Enrichment of Phospholipids from Biological Matrices with Zirconium Oxide-Modified Silica Sorbents

Size: px
Start display at page:

Download "Enrichment of Phospholipids from Biological Matrices with Zirconium Oxide-Modified Silica Sorbents"

Transcription

1 Enrichment of Phospholipids from Biological Matrices with Zirconium Oxide-Modified Silica Sorbents Xiaoning Lu, Jennifer E. Claus, and David S. Bell Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA T sigma-aldrich.com/analytical

2 Introduction There has been increasing interest in detection and structural characterization of phospholipids in biological matrices such as plasma and tissues. Phospholipids are the building blocks of cellular membranes and play important roles in various cellular responses. Studies have also shown the alteration of phospholipids in blood has been associated with various diseases such as heart attack and vascular diseases and, therefore, are believed potential molecular markers for diseases. Phospholipids are commonly extracted from biological matrices by liquid-liquid extraction (LLE) methods. However, the LLE methods tend to be time consuming and labor-intensive. The present work describes a simple and rapid solid phase extraction (SPE) procedure for extraction and enrichment of phospholipids from biological matrices using zirconium oxide-modified silica particles. In the procedure, biological samples such as plasma and serum were subjected to protein precipitation with organic solvents followed by filtration or centrifugation to remove the protein precipitates. The resulting organic extracts were then loaded onto the zirconium oxide-modified silica particles, either packed in 96-well plates or a single cartridge. 2

3 Introduction (contd.) After wash, the phospholipids are eluted from the particles by 5% ammonium hydroxide in the organic solvents. The entire procedure generally takes less than 15 min and multiple samples (e.g. 96 samples) can be processed in parallel with 96-well plates packed with the sorbents. The applicability of the method has been demonstrated by the extraction of seven classes of phospholipids with varied polar head groups including phosphatidic acid, phosphatidylcholine, phosphatidylethanolamine, phosphatidylglycerol, phosphatidylinositol, phosphatidylserine, and sphingomyelin from plasma samples. 3

4 Phospholipid Structure O R1O O P OR2 O - OR3 Phospholipid Phospholipid R1 R2 R3 Phosphatidic Acid (PA) Fatty Acyl Fatty Acyl H Phosphatidylcholine (PC) Fatty Acyl Fatty Acyl -CH 2 -CH 2 -N +(CH 3 ) 3 Phosphatidylethanolamine (PE) Fatty Acyl Fatty Acyl -CH 2 -CH 2 -NH 2 Phosphatidylglycerol (PG) Fatty Acyl Fatty Acyl -CH 2 -CH(OH)-CH 2 (OH) OH OH Phosphatidylinositol (PI) Fatty Acyl Fatty Acyl HO OH OH Phosphatidylserine (PS) Fatty Acyl Fatty Acyl Lysophosphatidylcholine (LPC) Fatty Acyl H -CH 2 -CH(COOH)-NH 2 -CH 2 -CH 2 -N +(CH 3 ) 3 4

5 Experimental 2.1. Phospholipid Standards and MRM Transitions # PL short name Full name Molecular Formula MW (mono) ESI ions (+) MRM(+) 1 PA 16:0/16:0 2 PC 16:0/16:0 3 PE 18:1/18:1 1,2-Dipalmitoyl-sn-glycero-3- phosphate acid C35H69O8P ([M+NH4]+) 666.7/551.7 (NL115) 1,2-Dipalmitoyl-rac-glycero-3- phosphocholine C40H81NO8P (M+) 734.6/ ,2-Dioleoyl-sn-glycero /603.7 phosphoethanolamine C41H78NO8P [M+H]+ (NL141) 4 PG 16:0/18:1 5 PI 38:4 Not sure 6 PS 16:0/16:0 1-Palmitoyl-2-oleoyl-sn-glycero-3- phospho-rac-(1-glycerol) C40H77O10P C47H83O13P (18:2/20:2) C47H83O13P (18:0/20:4) ,2-Dipalmitoyl-sn-glycero-3- phospho-l-serine C38H74NO10P , [M+NH4] , [M+H] , [M+NH4] , [M+H] , [M+Na] /577.7 (NL172) 904.8/627.8 (NL277) 736.7/551.7 (NL185) 7 SM 16:0 N-Palmitoyl-D-sphingomyelin C39H80N2O6P (M+) 703.7/

6 Experimental (contd.) 2.2. LC-MS Analysis LC-MS conditions: instrument: Agilent 1200 HPLC with 3200 QTRAP (AB/Sciex ) HPLC column: Ascentis C18, 15 cm x 2.1 mm (Sigma Cat. No U) mobile phase (isocratic): 20% water/30% methylene chloride/50% methanol, with 10 mm ammonium formate or 5 mm ammonium methylmalonic acid (MMA) flow rate: 0.2 ml/min temp.: 35 C det.: MS, ESI(+), MRM mode column pressure: 1277 psi (88 bar) injection: 2 µl 6

7 Experimental (contd.) 2.3. Phospholipid Enrichment Procedure The phospholipid enrichment using HybridSPE 96-well plates was performed either in the on-line or off-line modes. In the on-line mode, the biological sample, e.g. plasma and serum, are protein crashed with organic solvents inside the wells of the plates. Whereas in the off-line mode, the protein crashing step is carried out outside the plates. In either mode, a solvent-toplasma ratio of 9:1 was employed for optimum protein crashing results and maximum extraction of phospholipids from the plasma. After sample loading, the samples are pulled through the plate via vacuum typically at 10 in Hg for 3 min. Extended vacuum may be needed for more viscous samples. A washing was then performed to remove the loosely or non-specific bindings to the sorbents. Finally, the phospholipids were eluted off the sorbents with 5% ammonium hydroxide (NH 4 OH) in organic solvents. The collected samples were directly injected for LC-MS analysis. Alternatively, the samples were dried with nitrogen and reconstituted into the LC-MS mobile phase. 7

8 Flow Chart of the Phospholipid Enrichment Procedure On-line Flow Chart µl Plasma 900 µl Solvent Off-line Flow Chart 100 µl Plasma 900 µl Solvent +Centrifugation (e.g g x 3 min.) Sample loading Mixing (e.g. by vortex) Vacuum (10 in Hg for 3 min.) Vacuum (10 in Hg for 3 min.) Flow thru Washing (optional) Flow thru Washing (optional) Elution: NH 4 OH in solvent Wash solvent Recovered Phospholipids Elution: NH 4 OH in solvent Wash solvent Recovered Phospholipids 8

9 Results and Discussion 3.1. LC-MS of the Phospholipids SM16:0/16:0 PC16:0/16: / /184.2 PS16:0/16: /551.7 PE18:1/18: /603.7 PI18:0/20: /627.8 PG16:0/18: /577.7 PA16:0/16: /551.7 MP with 5 mm (NH 4 ) 2 MMA MP with 10 mm NH 4 FA Time, min 9

10 Results and Discussion (contd.) 3.2. Phospholipid Enrichment Mechanism Lewis Base Hydroxide Phosphate Fluoride Citrate Sulfate Acetate Formate Chloride rel. Strength on Zirconia Strongest Weakest 10

11 Results and Discussion (contd.) 3.3. Phospholipid Enrichment 96-well plate 5 µm PTFE (Teflon) Frit Zirconia coated silica particles, 50 mg 0.2 µm filter membrane 11

12 Results and Discussion (contd.) 3.4. Recovery of Phospholipid Enrichment Recovery of neat phospholipid standards and phospholipids spiked in rabbit plasma. MRM Neat Flow-through Neat Elute Spike Flow-through Spike Elute PA 16:0/16: / % 1% 0% 0% PC 16:0/16: / % 96% 0% 154% PE 18:1/18: / % 94% 0% 106% PG 16:0/18: / % 34% 0% 41% PI 38: / % 0% 0% 0% PS 16:0/16: / % 19% 0% 26% SM 16: / % 94% 0% 793% Experimental: Neat phospholipid standards of 200 ng/ml each were prepared in water/methanol/dichloromethane (10/45/45). The rabbit plasma were protein crashed with methanol/dichloromethane (50/50) at ratio of 1:9, and then centrifuged at 6000 g for 3 min. The resulting supernatant were collected and spiked with phospholipid standards at 200 ng/ml. In either case, 400 μl of the neat or spiked phospholipid samples were loaded onto the HybridSPE plate and pulled through under vacuum at 10 in Hg for 3 min. The flowthrough were collected for LC-MS analysis. The bound phospholipid on the sorbent were eluted with 400 μl of 5% ammonium hydroxide in methanol/dichloromethane (1:1), and 12 collected for LC-MS analysis.

13 Results All the tested phospholipids were completely captured by the plate, as indicated by the fact that no phospholipids were detected in the flow-through of the neat or spiked phospholipid samples. PA and PI strongly bind to the phase and could not be eluted off the plate by the explored elution solvent, as reflected by the zero recovery of neat or spiked phospholipids. PC, PE, and SM were readily eluted and completely recovered from the plate. Relatively lower recovery for PG (30%-40%) and PS (20-30%) were observed, indicating the elution is not complete. The high recoveries of spiked PC (154%) and SM (793%) from rabbit plasma were due to the presence of endogenous PC and SM in the rabbit plasma. 13

14 Conclusion A generic phospholipid enrichment method has been devised for 7 classes of phospholipids except PA and PI, which could not be eluted off the zirconia sorbents. The method utilizes the selective affinity of zirconia toward phosphate groups of the different phospholipids. The method can be performed with high throughput, with the aid of a 96-well plate. In addition, an LC-MS method has been established for separation and detection of phospholipids on a C18 reversed-phase column with high separation efficiency (e.g. nice peak shapes). Further work will be necessary to elaborate conditions for elution of PA and PI off the zirconia sorbents. Ascentis and HybridSPE are registered trademarks of Sigma-Aldrich Co. LLC. 14

Alternative Method of Hybrid SPE Sample Preparation for High Recovery LC-MS Analysis of Pharmaceutical Compounds in Plasma

Alternative Method of Hybrid SPE Sample Preparation for High Recovery LC-MS Analysis of Pharmaceutical Compounds in Plasma Alternative Method of Hybrid SPE Sample Preparation for High Recovery LC-MS Analysis of Pharmaceutical Compounds in Plasma Authors: Xiaoning Lu, and Michael Ye Presenter: Carmen T. Santasania Supelco/Sigma/Aldrich

More information

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS Olga Shimelis, An Trinh, and Michael Ye Supelco, Div. of Sigma-Aldrich, Bellefonte, PA T407125 Introduction Molecularly

More information

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids Jeremy Netto, 1 Stephen Wong, 1 Federico Torta, 2 Pradeep Narayanaswamy, 2

More information

Impact of Reversed-Phase Chiral Chromatography on the LC-MS Analysis of Drugs in Biological Fluids

Impact of Reversed-Phase Chiral Chromatography on the LC-MS Analysis of Drugs in Biological Fluids Impact of Reversed-Phase Chiral Chromatography on the LC-MS Analysis of Drugs in Biological Fluids David S. Bell, Carmen T. Santasania, Jennifer Claus, Wayne K. Way, and Craig R. Aurand Supelco, Div. of

More information

LC/MS Method for Comprehensive Analysis of Plasma Lipids

LC/MS Method for Comprehensive Analysis of Plasma Lipids Application Note omics LC/MS Method for Comprehensive Analysis of Plasma s Authors Tomas Cajka and Oliver Fiehn West Coast Metabolomics Center, University of California Davis, 451 Health Sciences Drive,

More information

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis Application Note AN784 Extraction of Multiple Mycotoxins from Nuts Using ISOLUTE Myco Page 1 Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis This application note

More information

Transfer and Speed-up of Methods to Fused-Core Particle Columns EPA Method 1694

Transfer and Speed-up of Methods to Fused-Core Particle Columns EPA Method 1694 Transfer and Speed-up of Methods to Fused-Core Particle Columns EPA Method 1694 Ed Mauney, Hillel K. Brandes, William H. Campbell, Wayne K. Way, Richard A. Henry, David S. Bell, and Carmen T. Santasania

More information

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS By Shun-Hsin Liang and Frances Carroll Abstract Vitamin K₁ and K₂ analysis is typically complex and time-consuming because these

More information

Systematic Evaluation of Solid Phase Extraction (SPE) Chemistries for the Determination of Acidic, Neutral, and Basic Drugs

Systematic Evaluation of Solid Phase Extraction (SPE) Chemistries for the Determination of Acidic, Neutral, and Basic Drugs Systematic Evaluation of Solid Phase Extraction (SPE) Chemistries for the Determination of Acidic, Neutral, and Basic Drugs Laura Marlatt, An Trinh, and David S. Bell Supelco, 595 North Harrison Road,

More information

The use of mass spectrometry in lipidomics. Outlines

The use of mass spectrometry in lipidomics. Outlines The use of mass spectrometry in lipidomics Jeevan Prasain jprasain@uab.edu 6-2612 utlines Brief introduction to lipidomics Analytical methodology: MS/MS structure elucidation of phospholipids Phospholipid

More information

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis Application Note AN782 Extraction of Multiple Mycotoxins from Grain Using ISOLUTE Myco Page 1 Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis This application

More information

LC-Based Lipidomics Analysis on QTRAP Instruments

LC-Based Lipidomics Analysis on QTRAP Instruments LC-Based Lipidomics Analysis on QTRAP Instruments Junhua Wang and Paul RS Baker SCIEX LC-Based Lipidomics Analysis Topics Covered Lipid extraction techniques Hydrophilic Interaction Chromatography (HILIC)

More information

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS Christophe Deckers, M.Sc. Sample prep Application Scientist Types of ``Interferences``

More information

Mass Spectrometry based metabolomics

Mass Spectrometry based metabolomics Mass Spectrometry based metabolomics Metabolomics- A realm of small molecules (

More information

Performance of an ultra low elution volume 96-well plate

Performance of an ultra low elution volume 96-well plate Performance of an ultra low elution volume 96-well plate Claude R. Mallet, Ziling Lu, Jeff R. Mazzeo, Uwe D. Neue Waters Corporation PittCon 2003 March 10-14 2003 Orlando, Florida Today s Challenges Faced

More information

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Application Note Food Safety Authors Chenhao Zhai Agilent

More information

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis Application Note AN807 Extraction of Aflatoxin M From Infant Formula Using ISLUTE Myco Page Extraction of Aflatoxin M From Infant Formula Using ISLUTE Myco SPE Columns prior to LC-MS/MS Analysis This application

More information

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis Comprehensive Study of SLE as a Sample Preparation Tool for Bioanalysis Wan Wang, Warren Chen, Jerry Wang Bonna-Agela Technologies 179 Southern Street, West TEDA, Tianjin, China Abstract A simple, fast,

More information

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis UCT Part Numbers ECHLD126-P ENVIRO-CLEAN HLDVB mg/6ml SPE cartridge PE Frit SLDAID21-18UM Selectra DA UHPLC column ( 2.1 mm, 1.8

More information

Supporting Information (SI)

Supporting Information (SI) Electronic Supplementary Material (ESI) for Analyst. This journal is The Royal Society of Chemistry 2015 Supporting Information (SI) Title: Optimization of Metabolite Extraction of Human Vein Tissue for

More information

Application Note. Author. Abstract. Introduction. Food Safety

Application Note. Author. Abstract. Introduction. Food Safety Determination of β2-agonists in Pork with SPE eanup and LC-MS/MS Detection Using Agilent BondElut PCX Solid-Phase Extraction Cartridges, Agilent Poroshell 120 column and Liquid Chromatography-Tandem Mass

More information

SUPPLEMENTARY DATA. Materials and Methods

SUPPLEMENTARY DATA. Materials and Methods SUPPLEMENTARY DATA Materials and Methods HPLC-UV of phospholipid classes and HETE isomer determination. Fractionation of platelet lipid classes was undertaken on a Spherisorb S5W 150 x 4.6 mm column (Waters

More information

Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results

Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results Presenter: Diana Wang Applications Engineer SPP-Support@agilent.com 1 Today s Agenda - Why sample

More information

Application Note. Abstract. Authors. Pharmaceutical

Application Note. Abstract. Authors. Pharmaceutical Analysis of xycodone and Its Metabolites-oroxycodone, xymorphone, and oroxymorphone in Plasma by LC/MS with an Agilent ZRBAX StableBond SB-C18 LC Column Application ote Pharmaceutical Authors Linda L.

More information

Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS

Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS UCT Part Numbers CSXCE106 Clean Screen XCEL I 130mg / 6mL SPE Cartridge SPHACE5001-5 Select

More information

The challenging extraction of non-polar contaminants out of a non-polar vegetable oil sample

The challenging extraction of non-polar contaminants out of a non-polar vegetable oil sample The challenging extraction of non-polar contaminants out of a non-polar vegetable oil sample Presented by Michael Ye ExTech, 2014 sigma-aldrich.com/analytical 1 Agenda 1. Background; why and how analyze

More information

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS) High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS) Michael Rummel, Matthew Trass, Michael Campognone, and Sky Countryman Phenomenex, Inc., 411 Madrid Avenue, Torrance,

More information

Ion Exchange and Reversed Phase interactions in selective Bio-SPME extractions of designer drugs

Ion Exchange and Reversed Phase interactions in selective Bio-SPME extractions of designer drugs Ion Exchange and Reversed Phase interactions in selective Bio-SPME extractions of designer drugs Frank Michel, Craig R. Aurand, Robert E. Shirey, Yong Chen, Leonard M. Sidisky sigma-aldrich.com/analytical

More information

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W Simple, Fast and Accurate Quantitation of Human Plasma Vitamins and Their Metabolites by Protein Precipitation Combined with Columns Using HPLC-UV, HPLC-FLD or LC/MS/MS Shuguang Li, Jason Anspach, Sky

More information

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup Application Note Forensic Testing Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup Author Limian Zhao Agilent Technologies, Inc.

More information

LC-MS/MS Method for the Determination of Tenofovir from Plasma

LC-MS/MS Method for the Determination of Tenofovir from Plasma LC-MS/MS Method for the Determination of Tenofovir from Plasma Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 687 Key Words SPE, SOLA CX, Hypersil GOLD, tenofovir Abstract

More information

Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates

Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates Application Note Clinical Research Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates Authors Limian Zhao and Megan Juck Agilent Technologies, Inc. Abstract

More information

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes Eric S. Grumbach,, Diane M. Diehl and Jeffrey R. Mazzeo 2003 Waters Corp. Outline Introduction HILIC Definition

More information

SPME-LC Fibers for a Variety of Applications

SPME-LC Fibers for a Variety of Applications SPME-LC Fibers for a Variety of Applications Robert Shirey 1, Craig Aurand 1, Dajana Vuckovic 2, Katherine Stenerson 1, Yong Chen 1, Leonard Sidisky 1, and Janusz Pawliszyn 2 1 Supelco, Div. of Sigma-Aldrich,

More information

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts Application Note Clinical Research Authors Mark Sartain and Theodore Sana Agilent Technologies, Inc. Santa Clara, California, USA Introduction

More information

Overcoming the Challenges of Sample Preparation and LC/MS/MS Method Development for Clinical Applications

Overcoming the Challenges of Sample Preparation and LC/MS/MS Method Development for Clinical Applications Overcoming the Challenges of Sample Preparation and LC/MS/MS Method Development for Clinical Applications Speaker: Sky Countryman, Manager of PhenoLogix and Applied Technologies, Phenomenex Webinar Host:

More information

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis Michael S Young and Kim Van Tran Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Efficient, timesaving multiclass/ multiresidue methodology

More information

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid Application Note Clinical Research Authors Derick Lucas and Limian Zhao Agilent Technologies, Inc. Abstract Lipids from

More information

Rapid and Robust Detection of THC and Its Metabolites in Blood

Rapid and Robust Detection of THC and Its Metabolites in Blood Rapid and Robust Detection of THC and Its Metabolites in Blood Application Note Forensics/Doping Control Author Stephan Baumann Agilent Technologies, Inc. Santa Clara CA 95051 USA Abstract A robust method

More information

Surviving Matrix Effects Experiments. Grace van der Gugten St. Paul s Hospital, Vancouver, BC, Canada

Surviving Matrix Effects Experiments. Grace van der Gugten St. Paul s Hospital, Vancouver, BC, Canada Surviving Matrix Effects Experiments Grace van der Gugten St. Paul s Hospital, Vancouver, BC, Canada gvandergugten@providencehealth.bc.ca Outline What are matrix effects Post Column Infusion Phospholipid

More information

Relative Quantitation of Human Polymorphonuclear Leukocyte Cell Membrane GPEtn Lipids

Relative Quantitation of Human Polymorphonuclear Leukocyte Cell Membrane GPEtn Lipids Relative Quantitation of Human Polymorphonuclear Leukocyte Cell Membrane GPEtn Lipids Using the QTRAP System with mtraq Reagents Karin A. Zemski-Berry 1, John M. Hevko 2, and Robert C. Murphy 1 1 Department

More information

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS Application Note Forensic Analysis Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS Authors Joan Stevens and Limian Zhao Agilent Technologies,

More information

The HPLC Preparative Scale-Up of Soybean Phospholipids Application

The HPLC Preparative Scale-Up of Soybean Phospholipids Application The HPLC Preparative Scale-Up of Soybean Phospholipids Application Food and Flavors Authors Cliff Woodward and Ron Majors Agilent Technologies, Inc. 285 Centerville Road Wilmington, DE 1988-161 USA Abstract

More information

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis Providing a Universal, ne-step Alternative to Liquid-Liquid Extraction in Bioanalysis Jessalynn P. Wheaton, Erin E. Chambers, and Kenneth J. Fountain APPLICATIN BENEFITS n Simple, one-step sample preparation

More information

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer A Robustness Study for the Agilent 7 LC-MS/MS Mass Spectrometer Application Note Clinical Research Authors Linda Côté, Siji Joseph, Sreelakshmy Menon, and Kevin McCann Agilent Technologies, Inc. Abstract

More information

Phospholipid characterization by a TQ-MS data based identification scheme

Phospholipid characterization by a TQ-MS data based identification scheme P-CN1716E Phospholipid characterization by a TQ-MS data based identification scheme ASMS 2017 MP-406 Tsuyoshi Nakanishi 1, Masaki Yamada 1, Ningombam Sanjib Meitei 2, 3 1 Shimadzu Corporation, Kyoto, Japan,

More information

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No. Page 1 of 10 Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection EPL-BAS Method No. 205G881B Method Summary: Residues of 6-CPA are

More information

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 709 Key Words SPE, SOLA

More information

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Dominic Foley, Michelle Wills, and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION

More information

Fig. 1: Chemical structure of arachidonic acid COOH CH 3

Fig. 1: Chemical structure of arachidonic acid COOH CH 3 Elimination of Matrix Effects Using Mixed-mode SPE Plate for High Throughput Analysis of Free Arachidonic Acid in Plasma by LC-MS/MS Wan Wang, Suzi Qin, Linsen Li, Warren Chen, Jerry Wang 179, Southern

More information

Authors. Abstract. Introduction. Environmental

Authors. Abstract. Introduction. Environmental Determination of Ultratrace Amitrol in Water Samples by in situ Derivatization-Solid Phase Extraction-Liquid Chromatography-Mass Selective Detector Application Environmental Authors Gerd Vanhoenacker,

More information

SOLID PHASE EXTRACTION: Fundamentals and Recent Developments

SOLID PHASE EXTRACTION: Fundamentals and Recent Developments SOLID PHASE EXTRACTION: Fundamentals and Recent Developments Olga I. Shimelis, Ph.D Principal R&D Scientist sigma-aldrich.com/analytical Time Spent on the Analytical Process R.E. Majors, LC/GC Magazine,

More information

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis Application Note AN890 Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE SLE+ Page Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE SLE+ Prior to LC/MS-MS

More information

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases Clinical, Forensic & Toxicology Applications Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases By Shun-Hsin

More information

LC/MS/MS Analysis of Metabolites of Synthetic Cannabinoids JWH-018 and JWH-073 in Urine

LC/MS/MS Analysis of Metabolites of Synthetic Cannabinoids JWH-018 and JWH-073 in Urine Abstract A liquid chromatographic method was developed to resolve a comprehensive set of metabolites of JWH-018 and JWH-073. In addition to the chromatographic analysis method, an extraction method was

More information

Determination of Pharmaceutical Residues in Bovine Milk via LC MS/MS Following Solid Phase Extraction

Determination of Pharmaceutical Residues in Bovine Milk via LC MS/MS Following Solid Phase Extraction Determination of Pharmaceutical Residues in Bovine Milk via LC MS/MS Following Solid Phase Extraction Application Note FB0113 Keywords SPE (Solid Phase Extraction), Bovine Milk, JECFA (Joint FAO/WHO Expert

More information

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ Application Note Clinical Research Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ Authors Jennifer Hitchcock 1, Lauren Frick 2, Peter Stone 1, and Vaughn Miller 2 1 Agilent

More information

Hydrolysis of Acylglycerols and Phospholipids of Milled Rice Surface Lipids During Storage 1

Hydrolysis of Acylglycerols and Phospholipids of Milled Rice Surface Lipids During Storage 1 RICE QUALITY AND PROCESSING Hydrolysis of Acylglycerols and Phospholipids of Milled Rice Surface Lipids During Storage 1 H.S. Lam and A. Proctor ABSTRACT The relative contribution of acylglycerols and

More information

Supplementary Information. Effects of Perfluorooctanoic Acid on Metabolic Profiles in Brain and Liver of Mouse by a

Supplementary Information. Effects of Perfluorooctanoic Acid on Metabolic Profiles in Brain and Liver of Mouse by a Supplementary Information Effects of Perfluorooctanoic Acid on Metabolic Profiles in Brain and Liver of Mouse by a High-throughput Targeted Metabolomics Approach Nanyang Yu, Si Wei, *, Meiying Li, Jingping

More information

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis

Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE. PLD+ Prior to LC-MS/MS Analysis Application Note AN842.V. Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE PLD+ Prior to LC-MS/MS Analysis Page Extraction of 25-hydroxy Vitamin D from Serum Using ISOLUTE PLD+ Prior to LC-MS/MS

More information

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column Clinical, Forensic & Toxicology Applications 4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column By Amanda Rigdon Abstract A rapid, sensitive method

More information

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid Joanne ( 乔安妮 ) Mather Senior Scientist Waters Corporation Data courtesy of Erin Chambers and Mary

More information

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using 500 MG CLEAN UP CCX2 and LC-MS/MS

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using 500 MG CLEAN UP CCX2 and LC-MS/MS Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using 500 MG CLEAN UP CCX2 and LC-MS/MS UCT Part Numbers: CUCCX256 - Clean-Up CCX2 (C8 + Carboxylic Acid) 500mg/6mL

More information

Glycerolipid Analysis. LC/MS/MS Analytical Services

Glycerolipid Analysis. LC/MS/MS Analytical Services Glycerolipid Analysis LC/MS/MS Analytical Services Molecular Characterization and Quantitation of Glycerophospholipids in Commercial Lecithins by High Performance Liquid Chromatography with Mass Spectrometric

More information

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using CLEAN UP CCX2 and LC-MS/MS

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using CLEAN UP CCX2 and LC-MS/MS Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using CLEAN UP CCX2 and LC-MS/MS UCT Part Numbers CUCCX256 Clean Up CCX2 (C8 + Carboxylic Acid) 500mg / 6mL SPE Cartridge

More information

Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS

Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS UCT Part Numbers SSHLD063 Styre Screen HLD 60 mg, 6 ml column SPHPHO6001-5 Select ph Buffer Pouches 100 mm Phosphate

More information

Filtration Approaches to Sample Preparation and Cleanup. Golnar Javadi Application Engineer Sample Preparation Products

Filtration Approaches to Sample Preparation and Cleanup. Golnar Javadi Application Engineer Sample Preparation Products Filtration Approaches to Sample Preparation and Cleanup Golnar Javadi Application Engineer Sample Preparation Products Spp-support@agilent.com 1 Today s Agenda How does filtration work a brief overview

More information

Extraction of 11-nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE. SLE+ Prior to GC/MS Analysis

Extraction of 11-nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE. SLE+ Prior to GC/MS Analysis Application Note AN84 Extraction of -nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE SLE+ Page Extraction of -nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE SLE+

More information

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research Stephen Balloch and Gareth Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity afforded by mass selective detection Wide linear measuring range Simple, inexpensive sample

More information

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography Application ote Food Safety Authors Chen-Hao Zhai and Yun Zou Agilent Technologies Co. Ltd.

More information

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC Mia Summers and Kenneth J. Fountain Waters Corporation, Milford, MA, USA APPLICATION BENEFITS Updating USP Methods from HPLC to UPLC using sub-2 µm columns 85% decrease in analysis time, faster throughput

More information

EXPERIMENT 13: Isolation and Characterization of Erythrocyte

EXPERIMENT 13: Isolation and Characterization of Erythrocyte EXPERIMENT 13: Isolation and Characterization of Erythrocyte Day 1: Isolation of Erythrocyte Steps 1 through 6 of the Switzer & Garrity protocol (pages 220-221) have been performed by the TA. We will be

More information

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS Introduction Drug analysis from whole blood is gaining popularity due to a more complete measurement

More information

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety Determination of Aflatoxins in Food by LC/MS/MS Application Food Safety Authors Masahiko Takino Agilent Technologies 9-1 Takakura-Cho Hachiouji-Shi, Tokyo Japan Toshitsugu Tanaka Kobe Institute of Health

More information

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS Mark Ritchie and Evelyn Goh Waters Pacific Pte Ltd., Singapore A P P L I C AT ION B E N E F I T S Delivers a rapid 10-min MRM method

More information

Studies on Stationary Phase Selectivity for Solid-Core Particles

Studies on Stationary Phase Selectivity for Solid-Core Particles Studies on Stationary Phase Selectivity for Solid-Core Particles Richard A. enry, Wayne K. Way, Carmen T. Santasania, and David S. Bell Supelco, Div. of Sigma-Aldrich, Bellefonte, PA 6823 USA T4060 sigma-aldrich.com

More information

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ Anthony Sullivan, LC/MS Product Specialist Melanie Mülek and Christoph Müller LC-MS Applications Specialists Hewlett-Packard-Str.

More information

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent LC/MS Triple Quadrupole Application Food Safety Authors Yanyan Fang Agilent Technologies (Shanghai),

More information

Suk Hoo Yoon Korea Food Research Institute 1/42

Suk Hoo Yoon Korea Food Research Institute 1/42 Development of Phospholipases to Produce Structured Phospholipids Suk Hoo Yoon Korea Food Research Institute 1/42 Phospholipase D H H C O R R Z Fatty acyl chain -H Phosphatidic acid (PA) R O C H O - -CH

More information

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS TECHNICAL NOTE 21882 Robust extraction, separation, and quantitation of structural isomer steroids human plasma by SPE-UHPLC-MS/MS Authors Jon Bardsley 1, Kean Woodmansey 1, and Stacy Tremintin 2 1 Thermo

More information

Quantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Quantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup Application Note Forensic Testing Quantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup Author Limian Zhao Agilent Technologies, Inc. Abstract

More information

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) The target compound to be determined is 2, 4, 5-T. 1. Instrument Liquid Chromatograph-tandem mass spectrometer (LC-MS/MS)

More information

Taking the Complexity out of SPE Method Development

Taking the Complexity out of SPE Method Development Taking the Complexity out of SPE Method Development Includes guidelines for the following: P SPE Format Selection P SPE Sorbent Selection P SPE Methods P SPE Volume Guidelines P Common Laboratory Conversions

More information

Agilent 6470 Triple Quadrupole LC/MS System and EMR-Lipid. The Combination for Confident Pesticide Quantitation. Christoph Mueller

Agilent 6470 Triple Quadrupole LC/MS System and EMR-Lipid. The Combination for Confident Pesticide Quantitation. Christoph Mueller Agilent 670 Triple Quadrupole LC/MS System and EMR-Lipid The Combination for Confident Pesticide Quantitation Christoph Mueller Agilent Technologies LCMS Application Scientist EMEAI Food Market Program

More information

PHOTOCATALYTIC DECONTAMINATION OF CHLORANTRANILIPROLE RESIDUES IN WATER USING ZnO NANOPARTICLES. DR. A. RAMESH, Ph.D, D.Sc.,

PHOTOCATALYTIC DECONTAMINATION OF CHLORANTRANILIPROLE RESIDUES IN WATER USING ZnO NANOPARTICLES. DR. A. RAMESH, Ph.D, D.Sc., PHOTOCATALYTIC DECONTAMINATION OF CHLORANTRANILIPROLE RESIDUES IN WATER USING ZnO NANOPARTICLES DR. A. RAMESH, Ph.D, D.Sc., raamesh_a@yahoo.co.in 1 OBJECTIVES Determination of persistence and photolysis

More information

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18 Analyze Barbiturates in Urine with Agilent 6 LC/MS/MS and Poroshell EC-C8 Application ote Forensic Toxicology Authors Elijah Steinbauer and Pat Friel Toxicology Laboratory at the Veterans Administration

More information

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods Application Note Food Testing & Agriculture Authors Nick Byrd and Danielle Sweeney Campden BRI Gloucestershire UK

More information

Lipid Class Separation Using UPC 2 /MS

Lipid Class Separation Using UPC 2 /MS Michael D. Jones, 1,3 Giorgis Isaac, 1 Giuseppe Astarita, 1 Andrew Aubin, 1 John Shockcor, 1 Vladimir Shulaev, 2 Cristina Legido-Quigley, 3 and Norman Smith 3 1 Waters Corporation, Milford, MA, USA 2 Department

More information

Challenges in Developing an Ultra-Sensitive Bioanalytical Method for Ethinylestradiol in Human Plasma

Challenges in Developing an Ultra-Sensitive Bioanalytical Method for Ethinylestradiol in Human Plasma Challenges in Developing an Ultra-Sensitive Bioanalytical Method for Ethinylestradiol in uman Plasma Jessalynn P. Wheaton, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, 34 Maple Street,

More information

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1 Development of an SPE-LC/MS/MS Assay for the Simultaneous Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid in Support of Alzheimer s Research Dr. Erin E. Chambers Waters Corporation Presented

More information

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE) Removal of Triton X- from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE) Jonathan P. Danaceau, Erin Chambers, and Kenneth J. Fountain Waters Corporation, 34 Maple Street, Milford, MA USA

More information

Improved Isolation and Analysis of Mycotoxins from Cereals, Beer and Wine

Improved Isolation and Analysis of Mycotoxins from Cereals, Beer and Wine Improved Isolation and Analysis of Mycotoxins from Cereals, Beer and Wine Application Note Authors Elisabeth Korte, Maha Rudrabhatla, Max Erwine, Jason S Wood Agilent Technologies, Inc. Introduction Trichothecenes

More information

Phospholipid Removal Solutions

Phospholipid Removal Solutions Remove Phospholipids Remove Protein No Method Development www.phenomenex.com/phree One Quick Method, Three Big Advantages 1 2 Remove Proteins that can clog HPLC columns Eliminate Phospholipids that cause

More information

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology Application Note Small Molecule Pharmaceuticals and Generics Authors Ritu Arora, William udson, Ben Yong, and Paul Boguszewski

More information

Identification and Quantification of Psychedelic Phenethylamine 2C Series using SPE and LC-MS/MS

Identification and Quantification of Psychedelic Phenethylamine 2C Series using SPE and LC-MS/MS Identification and Quantification of Psychedelic Phenethylamine 2C Series using SPE and LC-MS/MS UCT Part Numbers: CSXCE106-Clean Screen XCEL I 130mg/6mL SLDA100ID21-3UM - Selectra DA HPLC column, 100

More information

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology Claude Mallet, 1 Jennifer Simeone, 2 Paul Rainville 3 1 Workflow Integration Group, Separations Technologies,

More information

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System Mary E. Lame, Erin E. Chambers, Sukhdev S. Bangar, and Kenneth J. Fountain

More information