1 ORIGINAL ARTICLE Biodegradation of orthodontic metallic brackets and associated implications for friction Saulo Regis Jr, a Paulo Soares, b Elisa S. Camargo, c Odilon Guariza Filho, c Orlando Tanaka, d and Hiroshi Maruo c Curitiba, Parana, Brazil Introduction: This study aimed to assess the effect of clinical exposure on the surface morphology, dimensions, and frictional behavior of metallic orthodontic brackets. Methods: Ninety-five brackets, of 3 commercial brands, were retrieved from patients who had finished orthodontic treatment. As-received brackets, matched by type and brand, were used for comparisons. Surface morphology and precipitated material were analyzed by optical and scanning electron microscopy and x-ray microanalysis. Bracket dimensions were measured with a measuring microscope. Resistance to sliding on a stainless steel wire was assessed. Results: Retrieved brackets showed surface alterations from corrosion, wear, and plastic deformation, especially in the external slot edges. Film deposition over the alloy surface was observed to a variable extent. The main elements in the film were carbon, oxygen, calcium, and phosphorus. The as-received brackets showed differences (P\0.05) in the slot sizes among brands, and 1 brand showed a 3% increase in the retrieved brackets slots. The frictional behavior differed among brands. Retrieved brackets of 2 brands showed 10% to 20% increases in resistance to sliding. Conclusions: Metallic brackets undergo significant degradation during orthodontic treatment, possibly with increased friction. At present, it is difficult to predict the impact of these changes on the clinical performance of orthodontic components. (Am J Orthod Dentofacial Orthop 2011;140:501-9) The degradation of metallic materials placed in patients has long been a concern of biomaterials science. Laboratory experiments that simulate in-vivo degradation of metal implants have made it possible to estimate the effect of specific parameters but lack the consistency required to represent the complexity of the oral environment. 1-5 Variations in temperature and ph caused by diet, decomposition of foods and cell debris, and oral florae and their by-products are important factors to consider when evaluating the clinical behavior of dental materials and comparing them with other biomaterials. In particular, orthodontic accessories are under masticatory forces and multi-axial loads from the activation of the wire in the bracket slot. 1 From the Pontifical Catholic University of Parana, Curitiba, Parana, Brazil. a Postgraduate student, Graduate Dentistry Program in Orthodontics. b Associate professor, Mechanical Engineering Department. c Associate professor, Graduate Dentistry Program in Orthodontics. d Professor, Graduate Dentistry Program in Orthodontics. The authors report no commercial, proprietary, or financial interest in the products or companies described in this article. Reprint requests to: Hiroshi Maruo, Graduate Dentistry Program in Orthodontics, Pontifical Catholic University of Parana, R. Imaculada Conceiç~ao, 1155, Curitiba, Parana, , Brazil; , h.maruo@gmail.com. Submitted, May 2010; revised and accepted, January /$36.00 Copyright Ó 2011 by the American Association of Orthodontists. doi: /j.ajodo In-vivo aged orthodontic components show signs of degradation such as morphologic changes and surface alterations from corrosion, wear, and formation of integuments. 4,6-12 Concerns regarding the clinical impact of these alterations include (1) elements released into the oral environment and their implications on biocompatibility, and (2) impairment of the performance of the orthodontic appliance. 13 Orthodontic metallic materials are usually composed of alloys including several base metals such as nickel, chromium, cobalt, iron, molybdenum, and titanium. Concerning biocompatibility, nickel stands out among the other elements. Carcinogenic, 14 mutagenic, 15 and cytotoxic properties 16,17 have been attributed to it, although no local or systemic clinical condition is clearly related to the forms of nickel used in dentistry. 13 The lone exception is hypersensitivity reactions. 18 Nickel might also elicit periodontal reactions such as gingival hyperplasia in orthodontic patients. This condition is hardly distinguishable from microbial-induced gingival overgrowth. 19 Surface alterations in orthodontic devices might compromise the appliance s esthetics, 20 increase microbial adhesion, 11,21 modify bracket-wire activations such as torque expression, 22 cause fractures during clinical use, 23 and influence the magnitude of friction between the bracket and the wire. 1,20,24 Many situations in an orthodontic 501

2 502 Regis et al practice require sliding teeth through a properly contoured archwire, and friction might cause significant loss of the applied force. 10,24 Many investigations have focused on the frictional behavior of as-received orthodontic materials Nevertheless, the alterations that orthodontic brackets undergo during treatment and their impact on clinical performance, especially friction, are mostly unknown. Therefore, we aimed to evaluate the surface morphology, dimensional stability, and frictional behavior of orthodontic metallic brackets retrieved after full orthodontic treatment, compared with as-received brackets from the manufacturers. MATERIAL AND METHODS The sample consisted of brackets with a slot size of mm ( in), as described by the manufacturer, retrieved from patients who had finished orthodontic treatment in the private practice of 2 experienced orthodontists (E.S.C., O.T.). The brackets were carefully debonded with a direct bond bracket remover pliers and kept in receptacles with distilled water. They were brushed with an electric brush for 10 seconds and rinsed with distilled water to remove any loosely attached integuments. They were then kept in self-sealed sterilizing packs until analysis. The following information was registered: patient identification, date of placement and date of removal of the appliance. Brackets with obvious distortions or calcifications that hindered the engagement of a mm ( in) cross-sectional wire were discarded. A total of 95 brackets of different types (for premolars, canines, and incisors from both arches) were selected. The sample comprised 3 brands: 32 Mini Standard Edgewise stainless steel brackets (American Orthodontics, Sheboygan, Wis), 34 Kirium Edgewise stainless steel brackets (3M Abzil, Sumare, Brazil), and 29 NuEdge preadjusted Roth prescription brackets (TP Orthodontics, LaPorte, Ind) made of copper-chromium alloy. The brackets were retrieved from 7 patients (mean age, 18 years 9 months), with a mean treatment time of 41 months. Stainless steel and nickel-titanium wires used in the orthodontic treatments were ligated with elastomeric and metallic ligatures. A group of brackets as-received from the manufacturers, matched by types and brands to the sample, was used for comparisons. They were submitted to the same procedures as the retrieved specimens. An optical reflected light microscope (BX60; Olympus Optical, Tokyo, Japan) was used to evaluate the surface morphology of the as-received and the retrieved brackets for areas of corrosion, signs of wear, plastic deformations, and adherent materials. The images were acquired in a bright field at various magnifications ( times). Fifteen retrieved brackets, equally distributed among brands, and their as-received counterparts were selected based on reflected light images for analysis in a scanning electron microscope (JSM-6360LV; Jeol, Tokyo, Japan) and an energy dispersive x-ray spectrometer. Secondary electron images and backscattered electron images were acquired at various magnifications ( times) by using a 20-kV accelerating voltage. The images allowed for assessment of the micromorphologic characteristics of the slot surfaces. Areas of interest were submitted to microanalysis for element assessment. Spectra were acquired with the same accelerating voltage at different magnifications and a 100-second acquisition time. The slot sizes (left and right, vertical dimension) and the internal widths between the tie-wings (cervical and occlusal, horizontal dimension) of the remaining 80 retrieved brackets and their as-received counterparts were measured in a measuring microscope (MM-40; Nikon, Yokohama, Japan). Measurements were made by 1 operator using a holder with a wire of a crosssection of mm ( in). The wire was used to position the brackets slots perpendicular to the microscope table. To evaluate the method error, the measurements of 12 specimens were repeated after a 1-week interval. No statistically significant difference was found between repeated measurements according to a paired-samples t test (P ). The sliding resistance analysis with stainless steel wires was conducted on both as-received and retrieved brackets in a universal testing machine (DL-500; EMIC, S~ao Jose dos Pinhais, Brazil), with a device especially designed for this experiment (Fig 1). Test specimens were obtained by bonding the brackets with a cyanoacrylate adhesive to a mm acrylic plate with a holder in a standardized way. This guaranteed that the bracket slots stayed parallel to the testing machine s vertical axis. Stainless steel wire segments (Shiny Bright, TP Orthodontics) with a cross-section of mm ( in) and a length of 11.5 cm were used. The wires were cleaned with 70% alcohol. They were ligated in their middle portion to the brackets with 3.0-mm elastomeric ligatures (Mini Stix noncoated, TP Orthodontics) immediately before the test to standardize the ligation force. Each segment was used only once. Test specimens were mounted in the device and assembled in the test machine. A 300 g weight was attached to the lower extremity of the wire to keep it under tension. The wire was pulled along the bracket at a rate of 5 mm per minute for 1 minute. The force levels were registered by a 10 kgf load cell. The sliding resistance was calculated by averaging the forces registered between the first and fifth millimeters of displacement, October 2011 Vol 140 Issue 4 American Journal of Orthodontics and Dentofacial Orthopedics

3 Regis et al 503 Fig 1. Device with test specimen mounted for the sliding resistance test, seen with greater magnification in detail. disregarding the initial static friction. Since the retrieved brackets were of different types, even for the same brand, it was of no use to consider the mean sliding resistance for each brand, but only the differences between retrieved and as-received brackets. For means of comparison among the different brands, the percentage difference in the sliding resistance between retrieved and as-received brackets was calculated with the following equation: DifSR(%) 5 [(SRretrieved SRas-received)/ SRas-received] 3 100, where DifSR is the percentage difference in sliding resistance, and SR is sliding resistance. Statistical analysis Kolmogorov-Smirnov and Levene tests were used to evaluate the normality of data distribution and the homogeneity of variance, respectively. Only the tie-wings widths and the percentage differences in the sliding resistance variables had nonnormal distributions. Only the as-received brackets sliding resistance had a homogeneous variance. The effect of usage on the tie-wings widths was investigated with the Mann-Whitney test. Slot dimensions were compared among brands and usage Fig 2. Optical microscopy images (bright field, 200 times magnification) of the slots of: A, NuEdge; B, Mini Standard Edgewise; and C, Kirium retrieved brackets. These images illustrate the wear and plastic deformation (arrows) of the external slot edges. conditions (as-received or retrieved) with analysis of variance (ANOVA) and the Games-Howell post-hoc test. The Kruskall-Wallis multiple comparisons test was used to evaluate differences among brands and patients in the percentage difference in the sliding resistance. A pairedsamples t test was used to investigate the effect of usage on sliding resistance for each brand in a 2-tailed model. The level of significance for all tests was set at a RESULTS Surface modifications with signs of corrosion, plastic deformation, and wear were seen in the retrieved brackets (Fig 2) when compared with the as-received brackets during optical microscopic evaluation. Wear American Journal of Orthodontics and Dentofacial Orthopedics October 2011 Vol 140 Issue 4

4 504 Regis et al Fig 3. Scanning electron microscope images of Kirium brackets: A, slot of an as-received bracket (backscattered electron image, 100-times magnification); B, slot of a retrieved bracket (secondary electron image, 100-times magnification); C, in the retrieved brackets, there were smooth milling grooves in the lateral walls of the slot (backscattered electron image, 100-times magnification); D, delamination debris on a retrieved bracket (secondary electron image, 300-times magnification); E, a precipitate of a retrieved bracket containing silica and barium (backscattered electron image, 300-times magnification). and deformation were especially seen on the external edges of the slot bottom surface. The retrieved specimens had film deposition and islands of aggregated materials to a variable extent. The variability occurred even with brackets removed from the same patient, with no individual pattern in the amount of integument formed. Mini Standard Edgewise retrieved brackets slots varied in areas with pits and film deposition or areas with extensive signs of wear and deformation. This last pattern was more frequent in retrieved Kirium brackets. NuEdge brackets had a rough surface, with pits and crevices in a parallel arrangement, in both as-received and retrieved specimens. These originally roughened surfaces, a coating applied by the manufacturer, and the presence of integuments made it difficult to evaluate specific usage-related alterations in NuEdge brackets. The same morphologic patterns were seen in greater detail in the scanning electron microscope images (Figs 3-5). The precipitated film exhibited a dark phase in the backscattered electron images, contrasting with the bright phase of the brackets alloy. This finding suggested the presence of elements with a low atomic number in the film. Carbon, oxygen, calcium, and phosphorus were the main elements detected in the film by energy dispersive x-ray spectrometry microanalysis. However, nitrogen, sulfur, sodium, potassium, and aluminum were eventually found. In some areas, the precipitates masked the topographic features of the underlying alloy surface (Fig 6). Greater magnifications showed details of the arrangement of pits and crevices in the NuEdge specimens (Fig 5, D and E). The coating applied by the manufacturer in the as-received NuEdge brackets was seen in backscattered electron images as a dark-phased, noncontinuous pellicle. The same distribution was not seen in the retrieved specimens, in which the integuments were limited to slots and gaps. The original coating was visually indistinguishable from the biofilm formed during oral exposure. Some materials of atypical composition were found in retrieved brackets, including a 150-mm precipitate containing silica, aluminum, barium, iron, carbon, and oxygen (Fig 3, E). Also included were silver and silica incrustations that had a brighter phase than the cobalt-chromium alloy (Fig 6). A comparison between as-received and retrieved brackets found no significant difference (P.0.05) in the tie-wings internal widths (Table I). Table II presents the slot dimensions according to usage and brands. NuEdge brackets had significantly greater slot dimensions than the other brands (P \0.05). Differences between retrieved and as-received brackets October 2011 Vol 140 Issue 4 American Journal of Orthodontics and Dentofacial Orthopedics

5 Regis et al 505 Fig 4. Scanning electron microscope images of Mini Standard Edgewise brackets: A, slot of an asreceived bracket (backscattered electron image, 100-times magnification); B, slot of a retrieved bracket (backscattered electron image, 100-times magnification); C, slot lateral wall of a retrieved bracket, with plastic deformation (bright arrow) and film deposition (backscattered electron image, 300-times magnification); D, islands of precipitates (300 times magnification) and E, pitting corrosion (dark arrows; 100-times magnification) after intraoral exposure. were seen only for the right slot of the Kirium brackets (P \0.05). Sliding resistance, evaluated separately for each brand according to usage condition by t tests in a 2-tailed model, had P values of 0.054, 0.103, and 0.08 for NuEdge, Mini Standard Edgewise, and Kirium, respectively. Since various bracket types were tested (premolar, canine, and incisor for both arches), it was not viable to calculate the sliding resistance means and compare them among brands. However, the percentage difference in the sliding resistance, as a ratio of sliding resistance alteration between retrieved and as-received brackets, could be compared among brands. The percentage difference in the sliding resistance had mean values of 17.99% (SD, 36.50%) for NuEdge brackets, 13.62% (SD, 34.26%) for Kirium, and 3.10% (SD, 31.82%) for Mini Standard Edgewise. Differences were statistically significant between Mini Standard Edgewise and the other brands (P \0.05). No significant differences were found for the percentage difference in the sliding resistance between patients (P ). DISCUSSION Microscopic analysis suggested that the orthodontic brackets underwent significant alterations in variable ways during treatment. Retrieved orthodontic accessories have shown diverse behaviors in the literature: some had significant alterations, 4,6,9-12,29 and others showed no differences from as-received. 7,8,29-32 These conflicting results might be attributed to the diversity of the materials analyzed (brackets, archwires, headgear components), the time of exposure to the oral environment, and the patient s characteristics. Additionaly, corrosion susceptibility is influenced by the composition of the alloy, its microstructure, manufacturing procedures and their impact on the internal stress of the material, and the formation of a surface passivation film. 33 Since the brackets evaluated in this study had different compositions and brands, dissimilarities in their surface morphology after orthodontic treatment could be related to each alloy s tribological, physical, and electrochemical properties, and to manufacturing procedures. The precipitated film over the retrieved brackets surfaces was also found on orthodontic wires 9,30 and headgear components. 10 The overlap of the calcium and phosphorus distributions (Fig 6) is consistent with crystalline particle formation. This finding was also described by Eliades et al. 9,10 These authors additionally examined the molecular composition of the precipitated film. They found amide, alcohol, and carbonate as the main American Journal of Orthodontics and Dentofacial Orthopedics October 2011 Vol 140 Issue 4

6 506 Regis et al Fig 5. Scanning electron microscope images of NuEdge brackets: A, slot of an as-received bracket (backscattered electron image, 100-times magnification); B, slot of a retrieved bracket (backscattered electron image, 100-times magnification); C, as-received bracket slot with deformation (secondary electron image, 100-times magnification); D, internal edge between the slot and the middle portion of the bracket, showing traces of the mechanical action in the slot (left) in contrast to the rough surface (right) that does not come into contact with the wires (backscattered electron image, 1000-times magnification); E, retrieved bracket slot surface showing the linear distribution of crevices typically seen in NuEdge brackets. Also seen is an area with signs of wear and plastic deformation (secondary electron image, 300-times magnification). organic constituents. They also identified the presence of chlorine, potassium, and sodium in a uniform distribution. Such findings suggest the formation of a biofilm composed of a proteinaceous matrix and scattered crystalline particles. The differing properties of the bracket materials and differing retrieval protocols might account for the dissimilarities in the integuments found in these studies.the cleansing procedures we used were intended to simulate a patient s dental hygiene and to remove loosely attached material. The precipitate containing silica and barium might be attributed to slot contamination by composite resins or adhesives, which have similar compositions. 34 Although mass transfer can occur with the sliding of a metallic surface, the silver-containing incrustations found on retrieved brackets (Fig 6) could not be attributed to any clinical findings. 26 Calcium and aluminum inclusions found on biomedically applied materials were strongly related to corrosion. Other sources of inclusions could be raw-material impurities and inclusions not dissolved in the melt. 35 The new brackets slot dimensions varied among brands, with a peak of mm ( in; 12%) beyond the nominal value in the NuEdge specimens. When evaluating the same dimensions, Oh et al 25 also found differences among brands, although of smaller magnitudes (0.018 mm). The evaluated brackets had acceptable dimensional stability during orthodontic treatment. Only the Kirium retrieved brackets had significant differences compared with their as-received counterparts. Specifically, there was a mm ( in; 3%) increase in a slot size. These findings agree with those of Fischer-Brandies et al, 36 who found indentations and a mm slot widening on stainless steel brackets tested with wire torque activations. They attributed the results to the low stiffness of the bracket material compared with the wires. The same mechanism could be related to the wear and deformation of the slot edges of the retrieved brackets in this study. Alterations caused by debonding were minimized by using a procedure that applied force in the bracket base, preserving the area of interest. Considering the slot variations among brands, the observed changes in used Kirium brackets might not have clinical significance. These results did not allow the rejection of the null hypothesis of no difference in sliding resistance between October 2011 Vol 140 Issue 4 American Journal of Orthodontics and Dentofacial Orthopedics

7 Regis et al 507 Fig 6. MEV and energy dispersive x-ray spectrometry images of a NuEdge bracket with film deposition and silver-containing particles: A, (100-times magnification) backscattered electron images of the slot, where silver particles appear as the brighter phase; B and C, backscattered electron and secondary electron images, respectively, of the marked area in A with greater magnification (1000 times); D, energy dispersive x-ray spectrometry map of cobalt, with the bright dots indicating that element, and chromium had a similar distribution; E, energy dispersive x-ray spectrometry map of calcium demonstrating a greater concentration in the area corresponding to the precipitated film images in B and C, with phosphorus having a similar distribution; F, energy dispersive x-ray spectrometry map of silver. Molybdenum and silica had more uniform distributions, although they were concentrated in the same areas as the silver. Table I. Comparison of the internal horizontal widths of the tie-wings of as-received and retrieved brackets Mean SD Significance* Cervical tie-wings internal width (mm) As-received Retrieved Occlusal tie-wings internal width (mm) As-received Retrieved *Mann-Whitney test results. as-received and retrieved brackets by using a 2-tailed model for each brand separately. However, the percentage difference in the sliding resistance indicates significantly different tendencies when comparing Kirium (13.62%; SD, 34.26%) and NuEdge (17.99%; SD, 36.50%) brackets with Mini Standard Edgewise brackets ( 3.10%; SD, 1.82%). Considering that the percentage differences in the sliding resistance results indicate the sliding resistance changes in each brand, a 1-tailed model could be adopted for statistical testing. In this scenario, by using a t test for paired samples to compare sliding resistance values between as-received and retrieved brackets separately for each brand, the P values were for NuEdge, 0.04 for Kirium, and for Mini Standard Edgewise. Therefore, Kirium and NuEdge brackets had greater mean friction, whereas the Mini Standard Edgewise brackets had no significant alteration. Table II. Heights (mm) of right and left slots according to brand and use n Mean SD Height of left slot NU as-received a NU retrieved a MS as-received b MS retrieved b KR as-received b,c KR retrieved c,d Height of right slot NU as-received a NU retrieved a MS as-received b MS retrieved b KR as-received b,c KR retrieved d Similar letters identify groups without significant differences (P $0.05), according to the Games-Howell multiple comparisons test. NU, NuEdge; MS, Mini Standard Edgewise; KR, Kirium. Although some authors have hypothesized that an increase in friction occurs due to surface alterations, 1,20,24 Berg et al 37 suggested a lubricating effect for salivary pellicles over the sliding surfaces in the oral environment. The percentage difference in the sliding resistance variable had great variability for each brand, with high standard devation values. Oh et al 25 found similar results in an in-vitro setup, where variations in American Journal of Orthodontics and Dentofacial Orthopedics October 2011 Vol 140 Issue 4

8 508 Regis et al the frictional resistances ranged up to 2.5 times among brackets of the same brand. Eliades and Bourauel 1 reported preliminary results of an ongoing study showing that retrieved nickel-titanium and stainless steel wires might have approximately 10% to 20% increases in frictional forces, with great fluctuations over the measured displacement. They attributed this finding to the complexes precipitated intraorally. Wichelhaus et al 38 also verified a significant increase in friction of nickeltitanium wires after clinical use. We found no similar evaluations of retrieved brackets. The frictional test used in this study does not fully represent oral conditions. Second- and third-order inclinations, dental-arch convexity, binding effects between bracket and archwire, and material-related frictional implications were not part of our goals. Nevertheless, this study provides evidence of the effects of clinical use and associated phenomena on orthodontic components classic friction. The different behaviors of the brackets we evaluated raises doubt about which factors determined the observed alterations. Further investigations with other commercial brands, standardized treatment mechanics, and controlled durations of oral exposure might clarify the effect of specific factors on orthodontic accessories friction during clinical use. Our results suggest that care must be taken when extrapolating conclusions obtained in research with as-received materials to long-term clinical scenarios. CONCLUSIONS Clinical use causes surface alterations in metallic orthodontic brackets, with distinct patterns of alterations for different brands. Differences in morphology after use are smaller than those found in as-received brackets among brands. Distinct frictional behaviors were observed for each bracket brand with clinical use. There were 10% to 20% increases between retrieved and asreceived brackets in NuEdge and Kirium, whereas the Mini Standard Edgewise brackets remained unaffected. REFERENCES 1. Eliades T, Bourauel C. Intraoral aging of orthodontic materials: the picture we miss and its clinical relevance. Am J Orthod Dentofacial Orthop 2005;127: Eliades T, Gioka C, Zinelis S, Eliades G, Makou M. Plastic brackets: hardness and associated clinical implications. World J Orthod 2004;5: Cortizo MC, De Mele MF, Cortizo AM. Metallic dental material biocompatibility in osteoblastlike cells: correlation with metal ion release. Biol Trace Elem Res 2004;100: Eliades T, Zinelis S, Eliades G, Athanasiou AE. Nickel content of as-received, retrieved, and recycled stainless steel brackets. Am J Orthod Dentofacial Orthop 2002;122: Kao CT, Ding SJ, Min Y, Hsu TC, Chou MY, Huang TH. The cytotoxicity of orthodontic metal bracket immersion media. Eur J Orthod 2007;29: Widu F, Drescher D, Junker R, Bourauel C. Corrosion and biocompatibility of orthodontic wires. J Mater Sci Mater Med 1999;10: Eliades T, Zinelis S, Eliades G, Athanasiou AE. Characterization of as-received, retrieved, and recycled stainless steel brackets. J Orofac Orthop 2003;64: Eliades T, Zinelis S, Papadopoulos MA, Eliades G, Athanasiou AE. Nickel content of as-received and retrieved NiTi and stainless steel archwires: assessing the nickel release hypothesis. Angle Orthod 2004;74: Eliades T, Eliades G, Athanasiou AE, Bradley TG. Surface characterization of retrieved NiTi orthodontic archwires. Eur J Orthod 2000; 22: Eliades T, Eliades G, Watts DC. Intraoral aging of the inner headgear component: a potential biocompatibility concern? Am J Orthod Dentofacial Orthop 2001;119: Harzer W, Schroter A, Gedrange T, Muschter F. Sensitivity of titanium brackets to the corrosive influence of fluoride-containing toothpaste and tea. Angle Orthod 2001;71: Daems J, Celis JP, Willems G. Morphological characterization of as-received and in vivo orthodontic stainless steel archwires. Eur J Orthod 2009;31: House K, Sernetz F, Dymock D, Sandy JR, Ireland AJ. Corrosion of orthodontic appliances should we care? Am J Orthod Dentofacial Orthop 2008;133: Lewis CG, Sunderman FW Jr. Metal carcinogenesis in total joint arthroplasty. Animal models. Clin Orthop Relat Res 1996;(329 Suppl):S Kasprzak KS, Bialkowski K. Inhibition of antimutagenic enzymes, 8-oxo-dGTPases, by carcinogenic metals. Recent developments. J Inorg Biochem 2000;79: Faccioni F, Franceschetti P, Cerpelloni M, Fracasso ME. In vivo study on metal release from fixed orthodontic appliances and DNA damage in oral mucosa cells. Am J Orthod Dentofacial Orthop 2003;124: Uo M, Watari F, Yokoyama A, Matsuno H, Kawasaki T. Tissue reaction around metal implants observed by x-ray scanning analytical microscopy. Biomaterials 2001;22: Eliades T, Athanasiou AE. In vivo aging of orthodontic alloys: implications for corrosion potential, nickel release, and biocompatibility. Angle Orthod 2002;72: Gursoy UK, Sokucu O, Uitto VJ, Aydin A, Demirer S, Toker H, et al. The role of nickel accumulation and epithelial cell proliferation in orthodontic treatment-induced gingival overgrowth. Eur J Orthod 2007;29: Bourauel C, Fries T, Drescher D, Plietsch R. Surface roughness of orthodontic wires via atomic force microscopy, laser specular reflectance, and profilometry. Eur J Orthod 1998;20: Chin MY, Sandham A, de Vries J, van der Mei HC, Busscher HJ. Biofilm formation on surface characterized micro-implants for skeletal anchorage in orthodontics. Biomaterials 2007;28: Gioka C, Eliades T. Materials-induced variation in the torque expression of preadjusted appliances. Am J Orthod Dentofacial Orthop 2004;125: Yokoyama K, Hamada K, Moriyama K, Asaoka K. Degradation and fracture of Ni-Ti superelastic wire in an oral cavity. Biomaterials 2001;22: Drescher D, Bourauel C, Schumacher HA. Frictional forces between bracket and arch wire. Am J Orthod Dentofacial Orthop 1989;96: October 2011 Vol 140 Issue 4 American Journal of Orthodontics and Dentofacial Orthopedics

9 Regis et al Oh KT, Choo SU, Kim KM, Kim KN. A stainless steel bracket for orthodontic application. Eur J Orthod 2005;27: Kusy RP, Whitley JQ. Effects of surface roughness on the coefficients of frictionin model orthodontic systems. J Biomech 1990;23: Kusy RP, Whitley JQ, de Araujo Gurgel J. Comparisons of surface roughnesses and sliding resistances of 6 titanium-based or TMA-type archwires. Am J Orthod Dentofacial Orthop 2004;126: Cacciafesta V, Sfondrini MF, Scribante A, Klersy C, Auricchio F. Evaluation of friction of conventional and metal-insert ceramic brackets in various bracket-archwire combinations. Am J Orthod Dentofacial Orthop 2003;124: Alcock JP, Barbour ME, Sandy JR, Ireland AJ. Nanoindentation of orthodontic archwires: the effect of decontamination and clinical use on hardness, elastic modulus and surface roughness. Dent Mater 2009;25: Edie JW, Andreasen GF, Zaytoun MP. Surface corrosion of nitinol and stainless steel under clinical conditions. Angle Orthod 1981; 51: Grimsdottir MR, Hensten-Pettersen A. Surface analysis of nickel-titanium archwire used in vivo. Dent Mater 1997;13: Petoumeno E, Kislyuk M, Hoederath H, Keilig L, Bourauel C, Jager A. Corrosion susceptibility and nickel release of nickel titanium wires during clinical application. J Orofac Orthop 2008;69: Lin MC, Lin SC, Lee TH, Huang HH. Surface analysis and corrosion resistance of different stainless steel orthodontic brackets in artificial saliva. Angle Orthod 2006;76: Soderholm KJ, Yang MC, Garcea I. Filler particle leachability of experimental dental composites. Eur J Oral Sci 2000;108: Hastings GW. Implant retrieval: material and biological analysis. Biomaterials 1980;1: Fischer-Brandies H, Orthuber W, Es-Souni M, Meyer S. Torque transmission between square wire and bracket as a function of measurement, form and hardness parameters. J Orofac Orthop 2000;61: Berg IC, Rutland MW, Arnebrant T. Lubricating properties of the initial salivary pellicle an AFM study. Biofouling 2003;19: Wichelhaus A, Geserick M, Hibst R, Sander FG. The effect of surface treatment and clinical use on friction in NiTi orthodontic wires. Dent Mater 2005;21: American Journal of Orthodontics and Dentofacial Orthopedics October 2011 Vol 140 Issue 4

Galvanic Corrosion among Different Combination of Orthodontic Archwires and Stainless Steel Brackets

Galvanic Corrosion among Different Combination of Orthodontic Archwires and Stainless Steel Brackets Original Research Galvanic Corrosion among Different Combination of Orthodontic Archwires and Stainless Steel Brackets Farzin Heravi 1, Nima Mokhber 2, Elnaz Shayan 3 1 Dental Materials Research Center,

More information

INFLUENCE OF ARTIFICIAL SALIVA ON NiTi ORTHODONTIC WIRES: A STUDY ON THE SURFACE CHARACTERIZATION

INFLUENCE OF ARTIFICIAL SALIVA ON NiTi ORTHODONTIC WIRES: A STUDY ON THE SURFACE CHARACTERIZATION Materials Science, Vol. 47, No. 6, May, 2012 (Ukrainian Original Vol. 47, No. 6, November December, 2011) INFLUENCE OF ARTIFICIAL SALIVA ON NiTi ORTHODONTIC WIRES: A STUDY ON THE SURFACE CHARACTERIZATION

More information

History of NiTinol. Ni = Nickel Ti = Titanium N = Naval O = Ordnance L = Laboratory

History of NiTinol. Ni = Nickel Ti = Titanium N = Naval O = Ordnance L = Laboratory History of NiTinol Ni = Nickel Ti = Titanium N = Naval O = Ordnance L = Laboratory NiTinol was discovered in a laboratory of the U.S. Marines in 1962 (US-Naval Ordnance Laboratory). The U.S. Marines wanted

More information

ABSTRACT INTRODUCTION /jp-journals

ABSTRACT INTRODUCTION /jp-journals ORIGINAL RESEARCH 10.5005/jp-journals-10024-2102 An in vitro Evaluation of Friction Characteristics of Conventional Stainless Steel and Self-ligating Stainless Steel Brackets with different Dimensions

More information

The effect of surface treatment and clinical use on friction in NiTi orthodontic wires

The effect of surface treatment and clinical use on friction in NiTi orthodontic wires Dental Materials (2005) 21, 938 945 www.intl.elsevierhealth.com/journals/dema The effect of surface treatment and clinical use on friction in NiTi orthodontic wires Andrea Wichelhaus a, *, Marc Geserick

More information

Frictional Properties of Self-Ligating Brackets and Low-Friction Ligatures

Frictional Properties of Self-Ligating Brackets and Low-Friction Ligatures ISSN 1949-0119 2012 Science Publications Frictional Properties of Self-Ligating Brackets and Low-Friction Ligatures Maria Francesca Sfondrini, Danilo Fraticelli, Federico Rosti, Andrea Scribante and Paola

More information

Effect of Sodium Fluoride Mouthwash on the Frictional Resistance of Orthodontic Wires

Effect of Sodium Fluoride Mouthwash on the Frictional Resistance of Orthodontic Wires Original Article Effect of Sodium Fluoride Mouthwash on the Frictional Resistance of Orthodontic Wires Allahyar Geramy 1, Tabassom Hooshmand 2, Tahura Etezadi 3 1 Professor, Dental Research Center, Dentistry

More information

Introducing the Clarity SL Self-Ligating Appliance System

Introducing the Clarity SL Self-Ligating Appliance System Introducing the Clarity SL Self-Ligating Appliance System Lunch and Learn Agenda Clarity SL Brackets Overview Consultation Benefits Bracket Positioning Benefits Bonding Benefits Leveling & Aligning Benefits

More information

Comparative analysis of load/deflection ratios of conventional and heat-activated rectangular NiTi wires

Comparative analysis of load/deflection ratios of conventional and heat-activated rectangular NiTi wires Comparative analysis of load/deflection ratios of conventional and heat-activated rectangular NiTi wires Fabio Schemann-Miguel 1, Flávio Cotrim-Ferreira 2, Alessandra Motta Streva 3, Alexander Viégas de

More information

Frictional Resistance in Various Ceramic Brackets using Archwires of different Alloys, Sizes and Cross sections: A Comparative Study

Frictional Resistance in Various Ceramic Brackets using Archwires of different Alloys, Sizes and Cross sections: A Comparative Study Research Article Frictional Resistance in Various Ceramic Brackets using Archwires of different Alloys, Sizes and Cross sections: Dr Yash Raj Bahadur. 1 Dr Deepak Kumar Agarwal, 2 Dr Ankur Gupta, 3 Dr

More information

Tahereh Hosseinzadeh Nik 1, Tabassom Hooshmand 2*, Habibeh Farazdaghi 3, Arash Mehrabi 1 and Elham S Emadian Razavi 1

Tahereh Hosseinzadeh Nik 1, Tabassom Hooshmand 2*, Habibeh Farazdaghi 3, Arash Mehrabi 1 and Elham S Emadian Razavi 1 Hosseinzadeh Nik et al. Progress in Orthodontics 2013, 14:48 RESEARCH Open Access Effect of chlorhexidine-containing prophylactic agent on the surface characterization and frictional resistance between

More information

Cross-section dimensions and mechanical properties of esthetic orthodontic coated archwires

Cross-section dimensions and mechanical properties of esthetic orthodontic coated archwires ORIGINAL ARTICLE Cross-section dimensions and mechanical properties of esthetic orthodontic coated archwires Dayanne Lopes da Silva, a Claudia Trindade Mattos, b Eduardo Franzotti Sant Anna, c Ant^onio

More information

INDICATIONS. Fixed Appliances are indicated when precise tooth movements are required

INDICATIONS. Fixed Appliances are indicated when precise tooth movements are required DEFINITION Fixed Appliances are devices or equipments that are attached to the teeth, cannot be removed by the patient and are capable of causing tooth movement. INDICATIONS Fixed Appliances are indicated

More information

Friction between different wire bracket combinations in vitro evaluation

Friction between different wire bracket combinations in vitro evaluation www.scielo.br/jaos Friction between different wire bracket combinations in vitro evaluation Tatiana Kelly da Silva FIDALGO 1, Matheus Melo PITHON 2, José Vinicius Bolognesi MACIEL 2, Ana Maria BOLOGNESE

More information

Effects of a diamond-like carbon coating on the frictional properties of orthodontic wires

Effects of a diamond-like carbon coating on the frictional properties of orthodontic wires Original Article Effects of a diamond-like carbon coating on the frictional properties of orthodontic wires Takeshi Muguruma a ; Masahiro Iijima b ; William A. Brantley c ; Itaru Mizoguchi d ABSTRACT Objective:

More information

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective Victory Series Active Self-Ligating Brackets Reliable and Effective Victory Series Active Self-Ligating Brackets Reliability Victory Series SL Brackets are widely accepted as an industry benchmark for

More information

Evaluation of Torsional Strength of Ceramic Brackets Produced by Arch Wire Twisting Moment

Evaluation of Torsional Strength of Ceramic Brackets Produced by Arch Wire Twisting Moment Evaluation of Torsional Strength of Produced by Arch Wire Twisting Moment Raghavendra S. Vhanbatte, 1 Rashmi V. Uddanwadiker, 2 Pallavi Patil 3 1 Postgraduate student, Department of Mechanical Engineering,

More information

Torque efficiency of different archwires in and inch conventional brackets

Torque efficiency of different archwires in and inch conventional brackets Original Article Torque efficiency of different archwires in 0.018- and 0.022-inch conventional brackets Iosif Sifakakis a ; Nikolaos Pandis b ; Margarita Makou c ; Theodore Eliades d ; Christos Katsaros

More information

GAC Archwire Collection

GAC Archwire Collection GAC Archwire Collection All the Metal That Matters Copper to Nickel and All Points In-Between Your choice in archwires is a critical decision in your quest for excellence. GAC has been recognized as a

More information

Comparison of mechanical properties of beta-titanium wires between leveled and unleveled brackets: an in vitro study

Comparison of mechanical properties of beta-titanium wires between leveled and unleveled brackets: an in vitro study Insabralde et al. Progress in Orthodontics 2014, 15:42 RESEARCH ARTICLE Comparison of mechanical properties of beta-titanium wires between leveled and unleveled brackets: an in vitro study Open Access

More information

There has been increased use of self-ligating. Friction between various self-ligating brackets and archwire couples during sliding mechanics

There has been increased use of self-ligating. Friction between various self-ligating brackets and archwire couples during sliding mechanics ORIGINAL ARTICLE Friction between various self-ligating brackets and archwire couples during sliding mechanics Sennay Stefanos, a Antonino G. Secchi, b Guy Coby, c Nipul Tanna, a and Francis K. Mante d

More information

Measurements of the torque moment in various archwire bracket ligation combinations

Measurements of the torque moment in various archwire bracket ligation combinations European Journal of Orthodontics 34 (2012) 374 380 doi:10.1093/ejo/cjr022 Advance Access Publication 13 May 2011 The Author 2011. Published by Oxford University Press on behalf of the European Orthodontic

More information

Characterizing constraining forces in the alignment phase of orthodontic treatment

Characterizing constraining forces in the alignment phase of orthodontic treatment Original Article Characterizing constraining forces in the alignment phase of orthodontic treatment Christopher G. Gibson a ; Feng-Chang Lin b ; Ceib Phillips c ; Alex Edelman d ; Ching-Chang Ko e ABSTRACT

More information

A Comparison of Frictional Resistance between Active and Passive Self-ligating Brackets with Conventional Bracket Systems

A Comparison of Frictional Resistance between Active and Passive Self-ligating Brackets with Conventional Bracket Systems 10.5005/jp-journals-10015-1102 H ORIGINAL Jyothikiran et RESEARCH al A Comparison of Frictional Resistance between Active and Passive Self-ligating Brackets with Conventional Bracket Systems 1 H Jyothikiran,

More information

Advantages. Fantastic. systems. Together with Unitek Lateral Development Archwires, this system puts you in control to

Advantages. Fantastic. systems. Together with Unitek Lateral Development Archwires, this system puts you in control to 3M Self-Ligating Appliances Lateral Development System Treatment Advantages Fantastic Finishes For orthodontic professionals who choose lateral development as a desired treatment outcome, 3M now offers

More information

Advantages. Fantastic. Treatment. Finishes. 3M Self-Ligating Appliances Lateral Development System. Unitek Lateral Development Archwires

Advantages. Fantastic. Treatment. Finishes. 3M Self-Ligating Appliances Lateral Development System. Unitek Lateral Development Archwires 3M Self-Ligating Appliances Lateral Development System Treatment Advantages Fantastic Finishes Adhesive Coated Appliance System Adhesive Coated Appliance System Unitek Lateral Development Archwires For

More information

In Vitro Frictional Forces Generated by Three Different Ligation Methods

In Vitro Frictional Forces Generated by Three Different Ligation Methods Original Article In Vitro Frictional Forces Generated by Three Different Ligation Methods Paola Gandini a ; Linda Orsi b ; Chiara Bertoncini c ; Sarah Massironi d ; Lorenzo Franchi e ABSTRACT Objective:

More information

Frictional resistance in monocrystalline ceramic brackets with conventional and nonconventional elastomeric ligatures

Frictional resistance in monocrystalline ceramic brackets with conventional and nonconventional elastomeric ligatures Galvão et al. Progress in Orthodontics 2013, 14:9 RESEARCH Open Access Frictional resistance in monocrystalline ceramic brackets with conventional and nonconventional elastomeric ligatures Mariana Bulhoes

More information

Canine Extrusion Technique with SmartClip Self-Ligating Brackets

Canine Extrusion Technique with SmartClip Self-Ligating Brackets Canine Extrusion Technique with SmartClip Self-Ligating Brackets Dr. Luis Huanca Ghislanzoni Dr. Luis Huanca received his DDS in 2006 and the MS and Specialist in Orthodontics in 2009 from the University

More information

Variations in surface roughness of seven orthodontic archwires: an SEM-profilometry study

Variations in surface roughness of seven orthodontic archwires: an SEM-profilometry study Original Article THE KOREAN JOURNAL of ORTHODONTICS pissn 2234-7518 eissn 2005-372X http://dx.doi.org/10.4041/kjod.2012.42.3.129 Variations in surface roughness of seven orthodontic archwires: an SEM-profilometry

More information

Introduction. Formerly known as Susan David

Introduction. Formerly known as Susan David European Journal of Orthodontics 20 (1998) 589 596 1998 European Orthodontic Society A comparative in vitro study of the frictional characteristics of two types of self-ligating brackets and two types

More information

Aesthetic Brackets and Accessories

Aesthetic Brackets and Accessories Aesthetic Brackets and Accessories Our Strength is our Service Ortho-Care s version of the MBT prescription shown in this chapter is not claimed to be a duplication, and does not in any way imply endorsement

More information

With judicious treatment planning, the clinical

With judicious treatment planning, the clinical CLINICIAN S CORNER Selecting custom torque prescriptions for the straight-wire appliance Earl Johnson San Francisco, Calif Selecting custom torque prescriptions based on the treatment needs of each patient

More information

ORTHODONTIC WIRES. 1. Commercial name: Contour NiTi Memoflex NiTi Cuper NiTi Stainless Steel TMA Archwire Sequence Springs

ORTHODONTIC WIRES. 1. Commercial name: Contour NiTi Memoflex NiTi Cuper NiTi Stainless Steel TMA Archwire Sequence Springs PAGE 1 of 6 1. Commercial name: Contour NiTi Memoflex NiTi Cuper NiTi Stainless Steel TMA Archwire Sequence Springs 2. Manufacturer: Aditek do Brasil Ltda. Address: Cesário Mota, 14 ZIP: 14140-000 - Cravinhos

More information

The Tip-Edge appliance and

The Tip-Edge appliance and Figure 1: Internal surfaces of the edgewise archwire slot are modified to create the Tip-Edge archwire slot. Tipping surfaces (T) limit crown tipping during retraction. Uprighting surfaces (U) control

More information

GREAT LAKES BRACKETS. Treat with Quality. Treat with Confidence.

GREAT LAKES BRACKETS. Treat with Quality. Treat with Confidence. GREAT LAKES BRACKETS Treat with Quality. Treat with Confidence. BRACKET SYSTEMS DESIGNED BY ORTHODONTISTS, FOR ORTHODONTISTS, GREAT LAKES BRACKETS DELIVER. EVERY SINGLE BRACKET IS HAND INSPECTED AND ENGINEERED

More information

The effect of buccal lingual slot dimension size on third-order torque response

The effect of buccal lingual slot dimension size on third-order torque response European Journal of Orthodontics, 2017, 209 214 doi:10.1093/ejo/cjw043 Advance Access publication 3 June 2016 Original Article The effect of buccal lingual slot dimension size on third-order torque response

More information

BRACKET CHOICES SYSTEM CHOICES. Empower 2 reconvertible molar tubes provide versatility without compromise.

BRACKET CHOICES SYSTEM CHOICES. Empower 2 reconvertible molar tubes provide versatility without compromise. IT S ALL ABOUT CHOICE Empower is the industry s most complete self ligating bracket system, bringing you choice like you ve never had before. The Empower family includes bracket and system options that

More information

Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience

Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience Lotus Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience Designed by an Orthodontist The initial concept for the Lotus Plus Self-Ligating Bracket System started with a simple

More information

The development of modern Oral

The development of modern Oral RESEARCH In Vitro Evaluation of the Removal Force of Abutments in Frictional Dental Implants João César Zielak, DDS, MSc, PhD* Murilo Rorbacker, DDS Rodrigo Gomes, DDS Celso Yamashita, DDS, MSc, PhD Carla

More information

Friction force on brackets generated by stainless steel wire and superelastic wires with and without IonGuard

Friction force on brackets generated by stainless steel wire and superelastic wires with and without IonGuard O r i g i n a l A r t i c l e Friction force on brackets generated by stainless steel wire and superelastic wires with and without IonGuard Luiz Carlos Campos Braga*, Mario Vedovello Filho**, Mayury Kuramae***,

More information

Evaluation of methods of archwire ligation on frictional resistance

Evaluation of methods of archwire ligation on frictional resistance European Journal of Orthodontics 26 (2004) 327 332 European Journal of Orthodontics vol. 26 no. 3 European Orthodontic Society 2004; all rights reserved. Evaluation of methods of archwire ligation on frictional

More information

The role of friction in orthodontics

The role of friction in orthodontics special article The role of friction in orthodontics Mariana Ribeiro Pacheco 1, Wellington Corrêa Jansen 2, Dauro Douglas Oliveira 3 Introduction: Sliding mechanics is widely used during orthodontic treatment.

More information

.::...;;;,:..=.::..==;::..;=== RESEARCH ARTICLE -

.::...;;;,:..=.::..==;::..;=== RESEARCH ARTICLE - (J Ind Orthocl Soc 2002; 35:107-112).::...;;;,:..=.::..==;::..;===------------ RESEARCH ARTICLE - Evaluation of frictional forces of different archwire materials against preadjusted edgewise stainless

More information

Nickel and Chromium Ions Release from Stainless Steel Bracket Immersed in Fluoridated Mouthwash

Nickel and Chromium Ions Release from Stainless Steel Bracket Immersed in Fluoridated Mouthwash from Stainless Steel Bracket Immersed in Fluoridated Mouthwash Ida Bagus Narmada 1 *, Ria Anbar Baya 2, Thalca Hamid 3 1. Departement of Orthodontics, Faculty of Dental Medicine, Universitas Airlangga,

More information

Valeri Petrov 1, Ivana Ilievska 2, Laura Andreeva 1, Valdek Mikli 3, Andrzej Zaleski 4, Angelina Stoyanova-Ivanova 2

Valeri Petrov 1, Ivana Ilievska 2, Laura Andreeva 1, Valdek Mikli 3, Andrzej Zaleski 4, Angelina Stoyanova-Ivanova 2 DOI: 10.18044/Medinform.201741.559 Studying the Effects of Autoclaving Process on Orthodontic Archwires Valeri Petrov 1, Ivana Ilievska 2, Laura Andreeva 1, Valdek Mikli 3, Andrzej Zaleski 4, Angelina

More information

Surface roughness of three types of modern plastic bracket slot floors and frictional resistance

Surface roughness of three types of modern plastic bracket slot floors and frictional resistance Original Article Surface roughness of three types of modern plastic bracket slot floors and frictional resistance Sung-Hwan Choi a ; Da-Young Kang b ; Chung Ju Hwang c ABSTRACT Objective: To quantitatively

More information

Experience with Contemporary Tip-Edge plus Technique A Case Report.

Experience with Contemporary Tip-Edge plus Technique A Case Report. IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861. Volume 13, Issue 3 Ver. I. (Mar. 2014), PP 12-17 Experience with Contemporary Tip-Edge plus Technique A Case

More information

An Accurate Methodology to detect Leaching of Nickel and Chromium Ions in the Initial Phase of Orthodontic Treatment: An in vivo Study

An Accurate Methodology to detect Leaching of Nickel and Chromium Ions in the Initial Phase of Orthodontic Treatment: An in vivo Study JCDP ORIGINAL RESEARCH Leaching 10.5005/jp-journals-10024-1828 of Nickel and Chromium Ions An Accurate Methodology to detect Leaching of Nickel and Chromium Ions in the Initial Phase of Orthodontic Treatment:

More information

Frictional Forces of a Fixed, Esthetic Orthodontic Appliance System

Frictional Forces of a Fixed, Esthetic Orthodontic Appliance System UNLV Theses, Dissertations, Professional Papers, and Capstones 12-1-2012 Frictional Forces of a Fixed, Esthetic Orthodontic Appliance System Brady Michael Nance University of Nevada, Las Vegas, brady.nance@sdm.unlv.edu

More information

Delta Force. Bracket System. Putting you in the driver s seat for ultimate control

Delta Force. Bracket System. Putting you in the driver s seat for ultimate control Delta Force Bracket System Putting you in the driver s seat for ultimate control Variable Force Orthodontics The Delta Force Bracket System incorporates an advanced design that allows you to control the

More information

A New Low-Friction Ligation System

A New Low-Friction Ligation System 2005 JCO, Inc. May not be distributed without permission. www.jco-online.com A New Low-Friction Ligation System ARTURO FORTINI, MD, DDS MASSIMO LUPOLI, MD, DDS VITTORIO CACCIAFESTA, DDS, MS Many factors

More information

Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System.

Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System. SM 3M Health Care Academy Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System. Dr. Gianluigi Fiorillo Dr. Gianluigi Fiorillo received

More information

Effect of Artificial Salvia on Force Values of Two Types of Nickel Titanium Orthodontic

Effect of Artificial Salvia on Force Values of Two Types of Nickel Titanium Orthodontic ISSN: 1812 1217 Effect of Artificial Salvia on Force Values of Two Types of Nickel Titanium Orthodontic Arch Wires Afrah K Al -Hamadani BDS, MSc (Asst Prof.) Enas T Al-Jwary BDS, MSc (Lec.) Dept of Pedod,

More information

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282 tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282 266 . innovative comprehensive efficient. Dentaurum Online Shop shop.dentaurum.com

More information

Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction

Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction Original Article Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction Michael Chung a ; Robert J. Nikolai b ; Ki Beom Kim c ; Donald R. Oliver c ABSTRACT Objective:

More information

An evaluation of two types of nickel-titanium wires in terms of micromorphology and nickel ions release following oral environment exposure

An evaluation of two types of nickel-titanium wires in terms of micromorphology and nickel ions release following oral environment exposure Ghazal et al. Progress in Orthodontics (2015) 16:9 DOI 10.1186/s40510-015-0081-1 RESEARCH Open Access An evaluation of two types of nickel-titanium wires in terms of micromorphology and nickel ions release

More information

Dental Materials Journal 2015; 34(3): Keisuke TOCHIGI, Souichiro ODA and Kazuhito ARAI INTRODUCTION

Dental Materials Journal 2015; 34(3): Keisuke TOCHIGI, Souichiro ODA and Kazuhito ARAI INTRODUCTION Dental Materials Journal 2015; 34(3): 388 393 Influences of archwire size and ligation method on the force magnitude delivered by nickel-titanium alloy archwires in a simulation of mandibular right lateral

More information

A randomized clinical trial to compare three methods of orthodontic space closure

A randomized clinical trial to compare three methods of orthodontic space closure Journal of Orthodontics, Vol. 29, 2002, 31 36 SCIENTIFIC SECTION A randomized clinical trial to compare three methods of orthodontic space closure V. Dixon, M. J. F. Read, K. D. O Brien, H. V. Worthington

More information

Efficient Bonding Protocol for the Insignia Custom Bracket System

Efficient Bonding Protocol for the Insignia Custom Bracket System Efficient Bonding Protocol for the Insignia Custom Bracket System Abstract The Insignia appliance is reverse-engineered from a digital set-up of the prescribed dental alignment. Each bracket configuration,

More information

The Ultimate in Crystal Clear Aesthetics

The Ultimate in Crystal Clear Aesthetics The Ultimate in Crystal Clear Aesthetics SEM x500 SEM x30 SEM x25 SEM x30 Patient Aesthetics The macro photograph shows the clarity of the flawless PURE brackets. Made from pure monocrystalline sapphire

More information

Finite-element analysis of NiTi wire deflection during orthodontic levelling treatment

Finite-element analysis of NiTi wire deflection during orthodontic levelling treatment IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Finite-element analysis of NiTi wire deflection during orthodontic levelling treatment To cite this article: M F Razali et al

More information

SELF LIGATING BRACKETS. 1. Commercial name: EasyClip EasyClip+ Lotus Self-ligating single buccal tubes

SELF LIGATING BRACKETS. 1. Commercial name: EasyClip EasyClip+ Lotus Self-ligating single buccal tubes PAGE 1 of 6 1. Commercial name: EasyClip EasyClip+ Lotus Self-ligating single buccal tubes 2. Manufacturer: Aditek do Brasil Ltda. Address: Cesário Mota, 14 ZIP: 14140-000 - Cravinhos SP - Brazil Phone:

More information

A finite element analysis of the effects of archwire size on orthodontic tooth movement in extraction space closure with miniscrew sliding mechanics

A finite element analysis of the effects of archwire size on orthodontic tooth movement in extraction space closure with miniscrew sliding mechanics Kawamura and Tamaya Progress in Orthodontics (2019) 20:3 https://doi.org/10.1186/s40510-018-0255-8 RESEARCH Open Access A finite element analysis of the effects of archwire size on orthodontic tooth movement

More information

The Science Behind The System Clinical Abstracts, Volume 2

The Science Behind The System Clinical Abstracts, Volume 2 The Science Behind The System Clinical Abstracts, Volume 2 Visit www.damonsystem.com for Damon product information, clinical procedures, seminar registrations and more! 1717 West Collins Avenue, Orange,

More information

Case Report Unilateral Molar Distalization: A Nonextraction Therapy

Case Report Unilateral Molar Distalization: A Nonextraction Therapy Case Reports in Dentistry Volume 2012, Article ID 846319, 4 pages doi:10.1155/2012/846319 Case Report Unilateral Molar Distalization: A Nonextraction Therapy M. Bhanu Prasad and S. Sreevalli Department

More information

Orthodontic bracket slot dimensions as measured from entire bracket series

Orthodontic bracket slot dimensions as measured from entire bracket series Original Article Orthodontic bracket slot dimensions as measured from entire bracket series Paul Brown a ; Warren Wagner b ; Hyden Choi c ABSTRACT Objective: To measure the slot dimensions of an entire

More information

System Orthodontic Treatment Program By Dr. Richard McLaughlin, Dr. John Bennett and Dr. Hugo Trevisi

System Orthodontic Treatment Program By Dr. Richard McLaughlin, Dr. John Bennett and Dr. Hugo Trevisi A Clinical Review of the MBT Versatile+ Appliance System Orthodontic Treatment Program By Dr. Richard McLaughlin, Dr. John Bennett and Dr. Hugo Trevisi Treatment Philosophy of the MBT Appliance System

More information

Controlled Space Closure with a Statically Determinate Retraction System

Controlled Space Closure with a Statically Determinate Retraction System Original Article Controlled Space Closure with a Statically Determinate Retraction System Kwangchul Choy, DDS, MS, PhD a ; Eung-Kwon Pae, DDS, MSc, PhD b ; Kyung-Ho Kim, DDS, MS, PhD c ; Young Chel Park,

More information

FRICTION PRODUCED BY ESTHETIC BRACKETS WITH VARYING LIGATION CHRISTINE KNOX ABENOJA

FRICTION PRODUCED BY ESTHETIC BRACKETS WITH VARYING LIGATION CHRISTINE KNOX ABENOJA FRICTION PRODUCED BY ESTHETIC BRACKETS WITH VARYING LIGATION by CHRISTINE KNOX ABENOJA ANDRE FERREIRA, COMMITTEE CHAIR LIONEL SADOWSKY JOHN BURGESS MARK LITAKER FIROZ RAHEMTULLA A THESIS Submitted to the

More information

Clarity. Form meets function: Discover the science behind a versatile self-ligating bracket system. Aesthetic Orthodontic Solutions

Clarity. Form meets function: Discover the science behind a versatile self-ligating bracket system. Aesthetic Orthodontic Solutions Clarity Aesthetic Orthodontic Solutions Form meets function: Discover the science behind a versatile self-ligating bracket system Armineh Khachatoorian Scientific Affairs Manager 3M Oral Care Rethink your

More information

Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D.

Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D. Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D. Abstract The ribbon arch previously had great popularity and utility early in the 20 th century, but lost its appeal as edgewise

More information

Space Closure Biomechanics Applied Using The MBT System Technique

Space Closure Biomechanics Applied Using The MBT System Technique Space Closure Biomechanics Applied Using The MBT System Technique by Hugo Trevisi, D.D.S. Dr. Hugo Trevisi, São Paulo, Brazil Dr. Hugo Trevisi received his dental degree in 1974 at Lins College of Dentistry

More information

Case Report Orthodontic Replacement of Lost Permanent Molar with Neighbor Molar: A Six-Year Follow-Up

Case Report Orthodontic Replacement of Lost Permanent Molar with Neighbor Molar: A Six-Year Follow-Up Hindawi Case Reports in Dentistry Volume 2017, Article ID 4206435, 9 pages https://doi.org/10.1155/2017/4206435 Case Report Orthodontic Replacement of Lost Permanent Molar with Neighbor Molar: A Six-Year

More information

A Comparative Study between Bond Strength of Rebonded and Recycled Orthodontic Brackets

A Comparative Study between Bond Strength of Rebonded and Recycled Orthodontic Brackets 1 A Comparative Study between Bond Strength of Rebonded and Recycled Orthodontic Brackets ABSTRACT F. Heravi DDS MS*, R. Naseh DDS MS** Introduction: A large number of orthodontists prefer to rebond the

More information

Crowding and protrusion treated by unusual extractions

Crowding and protrusion treated by unusual extractions SM 3M Health Care Academy Crowding and protrusion treated by unusual extractions Gianluigi Fiorillo, DDS Dr. Gianluigi Fiorillo received his degree in Dentistry from La Sapienza University of Rome in 1992

More information

A clinical comparison of three aligning archwires in terms of alignment efficiency: A prospective clinical trial

A clinical comparison of three aligning archwires in terms of alignment efficiency: A prospective clinical trial Original Article A clinical comparison of three aligning archwires in terms of alignment efficiency: A prospective clinical trial Reem Sh. Abdelrahman a ; Kazem S. Al-Nimri b ; Emad F. Al Maaitah c ABSTRACT

More information

Effects of three different mouthwashes on the surface characteristics of nickel-titanium and Stainless steel archwires in orthodontics

Effects of three different mouthwashes on the surface characteristics of nickel-titanium and Stainless steel archwires in orthodontics Original Research Effects of three different mouthwashes on the surface characteristics of nickel-titanium and Stainless steel archwires in orthodontics Maryam Omidkhoda 1, Maryam Poosti 2, Zoha Sahebnasagh

More information

THE MBT VERSATILE+ APPLIANCE SYSTEM

THE MBT VERSATILE+ APPLIANCE SYSTEM THE MBT VERSATILE+ APPLIANCE SYSTEM McLaughlin, Bennett, Trevisi The MBT Versatile+ Appliance System THE DEVELOPMENT OF A TREATMENT MECHANICS AND APPLIANCE PHILOSOPHY The first fully programmed preadjusted

More information

Comparison of Frictional Resistance in Conventional Brackets with Different Stainless Steel Wires

Comparison of Frictional Resistance in Conventional Brackets with Different Stainless Steel Wires ORIGINAL RESEARCH Comparison of Frictional Resistance in Conventional 10.5005/jp-journals-10024-2394 Brackets with Different SS Wires Comparison of Frictional Resistance in Conventional Brackets with Different

More information

Quality. Elegance. Innovation.

Quality. Elegance. Innovation. Quality. Elegance. Innovation. Triumph Orthodontic Instruments Quality and engineering come together for superior performance The Triumph Orthodontic Instrument line provides the orthodontic professional

More information

Honing Damon System Mechanics for the Ultimate in Efficiency and Excellence Jeff Kozlowski, DDS East Lyme, CT

Honing Damon System Mechanics for the Ultimate in Efficiency and Excellence Jeff Kozlowski, DDS East Lyme, CT Honing Damon System Mechanics for the Ultimate in Efficiency and Excellence Jeff Kozlowski, DDS East Lyme, CT Dr. Kozlowski received his DDS degree and certificate in orthodontics from the State University

More information

Correlation Between Frictional Force and Surface Roughness of Orthodontic Archwires

Correlation Between Frictional Force and Surface Roughness of Orthodontic Archwires SCANNING VOL. 37, 399 405 (2015) Wiley Periodicals, Inc. Correlation Between Frictional Force and Surface Roughness of Orthodontic Archwires SAMJIN CHOI, 1 EUN-YOUNG HWANG, 2 HUN-KUK PARK, 1 AND YOUNG-GUK

More information

Tooth movement can occur when the applied

Tooth movement can occur when the applied ONLINE ONLY Comparison of frictional forces during the initial leveling stage in various combinations of self-ligating brackets and archwires with a custom-designed typodont system Tae-Kyung Kim, a Ki-Dal

More information

Effect of Recycling and Autoclave Sterilization on the Unloading Forces of NiTi Closed-Coil Springs: An In Vitro Study

Effect of Recycling and Autoclave Sterilization on the Unloading Forces of NiTi Closed-Coil Springs: An In Vitro Study Momeni Danaei Sh., et al. J Dent Shiraz Univ Med Sci., Dec. 0; (): 8-90. Original Article Effect of Recycling and Autoclave Sterilization on the Unloading Forces of NiTi Closed-Coil Springs: An In Vitro

More information

Mechanical Properties of Different Aesthetic Archwires

Mechanical Properties of Different Aesthetic Archwires Original Article Mechanical Properties of Different Aesthetic Archwires Mehmet Akın, DDS, PhD; 1 Zehra İleri, DDS, PhD; 1 Sertaç Aksakallı, DDS, PhD; 2, * and Faruk Ayhan Basxçiftçi, DDS, PhD 3 ABSTRACT

More information

CLASS II CORRECTION SIMPLIFIED

CLASS II CORRECTION SIMPLIFIED C L A S S I I C O R R E C T O R CLASS II CORRECTION SIMPLIFIED PowerScope 2 delivers simple, efficient Class II correction like you ve never seen before. The latest generation of PowerScope features a

More information

Device Technical File

Device Technical File Device Technical File ContactPro Matrix System Instructions for Use 1.0 Product Description A. For placement of restorations in the posterior area of the oral cavity, the Microbrush ContactPro Sectional

More information

Is nickel titanium superior to multistranded stainless steel wire in aligning crowded lower anteriors? A comparative in-vivo study.

Is nickel titanium superior to multistranded stainless steel wire in aligning crowded lower anteriors? A comparative in-vivo study. IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861. Volume 9, Issue 3 (Jul.- Aug. 2013), PP 47-51 Is nickel titanium superior to multistranded stainless steel

More information

A comparative study of the static and kinetic frictional resistance of titanium molybdenum alloy archwires in stainless steel brackets

A comparative study of the static and kinetic frictional resistance of titanium molybdenum alloy archwires in stainless steel brackets European Journal of Orthodontics 26 (2004) 105 111 European Journal of Orthodontics vol. 26 no. 1 European Orthodontic Society 2004; all rights reserved. A comparative study of the static and kinetic frictional

More information

Instruments Summary. Orthodontic Products - Catalog 03

Instruments Summary. Orthodontic Products - Catalog 03 03 Instruments Instruments Summary Hu-Friedy Instruments 167 Distal end cutters 168 Cutters 168 Utility pliers 169 Lingual pliers 170 Forming pliers 171 Instruments 172 Holders 173 Mosquito and Mathieu

More information

Acknowledgments Introduction p. 1 Objectives p. 1 Goals p. 2 History of Dental Materials p. 3 The Oral Environment p. 4 Characteristics of the Ideal

Acknowledgments Introduction p. 1 Objectives p. 1 Goals p. 2 History of Dental Materials p. 3 The Oral Environment p. 4 Characteristics of the Ideal Preface p. v Acknowledgments p. vii Introduction p. 1 Objectives p. 1 Goals p. 2 History of Dental Materials p. 3 The Oral Environment p. 4 Characteristics of the Ideal Dental Material p. 5 Quality Assurance

More information

The management of impacted

The management of impacted Using a rigid hook and spring auxiliary slid onto the archwire to direct eruption of impacted teeth BY S. JAY BOWMAN, DMD, MSD, AND ALDO CARANO, DR ODONT, MS, SPEC ORTHOD Figure 1: A 12-year-old female

More information

Forsus Class II Correctors as an Effective and Efficient Form of Anchorage in Extraction Cases

Forsus Class II Correctors as an Effective and Efficient Form of Anchorage in Extraction Cases Forsus Class II Correctors as an Effective and Efficient Form of Anchorage in Extraction Cases by Lisa Alvetro, DDS, MSD After receiving her DDS summa cum laude from Ohio State University, Dr. Alvetro

More information

TURN CLASS II INTO SIMPLE CLASS I PATIENTS.

TURN CLASS II INTO SIMPLE CLASS I PATIENTS. TURN CLASS II INTO SIMPLE CLASS I PATIENTS. THE CARRIERE MOTION TM APPLIANCE fast gentle natural The Carriere Philosophy. Fast. Shortens overall treatment time by up to four months as it treats Class II

More information

A Collection of High Quality Orthodontic Instruments

A Collection of High Quality Orthodontic Instruments A Collection of High Quality Orthodontic Instruments User friendly ergonomic design Tenvis instruments starts with a great design, the finest materials and a commitment to quality. Meticulously handcrafted

More information

The Loading And Unloading Properties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System

The Loading And Unloading Properties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System The Loading And Unloading roperties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System Authors: Thomas Mathew MDS Orthodontics, Lecturer, Faculty of Dentistry,

More information

MATERIAL CHARACTERISTICS OF THE ORTHODONTIC ARCHWIRES

MATERIAL CHARACTERISTICS OF THE ORTHODONTIC ARCHWIRES DAAAM INTERNATIONAL SCIENTIFIC BOOK 20 pp. 30-308 CHAPTER 2 MATERIAL CHARACTERISTICS OF THE ORTHODONTIC ARCHWIRES SZUHANEK, C. Abstract Our aim was to investigate the mechanical properties of coated superelastic

More information

The Difference is More than Cosmetic

The Difference is More than Cosmetic The Difference is More than Cosmetic Attract new patients and build your practice. Exclusive Personalized Color- Matching Technology virtually invisible Optimum appearance throughout treatment with non-staining

More information

Presented by Sean Day, Stanley Gelin, Kristin O Connor, Chris Sullivan, and Brian Wilson

Presented by Sean Day, Stanley Gelin, Kristin O Connor, Chris Sullivan, and Brian Wilson Different corrosive effects on hydroxyapatite nanocrystals and amine fluoride-based mouthwashes on dental titanium brackets: a comparative in vitro study Presented by Sean Day, Stanley Gelin, Kristin O

More information