1 Optimizing DPOAEs for Serial Monitoring: From Nitty-gritty Solutions to Source Separation Dawn Konrad-Martin, PhD, CCC-A VA National Center for Rehabilitative Auditory Research (NCRAR) Portland, Oregon VA Medical Center One of 13 National VA RR&D Research Centers of Excellence Acknowledgements NCRAR Collaborators Garnett McMillan Marilyn Dille Kelly Reavis Wendy Helt Peter Jacobs Other Collaborators Lori Dreisbach at UCSD Lynne Marshall at Naval Submarine Medical Research Laboratory Research supported by grants from VA Rehabilitation Research & Disorders Service Cody Gohean LEARNING OBJECTIVES List benefits/limitations of various DPOAE recording paradigms Estimate your own test-retest variability Describe how to optimize test-retest reliability Discuss how changes in DPOAE fine structure might impact serial monitoring results PART 1 OUTLINE Define Screening Goals Why Screen Using OAEs? DPOAE Test Types Criteria for Valid DPOAEs Criteria for Significant DPOAE Changes 1

2 DEFINE GOALS DEFINE GOALS Screening and diagnosis have DIFFERENT goals Goal of screening is to determine which patients need follow-up diagnostic testing Screening for hearing changes (option A) and subclinical damage (option B) involve different analytic strategies PROTOCOL OPTION A PROTOCOL OPTION A Option A requires understanding the operating characteristics of the screening test i.e., Sensitivity and Specificity based on a comparison of DPOAE shifts to audiogram shifts in a sample of exposed subjects A shift of X db results correctly identifies hearing shift in X % of exposed ears PROTOCOL OPTION B PROTOCOL OPTION B Option B requires comparison to normative reference limits for screening test shifts i.e., norms show no more than X db shift in DPOAE level in 9% of unexposed subjects, so shifts larger than this are a screen fail DPOAE used as proxy for OHC function (not used as an estimate of hearing shift) Photos courtesy of Dr. Marc Lenoir, from Promenade around the cochlea EDU website: by Remy Pujol et al., INSERM and University of Montpellier 2

3 DEFINE GOALS Audiogram is the current gold standard measure of ototoxicity in ototoxicity monitoring programs Audiogram is the current legal standard for noise damage in hearing conservation programs Follow-up for both screening options will involve audiometry NCRAR Ototoxicity Screening Protocol Pass Screening Audiogram Fail Responsive Screening Pass Audiogram Full <= 8 khz Audiogram (Fig. 4) DPOAE Fail Audiogram and/or DPOAE Assess Pre-Exposure Risk Exposure Pass DPIO Model Audiogram Shift Fail Option B Non- Responsive Screening Pass DPOAE Model of Audiogram <= 8 khz Shift The goal appropriate for MOST screening applications is Option B. Option A is generally reserved for patients who cannot take a behavioral screening test. DPOAE Fail WHY SCREEN WITH DPOAES? WHY SCREEN WITH DPOAES? Objective measure that tests outer hair cell (OHC) system function OHC system must be normal for hearing thresholds to be normal Initial damage from ototoxins and noise impacts OHCs DPOAE changes may signal hearing changes, or more subtle damage effects WHY SCREEN WITH DPOAES? DPOAEs were affected in more ears than conventional audiometry (CA) Noise: Lapsley Miller & Marshall, 27; Helleman & Dreschler, 212 Aminoglycosides: Katbamna et al., 1999; Stavroulaki et al., 22; Mulheran & Degg, 1997; Cisplatin: Stavroulaki et al., 21 DPOAEs were affected in more ears than CA, but in a similar number of ears compared with HFA Cisplatin: Ress et al., 1999 DPOAEs were affected in fewer ears and later than HFA Cisplatin, Carboplatin or both: Knight et al., 27 WHY SCREEN WITH DPOAES? DPOAE changes can occur earlier in DPs vs hearing threshold DPOAEs can be more reliable in certain populations Especially among the very sick or the very young Testing can be less time consuming Depending on DPOAE test type and number of frequencies tested 3

4 DPOAE TEST TYPES DPOAE TEST TYPES (1) Frequency sweep (DP-gram) Typical clinical measurement Correlated with hearing threshold Coarse- or Fine-frequency step sizes (2) Level sweep (I/O function) Slope estimates nonlinear BM growth (3) Fine frequency-ratio sweep (group delay or travel time) Delay is associated with tuning 1. DPOAE Frequency Sweep (DP-Gram) Lower levels give larger shifts f 2 Varied L 1 & L 2 Fixed f 2 /f 1 Fixed - Plots frequency on the x-axis & DPOAE level on y-axis - Course- or Fine-Frequency Step Size DPOAE level DATA FROM LORI DREISBACH Figure 1 from Marshall, Lapsley Miller, & Heller (21) Noise & Health, 3(12), 43-6 Also see review in Gorga et al. (27) 2. DPOAE Level Sweep (I/O function) Input-Output (I/O) Function f 2 Fixed L 1, L 2, or Both L 1 & L 2 Varied f 2 /f 1 Fixed -Plot L 1 (or L 2 ) on x-axis & DPOAE level on y-axis - Can determine threshold and slope of function Linear System Input Spectrum Amplitude Frequency Output Input Output Spectrum Amplitude Frequency threshold Nonlinear System Amplitude Frequency Output Input Amplitude Frequency Distortion products SLIDE FROM KIM SCHAIRER 4

5 3. DPOAE Frequency-Ratio Sweep (Group delay/travel Time) f 2 Fixed L 1 & L 2 Fixed f 2 /f 1 Varied (e.g. 1.1 to 1.3) Each measurement made has an accompanying amplitude and phase DPOAE freq on x-axis; DPOAE phase (time) on y-axis CRITERIA FOR SIGNIFICANT DPOAE CHANGE: ROC Curves (Protocol Option A) Reference Limits (Protocol Option B) Phase can be translated into time using the formula: Time (ms) = Δφ / 2π * Δf 2 Reproducibility (repeatability) of a measurement over time is paramount for serial monitoring Important to know and control for expected variance in performance to determine a real change 8 db PROTOCOL OPTION A And how well did you say this predicts the audiogram??? SCREENING TEST: Symbols CHANGED NOT CHANGED Some Definitions GOLD STANDARD: Hearing Shift CHANGED Hits (Sensitivity) False Negatives (Misses) NOT CHANGED False positives True Negatives (Specificity) % % 5

6 Receiver Operating Characteristic (ROC) curve FAKE DATA FOR ILLUSTRATIVE PURPOSES ONLY True Hearing Status Shift in Crashing Hearing No Hearing Symbols Threshold Change Change > 1 db (hearing change) (TP = 1.) (FP = 1.) < 1 db (no hearing change) (FN =.) (TN =.) Total For any clinical test, there is always a trade-off between being sensitive and specific. Total > 3 db (hearing change) 13 (TP =.87) 2 (FP =.57 ) < 3 db (no hearing change) 2 (FN =.13 ) 15 (TN =.43) Total > 6 db (hearing change) (TP =.67) 3 (FP =.9) < 6 db (no hearing change) 5 (FN =.33) 32 (TN =.91) Total > 12 db (hearing change) < 12 db (no hearing change) 5 1 (TP =.33) (FP =.3) (FN =.67) (TN =.97) Total Reavis et al, Ear & Hear 31, ears of 24 patients treated with cisplatin Mean age 58.5 years 3.4 monitoring visits on average Received ~ 4 mg of cisplatin over 42 days Half the subjects experienced ASHAsignificant hearing changes Reavis et al, Ear & Hear 31, 211 Measured hearing and DPOAEs before and during chemotherapy Used DPOAEs in a model to identify ( predict ) HL at SAME monitor visit Outcome variable was HL (yes,no) in the screening audiogram, called the SRO Predictors were DPOAE metrics alone or together w ototoxicity risk factors Reavis et al, Ear & Hear 31, 211 Reavis et al, Ear & Hear 31, 211 DP-gram search for highest DP freq f2= 1- khz; f2/f1=1.22; L1, L2=65, 59 DP I/O s at four highest frequencies tested in 1/3 rd octave steps f2=35-6 db SPL, L1 optimized using covaried paradigm (Kummer et al., 1998) Summary measures from I/O functions used as predictors 6

7 Better False Positive Rate When Risk Factors Are Included (blue) Pre-treatment Susceptibility Assessment (green) DPOAE-based Hearing Shift Assessment (black) Susceptibility + DPOAE-based Hearing Shift Assessment (red) Perfect accuracy compared with gold standard hearing test HOW WOULD DP-BASED PREDICTION MODEL WORK? At each monitoring visit, make DPOAE measurements Into the formula provided, enter DPOAE test-retest difference Enter relevant baseline patient factor information & exposure information Formula spits out likelihood that hearing has changed Formula is currently being validated PROTOCOL OPTION B PROTOCOL OPTION B Normative Reference Limits Use established norms for DPOAE changes (e.g., in level shifts) to determine if OHC function has changed or is stable Develop your own clinic norms to ensure your results are comparable to published data PROTOCOL OPTION B Standard error of measurement difference (SEM) Typically 2 X SEM = ~5 db for f 2 from 1-4 khz (Franklin et al. 1992; Beattie et al., 23) Average amplitude difference plus 2 SD 6 db for most frequencies from 1-6 khz (Roede et al., 1993) Cumulative distributions > 95% of ears had test-retest change of < 6 db for f 2 from 1 - khz (NCRAR; Reavis et al., 28) A Meta-Analysis of Reference Limits McMillan et al, in preparation 7

8 PROTOCOL OPTION B Reliability decreases with increasing time between measurements Weak emissions (i.e. near DPOAE threshold) less reliable than robust emissions Emissions at very low frequencies less reliable if averaging insufficient to achieve high SNR Emissions above 4 khz slightly less reliable than those from 2-4 khz Gorga et al., 212; Helleman & Dreschler, 212; our data in preparation HOW WOULD DP-BASED REFERNCE LIMITS WORK? DPOAE decrements or enhancements of >5-6 db indicates a screen failure DPOAE level differences between groups of noise-exposed and non-noise-exposed ears with normal hearing are often less than this, so Need to improve retest reliability! PART 2 OUTLINE BREAK TIME!!! Case Studies Improving Retest Reliability Determining Retest Reliability in YOUR Clinic DPOAE Source Generation OAE Source Generation Which One is Least Like The Others? DPOAE SFOAE Acoustic response measured in the ear canal Evoked using two-tone stimulation (f1 < f2) At least 2 sources (nonlinear distortion & coherent linear reflection) TEOAE SOAE 8

9 Noise floor Usually average amplitude in several frequency bins above and below 2f 1 -f 2 bin Noise is worse at low frequencies than high Signal-to-noise ratio (SNR) db difference between SPL at 2f 1 -f 2 and the estimated noise System distortion Some Definitions Distortion from the equipment that occurs at the 2f1-f2 frequency Greatest at high frequencies and high primary levels Determined the highest frequencies that provided a DPOAE level (Ldp) that was 6 db above the noise (Ndp) DP Gram Ldp Ndp Monitored this octave range in fine resolution F2/F1 step sizes (1/48 th ) at baseline and at each treatment. Pure-tone Thresholds Subject 19 (LE) Fine Resolution Step Size 2 2 Ldp Ndp 3 4 cumulative dose (mg) DPOAE Levels Subject 14 (RE) Subject 38 (LE) Pure-tone Thresholds Pure-tone Thresholds dose (mg) dose (mg) DPOAE Levels DPOAE Levels 9

10 What region of the cochlea do ototoxic insults damage initially? frequency DPOAEs change first What region of the cochlea does industrial noise damage initially? khz DPOAEs change first IMPROVING RETEST RELIABILITY OAEs can change in regions with & without associated hearing changes Decrements & enhancements indicate damage Inside Your Probe Speaker 2 Plane of mic mic port Speaker 1 Drawing by S. Blatrix from "promenade around the cochlea" EDU website by Rémy Pujol et al., INSERM and University Montpellier 1 IMPROVING RETEST RELIABILITY Instrumentation-related sources of variability Noise & system distortion -Subject noise - Ambient noise - Electrical noise (wires are antennae) Probe fit - Affect both noise floor and system distortion In-the-Ear Calibration Measurement parameters - Level and frequency Tester(s) - Using these tips, practice makes perfect Instrumentation Reducing noise Keep test ear away from noise sources in test environment Maintain good probe tip seal Increase averaging time Average until noise floor is the level of your system distortion or artifact-free averaging time reaches 32 or even 64 seconds Ensure pre-amp batteries have good charge Microphone cable acts as an antenna and will transmit noise when rubbed or moved keep it away from other cables/equipment Instrumentation Improving fit of probe & probe tip Firm vs loose placement Ports facing tympanic membrane vs ports blocked by canal wall Wax issues Let foam tips expand fully before test Sound delivery tubes straight Anchor the probe so it is not pulling out Cable from microphone immobile, placed where it won t wiggle or vibrate

11 What is the problem with traditional calibration techniques for measuring OAEs? Errors in stimulus levels due to standing waves in the ear canal What will the result be? Exact primary levels generating the emission are unknown in certain frequency regions Practice to Improve Tester Reliability What can I do to address problem? For serial monitoring - Ensure probe placement is consistent across tests!!! 1. Establish a conservative estimate of your system distortion a) Measure with your own OAE equipment and paradigm in a coupler and, if you can, in ears with severe-profound HL b) Use this to revise criteria for a valid response (see rules for this) 2. Determine retest reliability in YOUR clinic and use it to establish criteria for a significant emission shift (SES) Determine reliability (SEM) for your group, tester, stimulus paradigm, equipment, setting. Compare to other people's. 3. Plot your test-retest reliability data look for trends (e.g., are certain frequencies/levels giving you outliers?) 4. Validity/Sanity check using +SES and SES Control group Exposed group Slide adapted from Lynne Marshall 1a) Establish conservative estimate of your system distortion Make DPOAE measurements in a coupler using test protocol Use standard coupler (e.g. 2cc coupler meeting IEC 711 specifications, such as the B&K ear simulator Type 4157) Make of these measurements; Mean plus 2 SDs to use as a conservative estimate of system distortion Perform frequent system distortion measurements to assess system performance over time 1b) Revise Criteria For Valid DPOAEs Favorable SNR (e.g., 6 db, or db in noisy environment) OAE amplitude is larger compared to conservative estimate of YOUR system distortion Middle ear function is stable DPOAE must be valid to be reliable 2. DETERMINING RELIABILITY IN YOUR CLINIC 11

12 2. DETERMINING RELIABILITY IN YOUR CLINIC Determine Clinician Retest Error Rate: 1. Measure DPOAE in 2 subjects at T= days for each frequency/level combination in your your test paradigm 2. Repeat (re-test) the test in the same 2 subjects at T=14 days 3. Calculate the average retest difference = Retest Amplitude - Test Amplitude = T amp T14 amp < 6 db 2. DETERMINING RELIABILITY IN YOUR CLINIC For a given stimulus condition, if no more than 1 / 2 subjects has a retest error > 6 db, then we are reasonably certain that for monitoring purposes, the retest error rate will be controlled below 5% If more than 1 error out of 2 subjects, then refer to tips and tricks for making reliable measures. Repeat with a different 2 subjects 3. Plot your retest differences to check for patterns & outliers Delta for Condition A Delta for Condition B For each frequency/condition, calculate post-pre delta and plot by subject (arbitrary numbers used in example) Look for conditions with increased variability (B) And for conditions with extreme outliers (C) Examine data for measurement problems (high noise levels, bad stimulus levels, bad calibrations, transcription errors) Slide from Lynne Marshall Delta for Condition C Validity/Sanity Check Using Significant Emission Shifts Plot SES data for control group +SES vs SES should be few and randomly scattered across frequency Look for outliers & patterns Find sources of problems Were these associated with bad measurements? Plot SES data for noise-exposed group Indications of noise-induced change from Helleman & Dreschler, 212 SES rate higher than for control group More +SES than SES More SES- in 3-6 khz range than other frequencies THANK YOU FOR LISTENING Dawn.martin@va.gov 12

Optimizing DPOAEs for Serial Monitoring: From Nitty-gritty Solutions to Source Separation

Optimizing DPOAEs for Serial Monitoring: From Nitty-gritty Solutions to Source Separation Optimizing DPOAEs for Serial Monitoring: From Nitty-gritty Solutions to Source Separation Dawn Konrad-Martin, PhD, CCC-A VA National Center for Rehabilitative Auditory Research (NCRAR) Portland, Oregon

More information

Otoacoustic Emissions As A Test Of Noise-Induced Hearing Loss. Brenda L Lonsbury-Martin PhD

Otoacoustic Emissions As A Test Of Noise-Induced Hearing Loss. Brenda L Lonsbury-Martin PhD Otoacoustic Emissions As A Test Of Noise-Induced Hearing Loss Brenda L Lonsbury-Martin PhD Department of Otolaryngology--Head & Neck Surgery Loma Linda University Medical Center blonsbury-martin@llu.edu

More information

The clinical Link. Distortion Product Otoacoustic Emission

The clinical Link. Distortion Product Otoacoustic Emission Distortion product otoacoustic emissions: Introduction Michael P. Gorga, Ph.D. Boys Town National Research Hospital Work supported by the NIH Collaborators on BTNRH OAE Projects Stephen Neely Kathy Beauchaine

More information

FALSE POSITIVE DP GRAMS

FALSE POSITIVE DP GRAMS FALSE POSITIVE DP GRAMS Prof.Dr / Mona mourad Dr / Samir Asal Introduction Distortion-Product Otoacoustic Emissions (DPOAEs) are acoustic energy in the ear canal arising from the non- linear interaction

More information

Clinical applications of otoacoustic emissions in Industry. Prof. Dr. B. Vinck, MSc, PhD University of Ghent, Belgium

Clinical applications of otoacoustic emissions in Industry. Prof. Dr. B. Vinck, MSc, PhD University of Ghent, Belgium Clinical applications of otoacoustic emissions in Industry Prof. Dr. B. Vinck, MSc, PhD University of Ghent, Belgium The 25th anniversary of the discovery of otoacoustic emissions (OAEs) [sounds that can

More information

Emissions and Hearing Thresholds After a Six-Month Deployment on an Aircraft Carrier

Emissions and Hearing Thresholds After a Six-Month Deployment on an Aircraft Carrier #772 Changes in Evoked Otoacoustic Emissions and Hearing Thresholds After a Six-Month Deployment on an Aircraft Carrier Lynne Marshall, Judi A. Lapsley Miller, Laurie M. Heller, Linda M. Hughes, and Linda

More information

OtoAcoustic Emissions (OAE s)

OtoAcoustic Emissions (OAE s) OtoAcoustic Emissions (OAE s) Phenomenon and applications in audiological diagnostics Measurement procedures TEOAE and DPOAE Physiological backgound, functional models Acknowledgment: several illustrations

More information

Improving the diagnostic power of otoacoustic emissions. Arturo Moleti Physics Department University of Roma Tor Vergata

Improving the diagnostic power of otoacoustic emissions. Arturo Moleti Physics Department University of Roma Tor Vergata Improving the diagnostic power of otoacoustic emissions Arturo Moleti Physics Department University of Roma Tor Vergata The human ear Ear canal: resonant cavity Middle ear: impedance adapter and pressure

More information

Ototoxicity risk assessment combining distortion product otoacoustic emissions with a cisplatin dose model a)

Ototoxicity risk assessment combining distortion product otoacoustic emissions with a cisplatin dose model a) Ototoxicity risk assessment combining distortion product otoacoustic emissions with a cisplatin dose model a) Marilyn F. Dille b and Garnett P. McMillan National Center for Rehabilitative Auditory Research,

More information

Advanced otoacoustic emission detection techniques and clinical diagnostics applications

Advanced otoacoustic emission detection techniques and clinical diagnostics applications Advanced otoacoustic emission detection techniques and clinical diagnostics applications Arturo Moleti Physics Department, University of Roma Tor Vergata, Roma, ITALY Towards objective diagnostics of human

More information

A Review of the Effectiveness of Otoacoustic Emissions for Evaluating Hearing Status After Newborn Screening

A Review of the Effectiveness of Otoacoustic Emissions for Evaluating Hearing Status After Newborn Screening Otology & Neurotology 34:1058Y1063 Ó 2013, Otology & Neurotology, Inc. A Review of the Effectiveness of Otoacoustic Emissions for Evaluating Hearing Status After Newborn Screening Thomas Janssen ENT-Department,

More information

NSMRL TECHNICAL REPORT #TR1231 February 26, 2004

NSMRL TECHNICAL REPORT #TR1231 February 26, 2004 Naval Submarine Medical Research Laboratory NSMRL TECHNICAL REPORT #TR131 February 6, 4 STIMULUS-FREQUENCY OTOACOUSTIC EMISSIONS VALIDITY & RELIABILITY OF SFOAES IMPLEMENTED ON MIMOSA ACOUSTICS SFOAE MEASUREMENT

More information

Trajectory of the Aging Cochlea

Trajectory of the Aging Cochlea Trajectory of the Aging Cochlea Sumitrajit (Sumit) Dhar Professor & Chair Roxelyn & Richard Pepper Department of Communication Sciences and Disorders Fellow, Hugh Knowles Center for Hearing Science Northwestern

More information

AUDL GS08 and GAV1: 2013 Final exam page 1/13. You must complete all sections. Label all graphs. Show your work!

AUDL GS08 and GAV1: 2013 Final exam page 1/13. You must complete all sections. Label all graphs. Show your work! AUDL GS08 and GAV1: 2013 Final exam page 1/13 You must complete all sections. Label all graphs. Show your work! Section A: Short questions concerning Signals & Systems A1. Give the sound pressure levels

More information

mously with greater accuracy for mid and high frequencies, compared to the accuracy that is achieved at lower frequencies.

mously with greater accuracy for mid and high frequencies, compared to the accuracy that is achieved at lower frequencies. Further efforts to predict pure-tone thresholds from distortion product otoacoustic emission inputõoutput functions Michael P. Gorga, a) Stephen T. Neely, Patricia A. Dorn, and Brenda M. Hoover Boys Town

More information

Technical Report: Distortion Product Otoacoustic Emissions That Are Not Outer Hair Cell Emissions DOI: /jaaa

Technical Report: Distortion Product Otoacoustic Emissions That Are Not Outer Hair Cell Emissions DOI: /jaaa J Am Acad Audiol 20:306 310 (2009) Technical Report: Distortion Product Otoacoustic Emissions That Are Not Outer Hair Cell Emissions DOI: 10.3766/jaaa.20.5.3 Shlomo Silman*{{ Michele B. Emmer*{ Carol A.

More information

OAE Test System. Screener PLUS. Diagnostic PLUS. with 4 frequency DPOAE testing Protocols

OAE Test System. Screener PLUS. Diagnostic PLUS. with 4 frequency DPOAE testing Protocols Screener PLUS with 4 frequency DPOAE testing Protocols Diagnostic PLUS with 4, 6 and 12 frequency DPOAE testing Protocols *TEOAE upgrade OAE Test System Physicians Otoacoustic emissions testing is an ideal

More information

Assessing Sensorineural Hearing Loss Using Various Transient-Evoked Otoacoustic Emission Stimulus Conditions

Assessing Sensorineural Hearing Loss Using Various Transient-Evoked Otoacoustic Emission Stimulus Conditions Assessing Sensorineural Hearing Loss Using Various Transient-Evoked Otoacoustic Emission Stimulus Conditions Daniel B. Putterman, 1,2 Douglas H. Keefe, 3 Lisa L. Hunter, 4 Angela C. Garinis, 1,2 Denis

More information

I. INTRODUCTION. J. Acoust. Soc. Am. 111 (1), Pt. 1, Jan /2002/111(1)/271/14/$ Acoustical Society of America

I. INTRODUCTION. J. Acoust. Soc. Am. 111 (1), Pt. 1, Jan /2002/111(1)/271/14/$ Acoustical Society of America The use of distortion product otoacoustic emission suppression as an estimate of response growth Michael P. Gorga, a) Stephen T. Neely, Patricia A. Dorn, and Dawn Konrad-Martin Boys Town National Research

More information

Emissions are low-intensity sounds that may be detected in the external ear canal by a microphone

Emissions are low-intensity sounds that may be detected in the external ear canal by a microphone OAEs Emissions are low-intensity sounds that may be detected in the external ear canal by a microphone OAE is a pre-neural phenomenon They can be measured even when the 8 th cranial nerve is severely damaged

More information

Handheld OAE-Tympanometry Combination System

Handheld OAE-Tympanometry Combination System Handheld OAE-Tympanometry Combination System The ONLY handheld OAE-TYMP combination testing device in the world! www.maico-diagnostics.com Visual Evaluation Ear Canal Middle Ear Evaluation Eardrum Cochlea

More information

Role of Distortion Product Otoacoustic Emissions (DPOAES) In Detecting Early Hearing Impairment in Individuals With Normal Pure Tone Audiometry (PTA)

Role of Distortion Product Otoacoustic Emissions (DPOAES) In Detecting Early Hearing Impairment in Individuals With Normal Pure Tone Audiometry (PTA) ORIGINAL ARTICLE Role of Distortion Product Otoacoustic Emissions (DPOAES) In Detecting Early Hearing Impairment in Individuals With Normal Pure Tone Audiometry (PTA) Anil Suri, Divaya Gupta, Deepak Kotwal,

More information

General about Calibration and Service on Audiometers and Impedance Instruments

General about Calibration and Service on Audiometers and Impedance Instruments General about Calibration and Service on Audiometers and Impedance Instruments Calibration and Service of Audiometers - page 1 of 14 Rev.4 2008-05-21/JF General about Calibration and Service on Audiometers

More information

ERO SCAN. OAE Test System. Screener. Diagnostic. with 4 frequency DPOAE testing Protocols

ERO SCAN. OAE Test System. Screener. Diagnostic. with 4 frequency DPOAE testing Protocols ERO SCAN Screener with 4 frequency DPOAE testing Protocols Diagnostic with 4, 6 and 12 frequency DPOAE testing Protocols *TEOAE upgrade OAE Test System ERO SCAN - OAE Test System Physicians Otoacoustic

More information

A Guide to. Otoacoustic Emissions (OAEs) for Physicians.

A Guide to. Otoacoustic Emissions (OAEs) for Physicians. A Guide to Otoacoustic Emissions (OAEs) for Physicians www.maico-diagnostics.com Introduction Hearing loss is not uncommon in children and adults. According to recent estimates, 37.5 million people in

More information

Annelies Bockstael, Hannah Keppler, and Dick Botteldooren

Annelies Bockstael, Hannah Keppler, and Dick Botteldooren Improved hearing conservation in industry: More efficient implementation of distortion product otoacoustic emissions for accurate hearing status monitoring Annelies Bockstael, Hannah Keppler, and Dick

More information

Hearing Evaluation: Diagnostic Approach

Hearing Evaluation: Diagnostic Approach Hearing Evaluation: Diagnostic Approach Hearing Assessment Purpose - to quantify and qualify in terms of the degree of hearing loss, the type of hearing loss and configuration of the hearing loss - carried

More information

Acoustics, signals & systems for audiology. Psychoacoustics of hearing impairment

Acoustics, signals & systems for audiology. Psychoacoustics of hearing impairment Acoustics, signals & systems for audiology Psychoacoustics of hearing impairment Three main types of hearing impairment Conductive Sound is not properly transmitted from the outer to the inner ear Sensorineural

More information

Prescribe hearing aids to:

Prescribe hearing aids to: Harvey Dillon Audiology NOW! Prescribing hearing aids for adults and children Prescribing hearing aids for adults and children Adult Measure hearing thresholds (db HL) Child Measure hearing thresholds

More information

Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA

Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA 2017-10 Presenter Peter Wulf-Andersen Engineering degree in Acoustics Co-founder

More information

Neuro-Audio Version 2010

Neuro-Audio Version 2010 ABR PTA ASSR Multi-ASSR OAE TEOAE DPOAE SOAE ECochG MLR P300 Neuro-Audio Version 2010 one device for all audiological tests Auditory brainstem response (ABR)/Brainstem evoked response audiometry (BERA)

More information

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE Copyright SFA - InterNoise 2000 1 inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering 27-30 August 2000, Nice, FRANCE I-INCE Classification: 6.2 THE INFLUENCE OF

More information

Utility of Standard DPOAEs in the Evaluation of the Normal-Hearing Tinnitus Patient

Utility of Standard DPOAEs in the Evaluation of the Normal-Hearing Tinnitus Patient Utility of Standard DPOAEs in the Evaluation of the Normal-Hearing Tinnitus Patient Background Shiomi et al, 1997 (Journal of the Association of Research In Otology): In comparison to normal-hearing and

More information

The following describes general use of ELIOS device

The following describes general use of ELIOS device General use of the device The following describes general use of ELIOS device 1. Introduction of graphical user interface Presence indicator of SD card Charging indicator General Informations Screen title

More information

Jacob Sulkers M.Cl.Sc (AUD) Candidate University of Western Ontario: School of Communication Sciences and Disorders

Jacob Sulkers M.Cl.Sc (AUD) Candidate University of Western Ontario: School of Communication Sciences and Disorders Critical Review: The (Sound) Wave of the Future: Is Forward Pressure Level More Accurate than Sound Pressure Level in Defining In Situ Sound Levels for Hearing Aid Fitting? Jacob Sulkers M.Cl.Sc (AUD)

More information

More robust estimates for DPOAE level at audiometric frequencies

More robust estimates for DPOAE level at audiometric frequencies Psychological and Physiological Acoustics (others): Paper ICA216-88 More robust estimates for DPOAE level at audiometric frequencies Dorte Hammershøi (a), Rodrigo Ordoñez (b), Anders Tornvig Christensen

More information

FM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC.

FM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC. T E S T I N G FM SYSTEMS with the FONIX 6500-CX Hearing Aid Analyzer (Requires software version 4.20 or above) FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 (503) 620-2722 (800) 547-8209

More information

INTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs.

INTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs. INTRODUCTION TO PURE TONE AUDIOMETERY (AUDIOMETER & TESTING ENVIRONMENT) By Mrs. Wedad Alhudaib with many thanks to Mrs. Tahani Alothman Topics : This lecture will incorporate both theoretical information

More information

OAE Test System. by Maico Diagnostics ERO SCAN. EroScan Pictured. OAE Hearing Screener DPOAE/TEOAE Test Systems.

OAE Test System. by Maico Diagnostics ERO SCAN. EroScan Pictured. OAE Hearing Screener DPOAE/TEOAE Test Systems. OAE Test System by Maico Diagnostics ERO SCAN EroScan Pictured OAE Hearing Screener DPOAE/TEOAE Test Systems www.maico-diagnostics.com Ero Scan - OAE Test System Visual Evaluation Middle Ear Evaluation

More information

Hearing Sound. The Human Auditory System. The Outer Ear. Music 170: The Ear

Hearing Sound. The Human Auditory System. The Outer Ear. Music 170: The Ear Hearing Sound Music 170: The Ear Tamara Smyth, trsmyth@ucsd.edu Department of Music, University of California, San Diego (UCSD) November 17, 2016 Sound interpretation in the auditory system is done by

More information

Music 170: The Ear. Tamara Smyth, Department of Music, University of California, San Diego (UCSD) November 17, 2016

Music 170: The Ear. Tamara Smyth, Department of Music, University of California, San Diego (UCSD) November 17, 2016 Music 170: The Ear Tamara Smyth, trsmyth@ucsd.edu Department of Music, University of California, San Diego (UCSD) November 17, 2016 1 Hearing Sound Sound interpretation in the auditory system is done by

More information

A Guide to. Otoacoustic Emissions (OAEs) for Otolaryngologists.

A Guide to. Otoacoustic Emissions (OAEs) for Otolaryngologists. A Guide to Otoacoustic Emissions (OAEs) for Otolaryngologists www.maico-diagnostics.com Introduction Hearing loss is not uncommon in children and adults. According to recent estimates, 37.5 million people

More information

ABR PTA ASSR Multi-ASSR OAE TEOAE DPOAE SOAE VEMP ECochG MLR P300

ABR PTA ASSR Multi-ASSR OAE TEOAE DPOAE SOAE VEMP ECochG MLR P300 ABR PTA ASSR Multi-ASSR OAE TEOAE DPOAE SOAE VEMP ECochG MLR P300 Neuro-Audio one device for all audiological tests Auditory brainstem response (ABR)/Brainstem evoked response audiometry (BERA) (air and

More information

Comparison of Distortion Product Otoacoustic Emission Amplitude between Normal Hearing Male and Female Subjects

Comparison of Distortion Product Otoacoustic Emission Amplitude between Normal Hearing Male and Female Subjects Comparison of Distortion Product Otoacoustic Emission Amplitude between Normal Hearing Male and Female Subjects Shawkt Al-Tamimi MD*, Mefleh Alsarhan MD*, Hussein Alqasem PhD*, Mohammed Rawashdeh MD*,

More information

2) What is the most robust distortion product evoked with DPOAEs? A. F1 B. F2 C. 2F1-F2 D. 2F2 F1

2) What is the most robust distortion product evoked with DPOAEs? A. F1 B. F2 C. 2F1-F2 D. 2F2 F1 Instructions: Look at the following questions below in their respective categories, and check your answers against the Answer Key on the last page. Contact the SAA Programs Subcommittee Chair with any

More information

Estimating auditory filter bandwidth using distortion product otoacoustic emissions

Estimating auditory filter bandwidth using distortion product otoacoustic emissions Downloaded from vbn.aau.dk on: januar 27, 219 Aalborg Universitet Estimating auditory filter bandwidth using distortion product otoacoustic emissions Rukjær, Andreas Harbo; Hauen, Sigurd van; Ordoñez Pizarro,

More information

Chapter 3. Sounds, Signals, and Studio Acoustics

Chapter 3. Sounds, Signals, and Studio Acoustics Chapter 3 Sounds, Signals, and Studio Acoustics Sound Waves Compression/Rarefaction: speaker cone Sound travels 1130 feet per second Sound waves hit receiver Sound waves tend to spread out as they travel

More information

HCS 7367 Speech Perception

HCS 7367 Speech Perception Long-term spectrum of speech HCS 7367 Speech Perception Connected speech Absolute threshold Males Dr. Peter Assmann Fall 212 Females Long-term spectrum of speech Vowels Males Females 2) Absolute threshold

More information

Low-level otoacoustic emissions may predict susceptibility to noise-induced hearing loss a)

Low-level otoacoustic emissions may predict susceptibility to noise-induced hearing loss a) Low-level otoacoustic emissions may predict susceptibility to noise-induced hearing loss a) Judi A. Lapsley Miller, b Lynne Marshall, Laurie M. Heller, c and Linda M. Hughes Naval Submarine Medical Research

More information

ERO SCAN. Otoacoustic Emission Testing

ERO SCAN. Otoacoustic Emission Testing ERO SCAN Otoacoustic Emission Testing ERO SCAN OAE Testing for all ages Newborns School Children Toddlers Adults ERO SCAN Screening Version The ERO SCAN with screening function is the smart choice for

More information

An efficient test protocol for identification of a limited, sensitive frequency test range for early detection of ototoxicity

An efficient test protocol for identification of a limited, sensitive frequency test range for early detection of ototoxicity Journal of Rehabilitation Research and Development Vol. 39 No. 5, September/October 2002 Pages 567 574 An efficient test protocol for identification of a limited, sensitive frequency test range for early

More information

Can components in distortion-product otoacoustic emissions be separated?

Can components in distortion-product otoacoustic emissions be separated? Can components in distortion-product otoacoustic emissions be separated? Anders Tornvig Section of Acoustics, Aalborg University, Fredrik Bajers Vej 7 B5, DK-922 Aalborg Ø, Denmark, tornvig@es.aau.dk David

More information

ECHOLAB. Screening and Clinical Modality. User Programmable Protocols. Otoacoustic Emissions System TEOAE - DPOAE - ABR - AABR

ECHOLAB. Screening and Clinical Modality. User Programmable Protocols. Otoacoustic Emissions System TEOAE - DPOAE - ABR - AABR ECHOLAB Otoacoustic Emissions System - DPOAE - ABR - AABR Screening and Clinical Modality User Programmable Protocols ECHOLAB Otoacoustic Emissions System - DPOAE - ABR - AABR All of Labat s technology

More information

Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial

Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial Enhancing the Initial Hearing Aid Fitting Michael Block, Ph.D. Session 0228 Importance of a Good Start The consequences of doing a poor job of selling the hearing-impaired client on a positive treatment

More information

Basic Audiogram Interpretation

Basic Audiogram Interpretation Basic Audiogram Interpretation Audiogram - graph showing Frequency on Horizontal axis db Hearing Level on Vertical axis db level increases as we move down on graph Audiogram displays the individuals pure

More information

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 보청기의전기음향적성능분석 Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 차례 I. 성능분석이필요한이유 II. 성능분석에필요한장비의구성 III. 성능분석시구성품의위치 IV. 성능분석의기준 V. 성능분석의내용 VI. 성능분석의절차 I. 성능분석이필요한이유 보청기의품질관리보청기의적합 - frequency

More information

Proposed Recommended Procedure for the Use of OAEs in Hearing Conservation: a Delphi Exercise

Proposed Recommended Procedure for the Use of OAEs in Hearing Conservation: a Delphi Exercise Proposed Recommended Procedure for the Use of OAEs in Hearing Conservation: a Delphi Exercise ENT/SAAA/SASLHA conference Bloemfontein 30 September 2013 Acknowledgments Alison Codling-Health and Safety

More information

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 122 Annex C Edition 1999 (E) Equipment for speech audiometry Annex C Test report format Appareils pour l audiométrie vocale Annexe C Format du rapport d essai OIML R

More information

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04 AURICAL Plus with DSL v. 5.0b Quick Guide 0459 Doc no. 7-50-0900/04 Copyright notice No part of this Manual or program may be reproduced, stored in a retrieval system, or transmitted, in any form or by

More information

Issues faced by people with a Sensorineural Hearing Loss

Issues faced by people with a Sensorineural Hearing Loss Issues faced by people with a Sensorineural Hearing Loss Issues faced by people with a Sensorineural Hearing Loss 1. Decreased Audibility 2. Decreased Dynamic Range 3. Decreased Frequency Resolution 4.

More information

SUBJECT: Physics TEACHER: Mr. S. Campbell DATE: 15/1/2017 GRADE: DURATION: 1 wk GENERAL TOPIC: The Physics Of Hearing

SUBJECT: Physics TEACHER: Mr. S. Campbell DATE: 15/1/2017 GRADE: DURATION: 1 wk GENERAL TOPIC: The Physics Of Hearing SUBJECT: Physics TEACHER: Mr. S. Campbell DATE: 15/1/2017 GRADE: 12-13 DURATION: 1 wk GENERAL TOPIC: The Physics Of Hearing The Physics Of Hearing On completion of this section, you should be able to:

More information

Ear Exam and Hearing Tests

Ear Exam and Hearing Tests Ear Exam and Hearing Tests Test Overview A thorough evaluation of a person's hearing requires an ear exam and hearing tests. In children, normal hearing is important for language to develop correctly.

More information

Educational Module Tympanometry. Germany D Germering

Educational Module Tympanometry. Germany D Germering Educational Module anometry PATH medical Germany D-82110 Germering Our educational modules 1 are made for providing information on how the hearing organ works and which test procedures are used to test

More information

Signals, systems, acoustics and the ear. Week 1. Laboratory session: Measuring thresholds

Signals, systems, acoustics and the ear. Week 1. Laboratory session: Measuring thresholds Signals, systems, acoustics and the ear Week 1 Laboratory session: Measuring thresholds What s the most commonly used piece of electronic equipment in the audiological clinic? The Audiometer And what is

More information

Introduction. IAPA: June 04 1

Introduction. IAPA: June 04 1 Introduction Conflicting views on the prevalence and nature of otoacoustic emission [OAE] abnormalities in ARNSHL families (Morell et al, 1998; Cohn & Kelley, 1999). Detailed study of OAEs in greater number

More information

Fine structures in hearing thresholds and distortion product otoacoustic emissions

Fine structures in hearing thresholds and distortion product otoacoustic emissions Proceedings oth International Congress on Acoustics, ICA 21 23 27 August 21, Sydney, Australia Fine structures in hearing thresholds and distortion product otoacoustic emissions Dorte Hammershøi, Rodrigo

More information

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system The information in this document contains general descriptions of the technical options available, which do not always have to be present in individual cases and are subject to change without prior notice.

More information

Comparison of Screening Otoacoustic Emission Devices for Pre-school Aged Children

Comparison of Screening Otoacoustic Emission Devices for Pre-school Aged Children Illinois State University ISU ReD: Research and edata AuD Capstone Projects - Communication Sciences and Disorders Communication Sciences and Disorders Spring 5-11-2017 Comparison of Screening Otoacoustic

More information

The Effect of Analysis Methods and Input Signal Characteristics on Hearing Aid Measurements

The Effect of Analysis Methods and Input Signal Characteristics on Hearing Aid Measurements The Effect of Analysis Methods and Input Signal Characteristics on Hearing Aid Measurements By: Kristina Frye Section 1: Common Source Types FONIX analyzers contain two main signal types: Puretone and

More information

3M Center for Hearing Conservation

3M Center for Hearing Conservation 3M Center for Hearing Conservation Key Terms in Occupational Hearing Conservation Absorption A noise control method featuring sound-absorbing materials that are placed in an area to reduce the reflection

More information

Clare Forshaw Corporate Medical Unit, Health & Safety Executive. Bootle

Clare Forshaw Corporate Medical Unit, Health & Safety Executive. Bootle Report of an International Expert Symposium on the usefulness of Otoacoustic Emissions (OAE) Testing in Occupational Health Surveillance 8-9 th February 2011 Manchester, UK Clare Forshaw Corporate Medical

More information

Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids

Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids Acoustical Measurements and Instrumentation: Paper ICA2016-388 Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids Zargos Neves Masson (a),

More information

Wheeler, K.S. M.Cl.Sc. (Aud) Candidate School of Communication Sciences and Disorders, U.W.O

Wheeler, K.S. M.Cl.Sc. (Aud) Candidate School of Communication Sciences and Disorders, U.W.O Copyright 2007 by Wheeler, K.S. Critical Review: Is there evidence that auditory steady-state response measures provide a more accurate electrophysiological estimate of behavioural thresholds in infants

More information

Speaker s Notes: AB is dedicated to helping people with hearing loss hear their best. Partnering with Phonak has allowed AB to offer unique

Speaker s Notes: AB is dedicated to helping people with hearing loss hear their best. Partnering with Phonak has allowed AB to offer unique 1 General Slide 2 Speaker s Notes: AB is dedicated to helping people with hearing loss hear their best. Partnering with Phonak has allowed AB to offer unique technological advances to help people with

More information

Chapter 6: Discussion and Interpretation of Results

Chapter 6: Discussion and Interpretation of Results Chapter 6: Discussion and Interpretation of Results The development of objective procedures in audiology came a long way since the 1920s. With the aid of modern technology, audiologists can measure the

More information

UPDATE ON ELECTROPHYSIOLOGIC MEASURES OF HEARING SENSITIVITY IN INFANTS

UPDATE ON ELECTROPHYSIOLOGIC MEASURES OF HEARING SENSITIVITY IN INFANTS UPDATE ON ELECTROPHYSIOLOGIC MEASURES OF HEARING SENSITIVITY IN INFANTS Why do we need frequency-specific stimuli? Percent of Audiograms 50 Pittman & Stelmachowicz (2002) 40 30 20 10 Adults Children 0

More information

MEASUREMENTS AND EQUIPMENT FOR AUDIOLOGICAL EVALUATIONS

MEASUREMENTS AND EQUIPMENT FOR AUDIOLOGICAL EVALUATIONS MEASUREMENTS AND EQUIPMENT FOR AUDIOLOGICAL EVALUATIONS Audiometric results will be analysed according to the following guidelines indicated in literature. 1. Otoscopic Examination The different structures

More information

Terri-Lynn Gagnon M.Cl.Sc (AUD) Candidate University of Western Ontario: School of Communication Sciences and Disorders

Terri-Lynn Gagnon M.Cl.Sc (AUD) Candidate University of Western Ontario: School of Communication Sciences and Disorders Critical Review: Are Otoacoustic Emissions Effective for Characterizing Subclinical Auditory Impairment in Normal Hearing Individuals with Type I Diabetes Mellitus? Terri-Lynn Gagnon M.Cl.Sc (AUD) Candidate

More information

by Maico Diagnostics

by Maico Diagnostics Diagnostic OAE-Tympanometry Combination System by Maico Diagnostics ERO SCAN Pro 3 tests with one probe fit Electronic medical records (EMR), ready Link results to patient's electronic chart with EroScan

More information

Testing Digital Hearing Aids

Testing Digital Hearing Aids Testing Digital Hearing Aids with the FONIX 6500-CX Hearing Aid Analyzer Frye Electronics, Inc. Introduction The following is a quick guide for testing digital hearing aids using the FONIX 6500-CX. All

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Equipment for speech audiometry Appareils pour l audiométrie vocale OIML R 122 Edition 1996 (E) CONTENTS Foreword... 3

More information

Swept - Tone Evoked Otoacoustic Emissions: Stimulus Calibration and Equalization

Swept - Tone Evoked Otoacoustic Emissions: Stimulus Calibration and Equalization University of Miami Scholarly Repository Open Access Theses Electronic Theses and Dissertations 2011-12-19 Swept - Tone Evoked Otoacoustic Emissions: Stimulus Calibration and Equalization Todor Mihajloski

More information

Two Modified IEC Ear Simulators for Extended Dynamic Range

Two Modified IEC Ear Simulators for Extended Dynamic Range Two Modified IEC 60318-4 Ear Simulators for Extended Dynamic Range Peter Wulf-Andersen & Morten Wille The international standard IEC 60318-4 specifies an occluded ear simulator, often referred to as a

More information

Hearing Screening, Diagnostics and Intervention

Hearing Screening, Diagnostics and Intervention JCIH Newborn Hearing Screening Guidelines 1-3-6 Model By 1 month Screenhearing Hearing Screening, Diagnostics and Intervention By 3 months: Evaluate hearing and complete diagnosticaudiology and otolaryngology

More information

Reliable, flexible & precise

Reliable, flexible & precise OtoRead Portable OAE Reliable, flexible & precise l e a d i n g d i a g n o s t i c s o l u t i o n s OtoRead Portable OAE Reliable, flexible & precise The OtoRead is a fast, automatic handheld OAE instrument

More information

Testing FM Systems with FONIX FP40 Analyzers

Testing FM Systems with FONIX FP40 Analyzers Testing FM Systems with FONIX FP40 Analyzers FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 USA (503) 620-2722 (800) 547-8209 Fax: (503) 639-0128 http://www.frye.com e-mail: general@frye.com

More information

functions grow at a higher rate than in normal{hearing subjects. In this chapter, the correlation

functions grow at a higher rate than in normal{hearing subjects. In this chapter, the correlation Chapter Categorical loudness scaling in hearing{impaired listeners Abstract Most sensorineural hearing{impaired subjects show the recruitment phenomenon, i.e., loudness functions grow at a higher rate

More information

Testing FM Systems on the 7000 Hearing Aid Test System

Testing FM Systems on the 7000 Hearing Aid Test System Testing FM Systems on the 7000 Hearing Aid Test System Introduction Testing FM Systems on the 7000 Hearing Aid Test System This workbook describes how to test FM systems with the 7000 Hearing Aid Test

More information

SOLUTIONS Homework #3. Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03

SOLUTIONS Homework #3. Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03 SOLUTIONS Homework #3 Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03 Problem 1: a) Where in the cochlea would you say the process of "fourier decomposition" of the incoming

More information

Simultaneous Measurement of Noise-Activated Middle-Ear Muscle Reflex and Stimulus Frequency Otoacoustic Emissions

Simultaneous Measurement of Noise-Activated Middle-Ear Muscle Reflex and Stimulus Frequency Otoacoustic Emissions : JARO 7: 125 139 (2006) DOI: 10.1007/s10162-006-0028-9 JARO Journal of the Association for Research in Otolaryngology Simultaneous Measurement of Noise-Activated Middle-Ear Muscle Reflex and Stimulus

More information

Newborn Screening and Middle Ear Problems

Newborn Screening and Middle Ear Problems Lisa L. Hunter 1, Douglas H. Keefe 2, M. Patrick Feeney 3;4, David K. Brown 1, Jareen Meinzen-Derr and Denis F. Fitzpatrick 2 1 Cincinnati Children's Hospital Medical Center, 2 Boys Town National Research

More information

Otoacoustic Emissions

Otoacoustic Emissions Understanding and Using Otoacoustic Emissions by David T. Kemp The incredible turned out to be true! Dr. David T. Kemp Professor of Auditory Biophysics University College London Otoacoustic emission is

More information

Signals, systems, acoustics and the ear. Week 5. The peripheral auditory system: The ear as a signal processor

Signals, systems, acoustics and the ear. Week 5. The peripheral auditory system: The ear as a signal processor Signals, systems, acoustics and the ear Week 5 The peripheral auditory system: The ear as a signal processor Think of this set of organs 2 as a collection of systems, transforming sounds to be sent to

More information

QUICK START GUIDE SENTIERO OAE, TYMP AND AUDIOMETRY SCREENING ( and )

QUICK START GUIDE SENTIERO OAE, TYMP AND AUDIOMETRY SCREENING ( and ) QUICK START GUIDE SENTIERO OAE, TYMP AND AUDIOMETRY SCREENING (8-04-15907 and 8-04-15908) DEVICE CONNECTORS On/off button Headphones Power supply Air pump USB cable START MENU CONFIRM PRESET Label printer

More information

Mechanical Properties of the Cochlea. Reading: Yost Ch. 7

Mechanical Properties of the Cochlea. Reading: Yost Ch. 7 Mechanical Properties of the Cochlea CF Reading: Yost Ch. 7 The Cochlea Inner ear contains auditory and vestibular sensory organs. Cochlea is a coiled tri-partite tube about 35 mm long. Basilar membrane,

More information

(OAEs) for. Steven D. Smith, Au.D.

(OAEs) for. Steven D. Smith, Au.D. A Guide to Otoacoustic Emissions (OAEs) for School Nurses Steven D. Smith, Au.D. Director of Audiology, Director of Physicians Hearing & Balance Center Drs. Kitchens, Chapman, & Anderson, PA, Montgomery,

More information

A Sound Foundation Through Early Amplification

A Sound Foundation Through Early Amplification 3 A Sound Foundation Through Early Amplification Proceedings of the 7th International Conference 2016 The strategic application of otoacoustic emissions to infants and children Carolina Abdala, Ph.D. Abstract

More information

! Can hear whistle? ! Where are we on course map? ! What we did in lab last week. ! Psychoacoustics

! Can hear whistle? ! Where are we on course map? ! What we did in lab last week. ! Psychoacoustics 2/14/18 Can hear whistle? Lecture 5 Psychoacoustics Based on slides 2009--2018 DeHon, Koditschek Additional Material 2014 Farmer 1 2 There are sounds we cannot hear Depends on frequency Where are we on

More information