The Screw Loosening and Fatigue Analyses of Three Dimensional Dental Implant Model

Size: px
Start display at page:

Download "The Screw Loosening and Fatigue Analyses of Three Dimensional Dental Implant Model"

Transcription

1 The Screw Loosening and Fatigue Analyses of Three Dimensional Dental Implant Model M. Wierszycki 1,2, W. Kąkol 2, T. Łodygowski 1 1 Institute of Structural Engineering, Poznań University of Technology, ul. Piotrowo 5, Poznań, Poland 2 BudSoft, ABAQUS Inc. Representative, ul. Św. Marcin 58/64, Poznań, Poland Abstract: Dental restorations with the application of implants are very effective and commonly used in dental treatment. However, for more than fifty percent of patients, diverse complications can be observed. In some cases, these problems are caused by mechanical reasons such as loosening of the retaining screws or fracture and cracking of the dental implant components. We would like to focus on the complex biomechanical relations in dental implant simulations. A cyclic scheme of physiological occlusal loading and fatigue changes of dental material, bone loss phenomenon and changeability of boundary conditions are investigated. The implant structure poses a purely mechanical problem, but the loads and boundary conditions have already more complex, biomechanical grounds. The changes of the bone, especially bone loss phenomenon, have a great influence on implant behavior, stress distribution and therefore fatigue damage. The computer simulations of implant behavior employing of ABAQUS and fe-safe software enable us to consider all these phenomena. For the full simulation of implant structure behavior, a 3D model, which includes a spiral thread, is necessary. Only the 3D model allows a full simulation of kinematics of the implant, describing the multiaxial state of stress and, in consequence, the possibility of screw loosening. In the case of screw loosening simulation, the modeling of tightening is a crucial task. It is carried out in such a way that it describes a real physical process. The friction dissipated energy has been used as a measure of the retaining screw loosening resistance. The robustness, exactness and efficiency of complex contact analyses carried out with ABAQUS enables us to obtain valuable results and also serve for practical applications of these results in modern prosthodontics. Keywords: fatigue analysis, biomechanics, dental implant 1. Introduction Nowadays, the finite element method plays an important role in solving engineering problems in many fields of science and industry and can be successfully applied also in simulations of biomechanical systems (Będziński, 1997, Geng et al., 2001, Sakaguchi et al. 1993). The method 2006 ABAQUS Users Conference 527

2 is a well established one, used in biomechanical simulations for over 20 years so far. It allows taking into consideration the key features like material inhomogeneity and anisotropic mechanical properties of tissues as well as a very complicated geometry of human body parts (Swatrz et al., 1991). Furthermore, it is proved (Hędzelek et al., 2003) that function, failure, prediction of changes and remodeling of the biostructures are related to stress and strain fields in tissues and they may be calculated with the help of FEA. Thus, FEA is an efficient tool for testing biomechanical sets, like dental implants, but it is still often very difficult to obtain useful and valuable results of these kinds of problems. The main reason for this is the complexity of biostructures and the complexity of numerical simulations stemming from that. 2. Models of dental implant Implants are commonly applied treatment method of dental restorations. Unfortunately, numerous clinical observations point to the occurrence of both early and late complications. In many cases, these problems are caused by mechanical fractures of the implants themselves (Hędzelek et al., 2003, Kąkol et al., 2002). The most frequent complications are loosening of the connecting screw, fracture and cracking of the dental implant parts. While loosening of the connecting screw causes mostly a patient s discomfort in the implant usage, cracking leads to much more serious complications and makes further treatment extremely difficult. To understand the reasons of observed mechanical complications, it is necessary to know stress and strain fields in implant components as well as changes in boundary conditions. The cyclic loads and the character of the fracture indicate material fatigue as the basic cause of this fracture (Kocańda, 1985). The authors would like to confirm this proposition by means of numerical simulations of a dental implant system (Kąkol et al., 2003). The simulations were carried out with ABAQUS/Standard finite element program. For the fatigue calculations the fe-safe program was used (Safe Technology Ltd). a) b) Figure 1. Models of dental implant: a) axisymmetric with part of jaw bone, b) three dimensional with spiral thread ABAQUS Users Conference

3 In the analysis of a dental implant, the implant structure is not trivial, however it is a purely mechanical problem. The loads and boundary conditions have already more complex, biomechanical reasons. The most significant complexities of finite element stress-strain and fatigue analyses are geometry and mesh preparation, fatigue material characteristics, implant assembly, physiological changes of loads and finally definition of boundary conditions describing it as a bone. 2.1 Geometry The numerical models were created on the basis of technical documentation of the commercial implantological system OSTEOPLANT. It is a commonly used system, consisting of an implant and abutment with a nonrotational hexagonal connection, assembled by a screw. For the full simulation of implant structure behavior, the geometrically complex three dimensional model is necessary. This model, which includes a spiral thread, enables taking into consideration a few important aspects such as full simulation of kinematics of an implant, describing the multiaxial state of stress and, the most important, the possibility of simulation of screw loosening. The most interesting result will be the relationship between torque moment, friction coefficient and loosening or fatigue life of the screw under cyclic loads. Unfortunately, a three-dimensional model is very large (ca. 500K dof). Due to the fact that most parts of an implant are axisymmetric, an axisymmetric concept of modeling is a good idea of simplification. ABAQUS/Standard offers axisymmetric solid CAXA elements, which are capable of modeling nonlinear asymmetric deformation, as well. The CAXA elements are intended for an analysis of hollow bodies, such as pipes, but they may also be used to model solid bodies with some limitations. These elements use standard isoparametric interpolation in the radial symmetry axis plane, combined with the Fourier interpolation with respect to the angle of revolution. The asymmetric deformation is assumed to be symmetric with respect to the radial symmetry axis plane at an angle equal 0 or π (ABAQUS Manuals, 2005). An axisymmetric model of an implant was created with the application of CAXA elements (Fig. 1). This approach reduces the geometry description of an implant model from a three dimensional to a two dimensional one. The threads of screw and implant body were simplified to axisymmetric, parallel rings. Thanks to this concept, the size of the problem (ca. 75K dof) and the cost of the calculation were significantly reduced. 2.2 Material properties The part of jaw is composed of two kinds of bones: the cancellous bone and the cortical one (Fig. 1). The problem of describing physical law of a bone is very complex. The mechanical characteristics and internal microstructure of cortical and cancellous bones are nonhomogeneous, anisotropic and variable in time. The changes of bone characteristics are caused by the phenomenon of remodeling tissues. It is very difficult to take these aspects into consideration in implant models. In finite element analysis, many concepts of description of mechanical properties of a bone could be applied. Starting from the very simple, isotropic, going on to the more complicated, transversely isotropic or orthotropic and ending with a very complex, anisotropic ones. The assumed material characteristics of the jaw bones are linear, homogeneous and isotropic (Tab. 3). This simplification is justifiable due to the role which the bone plays in fatigue analysis of an implant. The most important here is the influence of boneloss around an implant as well as bone flexibility on implant boundary conditions and implant fatigue life (Hędzelek et al., 2003) ABAQUS Users Conference 529

4 The implant is made of medical titanium alloy, the mechanical properties of which are nonlinear. Their description was based on the certificate of conformity (ASTM F136-98, ISO 5832 PT 2-93). Table 1. Material properties of implant model components. Implant's body (Titanium Grade 4) Screw (6AL-4V-ELI) Abutment (6AL-4V-ELI) Young's modulus [Mpa] Poisson ratio Yield [MPa] 615, ,8 Tensile Yield [MPa] 742, ,4 2.3 Mechanical assembly The implant system is seemingly simple, but in fact it is quite a complex mechanical system (Kąkol et al., 2002, Sakaguchi et al., 1993). An important aspect of implant assembly is the modeling of tightening. For this purpose, it is necessary to define the contact surfaces between a root, an abutment and a screw. The friction characteristic on these surfaces is one of the key parameters influencing preload axial force, reduction of implant components mobility, resistance of screw loosening but also fatigue life of a whole implant. For the friction coefficient, a value ranging between 0.1 (as in a specially finished surface) and 0.5 (as in dry titanium to titanium friction) may be found in literature. In the present analyses, three different friction coefficients (0.1, 0.2 and 0.5) were considered. The friction characteristic is one of the key parameters influencing preload axial fores V, reduction of mobility of implant components and screw loosening resistance. The first step of the simulation is the tightening of the screw. In the case of the axisymmetric model, the middle part of the screw was subjected to temperature loading in order to simulate this. The thermal expansion property of the screw material was orthotropic. It was defined in such a way, that shrinking occurred only along the screw axis. The value of axial force in a tightened screw was calculated from the empirical equation. It is dependent on friction coefficient and torque moment. This force (ranging from 80 to 850 N) was changed into a temperature field (Kąkol et al., 2003). In the case of the three-dimensional model, the simulation of tightening can be defined in such a way that it describes a real physical process. However it creates a very complex contact problem. 2.4 Loads The external loads of an implant model were applied in the second step of the simulation. The values and directions of forces were taken on the grounds of a physiologically proven scheme. To estimate the least favorable distribution of stress, only the maximal realistic occlusal forces were taken into account. The loading of an implant is never axial. The vertical components of it are estimated at 600 N and the horizontal ones at 100 N. For the fatigue calculations, it is necessary to define the character of load changeability in the shape of a curve load-time, the socalled load signal. It is not easy to assume these characteristics and typical values of occlusal forces. In an applied low-cycled scheme of 24-hour loads, the average values were 60 N (Hędzelek et al., 2003, Kąkol et al., 2003) ABAQUS Users Conference

5 2.5 Boundary conditions In the first stage of implant analysis, all degrees of freedom at the bottom part of implant body were fixed. This assumption seemed to have its explanation in dental practice, where no movements of implants under physiological load are acceptable. However, the difference between infinitely stiff fixing and even low flexibility is significant, especially in the cyclic loading and fatigue damage context. In the next stage of implant analysis, the boundary conditions of implants are modeled as a small part of the jaw bone. The geometry of a small part of jaw surrounding the implant is very simplified but it enables us to take into consideration the changes in implant fixing conditions. The changing flexibility of the bone and boneloss phenomenon are also very important, especially because boneloss has a significant influence on implant behavior, stress distribution and therefore fatigue damage. The degree of encasement and osseointegration of the implant may not be 100%. It dependents on bone quality, stresses developed during healing and function, and the location of the implant in the jaw. This percentage may decrease to as low as 50%. This is caused by remodeling of bone phenomenon. In these analyses three levels of osseointegration were considered. In the case of the first level, the implant body is fully fixed in the jaw bone. In the next two, the degree of implant body embedding decreases to 75 and 50% respectively (Fig. 2) (Hędzelek et al., 2003). a) b) c) Figure 2. Boundary conditions of implant model levels of osseointegration: a) 100%, b) 75%, c) 50%. 3. Behavior of screw Full simulation of screw loosening can be performed only using a 3D implant model. Simulation of tightening can be performed in such a way that it describes a real physical process ABAQUS Users Conference 531

6 3.1 Simulation of tightening The simulation of tightening defined in such a way that it describes a real physical process requires a complex contact analysis. If this analysis is performed using the implicit code, as ABAQUS/Standard, a lot of small increments and equilibrium iterations are usually required to reach a converged solution. In some cases it is very difficult to obtain any converged solution. For simulations of screw tightening, the quasi-static analysis using the explicit code, like ABAQUS/ Explicit, can be carried out, alternatively. The results presented on Fig. 3, Fig. 4 and Fig. 5 were obtained from the analyses performed with ABAQUS/Standard. The friction coefficient was assumed equal to 0.2. The analyses were performed using the displacement control technique. The screw was subjected to a rotational displacement equal to 1 radian, which in turn was applied to the reference point of a rigid body defined on the top surface nodes of the screw head. The Fig. 3 shows the time-history of the tightening moment. The expected zig-zag type curve was recorded at the point of applied displacement. Figure 3. Tightening moment [Ncm]. Fig. 4 shows nearly the linear evolution of normal force in the screw during a tightening process while Fig. 5 shows the comparison of total energies involved in the process, namely internal energy (ALLIE), plastic dissipation energy (ALLPD) and energy dissipated through frictional effects (ALLFD) ABAQUS Users Conference

7 Figure 4. Normal force in screw [N] Figure 5. Tightening energy balance [mj] ABAQUS Users Conference 533

8 3.2 Screw loosening phenomenon In the first step of simulation the implant body was tightened. Two values of tightening moment (15 Ncm and 35 Ncm) and two friction coefficients (0.1 and 0.2) were taken into considerations. In the second step of simulation the torsional loading (asymmetric force) was applied to the top of the abutment to initiate a small movement of the implant system. To measure a resistance to screw loosening the frictional dissipation energy accumulated over a whole process was chosen. The energy dissipated by contact friction forces between the contact surfaces is as follows: where: v velocity field t f frictional traction S boundary (contact surfaces) E F = vt f ds S In the Fig. 6 the energy ALLFD is plotted for two different friction coefficients of a screw threads. It is seen that this energy can be a measure of work to be performed during the screw loosening process. Further investigations in this subject will be carried out. Figure 6. Comparison of ALLFD energy for different friction coefficients ABAQUS Users Conference

9 4. Fatigue analysis Fatigue calculations were carried out in the fe-safe program, which uses advanced multiaxial fatigue algorithms incorporating a multiaxial plasticity model to estimate the life of fatigue. These algorithms are based on the stress results obtained from the finite element analysis (Fig. 7), variations in loading, hysteresis loop cycle closure, and cyclic material properties. Elastic stresses from the FEA model are translated into elastic-plastic stresses by means of a biaxial Neuber's rule and cyclic material properties. In order to estimate failure-free term of an implant, a designed life is defined. Fe-safe calculates the factor (FOS Factor Of Strength) by which the stresses at each node can be increased or reduced to give the required life. The above is the most interesting and vivid for our case. During a single analysis concerning each node separately, a 6-stress tensor is used to calculate the principal stresses and strains and their orientation. A stress concentration factor and scale factor are applied at this stage. A rainflow cycle counting algorithm is used to extract fatigue cycles. For biaxial fatigue methods in turn a critical plane procedure is used to calculate the orientation of the most damaged plane at the node (ABAQUS/Safe Manual, Bishop, 2000, Draper, 1999). Fatigue life of an implant screw was calculated for nine separate cases of loading, three cases of boundary condition schemes and three cases of jaw cancellous bone density. For all of these cases, the same cyclic scheme of loading was assumed. A twenty-four-hour changeability scheme was assumed as a signal, while the number of days corresponding to four years was assumed as the number of cycles. The FOS distribution analysis for particular cases indicates the axial forces in the screw and the changes in the scheme of boundary conditions, which have the greatest influence on fatigue changes. For different bone density and at the same time divergent stiffness of boundary conditions, significant differences of stress distributions present in the screw are noticeable. Yet, it does not lead to serious fatigue changes. For axial forces above 600 N, there is a noticeable increase in the areas endangered by fatigue failure. The degree of required stress reduction reaches ca. 30%. In the most unfavorable load case, the maximal axial force value is the result of a high torque moment and a very small friction coefficient on a screw thread. It is important to pay attention to the danger of increasing the tightening forces in an implant screw. In two-part implants, this high tightening force is motivated by biological and medical aspects. However, the increase in torsion moments and decrease in friction coefficients pose a danger for fatigue life of implant components. 4.1 Moment and friction vs. FOS Fatigue life of the implant screw was calculated for nine cases of loading. For all of these cases the same cyclic scheme of loading was assumed.(table 2). A twenty-four-hour changeability scheme was assumed as a signal while the number of days corresponding to four years was assumed as the number of cycles ABAQUS Users Conference 535

10 Table 2. List of load cases and friction coefficient. Load case Torque moment [Ncm] Friction Coefficient (μ) Axial force [N] The FOS distribution analysis for particular cases indicates that fatigue changes for axial forces up to 200 N occur only in the neighborhood of the first thread twist. For stronger forces, however, the notch under the screw head is also a dangerous area. In all of those cases, stresses exceed the save range maximally in 5 10 %. For axial forces above 600 N (loading case 6 and 7) there is a clearly seen increase of fatigue failure endangered areas. Next points of fatigue changes appear. At the same time, the degree of required stress reduction reaches 30 %. In the most unfavorable load case, the maximal axial force value is the result of both the greatest torque moment and the smallest friction coefficient. The results obtained on this stage of fatigue analysis enable us to formulate some estimate of fatigue characteristic of the implant work. It is important to pay attention to the danger of increasing the tightening forces in implant screw. In two-part implants it is motivated by biological and treatment aspects. However, the increase of torsion moments and decrease of friction coefficient poses a danger for fatigue life of implant components. 4.2 Stiffness and levels of osseointegration vs. FOS The jaw bone is undergoing constantly changes processes; during few months the jaw bone can be completely renewed. This process is called as a bone remodeling. At the continuum level we distinguish, based on porosity, cortical bone and trabecular bone. The outer layers of a bone are of cortical type while inner layers of trabecular type with varying density. It can happen that a whole trabecular part is replaced by cortical one during a remodeling process. To choose an appropriate model of jaw bone remodeling is difficult task. Instead of employing any bio-remodeling theory we performed a series of simulations which addressed similar effects on a cortical as well as trabecular jaw bone with varying density and levels of osseointegration and their influence on the fatigue life of the implant system. Four different Young s modulus for a trabecular part were assumed to simulate a change of bone s stiffness, starting with a value close to a real measured one, through a small stiffness (hundreds of MPa) and ending with similar to stiffness of cortical part. The values used in simulations are listed in Table ABAQUS Users Conference

11 Table 3. Stiffness of jaw bone. Bone Young's modulus [Mpa] Cortical all schemes Cancellous 1st scheme nd scheme rd scheme th scheme 690 In the case of overloading a bone is undergoing a resorption process which leads to reduction in bone mineral content and can lead to an osteoporosis disease, as well. It is very difficult to take into account such phenomenon in the fatigue analysis. Again, instead of employing a theory of a removal by a bone itself, to estimate a fatigue life of the implant system, we performed simulations for four different levels of osseointegration. The simulations were performed for four different stiffness of cortical bone. The top level of cortical bone, according to the Fig. 2, was assumed as: a) entirely embedded to a bone (100%) b) 3mm below conical implant surface (75%) c) 6mm below conical implant surface (50%). A choice of levels of osseointegration expressed in assumed geometry (Fig. 2) was based on the following assumptions: a) the first level is a target position after a restoration treatment and was taken in all our previous simulations b) the second level corresponds to such a bone removal which uncover the first outer thread and where hexagonal abutment changes into inner thread c) the third level corresponds to such boundary conditions of a implant body for which bending is acting on its weakest cross-section ABAQUS Users Conference 537

12 a) b) Figure 7. The Mises equivalent stresses distribution for three different levels of osseointegration: a) 100%, b) 75%, c) 50%. c) ABAQUS Users Conference

13 4.3 Verification The confirmation of numerical results is a case report of implant fracture and examination of fractured surface. The mechanical failure occurred in 55-year-old male patient after one-year period of using single crown replacing the first maxillar premolar. The fracture line passed through upper part of implant body and abutment screw. The failure was located transversally to the long axis between smooth and threaded part of the fixture. Radiograph showed also the bone loss down to the third thread of the implant. Scanning electron microscope examination presented fatigue striations, indicating the advancement of the crack front under cyclic loading. Fractography revealed obliteration and incrustation of some of the fracture features probably because of postfracture wear of contacting surfaces as a result of the component remaining joined together by abutment screw. On Fig. 8 (from Scanning Electron Microscope) fracture surface of dental implant is presented. The characteristic of fatigue damage changes like fatigue bands, called striations, can be observed. Fatigue striations are microscopic features on a fatigue fracture surface that identify one propagation cycle of a fatigue crack. It was possible to identify area with plastic deformation corresponding to final fracture. a) b) Figure 8. Fatigue fracture surface of dental implant (SEM pictures): a) 150x, b) 900x. 5. Conclusions FEA simulation of the behavior of a dental implant system was presented. It was shown that computational modeling and 3D simulation using ABAQUS and fe-safe software enable the realistic prediction of its behavior under restoration treatment as well as service loads. For screw loosening simulation, the modeling of tightening is a crucial task. It has to be carried out in such a way that describes a real physical process as much as possible. Only a 3D modeling 2006 ABAQUS Users Conference 539

14 allows a full simulation of kinematics of the implant, describing the multi-axial state of stress and, in consequence, the possibility of screw loosening. On the basis of the results of fatigue analysis, it can be claimed that the material fatigue is the basic reason of the observed complications. The application as a measure of loosening resistance, the energy dissipated through frictional effects, can lead to practical recommendations in prosthodontics. It is hoped that using presented approach an optimal implant shape can be derived further. 6. References 1. ABAQUS Analysis User's Manual, ABAQUS, Inc. Pawtucket, ABAQUS Theory Manual, ABAQUS, Inc. Pawtucket, ABAQUS/Safe User's Manual, HKS, Inc. Pawtucket, Będziński R., Biomechanika inżynierska (in Polish), Oficyna Wydawnicza Politechniki Wrocławskiej, Wrocław, Bishop N. W. M., Sherratt F., Finite Element Based Fatigue Calculations, NAFEMS, Glasgow, Draper J., Modern metal fatigue analysis, HKS, Inc. Pawtucket, Hędzelek W., Zagalak R., Łodygowski T., Wierszycki M., The effect of marginal bone loss and bone density on the risk of late implant components failures, 27th Annual Conf. of the European Prosthodontic Association, Geneva, September 4-6, Kąkol W., Łodygowski T., Wierszycki M., Hędzelek W., Zagalak R., Numerical Analysis of the Behavior of Dental Implant, ABAQUS Users' Conference 2002, Newport, Rhode Island, Koa C-C., Swiftb J. Q., DeLonga R., Douglasa W. H., Kima Y., Anc K-N., Changd C-H., Huangd H-L., An intra-oral hydraulic system for controlled loadingof dental implants, Journal of Biomechanics 35, pp , Kocańda S., Zmęczeniowe pękanie metali (in Polish), Wydawnictwo Naukowo-Techniczne, Warszawa Merz B. R., Hunenbart S., Belser U. C., Mechanics of the implant-abutment connection: an 8- degree taper compared to a butt joint connection, The International Journal of Oral & Maxillofacial Implants, 15 (4), pp , Natali A. N., Pavan P. G., Schileo E., Dettin M., Bagno A., Di Bello C., Dental implant osseo-integration: a coupled biochemical and biomechanical approach, Acta of Bioengineering and Biomechanics, 13th Conference of the European Society of Biomechanics, Vol. 4 Supp. 1, Wrocław, Poland, pp , Patterson E. A., Johns R. B., Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants, The International Journal of Oral & Maxillofacial Implants, 7 (1), pp , ABAQUS Users Conference

15 14. Sakaguchi R. L., Borgersen S. E., Nonlinear finite element contact analysis of dental implant components, The International Journal of Oral & Maxillofacial Implants, 7 (1), pp , Acknowledgements The research was financed from the budget resources meant for science in the years 2005 and The financial support was provided by the Ministry of Science and Information Society Technologies Grant No. 3 T11F , BW /06. A part of the calculations was carried out in the Poznań Supercomputing and Networking Center ABAQUS Users Conference 541

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. Chandrasenan.P 1, Vishnu.G 2, Akshay K Nair 3 1M Tech student, Department of Mechanical

More information

Research Article Hollow Abutment Screw Design for Easy Retrieval in Case of Screw Fracture in Dental Implant System

Research Article Hollow Abutment Screw Design for Easy Retrieval in Case of Screw Fracture in Dental Implant System Hindawi Journal of Healthcare Engineering Volume 2017, Article ID 4842072, 6 pages https://doi.org/10.1155/2017/4842072 Research Article Hollow Abutment Screw Design for Easy Retrieval in Case of Screw

More information

STUDYING THE EFFECT OF THE PRELOAD INDUCED BY SCREW TIGHTENING ON STRESS DISTRIBUTION OF A DENTAL IMPLANT

STUDYING THE EFFECT OF THE PRELOAD INDUCED BY SCREW TIGHTENING ON STRESS DISTRIBUTION OF A DENTAL IMPLANT Belagh & Kashani, 2017 Volume 3 Issue 3, pp.195-211 Date of Publication: 19 th December, 2017 DOI-https://dx.doi.org/10.20319/mijst.2017.33.195211 This paper can be cited as: Belagh, A.B.F & Kashani, H.M.

More information

APPLICATION OF COMPOSITE FRACTURE MECHANICS TO BONE FRACTURE ANALYSIS USING ABAQUS XFEM

APPLICATION OF COMPOSITE FRACTURE MECHANICS TO BONE FRACTURE ANALYSIS USING ABAQUS XFEM APPLICATION OF COMPOSITE FRACTURE MECHANICS TO BONE FRACTURE ANALYSIS USING ABAQUS XFEM Presenter: David Reid Dassault Systemes UK Ltd SIMULIA david.reid@3ds.com 01925 885971 07825 308031 Work by: Xiaoliang

More information

While the high rate of success of dental

While the high rate of success of dental Mechanics of the Implant-Abutment Connection: An 8-Degree Taper Compared to a Butt Joint Connection Beat R. Merz, Dr sc techn, MBA 1 /Stephan Hunenbart, Dipl Eng TU 2 / Urs C. Belser, DMD, Prof Dr med

More information

Dental Implants Resistance: Computational Analysis

Dental Implants Resistance: Computational Analysis Dental Implants Resistance: Computational Analysis Soraia Isabel Correia Ribeiro Soraia.ribeiro@tecnico.ulisboa.pt Instituto Superior Técnico, Lisbon October 2016 Abstract The overall goal of the present

More information

The University of Sydney Slide 1

The University of Sydney Slide 1 The University of Sydney Slide 1 SIMULATION DRIVEN BIOMEDICAL DESIGN Lecture 4 Presented by Dr Paul Wong AMME4981/9981 Semester 1, 2016 The University of Sydney Slide 2 Simulation Types There is more to

More information

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants INTERNATIONAL STANDARD ISO 14801 Second edition 2007-11-15 Dentistry Implants Dynamic fatigue test for endosseous dental implants Art dentaire Implants Essai de fatigue dynamique pour implants dentaires

More information

Dental Implants: Hex or Conical Connection?

Dental Implants: Hex or Conical Connection? Dental Implants: Hex or Conical Connection? ARDII Inaugural Global Symposium Toronto, May 17-19, 2018 Samantha Webster & Dr. Kornel Ehmann Northwestern University Department of Mechanical Engineering Dr.

More information

(c) B. Ravi, IIT Bombay 1

(c) B. Ravi, IIT Bombay 1 Collaborative Engineering Evaluation, Testing TKP Safety, TKP System Mechanism FEA Physical testing Fatigue Wear What How Evaluate Testing Judgement Criteria's Testing Conditions Parameters Measured OrthoCAD

More information

Finite-element study of vibration effect to fracture healing of a human tibia

Finite-element study of vibration effect to fracture healing of a human tibia Finite-element study of vibration effect to fracture healing of a human tibia Leonid Maslov 1, Jean-Baptiste Etheve 2, Nikolay Sabaneev 3 Ivanovo State Power Engineering University, Ivanovo, Russia 1 Corresponding

More information

The fundamental biomechanics of oral implants. Dr. Szűcs Attila

The fundamental biomechanics of oral implants. Dr. Szűcs Attila The fundamental biomechanics of oral implants Dr. Szűcs Attila Physical elements I. Mechanics solving of technical problems by physical laws»» investigation of prosthesis and the connection to abutments

More information

Anuradha Borse 1* Rubina Tabassum 2 Omkar Shetty 3

Anuradha Borse 1* Rubina Tabassum 2 Omkar Shetty 3 Research AHB Article Adv Hum Biol 2014; 4(2):22-31. A Comparative Evaluation of Stress Distribution at the Implant Abutment Interface between Platforms Switched Morse Taper and Pseudo Platform Switched

More information

Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element analysis

Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element analysis ORIGINAL ARTICLE https://doi.org/15125/jkaoms.2018.44.2.59 pissn 224-7550 eissn 224-590 Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element

More information

A DAMAGE/REPAIR MODEL FOR ALVEOLAR BONE REMODELING

A DAMAGE/REPAIR MODEL FOR ALVEOLAR BONE REMODELING A DAMAGE/REPAIR MODEL FOR ALVEOLAR BONE REMODELING M. Mengoni 1 and J.-P. Ponthot 2 1. ABSTRACT Tooth movements obtained through orthodontic appliances result from a complex biochemical process of bone

More information

NobelActive. The NobelActive technical story

NobelActive. The NobelActive technical story NobelActive The NobelActive technical story The NobelActive technical story NobelActive origins The life cycle of NobelActive began following years of research and development on a self-drilling and bone-condensing

More information

Influence of Different Angulations of Force in Stress Distribution in Implant Retained Finger Prosthesis: A Finite Element Study

Influence of Different Angulations of Force in Stress Distribution in Implant Retained Finger Prosthesis: A Finite Element Study World Journal of Medical Sciences 10 (2): 217-221, 2014 ISSN 1817-3055 IDOSI Publications, 2014 DOI: 10.5829/idosi.wjms.2014.10.2.82275 Influence of Different Angulations of Force in Stress Distribution

More information

Topology optimisation of hip prosthesis to reduce stress shielding

Topology optimisation of hip prosthesis to reduce stress shielding Computer Aided Optimum Design in Engineering IX 257 Topology optimisation of hip prosthesis to reduce stress shielding S. Shuib 1, M. I. Z. Ridzwan 1, A. Y. Hassan 1 & M. N. M. Ibrahim 2 1 School of Mechanical

More information

Modelling of temporomandibular joint and FEM analysis

Modelling of temporomandibular joint and FEM analysis Acta of Bioengineering and Biomechanics Vol. 8, No. 1, 2006 Modelling of temporomandibular joint and FEM analysis MARTINA FRIOVÁ, ZDENK HORÁK, SVATAVA KONVIKOVÁ Laboratory of Biomechanics, Department of

More information

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties Lin Chen, Shen Jingfeng & Chen Bing School of mechanical engineering, University of Shanghai for Science & Technology,

More information

Assessing potential cracking zones in embankment dams

Assessing potential cracking zones in embankment dams Southern Cross University epublications@scu 23rd Australasian Conference on the Mechanics of Structures and Materials 2014 Assessing potential cracking zones in embankment dams Ke He University of New

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 4,100 116,000 120M Open access books available International authors and editors Downloads Our

More information

An effort is made to analyse the stresses experienced by the human femur. In order

An effort is made to analyse the stresses experienced by the human femur. In order Finite Element Analysis of Human Femur Bone U N Mughal 1, H A Khawaja 2*, M Moatamedi 1, M Souli 3 1. Narvik University College, Norway 2. UiT-The Arctic University of Norway, Norway 3. University of Lille,

More information

6 th International Conference on Trends in Agricultural Engineering 7-9 September 2016, Prague, Czech Republic

6 th International Conference on Trends in Agricultural Engineering 7-9 September 2016, Prague, Czech Republic MEASUREMENT OF PRESSURE DISTRIBUTION IN KNEE JOINT REPLACEMENT J. Volf 1, V. Novák 1, V. Ryzhenko 1, D. Novák 2 1 Faculty of Engineering, Czech University of Life Sciences Prague, Czech Republic 2 Faculty

More information

Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant

Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant International Journal of Mechanical & Mechatronics Engineering IJMME-IJENS Vol:15 No:06 84 Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant

More information

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013 Finite Element Analysis of Radius and Ulna Eli Pavlatos April 24, 2013 Outline Review bone biomechanics bone behavior during fracture Stress analysis concepts properties of bone for models Fracture simulations

More information

Biomechanics of Fractures and Fixation

Biomechanics of Fractures and Fixation Biomechanics of Fractures and Fixation Theodore Toan Le, MD Original Author: Gary E. Benedetti, MD; March 2004 New Author: Theodore Toan Le, MD; Revised October 09 Basic Biomechanics Material Properties

More information

Effect of Screw Profile on Stress Distribution Pattern of Dental Implants by FEA

Effect of Screw Profile on Stress Distribution Pattern of Dental Implants by FEA ISSN 2395-1621 Effect of Screw Profile on Stress Distribution Pattern of Dental Implants by FEA #1 RaiphaleGovind, #2 Shinde B.M. 1 govindaraiphale@gmail.com*1 2bharat.shinde@zealeducation.com 1 (M.E Student

More information

Summer Workshop of Applied Mechanics. Theoretical Analysis of Total Hip Joint Replacement of Sandwich design by FEM

Summer Workshop of Applied Mechanics. Theoretical Analysis of Total Hip Joint Replacement of Sandwich design by FEM Summer Workshop of Applied Mechanics June 2002 Department of Mechanics Faculty of Mechanical Engineering Czech Technical University in Prague Theoretical Analysis of Total Hip Joint Replacement of Sandwich

More information

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis 1 Research Article Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis Aswin Yodrux* Department of Materials Handling and Logistics for Engineering,

More information

Stress Analysis of an Endosseus Dental Implant by BEM and FEM

Stress Analysis of an Endosseus Dental Implant by BEM and FEM Send Orders of Reprints at reprints@benthamscience.org The Open Mechanical Engineering Journal, 2012, 6, 115-124 115 Stress Analysis of an Endosseus Dental Implant by BEM and FEM R. Citarella *,1, E. Armentani

More information

Abstract. Keywords: Fatigue Failure, Surface Roughness, Fatigue Strength Factor, Bio-Compatible Dental Implant, Electron Beam Melting (EBM)

Abstract. Keywords: Fatigue Failure, Surface Roughness, Fatigue Strength Factor, Bio-Compatible Dental Implant, Electron Beam Melting (EBM) THE NUMERICAL MODELING OF FATIGUE PROPERTIES OF A BIO- COMPATIBLE DENTAL IMPLANT PRODUCED BY ELECTRON BEAM MELTING (EBM) M. Jamshidinia 1, F. Kong 1, R. Kovacevic 1, * Research Center for Advanced Manufacturing

More information

STRESS ANALYSIS OF DENTAL IMPLANT INSERTED IN THE MANDIBLE

STRESS ANALYSIS OF DENTAL IMPLANT INSERTED IN THE MANDIBLE Journal of MECHANICAL ENGINEERING Strojnícky časopis, VOL 68 (2018), NO 1, 25-32 STRESS ANALYSIS OF DENTAL IMPLANT INSERTED IN THE MANDIBLE ŁAGODA Agnieszka 1, NIESŁONY Adam 2 1 Faculty of Mechanical Engineering,

More information

ARCHIVES OF CIVIL ENGINEERING, LX, 2, 2014

ARCHIVES OF CIVIL ENGINEERING, LX, 2, 2014 DOI: 10.2478/ace-2014-0012 ARCHIVES OF CIVIL ENGINEERING, LX, 2, 2014 NUMERICAL PREDICTION OF THE EFFECTS OF MINIATURE CHANNEL SHAPED SCRATCHES ON THE FRACTURE BEHAVIOUR OF WIRES FOR CIVIL ENGINEERING

More information

Advanced FE Modeling of Absorbable PLLA Screws

Advanced FE Modeling of Absorbable PLLA Screws Advanced FE Modeling of Absorbable PLLA Screws Jorgen Bergstrom, Ph.D., David Quinn, Ph.D., Eric Schmitt jbergstrom@veryst.com, LLC September 14, 2011 Introduction Anterior cruciate ligament (ACL) reconstruction

More information

Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method

Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method Research Article Biomechanical Analysis of Hip Joint Arthroplasties using CT-Image Based Finite Element Method Mitsugu Todo * Research Institute for Applied Mechanics, Kyushu University, Kasuga, Japan

More information

Comparative Study of Fixation Devices for Intertrochanteric Fractures

Comparative Study of Fixation Devices for Intertrochanteric Fractures Comparative Study of Fixation Devices for Intertrochanteric Fractures C. Sticlaru * A. Davidescu Politehnica University of Timişoara Politehnica University of Timişoara Timişoara, România Timişoara, România

More information

BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA

BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2008, 152(2):1. T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova 1 BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL

More information

The Development and Analysis of an FEA Simulation on an Artificial Knee

The Development and Analysis of an FEA Simulation on an Artificial Knee ME 461: Finite Element Analysis Fall 2015 The Development and Analysis of an FEA Simulation on an Artificial Knee Daniel Offley, Joseph Martin, Harrison Wostein Department of Mechanical and Nuclear Engineering,

More information

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents. Blucher Mechanical Engineering Proceedings May 2014, vol. 1, num. 1 www.proceedings.blucher.com.br/evento/10wccm AN ANALYSIS OF THE CONTACT BETWEEN THE STENT AND THE ARTERY USING TUBE HIDROFORMING SIMULATION

More information

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 DOI 10.1007/s13191-012-0189-7 ORIGINAL ARTICLE Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A

More information

Comparative study of single-thread, double-thread, and triple-thread dental implant: a three-dimensional finite element analysis

Comparative study of single-thread, double-thread, and triple-thread dental implant: a three-dimensional finite element analysis ISSN 1 746-7233, England, UK World Journal of Modelling and Simulation Vol. 3 (2007) No. 4, pp. 310-314 Comparative study of single-thread, double-thread, and triple-thread dental implant: a three-dimensional

More information

ACCURATE NUMERICAL ANALYSIS OF GEAR STRENGTH BASED ON FINITE ELEMENT METHOD

ACCURATE NUMERICAL ANALYSIS OF GEAR STRENGTH BASED ON FINITE ELEMENT METHOD 31 st December 212. Vol. 46 No.2 25-212 JATIT & LLS. All rights reserved. ACCURATE NUMERICAL ANALYSIS OF GEAR STRENGTH BASED ON FINITE ELEMENT METHOD XUEYI LI, CHAOCHAO LI, DAQIAN GENG, SHOUBO JIANG, BINBING

More information

STRESS ANALYSIS OF THE RADIAL HEAD REPLACEMENTS IN AN ELBOW ARTICULATION

STRESS ANALYSIS OF THE RADIAL HEAD REPLACEMENTS IN AN ELBOW ARTICULATION Engineering MECHANICS, Vol. 15, 2008, No. 5, p. 319 327 319 STRESS ANALYSIS OF THE RADIAL HEAD REPLACEMENTS IN AN ELBOW ARTICULATION Pavel Hlavoň*, Vladimír Fuis*, Zdeněk Florian** The paper is focused

More information

Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue

Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue Megan McCain ES240 FEM Final Project December 19, 2006 Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue Cardiac myocytes are the cells that constitute the working muscle of the heart.

More information

Transactions on Biomedicine and Health vol 3, 1996 WIT Press, ISSN

Transactions on Biomedicine and Health vol 3, 1996 WIT Press,   ISSN Evaluation of forearm fixation plate design using finite element methods S.H. Saidpour, I.M. Flitta, X. Velay School ofdesign Engineering and Computing, Department of Product Design and Manufacture, Bournemouth

More information

Design and Analysis of Dental Surgical Fixture

Design and Analysis of Dental Surgical Fixture Volume 119 No. 10 2018, 1951-1956 ISSN: 1311-8080 (printed version); ISSN: 1314-3395 (on-line version) url: http://www.ijpam.eu ijpam.eu Design and Analysis of Dental Surgical Fixture Suhas H.G 1,*, Rajkumar

More information

Introduction to Biomedical Engineering

Introduction to Biomedical Engineering Introduction to Biomedical Engineering FW 16/17, AUT Biomechanics of tendons and ligaments G. Rouhi Biomechanics of tendons and ligaments Biomechanics of soft tissues The major soft tissues in musculoskeletal

More information

Finite Element Analysis of Dental Implant as Orthodontic Anchorage

Finite Element Analysis of Dental Implant as Orthodontic Anchorage JCDP 10.5005/jp-journals-10024-1044 ORIGINAL RESEARCH Finite Element Analysis of Dental Implant as Orthodontic Anchorage Finite Element Analysis of Dental Implant as Orthodontic Anchorage Anirban Sarmah,

More information

THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION REFERENCE REFLECTORS AND TRANSDUCERS FOR COMPRESSOR BLADE DOVETAILS

THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION REFERENCE REFLECTORS AND TRANSDUCERS FOR COMPRESSOR BLADE DOVETAILS International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada THE DEVELOPMENT AND MANUFACTURE OF FIXED- ULTRASONIC INSPECTION

More information

Journal of Biomechanical Science and Engineering

Journal of Biomechanical Science and Engineering 0123456789 Bulletin of the JSME Vol.9, No.2, 2014 Journal of Biomechanical Science and Engineering Finite element analysis of hip joint cartilage reproduced from real bone surface geometry based on 3D-CT

More information

Dental Research Journal

Dental Research Journal Dental Research Journal Original Article The influence of thread geometry on biomechanical load transfer to bone: A finite element analysis comparing two implant thread designs Manisha G Herekar, Viraj

More information

Professor Stephen D. Downing

Professor Stephen D. Downing Professor Stephen D. Downing Department of Mechanical Science and Engineering University of Illinois at Urbana-Champaign 2011-2013 Stephen Downing, All Rights Reserved 1. Comparison to Wrought Metals 2.

More information

Simulation of bone indentation

Simulation of bone indentation Modelling in Medicine and Biology VII 113 Simulation of bone indentation S. Kasiri 1, G. Reilly 1,2 & D. Taylor 1 1 Trinity Centre for Bioengineering, Trinity College Dublin, Ireland 2 Institute of Technology

More information

Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements

Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements Daniel Kluess, Sarah Bierbaum, Wolfram Mittelmeier, Rainer Bader,

More information

Numerical analysis of the embedded abutments of integral bridges

Numerical analysis of the embedded abutments of integral bridges Numerical analysis of the embedded abutments of integral bridges Ming XU Researcher, Dept. of Civil & Env. Engrg., University of Southampton, UK Summary Alan G. BLOODWORTH Lecturer, Dept. of Civil & Env.

More information

BASIC Level 1 Practitioner Certification

BASIC Level 1 Practitioner Certification Chapter 5 Place the Prosthesis Because dental implants are now an important and distinct part of prosthodontic dentistry, general dentists must be well versed in all aspects of the treatment option. The

More information

Numerical Analysis of Coronary Stent for Diverse Materials

Numerical Analysis of Coronary Stent for Diverse Materials Numerical Analysis of Coronary Stent for Diverse Materials Vasantha Kumar Assistant Professor Department of Mechanical Engineering Bearys Institute of Technology Mangalore, Karnataka, India Abstract Coronary

More information

Practical Aspects of Finite Element Method Applications in Dentistry

Practical Aspects of Finite Element Method Applications in Dentistry 10.1515/bjdm-2017-0011 BALKAN JOURNAL OF DENTAL MEDICINE ISSN 2335-0245 STOMATOLOGICAL SOCIETY Practical Aspects of Finite Element Method Applications in Dentistry SUMMARY The use of numerical methods,

More information

Comparative study of the contact pressures in hip joint models with femoroacetabular impingment with different cephalic deformities

Comparative study of the contact pressures in hip joint models with femoroacetabular impingment with different cephalic deformities Comparative study of the contact pressures in hip joint models with femoroacetabular impingment with different cephalic deformities Iryna Havenko Instituto Superior Técnico, Universidade de Lisboa, Portugal

More information

Failure criteria for Adhesives. Sainath Kadam, 3mE 19 oktober 2014

Failure criteria for Adhesives. Sainath Kadam, 3mE 19 oktober 2014 Failure criteria for Adhesives Sainath Kadam, 3mE 19 oktober 2014 () Failure criteria for Adhesives 1 / 59 Outline 1 Introduction 2 Testing 3 Data analysis 4 FEM 5 Results and Conclusions () Failure criteria

More information

Three dimensional finite element analysis of immediate loading mini over denture implants with and without acrylonitrile O ring

Three dimensional finite element analysis of immediate loading mini over denture implants with and without acrylonitrile O ring Original Research Three dimensional finite element analysis of immediate loading mini over denture implants with and without acrylonitrile O ring Srinivasan Jayaraman, Sreekanth Mallan 1, Babu Rajan, Murugan

More information

CHAPTER 5 MODELING OF THE BRIDGE

CHAPTER 5 MODELING OF THE BRIDGE 62 CHAPTER 5 MODELING OF THE BRIDGE 5.1 MODELING SAP2000, a nonlinear software package was used for modeling and analysing the study bridge. The following list provides details about the element type used

More information

Effect of Screws Placement on Locking Compression Plate for Fixating Medial Transverse Fracture of Tibia

Effect of Screws Placement on Locking Compression Plate for Fixating Medial Transverse Fracture of Tibia Effect of Screws Placement on Locking Compression Plate for Fixating Medial Transverse Fracture of Tibia Ehsan Taheri *1, Behrooz Sepehri 2, Reza Ganji 3, Cyrus Nasirai 4 1 Mashhad Branch, Islamic Azad

More information

ABSTRACT INTRODUCTION

ABSTRACT INTRODUCTION ABSTRACT Purpose: Evaluation of stress distribution maxilla under parafunctional loading on short dental implant using 3-dimensional (3D) finite element models. Material and method: 3D maxillary and mandibular

More information

2008 International ANSYS Conference

2008 International ANSYS Conference 2008 International ANSYS Conference Simulation of Mandible Biomechanics after Fixation of Subcondylar Fractures Dana Coombs Analytical Engineer Synthes USA 2008 ANSYS, Inc. All rights reserved. 1 ANSYS,

More information

Very small abutment head easy and secure handling. Ankylos. The SmartFix concept. Prosthetic solution on angled implants

Very small abutment head easy and secure handling. Ankylos. The SmartFix concept. Prosthetic solution on angled implants Very small abutment head easy and secure handling Ankylos The SmartFix concept Prosthetic solution on angled implants Stable prosthetic fit The area supporting the prosthesis is extended distally by the

More information

EVALUATION OF STRESS DISTRIBUTION AND MICROMOTION OF DENTAL IMPLANT: IN VIVO CASE STUDY

EVALUATION OF STRESS DISTRIBUTION AND MICROMOTION OF DENTAL IMPLANT: IN VIVO CASE STUDY EVALUATION OF STRESS DISTRIBUTION AND MICROMOTION OF DENTAL IMPLANT: IN VIVO CASE STUDY Maya Genisa 1, Zainul Ahmad Rajion 1, Solehuddin Shuib 2, Dasmawati Mohamad 1 and Abdullah Pohchi 1 1 School of Dental

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that JTL America 3205 Clairmont Ct., Ste. B Fort

More information

Case study for quality control & quality assurance in metal additive manufacturing: A dental bridge

Case study for quality control & quality assurance in metal additive manufacturing: A dental bridge Case study for quality control & quality assurance in metal additive manufacturing: A dental bridge Keywords: Dental bridge, quality in dentistry Table of Contents Introduction... 2 Product description...

More information

Effect of the femoral stem size on femur bone quality towards THR

Effect of the femoral stem size on femur bone quality towards THR Original article: Effect of the femoral stem size on femur bone quality towards THR Palash Kumar Maji a,*, Amit Roy Chowdhury b, Debasis Datta b, S Karmakar a, Subhomoy Chatterjee b and A K Prasad a a

More information

Mechanical Aspects of an Interference Screw Placement in ACL Reconstruction

Mechanical Aspects of an Interference Screw Placement in ACL Reconstruction Mechanical Aspects of an Interference Screw Placement in ACL Reconstruction Mahmoud Chizari 1, Mohammad Alrashidi 2, Khaled Alrashdan 2, Ibrahim Yildiz 3, Jamaluddin Mahmud 4 1 School of Engineering and

More information

GMI FRONTIER implant system. Surgical procedures guide

GMI FRONTIER implant system. Surgical procedures guide GMI implant system Surgical procedures guide ABOUT THIS MANUAL This surgical procedures guide or surgical manual for the GMI implant system is designed solely to provide instructions for using GMI products,

More information

IMPLEMENTATION OF CONTACT DEFINITIONS CALCULATED BY FEA TO DESCRIBE THE HEALING PROCESS OF BASAL IMPLANTS

IMPLEMENTATION OF CONTACT DEFINITIONS CALCULATED BY FEA TO DESCRIBE THE HEALING PROCESS OF BASAL IMPLANTS Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2008, 152(1):169 173. S. Ihde, T. Goldmann, L. Himmlova, Z. Aleksic, J. Kuzelka 169 IMPLEMENTATION OF CONTACT DEFINITIONS CALCULATED BY FEA TO DESCRIBE

More information

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter Design Philosophy Cementless Stems of 12/14 Taper Stem Length Neck shaft Angle STEM-N3 STEM-N4 STEM-N5 STEM-N6 STEM-N7 STEM-N8 STEM-N9 123 mm 125 mm 130 mm 135 mm 140 mm 145 mm 150 mm 132 Cement Polished

More information

Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur

Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur Finite Element Analysis of Proximal Femur Nail for Subtrochanteric Fractured Femur Sowmianarayanan.S Assistant Consultant, Tata Consultancy Services Limited Prof. A.Chandrasekaran Professor, Dept. of Orthopaedic

More information

Zentrum für Zahn-, Mundund Kieferheilkunde

Zentrum für Zahn-, Mundund Kieferheilkunde Numerical Investigations On The Mechanical Behaviour Of The Novel Bar Sys tem SFI-Bar By Cendres+Métaux Report to the company CENDRES+MÉTAUX SA Bözingenstrasse 122 CH-2501 Biel/Bienne Dipl.-Math. Ludger

More information

Numerical evaluation of bone remodelling associated with trans-femoral osseointegration implant - a 68 month follow-up study

Numerical evaluation of bone remodelling associated with trans-femoral osseointegration implant - a 68 month follow-up study Title Page & Abstract Numerical evaluation of bone remodelling associated with trans-femoral osseointegration implant - a 68 month follow-up study D H Xu 1, A D Crocombe 2, W Xu* 2 1 Department of Orthopaedics

More information

FEM and BEM Stress Analysis of Mandibular Bone Surrounding a Dental Implant

FEM and BEM Stress Analysis of Mandibular Bone Surrounding a Dental Implant 282 The Open Mechanical Engineering Journal, 215, 9, 282-292 Send Orders for Reprints to reprints@benthamscience.ae Open Access FEM and BEM Stress Analysis of Mandibular Bone Surrounding a Dental Implant

More information

MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES

MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES Prof.dr.eng. Mirela TOTH-TAŞCĂU, Lecturer.dr.eng. Lucian RUSU, PhD.eng. Flavia BĂLĂNEAN, PhD.eng. Cristian TOADER-PASTI, Politehnica

More information

Three Dimensional Finite Element Analysis to Detect Stress Distribution in Spiral Implants and Surrounding Bone

Three Dimensional Finite Element Analysis to Detect Stress Distribution in Spiral Implants and Surrounding Bone Original Article Three Dimensional Finite Element Analysis to Detect Stress Distribution in Spiral Implants and Surrounding Bone Matteo Danza*, Annalisa Palmieri**, Francesca Farinella***, Giorgio Brunelli****,

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that JTL America 3205 Clairmont Ct., Ste. B Fort

More information

Student Presentation # 3 Biomechanics

Student Presentation # 3 Biomechanics Student Presentation # 3 Biomechanics Anu Phogat Peter Ho 1 BIOMECHANICS OF DENTAL IMPLANTS 2 STRUCTURE OF TOOTH 3 INTERESTING FACTS According to the National Institutes of Health,by age 50, the average

More information

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 6, Issue 12 (May 2013), PP. 119-123 A Mathematical Model to Calculate Contact Stresses

More information

Multi-scale Finite Element Analysis on Biomechanical Response of Functionally Graded Dental Implant / Mandible System

Multi-scale Finite Element Analysis on Biomechanical Response of Functionally Graded Dental Implant / Mandible System Journal of the Serbian Society for Computational Mechanics / Vol. 1 / No. 1, 2007 / pp. 164-173 Multi-scale Finite Element Analysis on Biomechanical Response of Functionally Graded Dental Implant / Mandible

More information

Paper submission. Denture repair with the application of a magnetic attachment to the inner crown of a telescopic crown: A 3-year follow-up case

Paper submission. Denture repair with the application of a magnetic attachment to the inner crown of a telescopic crown: A 3-year follow-up case Paper submission Denture repair with the application of a magnetic attachment to the inner crown of a telescopic crown: A 3-year follow-up case A. Izumida Department of Comprehensive Dentistry, Tohoku

More information

Installation of force transducers

Installation of force transducers Installation of force transducers Force transducers have reached a high technical standard in recent years. Transducers achieve part of their high accuracy due to improved possibilities for calibration.

More information

Original Article Effect of Anti-rotational Design and Countertorque Device on Peri-implant Bone Stress During and After Preload A Preliminary Results

Original Article Effect of Anti-rotational Design and Countertorque Device on Peri-implant Bone Stress During and After Preload A Preliminary Results Effect of Anti-rotational Design and Countertorque Device on Peri-implant Bone Stress During and After Preload A Preliminary Results Ming-Su Lee, DDS, MS Attending Doctor, Department of Dentistry, National

More information

Journal of Dental Research, Dental Clinics, Dental Prospects. Original Article

Journal of Dental Research, Dental Clinics, Dental Prospects. Original Article Journal of Dental Research, Dental Clinics, Dental Prospects Original Article Comparison of the Effect of Three Abutment-implant Connections on Stress Distribution at the Internal Surface of Dental Implants:

More information

Comparative study of 12 thread shapes of dental implant designs: a three-dimensional finite element analysis

Comparative study of 12 thread shapes of dental implant designs: a three-dimensional finite element analysis ISSN 1 746-7233, England, UK World Journal of Modelling and Simulation Vol. 2 (2006) No. 2, pp. 134-140 Comparative study of 12 thread shapes of dental implant designs: a three-dimensional finite element

More information

Advanced Integration of Non-Destructive Testing during Laboratory Fatigue Test of Airframe Metal Parts

Advanced Integration of Non-Destructive Testing during Laboratory Fatigue Test of Airframe Metal Parts ECNDT 2006 - Th.2.1.4 Advanced Integration of Non-Destructive Testing during Laboratory Fatigue Test of Airframe Metal Parts Jiří BĚHAL jr., Jiří VRHEL Aeronautical Research and Test Institute, Prague,

More information

MECHANISM OF LOAD TRANSFER ALONG THE BONE- DENTAL IMPLANT INTERFACE

MECHANISM OF LOAD TRANSFER ALONG THE BONE- DENTAL IMPLANT INTERFACE MECHANISM OF LOAD TRANSFER ALONG THE BONE- DENTAL IMPLANT INTERFACE A Thesis Presented by Samira Faegh to The Department of Mechanical and Industrial Engineering In partial fulfillment of the requirements

More information

Numerical predicting of contact and pressure sore of lower extremity parts caused by prosthetic and orthotic

Numerical predicting of contact and pressure sore of lower extremity parts caused by prosthetic and orthotic MACROJOURNALS The Journal of MacroTrends in Health and Medicine Numerical predicting of contact and pressure sore of lower extremity parts caused by prosthetic and orthotic Reza Fakhrai, Bahram Saadatfar,

More information

EXTERNAL FIXATION SYSTEM

EXTERNAL FIXATION SYSTEM EXTERNAL FIXATION SYSTEM The Apex Pin Fixation System is a one step procedure reducing insertion time and reducing insertion temperature. There are three types of fixation pins: Self-drilling / Self-tapping

More information

Numerical Study of the Behavior of a Zirconia Dental Prosthesis with Prior Defect

Numerical Study of the Behavior of a Zirconia Dental Prosthesis with Prior Defect Journal of the Serbian Society for Computational Mechanics / Vol. 11 / No. 1, 2017 / pp 1-10 (DOI: 10.24874/jsscm.2017.11.01.01) Numerical Study of the Behavior of a Zirconia Dental Prosthesis with Prior

More information

AnyRidge Internal System

AnyRidge Internal System Type of Normal Thin Low Loose bone Type of and selection of fixture With the normal ridge, Standard implant is more than enough because there are enough quantity and Normal quality of bone. Type of and

More information

A computer-based simulation of vacuum extraction during childbirth

A computer-based simulation of vacuum extraction during childbirth A computer-based simulation of vacuum extraction during childbirth Rudy Lapeer, Zelimkhan Gerikhanov and Vilius Audinis School of Computing Sciences University of East Anglia Norwich NR4 7TJ, UK Abstract:

More information

comprehensive Design Rationale shoulder system Knees Hips Extremities Cement and Accessories PMI TEchnology

comprehensive Design Rationale shoulder system Knees Hips Extremities Cement and Accessories PMI TEchnology comprehensive shoulder system Design Rationale Knees Hips Extremities Cement and Accessories PMI TEchnology . INTRODUCTION The Comprehensive Shoulder System is an evolutionary design based on the successful

More information

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments Key Engineering Materials Online: 2005-06-15 ISSN: 1662-9795, Vols. 288-289, pp 603-606 doi:10.4028/www.scientific.net/kem.288-289.603 2005 Trans Tech Publications, Switzerland Torsion Property and Cyclic

More information

Available online at ScienceDirect

Available online at   ScienceDirect Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 ( 2014 ) 1251 1257 24th DAAAM International Symposium on Intelligent Manufacturing and Automation, 2013 Comparative Study

More information