a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

Size: px
Start display at page:

Download "a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK"

Transcription

1 Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman a, Stephen Steel a and Andrew Shennan b Background The Omron 705IT (HEM-759-E, Omron Corporation, Kyoto, Japan) is an automated oscillometric upper arm blood pressure monitor for the professional and home use markets. The aim of this study was to validate the accuracy of this device according to the British Hypertension Society and the Association for the Advancement of Medical Instrumentation SP10 validation criteria. Methods Study participants were recruited until a total of 85 were obtained that filled the blood pressure categories specified by the British Hypertension Society protocol. Recruitment to the study was from the general medical and specialist clinics and from the staff at Guy s & St Thomas Hospital in London, UK. Nine sequential same-arm blood pressure readings were taken from each participant by two trained observers, alternating between mercury reference sphygmomanometers and the Omron 705IT (HEM-759-E). The differences between the reference and test device readings, for both systolic and diastolic pressures, were compared with British Hypertension Society and Association for the Advancement of Medical Instrumentation criteria to determine the outcome of the study. Results The Omron 705IT (HEM-759-E) is graded A for systolic and A for diastolic blood pressures according to the British Hypertension Society criteria. The mean (standard deviation) of the difference between the observer and the device measurements was 0.60 (6.0) mmhg for systolic and 3.15 (6.6) mmhg for diastolic pressures, respectively. The device, therefore, also satisfies the Association for the Advancement of Medical Instrumentation SP10 standard, that requires differences of less than ± 5 (8) mmhg. Conclusions The Omron 705IT (HEM-759-E) achieved an A/A performance classification under the British Hypertension Society criteria and passes the Association for the Advancement of Medical Instrumentation requirements for the study population. It can be recommended for professional and home-use in an adult population. Blood Press Monit 11:27 32 c 2006 Lippincott Williams & Wilkins. Blood Pressure Monitoring 2006, 11:27 32 Keywords: automated devices, blood pressure monitoring, oscillometric measurement, validation studies a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Correspondence and requests for reprints to Andrew Coleman, Medical Physics Department, The Rayne Institute, Guy s & St Thomas NHS Foundation Trust, London SE1 7EH, UK Tel: + 44 (0) ; fax: + 44 (0) ; andrew.coleman@gstt.sthames.nhs.uk Sponsorship: The study was funded by Omron Healthcare Europe B.V. Kruisweg 577, 2132 NA Hoofddorp, The Netherlands. Conflicts of interest: None Received 15 April 2005 Revised 17 August 2005 Accepted 17 August 2005 Introduction The traditional clinical use of mercury sphygmomanometers is in decline with increasing numbers of automated blood pressure (BP) monitors available [1]. Although there are now estimated to be over 400 automated devices on the UK market alone, only eighteen automated devices have been validated for accuracy against one or more of the recognized standards published by the British Hypertension Society (BHS) [2], European Society of Hypertension (ESH) [3] and the Association for the Advancement of Medical Instrumentation (AAMI) [4]. Of the eighteen validated devices, only four are recommended for clinical use, on the basis of a critical assessment of the published studies, c 2006 Lippincott Williams & Wilkins as listed by the dables Educational Trust website [5,6]. This study has examined the accuracy of the Omron 705IT following the BHS protocol. The Omron 705IT (HEM-759-E) is a widely used oscillometric BP monitor that is currently marketed for both professional and home use throughout Europe. It has previously been validated using the ESH protocol [7], which involves intercomparison of same-arm BP measurements with a reference mercury sphygmomanometer in a relatively small sample of study participants (n = 33) having systolic blood pressures (SBPs) in a range up to 180 mmhg. The study described here has followed

2 28 Blood Pressure Monitoring 2006, Vol 11 No 1 the BHS protocol to validate the Omron 705IT for clinical use. In common with the AAMI protocol, the BHS specifies a larger number of participants (n = 85) and a wider SBP range, extending above 180 mmhg. These requirements make the BHS and AAMI protocols particularly difficult and time consuming to complete, but also give them greater statistical power, which is appropriate for the evaluation of a clinically used device. Methods The tested device The Omron 705IT (HEM-759-E) is an electronic oscillometric device with a capacitive pressure sensor designed to measure pressure in the range mmhg and pulse rates in the range beats per min. The specified accuracy for the pressure scale is ± 3 mmhg. The device has a memory, which allows recording of 28 measurements, and a data output port, which can download data to a PC and a printer. Systolic and diastolic BPs are displayed, along with pulse rate on a digital readout. Inflation is controlled by an automatic pumping system and deflation is controlled by an automatic pressure release valve. The device is claimed to have the same measurement algorithm as the newer Omron M6 (HEM-7001-E, Omron Corporation, Kyoto, Japan), which has some additional features. The unit is powered by either a mains adaptor or four 1.5 V alkaline batteries (type LR6 AA). It weighs 380 g (without batteries) and measures 115 mm 117 mm 71 mm (length, width and height, respectively) and is supplied with a cuff of dimensions 140 mm 480 mm (width and length, respectively). The standard cuff is applicable to arm circumferences in the range cm. An optional additional large cuff of dimensions 174 mm 617 mm is available for arm circumferences in the range cm. Study design We have adhered to the BHS validation protocol with two modifications: (1) In common with previously published BHS validations [8,9], we have restricted testing in the low SBP range such that instead of eight participants with an SBP of < 90 mmhg and 20 participants with an SBP in the range mmhg, we recruited eight participants with an SBP r 100 mmhg and 20 participants whose SBP fell in the range mmhg. (2) We have used a semi-automated static pressure calibration of the device, as opposed to the manual dynamic pressure calibration approach described in the protocol. This modification is detailed in the following section. Before-use calibration Three identical new Omron 705IT (HEM-759-E) test devices were donated by the manufacturer for the study. Written confirmation that these represented standard production models was obtained. The pressure scales of these three test devices were calibrated on receipt. A semi-automated static pressure calibration was substituted for the manual dynamic pressure described in the protocol. A calibrated Druck 520 (GE Druck, Leicester, UK) pressure actuator, interfaced to a PC, was used to apply and maintain accurately known static pressures to the test device. The pressure scale of this calibrated actuator reads out in steps of 0.01 mmhg and the uncertainty in the absolute applied pressure is ± 0.15 mmhg at the 95% confidence level relative to the UK National Standard. Thirty different static pressures, taken from tabulated random values of pressure specified in the BHS protocol, were automatically applied to the test device in sequence at 30-s intervals. The corresponding pressure readings from the digital scale of the Omron 705IT (HEM-759-E), operated in manual mode, were recorded by an observer who was blinded from the remotely applied pressure values. All 30 control and test measurement pairs were within 3 mmhg for each of the three devices as required. In-use field assessment and after-use calibration The three test devices were then placed in hospital clinics and the static calibration repeated after each had undergone at least 400 inflations in the clinic. Again, all 30 control and test measurements remained within 3 mmhg on each device. One device was arbitrarily selected for inclusion in the subsequent clinical phase of the validation. Static device validation Recruitment of study participants Ethical approval was obtained for the study. Participants were recruited to the trial if they provided signed consent. Recruitment was at random from the population of patients attending routine outpatient clinics in a large teaching hospital and from staff responding to an advertisement posted on the hospital intranet. Recruitment of participants continued until all specified BP categories were filled (n = 76). A further nine participants were recruited to make up the total of 85, without regard to any specified BP category, as allowed under the BHS protocol. Participants excluded were those with atrial fibrillation, frequent extra systoles and sufficiently weak Korotkoff sounds that acceptable auscultation was impossible. In addition, participants were excluded if an incomplete set of BP measurements were obtained. This occurred in participants who could not tolerate the repeated arm compressions or who became anxious during the course of the measurements.

3 Validation of Omron 705IT Coleman et al. 29 Observer training and assessment The study team consisted of a medically qualified expert, who provided clinical oversight, along with two clinical scientists and a technician, who recruited patients and acted as observers, undertaking the clinical intercomparison between the automated and mercury reference devices. All team members were full-time hospital staff and undertook the validation as part of their designated duties. A healthcare assistant was employed to act as an additional observer when required. All those involved as observers were trained using the CD-ROM tutorial detailed on the British Hypertension Society web site [10]. Validation procedure The participants were seated in a warm, quiet room close to the clinic, and measurements started after consent was obtained, allowing at least 5 min rest, during part of which participants were asked their age, height, and weight for the record. The arm circumference was also measured and recorded to allow the appropriate cuff size to be selected. All pressures were measured with the patient seated, with the upper arm at heart level. Sequential same-arm measurements of SBP and DBP were independently recorded by two trained observers simultaneously using two Accoson Dekamet Mk3 mercury sphygmomanometers (manufactured by A. C. Cossor & Son Ltd, London, UK) and the Omron 705IT test device. Care was taken to ensure that observers were blinded to each other s readings. This was achieved by positioning the reference devices so that each observer could view only the scale of their own device and by ensuring they completed separate forms for recording the results. Measurements were carried out and recorded in the protocol-specified sequence, with more than 30 s intervals between measurements to minimize venous congestion, but less than 1 min to minimize variability in BP: BPA mercury (entry value used to categorize participants) BPB device (to check the automated device, not included in the analysis) BP1 mercury (observers 1 and 2) BP2 device (supervisor with test device) BP3 mercury (observers 1 and 2) BP4 device (supervisor with test device) BP5 mercury (observers 1 and 2) BP6 device (supervisor with test device) BP7 mercury (observers 1 and 2) Participants were allocated into BP categories on the basis of measurement BPA, which was subsequently discarded. Measurement BPB was also discarded, and served only to confirm that the automated test device was capable of registering a reading. The measurements BP1 to BP7 were retained for the analysis of the accuracy of the test device. Data analysis The forward and backward differences between the device and the observer measurements are calculated (in mmhg) in each case. The resulting pairs of forward and backward device observer differences are as follows: BP1 BP2, BP2 BP3 BP3 BP4, BP4 BP5 BP5 BP6, BP6 BP7 The three device observer difference values most favourable to the device (i.e. the smallest value of each listed pair, ignoring the sign) are selected for the analysis, giving a total of three device observer differences for each participant, and generating 255 difference values for the complete study of 85 participants. This analysis is done separately for both SBP and DBP. The 255 device observer differences derived from this procedure are then categorized, again ignoring the sign, to determine the percentage with values r 5, r 10 and r 15 mmhg. The protocol specifies the minimum number of device observer comparison values allowed within these classes for the test device to achieve an A, B, C or D rating. The percentage in all three categories should be equal to, or exceed, the specified minimum percentage. An A rating, for example, requires at least 60% of device observer difference values to be in the r 5 mmhg error category, at least 85% in the r 10 mmhg category and at least 95% to be in the r 15 mmhg category. This analysis is done separately for each observer and for both systolic and diastolic BP. The final grade for each systolic and diastolic BP is the better of the grades obtained by the two observers. Results Study participants A total of 143 participants were recruited to fill all the protocol-specified BP categories. Participants excluded consisted of five whose Korotkoff sounds were too weak, four in whom arrhythmias were observed, three who opted out of the study before a complete measurement set was obtained, and two for whom the test device gave repeated error readings. Forty-four participants were excluded because the relevant BP category was already full. Participants were selected consecutively after testing to best fit the required categories. The mean age of the 85 participants included in the study was 47.2 ± 14.9 years (range years), of which 38 were men and 47 women. The mean arm circumference was 30.5 ± 4.0 cm, with a range of cm.

4 30 Blood Pressure Monitoring 2006, Vol 11 No 1 Table 1 Grading criteria, mean and mean differences for test device and a sample analysis for overall pressure levels for both observers Differences between standard and test device (mmhg) Grade r 5 r 10 r 15 Mean ± SD (mmhg) Mean ± SD of differences (mmhg) Observer 1 SBP A ± ± 6.0 DBP A ± ± 6.6 Observer 2 SBP A ± ± 6.3 DBP B ± ± 6.6 Final grade SBP A ± ± 6.0 DBP A ± ± 6.6 Observer comparison SBP DBP Pressure range: SBP, mmhg; DBP, mmhg. On the basis of 255 difference values per observer for SBP, DBP. SBP, systolic blood pressure; DBP, diastolic blood pressure. Observer agreement Observer agreement levels are presented in Table 1. The BHS protocol specifies that at least 80% of measurements by the observers should be within 5 mmhg of each other, and 95% within 10 mmhg. The agreement between observers is well within these limits. Observer device agreement The overall results of the validation are given for both observers in Table 1. The percentage of device observer differences falling in the specified categories is given along with the corresponding grades for SBP and DBP for each observer. The grades are based on the BHS criteria. The mean value and standard deviation of the device and observer values and the mean and standard deviation of the device observer differences are also given. The mean and standard deviation of the device observer differences are 0.6 ± 6.0 mmhg for SBP and 3.15 ± 6.6 mmhg for DBP for the better observer. The data set, therefore, satisfies the AAMI SP10: 2002 standard requirement that SBP and DBP device observer differences be less than 5 ± 8 mmhg. The BHS protocol allows the selection of the results from the best observer, in this case observer 1 for both systolic and diastolic pressures. The Omron 705IT (HEM-759- E), therefore, achieves an overall A rating (the final grade as shown in Table 1) for both systolic and diastolic pressures. Figure 1a and b shows the Bland Altman plots [9] corresponding to the better observer measurements (observer 1) for diastolic and systolic pressures, respectively. Here, the 255 values of the difference between the device and observer (in mmhg), using the better observer values, are plotted against the mean value of the device and observer readings (in mmhg). Reference lines are given at differences of 0, ± 5, ± 10 and ± 15 mmhg. These graphs show a random scatter centred about the mean device observer difference. Fig. 1 (a) Mean diastolic pressure: device and observer 1 (mmhg) Difference: device observer 1 (mmhg) (b) Difference: device observer 1 (mmhg) Mean systolic pressure: device and observer 1 (mmhg) Plots of the pressure difference between the better observer and the test device, and the mean pressure for the test device and that observer, in 85 participants for (a) diastolic pressure and (b) systolic pressure (the number of data points is 255 for each). The two vertical lines divide the pressure scale into low, medium and high regions as defined in Table 2. They also show that there are no trends in the data. These are both required features of the Bland Altman plots for the results to be generalizable. Table 2 gives a breakdown of the results by pressure range. Applying the BHS criteria within these ranges, the Omron 705IT (HEM-759-E) achieves an A/A rating in

5 Validation of Omron 705IT Coleman et al. 31 Table 2 Grading criteria, mean and mean differences for test device at high, medium and low pressure levels using the better observer Differences between standard and test device (mmhg) Grade r 5 r 10 r 15 n Low pressure range ( < 130/80 mmhg) SBP A DBP A Medium pressure range ( / mmhg) SBP A DBP A High pressure range ( > 160/100 mmhg) SBP A DBP C SBP, systolic blood pressure; DBP, diastolic blood pressure, n, number of participants. the low and medium pressure regions, and an A/C rating in the high-pressure region ( > 160/100 mmhg). This analysis provides valuable information suggesting reduced accuracy at high diastolic pressures. These results, however, do not influence the overall A/A rating for the device. Discussion The Omron 705IT (HEM-759-E) achieved an overall A grade for both systolic and diastolic BP according to the BHS protocol. It also fulfilled the AAMI criteria. It should be pointed out that the participant population, designed here to meet BHS protocol, differed slightly from AAMI standard requirements. Specifically, the AAMI standard requires participants representing an adult population to be distributed such that at least 10% have SBP and DBP readings below 100 and 60 mmhg, respectively, and 10% of participants have SBP and DBPs above 160 and 100 mmhg, respectively. Twenty-eight out of 85 (32.9%) participants had SPB and DBPs above the upper thresholds, which is more than the 10% minimum, as required, but only eight participants (9.4%) had SBP and DBPs below the lower thresholds, which represents one participant less than required in this range. The AAMI standard for adult populations also specifies the arm circumference distribution requirements such that at least 10% of participants have an arm circumference above 35 cm, and 10% have an arm circumference below 25 cm. The distribution of participants used in this study satisfies the larger arm circumference requirement, with 11 out of 85 (12.9%) above 35 cm, but does not achieve the smaller arm circumference requirement, having seven participants (8.2%) below 25 cm, two participants less than required. The device can be considered to have passed the AAMI criteria for the study population, which, in practice, deviates only slightly from that specified in the standard to represent an adult population. On the basis of both BHS and AAMI requirements, the Omron 705IT (HEM-759-E) can, therefore, be recommended for professional as well as home-use in an adult population. Only three automated devices currently recommended by the International Advisory Body, Dable Educational Trust, have achieved a BHS A/A grade. This device will usefully increase the currently limited choice of validated automated devices. The device performed well in all ranges except in the diastolic high-pressure range, for which it received a C grade. It is not uncommon for oscillometric monitors to have a tendency to underestimate the systolic pressure. It has been suggested that this could be a function of the vascular compliance, which will be reduced in participants with arterial disease, causing alterations in the waveforms, rendering the algorithms inaccurate [10]. It is also possible that this apparent poorer performance at high pressures is an artefact resulting from the larger fluctuations that occur in BP at higher pressures. This explanation would suggest that larger device observer differences may occur during the > 30-s intervals between sequential mercury and test device readings at higher BPs. One modification to the BHS protocol in this study involved changing the low SBP range from eight participants with SBP < 90 mmhg to eight participants with SBP in the range r 100 mmhg and 20 participants with an SBP in the range mmhg to 20 participants whose SBP fell in the range mmhg. In practice, it is extremely difficult to find participants with a systolic pressure of less than 90 mmhg. This is an accepted modification to the BHS used in other published studies [10,11]. The rationale for this modification is that, given the rarity of such pressures in the general population, it is unlikely that the device will be used clinically over these low ranges. The total number of participants recruited, 143, was larger than that reported in trials completed 2 3 years ago using the same hospital clinics (126 and 95) [10,11]. This resulted from the need to continue unproductive recruitment for a long period in order to find the 20 patients in the DBP category, mmhg, and SBP category, mmhg while other categories were full. The reason for this increased recruitment difficulty in these BP categories is not known, but may reflect the increasingly effective control of BP in the local clinic populations. A further modification to the BHS protocol in this study has been to replace the cumbersome manual dynamic pressure calibration with a simple, fast and accurate semiautomated static pressure calibration. The PCcontrolled electronic calibrated reference pressure actuator has an accuracy exceeding that given by visual readings of a mercury column by at least one order of

6 32 Blood Pressure Monitoring 2006, Vol 11 No 1 magnitude. This approach is particularly appropriate as the digital scales of most modern BP monitors are capable of higher precision than the mercury reference devices [12]. A static calibration, as used here, does not confirm that the dynamic response of the test device mimics that of the falling column of the mercury reference device as intended by the BHS protocol. The dynamic response of the test device, however, is compared with that of the mercury sphygmomanometer as part of the clinical validation, where the systolic and diastolic pressures are determined from a continuously falling pressure. This redundancy of the initial dynamic calibrations in the BHS protocol is reflected in the design of the newer ESH protocol, which no longer specifies the need for calibration of the test device. A static calibration accuracy of better than ± 3 mmhg, however, is a requirement of the European Specification for Sphygmomanometers [13], and the calibration performed in this study usefully serves to confirm that the test device satisfies this standard before embarking on the expensive and timeconsuming clinical phase of the validation. The before and after clinical-use calibrations also serve to test the long-term stability of this calibration, which has important implications for its suitability for clinical use [12]. Acknowledgement Appreciation is expressed to Anne-Marie Reinders for her advice. References 1 O Brien E. Replacing the mercury sphygmomanometer. BMJ 2000; 320: O Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, Altman DG, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11:S2543 S O Brien E, Pickering T, Asmar R, Myers M, Parati G, Staessen J, et al. Working Group on blood pressure monitoring of the European Society of Hypertension international protocol for validation of blood pressure devices in adults. Blood Press Monit 2002; 7: Association for the Advancement of Medical Instrumentation. American National Standard for manual, electronic, or automated sphygmomanometers. ANSI/AAMI SP10: Arlington, Virginia: AAMI; O Brien E. A web site for blood pressure measuring devices: dableducational.com. Blood Press Monit 2003; 8: Dabl Educational web site. [Accessed February 2005]. 7 El Assaad MA, Topouchian JA, Asmar RA. Evaluation of two devices for selfmeasurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT. Blood Press Monit 2003; 8: British Hypertension Society web site. [Accessed February 2004]. 9 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: Jones CR, Taylor K, Chowienczyk P, Poston L, Shennan AH. A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor. Blood Press Monit 2000; 5: Cuckson AC, Reinders A, Shabeeh H, Shennan A. Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol. Blood Press Monit 2002; 7: Coleman AJ, Steel SD, Ashworth M, Vowler SL, Shennan A. The accuracy of the pressure scale of sphygmomanometers in clinical use within primary care. Blood Press Monit 2005; 10: BS EN1060-1:1996. Specification for non-invasive sphygmomanometers. Part 1: general requirements. Brussels; 1996.

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009 42 Original article Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP, the Omron HBP T and the Pic Indolor Professional Roland

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 16 Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-71-E)

More information

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Original article http://dx.doi.org/10.3345/kjp.2011.54.11.463 Korean J Pediatr 2011;54(11):463-469 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 223 Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol Paolo Verdecchia, Fabio Angeli, Roberto Gattobigio,

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

Validation Study Result Form

Validation Study Result Form Validation Study Result Form Please complete Section 1 to Section 5 of this form and return it to dabl Educational Trust with copies of the validation Study Ref. plots. In Section 3 to Section 5, complete

More information

Maker Spacelabs Healthcare Manufacturer Spacelabs Healthcare. Brand Spacelabs Model OnTrak

Maker Spacelabs Healthcare Manufacturer Spacelabs Healthcare. Brand Spacelabs Model OnTrak Validation Study Result Form Please complete Section 1 to Section 3 of this form and return it to dabl Educational Trust with copies of the validation plots and a digital photograph of the device used

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

Home blood pressure measurement with oscillometric upper-arm devices

Home blood pressure measurement with oscillometric upper-arm devices REVIEW Home blood pressure measurement with oscillometric upper-arm devices R.L. Braam, Th. Thien *# Department of Internal Medicine, University Medical Centre St Radboud, PO Box 9101, 6500 HB Nijmegen,

More information

Validation Study Result Form

Validation Study Result Form Validation Study Result Form Please complete Section 1 to Section 3 of this form and return it to dabl Educational Trust with copies of the validation plots. The requirements for each of these sections

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the Omron M6 Comfort IT (HEM-7322U-E) with the Omron M6 Comfort (HEM-7321-E) Devices Omron M6 Comfort IT (HEM-7322U-E) Omron M6 Comfort (HEM-7321-E) Pictures Display Validation ESH 2010 Device

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval BPU22 with the AViTA BPM63S Devices Item 9 HARTMANN Veroval BPU22 AViTA BPM63S Pictures Display Image Validation Equivalence ESH 2010 ESH 2002 BHS AAMI Category Arm Type

More information

Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus Devices Microlife A6 PC (BP3GU1-8Y) 9 BP3BT0-A 9 Microlife BP A100 Plus 9 Image 10 10 10 Validation

More information

Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus Devices Microlife A2 Basic (BP3GQ1-3P) 9 BP3BT0-A 9 Microlife BP A100 Plus 9 Image 10 10 10 Validation

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

Device Judgement Recommended

Device Judgement Recommended dabl Educational Trust Validation Checklist Device Judgement Recommended Manufacturer IEM Device Stabil O Graph Category Upper Arm (Self) Mode Oscillometric Protocol Reference BHS + ESH Westhoff TH, Schmidt

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval upper arm blood pressure monitor with the AViTA BPM63S Devices Pictures HARTMANN Veroval upper arm blood pressure monitor AViTA BPM63S

More information

There is convincing evidence in clinical studies

There is convincing evidence in clinical studies AJH 1998;11:1413 1417 Reliability of Reporting Self-Measured Blood Pressure Values by Hypertensive Patients Thomas Mengden, Rosa Maria Hernandez Medina, Belen Beltran, Elena Alvarez, Karin Kraft, and Hans

More information

Validation Study Registration Form

Validation Study Registration Form TrUSt Validation Study Registration Form This form should be completed by the manufacturer or on behalf of the manufacturer. Study Investigator Details Name Address Mrs.Annemarie de Greef and Professor

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the SAFETY PRONTEX INTEGRA (Nissei DS-B33-03) with the Nissei DSK-1011 Devices Item 9 SAFETY PRONTEX INTEGRA (Nissei DS-B33-03) Nissei DSK-1011 Pictures Display Image Validation ESH IP2010

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval BPW22 with the AViTA BPM15S Devices Item 9 HARTMANN Veroval BPW22 AViTA BPM15S Pictures Display Image Validation Equivalence ESH 2010 ESH 2002 BHS AAMI Category Wrist

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

The increasing awareness of hypertension as a serious

The increasing awareness of hypertension as a serious AJH 2002; 15:787 792 Evaluation of the Performance of a Wrist Blood Pressure Measuring Device With a Position Sensor Compared to Ambulatory 24-Hour Blood Pressure Measurements Sakir Uen, Burkard Weisser,

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval wrist blood pressure monitor with the AViTA BPM15S Devices HARTMANN Veroval wrist blood pressure monitor AViTA BPM15S Pictures Display

More information

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator MSc Thesis in Clinical Science and Technology Sara Rose Newell Department of Health Science

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008 Analytical methods and statistical analyses 285 PA.NET International Quality Certification Protocol for blood pressure monitors Stefano Omboni a,b, Carlo Costantini a,b,c, Claudio Pini b,d, Roberto Bulegato

More information

Special articles from the ESH working group on blood pressure monitoring 23

Special articles from the ESH working group on blood pressure monitoring 23 Special articles from the ESH working group on blood pressure monitoring 23 European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices

More information

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

!#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 !"#$ Blood Pressure Monitor Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 Introduction

More information

Primary hypertension in adults

Primary hypertension in adults Primary hypertension in adults NICE provided the content for this booklet which is independent of any company or product advertised Hypertension Welcome NICE published an updated guideline on the diagnosis

More information

Declaration of Equivalence Form

Declaration of Equivalence Form Declaration of Equivalence Form DECLARATION OF BLOOD PRESSURE MEASURING DEVICE EQUIVALENCE 2013 SECTION A - Please complete all items. I Bill Huang,, Name of a Company Director A SIGNED COPY WILL BE POSTED

More information

Data Collection Worksheet

Data Collection Worksheet Data Collection Worksheet 1. Has a doctor or nurse ever said that you have: High blood pressure or hypertension? [ ] No [ ] Yes [ ] Not Sure 1.a. If Yes, then at what age were you first told this? Age

More information

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? O R I G I N A L A R T I C L E Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? I. Aksoy, J. Deinum, J.W.M. Lenders, Th. Thien *# Department of

More information

greater than 90% of patients having blood pressure measured consistently according to the protocol.

greater than 90% of patients having blood pressure measured consistently according to the protocol. Clinical methods and pathophysiology 1 Reliability and validity of blood pressure measurement in the Secondary Prevention of Small Subcortical Strokes study Pablo E. Pérgola a, Carole L. White b, John

More information

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home.

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home. Welch Allyn ABPM 6100S Comprehensive blood pressure monitoring that brings patient comfort home. Features and Benefits of the Welch Allyn ABPM 6100S Pictorial application instructions, artery markers and

More information

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice.

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice. AJH 26; 19:81 89 BP Measurement Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice Tony C. Y. Pang and Mark A. Brown Background: To determine the extent of discrepancies between

More information

In clinical practice, blood pressure (BP) treatment

In clinical practice, blood pressure (BP) treatment Original Paper Does Hidden Undercuffing Occur Among Obese Patients? Effect of Arm Sizes and Other Predictors of the Difference Between Wrist and Upper Arm Blood Pressures Hardik Doshi, MD; Alan B. Weder,

More information

MX3 Plus English. Introduction. Table of contents. 1 How to obtain meaningful blood pressure readings

MX3 Plus English. Introduction. Table of contents. 1 How to obtain meaningful blood pressure readings X3 Plus English X3 Plus Introduction The ORON X3 Plus measures your blood pressure and pulse simply and quickly from the arm, without the use of an inflation bulb or stethoscope. The measurement values

More information

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon (2009) 23, 794 800 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Unreliable oscillometric blood pressure measurement: prevalence, repeatability

More information

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood SPECIAL ARTICLE Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Eoin O Brien a,thomas

More information

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License Zheng DC, Liu CY, Amoore J, Mieke S, Murray A. Need for Re-validation of Automated Blood Pressure Devices for use in Unstable Conditions. In: 215 Computing in Cardiology Conference (CinC). 215, Nice, France:

More information

The introduction of oscillometric automated blood pressure

The introduction of oscillometric automated blood pressure Atrial Fibrillation Impact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring Nikolaos Pagonas,* Sven Schmidt,* Jörg Eysel, Friederike Compton, Clemens Hoffmann, Felix Seibert,

More information

Use and Interpretation of Home Blood Pressure Monitoring

Use and Interpretation of Home Blood Pressure Monitoring Use and Interpretation of Home Blood Pressure Monitoring BLOOD PRESSURE in year 2015 PARAMETERS TODAY Gianfranco Parati University of Milano-Bicocca & Ospedale S.Luca, IRCCS, Istituto Auxologico Italiano

More information

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure? MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents 01... Blood Pressure Monitor Intended Use 02... 1. What is blood pressure? 02... 2. Why is it useful measure blood pressure at home?... A. WHO blood pressure

More information

M1 Classic. Digital Blood Pressure Monitor. Instruction manual page 1 SYS. mmhg DIA. mmhg. PULSE/min

M1 Classic. Digital Blood Pressure Monitor. Instruction manual page 1 SYS. mmhg DIA. mmhg. PULSE/min 1 Classic Digital Blood Pressure onitor Instruction manual page 1 1 Classic English Introduction The ORON 1 Classic measures your blood pressure and pulse simply and quickly from the arm, without the use

More information

Please do not hesitate to contact our dedicated service unit for more details regarding this program at

Please do not hesitate to contact our dedicated service unit for more details regarding this program at The Blue Cross and Blue Shield Federal Employee Program (FEP) and the American Medical Association (AMA) have come together in a collaborative effort to provide physicians with resources designed to improve

More information

doi: /fampra/cmw053

doi: /fampra/cmw053 doi: 10.1093/fampra/cmw053 Manuscript type Original research Title Comparison of blood pressure measurements on the bare arm, over a sleeve, and over a rolled-up sleeve in the elderly Short title Blood

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Microlife WatchBP Home A

Microlife WatchBP Home A Microlife The Microlife is designed to provide reliable and unbiased blood pressure measurement results, and strictly follows European Society of Hypertension (ESH) and American Heart Association (AHA)

More information

Automated versus manual blood pressure measurement: A randomized crossover trial

Automated versus manual blood pressure measurement: A randomized crossover trial Automated versus manual blood pressure measurement: A randomized crossover trial Author Heinemann, Megan, Sellick, Ken, Rickard, Claire, Reynolds, Pam, McGrail, Matthew Published 2008 Journal Title International

More information

BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN?

BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN? BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN? DECLARATION OF INTEREST Medical Director and Chairman, Advisory Board dabl Ltd., Dublin, Ireland. BLOOD PRESSURE MEASUREMENT

More information

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation (2009), 1 5 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT Blood Pressure Measurement Can Isik Electrical Engineering and Computer Science Department Syracuse University Syracuse, NY 13244 cisik@syr.edu Abstract Blood pressure is one of the fundamental vital signs,

More information

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA)

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) THE NEW ARMENIAN MEDICAL JOURNAL Vol.5 (2011), Nо 2, p.29-34 DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) Zelveian P.H. 1, 2, Podosyan G.A. 2 1 Institute

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials. AJH 1995; 8:311-315 BRIEF COMMUNICATIONS Lack of Placebo Effect on Ambulatory Blood Pressure Giuseppe Mancia, Stefano Omboni, Gianfranco Parati, Antonella Ravogli, Alessandra Villani, and Alberto Zanchetti

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

2. Measure a subject's blood pressure and heart rate both at rest and during exercise. Lab Activity 11 The Cardiovascular System Student Learning Objectives After completing this lab, you should be able to: 1. Define, explain and correctly use the key terms. 2. Measure a subject's blood

More information

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER Naser KA. Teaching Hospital Peradeniya, Peradeniya, Sri Lanka Zawahir S

More information

RX-3. Blood Pressure Monitor. Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81

RX-3. Blood Pressure Monitor. Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 RX-3 Blood Pressure Monitor Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 IM-HEM-640-E-01-08/08

More information

THE ACCURATE ASSESSMENT

THE ACCURATE ASSESSMENT ORIGINAL INVESTIGATION Consistency of Blood Pressure Differences Between the Left and Right s Kazuo Eguchi, MD; Mona Yacoub, MD; Juhee Jhalani; William Gerin, PhD; Joseph E. Schwartz, PhD; Thomas G. Pickering,

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the Microlife BP A150 AFIB with the Microlife BP A100 Plus and Microlife BP A100 Device Equivalence Evaluation Form Devices Microlife BP A150 AFIB 9 Microlife BP A100 Plus 9 Microlife BP

More information

Alternatives to the mercury sphygmomanometer

Alternatives to the mercury sphygmomanometer Original Article Alternatives to the mercury sphygmomanometer Susan Buchanan a, *, Peter Orris b and Joshua Karliner c a Environmental and Occupational Health Sciences Division, University of Illinois

More information

An Automated System for Blood Pressure Determination During Exercise

An Automated System for Blood Pressure Determination During Exercise An Automated System for Blood Pressure Determination During Exercise SEPHEN P. GLASSER, M.D., AND MAYNARD R. RAMSEY III, M.D., PH.D. with the technical assistance ofblanche Price SUMMARY Automation of

More information

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015 Original article 1 Clinical practice of ambulatory versus home blood pressure monitoring in hypertensive patients Jorge A. Paolasso, Florencia Crespo, Viviana Arias, Eduardo A. Moreyra, Ariel Volmaro,

More information

Céline Guggiari a, Christophe Büla a, Katia Iglesias b and Bernard Waeber b,c

Céline Guggiari a, Christophe Büla a, Katia Iglesias b and Bernard Waeber b,c Published in Blood Pressure Monitoring 19, issue 11, 32-37, 2014 which should be used for any reference to this work 1 Measurement with an automated oscillometric wrist device with position sensor leads

More information

boso ABI-system 100 Early detection of PAOD fast, precise, without Doppler Premium quality for doctors and hospitals

boso ABI-system 100 Early detection of PAOD fast, precise, without Doppler Premium quality for doctors and hospitals boso ABI-system 100 Early detection of PAOD fast, precise, without Doppler Premium quality for doctors and hospitals Product of tradition Premium quality for doctors and hospitals For decades, boso has

More information

The Value of a BP Determination Method Using a Novel Non-Invasive BP Device against the Invasive Catheter Measurement

The Value of a BP Determination Method Using a Novel Non-Invasive BP Device against the Invasive Catheter Measurement The Value of a BP Determination Method Using a Novel Non-Invasive BP Device against the Invasive Catheter Measurement Jinsong Xu 1., Yanqing Wu 1., Hai Su 1 *, Weitong Hu 1, Juxiang Li 1, Wenying Wang

More information

B lood pressure is the pressure exerted by the circulating blood against vessel walls, and is a vital sign in clinical

B lood pressure is the pressure exerted by the circulating blood against vessel walls, and is a vital sign in clinical OPEN SUBJECT AREAS: BIOMEDICAL ENGINEERING ELECTRICAL AND ELECTRONIC ENGINEERING Tissue-Informative Mechanism for Wearable Non-invasive Continuous Blood Pressure Monitoring Sung Hun Woo 1, Yun Young Choi

More information

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

PDF hosted at the Radboud Repository of the Radboud University Nijmegen PDF hosted at the Radboud Repository of the Radboud University Nijmegen The following full text is a publisher's version. For additional information about this publication click this link. http://hdl.handle.net/2066/53625

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License Pan F, Zheng DC, He PY, Murray A. Does the Position or Contact Pressure of the Stethoscope Make Any Difference to Clinical Blood Pressure Measurements: An Observational Study. Medicine 214, 93(29). Copyright:

More information

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room. Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room Abstract Ahmad Mirdamadi (1), Mostafa Etebari (2) Original Article BACKGROUND:

More information

Arm position and blood pressure: a risk factor for hypertension?

Arm position and blood pressure: a risk factor for hypertension? (3) 17, 389 395 & 3 Nature Publishing Group All rights reserved 09-9240/03 $25.00 www.nature.com/jhh ORIGINAL ARTICLE Arm position and blood pressure: a risk factor for hypertension? A Mourad, S Carney,

More information

No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death

No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death STROKE statistics No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death Every 6 seconds a person in the world has a STROKE regardless of age and gender. Worldwide 15 million people suffer

More information

Comparison of blood pressure measured at the arm, ankle and calf

Comparison of blood pressure measured at the arm, ankle and calf doi:10.1111/j.1365-2044.2008.05633.x Comparison of blood pressure measured at the arm, ankle and calf C. Moore, 1 A. Dobson, 2 M. Kinagi 2 and B. Dillon 3 1 SpR Anaesthetics, 2 Consultant Anaesthetist,

More information

Home monitoring of blood glucose

Home monitoring of blood glucose Home Monitoring of Glucose and Blood Pressure JAMES R. TAYLOR, PharmD, CDE, and KENDALL M. CAMPBELL, MD, University of Florida, Gainesville, Florida Home monitoring of blood glucose and blood pressure

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

apneabp card(x)plore ABPM-04 Ambulatory Blood Pressure Monitors & Combinations

apneabp card(x)plore ABPM-04 Ambulatory Blood Pressure Monitors & Combinations apneabp card(x)plore ABPM-04 Ambulatory Blood Pressure Monitors & Combinations Meditech Cardiovascular Monitoring Solutions Meditech is one of the worldwide technology leaders in the area of manufacturing

More information

The Smart Blood Pressure Monitor. Instruction manual

The Smart Blood Pressure Monitor. Instruction manual The Smart Blood Pressure Monitor Instruction manual The Smart Blood Pressure Monitor Read this manual before use. Keep it for future reference. Thank you for purchasing the Withings Blood Pressure Monitor.

More information

Importance Of Taking Manual Blood Pressure In Both Arms

Importance Of Taking Manual Blood Pressure In Both Arms Importance Of Taking Manual Blood Pressure In Both Arms Its importance to health is now well-recognised, in that high blood pressure can lead to Have a look at our slide show on how to take a blood pressure

More information

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust.

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust. AN INTRODUCTION TO DOPPLER Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust. THE DOPPLER EFFECT The Doppler Principle was described by Physicist and mathematician

More information

STATE OF THE ART BP ASSESSMENT

STATE OF THE ART BP ASSESSMENT STATE OF THE ART BP ASSESSMENT PROFESSOR MOLECULAR PHARMACOLOGY CONWAY INSTITUE UNIVERSITY COLLEGE DUBLIN CO-CHAIRMAN BLOOD PRESCSURE MANAGEMENT IN LOW RESOURCE SETTINGS CENTRE FOR INTERNATIONAL HUMANITARIAN

More information

Automatic oscillometric blood pressure devices are frequently

Automatic oscillometric blood pressure devices are frequently Arterial Stiffness as Underlying Mechanism of Disagreement Between an Oscillometric Blood Pressure Monitor and a Sphygmomanometer Nicole M. van Popele, Willem Jan W. Bos, Nicole A.M. de Beer, Deirdre A.M.

More information