1 Effect of Environment on Fatigue Failure of Controlled Memory Wire Nickel-Titanium Rotary Instruments Ya Shen, DDS, PhD,* Wei Qian, DDS, PhD,* Houman Abtin, BDS,* Yuan Gao, DDS, PhD, and Markus Haapasalo, DDS, PhD* Abstract Introduction: This study examined the fatigue behavior of 2 types of nickel-titanium (NiTi) instruments made from a novel controlled memory NiTi wire (CM wire) under various environment conditions. Methods: Three conventional superelastic NiTi instruments of ProFile (Dentsply Maillefer, Ballaigues, Switzerland), Typhoon (Clinician s Choice Dental Products, New Milford, CT), and DS-SS NEYY (Clinician s Choice Dental Products) and 2 new CM wire instruments of Typhoon CM and DS-SS NEYY CM were subjected to rotational bending at the curvature of 35 in air, deionized water, 17% EDTA, or deionized water after immersion in 6% sodium hypochlorite for 25 minutes, and the number of revolutions of fracture (N f ) was recorded. The fracture surface of all fragments was examined by a scanning electron microscope. The crack-initiation sites and the percentage of dimple area to the whole fracture cross-section were noted. Results: Two new CM Wire instruments yielded an improvement of >4 to 9 times in N f than conventional NiTi files with the same design under various environments (P <.05). The fatigue life of 3 conventional superelastic NiTi instruments was similar under various environments, whereas the N f of 2 new CM Wire instruments was significantly longer in liquid media than in air (P <.05). The vast majority of CM instruments showed multiple, whereas most instruments made from conventional NiTi wire had one crack. The values of the area fraction occupied by the dimple region were significantly smaller on CM NiTi instruments than in conventional NiTi instruments under various environments (P <.05). Conclusions: Within the limitations of this study, the type of NiTi metal alloy (CM files vs conventional superelastic NiTi files) influences the cyclic fatigue resistance under various environments. The fatigue life of CM instruments is longer in liquid media than in air. (J Endod 2012;38: ) Key Words Controlled memory, EDTA, fatigue, nickel-titanium instrument, sodium hypochlorite The advent of nickel-titanium (NiTi) rotary instruments has revolutionized root canal treatment by reducing operator fatigue and the time required to finish the preparation and minimizing procedural errors associated with hand instrumentation (1). Since their introduction (2), an increasing number of NiTi rotary systems have been marketed by various manufacturers. These systems differ from one another in the design of the cutting blades, body taper, and configuration of the file tip. Despite their increasing popularity, a major concern with the use of NiTi rotary instruments is the possibility of unexpected separation in use (3, 4). Two mechanisms that may lead to NiTi rotary fracture, namely cyclic fatigue and torsional overloading, have been described (5). Cyclic fatigue is a result of rotation around a curve with the consequence of repeated extension and compression of metal and, finally, work hardening followed by fracture (3, 6). The combination of chelating agents and sodium hypochlorite (NaOCl) has been advocated as an effective irrigation regimen to remove the organic and inorganic matter during the instrumentation phase (7 9). The solution of 17% EDTA was thought to chemically soften the root canal dentin and dissolve the smear layer (8, 9). Additionally, manufacturers of NiTi instruments recommend paste-type EDTA use as a lubricant during rotary root canal preparation, presumably to reduce the risk of instrument separation. The corrosion of endodontic instruments can occur during chemomechanical preparation, cleaning procedures, chemical disinfection, or sterilization (10 13). The fatigue process begins with crack initiation at the material surface. It is well known that environmental conditions can modify both the crack initiation and propagation processes. The resistance to corrosion of root canal instruments could influence their clinical behavior. It is likely that pitting or crevice corrosion might occur initially and then increase the probability of fatigue failure, thereby altering the fracture mechanism from conventional fatigue failure to corrosion failure. Although the corrosion behavior of NaOCl on NiTi instruments has been studied (12 14), the effect of EDTA on the fatigue behavior of superelastic NiTi instruments has not been reported hitherto. NiTi alloy belongs to the family of intermetallic alloys. This means that alloy can exist in the following various crystallographic forms, with distinct phases and different mechanical properties: austenitic, transformation and martensitic (15). To improve the fracture resistance of NiTi files, manufacturers have developed new manufacturing processes (16 19). Recently, thermal treatment of NiTi alloy is used to optimize the mechanical properties of this alloy (16, 18 23). NiTi rotary instruments (Typhoon From the *Division of Endodontics, Department of Oral Biological and Medical Sciences, Faculty of Dentistry, The University of British Columbia, Vancouver, British Columbia, Canada; and State Key Laboratory of Oral Diseases, West China College and Hospital of Stomatology, Sichuan University, Sichuan, Chengdu, China. Address requests for reprints to Dr Markus Haapasalo, Division of Endodontics, UBC Faculty of Dentistry, 2199 Wesbrook Mall, Vancouver, BC, Canada V6T 1Z3. address: markush@dentistry.ubc.ca /$ - see front matter Copyright ª 2012 American Association of Endodontists. doi: /j.joen Shen et al. JOE Volume 38, Number 3, March 2012

2 CM; Clinician s Choice Dental Products, New Milford, CT) made from a NiTi wire subjected to proprietary thermomechanical processing (CM wire) have been introduced into the market. The CM wire technology makes the NiTi instruments more resistant to cyclic fatigue than non CM wire NiTi instruments (23). The purpose of this study to examine the fatigue behavior of NiTi instruments with 2 different designs made from a novel controlled memory NiTi alloy and conventional NiTi alloy when immersed in various media. Materials and Methods The fatigue testing protocol has been described previously and was reproduced throughout the experimental period (13, 14, 23). Briefly, each NiTi instrument was constrained to a curvature by 3 rigid, stainless-steel pins; a calibrated digital photograph of the curvature was taken. The instrument was then allowed to rotate at 300 rpm (as recommended by the manufacturers) until fracture. The fatigue life, or the total number of revolutions to failure, N f, was recorded. NiTi rotary instruments of.04 taper and size 25 ProFile (PF; Dentsply Maillefer, Ballaigues, Switzerland), Typhoon (TYP; Clinician s Choice Dental Products, New Milford, CT), Typhoon CM (TYP CM; Clinician s Choice Dental Products), DS-SS NEYY (NEYY, Clinician s Choice Dental Products), and DS-SS NEYY CM (NEYY CM, Clinician s Choice Dental Products) were subjected to rotational bending at the curvature of 35 with an 8-mm radius in air (relative humidity, 65%), deionized water, 17% EDTA with a ph of 7.0 (Fisher Scientific, Ottawa, Canada), or deionized water at the temperature of 23 2 C after the working part (16 mm) of the instrument was immersed in 6% NaOCl (Ultra bleach; Shoppers Drug Mart Pharmaprix, Toronto, Canada) at room temperature for 25 minutes. Only a 16-mm length from the tip of the instrument was immersed in the liquid medium during the test to avoid galvanic action between the instrument and its handle. All tested CM wire instruments from Clinician s Choice Dental Products were prototypes, but according to the manufacturer they are identical with Typhoon instruments that will be available soon commercially. Each group included 12 instruments. After the test, the detached fragment was collected, rinsed briefly with deionized water, and mounted so that the fracture surface was parallel to the microscope stage for detailed examination under scanning electron microscopy (SEM; Stereoscan 260; Cambridge Instruments, Cambridge, UK) operating at 5 to 8 kv. The fracture surface of all detached fragments was examined under SEM at various magnifications to identify the crack (s) and to measure the area fraction occupied by the dimple region. A detailed description has been provided elsewhere (23, 24). Data of the number of crack (s) for each specimen were examined by using the Fisher exact test for differences between different alloys and file designs. The region in which the dimple area could be found was outlined on the photomicrograph for each specimen, which was measured with ImageJ 1.4 g software (National Institutes of Health, Bethesda, MD) on each photomicrograph. Pearson correlation coefficients were calculated between N f and the area occupied by the dimple region. The results were analyzed by using univariate analysis or post hoc analysis in software (SPSS for Windows 11.0, SPSS, Chicago, IL) when necessary at a significance level of P <.05. Results A total of 240 NiTi rotary instruments were tested in air, deionized water, 17% EDTA, or deionized water after immersion in 6% NaOCl for 25 minutes (Table 1). The fatigue life of 3 conventional superelastic NiTi instruments (ie, PF, TYP, and NEYY) was similar under various environments (univariate analysis, P >.05), whereas the N f of 2 new CM wire instruments (ie, TYP CM and NEYY CM) was significantly longer in aqueous media than in air (univariate analysis, P <.05). Two CM wire instruments yielded an improvement of >4 to 9 times in N f. For instruments manufactured from conventional NiTi wire, there was no change in N f (post hoc analysis, P <.05). The N f of 2 CM instruments was similar in the different liquid media (ie, water, water after immersion in 6% NaOCl for 25 minutes, and under 17% EDTA; P >.05). The square (NEYY CM) versus triangular (TYP CM) configuration showed a significantly different lifetime on CM Wire under all environments (univariate analysis, P <.05). NEYY instruments had a significantly longer N f than the TYP and ProFile instruments under all environments (univariate analysis, P <.05). Features of the fracture surface were very similar for both CM wire instruments. The vast majority of CM wire instruments showed multiple, whereas most instruments made from conventional NiTi wire had one crack (Table 2; Fisher exact test, P <.05). The values of the area occupied by the dimple region as part of the total surface area of the fractured cross-section were significantly smaller on CM wire NiTi instruments compared with conventional NiTi instruments with the same design under all conditions (Fig. 1; post hoc analysis, P <.05). The N f was highly correlated with the values of the area occupied by the dimple region (r 2 = 0.86). None of the tested materials was susceptible to pitting or crevice corrosion in aqueous media (water, 17% EDTA, or water after immersion in 6% NaOCl for 25 minutes) as evaluated under SEM. Discussion The main aim of this study was to assess the resistance to cyclic fatigue of the novel CM instruments in various solutions. Clinically, irrigant solutions are stored at room temperature and are used in teeth that have been isolated by rubber dam. Only the working part (16 mm from the tip) of instruments was immersed in solution to avoid galvanic corrosion phenomena that are induced when 2 dissimilar metals are TABLE 1. The N f and the Area/Total Cross-section Area on the Fractured Instrument (%) for Each Brand in Various Environments Water after immersion Air Water 17% EDTA in 6% NaOCl N f area (%) ProFile TYP NEYY TYP CM NEYY CM There was a significant difference in the N f and the dimple area/total cross-section between TYP and TYP CM at 4 conditions (P <.05) and between NEYY and NEYY CM at 4 conditions (P <.05). There was a significant difference in the N f of TYP CM between air with aqueous media (P <.05). There was a significant difference in the N f of NEYY CM between air with liquid media (P <.05). JOE Volume 38, Number 3, March 2012 Fatigue of CM NiTi Instrument in Various Media 377

3 TABLE 2. The Number of Crack Origins (% of specimens with at least initiation site) as Observed on the Fracture of Various Instruments Fatigued at Various Environments Air Water 17% EDTA Water after immersion in 6% NaOCl ProFile 7 (58) 5 (42) 10 (83) 2 (17) 8 (67) 4 (33) 10 (83) 2 (17) TYP 11 (92) 1 (8) 12 (100) 0 11 (92) 1 (8) 12 (100) 0 NEYY 9 (75) 3 (25) 10 (83) 2 (17) 9 (75) 3 (25) 10 (83) 2 (17) TYP CM 6 (50) 6 (50) 0 12 (100) 0 12 (100) 0 12 (100) NEYY CM 1 (8) 11 (92) 0 12 (100) 0 12 (100) 0 12 (100) coupled in a corrosive electrolyte (25). EDTA was chosen as the medium in which the test was performed to simulate the clinical situation. NaOCl is the most common irrigant used in root canal treatment. NiTi instruments come into contact with NaOCl during disinfection or when the solution is present in the pulp chamber and root canal during instrumentation. NaOCl was found to be highly corrosive to the stainless-steel pins in a pilot study, necessitating very frequent changes of these pins if the test were to be performed in such conditions. During this study and pilot experiments, it was noticed that the fatigue life of CM instruments was around 25 minutes (7,500 cycles) in water. Therefore, the contact time of all NiTi instruments with higher concentrations (6%) was 25 minutes for static immersion before fatigue testing. The different crystal structures observed in NiTi alloys impart different properties. For instance, the martensite phase is less stiff and more pliable, possessing an elastic modulus of 31 to 35 GPa compared with 84 to 98 GPa for austenite (26). Differential scanning calorimetric analyses have found that the conventional superelastic NiTi wire has the austenite structure at room temperature, whereas CM wire is a mixture of martensite and austenite structure (27). The results of the present study indicated that 2 NiTi instruments made from CM wire were nearly 400% to 600% more resistant to fatigue failure than instruments made from conventional NiTi wire with the same design under various envionments. It is consistent with the view that the martensitic form of NiTi has remarkable fatigue resistance; the crack growth in martensite is slower than in austensite and that stress-induced formation of martensite reduces crack growth rate (28). Both the cross-section factor and material property had a substantial impact on fatigue lifetime (23, 29). This was confirmed by the present findings that square-configuration (ie, NEYY and NEYY CM) Figure 1. A photomicrograph of a fracture surface of (A) NEYY, (B) NEYY CM, (C) Typhoon, and (D) Typhoon CM instruments with the region of fatigue crack propagation and dimple area outlined (dotted line) with (arrows) in deionized water after immersion in 6% NaOCl for 25 minutes. 378 Shen et al. JOE Volume 38, Number 3, March 2012

4 instruments had a significantly longer N f than triangular-configuration (ie, TYP and TYP CM) instruments under all environments. The fatigue of NiTi alloys is sensitive to temperature, both locally and environmentally. A function of the aqueous media in metal fatigue behavior is to carry the heat away from the metal-to-metal contact in order to reduce their operating temperature. On repeated loading, the latent heat of the stress-induced martensitic transformation released can elevate the local temperature, leading to a shortened fatigue life (30). The fatigue behavior of conventional superelastic NiTi instruments testing in air produced similar results as those in aqueous media. The short fatigue life (<1,000 cycles) on conventional superelastic NiTi instruments makes the material withstand few load cycles and is not long enough for heat attack. Hence, the absolute number of revolutions survived in aqueous media does not differ much from that in the air for conventional NiTi instruments. On the other hand, an aqueous medium, water or EDTA, would serve as an effective heat sink for the long fatigue life of the CM instrument; the low-cycle fatigue life therein was distinguishable from that in air. Therefore, it may not be surprising that the fatigue life of CM wire instruments in media (from 5,100 to 7,800 cycles [17 26 minutes]) was more than twice as long as that in air (from 2,400 to 3,600 cycles [8 12 minutes]) according to this study. Notice that when NiTi instruments were tested in a dry condition, there was no coolant or lubricant present. The results of the present study indicated that NiTi instruments made from CM wire were nearly 300% to 800% more resistant to fatigue failure than instruments made from conventional NiTi wire in a dry condition. The sources of heat in a dry condition may include friction between the instrument and pins and internal friction from movement of the interface between phases inside the material. This may partly explain that the standard deviations of the fatigue life of CM wire instruments were higher when tested in air than those in liquid media. Also, for the same reason, the mean length of the fractured segment of all instruments in liquid media was more consistent than in air (data not shown). NiTi alloy has proved to be one of the most biocompatible materials, and it is extremely resistant to corrosion (31). Tobushi et al (32) found that the influence of corrosion fatigue of NiTi wire in water did not appear in the region of low-cycle fatigue. This was confirmed by the present findings. Additionally, a neutral EDTA solution reduces the mineral and noncollagenous protein component of dentin, leading to surface softening but not to erosion of the surface dentin layer during the irrigation (8, 9). The results here indicated that the fatigue lives of all NiTi instruments cycled either under water or 17% EDTA were comparable. These findings are in agreement with previous data (33, 34); the authors attributed the resistance of NiTi in EDTA solution to the ability of EDTA to form the passive film on the metal surface (35). The influence of NaOCl on the corrosion of NiTi instruments has been studied (12 14); despite minor signs of corrosion, they did not appear to involve any significant alteration of the mechanical properties and performance of the instruments. Immersion in 5% NaOCl at a high temperature (37 C and 50 C) for 5 minutes did not reduce the cyclic fatigue resistance of conventional NiTi rotary instruments significantly (36, 37). Such findings lend support to our study here; immersed CM NiTi and conventional NiTi instruments in 6% NaOCl solution for 25 minutes did not reduce the cyclic fatigue fracture resistance. The fractographic appearance of instruments fatigued in NaOCl was very similar to those fatigued in air and water. In addition, no pitting corrosion was formed on the surface (ie, the periphery of the fracture cross-section). The data obtained here cannot be directly extrapolated to clinical conditions. Therefore, conclusions from the present study must be drawn with caution. The fatigue life of a component can be expressed as the number of loading cycles required to initiate a fatigue crack and to propagate the crack to critical size. With continued cyclic loading, the growth of the dominant crack or cracks will continue until the remaining uncracked section of the component no longer can support the load. At this point, the fracture toughness is exceeded, and the remaining cross-section of the material experiences rapid fracture. This rapid overload fracture is the third stage of fatigue failure, which is occupied by the dimple region. In the present study, the area occupied by the dimple region as part of the total surface area of the fracture cross-section was measured for each specimen. The values of this area were significantly smaller on instruments made from CM wire than on instruments made from conventional NiTi wire. Hence, it is not surprising that CM files had fatigue resistance superior to files made from conventional NiTi alloy. To our knowledge, this study is the first to focus on the evaluation of fatigue behavior of the novel CM instruments under various environments. Microstructure often plays an important role in controlling the corrosion behavior of materials. The austenitic NiTi was electrochemically more active than the mixture of austenite and martensite (38). However, both NiTi instruments with austenite structure (ie, PF, TYP, and NEYY) and NiTi instruments with a mixture of martensite and austenite structures (ie, TYP CM and NEYY CM) were not susceptible to corrosion in deionized water, deionized water after immersion in 6% NaOCl for 25 minutes, or 17% EDTA at room temperature. Within the limitations of this study, the type of NiTi metal alloy (CM instruments vs conventional superelastic NiTi instruments) influences cyclic fatigue resistance under various environments. The fatigue life of CM instruments is longer in liquid media than in air. Acknowledgments The authors thank Dentsply Tulsa Dental Specialties and Clinician s Choice Dental Products for donating the files used in this study. The authors deny any conflicts of interest related to this study. References 1. Peters OA. Current challenges and concepts in the preparation of root canal systems: a review. J Endod 2004;35: Walia HM, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of Nitional investigation root canal files. J Endod 1988;14: Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 1997;23: Shen Y, Peng B, Cheung GS. Factors associated with the removal of fractured NiTi instruments from root canal systems. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2004;98: Shen Y, Cheung GS, Peng B, Haapasalo M. Defects in nickel-titanium instruments after clinical use. Part 2: fractographic analysis of fractured surface in a cohort study. J Endod 2009;35: Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod 2009;35: Baumgartner JC, Mader CL. A scanning electron microscopic evaluation of four root canal irrigation regimens. J Endod 1987;13: Zehnder M. Root canal irrigants. J Endod 2006;32: Haapasalo M, Shen Y, Qian W, Gao Y. Irrigation in endodontics. Dent Clin North Am 2010;54: O Hoy PYZ, Messer HH, Palamara JE. The effect of cleaning procedures on fracture properties and corrosion of NiTi files. Int Endod J 2003;36: Parashos P, Messer HH. Rotary NiTi instrument fracture and its consequences. J Endod 2006;32: Peters OA, Roehlike JO, Baumann MA. Effect of immersion in sodium hypochlorite on torque and fatigue resistance of nickel-titanium instruments. J Endod 2007;33: Praisarnti C, Chang JW, Cheung GS. Electropolishing enhances the resistance of nickel-titanium rotary files to corrosion-fatigue failure in hypochlorite. J Endod 2010;36: JOE Volume 38, Number 3, March 2012 Fatigue of CM NiTi Instrument in Various Media 379

5 14. Cheung GS, Shen Y, Darvell BW. Effect of environment on low-cycle fatigue of a nickel-titanium instrument. J Endod 2007;33: Wang FE, Pickart SJ, Alperia HA. Mechanism of the NiTi martensitic transformation and the crystal structures of NiTi-II and NiTi-III phases. J Appl Phys 1972;43: Gambarini G, Grande N, Plotino G, et al. Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods. J Endod 2008;34: Alapati SB, Brantley WA, Iijima M, et al. Metallurgical characterization of a new nickel-titanium wire for rotary endodontic instruments. J Endod 2009;35: Condorelli GG, Bonaccorso A, Smecca E, Sch afer E, Cantatore G, Tripi TR. Improvement of the fatigue resistance of NiTi endodontic files by surface and bulk modifications. Int Endod J 2010;43: Gambarini G, Plotino G, Grande NM, Al-Sudani D, De Luca M, Testarelli L. Mechanical properties of nickel-titanium rotary instruments produced with a new manufacturing technique. Int Endod J 2011;44: Johnson E, Lloyd A, Kuttler S, Namerow K. Comparison between a novel nickeltitanium alloy and 508 nitinol on the cyclic fatigue life of ProFile 25/.04 rotary instruments. J Endod 2008;34: Kramkowski TR, Bahcall J. An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT Series X rotary nickel-titanium files. J Endod 2009;35: Gao Y, Shotton V, Wilkinson K, Phillips G, Johnson WB. Effects of raw material and rotational speed on the cyclic fatigue of ProFile Vortex rotary instruments. J Endod 2010;36: Shen Y, Qian W, Abtin H, Gao Y, Haapasalo M. Fatigue testing of controlled memory wire nickel-titanium rotary instruments. J Endod 2011;37: Cheung GS, Darvell BW. Fatigue testing of a NiTi rotary instrument. Part 2: fractographic analysis. Int Endod J 2007;40: Angelini E, Zucchi F, Caputo A. Degradation processes on metallic surfaces. In: Barbucci R, ed. Integrated Biomaterial Science. Dordrecht: Kluwer Academic Plenum Publishers; 2002: Kusy RP. A review of contemporary archwires: their properties and characteristics. Angle Orthod 1997;67: Shen Y, Zhou H, Zheng Y, Campbell L, Peng B, Markus H. Metallurgical characterization of controlled memory wire nickel-titanium rotary instruments. J Endod 2011;37: Mckelvey AL, Ritchie RO. Fatigue-crack growth behavior in the superelastic and shape-memory alloy Nitinol. Metall Mater Trans A 2001;32A: Zhang EW, Cheung GSP, Zheng YF. A mathematical model for describing the mechanical behaviour of root canal instruments. Int Endod 2011;44: Prymak O, Klocke A, Kahl-Nieke B, Epple M. Fatigue of orthodontic nickel-titanium (NiTi) wires in different fluids under constant mechanical stress. Mater Sci Eng A 2004;378: Hodgson DE, Wu MH, Biermann RJ. Shape memory alloys. In: Davis JR, Allen P, Lampman SR, et al., eds. Properties and Selection: Nonferrous Alloys and Special- Purpose Materials. 10th edn. Materials Park, OH: ASM International; 1990: Tobushi H, Nakahara T, Shimeno Y, Hashimoto T. Low-cycle fatigue of TiNi shape memory alloy and formulation of fatigue life. J Eng Mater Technol 2000;122: Darabara M, Bourithis L, Zinelis S, Papadimitriou GD. Susceptibility to localized corrosion of stainless steel and NiTi endodontic instruments in irrigating solutions. Int Endod J 2004;37: Bonaccorso A, Tripi TR, Rondelli G, Condorelli GG, Cantatore G, Sch afer E. Pitting corrosion resistance of nickel-titanium rotary instruments with different surface treatments in seventeen percent ethylenediaminetetraacetic Acid and sodium chloride solutions. J Endod 2008;34: Reinhard G, Radtke M, Rammelt U. The role of the salts of weak acids in the chemical passivation of iron and steel in aqueous solutions. Corrosion Science 1992;33: Berutti E, Angelini E, Rigolone M, Migliaretti G, Pasqualini D. Influence of sodium hypochlorite on fracture properties and corrosion of ProTaper Rotary instruments. Int Endod J 2006;39: Pedulla E, Grande NM, Plotino G, Pappalardo A, Rapisarda E. Cyclic fatigue resistance of three different nickel-titanium instruments after immersion in sodium hypochlorite. J Endod 2010;37: Wang J, Li N, Han EH, Ke W. Effect of ph, temperature and Cl- concentration on electrochemical behavior of NiTi shape memory alloy in artificial saliva. J Mater Sci Mater Med 2006;17: Shen et al. JOE Volume 38, Number 3, March 2012

The introduction of nickel-titanium (NiTi) to endodontics over 2 decades ago has

The introduction of nickel-titanium (NiTi) to endodontics over 2 decades ago has ProFile Vortex Instruments after Clinical Use: A Metallurgical Properties Study Ya Shen, DDS, PhD,* Jeffrey M. Coil, DMD, MSD, PhD,* Hui-min Zhou, DDS, PhD, Esther Tam, DDS,* Yu-feng Zheng, PhD, and Markus

More information

The nickel-titanium (NiTi) alloy, which consists of nickel and titanium in a nearly

The nickel-titanium (NiTi) alloy, which consists of nickel and titanium in a nearly Metallurgical Characterization of Controlled Memory Wire Nickel-Titanium Rotary Instruments Ya Shen, DDS, PhD,* Hui-min Zhou, PhD, Yu-feng Zheng, PhD, Les Campbell, DDS,* Bin Peng, DDS, PhD, and Markus

More information

Effect of autoclave sterilization on the cyclic fatigue resistance of thermally treated Nickel Titanium instruments

Effect of autoclave sterilization on the cyclic fatigue resistance of thermally treated Nickel Titanium instruments doi:10.1111/iej.12550 Effect of autoclave sterilization on the cyclic fatigue resistance of thermally treated Nickel Titanium instruments D. Zhao 1, Y. Shen 2,3, B. Peng 1 & M. Haapasalo 2 1 Department

More information

Over the past 2 decades, nickel-titanium (NiTi) instruments have become an important

Over the past 2 decades, nickel-titanium (NiTi) instruments have become an important Current Challenges and Concepts of the Thermomechanical Treatment of Nickel-Titanium Instruments Ya Shen, DDS, PhD,* Hui-min Zhou, DDS, PhD, Yu-feng Zheng, PhD, Bin Peng, DDS, PhD, and Markus Haapasalo,

More information

A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments

A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments doi:10.1111/j.1365-2591.2010.01838.x A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments G. S. P. Cheung 1, E. W. Zhang 2 & Y. F. Zheng 2 1 Area

More information

Surface properties of nickel-titanium rotary instruments

Surface properties of nickel-titanium rotary instruments MATERIALS AND INSTRUMENTS 45 Antonio Bonaccorso, Teresa Roberta Tripi, Giuseppe Cantatore, Guglielmo Guido Condorelli Surface properties of nickel-titanium rotary instruments Key words autoclave sterilisation,

More information

HyFlex nickel titanium rotary instruments after clinical use: metallurgical properties

HyFlex nickel titanium rotary instruments after clinical use: metallurgical properties doi:10.1111/iej.12049 HyFlex nickel titanium rotary instruments after clinical use: metallurgical properties Y. Shen 1, J. M. Coil 1, H. Zhou 2, Y. Zheng 3 & M. Haapasalo 1 1 Division of Endodontics, Department

More information

Hend M. Abou El Nasr * and Shehab El Din M. Saber ** ABSTRACT صور 4 لون

Hend M. Abou El Nasr * and Shehab El Din M. Saber ** ABSTRACT صور 4 لون رقم املقالة الترقيم عدد الصفحات ص األلوان صور 1 لون صور 4 لون مالحظات 4 137 Metallurgical and Mechanical Properties of Nickel Titanium Instruments Made of Fire-Wire, Controlled Memory Wire, M-Wire, and

More information

Comparison of cyclic fatigue resistance of original and counterfeit rotary instruments

Comparison of cyclic fatigue resistance of original and counterfeit rotary instruments Ertas et al. BioMedical Engineering OnLine 2014, 13:67 RESEARCH Open Access Comparison of cyclic fatigue resistance of original and counterfeit rotary instruments Huseyin Ertas 1*, Ismail Davut Capar 1,

More information

Microstructural and mechanical characterizations of new Ni-Ti endodontic instruments

Microstructural and mechanical characterizations of new Ni-Ti endodontic instruments MATEC Web of Conferences 33, 03005 ( 2015) DOI: 10.1051/ matecconf/ 20153303005 C Owned by the authors, published by EDP Sciences, 2015 Microstructural and mechanical characterizations of new Ni-Ti endodontic

More information

The use of sodium hypochlorite (NaOCl) to irrigate root canals is currently the gold

The use of sodium hypochlorite (NaOCl) to irrigate root canals is currently the gold Effect of Immersion in Sodium Hypochlorite on Torque and Fatigue Resistance of Nickel-Titanium Instruments Ove A. Peters, PD, Dr. med dent MS, Jan O. Roehlike, Dr. med dent, Michael A. Baumann, Prof. Dr.

More information

Effect of Cryogenic Treatment on Fracture Resistance of Nickel Titanium Rotary Instruments a Systematic Review

Effect of Cryogenic Treatment on Fracture Resistance of Nickel Titanium Rotary Instruments a Systematic Review IJCRR Section: Healthcare Sci. Journal Impact Factor 4.016 ICV: 71.54 Review Article Effect of Cryogenic Treatment on Fracture Resistance of Nickel Titanium Rotary Instruments a Systematic Review Kumar

More information

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments Key Engineering Materials Online: 2005-06-15 ISSN: 1662-9795, Vols. 288-289, pp 603-606 doi:10.4028/www.scientific.net/kem.288-289.603 2005 Trans Tech Publications, Switzerland Torsion Property and Cyclic

More information

Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments

Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments Research article ISSN 2234-7658 (print) / ISSN 2234-7666 (online) Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments Koray Yılmaz 1, Taha Özyürek 2 * 1 Samsun Oral and Dental

More information

thinking beyond cyclic fatigue Testing standards for endo files vary and there are other qualities to be concerned with, too

thinking beyond cyclic fatigue Testing standards for endo files vary and there are other qualities to be concerned with, too thinking beyond cyclic fatigue Testing standards for endo files vary and there are other qualities to be concerned with, too In endodontic literature, significant attention has been devoted to and much

More information

Dental Research Journal

Dental Research Journal Dental Research Journal Original Article A novel approach to determine the effect of irrigation on temperature and failure of Ni-Ti endodontic rotary files Sayed Ali Mousavi 1, Vida Kargar-Dehnavi 2, Sayed

More information

Microestructure alterations of rotary files after multiple simulated operative procedures

Microestructure alterations of rotary files after multiple simulated operative procedures Journal section: Clinical and Experimental Dentistry Publication Types: Research doi:10.4317/medoral.15.e658 Microestructure alterations of rotary files after multiple simulated operative procedures Arlinda

More information

An In Vitro Comparison of Cyclic Fatigue of Profile Vortex and Endosequence Rotary Nickel- Titanium Files

An In Vitro Comparison of Cyclic Fatigue of Profile Vortex and Endosequence Rotary Nickel- Titanium Files Virginia Commonwealth University VCU Scholars Compass Theses and Dissertations Graduate School 2011 An In Vitro Comparison of Cyclic Fatigue of Profile Vortex and Endosequence Rotary Nickel- Titanium Files

More information

Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments?

Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments? doi:10.1111/iej.12166 Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments? P. Kiefner 1, M. Ban 2 & G. De-Deus 3 1 Praxis f ur Zahnheilkunde/Endodontie, Stuttgart;

More information

Influence of sodium hypochlorite on fracture properties and corrosion of ProTaper Rotary instruments

Influence of sodium hypochlorite on fracture properties and corrosion of ProTaper Rotary instruments doi: 10.1111/j.1365-2591.2006.01134.x Influence of sodium hypochlorite on fracture properties and corrosion of ProTaper Rotary instruments E. Berutti 1, E. Angelini 2, M. Rigolone 1, G. Migliaretti 3 &

More information

REPORT DOCUMENTATION PAGE

REPORT DOCUMENTATION PAGE REPORT DOCUMENTATION PAGE Form Approved OMB No. 0704-0188 Public reporting burden for this collection of information is estimated to average 1 hour per response, including the time for reviewing instructions,

More information

Tools for Endodontic Success

Tools for Endodontic Success CLEANING & SHAPING Tools for Endodontic Success A Look Inside: C-Files Gates Glidden Drills K-Files Mid-Sized K Files To Dye For Flexicut Files Sodium Hypochlorite Hedstrom Files C.L. Canal Lubricant Reamers

More information

CIC Edizioni Internazionali. In vitro resistance to fracture of two nickel-titanium rotary instruments made with different thermal treatments

CIC Edizioni Internazionali. In vitro resistance to fracture of two nickel-titanium rotary instruments made with different thermal treatments Original article In vitro resistance to fracture of two nickel-titanium rotary instruments made with different thermal treatments Gabriele Miccoli DDS Gianfranco Gaimari DDS, PhD Marco Seracchiani DDS

More information

Differential Scanning Calorimetric (DSC) Study of New and Sterilized Nickel-Titanium Rotary Endodontic Instruments

Differential Scanning Calorimetric (DSC) Study of New and Sterilized Nickel-Titanium Rotary Endodontic Instruments Marquette University e-publications@marquette Master's Theses (2009 -) Dissertations, Theses, and Professional Projects Differential Scanning Calorimetric (DSC) Study of New and Sterilized Nickel-Titanium

More information

Effect From Surface Treatment of Nickel-Titanium Rotary Files on the Fracture Resistance

Effect From Surface Treatment of Nickel-Titanium Rotary Files on the Fracture Resistance SCANNING VOL. 37, 82 87 (2015) Wiley Periodicals, Inc. Effect From Surface Treatment of Nickel-Titanium Rotary Files on the Fracture Resistance BO HOON KIM, 1 JUNG-HONG HA, 2 WOO CHEOL LEE, 3 SANG-WON

More information

Analysis of Surface Characteristics of ProTaper Universal and ProTaper Next Instruments by Scanning Electron Microscopy

Analysis of Surface Characteristics of ProTaper Universal and ProTaper Next Instruments by Scanning Electron Microscopy Journal section: Operative Dentistry and Endodontics Publication Types: Research doi:10.4317/jced.54049 http://dx.doi.org/10.4317/jced.54049 Analysis of Surface Characteristics of ProTaper Universal and

More information

Last Generation Endodontic Instruments. A Study of their. Drs. Siragusa and Racciatti are affiliated with the Department of

Last Generation Endodontic Instruments. A Study of their. Drs. Siragusa and Racciatti are affiliated with the Department of Last Generation Endodontic Instruments. A Study of their Efficacy in Apical Shaping M. Siragusa, DDS, PhD, and G. Racciatti, DDS Drs. Siragusa and Racciatti are affiliated with the Department of Endodontics,

More information

Cyclic fatigue resistance and fractographic analysis of RaCe and ProTaper rotary NiTi instruments

Cyclic fatigue resistance and fractographic analysis of RaCe and ProTaper rotary NiTi instruments ORIGINAL ARTICLE Shahram Azimi1* DDS, MS, Parisa Delvari1 DDS, Hamid Cyrus Hajarian2 MD, DDS, MS, FICS, Mohammad Ali Saghiri3 BS, M.S, PhD, Kasra Karamifar4 DDS, Mehrdad Lotfi5 DMD, MSc Cyclic fatigue

More information

The Effect of Surface Defects on the Cyclic Fatigue Fracture of HEROShaper Ni-Ti rotary files in a Dynamic Model: A Fractographic Analysis

The Effect of Surface Defects on the Cyclic Fatigue Fracture of HEROShaper Ni-Ti rotary files in a Dynamic Model: A Fractographic Analysis 대한치과보존학회지 : Vol. 32, No. 2, 2007 The Effect of Surface Defects on the Cyclic Fatigue Fracture of HEROShaper Ni-Ti rotary files in a Dynamic Model: A Fractographic Analysis Jung-Kyu Lee 1, Eui-Sung Kim

More information

INFLUENCE OF ARTIFICIAL SALIVA ON NiTi ORTHODONTIC WIRES: A STUDY ON THE SURFACE CHARACTERIZATION

INFLUENCE OF ARTIFICIAL SALIVA ON NiTi ORTHODONTIC WIRES: A STUDY ON THE SURFACE CHARACTERIZATION Materials Science, Vol. 47, No. 6, May, 2012 (Ukrainian Original Vol. 47, No. 6, November December, 2011) INFLUENCE OF ARTIFICIAL SALIVA ON NiTi ORTHODONTIC WIRES: A STUDY ON THE SURFACE CHARACTERIZATION

More information

Comparative analysis of various corrosive environmental conditions for NiTi rotary files

Comparative analysis of various corrosive environmental conditions for NiTi rotary files Comparative analysis of various corrosive environmental conditions for NiTi rotary files Comparative analysis of various corrosive environmental conditions for NiTi rotary files Ji-Wan Yum, Jeong-Kil Park,

More information

Nickel-titanium (NiTi) endodontic files are particularly helpful for successful shaping

Nickel-titanium (NiTi) endodontic files are particularly helpful for successful shaping SEM Observations and Differential Scanning Calorimetric Studies of New and Sterilized Nickel-Titanium Rotary Endodontic Instruments Georgia B. Alexandrou,* Konstantinos Chrissafis, Leonidas P. Vasiliadis,*

More information

Electrochemical corrosion assessment of RaCe and Mtwo rotary nickle-titanium instruments after clinical use and sterilization

Electrochemical corrosion assessment of RaCe and Mtwo rotary nickle-titanium instruments after clinical use and sterilization Journal section: Clinical and Experimental Dentistry Publication Types: Research doi:10.4317/medoral.17413 http://dx.doi.org/doi:10.4317/medoral.17413 Electrochemical corrosion assessment of RaCe and Mtwo

More information

WaveoneandoneShapeFilesSurvivalinSeverelyCurvedArtificialCanals

WaveoneandoneShapeFilesSurvivalinSeverelyCurvedArtificialCanals : J Dentistry and Otolaryngology Volume 14 Issue 4 Version 1.0 Year 2014 Type: Double Blind Peer Reviewed International Research Journal Publisher: Global Journals Inc. (USA) Online ISSN: 2249-4618 & Print

More information

Rotary Fatigue Testing to Determine the Fatigue Life of NiTi alloy Wires: An Experimental and Numerical Analisys

Rotary Fatigue Testing to Determine the Fatigue Life of NiTi alloy Wires: An Experimental and Numerical Analisys Available online at www.sciencedirect.com Procedia Engineering 00 (2016) 000 000 www.elsevier.com/locate/procedia 1st International Conference on Structural Integrity Rotary Fatigue Testing to Determine

More information

Susceptibility to localized corrosion of stainless steel and NiTi endodontic instruments in irrigating solutions

Susceptibility to localized corrosion of stainless steel and NiTi endodontic instruments in irrigating solutions Susceptibility to localized corrosion of stainless steel and NiTi endodontic instruments in irrigating solutions M. Darabara 1, L. Bourithis 1, S. Zinelis 2 & G. D. Papadimitriou 1 1 School of Mining and

More information

Comparison of the flexibility and torsional resistance of nickel-titanium rotary instruments.

Comparison of the flexibility and torsional resistance of nickel-titanium rotary instruments. original article Comparison of the flexibility and torsional resistance of nickel-titanium rotary instruments Weber Schmidt Pereira Lopes 1 Hélio Pereira Lopes 2 Carlos Nelson Elias 3 Marcelo Mangelli

More information

Glassman Webinar PPT 7/18/2014. Nickel Titanium Clinical Performance (2012) Depends on: three main factors :

Glassman Webinar PPT 7/18/2014. Nickel Titanium Clinical Performance (2012) Depends on: three main factors : Nickel Titanium Clinical Performance (2012) Depends on: three main factors : 1. Design 2. Metallurgy-Alloy/Manufacturing 3. Methods of use All of them are nowadays strictly interconnected, allowing manufacturers

More information

Lifespan of One Shape Files used in Severely Curved Canals in Resin Blocks and Extracted Teeth

Lifespan of One Shape Files used in Severely Curved Canals in Resin Blocks and Extracted Teeth Lifespan of One Shape Files used in Severely Curved Canals in Resin Blocks and Extracted Teeth Emilia Karova Assistant Professor, PhD, Department of Conservative Dentistry, Faculty of Dental Medicine,

More information

Cyclic Fatigue Resistance of Three Single-use NiTi Instruments after Immersion in Sodium Hypochlorite

Cyclic Fatigue Resistance of Three Single-use NiTi Instruments after Immersion in Sodium Hypochlorite IJEDS Cyclic Fatigue Resistance of Three Single-use NiTi Instruments after 10.5005/jp-journals-10029-1074 Immersion in Sodium Hypochlorite ORIGINAL ARTICLE Cyclic Fatigue Resistance of Three Single-use

More information

Nickel-titanium (NiTi) rotary instruments were introduced to improve root canal

Nickel-titanium (NiTi) rotary instruments were introduced to improve root canal Use of Nickel-Titanium Rotary PathFile to Create the Glide Path: Comparison With Manual Preflaring in Simulated Root Canals Elio Berutti, MD, DDS,* Giuseppe Cantatore, MD, DDS, Arnaldo Castellucci, MD,

More information

Comparative analysis of torsional and bending behavior through finiteelement models of five Nickel-Titanium endodontic instruments

Comparative analysis of torsional and bending behavior through finiteelement models of five Nickel-Titanium endodontic instruments Submitted to Journal of Endodontics, July 2009 Comparative analysis of torsional and bending behavior through finiteelement models of five Nickel-Titanium endodontic instruments Valérie Chevalier *, Reza

More information

Effective cleaning and shaping of the root

Effective cleaning and shaping of the root A new concept in canal preparation Dr. Ghassan Yared discusses canal preparation with only one reciprocating instrument without prior hand filing Effective cleaning and shaping of the root canal system

More information

Geometric and dimensional characteristics of simulated curved canals prepared with protaper instruments

Geometric and dimensional characteristics of simulated curved canals prepared with protaper instruments www.scielo.br/jaos Geometric and dimensional characteristics of simulated curved canals prepared with protaper instruments Renata de Castro MARTINS 1, Maria Guiomar de Azevedo BAHIA 2, Vicente Tadeu Lopes

More information

Fracture and deformation of ProTaper Next instruments after clinical use

Fracture and deformation of ProTaper Next instruments after clinical use Journal section: Operative Dentistry and Endodontics Publication Types: Research doi:10.4317/jced.54910 http://dx.doi.org/10.4317/jced.54910 instruments after clinical use Gema Fernández-Pazos, Benjamín

More information

Torsion and Bending Properties of EdgeEndo Files

Torsion and Bending Properties of EdgeEndo Files Marquette University e-publications@marquette Master's Theses (2009 -) Dissertations, Theses, and Professional Projects Torsion and Bending Properties of EdgeEndo Files Chad Hansen Marquette University

More information

This in-vitro investigation was designed to form a simulated clinical use using RaCe

This in-vitro investigation was designed to form a simulated clinical use using RaCe Cairo Dental Journal (25) Number (3), 367:374 September, 2009 A Comparative Study of Deformation and Fracture of Two Rotary Nikel Titanium Instruments after Simulated Clinical Use (In vitro) El-Maghraby

More information

WaveOne Gold reciprocating instruments: clinical application in the private practice: Part 2

WaveOne Gold reciprocating instruments: clinical application in the private practice: Part 2 C L I N I C A L WaveOne Gold reciprocating instruments: clinical application in the private practice: Part 2 Peet van der Vyver 1 and Martin Vorster 2 1 Department of Odontology, School of Dentistry, University

More information

In vitro evaluation of root canal preparation with plastic endodontic rotary finishing file - A SEM study

In vitro evaluation of root canal preparation with plastic endodontic rotary finishing file - A SEM study Original Research In vitro evaluation of root canal preparation with plastic endodontic rotary finishing file - A SEM study ASHUTOSH * ASEEM P. TIKKU ** ANIL CHANDRA ** PROMILA VERMA *** RAKESH K. YADAV

More information

AN INVESTIGATION OF PHASE TRANSFORMATION MECHANISMS FOR NICKEL-TITANIUM ROTARY ENDODONTIC INSTRUMENTS DISSERTATION

AN INVESTIGATION OF PHASE TRANSFORMATION MECHANISMS FOR NICKEL-TITANIUM ROTARY ENDODONTIC INSTRUMENTS DISSERTATION AN INVESTIGATION OF PHASE TRANSFORMATION MECHANISMS FOR NICKEL-TITANIUM ROTARY ENDODONTIC INSTRUMENTS DISSERTATION Presented in Partial Fulfillment of the Requirements for the Degree Doctor of Philosophy

More information

Comparison of cyclic fatigue resistance of XP-endo Shaper, HyFlex CM, FlexMaster and Race instruments

Comparison of cyclic fatigue resistance of XP-endo Shaper, HyFlex CM, FlexMaster and Race instruments Journal of Dental Research, Dental Clinics, Dental Prospects Original Article Comparison of cyclic fatigue resistance of XP-endo Shaper, HyFlex CM, FlexMaster and Race instruments Mehmet Adiguzel* Ipek

More information

A WHOLE NEW DIMENSION OF TOTAL PERFORMANCE.

A WHOLE NEW DIMENSION OF TOTAL PERFORMANCE. A WHOLE NEW DIMENSION OF TOTAL PERFORMANCE. GT Series X Endo System HAS IT ALL. In the past, rotary file design focused on either speed, respecting canal morphology or resisting file separation. One focus,

More information

Evaluation of selected mechanical properties of NiTi rotary glide path files manufactured from controlled memory wires

Evaluation of selected mechanical properties of NiTi rotary glide path files manufactured from controlled memory wires Dental Materials Journal 2018; : Evaluation of selected mechanical properties of NiTi rotary glide path files manufactured from controlled memory wires Miki NISHIJO 1, Arata EBIHARA 1, Daisuke TOKITA 1,

More information

IN VITRO EVALUATION OF DENTINAL

IN VITRO EVALUATION OF DENTINAL ORIGINAL RESEARCH IN VITRO EVALUATION OF DENTINAL MICROCRACK FORMATION DURING ROOT CANAL PREPARATIONS BY DIFFERENT NITI SYSTEMS Rahman H*, Chandra A**, Singh S # * Assistant Professor, Department of Conservative

More information

Failure of ProFile instruments used with air, high torque control, and low torque control motors

Failure of ProFile instruments used with air, high torque control, and low torque control motors Failure of ProFile instruments used with air, high torque control, and low torque control motors Ghassan Yared, DDS, MSc, a and Philippe Sleiman, DDS, b Toronto, Ontario, Canada, and Beirut, Lebanon UNIVERSITY

More information

Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students

Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students Journal section: Clinical and Experimental Dentistry Publication Types: Research doi:10.4317/jced.50735 http://dx.doi.org/10.4317/jced.50735 Failure of ProTaper rotary Ni-Ti used by undergraduate students

More information

Comparison of the Effects of Three Different Nickel titanium Rotary Instruments on the Fracture Resistance of Obturated Roots: An in vitro Study

Comparison of the Effects of Three Different Nickel titanium Rotary Instruments on the Fracture Resistance of Obturated Roots: An in vitro Study Saimanaaz AH Shaikh et al ORIGINAL RESEARCH 10.5005/jp-journals-10031-1191 Comparison of the Effects of Three Different Nickel titanium Rotary Instruments on the Fracture Resistance of Obturated Roots:

More information

Improve Your Root Canal Preparations with Adaptive Motion Technology

Improve Your Root Canal Preparations with Adaptive Motion Technology Improve Your Root Canal Preparations with Adaptive Motion Technology During root canal preparations, clinicians may experience a variety of challenges. However, there are several difficulties that have

More information

Corresponding Author:Dr.Sneha Vaidya 3

Corresponding Author:Dr.Sneha Vaidya 3 IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 16, Issue 11 Ver. VII (Nov. 2017), PP 75-81 www.iosrjournals.org Efficacy of Endoactivator Irrigation

More information

Factors Affecting Intracanal Instrument Fracture

Factors Affecting Intracanal Instrument Fracture Factors Affecting Intracanal Instrument Fracture 2 Christos Boutsioukis and Theodor Lambrianidis 2.1 Introduction Root canal treatment (RCT) may require the use of a variety of instruments including files

More information

ENDODONTIC IRRIGATION SYSTEM. Saves time and improves outcomes

ENDODONTIC IRRIGATION SYSTEM. Saves time and improves outcomes ENDODONTIC IRRIGATION SYSTEM Saves time and improves outcomes ENDODONTIC IRRIGATING SOLUTIONS When it comes to irrigating, CanalPro Endodontic Solutions are your best choice for successfull treatment.

More information

Endodontic instrument fracture: causes and prevention

Endodontic instrument fracture: causes and prevention Endodontic instrument fracture: causes and prevention M. B. McGuigan, 1 C. Louca 1 and H. F. Duncan* 2 IN BRIEF Reveals incidence of stainless steel and nickel-titanium file fracture. Considers the instrument,

More information

Instructions for use: FlexMaster. the rotary NiTi-System for every case. Endo Easy Efficient

Instructions for use: FlexMaster. the rotary NiTi-System for every case. Endo Easy Efficient Instructions for use: FlexMaster the rotary NiTi-System for every case Endo Easy Efficient 2 Content 1. Why change to the FlexMaster system? 3 2. Why nickel-titanium? 3 3. Taper, what does it mean? 4 4.

More information

Creating a glide path for rotary NiTi instruments: part two

Creating a glide path for rotary NiTi instruments: part two Clinical Creating a glide path for rotary NiTi instruments: part two Peet van der Vyver 1 Introduction In part one of this series the author discussed the rationale for the preparation of a glide path

More information

Fracture Resistance of Wave-one Rotary File Used In Reciprocating Motion

Fracture Resistance of Wave-one Rotary File Used In Reciprocating Motion Original Research Fracture Resistance of Wave-one Rotary File Used In Reciprocating Motion AUTHORS: Manish Shah*, Mahendra Patel**, Aesha Amin***, Pranav Kharod****, Chintan Joshi***** ABSTRACT: Introduction:

More information

Case Report. July 2015; Vol. 12, No. 7. Vineet Agrawal 1, Sonali Kapoor 2, Mukesh Patel 3

Case Report.  July 2015; Vol. 12, No. 7. Vineet Agrawal 1, Sonali Kapoor 2, Mukesh Patel 3 Case Report Ultrasonic Technique to Retrieve a Rotary Nickel-Titanium File Broken Beyond the Apex and a Stainless Steel File from the Root Canal of a Mandibular Molar: A Case Report Vineet Agrawal 1, Sonali

More information

Evolution of Nickel titanium Alloys in Endodontics

Evolution of Nickel titanium Alloys in Endodontics REVIEW ARTICLE 10.5005/jp-journals-10024-2181 1 Hani F Ounsi, 2 Wadih Nassif, 3 Simone Grandini, 4 Ziad Salameh, 5 Prasanna Neelakantan, 6 Sukumaran Anil ABSTRACT To improve clinical use of nickel titanium

More information

Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments

Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments doi:10.1111/j.1365-2591.2011.01998.x Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments E. S. J. Pereira 1, I. F. C. Peixoto

More information

A comparative scanning electron microscopic study of the effect of three different rotary instruments on smear layer formation

A comparative scanning electron microscopic study of the effect of three different rotary instruments on smear layer formation 55 Journal of Oral Science, Vol. 51, No. 1, 55-60, 2009 Original A comparative scanning electron microscopic study of the effect of three different rotary instruments on smear layer formation Shahriar

More information

Evaluation of canal preparation with rotary and hand Niti files in curved root canals using Cone Beam Computed Tomography: An In Vitro Study

Evaluation of canal preparation with rotary and hand Niti files in curved root canals using Cone Beam Computed Tomography: An In Vitro Study Original Research Article Evaluation of canal preparation with rotary and hand Niti files in curved root canals using Cone Beam Computed Tomography: An In Vitro Study Neerja Sharma 1*, Shafia Sarfi 2,

More information

IN VITRO EVALUATION OF ROOT CANAL PREPARATION WITH TWO ROTARY INSTRUMENT SYSTEMS PRO TAPER AND HERO SHAPER

IN VITRO EVALUATION OF ROOT CANAL PREPARATION WITH TWO ROTARY INSTRUMENT SYSTEMS PRO TAPER AND HERO SHAPER DOI: 10.15386/cjmed-454 Original Research IN VITRO EVALUATION OF ROOT CANAL PREPARATION WITH TWO ROTARY INSTRUMENT SYSTEMS PRO TAPER AND HERO SHAPER CAROLA PENTELESCU, LOREDANA COLCERIU, OVIDIU PASTRAV,

More information

Working length changes following straight-line access and different coronal flaring methods

Working length changes following straight-line access and different coronal flaring methods ORIGINAL ARTICLE Working length changes following straight-line access and different coronal flaring methods Shiva Sadeghi 1 * DDS, MS and Zeinab Doago 2, DDS 1. Assistant Professor of Endodontics, Dental

More information

in simulated curved canals

in simulated curved canals Journal section: Clinical and Experimental Dentistry Publication Types: Research doi:10.4317/medoral.16.e454 http://dx.doi.org/doi:10.4317/medoral.16.e454 Shaping ability of in simulated curved canals

More information

Single File Reciprocating Technique Using Conventional Nickel Titanium Rotary Endodontic Files

Single File Reciprocating Technique Using Conventional Nickel Titanium Rotary Endodontic Files SCANNING VOL. 35, 349 354 (2013) Wiley Periodicals, Inc. Single File Reciprocating Technique Using Conventional Nickel Titanium Rotary Endodontic Files SO YOUN JIN, 1 WOOCHEOL LEE, 2 MO K. KANG, 3 BOCK

More information

Comparison of the Different Techniques to Remove Fractured Endodontic Instruments from Root Canal Systems

Comparison of the Different Techniques to Remove Fractured Endodontic Instruments from Root Canal Systems Comparison of the Different Techniques to Remove Fractured Endodontic Instruments from Root Canal Systems Nimet Gencoglu a Dilek Helvacioglu b Abstract Objectives: To evaluate the success of certain methods

More information

Torsional Testing of Race, K3 and Profile Nickel Titanium Endodontic Files

Torsional Testing of Race, K3 and Profile Nickel Titanium Endodontic Files Virginia Commonwealth University VCU Scholars Compass Theses and Dissertations Graduate School 2004 Torsional Testing of Race, K3 and Profile Nickel Titanium Endodontic Files Sean D. Fessenden Virginia

More information

Original Study. Annals of Dental Specialty Vol. 6; Issue 1. Jan Mar

Original Study. Annals of Dental Specialty Vol. 6; Issue 1. Jan Mar Original Study A COMPARISON OF TRANSPORTATION OF THREE NICKEL-TITANIUM ENDODONTICS ROTARY SYSTEMS IN SEVERELY CURVED ROOT CANALS Donyavi Z, 1 Shahriari S, 2 Esmaeilzadeh M, 3 Ghanei S 4 1. Assistant Professor,

More information

Torque During Canal Instrumentation Using Rotary Nickel-Titanium Files

Torque During Canal Instrumentation Using Rotary Nickel-Titanium Files JOURNAL OF ENCODONTCS Copyright 2 by The American Association of Endodontists Printed in U.S.A. VOL. 26, No. 3, MARCH 2 Torque During Canal nstrumentation Using Rotary Nickel-Titanium Files Boonrat Sattapan,

More information

A comparison of cyclic fatigue between used and new Mtwo Ni Ti rotary instruments

A comparison of cyclic fatigue between used and new Mtwo Ni Ti rotary instruments doi: 10.1111/j.1365-2591.2006.01142.x A comparison of cyclic fatigue between used and new Mtwo Ni Ti rotary instruments G. Plotino 1, N. M. Grande 1, E. Sorci 2, V. A. Malagnino 2 & F. Somma 1 1 Department

More information

Safety and Simplicity!

Safety and Simplicity! Safety and Simplicity! A safe and precise system for your daily practice A safe and precise system for your daily practice CMA system consists of 4 nickel-titanium rotary endodontic instruments for root

More information

A Review of the Various Surface Treatments of NiTi Instruments

A Review of the Various Surface Treatments of NiTi Instruments A Review of the Various Surface Treatments of NiTi Instruments Zahed Mohammadi a, Mohammad Karim Soltani b, Sousan Shalavi c, Saeed Asgary a* a Iranian Center for Endodontic Research, Research Institute

More information

by Drs. Clifford J. Ruddle, Pierre Machtou and John D. West

by Drs. Clifford J. Ruddle, Pierre Machtou and John D. West DENTISTRY TODAY April 2013 THE SHAPING MOVEMENT 5TH GENERATION TECHNOLOGY by Drs. Clifford J. Ruddle, Pierre Machtou and John D. West Since the beginning of modern day endodontics, there have been numerous

More information

THE WORLD'S FIRST CORDLESS ULTRASONIC ENDO ACTIVATION DEVICE. Patent pending USA Engineered and Manufactured SCIENTIFIC EVIDENCE SUPPORTS SEE WHY

THE WORLD'S FIRST CORDLESS ULTRASONIC ENDO ACTIVATION DEVICE. Patent pending USA Engineered and Manufactured SCIENTIFIC EVIDENCE SUPPORTS SEE WHY Patent pending USA Engineered and Manufactured SEE WHY SCIENTIFIC EVIDENCE SUPPORTS ULTRASONIC ACTIVATION THE WORLD'S FIRST CORDLESS ULTRASONIC ENDO ACTIVATION DEVICE The EndoUltra produces ultrasonic

More information

THE WORLD'S FIRST CORDLESS ULTRASONIC ENDO ACTIVATION DEVICE. Patent pending USA Engineered and Manufactured

THE WORLD'S FIRST CORDLESS ULTRASONIC ENDO ACTIVATION DEVICE. Patent pending USA Engineered and Manufactured Patent pending USA Engineered and Manufactured SEE WHY SCIENTIFIC EVIDENCE SUPPORTS ULTRASONIC ACTIVATION THE WORLD'S FIRST CORDLESS ULTRASONIC ENDO ACTIVATION DEVICE The EndoUltra produces ultrasonic

More information

canal shaping capabilities of the ProTaper Universal and ProTaper Gold rotary instrument systems

canal shaping capabilities of the ProTaper Universal and ProTaper Gold rotary instrument systems Original Article http://dx.doi.org/10.1590/1678-77572016-0230 canal shaping capabilities of the ProTaper Universal and ProTaper Gold rotary instrument systems Abstract Jussaro Alves DUQUE 1 Rodrigo Ricci

More information

ORIGINAL PAPERS. Porównanie skuteczności opracowania sztucznych kanałów w kształcie litery L dwoma systemami maszynowymi: Hyflex CM i ProTaper Next

ORIGINAL PAPERS. Porównanie skuteczności opracowania sztucznych kanałów w kształcie litery L dwoma systemami maszynowymi: Hyflex CM i ProTaper Next ORIGINAL PAPERS Dent. d. Probl. 2015, 52, 1, 54 61 ISSN 1644-387X Copyright by Wroclaw dical University and Polish Dental Society Michał Łęski 1, A, B, Mateusz Radwański 2, B D 1, E, F, Halina Pawlicka

More information

H. Caballero-Flores, C. K. Nabeshima, E. Binotto & M. E. L. Machado. Introduction

H. Caballero-Flores, C. K. Nabeshima, E. Binotto & M. E. L. Machado. Introduction doi:10.1111/iej.12982 Fracture incidence of instruments from a single-file reciprocating system by students in an endodontic graduate programme: a cross-sectional retrospective study H. Caballero-Flores,

More information

CBCT Evaluation Of Glidepath Preparation By Different Rotary Nickel Titanium Glide Path...

CBCT Evaluation Of Glidepath Preparation By Different Rotary Nickel Titanium Glide Path... IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-853, p-issn: 2279-861.Volume 15, Issue 3 Ver. VIII (Mar. 216), PP 1-6 www.iosrjournals.org CBCT Evaluation Of Glidepath Preparation

More information

A Shift Up In Performance. Performance Refined

A Shift Up In Performance. Performance Refined A Shift Up In Performance. PROTAPER NEXT. Performance. You can count on the same ProTaper efficiency and variable taper performance you know and trust from ProTaper Universal. And now, that performance

More information

A NEW CONCEPT OF MANAGEMENT OF INTRACANAL SEPARATED INSTRUMENT BY ELECTROCHEMICAL DISSOLUTION: A REVIEW

A NEW CONCEPT OF MANAGEMENT OF INTRACANAL SEPARATED INSTRUMENT BY ELECTROCHEMICAL DISSOLUTION: A REVIEW Review Article International Journal of Dental and Health Sciences Volume 02,Issue 06 A NEW CONCEPT OF MANAGEMENT OF INTRACANAL SEPARATED INSTRUMENT BY ELECTROCHEMICAL DISSOLUTION: A REVIEW Akash Agrawal

More information

INSTRUCTIONS FOR USE AND CARE

INSTRUCTIONS FOR USE AND CARE DIAMONDS INSTRUCTIONS FOR USE AND CARE DIAMONDS, MULTI-USE, FRICTION GRIP Quala Diamonds are a rotary cutting device made of stainless steel and coated with diamond particles on the working end. It is

More information

MDJ The effects of canal preparation by different NiTi rotary Vol.:9 No.:2 2012

MDJ The effects of canal preparation by different NiTi rotary Vol.:9 No.:2 2012 The effects of canal preparation by different NiTi rotary instruments and reciprocating WaveOne file on the incidence of dentinal defects Dr. Iman M. Al-Zaka, B.D.S., M.Sc Abstract MDJ The purpose of this

More information

Dynamic and Cyclic Fatigue of Engine-Driven Rotary Nickel-Titanium Endodontic Instruments

Dynamic and Cyclic Fatigue of Engine-Driven Rotary Nickel-Titanium Endodontic Instruments JOURNAL OF ENDODONTICS rinted in U.S.A. Copyright 1999 by The American Association of Endodontists VOL. 25, NO. 6, JUNE 1999 Dynamic and Cyclic Fatigue of Engine-Driven Rotary Nickel-Titanium Endodontic

More information

Shaping Potential of Manual NiTi K-File and Rotary ProTaper and Analyzing the Final Outcome of Shaped Canals using CT

Shaping Potential of Manual NiTi K-File and Rotary ProTaper and Analyzing the Final Outcome of Shaped Canals using CT 10.5005/jp-journals-10024-1343 ORIGINAL RESEARCH Shaping Potential of Manual NiTi K-File and Rotary ProTaper and Analyzing the Final Outcome of Shaped Canals using CT Surbhi Kakar, Anil Dhingra, Himanshu

More information

Flute and Shank Dimensions of Reciprocating Instruments before and after Simulated Root Canal shaping

Flute and Shank Dimensions of Reciprocating Instruments before and after Simulated Root Canal shaping Felipe C Sampaio et al ORIGINAL RESEARCH 10.5005/jp-journals-10024-2016 Flute and Shank Dimensions of Reciprocating Instruments before and after Simulated Root Canal shaping 1 Felipe C Sampaio, 2 Ana PP

More information

Integra Miltex. Endodontic Files. Limit uncertainty with the brand you trust

Integra Miltex. Endodontic Files. Limit uncertainty with the brand you trust Limit uncertainty with the brand you trust Liberator Niti Rotary Files The Rotary File that Will Not Self-Thread Liberator rotary nickel-titanium endodontic files are the result of years of research and

More information

Comparative Evaluation of Taper Preparation Variability of Three Different Niti Files- An In Vitro CBCT Study

Comparative Evaluation of Taper Preparation Variability of Three Different Niti Files- An In Vitro CBCT Study Comparative Evaluation of Taper Preparation Variability of Three Different Niti Files- An In Vitro CBCT Study Dr. Tirthankar Bhaumik, BDS (UNB, W.B.) Post Graduate Student West Bengal University of Health

More information

1) INDICATIONS FOR USE These products have to be used only in hospital environments, clinics or dental offices by qualified dental personnel.

1) INDICATIONS FOR USE These products have to be used only in hospital environments, clinics or dental offices by qualified dental personnel. ProFile 04/06 & O.S. EN DENTAL USE ONLY DIRECTIONS FOR USE PROFILE 0) COMPOSITION The cutting part of these instruments is made of a nickel-titanium alloy. 1) INDICATIONS FOR USE These products have to

More information

Shaping ability of ProTaper Universal, WaveOne and ProTaper Next in simulated L-shaped and S-shaped root canals

Shaping ability of ProTaper Universal, WaveOne and ProTaper Next in simulated L-shaped and S-shaped root canals Wu et al. BMC Oral Health (2015) 15:27 DOI 10.1186/s12903-015-0012-z RESEARCH ARTICLE Open Access Shaping ability of ProTaper Universal, WaveOne and ProTaper Next in simulated L-shaped and S-shaped root

More information

EndoWave. Enhanced safety for root canal preparation. Thinking ahead. Focused on life.

EndoWave. Enhanced safety for root canal preparation. Thinking ahead. Focused on life. EndoWave Enhanced safety for root canal preparation Thinking ahead. Focused on life. Features that make a difference The triangular crosssection of the file minimizes the risk of fracture 3-point contact

More information