Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy. Abstract
|
|
- Felix Booker
- 6 years ago
- Views:
Transcription
1 Progress Report No. 2-3, March 31, 1999 The Home Automation and Healthcare Consortium Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy Prof. Kamal Youcef-Toumi Principal Investigator Vidi A. Saptari Graduate Research Assistant Abstract A modular Fourier-transform spectrometer was designed and built for the noninvasive blood glucose measurement research using the near-infrared absorption technique. A series of experiments were performed to evaluate the spectrometer s efficiency and its stability. The efficiency, which represents the photon loss through the spectrometer, is proportional to the achievable signal-to-noise ratio. The stability of the instrument is important for measurement repeatability, which would affect the sensitivity of the quantitative assessment of glucose concentration. A preliminary experiment on aqueous glucose solutions was performed, with the purpose of identifying the relevant spectral bands for glucose concentration prediction. It is found that there are two relatively strong bands where glucose molecules affect the water absorption spectrum. One is around 5950 cm around 7250 cm, where there is a glucose absorption band, and the other is, where the magnitude of water absorption is reduced by the presence of glucose molecules. Simple algorithm utilizing spectral information from these two bands may be sufficient for glucose concentration prediction. 1
2 I. Introduction This work is motivated by the profound need for a noninvasive way to measure glucose concentration in the blood. A noninvasive glucose monitoring device would provide a safer and a more convenient method to treat and control diabetes. The goal of diabetes therapy, within and outside hospital, is to approximate the 24-hour blood glucose profile of a normal individual, which necessitates continuous monitoring. Several optical techniques have received considerable attention over several years, which include Raman, absorption and polarimetry spectroscopy [1-4]. Among them, the absorption technique in the near-infrared region was selected for our research. As discussed in the previous reports, one of the advantages of the absorption technique is its relatively stronger signal-to-noise ratio. Furthermore, it requires simpler and less expensive instrumentation than the other optical methods. We have completed the development of a modular Fourier-transform spectrometer, which would be used for the subsequent experiments. The work was motivated by the need for a versatile and dedicated instrument for research in the field of noninvasive blood glucose sensing. Great emphasis is placed to ensure that each element is designed or chosen to give maximum signal-to-noise ratio in the near-infrared region. Overview of the system design is presented in the previous report [5]. In this report, we present the performance evaluation of the spectrometer on the basis of measurements chosen because of their impact on the instrument s capability for blood glucose concentration prediction. Identification of relevant wavelength ranges for the glucose analysis, with the experimental results is also presented in this paper. Finally, we discuss challenges and obstacles we anticipate. II. (i) System Description Theory of Operation Figure 1 shows a schematic of the FT spectrometer setup. A white light from a tungsten halogen source is collected and collimated by a concave mirror and a pair of lenses. An aperture is used to limit the amount of divergence of the light entering the interferometer. The interferometer is of Michelson type, comprising a beamsplitter and two perpendicular plane mirrors. One of the mirrors moves linearly in the axis as indicated in the diagram. This produces a variable path difference between the two beams reflected off of the mirrors. The recombined beam is then focused onto the sample and onto the detector. The detector measures the interferogram, which is stored in the computer. This signal is then Fourier-transformed to obtain the spectrum. A second Michelson interferometer with an HeNe reference laser, as indicated in the diagram, is used for controlling the velocity and position of the linear motion. In addition, it is also responsible for clocking the data acquisition at precise and repeatable locations of mirror retardation. This would enable repetitive scanning to improve the signal-to-noise ratio. 2
3 Reference laser Laser detector Beam-splitter (Interferometer 2) Plane Mirror Concave mirror Light source Objective lens Aperture Collimating lens +1mm Moving plane mirror Beam-splitter (Interferometer 1) Plane mirror Liquid sample Focusing lens Detector Figure 1: Fourier transform spectrometer layout (ii) System Implementation and Components Design Consideration From the literature and the previous preliminary experiments, it has been realized that signals due to the blood glucose variations in the physiological range are weak. Furthermore, in the noninvasive application, there is further loss of signal due to scattering. As a result, realization of accurate blood glucose quantification would likely be limited by the amount of signal-to-noise ratio achievable. Hence, it is important that each element of the spectrometer is designed to maximize the signal-to-noise ratio. The spectrometer uses a tungsten-halogen light source that provides a continuous spectrum from the visible through the near-infrared. An extended, thermoelectricallycooled InGaAs detector is selected for maximum signal-to-noise ratio in the 1000nm nm wavelength region. For minimum loss due to absorption, Calcium-fluoride lenses are used for light collection and collimation. The photograph below shows the setup of the system. The absorption spectrum in the near-infrared region is usually broad and overlapping. Therefore, high spectral resolution is not required. Since throughput is inversely related to the resolution, it is advantageous to match the spectrometer resolution to the required resolution [5]. Light Source 3
4 White Light InGaAs Photodetector Beamsplitter HeNe Laser Sample Photograph 1: Fourier-transform spectrometer setup III. (i) Spectrometer Characterization and Performance Evaluation Efficiency Efficiency here is defined as the power of radiation received by the detector as compared with the power emitted by the source. This represents the photon loss through the spectrometer, which may be due to the absorption by the lenses and the beamsplitter, the loss to the environment, as well as the loss due to misalignment of the interferometer. It is assumed to be a constant Sample value throughout the relevant wavelength range. The equation below shows its relation with the achievable signal-to-noise ratio [6]. where, U SNR = v ( T). Θ. v. t NEP 1/ 2 Voice-coil. ζactuator Voice-coil Actuator U v (T) is spectral density at wavenumber v from a black body source at a temperature T ( W / sr. cm 2. cm 1 ) 2 Θ is the throughput of the system ( cm. sr) v is the resolution of the spectrum ( cm ) t is the integrating time in seconds 4
5 ζ is the efficiency accounting for losses due to the optical components NEP stands for noise equivalent power, which is a sensitivity figure-of-merit of / the detector ( W. Hz 2 ) In the equation above, the source is assumed to be a black-body radiation. In the case of noninvasive measurement, the source term U v (T ) would be replaced by the radiance of the diffuse-reflected light from the sample, such as the skin. To assess the spectrometer efficiency, light from the tungsten-halogen source was directed onto the detector through a narrow-band filter at 1600nm. The radiation power at the detector, Ρ 1 at the output of the preamplifier, was recorded. The procedure was done at full throughput, limited by the area of the detector and the speed of the focusing optics. Using the same light source and band-pass filter, an interferogram was recorded using the spectrometer. The peak of the interferogram, Ρ 2, which corresponded to the radiation power when all the wavelengths interfere constructively, was recorded. Efficiency was then computed as Ρ 2 Ρ1, which was equal to This value is very good considering the complexity of the system, and is in good agreement with the rule-of-thumb estimations of Fourier-transform spectrometer efficiencies of 10% [7]. (ii) Stability Another criterion of importance is the spectrometer stability, which would translate into repeatability of the measurements. The weak glucose signal intrinsic to the absorption technique requires excellent stability of the spectrometer. The repeatability of the spectrum is especially important in the regions where glucose absorption bands are to be analyzed. Irreproducibility of measurements could arise from many sources, such as change in alignment, interferometer thermal expansion, change in sample temperature, scattering and variation due to the light source itself. To assess stability, water spectra were acquired every 3 minutes over a period of one hour. This duration and periodicity was chosen to simulate an actual experiment. Each spectrum was a result of a Fourier-transformation of 30 averaged, double-sided interferograms. The first spectrum was subtracted from each subsequent spectrum to get the difference spectra, which are plotted in figure 2. To examine the impact of these variations, the plot should be examined together with the difference spectra of glucose solutions, which is presented in the next section
6 Relative Magnitude Figure 2: Differential spectra of repetitive water measurements IV. Measurement of Glucose in Water A preliminary aqueous glucose measurement experiment was done with the purpose of identifying the relevant bands for glucose analysis, as well as examining the capability of the instrument for the task. Glucose solutions with high concentration were used in this study. The lowest concentration was 50 mmol / L, which corresponded to about 4 times as high as the upper physiological range. Figure 3 shows the transmission spectra of pure water and glucose solution. Figure 4 shows the differential spectra of glucose solutions with different concentrations with respect to the water transmission spectrum. It can be seen from the two plots that the presence of glucose affects the spectrum of water in two regions. One is between 5800 cm and 6100 cm where there is a glucose absorption band. The other is a narrower band at around 7250 cm, where there is, in fact, more transmission due to the presence of glucose. This might result from the displacement of water molecules by glucose molecules. In these regions, effects of glucose on water spectra are more pronounced than in the lower wavelength regions, which have been considered by several other investigators [2, 8]. However, water absorption here is much higher. Therefore, in order to probe these bands, a long-wave-pass filter with a cutoff around 1400nm was required. Otherwise, they would not be visible at all, as the shorter-wave radiation would dominate. The water-glucose differential spectra (figure 4) were obtained using the same throughput and detector amplifier gain setting as the experiment to obtain the stability differential spectra (figure 2). Therefore, the magnitudes of the spectra between the two figures can be directly compared. At 50 mmol / L, glucose altered the water spectrum by 6 units at around 6000 cm and by 3 units at 7250 cm. On the other hand, the extreme values of the variations of water spectra are 1.5 units at 6000 cm and 2.5 units at 7250 cm. This result suggests that for accurate lower glucose concentration prediction, stability needs to be improved. 6
7 Relative Magnitude cm Figure3: Water-glucose solution spectrum (blue: pure water ; magenta: 100 mmol / L glucose solution) Figure4: Water-glucose differential spectra (red: 50 mmol / L ; magenta: 100 mmol / L ; yellow: 200 mmol / L ; black: 500 mmol / L ) cm V. Research Plan 7
8 The experimental result shows that simple algorithm utilizing the two relevant bands might be sufficient for glucose concentration prediction. However, repeatability of the measurements needs to be improved before low concentration prediction could be achieved. Our short-term plan includes improvement of measurement repeatability. This involves identification of the source of error, which may be due to instrumental factor or the sample element. When this is accomplished, careful analysis of achievable sensitivity for waterglucose and blood-glucose measurement will be attempted. This should be investigated through both experimentation and theoretical studies. Additional problems with respect to the noninvasive application are signal loss due to light scattering in tissue and physiological variations such as skin temperature changes. Both degrade the signal-to-noise ratio. In this research, we aim to understand the origins of the signals and the sources of errors. This would enable appropriate modifications to the instrumentation design and/or procedural scheme to improve the accuracy of noninvasive blood glucose measurement. VI. References 1. Marbach R. et. al., Noninvasive Blood Glucose Assay by Near Infrared Diffuse Reflectance Spectroscopy of the Human Inner Lip, Applied Spectroscopy 47 (7): , Robinson M.R. et. al., Noninvasive Glucose Monitoring in Diabetic Patients: A Preliminary Evaluation, Clin. Chem. 38 (9): , Cote G.L. et. Al., Noninvasive Glucose Sensing Utilizing a Digital Close-Loop Polarimetric Approach IEEE Trans. On Biomed. Eng. 44(12): , Feld M.S. et. Al., Rapid, Noninvasive Concentration Measurements of Aqueous Biological Analytes by Near-Infrared Raman Spectroscopy, Applied Optics 35(1): , The Home Automation and Healthcare Consortium Progress Report Griffiths P.R., de Haseth J.A., Fourier Transform Infrared Spectrometry, John Wiley & Sons, Mattson D.R., Sensitivity of A Fourier Transform Infrared Spectrometer, Applied Spectroscopy 32(4): , Muller U.A. et. al., Non-invasive Blood Glucose Monitoring by Means of Near Infrared Spectroscopy: Methods for Improving the Reliability of the Calibration Models, Artif. Pancreas and Related Tech. In Diabetes and Endocrinology 20 (5): ,
In vivo Infrared Spectroscopy
In vivo Infrared Spectroscopy Zhe Xia March 29, 2017 Outline Introduction Theory Instrument Case study Conclusion Introduction In vivo: Latin for within the living, In vivo studies focus on the effects
More informationEvaluation of Measurement Sites for Noninvasive Blood Glucose Sensing with Near-Infrared Transmission Spectroscopy
Clinical Chemistry 45:9 1621 1627 (1999) Oak Ridge Conference Evaluation of Measurement Sites for Noninvasive Blood Glucose Sensing with Near-Infrared Transmission Spectroscopy Jason J. Burmeister * and
More informationVidi A. Saptari. Submitted to the Department of Mechanical Engineering in partial fulfillment of the requirements for the degree of.
A Spectroscopic System for Near Infrared Glucose Measurement by Vidi A. Saptari B.S., Carnegie Mellon University (1997) S.M., Massachusetts Institute of Technology (1999) Submitted to the Department of
More informationNon-invasive blood glucose measurement by near infrared spectroscopy: Machine drift, time drift and physiological effect
Spectroscopy 24 (2010) 629 639 629 DOI 10.3233/SPE-2010-0485 IOS Press Non-invasive blood glucose measurement by near infrared spectroscopy: Machine drift, time drift and physiological effect Simon C.H.
More informationDetailed Syllabus. Waves and Modern Physics Spring 2011
Detailed Syllabus Waves and Modern Physics Spring 2011 Prof. Antonio Badolato Date Lecture Topics; Chapters from Books (G = Giancoli) Homework (HW) and Recitation (R) Assignments 12 Jan, Wed Oscillations:
More informationNocturnal hypoglycemic alarm based on nearinfrared
University of Iowa Iowa Research Online Theses and Dissertations Spring 2014 Nocturnal hypoglycemic alarm based on nearinfrared spectroscopy Sanjeewa Rasika K. Ranasinghe Pathirajage University of Iowa
More informationDETECTION OF BREAST & CERVICAL CANCER USING RAMAN SPECTROSCOPY
DETECTION OF BREAST & CERVICAL CANCER USING RAMAN SPECTROSCOPY Binay Bhushan 1, Asima Pradhan 2 Post-Doctoral Fellow, Department of Physics, Indian Institute of Technology, Kanpur, India 1 Professor, Department
More informationNoninvasive near infrared spectroscopy on living tissue with multivariate calibration approaches
University of Iowa Iowa Research Online Theses and Dissertations Fall 2010 Noninvasive near infrared spectroscopy on living tissue with multivariate calibration approaches Chuannan Bai University of Iowa
More informationMonitoring Tumor Therapy
Monitoring Tumor Therapy Zenalux Biomedical, Inc. The Zenascope was used to monitor the physiological response to a vascular disrupting agent in mouse tumors. Changes in hemoglobin concentration, hemoglobin
More informationStress Wave Focusing Transducers
UCRL-K-130697 PREPRINT Stress Wave Focusing Transducers Steven R. Visuri, Richard A. London, Luiz Da Silva This paper was prepared for submittal to Optical Society of America, Spring Topical Meetings Orlando,
More informationULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah
ULTRASOUND IMAGING EE 472 F2018 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Diagnostic Ultrasound: Physics and Equipment, 2nd ed., by Peter R. Hoskins (Editor), Kevin Martin (Editor),
More informationLinear Ultrasonic Wave Propagation in Biological Tissues
Indian Journal of Biomechanics: Special Issue (NCBM 7-8 March 29) Linear Ultrasonic Wave Propagation in Biological Tissues Narendra D Londhe R. S. Anand 2, 2 Electrical Engineering Department, IIT Roorkee,
More informationTrans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By
Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By Dr. Syed Tufail Hussain Sherazi E-mail: tufail_sherazi@yahoo.com National Center of Excellence in
More informationDiploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus
Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus Page 1 of 7 11/18 Candidates are expected to cover all of the content of this syllabus when preparing
More informationThe Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics
The Physics of Ultrasound Pipe Organ 10-8000 Emission Dog 452-1080 Man 85-1100 Spectrum Bat 10,000-120,000 Porpoise 7000-120,000 Claus G. Roehrborn Professor and Chairman 10 20 Cycles per second Reception
More informationNoninvasive monitoring of glucose concentration using differential absorption low-coherence interferometry based on rapid scanning optical delay line
Journal of Physics: Conference Series Noninvasive monitoring of glucose concentration using differential absorption low-coherence interferometry based on rapid scanning optical delay line To cite this
More informationOn Different Wavelengths: The Spectrum of Retinal Imaging. On Different Wavelengths: The Spectrum of Retinal Imaging. Wavelength Specific Imaging
On Different Wavelengths: The Spectrum of Retinal Imaging Timothy J. Bennett, CRA, FOPS, OCT-C Penn State Hershey Eye Center Hershey, PA On Different Wavelengths: The Spectrum of Retinal Imaging Wavelengths
More informationIntrinsic Signal Optical Imaging
Intrinsic Signal Optical Imaging Introduction Intrinsic signal optical imaging (ISOI) is a technique used to map dynamics in single cells, brain slices and even and most importantly entire mammalian brains.
More informationbio-mof-1 DMASM Wavenumber (cm -1 ) Supplementary Figure S1 FTIR spectra of bio-mof-1, DMASMI, and bio-mof-1 DMASM.
bio-mof-1 Transmittance bio-mof-1 DMASM DMASMI 2000 1500 1000 500 Wavenumber (cm -1 ) Supplementary Figure S1 FTIR spectra of bio-mof-1, DMASMI, and bio-mof-1 DMASM. Intensity (a.u.) bio-mof-1 DMASM as
More informationSmart Probe for Uric Acid Detection
Smart Probe for Uric Acid Detection Rahul Pimpare 1, Samyak Ramteke 2, Pauroosh Kaushal 3 1 Student,Department of instrumentation & Control Engineering, D Y Patil College of Engineering Akurdi, Pune,India.
More informationPhotochemical Applications to the Study of Complexity Phospholipid Bilayer Environments
Virginia Commonwealth University VCU Scholars Compass Theses and Dissertations Graduate School 2006 Photochemical Applications to the Study of Complexity Phospholipid Bilayer Environments Christopher John
More informationPossibilities of time resolved studies at FLASH
Possibilities of time resolved studies at FLASH S. Düsterer Outline Pump-probe at FLASH FEL + FEL Optical lasers + FEL THz + FEL The future FEL + FEL split and delay Autocorrelator / beam splitter Uni
More informationIntroduction. Device Description. Dual-Band Spectral Output
Optimum Spectrum and Pulse Shape for Vascular Lesion Treatment: The Science Behind MaxG James Childs, Ph.D.; Andrei Erofeev, Ph.D.; Mikhail Smirnov, Ph.D.; and Gregory Altshuler, Ph.D., Sc.D. Introduction
More informationPermanent City Research Online URL:
Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement
More informationTechnical Discussion HUSHCORE Acoustical Products & Systems
What Is Noise? Noise is unwanted sound which may be hazardous to health, interfere with speech and verbal communications or is otherwise disturbing, irritating or annoying. What Is Sound? Sound is defined
More informationDevelopment of a Goniometer with Nanoradian Accuracy
Development of a Goniometer with Nanoradian Accuracy Basil Blank, Thomas Kupp, Eric Johnson, Alex Deyhim, Chi-Chang Kao*, Wolfgang Caliebe*, Gary Navrotski** Abstract Advanced Design Consulting, Inc. 126
More informationAcceptance testing of a large aperture dynamic wavefront sensor
Acceptance testing of a large aperture dynamic wavefront sensor David J. Fischer*, Alan Wertheimer, Laurie Hill Eastman Kodak Company, 1447 St. Paul St, Rochester, NY 14653 ABSTRACT A large aperture dynamic
More informationChromatography Vacuum Ultraviolet Spectroscopy
Application Note Differentiation and Determination Differentiation and Determination of Fatty Acid Methyl of Fatty Esters Acid by Gas Methyl Chromatography Esters by Vacuum Gas Ultraviolet Spectroscopy
More informationNon Invasive Estimation of Blood Glucose using Near Infra red Spectroscopy and Double Regression Analysis
Non Invasive Estimation of Blood Glucose using Near Infra red Spectroscopy and Double Regression Analysis Swathi Ramasahayam 1, Sri Haindavi K 2 Center for VLSI and Embedded Systems Technology 1,2,4, IIIT-Hyderabad,
More informationFive-Fold Reduction of Lasing Threshold near the First ΓL-Pseudogap of ZnO Inverse Opals arxiv: v1 [physics.
Five-Fold Reduction of Lasing Threshold near the First ΓL-Pseudogap of ZnO Inverse Opals arxiv:0907.0736v1 [physics.optics] 4 Jul 2009 Michael Scharrer 1, Heeso Noh 1,2, Xiaohua Wu 1, Mark A Anderson 1,
More informationChapter 12: Mass Spectrometry: molecular weight of the sample
Structure Determination: hapter 12: Mass Spectrometry- molecular weight of the sample; formula hapter 12: Infrared Spectroscopy- indicated which functional groups are present hapter 13: Nuclear Magnetic
More informationThe optical properties and spectral features of malignant skin melanocytes in the terahertz frequency range
Journal of Physics: Conference Series PAPER OPEN ACCESS The optical properties and spectral features of malignant skin melanocytes in the terahertz frequency range Related content - Development of the
More informationWHITE PAPER. Efficient Measurement of Large Light Source Near-Field Color and Luminance Distributions for Optical Design and Simulation
Efficient Measurement of Large Light Source Near-Field Color and Luminance Distributions for Optical Design and Simulation Efficient Measurement of Large Light Source Near-Field Color and Luminance Distributions
More informationMuscle OxyGen monitor. The Science Behind Moxy
Muscle OxyGen monitor The Science Behind Moxy How Does Moxy Monitor Work? Moxy Monitor works by shining near-infrared light onto the skin and detecting some of the light after it has travelled into the
More informationCHAPTER 3 MEASUREMENT OF GLUCOSE IN BLOOD SERUM
55 CHAPTER 3 MEASUREMENT OF GLUCOSE IN BLOOD SERUM 3.1 INTRODUCTION Glucose is the most important carbohydrate fuel in the body. In the fed state, the majority of circulating glucose comes from the diet
More informationDouble Stage Seeded FEL With Fresh Bunch Injection Technique at FERMI
Double Stage Seeded FEL With Fresh Bunch Injection Technique at FERMI Enrico Allaria on behalf of the FERMI commissioning team 1 Outline Seeded Free Electron Lasers Benefits and possibilities Seeded FEL
More informationBLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR
BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR P.Daarani 1 & A.Kavithamani 2 International Journal of Latest Trends in Engineering and Technology pp. 141-147 DOI: http://dx.doi.org/10.21172/1.ires.19
More informationFURTHER STUDY OF SPECTRUM FROM STOMACH CANCER SERUM EMISSION
FURTHER STUDY OF SPECTRUM FROM STOMACH CANCER SERUM EMISSION Xiaozhou Li 1, Huiqiang Jin 2, Ming Lei 3 1 Sence and Research Section, Liaoning University, Shenyang, 110036, China; 2 Physics Department,
More informationAvailable Online through Research Article
ISSN: 0975-766X Available Online through Research Article www.ijptonline.com SPECTROPHOTOMETRIC METHODS FOR THE DETERMINATION OF FROVATRIPTAN SUCCINATE MONOHYDRATE IN BULK AND PHARMACEUTICAL DOSAGE FORMS
More informationEfficient Measurement of Large Light Source Near-field Color and Luminance Distributions for Optical Design and Simulation
Efficient Measurement of Large Light Source Near-field Color and Luminance Distributions for Optical Design and Simulation Hubert Kostal*, Douglas Kreysar, Ronald Rykowski Radiant Imaging, Inc., 22908
More informationSound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound
CH.12. Dr.Rajaa Sound in medicine أ.م.د. رجاء سهيل جنم جامعة تكريت كلية Sound : It is the audible waves of frequency between 20 Hz and 20 khz. Infrasound : refers to the sound of frequency below the normal
More informationCompetitive Intelligence Report Technology & Patents
Competitive Intelligence Report Technology & Patents N o n i n v a s i v e d i a b e t e s m o n i t o r i n g 01 November - 15 November, 2018 7 Published Granted Inactive 5 8 0 25 2 1 Start ups Companies
More informationSAXS on lipid structures
Practical Course in Biophysics, Experiment R2b SAXS on lipid structures Summer term 2015 Room: Advisor: X-ray lab at LS Rädler, NU111 Stefan Fischer Tel: +49-(0)89-2180-1459 Email: stefan.f.fischer@physik.lmu.de
More informationSupporting Information
Supporting Information An efficient broadband and omnidirectional light-harvesting scheme employing the hierarchical structure based on ZnO nanorod/si 3 N 4 -coated Si microgroove on 5-inch single crystalline
More informationDevelopment of near-infrared absorption spectrometry system by using NIR wideband glass phosphor LED
Journal of Physics: Conference Series PAPER OPEN ACCESS Development of near-infrared absorption spectrometry system by using NIR wideband glass phosphor LED Recent citations - Luminescence properties of
More informationLow-Pressure Chromium Speciation in P2O5 using the Spectro ARCOS MV ICP and the ESI SC-DX chromfast System
Low-Pressure Chromium Speciation in P2O5 using the Spectro ARCOS MV ICP and the ESI SC-DX chromfast System Dion Tsourides AAFPCO Meeting Albuquerque, NM (Acknowledgements ESI) Introduction Chromium exists
More informationFT-Raman Surface Mapping of Remineralized Artificial Dental Caries
FT-Raman Surface Mapping of Remineralized Artificial Dental Caries I. Stangel, 1 R. Rubinovitz, 2 G.D. Arndt, 3 D. Byerly, 3 C. Theriot, 3 J.E. Kerr, 4 1 BioMat Sciences, Bethesda, MD; 2 Thermo Fisher
More informationUltrasound Measurements and Non-destructive Testing Educational Laboratory
Session 3548 Ultrasound Measurements and Non-destructive Testing Educational Laboratory Vladimir Genis, Horacio Sosa Goodwin College of Professional Studies, Drexel University, Philadelphia, 19104 Emil
More informationMEDAK DIST. ANDHRA PRADESH STATE, INDIA. Research Article RECEIVED ON ACCEPTED ON
Page67 Available Online through IJPBS Volume 1 Issue 2 APRIL- JUNE 2011 SIMPLE QUANTITATIVE METHOD DEVELOPMENT AND VALIDATION OF VALSARTAN IN PUREFORM AND PHARMACEUTICAL DOSAGE FORMS BYUV SPECTROSCOPY
More informationClinically Available Optical Topography System
Clinically Available Optical Topography System Clinically Available Optical Topography System 18 Fumio Kawaguchi Noriyoshi Ichikawa Noriyuki Fujiwara Yûichi Yamashita Shingo Kawasaki OVERVIEW: Progress
More informationSoy Lecithin Phospholipid Determination by Fourier Transform Infrared Spectroscopy and the Acid Digest/Arseno-Molybdate Method: A Comparative Study
Soy Lecithin Phospholipid Determination by Fourier Transform Infrared Spectroscopy and the Acid Digest/Arseno-Molybdate Method: A Comparative Study J.M. Nzai and A. Proctor* Department of Food Science,
More informationA Snapshot on Nuclear Cardiac Imaging
Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the
More informationAn active unpleasantness control system for indoor noise based on auditory masking
An active unpleasantness control system for indoor noise based on auditory masking Daisuke Ikefuji, Masato Nakayama, Takanabu Nishiura and Yoich Yamashita Graduate School of Information Science and Engineering,
More informationDetermination of Free Fatty Acids in Crude Palm Oil and Refined-Bleached-Deodorized Palm Olein Using Fourier Transform Infrared Spectroscopy
Determination of Free Fatty Acids in Crude Palm Oil and Refined-Bleached-Deodorized Palm Olein Using Fourier Transform Infrared Spectroscopy Y.B. Che Man a, *, M.H. Moh a, and F.R. van de Voort b a Department
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationUV Spectrophotometric Estimation of Alprazolam by Area Under Curve And First Order Derivative Methods in Bulk and Pharmaceutical Dosage Form
Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2016, 8 (5):105-110 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4
More information7 Grip aperture and target shape
7 Grip aperture and target shape Based on: Verheij R, Brenner E, Smeets JBJ. The influence of target object shape on maximum grip aperture in human grasping movements. Exp Brain Res, In revision 103 Introduction
More informationfor Heart-Health Scanning
Magnetocardiography for Heart-Health Scanning CardioMag Imaging, Inc. 1 Basic Principles of Magnetocardiography (MCG) The cardiac electric activity that produces a voltage difference on the body surface
More informationDevelopment and application of novel spectroscopic tools for breast cancer diagnosis
Development and application of novel spectroscopic tools for breast cancer diagnosis LBRC researchers: Ramachandra Dasari, Jeon Woong Kang, Niyom Lue, Rishikesh Pandey, Nicolas Spegazzini External technology
More informationX ray Spectra and Peak Power Control with isase. J. Wu (SLAC) and C. Pellegrini (UCLA/SLAC) May 15, 2013
X ray Spectra and Peak Power Control with isase J. Wu (SLAC) and C. Pellegrini (UCLA/SLAC) May 15, 2013 OUTLINE Review SASE FEL Short longitudinal coherent length leads to spiky temporal and spectral profiles
More informationDESIGN OF AUTOMATIC INSULIN DOSAGE INDICATOR FOR DIABETIC PATIENTS USING BMI
DESIGN OF AUTOMATIC INSULIN DOSAGE INDICATOR FOR DIABETIC PATIENTS USING BMI Ms. R. Kanimozhi PG Scholar Dr. S. Saravanakumar Assistant Professor Department of Electronics and Communication Engineering,
More informationCompact Gamma Camera for Detection of Prostate Cancer
Compact Gamma Camera for Detection of Prostate Cancer Presented at: Human Interest Panel Federal Laboratory Consortium Annual Conference Nashville, Tennessee Brookhaven National Laboratory and Hybridyne
More informationTemperature Effect on Near Infrared Spectrum of Glucose Monitoring
206 International Conference on Electronic Information Technology and Intellectualization (ICEITI 206) ISBN: 978--60595-364-9 Temperature Effect on Near Infrared Spectrum of Glucose Monitoring Renjie Xu,
More informationImproving the diagnostic power of otoacoustic emissions. Arturo Moleti Physics Department University of Roma Tor Vergata
Improving the diagnostic power of otoacoustic emissions Arturo Moleti Physics Department University of Roma Tor Vergata The human ear Ear canal: resonant cavity Middle ear: impedance adapter and pressure
More informationMultimodal Spectroscopic Tissue Scanner for diagnosis of ex vivo surgical specimens
Multimodal Spectroscopic Tissue Scanner for diagnosis of ex vivo surgical specimens Collaborating researcher: Dr. Maryann Fitzmaurice (Case Western Reserve University, Cleveland, OH), Dr. Tulio Valdez
More informationFor more information, please contact: or +1 (302)
Introduction Quantitative Prediction of Tobacco Components using Near-Infrared Diffuse Reflectance Spectroscopy Kristen Frano Katherine Bakeev B&W Tek, Newark, DE Chemical analysis is an extremely important
More informationCHAPTER 6. BLUE GREEN AND UV EMITTING ZnO NANOPARTICLES SYNTHESIZED THROUGH A NON AQUEOUS ROUTE
71 CHAPTER 6 BLUE GREEN AND UV EMITTING ZnO NANOPARTICLES SYNTHESIZED THROUGH A NON AQUEOUS ROUTE 6.1 INTRODUCTION Several techniques such as chemical vapour deposition, electrochemical deposition, thermal
More informationSYNAPT G2-S High Definition MS (HDMS) System
SYNAPT G2-S High Definition MS (HDMS) System High performance, versatility, and workflow efficiency of your MS system all play a crucial role in your ability to successfully reach your scientific and business
More information4 Development of an ESR online-method for the monitoring of in vitro fat digestion
4 Development of an ESR online-method for the monitoring of in vitro fat digestion 4.1 Introduction When regarding the oral administration of lipid-based nanocapsules, gastrointestinal digestion will play
More informationTHE RIGAKU JOURNAL VOL. 23 / 2006, A1-A10 CONSIDERATIONS REGARDING THE ALIGNMENT OF DIFFRACTOMETERS FOR RESIDUAL STRESS ANALYSIS
THE RIGAKU JOURNAL VOL. 23 / 2006, A1-A10 CONSIDERATIONS REGARDING THE ALIGNMENT OF DIFFRACTOMETERS FOR RESIDUAL STRESS ANALYSIS THOMAS R. WATKINS, O. BURL CAVIN, CAMDEN R. HUBBARD, BETH MATLOCK, AND ROGER
More informationAssessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection
Proceedings of the 2012 International Conference on Industrial Engineering and Operations Management Istanbul, Turkey, July 3 6, 2012 Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter
More informationJournal of Chemical and Pharmaceutical Research, 2014, 6(3): Research Article
Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2014, 6(3):1356-1360 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Spectral discrimination of gastric cancer and
More informationSUPPLEMENTARY INFORMATION
Terabit free-space data transmission employing orbital angular momentum multiplexing Jian Wang 1, 2 *, Jeng-Yuan Yang 1, Irfan M. Fazal 1, Nisar Ahmed 1, Yan Yan 1, Hao Huang 1, Yongxiong Ren 1, Yang Yue
More informationSpectral Imaging as a Modern Tool for Medical Diagnostics
Spectral Imaging as a Modern Tool for Medical Diagnostics Grade 3 Grade 2 Issa Ibraheem Andreas Kercek CTR Carinthian Tech Research AG A - 9524 Villach / Austria Grade 1 Grade 0 Healthy Contents Spectral
More informationFERMI: EUV and Soft X-Ray FELs with HGHG
1 FERMI: EUV and Soft X-Ray FELs with HGHG E. Allaria on behalf of the FERMI commissioning team Enrico Allaria Outline 2 Elettra and the FERMI FEL project FERMI parameters FEL-1 experimental results FEL
More informationDual channel photoplethysmography studies of cardio-vascular response to the body position changes
Dual channel photoplethysmography studies of cardio-vascular response to the body position changes R. Erts, I. Kukulis, J. Spigulis, L. Kere University of Latvia, Raina Blvd. 9, Riga, LV-, Latvia Renars.erts@lu.lv
More informationEUEC 2015 Experience using Cavity Enhanced Absorption Spectroscopy (CEAS) for HCl measurement on a coal-fired power plant
Steve Gibbons, Business Development - CEMS EUEC 2015 Experience using Cavity Enhanced Absorption Spectroscopy (CEAS) for HCl measurement on a coal-fired power plant Introduction Steve Gibbons Business
More informationSynchrotron Radiation Infrared Microscopy Analysis of Amyloid Fibrils in Alzheimer s Disease Model Mouse Brain Tissue Takayasu Kawasaki 1, Toyonari Ya
Synchrotron Radiation Infrared Microscopy Analysis of Amyloid Fibrils in Alzheimer s Disease Model Mouse Brain Tissue Takayasu Kawasaki 1, Toyonari Yaji 2, Koichi Tsukiyama 1, Toshiaki Ohta 2, and Kazuhiro
More informationPosterREPRINT EVALUATION OF A RF-ONLY STACKED RING BASED COLLISION CELL WITH AXIAL FIELD FOR THE LC-MS-MS ANALYSIS OF THE METABOLITES OF RABEPRAZOLE
OVERVIEW Comparison of the performance of a stacked ring based collision cell with axial field and a hexapole collision cell. Full scan LC-MS/MS-MS using precursor, product and neutral loss modes of operation
More informationQuadrupole and Ion Trap Mass Analysers and an introduction to Resolution
Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution A simple definition of a Mass Spectrometer A Mass Spectrometer is an analytical instrument that can separate charged molecules according
More informationINNOVATIONS IN SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY
INNOVATIONS IN SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY M.S. Maxim Antakov, Igor Burnaevsky National Research University Moscow Institute of Electronic Technology Biomedical Systems Department Moscow
More informationIES LM Test Report
IES LM 79 08 Test Report Test results reported for: B 3 DL report: LEAG002 010 Original issue date: 01 May 2012 Prepared for: Testing performed by: The Leader Leader Mfg., Inc 1003 7th Ave. 1540 Leader
More informationCOMPARISON OF DISPERSIVE AND FOURIER-TRANSFORM NIR INSTRUMENTS FOR MEASURING GRAIN
COMPARISON OF DISPERSIVE AND FOURIER-TRANSFORM INSTRUMENTS FOR MEASURING GRAIN AND FLOUR ATTRIBUTES P. R. Armstrong, E. B. Maghirang, F. Xie, F. E. Dowell ABSTRACT. Dispersive and Fourier-transform (FT)
More informationQuality Analysis of Reheated Oils by Fourier Transform Infrared Spectroscopy
International Conference on Electromechanical Control Technology and Transportation (ICECTT 05) Quality Analysis of Reheated Oils by Fourier Transform Infrared Spectroscopy Keying Zhao, a, Lei Shi,b and
More informationFood protein powders classification and discrimination by FTIR spectroscopy and principal component analysis
APPLICATION NOTE AN53037 Food protein powders classification and discrimination by FTIR spectroscopy and principal component analysis Author Ron Rubinovitz, Ph.D. Thermo Fisher Scientific Key Words FTIR,
More informationSPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS
SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS Vessela Tzvetanova Krasteva, Irena Ilieva Jekova Centre of Biomedical Engineering Prof. Ivan Daskalov - Bulgarian Academy of Sciences Acad.G.Bonchev
More informationScienceDirect. Comparative Study of Different Measurement Sites using NIR Based Non-invasive Glucose Measurement system
Available online at www.sciencedirect.com ScienceDirect Procedia Computer Science 70 (2015 ) 469 475 4 th International Conference on Eco-friendly Computing and Communication Systems (ICECCS) Comparative
More informationFREQUENCY COMPRESSION AND FREQUENCY SHIFTING FOR THE HEARING IMPAIRED
FREQUENCY COMPRESSION AND FREQUENCY SHIFTING FOR THE HEARING IMPAIRED Francisco J. Fraga, Alan M. Marotta National Institute of Telecommunications, Santa Rita do Sapucaí - MG, Brazil Abstract A considerable
More informationResearch Journal of Pharmaceutical, Biological and Chemical Sciences
Research Journal of Pharmaceutical, Biological and Chemical Sciences Oxygen Glucose and Heart rate Monitor: An non Invasive approach Anne Frank Joe A*, Veeramani Saikrupa, and Prajukti Panda Sathyabama
More informationLinkam Scientific Instruments
FTIR600 - Systems for IR Microscopy and Spectroscopy FT-IR is an increasingly popular method of sample analysis. With today s highly sophisticated instruments combining research quality optical features
More information2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry
Dr. Sanjeeva Srivastava 1. Fundamental of Mass Spectrometry Role of MS and basic concepts 2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry 2 1 MS basic concepts Mass spectrometry - technique
More informationDETECTION OF HEART ABNORMALITIES USING LABVIEW
IASET: International Journal of Electronics and Communication Engineering (IJECE) ISSN (P): 2278-9901; ISSN (E): 2278-991X Vol. 5, Issue 4, Jun Jul 2016; 15-22 IASET DETECTION OF HEART ABNORMALITIES USING
More informationPhotoacoustic Imaging and Therapy in Biomedicine. Nicholas Tobey and Grace Yook. Optical Engineering. Dr. Kasra Daneshvar
Photoacoustic Imaging 1 Photoacoustic Imaging and Therapy in Biomedicine Nicholas Tobey and Grace Yook Optical Engineering Dr. Kasra Daneshvar July 16, 2010 Photoacoustic Imaging 2 Abstract When a pulsed
More informationRaman Spectroscopy and imaging to explore skin and hair. What is the Raman scattering effect?
Diapositive 1 1 Raman Spectroscopy and imaging to explore skin and hair Sir C.V. Raman Nobel price in physics in 193 1 µm Diapositive 2 What is the Raman scattering effect? Incident monochromatic beam
More informationUnderwater Acoustic Measurements in Megahertz Frequency Range.
Underwater Acoustic Measurements in Megahertz Frequency Range. Current State and Prospects of Development in Russia Alexander M. Enyakov,, Many medical applications of underwater acoustic measurements
More informationCharacterization of skin tissue heterogeneity with near-infrared microspectroscopy and its effects on noninvasive measurements of glucose
University of Iowa Iowa Research Online Theses and Dissertations Fall 2011 Characterization of skin tissue heterogeneity with near-infrared microspectroscopy and its effects on noninvasive measurements
More informationAn electrooptical muscle contraction sensor. Alessio Chianura and Mario E Giardini. Medical and Biological Engineering and Computing 48 (2010)
An electrooptical muscle contraction sensor Alessio Chianura and Mario E Giardini Medical and Biological Engineering and Computing 48 (2010) 731-734 This is an author-generated version The final publication
More informationJuan Carlos Tejero-Calado 1, Janet C. Rutledge 2, and Peggy B. Nelson 3
PRESERVING SPECTRAL CONTRAST IN AMPLITUDE COMPRESSION FOR HEARING AIDS Juan Carlos Tejero-Calado 1, Janet C. Rutledge 2, and Peggy B. Nelson 3 1 University of Malaga, Campus de Teatinos-Complejo Tecnol
More informationFeng Xiujuan National Institute of Metrology (NIM),China
The acoustic calibration service in transportation at NIM Feng Xiujuan National Institute of Metrology (NIM),China 1. Calibration requirements 2. Calibration service at NIM 2.1 Microphone 2.2 Ultrasonic
More information