Receptor dose and Patient Dose in Radiographic exposures - a 15 year review
|
|
- Stuart Chapman
- 5 years ago
- Views:
Transcription
1 Receptor dose and Patient Dose in Radiographic exposures - a 15 year review D J Peet*, N Tyler, M Pryor, P Hollaway, C Strudley and L Leavesley Regional Radiation Protection Service, Royal Surrey County Hospital, Guildford, UK Abstract A patient dose programme has been established locally for the last 15 years across 109 hospitals and 250 X-ray rooms in line with the National Protocol [4], in conjunction with a programme to look at routine performance of these rooms. Routine performance checks initially looked primarily at film density and AEC performance but with the introduction of Computerised Radiography (CR) across UK hospitals and a revision of recommended procedures in the UK [3], the emphasis has shifted to assessing receptor dose under AEC control. Results show a wide variation in film density in the early years indicating sub optimal performance and dose. The spread was reduced over later years. The introduction of CR has led to a variety of approaches by the CR companies, X-ray companies and local sites. Receptor doses vary widely as a result. Large variations within hospitals were also observed. The doses over the last 15 years are reviewed and compared against diagnostic reference levels and with the performance of the imaging chain. Results show that patient dose programmes and optimisation strategies were having an impact, but the introduction of CR requires renewed efforts to ensure images and doses are optimised. KEYWORDS: Radiography; Patient dose; QC; optimisation; CR Introduction In the UK routine performance measurements and patient dosimetry programmes are common place for diagnostic radiology facilities. Routine performance measurements are required to be carried out under UK regulations requiring QC programmes to be in place (Ionising Radiations Regulation 1999) [1] Patient dose assessments are required under the Ionising Radiation (Medical Exposure) Regulations 2000/2006 [2]. The methodology for these is well documented with Recommended Standards for the Routine Performance Testing of Diagnostic X-ray Imaging Systems [3] and a National Protocol for Patient Dose Measurements in Diagnostic Radiology [4]. The UK has been active in the setting of national and local diagnostic reference levels with published guidance as to the methodology and approach for this [5]. The DH have adopted new National reference levels [6] in line with the latest published figures and National reviews of dose [7]. The National Dose database is kept under constant review with third quartile doses for common examinations last published in 2007 [8]. A patient dose programme has been established locally for the last 15 years across 109 hospitals and 250 X-ray rooms in line with the National Protocol, in conjunction with a programme to look at routine performance of these rooms. Routine performance checks initially looked primarily at film density and AEC performance, but with the introduction of CR across UK hospitals and a revision of recommended procedures in the UK, the emphasis has shifted to assessing receptor dose under AEC control. Method A database was developed in Guildford with the aim of increasing the efficiency of routine surveys, by generating a paper survey form which included previous survey data; by streamlining data input and automating repetitive calculations; and by producing a report form with a summary of the results which could be sent to each hospital. 1
2 The aim was also to store the information about the hospitals surveyed, the X-ray equipment within those hospitals and the results of annual Physics radiation protection and quality assurance surveys carried out on that equipment. The database was to be designed to be flexible and to store the survey results without defining an exact measurement set for each piece of equipment. This would enable analysis of the results and allow reviews of measurement techniques, X-ray equipment performance and patient dose and also allow flexibility in measurement and retain a degree of professional judgement over the measurements required. A separate dataset was also developed to store patient dose data using the same tombstone data from the main database to enable analysis of dose against the performance measurements in a particular location and in a particular year. Simple queries were written to collate the results of AEC performance with film density or receptor dose once a hospital had switched to CR. The doses for common examinations were also grouped into 3 year periods and collated. The dose for the AP lumbar spine was expected to correlate with the standard setting of the AEC system. The dose for the PA chest examination was also reviewed as this is an examination normally carried out under manual control and therefore exposure parameters are more likely to be adjusted in relation to resultant image quality. Results The optical density of films taken under AEC control during routine performance checks using standard lumbar spine settings are shown in Figure 1. Figure 1: The normalised number of systems tested against the optical densities measured. 60 Normalised Number of Measurements (%) Optical Density Range 2
3 The mean, standard deviation and number of optical density measurements are displayed in Table 1. Table 1: Variations in optical density measurements. Year Mean Optical Density Standard Deviation Number of Measurements Figure 2 shows the normalised number of measurements against receptor dose for the period of Figure 2: The normalised number of systems tested against receptor dose measured. 25 Normalised Number of Measurements (%) More Receptor Dose (µgy) The mean, standard deviation and number of receptor dose measurements are shown in Table 2. Standard deviation significantly increased in 2007 due to an increase in the number of installed units. Results from 2008 show that the spread of values has reduced following some optimisation. The numbers of measured systems are for the first six month period in 2008 and numbers are expected to be in the region of 1000 measurements for the 12 month period. 3
4 Table 2: Variations in receptor dose measurements. Year Mean Receptor Dose (µgy) Standard Deviation Number of Measurements Patient dose measurements over the same period are shown for AP lumbar spine and PA chest views in Figures 3 & 4 respectively. Mean values with standard error of the mean are plotted for each three year phase of measurements. This data has also been tabulated in Tables 3 & 4. Figure 3. Measured patient doses for lumber spine AP examinations between 1994 & 2008 Patient Doses for Lumbar Spine AP examinations over 15 years 12.0 Mean Entrance Surface Dose (mgy) Year Mean Dose - Film Mean Dose - CR National Reference Dose 1980s National Reference Dose 1995 National Reference Dose 2000 (NDRL) National Reference Dose
5 Figure 4. Measured patient doses for Chest PA examinations between 1994 & 2008 Patient Doses for Chest PA examinations over 15 years Mean Entrance Surface Dose (mgy) Year Mean Dose - Film National Reference Dose 1980s National Reference Dose 2005 Mean Dose - CR National Reference Dose 1995 & 2000 (NDRL) Table 3 Entrance surface dose measurements for AP lumbar spine examinations. Standard Mean dose (mgy) Year Deviation No. of records National Ref Doses (mgy) film CR film CR film CR 1980s
6 Table 4 Entrance surface dose measurements for PA Chest examinations. Standard Mean dose (mgy) Year Deviation No. of records National Ref Doses (mgy) film CR film CR film CR 1980s Figure 5 Screen manufacturers used 1995 to M Agfa Dupont Fuji Imation Kodak Figure 6 CR manufacturers systems in use Agfa Kodak Fuji Konica 6
7 Discussion Results of optical density measurements (figure 1) show that the QC programme was effective with a reduction in the standard deviation of measurements around the central target optical density of a single value in the range Following a peak in measurements in 2003 a sharp decline in the number of systems tested using optical density is noted which corresponds with the fast introduction of CR in our area. Measurements of receptor dose were initiated in 2005 following the significant implementation of CR and therefore the loss of optical density as a marker of AEC performance. A receptor dose of µGy has been used as a target and these results are showing that a considerable number of systems are achieving this. The spread is relatively consistent over the 4 years of data. Optimisation is however only underway in this area from the beginning of It is expected that in the following 2009 round of surveys there will be a significant decrease in the variation due to recommendations made in the 2008 surveys. Patient dose over the period shows a steady drop in doses which has been due to a number of optimisation factors including faster film screen combinations, new equipment and the standard use of AEC s The patient dose measurements for the AP lumbar spine view which is almost exclusively carried out under AEC control show that since the introduction of CR, mean doses remain close to those of film. It is hypothesised that this may increase by a small margin over the coming year as the systems are optimised for CR. Chest images predominantly use an out-of-bucky, medium kv, manual control exposure and are therefore not controlled by the AEC system. CR manufacturers data suggest that the equivalent speed for Kodak systems is closer to 300 or even 200 than the standard 400 speed film screen combination that had been in use in many hospitals. With the lower sensitivity of the CR systems image quality has been noted to adversely affected and therefore local optimisation has been carried out. This is clear in the dose data with the measured doses reported for CR systems being significantly greater than those still using film. Optimisation using film has been limited to an extent by the wide range of film types, screen types and their combinations. This has resulted in a wide variation in film screen speed and therefore corresponding dose and also a target density within a range of values. With this variation coupled with the processing variations due to such things as chemical types, temperatures and cycle times to name a few, the imaging chain variations were numerous. With the introduction of CR a number of these issues have been eliminated. CR readers and cassettes, for the same manufacturer, in general have the same response and therefore a number of variables have been removed. CR plates are coupled with a particular reader manufacturer removing the previous combinations possible and with the removal of chemistry, CR reader stability is also much less of an issue when compared to wet processing. Conclusions Optimisation has been taking place over the last 15 years with target values of optical density for AEC systems. At the same time doses for plain films have been reducing as the effect of a National dose measurement programme is felt. 7
8 The introduction of CR, if receptor doses are adjusted to the noise levels recommended by the manufacturer, is likely to result in an increase in dose over the next few years. To date AEC s have not been adjusted but are now being set at a receptor dose of 4mGy. In the UK National reference doses have not been adopted based on the latest dose review within the National dose database. In light of the changes in technology this appears to be a prudent approach as doses appear to be increasing in the short term following the introduction of CR. Further work is required to monitor this change. References [1] Health and Safety Commission Ionising Radiations Regulations 1999 (SI 1999 No 3232) (The Stationery Office, London). [2] Department of Health. The Ionising Radiation (Medical Exposure) Regulations Statutory Instruments 2000 No (The Stationery Office, London). [3] Institute of Physics and Engineering in Medicine (2005) Recommended Standards for the Routine Performance Characteristics of Diagnostic X-ray systems used in Medicine Report No 91 (IPEM, York) [4] Dosimetry Working Party of the Institute of Physical Sciences in Medicine (1992) National protocol for patient dose measurements in diagnostic radiology. (NRPB, Chilton) [5] Institute of Physics and Engineering in Medicine (2004) Guidance on the establishment and use of diagnostic reference levels for medical x-ray examinations Report No 88 (IPEM, York) [6] Department of Health. Guidance on the establishment and use of Diagnostic Reference Levels (DRLs) as the term is applied in the Ionising Radiation (Medical Exposure) Regulations 2000 DH DRL Working Party (2007) DH_ [7] Hart D, Hillier MC and Wall BF. Doses to Patients from Medical X-ray Examinations in the UK 2000 Review. (2002) NRPB-W14 (NRPB, Chilton). [8] Hart D, Hillier MC and Wall BF. Doses to Patients from Radiographic and Fluoroscopic X-ray imaging Procedures in the UK 2005 Review. (2007) HPA-RPD-029 (HPA, Chilton). 8
Doses to Patients arising from Dental X-ray Examinations in the UK, A Review of Dental X-ray Protection Service Data
HPA-RPD-22 Doses to Patients arising from Dental X-ray Examinations in the UK, 22 24 A Review of Dental X-ray Protection Service Data A D Gulson, T A Knapp and P G Ramsden ABSTRACT The Dental X-ray Protection
More informationA Standardised Approach to Optimisation
The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained
More informationPatient dose in routine X-ray examinations in Yazd state
Iran. J. Radiat. Res., 2004; 1(4): 199-204 Patient dose in routine X-ray examinations in Yazd state F. Bouzarjomehri Health Physics Dept. Shahid Sadoghi Univ. of Medical Sciences, Yazd, Iran ABSTRACT Background:
More informationShielding Calculation: Radiographic Room. Jerry Williams
Shielding Calculation: Radiographic Room Jerry Williams What you need to know Room use and layout DAP workload DAP averaged kv Distance to barrier Construction details Walls Ceilings/ floors Surrounding
More informationBREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0
BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2
More informationEstablishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology
Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics
More informationTHE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES
IJRRAS 9 (1) October 11 www.arpapress.com/volumes/vol9issue1/ijrras_9_1_16.pdf THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES Bamidele
More informationTHE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS
THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS Larysa Stadnyk, Pylypenko M., Shalepa O., Nosyk O. Grigorev Institute for Medical Radiology,
More informationLocal reference dose evaluation in conventional radiography examinations in Iran
JOURNAL OF PLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 Local reference dose evaluation in conventional radiography examinations in Iran M. Shirin Shandiz, 1,2 M.T. Bahreyni Toossi, 3a S.
More informationAlfred Health's CT quality control program
Alfred Health's CT quality control program Poster No.: R-0037 Congress: Type: Authors: 2014 CSM Scientific Exhibit A. Perdomo, Z. Brady, N. Tran, L. Hudson, K. Provis; PRAHRAN/ AU Keywords: Quality assurance,
More informationRadiation Protection Dosimetry Advance Access published September 13, 2006
Radiation Protection Dosimetry Advance Access published September 13, 26 Radiation Protection Dosimetry (26), 1 of 6 doi:1.193/rpd/ncl113 TECHNICAL NOTE RADIATION EXPOSURE AND DOSE EVALUATION IN INTRAORAL
More informationPATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS
Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,
More informationPopulation Dose from General X-ray and Nuclear Medicine: 2010
Population Dose from General X-ray and Nuclear Medicine: 2010 Medical Exposure Radiation Unit 1 Executive Summary of Results The Health Services Executive (HSE) has a legal obligation, under Article 12
More informationIRMER Ionising Radiation Medical Exposure Regulations
IRMER Ionising Radiation Medical Exposure Regulations Download for information at www.scottishdental.org IONISING RADIATION (MEDICAL EXPOSURE ) REGULATION 2000 (as Amended) (IR(ME)R AN EXPLANATION GUIDE
More informationThe New Jersey Radiographic Quality Assurance Program at 5 Years
The New Jersey Radiographic Quality Assurance Program at 5 Years Julie Timins, MD a, Paul Orlando, BS b, Jill Lipoti, PhD b Purpose: Five years ago, the New Jersey Bureau of Radiological Health decided
More informationP T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3
Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,
More informationDosimetry in digital mammography
Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol
More informationPatient dose measurement in common medical X-ray examinations in Iran
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Patient dose measurement in common medical X-ray examinations in Iran Behrouz Rasuli, 1a Ali Mahmoud-Pashazadeh, 2 Mohammad Ghorbani,
More informationEstablishment of local diagnostic reference levels for quality control in intraoral radiography
Oral Radiol (27) 33:38 44 DOI.7/s282-6-24-9 ORIGINAL ARTICLE Establishment of local diagnostic reference levels for quality control in intraoral radiography Maki Izawa Yasuo Harata Noriyoshi Shiba Nobuhide
More informationDiagnostic Reference Levels (DRLs)
Section 7 Diagnostic Reference Levels (DRLs) Establishment and Use of Diagnostic Reference Levels (DRLs) The National Radiation Safety Committee (NRSC), established in November 2007, is a statutory committee
More informationA STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY
Journal of Bangladesh Academy of Sciences, Vol. 3, No. 1, 7-17, 211 A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY MAHFUZA BEGUM 1, A. S. MOLLAH
More informationReport. Surveying dose levels for computed radiography in the UK. KCARE King s College Hospital Denmark Hill London SE5 9RS
Report Surveying dose levels for computed radiography in the UK KCARE King s College Hospital Denmark Hill London SE5 9RS Tel No: +44 (0)20 3299 1610 Fax No: +44 (0)20 3299 3314 www.kcare.co.uk Certificate
More informationEVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2
Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 7 doi:10.1093/rpd/ncn143 EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY
More informationResearch Article. Corresponding author H. Osman
Scholars Journal of Applied Medical Sciences (SJAMS) ISSN 2320-6691 Sch. J. App. Med. Sci., 2013; 1(5):511-515 Scholars Academic and Scientific Publisher (An International Publisher for Academic and Scientific
More informationSurvey of Mammography Practice: Initial Results from Serbia
Survey of Mammography Practice: Initial Results from Serbia Olivera Ciraj Bjelac *, Dusko Kosutic, Danijela Arandjic, Milojko Kovacevic Vinca Institute of Nuclear Sciences, Radiation and Environmental
More informationAppendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit
References References The accuracy, quality and relevance of these works are not guaranteed or uniform and more recent information may have superseded these works. This list is not exhaustive. It does
More informationSUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT
SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT The use of dental CBCT equipment must comply with all the regulations (IRR99 and IR(ME)R2000)
More informationBRITISH SOCIETY OF DENTAL AND MAXILLOFACIAL RADIOLOGY CORE CURRICULA IN DENTAL RADIOGRAPHY AND RADIOLOGY FOR THE DENTAL TEAM
BRITISH SOCIETY OF DENTAL AND MAXILLOFACIAL RADIOLOGY CORE CURRICULA IN DENTAL RADIOGRAPHY AND RADIOLOGY FOR THE DENTAL TEAM 2015 0 INTRODUCTION The original core curriculum in Dental Radiography and Radiology
More informationIntegrated Research Application System (IRAS)
Integrated Research Application System (IRAS) Question specific guidance Part B Section 3 Exposure to ionising radiation The following document collates all guidance for the questions in Part B: Section
More informationRadiation Protection Dosimetry Advance Access published April 22, 2008
Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 5 doi:10.1093/rpd/ncn144 EVALUATION OF EXPOSURE PARAMETERS IN PLAIN RADIOGRAPHY: A COMPARATIVE
More informationDETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA
DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,
More informationStep by Step Guidance on how to Conduct a Clinical Audit in. Dental Radiography Core Subject
Step by Step Guidance on how to Conduct a Clinical Audit in Dental Radiography Core Subject Aims: To describe the purpose of a clinical audit and its importance as part of a good practice quality assurance
More informationABSTRACT I. INTRODUCTION
2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography
More informationIAEA-TECDOC-796. Radiation doses in diagnostic radiology
IAEA-TECDOC-796 Radiation doses in diagnostic radiology The IAEA does The originating Section FOREWORD It is well recognized that diagnostic radiology is the largest contributor to the collective dose
More informationCT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell
CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction
More informationThe checklist for the use of Ionising Radiation in Primary Dental Care. Not fully compliant
The checklist for the use of Ionising Radiation in Primary Dental Care Health and Safety Executive (HSE) 1 The practice has registered with the HSE that radiographic equipment is being used on the premise
More informationCLINICAL AUDIT OF DENTAL RADIOGRAPHY
CLINICAL AUDIT OF DENTAL RADIOGRAPHY DENTAL RADIOGRAPHIC AUDIT. (50 Pairs of Bitewing radiographs, 25 Periapical radiographs and 10 OPG s.) INTRODUCTION A clinical audit was carried out at the practice
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationOPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY
March 2010 EAST AFRICAN MEDICAL JOURNAL 127 East African Medical Journal Vol. 87 No. 3 March 2010 OPTIMIsATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY G. K. Korir, BSc, MSc, PhD
More informationSurvey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations
Caspian Journal of Health Research 2016; 2(2):18-27 Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Hossein Rahanjam 1, Hossein
More informationIONISING RADIATION REGULATIONS 99
IONISING RADIATION REGULATIONS 99 & IRMER IONISING RADIATION MEDICAL EXPOSURE REGULATIONS BARBARA LAMB Specialist Radiographer Dental and maxillofacial radiography BarbaraHLamb@googlemail.com 07775994424
More informationDIAGNOSTIC RADIOGRAPHY ANSWERS For Medical Students. Clare Hawkins Superintendent Radiographer
DIAGNOSTIC RADIOGRAPHY ANSWERS For Medical Students Clare Hawkins Superintendent Radiographer 1. Which fish bones are visible on x- rays and why? Cole fish, gurnard, monk fish, plaice, red snapper, cod,
More informationTrends in Patients Exposure Doses during Radiographic Examination in Japan
Trends in Patients Exposure Doses during Radiographic Examination in Japan S. Suzuki 1*, Y. Asada 1, M. Kobayashi 2, K. Kobayashi 2, Y. Matsunaga 3, T. Otsuka 4 1 School of Health Sciences, Fujita Health
More informationDoses Received by Patients during Thorax X-Ray Examinations
Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 01, 45-51 Received: September 16, 01; Accepted: February 0, 013 Original Article Doses Received by Patients during Thorax X-Ray Examinations Nsikan
More informationEvaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography
Available online at http://www.biij.org/2006/3/e35 doi: 10.2349/biij.2.3.e35 biij Biomedical Imaging and Intervention Journal ORIGINAL PAPER Evaluation of radiation dose and image quality following changes
More informationEstablishing the Greek National Reference Levels for Interventional Cardiology procedures
Establishing the Greek National Reference Levels for Interventional Cardiology procedures George Simantirakis*, Christina Koukorava, Maria Kalathaki, Christos Pa
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training
More informationNational collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations
National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations Eva Godske Friberg *, Anders Widmark and Ingrid Helen Ryste Hauge Norwegian Radiation
More informationBICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:
BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:
More informationTesting of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution
Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi
More informationDoses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean
Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could
More informationCT Quality Control Manual FAQs
CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues
More informationInternational Journal of Radiation Research, January 2017
Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing
More informationInformation and Guidance
Health and Safety Executive (HSE) 1 The practice has registered with the HSE that radiographic equipment is being used on the premise and confirmed compliance with IRR17 by 5/2/2018. 2 HSE have confirmed
More informationLimited information related to: - Type and frequency of procedures; - Exposure levels involved.
Limited information related to: Outside Radiological Department - Type and frequency of procedures; - Exposure levels involved. Majority of the examinations performed outside the department. 2 Patients
More informationDiagnostic reference levels as a quality assurance tool
Australian Institute of Radiography The Radiographer 2009; 56 (3): 32 37 Literature review Diagnostic reference levels as a quality assurance tool KD Edmonds Medical Physics Section, Medical Radiation
More informationMeasurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran
Global Journal of Health Science; Vol. 7, No. 5; 2015 ISSN 1916-9736 E-ISSN 1916-9744 Published by Canadian Center of Science and Education Measurement of Entrance Skin Dose and Calculation of Effective
More informationEvaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray
Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray Poster No.: C-1492 Congress: ECR 2012 Type: Scientific Exhibit Authors: R. M. Lopes, L. H. Domingos ; Mortágua/PT, Azoia de
More informationImpact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals
B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy
More informationCumulative radiation CT dose in young traumatic patients: a single centre 5 years retrospective review
Cumulative radiation CT dose in young traumatic patients: a single centre 5 years retrospective review Poster No.: C-1604 Congress: ECR 2014 Type: Scientific Exhibit Authors: S. Salerno, C. Geraci, E.
More informationDiagnostic reference levels in intraoral dental radiography in Korea
Imaging Science in Dentistry 2012; 42 : 237-42 http://dx.doi.org/10.5624/isd.2012.42.4.237 Diagnostic reference levels in intraoral dental radiography in Korea Eun-Kyung Kim, Won-Jeong Han, Jin-Woo Choi,
More informationPatient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran
Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran Soraya Emamgholizadeh Minaei 1, Fariba Firouzi 1 2, 3,*, Hamid Reza Khosravi 1 Department of radiology, Shahid
More informationJMSCR Vol 06 Issue 12 Page December 2018
www.jmscr.igmpublication.org Impact Factor (SJIF): 6.379 Index Copernicus Value: 79.54 ISSN (e)-2347-176x ISSN (p) 2455-0450 DOI: https://dx.doi.org/10.18535/jmscr/v6i12.80 Diagnostic Reference Levels:
More informationRadiological Technology (Radiography): Emergence of a Profession
Radiological Technology (Radiography): Emergence of a Profession (Dr) Bev Snaith Consultant Radiographer, Mid Yorkshire Hospitals Honorary Research Fellow, University of Bradford UK Introduction Gratitude
More informationICRP Symposium on the International System of Radiological Protection
ICRP Symposium on the International System of Radiological Protection October 24-26, 2011 Bethesda, MD, USA M. Balonov, ICRP Committee 2, IRH, Russia P. Shrimpton, HPA, UK To be used just for regulatory
More informationPatient dose assessment of CT perfusion scanning at the RSCH
Patient dose assessment of CT perfusion scanning at the RSCH Lesley Leavesley, Emma Whitehead, Matthew Pryor, Debbie Peet Regional Radiation Protection Service Royal Surrey County Hospital, Guildford Overview
More informationNational COPD Audit Programme
National COPD Audit Programme Planning for every breath National Chronic Obstructive Pulmonary Disease (COPD) Audit Programme: Primary care audit (Wales) 2015 17 Data analysis and methodology Section 2:
More informationProf. Dr. Doğan BOR Ankara University Institute of Nuclear Science
PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural
More informationEntrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals
Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and
More informationIranian Journal of Medical Physics
Iranian Journal of Medical Physics ijmp.mums.ac.ir Dose Evaluation for Common Digital Radiographic s in Selected Hospitals in Pahang Malaysia Soo-Foon Moey 1*, Zubir Ahmad Shazli 1, Inayatullah Shah Sayed
More informationEssential CPD for the dental team
Essential CPD for the dental team Saturday 19 April 2008 and Saturday 17 May 2008 A two day meeting organised by the Royal Society of Medicine Venue: 1 Wimpole Street, London, W1G 0AE DAY 1 Infection Control
More informationBICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives
More informationPatient information leaflet. Royal Surrey County Hospital. NHS Foundation Trust. Having a Mammogram. Radiology Department
Patient information leaflet Royal Surrey County Hospital NHS Foundation Trust Having a Mammogram Radiology Department This leaflet provides you with information about our mammography service (breast x-ray).
More informationA. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.
Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1
More informationPilot study of patient and phantom breast dose measurements in Bulgaria
Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast
More informationEuropean Guidelines on DRLs for Paediatric Imaging
European Guidelines on DRLs for Paediatric Imaging Final complete draft 1 1 1 1 1 1 1 1 0 1 0 1 Contents PREFACE... EXECUTIVE SUMMARY... 1. Background.... Introduction.... Purpose and scope.... Definitions...
More informationDoses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?
Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.
More informationImpact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program
Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program An De Hauwere, Hubert Thierens Ghent University, Department of Medical Physics,
More informationPNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY
PNRA-RG-904.05 (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY For Further Details Directorate of Policies & Procedures PAKISTAN
More informationAssessing the standard dose to standard patients for x-ray investigations
Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The
More informationBICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives
More informationAudit Report. National Audit of Paediatric Radiology Services in Hospitals
Audit Report National Audit of Paediatric Radiology Services in Hospitals www.rcr.ac.uk 2 Contents Introduction 3 Standards 4 Material and methods 5 Results 6 Discussion 8 References 10 www.rcr.ac.uk 3
More informationThe Royal College of Radiologists, August 2000 This Publication is Copyright under the Berne Convention and the International Copyright Convention.
A Guide to Justification for Clinical Radiologists Ref No: BFCR(00)5 The Royal College of Radiologists 38 Portland Place London W1N 4JQ Telephone 020 7636 4432 Fax 020 7323 3100 Email: enquiries@rcr.ac.uk
More informationAsk EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT
Ask EuroSafe Imaging Tips & Tricks CT Working Group On the use of Diagnostic Reference Levels in CT Robert Bujila (Karolinska University Hospital, SE) Eileen Kelly (Galway University Hospitals, IE) Matthias
More informationPatient information leaflet. Royal Surrey County Hospital. NHS Foundation Trust. Having a Mammogram. Radiology Department
Patient information leaflet Royal Surrey County Hospital NHS Foundation Trust Having a Mammogram Radiology Department This leaflet provides you with information about our mammography service (breast x-ray).
More informationPatient effective dose evaluation for chest X-ray examination in three digital radiography centers
Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical
More informationPatient doses in general radiography examinations
peer reviewed ORIGINAL ARTICLE Patient doses in general radiography examinations Wambani JS 1 MBChB, M. Med (Rad), Paediatric (Rad) Onditi EG 2 MBChB, M.Med (Rad) Korir GK 3 B.Sc, MSc (Nuclear Science),
More informationSurvey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria Bidemi I.
More information4:ll.J1 ~I 41 ~.J,II,,1'1 'IJ".:J ~.J,II...:.CJI WORLD HEALTH ORGANIZATION ORGANISA'TION MONDIALE DE LA SANT~
WORLD HEALTH ORGANIZATION REGIONAL OFFICE FOR THE EASTERN MEDITERRANEAN 4:ll.J1 ~I 41 ~.J,II,,1'1 'IJ".:J ~.J,II...:.CJI.. ORGANISA'TION MONDIALE DE LA SANT~ BUREAU REGIONAL DE LA MEDITERRANEE ORIENTALE
More informationEstimation of Radiation Absorbed Dose to Patients Undergoing Chest X-ray Examination in Four Government Own Hospitals in Nigeria
International Journal of Biophysics 2017, 7(2): 24-32 DOI: 10.5923/j.biophysics.20170702.02 Estimation of Radiation Absorbed Dose to Patients Undergoing Chest X-ray Examination in Four Government Own Hospitals
More informationRadiation exposure of the Yazd population from medical conventional X-ray examinations
Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical
More informationThe Radiology Aspects
REQUIREMENTS FOR INTERNATIONAL ACCREDITATION OF BREAST CENTERS/UNITS The Radiology Aspects Miri Sklair-Levy, Israel RADIOLOGY GUIDELINES FOR QUALITY ASSURANCE IN BREAST CANCER SCREENING AND DIAGNOSIS Radiologists
More informationPatti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety
Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February 2008. Radiation Safety Sub -headings Background Radiation Effects of Radiation Safe Levels Effective Doses ALARA Principle Radiation
More informationIntroduction 1. Executive Summary 5
Roman_pages 20-09-2005 21:01 Pagina IX Table of contents Introduction 1 Executive Summary 5 1. Epidemiological guidelines for quality assurance in breast cancer screening 15 1.10 Introduction 17 1.20 Local
More informationMA 2030 Radiography Skills for Medical Assistants
South Central College MA 2030 Radiography Skills for Medical Assistants Common Course Outline Course Information Description This course takes a comprehensive look at the skills and processes needed to
More informationRadiation Doses to Patients Undergoing Selected Radiographic Examinations in Three Hospitals South-West, Nigeria
International Journal of Chemical and Biomedical Science 2017; 3(3): 26-31 http://www.aascit.org/journal/ijcbs ISSN: 2472-9574 (Print); ISSN: 2472-9590 (Online) Radiation Doses to Patients Undergoing Selected
More informationStudent Perception of Optimisation compared with the Evidence of Optimisation in Paediatric Radiography Case Studies
Student Perception of Optimisation compared with the Evidence of Optimisation in Paediatric Radiography Case Studies Jennifer Doyle, Allison McGee, Kate Matthews Diagnostic Imaging School of Medicine and
More informationA Study of dose Distribution and Radiation Protection in Industrial Radiography in Ireland
A Study of dose Distribution and Radiation Protection in Industrial Radiography in Ireland L. Currivan, J. T. Duffy, D. Spain and D. Pollard Radiological Protection Institute of Ireland, 3 Clonskeagh Square,
More informationEthics in Medical Radiological Protection
Ethics in Medical Radiological Protection Dr. Claire Cousins, Chair ICRP Consultant Vascular & Interventional Radiologist, Addenbrooke s Hospital, Cambridge, UK Ethics in Medical Radiological Protection
More informationIntroduction and Background
CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section
More information