Predictability and accuracy of IOL formulas in high myopia
|
|
- Alexandra Moore
- 6 years ago
- Views:
Transcription
1 ARTICLE Predictability and accuracy of IOL formulas in high myopia Mohamed Yasser Sayed Saif, MD 1 ; Mohamed Othman Abdel Khalek, MD 1 ; Ahmed Tamer Sayed Saif, MD 2 ; Passant Sayed Saif, MD 3 ; Sherif Kamel Safina, MSc 4 PURPOSE: To study the predictability and accuracy of different intraocular lens calculation formulas in high myopia patients with axial length (AL) 26. mm or spherical equivalent (SE) refraction of more than 6 diopters, undergoing phacoemulsification. SETTING: Multicenter study in Beni Suef University hospital, Fayoum university hospital and MUST hospital in Egypt. METHODS: This prospective comparative study comprised 21 eyes of cataract patients divided into group A: AL less than 26 mm; and group B: AL more than 26 mm. All eyes had SE refraction of more than 6 diopters. A-scan ultrasound biometry and autokeratometry were performed. The IOL power was calculated using the Haigis, SRK-T, Holladay, Hoffer-Q, Binkhorst and SRK-II formulas. Differences between achieved postoperative refraction and target refraction was calculated using different formulas. P value <. was considered statistically significant. RESULTS: Mean age was 4. ± 8. years and 4. ± 11. years for group A and B, respectively. Mean AL was ± 2.42 mm. The average k reading was ± 1.82 and ± 2.74; pre-se was 9.78 ± 3.24 D and 12. ± 4.21 D (from. to D); mean were Haigis.46 ±.8 D and +.12 ±.73 D, Holladay.34 ±.83 D and.66 ±.77 D, SRK/T.23 ±.6 D and.24 ±.8 D, Hoffer-Q.1 ±.76 D and.69 ±.74 D, Binkhorst.7 ±.78 D and.69 ±.73, and SRK-II.4 ±.74 D and.46 ± 1.6 D. CONCLUSION: The Haigis formula showed the best result in myopic eyes with axial length of more than 26 mm, while SRK-II was best in axial lengths of less than 26 mm. J Emmetropia 16; 1: Pathologic myopia is classified as a high, progressive myopic refractive error that generally presents in early in childhood. In pathologic myopia, the spherical equivalent (SE) is more than 6. diopters, or axial length (AL) more than 26. mm 1. Patients with high axial myopia frequently present lens opacities and develop cataracts at an earlier age than emmetropes 2. Intra-Ocular Lens (IOL) power calculation in high myopes is challenging for cataract surgeons 3. Submitted: 9//1 Revised: 12/3/1 Accepted: 2/17/16 1 Beni-Suef University, Beni-Suef, Egypt. 2 Fayoum University, Fayoum, Egypt. 3 Misr University for Science and Technology, October, Egypt. 4 Embaba Hospital, Giza, Egypt. Financial Disclosure: The authors have no commercial or proprietary interest in the products mentioned herein. Corresponding Author: Mohamed Yasser Sayed Saif Beni-Suef University Salah Salem st, Beni Suef, Egypt ysaif@med.bsu.edu.eg The two main problems encountered are correct measurement of the AL to be used in calculations, due to posterior staphyloma, and lack of accuracy of available intraocular lens calculation formulas for cases of high myopia 4,. The aim of this study is compare and evaluate the accuracy of different intraocular lens calculation formulas in patients with high myopia. METHODS This was a prospective comparative study approved by the board of the Beni Suef ophthalmology and the research ethics committee. The study adhered to the tenets of the Declaration of Helsinki. Patients This study comprised 21 eyes of patients scheduled to undergo phacoemulsification in the Beni Suef University hospital, Fayoum university hospital and MUST hospital who presented with cataracts, AL 26 mm or refraction SE of over 6 diopters. 16 SECOIR Sociedad Española de Cirugía Ocular Implanto-Refractiva ISSN:
2 18 ASSESSMENT OF IOL FORMULAS IN HIGH MYOPIA Patients with refraction SE less than 6 D or AL less than 26 mm, history of previous ocular surgery, posterior segment disorders, eventful cataract surgery, keratoconus, endothelial dystrophy, and glaucoma were excluded. Procedures Preoperative examination included best corrected visual acuity, slit lamp biomicroscopy, indirect ophthalmoscopy, and Goldman applanation tonometry. In all cases, AL was measured using A-scan ultrasound biometry (E-Z SCAN AB Ophthalmic ultrasound scanner; Sonomed Escalon, Lake Success, NY, USA). Corneal power (k) was measured using autokeratometry (Topcon autokeratometer KR 89; Topcon Medical Systems, Inc. Paramus, NJ, USA). Preoperative IOL power calculations were performed using Haigis, SRK-T, Holladay, Hoffer-Q, Binkhorst and SRK-II formulas in all patients. The IOL formula that provided a lens power with the foregoing postoperative refraction was selected. Phacoemulsification was performed and AcrySof foldable IOL (Alcon, Fort Worth, TX, USA) was implanted within the capsular bag. Patients were followed up for 6 weeks postoperatively at the following intervals: 1 week, 3 weeks, and 6 weeks. At each visit, the patient underwent a complete examination of the anterior segment using slit lamp, cycloplegic refraction with retinoscopy, and auto refractometry. The following calculations were used: spherical equivalent (SE) = sphere + half the cylinder; error formula = post-operative SE predicted formula error. Statistical analysis At the end of this study, data were statistically described in terms of mean ± standard deviation (SD), median, minimum, maximum, correlation and percentages. Anova was used to compare numerical variables between the study groups. All statistical calculations were done using IBM SPSS Statistics 21 (Statistical Package for the Social Science). P-values of less than. were considered significant. RESULTS In total, 21 eyes of myopic patients with a mean age of ±.1711 years were included, 12 from men (8.14%) and 9 from women (41.86%). Patients were divided into 2 groups: A) axial length less than 26. mm; B) axial length more than 26.1 mm. The mean age was 4. ± 8. years for group A and 4. ± 11. years for group B, with no statistical significance between both groups. The age range was from 32 to 7 years; there were 1 (6.98%) women in group A and 7 (34.88%) in group B, as shown in Table 1. Table 1. Pre and postoperative results. Group A n = Group B n = 16 Group Mean Median SD Maximum Minimum Age (years) A B Pre-VA A..... B Post-VA A B Pre-IOP A B Post-IOP A B Average K A B Lens thickness A B ACD A B Vitreous length A B Power A B Pre-SE A B Post-SE A B ACD, anterior chamber depth; IOP, intraocular pressure; SD, standard deviation; SE, spherical equivalent; VA, visual acuity.
3 ASSESSMENT OF IOL FORMULAS IN HIGH MYOPIA 19 The mean preoperative VA in both group was.2 ±.1, which improved to.7 in group A and.8 in group B, as shown in Table 1. The mean preoperative IOP was 17. ± 3. mmhg and 16. ± 2. mmhg in both groups, which decreased to 14. ± 3. mmhg and 14. ± 2. mmhg, respectively, as shown in Table 1. The mean axial length was ± 2. mm, with a range of mm to 33.2 mm. The mean K reading was ± 1.82 D and ± 2.74 D in groups A and B, with a range of 34.7 D to D. The SE refraction was 9.78 ± 3.24 D for group A, and 12. ± 4.21 D for group B, with a range of. to D. The postoperative mean SE was.98 and.4, respectively, with a range of +3. to 4.7, as shown in Table 1. The error of each formula was calculated as the difference between the target or desired postoperative refraction and the actual or achieved refraction. The statistical differences between different formulas is shown in Table 2. As regards the mean error of deviation, Hoffer-Q formula in group A is the best, while the Haigis formula is the best in group B, as shown in Table 2. There was no statistically significant difference between both groups as regards the Haigis formula, while the SRK-T, Holladay, Hoffer-Q, Binkhorst and SRK-II formulas showed a statistically significance between both groups, as shown in Tables 3 and 4. Table 2. Statistical analysis of in the formulas. Group A n = Group B n = 16 Group Mean Median SD Maximum Minimum Haigis A B SRK T A B SRKII A B Holladay A B Hoffer Q A B Binkhorst A B SD, standard deviation. Table 3. Analysis of differences between groups for all formulas. Mean SD Maximum Minimum P value Haigis SRK-T * SRK-II * Holladay * Hoffer Q * Binkhorst * SD, standard deviation; *Significant differences. Table 4. Results of comparison between both groups for all formulas using other statistical tests. Haigis SRK-T SRK-II Holladay Hoffer Q Binkhorst Mann-Whitney U Wilcoxon W Z Asymp. Sig. (2-tailed).683
4 ASSESSMENT OF IOL FORMULAS IN HIGH MYOPIA > < > < Haigis SRK-T SRK-II Holladay Figure 1. The postoperative error in group A with axial length less than 26 mm. (y = number of eyes). Haigis SRK-T SRK-II Holladay Figure 2. The postoperative error in group B with axial length more than 26 mm (y = number of eyes). In Group A: SRK-II showed the best result, followed by the Holladay, SRK-T and Hoffer-Q, and Haigis and Binkhorst formulas. The Haigis formula showed more tendancy toward myopic shift, as shown in Figure 1. For Group B: the Haigis formula showed the most accurate results, followed by the SRK-T, Holladay, Binkhorst and Hoffer-Q formulas. The SRK-II showed myopic shift and the poorest results in this group, as shown in Figure 2. DISCUSSION This study is a prospective comparative analysis of 21 eyes presenting with cataracts, with AL 26 mm or SE of more than 6 diopters. The mean postoperative refractive SE when implanting a plus power IOLs was.3 ±.1 D (p <.1). A Pentacam examination was performed in patients with a K reading above 46, revealing no keratoconus or suspected keratoconus. Patients with short axial lens showed an SE refraction ranging from 6.63 to D. However, the Haigis formula showed the least deviation in group B (axial length of over 26 mm), while SRK-T showed the least deviation in group A (axial length of less than 26 mm). In, Raouf et al. 6 studied 3 eyes to compare 3 IOL calculation formulas in high myopia, and found no statically significant difference between them. In this study, the SRK/T formula caused the lowest mean error (.17D), and the Haigis formula showed a higher mean error (.21D), with post-operative hyperopic shift. The foregoing authors found the SRK/T formula to be the most accurate in high myopia compared with Haigis, and that the Haigis formula tended to cause a hyperopia shift. Our study, meanwhile, shows the Haigis formula to be the most accurate, with minimal post-operative myopia shift. These contrasting results may be due to the grouping criteria: Raouf et al. divides eyes into 3 subgroups according to AL, with fewer cases and a difference in A-scan biometry and keratometry. Saif et al. in 7, analyzed 3 formulas, SRK/T, SRK-II and Holliday s, in IOL calculations, and found the SRK/T formula to be better in high myopia. They did not evaluate the Haigis, Hoffer-Q or Binkhorst formulas, and the results were based on spherical error. Jia Wang et al. 8, studied eyes divided into 3 groups according to AL (Group 1 = 33 eyes, Group 2 = 92 eyes, and Group 3 = 7 eyes). The Haigis had a significantly lower mean error in all patients compared with the other formulas (<.). In Groups 1 to 3, the mean error calculated by the Haigis was either significantly lower or comparable to that calculated by the other formulas. Compared with other formulas using IOL Master biometric data, the Haigis formula yielded superior refractive results in eyes with various AL. This is consistent with our findings, in which the Haigis formula was the most accurate in higher axial length, and the least accurate in short axial length. Ganesh et al. in 4 9, compared the accuracy of IOL power calculation in high myopia using the Haigis and SRK II formulas, and found the Haigis formula to be very effective in eyes with AL ranging from 2 to 32 mm. However, a comparison of the SRK II and Haigis formulas did not clarify the accuracy of the Haigis formula, as all recent studies have shown that the SRK II formula does not accurately predict IOL power in high axial myopia patients. Roessler et al. showed the Haigis to be superior to the Holladay 1 and SRK/T formulas in predicting postoperative refractive outcomes in 37 eyes with AL > 26. mm. Terzi et al. 11 studied extremely high myopic eyes and extremely high hyperopic eyes, finding that the Haigis performed better than the Holladay 2, Hoffer Q and SRK/T formulas in 44 eyes with AL > 26 mm undergoing myopic refractive lens exchange. Bang et al. 12 concluded that the Haigis formula was the most accurate in predicting postoperative refractive error compared with the Holladay 1, Holladay 2, Hoffer Q and SRK-T formulas in 3 eyes with AL > 27 mm.
5 ASSESSMENT OF IOL FORMULAS IN HIGH MYOPIA 21 In conclusion, the Haigis formula showed the best result in high myopic axial cataractus eyes with minimal shift to myopia, while the SRK-II formula show the best results for short axial length. REFERENCES 1. Friedman NJ, Kaiser PK. Essentials of Ophthalmology. Philadelphia, PA: Elsevier Inc; 7: Kaufman BJ, Sugar J. Discrete nuclear sclerosis in young patients with myopia. Arch Ophthalmol. 1996; 114: Olsen T, Corydon L, Gimbel H. Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg. 199; 21: Hafez A. Accuracy of A and B modes in myopic biometry. Bull Ophthalmol Soc Egypt, 4; 97: Ghanem AA, El-Sayed HM. Accuracy of intraocular lens power calculation in high myopia. Oman Journal of Ophthalmology ; 3: El-Nafees R, Moawad A, Kishk H, Gaafar W. Intra-ocular lens power calculation in patients with high axial myopia before cataract surgery. Saudi J Ophthalmol. ; 24: Saif SEH, Ahmed MH, Amer SK, Saif MYS. Comparison between IOL Formulae. Bull Ophthalmol Soc Egypt. ; 98: Wang JK, Shang SW. Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths. Int J Ophthalmol, 13; 6: Eom Y, Kang S, Song JS, Kim YY, Kim HM. Comparison of Hoffer Q and Haigis Formulae for Intraocular Lens Power Calculation According to the Anterior Chamber Depth in Short Eyes. American Journal of Ophthalmology. 14; 17: e2.. Roessler GF, Dietlein TS, Plange N, et al. Accuracy of intraocular lens power calculation using partial coherence interferometry in patients with high myopia. Ophthalmic Physiol Opt. 12; 32: Terzi E, Wang L, Kohnen T. Accuracy of modern intraocular lens power calculation formulas in refractive lens exchange for high myopia and high hyperopia. J Cataract Refract Surg. 9; 3: Bang S, Edell E, Yu Q, Pratzer K, Stark W. Accuracy of intraocular lens calculations using the IOL Master in eyes with long axial length and a comparison of various formulas. Ophthalmology. 11; 118:3-6. First author: Mohamed Yasser Sayed Saif, MD Beni-Suef University. Beni Suef, Egypt.
Trabeculectomy is an effective method for lowering
ORIGINAL STUDY Refractive Outcome of Cataract Surgery in Eyes With Prior Trabeculectomy: Risk Factors for Postoperative Myopia Oliver L. Yeh, MD, Karine D. Bojikian, MD, Mark A. Slabaugh, MD, and Philip
More informationComparison of the Effects of Bright Light, Phenylephrine 2.5%, Tropicamide 1%, and Pilocarpine 2% on Anterior Chamber Depth and IOL Power
Comparison of the Effects of Bright Light, Phenylephrine 2.5%, Tropicamide 1%, and Pilocarpine 2% on Anterior Chamber Depth and IOL Power Mona Sane MD, Vincent Imbrogno DO, Hoon Jung MD The authors have
More informationIOL Power Calculation for Children
1 IOL Power Calculation for Children Rupal H. Trivedi, MD MSCR M. Edward Wilson, MD The authors have no financial interest in the subject matter of this presentation. Intraocular lens (IOL) implantation
More informationAccuracy of Intraocular Lens Calculation Formulas
Accuracy of Intraocular Lens Calculation Formulas Ronald B. Melles, MD, 1 Jack T. Holladay, MD, MSEE, 2 William J. Chang, MD 1 Purpose: To compare the accuracy of intraocular lens (IOL) calculation formulas
More informationAccuracy of Biometry for Intraocular Lens Implantation Using the New Partial Coherence Interferometer, AL-scan
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2014;28(6):444-450 http://dx.doi.org/10.3341/kjo.2014.28.6.444 Original Article Accuracy of Biometry for Intraocular Lens Implantation Using the New
More informationComparative Efficacy of the New Optical Biometer on Intraocular Lens Power Calculation (AL-Scan versus IOLMaster)
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(3):241-248 https://doi.org/10.3341/kjo.2017.0063 Original Article Comparative Efficacy of the New Optical Biometer on Intraocular Lens Power
More informationPerformance of the SRK/T formula using A-Scan ultrasound biometry after phacoemulsification in eyes with short and long axial lengths
Karabela et al. BMC Ophthalmology (2016) 16:96 DOI 10.1186/s12886-016-0271-8 RESEARCH ARTICLE Open Access Performance of the SRK/T formula using A-Scan ultrasound biometry after phacoemulsification in
More informationMyopic Shift After Intraocular Lens Implantation in Children Less Than Two Years of Age
Original Article Myopic Shift After Intraocular Lens Implantation in Children Less Than Two Years of Age Suma Ganesh 1, Reena Gupta 2, Sumita Sethi 3, Chandra Gurung 4, Raman Mehta 5 1,5 Dr. Shroff s Charitable
More informationComparison of Intraocular Lens Power Calculation Methods Following Myopic Laser Refractive Surgery: New Options Using a Rotating Scheimpflug Camera
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(6):497-505 https://doi.org/10.3341/kjo.2018.0008 Original Article Comparison of Intraocular Lens Power Calculation Methods Following Myopic
More informationLearn Connect Succeed. JCAHPO Regional Meetings 2017
Learn Connect Succeed JCAHPO Regional Meetings 2017 Faculty Biometry and IOL Calculations ASCRS and ASOA Symposium and Congress Los Angeles, CA Daniel H. Chang, M.D. - Empire Eye and Laser Center Bakersfield,
More informationRefractive Changes after Removal of Anterior IOLs in Temporary Piggyback IOL Implantation for Congenital Cataracts
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2013;27(2):93-97 http://dx.doi.org/10.3341/kjo.2013.27.2.93 Original Article Refractive Changes after Removal of Anterior IOLs in Temporary Piggyback
More informationInaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure Glaucoma
Original Article DOI 10.3349/ymj.2009.50.2.206 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 50(2):206-210, 2009 Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure
More informationIntraoperative biometry for intraocular lens (IOL) power calculation at silicone oil removal
European Journal of Ophthalmology / Vol. 13 no. 7, 2003 / pp. 622-626 Intraoperative biometry for intraocular lens (IOL) power calculation at silicone oil removal S.M. EL-BAHA, A. EI-SAMADONI, H.F. IDRIS,
More informationPrediction Error After Lens Implantation in Children With Axial Length Less than 22 mm Below 2 Yrs
E-ISSN 2454-2784 Original Article Prediction Error After Lens Implantation in Children With Axial Length Less than 22 mm Below 2 Yrs Reena Gupta 1, Suma Ganesh 2, Chekitaan Singh 3, A.K Khurana 1 1 Regional
More informationCataract Surgery in the Patient with a History of LASIK or PRK
Cataract Surgery in the Patient with a History of LASIK or PRK #56996-RS April 2018 Sebastian Lesniak, MD Matossian Eye Associates None Disclosures Bio Matossian Eye Associates, Hopewell NJ, 7/2015 Present
More informationComparison of Intraocular Lens Power Calculation Using the Binkhorst and SRK Formulae: A Clinical Study
Comparison of Intraocular Lens Power Calculation Using the Binkhorst and SRK Formulae: A Clinical Study Pages with reference to book, From 38 To 40 Manzoor Hussain, Jehangir Durrani ( Department of Ophthalmology,
More informationEfficacy and Safety of Cataract Extraction with Negative Power Intraocular Lens Implantation
The Open Ophthalmology Journal, 2008, 2, 15-19 15 Efficacy and Safety of Cataract Extraction with Negative Power Intraocular ens Implantation Michael A. Kapamajian 1 and Kevin M. Miller *,2,# 1 From the
More informationImmersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification
Original Article Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification with Foldable IOL Irum Abbas, Atif Mansoor Ahmad, Tahir Mahmood Pak J Ophthalmol 2009, Vol. 25 No.
More informationPrediction accuracy of intraocular lens power calculation methods after laser refractive surgery
Wu et al. BMC Ophthalmology (2017) 17:44 DOI 10.1186/s12886-017-0439-x RESEARCH ARTICLE Open Access Prediction accuracy of intraocular lens power calculation methods after laser refractive surgery Yubo
More informationComparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification
Received: 5.5.2007 Accepted: 22.1.2008 Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification Heidarali Moeini*, Fateme Eslami**, Akram Rismanchian*,
More informationNature and Science 2016;14(9)
Comparison between Implantable Contact Lens (ICL) versus Acry Sof Cachet Phakic Intraocular Lenses in Correction of Moderate to High Myopia Khaled Nada, M. Sc.; Mohammed Attia, M. D and Ashraf El Habbak,
More informationComparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia
Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia Seyed Mohammad Reza Taheri, MD 1 Azita Kheiltash, MD, MPH 2 Hassan Hashemi, MD 1,3 Abstract Purpose: To
More informationRefractive Dilemma. Challenging Case
Challenging Case Refractive Dilemma Section Editor: Alireza Baradaran-Rafii, MD Case presentation A 21-year old man was referred to an ophthalmology clinic insisting on getting rid of his glasses which
More informationAuthor s Affiliation. Original Article. Comparison of Biometry in Phakic and Dense Modes. Muhammad Suhail Sarwar. Unaiza Mariam
Comparison of Biometry in Phakic and Dense Modes Original Article Purpose: To compare biometry readings in phakic and dense modes. Author s Affiliation Muhammad Suhail Sarwar Unaiza Mariam Correspondence
More informationPartial Coherence Interferometry as a Technique to Measure the Axial Length of the Eye Archived Medical Policy
Partial Coherence Interferometry as a Technique to Measure the Axial Length of the Eye Applies to all products administered or underwritten by Blue Cross and Blue Shield of Louisiana and its subsidiary,
More informationFactors Affecting the Accuracy of Intraocular Lens Power Calculation with Lenstar
Original Article www.cmj.ac.kr Factors Affecting the Accuracy of Intraocular Lens Power Calculation with Lenstar Tae Hee Lee, Mi Sun Sung, Lian Cui, Ying Li and Kyung Chul Yoon* Department of Ophthalmology,
More informationPostoperative refraction changes in phacoemulsification cataract surgery with implantation of different types of intraocular lens
371-376 Iwase:Shoja 7-04-2008 16:57 Pagina 371 European Journal of Ophthalmology / Vol. 18 no. 3, 2008 / pp. 371-376 Postoperative refraction changes in phacoemulsification cataract surgery with implantation
More informationMegalocornea is a non-progressive, uniformly
Case Report 191 Anterior Megalophthalmos Chien-Kuang Tsai, MD; Ing-Chou Lai, MD; Hsi-Kung Kuo, MD; Mei-Chung Teng, MD; Po-Chiung Fang, MD We describe a 36-year-old female who suffered from presenile cataract
More informationStructural changes of the anterior chamber following cataract surgery during infancy
Structural changes of the anterior chamber following cataract surgery during infancy Matthew Nguyen, Emory University Marla Shainberg, Emory University Allen Beck, Emory University Scott Lambert, Emory
More informationTHE PENTACAM AXL. Improving Cataract Surgery Outcomes. Optical biometry and anterior segment tomography in one device
Insert to November/December 2016 Sponsored by OCULUS THE PENTACAM AXL Improving Cataract Surgery Outcomes Optical biometry and anterior segment tomography in one device A New Way to Calculate IOL Power
More informationExtended Depth Of Focus IOL For Presbyopia Correction
A closer look at Extended Depth Of Focus IOL For Presbyopia Correction Abdallah K. Hassouna, MD, PhD. Professor of Ophthalmology, Ain Shams University Consultant Ophthalmic Surgeon, Al Watany Eye Hospital
More informationCataract Surgery in Patients with a Previous History of KAMRA Inlay Implantation: A Case Series
Ophthalmol Ther (2017) 6:207 213 DOI 10.1007/s40123-017-0088-4 CASE SERIES Cataract Surgery in Patients with a Previous History of KAMRA Inlay Implantation: A Case Series Majid Moshirfar. Tyler S. Quist.
More informationINTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany
European Journal of Ophthalmology / Vol. 17 no. 3, 2007 / pp. 363-367 Anterior segment optical coherence tomography for evaluation of changes in anterior chamber angle and depth after intraocular lens
More informationMyopic Shift after Implantation of a Novel Diffractive Trifocal Intraocular Lens in Korean Eyes
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(1):16-22 https://doi.org/10.3341/kjo.2017.0060 Original Article Myopic Shift after Implantation of a Novel Diffractive Trifocal Intraocular
More informationComparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy PATIENTS AND METHODS. Patients
Special Issue Comparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy Jack X. Ma, 1 Maolong Tang, 2 Li Wang, 3 Mitchell P. Weikert, 3 David Huang, 2 and Douglas D. Koch
More informationCataract and cornea. Miltos O. Balidis PhD, FEBOphth,ICOphth ATHENS
Cataract and cornea Miltos O. Balidis PhD, FEBOphth,ICOphth CATARACT and Stromal opacities Keratoplasty Keratoconus Endothelial pathology Scars PTK Trypan blue 0.01%. Work at the transparent side of cornea
More informationCATARACT SURGERY AFTER RADIAL KERATOTOMY
AFTER RADIAL KERATOTOMY How to avoid disappointment. BY BERNARD MATHYS, MD CATARACT SURGERY Radial keratotomy (RK; Figure 1) was a popular refractive surgical procedure to correct myopia in the 1970s and
More informationAnterior movement of the crystalline lens in the process of accommodation in children
European Journal of Ophthalmology / Vol. 17 no. 4, 2007 / pp. 515-520 Anterior movement of the crystalline lens in the process of accommodation in children B.J. KALUZNY Department of Ophthalmology, Collegium
More informationEffects of Choroidal Thickness on Refractive Outcome Following Cataract Surgery in Primary Angle Closure
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(5):382-390 https://doi.org/10.3341/kjo.2017.0129 Original Article Effects of Choroidal Thickness on Refractive Outcome Following Cataract Surgery
More informationSurgical induced astigmatism in femtosecond laser assisted cataract surgery
ARTICLE Surgical induced astigmatism in femtosecond laser assisted cataract surgery Laureano Álvarez-Rementería, MD 1, Vanessa Blázquez, OD 1, Inés Contreras, MD, PhD 1,2 PURPOSE: To report the surgical
More informationThe Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive Surgery
접수번호 :08-024 Korean Journal of Ophthalmology 2009;23:1-5 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.1.1 The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive
More informationRefractive lens exchange with distance-dominant diffractive bifocal intraocular lens implantation
Graefes Arch Clin Exp Ophthalmol (2010) 248:1507 1514 DOI 10.1007/s00417-010-1345-6 REFRACTIVE SURGERY Refractive lens exchange with distance-dominant diffractive bifocal intraocular lens implantation
More informationLaser Refractive Cataract Surgery with the LenSx Laser
Laser Refractive Cataract Surgery with the LenSx Laser a Novartis company 1 LenSx Laser Important Safety Information Caution: United States Federal Law restricts this device to sale and use by or on the
More informationIntraocular Lens Power Estimation in Combined Phacoemulsification and Pars Plana Vitrectomy in Eyes with Epiretinal Membranes: A Case-Control Study
Original Article http://dx.doi.org/10.3349/ymj.2015.56.3.805 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 56(3):805-811, 2015 Intraocular Lens Power Estimation in Combined Phacoemulsification and Pars
More informationAssessing & co-managing cataract: Prof Charles NJ McGhee
Assessing & co-managing cataract: Prof Charles NJ McGhee This handout covers most of the PowerPoint slide details used in cataract lectures 1&2 and may be useful for notetaking during the lectures. Greater
More informationAstigmatic Outcomes of Temporal versus Nasal Clear Corneal Phacoemulsification
Astigmatic Outcomes of Temporal versus Nasal Clear Corneal Phacoemulsification Mohammad Pakravan, MD; Homayoun Nikkhah, MD; Shahin Yazdani, MD Camelia Shahabi, MD; Massih Sedigh-Rahimabadi, MD Labbafinejad
More informationAiming For Emmetropia after Cataract Surgery
THE EMMETROPIA IRAQI POSTGRAUATE AFTER MEICAL CATARACT JOURNAL SURGERY Aiming For Emmetropia after Cataract Surgery Ahmed Majeed Rasheed ABSTRACT: BACKGROUN: Emmetropia after cataract removal and posterior
More informationJournal of Ophthalmology & Clinical Research
Research Article Primary and Secondary Intraocular Lens Implantation in Congenital Cataract Surgery: A Retrospective Comparative Study of the Visual Outcomes Alhawsawi Abrar 1*, Alhibshi Nizar 2, Maniyar
More informationThe Effect of Phacoemulsification with Posterior Chamber Intraocular Lens Implantation on Intraocular Pressure and Anterior Chamber Depth
The Effect of Phacoemulsification with Posterior Chamber Intraocular Lens Implantation on Intraocular Pressure and Anterior Chamber Depth Mohammad Javad Mohamadi, MD 1 Reza Soltani Moghadam, MD 1 Hasan
More informationKeratometric measurements and IOL calculations in pseudophakic post-dsaek patients
Xu et al. BMC Ophthalmology (2018) 18:268 https://doi.org/10.1186/s12886-018-0931-y RESEARCH ARTICLE Open Access Keratometric measurements and IOL calculations in pseudophakic post-dsaek patients Ke Xu,
More informationPublished in: British Journal of Ophthalmology. Document Version: Peer reviewed version
Clear lens extraction for the management of primary angle closure glaucoma: surgical technique and refractive outcomes in the EAGLE cohort Day, A., Cooper, D., Burr, J. M., Foster, P., Friedman, D., Gazzard,
More informationClinical study of traumatic cataract and its management
Clinical study of traumatic cataract and its management Original article Manjula Mangane, M.R. Pujari, Chethan N. Murthy Department of Ophthalmology, Basaweshwar Teaching and General Hospital, Gulbarga,
More informationAnterior Chamber Depth Change Following Cataract Surgery in Pseudoexfoliation Syndrome; a Preliminary Study
Original Article Anterior Chamber Depth Change Following Cataract Surgery in Pseudoexfoliation Syndrome; a Preliminary Study Mohammad Reza Fallah Tafti, MD; Hossein Abdollah Beiki, MD; S. Farzad Mohammadi,
More informationAppendix Table 1. Ophthalmic drugs approved by the US Food and Drug Administration,
SUPPLEMENTARY DATA Appendix Table 1. Ophthalmic drugs approved by the US Food and Drug Administration, 2002-2012 Approval Year Name Indication Pivotal Trial Design Randomized Comparator Masked Post-Approval
More informationPediatric intraocular lens power calculations
REVIEW C URRENT OPINION Pediatric intraocular lens power calculations Mary A. O Hara Purpose of review To implant an appropriate intraocular lens (IOL) in a child, we must measure the eye well, calculate
More informationThe visual outcome after implantation of the Multifocal Intra Ocular Lens. Dr.Bhargav Dave National Institute of Ophthalmology Pune
The visual outcome after implantation of the Multifocal Intra Ocular Lens Dr.Bhargav Dave National Institute of Ophthalmology Pune 1 The era of cataract surgery has come leaps and bounds since the inception
More informationIntrascleral-fixated intraocular lenses for aphakic correction in the absence of capsular support
European Journal of Ophthalmology / Vol. 17 no. 5, 2007 / pp. 714-719 Intrascleral-fixated intraocular lenses for aphakic correction in the absence of capsular support R.A. AZNABAYEV, I.S. ZAIDULLIN, M.S.H.
More informationSTUDY OF EFFECTIVENESS OF LENS EXTRACTION AND PCIOL IMPLANTATION IN PRIMARY ANGLE CLOSURE GLAUCOMA Sudhakar Rao P 1, K. Revathy 2, T.
STUDY OF EFFECTIVENESS OF LENS EXTRACTION AND PCIOL IMPLANTATION IN PRIMARY ANGLE CLOSURE GLAUCOMA Sudhakar Rao P 1, K. Revathy 2, T. Sreevathsala 3 HOW TO CITE THIS ARTICLE: Sudhakar Rao P, K. Revathy,
More informationLearn Connect Succeed. JCAHPO Regional Meetings 2015
Learn Connect Succeed JCAHPO Regional Meetings 2015 Pediatric Cataracts: Complicated Cases and Controversies M. Edward Wilson, M.D. N. Edgar Miles Professor of Ophthalmology and Pediatrics Storm Eye Institute
More informationPearls for the Refractive Technician Fadiah Alkhawaldeh, IMBA, COT, ROUB
Pearls for the Refractive Technician Fadiah Alkhawaldeh, IMBA, COT, ROUB Cleveland Clinic Cole Eye Institute OOS, Columbus, OH February, 2014 alkhawf@ccf.org NO FINANCIAL DISCLOSURES A Puzzle of an Eye
More informationIOL Master. Photorefractive keratectomy
89/11/3 : 89/9/3 : 1389 /119 / IOL Master * Photorefractive keratectomy 3 2 1 2 4 Photorefractive keratectomy(prk) (IOL) :. IoL-Master.. (clinical trial) :. 30 20 35.. PRK. IOL Master ( GM300,china) IOL.
More informationBiometric parameters, corneal astigmatism and ocular comorbidity in cataract surgery patients
Medical sciences (2018) 1 10 Biometric parameters, corneal astigmatism and ocular comorbidity in cataract surgery patients Aistė Varoniukaitė¹, Laurita Vaičaitė¹, Loreta Kuzmienė² 1 Faculty of Medicine,
More informationA Study to Reach the Proper Measurement of the White to White Diameter Using both the Caliper and IOL Master for ICL Suitable Size
Current Science International Volume : 06 Issue : 01 Jan.- Mar. 2017 Pages: 41-49 A Study to Reach the Proper Measurement of the White to White Diameter Using both the Caliper and IOL Master for ICL Suitable
More informationMEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment
MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND,, PAGE: 1 OF: 5 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product, including
More informationAnterior segment imaging
Article Date: 11/1/2016 Anterior segment imaging AS OCT vs. UBM vs. endoscope; case based approaches BY BENJAMIN BERT, MD, FACS AND BRIAN FRANCIS, MD, MS Currently, numerous imaging modalities are available
More informationComplication and Visual Outcome after Peadiatric Cataract Surgery with or Without Intra Ocular Lens Implantation
Original Article Complication and Visual Outcome after Peadiatric with or Without Intra Ocular Lens Implantation Mazhar-ul-Hasan, Umair A. Qidwai, Aziz-ur-Rehman, Nasir Bhatti, Rashid H. Alvi Pak J Ophthalmol
More informationIssue 15 The following key clinical peer reviewed journals will be reviewed: MONTHLY RESEARCH UPDATE 151(3) American Journal of Ophthalmology 129(5)
Welcome to Bausch and Lomb s monthly research update. With our background in clinical ophthalmic research, mainly of the anterior eye, Bausch and Lomb have asked us to produce an independent report of
More information9/25/2017 CASE. 67 years old On 2 topical meds since 3 years. Rx: +3.0 RE LE
CASE 67 years old On 2 topical meds since 3 years Rx: +3.0 /-0.5@65 RE +2.5/-0.5@115 LE IOP : 17 RE 19 LE CD: 0.5 RE 0.6 LE 1 67 years old On 2 topical meds since 3 years Rx: +3.0 /-0.5@65 RE +2.5/-0.5@115
More informationManagement of Corneal Astigmatism by Limbal Relaxing Incisions during Cataract Surgery
Management of Corneal Astigmatism by Limbal Relaxing Incisions during Cataract Surgery Mohammad Ali Zare, MD 1 Mehdi Hosseini Tehrani, MD 2 Mohsen Gohari, MD 3 Mahmood Jabbarvand, MD 1 Mohammad Naser Hashemian,
More informationThree-year Follow-up of the Artisan Phakic Intraocular Lens for Hypermetropia
Three-year Follow-up of the Artisan Phakic Intraocular Lens for Hypermetropia Ruchi Saxena, MS, 1 Monika Landesz, MD, PhD, 2 Bastiaantje Noordzij, 1 Gregorius P. M. Luyten, MD, PhD 1 Purpose: We report
More informationBiometric Measurements: Average Intra-Ocular Lens Power for Sudanese Patients with Senile Cataract
Optics 2015; 4(4): 25-30 Published online September 28, 2015 (http://www.sciencepublishinggroup.com/j/optics) doi: 10.11648/j.optics.20150404.11 ISSN: 2328-7780 (Print); ISSN: 2328-7810 (Online) Biometric
More informationRESEARCH PROJECT. Retina and Vitreous Service, Hadassah Ein Keren Hospital, Jerusalem-Israel
RESEARCH PROJECT IOL CALCULATION FOR HIGHLY MYOPIC EYES, COMPARISON OF DIFFERENT PUBLISHED METHODS. HADASSAH EIN KEREN HOSPITAL, JERUSALEM, ISRAEL, 2014 AUTHOR: Carlos Idrobo MD, Retina and Vitreous Service,
More informationOriginal article. Preliminar treatment of astigmatism during phacoemulsification. Sharma BR 1, Kumar A 2 1
Original article Preliminar eliminary experiences with limbal relaxing incision for treatment of astigmatism during phacoemulsification Sharma BR 1, Kumar A 2 1 Associate Professor, 2 Ophthalmologist,
More informationASCRS 2016 Instructional Course Mastering Femtosecond Laser Assisted Phacoemulsification. An Evidence-Based Review
ASCRS 2016 Instructional Course 07-410 Mastering Femtosecond Laser Assisted Phacoemulsification An Evidence-Based Review TIMOTHY V ROBERTS MBBS (NSW), MMed (Syd), FRANZCO, FRACS, GAICD Vision Eye Institute,
More informationComparison of Preoperative and Postoperative Ocular Biometry in Eyes with Phakic Intraocular Lens Implantations
Original Article http://dx.doi.org/10.3349/ymj.2013.54.5.1259 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 54(5):1259-1265, 2013 Comparison of Preoperative and Postoperative Ocular Biometry in Eyes
More informationPosterior Chamber Intraocular Lens Implantation in Pediatric Cataract with Microcornea and/or Microphthalmos
Posterior Chamber Intraocular Lens Implantation in Pediatric Cataract with Microcornea and/or Microphthalmos Young Suk Yu, MD 1,2, Seong-Joon Kim, MD 1,2, Ho Kyoung Choung, MD 3 Department of Ophthalmology,
More informationAcrySof ReSTOR Multifocal versus AcrySof SA60AT Monofocal Intraocular Lenses: A Comparison of Visual Acuity and Contrast Sensitivity
AcrySof ReSTOR Multifocal versus AcrySof SA6AT Monofocal Intraocular Lenses: A Comparison of Visual Acuity and Contrast Sensitivity Hassan Hashemi, MD,2 Hamid Reza Nikbin, MD 2 Mehdi Khabazkhoob, MSc 2
More informationAnterior chamber depth measurement: interchangeability between Pentacam and other devices
UPDATE/REVIEW Anterior chamber depth measurement: interchangeability between Pentacam and other devices Alberto Domínguez-Vicent, MSc 1 ; Daniel Monsálvez-Romín, OD 1 ; Vicent Sanchis, MSc 1 ; Santiago
More informationREFRACTIVE LENS SURGERY: WHEN AND WHY?
REFRACTIVE LENS SURGERY: WHEN AND WHY? Symposium of the Hellenic Society of Intraocular Implant and Refractive Surgery Cairo, 14/3/2018 Pandelis A. Papadopoulos, MD, PhD, FEBO, FEBOS-CR Director, Ophthalmology
More informationEvaluation of the Effect of Diabetes Mellitus on Corneal Biomechanics
The Egyptian Journal of Hospital Medicine (January 2019) Vol. 74 (4), Page 782-788 Abd Elmagid Mohamed Tag Eldin, Abdulmoez Haddad Ahmed, Amr Khalil Abdelmonem* Ophthalmology department, Faculty of Medicine,
More informationEffects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes. Abstract
Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes Javad Amoozadeh, MD 1 Nima Mirzaee Rad, MD 2 Amir Houshang Beheshtnejad, MD 3 Ahmad Kheirkhah, MD 1 Hassan Hashemi, MD 4,5
More informationA Clinical Evaluation of Multifocal versus Monofocal Intraocular Lenses after Cataract Extraction in a Tertiary Care Hospital in India
EUROPEAN ACADEMIC RESEARCH Vol. II, Issue 1/ April 2014 ISSN 2286-4822 www.euacademic.org Impact Factor: 3.1 (UIF) DRJI Value: 5.9 (B+) A Clinical Evaluation of Multifocal versus Monofocal Intraocular
More informationMicrocoaxial Phacoemulsification versus Conventional Phacoemulsification: A Prospective Study
Microcoaxial Phacoemulsification versus Conventional Phacoemulsification: A Prospective Study Seyed Javad Hashemian, MD 1 Farzad Pakdel, MD 2 Alireza Foroutan, MD 2 Mahmood Joshaghani, MD 2 Jafar Ghaempanah,
More informationClinical Evaluation of the BunnyLens IOL
Clinical Evaluation of the BunnyLens IOL Introduction: BunnyLens is a foldable Hydrophlic Acrylic IOL with four ear shaped haptic design. The lens design offers many advantages in terms of: 1. Centration
More informationMultifocal Toric Swiss Army Knife For the Cataract and Refractive Surgeon
Multifocal Toric Swiss Army Knife For the Cataract and Refractive Surgeon Leonidas Mavroudis Thessaloniki Orasi Day Case Microsurgery Clinic Athens 2018 Disclaimer : Alcon Leonidas Mavroudis Thessaloniki
More informationOphthalmic A - Scan ultrasonography studies in myopia
Original article Ophthalmic A - Scan ultrasonography studies in Manjula Mangane, Vijay S. Lodha, Chethan Murthy Department of Ophthalmology, Basaweshwar teaching and general hospital, Gulbarga, Karnataka,
More informationGetting Started GUIDE. TECNIS Symfony IOL and TECNIS Symfony TORIC IOL. TECNIS SYMFONY IOL pg 1
Getting Started GUIDE TECNIS Symfony IOL and TECNIS Symfony TORIC IOL TECNIS SYMFONY IOL pg 1 IN THIS GUIDE LENS SPECIFICATIONS pg 3 PREOPERATIVE pg 4-5 TECNIS Symfony IOL combines two complementary and
More informationStudies on the use of 3.5-mm to 4-mm corneal incisions in
C L I N I C A L T R I A L S Astigmatic Change Induced by 2.8-mm Corneal Incisions for Cataract Surgery Jaime Tejedor 1,2 and José A.Pérez-Rodríguez 1 From the 1 Department of Ophthalmology, Hospital Ramón
More informationMultifocalS: A Reference Guide
MultifocalS: A Reference Guide Your complete resource for advanced technology IOL implantation, from patient selection to post-op evaluation For best results, use this comprehensive guide from start to
More informationAccuracy of intraocular lens power calculation in pediatric cataracts with less than a 20 mm axial length of the eye
Original article Accuracy of intraocular lens power calculation in pediatric cataracts with less than a 20 mm axial length of the eye Purushottam Joshi, 1 Raman Mehta, 2 Suma Ganesh 2 1 Mechi Eye Hospital,
More informationBiometry and intraocular lens power calculation Alexander C. Lee, Mujtaba A. Qazi and Jay S. Pepose
Biometry and intraocular lens power calculation Alexander C. Lee, Mujtaba A. Qazi and Jay S. Pepose Pepose Vision Institute, and the Department of Ophthalmology and Visual Sciences, Washington University
More informationRetinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography
Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo
More informationCataract Surgery by Phacoemulsification Surgical and Visual Outcome with Foldable and Non-Foldable Lenses
www.jmscr.igmpublication.org Impact Factor 3.79 ISSN (e)-2347-176x Cataract Surgery by Phacoemulsification Surgical and Visual Outcome with Foldable and Non-Foldable Lenses Authors Dr. Chitra Pande 1,
More informationEUROTIMES ESCRS SUPPLEMENT FEBRUARY 2007
Pentacam opens eyes to new diagnostic possibilities Scheimpflug based anterior segment tomography EUROTIMES ESCRS SUPPLEMENT FEBRUARY 2007 XXIV Congress of the ESCRS London, September 2006 Introduction
More informationLearn Connect Succeed. JCAHPO Regional Meetings 2017
Learn Connect Succeed JCAHPO Regional Meetings 2017 Cataract Surgery in 2017 DARBY D. MILLER, MD MPH CORNEA, CATARACT AND REFRACTIVE SURGERY ASSISTANT PROFESSOR OF OPHTHALMOLOGY MAYO CLINIC FLORIDA Natural
More informationIOL Calculations in Short, Long, and Postrefractive Eyes
IOL Calculations in Short, Long, and Postrefractive Eyes Ildamaris Montes de Oca, MD Sabite E. Gökce, MD Katie Hallahan, MD Li Wang, MD, PhD Douglas D. Koch, MD Intraocular lens (IOL) Calculations: Fundamentals
More informationAssessment of Retinal Nerve Fiber Layer Changes by Cirrus High-definition Optical Coherence Tomography in Myopia
Divya Singh et al Original REASEARCH 10.5005/jp-journals-10028-1223 Assessment of Retinal Nerve Fiber Layer Changes by Cirrus High-definition Optical Coherence Tomography in Myopia 1 Divya Singh, 2 Sanjay
More informationORIGINAL ARTICLE. HIGH VOLUME CAMP SURGERIES A CLINICAL STUDY D. N. Prakash, K, Sathish, Sankalp Singh Sharma, Soujanya. K, Savitha Patil.
HIGH VOLUME CAMP SURGERIES A CLINICAL STUDY D. N. Prakash, K, Sathish, Sankalp Singh Sharma, Soujanya. K, Savitha Patil. 1. Assistant Professor. Department of Ophthalmology, MMC & RI, Mysore, 2. Associate
More informationBilateral Microphthalmos Associated with Papillomacular Fold, Severe Hyperopia and Steep Cornea
Bilateral Microphthalmos Associated with Papillomacular Fold, Severe Hyperopia and Steep Cornea Mojtaba Abrishami, MD 1 Alireza Maleki, MD 2 Ali Hamidian-Shoormasti, MD 3 Mostafa Abrishami, MD 4 Abstract
More information