Performance characteristics of the Agilent 1100 Series preparative pump. Technical Note. Abstract

Size: px
Start display at page:

Download "Performance characteristics of the Agilent 1100 Series preparative pump. Technical Note. Abstract"

Transcription

1 Performance characteristics of the Agilent Series preparative pump Technical Note Abstract The Agilent Series preparative pump as part of the Agilent Series purification system is especially designed for flow rates up to ml/min at a maximum backpressure of 4 bar. In this Technical Note we show the performance characteristics of the preparative pump for flow rates ranging from below ml/min up to ml/min.

2 Introduction The Agilent Series preparative pump is an isocratic high performance pump with two parallel pistons. It offers flow rates up to ml/min at 4 bar without the need to change pump heads. Two pumps are connected via a lowvolume mixing device to give a high-pressure gradient mixing system with an internal delay volume of only about.7 ml. Additional features of the pump include an automated seal wash using a peristaltic pump and an automated electro-magnetic purge valve. Although the pumping system is designed for flow rates up to ml/min it also shows very good performance at lower flow rates 2. In this Technical Note we show the performance characteristics of a binary gradient system over a flow rate range starting below ml/min up to ml/min. The following performance parameters were measured: composition ripple, composition precision, composition accuracy, composition linearity, flow accuracy, flow precision (isocratic and gradient performance), and precision of retention times. Note: All experiments were carried out on a system that is used for the daily work in Agilent s laboratory. Depending on the working hours of your system the results for the tests may vary slightly. Equipment All experiments were performed on an Agilent Series system containing the following modules: Two Agilent Series preparative pumps Agilent Series preparative autosampler Agilent Series diode array detector The system was controlled using the Agilent ChemStation (rev. A.9.). Results. Composition ripple, precision, accuracy and linearity To perform reproducible separations based on gradient runs it is important that the pump mixes the solvents accurately and precisely. For fast and direct method scale-up it is important that the gradient formation is consistently good over the complete flow rate range. If the preparative pumps are used in an Agilent Series purification system precise retention times from run to run are especially important for timebased fraction collection or peakbased fraction collection over a certain time interval of the purification run. Gradient composition test set-up Method and data analysis from the gradient composition test of the ChemStation OQ/PV was used to determine the parameters composition ripple, composition precision, composition accuracy and composition linearity. Solvent A was water, solvent B was water with.5 or 5 % acetone as tracer. The test was performed at flow rates of, 2, 5,, 25, 5 and ml/min. Different flow cells with pathlengths of, 3 and.3 mm were used in the diode array detector. A restriction capillary was used for all experiments to keep the backpressure at about bar. The autosampler was removed from the system for the composition ripple, composition precision, composition accuracy and composition linearity tests. Table gives an overview of the gradient composition test parameters. Flow cell [mm] Backpressure [bar] % Acetone Table Test parameters for composition, ripple, precision, accuracy and linearity tests 2

3 Results of gradient composition test Figure shows the results for the composition ripple, accuracy, precision and linearity tests. The marked areas show the limits as specified in the ChemStation OQ/PV for the Agilent Series quaternary pump. All values are within the limits except for the composition ripple. The variances in the values are due to different pressures, flow cell pathlengths and acetone concentration. This shows that the pump stays within the limits even if these parameters are changed. The performance criterion of the composition ripple is not fulfilled due to the missing pressure damper of the gradient pumping system. A pressure damper was not built into the preparative pump to keep the delay volume as small as possible. As mentioned before, the experiments were carried out with a restriction capillary instead of a chromatographic column. A preparative column in the system works as a dampening mechanism and reduces the ripple drastically. Therefore, a composition ripple experiment was repeated using a column instead of a restriction capillary. Table 2 shows that the ripple is reduced by a factor of when using a column appropriate for the flow rate. Note: The gradient composition test must never be carried out with a column. A restriction capillary must be used. The experiment with the column was only done to show the influence of the column as a dampening mechanism in a system containing the preparative pumps. Flow rate Ripple Flow cell Backpressure % Acetone [ml/min] [% B] [mm] [bar] Restriction capillary Column (2.2 5 mm) Table 2 Composition ripple with restriction capillary and with column Limit < Ripple [% B] Accuracy [% B] 2 Limit < Limit < Precision [% B] Limit < Correlation Figure Gradient composition test showing composition ripple, accuracy, precision and linearity (correlation) 3

4 2. Flow accuracy Flow accuracy is a parameter which is not important for run-torun precision, however, it is important for proper method performance after scale-up experiments. Flow accuracy depends on the system backpressure and, on the compensation algorithms of the preparative pump firmware ml/min 5 ml/min 25 ml/min 5 ml/min ml/min Flow accuracy Flow accuracy test set-up The flow accuracy was measured by collecting the solvent (water) over five minutes. The collected volume was determined by weighing the collected water and dividing the amount by the density. The test was performed at flow rates of, 2, 5,, 25, 5 and ml/min. The accuracy for each flow rate was measured at three different backpressures (< bar, > and < 2 bar, > 2 bar). Restriction capillaries were used instead of a chromatographic column for all experiments. Results of flow accuracy test Figure 2 shows the results of the flow accuracy tests. The flow accuracy is below 5 % for all flow rates and pressure settings except for ml/min and high backpressure. This ensures fast and easy scale-up over the complete flow rate range without time-consuming method redevelopment as shown in an earlier Application Note Flow precision isocratic Flow precision is important for run-to-run precision of retention times. Since the pump flow is controlled by the compensation algorithms of the preparative pump firmware the flow precision is backpressure dependent. Flow precision test set-up The flow precision was measured by collecting the solvent (water) over 5 minutes. The collected volume was determined by weighing the collected water and dividing the amount by the density. The test was performed at a flow rate of 25 ml/min. The precision was measured at three different backpressures (< bar, > and < 2 bar, > 2 bar). Pressure [bar] Figure 2 Flow accuracy for different flow rates and backpressures Five runs were performed at each pressure to determine the precision. Restriction capillaries were used instead of a chromatographic column for all experiments. Results of flow precision test Figure 3 shows the results for the isocratic flow precision tests. The relative standard deviation of the flow precision is marked. The relative standard deviation is below.3 % for the different backpressures. 4

5 4. Flow precision gradient Flow precision test set-up The flow precision was measured by collecting the solvent (water, methanol) over seven minutes. The collected volume was measured. The test was performed at a flow rate of ml/min. Gradient: % water for min, % water to % MeOH in 5 min, % MeOH for min. The precision was measured at three different backpressure ranges. Five runs were performed at each pressure to determine the precision. Restriction capillaries were used for all experiments instead of a chromatographic column. Flow [ml/min] Flow Precision RSD.2 % RSD.2 % Pressure [bar] RSD.9 % Figure 3 Relative standard deviation of the isocratic flow precision at different backpressure settings Results of flow precision test Figure 4 shows the results of the gradient flow precision tests. The relative standard deviation of the flow precision, shown with the black error bars, is below.4 % for the different backpressure settings. Flow [ml/min] RSD.25 % Flow Precision RSD. % RSD.4 % Max. pressure [bar] Figure 4 Relative standard deviation of the gradient flow precision at different backpressure settings 5

6 5. Precision of retention time Precision of retention time is an important characteristic that influences qualitative and quantitative results in analytical HPLC. In preparative HPLC this parameter is important for repetitive purification runs using time-based fraction collection. The retention times have to be stable within a certain time window. Precision of retention time test setup Precision of retention times was measured by ten repetitive injections from the same sample containing two major and one minor compound (figure 5). The precision of retention time was calculated using the Chem- Station sequence summary report. The experiments were carried out on columns with inner diameters of 3, 4.6, 9.4, 2.2 and 5 mm with flow rates of.35,.85, 3.5, 8 and ml/min, respectively. Mobile phase A was water, mobile phase B was methanol Gradient: 5 % B to 25 % B in min, 25 % B for 9.9 min, 25 % B to 5 % B in. min. Figure 6 shows the relative standard deviation of the retention times measured for consecutive runs for a sample that contained three compounds. For flow rates above ml/min and up to ml/min the relative standard deviation of the retention times is.3 % or below, which is excellent for a preparative pump. mau Flow rate: 3.5 ml/min consecutive runs Time [min] Figure 5 Precision of retention time test at 3.5 ml/min RSD [%] Precision of Retention Time. Figure 6 Relative standard deviation for consecutive runs Peak Peak 2 Peak 3 6

7 Conclusion In this Technical Note we showed the performance of the Agilent Series preparative pump over the flow rate range of.35 up to ml/min. We showed that the performance below 5 ml/min is almost as good as for an analytical pump, for example, the Agilent Series quaternary pump. Although the pump performance is still very good at such low flow rates, other parts of the system, especially the capillaries with a large inner diameter required for higher flow rates, will decrease the overall system performance. Therefore, Agilent Technologies offers dedicated systems from capillary to preparative scale rather than a single system covering the complete flow rate range. This ensures best performance for a specific application rather than low performance over a wide application range. References. New perspectives in purification with HPLC and HPLC/MS, Agilent Technolgies Brochure, 2, publication number EN 2. Performance of the Agilent Series preparative pump at low flow rates", Agilent Technologies Technical Note, 2, publication number EN 3. Scale-up from an analytical scale to a preparative scale method with the Agilent Series purification system PS, Agilent Technologies Application Note, 22, publication number EN 7

8 Copyright 22 Agilent Technologies All Rights Reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed under the copyright laws. Published June, 22 Publication Number EN

Chiral Multicolumn Method Development on the Agilent 1260 Infinity II SFC System

Chiral Multicolumn Method Development on the Agilent 1260 Infinity II SFC System Chiral Multicolumn Method Development on the Agilent 126 Infinity II SFC System Application Note Small Molecule Pharmaceuticals Author Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Abstract

More information

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate Application Note Pharmaceutical QA/QC Author Siji Joseph Agilent Technologies, Inc. Bangalore,

More information

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System Application Note Food Testing and Agriculture Mass-Based Purification of Natural Product Impurities Using an Agilent 126 Infinity II Preparative LC/MSD System Authors Florian Rieck and Jörg Hippler Agilent

More information

Agilent 1260 Infinity Analytical SFC System

Agilent 1260 Infinity Analytical SFC System Agilent 1260 Infinity Analytical SFC System Installation Instructions: Checkout Procedure This technical note describes the checkout procedures for the Agilent 1260 Infinity Analytical SFC System. Contents

More information

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns Maximizing Efficiency Using Agilent InfinityLab Poroshell 12 Columns 1, Plates in Less Than Minutes Using Coupled Column Technology Application Note Food, Environmental, Chemical, Pharmaceutical Authors

More information

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine- Analysis of Pharmaceutical Substances Using HPLC and UHPLC Methods Backward Compatibility for Validated Analytical Procedures and Methods Comparing the Infi nity LC and the Infi nity II LC Application

More information

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System Application Note Small Molecule Pharmaceutical Chiral Screening for SFC and UHPLC with the Agilent 16 Infinity II SFC/UHPLC Hybrid System Author Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

More information

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 1 PFP Application Note Clinical Research Author Rongjie Fu Agilent Technologies (Shanghai) Co. Ltd Abstract

More information

Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat

Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat Performance Characteristics of the Agilent 9 Infinity II Multicolumn Thermostat Technical Overview Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview

More information

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC Application Note Environmental Author Melanie Metzlaff Agilent Technologies, Inc. Waldbronn, Germany Abstract

More information

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system A practical guide on how to use parameters to increase peak capacity Application Note Pharmaceutical and Chemical Author

More information

Ultrafast analysis of synthetic antioxidants in vegetable oils using the Agilent 1290 Infinity LC system

Ultrafast analysis of synthetic antioxidants in vegetable oils using the Agilent 1290 Infinity LC system Ultrafast analysis of synthetic antioxidants in vegetable oils using the Agilent 19 Infinity LC system Application Note Food Authors Standard solution Gerd Vanhoenacker, Frank David, Pat Sandra Research

More information

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice Application Note Author Food Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India 8 6 4 2

More information

Agilent 1260 Infinity Analytical Hybrid SFC/UHPLC system

Agilent 1260 Infinity Analytical Hybrid SFC/UHPLC system Agilent 1260 Infinity Analytical Hybrid SFC/UHPLC system Shared use of Columns - Technical Note This technical note describes the methods required to perform automated column switching in SFC/LC and SFC/LC/MS

More information

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC Analysis of Counterfeit Antidiabetic Drugs by UPLC with the Agilent 1 Infinity Mobile LC Application Note Small Molecule Pharmaceuticals & Generics Author Sonja Schneider Agilent Technologies, Inc. Waldbronn,

More information

Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution

Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution Application Note Food Testing Author Edgar Naegele

More information

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification Application ote Food Testing and Agriculture Analysis of Vitamins Using an SF/UPL ybrid System with a Triple Quadrupole L/MS for Quantification ounts 4.6.4...8.6.4....4.6.8...4.6.8 Acquisition time (min

More information

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil Application Note Authors Xueqi Li, Selina Wang UC Davis Olive Center Robert

More information

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application Pharmaceutical Author Cliff Woodward Agilent Technologies, Inc. 285 Centerville

More information

USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC

USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC Application Note Manufacturing QA/QC Authors Syed S. Lateef, Siji Joseph Agilent Technologies Bangalore, India 500 400 Pravastatin

More information

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column Application Note Pharmaceutical and Food Testing Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column Author Lu Yufei Agilent Technologies, Inc. Abstract A liquid chromatographic

More information

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column APPLICATION NOTE 21673 Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column Authors Aaron Lamb and Brian King, Thermo Fisher Scientific, Runcorn, UK

More information

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 129 Infinity LC System Application Note Pharmaceutical and Chemical Analysis

More information

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages APPLICATION NOTE 21671 Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages Authors Jon Bardsley, Thermo Fisher Scientific, Runcorn, UK Keywords Vanquish Flex, Acclaim PolarAdvantage

More information

Agilent 1220 Infinity LC System Performance Specifications

Agilent 1220 Infinity LC System Performance Specifications IET International Equipment Trading, Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 1220 Infinity LC System Performance

More information

Application Note. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling. Abstract.

Application Note. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling. Abstract. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling Application Note Food 1 Agilent 12 Infinity 1 Author Siji Joseph Agilent Technologies, Inc. Bangalore,

More information

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS FORENSICS AND TOXICOLOGY ANALYSIS DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS Solutions for Your Analytical Business Markets and Applications

More information

QuickSplitTM Flow Splitters

QuickSplitTM Flow Splitters QuickTM Flow ters between 2 or more detectors such as Mass Spectroscopy and evaporative light scattering, UV, or IR. This is especially important when one of the detectors is destructive, such as Mass

More information

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns Converting a CHP Method for Insulin to Agilent Poroshell Columns Application Note Biopharm Author Rongjie Fu Agilent Technologies (Shanghai) Co., Ltd. 4 Ying Lun Road Shanghai 3 China Abstract A regulatory

More information

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer A Robustness Study for the Agilent 7 LC-MS/MS Mass Spectrometer Application Note Clinical Research Authors Linda Côté, Siji Joseph, Sreelakshmy Menon, and Kevin McCann Agilent Technologies, Inc. Abstract

More information

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products High-Performance Anion-Exchange Chromatography coupled with Pulsed Electrochemical Detection as a powerful tool to evaluate lactose content in lactose-free labeled products Application Note Authors: C.

More information

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 9 Infinity II/63B LC/TOF system Application Brief Authors Mike Adams, Karen Kaikaris, and A. Roth CWC Labs Joan Stevens, Sue D

More information

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Keywords Introduction Application Note PHA0413 High Pressure Liquid Chromatography

More information

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract JJC Jordan Journal of Chemistry Vol. 4 No.4, 2009, pp. 357-365 Development and Validation of Analytical Method for Fluconazole and Fluconazole Related Compounds (A, B, and C) in Capsule Formulations by

More information

Edgar Naegele. Abstract

Edgar Naegele. Abstract Simultaneous determination of metabolic stability and identification of buspirone metabolites using multiple column fast LC/TOF mass spectrometry Application ote Edgar aegele Abstract A recent trend in

More information

QuickSplit TM Flow Splitters

QuickSplit TM Flow Splitters Quick TM Flow ters between 2 or more detecrs such as Mass Spectroscopy and evaporative light scattering, UV, or IR. This is especially important when one of the detecrs is destructive, such as Mass Spec

More information

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit Application Note General Chromatography, Food Industry Authors John W. Henderson Jr., Judy Berry, Anne Mack, William Long Agilent

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(2):770-775 Validation of Rapid Liquid Chromatographic

More information

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE WORLD JOURNAL OF PHARMACY AND PHARMACEUTICAL SCIENCES Ramalakshmi et al. SJIF Impact Factor 6.647 Volume 7, Issue 2, 1010-1018 Research Article ISSN 2278 4357 METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC

More information

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW 132 CHAPTER 6 DEVELOPMENT AND VALIDATION OF A STABILITY-INDICATING RP-HPLC METHOD FOR SIMULTANEOUS DETERMINATION OF PARACETAMOL, TRAMADOL HYDROCHLORIDE AND DOMPERIDONE IN A COMBINED DOSAGE FORM 6.1 INTRODUCTION

More information

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079 Ultra High Performance Liquid Chromatograph Nexera X2 C196-E79 Maximizing the Potential of UHPLC/HPLC Analyses Intelligence Intelligent data solutions, including a new i -PDeA peak calculation method for

More information

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology Determination of Impurities and Related Substances for Acetylsalicylic Acid (Ph. Eur. Monograph 9): Increased Sensitivity and Resolution, Faster Analysis, and Reduced Solvent Usage Using Kinetex 2.6 µm

More information

Forensic Analysis of Blood Alcohol Concentration

Forensic Analysis of Blood Alcohol Concentration Application Note Forensics Forensic Analysis of Blood Alcohol Concentration Using the Agilent 886 GC with Agilent J&W DB BAC1 UI and Agilent J&W DB-BAC2 UI columns and the Agilent 7697A headspace sampler

More information

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies. Using Software Tools to Improve the Detection of Impurities Application Note Christine Miller Introduction The analysis of raw materials and finished products for or impurities presents a challenge in

More information

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement Application Note Food Testing & Agriculture Author Edgar Naegele Agilent Technologies,

More information

2D-LC as an Automated Desalting Tool for MSD Analysis

2D-LC as an Automated Desalting Tool for MSD Analysis 2D-LC as an Automated Desalting Tool for MSD Analysis Direct Mass Selective Detection of a Pharmaceutical Peptide from an MS-Incompatible USP Method Application Note Biologics and Biosimilars Author Sonja

More information

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets Journal of PharmaSciTech 0; ():- Research Article Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets * Sayyed Hussain,

More information

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 129 Infinity II LC Application Note Small Molecule Pharmaceuticals Authors Gerd Vanhoenacker, Mieke Steenbeke, Koen Sandra,

More information

ISSN: ; CODEN ECJHAO E-Journal of Chemistry 2011, 8(3),

ISSN: ; CODEN ECJHAO E-Journal of Chemistry  2011, 8(3), ISSN: 0973-4945; CODEN ECJHAO E- Chemistry http://www.e-journals.net 2011, 8(3), 1275-1279 Simultaneous Determination of Paracetamol, Phenylephrine Hydrochloride, Oxolamine Citrate and Chlorpheniramine

More information

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD)

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD) LIFE SCIENCES AR ENOUGH? HOW DO YOU KNOW RAW MATERIALS ARE PURE? HOW DO YOU EVALUATE PRODUCT PACKAGING? HOW DO YOU DO YOU KNOW WHAT ANALYTICAL TECHNIQUE TO USE? HOW DO YOU SIMULTANEOUSLY TEST FOR TWO BYPRODUCTS?

More information

Analysis of bisphenol A leaching from baby feeding bottles

Analysis of bisphenol A leaching from baby feeding bottles Agilent Application Solution Analysis of bisphenol A leaching from baby feeding bottles Application Note Consumer Products Author Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India Abstract

More information

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS Application Note Clinical Research Authors Frank David and Bart Tienpont, Research Institute for Chromatography,

More information

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection Application Note Cannabis Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection Authors Mike Adams, Annette Roth, Sue D

More information

Development and validation of liquid chromatographic method for trazodone hydrochloride

Development and validation of liquid chromatographic method for trazodone hydrochloride Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2012, 4 (6):1657-1664 (http://scholarsresearchlibrary.com/archive.html) ISS 0975-5071 USA CODE: DPLEB4

More information

STANDARD OPERATING PROTOCOL (SOP)

STANDARD OPERATING PROTOCOL (SOP) 1 STANDARD PERATING PRTCL (SP) Subject: To determine the Contents of isoflavone in soybean products by HPLC Analysis Project/Core No.: Core B Total 9 Pages SP No.: CB0102 Modified Date: 05/15/01 BTANICAL

More information

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Dominic Foley, Michelle Wills, and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION

More information

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS. A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS. Martin R Green*, Keith Richardson, John Chipperfield, Nick Tomczyk, Martin

More information

InfinityLab Purification Capillary Kits

InfinityLab Purification Capillary Kits InfinityLab Purification Capillary Kits Technical Note This note describes the capillary connections for Agilent preparative systems. Agilent Technologies Contents Abbreviations 3 List of Modules 4 Overview

More information

Lipidomic Analysis by UPLC-QTOF MS

Lipidomic Analysis by UPLC-QTOF MS Lipidomic Analysis by UPLC-QTOF MS Version: 1 Edited by: Oliver Fiehn Summary Reagents and Materials Protocol Summary:Lipidomic analysis by UPLC-QTOF mass spectrometry Reagents and Materials: Reagent/Material

More information

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC METHOD APPROVALS Norvin Pharma Inc. Author Analytical Laboratory Approver Analytical Laboratory Group Leader Approver Manager Quality Control Chemistry

More information

Determination of enantiomeric purity of Timolol Maleate by Supercritical Fluid Chromatography

Determination of enantiomeric purity of Timolol Maleate by Supercritical Fluid Chromatography Determination of enantiomeric purity of Timolol Maleate by Supercritical Fluid Chromatography Dr. Damian Connolly, (PMBRC, WIT) & Mr. Adrian Marley (Allergan Pharmaceuticals, Westport) PMBRC 3 rd Analytical

More information

ANALYTICAL SCALE AGILENT LCMS SYSTEMS OPERATIONAL QUALIFICATION

ANALYTICAL SCALE AGILENT LCMS SYSTEMS OPERATIONAL QUALIFICATION ANALYTICAL SCALE AGILENT LCMS SYSTEMS OPERATIONAL QUALIFICATION Overview This document describes the test program for qualifying Agilent LCMS systems. The following tables list all MS source-specific OQ

More information

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column Application Note Clinical Research LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell EC C8 Column Authors Rongjie Fu and Zhiming Zhang Agilent Technologies

More information

Amudha S et al., Asian Journal of Pharmthiaceutical Technology & Innovation, 04 (21); 2016; Research Article

Amudha S et al., Asian Journal of Pharmthiaceutical Technology & Innovation, 04 (21); 2016; Research Article Asian Journal of Pharmaceutical Technology & Innovation ISSN: 2347-8810 Research Article Received on: 09-11-2016 Accepted on: 20-11-2016 Published on: 15-12-2016 Corresponding Author: * Amudha S, Dept.

More information

RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms

RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms Asian Journal of Chemistry Vol. 22, No. 7 (2010), 5067-5071 RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms A. LAKSHMANA RAO*, G. TARAKA RAMESH and J.V.L.N.S. RAO Department of Pharmaceutical

More information

Agilent 7700x ICP-MS

Agilent 7700x ICP-MS Agilent 7700x ICP-MS Metals analysis Mid ppt to low ppb detection range Option to detect in four different gas modes to gain optimum detection of multiple metals or interferences in a sample Autosampler

More information

Comparison of conventional HPLC with UPLC method for determination of albuterol sulfate and it s impurities in pharmaceutical formulation

Comparison of conventional HPLC with UPLC method for determination of albuterol sulfate and it s impurities in pharmaceutical formulation Oriental Journal of Chemistry Vol. 24(2), 537-544 (2008) Comparison of conventional HPLC with UPLC method for determination of albuterol sulfate and it s impurities in pharmaceutical formulation P.N. DALVI,

More information

Journal of Chemical and Pharmaceutical Research, 2013, 5(1): Research Article

Journal of Chemical and Pharmaceutical Research, 2013, 5(1): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2013, 5(1):180-184 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 A simple and sensitive RP-HPLC method for estimation

More information

application Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS

application Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS application LC/MS Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS Introduction Many kinds of natural colors are used in beverages,

More information

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW 51 CHAPTER 2 SIMULTANEOUS ESTIMATION OF PIOGLITAZONE, GLIMEPIRIDE AND GLIMEPIRIDE IMPURITIES IN COMBINATION DRUG PRODUCT BY A VALIDATED STABILITY-INDICATING RP-HPLC METHOD 2.1 INTRODUCTION OF DOSAGE FORM

More information

Fig.1. Denatonium benzoate (DB) chemical structure

Fig.1. Denatonium benzoate (DB) chemical structure ILIADE code 280 CLEN Method Determination of Denatonium Benzoate in Alcoholic Products by HPLC-UV 1 Scope The purpose of this method is verification of fulfilment of the legislative requirements on denatured

More information

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual Page 1 / 14 Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual ZV-3051-0200-15 200 Page 2 / 14 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST

More information

Extended Application Note

Extended Application Note Extended Application Note Comparison of 4 μm vs. 2.o Columns in Analysis of Drospirenone/Ethinyl Estradiol Tablet APP A-339 www.mtc-usa.com 1-732-578-1777 INTRODUCTION The formulation of drospirenone and

More information

Measuring Lipid Composition LC-MS/MS

Measuring Lipid Composition LC-MS/MS Project: Measuring Lipid Composition LC-MS/MS Verification of expected lipid composition in nanomedical controlled release systems by liquid chromatography tandem mass spectrometry AUTHORED BY: DATE: Sven

More information

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography Application Note Metabolomics Author Yuqin Dai Agilent Technologies, Inc. 31 Stevens Creek Blvd,

More information

Application Note. Separation of clindamycin phosphate and process impurities. Summary. Introduction. Category Pharmaceutical analysis Matrix

Application Note. Separation of clindamycin phosphate and process impurities. Summary. Introduction. Category Pharmaceutical analysis Matrix Application Note Separation of clindamycin phosphate and process impurities Category Pharmaceutical analysis Matrix Drugs Method UHPLC Keywords Antibiotics, process impurities, quality control Analytes

More information

Iodine HPLC Assay. Catalog Number: IOD34-H Tests For Research Use Only. Not for use in diagnostic procedures.

Iodine HPLC Assay. Catalog Number: IOD34-H Tests For Research Use Only. Not for use in diagnostic procedures. Package Insert Iodine HPLC Assay 100 Tests For Research Use Only. Not for use in diagnostic procedures. v. 1.0 Eagle Biosciences, Inc. 20A Northwest Blvd., Suite 112, Nashua, NH 03063 Phone: 866-419-2019

More information

Application Note. Abstract. Author. Biotherapeutics & Biosimilars. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Abstract. Author. Biotherapeutics & Biosimilars. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany Sensitive and Reproducible Glycan Analysis of Human Immunoglobulin G The Agilent 1260 Infi nity Bio-inert Quaternary LC System with an Agilent AdvanceBio 2.7 µm Glycan Mapping Column and Fluorescence Detection

More information

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS Int. J. Chem. Sci.: 10(4), 2012, 2039-2047 ISSN 0972-768X www.sadgurupublications.com DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(1):138-144 Simultaneous RP HPLC determination of Latanoprost

More information

VITAMINES A / E IN PLASMA BY UV - FAST CODE Z18610

VITAMINES A / E IN PLASMA BY UV - FAST CODE Z18610 VITAMINES A / E IN PLASMA BY UV - FAST CODE Z18610 BIOCHEMISTRY Vitamins A and E are two fat-soluble vitamins indispensable to animals life for a correct growth and development. Vitamin A exists in three

More information

Validation of Quantitative Method for Glycidol Fatty Acid Esters (GEs) in Edible Oils

Validation of Quantitative Method for Glycidol Fatty Acid Esters (GEs) in Edible Oils 102nd AOCS Annual Meeting & Expo Validation of Quantitative Method for Glycidol Fatty Acid Esters (GEs) in Edible Oils Hiroki Shiro, Naoki Kondo and Yoshinori Masukawa * Tochigi Research Labs Kao corporation

More information

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page: Research Article CODEN: AJPAD7 ISSN: 2321-0923 Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page: www.ajpamc.com ANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF GEFITINIB

More information

FLUNITRAZEPAM Latest Revision: January 24, 2006

FLUNITRAZEPAM Latest Revision: January 24, 2006 FLUNITRAZEPAM Latest Revision: January 24, 2006 1. SYNONYMS CFR: Flunitrazepam CAS #: 1622-62-4 Other Names: 5-(2-Fluorophenyl)-1,3-dihydro-1-methyl-7-nitro-2H- 1,4-benzodiazepin-2-one Flunitrax Hipnosedon

More information

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ FORENSICS MARKET A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ Desomorphine, also known by its street name

More information

Pelagia Research Library

Pelagia Research Library Available online at www.pelagiaresearchlibrary.com Der Pharmacia Sinica, 2015, 6(1):6-10 ISSN: 0976-8688 CODEN (USA): PSHIBD Validated RP-HPLC method for simultaneous estimation of metformin hydrochloride

More information

Analysis of HMF by HPLC

Analysis of HMF by HPLC COST Action 927 Training School Building Skills on the Analysis of Thermal Process Contaminants in Foods 22-26 October 2007, Ankara Analysis of HMF by HPLC Vural Gökmen O O OH Background O COOH O R 2 Carbonyl

More information

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2 JPSBR: Volume 5, Issue 2: 2015 (151-155) ISS. 2271-3681 A ew Gradient RP- HPLC Method for Quantitative Analysis of : (3-luoro-4- Morpholin-4-yl-Phenyl)-Carbamic Acid Methyl Ester and its Related Substances

More information

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual Page 1 / 12 25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual ZV-4027-0500-10 500 2-8 C Page 2 / 12 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. IVD SYMBOLS...

More information

Scholars Research Library

Scholars Research Library Available online at www.scholarsresearchlibrary.com Der Pharmacia Lettre, 2011, 3(3): 350-356 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4 A new validated RP-HPLC

More information

SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM

SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM SUPPORTING INFORMATION FOR: CONCENTRATIONS OF POLYBROMINATED DIPHENYL ETHERS, HEXABROMOCYCLODODECANES AND TETRABROMOBISPHENOL-A IN BREAST MILK FROM UNITED KINGDOM WOMEN DO NOT DECREASE OVER TWELVE MONTHS

More information

Journal of Pharmacreations

Journal of Pharmacreations Journal of Pharmacreations ISSN: 2348-6295 Pharmacreations Vol.5 Issue 1 Jan- Mar- 2018 Journal Home page: www.pharmacreations.com Research article Open Access Method development and validation of apixaban

More information

Agilent Anion-Exchange Media for Proteins - Loading vs Resolution - Effect of Flow Rate and Example Protein Separations

Agilent Anion-Exchange Media for Proteins - Loading vs Resolution - Effect of Flow Rate and Example Protein Separations Agilent Anion-Exchange Media for Proteins - Loading vs Resolution - Effect of Flow Rate and Example Protein Separations Technical Overview Introduction PL-SAX is a hydrophilic strong anion-exchange chromatographic

More information

Volume 2 (6), 2014, Page CODEN (USA)-IJPRUR, e-issn: International Journal of Pharma Research and Health Sciences

Volume 2 (6), 2014, Page CODEN (USA)-IJPRUR, e-issn: International Journal of Pharma Research and Health Sciences CODEN (USA)-IJPRUR, e-issn: 2348-6465 International Journal of Pharma Research and Health Sciences Available online at www.pharmahealthsciences.net Original Article Simultaneous Determination of Salbutamol

More information

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS Tanya Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Some of the most challenging methods

More information

Achiral SFC: Development of an Orthogonal SFC Method for Mometasone Furoate Impurity Analysis

Achiral SFC: Development of an Orthogonal SFC Method for Mometasone Furoate Impurity Analysis Achiral SFC: Development of an Orthogonal SFC Method for Mometasone Furoate Impurity Analysis Zhenyu Wang Merck Research Laboratories NJCG 2013 Outline SFC chronicle The journey of a half century Re-birth

More information

Journal of Chemical and Pharmaceutical Research, 2017, 9(9): Research Article

Journal of Chemical and Pharmaceutical Research, 2017, 9(9): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2017, 9(9):70-80 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Development and Validation of Stability Indicating

More information

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS TOLBUTAMIDE IN PURE AND PHARMACEUTICAL FORMULATIONS

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS TOLBUTAMIDE IN PURE AND PHARMACEUTICAL FORMULATIONS Int. J. Chem. Sci.: 11(4), 2013, 1607-1614 ISSN 0972-768X www.sadgurupublications.com DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS TOLBUTAMIDE IN PURE AND PHARMACEUTICAL FORMULATIONS

More information

Int. J. Pharm. Sci. Rev. Res., 30(1), January February 2015; Article No. 32, Pages:

Int. J. Pharm. Sci. Rev. Res., 30(1), January February 2015; Article No. 32, Pages: Research Article Development and Validation of Stability Indicating RP-HPLC Method for the Estimation of Cilnidipine in Bulk and Pharmaceutical Dosage Form. Atul Kadam* 1, Dr. (Mrs.) Purnima Hamrapurkar

More information

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE IN TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG ANALYTICAL METHOD VALIDATION REPORT FOR ASSAY 43 2.1 Introduction Analytical

More information