pqct Measurement of Bone Parameters in Young Children

Size: px
Start display at page:

Download "pqct Measurement of Bone Parameters in Young Children"

Transcription

1 Journal of Clinical Densitometry, vol. 3, no. 1, 9 14, Spring 2000 Copyright 2000 by Humana Press Inc. All rights of any nature whatsoever reserved /00/3:9 14/$11.50 Original Article pqct Measurement of Bone Parameters in Young Children Validation of Technique Teresa L. Binkley, MS 1 and Bonny L. Specker, PHD 1,2 1 Ethel Austin Martin Program in Human Nutrition, South Dakota State University, Brookings, SD; and 2 Department of Pediatrics, University of South Dakota School of Medicine, Brookings, SD Abstract Dual-energy X-ray absorptiometry (DXA) measures areal bone mineral density (BMD) and is affected by bone size. Peripheral quantitative computed tomography (pqct) measures volumetric density and should not be affected by bone size. We hypothesized that pqct could be used to measure geometric parameters in the tibia and that bone size, not density, would correlate with weight and height in 3 and 4 yr olds. Phantom measurements indicate that accurate results for cortical volumetric bone mineral density can be obtained at cortical thickness > 2 mm with voxel/speed sizes of 0.40/20 mm/s. Correlations between pqct measured geometric bone parameters and phantom calculations were observed (all r > 0.96). Baseline data from an ongoing trial of 101 preschool children (53 male) were used to correlate bone parameters and anthropometrics. Total cross-sectional area, cortical area, and cortical thickness correlated with weight (r = 0.54, p < 0.001; r = 0.52, p < 0.001; r = 0.30, p = 0.002) and height (r = 0.41, p < 0.001; r = 0.55, p < 0.001; r = 0.41, p < 0.001). Because of the small cortical thickness at this age (mean = 1.2 mm), the cortical density was not analyzed. In a regression model including height and weight, weight was the only predictor of total cross-sectional area (p < 0.001); cortical thickness was predicted by height (p = 0.006). Both height (p = 0.005) and weight (p = 0.05) predicted the cortical area. In summary, pqct accurately measures the total cross-sectional area, cortical area, and cortical thickness in children age 3 4 yr. Key Words: Peripheral quantitative computed tomography (pqct); cortical density; total cross-sectional area; cortical area; cortical thickness. Introduction Optimal formation and mineralization of the skeletal system during childhood is thought to be important in preventing osteoporosis during adult life (1). Dual-energy X-ray absorptiometry (DXA) is the most widely used technique for measuring bone density in clinical trials; however, methods to Received 07/07/99; Revised 08/20/99; Accepted 09/25/99. Address correspondence to Teresa L. Binkley, Box 2204, South Dakota State University, Brookings, SD Teresa_Binkley@sdstate.edu 9 calculate bone mineral density (BMD) using DXA are dependent on bone area (cm 2 ), not volume (cm 3 ). Peripheral quantitative computed tomography (pqct) measures BMD using a volume (cm 3 ) and has the ability to differentiate cortical bone from trabecular bone. Parameters such as cortical density, total cross-sectional bone area, cortical bone area, and cortical thickness can be assessed using pqct. Although these measurements are expected to be beneficial in the study of bone growth in young children, the question that arises is the extent to which this technique can measure smaller bones.

2 10 Binkley and Specker The purpose of this study was to use a phantom of known size and density to validate the performance of a pqct densitometer, as well as to test the precision of the instrument in children. Also, baseline data taken from a longitudinal study now underway at South Dakota State University (Brookings, SD) were used to correlate various pqct bone parameter measurements with anthropometric measurements in 3- and 4-yr-old children. Materials and Methods The pqct densitometer used in all measurements was an XCT 2000 (Norland Medical Systems, Inc., Fort Atkinson, WI). This unit has six detectors, with one CT motion or slice equal to 90 projection angles, considered to be 1 block when scanning (range = 1 6 blocks). The size of the field of view in the slice plane can be set by the operator. The sampling resolution, also selected by the user, is referred to as a voxel size (range = mm 2 ). The CT scan speed, in millimeters per second (mm/s), can also be predetermined by the user (range = mm/s). The number of blocks, voxel size, and scan speed are all parameters fixed at the time the scan is taken. Once the scan is completed, computer software offers numerous settings for analysis. A scan can be analyzed any number of times with these various settings. Analysis of all scans was done using research software version 5.40 Rev. B, which allows modes (options) for the following: detecting the outer bone edge (Contour Mode), defining the way subcortical and trabecular bone are separated (Peel Mode), and separating cortical bone from trabecular bone (CortMode). The CALCBD function of the software is used to obtain results for total cross-sectional area in millimeters squared and is a function of Contour Mode and Peel Mode selection. All of the mode settings used in our study were selected based on recommendations by the manufacturer and consensus of users working with this particular densitometer (Norland pqct Users Meeting, Sun Valley, ID; August 1998) To detect the outer bone edge, we chose Contour Mode 2. This mode utilizes an algorithm that sets a density threshold automatically and searches to find voxels that meet specified conditions to be marked as the outer bone edge. For separating subcortical from trabecular bone, Peel Mode 2, which implements an operatorselected threshold density of 400 mg/cm 3,was used. Any voxel with a density higher than this threshold is considered subcortical or cortical, whereas voxels with lower values are designated as trabecular. Another function, CORTBD, gives results for cortical density, cortical area, and cortical thickness. CortMode 1 was the setting for separating cortical from trabecular bone. In this mode, a density threshold is entered by the operator and voxels with a lower density are removed, leaving the cortical shell. The default threshold of 710 mg/cm 3 was used. Cortical thickness can be assessed using two different software procedures, either an iterative contour detection, which is a direct measurement of cortical thickness based on the true shape of the bone, or a threshold algorithm, often referred to as the circular ring model (2). In our study, the circular ring model for thickness with a threshold of 710 mg/cm 3 was used. Validation of the XCT 2000 was performed using a phantom supplied by Dr. Thomas Hangartner, Director, BioMedical Imaging Laboratory (Dayton, OH). This apparatus consisted of a Lucite cylinder that held an insert with a graduated thickness ranging from 0.3 mm to 4 mm. The inner diameter of the insert was 20 mm throughout, with the wall thickness equal to 4, 3, 2, 1, and 0.3 mm (see Fig. 1). To vary the density, two inserts were used, one of aluminum, the other of a poly(vinyl chloride) (PVC) material. Both the Lucite tube and insert were filled with distilled water at room temperature. For each Fig. 1. Phantom sketch showing a constant inner opening diameter and graduated wall thickness.

3 pqct Measurement of Bone Parameters in Children 11 thickness, on each insert, three scans were done using both one and two blocks with the following voxel/speed combinations: voxel = 0.4/speed = 20; voxel=0.3/speed = 15; voxel = 0.2/speed = 10. This amounted to a total of 18 scans at each thickness. A reproducibility study of the XCT 2000 involved duplicate scans on 11 children ages 3 5 yr. The radiation dose for each scan was 8 mrem (central dose supplied by the manufacturer) and written parental consent was obtained to do two scans at the 20% distal tibia site of the left leg. This site is located by calculating 20% of the total distance between the tibial tuberosity and the medial malleolus and marking that calculated distance from the distal end. The patient s leg was repositioned between the scans without the use of a scout view. Settings for the scan were voxel=0.4, speed = 20, and blocks = 1. Calculations for total cross-sectional area, cortical area, and cortical thickness were obtained using the software settings stated previously. Coefficients of variation (CV) were calculated on duplicate scans (3) as where CV = Sd of difference between two scans Overall mean for that measurement SD = sq rt ( sum (difference between two ) scans)2 2N either phantom at any voxel/speed combination. In both the PVC and aluminum phantom, cortical density measurements decreased significantly between the 1.0- and 0.3-mm wall thickness (p < 0.001). In the aluminum phantom, there was no significant difference between mean density readings for any speed when the wall thickness was greater than 1 mm (mean density = mg/cm 3 ). For the PVC insert, the smaller wall thickness had more of an effect on the density reading than the larger wall thickness, tested by the interaction of wall thickness and speed (p < 0.001) (see Fig. 2). For other significant differences in the mean density readings of the PVC insert, see Table 1. The total cross-sectional area and ring area were calculated for each thickness of the phantom and then compared to total cross-sectional area and cortical area readings. Calculated total cross-sectional area versus measured total cross-sectional area showed a strong correlation for both PVC (r = 0.962) and aluminum (r = 0.978), as did calculated ring area versus measured cortical area (r = PVC; r = Al). Phantom thickness versus measured cortical thickness also showed a strong correlation (r = PVC; r = Al) (see Fig. 3). Neither speed nor number of blocks had an effect on the ability of the machine to measure total cross-sectional area, cortical area, or cortical thickness. Baseline data from an ongoing exercise and calcium trial of 101 preschool children (53 male) were collected by pqct scan at the 20% distal tibia site of the left leg. Written parental consent was obtained before scanning. Settings for the scan were voxel = 0.4, speed = 20, and blocks = 1. Calculations for cortical density, total cross-sectional area, cortical area, and cortical thickness were obtained with the same software settings as stated previously. Height (cm) and weight (kg) were measured at the time of the scan. For tests involving children, protocols were approved by the Human Subjects Committee at South Dakota State University. Results Validation Results The number of blocks used (one vs two) to obtain the scan had no effect on cortical density readings for Fig. 2. PVC phantom density reading versus phantom wall thickness for scan speeds of ( ) 10 mm/s; ( ) 15 mm/s; ( ) 20 mm/s.

4 12 Binkley and Specker Table 1 Mean Density Readings (mg/cm 3 ) for Each PVC Phantom Wall Thickness Mean (SD) density reading per wall thickness Voxel/speed 0.3 mm 1 mm 2 mm 3 mm 4 mm Significant differences a 0.20/10 mm/s 853 (12) 952 (4) 963 (3) 963 (6) 956 (7) 0.3 mm 1, 2, 3, 4 mm 0.30/15 mm/s 778 (6.7) 926 (2) 952 (2) 951 (1) 945 (2) 0.3 mm 1, 2, 3, 4 mm 1 mm 0.3, 2, 3, 4 mm 4 mm 0.3, 1, 2 mm 0.40/10 mm/s 740 (7.0) 903 (3) 935 (1) 941 (1) 940 (2) 0.3 mm 1, 2, 3, 4 mm 1 mm 0.3, 2, 3, 4 mm a Significant difference in means (p < 0.05). Reproducibility Results Coefficient of variation for pqct bone parameters from duplicate scans in children aged 3 5 yr ranged from % (see Table 2). The child with the lowest mean cortical thickness (0.924 mm) had the largest discrepancy in cortical thickness measurements. When this child was excluded, the coefficients of variation were 3.1% for total area, 4.5% for cortical area, and 6.8% for cortical thickness. There was no obvious reason for the discrepant results, as both scans appeared acceptable. Preliminary Data Results Total cross-sectional area, cortical area, and cortical thickness were correlated with weight (r = 0.54, p < 0.001; r = 0.52, p < 0.001; r = 0.30, p = 0.002) and height (r = 0.41, p < 0.001; r = 0.55, p < 0.001; r = 0.40, p < 0.001). Because of the small cortical thickness at this age (mean = 1.2 mm), the relationship between cortical density and height/weight was not analyzed. There were no differences between females and males in mean ± SEM in total cross-sectional area (179.1 ± 4.5 vs ± 4.4; p = 0.57), cortical area (50.7 ± 1.8 vs 53.2 ± 1.7; p = 0.29), or cortical thickness (1.19 ± 0.05 vs 1.26 ± 0.05; p = 0.29). In a regression model with height and weight as predictors, height was the only significant predictor of cortical thickness (p = 0.006), whereas total crosssectional area was predicted by weight (p < 0.001). Both height (p = 0.005) and weight (p = 0.05) predicted the cortical area (see Fig. 4). Discussion Although dual-energy X-ray absorptiometry (DXA) is considered the method of choice for evaluating bone mineral status, pqct offers two possible advantages for measurement of bone in growing children. One advantage is that the true volumetric bone mineral density is measured. The other improvement is the ability to measure total cross-sectional and cortical area of bone and cortical thickness. In young children, volumetric density may be a better expression for BMD than that given by the areal density measurement from DXA, which is influenced by bone size (4). The results of our study show that pqct, using the XCT 2000 clinical model, can be used to measure cortical density accurately when the cortical thickness of the bone is 2 mm or greater with voxel/speed sizes of 0.40/20 mm/s. A study by Hangartner and Gilsanz found the minimal thickness to be mm for accurate density readings using the same phantom on two different QCT machines (5). Unfortunately, the mean cortical thickness of the 20% distal tibia in 3- and 4- yr-old children (mean = 1.2 mm) is too small to get an accurate pqct cortical density reading with the above voxel/speed settings. The other benefit of pqct, the ability to differentiate cortical bone, makes geometric assessments possible. Although DXA is beneficial in obtaining total bone mineral context (BMC) changes during growth (6), pqct may be helpful in determining particular changes in bone parameters such as total

5 pqct Measurement of Bone Parameters in Children 13 Table 2 Coefficient of Variation for Repeated Scans in Children Aged 3 5 yr % CV (N = 11) Mean (N = 11) Total area (cm 2 ) Cortical area (cm 2 ) Cortical thickness (mm) Fig. 3. Phantom pqct results for PVC insert ( ) and aluminum insert ( ) compared to calculated or known values ( ) for each wall thickness. Top panel: total cross-sectional area results versus calculated crosssectional area for each wall thickness (r = PVC; r = Al); middle panel: cortical area results versus calculated ring area for each wall thickness (r = PVC; r = Al); bottom panel: thickness results versus known wall thickness (r = PVC; r = Al). cross-sectional area, cortical area, and cortical thickness. As growth occurs, changes in these bone parameters may be observed. When testing various scan speeds, we found that scan speed had a small effect on density at the smaller phantom wall thickness. Although there were no significant differences in measured density in either the PVC or aluminum insert at thickness sites 1 mm or greater using the 10 mm/s speed, there is a feasibility issue when measuring young children because of the time it takes to complete a scan at this speed. At a speed setting of 20 mm/s, it takes approximately 104 s to complete a scan at the 20% distal tibia site, whereas at the same site, using a speed of 10 mm/s, it would take about 181 s. In a study such as ours, where the subjects are 3 and 4 yr olds, time becomes a critical factor in obtaining an acceptable scan, and we have chosen the 20-mm/s speed. Preliminary results from the preschool study show that the mean cortical thickness of the 20% distal tibia site is too small to accurately analyze cortical density. Other bone parameters measured at this age can be measured using pqct. We found no difference in bone size because of gender. Inconsistent findings have been reported as to whether bone size differs with gender at this age, which may be caused in part by the use of different absorptiometers, differences in sites measured, or sample sizes insufficient to detect differences (7 10). Findings from this study indicate that pqct can be used to accurately measure total cross-sectional area, cortical area, and cortical thickness in the bones of young children 3 to 4 yr of age at a scan speed of 20 mm/s and voxel size of 0.40 mm. Total cross-sectional area, cortical area, and cortical thickness increase with increasing height and weight, but no gender differences in bone size were noted. Acknowledgments The authors gratefully acknowledge G. Niemeyer and D. Schiferl, Norland/Stratec Medical Systems,

6 14 Binkley and Specker Fort Atkinson, WI, for their technical expertise. Also, many thanks to T. Hangartner, Biomedical Imaging Laboratory, Dayton, OH, for allowing the use of his phantom. Fig. 4. pqct geometric bone parameter results from 3- to 4-yr-old children scanned at the 20% distal tibia site (N = 101). Top panel: total cross-sectional area versus weight (r = 0.54, p < 0.001); middle panel: cortical area versus height (r = 0.55, p < 0.001); bottom panel: cortical thickness versus height (r = 0.41, p < 0.001). References 1. Schonau E The development of the skeletal system in children and the influence of muscular strength. Horm Res 49: Louis O, Willnecker J, Soykens S, Van den Winkel P, Osteaux M Cortical thickness assessed by peripheral quantitative computed tomography: accuracy evaluated on radius specimens. Osteoporos Int 5: Mueller WH, Martorell R Reliability and Accuracy of Measurement. In: Anthropometric Standardization Reference Manual, Lohman T, Roche A, Martorell R, eds. Human Kinetics Books, Champaign, IL, Prentice A, Parsons T, Cole T Uncritical use of bone mineral density in absorptiometry may lead to size related artifacts in the identification of bone mineral determinants. Am J Clin Nutr 60: Hangartner T, Gilsanz V Evaluation of cortical bone by computed tomography. J Bone Miner Res 11: Brunton A, Weiler H, Atkinson S Improvement in the accuracy of dual energy x-ray absorptiometry for whole body an regional analysis of body composition: validation of using piglets and methodologic considerations in infants. Pediatr Res 41: Gilsanz V, Boechat MI, Roe TF, Loro ML, Sayre JW, Goodman WG Gender differences in vertebral body sizes in children and adolescents. Radiology 190: Henderson RC Assessment of bone mineral content in children. J Pediatr Orthop 11: Ruprich RC, Specker BL, Lieuw-A-Fa M, Ho M Gender and race differences in bone mass during infancy. Calcif Tissue Int 58: Southard RN, Morris JD, Mahan JD, et al Bone mass in healthy children: measurement with quantitative DXA. Radiology 179:

CLINICAL DENSITOMETRY

CLINICAL DENSITOMETRY JOURNAL OF CLINICAL DENSITOMETRY THE OFFICIAL JOURNAL OF THE INTERNATIONAL SOCIETY FOR CLINICAL DENSITOMETRY EDITOR-IN-CHIEF CLIFFORD J. ROSEN, MD HUMANA PRESS VOLUME 5 NUMBER 4 WINTER 2002 ISSN: 1094

More information

Muscle-bone relationships in the lower leg of healthy pre-pubertal females and males

Muscle-bone relationships in the lower leg of healthy pre-pubertal females and males J Musculoskelet Neuronal Interact 2008; 8(3):239-243 Original Article Hylonome Muscle-bone relationships in the lower leg of healthy pre-pubertal females and males T.L. Binkley and B.L. Specker Ethel Austin

More information

L.W. Sun 1,2, G. Beller 1, D. Felsenberg 1. Introduction. Original Article. Abstract

L.W. Sun 1,2, G. Beller 1, D. Felsenberg 1. Introduction. Original Article. Abstract J Musculoskelet Neuronal Interact 2009; 9(1):18-24 Original Article Hylonome Quantification of bone mineral density precision according to repositioning errors in peripheral quantitative computed tomography

More information

Quantitative Computed Tomography 4 Introduction

Quantitative Computed Tomography 4 Introduction Quantitative Computed Tomography 4 Introduction Quantitative Computed Tomography (QCT) is a well recognised technique for the measurement of bone mineral density (BMD) in the lumbar spine1 and forearm2.

More information

Bone density and size in ambulatory children with cerebral palsy

Bone density and size in ambulatory children with cerebral palsy DEVELOPMENTAL MEDICINE & CHILD NEUROLOGY ORIGINAL ARTICLE Bone density and size in ambulatory children with cerebral palsy TISHYA AL WREN 1 DAVID C LEE 1 ROBERT M KAY 1 FREDERICK J DOREY 2 VICENTE GILSANZ

More information

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7 STRUCTURED SELF ASSESSMENT CONTENT SPECIFICATIONS SSA LAUNCH DATE: JANUARY 1, 2018 Bone Densitometry The purpose of continuing qualifications requirements (CQR) is to assist registered technologists in

More information

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017 EXAMINATION CONTENT SPECIFICATIONS Bone Densitometry The purpose of the bone densitometry examination is to assess the knowledge and cognitive skills underlying the intelligent performance of the tasks

More information

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017 STRUCTURED EDUCATION REQUIREMENTS Bone Densitometry The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge that, when coupled

More information

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p.

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p. Dedication Preface Acknowledgments Continuing Education An Introduction to Conventions in Densitometry p. 1 Densitometry as a Quantitative Measurement Technique p. 2 Accuracy and Precision p. 2 The Skeleton

More information

Investigating the Effects of Motion Streaks on pqct-derived Leg Muscle Density and Its Association With Fractures

Investigating the Effects of Motion Streaks on pqct-derived Leg Muscle Density and Its Association With Fractures Journal of Clinical Densitometry: Assessment & Management of Musculoskeletal Health, vol. 21, no. 1, 130 139, 2018 Copyright 2016 by The International Society for Clinical Densitometry 1094-6950/21:130

More information

Peripheral quantitative computed tomography of the distal radius in young subjects new reference data and interpretation of results

Peripheral quantitative computed tomography of the distal radius in young subjects new reference data and interpretation of results J Musculoskelet Neuronal Interact 2005; 5(2):119-126 Original Article Hylonome Peripheral quantitative computed tomography of the distal radius in young subjects new reference data and interpretation of

More information

The Bone Densitometry Examination

The Bone Densitometry Examination The Bone Densitometry Examination The purpose of The American Registry of Radiologic Technologist (ARRT ) Bone Densitometry Examination is to assess the knowledge and cognitive skills underlying the intelligent

More information

CT Imaging of skeleton in small animals. Massimo Marenzana

CT Imaging of skeleton in small animals. Massimo Marenzana CT Imaging of skeleton in small animals Massimo Marenzana Introduction Osteoporosis is a disease in which bones become fragile and more likely to break. It can be defined as a systemic skeletal disease

More information

2013 ISCD Combined Official Positions

2013 ISCD Combined Official Positions 2013 ISCD Combined Oicial Positions Oicial Positions of the International Society for Clinical Densitometry The International Society for Clinical Densitometry (ISCD) is a not-for-profit multidisciplinary

More information

Norland Densitometry A Tradition of Excellence

Norland Densitometry A Tradition of Excellence Norland Densitometry A Tradition of Excellence Norland DXA Bone Density Measurement Osteoporosis is a disease marked by reduced bone strength leading to an increased risk of fractures. About 54 million

More information

2013 ISCD Official Positions Adult

2013 ISCD Official Positions Adult 2013 ISCD Official Positions Adult These are the Official Positions of the ISCD as updated in 2013. The Official Positions that are new or revised since 2007 are in bold type. Indications for Bone Mineral

More information

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302)

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302) PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302) Date of Interview/Examination (MM/DD/YYYY): A downloadable PDF

More information

Thickness Computation Under In-Vivo Trabecular Bone CT Imaging

Thickness Computation Under In-Vivo Trabecular Bone CT Imaging Thickness Computation Under In-Vivo Trabecular Bone CT Imaging Gokul. S PG Scholar,M.E - II year Department of Computer Science and Engineering Jansons Institute of Technology Coimbatore, Tamilnadu Aarthi.K

More information

The effect of vertebral rotation of the lumbar spine on dual energy X-ray absorptiometry measurements: observational study

The effect of vertebral rotation of the lumbar spine on dual energy X-ray absorptiometry measurements: observational study JCY Cheng HL Sher X Guo VWY Hung AYK Cheung Key words: Absorptiometry, Bone density; Densitometry, X-ray; Lumbar vertebrae; Scoliosis "#$%& "# X HKMJ 2001;7:241-5 The Chinese University of Hong Kong, Prince

More information

Bone Densities and Bone Size at the Distal Radius in Healthy Children and Adolescents: A Study Using Peripheral Quantitative Computed Tomography

Bone Densities and Bone Size at the Distal Radius in Healthy Children and Adolescents: A Study Using Peripheral Quantitative Computed Tomography Bone Vol. 28, No. 2 February 2001:227 232 Bone Densities and Bone Size at the Distal Radius in Healthy Children and Adolescents: A Study Using Peripheral Quantitative Computed Tomography C. M. NEU, 1,2

More information

DXA Best Practices. What is the problem? 9/29/2017. BMD Predicts Fracture Risk. Dual-energy X-ray Absorptiometry: DXA

DXA Best Practices. What is the problem? 9/29/2017. BMD Predicts Fracture Risk. Dual-energy X-ray Absorptiometry: DXA BMD Predicts Fracture Risk Ten Year Fracture Probability (%) 50 40 30 20 10 Age 80 70 60 50 E. Michael Lewiecki, MD Director, New Mexico Clinical Research & Osteoporosis Center Director, Bone TeleHealth

More information

ADOLESCENT OBESITY: IS IT BAD FOR THE BONES

ADOLESCENT OBESITY: IS IT BAD FOR THE BONES ADOLESCENT OBESITY: IS IT BAD FOR THE BONES Babette S. Zemel, PhD Director, Nutrition And Growth Laboratory Division Of Gastroenterology, Hepatology And Nutrition The Children s Hospital Of Philadelphia

More information

Accuracy of DEXA scanning & other methods for determining BMD.

Accuracy of DEXA scanning & other methods for determining BMD. BMD- Measurement Site Accuracy of DEXA scanning & other methods for determining BMD. Ann Larkin In general, densitometry techniques can be performed in either the axial or the appendicular skeleton, depending

More information

Comparison of Muscle Image Acquisition & Analysis Techniques

Comparison of Muscle Image Acquisition & Analysis Techniques Comparison of Muscle Image Acquisition & Analysis Techniques Toronto General Research Institute Andy Kin On Wong PhD Assistant UHN Scientist Scanco User Meeting, September 20, 2016 2:15 pm Muscle & Bone

More information

THE PENNSYLVANIA STATE UNIVERSITY SCHREYER HONORS COLLEGE DEPARTMENT OF SCIENCE

THE PENNSYLVANIA STATE UNIVERSITY SCHREYER HONORS COLLEGE DEPARTMENT OF SCIENCE THE PENNSYLVANIA STATE UNIVERSITY SCHREYER HONORS COLLEGE DEPARTMENT OF SCIENCE VALIDATING THE USE OF PERIPHERAL BONE SITES TO PREDICT THE RISK OF FEMORAL NECK FRACTURE RACHEL ALBA SPRING 2017 A thesis

More information

Bone Investigational Toolkit BIT. Biomechanical Bone Integrity Assessment

Bone Investigational Toolkit BIT. Biomechanical Bone Integrity Assessment Bone Investigational Toolkit BIT TM Biomechanical Bone Integrity Assessment The Bone Densitometry Problem Modeling of cortical shell failure with third-party engineering models. Bone densitometry is widely

More information

Clinical Study Comparison of QCT and DXA: Osteoporosis Detection Rates in Postmenopausal Women

Clinical Study Comparison of QCT and DXA: Osteoporosis Detection Rates in Postmenopausal Women International Endocrinology Volume 3, Article ID 895474, 5 pages http://dx.doi.org/.55/3/895474 Clinical Study Comparison of QCT and DXA: Osteoporosis Detection Rates in Postmenopausal Women Na Li, Xin-min

More information

The bone mass concept: problems in short stature

The bone mass concept: problems in short stature European Journal of Endocrinology (2004) 151 S87 S91 ISSN 0804-4643 The bone mass concept: problems in short stature E Schoenau, C Land, A Stabrey, T Remer 1 and A Kroke 1 Children s Hospital, University

More information

Documentation, Codebook, and Frequencies

Documentation, Codebook, and Frequencies Documentation, Codebook, and Frequencies Dual-Energy X-ray Absorptiometry Femur Bone Measurements Examination Survey Years: 2005 to 2006 SAS Transport File: DXXFEM_D.XPT January 2009 NHANES 2005 2006 Data

More information

Bone Mineral Densitometry with Dual Energy X-Ray Absorptiometry

Bone Mineral Densitometry with Dual Energy X-Ray Absorptiometry Bone Mineral Densitometry with Dual Energy X-Ray Absorptiometry R Gilles, Laurentius Ziekenhuis Roermond 1. Introduction Osteoporosis is characterised by low bone mass, disruption of the micro-architecture

More information

9 Quality Assurance in Bone Densitometry section

9 Quality Assurance in Bone Densitometry section 9 Quality Assurance in Bone Densitometry section Introduction Bone densitometry is frequently used to determine an individual's fracture risk at a particular point in time but may also be used to assess

More information

SPINAL CORD INJURY (SCI) with a motor complete lesion

SPINAL CORD INJURY (SCI) with a motor complete lesion 498 Fracture Threshold in the Femur and Tibia of People With Spinal Cord Injury as Determined by Peripheral Quantitative Computed Tomography Prisca Eser, PhD, Angela Frotzler, MSc, Yvonne Zehnder, MD,

More information

Bone Densitometry Equipment Operator

Bone Densitometry Equipment Operator Bone Densitometry Equipment Operator The purpose of the Bone Densitometry Equipment Operator Examination, which is made available to state licensing agencies, is to assess the knowledge and cognitive skills

More information

Omnisense: At Least As Good As DXA

Omnisense: At Least As Good As DXA Omnisense: At Least As Good As DXA The following document summarizes a series of clinical studies that have been conducted to compare between different qualities of the Sunlight support the claim that

More information

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi Clinical Utility of Bone Densitometry Diagnosis (DXA)

More information

Author's response to reviews

Author's response to reviews Author's response to reviews Title:The association between lean mass and bone mineral content in the high disease activity group of adult patients with juvenile idiopathic arthritis Authors: Kristyna Brabnikova

More information

Standard Operating Procedure TCRC Dual-Energy X-ray Absorptiometry (DXA)

Standard Operating Procedure TCRC Dual-Energy X-ray Absorptiometry (DXA) 1. RELEVANCE a. This SOP outlines the instructions to completing Duel Energy X-Ray Absorptiometry (DXA) including: scanning, analysis, review and filing. 2. SCOPE a. This SOP applies to all TCRC RDs. 3.

More information

This is a repository copy of Microarchitecture of bone predicts fractures in older women.

This is a repository copy of Microarchitecture of bone predicts fractures in older women. This is a repository copy of Microarchitecture of bone predicts fractures in older women. White Rose Research Online URL for this paper: http://eprints.whiterose.ac.uk/130351/ Version: Accepted Version

More information

BONE MINERAL DENSITY OF BANGLADESHI PEOPLE

BONE MINERAL DENSITY OF BANGLADESHI PEOPLE 6 DAFFODIL INTERNATIONAL UNIVERSITY JOURNAL OF SCIENCE BANU: BONE AND MINERAL TECHNOLOGY, DENSITY VOLUME OF BANGLADESHI 4, ISSUE 2, JULY PEOPLE 2009 BONE MINERAL DENSITY OF BANGLADESHI PEOPLE Dilruba Akhter

More information

Interchangeability of pencil-beam and fan-beam dual-energy X-ray absorptiometry measurements in piglets and infants 1 3

Interchangeability of pencil-beam and fan-beam dual-energy X-ray absorptiometry measurements in piglets and infants 1 3 Interchangeability of pencil-beam and fan-beam dual-energy X-ray absorptiometry measurements in piglets and infants 1 3 Winston WK Koo, Mouhanad Hammami, and Elaine M Hockman ABSTRACT Background: Compared

More information

Densitometry Techniques

Densitometry Techniques 2 Densitometry Techniques CONTENTS PLAIN RADIOGRAPHY IN THE ASSESSMENT OF BONE DENSITY QUALITATIVE MORPHOMETRY QUALITATIVE SPINAL MORPHOMETRY THE SINGH INDEX QUANTITATIVE MORPHOMETRIC TECHNIQUES CALCAR

More information

Advance in Knowledge This study provides normative reference values for vertebral trabecular bone density obtained by using quantitative CT in healthy

Advance in Knowledge This study provides normative reference values for vertebral trabecular bone density obtained by using quantitative CT in healthy Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, use the Radiology Reprints form at the end of this article.

More information

DXA IS THE MOST commonly used method of assessing

DXA IS THE MOST commonly used method of assessing JOURNAL OF BONE AND MINERAL RESEARCH Volume 24, Number 8, 2009 Published online on February 16, 2009; doi: 10.1359/JBMR.090224 Ó 2009 American Society for Bone and Mineral Research Contribution of the

More information

Product: Denosumab (AMG 162) Clinical Study Report: month Primary Analysis Date: 21 November 2016 Page 1

Product: Denosumab (AMG 162) Clinical Study Report: month Primary Analysis Date: 21 November 2016 Page 1 Date: 21 November 2016 Page 1 2. SYNOPSIS Name of Sponsor: Amgen Inc., Thousand Oaks, CA, USA Name of Finished Product: Prolia Name of Active Ingredient: denosumab Title of Study: Randomized, Double-blind,

More information

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers*

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers* JOURNAL OF BONE AND MINERAL RESEARCH Volume 13, Number 10, 1998 Blackwell Science, Inc. 1998 American Society for Bone and Mineral Research Bone Mineral and Body Composition Measurements: Cross-Calibration

More information

Metabolism Core. History: Services: Metabolism Core Published on Nutrition & Obesity Research Center (

Metabolism Core. History: Services: Metabolism Core Published on Nutrition & Obesity Research Center ( Metabolism Core The Metabolism Core was designed to provide state-of-the-art assessments of human energy expenditure, substrate metabolism, body composition, body fat distribution, and bone quality; to

More information

LUMBAR IS IT IMPORTANT? S. Tantawy,, M.D.

LUMBAR IS IT IMPORTANT? S. Tantawy,, M.D. بسم االله الرحمن الرحيم DEXA LATERAL LUMBAR IS IT IMPORTANT? By S. Tantawy,, M.D. Osteopenia,, bone mineral deficiency in the absence of fracture, is an indicator of the bone structural integrity and compared

More information

Interpreting DEXA Scan and. the New Fracture Risk. Assessment. Algorithm

Interpreting DEXA Scan and. the New Fracture Risk. Assessment. Algorithm Interpreting DEXA Scan and the New Fracture Risk Assessment Algorithm Prof. Samir Elbadawy *Osteoporosis affect 30%-40% of women in western countries and almost 15% of men after the age of 50 years. Osteoporosis

More information

Patients with diffuse idiopathic skeletal hyperostosis do not have increased peripheral bone mineral density and geometry

Patients with diffuse idiopathic skeletal hyperostosis do not have increased peripheral bone mineral density and geometry RHEUMATOLOGY Rheumatology 2010;49:977 981 doi:10.1093/rheumatology/keq014 Advance Access publication 15 February 2010 Concise report Patients with diffuse idiopathic skeletal hyperostosis do not have increased

More information

QCT and CT applications in Osteoporosis Imaging

QCT and CT applications in Osteoporosis Imaging Q appli in Osteoporosis Imaging Thomas M. Link, MD, PhD Department of Radiology Biomedical Imaging University of California, San Francisco Goals 1. To identify advantages disadvantages of Q compared to

More information

Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA

Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA ORIGINAL ARTICLE Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA Leila Amiri 1, Azita Kheiltash 2, Shafieh Movassaghi 1, Maryam Moghaddassi 1, and Leila Seddigh 2 1 Rheumatology

More information

Mineral Density Of Subchondral Bone May Be Quantitatively Evaluated Using A Clinical Cone Beam Computed Tomography Scanner

Mineral Density Of Subchondral Bone May Be Quantitatively Evaluated Using A Clinical Cone Beam Computed Tomography Scanner Mineral Density Of Subchondral Bone May Be Quantitatively Evaluated Using A Clinical Cone Beam Computed Tomography Scanner Mikael J. Turunen, PhD 1, Juha Töyräs, PhD 1, Harri Kokkonen, PhD 2, Jukka S.

More information

Lunar Prodigy Advance

Lunar Prodigy Advance GE Medical Systems Lunar Prodigy Advance Direct-Digital Densitometry imagination at work Your practice needs to move fast, yet you want peace of mind. A partnership is a journey - expertise, support and

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements Standards Revisions for Organizations Providing Fluoroscopy Services The Joint Commission has approved the following revisions for prepublication. While revised requirements

More information

Quality Control of DXA System and Precision Test of Radio-technologists

Quality Control of DXA System and Precision Test of Radio-technologists J Bone Metab 2014;21:2-7 http://dx.doi.org/10.11005/jbm.2014.21.1.2 pissn 2287-6375 eissn 2287-7029 Review Article Quality Control of DXA System and Precision Test of Radio-technologists Ho-Sung Kim 1,

More information

Effect of a single botulinum toxin injection on bone development in growing rabbits

Effect of a single botulinum toxin injection on bone development in growing rabbits J Musculoskelet Neuronal Interact 2006; 6(3):264-268 Original Article Hylonome Effect of a single botulinum toxin injection on bone development in growing rabbits F. Rauch and R. Hamdy Genetics Unit, Shriners

More information

CLINIQCT NO-DOSE CT BONE DENSITOMETRY FOR ROUTINE AND SPECIALIST USE.

CLINIQCT NO-DOSE CT BONE DENSITOMETRY FOR ROUTINE AND SPECIALIST USE. CLINIQCT NO-DOSE CT BONE DENSITOMETRY FOR ROUTINE AND SPECIALIST USE Clinically superior BMD solutions for physicians DXA equivalent hip measurements Dual-use of standard abdominal or pelvic CT studies

More information

THE DIAGNOSIS OF OSTEOPOROSIS BY MEASURING LUMBAR VERTEBRAE DENSITY WITH MDCT: A COMPARATIVE STUDY WITH QUANTITATIVE COMPUTERIZED TOMOGRAPHY (QCT)

THE DIAGNOSIS OF OSTEOPOROSIS BY MEASURING LUMBAR VERTEBRAE DENSITY WITH MDCT: A COMPARATIVE STUDY WITH QUANTITATIVE COMPUTERIZED TOMOGRAPHY (QCT) Acta Medica Mediterranea, 2013, 29: 775 THE DIAGNOSIS OF OSTEOPOROSIS BY MEASURING LUMBAR VERTEBRAE DENSITY WITH MDCT: A COMPARATIVE STUDY WITH QUANTITATIVE COMPUTERIZED TOMOGRAPHY (QCT) KEMAL KARA 1,

More information

Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Children s Healthcare of Atlanta

Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Children s Healthcare of Atlanta Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Technical: Patient positioning Performance of the scan Analysis of the data Theoretical: Identification

More information

Greater Polar Moment of Inertia at the Tibia in Athletes Who Develop Stress Fractures

Greater Polar Moment of Inertia at the Tibia in Athletes Who Develop Stress Fractures Greater Polar Moment of Inertia at the Tibia in Athletes Who Develop Stress Fractures Lee A. Weidauer,* PhD, Teresa Binkley, PhD, Matt Vukovich, PhD, and Bonny Specker, PhD, Investigation performed at

More information

Bone mass and density response to a 12-month trial of calcium and vitamin D supplement in preadolescent girls

Bone mass and density response to a 12-month trial of calcium and vitamin D supplement in preadolescent girls J Musculoskel Neuron Interact 2003; 3(1):63-70 Original Article Hylonome Bone mass and density response to a 12-month trial of calcium and vitamin D supplement in preadolescent girls L.J. Moyer-Mileur

More information

Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13

Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13 Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13 Stephen J Pintauro, Tim R Nagy, Christa M Duthie, and Michael

More information

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status Journal of Clinical Densitometry, vol. 10, no. 3, 239e243, 2007 Ó Copyright 2007 by The International Society for Clinical Densitometry 1094-6950/07/10:239e243/$32.00 DOI: 10.1016/j.jocd.2007.03.002 Original

More information

Osteoporosis New Insights

Osteoporosis New Insights Osteoporosis New Insights by JOHN KINNEY Synchrotron radiation is giving us new insights into causes and treatments for osteoporosis. OSTEOPOROSIS IS A DISEASE characterized by fragile bones that results

More information

# % # & # # ( # ) # % # # ) +,#. # ) / ) 3 ) ) 7 55, # 09 /:2 ;88% 111 <!= <01 9<1:<:1:==

# % # & # # ( # ) # % # # ) +,#. # ) / ) 3 ) ) 7 55, # 09 /:2 ;88% 111 <!= <01 9<1:<:1:== ! # % # & # # ( # ) # % # # ) +,#. # ) /01 2 + ) 3 ) 4 5 6 ) 7 55, 5 8 3 # 09 /:2 ;88% 111 This is the accepted version of the following article: Joshua N. Farr, Wei Zhang, Shaji

More information

Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry

Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry J Bone Metab 2017;24:105-109 https://doi.org/10.11005/jbm.2017.24.2.105 pissn 2287-6375 eissn 2287-7029 Original Article Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Birth Date: 40.2 years Height / Weight: 158.0 cm 52.0 kg Sex / Ethnic: Female Patient ID: Total Body Tissue Quantitation Composition Reference: Total Tissue 50% 40% 30% 20% 20 30 40 50 60 70 80 90 100

More information

Prodigy. from GE Healthcare. Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world. gehealthcare.

Prodigy. from GE Healthcare. Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world. gehealthcare. Prodigy from GE Healthcare Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world gehealthcare.com Prodigy High performance, efficient and reliable DXA system

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Changes in volumetric BMD of radius and tibia upon antidepressant drug administration in young depressive patients

Changes in volumetric BMD of radius and tibia upon antidepressant drug administration in young depressive patients J Musculoskelet Neuronal Interact 2012; 12(4):224-229 Original Article Hylonome Changes in volumetric BMD of radius and tibia upon antidepressant drug administration in young depressive patients L. Winterhalder

More information

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Birth Date: 24.7 years Height / Weight: 8.0 cm 79.0 kg Sex / Ethnic: Male Patient ID: Total Body Tissue Quantitation Composition Reference: Total Tissue 40% 30% 20% 0% 20 30 40 50 60 70 80 90 00 Centile

More information

Clinical Densitometry

Clinical Densitometry Volume 8 Number 3 Fall 2005 ISSN: 1094 6950 Journal of Clinical Densitometry The Official Journal of The International Society for Clinical Densitometry Editor-in-Chief Paul D. Miller, MD HumanaJournals.com

More information

Patient Guide. Intramedullary Skeletal Kinetic Distractor For Tibial and Femoral Lengthening

Patient Guide. Intramedullary Skeletal Kinetic Distractor For Tibial and Femoral Lengthening Patient Guide Intramedullary Skeletal Kinetic Distractor For Tibial and Femoral Lengthening Introduction You have decided to have a limb lengthening operation. The surgery you have chosen uses a device

More information

Nutritional Aspects of Osteoporosis Care and Treatment Cynthia Smith, FNP-BC, RN, MSN, CCD Pars Osteoporosis Clinic, Belpre, Ohio

Nutritional Aspects of Osteoporosis Care and Treatment Cynthia Smith, FNP-BC, RN, MSN, CCD Pars Osteoporosis Clinic, Belpre, Ohio Osteoporosis 1 Nutritional Aspects of Osteoporosis Care and Treatment Cynthia Smith, FNP-BC, RN, MSN, CCD Pars Osteoporosis Clinic, Belpre, Ohio 1) Objectives: a) To understand bone growth and development

More information

To understand bone growth and development across the lifespan. To develop a better understanding of osteoporosis.

To understand bone growth and development across the lifespan. To develop a better understanding of osteoporosis. Nutrition Aspects of Osteoporosis Care and Treatment t Cynthia Smith, FNP-BC, RN, MSN, CCD Pars Osteoporosis Clinic, Belpre, OH. Objectives To understand bone growth and development across the lifespan.

More information

Purpose. Methods and Materials

Purpose. Methods and Materials Prevalence of pitfalls in previous dual energy X-ray absorptiometry (DXA) scans according to technical manuals and International Society for Clinical Densitometry. Poster No.: P-0046 Congress: ESSR 2014

More information

Micro-CT imaging of surgical screw tightening in human trabecular bone

Micro-CT imaging of surgical screw tightening in human trabecular bone Micro-CT imaging of surgical screw tightening in human trabecular bone E. Perilli 1, M. Ryan 1, R. Ab-Lazid 1, J.J. Costi 1, K.J. Reynolds 1 1 Medical Device Research Institute, School of Computer Science,

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 48.2 years Height / Weight: 150.0 cm 72.0 kg Sex / Ethnic: Female

More information

UKnowledge. University of Kentucky. Catherine E. O'Brien University of Arkansas at Little Rock. Gulnur Com University of Southern California

UKnowledge. University of Kentucky. Catherine E. O'Brien University of Arkansas at Little Rock. Gulnur Com University of Southern California University of Kentucky UKnowledge Barnstable Brown Kentucky Diabetes and Obesity Center Faculty Publications Diabetes and Obesity 1-11-2018 Peripheral Quantitative Computed Tomography Detects Differences

More information

Observations on the Geometric Distribution of Micro-structural Features in Cortical Bone

Observations on the Geometric Distribution of Micro-structural Features in Cortical Bone Observations on the Geometric Distribution of Micro-structural Features in Cortical Bone I. S. Hage, R. F. Hamade Abstract This paper reports on observations of the distribution of micro-structural features

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 43.4 years Height / Weight: 170.0 cm 66.0 kg Sex / Ethnic: Female

More information

DEXA Scores and Bone Density Measured on Routine CT Scans

DEXA Scores and Bone Density Measured on Routine CT Scans DEXA Scores and Bone Density Measured on Routine CT Scans Poster No.: R-0072 Congress: 2015 ASM Type: Scientific Exhibit Authors: C. Annabattula, P. Phadke; Sydney/AU Keywords: Screening, Nuclear medicine

More information

Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems?

Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems? Osteoporos Int (2010) 21:1227 1236 DOI 10.1007/s00198-009-1062-3 ORIGINAL ARTICLE Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems? B. Fan & Y. Lu

More information

Least significant changes and monitoring time intervals for high-resolution pqct-derived bone outcomes in postmenopausal women

Least significant changes and monitoring time intervals for high-resolution pqct-derived bone outcomes in postmenopausal women J Musculoskelet Neuronal Interact 2015; 15(2):190-196 Original Article Hylonome Least significant changes and monitoring time intervals for high-resolution pqct-derived bone outcomes in postmenopausal

More information

BONE MASS is lower in women than in men, and this

BONE MASS is lower in women than in men, and this 0021-972X/97/$03.00/0 Vol. 82, No. 5 Journal of Clinical Endocrinology and Metabolism Printed in U.S.A. Copyright 1997 by The Endocrine Society Differential Effect of Gender on the Sizes of the Bones in

More information

Assessing Bone Quality in Terms of Bone Mineral Density, Buckling Ratio and Critical Fracture Load

Assessing Bone Quality in Terms of Bone Mineral Density, Buckling Ratio and Critical Fracture Load J Bone Metab 01;1:- http://dx.doi.org/10.1100/jbm.01.1.. pissn - eissn -0 Original Article Assessing Bone Quality in Terms of Bone Mineral Density, Buckling Ratio and Critical Fracture Load Anitha D 1,

More information

Bone Densitometry. What is a Bone Density Scan (DXA)? What are some common uses of the procedure?

Bone Densitometry. What is a Bone Density Scan (DXA)? What are some common uses of the procedure? Scan for mobile link. Bone Densitometry What is a Bone Density Scan (DXA)? Bone density scanning, also called dual-energy x-ray absorptiometry (DXA) or bone densitometry, is an enhanced form of x-ray technology

More information

PROPHECY. Preoperative Navigation Guides ANKLE CT SCAN PROTOCOL

PROPHECY. Preoperative Navigation Guides ANKLE CT SCAN PROTOCOL PROPHECY Preoperative Navigation Guides ANKLE CT SCAN PROTOCOL 90 FIGURE 1 Examples FIGURE 1 Examples of neutral ankle positioning. PROPHECY Ankle CT Scan Protocol PROPHECY INBONE and PROPHECY INFINITY

More information

The Development of Metaphyseal Cortex Implications for Distal Radius Fractures During Growth

The Development of Metaphyseal Cortex Implications for Distal Radius Fractures During Growth JOURNAL OF BONE AND MINERAL RESEARCH Volume 16, Number 8, 2001 2001 American Society for Bone and Mineral Research The Development of Metaphyseal Cortex Implications for Distal Radius Fractures During

More information

Automated measurement method of bone mineral density of vertebral bones and spinal curvature using sagittal plane in X-ray torso CT images

Automated measurement method of bone mineral density of vertebral bones and spinal curvature using sagittal plane in X-ray torso CT images 501-1194 1-1 501-1194 1-1 501-1194 1-1 501-1194 1-1 501-1194 1-1 2009 2 9 2009 6 4 Automated measurement method of bone mineral density of vertebral bones and spinal curvature using sagittal plane in X-ray

More information

Building Bone Density-Research Issues

Building Bone Density-Research Issues Building Bone Density-Research Issues Helping to Regain Bone Density QUESTION 1 What are the symptoms of Osteoporosis? Who is at risk? Symptoms Bone Fractures Osteoporosis 1,500,000 fractures a year Kyphosis

More information

MECHANICAL PROPERTIES OF A SINGLE CANCELLOUS BONE TRABECULAE TAKEN FROM BOVINE FEMUR

MECHANICAL PROPERTIES OF A SINGLE CANCELLOUS BONE TRABECULAE TAKEN FROM BOVINE FEMUR Advanced Materials Development and Performance (AMDP2011) International Journal of Modern Physics: Conference Series Vol. 6 (2012) 349-354 World Scientific Publishing Company DOI: 10.1142/S201019451200342X

More information

Bone Mineral Density of Healthy Turkish Children and Adolescents

Bone Mineral Density of Healthy Turkish Children and Adolescents Journal of Clinical Densitometry, vol. 9, no., 8 90, 006 Ó Copyright 006 by The International Society for Clinical Densitometry 09-690/06/9:8 90/$.00 DOI: 0.06/j.jocd.00.08.00 Original Article Bone Mineral

More information

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT O S T E O P O R O S I S A S S E S S M E N T Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT Identify Patients at Risk The evaluation of bone health has become an essential

More information

New Dual-energy X-ray Absorptiometry Machines (idxa) and Vertebral Fracture Assessment

New Dual-energy X-ray Absorptiometry Machines (idxa) and Vertebral Fracture Assessment Case 1 New Dual-energy X-ray Absorptiometry Machines (idxa) and Vertebral Fracture Assessment (VFA) History and Examination Your wealthy friend who is a banker brings his 62-year-old mother to your office

More information

Objectives. Discuss bone health and the consequences of osteoporosis on patients medical and disability status.

Objectives. Discuss bone health and the consequences of osteoporosis on patients medical and disability status. Objectives Discuss bone health and the consequences of osteoporosis on patients medical and disability status. Discuss the pathophysiology of osteoporosis and major risk factors. Assess the major diagnostic

More information

support optimum bone mass accrual during rapid weight loss to prevent fracture and early onset osteoporosis.

support optimum bone mass accrual during rapid weight loss to prevent fracture and early onset osteoporosis. The BOW Study (Bone in Weight Loss): The effect of profound weight loss following intragastric balloon placement on skeletal integrity in severely obese adolescents. Background: The effect of profound

More information