FM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC.

Size: px
Start display at page:

Download "FM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC."

Transcription

1 T E S T I N G FM SYSTEMS with the FONIX 6500-CX Hearing Aid Analyzer (Requires software version 4.20 or above) FRYE FRYE ELECTRONICS, INC. P.O. Box Tigard, OR (503) (800) Fax: (503) sales@frye.com

2 Contents Introduction General Fitting Protocol Curve Labeling System Arrangement of This Workbook Section 1: Coupler Measurements with FM Systems 1.1 Coupler Setup in the sound chamber How to Attach the 2-cc Coupler Hearing Aid Setup FM System Environmental Microphone Setup Omni-directional: FM-only Test aid alone with typical input EHA Test FM microphone with typical input EFM Test aid alone with high-level input EHA Test FM system with high-level input EFM Omni-directional: FM+ENV Single frequency output of omni-directional hearing aid in the sound chamber Single frequency output of omni-directional FM mic at 65 db SPL Verify increase in signal level Test aid alone with typical input EHA Test FM+ENV microphones with typical input to FM microphone EFE Test environmental microphone response with typical input EEF Test aid alone with high-level input EHA Test with FM system using a high-level input EFE Test environmental mic response with high-level input EEF Coupler Setup in the Sound Field Hearing Aid Setup FM System Setup Environmental microphones on body-worn receivers Directional: Testing FM-only Test aid alone with typical input EHA Test FM microphone with typical input EFM Test aid alone with high-level input EHA Test FM system with high-level input EFM Directional: Testing FM+ENV Single frequency output of hearing aid Single frequency output of FM mic at 65 db SPL Verify increase in signal level Test aid with typical input EHA

3 1.6.5 Test FM+ENV microphones with typical input to FM microphone EFE Test environmental mic response with typical input EEF Test aid with high-level input EHA Test with FM system using a high-level input EFE Test environmental mic response with high-level input EEF Section 2: Real-Ear Evalutation with FM Systems 2.1 Setup Hearing Aids FM System Setup for Testing Environmental Microphones Environmental microphones on body-worn receivers Environmental microphone on a hearing aid Testing FM-Only Test aid with typical input RHA Test FM system with typical input RFM Test aid with high-level input RHA Test FM system with high-level input RFM Testing FM+ENV Single frequency output of hearing aid Single frequency response of FM mic at 65 db SPL Verify increase in signal level Test aid with typical input RHA Test FM system with typical input RFM Test environmental mic with typical input REF Test aid with high-level input RHA Test FM system with high-level input RFE Test environmental mic with high-level input REF Glossary:

4 Introduction This workbook contains procedures for testing FM systems with the 6500-CX hearing aid analyzer using coupler and real-ear measurement methods. An FM system consists of two main parts: a transmitter and a receiver. The transmitter microphone can come in many different sizes and shapes, and it can be omni-directional or directional. The body-worn FM receiver can be coupled to a button earphone or interfaced with the personal hearing aid via a direct audio input boot, silhouette coil, or via a neck loop. The ear-level FM receivers are housed within the BTE aid or attached to the base of the BTE via a boot connection. No matter the combination, all FM systems can be tested. These test procedures should be performed on an FM system at least once a year. For children under the age of 5, these procedures should be performed every six months. As with most hearing aids, both the gain and the maximum output (SSPL) of an FM system should be adjusted to meet the listener s auditory requirements. There are several issues to consider when comparing the hearing aid and FM performance. With FM systems, the pickup microphone is normally worn at the chest, 6-8 inches (15-20 cm) under the mouth of the talker, whereas with hearing aids, the pickup microphone is worn by the listener, typically at a minimum of 3 ft. (1 m) from the talker. This difference affects both the level and the frequency content of typical input signals. With FM systems, typical speech input levels range from 75 to 85 db SPL, as compared with 60 to 70 db SPL for hearing aids. With the chest-worn location of the pickup microphone, the high frequencies are relatively lower in level (about 5 db at 5 khz) and the low frequencies are relatively higher in level (about 5 db at 500 Hz) as compared with directly in front of the talker s mouth. The increased vocal effort by teachers in a classroom situation may create further changes to the speech spectrum at the input to an FM system, as compared with the speech spectrum at the input of a hearing aid (Cornelisse et al., 1991). FM systems often have an automatic gain control (AGC) or other nonlinear characteristic that could interact with that of the hearing aid being used to deliver the FM signal if such a hearing aid is being used in this fashion. The frequency response of an FM system is often not as adjustable as that of a hearing aid, limiting the range of electroacoustic modifications available. The procedures recommended in this chapter are based on the ASHA guidelines for fitting and monitoring FM systems (1999), available at: Introduction 1

5 General Fitting Protocol For the purposes of this workbook, we will assume that the hearing aid being tested has been fitted appropriately for the client s loss. We will adjust the FM system to the output of the hearing aid. Goals: For students with mild to moderately-severe losses (<80 db HL), the goal is to match the output of the hearing aid and the FM system when typical input signals are used. With typical inputs, the FM output will be the same intensity as the hearing aid output. For students with severe to profound losses (>80 db HL), the goal is to match the gain of the hearing aid and the FM system when typical input signals are used. With typical input signals, the FM output will be about 10 db more intense than the hearing aid output. Testing Order: Start with 2-cc coupler tests and adjustments if possible. Follow up with real-ear verification. If time doesn t permit both test conditions, omit the 2-cc coupler tests. For both 2-cc coupler and real-ear tests: 1. Test the user s hearing aid with typical input signal levels. 2. Test the FM system with typical input signal levels, and make any necessary adjustments. 3. Test the user s hearing aid with high input signal levels. 4. Test the FM system with high input signal levels, and make any necessary adjustments. Curve Labeling System In order to efficiently record and refer to the various measurements, we have adopted a curve labeling system created by Dr. Linda Thibodeau. This system is consistent with the type of measurement made and the SPL of the test signal used. The first letter indicates whether it was a coupler or a real-ear measurement. All coupler measurement labels begin with an E for Electroacoustic. All real-ear measurement labels begin with an R for Real-Ear. The second and third letters indicate the type of test taken. HA means a hearing aid test FM means an FM-only test FE means an FM+ENV test with the signal delivered to the FM microphone EF means an FM+ENV test with the signal delivered to the environmental microphone 2 Testing FM Systems with FONIX FP40 Analyzers

6 The numbers at the end indicate the sound pressure level of the test signal used. Examples: EHA65 is the electroacoustic coupler measurement of the hearing aid with the signal input at 65 db SPL EFM80 is the electroacoustic coupler measurement of the FM microphone with the signal input at 80 db SPL RFE80 is the real-ear measurement of the FM+ENV with the signal delivered to the FM microphone at 80 db SPL See the Glossary for a list of all the labels and their definitions. Arrangement of This Workbook There are two main sections in this workbook. Section 1 describes sets of coupler measurements of FM systems. Section 2 describes corresponding sets of real-ear measurements of FM systems. Section 1 contains four different sets of testing procedures, but you only need to follow the testing procedure suitable for the FM system being tested. Section 2 contains two different sets of testing procedures. Likewise, you only need to follow the real-ear procedures suitable for the FM system. Ideally, you should perform one set of coupler measurements and one set of real-ear measurements on each FM system. It is assumed that you will perform the specified tests in sequential order. For instance, the instructions in Section 1.2.2, Testing the FM response with typical input, assume that you have done the steps in Section 1.2.1, Testing the hearing aid response with typical input. If it is more convenient for you to alter the testing order, just be aware that some of the key presses will change. Introduction 3

7 1: Coupler Measurements with FM Systems Coupler measurements of FM systems allow you to adjust the system without needing your client present. The goal is to set the FM system controls so that the user receives the FM signal at an appropriate level with reference to the acoustic signal the user receives from his personal hearing aid. There are four different sets of measurements described in this section. Use the set that is appropriate for the FM system setup. If the FM receiver and hearing aid are omni-directional and will fit inside the sound chamber, and: The FM receiver is primarily set to receive FM-only signals, follow the procedures described in Section 1.2, Omni-directional: Testing FM-only. The FM receiver is primarily set to receive FM with environmental signals (FM+ENV), follow the procedures described in Section 1.3, Omni-directional: Testing FM+ENV. If the FM receiver or hearing aid is directional or cannot fit inside the sound chamber, and The FM receiver is primarily set to receive FM-only signals, follow the procedures described in Section 1.5, Directional: Testing FM-only. The FM receiver is primarily set to receive FM with environmental signals (FM+ENV), follow the procedures described in Section 1.6, Directional: Testing FM+ENV. For all procedures, the hearing aid should be set to user settings. 1.1 Coupler Setup in the sound chamber This section describes how to attach a 2-cc coupler to the FM receiver, and how to set up an FM system for coupler measurements How to Attach the 2-cc Coupler A variety of output devices can be used with FM systems. These include button earphones, BTE cases with microphone/earphones, BTE aids with an internal FM receiver, an external FM receiver boot, or options for coupling to the personal BTE hearing aid via direct audio inputs, silhouette coils, or neck loops. This section contains a description of how to arrange the 2-cc coupler with each type of output device. 4 Testing FM Systems with FONIX 6500-CX Analyzers

8 FM Receiver FM Receiver FM Receiver Foam Pad Foam Pad Foam Pad to Analyzer to Analyzer to Analyzer Figure 1.1.1A Figure 1.1.1B Figure 1.1.1C Button earphone Internal FM receiver Silhouette coil External FM boot receiver Direct audio input Button earphone (Figure 1.1.1A) 1. Remove the ear-level adapter from the HA-2 coupler, if necessary. 2. Snap on the button earphone to the HA-2 coupler. 3. Place the coupler/hearing-aid assembly on a foam pad outside the text box, 2 ft. (60 cm) from the transmitter during testing. BTE case with microphone/earphone or BTE aid with internal FM receiver, external FM boot receiver, or direct audio input (Figure 1.1.1B) 1. Attach the ear-level adapter to the HA-2 coupler, if necessary. 2. Attach the BTE hearing aid to the HA-2 coupler. 3. Place the coupler/hearing-aid assembly on a foam pad outside the test box, 2 ft. (60 cm) from the transmitter during testing. BTE aid with silhouette coil (Figure 1.1.1C) 1. Attach the ear-level adapter to the HA-2 coupler, if necessary. 2. Attach the BTE hearing aid to the HA-2 coupler. 3. Place the coupler/hearing-aid assembly on a foam pad outside the text box, 2 ft. (60 cm) from the transmitter during testing. Choose a location free of stray magnetic fields (away from video monitors and other electrical devices). It may be necessary to turn the analyzer s video monitor off during testing. 4. Align the silhouette coil underneath the hearing aid, simulating the way it would be worn on the ear. Section 1: Coupler Tests of FM Systems 5

9 BTE aid with neck loop (Figure 1.1.1D) 1. Within the range of the coupler microphone cable, choose a location to seat the listener that is free of stray magnetic fields (away from video monitors and other electrical devices), while also maintaining the required separation (2 ft./60 cm) from the transmitter. It may be necessary to turn the analyzer s video monitor off during testing. 2. Place the neck loop around the listener s neck or around the neck of a person of similar size. 3. Attach the listener s hearing aid to the HA-2 coupler. 4. While holding the coupler in hand, place the hearing aid at the listener s ear, as typically worn. 5. Take care not to move or touch the coupler microphone while testing. FM Receiver to Analyzer Figure 1.1.1D Neck loop 6 Testing FM Systems with FONIX 6500-CX Analyzers

10 1.1.2 Hearing Aid Setup 1. Make sure the sound chamber is leveled. 2. Attach the hearing aid, set to user settings, to the HA-2 coupler in the usual way. Be sure the FM transmitter is off and the aid is set to receive the signal from the hearing aid microphone. See Figure Insert the coupler microphone into the coupler. 4. Place the assembly into the sound chamber. 5. Close and latch the sound chamber door. Figure Hearing aid setup with FM boot FM System 1. Make sure the sound chamber is leveled. 2. Set the FM receiver to receive FM-only. 3. Insert the coupler microphone into the HA-2 coupler and attach the output of the FM receiver as explained in Section 1.1.1, and place at least 2 ft. (60 cm) from all transmitter parts. See Figure 1.1.3A. 4. Place the FM transmitting microphone at the test position in the chamber. 5. Close the lid and latch the chamber. Section 1: Coupler Tests of FM Systems 7

11 Test Chamber FM Mic (or Environmental Mic) 6500 Main Electronics Module Foam Pad Receiver 2 ft. (60 cm) Transmitter Figure 1.1.3A FM setup Isolating the environmental microphone When the FM system is usually used in FM+ENV mode, the testing procedures call for the environmental microphone to be turned on as well at the FM microphone. In most quiet testing environments, this will call for the FM setup described above. However, if you are testing in a noisy environment, you may want to isolate the environmental microphone from the room noise. To check if you need to provide sound isolation for the environmental microphone when testing the FM microphone in FM+ENV mode, do this procedure: 1. Set the FM microphone up for testing as described above. 2. Set the FM receiver to receive FM only. 3. Observe the RMS OUT for this FM-only signal. 4. Set the FM receiver to receive FM+ENV. 5. Observe the RMS OUT for the FM+ENV setting. If the RMS OUT is more than 3 db higher for FM+ENV than for FM-only, then you will need to isolate the environmental microphone when testing the FM microphone in FM+ENV mode. If the RMS OUT for both signals is within 3 db, no additional setup is necessary. 6. To isolate the environmental microphone, it is preferable to use a spare sound chamber if one is available to you. Otherwise, take a shoebox, line it with packing foam, and place the environmental microphone inside, covering it with some more foam. See Figure 1.1.3B. This should isolate the room noise from the environmental microphone. 8 Testing FM Systems with FONIX 6500-CX Analyzers

12 Isolate environmental microphone Shoebox lined with foam FM mic Figure 1.1.3B Isolating the environmental microphone Environmental Microphone Setup Environmental microphones on ear-level FM receivers 1. Make sure the sound chamber is leveled. 2. Attach the ear-level FM receiver to the HA-2 coupler and set it to receive input from the environmental microphone only. 3. Insert the coupler microphone into the HA-2 coupler. 4. Place the ear-level receiver and coupler at the test position in the sound chamber. 5. Close and latch the sound chamber lid. 6. You are now ready to test the FM system with the signal delivered to the environmental microphone. Section 1: Coupler Tests of FM Systems 9

13 Lapel environmental microphones on Body-worn FM Receivers 1. Make sure the sound chamber is leveled. 2. Attach the output transducer from the FM receiver to the HA-2 coupler outside the sound chamber as described in Section Insert the coupler microphone into the HA-2 coupler. 4. Place the lapel microphone from the FM receiver at the test position in the sound chamber. See Figure Close and latch the lid of the sound chamber. 6. You are now ready to test the FM system with the signal delivered to the environmental microphone. Figure Testing lapel receivers 10 Testing FM Systems with FONIX 6500-CX Analyzers

14 1.2 Omni-directional: FM-only This section contains a description of the four different coupler measurements of FM systems when the client primarily uses the system in FM-only mode. These measurements are: EHA65: typical electroacoustic response of the aid EFM80: typical electroacoustic response of the FM transmitter/receiver EHA90: saturation electroacoustic response of the aid EFM90: saturation electroacoustic response of the FM transmitter/receiver Test aid alone with typical input EHA65 1. Set up the hearing aid according to the hearing aid setup described in Section Press [RESET] to enter Composite Weighted Gain Mode. 3. Press [WEIGHT] to enter Composite Weighted Power Mode. The display will be labeled as OUTPUT SPL. 4. Use [ ] to set RMS Source to 65 db SPL. 5. Once the curve is stable, press [FREEZE] to pause the measurement. If the 6500-CX has the Multi-Curve Option, press [START] (not [CONTINUE]!) to save the curve into the Multi-Curve stack. For units without Multi-Curve, press [CRT] to print the measurement. 6. This will be referred to as the Electroacoustic Hearing Aid Response or EHA65. See Figure Figure EHA65 hearing aid response at 65 db SPL Section 1: Coupler Tests of FM Systems 11

15 1.2.2 Test FM microphone with typical input EFM80 1. Set up the FM system according to the FM system setup in Section Use [ ] to adjust the signal level to 80 db SPL. 3. This curve will be referred to as the Electroacoustic FM Response or EFM80. (Since the current running curve is always stored as CRV 1 in the Multi-Curve buffer, there is no need to press [FREEZE] and [START] to save the curve as you did in Section ) 4. To view the EFM80 and EHA65 on the same screen requires the Multi-Curve Option: a. Press [MULTI]. b. Highlight MULTIPLE CURVES. c. Press [START]. d. If necessary, use [<, >] to display CRV 1 in the first slot. This curve is EFM80. e. Use [ ] to select the second slot. f. If necessary, use [<, >] to display CRV 2 in the second slot. This curve is EHA65. See Figure If you don t have the Multi-Curve Option, press [CRT] to print out EFM80 and compare it to EHA65. Figure Comparison of EHA65 (CRV2), the hearing aid response at 65 db SPL and EFM80 (CRV1), the FM response at 80 db SPL 12 Testing FM Systems with FONIX 6500-CX Analyzers

16 5. If you have a client with a mild to moderate hearing loss (less than 80 db HL), the output of the two curves should be within 5 db of each other. If you have a client with a severe to profound loss (greater than 80 db HL), the output of EFM80 should be 7-10 db higher than the output of EHA65. If you have to adjust the system to obtain a better result, first adjust the FM receiver volume control if possible. When using FM systems coupled to personal hearing aids adjust the settings on the hearing aid as a final step if necessary. If this must be done, the teacher, parents, and therapists must be made aware that when the FM system is used, the settings on the hearing aid must be changed for optimum performances Test aid alone with high-level input EHA90 1. Turn off the FM signal, and set up the hearing aid for testing as described in the hearing aid setup in Section Press [SINE/COMPOSITE] button to enter Pure Tone Mode. The analyzer will automatically set itself to a signal level of 90 db SPL. 3. Press [START] to begin the pure-tone sweep. The response will be stored in the Multi-Curve stack when the measurement is complete. For instruments without the Multi-Curve option, press [CRT] to print a hardcopy. Figure EHA90 hearing aid response at 90 db SPL 4. This response will be referred to as the Electroacoustic Hearing Aid Saturation Response or EHA90. See Figure Section 1: Coupler Tests of FM Systems 13

17 1.2.4 Test FM system with high-level input EFM90 1. Set up the FM system for testing according to the FM setup in Section The FM receiver should be set to receive FM-only signals. 2. Press [START] to begin the pure-tone sweep. The response will be stored in the Multi-Curve stack when the measurement is complete. For instruments without the Multi-Curve option, press [CRT] to print a hardcopy. 3. This response will be referred to as the Electroacoustic FM Saturation Response or EFM90. Figure Comparison of EHA90 (CRV2), the hearing aid response at 90 db SPL with EFM90 (CRV1), the FM response at 90 db 4. To view both curves on the same screen requires the Multi-Curve Option: a. Press [MULTI]. b. Highlight MULTIPLE CURVES. c. Press [START]. d. If necessary, use [<, >] to display CRV 1 in the first slot. This curve is EFM90. e. Use [ ] to select the second slot. f. If necessary, use [<, >] to display CRV 2 in the second slot. This curve is EHA Testing FM Systems with FONIX 6500-CX Analyzers

18 g. You may want to press [ ] and [FREEZE] to turn off the display of CRV 3, since it is not relevant to this comparison. Similarly, press [ ] and [FREEZE] again to turn off the display of CRV 4. If you don t have the Multi-Curve Option, press [CRT] to print out EFE90 and compare it to EHA If EFM90 is greater than 5 db above or 10 db below EHA90, adjust the maximum output control if possible on the FM receiver or on the hearing aid if necessary. If the hearing aid is adjusted, the teacher, parents, and therapists must be made aware that when the FM system is used, the settings on the hearing aid must be changed for optimum performances. For mild to moderate losses it may be acceptable if the EFM90 is 5-10 db below EHA90. See Figure Section 1: Coupler Tests of FM Systems 15

19 1.3 Omni-directional: FM+ENV This section contains a description of the different coupler measurements of omni-directional FM systems when the client primarily uses the system with the environmental microphone added to the FM signal (FM+ENV mode). These measurements are: Response of aid with 65 db SPL signal at 1 khz Response of FM transmitter/receiver with 65 db SPL signal at 1 khz Verify increase of FM transmitter/receiver when signal level is increased to 80 db SPL. EHA65: typical electroacoustic response of the aid EFE80: typical electroacoustic response of the FM transmitter/receiver EEF65: typical electroacoustic response of environmental microphone EHA90: saturation electroacoustic response of the aid EFE90: saturation electroacoustic response of the FM transmitter/receiver EEF90: saturation electroacoustic response of environmental microphone Single frequency output of omni-directional hearing aid in the sound chamber 1. Set up the hearing aid for testing as described in the hearing aid setup in Section The FM transmitter should be off, and the hearing aid set to receive input from the hearing aid microphone. 2. Press [RESET]. 3. Press [SINE/COMPOSITE] to put the analyzer in PURE TONE MODE. 4. If necessary, use [<, >] to adjust the frequency of the single tone signal to 1000 Hz. Figure Hearing aid output at 65 db SPL at 1000 Hz 16 Testing FM Systems with FONIX 6500-CX Analyzers

20 5. Use [ ] to adjust the level of the signal to 65 db SPL. 6. Make a note of the AID OUT on the screen. This is the HEARING AID OUTPUT. See Figure Single frequency output of omni-directional FM mic at 65 db SPL 1. Set up the FM microphone for testing as described in the FM setup in Section The FM receiver should be set to receive FM+ENV. 2. If you followed the steps of Section 1.3.1, the analyzer should be in PURE TONE MODE with a 65 db SPL signal at 1000 Hz. 3. Look at the AID OUT on the screen. This is the FM OUTPUT. 4. Adjust the FM receiver volume so that the FM OUTPUT is matched to the HEAR- ING AID OUTPUT found in Section This matches the gain of the FM system and the hearing aid Verify increase in signal level 1. Leave the FM system and analyzer set up as described in Section with the aid or FM receiver attached to the 2-cc coupler and set to receive FM+ENV. 2. Press [ ] repeatedly to increase the signal level to 80 db SPL. 3. Look at the AID OUT on the screen. This is the new FM OUTPUT. 4. Compare the HEARING AID OUTPUT found in Section to this new FM OUT- PUT, and adjust the FM receiver volume control according to the following: If the increase is: 10 db, no further adjustments are necessary. 15 db, reduce the FM receiver volume control so that the FM OUTPUT falls by 5 db. 5 db, increase the FM receiver volume control so that the FM OUTPUT increases by 7 or 8 db. 0 db, the FM transmitter must have a very low compression threshold, so increase the FM receiver volume control so that the FM OUTPUT increases by 5 db Test aid alone with typical input EHA65 1. Set up the hearing aid for testing as described in Section The FM transmitter should be off, and the hearing aid set to receive input from the hearing aid microphone. 2. Press [RESET] to enter Composite Weighted Gain Mode. 3. Press [WEIGHT] to enter Composite Weighted Power Mode. 4. Use [ ] to set RMS Source to 65 db SPL. Section 1: Coupler Tests of FM Systems 17

21 Figure EHA65 hearing aid response at 65 db SPL 5. Once the curve is stable, press [FREEZE] to pause the measurement. If the 6500-CX has the Multi-Curve Option, press [START] (not [CONTINUE]!) to save the curve into the Multi-Curve stack. For units without Multi-Curve, press [CRT] to print the measurement. 6. This will be referred to as the Electroacoustic Hearing Aid Response or EHA65. See Figure Test FM+ENV microphones with typical input to FM microphone EFE80 1. Set up the FM system according to Section The FM receiver should be set to receive FM+ENV. 2. Use [ ] to adjust the signal level to 80 db SPL. 3. This curve will be referred to as the Electroacoustic FM Response or EFE To view the EFE80 and EHA65 on the same screen requires the Multi-Curve Option: a. Press [MULTI]. b. Highlight MULTIPLE CURVES. c. Press [START]. d. If necessary, use [<, >] to display CRV 1 in the first slot. This curve is EFE80 or the Electroacoustic FM and Environmental Response with 80 db input to the FM microphone. e. Use [ ] to select the second slot. 18 Testing FM Systems with FONIX 6500-CX Analyzers

22 f. If necessary, use [<, >] to display CRV 2 in the second slot. This curve is EHA65. If you don t have the Multi-Curve Option, press [CRT] to print out EFE80 and compare it to EHA65. Figure Comparison of EHA65 (CRV2), the hearing aid response at 65 db SPL and EFM80 (CRV1), the FM response at 80 db SPL 5. Compare EHA65 (CRV 2) to EFE80 (CRV 1) to confirm that the desired FM advantage is maintained over the 500 to 2000 Hz range. In general, that advantage should be 7-10 db. See Figure For units with Multi-Curve, you will want to save EFE80 for a comparison in the next section. To do this: a. Press [CONTINUE] twice to exit from Multi-Curve. b. Press [FREEZE] to save the current EFE80 measurement. c. Press [START] to save the curve to the Multi-Curve stack. EFE80 has just become CRV 2 and EHA65 has just become CRV 3. You will use them in the next section. Section 1: Coupler Tests of FM Systems 19

23 1.3.6 Test environmental microphone response with typical input EEF65 1. Set up the environmental microphone for testing as described in Section Set the FM receiver to receive FM+ENV signals. 2. Use [ ] to set the source level to 65 db SPL and wait for the response to stabilize. 3. This measurement will be referred to as EEF65, or the Electroacoustic FM and Environmental Response with 65 db SPL input to the environmental microphone. 4. Compare EEF65 to EHA65 (from Section 1.3.4) and EFE80 (from Section 1.3.5). If you have Multi-Curve, do the following: a. Press [MULTI]. b. Highlight MULTIPLE CURVES. c. Press [START]. d. CRV 1 will be selected for the first Multi-Curve slot. This is the EEF65 response. CRV 2 should be selected for the second Multi-Curve slot. This is the EFE80 response. Use [ ] to select the third Multi-Curve slot. e. Use [<, >] to select CRV 3 for the third Multi-Curve slot. This is the EHA65 response. f. Compare EEF65 (CRV 1), EFE80 (CRV 2), and EHA65 (CRV 3). For a successful fit, the EHA65 and EEF65 should be similar, and the EFE80 should be 7-10 db higher. See Figure Figure Comparison of EHA65 (CRV1), EFM80 (CRV2), and EEF65, the environmental microphone response at 65 db SPL 20 Testing FM Systems with FONIX 6500-CX Analyzers

24 1.3.7 Test aid alone with high-level input EHA90 1. Turn off the FM signal, and set up the hearing aid for testing as described in Section Press [SINE/COMPOSITE] button to enter Pure Tone Mode. The analyzer will automatically set itself to a signal level of 90 db SPL. 3. Press [START] to begin the pure-tone sweep. The response will be stored in the Multi-Curve stack when the measurement is complete. For instruments without the Multi-Curve option, press [CRT] to print a hardcopy. 4. This response will be referred to as the Electroacoustic Hearing Aid Saturation Response or EHA90. See Figure Figure EHA90 the hearing aid response at 90 db SPL Test with FM system using a high-level input EFE90 1. Set up the FM system for testing according to Section The FM receiver should be set to receive FM+ENV. 2. Press [START] to begin the pure-tone sweep. The response will be stored in the Multi-Curve stack when the measurement is complete. For instruments without the Multi-Curve option, press [CRT] to print a hardcopy. 3. This response will be referred to as EFE90 or the Electroacoustic FM and Environmental Response with 90 db SPL input to the FM microphone. Section 1: Coupler Tests of FM Systems 21

25 Figure Comparison of EHA90, the hearing aid response at 90 db SPL, and EFM 90, the FM response at 90 db SPL 4. To view both curves on the same screen requires the Multi-Curve Option: a. Press [MULTI]. b. Highlight MULTIPLE CURVES. c. Press [START]. d. If necessary, use [<, >] to display CRV 1 in the first slot. This curve is EFE90. e. Use [ ] to select the second slot. f. If necessary, use[<, >] to display CRV 2 in the second slot. This curve is EHA90. g. You may want to press [ ] and [FREEZE] to turn off the display of CRV 3, since it is not relevant to this comparison. Similarly, press [ ] and [FREEZE] again to turn off the display of CRV 4. If you don t have the Multi-Curve Option, press [CRT] to print out EFE90 and compare it to EHA If EFE90 is greater than 5 db above EHA90, adjust the FM system to obtain a better match. As a last resort, adjust the settings on the hearing aid. If this must be done, the teacher, parents, and therapists must be made aware that when the FM system is used, the settings on the hearing aid must be changed for optimum performances. See Figure Testing FM Systems with FONIX 6500-CX Analyzers

26 1.3.9 Test environmental mic response with high-level input EEF90 1. Set up the environmental microphone for testing as described in Section Set the FM receiver to FM+ENV. 2. Press [START/STOP] to run a pure-tone measurement at 90 db SPL. 3. This measurement will be referred to as EEF90 or the Electroacoustic FM and Environmental Response with 90 db SPL input to the environmental microphone. 4. To view both curves on the same screen requires the Multi-Curve Option: a. Press [MULTI]. b. Highlight MULTIPLE CURVES. c. Press [START]. d. EEF90 (CRV 1) should be displayed in the first slot, and EFE90 (CRV 2) should be displayed in the second slot. e. Press [ ] to select the third slot. f. If necessary, press [<, >] to select CRV 3 for the third slot. This is the EHA90 measurement. 5. Compare EEF90 (CRV 1) to EFE90 (CRV 2) and EHA90 (CRV 3). All three curves should match. See Figure Figure Comparison of EHA90, EFM90, and EFF90, the environmental microphone response at 90 db SPL Section 1: Coupler Tests of FM Systems 23

27 1.4 Coupler Setup in the Sound Field When testing a FM system with a large or directional microphone, the measurements are performed in the sound field. Therefore, the corresponding hearing aid tests will also have to be performed in the sound field. This requires a little creativity; you may have to adjust the following recommended procedure to fit the particular equipment that you are testing Hearing Aid Setup 1. Enter the Real Ear Mode by pressing [START/STOP] on the remote module. All further key presses described in this section will be performed on the remote module. 2. Affix the reference microphone with Fun-Tak to the platform at the top of the floor stand, 6 inches (15 cm) in front of the sound field speaker, with the microphone pointing toward the speaker. See Figure 1.4.1A. (If you don t have a floor stand, the back of a chair can suffice. But be sure the chair is facing to the side of the setup to avoid reflections from the back of the chair.) 6 inches Reference mic Probe mic (not used) Figure 1.4.1A Sound field leveling setup 3. Stand clear of the test area, holding the remote module. 4. Press [LEVEL] on the remote module. 5. Remove the reference microphone and the platform from the floor stand, leaving the floor stand in its position in front of the sound field speaker. 6. Attach the hearing aid to the HA-2 coupler. 24 Testing Testing FM Systems FM Systems with with FONIX FONIX 6500-CX FP40 Analyzers

28 7. Thread a probe tube through the black probe tube adapter with the tube first entering the ridged end of the adapter. The tube should extend 1-2 mm past the other end of the adapter. Use Fun-Tak to secure the probe tube to the adapter. See Figure 1.4.1B. Figure 1.4.1B Thread probe tube through probe mic adapter 8. Insert the adapter into the HA-2 coupler, and attach the probe tube to the probe microphone. See Figure 1.4.1C. Probe mic adapter Probe mic Figure 1.4.1C Attach coupler to probe mic Section 1: Coupler Tests of FM Systems 25

29 9. Securely attach the assembly to the floor stand using Fun-Tak. The body of the hearing aid should be vertical with the hearing aid microphone pointed towards the sound field speaker. See Figure 1.4.1D. Figure D Setup for testing a hearing aid in the sound field using a coupler 10. You are now ready to test the hearing aid FM System Setup Since the tests of the FM transmitter/receiver in this chapter always follow the tests of the hearing aid, it is assumed in this setup that you have already entered Real Ear Mode and leveled the sound field speaker. If you are testing out of order, follow the instructions found in Section 1.4.1, Steps 1-4, to level the sound field before testing. 1. Place the FM transmitting microphone on the floor stand six inches (15 cm) from the sound field speaker. Keep all sound-entry ports clear of the platform. Place the transmitter and its cord behind the transmitting microphone. 2. Thread a probe tube through the black probe tube adapter with the tube first entering the ridged end of the adapter. The tube should extend 1-2 mm past the other end of the adapter. Use Fun-Tak to secure the probe tube to the adapter. See Figure 1.4.1B. 3. Insert the adapter into the HA-2 coupler, and attach the probe tube to the probe microphone. See Figure 1.4.1C. 4. Attach the FM receiver to the HA-2 coupler as described in Section Testing FM Systems with FONIX 6500-CX Analyzers

30 5. Place the assembly 2 ft (60 cm) from the transmitting microphone, out of the sound field of the speaker. See Figure Stand clear of the test area, holding the remote module. You are now ready to test the FM transmitter/receiver. FM mic Reference mic (not used) Probe mic FM transmitter FM receiver attached Figure FM setup system Isolating the environmental microphone The testing procedures in this section call for the environmental microphone to be turned on as well at the FM microphone. In most quiet testing environments, this will call for the FM setup described in Section However, if you are testing in a noisy environment, you may want to isolate the environmental microphone from the room noise. To check if you need to provide sound isolation for the environmental microphone when testing the FM microphone in FM+ENV mode, do this procedure: 1. Set the FM microphone up for testing as described in Section Set the FM receiver to receive FM only. 3. Press [START/STOP] on the remote module to start a composite measurement. When the measurement stabilizes, press [START/STOP] again. 4. Observe the RMS OUT for this FM-only signal. 5. Set the FM receiver to receive FM+ENV. 6. Repeat Step 3. Section 1: Coupler Tests of FM Systems 27

31 7. Observe the RMS OUT for the FM+ENV setting. If the RMS OUT is more than 3 db higher for FM+ENV than for FM-only, then you will need to isolate the environmental microphone when testing the FM microphone in FM+ENV mode. If the RMS OUT for both signals is within 3 db, no additional setup is necessary. 8. To isolate the environmental microphone, it is preferable to use a spare sound chamber if one is available to you. Otherwise, take a shoebox, line it with packing foam, and place the environmental microphone inside, covering it with some more foam. See Figure 1.1.3B. This should isolate the room noise from the environmental microphone Environmental microphones on body-worn receivers 1. Enter Real Ear Mode by pressing [START/STOP] on the remote module or [PROBE] on the 6500-CX main module. 2. Affix the FM receiver to the platform stand in the manner shown in Figure 1.4.3A. Reference Microphone Figure 1.4.3A Attaching reference microphone 3. Using Fun-Tak, affix the reference microphone to the receiving unit so the entrance to the reference microphone is very close to that of the environmental microphone. 4. Place the FM receiver six inches (15 cm) from, and facing, the loudspeaker (Refer to Figure 1.4.3B) and set it to receive input from the environmental microphone only. 5. Choose a place to stand during testing, preferably away from the loudspeaker and microphone, holding the remote module. Stand precisely in this place during both leveling and testing. 6. Press [LEVEL] on the remote module. This levels the test signal at the place where the FM receiver microphone will be during testing. 7. Leave the reference microphone in place during testing. 28 Testing FM Systems with FONIX 6500-CX Analyzers

32 8. Thread the probe tube through the probe tube adapter so that it extends 3 mm past the end. Use Fun-Tak to hold the probe tube in place. See Figure 1.4.1B. 9. Insert the adapter and probe tube (attached to the probe microphone) into the HA-2 coupler, and attach to the button earphone, direct audio input hearing aid, or aid coupled with neck loop or silhouette, as explained in Section You are now ready to test the FM system with the signal delivered to the environmental microphone. Figure 1.4.3B Leveling the sound field Figure 1.4.3C Testing with body-worn receiver Section 1: Coupler Tests of FM Systems 29

33 1.5 Directional: Testing FM-only This section contains a description of the four different coupler measurements of FM systems when the client primarily uses the system in FM-only mode. These measurements are: EHA65: typical electroacoustic response of the aid EFM80: typical electroacoustic response of the FM transmitter/receiver EHA90: saturation electroacoustic response of the aid EFM90: saturation electroacoustic response of the FM transmitter/receiver Test aid alone with typical input EHA65 1. Follow the hearing aid setup instructions found in Section to set up the hearing aid for testing the sound field, enter Real Ear Mode, and level the sound field speaker. The FM transmitter should not be on and the aid should be set to receive the hearing aid microphone. 2. Press [MENU] on the remote module. 3. Press [ ] to highlight CREATE TARGET. 4. Press [<, >] to select SPL TARGET. 5. Press [START/STOP] twice. This will open the SPL Test Screen. Figure EHA65 the hearing aid response at 65 db SPL 6. Look in the bottom left corner of your screen. If it says AUTO, press [UNAIDED] on the remote module. This will put you in MANUAL Mode. 30 Testing FM Systems with FONIX 6500-CX Analyzers

34 7. Press [MENU] to enter the SPL Menu. 8. Select COMP for AIDED CURVE 1 and AIDED CURVE Press [MENU] to return to the SPL Test Screen. 10. If necessary, select CRV 1 with [ ]. 11. Use [<, >] to select a source level of 65 db SPL. 12. Press [START/STOP] to begin the measurement. 13. Press [START/STOP] once the measurement has stabilized. This curve is the Electroacoustic Hearing Aid Response or EHA65. See Figure Test FM microphone with typical input EFM80 1. Set up the FM system for testing according to FM setup described in Section Use [ ] to select CRV 2. 3 Use [<, >] to set the RMS source to 80 db SPL. 4. Press [START/STOP] to start the measurement. 5. Once the measurement stabilizes, press [START/STOP] again to stop the measurement. Figure Comparison of EHA65 (CRV1), the hearing aid response at 65 db SPL and EFM80 (CRV2), the FM response at 80 db SPL Section 1: Coupler Tests of FM Systems 31

35 6. CRV 2 is now the Electroacoustic FM Response or EFM If you have a client with a mild to moderate hearing loss (less than 80 db HL), the output of the two curves should be within 5 db of each other. If you have a client with a severe to profound loss (greater than 80 db HL), the output of EFM80 should be 7-10 db higher than the output of EHA65. If you have to adjust the system to obtain a better result, first adjust the FM receiver volume control if possible. When using FM systems coupled to personal hearing aids adjust the settings on the hearing aid as a final step if necessary. If this must be done, the teacher, parents, and therapists must be made aware that when the FM system is used, the settings on the hearing aid must be changed for optimum performances. See Figure Test aid alone with high-level input EHA90 1. Set up the hearing aid for testing in the sound field by following the directions found in the hearing aid setup in Section If you ve already leveled the sound field speaker, you can skip to step 6 in that section. 2. Press [MENU] to enter the SPL Menu. 3. Select PURE TONE for AIDED CURVE 1 and AIDED CURVE Press [MENU] to return to the SPL Test Screen. 5. Press [ ] to select CRV 1. Figure EHA90 the hearing aid response at 90 db SPL 32 Testing FM Systems with FONIX 6500-CX Analyzers

36 6. Use [>] to select a source level of 90 db SPL. 7. Press [START/STOP] to begin the sweep. 8. This response will be referred to as the Electroacoustic Hearing Aid Saturation Response or EHA90. See Figure Test FM system with high-level input EFM90 1. Set up the FM system for testing according to the FM setup described in Section Use [ ] to select CRV Use [>] to select a source level of 90 db SPL. 4. Press [START/STOP] to make the measurement. 5. CRV 2 is now the Electroacoustic FM Saturation Response or EFM Compare EHA90 (CRV 1) to EFM90 (CRV 2). If EFM90 is greater than 5 db above or 10 db below EHA90, adjust the maximum output control if possible on the FM receiver or on the hearing aid if necessary. If the hearing aid is adjusted, the teacher, parents, and therapists must be made aware that when the FM system is used, the settings on the hearing aid must be changed for optimum performances. For mild to moderate losses it may be acceptable if the EFM90 is 5-10 db below EHA90. See Figure Figure Comparison of EHA90 (CRV1), the hearing aid response at 90 db SPL with EFM90 (CRV2), the FM response at 90 db SPL Section 1: Coupler Tests of FM Systems 33

37 1.6 Directional: Testing FM+ENV This section contains a description of the different coupler measurements of FM systems when the client primarily uses the system with the environmental microphone added to the FM signal (FM+ENV mode). These measurements are: Response of aid with 65 db SPL signal at 1 khz Response of FM transmitter/receiver with 65 db SPL signal at 1 khz Verify increase of FM transmitter/receiver when signal level is increased to 80 db SPL. EHA65: typical electroacoustic response of the aid EFE80: typical electroacoustic response of the FM transmitter/receiver EEF65: typical electroacoustic response of environmental microphone EHA90: saturation electroacoustic response of the aid EFE90: saturation electroacoustic response of the FM transmitter/receiver EEF90: saturation electroacoustic response of environmental microphone Single frequency output of hearing aid 1. Set up the aid for testing according to the hearing aid setup described in Section The FM transmitter should be off and the hearing aid set to receive input from the hearing aid microphone. Make sure you level the sound field speaker. 2. Press [MENU]. 3. Use [ ] to highlight MODE. 4. Use [<, >] to select PURE TONE. 5. Press [MENU] to exit from the Quik-Probe II Menu. Figure Hearing aid output at 65 db SPL at 1000 Hz 34 Testing FM Systems with FONIX 6500-CX Analyzers

38 6. Press [START/STOP] to begin a single frequency signal. 7. If necessary, use [<, >] to adjust the frequency to 1000 Hz. 8. Use [, ] to adjust the signal level to 65 db SPL. 9. Note the MIC OUT on the screen. This is the HEARING AID OUTPUT. See Figure Press [START/STOP] to turn off the signal Single frequency output of FM mic at 65 db SPL 1. Set up the FM system for testing according to Section The FM receiver should be set to receive FM+ENV. 2. If you followed the procedure described in Section 1.6.1, you should be in Real Ear Mode, with the analyzer ready to deliver a pure-tone signal of 1000 Hz at 65 db SPL. 3. Press [START/STOP] to start the measurement. 4. Note the AID OUT on the screen. This is the FM OUTPUT. 5. Adjust the FM receiver volume so that the FM OUTPUT is matched to the HEAR- ING AID OUTPUT found in Section This matches the gain of the FM system and the hearing aid. 6. Press [START/STOP] to turn off the signal Verify increase in signal level 1. Leave the FM system and the analyzer set up as described in Section Press [START/STOP] to start the single frequency measurement. 3. Press [ ] repeatedly to increase the signal level to 80 db SPL. 4. Note the AID OUT on the screen. This is the FM OUTPUT. 5. Compare the HEARING AID OUTPUT found in Section to this new FM OUT- PUT, and adjust the FM receiver volume control according to the following: If the increase is: 10 db, no further adjustments are necessary. 15 db, reduce the FM receiver volume control so that the FM OUTPUT falls by 5 db. 5 db, increase the FM receiver volume control so that the FM OUTPUT increases by 7 or 8 db. 0 db, the FM transmitter must have a very low compression threshold, so increase the FM receiver volume control so that the FM OUTPUT increases by 5 db. Section 1: Coupler Tests of FM Systems 35

Testing FM Systems on the 7000 Hearing Aid Test System

Testing FM Systems on the 7000 Hearing Aid Test System Testing FM Systems on the 7000 Hearing Aid Test System Introduction Testing FM Systems on the 7000 Hearing Aid Test System This workbook describes how to test FM systems with the 7000 Hearing Aid Test

More information

Testing FM Systems with FONIX FP40 Analyzers

Testing FM Systems with FONIX FP40 Analyzers Testing FM Systems with FONIX FP40 Analyzers FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 USA (503) 620-2722 (800) 547-8209 Fax: (503) 639-0128 http://www.frye.com e-mail: general@frye.com

More information

Using Multi-Curve in Real-Ear Measurements on the 6500-CX Hearing Aid Analyzer

Using Multi-Curve in Real-Ear Measurements on the 6500-CX Hearing Aid Analyzer Using Multi-Curve in Real-Ear Measurements on the 6500-CX Hearing Aid Analyzer FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 (503) 620-2722 (800) 547-8209 Fax: (503) 639-0128 www.frye.com

More information

Testing Digital Hearing Aids

Testing Digital Hearing Aids Testing Digital Hearing Aids with the FONIX 6500-CX Hearing Aid Analyzer Frye Electronics, Inc. Introduction The following is a quick guide for testing digital hearing aids using the FONIX 6500-CX. All

More information

Testing Digital Hearing Aids

Testing Digital Hearing Aids Testing Digital Hearing Aids with the FONIX FP40 Hearing Aid Analyzer FRYE 7 Introduction The following is a quick guide for testing digital hearing aids. All digital aids can be tested, but some of the

More information

Step-by-Step RECD Guide

Step-by-Step RECD Guide Precision Audiometric Instruments www.medrx-usa.com Step-by-Step RECD Guide The RECD task involves 4 steps: 1 - Complete Calibration of the Speakers and Probe Tube 2 - Measure an Ear Response 3 - Perform

More information

INTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs.

INTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs. INTRODUCTION TO PURE TONE AUDIOMETERY (AUDIOMETER & TESTING ENVIRONMENT) By Mrs. Wedad Alhudaib with many thanks to Mrs. Tahani Alothman Topics : This lecture will incorporate both theoretical information

More information

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04 AURICAL Plus with DSL v. 5.0b Quick Guide 0459 Doc no. 7-50-0900/04 Copyright notice No part of this Manual or program may be reproduced, stored in a retrieval system, or transmitted, in any form or by

More information

ELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT

ELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT ELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT Astrid Haastrup, GN ReSound Mona Dworsack-Dodge, AuD, GN Otometrics Abstract With ReSound s 2.4 GHz wireless

More information

ELECTROACOUSTIC EVALUATION OF THE RESOUND MULTI MIC WITH OTOMETRICS AURICAL HIT

ELECTROACOUSTIC EVALUATION OF THE RESOUND MULTI MIC WITH OTOMETRICS AURICAL HIT ELECTROACOUSTIC EVALUATION OF THE RESOUND MULTI MIC WITH OTOMETRICS AURICAL HIT Astrid Haastrup, GN ReSound Mona Dworsack-Dodge, AuD, GN Otometrics Abstract With ReSound s 2.4 GHz wireless technology and

More information

Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial

Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial Enhancing the Initial Hearing Aid Fitting Michael Block, Ph.D. Session 0228 Importance of a Good Start The consequences of doing a poor job of selling the hearing-impaired client on a positive treatment

More information

ChildFit. Widex Baby. Compass quick guide

ChildFit. Widex Baby. Compass quick guide ChildFit When working with very young children, it is not always possible to obtain all of the information necessary for an accurate hearing aid fitting. To that end, Widex has developed the ChildFit procedure

More information

A Sound Foundation Through Early Amplification

A Sound Foundation Through Early Amplification 11 A Sound Foundation Through Early Amplification Proceedings of the 7th International Conference 2016 Hear well or hearsay? Do modern wireless technologies improve hearing performance in CI users? Jace

More information

April 23, Roger Dynamic SoundField & Roger Focus. Can You Decipher This? The challenges of understanding

April 23, Roger Dynamic SoundField & Roger Focus. Can You Decipher This? The challenges of understanding Roger Dynamic SoundField & Roger Focus Lindsay Roberts, Au.D. Pediatric and School Specialist PA, WV, OH, IN, MI, KY & TN Can You Decipher This? The challenges of understanding Meaningful education requires

More information

How to use AutoFit (IMC2) How to use AutoFit (IMC2)

How to use AutoFit (IMC2) How to use AutoFit (IMC2) How to use AutoFit (IMC2) 1 AutoFit is a beneficial feature in the Connexx Fitting Application that automatically provides the Hearing Care Professional (HCP) with an optimized real-ear insertion gain

More information

Tune in on life with SCOLAbuddy. A new fm receiver from Widex

Tune in on life with SCOLAbuddy. A new fm receiver from Widex Tune in on life with SCOLAbuddy A new fm receiver from Widex Meet your new buddy The latest member of the Scola FM family The successful Widex SCOLA FM product series has acquired a new family member:

More information

VPAT Summary. VPAT Details. Section Telecommunications Products - Detail. Date: October 8, 2014 Name of Product: BladeCenter HS23

VPAT Summary. VPAT Details. Section Telecommunications Products - Detail. Date: October 8, 2014 Name of Product: BladeCenter HS23 Date: October 8, 2014 Name of Product: BladeCenter HS23 VPAT Summary Criteria Status Remarks and Explanations Section 1194.21 Software Applications and Operating Systems Section 1194.22 Web-based Internet

More information

Using the FONIX 7000 to Verify Coupler and Real-Ear Performance to Adhere to the AAA Guidelines for the Audiologic Management

Using the FONIX 7000 to Verify Coupler and Real-Ear Performance to Adhere to the AAA Guidelines for the Audiologic Management Using the FONIX 7000 to Verify Coupler and Real-Ear Performance to Adhere to the AAA Guidelines for the Audiologic Management of Adult Hearing Impairment Elizabeth Baum, B.A. Au.D. Student Program in Audiology

More information

AMBCO 1000+P AUDIOMETER

AMBCO 1000+P AUDIOMETER Model 1000+ Printer User Manual AMBCO 1000+P AUDIOMETER AMBCO ELECTRONICS 15052 REDHILL AVE SUITE #D TUSTIN, CA 92780 (714) 259-7930 FAX (714) 259-1688 WWW.AMBCO.COM 10-1004, Rev. A DCO 17 008, 11 13 17

More information

Roger TM. for Education Bridging the understanding gap

Roger TM. for Education Bridging the understanding gap Roger TM for Education Bridging the understanding gap Active participation in the modern classroom For all children, especially those with hearing difficulties, it s important to hear well in school for

More information

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 보청기의전기음향적성능분석 Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 차례 I. 성능분석이필요한이유 II. 성능분석에필요한장비의구성 III. 성능분석시구성품의위치 IV. 성능분석의기준 V. 성능분석의내용 VI. 성능분석의절차 I. 성능분석이필요한이유 보청기의품질관리보청기의적합 - frequency

More information

The following describes general use of ELIOS device

The following describes general use of ELIOS device General use of the device The following describes general use of ELIOS device 1. Introduction of graphical user interface Presence indicator of SD card Charging indicator General Informations Screen title

More information

TRAINING MANUAL AVANT HIT+ Hearing Instrument Test Chamber.

TRAINING MANUAL AVANT HIT+ Hearing Instrument Test Chamber. TRAINING MANUAL AVANT Hearing Instrument Test Chamber HIT+ www.medrx-int.com Contents Getting to know your HIT+... 3 AVANT HIT Software Overview.. 4 AVANT HIT Software Options.... 5 Preparing for Testing...

More information

Carol De Filippo. Viet Nam Teacher Education Institute June 2010

Carol De Filippo. Viet Nam Teacher Education Institute June 2010 Carol De Filippo Viet Nam Teacher Education Institute June 2010 The Problem Noise Distance Reverberation FM Systems Loop Systems Sound Field Systems Classrooms are big, NOISY places! Outdoor Traffic, trains,

More information

Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids

Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids Acoustical Measurements and Instrumentation: Paper ICA2016-388 Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids Zargos Neves Masson (a),

More information

SUMMARY TABLE VOLUNTARY PRODUCT ACCESSIBILITY TEMPLATE

SUMMARY TABLE VOLUNTARY PRODUCT ACCESSIBILITY TEMPLATE Date: 1 August 2009 Voluntary Accessibility Template (VPAT) This Voluntary Product Accessibility Template (VPAT) describes accessibility of Polycom s Polycom CX200, CX700 Desktop IP Telephones against

More information

Digital hearing aids are still

Digital hearing aids are still Testing Digital Hearing Instruments: The Basics Tips and advice for testing and fitting DSP hearing instruments Unfortunately, the conception that DSP instruments cannot be properly tested has been projected

More information

ADHEAR The new bone-conduction hearing aid innovation

ADHEAR The new bone-conduction hearing aid innovation ADHEAR The new bone-conduction hearing aid innovation MED-EL has world-wide launched a new kind of hearing aid, ADHEAR, for people who have an hearing impairment and want to prevent surgery. This little

More information

FONIX FP35. Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Visible Speech

FONIX FP35. Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Visible Speech FONIX FP35 Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Move your hearing aid fittings into the world of the touch screen! The new FP35 Touch adds a new interactive dimension to hearing

More information

binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way

binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way www.bestsound-technology.com binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way Richard Schultz-Amling Sivantos, 2015. 1 Introduction Since the release

More information

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 122 Annex C Edition 1999 (E) Equipment for speech audiometry Annex C Test report format Appareils pour l audiométrie vocale Annexe C Format du rapport d essai OIML R

More information

Validation Studies. How well does this work??? Speech perception (e.g., Erber & Witt 1977) Early Development... History of the DSL Method

Validation Studies. How well does this work??? Speech perception (e.g., Erber & Witt 1977) Early Development... History of the DSL Method DSL v5.: A Presentation for the Ontario Infant Hearing Program Associates The Desired Sensation Level (DSL) Method Early development.... 198 Goal: To develop a computer-assisted electroacoustic-based procedure

More information

Summary Table: Voluntary Product Accessibility Template

Summary Table: Voluntary Product Accessibility Template Date: August 16 th, 2011 Name of Product: Cisco Unified Wireless IP Phone 7921G, 7925G, 7925G-EX and 7926G Contact for more information: Conrad Price, cprice@cisco.com Summary Table: Voluntary Product

More information

Software Training Manual

Software Training Manual Software Training Manual Copyright June 2004 Avant Training Manual Rev 1 Effective Date June 1, 2004 INTRODUCTION Congratulations! Your decision to incorporate the MedRx Avant into your daily practice

More information

Presenter. Community Outreach Specialist Center for Sight & Hearing

Presenter. Community Outreach Specialist Center for Sight & Hearing Presenter Heidi Adams Heidi Adams Community Outreach Specialist Center for Sight & Hearing Demystifying Assistive Listening Devices Based on work by Cheryl D. Davis, Ph.D. WROCC Outreach Site at Western

More information

SUMMARY TABLE VOLUNTARY PRODUCT ACCESSIBILITY TEMPLATE

SUMMARY TABLE VOLUNTARY PRODUCT ACCESSIBILITY TEMPLATE Date: 2 November 2010 Updated by Alan Batt Name of Product: Polycom CX600 IP Phone for Microsoft Lync Company contact for more Information: Ian Jennings, ian.jennings@polycom.com Note: This document describes

More information

personalization meets innov ation

personalization meets innov ation personalization meets innov ation Three products. Three price points. Premium innovations all around. Why should a truly personalized fit be available only in a premium hearing instrument? And why is it

More information

The Effect of Analysis Methods and Input Signal Characteristics on Hearing Aid Measurements

The Effect of Analysis Methods and Input Signal Characteristics on Hearing Aid Measurements The Effect of Analysis Methods and Input Signal Characteristics on Hearing Aid Measurements By: Kristina Frye Section 1: Common Source Types FONIX analyzers contain two main signal types: Puretone and

More information

Summary Table Voluntary Product Accessibility Template. Supports. Please refer to. Supports. Please refer to

Summary Table Voluntary Product Accessibility Template. Supports. Please refer to. Supports. Please refer to Date Aug-07 Name of product SMART Board 600 series interactive whiteboard SMART Board 640, 660 and 680 interactive whiteboards address Section 508 standards as set forth below Contact for more information

More information

CROS System Initial Fit Protocol

CROS System Initial Fit Protocol CROS System Initial Fit Protocol Our wireless CROS System takes audio from an ear level microphone and wirelessly transmits it to the opposite ear via Near-Field Magnetic Induction (NFMI) technology, allowing

More information

iclicker2 Student Remote Voluntary Product Accessibility Template (VPAT)

iclicker2 Student Remote Voluntary Product Accessibility Template (VPAT) iclicker2 Student Remote Voluntary Product Accessibility Template (VPAT) Date: May 22, 2017 Product Name: i>clicker2 Student Remote Product Model Number: RLR14 Company Name: Macmillan Learning, iclicker

More information

Procedures for the setting up of fm radio systems for use with hearing aids

Procedures for the setting up of fm radio systems for use with hearing aids Procedures for the setting up of fm radio systems for use with hearing aids David Evans david@connevans.com Please send an email to fmadvantage@connevans.com if you would like to receive updates and associated

More information

Summary Table Voluntary Product Accessibility Template. Supports. Not Applicable. Not Applicable- Not Applicable- Supports

Summary Table Voluntary Product Accessibility Template. Supports. Not Applicable. Not Applicable- Not Applicable- Supports PLANTRONICS VPAT 1 Product: Telephony Call Center Hearing Aid Compatible (HAC) Headsets Summary Table Section 1194.21 Software Applications and Operating Systems Section 1194.22 Web-based internet information

More information

#0086.

#0086. USER MANUAL Contents Introduction... 3 AVANT POLAR HIT Software Overview.. 4 AVANT POLAR HIT Software Options. 5 Preparing for Testing.. 7 Rotational Table Control.. 10 Making HIT Measurements 11 Sample

More information

Radio Aid Systems. 14 School &

Radio Aid Systems. 14 School & Radio Aid Systems Radio Aid Systems from Connevans For children and young people with a hearing loss, listening in noisy and poor acoustic environments is particularly challenging. It is difficult, tiring

More information

Quick Guide Binaural REM

Quick Guide Binaural REM Quick Guide Binaural REM The purpose of this document is to provide a quick guide for the Binaural REM feature found in the REM440 Real-Ear Measurement module in the Affinity 2.0 and Callisto Suites. This

More information

Summary Table Voluntary Product Accessibility Template

Summary Table Voluntary Product Accessibility Template PLANTRONICS VPAT 4 Product: Wireless Headsets (Non-HAC) Operated with dedicated Base assembly Over the Ear Voice Tube: CS70, CS70/HL10 Lifter Over the Ear Noise-Canceling: 510S, 510S/HL10 Lifter Summary

More information

Apple emac. Standards Subpart Software applications and operating systems. Subpart B -- Technical Standards

Apple emac. Standards Subpart Software applications and operating systems. Subpart B -- Technical Standards Apple emac Standards Subpart 1194.21 Software applications and operating systems. 1194.22 Web-based intranet and internet information and applications. 1194.23 Telecommunications products. 1194.24 Video

More information

Product Model #:ASTRO Digital Spectra Consolette W7 Models (Local Control)

Product Model #:ASTRO Digital Spectra Consolette W7 Models (Local Control) Subpart 1194.25 Self-Contained, Closed Products When a timed response is required alert user, allow sufficient time for him to indicate that he needs additional time to respond [ N/A ] For touch screen

More information

Hearing Solutions Catalog Winter Supplement January 2009

Hearing Solutions Catalog Winter Supplement January 2009 Hearing Solutions Catalog Winter Supplement January 2009 LX Virtue The following information is a supplement containing information for products released January 1, 2009. It should be used along with

More information

Phonak Baseo Q (Q15) Product information

Phonak Baseo Q (Q15) Product information Phonak Baseo Q (Q15) TM Product information Phonak Baseo Q15 is the comprehensive four-channel, entry-level BTE family for mild to profound hearing losses, available in four models M, P, SP and UP. Powered

More information

Summary Table Voluntary Product Accessibility Template. Supporting Features. Supports. Supports. Supports. Supports

Summary Table Voluntary Product Accessibility Template. Supporting Features. Supports. Supports. Supports. Supports Date: March 31, 2016 Name of Product: ThinkServer TS450, TS550 Summary Table Voluntary Product Accessibility Template Section 1194.21 Software Applications and Operating Systems Section 1194.22 Web-based

More information

RECD-Measurements with Connexx 6.3 and Unity 2

RECD-Measurements with Connexx 6.3 and Unity 2 RECD-Measurements with Connexx 6.3 and Unity 2 Vera Spitzlei, Corporate Audiology Siemens Audiologische Technik GmbH For internal use only / Copyright Siemens AG 2009. All rights reserved. Overview Why

More information

Voluntary Product Accessibility Template (VPAT)

Voluntary Product Accessibility Template (VPAT) Voluntary Product Accessibility Template (VPAT) Date: January 25 th, 2016 Name of Product: Mitel 6730i, 6731i, 6735i, 6737i, 6739i, 6753i, 6755i, 6757i, 6863i, 6865i, 6867i, 6869i, 6873i Contact for more

More information

Section 5 - The use of personal FM systems with soundfield systems

Section 5 - The use of personal FM systems with soundfield systems Section 5 - The use of personal FM systems with soundfield systems QS13 Where soundfield systems are used in conjunction with personal FM systems, equipment must be selected and set up to ensure that the

More information

Product Model #: Digital Portable Radio XTS 5000 (Std / Rugged / Secure / Type )

Product Model #: Digital Portable Radio XTS 5000 (Std / Rugged / Secure / Type ) Rehabilitation Act Amendments of 1998, Section 508 Subpart 1194.25 Self-Contained, Closed Products The following features are derived from Section 508 When a timed response is required alert user, allow

More information

Roger TM Plug in, turn on and teach. Dynamic SoundField

Roger TM Plug in, turn on and teach. Dynamic SoundField Roger TM Plug in, turn on and teach Dynamic SoundField Why soundfield matters Hearing well in a classroom setting is important for students to truly excel and have the best learning experience. Modern

More information

Avaya IP Office 10.1 Telecommunication Functions

Avaya IP Office 10.1 Telecommunication Functions Avaya IP Office 10.1 Telecommunication Functions Voluntary Product Accessibility Template (VPAT) Avaya IP Office is an all-in-one solution specially designed to meet the communications challenges facing

More information

Two Modified IEC Ear Simulators for Extended Dynamic Range

Two Modified IEC Ear Simulators for Extended Dynamic Range Two Modified IEC 60318-4 Ear Simulators for Extended Dynamic Range Peter Wulf-Andersen & Morten Wille The international standard IEC 60318-4 specifies an occluded ear simulator, often referred to as a

More information

Communications Accessibility with Avaya IP Office

Communications Accessibility with Avaya IP Office Accessibility with Avaya IP Office Voluntary Product Accessibility Template (VPAT) 1194.23, Telecommunications Products Avaya IP Office is an all-in-one solution specially designed to meet the communications

More information

iclicker+ Student Remote Voluntary Product Accessibility Template (VPAT)

iclicker+ Student Remote Voluntary Product Accessibility Template (VPAT) iclicker+ Student Remote Voluntary Product Accessibility Template (VPAT) Date: May 22, 2017 Product Name: iclicker+ Student Remote Product Model Number: RLR15 Company Name: Macmillan Learning, iclicker

More information

Set-up Instructions. MANUFACTURED BY COMMUNICATIONS-APPLIED TECHNOLOGY CO., INC. RESTON, VA. CAGE CODE: 0EEY2

Set-up Instructions. MANUFACTURED BY COMMUNICATIONS-APPLIED TECHNOLOGY CO., INC. RESTON, VA. CAGE CODE: 0EEY2 Set-up Instructions C-AT Level A/B Radio Interface with: Push-to-Talk Assembly, Shaped Audio Throat Microphone Bi-aural Speaker Assembly (two configurations) MANUFACTURED BY COMMUNICATIONS-APPLIED TECHNOLOGY

More information

Personal Listening Solutions. Featuring... Digisystem.

Personal Listening Solutions. Featuring... Digisystem. Personal Listening Solutions Our Personal FM hearing systems are designed to deliver several important things for pupils and students with a hearing impairment: Simplicity to set up and to use on a daily

More information

EUHA Guideline Wireless remote microphone systems configuration, verification and measurement of individual benefit

EUHA Guideline Wireless remote microphone systems configuration, verification and measurement of individual benefit EUHA Guideline Wireless remote microphone systems configuration, verification and measurement of individual benefit Guideline 04-06 - v1.0 - issued 9 May 2017 Audiology Expert Group Beate Gromke (Chairwoman),

More information

Verification of soft speech amplification in hearing aid fitting: A comparison of methods

Verification of soft speech amplification in hearing aid fitting: A comparison of methods Verification of soft speech amplification in hearing aid fitting: A comparison of methods Sarah E. Dawkins, B.A. AuD Research Project April 5, 2007 University of Memphis Project Advisor Robyn M. Cox, PhD.

More information

Digital. hearing instruments have burst on the

Digital. hearing instruments have burst on the Testing Digital and Analog Hearing Instruments: Processing Time Delays and Phase Measurements A look at potential side effects and ways of measuring them by George J. Frye Digital. hearing instruments

More information

Beltone Electronics 2601 Patriot Boulevard Glenview, IL U.S.A. (800)

Beltone Electronics 2601 Patriot Boulevard Glenview, IL U.S.A. (800) Beltone Electronics 261 Patriot Boulevard Glenview, IL 626 U.S.A. (8) 621-1275 Beltone Canada 31 Supertest Road Toronto, Ontario M3J 2M4 Canada (416) 736-4444 www.beltone.com 21, Beltone Electronics 521993

More information

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system The information in this document contains general descriptions of the technical options available, which do not always have to be present in individual cases and are subject to change without prior notice.

More information

Voluntary Product Accessibility Template (VPAT)

Voluntary Product Accessibility Template (VPAT) (VPAT) Date: Product Name: Product Version Number: Organization Name: Submitter Name: Submitter Telephone: APPENDIX A: Suggested Language Guide Summary Table Section 1194.21 Software Applications and Operating

More information

Note: This document describes normal operational functionality. It does not include maintenance and troubleshooting procedures.

Note: This document describes normal operational functionality. It does not include maintenance and troubleshooting procedures. Date: 18 Nov 2013 Voluntary Accessibility Template (VPAT) This Voluntary Product Accessibility Template (VPAT) describes accessibility of Polycom s C100 and CX100 family against the criteria described

More information

Date: April 19, 2017 Name of Product: Cisco Spark Board Contact for more information:

Date: April 19, 2017 Name of Product: Cisco Spark Board Contact for more information: Date: April 19, 2017 Name of Product: Cisco Spark Board Contact for more information: accessibility@cisco.com Summary Table - Voluntary Product Accessibility Template Criteria Supporting Features Remarks

More information

DSM PRO. Software Training Manual. Copyright November 2003

DSM PRO. Software Training Manual. Copyright November 2003 DSM PRO Software Training Manual Copyright November 2003 DSM PRO Training Manual Rev 1 Effective November 2003 INTRODUCTION Congratulations! Your decision to incorporate the MedRx DSM PRO into your daily

More information

Summary Table Voluntary Product Accessibility Template. Supporting Features. Not Applicable. Supports. Not Applicable. Supports

Summary Table Voluntary Product Accessibility Template. Supporting Features. Not Applicable. Supports. Not Applicable. Supports Page 1 of 5 PLANTRONICS Over the Head USB Noise cancelling Stereo: Operated with DA60 In-line Stereo QD W/ D261N Over the Head USB Noise cancelling Mono/ Binaural: Operated with DA60 In-line Mono QD W/

More information

Note: This document describes normal operational functionality. It does not include maintenance and troubleshooting procedures.

Note: This document describes normal operational functionality. It does not include maintenance and troubleshooting procedures. Date: 26 June 2017 Voluntary Accessibility Template (VPAT) This Voluntary Product Accessibility Template (VPAT) describes accessibility of Polycom s CX5100 Unified Conference Station against the criteria

More information

Voluntary Product Accessibility Template (VPAT)

Voluntary Product Accessibility Template (VPAT) Avaya Vantage TM Basic for Avaya Vantage TM Voluntary Product Accessibility Template (VPAT) Avaya Vantage TM Basic is a simple communications application for the Avaya Vantage TM device, offering basic

More information

MedRx HLS Plus. An Instructional Guide to operating the Hearing Loss Simulator and Master Hearing Aid. Hearing Loss Simulator

MedRx HLS Plus. An Instructional Guide to operating the Hearing Loss Simulator and Master Hearing Aid. Hearing Loss Simulator MedRx HLS Plus An Instructional Guide to operating the Hearing Loss Simulator and Master Hearing Aid Hearing Loss Simulator The Hearing Loss Simulator dynamically demonstrates the effect of the client

More information

Introduction. What s new in Genie 2? New Fitting features. Support for new hearing aids:

Introduction. What s new in Genie 2? New Fitting features. Support for new hearing aids: Guide 207. Introduction Genie 2 and Genie have been updated with a variety of new features and enhancements. Please find a quick overview of these below: What s new in Genie 2? Support for new hearing

More information

Summary Table Voluntary Product Accessibility Template

Summary Table Voluntary Product Accessibility Template PLANTRONICS VPAT 3 Product: Wireless Hearing Aid Compatible (HAC) Headsets Operated with dedicated Base assembly Over the Head Voice Tube: Pulsar 590, CS351 Over the Head Noise-Canceling: CS351N Convertible

More information

2.0. Desktop Fitting Guide getting started. Preparation of the hearing instruments

2.0. Desktop Fitting Guide getting started. Preparation of the hearing instruments Desktop Fitting Guide getting started This Guide provides you a detailed introduction to hearing instrument fitting with Phonak Target 2.0. Please also find the [News] in the Phonak Target start screen.

More information

Suitcase Lab Measurement of Digital Cellphone Interference Levels on Hearing Aids

Suitcase Lab Measurement of Digital Cellphone Interference Levels on Hearing Aids J Am Acad Audiol 12 : 281-285 (2001) Suitcase Lab Measurement of Digital Cellphone Interference Levels on Hearing Aids Mead C. Killion* Harry Teder' Russ Thoma* Abstract A low-cost, "real-life" method

More information

Electro-Acoustic Stimulation (EAS) Naída CI Q90 EAS System. Naída CI Q90 EAS System Components. Naída CI Q90 Acoustic Earhook

Electro-Acoustic Stimulation (EAS) Naída CI Q90 EAS System. Naída CI Q90 EAS System Components. Naída CI Q90 Acoustic Earhook Electro-Acoustic Stimulation (EAS) EAS is a general term that refers to the simultaneous delivery of electric stimulation (cochlear implant) and acoustic amplification (hearing aid) to the same ear. Naída

More information

Konftel 300Mx. Voluntary Product Accessibility Template (VPAT)

Konftel 300Mx. Voluntary Product Accessibility Template (VPAT) Konftel 300Mx Voluntary Product Accessibility Template (VPAT) The Konftel 300Mx is a sophisticated speakerphone, intended for use by groups of up to ten people in conference room and meeting room environments.

More information

General about Calibration and Service on Audiometers and Impedance Instruments

General about Calibration and Service on Audiometers and Impedance Instruments General about Calibration and Service on Audiometers and Impedance Instruments Calibration and Service of Audiometers - page 1 of 14 Rev.4 2008-05-21/JF General about Calibration and Service on Audiometers

More information

Phonak Fast Facts. Audéo B-Direct

Phonak Fast Facts. Audéo B-Direct Phonak Fast Facts Audéo B-Direct Phonak is introducing Audéo B-Direct, a new extension of the successful RIC hearing aid family for mild to severe hearing loss. Powered by Phonak Belong technology, Phonak

More information

Nucleus CP810 Sound Processor User Guide

Nucleus CP810 Sound Processor User Guide Nucleus CP810 Sound Processor User Guide Symbols Note Important information or advice. Can save inconvenience. Caution (no harm) Special care to be taken to ensure safety and effectiveness. Could cause

More information

Avaya Model 9611G H.323 Deskphone

Avaya Model 9611G H.323 Deskphone Avaya Model 9611G H.323 Deskphone Voluntary Product Accessibility Template (VPAT) The statements in this document apply to Avaya Model 9611G Deskphones only when they are configured with Avaya one-x Deskphone

More information

Custom instruments. Insio primax User Guide. Hearing Systems

Custom instruments. Insio primax User Guide. Hearing Systems Custom instruments Insio primax User Guide Hearing Systems Content Welcome 4 Your hearing instruments 5 Instrument type 5 Getting to know your hearing instruments 5 Components and names 6 Controls 8 Settings

More information

AMBCO AUDIOMETER

AMBCO AUDIOMETER Model 1000+ User Manual AMBCO 1000+ AUDIOMETER AMBCO ELECTRONICS 15052 REDHILL AVE SUITE #D TUSTIN, CA 92780 (714) 259-7930 FAX (714) 259-1688 WWW.AMBCO.COM 10-1003, Rev. A DCO 17 008, 11 13 17 Model 1000+

More information

Neuro-Audio Version 2010

Neuro-Audio Version 2010 ABR PTA ASSR Multi-ASSR OAE TEOAE DPOAE SOAE ECochG MLR P300 Neuro-Audio Version 2010 one device for all audiological tests Auditory brainstem response (ABR)/Brainstem evoked response audiometry (BERA)

More information

SOFTWARE INSTRUCTION MANUAL

SOFTWARE INSTRUCTION MANUAL Audiometer SOFTWARE INSTRUCTION MANUAL MedRx AVANT Audiometer Effective August 2005 Software Training Manual Rev 1 TABLE OF CONTENTS TABLE OF CONTENTS... 2 GETTING STARTED... 4 INSTALLATION SOFTWARE...

More information

CLASSROOM AMPLIFICATION: WHO CARES? AND WHY SHOULD WE? James Blair and Jeffery Larsen Utah State University ASHA, San Diego, 2011

CLASSROOM AMPLIFICATION: WHO CARES? AND WHY SHOULD WE? James Blair and Jeffery Larsen Utah State University ASHA, San Diego, 2011 CLASSROOM AMPLIFICATION: WHO CARES? AND WHY SHOULD WE? James Blair and Jeffery Larsen Utah State University ASHA, San Diego, 2011 What we know Classroom amplification has been reported to be an advantage

More information

The Devil is in the Fitting Details. What they share in common 8/23/2012 NAL NL2

The Devil is in the Fitting Details. What they share in common 8/23/2012 NAL NL2 The Devil is in the Fitting Details Why all NAL (or DSL) targets are not created equal Mona Dworsack Dodge, Au.D. Senior Audiologist Otometrics, DK August, 2012 Audiology Online 20961 What they share in

More information

Phonak Wireless Communication Portfolio Product information

Phonak Wireless Communication Portfolio Product information Phonak Wireless Communication Portfolio Product information The Phonak Wireless Communications Portfolio offer great benefits in difficult listening situations and unparalleled speech understanding in

More information

Oscilla - excellence since 1960

Oscilla - excellence since 1960 Oscilla - excellence since 1960 Oscilla USB310 / 330 PC-based Screening Audiometer 11 frequencies: 125-8000 Hz (selectable in USB330) Left, right and binaural Pulse tone Warble tone (USB330) 4 automatic

More information

Providing Effective Communication Access

Providing Effective Communication Access Providing Effective Communication Access 2 nd International Hearing Loop Conference June 19 th, 2011 Matthew H. Bakke, Ph.D., CCC A Gallaudet University Outline of the Presentation Factors Affecting Communication

More information

AA-M1C1. Audiometer. Space saving is achieved with a width of about 350 mm. Audiogram can be printed by built-in printer

AA-M1C1. Audiometer. Space saving is achieved with a width of about 350 mm. Audiogram can be printed by built-in printer Audiometer AA-MC Ideal for clinical practice in otolaryngology clinic Space saving is achieved with a width of about 350 mm Easy operation with touch panel Audiogram can be printed by built-in printer

More information

Accessibility Standards Mitel MiVoice 8528 and 8568 Digital Business Telephones

Accessibility Standards Mitel MiVoice 8528 and 8568 Digital Business Telephones Accessibility Standards Mitel products are designed with the highest standards of accessibility. Below is a table that outlines how Mitel MiVoice 8528 and 8568 digital business telephones conform to section

More information