Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures

Size: px
Start display at page:

Download "Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures"

Transcription

1 Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures Introduction Natural organic matter (NOM) as well as crude oil and crude oil fractions are very complex mixtures of organic compounds consisting of various elemental compositions and chemical structures. These mixtures can vary extremely depending on their origin and complexity. Natural organic matter consists mainly of oxygen containing compounds. However, also nitrogen and sulfur are presented in low amounts in NOM. Therefore, high dynamic range and mass accuracy is needed to detect and identify these compounds. Beside the roughly 9% of hydrocarbon compounds in crude oil, the remaining polar mainly polar compounds are of major interest which are mainly in the resin and asphaltene fractions. These compounds contain different amount of sulfur, nitrogen and oxygen which results in an extremely complex mixture difficult to analyze. Due to the complexity of oil, chromatography can be used to separate compound classes. However, separation of compounds by liquid chromatography is only of limited use. Therefore, mass spectrometers providing extremely high resolving power and mass accuracy like Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometers are needed to analysis these samples [1-2]. Authors Dr. Matthias Witt Bruker Daltonik GmbH, Bremen, Germany Keywords Instrumentation and Software Complex Mixture solarix XR Petroleomics DataAnalysis 4. Crude Oil ESI Source NOM APPI II Source FTMS Composer

2 Still, for FTMS the resolving power is limited by a) analyzer pressure, b) peak coalencense effects and c) non-perfect trapping potential resulting in a de-phasing of the ion cyclotron motion. The solarix XR has been developed to be able to study these kind of mixtures using only a 7T magnet. The non-perfect trapping has been overcome by the develop-ment of a new cell with parabolic trapping potentials, called dynamically harmonized analyzer cell (Fig. 1) [3-5]. It has been shown that extremely high resolving power of more than 2.. can be achieved with this new analyzer cell, however, these measurements were performed with a limited using quadrupole isolation. Nevertheless, very high resolving power of 8. at m/z have been achieved in magnitude mode in full broad band detection of very complex mixtures like crude oil and NOM without any quadrupole isolation using a solarix XR system. Even higher mass resolution of more than 1.2. can be achieved of complex mixtures using the absorption mode processing (AMP) [6]. NOM as well as oil samples have been used to test the performance of the new solarix XR FTMS system with the new dynamically harmonized analyzer cell. Suwannee river fulvic acid (SRFA) as a NOM standard has been measured by electrospray ionization. The mixture consists mainly of oxygen containing compounds with different number of carboxyl, carbonyl and hydroxyl functional groups. However, due to the origin of the sample sodium is present in the sample. ICR paracell Therefore, sodium adducts beside protonated species are detected resulting in more than 5 mass peaks in the spectrum. Different crude oil samples and bitumen as well as a residual fuel oil have been measured by APPI to test the performance of the new FT-ICR system concerning sample complexity. The APPI spectra in positive ion mode are extremely rich in content of peaks due to the formation of radical cations and protonated species. Hence, these spectra represent one of the most complex ones of small molecule mixtures. The spectra have been mainly analyzed concerning achieved mass resolving power and mass accuracy. Additionally, CASI measurements of complex mixtures have been carried out to compare CASI and full broad band results. Experimental Sample preparation The NOM standard SRFA has been purchased by the International Humic Substance Society (IHSS order No. 2S11). SRFA has been measured without any further purification. 2 µg ml -1 solution of SRFA in 5% methanol/5% water was directly sprayed into the electrospray ion source using a syringe pump. One residual fuel oil purchased from TCI (TCI order number S25) as well as two crude oils and one bitumen sample kindly provided by the company SINOPEC, China, were analyzed. Solutions of the oils were prepared without any further purification. 1 mg of the oil were dissolved in 2 µl dichloromethane. These sample stock solutions were diluted 1:5 with 5% methanol, 5% toluene for atmospheric pressure photoionization (APPI) measurements. Mass analysis Mass spectra were acquired with a Bruker solarix XR Fourier transform mass spectrometer (Bruker Daltonik GmbH, Bremen, Germany) equipped with a 7T refrigerated actively shielded superconducting magnet (Bruker Biospin, Wissembourg, France) and the new dynamically harmonized analyzer cell. The samples were ionized in positive ion mode using ESI and APPI (Bruker Daltonik GmbH, Bremen, Germany). Figure 1: New dynamically harmonized analyzer cell with parabolic trapping plates. Sample solutions were continuously infused using a syringe pump. A flow rate of 12 µl h -1 and 6 µl h -1 were used for ESI and APPI, respectively. The size of the acquired data sets was 8 MW resulting in a resolving power of 8 at m/z. The detection mass range was set to m/z 15. scans were acquired for each mass spectrum. A sine apodization was performed before Fourier transformation.

3 Mass calibration The mass spectra were calibrated externally with arginine clusters in positive ion mode using a linear calibration. A 1 µg ml -1 solution of arginine in 5% methanol was used to generate the clusters. To improve the mass accuracy the spectra were recalibrated internally with a known homologous series. Molecular formula calculation The mass formula calculation of the SRFA spectrum was done in DataAnalysis 4. (Bruker Daltonik GmbH, Bremen, Germany) using a maximum formula of C n H h O 3 Na, electron configurations even and a mass tolerance of.5 ppm. The calculation of molecular formulae of the oil samples was done in DataAnalysis 4. using a maximum formula of C n H h N 3 O 3 S 3 and electron configurations odd and even due to the formation and detection of radical cations and protonated species. The mass tolerance was set to.5 ppm. Compound classes were calculated with the Composer software (Sierra Analytics, Modesto, CA, USA. Results and Discussion The performance of the solarix XR system has been demonstrated by measurement of different very complex samples. The mass spectrum of SRFA measured in positive ion mode using electrospray ionization is shown in Figure 2. The mass spectrum has more than 5 mass peaks. More than 27 molecular formulae containing only carbon, hydrogen, oxygen and maximum of one sodium could be calculated with a root mean square (RMS) mass error less than 5. The mass error plot is shown in Figure 3. A residual fuel oil was measured to show the performance of the solarix XR for the petroleomics applications. The mass error plot of the ESI spectrum of the residual fuel oil is shown in Figure 4. Mainly compound class N and NO have been detected resulting in low number of peaks detected by electrospray ionization. Roughly molecular formulae have been calculated with a RMS error of 65. An even low RMS error of only 13 could be achieved of the compound class N1 considering only mass peaks above with a relative abundance of more than 1%. Mass error plot of ESI spectrum Figure 3: Mass error plot of ESI spectrum of SRFA in positive ion mode peaks were considered for the calculation ( 13 C peaks were excluded). A RMS mass error of only 49 was calculated. Mass error plot of ESI spectrum 45-1 avg. Deviation: RMS error: peaks considered for calculation avg. Deviation: 2 RMS error: peaks considered for calculation Broad band electrospray spectrum Resolving power: 8. (at m/z (at) m/z ) NOM standard NOM complex mixture standard Suwannee river fulvic acid Suwannee river fulvic acid Figure 4: Mass error plot of ESI spectrum of a residual fuel oil in positive ion mode. Nearly peaks were considered for the calculation ( 13 C peaks were excluded). A RMS mass error of only 65 was calculated m/z Figure 2: Broad band electrospray spectrum of Suwannee river fulvic acid from the IHSS measured in positive ion mode. More than 5 mass peaks could be detected.

4 More challenging samples are crude oils with high amount of sulfur when analyzed by APPI. These samples are extremely complex and APPI can detect sulfur containing compound with one, two or even three sulfur atoms. Additionally, radical cations as well as protonated species are formed resulting in spectra with much more peaks than detected by ESI. Two crude oil and a bitumen sample have been analyzed by APPI in positive ion mode. The APPI spectra are shown in Figure 5. Broad band APPI spectra Crude oil, low sulfur Crude oil, medium sulfur Bitumen, high sulfur m/z The average mass is shifting to higher mass with increase of the sulfur content. However, also the complexity is increasing resulting in a higher mass error. The error plots of the three samples are shown in Figure 6. The number of unambiguously assigned molecular formulae is increasing from roughly 5 to nearly 8 with the amount of sulfur in the complex sample. The RMS mass error shows the expected trend with an increase from 12 to 139 based on the complexity of the sample. The difference of the RMS mass error of peaks with a relative abundance of more than 1% is even more significant. The error increases from 2 to 4 with higher sulfur content in crude oil and nearly doubles for the extremely complex bitumen sample. Nevertheless, a resolving power of 8. at m/z could still be achieved even of the most complex bitumen sample. The mass spectra of extremely complex mixtures can be further improved by CASI (continuous accumulation of selected ions). Ions only of a limited mass range (typically Da isolation mass window) are selected with a quadrupole. These selected ions are accumulated in the storage cell for detection. The dynamic range of the mass spectrum can be improved with this technique. Also mass errors are lowered due to the reduced number of detected peaks compared to the broad mass spectrum. The mass error plots of the broad band and CASI measurements are shown in Figure 7. Figure 5: Broad band APPI spectra of two crude oil and one bitumen sample with different amount of sulfur in positive ion mode. Mass error plots a) Low sulfur containing sample (crude oil) b) Medium sulfur containing sample (crude oil) abs. avg. Deviation: 2, RMS error: 12 abs. avg. Deviation: 22, RMS error: considered for calculation 6518 considered for calculation RMS error: 2 (peaks > 1%) 45 RMS error: 4 (peaks > 1%) c) High sulfur containing sample (bitumen) abs. avg. Deviation: 51, RMS error: considered for calculation RMS error: 75 (peaks > 1%) Figure 6: Mass error plots of a) crude oil with low amount of sulfur, b) crude oil with medium amount of sulfur and c) bitumen with high amount of sulfur. Complexity of mass spectra increases with amount of sulfur resulting in a) detection of more mass peaks for error calculation and b) increase of mass error.

5 Mass error plots of bitumen sample a) Broad band, masses m/z abs. avg. Deviation: 18, RMS error: considered for calculation b) Broad band, masses m/z abs. avg. Deviation: 4, RMS error: considered for calculation RMS error: 46 (peaks > 1%) RMS error: 2 (peaks > 1%) RMS RMS error: error: 426 (peaks (peaks > 1%) > 1%) C peaks considered 13 C peaks considered 2 Peaks > 1% considered Peaks > 1% considered Figure 7: Mass error plots of bitumen sample with high amount of sulfur considering masses between m/z 38 and m/z 48 of a) broad band and b) CASI measurement. RMS mass error is further reduced from 99 to 67 using CASI instead of broad band detection. The mass error is further reduced from 99 to 67 considering only peaks between m/z 38 and m/z 48 using CASI instead of broad band mass detection. Only unambiguously assigned mass peaks with a relative abundance of more than 1% have been considered for the error plot calculation to compare similar peaks in the mass spectrum. Therefore, the number of peaks used for the error plots increases only by about 2% and the improvement in the dynamic range is not really visible here. The effect of improved dynamic range by CASI is shown in Figure 8. Peaks not visible in broad band have been observed by CASI marked with the red arrows. Therefore, this technique can be used to detect very low abundant species in extremely complex mixtures not detectable in broad band mode. Zoom in of CASI spectrum broad band CASI m/z Figure 8: Zoom in of a) broad band and b) CASI spectrum to show the improvement of dynamic range using CASI. Peaks marked with arrows could only be detected with the CASI technique.

6 Bruker Daltonics is continually improving its products and reserves the right to change specifications without notice. Bruker Daltonics , FTMS-46, #7442 Conclusion Complex mixtures like NOM and crude oil have been successfully analyzed on the molecular level with the new solarix 7T XR system. Signal transients of more than 6 seconds could be achieved resulting in a resolving power of about 8. at m/z in magnitude mode. RMS mass errors of less than 5 could be obtained for SRFA. Low RMS mass errors below 15 were observed for extremely complex mixtures like crude oil and bitumen. A resolving power of 8. could be achieved for very complex APPI spectra of these samples in broad band detection mode. Acknowledgments The author would like to thank Dr. Maowen Li from the company SINOPEC, China, for providing the crude oil and bitumen samples. Many thanks to Roland Jertz, Jochen Friedrich and Claudia Kriete from Bruker Daltonik GmbH (Bremen) for their support during sample measurements and generating mass error plots. References [1] Marshall, A. G.; Rodgers, R. P. Acc. Chem. Res. 24, 37, [2] Purcell, J. M.; Hendrickson, C. L.; Rodgers, R. P.; Marshall, A. G. Anal. Chem. 26, 78, [3] Boldin, I.A., Nikolaev, E.N. Rapid Commun. Mass Spectrom. 211, 25, [4] Nikolaev, E.N., Boldin, I.A., Jertz, R., Baykut, G. A. J. Am. Soc. Mass Spectrom. 211, 22, [5] Nikolaev, E.N., Jertz, R., Grigoryev, A., Baykut, G. A. Anal. Chem. 212, 84, [6] Qi, Y., Witt, M., Jertz, R. Baykut, G., Barrow, M.P., Nikolaev, E.N., O Connor, P.B. Rapid Commun. Mass Spectrom. 212, 26, For research use only. Not for use in diagnostic procedures. Bruker Daltonik GmbH Bremen Germany Phone +49 () Fax +49 () sales@bdal.de Bruker Daltonics Inc. Billerica, MA USA Phone +1 (978) Fax +1 (978) ms-sales@bdal.com Fremont, CA USA Phone +1 (51) Fax +1 (51) ms-sales@bdal.com

Changes of Petroleum Acid Distribution Characterized by FT-ICR MS in Heavy Acidic Crude Oil after True Boiling Point Distillation

Changes of Petroleum Acid Distribution Characterized by FT-ICR MS in Heavy Acidic Crude Oil after True Boiling Point Distillation Scientific Research China Petroleum Processing and Petrochemical Technology 2014, Vol. 16, No. 1, pp 8-12 March 30, 2014 Changes of Petroleum Acid Distribution Characterized by FT-ICR MS in Heavy Acidic

More information

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices Application ote #LCMS-2 esquire series Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices Introduction The simple monitoring

More information

FOURIER TRANSFORM MASS SPECTROMETRY

FOURIER TRANSFORM MASS SPECTROMETRY FOURIER TRANSFORM MASS SPECTROMETRY https://goo.gl/vx3ogw FT-ICR Theory Ion Cyclotron Motion Inward directed Lorentz force causes ions to move in circular orbits about the magnetic field axis Alan G. Marshall,

More information

A pillar[2]arene[3]hydroquinone which can self-assemble to a molecular zipper in the solid state

A pillar[2]arene[3]hydroquinone which can self-assemble to a molecular zipper in the solid state A pillar[2]arene[3]hydroquinone which can self-assemble to a molecular zipper in the solid state Mingguang Pan, Min Xue* Department of Chemistry, Zhejiang University, Hangzhou 310027, P. R. China Fax:

More information

Recent Advances in High Resolution Mass Spectrometry in Environmental and Petrochemical Analyses

Recent Advances in High Resolution Mass Spectrometry in Environmental and Petrochemical Analyses Recent Advances in High Resolution Mass Spectrometry in Environmental and Petrochemical Analyses Future Fuels Institute, Florida State University National High Magnetic Field Laboratory Petroleomics by

More information

Supplementary information. Additional methods: Elemental formula assignments

Supplementary information. Additional methods: Elemental formula assignments Impact of instrument and experiment parameters on reproducibility and repeatability of peaks within ultrahigh resolution ESI FT ICR mass spectra of natural organic matter Melissa C. Kido Soule 1, Krista

More information

MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH. December 19,

MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH. December 19, MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH December 19, 2014 1 The principle of MALDI imaging Spatially resolved mass spectra are recorded Each mass signal represents

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (10) 86 2R QUALIFICATION OF EQUIPMENT ANNEX 7: QUALIFICATION OF MASS SPECTROMETERS Full document title and reference Document

More information

Components of a Mass Spectrometer

Components of a Mass Spectrometer Components of a Mass Spectrometer Sample Introduction Inlet GC LC Direct Insertion (Syringe/Probe) Ionization Ion Separation Ion Detection Ion Source EI,CI,,, MALDI Mass Analyzer Under vacuum TOF, Quadrupole,

More information

Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS

Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS Bruker Daltonics Application Note # MT-93/FTMS-38 Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS Introduction Matrix Assisted Laser Desorption Ionization (MALDI)

More information

Essential Lipidomics Experiments using the LTQ Orbitrap Hybrid Mass Spectrometer

Essential Lipidomics Experiments using the LTQ Orbitrap Hybrid Mass Spectrometer Application Note: 367 Essential Lipidomics Experiments using the LTQ rbitrap Hybrid Mass Spectrometer Thomas Moehring 1, Michaela Scigelova 2, Christer S. Ejsing 3, Dominik Schwudke 3, Andrej Shevchenko

More information

Application Note # MT-111 Concise Interpretation of MALDI Imaging Data by Probabilistic Latent Semantic Analysis (plsa)

Application Note # MT-111 Concise Interpretation of MALDI Imaging Data by Probabilistic Latent Semantic Analysis (plsa) Application Note # MT-111 Concise Interpretation of MALDI Imaging Data by Probabilistic Latent Semantic Analysis (plsa) MALDI imaging datasets can be very information-rich, containing hundreds of mass

More information

Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance

Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance Bruker Daltonics Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance The new ultraflextreme exceeds all current expectations of MALDI-TOF/TOF technology: A proprietary khz smartbeam-ii TM MALDI

More information

Time (min) Supplementary Figure 1: Gas decomposition products of irradiated DMC.

Time (min) Supplementary Figure 1: Gas decomposition products of irradiated DMC. 200000 C 2 CH 3 CH 3 DMC 180000 160000 140000 Intensity 120000 100000 80000 60000 40000 C 2 H 6 CH 3 CH 2 CH 3 CH 3 CCH 3 EMC DEC 20000 C 3 H 8 HCCH 3 5 10 15 20 25 Time (min) Supplementary Figure 1: Gas

More information

New Instruments and Services

New Instruments and Services New Instruments and Services Liwen Zhang Mass Spectrometry and Proteomics Facility The Ohio State University Summer Workshop 2016 Thermo Orbitrap Fusion http://planetorbitrap.com/orbitrap fusion Thermo

More information

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer APPLICATION NOTE 72819 Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer Author Maria Grübner; Thermo Fisher Scientific, Germering,

More information

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Keywords Introduction Application Note PHA0413 High Pressure Liquid Chromatography

More information

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Introduction, Ionization Methods Mass Analyzers, Ion Activation Methods

More information

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography Application Note LCMS-109 A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography Time of Flight (LC-TOF) Mass Spectrometry Introduction Many clinical

More information

2013 Waters Corporation 1

2013 Waters Corporation 1 High Definition Mass Spectrometry (HDMS) application of ion mobility in oil and petroleum analysis Patrice Lemire Waters Corporation WCTOW Vancouver, May 13, 2013 2013 Waters Corporation 1 Introduction

More information

Characterization of an Unknown Compound Using the LTQ Orbitrap

Characterization of an Unknown Compound Using the LTQ Orbitrap Characterization of an Unknown Compound Using the LTQ rbitrap Donald Daley, Russell Scammell, Argenta Discovery Limited, 8/9 Spire Green Centre, Flex Meadow, Harlow, Essex, CM19 5TR, UK bjectives unknown

More information

NCUT National Centre for Upgrading Technology

NCUT National Centre for Upgrading Technology NCUT National Centre for Upgrading Technology a Canada Alberta alliance for bitumen and heavy oil research Comparison of the Reactivity of Naphthenic Acids in Athabasca bitumen and San Joaquin Valley Parviz

More information

FOURIER TRANSFORM MASS SPECTROMETRY

FOURIER TRANSFORM MASS SPECTROMETRY FOURIER TRANSFORM MASS SPECTROMETRY FT-ICR Theory Ion Cyclotron Motion Inward directed Lorentz force causes ions to move in circular orbits about the magnetic field axis Alan G. Marshall, Christopher L.

More information

FOURIER TRANSFORM MASS SPECTROMETRY

FOURIER TRANSFORM MASS SPECTROMETRY FOURIER TRANSFORM MASS SPECTROMETRY FT-ICR Theory Ion Cyclotron Motion Inward directed Lorentz force causes ions to move in circular orbits about the magnetic field axis Alan G. Marshall, Christopher L.

More information

New Instruments and Services

New Instruments and Services New Instruments and Services http://planetorbitrap.com/orbitrap fusion Combining the best of quadrupole, Orbitrap, and ion trap mass analysis in a revolutionary Tribrid architecture, the Orbitrap Fusion

More information

Topic 6 Structure Determination Revision Notes

Topic 6 Structure Determination Revision Notes 1) Introduction Topic 6 Structure Determination Revision Notes Mass spectrometry, infrared spectroscopy and NMR spectroscopy can be used to determine the structure of unknown compounds 2) Mass spectrometry

More information

Comparison of mass spectrometers performances

Comparison of mass spectrometers performances Comparison of mass spectrometers performances Instrument Mass Mass Sensitivity resolution accuracy Quadrupole 1 x 10 3 0.1 Da* 0.5-1.0 pmol DE-MALDI 2 x 10 4 20 ppm 1-10 fmol peptide 1-5 pmol protein Ion

More information

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform Abstract Targeted proteomics for biomarker verification/validation

More information

Amadeo R. Fernández-Alba

Amadeo R. Fernández-Alba % of compounds % of compounds % of compounds % of compounds Amadeo R. Fernández-Alba LC-Orbitrap QExactive Focus Instrumental LOQ 1% 9% 8% 7% 6% 5% 4% 3% 2% 1% %.1 mg/g.2 mg/g Tomato.5 mg/g ddms2 Target

More information

CONTENT. i iv ix. SVKM s NMIMS, School of Pharmacy and Technology Management

CONTENT. i iv ix. SVKM s NMIMS, School of Pharmacy and Technology Management CONTENT Chapter No. Title Page No. Abbreviations List of Figures List of Tables 1 Introduction 1 1.1 Practical aspects 4 1.2 Stability-indicating assay method (SIAM) 5 1.3 Regulatory aspects 6 1.4 Techniques

More information

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Jenny Chen, Hongxia Wang, Zhiqi Hao, Patrick Bennett, and Greg Kilby Thermo Fisher

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2015 Electronic Supplementary Information ovel pseudo[2]rotaxanes constructed by selfassembly of dibenzyl

More information

Ozonolysis of phospholipid double bonds during electrospray. ionization: a new tool for structure determination

Ozonolysis of phospholipid double bonds during electrospray. ionization: a new tool for structure determination Ozonolysis of phospholipid double bonds during electrospray ionization: a new tool for structure determination Michael C. Thomas, Todd W. Mitchell, Stephen J. Blanksby Departments of Chemistry and Biomedical

More information

Chapter 12: Mass Spectrometry: molecular weight of the sample

Chapter 12: Mass Spectrometry: molecular weight of the sample Structure Determination: hapter 12: Mass Spectrometry- molecular weight of the sample; formula hapter 12: Infrared Spectroscopy- indicated which functional groups are present hapter 13: Nuclear Magnetic

More information

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry Dr. Sanjeeva Srivastava 1. Fundamental of Mass Spectrometry Role of MS and basic concepts 2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry 2 1 MS basic concepts Mass spectrometry - technique

More information

Bruker Daltonics. autoflex III smartbeam. The Standard in MALDI-TOF Performance MALDI-TOF/TOF. think forward

Bruker Daltonics. autoflex III smartbeam. The Standard in MALDI-TOF Performance MALDI-TOF/TOF. think forward Bruker Daltonics autoflex III smartbeam The Standard in MALDI-TOF Performance think forward MALDI-TOF/TOF Designed for a Routine High Level of Performance The autoflex III smartbeam brings the power of

More information

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies. Using Software Tools to Improve the Detection of Impurities Application Note Christine Miller Introduction The analysis of raw materials and finished products for or impurities presents a challenge in

More information

Mass Spectrometry. Actual Instrumentation

Mass Spectrometry. Actual Instrumentation Mass Spectrometry Actual Instrumentation August 2017 See also http://www.uni-bielefeld.de/chemie/analytik/ms f additional infmation 1. MALDI TOF MASS SPECTROMETRY ON THE ULTRAFLEX 2 2. ESI MASS SPECTROMETRY

More information

Application Note # ET-17 / MT-99 Characterization of the N-glycosylation Pattern of Antibodies by ESI - and MALDI mass spectrometry

Application Note # ET-17 / MT-99 Characterization of the N-glycosylation Pattern of Antibodies by ESI - and MALDI mass spectrometry Bruker Daltonics Application Note # ET-17 / MT-99 Characterization of the N-glycosylation Pattern of Antibodies by ESI - and MALDI mass spectrometry Abstract Analysis of the N-glycosylation pattern on

More information

Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev

Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev Institute for Biochemical Physics, Rus. Acad. Sci., Moscow, Russia. Institute for Energy Problems

More information

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer Abstract This study demonstrates a sensitive, rapid and reliable research

More information

Application Note # MT-91. High Quality MALDI Imaging of Proteins and Peptides in Small Rodent Organ Tissues. Bruker Daltonics.

Application Note # MT-91. High Quality MALDI Imaging of Proteins and Peptides in Small Rodent Organ Tissues. Bruker Daltonics. Bruker Daltonics Application Note # MT-91 18385 Da 6230 Da 7843 Da High Quality MALDI Imaging of Proteins and Peptides in Small Rodent Organ Tissues New developments in MALDI instrumentation, laser technology

More information

Oxidative Metabolism of Amodiaquine using the ROXY EC System

Oxidative Metabolism of Amodiaquine using the ROXY EC System Application Note Drug Metabolism Electrochemical Reactions upfront MS EC/MS Proteomics & Protein Chemistry S-S bond reduction HDX Peptide bond cleavage Na+, K+ removal Drug-protein binding Lipidomics &

More information

Determination of Amantadine Residues in Chicken by LCMS-8040

Determination of Amantadine Residues in Chicken by LCMS-8040 Liquid Chromatography Mass Spectrometry Determination of Amantadine Residues in Chicken by LCMS-8040 A method for the determination of amantadine in chicken was established using Shimadzu Triple Quadrupole

More information

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement Application Note Food Testing & Agriculture Author Edgar Naegele Agilent Technologies,

More information

SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM

SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM UNITED KINGDOM WOMEN DO NOT DECREASE OVER TWELVE MONTHS

More information

SYNAPT G2-S High Definition MS (HDMS) System

SYNAPT G2-S High Definition MS (HDMS) System SYNAPT G2-S High Definition MS (HDMS) System High performance, versatility, and workflow efficiency of your MS system all play a crucial role in your ability to successfully reach your scientific and business

More information

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application Pharmaceutical Author Cliff Woodward Agilent Technologies, Inc. 285 Centerville

More information

DEPARTMENT OF CHEMISTRY AND CHEMICAL ORGANIC CHEMISTRY II 202-BZG-05 03

DEPARTMENT OF CHEMISTRY AND CHEMICAL ORGANIC CHEMISTRY II 202-BZG-05 03 DEPARTMENT OF CHEMISTRY AND CHEMICAL TECHNOLOGY ORGANIC CHEMISTRY II 202-BZG-05 03 TEST 1 11 MARCH 2010 INSTRUCTOR: I. DIONNE PRINT YOUR NAME: Answers INSTRUCTIONS: Answer all questions in the space provided.

More information

How to Use TOF and Q-TOF Mass Spectrometers

How to Use TOF and Q-TOF Mass Spectrometers How to Use TOF and Q-TOF Mass Spectrometers October 2011 What do TOF and Q-TOF offer? TOF Fast scanning of full spectrum High resolution full scan spectra Accurate mass measurements Q-TOF Fast scanning

More information

Mass Spectrometry Infrastructure

Mass Spectrometry Infrastructure Mass Spectrometry Infrastructure Todd Williams, Ph.D. Director KU Mass Spectrometry and Analytical Proteomics Laboratory Mass Spectrometry Lab B025 Malott Hall Mission The Mass Spectrometry and analytical

More information

1. Sample Introduction to MS Systems:

1. Sample Introduction to MS Systems: MS Overview: 9.10.08 1. Sample Introduction to MS Systems:...2 1.1. Chromatography Interfaces:...3 1.2. Electron impact: Used mainly in Protein MS hard ionization source...4 1.3. Electrospray Ioniztion:

More information

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Methods in Mass Spectrometry Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Sample Engineering Chemistry Biology Life Science Medicine Industry IDF / Police Sample

More information

Supporting Information

Supporting Information Supporting Information Self-hydroxylation of perbenzoic acids at a nonheme iron(ii) center Na Young Oh, a Mi Sook Seo, a Mi Hee Lim, a Mark B. Consugar, b Mi Joo Park, a Jan-Uwe Rohde, b Jaehong Han, a

More information

An optical dosimeter for the selective detection of gaseous phosgene with ultra-low detection limit

An optical dosimeter for the selective detection of gaseous phosgene with ultra-low detection limit Supporting information for An optical dosimeter for the selective detection of gaseous phosgene with ultra-low detection limit Alejandro P. Vargas, Francisco Gámez*, Javier Roales, Tània Lopes-Costa and

More information

Key Words:- Crude oil, Egyptian central Gulf of Suez, Chemical fingerprint, Capillary gas chromatography, Carbon number distribution.

Key Words:- Crude oil, Egyptian central Gulf of Suez, Chemical fingerprint, Capillary gas chromatography, Carbon number distribution. Chemical Fingerprinting of some Egyptian crude oils by gas chromatography Ashraf Yehia El-Naggar Chemistry department, Faculty of Science, Taif University, Kingdom of Saudi Arabia Egyptian Petroleum Research

More information

High-sensitivity Orbitrap mass analysis of intact macromolecular assemblies. R. J. Rose, E. Damoc, E. Denisov, A. Makarov, A. J. R.

High-sensitivity Orbitrap mass analysis of intact macromolecular assemblies. R. J. Rose, E. Damoc, E. Denisov, A. Makarov, A. J. R. High-sensitivity Orbitrap mass analysis of intact macromolecular assemblies R. J. Rose, E. Damoc, E. Denisov, A. Makarov, A. J. R. Heck SUPPLEMENTARY INFORMATION HCD multipole C -trap Transport octapole

More information

Supplementary Information: Liquid-liquid phase coexistence in lipid membranes observed by natural abundance 1 H 13 C solid-state NMR

Supplementary Information: Liquid-liquid phase coexistence in lipid membranes observed by natural abundance 1 H 13 C solid-state NMR Electronic Supplementary Material (ESI) for Physical Chemistry Chemical Physics. This journal is the wner Societies 28 Supplementary Information: Liquid-liquid phase coexistence in lipid membranes observed

More information

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Marta Kozak Clinical Research Applications Group Thermo Fisher Scientific San Jose CA Clinical Research use only, Not for

More information

Identification of Aromatic Fatty Acid Ethyl Esters

Identification of Aromatic Fatty Acid Ethyl Esters Chapter 3.2 Identification of Aromatic Fatty Acid Ethyl Esters The only use of gas chromatography is not sufficient to determine which compounds are eluting from the catalytic bed. At the beginning of

More information

CAMAG TLC-MS INTERFACE

CAMAG TLC-MS INTERFACE CAMAG TLC-MS INTERFACE 93.1 249.2 40 30 97.1 20 10 250.2 0 200 400 m/z WORLD LEADER IN PLANAR-CHROMATOGRAPHY Identification and elucidation of unknown substances by hyphenation of TLC / HPTLC and MS The

More information

LECTURE 3. Ionization Techniques for Mass Spectrometry

LECTURE 3. Ionization Techniques for Mass Spectrometry LECTURE 3 Ionization Techniques for Mass Spectrometry Jack Henion, Ph.D. Emeritus Professor, Analytical Toxicology Cornell University Ithaca, NY 14850 Lecture 3, Page 1 Contents Electron ionization (EI)

More information

Mass Spectrometry Introduction

Mass Spectrometry Introduction Mass Spectrometry Introduction Chem 744 Spring 2013 What MS is and is not MS is NOT a spectroscopic method. Molecules are not absorbing EM radiation MS is the generation, separation and characterization

More information

Mass Spectral Fragmentation Studies of Coumarin-Type Compounds Using GC High-Resolution MS

Mass Spectral Fragmentation Studies of Coumarin-Type Compounds Using GC High-Resolution MS The Open Analytical Chemistry Journal, 2011, 5, 27-36 27 Open Access Mass Spectral Fragmentation Studies of Coumarin-Type Compounds Using GC High-Resolution MS Viorica Lopez-Avila * and George Yefchak

More information

Terminology. Metabolite: substance produced or used during metabolism such as lipids, sugars and amino acids

Terminology. Metabolite: substance produced or used during metabolism such as lipids, sugars and amino acids Terminology Metabolite: substance produced or used during metabolism such as lipids, sugars and amino acids Metabolome: the quantitative complement of all the low molecular weight molecules present in

More information

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Application Note: 346 MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Gargi Choudhary and Diane Cho, Thermo Fisher Scientific, San Jose, CA Wayne Skinner and

More information

Mass Spectrometric Analysis of Eight Common Chemical Explosives Using Ion Trap Mass Spectrometer

Mass Spectrometric Analysis of Eight Common Chemical Explosives Using Ion Trap Mass Spectrometer Mass Spectrometric Analysis of Explosives Bull. Korean Chem. Soc. 2013, Vol. 34, No. 12 3659 http://dx.doi.org/10.5012/bkcs.2013.34.12.3659 Mass Spectrometric Analysis of Eight Common Chemical Explosives

More information

ph Switchable and Fluorescent Ratiometric Squarylium Indocyanine Dyes as Extremely Alkaline Sensors

ph Switchable and Fluorescent Ratiometric Squarylium Indocyanine Dyes as Extremely Alkaline Sensors ph Switchable and Fluorescent Ratiometric Squarylium Indocyanine Dyes as Extremely Alkaline Sensors Jie Li, Chendong Ji, Wantai Yang, Meizhen Yin* State Key Laboratory of Chemical Resource Engineering,

More information

Edgar Naegele. Abstract

Edgar Naegele. Abstract Simultaneous determination of metabolic stability and identification of buspirone metabolites using multiple column fast LC/TOF mass spectrometry Application ote Edgar aegele Abstract A recent trend in

More information

Supporting information

Supporting information Supporting information Figure legends Supplementary Table 1. Specific product ions obtained from fragmentation of lithium adducts in the positive ion mode comparing the different positional isomers of

More information

Characterizing fatty acids with advanced multinuclear NMR methods

Characterizing fatty acids with advanced multinuclear NMR methods Characterizing fatty acids with advanced multinuclear NMR methods Fatty acids consist of long carbon chains ending with a carboxylic acid on one side and a methyl group on the other. Most naturally occurring

More information

Heparin Sodium ヘパリンナトリウム

Heparin Sodium ヘパリンナトリウム Heparin Sodium ヘパリンナトリウム Add the following next to Description: Identification Dissolve 1 mg each of Heparin Sodium and Heparin Sodium Reference Standard for physicochemical test in 1 ml of water, and

More information

Vitamin D3 and related compounds by ESI and APCI

Vitamin D3 and related compounds by ESI and APCI Liquid Chromatography Mass Spectrometry SSI-LCMS-9 Vitamin D and related compounds by ESI and APCI LCMS-8 Summary Vitamin D and related compounds were measured by LC-ESI/APCI-MS-MS. Background Accurate

More information

Choosing the metabolomics platform

Choosing the metabolomics platform Choosing the metabolomics platform Stephen Barnes, PhD Department of Pharmacology & Toxicology University of Alabama at Birmingham sbarnes@uab.edu Challenges Unlike DNA, RNA and proteins, the metabolome

More information

Proficiency testing performance of Turkish laboratories on determination of relative composition of fatty acids in sunflower oil

Proficiency testing performance of Turkish laboratories on determination of relative composition of fatty acids in sunflower oil ORIGINAL ARTICLE J. Chem. Metrol. 11:2 (2017) 40-45 Proficiency testing performance of Turkish laboratories on determination of relative composition of fatty acids in sunflower oil Hasibe Yilmaz 1*, Simay

More information

CHEM 242 UV-VIS SPECTROSCOPY, IR SPECTROSCOPY, CHAP 13A ASSIGN AND MASS SPECTROMETRY C 8 H 17 A B C

CHEM 242 UV-VIS SPECTROSCOPY, IR SPECTROSCOPY, CHAP 13A ASSIGN AND MASS SPECTROMETRY C 8 H 17 A B C CEM 242 UV-VIS SPECTRSCPY, IR SPECTRSCPY, CAP 13A ASSIGN AND MASS SPECTRMETRY 1. Which choice matches the max of each steroid in the order given? C 8 17 C 8 17 C 8 17 A B C A. 209 nm 241 nm 284 nm B. 241

More information

CEE 697z Organic Compounds in Water and Wastewater

CEE 697z Organic Compounds in Water and Wastewater Print version CEE 697z Organic Compounds in Water and Wastewater NOM and MS Methods Lecture #10 NOM Characterization Analytical Tests elemental analysis spectral properties functional group chemistry Separation/Fractionation

More information

CEE 697z Organic Compounds in Water and Wastewater

CEE 697z Organic Compounds in Water and Wastewater Print version CEE 697z Organic Compounds in Water and Wastewater NOM and MS Methods Lecture #10 NOM Characterization Analytical Tests elemental analysis spectral properties functional group chemistry Separation/Fractionation

More information

Sulfate Radical-Mediated Degradation of Sulfadiazine by CuFeO 2 Rhombohedral Crystal-Catalyzed Peroxymonosulfate: Synergistic Effects and Mechanisms

Sulfate Radical-Mediated Degradation of Sulfadiazine by CuFeO 2 Rhombohedral Crystal-Catalyzed Peroxymonosulfate: Synergistic Effects and Mechanisms Supporting Information for Sulfate Radical-Mediated Degradation of Sulfadiazine by CuFeO 2 Rhombohedral Crystal-Catalyzed Peroxymonosulfate: Synergistic Effects and Mechanisms Submitted by Yong Feng, Deli

More information

Supporting Information. First synthetic entry to the trimer stage of 5,6-dihydroxyindole polymerization: orthoalkynylaniline-based

Supporting Information. First synthetic entry to the trimer stage of 5,6-dihydroxyindole polymerization: orthoalkynylaniline-based Supporting Information First synthetic entry to the trimer stage of 5,6-dihydroxyindole polymerization: orthoalkynylaniline-based access to the missing 2,7 :2,7 -triindole Luigia Capelli, Paola Manini,*

More information

Supporting Information for. Boronic Acid Functionalized Aza-Bodipy (azabdpba) based Fluorescence Optodes for the. analysis of Glucose in Whole Blood

Supporting Information for. Boronic Acid Functionalized Aza-Bodipy (azabdpba) based Fluorescence Optodes for the. analysis of Glucose in Whole Blood Supporting Information for Boronic Acid Functionalized Aza-Bodipy (azabdpba) based Fluorescence Optodes for the analysis of Glucose in Whole Blood Yueling Liu, Jingwei Zhu, Yanmei Xu, Yu Qin*, Dechen Jiang*

More information

AB Sciex QStar XL. AIMS Instrumentation & Sample Report Documentation. chemistry

AB Sciex QStar XL. AIMS Instrumentation & Sample Report Documentation. chemistry Mass Spectrometry Laboratory AIMS Instrumentation & Sample Report Documentation AB Sciex QStar XL chemistry UNIVERSITY OF TORONTO AIMS Mass Spectrometry Laboratory Department of Chemistry, University of

More information

High-Throughput Analysis of Oligonucleotides using Automated Electrospray Ionization Mass Spectrometry

High-Throughput Analysis of Oligonucleotides using Automated Electrospray Ionization Mass Spectrometry High-Throughput Analysis of Oligonucleotides using Automated Electrospray Ionization Mass Spectrometry Mark E. Hail 1, Brian Elliott 2, Kerry Nugent 3, Jeffrey L. Whitney 1, and David J. Detlefsen 1 1

More information

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By Dr. Syed Tufail Hussain Sherazi E-mail: tufail_sherazi@yahoo.com National Center of Excellence in

More information

Singlet oxygen photosensitisation by the fluorescent probe Singlet Oxygen Sensor Green

Singlet oxygen photosensitisation by the fluorescent probe Singlet Oxygen Sensor Green Singlet oxygen photosensitisation by the fluorescent probe Singlet Oxygen Sensor Green Xavier Ragàs, Ana Jiménez-Banzo, David Sánchez-García, Xavier Batllori and Santi Nonell* Grup d Enginyeria Molecular,

More information

Increased Identification Coverage and Throughput for Complex Lipidomes

Increased Identification Coverage and Throughput for Complex Lipidomes Increased Identification Coverage and Throughput for Complex Lipidomes Reiko Kiyonami, David Peake, Yingying Huang, Thermo Fisher Scientific, San Jose, CA USA Application Note 607 Key Words Q Exactive

More information

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection Application Note Cannabis Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection Authors Mike Adams, Annette Roth, Sue D

More information

Single Quadrupole Compact Mass Spectrometer for Liquid Sample

Single Quadrupole Compact Mass Spectrometer for Liquid Sample Single Quadrupole Compact Mass Spectrometer for Liquid Sample YL INSTRUMENT CO., LTD. YL9900 MS quickly provides essential information and improves the chemist's workflow. Mordern organic labs have become

More information

Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution

Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution A simple definition of a Mass Spectrometer A Mass Spectrometer is an analytical instrument that can separate charged molecules according

More information

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Identification of Steroids in Water by Ion Trap LC/MS/MS Application Identification of Steroids in Water by Ion Trap LC/MS/MS Application Environmental Author Paul Zavitsanos and Christine Miller Agilent Technologies, Inc. Centerville Road Wilmington, DE 9- USA Abstract

More information

More structural information with MS n

More structural information with MS n PRODUCT SPECIFICATIONS The LTQ XL linear ion trap mass spectrometer More structural information with MS n The LTQ XL linear ion trap mass spectrometer delivers more structural information faster and with

More information

Supporting information for the article

Supporting information for the article Supporting information for the article S1 Exceptional behavior of Ni 2 O 2 species revealed by ESI-MS and MS/MS studies in solution. Application of superatomic core to facilitate new chemical transformations

More information

Organic Chemistry Laboratory Fall Lecture 3 Gas Chromatography and Mass Spectrometry June

Organic Chemistry Laboratory Fall Lecture 3 Gas Chromatography and Mass Spectrometry June 344 Organic Chemistry Laboratory Fall 2013 Lecture 3 Gas Chromatography and Mass Spectrometry June 19 2013 Chromatography Chromatography separation of a mixture into individual components Paper, Column,

More information

New Mass Spectrometry Tools to Transform Metabolomics and Lipidomics

New Mass Spectrometry Tools to Transform Metabolomics and Lipidomics New Mass Spectrometry Tools to Transform Metabolomics and Lipidomics July.3.13 Ken Miller Vice President of Marketing, Life Sciences Mass Spectrometry 1 The world leader in serving science Omics & the

More information

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography Application Note Metabolomics Author Yuqin Dai Agilent Technologies, Inc. 31 Stevens Creek Blvd,

More information

An Alternative Approach: Top-Down Bioanalysis of Intact Large Molecules Can this be part of the future? Lecture 8, Page 27

An Alternative Approach: Top-Down Bioanalysis of Intact Large Molecules Can this be part of the future? Lecture 8, Page 27 An Alternative Approach: Top-Down Bioanalysis of Intact Large Molecules Can this be part of the future? Lecture 8, Page 27 Top-down HRAM Bioanalysis of Native Proteins/Molecules Relative Abundance 100

More information

LC-MS Analysis of Botanicals

LC-MS Analysis of Botanicals Botanical workshop UAB, Sept 11, 26 LC-MS Analysis of Botanicals Jeevan K. Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB Purdue-UAB Botanicals Center for Age-related Disease Applications

More information

Supporting Information

Supporting Information Supporting Information Wiley-VCH 2008 69451 Weinheim, Germany Supporting Information Enantioselective Cu-catalyzed 1,4-Addition of Various Grignard Reagents to Cyclohexenone using Taddol-derived Phosphine-Phosphite

More information

Supporting information for. Determination of sub nm levels of low molecular mass (LMM) thiols in natural waters by liquid

Supporting information for. Determination of sub nm levels of low molecular mass (LMM) thiols in natural waters by liquid Supporting information for Determination of sub nm levels of low molecular mass (LMM) thiols in natural waters by liquid chromatography tandem mass spectrometry after derivatization with p-(hydroxymercuri)

More information

CHM 424L Organic Laboratory, Dr. Laurie S. Starkey Introduction to Mass Spectrometry

CHM 424L Organic Laboratory, Dr. Laurie S. Starkey Introduction to Mass Spectrometry CM 424L rganic Laboratory, Dr. Laurie S. Starkey Introduction to Mass Spectrometry Mass spectrometry is used to determine a sample's molecular mass and molecular formula. Some structural information can

More information