Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

Size: px
Start display at page:

Download "Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation"

Transcription

1 Laboratory Hematology 8: Carden Jennings Publishing Co.. Ltd. Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation T. SCHISANO, L. VAN HOVE Abbott Diagnostics, Santa Clara, California ABSTRACT Reticulocyte enumeration has been shown to be a reliable indicator of changes in erythropoietic activity in the bone marrow. For many years the clinical laboratory has relied on the manual slide method for enumerating reticulocytes; this method is both time consuming and imprecise. With the advent of automated and semiautomated reticulocyte methods, the clinical laboratory has the ability to provide accurate and precise reticulocyte results in a timely and efficient manner. Abbott has 3 systems that provide an automated reticulocyte assay they are the Cell- Dyn 4000, Cell-Dyn 3700, and the Cell-Dyn The Abbott Cell-Dyn 4000 is a fully automated random-access hematology analyzer, for use in a high-throughput clinical setting, that provides basic hematologic analysis with a rapid and sensitive reticulocyte method. The Cell-Dyn 4000 system directly quantifies reticulocytes using fluorescence technology with an RNA-bound dye complex. Since the release of the Cell-Dyn 4000, Abbott has added a supravital staining method using New Methylene Blue dye to the Cell-Dyn 3700 and recently to the Cell-Dyn We compared blood specimens from healthy and anemic individuals for correlations in reticulocyte percentage and absolute count using Abbott s Cell-Dyn 4000, Cell-Dyn 3700, and Cell-Dyn 3200 reticulocyte methods against the manual (microscopic) reticulocyte slide method as well as performing intermethod comparisons of the Cell-Dyn systems. In addition, we evaluated imprecision of each method and developed normal reference ranges for the Cell-Dyn and MAPSS are registered trademarks of Abbott Laboratories. Correspondence and reprint requests: Thomas V. Schisano, Abbott Diagnostics, 5440 Patrick Henry Dr, Santa Clara, CA Received February 8, 2002; accepted February 22, 2002 reticulocyte parameters based on results from in-house donors at Abbott Diagnostics in Santa Clara. The data revealed that the Abbott Cell-Dyn methods are accurate, showing excellent correlation between the manual (microscopic) method and intermethod assays; the Abbott methods also produced lower levels of imprecision than the manual (microscopic) reticulocyte slide method. The Abbott methods provide the clinician with a more reliable means of evaluating erythropoietic activity both in anemia and in bone marrow transplantation patients than does the tedious and relatively imprecise manual method. Lab Hematol. 2002;8:85-90 KEYWORDS: Reticulocyte Normal reference range INTRODUCTION Reticulocytes are immature erythrocytes in the bone marrow, at the end-stage of hemoglobin synthesis, released into the peripheral blood stream. Reticulocyte enumeration has been shown to be a reliable indicator of changes in erythropoetic activity in the bone marrow [1-8]. Early-stage reticulocytes contain greater amounts of ribosomes, mitochondria, and Golgi bodies compared to late-stage reticulocytes. This material is made easily visible for microscopic enumeration by precipitation with supravital stain techniques such as New Methylene Blue and with fluorescent staining dyes. The degree of precipitation and fluorescent staining is directly proportional to the amount of these cellular components. Until the advent of the automated procedures, the clinical laboratory had to rely on the manual technique, which has shown technical problems with both accuracy and precision [9,10]. The purpose of this study was to compare the clinical performance of the Abbott Cell-Dyn 4000 fluorescence method (Abbott Diagnostics, Santa Clara, CA)-along with the 85

2 86 T Schisano and L.Van Hove supravital staining method of the Cell-Dyn 3700 and Cell- Dyn 3200-against the manual (microscopic) New Methylene Blue method. We also compared the Abbott Cell-Dyn methods against each other. To establish the reticulocyte parameter normal reference ranges for each Cell-Dyn system, we obtained 67 whole blood specimens from healthy in-house donors at Abbott Diagnostics in Santa Clara. We obtained the central-core 95% of both the male and female groups by calculating the 2.5. and 97.5-percentile limits. For our abnormal sampling we obtained 41 specimens from our contracted clinical sites in the San Francisco Bay area. All samples were run in duplicate. Linear regression analysis between methods and paireddifference precision analysis were performed. METHODS Sample Selection Sixty-seven in-house donors from Abbott Diagnostics in Santa Clara who were considered in general good health were used for the normal population sampling for our normal-range evaluation using an institutional review board-approved protocol. We selected 33 males and 34 females between the ages of 20 and 60 years. Forty-one abnormal unlinked retention specimens collected from our regional contracted sites were used for the abnormal population sampling in this study. These abnormal specimens contained various anemia cases including iron deficiency, sickle cell anemia, thalassemia, anemias of chronic disease, and normocytic and macrocytic anemias. All samples were run in duplicate and were collected in tubes containing K 3 EDTA anticoagulant. Reticulocyte Counting Methods Abbott Cell-Dyn All blood samples were run using R8-2E version software. The Cell-Dyn 4000 integrates fully automated random-access reticulocyte analysis with basic hematology on a single analyzer platform. This integration allows the user the option of running a reticulocyte with a complete blood count (CBC) using only ul of whole blood. No added steps are required in sample preparation. The absolute and percentage reticulocyte, along with the IRF (immature reticulocyte fraction), are included. The Cell-Dyn 4000 reticulocyte method uses an argon laser and fluorescence flow cytometry by incorporating a DNA- and RNAbinding dye that is excited at 488 nm and emits enhanced fluorescence at 530 nm. The intensity of the fluorescence that is emitted from the RNA-bound dye complex is directly proportional to the RNA concentration of the reticulocyte. Abbott Cell-Dyn All blood samples were run using 1.1 version software. The Cell-Dyn 3700 provides a semiautomated reticulocyte count based on the National Committee for Laboratory Standards and the International Council for Standardization in Haematology (NCCLS/ICSH) H44-A guideline for supravital staining. The method involves a single dilution of blood using 20 ul of whole blood in the reticulocyte staining reagent thiazine dye New Methylene Blue N. Au incubation time of approximately 30 minutes at room temperature was used. The user then selected a single soft key to initiate the reticulocyte analysis. Both red blood cells and reticulocytes are measured by multiangle polarized scatter separation (MAPSS) technology (Abbott Diagnostics) using an HeNe (helium-neon) laser. Red blood cell information is collected for each cell and the red blood cells and reticulocytes are distinguished from other cells and interferences through the use of algorithm gating and threshold setpoints. Reticulocytes are reported in percentages. The IRF is provided and the instrument will automatically calculate the reticulocyte absolute value if a red blood cell (RBC) count is entered. Abbott Cell-Dyn All blood samples were run using prerelease software version 1.92, intended for research use only, and were rested in this study prior to the release of software version 2.0, which officially added the reticulocyte assay to the Cell-Dyn As with the Cell-Dyn 3700, the Cell- Dyn 3200 provides a semiautomated reticulocyte count based on the NCCLS/ICSH H44-A guideline for supravital staining. The method involves a single dilution of blood using 20 ul of whole blood in the reticulocyte staining reagent thiamine dye New Methylene Blue N. After an incubation time of approximately 30 minutes at room temperature, reticulocyte analysis of the sample is initiated when the user selects a single soft key. Both red blood cells and reticulocytes are measured by MAPSS technology using an HeNe laser. RBC information is collected for each cell and the red blood cells and reticulocytes are distinguished from other cells and interferences through the use of algorithm gating and threshold setpoints. Reticulocytes are reported in percent. The instrument will automatically calculate the reticulocyte absolute value if an RBC count is entered. Manual (Microscopic) Method. Manual reticulocyte counts were performed following the NCCLS guidelines. One thousand RBCs were counted. Duplicate slides were made for each sample and both slides were counted. Data analysis results were analyzed using duplicate run samples for correlations, paired difference precision, and percentile ranking using Excel 2000 (Microsoft, Redmond, WA). RESULTS Linear regression analysis using the manual (microscopic) reticulocyte counting method compared to the Abbott Cell- Dyn methods gave an r range between and for reticulocyte percentage counts. Intermethod correlation of the reticulocyte percentage counts observed when comparing the Abbott Cell-Dyn 4000 against the Cell-Dyn 3700 and Cell-Dyn 3200 gave an r range between and The Cell.-Dyn 3700 versus Cell-Dyn 3200 reticulocyte percentage counts gave an r of Intermethod correlation

3 Abbott Cell-Dyn Reticulocyte Method and Normal Reference Range Evaluation FIGURE I. Manual versus Cell-Dyn reticulocyte percentage. Regression analyses for the manual reticulocyte percentage (Retic %) method compared to the Cell-Dyn Systems: A, Cell-Dyn 4000; B, Cell-Dyn 3700; C, Cell-Dyn of the reticulocyte absolute counts observed comparing the Abbott Cell-Dyn 4000 versus the Cell-Dyn 3700 and Cell- Dyn 3200 gave an r range between and The Cell-Dyn 3700 versus Cell-Dyn 3200 reticulocyte absolute counts gave an r of Linear regression analysis of the microscopic manual reticulocyte percentage counting method against all the Abbott Cell-Dyn reticulocyte percentage (retic %) methods and the intermethod reticulocyte percentage correlations arc shown in Figures 1 and 2 and summarized in Tables 1 and 2. Correlations of the Abbott Cell-Dyn 4000 reticulocyte absolute count against the Cell-Dyn 3700, and Cell-Dyn 3200 and FIGURE 2. Cell-Dyn versus Cell-Dyn reticulocyte percentage Regression analyses for the reticulocyte percentage (Retic %) intermethod comparison of the Cell-Dyn systems: A, Cell- Dyn 4000 versus Cell-Dyn 3700; B, Cell-Dyn 4000 versus Cell-Dyn 3200; C, Cell-Dyn 3700 versus Cell-Dyn TABLE 1. Regression Summary: Manual versus Cell-Dyn Reticulocyte Percentage (Retic %) (N = 216)

4 88 T. Schisano and L.Van Hove TABLE 2. Regression Summary: Cell-Dyn versus Cell-Dyn Reticulocyte Percentage (Retic 96) (N = 216) TABLE 3. Regression Summary: Cell-Dyn versus Cell-Dyn Reticulocyte Absolute (N = 216) Cell-Dyn 3700 reticulocyte absolute counts against the Cell- Dyn 3200 are shown in Figure 3 and summarized in Table 3. The normal reference range evaluation showed excellent tracking between methods and established the normal reference range for reticulocyte parameters on the Abbott Cell- Dyn platforms. Our sample population of 67 in-house donors from Abbott Diagnostics, which contained 33 male and 34 female donors, represents a typical population sampling found in the San Francisco Bay Area of California. Table 4 contains data from the total population of 67 donors, run in duplicate for a total of 134 data points. Tables 5 and 6 summarize the data from 33 male and 34 female donors taken from the total group; they summarize the 95% central core of the normal population as shown by the 2.5- and 97.5-percentile limits. These results, shown in Tables 4 through 6, verified across all platforms what is reported in the literature [11]. The 33 male versus 34 female results in Tables 5 and 6 show slightly higher reticulocyte percentage ranges for each system in the female group, as would be expected [11]. Paired-difference precision showed a coefficient of variation (CV) at 95% between 6.06 and 7.66 for both reticulocyte percentage and absolute counts across all methods excluding the manual method, which showed a CV at 95% of for percentage reticulocytes. The paired-difference precision showed that each Cell-Dyn method has a low level of imprecision, and results are well within the manufacturer s specification whereas-as expected-the manual count has a higher level of imprecision, more than double that of the Cell-Dyn methods. The imprecision of the manual method has long been established [9-10]. Paired-difference precision is summarized in Table 7. TABLE 4. Total Population Normal Reference Range: 95% Central Core ( Percentile Limits) FlGURE 3. Cell-Dyn versus Cell-Dyn reticulocyte absolute count. Regression analyses for the reticulocyte absolute count (Retic Abs) intermethod comparison of the Cell-Dyn systems: A. Cell-Dyn 4000 versus Cell-Dyn 3700; B, Cell-Dyn 4000 versus Cell-Dyn 3200; C, Cell-Dyn 3700) versus Cell-Dyn 3200.

5 Abbott Cell-Dyn Reticulocyte Method and Normal Reference Range Evaluation 89 TABLE 5. Male Normal Reference Range: 95% Central Core ( Percentile Limits) TABLE 7. Paired Difference Precision: % Coefficient of Variation = DISCUSSION Our data showed good correlation between the Cell-Dyn systems and the manual (microscopic) method for reticulocyte percent. The intermethod correlations between analyzers were also good and showed a better correlation of the Cell- Dyn 3700 against the Cell-Dyn This correlation exists because both systems use the same technology and the same reticulocyte-staining reagent thiamine dye New Methylene Blue N. Even against the fluorescence flow cytometric method of the Cell-Dyn 4000, both the Cell-Dyn 3700 and Cell-Dyn 3200 correlated well. Reticulocyte enumeration has been well documented in the literature for its use in indicating changes in erythropoietic activity in the bone marrow, diagnosing anemia, and monitoring bone marrow transplantation patients. In the past, the clinical laboratory has relied upon the manual (slide) method to generate a reticulocyte count for the clinician. The manual method is time consuming and thus decreases the throughput in the clinical laboratory. It also has been shown, as in this evaluation, to be more imprecise than automated methods [9,10]. Attempts have been made by other hematology instrument manufacturers to automate the reticulocyte parameter and thus allow the clinical laboratory to maintain a relatively high throughput and improve precision. In the past, standalone reticulocyte analyzers were developed using flow cytometry to measure the staining effect on the reticulocyte with the fluorochrome dye Auramine-0. Although automated, these systems are not designed to allow for generation of multiple hematology parameters and require the addition of a red cell count from an outside source to generate the absolute reticulocyte count. Semiautomated supravital staining methods have been developed by various manufacturers that are run on multiparameter hematology analyzers (such as the Cell-Dyn 3700 and Cell-Dyn 3200) and allow a full CBC report with the whole blood sample. The Cell-Dyn 4000 was the first system to integrate fully automated random-access reticulocyte analysis with basic hematology on a single analyzer platform. We evaluated this technology with the semiautomated supravital staining method of the Cell-Dyn 3700 and the prereleased semiautomated supravital staining method of the Cell-Dyn Both the Cell-Dyn 3700 and Cell-Dyn 3200 Systems use MAPSS technology and easy access into the reticulocyte mode on a multi-hematology parameter platform. The results showed good correlation between systems, with good correlation to the manual reticulocyte method. Intermethod paired-difference precision was excellent and well within manufacturer s specification whereas-as expectedthe manual slide method showed a greater level of imprecision. The data confirm the sensitivity of the fluorescence method [8]. The goals of this cross-platform reticulocyte evaluation were to compare the clinical performance of each of the 3 Cell-Dyn Systems and to establish normal reference ranges for the reticulocyte parameters. This study accomplished these goals and showed that each system maintains a high level of accuracy with a low level of imprecision. Any clinical hematology laboratory can rely on each of these platforms to enhance the throughput of their reticulocyte assay and can use each system interchangeably in a multianalyzer environment provided that the Cell-Dyn system-specific normal ranges are used. The addition of the reticulocyte assay to the Cell-Dyn 3200 will enhance the capabilities of this platform and further improve on a well-established technology. REFERENCES TABLE 6. Female Normal Reference Range: 95% Central Core ( Percentile Limits)

6 90 T Schisano and L.Van Hove

CELL-DYN 3700 Strength in Technology, Proven Reliability

CELL-DYN 3700 Strength in Technology, Proven Reliability CELL-DYN 3700 Strength in Technology, Proven Reliability Optical WBC Technology Patented MAPSS Differential Multiple Technologies CELL-DYN 3700 Multiple Technologies One Superior Result Multiple Technologies

More information

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies CELL-DYN 3700 Strength in Technology, Proven Reliability Optical WBC Technology Patented M.A.P.S.S. Differential Multiple Technologies CELL-DYN 3700 Multiple Technologies One Superior Result Multiple Technologies

More information

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom ISLH Laboratory Hematology 7:75-80 2001 Carden Jennings Publishing Co., Ltd. Official Publication Comparison of the Automated Reticulocyte Counts and Immature Reticulocyte Fraction Measurements Obtained

More information

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD

More information

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa The Role of the FBC in clinical decision making History Examination Investigations Decision 70% FBC Laboratory

More information

The Evolving Role of Reticulated Platelets

The Evolving Role of Reticulated Platelets The Evolving Role of Kyle Riding, PhD, MLS(ASCP) CM University of Central Florida Session Objectives Platelets Recall the process involved in megakaryopoiesis and thrombocytopoiesis Explain the various

More information

Hematology Unit Lab 1 Review Material

Hematology Unit Lab 1 Review Material Hematology Unit Lab 1 Review Material - 2018 Objectives Laboratory instructors: 1. Assist students during lab session Students: 1. Review the introductory material 2. Study the case histories provided

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Lavender Top Management SUCCESS BEYOND FINANCES

Lavender Top Management SUCCESS BEYOND FINANCES Lavender Top Management Alfonso Ziccardi Ass.Lab.Operation Manager for AP/CP/Safety Officer New York Hospital Medical Center of Queens SUCCESS BEYOND FINANCES 1 . VISION Mission/ Vision To successfully

More information

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND ADVANCED HAEMATOLOGY BATTLE OF THE BANDS Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND The identification of immature neutrophils (bands, metamyelocytes,

More information

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Sysmex America White Paper One Nelson C. White Parkway,

More information

Performance evaluation of Celltac G: a new automated hematology analyzer

Performance evaluation of Celltac G: a new automated hematology analyzer Original Article Performance evaluation of Celltac G: a new automated hematology analyzer Masaaki Sugiyama 1, Tsukasa Kobayashi 1, Yuki Jisyage 2, Shigeko Yamamoto 2, Yutaka Nagai 2,3 and Hiroshi Kondo

More information

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision Lavender Top Management Alfonso Ziccardi Lab.Operation Manager for AP/CP/Safety Officer New York Hospital Medical Center of Queens SUCCESS BEYOND FINANCES Mission/ Vision. VISION To successfully develop

More information

Role of immature reticulocyte fraction in evaluation of aplastic anemia in cases of pancytopenia

Role of immature reticulocyte fraction in evaluation of aplastic anemia in cases of pancytopenia Original article Role of immature reticulocyte fraction in evaluation of aplastic anemia in cases of pancytopenia Dr. R. Sindhu, S.K. Behera, D.P. Mishra Department of Pathology, MKCG medical college,

More information

Hematology & Coagulation Practicum Objectives CLS - 647

Hematology & Coagulation Practicum Objectives CLS - 647 Hematology & Coagulation Practicum Objectives CLS - 647 The following objectives are to be completed by the student for successful completion of this clinical rotation. The objectives within the psychomotor

More information

QIAsymphony DSP Circulating DNA Kit

QIAsymphony DSP Circulating DNA Kit QIAsymphony DSP Circulating DNA Kit February 2017 Performance Characteristics 937556 Sample to Insight Contents Performance Characteristics... 4 Basic performance... 4 Run precision... 6 Equivalent performance

More information

History. n 1865 First description of reticulocytes by Erb, who. n 1881 Using supravital staining, Ehrlich demonstrated

History. n 1865 First description of reticulocytes by Erb, who. n 1881 Using supravital staining, Ehrlich demonstrated SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT JANUARY 2016 SEED HAEMATOLOGY The importance of reticulocyte detection Production of reticulocytes All blood cells emanate from a stem cell. Under pronounced

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: BD FACSCount Instrument System with FACSCount Control Kit and BD FACSCount Reagent Kit (Absolute CD4+, CD8+, and CD3+ Counts) Number:

More information

RED BLOOD CELLS IMPORTANCE OF FILM EXAMINATION: PART ONE

RED BLOOD CELLS IMPORTANCE OF FILM EXAMINATION: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk RED BLOOD CELLS IMPORTANCE OF FILM EXAMINATION: PART ONE Author : Mark Richer Categories : Vets Date : March 25, 2013 Mark

More information

Blood DLC, Retic count, PCV, Hb and ESR. Dr. Tamara Alqudah

Blood DLC, Retic count, PCV, Hb and ESR. Dr. Tamara Alqudah Blood DLC, Retic count, PCV, Hb and ESR Dr. Tamara Alqudah Differential Leukocyte Count (DLC) There are 5 main types of WBCs: 1. Neutrophils: 40-80% 2. Eosinophils: 1-6 % 3. Basophils: < 1-2% 4. Lymphocytes:

More information

Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies

Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies HEMATOPATHOLOGY Evaluation of Automated Reticulocyte Counts and Their Reliability in the Presence of Howell-Jolly Bodies KAREN G. LOFSNESS, MS, MONICA L. KOHNKE, BS, AND NANCY A. GEIER, BS An automated

More information

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S 111 Riv Med Lab - JLM, Vol. 4, N. 2, 2003 First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S H. Vandeputte a, D. Darre

More information

10/30/2015. XN Case Studies: Every Picture Tells a Story

10/30/2015. XN Case Studies: Every Picture Tells a Story XN Case Studies: Every Picture Tells a Story Jill Crist MT(ASCP)Field Product Specialist OBJECTIVES Explain how scattergrams and histogram pictures can provide great insight into abnormal hematology samples

More information

USING RETICULOCYTE INDICES TO IDENTIFY α-thalassemia A PRELIMINARY REPORT

USING RETICULOCYTE INDICES TO IDENTIFY α-thalassemia A PRELIMINARY REPORT USING RETICULOCYTE INDICES TO IDENTIFY α-thalassemia A PRELIMINARY REPORT Suphan Soogarun 1, Jamsai Suwansaksri 2 and Viroj Wiwanitkit 3 1 Department of Clinical Microscopy, Faculty of Allied Health Sciences,Chulalongkorn

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

This amount has to be newly synthesized each day because:

This amount has to be newly synthesized each day because: Hemoglubin synthesis The circulation blood of normal adult contain about 750 mg of hemoglobin and of this about 7 8 gm are degraded daily. This amount has to be newly synthesized each day because: 1) The

More information

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How?

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How? Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How? Dennis B. DeNicola, DVM, PhD, DACVP Chief Veterinary Educator IDEXX Laboratories, Inc. Westbrook, ME USA Adjunct Professor of Veterinary

More information

Biological Variations of Hematologic Parameters Determined by UniCel DxH 800 Hematology Analyzer

Biological Variations of Hematologic Parameters Determined by UniCel DxH 800 Hematology Analyzer Biological Variations of Hematologic Parameters Determined by UniCel DxH 800 Hematology Analyzer Pianhong Zhang, PhD; Huqiang Tang, MS, MPH; Keqing Chen, MS; Yingying Chen, BS; Dongsheng Xu, MD, PhD Context.

More information

Please contact Customer Service with any questions. Ph: option 5

Please contact Customer Service with any questions. Ph: option 5 Sarasota Memorial Laboratory Services is pleased to announce that we are now performing the following testing in house: Immature Platelet Fraction (IPF) The Immature Platelet Fraction (IPF) is used to

More information

PointCare NOW TM Technical Note

PointCare NOW TM Technical Note PointCare NOW TM Technical Note Introduction to our Technology And Summary of Evaluation and Performance data 1 of 19 Introduction The PointCare NOW is a portable, and very fast, diagnostic instrument

More information

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples Hematopathology / HEMOGLOBIN-BASED BLOOD SUBSTITUTES Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples Jolanta Kunicka, PhD, 1 Michael Malin, PhD, 1 David Zelmanovic,

More information

Automated Hematology Analyzer. Differentiate with the XT-4000i

Automated Hematology Analyzer. Differentiate with the XT-4000i XT-4000i Automated Hematology Analyzer Differentiate with the XT-4000i Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady

More information

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags. WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags www.sysmex.com/us Introduction For many decades, physicians have relied on

More information

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease

More information

Human Hemoglobin Colorimetric Detection Kit

Human Hemoglobin Colorimetric Detection Kit Human Hemoglobin Colorimetric Detection Kit CATALOG NO: IRAAKT2522 LOT NO: SAMPLE INTENDED USE The Hemoglobin detection kit is designed to quantitatively measure all forms of hemoglobin present in blood

More information

Mission Hb accurately detects both

Mission Hb accurately detects both Mission Digital Hb-Testing System: Save Cost by testing 2- Parameters with Single Strip real time ie: i.e: Hemoglobin & Hematocrit-[PCV]. Presented By : Remedy Healthcare Private Limited, New Delhi, [INDIA].

More information

Technical Bulletin No. 157

Technical Bulletin No. 157 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 157 January 11, 2017 Flow Cytometry Lymphocyte Subset Analysis Testing Platform Change Change Effective Date: February 6, 2017 Methods

More information

The Egyptian Journal of Hospital Medicine (July 2017) Vol.68 (3), Page

The Egyptian Journal of Hospital Medicine (July 2017) Vol.68 (3), Page The Egyptian Journal of Hospital Medicine (July 2017) Vol.68 (3), Page 1358-1363 Study of Automated Red Blood Cells Parameters in Correlation with Routine RBCs Morphology by Smear Review Tahany Ali El

More information

CASE STUDIES PERIPHERAL BLOOD AND BODY FLUIDS

CASE STUDIES PERIPHERAL BLOOD AND BODY FLUIDS CASE STUDIES PERIPHERAL BLOOD AND BODY FLUIDS WHERE TO START anemias hemoglobinopathies new and old parameters uncommon things fluids.benign and malignant.yuk! really annoying stuff and maybe some entertainment

More information

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine Do Your Flow Cytometric LDTs Conform to the ICSH ICCS Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine How should LDTs be validated? Accuracy Specificity Sensitivity

More information

Customer Information Literature List Platelets

Customer Information Literature List Platelets Customer Information Literature List Platelets Date: August 2015 Subject: Literature List Platelets Issued by: Scientific Customer Services Number: 150803 Note: Whether references are given in British

More information

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes A. M. CENCI *1, M. MACONI *2, and B. CASOLARI *1 *1 Laboratory of

More information

RayBio Annexin V-Cy5 Apoptosis Detection Kit

RayBio Annexin V-Cy5 Apoptosis Detection Kit RayBio Annexin V-Cy5 Apoptosis Detection Kit User Manual Version 1.0 Mar 20, 2014 (Cat#: 68C5-AnnV-S) RayBiotech, Inc. We Provide You With Excellent Support And Service Tel:(Toll Free)1-888-494-8555 or

More information

Myelodysplasia/Myeloproliferative Neoplasms (MDS/MPN) Post-HCT Data

Myelodysplasia/Myeloproliferative Neoplasms (MDS/MPN) Post-HCT Data Instructions for Myelodysplasia/Myeloproliferative Neoplasms (MDS/MPN) Post-HCT Data (Form 2114) This section of the CIBMTR Forms Instruction Manual is intended to be a resource for completing the Myelodysplasia/Myeloproliferative

More information

Hematology 101. Cindy Rogers, MT(ASCP) Diagnostics System Specialist

Hematology 101. Cindy Rogers, MT(ASCP) Diagnostics System Specialist Hematology 101 Cindy Rogers, MT(ASCP) Diagnostics System Specialist More Acronyms...» CBC» RBC» HGB» HCT» WBC» MPV» PLT» RDW» DIFF» H&H» Complete Blood Count» Red Blood Cell» Hemoglobin» Hematocrit» White

More information

Complete Blood Count (CBC)

Complete Blood Count (CBC) Complete Blood Count (CBC) Objectives 1. To estimate the number of RBC in blood sample 2. To estimate the number of total WBC in blood sample 3. To perform a differential count for a blood sample Introduction

More information

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT OriginalArticle Comparison of Leukocyte Differential Count by Beckman Coulter Unicel DxH800, Beckman Coulter LH780 and Sysmex XE-5000 Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. Department

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

HAEMATOLOGICAL EVALUATION OF ANEMIA. Sitalakshmi S Professor and Head Department of Clinical Pathology St John s medical College, Bangalore

HAEMATOLOGICAL EVALUATION OF ANEMIA. Sitalakshmi S Professor and Head Department of Clinical Pathology St John s medical College, Bangalore HAEMATOLOGICAL EVALUATION OF ANEMIA Sitalakshmi S Professor and Head Department of Clinical Pathology St John s medical College, Bangalore Learning Objectives Laboratory tests for the evaluation of anemia

More information

Literature List Body Fluids

Literature List Body Fluids Literature List Body Fluids Customer Information March 2018 Date: March 2018 Subject: Literature List body fluids Issued by: Scientific Customer Services Number: 180312 Note: Whether references are given

More information

8.2 Principles of Quantitative Hematologic Determinations (1)

8.2 Principles of Quantitative Hematologic Determinations (1) 3 8. FIELD CENTER HEMATOLOGY SERVICES 8.1 Clinical Significance Quantitation of the formed elements of the blood (erythrocytes -RBCs, leukocytes - WBCs, and platelets) is important in the ARIC study primarily

More information

A Novel Regression Based Model for Detecting Anemia Using Color Microscopic Blood Images

A Novel Regression Based Model for Detecting Anemia Using Color Microscopic Blood Images J. Software Engineering & Applications, 2010, 3, 756-760 doi:10.4236/jsea.2010.38087 Published Online August 2010 (http://www.scirp.org/journal/jsea) A Novel Regression Based Model for Detecting Anemia

More information

Automated and Standardized Counting of Mouse Bone Marrow CFU Assays

Automated and Standardized Counting of Mouse Bone Marrow CFU Assays Automated and Standardized Counting of Mouse Bone Marrow CFU Assays 2 Automated and Standardized Colony Counting Table of Contents 4 Colony-Forming Unit (CFU) Assays for Mouse Bone Marrow 5 Automated Assay

More information

Application Note. Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets

Application Note. Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets Sept 27 Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets Phillip Ruiz, MD, PhD*, Michael J. Borowitz, MD, PhD**, Henok Tilahun***,

More information

Customer Information Literature List Red Blood Cells

Customer Information Literature List Red Blood Cells Customer Information Literature List Red Blood Cells Date: October 2015 Subject: Literature List Red Blood Cells Issued by: Scientific Customer Services Number: 151019 Note: Whether references are given

More information

ACCESS hstni SCIENTIFIC LITERATURE

ACCESS hstni SCIENTIFIC LITERATURE ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval

More information

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation ISLH Laboratory Hematology 7:255 266 2001 Carden Jennings Publishing Co., Ltd. Official Publication White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation BERNARD

More information

Customer Information Literature List White Blood Cells

Customer Information Literature List White Blood Cells Customer Information Literature List White Blood Cells Date: September 2015 Subject: Literature List White Blood Cells Issued by: Scientific Customer Services Number: 150901 Note: Whether references are

More information

Identification and Enumeration of Nucleated Red Blood Cells in Peripheral Blood

Identification and Enumeration of Nucleated Red Blood Cells in Peripheral Blood Identification and Enumeration of Nucleated Red Blood Cells in Peripheral Blood Bernard J FERNANDES Department of Laboratory Medicine, Mount Sinai Hospital and University of Toronto, Toronto, 6 University

More information

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population Clinical immunology Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population EL BIETA GÓRSKA, URSZULA DEMKOW, ROMAN PIÑKOWSKI, BARBARA JAKUBCZAK,

More information

Technical Bulletin No. 162

Technical Bulletin No. 162 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 162 cobas 6800 HCV Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

Nationwide Children s Hospital

Nationwide Children s Hospital Implementing IPF and RET He: Clinical Applications in a Pediatric Setting Gretchen Dennis MT(ASCP) cm, MLS(ASCP) Technical Specialist Routine Hematology Nationwide Children s Hospital 1 We have been growing!

More information

Interpreting the CBC. Robert Miller PA Assistant Professor of Clinical Pediatrics and Family Medicine USC Keck School of Medicine Retired

Interpreting the CBC. Robert Miller PA Assistant Professor of Clinical Pediatrics and Family Medicine USC Keck School of Medicine Retired Interpreting the CBC Robert Miller PA Assistant Professor of Clinical Pediatrics and Family Medicine USC Keck School of Medicine Retired The CBC 3 Cell Lines RBCs WBCs Platelets Assess general health Make

More information

BD CBA on the BD Accuri C6: Bringing Multiplexed Cytokine Detection to the Benchtop

BD CBA on the BD Accuri C6: Bringing Multiplexed Cytokine Detection to the Benchtop BD CBA on the BD Accuri C6: Bringing Multiplexed Cytokine Detection to the Benchtop Maria Dinkelmann, PhD Senior Marketing Applications Specialist BD Biosciences, Ann Arbor, MI 23-14380-00 Cellular Communication

More information

The LaboratoryMatters

The LaboratoryMatters Laboratory Medicine Newsletter for clinicians, pathologists & clinical laboratory technologists. A Initiative. Complete Blood Count This issue highlights: CBC, while ubiquitous, is an excellent diagnostic

More information

Sponsored and reviewed by ICCS Quality and Standards Committee Title: Verification of PNH assay sensitivity through spiking experiment Written by:

Sponsored and reviewed by ICCS Quality and Standards Committee Title: Verification of PNH assay sensitivity through spiking experiment Written by: Sponsored and reviewed by ICCS Quality and Standards Committee Title: Verification of PNH assay sensitivity through spiking experiment Written by: A Illingworth, T Oldaker, R Sutherland, S Kotanchiyev,

More information

Continuing Education Questions

Continuing Education Questions FOCUS: INTERPRETING THE COMPLETE BLOOD COUNT Continuing Education Questions SUMMER 2017 1. A methodical approach to CBC interpretation that is aimed at medical laboratory professionals differs from one

More information

Annexin V-FITC Apoptosis Detection Kit with SYTOX

Annexin V-FITC Apoptosis Detection Kit with SYTOX ab14086 Annexin V-FITC Apoptosis Detection Kit with SYTOX Instructions for Use For the rapid, sensitive and accurate measurement of apoptosis in living cells. This product is for research use only and

More information

Approccio morfologico alle microangiopatie trombotiche

Approccio morfologico alle microangiopatie trombotiche Approccio morfologico alle microangiopatie trombotiche Gina Zini Polo Oncologia e Ematologia Policlinico A. Gemelli Università Cattolica S. Cuore - Roma 1 Thrombotic microangiopathies Occlusive microangiopathic

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Hematology Revision. By Dr.AboRashad . Mob

Hematology Revision. By Dr.AboRashad  . Mob 1 1- Hb A2 is consisting of: a) 3 ά chains and 2 γ chains b) 2 ά chains and 2 β chains c) 2 ά chains and 2 δ chains** d) 2 ά chains and 3 δ chains e) 3 ά chains and 2 δ chains 2- The main (most) Hb found

More information

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay TECHNICAL NOTE GSP Neonatal Phenylalanine kit 3308-0010/3308-001B GSP Neonatal Phenylalanine kit (3308-0010/3308-001B) is a fully Introduction automated enzymatic assay for the GSP system. Optimized for

More information

Associations Between Serum Transferrin Receptor Concentrations and Erythropoietic Activities According to Body Iron Status

Associations Between Serum Transferrin Receptor Concentrations and Erythropoietic Activities According to Body Iron Status Annals of Clinical & Laboratory Science, vol. 33, no. 3, 2003 279 Associations Between Serum Transferrin Receptor Concentrations and Erythropoietic Activities According to Body Iron Status Jong Weon Choi

More information

[2] [1] [2] (c) In the lungs, goblet cells secrete mucus. The mucus is then moved by cilia.

[2] [1] [2] (c) In the lungs, goblet cells secrete mucus. The mucus is then moved by cilia. 1 (a) (i) Explain what is meant by the term tissue.... [2] Name one type of epithelial tissue found in the lungs.... [1] (b) Explain why the lungs can be considered to be an organ............ [2] (c) In

More information

5/1/2017 DISCUSSION POINTS. Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN

5/1/2017 DISCUSSION POINTS. Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN DISCUSSION POINTS Importance of hematological immature cell indices Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN Investigate the evidence for clinical utility:

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

DetectX Hemoglobin High Sensitivity Colorimetric Detection Kit

DetectX Hemoglobin High Sensitivity Colorimetric Detection Kit DetectX Hemoglobin High Sensitivity Colorimetric Detection Kit 2 Plate Kit Catalog Number K013-HX1 10 Plate Kit Catalog Number K013-HX5 Species Independent Sample Types Validated: Serum and Plasma Please

More information

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, May 1997, p. 309 313 Vol. 4, No. 3 1071-412X/97/$04.00 0 Copyright 1997, American Society for Microbiology Evaluation of a Method for Counting Absolute Numbers

More information

THE CLASSIFICATION OF ANEMIA*

THE CLASSIFICATION OF ANEMIA* THE CLASSIFICATION OF ANEMIA* RUSSELL L. HADEN, M.D. SUMMARY A laboratory and clinical classification of anemia has been outlined. The results of the blood study have been correlated with the clinical

More information

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2 Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...

More information

EDUCATIONAL COMMENTARY BLOOD CELL IDENTIFICATION

EDUCATIONAL COMMENTARY BLOOD CELL IDENTIFICATION EDUCATIONAL COMMENTARY BLOOD CELL IDENTIFICATION Educational commentary is provided through our affiliation with the American Society for Clinical Pathology (ASCP). To obtain FREE CME/CMLE credits click

More information

M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and Clinical Immunology, MU Sofia

M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and Clinical Immunology, MU Sofia 50, 48, 2012, 3, ADVIA 2120.,.., ANALYSIS OF DIAGNOSTIC RED CELL CYTOGRAMS, GENERATED BY THE ADVIA 2120 HEMATOLOGY ANALYZER M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and

More information

Update on CD4+ T-Cell Counting Technologies

Update on CD4+ T-Cell Counting Technologies Update on CD4+ T-Cell Counting Technologies Thomas Denny, MSc, M.Phil Professor of Medicine Duke Human Vaccine Institute Duke University Medical School UK NEQAS 24 th June 2013, Sheffield 2 3 4 5 6 Levels

More information

Evaluation of the QBC Star centrifugal three-part differential haematology system

Evaluation of the QBC Star centrifugal three-part differential haematology system ORIGINAL ARTICLE 67 Evaluation of the QBC Star centrifugal three-part differential haematology system O. ERHABOR, G. RICHARDSON, I. MOHAMMED, C. THORNTON, J. BARK, M. HURST, D. HAMER and P. KINSELLA Blood

More information

Technical Bulletin No. 161

Technical Bulletin No. 161 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 161 cobas 6800 HIV-1 Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis

Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Hematopathology / LEFT SHIFT MODEL PREDICTS INFECTION Manual Differential Cell Counts Help Predict Bacterial Infection A Multivariate Analysis Michael J. Wile, MD, 1 Louis D. Homer, MD, PhD, 2 Stede Gaehler,

More information

Evidence-Based Hematological Solutions

Evidence-Based Hematological Solutions Evidence-Based Hematological Solutions Beyond the Routine CBC Objectives Describe novel hematology parameters and their derivation. Investigate the evidence for their clinical utility. Discuss how new

More information

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W Simple, Fast and Accurate Quantitation of Human Plasma Vitamins and Their Metabolites by Protein Precipitation Combined with Columns Using HPLC-UV, HPLC-FLD or LC/MS/MS Shuguang Li, Jason Anspach, Sky

More information

MHC MULTIMER PROFICIENCY PANEL 2017

MHC MULTIMER PROFICIENCY PANEL 2017 MHC MULTIMER PROFICIENCY PANEL 2017 August 2017 CONTACT Charlotte Halgreen ProficiencyPanel@immudex.com FOR MORE INFORMATION www.proficiencypanel.com MHC MULTIMER PROFICIENCY PANEL 2017 This report summarizes

More information

Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations

Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations Human Progenitor Cell Enumeration by Flow Cytometry Practical Considerations Bruce Briggs, M.S., S.I. (A.S.C.P.) OSU Medical Center Clinical Flow Cytometry Objectives Conceptual overview: How the Flow

More information

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 R. HERKLOTZ and A. R. HUBER Zentrum für Labormedizin, Kantonsspital Aarau, 5001 Aarau, Switzerland. The automated leukocyte(wbc)

More information

Changes of hematological references depends on storage period and temperature conditions in rats and dogs

Changes of hematological references depends on storage period and temperature conditions in rats and dogs ISSN 1738-6055 (Print) ISSN 2233-7660 (Online) Lab Anim Res 2016: 32(4), 241-248 https://doi.org/10.5625/lar.2016.32.4.241 Changes of hematological references depends on storage period and temperature

More information

RayBio Annexin V-FITC Apoptosis Detection Kit Plus

RayBio Annexin V-FITC Apoptosis Detection Kit Plus RayBio Annexin V-FITC Apoptosis Detection Kit Plus User Manual Version 1.0 May 25, 2014 RayBio Annexin V-FITC Apoptosis (Cat#: 68FT-AnnVP-) RayBiotech, Inc. We Provide You With Excellent upport And ervice

More information

DEVELOPMENTAL VALIDATION OF SPERM HY-LITER EXPRESS Jennifer Old, Chris Martersteck, Anna Kalinina, Independent Forensics, Lombard IL

DEVELOPMENTAL VALIDATION OF SPERM HY-LITER EXPRESS Jennifer Old, Chris Martersteck, Anna Kalinina, Independent Forensics, Lombard IL DEVELOPMENTAL VALIDATION OF SPERM HY-LITER EXPRESS Jennifer Old, Chris Martersteck, Anna Kalinina, Independent Forensics, Lombard IL Background and Introduction SPERM HY-LITER is the first immunofluorescent

More information

LEANING Away Waste and Improving Patient Care with Advances in Hematology

LEANING Away Waste and Improving Patient Care with Advances in Hematology LEANING Away Waste and Improving Patient Care with Advances in Hematology Selke Mantie, MLS (ASCP), CLS (CSMLS), SSGBC Disclosure I am receiving an honorarium from Sysmex for today s presentation Objectives

More information

Phagocytosis of FITC labelled opsonized and non-opsonized E. coli bacteria by monocytes and granulocytes in a whole blood assay

Phagocytosis of FITC labelled opsonized and non-opsonized E. coli bacteria by monocytes and granulocytes in a whole blood assay Phagocytosis of FITC labelled opsonized and non-opsonized E. coli bacteria by monocytes and granulocytes in a whole blood assay APPLICATION NOTE Author: Andreas Spittler, MD, Associate Professor for Pathophysiology

More information