1 AJH 1998;11: Reliability of Reporting Self-Measured Blood Pressure Values by Hypertensive Patients Thomas Mengden, Rosa Maria Hernandez Medina, Belen Beltran, Elena Alvarez, Karin Kraft, and Hans Vetter For self-measurement of blood pressure to be useful, patient reporting of test results must be reliable and accurate. Until now no study directly measured the accuracy and reliability of patients reporting of self-measured blood pressure values. Thirty hypertensive patients (69 11 years) were instructed to measure blood pressure at home over 14 days with the highly accurate Omron IC monitor and to keep a record of all readings in a patient logbook. To assess the reliability of the records, patients were not informed about the memory capacity of the device. We compared automatically stored blood pressure readings with the respective logbook entries to analyze deletion (under-reporting), addition (over-reporting), and precision of reporting of test results. The prevalent pattern was under-reporting, averaging 36% 24% (3% to 89%), which occurred significantly more than over-reporting (9% 11%; 0% to 38%). The precision of reporting (identical values at corresponding times) was 76% 34% (0% to 100%). This observer error did not affect group comparisons of automatically stored values and logbook entries, although the estimated limits of agreement were wide. Blood pressure control, duration of hypertension, age, or previous use of self-measurement and patterns of logbook entries were not found to be predictive of the patients reliability. Our results demonstrate a substantial observer error in the reporting of self-measured blood pressure values. This bias may be reduced by memory-equipped blood pressure devices. Am J Hypertens 1998;11: American Journal of Hypertension, Ltd. KEY WORDS: Self-measurement of blood pressure, observer bias, memory equipped blood pressure device, reliability. There is convincing evidence in clinical studies that multiple blood pressure measurements obtained by self-monitoring patients at home improve the precision and reproducibility of blood pressure measurements. 1,2 Furthermore, epidemiologic and pharmacologic blood pres- Received April 20, Accepted October 6, From the Department of Internal Medicine, Medizinische Poliklinik, Bonn, Germany. Address correspondence and reprint requests to Thomas Mengden, MD, Medizinische Universitäts-Poliklinik, University Clinic of Bonn, Wilhelmstrasse 35-37, Bonn, Germany. sure studies have shown an increasing trend for selfmonitoring. 3 6 The increasing trend is mainly patientdriven and electronic devices for self-monitoring are now widely available in drug stores and mail order catalogues. 7,8 Self-monitoring of blood pressure is only useful if patient-generated data are reliable and accurate. Until now, analysis of self-monitored blood pressure values depended on the accurate reporting of measurements in the patients logbooks. In view of many well-known observer errors in clinical blood pressure measurement, 9,10 it is astonishing that all studies performed with self-monitoring heavily relied on patient-reported values. No study directly mea by the American Journal of Hypertension, Ltd /98/$19.00 Published by Elsevier Science, Inc. PII S (98)

2 1414 MENGDEN ET AL AJH DECEMBER 1998 VOL. 11, NO. 12 sured the accuracy and reliability of the routine daily use of home monitoring and the quality of of the reporting of the results. 8 PATIENTS AND METHODS Patients Thirty consecutive patients with arterial hypertension (systolic blood pressure 140 mm Hg, diastolic blood pressure 90 mm Hg, according to the World Health Organization definition) were chosen from an outpatient clinic in the Department of Internal Medicine at the University of Bonn. Twenty-one patients were already taking antihypertensive medication (11 monotherapy and 10 combination therapy). The duration of arterial hypertension ranged from 0 to 25 years. Sixteen patients had already purchased a device for home monitoring on their own, but had not been sufficiently instructed in the technique by a physician. Procedures Self-measurement was performed by all patients with a successfully tested and previously calibrated fully-automatic, oscillometric device with automatic inflation and deflation and a digital display of blood pressure and heart rate (Omron Intensive Control, IC; Omron, Matsusaka, Japan). The Omron IC has two components, the Omron IC blood pressure monitor and the Omron IC software. The Omron IC monitor looks like a standard blood pressure monitor. The system used in the present study is highly accurate and includes the same algorithm as the Omron HEM-705 CP. 11 As a special feature, the Omron IC is equipped with a memory chip and an interface for connecting it to a computer. The software can be used for statistical analysis as well as to prepare a schedule for the patient to set the measuring times. After sufficient instruction in the correct technique of self-measurement by a physician in charge of the trial, patients were instructed to measure blood pressure at home once in the morning and once in the evening at preset times over a period of 2 weeks. The patients were not informed about the memory capacity of the device. They were instructed to keep a record of all blood pressure readings with the corresponding time of measurement in a patient logbook. The patients were not instructed to validate the Omron IC measurements. The instructions were standardized and always performed by the same physician (TM). Schedules of blood pressure measurement were individually prepared according to the usual living habits and intake of antihypertensive medication (6 to 8 am and 6 to 8 pm). At set times the patient was reminded to measure blood pressure by an acoustic signal from the blood pressure device. Analyses Four determinations were made for each patient: 1) the number of blood pressure recordings stored in the Omron IC system; 2) the number of blood pressure recordings in the patients logbook; 3) the number of blood pressure recordings in the Omron IC system with a recording of any value in the logbook at the corresponding time; and 4) the number of corresponding time recordings in the Omron IC system and logbook showing identical values. From these data three measurements were derived: 1) over-reporting, 2) under-reporting, and 3) precision. A measure of over-reporting was developed to assess the degree to which patients added the results of tests that were not actually performed. Over-reporting was defined as the number of blood pressure reportings that were not verified at a corresponding time in the Omron IC system expressed as percentage of blood pressure recordings in the logbook. A score of 0% indicated that every logbook entry was verified by a recording of any value in the Omron IC system at the corresponding time. To assess the degree to which patients omitted test results, an under-reporting score was developed. Under-reporting was defined as the number of readings in the Omron IC system that were not reported in the patient logbook expressed as percentage of the total number of blood pressure readings in the Omron IC system. A score of 0% would thus indicate that every recording of the Omron IC system was entered in the logbook. We assessed the accuracy of patients reporting of blood pressure readings by comparing them to the Omron IC recorded readings. Precision was defined as the number of recordings in the Omron IC system and logbook that were of identical value and reported at the corresponding time expressed as percentage of the number of corresponding readings found in the Omron IC system and the logbook. A score of 100% would thus indicate that every blood pressure value written in the logbook was exactly the same as the value recorded in the Omron IC system at the corresponding time. Data are expressed as means SD. The data were analyzed using paired t tests to compare mean logbook values with Omron IC system values. Over- and under-reporting scores were compared by a Wilcoxon range test for nonparametric variables. Linear regression analysis was used to examine the correlation between the Omron IC system-derived mean blood pressure values, duration of hypertension, or age and the reliability (percentage of underreporting, over-reporting, and precision). Linear regression analysis was also used to examine the correlation between under-reporting, over-reporting, and precision.

3 AJH DECEMBER 1998 VOL. 11, NO. 12 SELF-MEASURED BP REPORTING RELIABILITY 1415 TABLE 1. AVERAGE ( SD) FOR TOTAL NUMBER OF AUTOMATICALLY STORED MONITOR PRESSURES, TOTAL NUMBER OF LOGBOOK PRESSURES, AND TOTAL NUMBER OF ACCURATE LOGBOOK PRESSURES (CORRESPONDING TIME RECORDINGS IN THE OMRON IC SYSTEM AND LOGBOOK SHOWING IDENTICAL VALUES) Monitor Logbook RESULTS Accurate Logbook Mean SD Reporting and Precision The total number of automatically stored monitor pressures, total number of logbook pressures, and total number of accurate logbook pressures (corresponding time recordings in the Omron IC system and logbook showing identical values) are given in Table 1. All patients showed more or less extensive underreporting ranging from 3% to 89% (mean 36% 24%). Over-reporting scores were significantly lower than under-reporting scores (P.0001) and ranged from 0% to 35% (mean 9% 11%). Nineteen of the 30 patients showed a pattern of both omitting readings from and adding phantom values to logbook entries. There was no relationship between patterns of under- and over-reporting. Precision ranged from 0% to 100% (mean 76% 34%). Fifteen patients showed precision indices 90%. There was no significant correlation between precision and reporting. Thus neither omission nor addition of values to logbook entries was indicative of poor precision. The analysis of individual cases showed variable patterns of patient performance with reporting of selfmeasured blood pressures (Figure 1). Each of these three patients had already performed self-measurement at home before inclusion in the study and was familiar with reporting self-generated blood pressure data in the logbook. All three patients were taking antihypertensive medication during the study. We then tested whether there was a relationship between adequate blood pressure control, age, duration of hypertension, or previous use of self-measurement and reliability. We could not establish a significant correlation between blood pressure control, age, or duration of hypertension and the percentage of under-, over-reporting, or precision. Furthermore, previous use of self-measurement did not predict a good reliability of patients reporting. Comparison of Logbook Entries With Automatically Stored Values This observer error did not affect group comparisons of automatically stored values and logbook entries, although the estimated limits of agreement were wide. Thus, the mean of all subjects average logbook pressures (155 15/87 9mmHg) were very similar compared with the automatically stored monitor pressures (154 16/87 8mmHg;P NS). The correlation coefficients between logbook entries and Omron IC values were r 0.95 (P.0001) for systolic and r 0.95 (P.0001) for diastolic values, respectively. However, for individual patients we observed differences between logbook entries and Omron IC values ranging from 11 to 10 mm Hg for systolic blood pressure and 5 to 7 mm Hg for diastolic blood pressure. DISCUSSION The widespread use of self-measurement has resulted in a marked reliance on patient-reported blood pressure data. The reliability of patient-reported logbook data has not been investigated, although observer errors of blood pressure measurement are a general finding in blood pressure studies. 9,10 The problems encountered with the use of older devices consisted of terminal digit preference, expectation bias, and false cuff inflation and deflation. The development of accurate, fully automated oscillometric devices with a digital display has substantially reduced the observer errors of blood pressure measurement at home. The present study is the first one investigating reliability in terms of observer error in reporting selfmeasured blood pressure values. The system used in the present study, modified to store all blood pressure readings with the respective time of measurement, enabled us to directly address the issue of patient performance in documenting self-measured blood pressures. We found a high degree of variability in underreporting, over-reporting, and precision of these data. In principal, the main error consisted of omission of additional blood pressure readings performed at times different from the scheduled set points. Thus, the mean number of logbook entries was significantly lower than the total number of measurements actually performed and stored in the Omron IC system (30 10 v 49 28; P.01). Although in individual patients the omission of values from the logbook resulted in different blood pressure values (Figure 1B), this did not affect the group comparisons of blood pressure values. The very similar results for mean logbook pressures and monitor pressures may be explained by a pattern of rejecting both low and high blood pressure values that patients chose not to write down. However, we could not detect a systematic pattern of underreporting. In this context it is interesting that in a study with

4 1416 MENGDEN ET AL AJH DECEMBER 1998 VOL. 11, NO. 12 FIGURE 1. Comparison of systolic blood pressure values derived from logbook entries with the automatically stored values of the Omron IC software over the whole study period in three selected patients. The open circles with solid lines depict all readings from the Omron IC software. The closed squares represent patient-reported logbook entries. Panel A: Patient 6 showed a precision index of 0% combined with under-reporting of 26% and over-reporting of 5%. In spite of the low precision, differences between patient-reported and automatically stored values were not relevant (1.1/1.1 mm Hg). A detailed analysis revealed that this patient entered all blood pressure readings in the logbook two days later than actually performed, apparently because of mixing up dates. Panel B: Patient 30 also showed a low precision index of 18% with high under- and over-reporting. Mean logbook values were significantly lower than Omron IC values (9.5/3.9 mm Hg), because this patient omitted high blood pressure readings. Panel C: Patient 21 showed a similar degree of under-reporting (65%) compared to patient 30. The precision index was, however, 100% and there was no relevant difference between logbook entries and stored Omron IC values (0.1/0.0 mm Hg). insulin-dependent diabetic patients who monitored their blood glucose levels at home there was a tendency to report lower values than the device actually displayed. This was shown by inserting a memory chip in the device without the patients knowledge. 12 In this study the precision index of the reported blood glucose values was 74% and thus very similar to the index found in our study (76%). In contrast to the present study in the diabetic patients, over-reporting was the prevalent pattern of reporting and averaged 40% (under-reporting was 10%). The prevalent errors in the diabetic patients consisted of omission of high glucose readings and the addition of values lower than the true test results. 12 We thus identified a degree of unreliability among the study patients that may have profound effects on the usefulness of self-measured blood pressure. This inaccuracy could not be explained by inadequate technique or failure of the blood pressure device. It seems unlikely that the observer bias documented in the present study was due to basic patient characteristics, such as age or blurred vision, since observer bias in

5 AJH DECEMBER 1998 VOL. 11, NO. 12 SELF-MEASURED BP REPORTING RELIABILITY 1417 blood pressure measurement is a more general finding. 9,10 Digit preference was evident in physician measurements as well as patient self-measurement in the recently published Self-Measurement for the Assessment of the Response to Trandolapril (SMART) study. 3 The findings emerging from our observations have important implications for the future planning of epidemiologic and pharmacologic studies. In summary, the degree of reliability of hypertensive patients reporting of self-measured blood pressure values was both variable and unpredictable. We were not able to establish a relationship between the different patterns of reporting or to associate these patterns with overall blood pressure control or previous use of self-measurement. The use of memoryequipped blood pressure devices with the patients knowledge and the providers feedback may ameliorate the problems of observer bias and contribute to improved treatment of hypertension. ACKNOWLEDGMENT We thank Frank Wilmerstädt (Omron Health Care Europe) for technical support of the study. REFERENCES 1. Mengden T, Bättig B, Vetter W: Self-measurement of blood pressure improves the accuracy and reduces the number of subjects in clinical trials. J Hypertens 1991; 9(suppl 6): Mengden T, Binswanger B, Jeck T, et al: An evaluation of self-measured blood pressure in a study with a calcium channel antagonist v a -blocker. Am J Hypertens 1992;5: Chatellier G, Dutrey-Dupagne C, Vaur L, et al: Home self blood pressure measurement in general practice. The SMART study. Am J Hypertens 1996;9: Weisser B, Grüne S, Burger R, et al: The Dübendorf Study: a population-based investigation on normal values of blood-pressure self-measurement. J Human Hypertens 1994;8: Ménard J, Serruier D, Bautier P, et al: Crossover design to test antihypertensive drugs with self-recorded blood pressure. Hypertension 1988;11: Mengden T, Binswanger B, Grüne S, et al: The use of self-measured blood pressure determinations in assessing dynamics of drug compliance in a study with once daily morning vs evening amlodipine. J Hypertens 1993;11: Frost and Sullivan Ltd.: The European Market for Cardiovascular Diagnostic Equipment. London, June Report: E Pickering T: Recommendations for the use of home (self) and ambulatory blood pressure monitoring. Am J Hypertens 1996;9: Bruce NG, Shaper AG, Walker M, Wannamethee G: Observer bias in blood pressure studies. J Hypertens 1988;6: Kolb C, Huss R, Mengden T, Vetter W: How reliable is conventional blood pressure registration? Comparison with a semiautomatic device. Schweiz Rundsch Med Prax 1991;80(12): O Brien E, Mee F, Atkins N, Thomas M: Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society protocol: the Omron HEM-705CP, Philips HP5332, and Nissei DS-175. Blood Press Monit 1996; 1: Mazze RS, Shamoon H, Rasmantier R, et al: Reliability of blood glucose monitoring by patients with diabetes mellitus. Am J Med 1984;77:

The increasing awareness of hypertension as a serious

The increasing awareness of hypertension as a serious AJH 2002; 15:787 792 Evaluation of the Performance of a Wrist Blood Pressure Measuring Device With a Position Sensor Compared to Ambulatory 24-Hour Blood Pressure Measurements Sakir Uen, Burkard Weisser,

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

Home blood pressure measurement with oscillometric upper-arm devices

Home blood pressure measurement with oscillometric upper-arm devices REVIEW Home blood pressure measurement with oscillometric upper-arm devices R.L. Braam, Th. Thien *# Department of Internal Medicine, University Medical Centre St Radboud, PO Box 9101, 6500 HB Nijmegen,

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

Citation Acta medica Nagasakiensia. 1997, 42

Citation Acta medica Nagasakiensia. 1997, 42 NAOSITE: Nagasaki University's Ac Title Age and Gender Differences in White Author(s) Li, Zhang Ting Citation Acta medica Nagasakiensia. 1997, 42 Issue Date 1997-12-20 URL http://hdl.handle.net/10069/16086

More information

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? O R I G I N A L A R T I C L E Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? I. Aksoy, J. Deinum, J.W.M. Lenders, Th. Thien *# Department of

More information

Calculation of Trough-to-Peak Ratio in the Research Unit Setting

Calculation of Trough-to-Peak Ratio in the Research Unit Setting A]H 1996; 9:71S-75S Calculation of Trough-to-Peak Ratio in the Research Unit Setting Advantages and Disadvantages Henry L. Elliott and Peter A. Meredith The trough-to-peak ratio for the response to an

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

Blood Pressure Monitors Nissei

Blood Pressure Monitors Nissei We have made it easy for you to find a PDF Ebooks without any digging. And by having access to our ebooks online or by storing it on your computer, you have convenient answers with blood pressure monitors

More information

ORIGINAL INVESTIGATION. Is Isolated Home Hypertension as Opposed to Isolated Office Hypertension a Sign of Greater Cardiovascular Risk?

ORIGINAL INVESTIGATION. Is Isolated Home Hypertension as Opposed to Isolated Office Hypertension a Sign of Greater Cardiovascular Risk? ORIGINAL INVESTIGATION Is Isolated Home as Opposed to Isolated Office a Sign of Greater Cardiovascular Risk? Guillaume Bobrie, MD; Nathalie Genès, MD; Laurent Vaur, MD; Pierre Clerson, MD; Bernard Vaisse,

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

AJH 2000;13: by the American Journal of Hypertension, Ltd /00/$20.00

AJH 2000;13: by the American Journal of Hypertension, Ltd /00/$20.00 AJH 2000;13:632 639 Comparison of Three Blood Pressure Measurement Methods for the Evaluation of Two Antihypertensive Drugs: Feasibility, Agreement, and Reproducibility of Blood Pressure Response Stéphanie

More information

The applicability of home blood pressure measurement in clinical practice: A review of literature

The applicability of home blood pressure measurement in clinical practice: A review of literature REVIEW The applicability of home blood pressure measurement in clinical practice: A review of literature Willem J Verberk Abraham A Kroon Heidi A Jongen-Vancraybex Peter W de Leeuw University Hospital

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon (2009) 23, 794 800 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Unreliable oscillometric blood pressure measurement: prevalence, repeatability

More information

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009 42 Original article Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP, the Omron HBP T and the Pic Indolor Professional Roland

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 16 Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-71-E)

More information

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA Case 1 What should be your BP goal for an elderly (> 75 yrs of

More information

Primary hypertension in adults

Primary hypertension in adults Primary hypertension in adults NICE provided the content for this booklet which is independent of any company or product advertised Hypertension Welcome NICE published an updated guideline on the diagnosis

More information

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator MSc Thesis in Clinical Science and Technology Sara Rose Newell Department of Health Science

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

Dr Diana R Holdright. MD, FRCP, FESC, FACC, MBBS, DA, BSc. Consultant Cardiologist HYPERTENSION.

Dr Diana R Holdright. MD, FRCP, FESC, FACC, MBBS, DA, BSc. Consultant Cardiologist HYPERTENSION. Dr Diana R Holdright MD, FRCP, FESC, FACC, MBBS, DA, BSc. Consultant Cardiologist HYPERTENSION www.drholdright.co.uk Blood pressure is the pressure exerted on the walls of the arteries when the heart pumps;

More information

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE Pamela B. Morris, MD, FACC, FAHA, FASPC, FNLA Professor Medicine and Cardiovascular Diseases Director, Seinsheimer Cardiovascular Health

More information

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

!#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 !"#$ Blood Pressure Monitor Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 Introduction

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 223 Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol Paolo Verdecchia, Fabio Angeli, Roberto Gattobigio,

More information

Arm position and blood pressure: a risk factor for hypertension?

Arm position and blood pressure: a risk factor for hypertension? (3) 17, 389 395 & 3 Nature Publishing Group All rights reserved 09-9240/03 $25.00 www.nature.com/jhh ORIGINAL ARTICLE Arm position and blood pressure: a risk factor for hypertension? A Mourad, S Carney,

More information

가정혈압의활용 CARDIOVASCULAR CENTER. Wook Bum Pyun M.D., Ph.D. HOME BLOOD PRESSURE MONITORING. Ewha Womans University, school of Medicine

가정혈압의활용 CARDIOVASCULAR CENTER. Wook Bum Pyun M.D., Ph.D. HOME BLOOD PRESSURE MONITORING. Ewha Womans University, school of Medicine 가정혈압의활용 HOME BLOOD PRESSURE MONITORING CARDIOVASCULAR CENTER Wook Bum Pyun M.D., Ph.D. pwb423@ewha.ac.kr Ewha Womans University, school of Medicine Non-Invasive Blood Pressure Measurement 5-20% Resistant

More information

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION*

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Progress in Clinical Medicine 1 ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Keishi ABE** Asian Med. J. 44(2): 83 90, 2001 Abstract: J-MUBA was a large-scale clinical

More information

STATE OF THE ART BP ASSESSMENT

STATE OF THE ART BP ASSESSMENT STATE OF THE ART BP ASSESSMENT PROFESSOR MOLECULAR PHARMACOLOGY CONWAY INSTITUE UNIVERSITY COLLEGE DUBLIN CO-CHAIRMAN BLOOD PRESCSURE MANAGEMENT IN LOW RESOURCE SETTINGS CENTRE FOR INTERNATIONAL HUMANITARIAN

More information

DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI.

DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI. DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI. Franco Cipollini, Carlo Porta, Enrica Arcangeli, Carla Breschi, & Giuseppe Seghieri Azienda USL 3, Ambulatorio

More information

Home Blood Pressure Log

Home Blood Pressure Log Home Blood Pressure Log 2 You or your health care professional may want to monitor your blood pressure at home twice a day for a week or more before appointments to accurately confirm your average blood

More information

Dr Doris M. W Kinuthia

Dr Doris M. W Kinuthia Dr Doris M. W Kinuthia Objectives Normal blood pressures in children Measurement of blood pressure in children Aetiology of Hypertension in children Evaluation of children with hypertension Treatment of

More information

Concentration-Effect Relationships and Implications for Trough-to-Peak Ratio

Concentration-Effect Relationships and Implications for Trough-to-Peak Ratio AJH 1996;9:66S-70S Concentration-Effect Relationships and Implications for Trough-to-Peak Ratio Peter A. Meredith and Henry L. Elliott The guidelines on trough-to-peak ratio identified an index of the

More information

Many interventional trials have demonstrated

Many interventional trials have demonstrated AJH 1998;11:165 173 Short-Term Effects of Withdrawing Angiotensin Converting Enzyme Inhibitor Therapy on Home Self-Measured Blood Pressure in Hypertensive Patients Laurent Vaur, Guillaume Bobrie, Claire

More information

Accurate blood pressure (BP) measurement is essential

Accurate blood pressure (BP) measurement is essential AJH 2006; 19:468 474 BP Measurement A Comparison of Home Measurement and Ambulatory Monitoring of Blood Pressure in the Adjustment of Antihypertensive Treatment Teemu J. Niiranen, Ilkka M. Kantola, Risto

More information

Adherence to Hypertension Visit Protocol Description: % of patients with HTN who are being monitored through follow up visits according to guidelines, based on "days since last BP", which looks back from

More information

BP.06). 1.8% 11.7% ( P

BP.06). 1.8% 11.7% ( P AJH 2000;13:184 190 Electronic Pill-Boxes in the Evaluation of Antihypertensive Treatment Compliance: Comparison of Once Daily Versus Twice Daily Regimen Michel Andrejak, Nathalie Genes, Laurent Vaur,

More information

Non-invasive continuous blood pressure monitoring based on radial artery tonometry (T-Line TL-200pro device) in the intensive care unit

Non-invasive continuous blood pressure monitoring based on radial artery tonometry (T-Line TL-200pro device) in the intensive care unit Non-invasive continuous blood pressure monitoring based on radial artery tonometry (T-Line TL-200pro device) in the intensive care unit Bernd Saugel; Agnes S Meidert; Alexander Hapfelmeier; Florian Eyer;

More information

Hypertensive patients commonly use home. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors Steven A. Yarows and Robert D.

Hypertensive patients commonly use home. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors Steven A. Yarows and Robert D. AJH 2000;13:276 282 Measurement Variation Among 12 Electronic Home Blood Pressure Monitors Steven A. Yarows and Robert D. Brook Although there are AAMI and BHS standards for accuracy of electronic home

More information

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER Naser KA. Teaching Hospital Peradeniya, Peradeniya, Sri Lanka Zawahir S

More information

Strategies for Managing Hypertension

Strategies for Managing Hypertension Strategies for Managing Hypertension by Marsha K. Millonig, MBA, RPh President,Catalyst Enterprises, LLC E.L.F. Publications, Inc. is accredited by the Accreditation Council for Pharmacy Education as a

More information

Home monitoring of blood glucose

Home monitoring of blood glucose Home Monitoring of Glucose and Blood Pressure JAMES R. TAYLOR, PharmD, CDE, and KENDALL M. CAMPBELL, MD, University of Florida, Gainesville, Florida Home monitoring of blood glucose and blood pressure

More information

Accuracy and reliability of wrist-cuff devices for selfmeasurement

Accuracy and reliability of wrist-cuff devices for selfmeasurement Original article 629 Accuracy and reliability of wrist-cuff devices for selfmeasurement of blood pressure Masahiro Kikuya a, Kenichi Chonan a,yutaka Imai a, Eiji Goto b and Masao Ishii c, on behalf of

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials. AJH 1995; 8:311-315 BRIEF COMMUNICATIONS Lack of Placebo Effect on Ambulatory Blood Pressure Giuseppe Mancia, Stefano Omboni, Gianfranco Parati, Antonella Ravogli, Alessandra Villani, and Alberto Zanchetti

More information

Egyptian Hypertension Guidelines

Egyptian Hypertension Guidelines Egyptian Hypertension Guidelines 2014 Egyptian Hypertension Guidelines Dalia R. ElRemissy, MD Lecturer of Cardiovascular Medicine Cairo University Why Egyptian Guidelines? Guidelines developed for rich

More information

Use and Interpretation of Home Blood Pressure Monitoring

Use and Interpretation of Home Blood Pressure Monitoring Use and Interpretation of Home Blood Pressure Monitoring BLOOD PRESSURE in year 2015 PARAMETERS TODAY Gianfranco Parati University of Milano-Bicocca & Ospedale S.Luca, IRCCS, Istituto Auxologico Italiano

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the Omron M6 Comfort IT (HEM-7322U-E) with the Omron M6 Comfort (HEM-7321-E) Devices Omron M6 Comfort IT (HEM-7322U-E) Omron M6 Comfort (HEM-7321-E) Pictures Display Validation ESH 2010 Device

More information

High-dose monotherapy vs low-dose combination therapy of calcium channel blockers and angiotensin receptor blockers in mild to moderate hypertension

High-dose monotherapy vs low-dose combination therapy of calcium channel blockers and angiotensin receptor blockers in mild to moderate hypertension (2005) 19, 491 496 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE High-dose monotherapy vs low-dose combination therapy of calcium channel blockers

More information

Does Antihypertensive Drug Class Affect Day-to-Day Variability of Self-Measured Home Blood Pressure? The HOMED-BP Study

Does Antihypertensive Drug Class Affect Day-to-Day Variability of Self-Measured Home Blood Pressure? The HOMED-BP Study Does Antihypertensive Drug Class Affect Day-to-Day Variability of Self-Measured Home Blood Pressure? The HOMED-BP Study Kei Asayama, MD, PhD; Takayoshi Ohkubo, MD, PhD; Tomohiro Hanazawa, MS; Daisuke Watabe,

More information

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License Zheng DC, Liu CY, Amoore J, Mieke S, Murray A. Need for Re-validation of Automated Blood Pressure Devices for use in Unstable Conditions. In: 215 Computing in Cardiology Conference (CinC). 215, Nice, France:

More information

HIGH BLOOD PRESSURE. What is malignant or accelerated hypertension?

HIGH BLOOD PRESSURE. What is malignant or accelerated hypertension? HIGH BLOOD PRESSURE Is high blood pressure important? What is blood pressure? What is high blood pressure? What are the causes of high blood pressure? What are the symptoms of high blood pressure? What

More information

Nomogram of the Relation of Brachial-Ankle Pulse Wave Velocity with Blood Pressure

Nomogram of the Relation of Brachial-Ankle Pulse Wave Velocity with Blood Pressure 801 Original Article Nomogram of the Relation of Brachial-Ankle Pulse Wave Velocity with Blood Pressure Akira YAMASHINA, Hirofumi TOMIYAMA, Tomio ARAI, Yutaka KOJI, Minoru YAMBE, Hiroaki MOTOBE, Zydem

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the SAFETY PRONTEX INTEGRA (Nissei DS-B33-03) with the Nissei DSK-1011 Devices Item 9 SAFETY PRONTEX INTEGRA (Nissei DS-B33-03) Nissei DSK-1011 Pictures Display Image Validation ESH IP2010

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice (2005) 19, 801 807 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Prognostic significance of blood pressure measured in the office, at home and

More information

White coat and masked hypertension

White coat and masked hypertension White coat and masked hypertension Conflict of interest Support from Spacelabs, Microlife. Honoraria from Novartis, Elpen, Boeringer-Ingelheim, CANA, Lilly, MSD, Sanofi, Menarini, Ciezi, Astra-Zeneca.

More information

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA)

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) THE NEW ARMENIAN MEDICAL JOURNAL Vol.5 (2011), Nо 2, p.29-34 DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) Zelveian P.H. 1, 2, Podosyan G.A. 2 1 Institute

More information

Verapamil SR and trandolapril combination therapy in hypertension a clinical trial of factorial design

Verapamil SR and trandolapril combination therapy in hypertension a clinical trial of factorial design Br J Clin Pharmacol 1998; 45: 491 495 Verapamil SR and trandolapril combination therapy in hypertension a clinical trial of factorial design Juergen Scholze, 1 Peter Zilles 2 & Daniele Compagnone 2 on

More information

Antihypertensive efficacy of olmesartan compared with other antihypertensive drugs

Antihypertensive efficacy of olmesartan compared with other antihypertensive drugs (2002) 16 (Suppl 2), S24 S28 2002 Nature Publishing Group All rights reserved 0950-9240/02 $25.00 www.nature.com/jhh compared with other antihypertensive drugs University Clinic Bonn, Department of Internal

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

RESEARCH INTRODUCTION

RESEARCH INTRODUCTION Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial Martin G Myers, professor of medicine, 1 Marshall

More information

In clinical practice, blood pressure (BP) treatment

In clinical practice, blood pressure (BP) treatment Original Paper Does Hidden Undercuffing Occur Among Obese Patients? Effect of Arm Sizes and Other Predictors of the Difference Between Wrist and Upper Arm Blood Pressures Hardik Doshi, MD; Alan B. Weder,

More information

Causes of Poor BP control Rates

Causes of Poor BP control Rates Goals Of Hypertension Management in Clinical Practice World Hypertension League (WHL) Meeting Adel E. Berbari, MD, FAHA, FACP Professor of Medicine and Physiology Head, Division of Hypertension and Vascular

More information

Candesartan Antihypertensive Survival Evaluation in Japan (CASE-J) Trial of Cardiovascular Events in High-Risk Hypertensive Patients

Candesartan Antihypertensive Survival Evaluation in Japan (CASE-J) Trial of Cardiovascular Events in High-Risk Hypertensive Patients 1/5 This site became the new ClinicalTrials.gov on June 19th. Learn more. We will be updating this site in phases. This allows us to move faster and to deliver better services. Show less IMPORTANT: Listing

More information

The hypertensive effects of the renin-angiotensin

The hypertensive effects of the renin-angiotensin Comparison of Telmisartan vs. Valsartan in the Treatment of Mild to Moderate Hypertension Using Ambulatory Blood Pressure Monitoring George Bakris, MD A prospective, randomized, open-label, blinded end-point

More information

Phase 3 investigation of aprocitentan for resistant hypertension management. Investor Webcast June 2018

Phase 3 investigation of aprocitentan for resistant hypertension management. Investor Webcast June 2018 Phase 3 investigation of aprocitentan for resistant hypertension management Investor Webcast June 2018 The following information contains certain forward-looking statements, relating to the company s business,

More information

Hypertension represents one of the most

Hypertension represents one of the most AJH 1998;11:1376 1380 Changes in Blood Pressure and Heart Rate Variability During Dental Surgery Kiyoshi Matsumura, Keiko Miura, Yutaka Takata, Hideo Kurokawa, Minoru Kajiyama, Isao Abe, and Masatoshi

More information

Sponsor Novartis. Generic Drug Name. Valsartan and amlodipine Trial Indication(s) Hypertension Protocol Number CVAA489A2306 Protocol Title

Sponsor Novartis. Generic Drug Name. Valsartan and amlodipine Trial Indication(s) Hypertension Protocol Number CVAA489A2306 Protocol Title Sponsor Novartis Generic Drug Name Valsartan and amlodipine Trial Indication(s) Hypertension Protocol Number CVAA489A2306 Protocol Title A randomized, double-blind, multi-center, active-controlled, parallel

More information

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015 Original article 1 Clinical practice of ambulatory versus home blood pressure monitoring in hypertensive patients Jorge A. Paolasso, Florencia Crespo, Viviana Arias, Eduardo A. Moreyra, Ariel Volmaro,

More information

Validation Study Result Form

Validation Study Result Form Validation Study Result Form Please complete Section 1 to Section 3 of this form and return it to dabl Educational Trust with copies of the validation plots. The requirements for each of these sections

More information

ED: raji Op: hashmath sc: micheal : lww_mbp_

ED: raji Op: hashmath sc: micheal : lww_mbp_ ED: raji Op: hashmath sc: micheal : lww_mbp_20021 Original scientific paper 1 Home-based self-measurement of blood pressure: a proposal using new reference values (the PURAS study) Juan A. Divisón-Garrote,

More information

Assessment of techniques for measurement of blood pressure in infants and children

Assessment of techniques for measurement of blood pressure in infants and children Archives of Disease in Childhood, 1973, 48, 932. Assessment of techniques for measurement of blood pressure in infants and children A. M. ELSEED, E. A. SHINEBOURNE, and M. C. JOSEPH From the Paediatric

More information

Effects of Calcium Antagonists on Insulin Release

Effects of Calcium Antagonists on Insulin Release Effects of Calcium Antagonists on Insulin Release Prof. Mashori Ghulam Rasool, Pakistan Introduction Cardiovascular diseases are the main cause of death in many countries and pose health problems not only

More information

NO causes vasodilation by stimulating vascular smooth

NO causes vasodilation by stimulating vascular smooth Clinical Treatment Clinical Potential of Combined Organic Nitrate and Phosphodiesterase Type 5 Inhibitor in Treatment-Resistant Hypertension James J. Oliver, James W. Dear, David J. Webb Abstract NO donor

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL Douglas G. Kelling Jr., MD Carmella Gismondi-Eagan, MD, FACP George C. Monroe, III, MD Revised, April 8, 2012 The information contained in this protocol

More information

A Placebo-Controlled, Forced Titration Study. Yves Lacourcière and Roland Asmar for the Candesartan/Losartan study investigators

A Placebo-Controlled, Forced Titration Study. Yves Lacourcière and Roland Asmar for the Candesartan/Losartan study investigators AJH 1999;12:1181 1187 A Comparison of the Efficacy and Duration of Action of Candesartan Cilexetil and Losartan as Assessed by Clinic and Ambulatory Blood Pressure After a Missed Dose, in Truly Hypertensive

More information

5.2 Key priorities for implementation

5.2 Key priorities for implementation 5.2 Key priorities for implementation From the full set of recommendations, the GDG selected ten key priorities for implementation. The criteria used for selecting these recommendations are listed in detail

More information

Managing hypertension: a question of STRATHE

Managing hypertension: a question of STRATHE (2005) 19, S3 S7 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Managing hypertension: a question of STRATHE Department of Cardiovascular Disease,

More information

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation (2009), 1 5 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (  The story so far. Medical Electronics Dr Neil Townsend Michaelmas Term 2001 (wwwrobotsoxacuk/~neil/teaching/lectures/med_elec) Blood Pressure has been measured (in some way) for centuries However, clinical interpretation

More information

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home.

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home. Welch Allyn ABPM 6100S Comprehensive blood pressure monitoring that brings patient comfort home. Features and Benefits of the Welch Allyn ABPM 6100S Pictorial application instructions, artery markers and

More information

Slide notes: References:

Slide notes: References: 1 2 3 Cut-off values for the definition of hypertension are systolic blood pressure (SBP) 135 and/or diastolic blood pressure (DBP) 85 mmhg for home blood pressure monitoring (HBPM) and daytime ambulatory

More information

How well do office and exercise blood pressures predict sustained hypertension? A Dundee Step Test Study

How well do office and exercise blood pressures predict sustained hypertension? A Dundee Step Test Study (2000) 14, 429 433 2000 Macmillan Publishers Ltd All rights reserved 0950-9240/00 $15.00 www.nature.com/jhh ORIGINAL ARTICLE How well do office and exercise blood pressures predict sustained hypertension?

More information

Trough to peak ratio: current status and applicability

Trough to peak ratio: current status and applicability Journal of Human Hypertension (1998) 12, 55 59 1998 Stockton Press. All rights reserved 0950-9240/98 $12.00 REVIEW ARTICLE Trough to peak ratio: current status and applicability Department of Medicine

More information

Younger adults with a family history of premature artherosclerotic disease should have their cardiovascular risk factors measured.

Younger adults with a family history of premature artherosclerotic disease should have their cardiovascular risk factors measured. Appendix 2A - Guidance on Management of Hypertension Measurement of blood pressure All adults from 40 years should have blood pressure measured as part of opportunistic cardiovascular risk assessment.

More information

Appendix This appendix was part of the submitted manuscript and has been peer reviewed. It is posted as supplied by the authors.

Appendix This appendix was part of the submitted manuscript and has been peer reviewed. It is posted as supplied by the authors. Appendix This appendix was part of the submitted manuscript and has been peer reviewed. It is posted as supplied by the authors. Appendix to: Banks E, Crouch SR, Korda RJ, et al. Absolute risk of cardiovascular

More information

The introduction of oscillometric automated blood pressure

The introduction of oscillometric automated blood pressure Atrial Fibrillation Impact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring Nikolaos Pagonas,* Sven Schmidt,* Jörg Eysel, Friederike Compton, Clemens Hoffmann, Felix Seibert,

More information

The accurate measurement of blood pressure

The accurate measurement of blood pressure Position Paper ASH Position Paper: Home and Ambulatory Blood Pressure Monitoring When and How to Use Self (Home) and Ambulatory Blood Pressure Monitoring Thomas G. Pickering, MD, D Phil; 1 William B. White,

More information

Clinical Recommendations: Patients with Periodontitis

Clinical Recommendations: Patients with Periodontitis The American Journal of Cardiology and Journal of Periodontology Editors' Consensus: Periodontitis and Atherosclerotic Cardiovascular Disease. Friedewald VE, Kornman KS, Beck JD, et al. J Periodontol 2009;

More information