An assessment of the validity of a subjective hearing aid verification procedure

Size: px
Start display at page:

Download "An assessment of the validity of a subjective hearing aid verification procedure"

Transcription

1 An assessment of the validity of a subjective hearing aid verification procedure Whitney M. Vineyard, B. A. AuD Research Project April, 211 University of Memphis Project Advisor Robyn M. Cox, PhD.

2 Purpose Hearing aids are fit according to rules derived from evidence-based prescriptions. The best way to evaluate a hearing aid is using objective measurement, using real ear measures. However, many audiologists do not use real ear measurements to verify that the hearing aid is working according to fitting goals.

3 Research Questions 1. How well does a structured subjective test battery work to verify that a hearing aid is matching fitting goals? 2. How much time does it take to run a thorough subjective test battery? 3. What are the limitations of this kind of approach?

4 How fitting goals differ from NAL-NL1 NAL-NL1 prescription targets do not exactly align with fitting goals at MSHC, where we aim to: Match 55 db targets. Fall 5 db below 7 db targets. Verify MPO targets primarily through subjective verification. We use subjective verification procedures to verify that hearing aids are set to appropriate loudness levels.

5 Methods The test battery was performed using two different hearing aids on each subject. The test battery consists of three steps: 1. repeating speech sounds to assess soft speech audibility 2. rating average speech loudness and 3. rating loud sound comfort. The subject was asked to repeat sounds or rate loudness comfort. After completion of the test battery, the output of the hearing aid was compared to NAL-NL1 prescription targets which were generated by entering hearing thresholds into a Verifit system.

6 Participants Subjects were users of hearing aids with sensorineural hearing loss and no history of middle ear pathology or ear surgery. Subjects were recruited from the Memphis Speech and Hearing Center and Hearing Aid Research Laboratory. 2 Subjects 11 male, 9 female Average age: 76 years Average length of hearing aid use: 6.6 years. Average amount of wear per day: 11.2 hours.

7 Flow Charts In order to standardize the test battery, flow charts were developed. These charts clarified the adjustments that were to be made in the software based on the responses of the subject.

8 Step 1: subjective verification of soft speech audibility START: Administer 6 sound test. Did subject hear /a/,/i/, /m/ & /u/? Yes Did subject hear /sh/ correctly? Yes No No Is this > 3 rd repetition Yes Is this > 3 rd repetition Yes No No Turn up overall gain for soft sounds 5 db and readminister test. Turn up gain for soft sounds above 2 Hz 3 db and re-administer test. Turn up gain for soft sounds above 4 Hz 3 db and readminister test. No Is this > 3 rd repetiti on Yes No Did subject hear /s/ correctly? Yes Move on to STEP 2

9 Step 2: subjective verification of average speech loudness Administer test: ISTS track played at 65 db SPL to represent average speech Have the subject give a loudness category Refer to Table 1 for adjustment information based on level of experience Is the sound > the goal? No Is the sound < the goal? No Yes Yes Is this > 3 rd repetition Yes Is this > 3 rd repetition Yes No No Turn DOWN loud gain in all frequencies for average sound 3 db Turn UP loud gain in all frequencies for average sound 3 db Move on to STEP 3

10 Step 3a: subjective verification of loud sound comfortbroadband MPO Administer test: Broadband MPO Sound Have the subject give a loudness category Refer to Table 1 for adjustment information based on level of experience Is the sound > 6? No Is the sound < 6? No Yes Yes Is this > 3 rd repetition Yes Is this > 3 rd repetition Yes No No Turn DOWN overall MPO by 3 db Turn UP overall MPO by 3 db Move on to STEP 3b

11 Step 3b: subjective verification of loud sound comforthigh frequency MPO Administer test: High Frequency MPO Sound Have the subject give a loudness category Refer to Table 1 for adjustment information based on level of experience Is the sound > 6? No Is the sound < 6? No Yes Yes Is this > 3 rd repetition Yes Is this > 3 rd repetition Yes No No Turn DOWN MPO > 3kHz by 3 db Turn UP MPO > 3 khz by3 db Test complete

12 Starkey Inspire Software For Step 1, adjustments were made using the Soft Sounds Channel. For Step 2, adjustments were made using the Loud Sounds Channel. For Step 3, adjustments were made using the MPO arrows.

13 Siemens CONNEXX Software For Step 1, adjustments were made using the Soft sounds arrows. For Step 2, adjustments were made using the Speech arrows. After review, examiners realized that post-programming results were not in line with the goals of the verification battery. As a result, this protocol was restructured after 1 subjects. For Step 3, adjustments were made using the output limiting bar.

14 Siemens CONNEXX Software For Step 1, adjustments were made using the Master gain arrows. For Step 2, adjustments were made using the Loud sounds arrows. For Step 3, adjustments were made using the output limiting bar.

15 Audiometric Data, all test ears

16 Results 1. How well did the structured subjective test battery work to verify that a hearing aid is matching fitting goals? The accuracy of the fitting was evaluated by subtracting post-programming REAR from the individual s NAL-NL1 prescription targets across the frequency range.

17 db difference from target db difference from target Difference from target after verification, 55 db Starkey Siemens

18 db difference from target db difference from target Difference from target after verification, 7 db Starkey Siemens

19 db difference from target db difference from target Difference from target after verification, MPO Starkey Siemens

20 Results 1. How well did the structured subjective test battery work to verify that a hearing aid is matching fitting goals? Question: Did the subjects move closer to fitting goals or further away?

21 Difference from Rx target (db SPL) Difference from target before vs. after, db Starkey

22 Difference from Rx target (db SPL) Difference from target before vs. after, db Siemens

23 Difference from Rx target (db SPL) Difference from target before vs. after, db Starkey

24 Difference from Rx target (db SPL) Difference from target before vs. after, db Siemens

25 Difference from Rx target (db SPL) Difference from target before vs. after, MPO Starkey

26 Difference from Rx target (db SPL) Difference from target before vs. after, MPO Siemens

27 db difference from target db difference from target Difference from target before vs. after, 55 db Starkey First 1 Subjects Last 1 Subjects

28 db difference from target db difference from target Difference from target before vs. after, 55 db Siemens First 1 Subjects Last 1 Subjects

29 db difference from target db difference from target Difference from target before vs. after, 7 db Starkey First 1 Subjects Last 1 Subjects

30 db difference from target db difference from target Difference from target before vs. after, MPO Siemens First 1 Subjects Last 1 Subjects

31 Statistical Analysis T-tests and ANOVA tests were performed to evaluate the relationship between the preverification fitting and the post-verification fitting. Question: Were the adjustments that we made to the hearing aids statistically significant differences? Answer: Yes, for many frequencies a statistically significant change occurred. This means that the result was likely a product of our test battery, and unlikely due to chance.

32 Summary of Statistics Subjective Test Starkey Siemens Soft Speech Closer to target Closer to target Average Speech MPO Closer to db re: target fitting goal Further below target NO CHANGE Further below target For all frequencies where a significant change occurred, it brought the hearing aid output closer to fitting goals.

33 Summary of Results Results showed that for both hearing aids, soft speech moved closer to targets. Average speech results were mixed, with one hearing aid moving closer to db re: 7 db targets, and the other with no change. MPO moved further below targets (closer to fitting goals) for both hearing aids.

34 Results: 2. How much time does it take to run a thorough subjective test battery? For an average subject It took 8.5 minutes for the Starkey aid, and it took 7.5 minutes for the Siemens aid Siemens 8 Starkey

35 Results: 3. What are the limitations of this kind of approach? Results were dependant upon the motivation and cognitive ability of the subjects. Variability across the subject range increased after verification. It is important to really know what changes the software is making on the hearing aids! For example, the speech adjustment bar for one hearing aid software made changes across the entire dynamic range.

36 Conclusions Subjective verification has the ability to bring hearing aid wearers closer to fitting goals when real ear verification is not possible. Subjective verification worked better for one hearing aid than another. It takes about 8 minutes to perform the subjective test battery.

Verification of soft speech amplification in hearing aid fitting: A comparison of methods

Verification of soft speech amplification in hearing aid fitting: A comparison of methods Verification of soft speech amplification in hearing aid fitting: A comparison of methods Sarah E. Dawkins, B.A. AuD Research Project April 5, 2007 University of Memphis Project Advisor Robyn M. Cox, PhD.

More information

Manufacturers NAL-NL2 Fittings Fail Real-ear Verification

Manufacturers NAL-NL2 Fittings Fail Real-ear Verification hearingreview.com http://www.hearingreview.com/2015/02/manufacturers-nal-nl2-fittings-fail-real-ear-verification/ Manufacturers NAL-NL2 Fittings Fail Real-ear Verification Research March 2015 Hearing Review

More information

Audiology Today MarApr2011

Audiology Today MarApr2011 26 Stay on Target? By Jason A. Galster The focus on gain and targets has obscured the most important part of the real-ear measurement audibility. The topic of audibility is familiar; after all, a primary

More information

AccuQuest Spotlight: Successful Fittings with Oasis. Fitting Range

AccuQuest Spotlight: Successful Fittings with Oasis. Fitting Range AccuQuest Spotlight: Successful Fittings with Oasis When fitting a client with hearing instruments, the experience should be positive for both the client and you, the hearing care professional (HCP). The

More information

You regular readers of Page Ten

You regular readers of Page Ten PAGE 0 You regular readers of Page Ten probably recall that last month we had to interrupt Robyn Cox, PhD, and her inquiring friend in the middle of their discussion on verification procedures for the

More information

The Devil is in the Fitting Details. What they share in common 8/23/2012 NAL NL2

The Devil is in the Fitting Details. What they share in common 8/23/2012 NAL NL2 The Devil is in the Fitting Details Why all NAL (or DSL) targets are not created equal Mona Dworsack Dodge, Au.D. Senior Audiologist Otometrics, DK August, 2012 Audiology Online 20961 What they share in

More information

Fitting Decisions and their Impact on Hearing Aid User Benefit. Mallory Maine, AuD Audiologist, GN ReSound

Fitting Decisions and their Impact on Hearing Aid User Benefit. Mallory Maine, AuD Audiologist, GN ReSound Fitting Decisions and their Impact on Hearing Aid User Benefit Mallory Maine, AuD Audiologist, GN ReSound Agenda Common Fitting Oversights #1 Setting the coupler type in fitting software To set or not

More information

Verifying the Lyric Audibility Advantage

Verifying the Lyric Audibility Advantage Field Study News Spring 2014 Verifying the Lyric Audibility Advantage John Pumford, Au.D. Manager of Clinical Research Phonak Lyric Introduction Since 2008, Lyric has presented people with mild to moderate

More information

Optimising Outcomes for Speech Mapping

Optimising Outcomes for Speech Mapping Optimising Outcomes for Speech Mapping Jan Pollard Chief Audiologist Sonic Craig Lett Audiologist ECS Back to Basics What is the point of all of this? Optimising the fitting for best patient outcomes and

More information

Validation Studies. How well does this work??? Speech perception (e.g., Erber & Witt 1977) Early Development... History of the DSL Method

Validation Studies. How well does this work??? Speech perception (e.g., Erber & Witt 1977) Early Development... History of the DSL Method DSL v5.: A Presentation for the Ontario Infant Hearing Program Associates The Desired Sensation Level (DSL) Method Early development.... 198 Goal: To develop a computer-assisted electroacoustic-based procedure

More information

Mistakes on Real Ear Measures Clinicians Often Make. Michael Valente

Mistakes on Real Ear Measures Clinicians Often Make. Michael Valente Mistakes on Real Ear Measures Clinicians Often Make Michael Valente There are two national guidelines (ASHA, 1998; AAA 2006) stating the need to use REM to verify the performance of hearing aids in order

More information

Clinical Tips for Matching Real-Ear Prescriptive Targets

Clinical Tips for Matching Real-Ear Prescriptive Targets AudiologyNOW! 2010 San Diego, CA Clinical Tips for Matching Real-Ear Prescriptive Targets Gabrielle Filips, Au.D., FAAA August Hernandez, Au.D., FAAA 1 Discussion topics Who s performing probe-microphone

More information

Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial

Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial Enhancing the Initial Hearing Aid Fitting Michael Block, Ph.D. Session 0228 Importance of a Good Start The consequences of doing a poor job of selling the hearing-impaired client on a positive treatment

More information

Audiogram+: The ReSound Proprietary Fitting Algorithm

Audiogram+: The ReSound Proprietary Fitting Algorithm Abstract Hearing instruments should provide end-users with access to undistorted acoustic information to the degree possible. The Resound compression system uses state-of-the art technology and carefully

More information

Technical Topics. Where did my gain go? Thin Tube Open Fit BTE Verification. History of Thin Tubes: Summary of presentation by C. Staples and S.

Technical Topics. Where did my gain go? Thin Tube Open Fit BTE Verification. History of Thin Tubes: Summary of presentation by C. Staples and S. Technical Topics Where did my gain go? Thin Tube Open Fit BTE Verification Summary of presentation by C. Staples and S. Aiken History of Thin Tubes: Growth of new BTE market: First quarter of 2006 market:

More information

Marlene Bagatto The University of Western Ontario, London, Canada. European Pediatric Amplification Conference Istanbul, Turkey November 15, 2011

Marlene Bagatto The University of Western Ontario, London, Canada. European Pediatric Amplification Conference Istanbul, Turkey November 15, 2011 Marlene Bagatto The University of Western Ontario, London, Canada European Pediatric Amplification Conference Istanbul, Turkey November 15, 2011 Intervention for Childhood Hearing Loss Access to early

More information

In the days before programmable hearing

In the days before programmable hearing PAGE 1 In the days before programmable hearing aids, the patient was fitted to the best gain and output through the use of numerous screwdriver adjustments and even by trying different instruments until

More information

Audiogram+: GN Resound proprietary fitting rule

Audiogram+: GN Resound proprietary fitting rule Audiogram+: GN Resound proprietary fitting rule Ole Dyrlund GN ReSound Audiological Research Copenhagen Loudness normalization - Principle Background for Audiogram+! Audiogram+ is a loudness normalization

More information

Baker, A., M.Cl.Sc (AUD) Candidate University of Western Ontario: School of Communication Sciences and Disorders

Baker, A., M.Cl.Sc (AUD) Candidate University of Western Ontario: School of Communication Sciences and Disorders Critical Review: Effects of multi-channel, nonlinear frequency compression on speech perception in hearing impaired listeners with high frequency hearing loss Baker, A., M.Cl.Sc (AUD) Candidate University

More information

Building Practice Success Through Speechmap Real Ear Measurements. Chris Stokes-Rees

Building Practice Success Through Speechmap Real Ear Measurements. Chris Stokes-Rees Building Practice Success Through Speechmap Real Ear Measurements Chris Stokes-Rees Speaker Disclosure Relevant financial relationships: Audioscan (Applications Specialist) Conestoga College, Kitchener

More information

Feelin The Squeeze? 10/17/2016. Any of these been on your mind lately? Key Part of NAS 12 Recommendations

Feelin The Squeeze? 10/17/2016. Any of these been on your mind lately? Key Part of NAS 12 Recommendations Any of these been on your mind lately? The Art of Independent Practice In Today s Td Corporate/Commodity Culture Presenter David J. Smriga, M.A. Senior Audiology Consultant Audioscan Education MAC Conference

More information

DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015.

DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015. www.bestsound-technology.com DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015. First fit is an essential stage of the hearing aid fitting process and is a cornerstone of the ultimate

More information

Best practice protocol

Best practice protocol Best practice protocol April 2016 Pediatric verification for SoundRecover2 What is SoundRecover? SoundRecover is a frequency lowering signal processing available in Phonak hearing instruments. The aim

More information

Phonak Target. SoundRecover2 adult fitting guide. Content. The Connecting the hearing instruments. February 2018

Phonak Target. SoundRecover2 adult fitting guide. Content. The Connecting the hearing instruments. February 2018 Phonak Target February 2018 SoundRecover2 adult fitting guide The following fitting guide is intended for adults. For Pediatric fittings please see the separate Pediatric fitting guide. SoundRecover2 is

More information

Hearing Protection for the Workplace

Hearing Protection for the Workplace Hearing Protection for the Workplace Hear clearly and protect your ears throughout your working day. When you work in a noisy environment, hearing clearly is key to effective communication, success and

More information

Top 10 ideer til en god høreapparat tilpasning. Astrid Haastrup, Audiologist GN ReSound

Top 10 ideer til en god høreapparat tilpasning. Astrid Haastrup, Audiologist GN ReSound HVEM er HVEM Top 10 ideer til en god høreapparat tilpasning Astrid Haastrup, Audiologist GN ReSound HVEM er HVEM 314363 WRITE DOWN YOUR TOP THREE NUMBER 1 Performing appropriate hearing assessment Thorough

More information

binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way

binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way www.bestsound-technology.com binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way Richard Schultz-Amling Sivantos, 2015. 1 Introduction Since the release

More information

A comparison of manufacturer-specific prescriptive procedures for infants

A comparison of manufacturer-specific prescriptive procedures for infants A comparison of manufacturer-specific prescriptive procedures for infants By Richard Seewald, Jillian Mills, Marlene Bagatto, Susan Scollie, and Sheila Moodie Early hearing detection and communication

More information

Clinical fitting guide

Clinical fitting guide Fitting Super Power is now super easy A quick reference to the Super Power features of Max: Super Power pre-sets Frequency compression 2 Power Adaptation Manager (PAM) A Sonova brand Why are we making

More information

Audibility, discrimination and hearing comfort at a new level: SoundRecover2

Audibility, discrimination and hearing comfort at a new level: SoundRecover2 Audibility, discrimination and hearing comfort at a new level: SoundRecover2 Julia Rehmann, Michael Boretzki, Sonova AG 5th European Pediatric Conference Current Developments and New Directions in Pediatric

More information

Conversations on Verification Part I. Hearing Aid Fitting Errors in Oregon Ron Leavitt, Nikki Clark & Camille Jenkins

Conversations on Verification Part I. Hearing Aid Fitting Errors in Oregon Ron Leavitt, Nikki Clark & Camille Jenkins Conversations on Verification Part I Hearing Aid Fitting Errors in Oregon Ron Leavitt, Nikki Clark & Camille Jenkins DISCLOSURE STATEMENTS RON LEAVITT, AUD Financial Disclosures: Nothing to disclose Non-Financial

More information

Abstract.

Abstract. Combining Phase Cancellation, Frequency Shifting and Acoustic Fingerprint for Improved Feedback Suppression Josef Chalupper, Thomas A. Powers, Andre Steinbuss www.siemens.com Abstract Acoustic feedback

More information

Initial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit

Initial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit Initial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit Harvey B. Abrams 1,2 Theresa Hnath Chisolm,2,1 Megan McManus 2 Rachel A.McArdle 1,2 1 Bay Pines VA Healthcare

More information

Fitting Frequency Compression Hearing Aids to Kids: The Basics

Fitting Frequency Compression Hearing Aids to Kids: The Basics Fitting Frequency Compression Hearing Aids to Kids: The Basics Presenters: Susan Scollie and Danielle Glista Presented at AudiologyNow! 2011, Chicago Support This work was supported by: Canadian Institutes

More information

The REM Cookbook How to Correctly Perform Real Ear Measurements

The REM Cookbook How to Correctly Perform Real Ear Measurements www.sonici.com The REM Cookbook How to Correctly Perform Real Ear Measurements The REM Cookbook Terminology Why do audiologists perform real ear measurements? The main reason is to verify that speech sounds

More information

Muse Wireless CROS System

Muse Wireless CROS System Clinical Validation of the Muse Wireless CROS System Ashley Hughes, Au.D., & Chad Parrish Introduction Fitting individuals with unilateral hearing loss (UHL) or asymmetrical hearing loss can pose unique

More information

ChildFit. Widex Baby. Compass quick guide

ChildFit. Widex Baby. Compass quick guide ChildFit When working with very young children, it is not always possible to obtain all of the information necessary for an accurate hearing aid fitting. To that end, Widex has developed the ChildFit procedure

More information

Candidacy and Verification of Oticon Speech Rescue TM technology

Candidacy and Verification of Oticon Speech Rescue TM technology PAGE 1 TECH PAPER 2015 Candidacy and Verification of Oticon Speech Rescue TM technology Kamilla Angelo 1, Marianne Hawkins 2, Danielle Glista 2, & Susan Scollie 2 1 Oticon A/S, Headquarters, Denmark 2

More information

2/16/2012. Fitting Current Amplification Technology on Infants and Children. Preselection Issues & Procedures

2/16/2012. Fitting Current Amplification Technology on Infants and Children. Preselection Issues & Procedures Fitting Current Amplification Technology on Infants and Children Cindy Hogan, Ph.D./Doug Sladen, Ph.D. Mayo Clinic Rochester, Minnesota hogan.cynthia@mayo.edu sladen.douglas@mayo.edu AAA Pediatric Amplification

More information

Differences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: /jaaa

Differences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: /jaaa J Am Acad Audiol 24:660 670 (2013) Differences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: 10.3766/jaaa.24.8.3 Kristi Oeding* Michael Valente* Abstract Background:

More information

WIDEXPRESS THE WIDEX FITTING RATIONALE FOR EVOKE MARCH 2018 ISSUE NO. 38

WIDEXPRESS THE WIDEX FITTING RATIONALE FOR EVOKE MARCH 2018 ISSUE NO. 38 THE WIDEX FITTING RATIONALE FOR EVOKE ERIK SCHMIDT, TECHNICAL AUDIOLOGY SPECIALIST WIDEX GLOBAL RESEARCH & DEVELOPMENT Introduction As part of the development of the EVOKE hearing aid, the Widex Fitting

More information

[Connexx 7.2 with RexFit 7.2]

[Connexx 7.2 with RexFit 7.2] [Connexx 7.2 with RexFit 7.2] Programming Guide An easy fit is a successful fit. [Connexx 7.2] Table of Contents First Fit Guide... 4 First Fit Adjustment... 5 Firmware Updates... 6 Follow-Up Fittings

More information

Hearing-loss-dependent personalization

Hearing-loss-dependent personalization PAGE 1 HEARING LOSS DEPENDENT PERSONALIZATION TECH PAPER ULTIMO 2015 Hearing-loss-dependent personalization INTRODUCTION In the meeting between clinician and patient with severe to profound hearing loss,

More information

Issues faced by people with a Sensorineural Hearing Loss

Issues faced by people with a Sensorineural Hearing Loss Issues faced by people with a Sensorineural Hearing Loss Issues faced by people with a Sensorineural Hearing Loss 1. Decreased Audibility 2. Decreased Dynamic Range 3. Decreased Frequency Resolution 4.

More information

User groups Most sensorineural, conductive and mixed hearing losses in the categories of severe and profound.

User groups Most sensorineural, conductive and mixed hearing losses in the categories of severe and profound. P R O D U C T I N F O R M A T I O N S W I F T 0 + Fitting range db HL -0 0 0 0 0 0 0 Hz 0 00 0 000 00 00 General information Swift + is a very reliable and powerful programmable BTE instrument for people

More information

Walkthrough

Walkthrough 0 8. Walkthrough Simulate Product. Product selection: Same look as estore. Filter Options: Technology levels listed by descriptor words. Simulate: Once product is selected, shows info and feature set Order

More information

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system The information in this document contains general descriptions of the technical options available, which do not always have to be present in individual cases and are subject to change without prior notice.

More information

How to Use Own Voice Processing in Connexx 8. Rebecca Herbig, AuD

How to Use Own Voice Processing in Connexx 8. Rebecca Herbig, AuD How to Use Own Voice Processing in Connexx 8 Rebecca Herbig, AuD Introduction Signia Nx TM hearing aids solve the own voice issue with a sophisticated technological solution a realtime detection of the

More information

Testing Digital Hearing Aids

Testing Digital Hearing Aids Testing Digital Hearing Aids with the FONIX 6500-CX Hearing Aid Analyzer Frye Electronics, Inc. Introduction The following is a quick guide for testing digital hearing aids using the FONIX 6500-CX. All

More information

How to use AutoFit (IMC2) How to use AutoFit (IMC2)

How to use AutoFit (IMC2) How to use AutoFit (IMC2) How to use AutoFit (IMC2) 1 AutoFit is a beneficial feature in the Connexx Fitting Application that automatically provides the Hearing Care Professional (HCP) with an optimized real-ear insertion gain

More information

Best Practice Protocols

Best Practice Protocols Best Practice Protocols SoundRecover for children What is SoundRecover? SoundRecover (non-linear frequency compression) seeks to give greater audibility of high-frequency everyday sounds by compressing

More information

How to Use InSituGram

How to Use InSituGram How to Use InSituGram www.siemens.com Wendy Davies Abstract: Siemens InSituGram allows pure tone audiometry to be performed using the hearing instrument. The insitu audiogram will usually be very similar

More information

Phonak Target 4.3. Desktop Fitting Guide. Content. March 2016

Phonak Target 4.3. Desktop Fitting Guide. Content. March 2016 Phonak Target 4.3 March 2016 Desktop Fitting Guide This guide provides you with a detailed introduction to latest hearing instrument fitting with Phonak Target. www.phonakpro.com/target_guide Please also

More information

RECD-Measurements with Connexx 6.3 and Unity 2

RECD-Measurements with Connexx 6.3 and Unity 2 RECD-Measurements with Connexx 6.3 and Unity 2 Vera Spitzlei, Corporate Audiology Siemens Audiologische Technik GmbH For internal use only / Copyright Siemens AG 2009. All rights reserved. Overview Why

More information

Best practices in A few favorite resources: Clear areas of agreement: How about these features? The bottom line:

Best practices in A few favorite resources: Clear areas of agreement: How about these features? The bottom line: Today s hearing aids: Using technology to its best potential Susan Scollie, Ph.D. Associate Professor Child Amplification Laboratory School of Communication Sciences and Disorders National Centre for Audiology

More information

Beyond the Audiogram:

Beyond the Audiogram: Beyond the Audiogram: Diagnostic and Verification Tools to Improve Hearing Aid and Cochlear Implant Fittings Brad Ingrao, AuD David S. Andreaggi, AuD DISCLAIMER The information presented in this presentation

More information

International Journal of Health Sciences and Research ISSN:

International Journal of Health Sciences and Research  ISSN: International Journal of Health Sciences and Research www.ijhsr.org ISSN: 2249-9571 Original Research Article Effect of Compression Parameters on the Gain for Kannada Sentence, ISTS and Non-Speech Signals

More information

Single channel noise reduction in hearing aids

Single channel noise reduction in hearing aids Single channel noise reduction in hearing aids Recordings for perceptual evaluation Inge Brons Rolph Houben Wouter Dreschler Introduction Hearing impaired have difficulty understanding speech in noise

More information

Self-Assessment Scales for Pre-Fitting Testing

Self-Assessment Scales for Pre-Fitting Testing 6 Self-Assessment Scales for Pre-Fitting Testing To this point, we have discussed objective pure-tone, narrowband, and speech recognition tests that can be used during the assessment prior to the fitting

More information

EEL 6586, Project - Hearing Aids algorithms

EEL 6586, Project - Hearing Aids algorithms EEL 6586, Project - Hearing Aids algorithms 1 Yan Yang, Jiang Lu, and Ming Xue I. PROBLEM STATEMENT We studied hearing loss algorithms in this project. As the conductive hearing loss is due to sound conducting

More information

DEVELOPMENT OF NAL-NL2

DEVELOPMENT OF NAL-NL2 DEVELOPMENT OF NAL-NL2 Harvey Dillon, Gitte Keidser, Teresa Ching, Matt Flax, Scott Brewer The HEARing CRC & The National Acoustic Laboratories creating sound value value TM TM www.hearingcrc.org Prescribe

More information

ASR. ISSN / Audiol Speech Res 2017;13(3): / RESEARCH PAPER

ASR. ISSN / Audiol Speech Res 2017;13(3): /   RESEARCH PAPER ASR ISSN 1738-9399 / Audiol Speech Res 217;13(3):216-221 / https://doi.org/.21848/asr.217.13.3.216 RESEARCH PAPER Comparison of a Hearing Aid Fitting Formula Based on Korean Acoustic Characteristics and

More information

Prescribe hearing aids to:

Prescribe hearing aids to: Harvey Dillon Audiology NOW! Prescribing hearing aids for adults and children Prescribing hearing aids for adults and children Adult Measure hearing thresholds (db HL) Child Measure hearing thresholds

More information

ELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT

ELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT ELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT Astrid Haastrup, GN ReSound Mona Dworsack-Dodge, AuD, GN Otometrics Abstract With ReSound s 2.4 GHz wireless

More information

Beltone Electronics 2601 Patriot Boulevard Glenview, IL U.S.A. (800)

Beltone Electronics 2601 Patriot Boulevard Glenview, IL U.S.A. (800) Beltone Electronics 261 Patriot Boulevard Glenview, IL 626 U.S.A. (8) 621-1275 Beltone Canada 31 Supertest Road Toronto, Ontario M3J 2M4 Canada (416) 736-4444 www.beltone.com 21, Beltone Electronics 521993

More information

Make the world louder!

Make the world louder! Make the world louder! Proven ultra power amplification for extreme audibility Power and dynamic Feedback Management Sumo DM can fulfil the output needs of even your most demanding clients - even those

More information

Insio: A new standard in custom instruments

Insio: A new standard in custom instruments Insio: A new standard in custom instruments www.siemens.com Abstract: Insio custom hearing instruments introduce novel features throughout the complete fitting process in order to obtain the best possible

More information

Localization 103: Training BiCROS/CROS Wearers for Left-Right Localization

Localization 103: Training BiCROS/CROS Wearers for Left-Right Localization Localization 103: Training BiCROS/CROS Wearers for Left-Right Localization Published on June 16, 2015 Tech Topic: Localization July 2015 Hearing Review By Eric Seper, AuD, and Francis KuK, PhD While the

More information

C HAPTER T HIRTEEN. Comparison of Benefits Provided by Various Hearing Aid Technologies in Older Adults

C HAPTER T HIRTEEN. Comparison of Benefits Provided by Various Hearing Aid Technologies in Older Adults C HAPTER T HIRTEEN Comparison of Benefits Provided by Various Hearing Aid Technologies in Older Adults Larry E. Humes, Dana Kinney, Elizabeth Thompson Abstract This article presents two different sets

More information

DSM PRO. Software Training Manual. Copyright November 2003

DSM PRO. Software Training Manual. Copyright November 2003 DSM PRO Software Training Manual Copyright November 2003 DSM PRO Training Manual Rev 1 Effective November 2003 INTRODUCTION Congratulations! Your decision to incorporate the MedRx DSM PRO into your daily

More information

Preference for One or Two Hearing Aids Among Adult Patients

Preference for One or Two Hearing Aids Among Adult Patients Preference for One or Two Hearing Aids Among Adult Patients Robyn M. Cox, 1 Kathryn S. Schwartz, 1 Colleen M. Noe, 2 and Genevieve C. Alexander 2 Objectives: Most practitioners believe that use of two

More information

C HAPTER FOUR. Audiometric Configurations in Children. Andrea L. Pittman. Introduction. Methods

C HAPTER FOUR. Audiometric Configurations in Children. Andrea L. Pittman. Introduction. Methods C HAPTER FOUR Audiometric Configurations in Children Andrea L. Pittman Introduction Recent studies suggest that the amplification needs of children and adults differ due to differences in perceptual ability.

More information

Speech in Noise: Using Measurements in Fitting. Presenter: Lori Bunkholt Exclusive Networks Education and Training Starkey Hearing Technologies

Speech in Noise: Using Measurements in Fitting. Presenter: Lori Bunkholt Exclusive Networks Education and Training Starkey Hearing Technologies Speech in Noise: Using Measurements in Fitting Presenter: Lori Bunkholt Exclusive Networks Education and Training Starkey Hearing Technologies From the desk of our Attorney The views expressed in presentations

More information

The Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February :50

The Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February :50 The Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February 2009 09:50 The HF microphone as a low-tech solution for performing musicians and "ultra-audiophiles" Of the

More information

Why? Speech in Noise + Hearing Aids = Problems in Noise. Recall: Two things we must do for hearing loss: Directional Mics & Digital Noise Reduction

Why? Speech in Noise + Hearing Aids = Problems in Noise. Recall: Two things we must do for hearing loss: Directional Mics & Digital Noise Reduction Directional Mics & Digital Noise Reduction Speech in Noise + Hearing Aids = Problems in Noise Why? Ted Venema PhD Recall: Two things we must do for hearing loss: 1. Improve audibility a gain issue achieved

More information

For hearing aid noise reduction, babble is not just babble

For hearing aid noise reduction, babble is not just babble For hearing aid noise reduction, babble is not just babble HELEN CONNOR SØRENSEN * AND CHARLOTTE T. JESPERSEN Global Audiology, GN ReSound A/S, Ballerup, Denmark Most modern hearing aids provide single-microphone

More information

New CTA PSAP Quality Standard (Starting with Some Interesting Questions)

New CTA PSAP Quality Standard (Starting with Some Interesting Questions) New CTA PSAP Quality Standard (Starting with Some Interesting Questions) Mead C. Killion, PhD, ScD(hon) & Gail Gudmundsen, AuD Etymotic Research, Inc. ADA Audacity Conference San Diego 2016 11 12 1. What

More information

Interpreting Speech Results to Optimise Hearing aid Fittings

Interpreting Speech Results to Optimise Hearing aid Fittings Interpreting Speech Results to Optimise Hearing aid Fittings Josephine Marriage PhD josephine@chears.co.uk BAA Monday 18 th November 2013 How has the role of families changed with introduction of NHSP?

More information

Product Portfolio Supplement

Product Portfolio Supplement 2016 Product Portfolio Supplement I N T R O D U C I N G Muse BTE 13 The latest release in the Muse family Made for At Starkey Hearing Technologies, we ve always focused on audibility and speech understanding

More information

The historically low hearing aid adoption

The historically low hearing aid adoption Utility of Smartphone-based Hearing Aid Applications as a Substitute to Traditional Hearing Aids Study suggests that a smartphone app provides adequate temporary/starter solution BY AMYN M. AMLANI, PHD;

More information

The Situational Hearing Aid Response Profile (SHARP), version 7 BOYS TOWN NATIONAL RESEARCH HOSPITAL. 555 N. 30th St. Omaha, Nebraska 68131

The Situational Hearing Aid Response Profile (SHARP), version 7 BOYS TOWN NATIONAL RESEARCH HOSPITAL. 555 N. 30th St. Omaha, Nebraska 68131 The Situational Hearing Aid Response Profile (SHARP), version 7 BOYS TOWN NATIONAL RESEARCH HOSPITAL 555 N. 30th St. Omaha, Nebraska 68131 (402) 498-6520 This work was supported by NIH-NIDCD Grants R01

More information

Hearing Solutions Catalog Winter Supplement January 2009

Hearing Solutions Catalog Winter Supplement January 2009 Hearing Solutions Catalog Winter Supplement January 2009 LX Virtue The following information is a supplement containing information for products released January 1, 2009. It should be used along with

More information

Frequency refers to how often something happens. Period refers to the time it takes something to happen.

Frequency refers to how often something happens. Period refers to the time it takes something to happen. Lecture 2 Properties of Waves Frequency and period are distinctly different, yet related, quantities. Frequency refers to how often something happens. Period refers to the time it takes something to happen.

More information

Can you hear me now? Amanda Wolfe, Au.D. Examining speech intelligibility differences between bilateral and unilateral telephone listening conditions

Can you hear me now? Amanda Wolfe, Au.D. Examining speech intelligibility differences between bilateral and unilateral telephone listening conditions Examining speech intelligibility differences between bilateral and unilateral telephone listening conditions Amanda Wolfe, Au.D. Communicating on the telephone continues to be a point of difficulty for

More information

3M Center for Hearing Conservation

3M Center for Hearing Conservation 3M Center for Hearing Conservation Key Terms in Occupational Hearing Conservation Absorption A noise control method featuring sound-absorbing materials that are placed in an area to reduce the reflection

More information

Paediatric Amplification

Paediatric Amplification Paediatric Amplification The paediatric technical advisory group (PTAG) of the NZAS recommends following the protocols found in UNHSEIP Diagnostic and Amplification Protocols (January, 2016). These guidelines

More information

2.0. Desktop Fitting Guide getting started. Preparation of the hearing instruments

2.0. Desktop Fitting Guide getting started. Preparation of the hearing instruments Desktop Fitting Guide getting started This Guide provides you a detailed introduction to hearing instrument fitting with Phonak Target 2.0. Please also find the [News] in the Phonak Target start screen.

More information

CROS System Initial Fit Protocol

CROS System Initial Fit Protocol CROS System Initial Fit Protocol Our wireless CROS System takes audio from an ear level microphone and wirelessly transmits it to the opposite ear via Near-Field Magnetic Induction (NFMI) technology, allowing

More information

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 보청기의전기음향적성능분석 Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 차례 I. 성능분석이필요한이유 II. 성능분석에필요한장비의구성 III. 성능분석시구성품의위치 IV. 성능분석의기준 V. 성능분석의내용 VI. 성능분석의절차 I. 성능분석이필요한이유 보청기의품질관리보청기의적합 - frequency

More information

HCS 7367 Speech Perception

HCS 7367 Speech Perception Long-term spectrum of speech HCS 7367 Speech Perception Connected speech Absolute threshold Males Dr. Peter Assmann Fall 212 Females Long-term spectrum of speech Vowels Males Females 2) Absolute threshold

More information

Re/Habilitation of the Hearing Impaired. Better Hearing Philippines Inc.

Re/Habilitation of the Hearing Impaired. Better Hearing Philippines Inc. Re/Habilitation of the Hearing Impaired Better Hearing Philippines Inc. Nature of Hearing Loss Decreased Audibility Decreased Dynamic Range Decreased Frequency Resolution Decreased Temporal Resolution

More information

Four-Channel WDRC Compression with Dynamic Contrast Detection

Four-Channel WDRC Compression with Dynamic Contrast Detection Technology White Paper Four-Channel WDRC Compression with Dynamic Contrast Detection An advanced adaptive compression feature used in the Digital-One and intune DSP amplifiers from IntriCon. October 3,

More information

Desktop Fitting Guide for Phonak Brio 3

Desktop Fitting Guide for Phonak Brio 3 Phonak Target 5.3.3 Desktop Fitting Guide for Phonak Brio 3 The Phonak Target fitting software is intended to be used by qualified hearing care professionals to configure, program, and fit hearing aids

More information

THE OCCLUSION EFFECT: PHYSICAL MEASUREMENT AND SELF-REPORT. Karen Jarvis Advisor: Robyn Cox, Ph.D. April 2004

THE OCCLUSION EFFECT: PHYSICAL MEASUREMENT AND SELF-REPORT. Karen Jarvis Advisor: Robyn Cox, Ph.D. April 2004 THE OCCLUSION EFFECT: PHYSICAL MEASUREMENT AND SELF-REPORT Karen Jarvis Advisor: Robyn Cox, Ph.D. April 2004 THE OCCLUSION EFFECT Results from trapped energy in the ear canal Unoccluded, the ear canal

More information

Using digital hearing aids to visualize real-life effects of signal processing

Using digital hearing aids to visualize real-life effects of signal processing Using digital hearing aids to visualize real-life effects of signal processing By Francis Kuk, Anne Damsgaard, Maja Bulow, and Carl Ludvigsen Digital signal processing (DSP) hearing aids have made many

More information

Guidelines for determining hearing aid output, hearing aid features, and fitting parameters for children. Jace Wolfe, PhD

Guidelines for determining hearing aid output, hearing aid features, and fitting parameters for children. Jace Wolfe, PhD Guidelines for determining hearing aid output, hearing aid features, and fitting parameters for children Jace Wolfe, PhD Speech-Language Pathologists Joanna T. Smith, M.S., CCC-SLP, LSLS Cert. AVT Tamara

More information

WIDEXPRESS. no.30. Background

WIDEXPRESS. no.30. Background WIDEXPRESS no. january 12 By Marie Sonne Kristensen Petri Korhonen Using the WidexLink technology to improve speech perception Background For most hearing aid users, the primary motivation for using hearing

More information

Practical Grading System for Evaluating Cisplatin Ototoxicity in Children

Practical Grading System for Evaluating Cisplatin Ototoxicity in Children Practical Grading System for Evaluating Cisplatin Ototoxicity in Children K30 Journal Article Review Denise Nicholson, Au.D., Ph.D. candidate Senior Audiologist Dept of Audiology Article: Chang, K.W. &

More information

TEN. 1What exactly is, or was, the Independent Hearing Aid. 2So what was the. Five years later: An update on the IHAFF fitting protocol

TEN. 1What exactly is, or was, the Independent Hearing Aid. 2So what was the. Five years later: An update on the IHAFF fitting protocol P A G E TEN As WDRC hearing aids have continued to gain popularity through the 1990s, we have witnessed the introduction of several new prescriptive fitting methods to help us select the gain, output,

More information