Hemoglobin Raleigh as the cause of a falsely increased hemoglobin A 1C in an automated ion-exchange HPLC method

Size: px
Start display at page:

Download "Hemoglobin Raleigh as the cause of a falsely increased hemoglobin A 1C in an automated ion-exchange HPLC method"

Transcription

1 Clinical Chemistry 44: (1998) Endocrinology and Metabolism Hemoglobin Raleigh as the cause of a falsely increased hemoglobin A 1C in an automated ion-exchange HPLC method Dan Chen, 1 Dan L. Crimmins, 1 Fong Fu Hsu, 2 Frederik P. Lindberg, 3 and Mitchell G. Scott 1* Irreversible glycation of the hemoglobin A 0 (HbA 0 ) chain leads to the production of HbA 1C, which can be used to monitor long-term blood glucose control in patients with diabetes mellitus. HbA 1C is less positively charged than nonglycated HbA 0, and this decrease in charge is the basis of ion-exchange and electrophoretic methods that measure HbA 1C. We recently identified a sample that appeared to contain 46% HbA 1C by an automated ion-exchange HPLC method (Bio-Rad Variant ) but only 3.8% by an immunoinhibition latex agglutination method. A combination of traditional and mass spectrometric protein analysis and genomic DNA analysis of the Hb chain and genes revealed that the patient was heterozygotic for Hb-Raleigh, a variant containing a valine3alanine substitution at position 1 of the chain. The amino-terminal alanine in this variant Hb is posttranslationally modified by acetylation, leading to a charge difference similar to glycation and making the behavior of HbA 1C and Hb Raleigh virtually identical in the ion-exchange HPLC method. This observation suggests that it is important to confirm HbA 1C values in excess of 15%, especially if they are not consistent with the clinical picture, by an independent HbA 1C method such as immunoassay or boronic acid affinity chromatography. However, for this particular variant Hb, even these latter methods might be misleading, because the acetylated N-terminal amino acid of the Hb-Raleigh chain cannot be glycated. Washington University School of Medicine, 1 Division of Laboratory Medicine; 2 Mass Spectrometry Resource Center, Department of Medicine; and 3 Department of Medicine, Division of Infectious Diseases, St. Louis, MO *Address correspondence to this author at: Washington University School of Medicine, Division of Laboratory Medicine, Box 8118, St. Louis, MO Fax ; mscott@labmed.wustl.edu. Received January 5, 1998; revision accepted March 6, Hemoglobin A 1C (HbA 1C ) is widely accepted as the most reliable marker for monitoring long-term glucose control in diabetes (1). HbA 1C is the result of irreversible glycation of the N-terminal amino acid (valine) of the HbA 0 chain (2, 3). Glycation of the amino group of the N- terminal residue produces a loss of a positive charge, which is the basis for methods that quantitate HbA 1C by ion-exchange chromatography or gel electrophoresis (3). These methods can be affected by other modifications that alter the charge of Hb, such as carbamylation and acetylation (4 6), as well as certain Hb variants (7 12). Other approaches for measuring HbA 1C are based on immunoassays using antibodies specific for the glycated amino terminus of the chain (13) or on the affinity of glycated amino groups for boronate (5, 14, 15). An HPLC ionexchange method was the basis of measurement in the Diabetes Complications and Control Trial (DCCT) and thus has become the reference point for HbA 1C quantification (1). HPLC-based ion-exchange methods for HbA 1C have recently been automated (16), and interferences by Hb species such as HbF and HbS have been minimized. However, several reports have described artificially high HbA 1C results with hemoglobin variants such as Hb- Sherwood Forest (9) (Arg 104 Thr of the chain) and Hb-South Florida (10) (Val 1 Met of the chain) and others (16) when an automated ion-exchange HPLC method is used. We recently observed a HbA 1C value of 46% for a sample that was analyzed on the Bio-Rad Variant automated ion-exchange HPLC method for HbA 1C and investigated this, using a series of protein chemistry and molecular biology techniques. Materials and Methods HbA 1C measurement Whole-blood HbA 1C was measured by two independent methods: (a) cation exchange column-based HPLC on the Bio-Rad Variant analyzer (16) and (b) HbA 1C -specific 1296

2 Clinical Chemistry 44, No. 6, mouse monoclonal antibody-based inhibition of latex agglutinition. For the cation-exchange column-based HPLC, preparation and injection of the sample hemolysate were performed according to the manufacturer s recommendations (Bio-Rad Variant HbA 1C program; Bio- Rad, Diagnostic Group). The HbA 1C specific mouse monoclonal antibody-based inhibition of latex agglutination used a DCA 2000 Hemoglobin A 1C reagent kit (13), as specified by the manufacturer (Bayer Corporation). Whole-blood HbF was measured using the Bio-Rad Variant HPLC -thalassemia Short program. alkaline and citrate agar gel electrophoresis of Hb HB electrophoresis was performed using the Ciba Corning Alkaline and Citrate Systems (Chiron Corporation) according to the manufacturer s instructions. Hb and and chain purification Hb chains were purified by a variation of the procedure published by Bucci (17). Briefly, the globin fraction from the patient s erythrocyte lysate was precipitated at 20 C with 15 volumes of acetone containing 20 ml/l concentrated HCl. The precipitate was washed four times with acetone and lyophilized. Thirty milligrams of the globin fraction were dissolved in 5 ml of buffer (20 mmol/l sodium 2-[N-morpholino]ethanesulfonate, ph 6.6, 8 mol/l urea, 30 mmol/l 2-mercaptoethanol) and applied to a Mono-S FPLC column (Pharmacia) equilibrated with the same buffer. Protein was eluted with a mmol/l NaCl linear gradient in the same buffer, and the absorbance at 280 nm was monitored. Peaks were collected and dialyzed against 10 mmol/l NH 4 HCO 3,pH 7.8. isoelectric focusing Polyacrylamide gel electrophoresis-isoelectric focusing of the various chain preparations was performed using precast IEF 3 10 MiniPlus SepraGels according to the manufacturer s instructions (Integrated Separation Systems). liquid chromatography-mass spectrometry (lc-ms) and digestion of chain Purified Hb chains were digested with sequencing grade modified trypsin (Promega, Madison, WI) by adding 5 L of aqueous 1 mol/l dithiothreitol and 5 L ofa 1 g/l trypsin solution to 100 L of a 1 g/l solution of chain in 100 mmol/l NH 4 HCO 3, ph 7.8. The reaction mixture was incubated for 24 h at 37 C, after which 7.5 L of aqueous 100 ml/l trifluoroacetic acid was added to quench the reaction. Reaction products were stored at 4 C until use. Preparative HPLC was performed on Waters hardware, using Millennium 2.10 software (18). Samples were chromatographed on a Vydac C 18 reversedphase column (218TP54) purchased from the NEST Group (Southborough, MA), using a linear trifluoroacetic acid/ acetonitrile gradient as described previously (19). Amino acid analysis of selected fractions was performed using 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate derivatization after 6 mol/l HCl hydrolysis for 24 h at 110 C as described (20). A Finnigan MAT TSQ-7000 mass spectrometer (San Jose, CA) with an electrospray ionization source was used for molecular mass analysis. Samples were either infused at 5 L/min in a mixture of 500 ml/l methanol, 490 ml/l water, and 10 ml/l acetic acid or analyzed in an LC-MS mode after microbore chromatography on a Michrom Bioresources UMA system on a Reliasil mm C 18 reversed-phase column with minor modifications (21 23). Collision-induced dissociation (CID) experiments were used to obtain partial sequence information on isolated tryptic peptides (23). N-terminal Edman sequence analysis was performed on an ABI Model 477 sequencer (Perkin Elmer). pcr and dna sequencing of gene Genomic DNA was purified from the patient s leukocytes, using sodium dodecyl sulfate-proteinase K digestion (24). A 1.7-kb DNA fragment containing all three coding exons of the Hb chain gene was amplified by PCR (30 cycles of 30 s at 95 C, 30 s at 55 C, 120 s at 72 C). The primers were as follows: 5 of the Hb gene, TGGGCATA- AAAGTCAGGGCA; and 3 of the Hb gene, CAGAT- GCTCAAGGCCCTTCATA. The entire open reading frame was sequenced directly from the purified PCR product using the Perkin Elmer Dye Primer Cycle Sequencing system (Perkin Elmer) and an Applied Biosystems 373 DNA sequencer (25). In addition to the PCR primers, the following oligonucleotides were used in sequencing: EX2 3 CCTTCCTATGACATGAA-CTTA- ACCA; EX3 5 ACAACTACAATCCAGCTACCA; EX1 3 GGTAGA-CCACCAGCAGCCT; and EX2 5 GGCATGTGGAGACAGAGAAGACT. Results and Discussion identification of Hb chain with decreased positive charge HbA 1C analysis was performed using the Bio-Rad Variant method on a whole-blood sample from a 65-year-old white male with a history of diabetes and end-stage renal failure. A large peak constituting 46% of the patient s total Hb was identified as HbA 1C by the instrument software (not shown). Glycated Hb values for this patient at our institution had been 4 8% over the previous 2 years, as measured with a boronate affinity column method (5). The immunoinhibition latex agglutination method (13) gave a result of 3.8% HbA 1C for the current sample. To rule out HbS or HbF as the cause of the unusually high HbA 1C value from the HPLC method, the patient s blood sample was analyzed with the Bio-Rad Variant thalassemia short program, which distinguishes HbF, HbS, and HbA. No HbF or HbS was identified in the patient s Hb, but a large peak eluting earlier than A 0 was evident (not shown). This peak was not identified by the

3 1298 Chen et al.: Falsely increased hemoglobin A 1C -thalassemia Short program, but the accompanying package insert states that peaks in this area are usually HbA 1c. These results clearly demonstrated the presence of a Hb species less positively charged than HbA 0 ; however, the abundance was unlikely for HbA 1c. Electrophoretic analysis at ph 8.6 in a cellulose acetate gel failed to show any difference between normal and patient Hb, whereas electrophoresis at ph 6.2 in a citrate acetate gel detected an abnormal band with mobility similar to HbF (not shown). This also suggested the loss of a positively charged group with a pi 8.6. Such a positive charge loss would most readily be ascribed to modification of either an N-terminal amino-terminal group, a histidine, an arginine, or lysine or to a mutation leading to the loss of a positively charged residue. For example, a similar pattern of Hb electrophoresis and falsely positive HbA 1c was previously reported from a patient with the Hb variant, Hb Sherwood Forest, which has an Arg3Thr mutation in the chain (9). To identify whether the charge alteration resided in the or chain, preparative cation-exchange chromatography was performed under denaturing conditions (6 mol/l urea) to separate the and chains. This procedure revealed three peaks from the patient sample, with one chain peak (y) and one chain peak corresponding to the normal and peaks, but another chain peak (x) eluted earlier than expected (Fig. 1). The earlier eluting chain (x) peak was approximately equal in size to the normal (y) peak, suggesting that the patient is heterozygotic for a chain variant. Isoelectric focusing-polyacrylamide gel electrophoresis (Fig. 1, inset) of the two chain peaks confirmed a charge difference, because the earlier eluting (x) peak has a more acidic pi compared with the normal (y) peak. To further isolate the location of the mutation, the molecular masses of all three fractions were determined by electrospray ionization MS. The values were vs the expected for the chain and vs the expected for (y), but for the (x) peak. Together these results suggested the patient to be heterozygotic for a chain mutation that produces a decreased positive charge and an increased mass of 14 daltons. the mutation is the n-terminal valine to alanine To further identify the site(s) of the mutation, each chain from the patient (peaks x and y) was digested with trypsin, and the resulting peptide mixture chromatographed on reversed-phase media in both a preparative and an LC-MS mode. There was little difference in the preparative chromatogram between the samples except for two peaks, A and B, that eluted between 36 and 37 min (Fig. 2). Peak A was much greater for the variant (x) chain (Fig. 2), whereas the opposite was true for the normal (y) chain digest (not shown). LC-MS yielded molecular mass values of 952 Da for tryptic digest T1 (peak B) and 966 Da for tryptic digest T1 (peak A). Peak B thus represents the T1 peptide of the wild-type (y) chain (residues 1 8) and peak A is the T1 peptide of the mutant (x) chain, which is 14 Da larger. On the basis of the tryptic peptides, over 94.5% of the sequence of normal Fig. 1. Chromatograms of the patient s Hb and chains separated by cation exchange HPLC. Two peaks of similar size, X and Y, are observed along with the chain peak. The (y) and peaks co-elute with normal Hb and chains (not shown). Inset: ph 3 10 isoelectric focusing-polyacrylamide gel of (x) and (y) peaks from preparative HPLC. Lane 1, Hb (x); lane 2, Hb (y) peak, lane 3, Hb (x and y) peaks, and lane S, Hb standards.

4 Clinical Chemistry 44, No. 6, Fig. 2. Preparative C 18 reversed-phase HPLC analysis of the tryptic digest for the mutant Hb chain. Peaks were assigned (after MS analysis) and numbered according to the chronological order of tryptic fragments derived from the known protein sequence. Not identified were T6 VK, T7 AHGK, and T15 YH. Fractions were collected at 0.5-min intervals beginning at 30 min. Peak A of T1 is from Hb (x), whereas peak B of T1 is from Hb (y) contamination of the Hb (x) fraction. globin was found in the map of the (x) chain, virtually eliminating the possibility that other sites in the protein contain a mutation. Thus, the site of mutation in the patient s Hb chain was localized to the N-terminal eight residues. CID spectra for the normal T1 peptide with m/z ion of and the corresponding mutant TI peptide with m/z ion of allows immediate assignment of the mutation site to one of the first two residues (Fig. 3). The b 2 fragment ion [nomenclature from (26)] for the normal peptide (upper panel) gives the expected value of 237 Da, whereas the mutant spectra (lower panel) provides a value of 251 Da, (i.e., 14 Da), for the same N-terminal fragmentation. Because it was not immediately obvious how a 14-Da change within the first two amino acids could lead to a loss of a positive charge, we characterized the mutation at the gene level. The DNA sequence of the patient s Hb gene revealed a single base heterozygous mutation (T3C) at base 2 of the codon for the first amino acid, producing a Val3 Ala substitution. Amino acid analysis for each T1 peptide (containing the first eight amino acids of chains) confirmed that the substitution of the first amino acid valine by alanine in the mutated (x) chain was the only difference between the (x) and (y) chains. Nevertheless, the substitution of valine by alanine does not result in either loss of charge or the 14-Da difference observed between the (x) and (y) chains. Indeed, the mass difference for a Val3Ala change is 28 Da. Interestingly, standard Edman N-terminal sequence analysis of the mutant (x) chain gave no sequence information suggesting that the N terminus of the mutant (x) chain may be blocked because of a posttranslational modification. In fact, Hb-Raleigh has characteristics similar to those of this patient s Hb. In Hb-Raleigh, the chain amino-terminal valine is replaced by acetylalanine (27). Such a mutation is consistent with both the loss of positive charge of chain and the 14-Da difference we observed here. Furthermore, the acetylated N-terminal residue would explain why this protein was refractory to Edman degradation. Thus, the combination of protein biochemical studies and DNA sequence analysis determined that this patient is heterozygous for Hb-Raleigh and that HB-Raleigh produces a falsely increased HbA 1c in this automated HPLC method. Ion-exchange HPLC is a widely used methodology (1) in the measurement of normal and abnormal Hb components, and the difference in charge between HbA and HbA 1C as a result of glycation of the N-terminal amino acid of Hb chain is the basis of the Bio-Rad HPLC HbA 1C method. However, this cation-exchange columnbased method, as well as other similar methods, can

5 1300 Chen et al.: Falsely increased hemoglobin A 1C Fig. 3. CID fragmentation spectra for the unaffected (top panel) and mutant (lower panel) T1 peptides. Instrumental conditions for both spectra: 40 ev collision energy and 3s/scan. CID spectra of the 2 ions for each T1 peptide at voltages from 20 to 30 ev provided complimentary data (not shown). Peak nomenclature is taken from Biemann (26), with peaks assigned according to MS-Product from clearly misidentify as HbA 1C some Hb variants that have a loss of positive charge (7 12, 16), as well as carbamylated Hbs (4 6). In most cases, the markedly high HbA 1C will cause the laboratory to take note and investigate. HPLC and electrophoretic methods are available to clearly identify or rule out HbF or HbS. For variant Hb with decreased charge or carbamylated Hb, alternative glycohemoglobin methods, such as immunoassays that utilize antibodies specific to the glycated N-terminal values of the chain, can provide accurate values (13). Similarly, boronate affinity methods (5) can overcome the falsely increased values in ion-exchange methods caused by variant or carbamylated Hbs. Furthermore, although immunoassays, boronate affinity, and ion-exchange methods detect different modifications of the Hb molecule, extensive work towards standardization (15, 28) has made results reasonably comparable with the gold-standard ionexchange method used in the DCCT trials. Thus, when abnormally high values are identified, it is almost always possible to provide useful clinical information by alternative methods. The rare Hb variant Hb-Raleigh may, however, be an exception because of the chain N-terminal modification. Because this mutation produces an acetylated alanine at the N terminus, Hb-Raleigh chain cannot be glycosylated at the N terminus and thus will not react with the immunoassay methods. Furthermore, it would exhibit decreased reactivity with boronate affinity methods. Therefore, it may not be possible to utilize Hb glycosylation as a means to monitor glucose control in patients with Hb-Raleigh. For such individuals, determination of the glycosylation of other proteins, such as albumin (28, 29), may be necessary. To our knowledge, this conundrum is restricted only to Hb- Raleigh, Hb-Long Island, and HbA 2 -Niigata, which has a chain N-terminal acetylated alanine (30), and which was recently described as having an interference in HbA 1C methods similar to the interference described here (31).

6 Clinical Chemistry 44, No. 6, We thank Michael Berk and David Windus for clinical input and Eva Reyes for technical assistance. References 1. The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. New Engl J Med 1993;329: Ladenson JH, Chan KM, Kilzer P. Glycated hemoglobin and diabetes. Clin Chem 1985;31: Bunn H F, Haney D N, Gabbay KH, Gallop PM. Further identification of the nature and linkage of the carbohydrate in hemoglobin A 1c. Biochem Biophys Res Commun 1975;67: Weykamp CW, Penders TJ, Siebelder CW, Muskiet FA, van der Slik W. Interference of carbamylated and acetylated hemoglobins in assays of glycohemoglobin by HPLC, electrophoresis, affinity chromatography, and enzyme immunoassay. Clin Chem 1993;39: Scott MG, Hoffman JW, Meltzer VN, Siegfried BA, Chan K-M. Effects of azotemin on results of the boronate-agarose affinity and ion-exchange methods for glycated hemoglobin. Clin Chem 1984; 30: Lund L. Mourits-Andersen T. Sorensen PJ. Hemoglobin A 1c, uremia [letter]. Clin Nephrol 1988;29; Sosenko JM, Fluckiger R, Platt OS, Gabbay KH. Glycosylation of variant hemoglobins in normal and diabetic subjects. Diabetes Care 1980;3: Puukka R, Puukka M. Effect of hemoglobin F on measurements of hemoglobin A 1c with physicians office analyzers. Clin Chem 1994;40: Schnedl WJ, Reisinger EC, Pieber TR, Lipp RW, Schreiber F, Hopmeier P, Krejs GJ. Hemoglobin Sherwood Forest detected by high performance liquid chromatography. Am J Clin Path 1995; 104: Shah SC, Malone JI, Boisel JP, Kasper TJ. Hemoglobin South Florida, new variant with normal electrophoretic pattern mistaken for glycosylated hemoglobin. Diabetes 1986;35: Blanke S, Johnsen A, Wimberley PD. Hemoglobin Hafnia: alpha 2[beta 116 (G18)His 3 Gln]2; a new hemoglobin variant mistaken for glycosylated hemoglobin. Biochim Biophys Acta 1988; 955: Wajcman H, Blouquit Y, Riou J, Kister J, Poyart C, Soria J, Galacteros F. A new hemoglobin variant found during investigations of diabetes mellitus: Hb Pavie[alpha 135 (H18) Val to Glu]. Clin Chim Acta 1990;188: John WG. Hemoglobin A 1c measurement. new precise immunoassay method involving latex particle agglutination. Clin Chem 1996;42: Lee PD, Sherman LD, O Day MR, Rognerud CL, Ou CN. Comparisons of home blood glucose testing and glycated protein measurements. Diabetes Res Clin Practice 1992;16: Bodor GS, Little RR, Garrett N, Brown W, Goldstein DE, Nahm MH. Standardization of glycohemoglobin determinations in the clinical laboratory: three years of experience. Clin Chem 1992; 38: Halwachs-Baumann G, Katzensteiner S, Schnedl W, Pürstner P, Pieber T, Wilders-Truschnig M. Comparative evaluation of three assay systems for automated determination of hemoglobin A 1C. Clin Chem 1997; Bucci E. Preparation of isolated chains of human hemoglobin. Methods Enzymol 1981;76: Scott MG, Hock KG, Crimmins DL, Fracasso PM. HPLC method for monitoring SDZ PSC 833 in whole blood. Clin Chem 1997;43: Crimmins DL, Saylor M, Rush J, Thoma RS. Facile, in situ matrix-assisted laser desorption ionization-mass spectrometry analysis and assignment of disulfide pairings in heteropeptide molecules. Anal Biochem 1995;226: Crimmins DL, Cherian R. Increasing the sensitivity of 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate amino acid analysis; a simple solution. Anal Biochem 1997;244: Witkowska HE, Bitsch F, Shackleton CH. Expediting rare variant hemoglobin characterization by combined HPLC/electroscopy mass spectrometry. Hemoglobin 1993;17: Shackleton CH, Falick AM, Green BN, Witkowska HE. Electrospray mass spectrometry in the clinical diagnosis of variant hemoglobins. J Chromatograph 1991;562: Wada Y. In: Chapman J, ed. Protein and peptide analysis by mass spectrometry. Towata, NJ: Humana Press, 1996; Strauss WM. Preparation of genetic DNA from mammalian tissue. In: Ausubel FM, Brent R, Kingston RE, Moore DD, Seidel JG, Smith JA, Struhl, eds. Current protocols in molecular biology. New York: John Wiley and Sons, 1996: Rosenthal A, Chanock-Jones DS. New protocols for DNA sequencing with dye terminators. DNA Sequence 1992;3: Biemann K. Contribution of mass spectrometry to peptide and protein structure. Biomed Environment Mass Spect 1988;16: Moo-Penn WF, Bechtel KC, Schmidt RM, Johnson MH, Jue DL, Schmidt DE Jr, et al. Hemoglobin Raleigh(beta valine to acetylalanine). Structural and functional characterization. Biochemistry 1977;16: Goldstein DE, Little RR, Lorenz RA, Malone JI, Nathan D, Peterson CM. Tests of glycemia in diabetes. Diabetes Care 1995;18: Johnson RN, Metcalf PA, Baker JR. Fructosamine: a new approach to the estimation of serum glycoprotein, an index of diabetic control. Clin Chim Acta 1983;127: Harano T, Harano Kushida Y, Ueda S. HbA 2 -Niigata [ 1(NA1)VAL- ALA]: a new chain variant found in the Japanese population. Hemoglobin 1991;15: Watanabe T, Kato K, Yarnada D, Midorikawa S, Wakano S, Shiga M, et al. A non-diabetic case of hemoglobin variant (Hb-Niigata) with inappropriately high and low HbA 1C titers detected by different methods. Clin Chem 44:(in press), 1998.

Defining HbA1c: the indispensable decision to approach measurement standardization

Defining HbA1c: the indispensable decision to approach measurement standardization Defining HbA1c: the indispensable decision to approach measurement standardization Jan-Olof Jeppsson IFCC WG for HbA1c standardization Clinical Chemistry, Malmö University Hospital Sweden 6 November, 2007

More information

Hemoglobin A1c: Comparison of Pointe Scientific s 2-Part Direct Hemoglobin A1c with the Bio-Rad Variant II and the Tosoh G8

Hemoglobin A1c: Comparison of Pointe Scientific s 2-Part Direct Hemoglobin A1c with the Bio-Rad Variant II and the Tosoh G8 Hemoglobin A1c: Comparison of Pointe Scientific s 2-Part Direct Hemoglobin A1c with the Bio-Rad Variant II and the Tosoh G8 Joseph D. Artiss, Ph.D., FACB, Associate Professor of Pathology, Wayne State

More information

PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System

PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System Application Note LC/MS PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System Purpose This application note describes an automated workflow

More information

HEMOGLOBIN ELECTROPHORESIS DR ARASH ALGHASI SHAFA HOSPITAL-AHWAZ

HEMOGLOBIN ELECTROPHORESIS DR ARASH ALGHASI SHAFA HOSPITAL-AHWAZ HEMOGLOBIN ELECTROPHORESIS DR ARASH ALGHASI SHAFA HOSPITAL-AHWAZ Hemoglobin Hemoglobin (Hb), protein constituting 1/3 of the red blood cells Each red cell has 640 million molecules of Hb sites in the cells:

More information

Introduction to Peptide Sequencing

Introduction to Peptide Sequencing Introduction to Peptide equencing Quadrupole Ion Traps tructural Biophysics Course December 3, 2014 12/8/14 Introduction to Peptide equencing - athan Yates 1 Why are ion traps used to sequence peptides?

More information

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry Dr. Sanjeeva Srivastava 1. Fundamental of Mass Spectrometry Role of MS and basic concepts 2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry 2 1 MS basic concepts Mass spectrometry - technique

More information

Research Article Hb A1c Separation by High Performance Liquid Chromatography in Hemoglobinopathies

Research Article Hb A1c Separation by High Performance Liquid Chromatography in Hemoglobinopathies Scientifica Volume 2016, Article ID 2698362, 4 pages http://dx.doi.org/10.1155/2016/2698362 Research Article Hb A1c Separation by High Performance Liquid Chromatography in Hemoglobinopathies Vani Chandrashekar

More information

Haemoglobinopathies case studies 11 th Annual Sickle Cell and Thalassaemia Conference October 2017

Haemoglobinopathies case studies 11 th Annual Sickle Cell and Thalassaemia Conference October 2017 Haemoglobinopathies case studies 11 th Annual Sickle Cell and Thalassaemia Conference 11 13 October 2017 Chris Lambert Haematology Service Delivery Manager Viapath Laboratories Kings College Hospital HUMAN

More information

Identification of Haemoglobinopathies by LC/MS

Identification of Haemoglobinopathies by LC/MS Identification of Haemoglobinopathies by LC/ Mark Harrison; Senior Scientist, ThermoFisher Scientific Sarah Battle; Senior Biomedical Scientist, Royal Hallamshire Hospital Introduction There are over 2

More information

Long-Term Evaluation of Electrospray Ionization Mass Spectrometric Analysis of Glycated Hemoglobin

Long-Term Evaluation of Electrospray Ionization Mass Spectrometric Analysis of Glycated Hemoglobin Clinical Chemistry 47:2 316 321 (2001) Endocrinology and Metabolism Long-Term Evaluation of Electrospray Ionization Mass Spectrometric Analysis of Glycated Hemoglobin Norman B. Roberts, 1* Alieu B. Amara,

More information

Characterization of Disulfide Linkages in Proteins by 193 nm Ultraviolet Photodissociation (UVPD) Mass Spectrometry. Supporting Information

Characterization of Disulfide Linkages in Proteins by 193 nm Ultraviolet Photodissociation (UVPD) Mass Spectrometry. Supporting Information Characterization of Disulfide Linkages in Proteins by 193 nm Ultraviolet Photodissociation (UVPD) Mass Spectrometry M. Montana Quick, Christopher M. Crittenden, Jake A. Rosenberg, and Jennifer S. Brodbelt

More information

Application of a new capillary HPLC- ICP-MS interface to the identification of selenium-containing proteins in selenized yeast

Application of a new capillary HPLC- ICP-MS interface to the identification of selenium-containing proteins in selenized yeast Application of a new capillary HPLC- ICP-MS interface to the identification of selenium-containing proteins in selenized yeast Application note Food supplements Authors Juliusz Bianga and Joanna Szpunar

More information

Sandra Božičević dipl. ing. med. biochem, Marijana Vučić Lovrenčić, PhD Vuk Vrhovac University Clinic, Zagreb, Croatia

Sandra Božičević dipl. ing. med. biochem, Marijana Vučić Lovrenčić, PhD Vuk Vrhovac University Clinic, Zagreb, Croatia 16. PRE-ANALYTICAL, ANALYTICAL AND POST-ANALYTICAL FACTORS INFLU- ENCING SPECIFIC TESTS FOR DIAGNOSIS AND MONITORING OF DM-National network in quality assessment Prof. Dubravka Juretić, Ph.D. University

More information

BCH Graduate Survey of Biochemistry

BCH Graduate Survey of Biochemistry BCH 5045 Graduate Survey of Biochemistry Instructor: Charles Guy Producer: Ron Thomas Director: Glen Graham Lecture 7 Slide sets available at: http://hort.ifas.ufl.edu/teach/guyweb/bch5045/index.html David

More information

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels Jane Y. Yang, David A. Herold Department of Pathology, University of California San Diego, 9500 Gilman

More information

Nature Methods: doi: /nmeth Supplementary Figure 1

Nature Methods: doi: /nmeth Supplementary Figure 1 Supplementary Figure 1 Subtiligase-catalyzed ligations with ubiquitin thioesters and 10-mer biotinylated peptides. (a) General scheme for ligations between ubiquitin thioesters and 10-mer, biotinylated

More information

HPLC '88. Poster Presentation. Isolation of Thymosin B4 from Thymosin Fraction 5 by Reverse Phase HPLC

HPLC '88. Poster Presentation. Isolation of Thymosin B4 from Thymosin Fraction 5 by Reverse Phase HPLC Essentials in HPLC '88 Poster Presentation Isolation of Thymosin B4 from Thymosin Fraction 5 by Reverse Phase HPLC M. Badamchian, M.P. Strickler, M.J. Stone, A.L. Goldstein for Waters.bioresearchThe absolute,

More information

DIRECT ENZYMATIC HbA1c ASSAY

DIRECT ENZYMATIC HbA1c ASSAY DIRECT ENZYMATIC HbA1c ASSAY ACCURATE AND SIMPLE DIAZYME INNOVATIONS IN CLINICAL DIAGNOSTICS About Diazyme Diazyme Laboratories is a Division of General Atomics located in Poway, California. Diazyme uses

More information

TECHNICAL BULLETIN. R 2 GlcNAcβ1 4GlcNAcβ1 Asn

TECHNICAL BULLETIN. R 2 GlcNAcβ1 4GlcNAcβ1 Asn GlycoProfile II Enzymatic In-Solution N-Deglycosylation Kit Product Code PP0201 Storage Temperature 2 8 C TECHNICAL BULLETIN Product Description Glycosylation is one of the most common posttranslational

More information

What is hemoglobin A1c? An analysis of glycated hemoglobins by electrospray ionization mass spectrometry

What is hemoglobin A1c? An analysis of glycated hemoglobins by electrospray ionization mass spectrometry Clinical Chemistry 44:9 1951 1958 (1998) Endocrinology and Metabolism What is hemoglobin A1c? An analysis of glycated hemoglobins by electrospray ionization mass spectrometry Karen P. Peterson, 1* James

More information

Hemoglobinopathy and thalassemia detection

Hemoglobinopathy and thalassemia detection Hemoglobinopathy and thalassemia detection Traditional methods and a novel method capillary electrophoresis technology By Aigars Brants, PhD Published October 2011 In humans, two pairs of unlike globin

More information

Communication. Identification of Methionine N -Acetyltransferase from Saccharomyces cerevisiae

Communication. Identification of Methionine N -Acetyltransferase from Saccharomyces cerevisiae Communication THE JOURNAL OP BIOLOGICAL CHEMISTRY Vol. 265, No. 7, Issue of March 5, pp. 3603-3606,lSSO 0 1990 by The American Society for Biochemistry and Molecular Biology, Inc. Printed in U. S. A. Identification

More information

Antoine Bouchoux, Pierre-Emerson Cayemitte, Julien Jardin, Geneviève Gésan-Guiziou, and Bernard Cabane

Antoine Bouchoux, Pierre-Emerson Cayemitte, Julien Jardin, Geneviève Gésan-Guiziou, and Bernard Cabane Biophysical Journal, Volume 96 Supplementary Material Casein Micelle Dispersions under Osmotic Stress Antoine Bouchoux, Pierre-Emerson Cayemitte, Julien Jardin, Geneviève Gésan-Guiziou, and Bernard Cabane

More information

LANCE Eu-W1024 ITC Chelate & Europium Standard AD0013 Development grade

LANCE Eu-W1024 ITC Chelate & Europium Standard AD0013 Development grade AD0017P-4 (en) 1 LANCE Eu-W1024 ITC Chelate & Europium Standard AD0013 Development grade INTRODUCTION Fluorescent isothiocyanato-activated (ITC-activated) Eu-W1024 chelate is optimized for labelling proteins

More information

Tosoh Automated Glycohemoglobin Analyzer HLC-723GX

Tosoh Automated Glycohemoglobin Analyzer HLC-723GX Tosoh Automated Glycohemoglobin Analyzer HLC-723GX The Diabetes Epidemic and the role of HbA 1c Diabetes is recognised worldwide as a disease that is reaching epidemic proportions. (1) IDF region Adult

More information

Caution: For Laboratory Use. A product for research purposes only. Eu-W1024 ITC Chelate & Europium Standard. Product Number: AD0013

Caution: For Laboratory Use. A product for research purposes only. Eu-W1024 ITC Chelate & Europium Standard. Product Number: AD0013 TECHNICAL DATA SHEET Lance Caution: For Laboratory Use. A product for research purposes only. Eu-W1024 ITC Chelate & Europium Standard Product Number: AD0013 INTRODUCTION: Fluorescent isothiocyanato-activated

More information

HbA1c and Diabetes. Steven Weier Senior Lecturer School of Biomedical Sciences Faculty of Health QUT. CRICOS No J

HbA1c and Diabetes. Steven Weier Senior Lecturer School of Biomedical Sciences Faculty of Health QUT. CRICOS No J HbA1c and Diabetes Steven Weier Senior Lecturer School of Biomedical Sciences Faculty of Health QUT Queensland University of Technology History of HbA1c 1966: Holmquist and Schroeder identify five subtypes

More information

VaTx1 VaTx2 VaTx3. VaTx min Retention Time (min) Retention Time (min)

VaTx1 VaTx2 VaTx3. VaTx min Retention Time (min) Retention Time (min) a Absorbance (mau) 5 2 5 3 4 5 6 7 8 9 6 2 3 4 5 6 VaTx2 High Ca 2+ Low Ca 2+ b 38.2 min Absorbance (mau) 3 2 3 4 5 3 2 VaTx2 39.3 min 3 4 5 3 2 4. min 3 4 5 Supplementary Figure. Toxin Purification For

More information

Improving the Accuracy of Hemoglobin A 1c. : Your Help Is Needed

Improving the Accuracy of Hemoglobin A 1c. : Your Help Is Needed Improving the Accuracy of Hemoglobin A 1c : Your Help Is Needed Kristina Jackson Behan, PhD, MT(ASCP) (Clinical Laboratory Sciences Program, University of West Florida, Pensacola, FL) DOI: 10.1309/RTVTWPYPTPA5YAGU

More information

Complete Amino Acid Sequence of the Protease Inhibitor BWI-4a from Buckwheat Seeds

Complete Amino Acid Sequence of the Protease Inhibitor BWI-4a from Buckwheat Seeds Biochemistry (Moscow), Vol. 65, No. 10, 2000, pp. 1140-1144. Translated from Biokhimiya, Vol. 65, No. 10, 2000, pp. 1347-1352. Original Russian Text Copyright 2000 by Belozersky, Dunaevsky, Musolyamov,

More information

Supporting information

Supporting information Supporting information Figure legends Supplementary Table 1. Specific product ions obtained from fragmentation of lithium adducts in the positive ion mode comparing the different positional isomers of

More information

Lecture 3. Tandem MS & Protein Sequencing

Lecture 3. Tandem MS & Protein Sequencing Lecture 3 Tandem MS & Protein Sequencing Nancy Allbritton, M.D., Ph.D. Department of Physiology & Biophysics 824-9137 (office) nlallbri@uci.edu Office- Rm D349 Medical Science D Bldg. Tandem MS Steps:

More information

Detection and quantitation of hemoglobin (Hb) fractions

Detection and quantitation of hemoglobin (Hb) fractions Evaluation of a Dual Hemoglobin A 2 /A 1c Quantitation Kit on the Bio-Rad Variant II Automated Hemoglobin Analyzer John D. Lafferty, MLT, ART; Andrew G. McFarlane, MLT, ART; David H. K. Chui, MD, FRCP

More information

Supplementary Materials for

Supplementary Materials for advances.sciencemag.org/cgi/content/full/2/4/e1500980/dc1 Supplementary Materials for The crystal structure of human dopamine -hydroxylase at 2.9 Å resolution Trine V. Vendelboe, Pernille Harris, Yuguang

More information

Triptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions

Triptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions Electronic Supplementary Material (ESI) for Chemical Science. This journal is The Royal Society of Chemistry 2015 Triptycene-Based Small Molecules Modulate (CAG) (CTG) Repeat Junctions Stephanie A. Barros

More information

Mass Spectrometry. Mass spectrometer MALDI-TOF ESI/MS/MS. Basic components. Ionization source Mass analyzer Detector

Mass Spectrometry. Mass spectrometer MALDI-TOF ESI/MS/MS. Basic components. Ionization source Mass analyzer Detector Mass Spectrometry MALDI-TOF ESI/MS/MS Mass spectrometer Basic components Ionization source Mass analyzer Detector 1 Principles of Mass Spectrometry Proteins are separated by mass to charge ratio (limit

More information

130327SCH4U_biochem April 09, 2013

130327SCH4U_biochem April 09, 2013 Option B: B1.1 ENERGY Human Biochemistry If more energy is taken in from food than is used up, weight gain will follow. Similarly if more energy is used than we supply our body with, weight loss will occur.

More information

Author. Introduction. Abstract

Author. Introduction. Abstract Simultaneous Measurement of Sirolimus and Tacrolimus Concentrations in Blood by Semi-Automated Extraction and Liquid Chromatography-Electrospray Ionization Mass Spectrometry Application Author Gary L.

More information

SUPPLEMENTARY MATERIAL

SUPPLEMENTARY MATERIAL SUPPLEMENTARY MATERIAL Purification and biochemical properties of SDS-stable low molecular weight alkaline serine protease from Citrullus Colocynthis Muhammad Bashir Khan, 1,3 Hidayatullah khan, 2 Muhammad

More information

Effect of Hemoglobin Eon Glycosylated Hemoglobin Determinations Using Different Commercial Kits

Effect of Hemoglobin Eon Glycosylated Hemoglobin Determinations Using Different Commercial Kits Effect of Hemoglobin Eon Glycosylated Hemoglobin Determinations Using Different Commercial Kits M Musalmah, PhD*,] Normah, DCP**, W B Wan Mohamad,MRCP***, 0 Noor Salwah, BSc*, H Abdul Fatah, DTMP*, N A

More information

Supplementary Materials for

Supplementary Materials for advances.sciencemag.org/cgi/content/full/2/1/e1500678/dc1 Supplementary Materials for Chemical synthesis of erythropoietin glycoforms for insights into the relationship between glycosylation pattern and

More information

Agilent Protein In-Gel Tryptic Digestion Kit

Agilent Protein In-Gel Tryptic Digestion Kit Agilent 5188-2749 Protein In-Gel Tryptic Digestion Kit Agilent Protein In-Gel Tryptic Digestion Kit Instructions Kit Contents The Protein In-Gel Tryptic Digestion Kit includes sufficient reagents for approximately

More information

Peptide sequencing using chemically assisted fragmentation (CAF) and Ettan MALDI-ToF Pro mass spectrometry

Peptide sequencing using chemically assisted fragmentation (CAF) and Ettan MALDI-ToF Pro mass spectrometry application note Ett MALDI-ToF Pro Peptide sequencing using chemically assisted fragmentation (CAF) d Ett MALDI-ToF Pro mass spectrometry Key words: MALDI-ToF, PSD, peptide sequencing, chemically assisted

More information

Calibration of ion-exchange HPLC measurements of glycohemoglobin: effect on interassay precision

Calibration of ion-exchange HPLC measurements of glycohemoglobin: effect on interassay precision Clinical Chemistry 43:12 2353 2357 (1997) Endocrinology and Metabolism Calibration of ion-exchange HPLC measurements of glycohemoglobin: effect on interassay precision Erik Gerlo* and Frans Gorus To investigate

More information

Detecting and Reporting Alpha Thalassemia In Newborns

Detecting and Reporting Alpha Thalassemia In Newborns Detecting and Reporting Alpha Thalassemia In Newborns T. Davis, C. Moore, L. Nayak, M.C. Dorley, M. del Pilar Aguinaga, M. Chan, J. Ubaike, C. Yusuf Alpha Thalassemia Screening Status in the US Clinical

More information

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser Evaluation Report Eurolyser HbA1c Test Kit on Smart and Cube Analyser Locations Location 1: Klinisk kemi University Hospital MAS Malmö Operator: Dr. Benny Larsson Date: June/July 2007 Location 2: Eurolyser

More information

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid Joanne ( 乔安妮 ) Mather Senior Scientist Waters Corporation Data courtesy of Erin Chambers and Mary

More information

Unraveling Hemoglobinopathies with Capillary Electrophoresis

Unraveling Hemoglobinopathies with Capillary Electrophoresis Session Number 2002 Unraveling Hemoglobinopathies with Capillary Electrophoresis David F. Keren, M.D. Professor of Pathology Division Director, Clinical Pathology The University of Michigan dkeren@med.umich.edu

More information

Measuring Lipid Composition LC-MS/MS

Measuring Lipid Composition LC-MS/MS Project: Measuring Lipid Composition LC-MS/MS Verification of expected lipid composition in nanomedical controlled release systems by liquid chromatography tandem mass spectrometry AUTHORED BY: DATE: Sven

More information

Trypsin Mass Spectrometry Grade

Trypsin Mass Spectrometry Grade 058PR-03 G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name Trypsin Mass Spectrometry Grade A Chemically Modified, TPCK treated, Affinity Purified

More information

Page 1 of 5 Biochemistry I Fall 2017 Practice for Exam2 Dr. Stone Name

Page 1 of 5 Biochemistry I Fall 2017 Practice for Exam2 Dr. Stone Name Page 1 of 5 Biochemistry I Fall 2017 Practice for Exam2 Dr. Stone ame o answers will be provided. ere are some constants and equations that may be useful: K a = [+][A-]/[A] p = pka + log [A-]/[A] K a for

More information

Multiple-Choice Questions Answer ALL 20 multiple-choice questions on the Scantron Card in PENCIL

Multiple-Choice Questions Answer ALL 20 multiple-choice questions on the Scantron Card in PENCIL Multiple-Choice Questions Answer ALL 20 multiple-choice questions on the Scantron Card in PENCIL For Questions 1-10 choose ONE INCORRECT answer. 1. Which ONE of the following statements concerning the

More information

The distribution of log 2 ratio (H/L) for quantified peptides. cleavage sites in each bin of log 2 ratio of quantified. peptides

The distribution of log 2 ratio (H/L) for quantified peptides. cleavage sites in each bin of log 2 ratio of quantified. peptides Journal: Nature Methods Article Title: Corresponding Author: Protein digestion priority is independent of their abundances Mingliang Ye and Hanfa Zou Supplementary Figure 1 Supplementary Figure 2 The distribution

More information

Caution: For Laboratory Use. A product for research purposes only. Eu-W1284 Iodoacetamido Chelate & Europium Standard. Product Number: AD0014

Caution: For Laboratory Use. A product for research purposes only. Eu-W1284 Iodoacetamido Chelate & Europium Standard. Product Number: AD0014 TECHNICAL DATA SHEET Lance Caution: For Laboratory Use. A product for research purposes only. Eu-W1284 Iodoacetamido Chelate & Europium Standard Product Number: AD0014 INTRODUCTION: Iodoacetamido-activated

More information

Comparison of mass spectrometers performances

Comparison of mass spectrometers performances Comparison of mass spectrometers performances Instrument Mass Mass Sensitivity resolution accuracy Quadrupole 1 x 10 3 0.1 Da* 0.5-1.0 pmol DE-MALDI 2 x 10 4 20 ppm 1-10 fmol peptide 1-5 pmol protein Ion

More information

MBB 694:407, 115:511. Please use BLOCK CAPITAL letters like this --- A, B, C, D, E. Not lowercase!

MBB 694:407, 115:511. Please use BLOCK CAPITAL letters like this --- A, B, C, D, E. Not lowercase! MBB 694:407, 115:511 First Test Severinov/Deis Tue. Sep. 30, 2003 Name Index number (not SSN) Row Letter Seat Number This exam consists of two parts. Part I is multiple choice. Each of these 25 questions

More information

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform Abstract Targeted proteomics for biomarker verification/validation

More information

Improve Protein Analysis with the New, Mass Spectrometry- Compatible ProteasMAX Surfactant

Improve Protein Analysis with the New, Mass Spectrometry- Compatible ProteasMAX Surfactant Improve Protein Analysis with the New, Mass Spectrometry- Compatible Surfactant ABSTRACT Incomplete solubilization and digestion and poor peptide recovery are frequent limitations in protein sample preparation

More information

MALDI-TOF. Introduction. Schematic and Theory of MALDI

MALDI-TOF. Introduction. Schematic and Theory of MALDI MALDI-TOF Proteins and peptides have been characterized by high pressure liquid chromatography (HPLC) or SDS PAGE by generating peptide maps. These peptide maps have been used as fingerprints of protein

More information

Introduction to Biochemistry Midterm exam )ومن أحياها(

Introduction to Biochemistry Midterm exam )ومن أحياها( Introduction to Biochemistry Midterm exam 2016-2017 )ومن أحياها( 1. Which of the following amino (in a peptide chain) would probably be found at a beta bend or turn? a. lysine * b. Gly c. arg d. asn 2.

More information

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) The target compound to be determined is 2, 4, 5-T. 1. Instrument Liquid Chromatograph-tandem mass spectrometer (LC-MS/MS)

More information

PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX

PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX PHOSPHOPEPTIDE ANALYSIS USING IMAC SAMPLE PREPARATION FOLLOWED BY MALDI-MS and MALDI PSD MX E. Claude 1, E. Emekwue 2, M. Snel 1, T. McKenna 1 and J. Langridge 1 1: Waters Corporation, Manchester, UK 2:

More information

Effects of hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by twelve methods

Effects of hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by twelve methods Effects of hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by twelve methods Curt Rohlfing, University of Missouri Steven Hanson, University of Missouri Cas Weykamp, Location Queen Beatrix

More information

Babu Antharavally, Ryan Bomgarden, and John Rogers Thermo Fisher Scientific, Rockford, IL

Babu Antharavally, Ryan Bomgarden, and John Rogers Thermo Fisher Scientific, Rockford, IL A Versatile High-Recovery Method for Removing Detergents from Low-Concentration Protein or Peptide Samples for Mass Spectrometry Sample Preparation and Analysis Babu Antharavally, Ryan Bomgarden, and John

More information

v o = V max [S] rate = kt[s] e V max = k cat E t ΔG = -RT lnk eq K m + [S]

v o = V max [S] rate = kt[s] e V max = k cat E t ΔG = -RT lnk eq K m + [S] Exam 3 Spring 2017 Dr. Stone 8:00 Name There are 100 possible points on this exam. -ΔG / RT v o = V max [S] rate = kt[s] e V max = k cat E t ΔG = -RT lnk eq K m + [S] h rate forward = k forward [reactants]

More information

CHAPTER-6 IDENTIFICATION, AND CHARACTERISATION OF DEGRADATION IMPURITY IN VALSARTAN TABLETS

CHAPTER-6 IDENTIFICATION, AND CHARACTERISATION OF DEGRADATION IMPURITY IN VALSARTAN TABLETS 129 CHAPTER-6 IDENTIFICATION, AND CHARACTERISATION OF DEGRADATION IMPURITY IN VALSARTAN TABLETS 130 6.1. Introduction Valsartan is an orally active specific angiotensin II blocker effective in lowering

More information

Proteins are sometimes only produced in one cell type or cell compartment (brain has 15,000 expressed proteins, gut has 2,000).

Proteins are sometimes only produced in one cell type or cell compartment (brain has 15,000 expressed proteins, gut has 2,000). Lecture 2: Principles of Protein Structure: Amino Acids Why study proteins? Proteins underpin every aspect of biological activity and therefore are targets for drug design and medicinal therapy, and in

More information

Phosphorylation of proteins Steve Barnes Feb 19th, 2002 in some cases, proteins are found in a stable, hyperphosphorylated state, e.g.

Phosphorylation of proteins Steve Barnes Feb 19th, 2002 in some cases, proteins are found in a stable, hyperphosphorylated state, e.g. Phosphorylation of proteins Steve Barnes Feb 19th, 2002 in some cases, proteins are found in a stable, hyperphosphorylated state, e.g., casein more interestingly, in most other cases, it is a transient

More information

Properties of amino acids in proteins

Properties of amino acids in proteins Properties of amino acids in proteins one of the primary roles of DNA (but far from the only one!!!) is to code for proteins A typical bacterium builds thousands types of proteins, all from ~20 amino acids

More information

Can glycated albumin assist in management of diabetes mellitus? Jasna Lenicek Krleza, PhD. Children s Hospital Zagreb, Croatia

Can glycated albumin assist in management of diabetes mellitus? Jasna Lenicek Krleza, PhD. Children s Hospital Zagreb, Croatia Can glycated albumin assist in management of diabetes mellitus? Jasna Lenicek Krleza, PhD. Children s Hospital Zagreb, Croatia DIABETES MELLITUS COURSE SUBJECT 2001 2014 WHEN WE PREPARED PROGRAMME of THIS

More information

In-Gel Tryptic Digestion Kit

In-Gel Tryptic Digestion Kit INSTRUCTIONS In-Gel Tryptic Digestion Kit 3747 N. Meridian Road P.O. Box 117 Rockford, IL 61105 89871 1468.2 Number Description 89871 In-Gel Tryptic Digestion Kit, sufficient reagents for approximately

More information

Previous Class. Today. Detection of enzymatic intermediates: Protein tyrosine phosphatase mechanism. Protein Kinase Catalytic Properties

Previous Class. Today. Detection of enzymatic intermediates: Protein tyrosine phosphatase mechanism. Protein Kinase Catalytic Properties Previous Class Detection of enzymatic intermediates: Protein tyrosine phosphatase mechanism Today Protein Kinase Catalytic Properties Protein Phosphorylation Phosphorylation: key protein modification

More information

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column Application Note BioPharma Authors James Martosella, Phu Duong, and Alex Zhu Agilent Technologies, Inc. Abstract

More information

Evidence Based HbA1c Accuracy. Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology

Evidence Based HbA1c Accuracy. Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology Evidence Based HbA1c Accuracy Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology Fresh lycohaemoglobin This program is designed to assist laboratories to assess the accuracy of their hihba1c

More information

Biomolecular Mass Spectrometry

Biomolecular Mass Spectrometry Lipids ot different than other organic small molecules Carbohydrates Polymers of monosaccharides linked via glycosidic bonds (acetals/ ketals) many different combinationsvery interesting no time ucleic

More information

Heparin Sodium ヘパリンナトリウム

Heparin Sodium ヘパリンナトリウム Heparin Sodium ヘパリンナトリウム Add the following next to Description: Identification Dissolve 1 mg each of Heparin Sodium and Heparin Sodium Reference Standard for physicochemical test in 1 ml of water, and

More information

Chemical Nature of the Amino Acids. Table of a-amino Acids Found in Proteins

Chemical Nature of the Amino Acids. Table of a-amino Acids Found in Proteins Chemical Nature of the Amino Acids All peptides and polypeptides are polymers of alpha-amino acids. There are 20 a- amino acids that are relevant to the make-up of mammalian proteins (see below). Several

More information

Supplementary Figure-1. SDS PAGE analysis of purified designed carbonic anhydrase enzymes. M1-M4 shown in lanes 1-4, respectively, with molecular

Supplementary Figure-1. SDS PAGE analysis of purified designed carbonic anhydrase enzymes. M1-M4 shown in lanes 1-4, respectively, with molecular Supplementary Figure-1. SDS PAGE analysis of purified designed carbonic anhydrase enzymes. M1-M4 shown in lanes 1-4, respectively, with molecular weight markers (M). Supplementary Figure-2. Overlay of

More information

Supplementary Figure S1. Appearacne of new acetyl groups in acetylated lysines using 2,3-13 C 6 pyruvate as a tracer instead of labeled glucose.

Supplementary Figure S1. Appearacne of new acetyl groups in acetylated lysines using 2,3-13 C 6 pyruvate as a tracer instead of labeled glucose. Supplementary Figure S1. Appearacne of new acetyl groups in acetylated lysines using 2,3-13 C 6 pyruvate as a tracer instead of labeled glucose. (a) Relative levels of both new acetylation and all acetylation

More information

SUPPLEMENTARY DATA. Materials and Methods

SUPPLEMENTARY DATA. Materials and Methods SUPPLEMENTARY DATA Materials and Methods HPLC-UV of phospholipid classes and HETE isomer determination. Fractionation of platelet lipid classes was undertaken on a Spherisorb S5W 150 x 4.6 mm column (Waters

More information

assay Introduction Conclusions: The D-100 TM system proved to be a robust and reliable method for HbA 1c

assay Introduction Conclusions: The D-100 TM system proved to be a robust and reliable method for HbA 1c Clin Chem Lab Med 215; aop Stéphane Jaisson*, Nathalie Leroy, Emmanuelle Guillard, Aurore Desmons and Philippe Gillery Analytical performances of the D- TM hemoglobin testing system (Bio-Rad) for assay

More information

Rat Hemoglobin A1c (HbA1c) Kit Instructions

Rat Hemoglobin A1c (HbA1c) Kit Instructions V.3 Crystal Chem Rat Hemoglobin A1c (HbA1c) Kit Instructions For the quantitative determination of hemoglobin A1c (HbA1c) in rat whole blood Catalog #80300 96 Assays For research use only. Not for use

More information

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1 Development of an SPE-LC/MS/MS Assay for the Simultaneous Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid in Support of Alzheimer s Research Dr. Erin E. Chambers Waters Corporation Presented

More information

DetergentOUT Tween. DetergentOUT GBS10. OrgoSol DetergentOUT

DetergentOUT Tween. DetergentOUT GBS10. OrgoSol DetergentOUT 252PR 01 G-Biosciences, St Louis, MO. USA 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name DetergentOUT Detergent Removal Systems For the Removal of Detergents

More information

Multiplex Protein Quantitation using itraq Reagents in a Gel-Based Workflow

Multiplex Protein Quantitation using itraq Reagents in a Gel-Based Workflow Multiplex Protein Quantitation using itraq Reagents in a Gel-Based Workflow Purpose Described herein is a workflow that combines the isobaric tagging reagents, itraq Reagents, with the separation power

More information

Short, 2 point questions. Be brief, but not vague. Specfic details are needed.

Short, 2 point questions. Be brief, but not vague. Specfic details are needed. Biochemistry Exam II Fall 2012 Dr. Stone Name There are 11 short answer/ multiple choice questions worth 2 points each. There are six long answer questions worth a total of 68 points. Short, 2 point questions.

More information

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis Ying Qing Yu Waters Corporation, Milford, MA, USA APPLICATION BENEFITS

More information

BabyBio IMAC columns DATA SHEET DS

BabyBio IMAC columns DATA SHEET DS BabyBio IMAC columns DATA SHEET DS 45 655 010 BabyBio columns for Immobilized Metal Ion Affinity Chromatography (IMAC) are ready-to-use for quick and easy purification of polyhistidine-tagged (His-tagged)

More information

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Jenny Chen, Hongxia Wang, Zhiqi Hao, Patrick Bennett, and Greg Kilby Thermo Fisher

More information

DELFIA Tb-N1 DTA Chelate & Terbium Standard

DELFIA Tb-N1 DTA Chelate & Terbium Standard AD0029P-1 (en) 1 DELFIA Tb-N1 DTA Chelate & AD0012 Terbium Standard For Research Use Only INTRODUCTION DELFIA Tb-N1 DTA Chelate is optimized for the terbium labeling of proteins and peptides for use in

More information

California Association for Medical Laboratory Technology Distance Learning Course

California Association for Medical Laboratory Technology Distance Learning Course California Association for Medical Laboratory Technology Distance Learning Course Hemoglobin A1c Testing of Patients with Hemoglobinopathies By Amy Fang Graduate Student in Clinical Science California

More information

4th Multidimensional Chromatography Workshop Toronto (January, 2013) Herman C. Lam, Ph.D. Calibration & Validation Group

4th Multidimensional Chromatography Workshop Toronto (January, 2013) Herman C. Lam, Ph.D. Calibration & Validation Group 4th Multidimensional Chromatography Workshop Toronto (January, 2013) Herman C. Lam, Ph.D. Calibration & Validation Group MDLC for Shotgun Proteomics Introduction General concepts Advantages Challenges

More information

Trypsin digestion: The lyophilized powder of the reduced S-carboxymethylated ACID) BETWEEN NORMAL (B+) AND THE COMMON NEGRO VARIANT

Trypsin digestion: The lyophilized powder of the reduced S-carboxymethylated ACID) BETWEEN NORMAL (B+) AND THE COMMON NEGRO VARIANT A SINGLE AMINO ACID SUBSTITUTION (ASPARAGINE TO ASPARTIC ACID) BETWEEN NORMAL (B+) AND THE COMMON NEGRO VARIANT (A+) OF HUMAN GLUCOSE-6-PHOSPHATE DEHYDROGENASE* BY AKIRA YOSHIDA DIVISION OF MEDICAL GENETICS,

More information

(III) MALDI instrumentation

(III) MALDI instrumentation Dr. Sanjeeva Srivastava (I) Basics of MALDI-TF (II) Sample preparation In-gel digestion Zip-tip sample clean-up Matrix and sample plating (III) MALDI instrumentation 2 1 (I) Basics of MALDI-TF Analyte

More information

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID James Little, Eastman Chemical Company, Kingsport, TN Overview Phospholipids and

More information

Oligosaccharide Profiling of O-linked Oligosaccharides Labeled with 2 Aminobenzoic Acid (2-AA)

Oligosaccharide Profiling of O-linked Oligosaccharides Labeled with 2 Aminobenzoic Acid (2-AA) Oligosaccharide Profiling of O-linked Oligosaccharides Labeled with 2 Aminobenzoic Acid (2-AA) Elisabeth A. Kast and Elizabeth A. Higgins GlycoSolutions Corporation, Worcester, MA Data originally presented

More information

2. Which of the following amino acids is most likely to be found on the outer surface of a properly folded protein?

2. Which of the following amino acids is most likely to be found on the outer surface of a properly folded protein? Name: WHITE Student Number: Answer the following questions on the computer scoring sheet. 1 mark each 1. Which of the following amino acids would have the highest relative mobility R f in normal thin layer

More information

Mercaptoethanesulfonic acid as the reductive thiol-containing reagent employed for the derivatization of amino acids with o-phthaldialdehyde analysis

Mercaptoethanesulfonic acid as the reductive thiol-containing reagent employed for the derivatization of amino acids with o-phthaldialdehyde analysis Acta Univ. Sapientiae, Alimentaria, 1 (2008) 49 60 Mercaptoethanesulfonic acid as the reductive thiol-containing reagent employed for the derivatization of amino acids with o-phthaldialdehyde analysis

More information

Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring

Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring Application Note Authors Ning Tang, Christine A. Miller and Keith Waddell Agilent Technologies, Inc. Santa Clara, CA USA This

More information