbA 1c Testing Bio-Rad Laboratories HEMOGLOBIN Testing VARIANT II TURBO Link System Efficient streamlined

More information

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring and in Urine Xiaolei Xie, Joe DiBussolo, Marta Kozak; Thermo Fisher Scientific, San Jose, CA Application Note 627 Key Words, norbuprenorphine,

More information

PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX

PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX E. Claude 1, E. Emekwue 2, M. Snel 1, T. McKenna 1 and J. Langridge 1 1: Waters Corporation, Manchester, UK 2:

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Information Microwave-assisted Kochetkov amination followed by permanent charge derivatization: A facile strategy for glycomics Xin Liu a,b, Guisen Zhang* a, Kenneth Chan b and

More information

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Application Note: 346 MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Gargi Choudhary and Diane Cho, Thermo Fisher Scientific, San Jose, CA Wayne Skinner and

More information

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Measuring Phytosterols in Health Supplements by LC/MS Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA Overview Purpose: Develop a method for the extraction of phytosterols

More information

Ionization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s)

Ionization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s) Ionization Methods Neutral species Charged species Removal/addition of electron(s) M + e - (M +. )* + 2e - electron ionization Removal/addition of proton(s) M + (Matrix)-H MH + + (Matrix) - chemical ionization

More information

Infectious Disease Testing. ULTRA Product Line. Safety is not a Matter of Chance

Infectious Disease Testing. ULTRA Product Line. Safety is not a Matter of Chance Infectious Disease Testing ULTRA Product Line Safety is not a Matter of Chance ULTRA Product Line The best answer for HBV, HCV and HIV screening: a global automated solution for safe results. Monolisa

More information

BioPlex 2200 Infectious Disease Panels

BioPlex 2200 Infectious Disease Panels BioPlex 2200 System BioPlex 2200 Infectious Disease Panels An Expanding Multiplexed Assay Menu Lyme HIV Ag-Ab MMV IgM Syphilis Total & RPR MMRV EBV HSV-1 & HSV-2 EBV IgM ToRC IgM ToRC Leading the way with

More information

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis X. Liu, C. Pohl, Dionex Corporation, Sunnyvale, CA, USA ABSTRACT Although reversed-phase (RP) silica columns (e.g., C18 and

More information

Biological Mass Spectrometry. April 30, 2014

Biological Mass Spectrometry. April 30, 2014 Biological Mass Spectrometry April 30, 2014 Mass Spectrometry Has become the method of choice for precise protein and nucleic acid mass determination in a very wide mass range peptide and nucleotide sequencing

More information

TGX and TGX Stain-Free FastCast Acrylamide Kits

TGX and TGX Stain-Free FastCast Acrylamide Kits Electrophoresis TGX and TGX Stain-Free FastCast Acrylamide Kits The newest frontier for handcast gels It s Time to Expect More from Handcast Gels! FastCast kits provide unparalleled convenience and performance.

More information

PosterREPRINT A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION. Presented at ABRF 2002, Austin, Texas, USA, 9th - 12th March 2002.

PosterREPRINT A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION. Presented at ABRF 2002, Austin, Texas, USA, 9th - 12th March 2002. Introduction A NOVEL APPROACH TO MALDI-TOF-MS SAMPLE PREPARATION Ed Bouvier 2, Jeff Brown 1, Emmanuelle Claude 1, John L. Gebler 2, Weibin Chen 2, *Dominic Gostick 1, Kevin Howes 1, James Langridge 1,

More information

The study of phospholipids in single cells using an integrated microfluidic device

The study of phospholipids in single cells using an integrated microfluidic device Supporting Information: The study of phospholipids in single cells using an integrated microfluidic device combined with matrix-assisted laser desorption/ionization mass spectrometry Weiyi Xie,, Dan Gao,

More information

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c Bio-Rad Laboratories Hemoglobin Testing VARIANT II Hemoglobin Testing System For HbA 1c Bio-Rad CDM System CDM 5.1 VII Instrument Patient Report V2_A1c_NU Patient Data Sample ID: AD Patient ID: Name: Physician:

More information

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Jenny Chen, Hongxia Wang, Zhiqi Hao, Patrick Bennett, and Greg Kilby Thermo Fisher

More information

Matrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry

Matrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry Matrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry Time-of-Flight Mass Spectrometry. Basic principles An attractive feature of the time-of-flight (TOF) mass spectrometer is its

More information

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS New Solvent Grade Targeted for Trace Analysis by UHPLC-MS Subhra Bhattacharya, Deva H. Puranam, and Stephen C. Roemer Thermo Fisher Scientific Fisher Chemical, One Reagent Lane, Fair Lawn, NJ Material

More information

[application note] DIRECT TISSUE IMAGING AND CHARACTERIZATION OF PHOSPHOLIPIDS USING A MALDI SYNAPT HDMS SYSTEM

[application note] DIRECT TISSUE IMAGING AND CHARACTERIZATION OF PHOSPHOLIPIDS USING A MALDI SYNAPT HDMS SYSTEM DIRECT TISSUE IMAGING AND CHARACTERIZATION OF PHOSPHOLIPIDS USING A MALDI SYNAPT HDMS SYSTEM Emmanuelle Claude, Marten Snel, Thérèse McKenna, and James Langridge INTRODUCTION The last decade has seen a

More information

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase Subhra Bhattacharya and Stephen C. Roemer Thermo Fisher Scientific Global Chemicals, Fair Lawn, NJ, USA AbstrAct Liquid

More information

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine High-Throughput Quantitative LC-MS/MS Analysis of and 14 Benzodiazepines in Urine Bill Yu, Kristine Van Natta, Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 588 Key Words Opiates,

More information

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS New Solvent Grade Targeted for Trace Analysis by UHPLC-MS Subhra Bhattacharya, Deva H. Puranam, and Stephen C. Roemer Thermo Fisher Scientific Fisher Chemical, One Reagent Lane, Fair Lawn, NJ Material

More information

Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS

Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS Bruker Daltonics Application Note # MT-93/FTMS-38 Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS Introduction Matrix Assisted Laser Desorption Ionization (MALDI)

More information

TECHNICAL BULLETIN. R 2 GlcNAcβ1 4GlcNAcβ1 Asn

TECHNICAL BULLETIN. R 2 GlcNAcβ1 4GlcNAcβ1 Asn GlycoProfile II Enzymatic In-Solution N-Deglycosylation Kit Product Code PP0201 Storage Temperature 2 8 C TECHNICAL BULLETIN Product Description Glycosylation is one of the most common posttranslational

More information

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water Richard F. Jack, 1 Jinshui Che, 2 Lipika Basumallick, 1 and Jeffrey Rohrer 1 1 Thermo Fisher Scientific, Sunnyvale, CA, USA;

More information

Immunohematology. IH-QC Modular System. Select. Combine. Control.

Immunohematology. IH-QC Modular System. Select. Combine. Control. Immunohematology IH-QC Modular System Select. Combine. Control. IH-QC Modular System Select. Combine. Control. Transfusion guidelines recommend regular checking of test materials, test methods, local working

More information

Cardiac Assessment Controls

Cardiac Assessment Controls Bio-Rad Laboratories CARDIAC ASSESSMENT CONTROLS Cardiac Assessment Controls Value Assigned for Clinical Laboratory and Point of Care Test Systems Solving the Challenges of Troponin I Testing for the Central

More information

PyroMark Q24 CpG MGMT Handbook

PyroMark Q24 CpG MGMT Handbook December 2009 PyroMark Q24 CpG MGMT Handbook For quantification of CpG methylation in region +17 to +39 in the MGMT gene Sample & Assay Technologies QIAGEN Sample and Assay Technologies QIAGEN is the leading

More information

Protect Your Mass Spec

Protect Your Mass Spec Protect Your Mass Spec Sample Preparation Solutions for Biological Samples www.phenomenex.com/bioms Dirty Samples Could be Costing You Money and Sensitivity Biological samples contain a wide variety of

More information

508 the number of suicide deaths in deaths per 100,000 people was the suicide rate in Suicide deaths in 2013 by gender

508 the number of suicide deaths in deaths per 100,000 people was the suicide rate in Suicide deaths in 2013 by gender An overview of suicide statistics This document summarises information about suicide deaths in New Zealand covering up to 13. It does not attempt to explain causes of suicidal behaviour or causes of changes

More information

Supporting information for: Memory Efficient. Principal Component Analysis for the Dimensionality. Reduction of Large Mass Spectrometry Imaging

Supporting information for: Memory Efficient. Principal Component Analysis for the Dimensionality. Reduction of Large Mass Spectrometry Imaging Supporting information for: Memory Efficient Principal Component Analysis for the Dimensionality Reduction of Large Mass Spectrometry Imaging Datasets Alan M. Race,,, Rory T. Steven,, Andrew D. Palmer,,,

More information

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants Jeremy Matthews, 1 Alex Chen, 2 and Flavio Bedini 1 1 Thermo Fisher Scientific, Singapore; 2 Alpha Analytical Pte Ltd, Singapore Overview

More information

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No. Page 1 of 10 Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection EPL-BAS Method No. 205G881B Method Summary: Residues of 6-CPA are

More information

Chemical compatibility of the Pall QPoint TM Docking Station - Tap Assembly with common surface disinfectants

Chemical compatibility of the Pall QPoint TM Docking Station - Tap Assembly with common surface disinfectants Technical Report.78a Chemical compatibility of the Pall QPoint TM - Tap Assembly with common surface disinfectants QPoint - More than Filtration .0 Background The Pall QPoint Tap Water Filter Assembly

More information

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D Julie A. Horner 1, Marta Kozak 1, Subodh Nimkar 1, and August A. Specht 1 1 Thermo Fisher Scientific, San Jose,

More information

Operating Instructions

Operating Instructions Sequazyme C-Peptide Sequencing Kit Operating Instructions 1 Product Description The Sequazyme C-Peptide Sequencing Kit enables peptide digestion and the analysis of sequentially truncated peptides using

More information

PosterREPRINT AN AUTOMATED METHOD TO SELF-CALIBRATE AND REJECT NOISE FROM MALDI PEPTIDE MASS FINGERPRINT SPECTRA

PosterREPRINT AN AUTOMATED METHOD TO SELF-CALIBRATE AND REJECT NOISE FROM MALDI PEPTIDE MASS FINGERPRINT SPECTRA Overview AN AUTOMATED METHOD TO SELF-CALIBRATE AND REJECT NOISE FROM MALDI PEPTIDE MASS FINGERPRINT SPECTRA Jeffery M Brown, Neil Swainston, Dominic O. Gostick, Keith Richardson, Richard Denny, Steven

More information

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems Bio-Rad Laboratories Cardiac Assessment Controls Cardiac Assessment Controls Value Assigned for Clinical Laboratory and Point of Care Test Systems Solving the Challenges of Troponin I Testing for the Central

More information

Cardiac Assessment Controls A critical element of reliable cardiac testing

Cardiac Assessment Controls A critical element of reliable cardiac testing Cardiac Assessment Controls Cardiac Assessment Controls A critical element of reliable cardiac testing Choose the products best suited for your cardiac assessment needs Each year, 17.7 million people on

More information

Peptide sequencing using chemically assisted fragmentation (CAF) and Ettan MALDI-ToF Pro mass spectrometry

Peptide sequencing using chemically assisted fragmentation (CAF) and Ettan MALDI-ToF Pro mass spectrometry application note Ett MALDI-ToF Pro Peptide sequencing using chemically assisted fragmentation (CAF) d Ett MALDI-ToF Pro mass spectrometry Key words: MALDI-ToF, PSD, peptide sequencing, chemically assisted

More information

Supporting information to Amino-functional polyester dendrimers based on bis-mpa as nonviral vectors for sirna delivery

Supporting information to Amino-functional polyester dendrimers based on bis-mpa as nonviral vectors for sirna delivery Supporting information to Amino-functional polyester dendrimers based on bis-mpa as nonviral vectors for sirna delivery P. Stenström, D. Manzanares, Y. Zhang, V. Ceña and M. Malkoch* * To whom correspondence

More information

Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis

Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis Kevin Cooley Chemistry Supervisor: Kingsley Donkor 1. Abstract Triglycerides are composed of

More information

Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF

Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF Application Note 219 Joan Stevens, PhD; Luke Roenneburg; Kevin Fawcett (Gilson,

More information

Components of a Mass Spectrometer

Components of a Mass Spectrometer Components of a Mass Spectrometer Sample Introduction Inlet GC LC Direct Insertion (Syringe/Probe) Ionization Ion Separation Ion Detection Ion Source EI,CI,,, MALDI Mass Analyzer Under vacuum TOF, Quadrupole,

More information

Solving practical problems. Maria Kuhtinskaja

Solving practical problems. Maria Kuhtinskaja Solving practical problems Maria Kuhtinskaja What does a mass spectrometer do? It measures mass better than any other technique. It can give information about chemical structures. What are mass measurements

More information

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites Valérie Thibert 1, Norbert Dirsch 2, Johannes Engl 2, Martin Knirsch 2 1 Thermo Fisher Scientific,

More information

Third Party Controls for Your Dade Behring Systems

Third Party Controls for Your Dade Behring Systems Bio-Rad Laboratories Q U A L I T Y C O N T R O L Third Party Controls for Your Dade Behring Systems The Complete QC Solution from Bio-Rad Laboratories Bio-Rad has the control you need for your Dade Behring

More information

[ CARE AND USE MANUAL ] GlycoWorks Single Use Sample Preparation Kit CONTENTS

[ CARE AND USE MANUAL ] GlycoWorks Single Use Sample Preparation Kit CONTENTS GlycoWorks Single Use Sample Preparation Kit CONTENTS I. INTRODUCTION General Guideline of Sample Preparation from Glycoprotein to enrich FLR Labeled Glycans Using Reductive Amination Reaction. II. STORAGE

More information

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Introduction, Ionization Methods Mass Analyzers, Ion Activation Methods

More information

Non-Traditional Approaches to MALDI- TOF Mass Spectrometry Analysis of Low Molecular Weight Polymers

Non-Traditional Approaches to MALDI- TOF Mass Spectrometry Analysis of Low Molecular Weight Polymers Non-Traditional Approaches to MALDI- TOF Mass Spectrometry Analysis of Low Molecular Weight Polymers Justin R. Engle and S. Kim R. Williams Laboratory for Advance Separation Technologies Department of

More information

FOURIER STUDY GREYLOCK PRESS: CTS PRODUCT SAMPLE - FOURIER YES. Did the study achieve its main objective?

FOURIER STUDY GREYLOCK PRESS: CTS PRODUCT SAMPLE - FOURIER YES. Did the study achieve its main objective? FOURIER STUDY Did the study achieve its main objective? 2 15% 1 5% 9.8% YES FOURIER compared Repatha with placebo in patients who were taking a statin and had hardening or narrowing of the arteries and

More information

APPLICATIONS TN Determination of Sugars in Animal Feed using HPLC-ELSD

APPLICATIONS TN Determination of Sugars in Animal Feed using HPLC-ELSD Determination of Sugars in Animal Feed using HPLC-ELSD Laura Snow, Zeshan Aqeel, Allen Misa and Scott Krepich Phenomenex, Inc., Madrid Ave., Torrance, CA 9050 USA Laura Snow Scientist- Phenologix Outside

More information

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries FT-Raman Surface Mapping of Remineralized Artificial Dental Caries I. Stangel, 1 R. Rubinovitz, 2 G.D. Arndt, 3 D. Byerly, 3 C. Theriot, 3 J.E. Kerr, 4 1 BioMat Sciences, Bethesda, MD; 2 Thermo Fisher

More information

Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS

Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS Application Note 269 Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS INTRODUCTION Ion chromatography (IC) has been extensively used as the preferred separation technique

More information

Supporting information

Supporting information Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2014 Supporting information Glycan Reductive Isotope-coded Amino Acid Labeling (GRIAL) for Mass Spectrometry-based

More information

CAMAG TLC-MS INTERFACE

CAMAG TLC-MS INTERFACE CAMAG TLC-MS INTERFACE 93.1 249.2 40 30 97.1 20 10 250.2 0 200 400 m/z WORLD LEADER IN PLANAR-CHROMATOGRAPHY Identification and elucidation of unknown substances by hyphenation of TLC / HPTLC and MS The

More information

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis Ying Qing Yu Waters Corporation, Milford, MA, USA APPLICATION BENEFITS

More information

ab Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric)

ab Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric) Version 10b Last updated 19 December 2018 ab118970 Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric) For the measurement of Lipid Peroxidation in plasma, cell culture and tissue extracts.

More information

Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography

Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography Application Note 8 Determination of Ethyl Sulfate Impurity in INTRODUCTION Indinavir sulfate is a specific and potent inhibitor of HIV- protease and is widely used in the treatment of AIDS. Indinavir sulfate

More information

Mercury Speciation Determinations in Asian Dietary Supplements

Mercury Speciation Determinations in Asian Dietary Supplements Mercury Speciation Determinations in Asian Dietary Supplements Terri Christison, Deepali Mohindra, Frank Hoefler, and Linda Lopez, Thermo Fisher Scientific, Sunnyvale, California, USA Overview Purpose:

More information

Ionization Methods. Ionization Methods

Ionization Methods. Ionization Methods Ionization Methods Árpád Somogyi Associate Director CCIC, Mass Spectrometry and Proteomics Laboratory OSU August 17, 2015 Ionization Methods Neutral species Charged species Removal/addition of electron(s)

More information

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages Application Note Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages INTRODUCTION The determination of inorganic anions and cations and organic acids in non-alcoholic

More information

Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222

Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222 Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222 Joan Stevens, Ph.D.; Luke Roenneburg; Tim Hegeman; Kevin Fawcett

More information

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography Application Note 65 Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography INTRODUCTION Preservatives are commonly added to many food products, such as soda, fruit juice,

More information

Triptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions

Triptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions Electronic Supplementary Material (ESI) for Chemical Science. This journal is The Royal Society of Chemistry 2015 Triptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions Stephanie A. Barros

More information

Mass Spectrometry. Actual Instrumentation

Mass Spectrometry. Actual Instrumentation Mass Spectrometry Actual Instrumentation August 2017 See also http://www.uni-bielefeld.de/chemie/analytik/ms f additional infmation 1. MALDI TOF MASS SPECTROMETRY ON THE ULTRAFLEX 2 2. ESI MASS SPECTROMETRY

More information

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition Evelyn Goh Waters Pacific, Singapore APPLICATION BENEFITS This method allows for the simultaneous analysis of 12 water-soluble

More information

Mouse C-Peptide ELISA Kit

Mouse C-Peptide ELISA Kit Mouse C-Peptide ELISA Kit Cat.No: DEIA4507 Lot. No. (See product label) Size 96T Intended Use The Mouse C-Peptide ELISA kit is for the quantitative determination of c-peptide in mouse serum, plasma, and

More information

ab ORAC Assay Kit

ab ORAC Assay Kit Version 1 Last updated 10 April 2018 ab233473 ORAC Assay Kit For the measurement of ORAC activity in cell lysate, plasma, serum, tissue homogenates and food extracts. This product is for research use only

More information

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Methods in Mass Spectrometry Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Sample Engineering Chemistry Biology Life Science Medicine Industry IDF / Police Sample

More information

Apolipoprotein A-1 ELISA

Apolipoprotein A-1 ELISA Apolipoprotein A-1 ELISA For the quantitative determination of apolipoprotein A1 in serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical

More information

Mouse C3 (Complement Factor 3) ELISA Kit

Mouse C3 (Complement Factor 3) ELISA Kit Mouse C3 (Complement Factor 3) ELISA Kit Cat. No.:DEIA8289 Pkg.Size:96T Intended use The Mouse C3 (Complement Factor 3) ELISA Kit is a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring

More information

Dienes Derivatization MaxSpec Kit

Dienes Derivatization MaxSpec Kit Dienes Derivatization MaxSpec Kit Item No. 601510 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION

More information

Ordering Information. ProteOn XPR36 System. Protein Interaction Analysis

Ordering Information. ProteOn XPR36 System. Protein Interaction Analysis Protein Interaction Analysis Ordering Information ProteOn XPR36 System The ProteOn XPR36 system is an optical biosensor capable of simultaneously measuring 36 individual molecular interactions. It integrates

More information

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research Padhraic Rossiter, 1 Jaime Salcedo Dominguez, 1 Jennifer Warren, 1 Norma Breen, 1 Lisa Calton 2 1 Waters Corporation,

More information

Country-wise and Item-wise Exports of Animal By Products Value Rs. Lakh Quantity in '000 Unit: Kgs Source: MoC Export Import Data Bank

Country-wise and Item-wise Exports of Animal By Products Value Rs. Lakh Quantity in '000 Unit: Kgs Source: MoC Export Import Data Bank Country-wise and Item-wise Exports of Animal By Products Value Rs. Lakh Quantity in '000 Unit: Kgs Source: MoC Export Import Data Bank Country CAPEXIL Description HS Code Value 2010-2011 Quantity 2010-2011

More information

SYNAPT G2-S High Definition MS (HDMS) System

SYNAPT G2-S High Definition MS (HDMS) System SYNAPT G2-S High Definition MS (HDMS) System High performance, versatility, and workflow efficiency of your MS system all play a crucial role in your ability to successfully reach your scientific and business

More information

ANALYTICAL SCIENCES JULY 2010, VOL The Japan Society for Analytical Chemistry

ANALYTICAL SCIENCES JULY 2010, VOL The Japan Society for Analytical Chemistry ANALYTICAL SCIENCES JULY 2010, VOL. 26 743 2010 The Japan Society for Analytical Chemistry MALDI Mass Spectrometry Using 2,4,6-Trihydroxyacetophenone and 2,4-Dihydroxyacetophenone with Cyclodextrins: Suppression

More information

Terms and Conditions. VISA Global Customer Assistance Services

Terms and Conditions. VISA Global Customer Assistance Services Terms and Conditions VISA Global Customer Assistance Services Visa Global Customer Assistance Services (VGCAS) 1 The Visa Global Customer Assistance Services are co-ordinated by the Global Assistance Centre

More information

Core-Shell Technology for Proteins and Peptides

Core-Shell Technology for Proteins and Peptides Core-Shell Technology for Proteins and Peptides Better BioSeparations on HPLC and UHPLC Systems www.phenomenex.com/aeris Aeris Core-Shell Technology Core-Shell Particles Precision Engineered for Protein

More information

ab Hemoglobin Assay Kit (Colorimetric)

ab Hemoglobin Assay Kit (Colorimetric) Version 1 Last updated 8 May 2018 ab234046 Hemoglobin Assay Kit (Colorimetric) For the measurement of hemoglobin in various biological samples. This product is for research use only and is not intended

More information

William Wikoff : UC Davis (Committee chair) Pavel Aronov: Stanford University. John Asara: Harvard University. Vladimir Shulaev: University of Texas

William Wikoff : UC Davis (Committee chair) Pavel Aronov: Stanford University. John Asara: Harvard University. Vladimir Shulaev: University of Texas MRG: Metabolomics Research Group William Wikoff : UC Davis (Committee chair) Pavel Aronov: Stanford University John Asara: Harvard University it Vladimir Shulaev: University of Texas Chris Turck: Max Planck

More information

Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section

Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section Emmanuelle Claude, Mark Towers, and Kieran Neeson Waters Corporation, Manchester,

More information

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Valérie Thibert, Bénédicte Duretz Thermo Fisher Scientific, Courtaboeuf,

More information

ab65336 Triglyceride Quantification Assay Kit (Colorimetric/ Fluorometric)

ab65336 Triglyceride Quantification Assay Kit (Colorimetric/ Fluorometric) Version 10 Last updated 19 December 2017 ab65336 Triglyceride Quantification Assay Kit (Colorimetric/ Fluorometric) For the measurement of triglycerides in various samples. This product is for research

More information

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials D. Edwards 2, D. Lennartz 1, L.Shick 2 1 Thermo Fisher Scientific, Langerwehe, Germany 2 Thermo Fisher

More information

MENTAL HEALTH CARE. OECD HCQI Expert meeting 17 th of May, Rie Fujisawa

MENTAL HEALTH CARE. OECD HCQI Expert meeting 17 th of May, Rie Fujisawa MENTAL HEALTH CARE OECD HCQI Expert meeting 17 th of May, 2013 Rie Fujisawa Mental health indicators Any hospital readmissions for patients with schizophrenia Same hospital readmissions for patients with

More information

Supplementary Material (ESI) for Chemical Communications This journal is (c) The Royal Society of Chemistry 2011

Supplementary Material (ESI) for Chemical Communications This journal is (c) The Royal Society of Chemistry 2011 Supporting Information Experimental details 1. Materials Ferric chloride (FeCl 3 6H 2 O), Ethylene glycol (EG) and Ammonium hydroxide (NH 3 H 2 O) were obtained from Beijing Chemical Regent Co. Ltd. (Beijing,

More information