Vascular Enhancement and Image Quality of CT Venography: Comparison of Standard and Low Kilovoltage Settings

Size: px
Start display at page:

Download "Vascular Enhancement and Image Quality of CT Venography: Comparison of Standard and Low Kilovoltage Settings"

Transcription

1 Cardiopulmonary Imaging Original Research Fujikawa et al. Vascular Enhancement and Image Quality of CTV Cardiopulmonary Imaging Original Research Atsuko Fujikawa 1 Shin Matsuoka Kenji Kuramochi Tatsuo Yoshikawa Kunihiro Yagihashi Yasuyuki Kurihara Yasuo Nakajima Fujikawa A, Matsuoka S, Kuramochi K, et al. Keywords: CT venography, deep venous thrombosis, low kilovoltage setting, peak kilovoltage, radiation dose DOI: /AJR Received August 2, 2010; accepted after revision February 23, All authors: Department of Radiology, St. Marianna University School of Medicine, Sugao, Miyamae-Ku, Kawasaki City, Kanagawa , Japan. Address correspondence to A. Fujikawa (a2fujikawa@marianna-u.ac.jp). AJR 2011; 197: X/11/ American Roentgen Ray Society Vascular Enhancement and Image Quality of CT Venography: Comparison of Standard and Low Kilovoltage Settings OBJECTIVE. The objective of our study was to investigate the vascular enhancement and image quality of CT venography (CTV) with a lower peak kilovoltage (kvp) setting than the standard setting. MATERIALS AND METHODS. In this retrospective study, the clinical records of 100 consecutive patients with suspected pulmonary embolism were analyzed. All patients underwent pulmonary CT angiography and CTV of the abdomen, pelvis, and lower extremities using 64-MDCT with automatic tube current modulation: 50 patients underwent CT at 120 kvp, the standard kvp setting, and 50 patients were scanned at 100 kvp; we refer to these groups as the standard-kvp group and the low-kvp group, respectively. Vessel enhancement and image noise were assessed in the inferior vena cava (IVC), femoral vein, and popliteal vein. Two radiologists who were blinded to the kvp setting placed the regions of interest on vessels by consensus and assessed image quality using a 5-point visual scale. Effective dose was estimated using the dose-length product. The Wilcoxon rank test was used to evaluate differences between the two groups using statistics software (JMP, version 5.1). A p value of less than 0.05 was considered to indicate statistical significance. RESULTS. Mean vascular enhancement was significantly higher in the low-kvp group than in the standard-kvp group: IVC, ± 12.2 (SD) HU versus ± 17.4 HU, respectively; femoral vein, ± 18.0 HU versus ± 24.5 HU; and popliteal vein, ± 17.5 HU versus ± 26.0 HU. Although the images of the low-kvp group had significantly higher image noise, there were no significant differences in image quality in the IVC and popliteal vein. The mean effective dose for the low-kvp protocol was significantly lower than that for the standard-kvp protocol. CONCLUSION. Lowering the kvp setting for CTV examinations improved vascular enhancement while providing sufficient image quality. D eep venous thrombosis (DVT) and pulmonary embolism (PE) are considered to be parts of the same pathologic process. Since the combination of CT venography (CTV) and pulmonary CT angiography (CTA) was initially described in 1998 [1], several researchers have reported highly accurate diagnoses of DVT in patients with PE [1, 2]. This method enables the simultaneous diagnosis of PE and DVT without the administration of additional contrast material. The fact that the Prospective Investigation of Pulmonary Embolism Diagnosis (PIOPED) II study investigators reported a higher sensitivity (90%) for PE using combined pulmonary CTA and CTV than using pulmonary CTA alone (83%) seems to validate this method [2]. The disadvantage of combined pulmonary CTA and CTV is the increased radiation dose to which the patient is exposed. Several methods have been proposed to reduce the radiation dose from pulmonary CTA and cardiac CTA [3 5], of which lowering the peak kilovoltage (kvp) setting is very effective. Goodman et al. [6] reported that obtaining discontiguous images for CTV reduced radiation dose. Other investigators have reported that a reduction in the kvp setting from 120 to 100 kvp during pulmonary CTA results in an average reduction of the estimated effective dose of approximately 44% [7]. Furthermore, lowering the kvp setting has been suggested as a technique to improve contrast enhancement [7, 8]. However, to our knowledge, the efficacy of a low-kilovoltage protocol for CTV has not been previously analyzed. On 838 AJR:197, October 2011

2 Vascular Enhancement and Image Quality of CTV the basis of previous reports of the efficacy of lowering the kvp setting in pulmonary CTA, we hypothesized that lowering the kvp setting in CTV would improve vascular enhancement and reduce radiation exposure without a significant degradation of image quality. Thus, the purpose of our study was to compare contrast enhancement and image quality of veins on CTV images obtained using the standard kvp setting and CTV images obtained using a lower kvp setting and to determine whether CTV with a low kvp setting can improve vascular enhancement without degrading the image quality. Materials and Methods Subjects Our study comprised a retrospective evaluation of MDCT studies performed in routine clinical practice before and after changing the standard scanning protocol for the evaluation of PE and DVT at our institute. The change in protocol, which involved a reduction in kvp, was adopted because this technique was reported to improve contrast enhancement while also reducing radiation dose [7 11]. In our institution, combined pulmonary CTA and CTA have been performed using 100 kvp since September We could make the prospective aspect of this research, but this protocol change was not made for an academic purpose; instead, the change was initiated because of its utility in clinical practice. This retrospective study was performed with institutional review board approval, and the requirement for informed consent was waived. The clinical records (age, sex, body mass index [BMI]) of 100 consecutive patients (74 women and 26 men; mean age ± SD, 64 ± 15 years; age range, years) suspected of having PE were analyzed in this study. Fifty patients (36 women and 14 men; mean age, 64 ± 15 years; age range, years) were scanned using the 120- kvp protocol; we refer to this group as the standard-kvp group. Another group of 50 patients (38 women and 12 men; mean age, 64 ± 15 years; age range, years) underwent CT after the scanning protocol had been changed to 100 kvp; we refer to this group as the low-kvp group. Patients who had undergone amputation of a lower extremity and those younger than 20 years were excluded from this evaluation. In addition, we excluded cases with severe degradation of image quality caused by artifacts (e.g., artifacts from metallic implants, motion). MDCT All patients were scanned with a 64-MDCT scanner (Aquilion 64, Toshiba Medical Solutions). After pulmonary CTA was performed, CTV was performed in the craniocaudal direction from the diaphragm to the ankles during a single inspiratory breath-hold. CTV was performed approximately 3 minutes 30 seconds after the IV injection of contrast material began. The scanning parameters were 120 kvp (standard-kvp group) or 100 kvp (low-kvp group), a 0.5-mm collimation, a 0.5-second rotation time, a pitch of 53 or 64, 1-mm-thick reconstructions, and 0.8-mm reconstruction increments with automatic tube current modulation. Automatic tube current modulation (z-axis modulation with Real E.C. technique, Toshiba Medical Solutions) was used with the noise level set at 8 SD for the 120-kVp group and at 10 SD for the 100-kVp group. The noise level settings were based on a phantom study performed at our institution. A different tube current was needed to achieve the noise level at 8 SD as follows: 120 kvp, 50 ma; 100 kvp, 80 ma; and 80 kvp, 175 ma. When the noise level was set at 10 SD, the 100-kVp protocol needed 50 ma. Considering CT tube unloading, 100 kvp with the noise level set at 10 SD was valid. Although the signal-to-noise ratio (SNR) of the 100-kVp protocol with the noise level at 10 SD was higher than the SNR of the 120-kVp protocol with the noise level at 8 SD, the contrast-to-noise ratio of the 100-kVp protocol with the noise level at 10 SD was almost equal to that of the 120-kVp protocol with the noise level at 8 SD. The SNR of the 100-kVp protocol with the noise level at 8 SD was lower than at 10 SD, but the low noise level setting increases radiation dose. For all studies, nonionic contrast material (iohexol, Omnipaque 300, GE Healthcare) was administered via an antecubital vein with an automatic power injector. We used 150 ml of iohexol for patients who weighed more than 50 kg and 3 ml/kg of body weight of iohexol for patients who weighed 50 kg or less. The first 80 ml of contrast material was injected at a rate of 3 4 ml/s, and the remaining contrast material was injected at a rate of 1.5 ml/s. Individual contrast optimization was achieved using bolus tracking in the main pulmonary artery. Assessment of Vascular Enhancement and Image Quality All images were reviewed and interpreted on PACS workstations (Ziostation, Ziosoft). Image analysis was performed using transverse images with a window level of 100 HU and a window width of 450 HU. Vascular enhancement was evaluated quantitatively using attenuation values in the inferior vena cava (IVC) at the level of the left renal vein, right femoral vein at the level of the femoral head, and right popliteal vein at the level of the popliteal fossa. The mean attenuation values for each vein were measured using a region of interest (ROI). For the evaluation of vascular enhancement, two radiologists (one with 4 years of experience in chest CT and the other with 17 years of experience) who were blinded to the kvp setting placed the ROIs on vessels by consensus and manually drew an elliptic ROI within the inner edge of each vessel. If a clot in the vessel or severe streak artifacts from a metallic prosthesis were present, they measured the vein on the left side. Image noise was objectively quantified by measuring the SD of the attenuation value in an ROI of each vein. The same two radiologists scored overall image quality by consensus using a 5-point scale (Figs. 1 and 2) as follows: 1, nondiagnostic (no diagnosis possible); 2, poor (inadequate to allow diagnosis of the presence or absence of a clot); 3, fair (optimal enhancement sufficient for diagnosis); 4, good (optimal enhancement to allow confident diagnosis of the presence or absence of a clot); or 5, excellent (optimal enhancement superior to a score of 4 that allows confident diagnosis of the presence or absence of a clot). In addition, a 5-point scale for the evaluation of subjective image quality based on previous reports was adopted [7, 8, 12]. The radiologists who scored image quality were blinded to the kvp setting. Images were selected randomly from both patient groups. Regarding image noise in the abdomen, reviewers also analyzed image quality of the liver as follows: 1, nondiagnostic (no diagnosis possible); 2, poor (confidence in making diagnosis degraded); 3, fair (moderate but sufficient for diagnosis); 4, good; or 5, excellent (enabling excellent differentiation of even small structures). The reviewers were blinded to the kvp setting used to obtain the images. Fig year-old woman with liver cirrhosis who presented for imaging to evaluate for suspected pulmonary embolism. Axial 100-kVp CT venography image obtained at level of inferior vena cava was assigned image quality score of 5 (i.e., excellent, optimal enhancement superior to a score of 4 that allows confident diagnosis of presence or absence of clot superior to a score of 4). Image shows very tortuous splenic venous system. AJR:197, October

3 Fujikawa et al. Image noise in a low-kv technique is influenced by BMI [13]. We analyzed BMI versus noise in each patient group using the Spearman rank correction. BMI was correlated with noise in each group and in each ROI position. Assessment of Attenuation Values for Deep Venous Thrombosis The detection of DVT was based on CTV images. The mean attenuation value for each DVT was measured using an ROI. For the evaluation of vascular enhancement, we manually drew an elliptic ROI within the clots in each vessel. If multiple foci of DVT were present in the same patient, we chose the largest clot. Evaluation of Radiation Dose To compare radiation exposure from CTV performed using the standard 120-kVp protocol with that from CTV performed using the low 100-kVp protocol, we recorded the dose-length product (DLP) as a CT radiation dose descriptor. The DLP was provided by the scanner system. The DLPs of 35 patients in the low-kvp group were measured using both the 100 and 120 kvp settings. Estimated DLP measurements did not include pulmonary CTA. A Fig. 2 Axial 100-kVp CT venography images obtained at level of femoral vein. A, 47-year-old woman with suspected pulmonary embolism. Image quality score of 4 (good, optimal enhancement to allow confident diagnosis of presence or absence of clot) was assigned. B, 73-year-old woman with suspected pulmonary embolism. Image quality score of 3 (fair, optimal enhancement, sufficient for diagnosis) was assigned. C, 73-year-old woman with suspected pulmonary embolism. Image quality score of 2 (poor, inadequate to allow diagnosis of presence or absence of clot) as assigned. Statistical Analysis Statistical analysis was performed using commercially available software (JMP, version 5.1, SAS Institute). A p < 0.05 was considered to indicate a statistically significant difference. The Wilcoxon rank test was applied to determine whether venous enhancement in each vein showed significant differences between the standard-kvp group and the low-kvp group and was used to compare the scoring of image quality between groups. The Wilcoxon rank test was also used to compare patient characteristics including age, sex, BMI, the number of measurable veins, and the prevalence of venous thrombi between the two groups. In addition, differences in vascular enhancement, image noise, and image quality between patients with and those without DVT were also tested using the Wilcoxon rank test. The Spearman rank correction was applied to determine whether BMI was correlated with the noise in each group and in each ROI position. Results There were no statistically significant differences in age, sex, or BMI between the standard-kvp and low-kvp groups (Table 1). The IVC, femoral vein, and popliteal vein were measurable in all patients (standard-kvp group, n = 50; low-kvp group, n = 50). Four femoral veins (standard-kvp group, n = 2; low-kvp group, n = 2) and two popliteal veins (standard-kvp group, n = 1; lowkvp group, n = 1) were measured on the left side because of severe artifacts or a diffuse thrombus on the right side. Eleven patients were excluded from this evaluation. One exclusion was based on patient age and nine, on the presence of metallic artifact. The final case was excluded because the patient had diffuse clots in both legs. There were no exclusions because of amputation. TABLE 1: Patient Characteristics B There were 20 DVTs (40% of patients) and 14 PEs (28%) in the standard-kvp group and 19 DVTs (38% of patients) and 10 PEs (20%) in the low-kvp group. There was no significant difference in the prevalence of DVTs or PEs between the two groups (DVT, p = 0.83; PE, p = 0.34). All 33 DVTs (standard-kvp group, n = 18; low-kvp group, n = 15) were measurable. The IVC, femoral vein, and popliteal vein showed significantly higher attenuation values in the low-kvp group than in the standardkvp group (p < 0.001) (Table 2). The mean attenuation value in the IVC was ± 12.2 HU in the standard-kvp group and ± 17.4 HU in the low-kvp group; in the femoral vein, ± 18.0 HU in the standard-kvp group and ± 24.5 HU in the low-kvp group; and in the popliteal vein, ± 17.5 HU in the standard-kvp group and ± 26.0 HU in the low-kvp group. The mean attenuation value of the DVTs was 63.8 ± 3.8 HU in the standard-kvp group and 64.7 ± 6.1 HU in the low-kvp group. There was no significant difference between the two groups (p = 0.928). Image noise in all measured veins in the low-kvp group was significantly higher than Characteristic 120 kvp (n = 50) 100 kvp (n = 50) p Age (y) 1 Mean ± SD 64 ± ± 15 Sex (no. of patients) 0.64 Male-female ratio 14:36 12:38 Body mass index 0.92 Mean ± SD 22.9 ± ± 3.5 With pulmonary embolism (no. of patients) With deep venous thrombosis (no. of patients) C 840 AJR:197, October 2011

4 Vascular Enhancement and Image Quality of CTV TABLE 2: Vascular Enhancement in Standard Peak Kilovoltage (kvp) and Low-kVp Groups Vascular Enhancement (HU) Region of Interest 120 kvp (n = 50) 100 kvp (n = 50) p Inferior vena cava ± ± < Femoral vein ± ± < Popliteal vein ± ± < Note Vascular enhancement in all measured veins showed significantly higher attenuation values in the low-kvp group than in the standard-kvp group (p < 0.001). TABLE 3: Image Noise in Standard Peak Kilovoltage (kvp) and Low-kVp Groups Vascular Enhancement (HU) Region of Interest 120 kvp (n = 50) 120 kvp (n = 50) p Inferior vena cava ± ± 8.37 < Femoral vein ± ± 8.35 < Popliteal vein ± ± Note Image noise in the low-kvp group was significantly higher than that in the standard-kvp group. that in the standard-kvp group (Table 3). However, there were no significant differences in the subjective score of image quality at the level of the IVC (p = 0.245) or the popliteal vein (p = 0.487). At the level of the femoral vein, the visual score of the low-kvp group (mean ± SD, 3.92 ± 0.63) was significantly lower than that of the standard-kvp group (4.18 ± 0.63); however, this difference between the groups was relatively weak (p = ) (Table 4). The mean image quality score of the liver was 4.3 ± 0.7 in the standard-kvp group and 4.12 ± 0.6 in the low-kvp group. This difference between the two groups was significant (p = 0.02). BMI was correlated with noise in each ROI position in the 120-kVp group (IVC vs BMI, p = ; femoral vein vs BMI, p = ; popliteal vein vs BMI, p = ) and in the 100-kVp group (IVC vs BMI, p = ; femoral vein vs BMI, p = ; and popliteal vein vs BMI, p = ). The correlations between noise and BMI in each group were moderate to strong. The mean DLP in the low-kvp group (1202 ± mgy cm; range, mgy cm) was significantly lower than that in the standard-kvp group ( ± mgy cm; range, mgy cm) (p < 0.001). Discussion In our study, vascular enhancement on CTV was greater with the low kvp setting than with the standard kvp setting. Although the image noise was higher in the low-kvp group than in the standard-kvp group, the mean visual scores of image quality at the level of the IVC and the popliteal vein were not statistically different between the two groups. Moreover, the CT radiation dose of CTV was significantly lower in the low-kvp group than in the standard-kvp group. To our knowledge, there have been no previous investigations of the vascular enhancement, image quality, and radiation dose of CTV with a low kvp setting. TABLE 4: Image Quality Score in Standard Peak Kilovoltage (kvp) and Low-kVp Groups Image Quality Score Region of Interest 120 kvp (n = 50) 100 kvp (n = 50) p Inferior vena cava 4.54 ± ± Femoral vein 4.18 ± ± Popliteal vein 4.38 ± ± Note Regarding image quality scores for the inferior vena cava and popliteal vein, there was no significant difference in subjective scores between the two patient groups, whereas the image quality scores for the femoral vein deteriorated in the low-kvp group compared with the standard-kvp group (p = ). There are several advantages to reducing the kvp setting in CTV. Lowering the kvp setting is an effective method of reducing the radiation dose [14]. Moreover, lowering the kvp setting leads to increased vascular enhancement because the attenuation of iodine-based contrast agents increases with reduced x-ray energy distribution. This increase in attenuation occurs because of the high relative atomic number of iodine [11] as well as the k-edge in such energy levels. Indeed, in this study, the mean attenuation values in all measured veins with the low kvp setting were significantly higher than those in the standard-kvp group. These results are consistent with the findings of previous pulmonary CTA studies [7, 8, 12]. The contrast between clot and venous enhancement also could be improved by lowering the kvp setting because the mean attenuation value of DVTs did not differ significantly between the two groups. The disadvantage of a lower kvp setting is the increase in image noise. Boone et al. [15] evaluated the relationship between image noise and radiation dose reduction in CT and reported that image noise increases at lower tube current (ma) and lower kvp settings. In our study, image noise was significantly higher in the low-kvp group than in the standard-kvp group. However, no significant difference in image quality at the level of the IVC and the popliteal vein between the groups was observed. Previous studies also found no significant degradation in image quality after lowering the kvp setting for pulmonary CTA [7, 8]. We used 1-mm image reconstructions for CTV because that is routine practice at our institution. However, if we had used thicker images, the image quality would have been higher and the image noise would have been lower. Regarding image noise in the abdomen, we also analyzed the image quality score of the liver. Although the mean image quality score of the liver was statistically lower in the low-kvp group than in the standard-kvp group (p = 0.02), the mean image quality score of the low-kvp group was This high mean image quality score shows that the deterioration of image quality did not much influence diagnostic accuracy. Nakayama et al. [13] reported that image noise is mainly related to the characteristics of the CT scanner used and to the characteristics of individual patients, such as body habitus. They evaluated the relationship between SNR of the aorta and the patient s body weight at 120 and 90 kvp settings and AJR:197, October

5 Fujikawa et al. reported that a technique with a low kvp should be used in patients weighing less than 80 kg. In our study, only one patient in each group weighed more than 80 kg, and there was no statistically difference in BMI between two groups (p = 0.917). In our study, the correlations between noise and BMI in each group were moderate to strong. Noise is known to increase at low kvp settings, so we should be more concerned about BMI when using a low kvp setting. In contrast, in our study, there was a weak but significant difference in image quality at the femoral vein between the two groups. However, the mean image quality score for the 100-kVp group was 3.92, which was sufficient for diagnosis. Dual-energy CT enables improved detection of small PEs, particularly by emphasizing the presence of subsegmental pulmonary iodine mapping defects. Probably the same result would be seen on CTV. In addition, dual-energy CT can remove bones and intraluminal plaques from angiography datasets on the basis of spectral differentiation separating iodine from calcium and can reduce metal artifact. In our study, artifacts from bones and metallic prostheses were suspected to cause deterioration in image quality of the femoral vein; dual-energy CT would solve these problems. The use of CTV for the diagnosis of PE is controversial, in part because the additional pelvic radiation dose from CTV is of concern. Although some researchers have argued that imaging the lower extremities of patients with PE is not necessary [16 18], a moderate potential benefit is probable for high-risk patients with signs or symptoms of DVT or a history of DVT [6, 19 21]. Some reports have also suggested that CTV should not be routinely performed in all patients evaluated for PE but that CTV may be useful only in patients with a high probability of PE. Sonography is the primary imaging modality for imaging the lower extremities for the diagnosis of DVT [22]. Although sonography of the lower extremities has approximately the same diagnostic accuracy as CTV for the detection of DVT [23], the improvement in contrast enhancement obtained with the low kvp setting in CTV may enhance the diagnostic accuracy of CTV for DVT. Moreover, although the absence of radiation exposure in sonography may render this modality preferable to detect DVT, our data indicate that a radiation dose reduction can be achieved by lowering the kvp setting in CTV. In addition, recent studies have shown that radiation from CTV can be reduced by 75% or more using a modified CTV technique [6, 20, 24]. Goodman et al. [6] reported that radiation from CTV can be reduced by 75% or more using discontinuous imaging, reduced anatomic coverage, and automated milliampere (ma) adjustments. Begemann et al. [25] reported that a wider collimation can reduce radiation dose. Thus, CTV with a lower kvp setting in conjunction with such modified techniques could reduce radiation exposure. The disadvantages of sonography include dependence on the operator s skill and cost. Our study has several limitations. First, the sample size of our study was relatively small and our results were obtained from a retrospective analysis. To confirm these findings a prospective analysis with a large patient sample is required. Second, cardiac function may affect vascular enhancement at CTV. Garg et al. [26] reported that complex hemodynamic factors due to a combination of abnormal cardiac contractility and peripheral arterial disease result in poor delivery of contrast bolus to the deep venous system. In particular, massive PE leads to severe right cardiac failure. We did not consider the effect of cardiac function in each patient; however, there was no statically significant difference in the observed PEs and DVTs between the standard-kvp and low-kvp groups. Thus, we speculate that the influence of the difference in cardiac function between the two groups is minimal. Third, sample size was relatively small for the evaluation of the attenuation value of clots because the attenuation of clots has been reported to be variable [1, 27]. If we had tried to make a statistically significant observation about the density of clots between the two groups, a larger sample would have been needed. Baldt et al. [28] reported that the attenuation of clots showed a trend of negative correlation with days after the onset of disease. Clots imaged within 8 days of disease onset showed an average attenuation of 66 HU, whereas those that had been present more than 8 days showed an average attenuation of 55 HU. In consideration of these reports, a larger sample size would be needed to evaluate the attenuation of clots between the two groups. In conclusion, we confirmed that CTV with the low kvp setting improves vascular enhancement with acceptable image quality for the detection of DVT. We recommend the CTV protocol with a low kvp for the evaluation of DVT in clinical practice. References 1. Loud PA, Katz DS, Klippenstein DL, Shah RD, Grossman ZD. Combined CT venography and pulmonary angiography in suspected thromboembolic disease: diagnostic accuracy for deep venous evaluation. AJR 2000; 174: Stein PD, Fowler SE, Goodman LR, et al. Multidetector computed tomography for acute pulmonary embolism. N Engl J Med 2006; 354: Bischoff B, Hein F, Meyer T, et al. Impact of a reduced tube voltage on CT angiography and radiation dose: results of the PROTECTION I study. JACC Cardiovasc Imaging 2009; 2: Hurwitz LM, Yoshizumi TT, Goodman PC, et al. Radiation dose savings for adult pulmonary embolus 64-MDCT using bismuth breast shields, lower peak kilovoltage, and automatic tube current modulation. AJR 2009; 192: Kalva SP, Sahani DV, Hahn PF, Saini S. Using the K-edge to improve contrast conspicuity and to lower radiation dose with a 16-MDCT: a phantom and human study. J Comput Assist Tomogr 2006; 30: Goodman LR, Stein PD, Beemath A, et al. CT venography for deep venous thrombosis: continuous images versus reformatted discontinuous images using PIOPED II data. AJR 2007; 189: Heyer CM, Mohr PS, Lemburg SP, Peters SA, Nicolas V. Image quality and radiation exposure at pulmonary CT angiography with 100- or 120- kvp protocol: prospective randomized study. Radiology 2007; 245: Schueller-Weidekamm C, Schaefer-Prokop CM, Weber M, Herold CJ, Prokop M. CT angiography of pulmonary arteries to detect pulmonary embolism: improvement of vascular enhancement with low kilovoltage settings. Radiology 2006; 241: Wintermark M, Maeder P, Verdun FR, et al. Using 80 kvp versus 120 kvp in perfusion CT measurement of regional cerebral blood flow. AJNR 2000; 21: Huda W, Scalzetti EM, Levin G. Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 2000; 217: Sigal-Cinqualbre AB, Hennequin R, Abada HT, Chen X, Paul JF. Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 2004; 231: Matsuoka S, Hunsaker AR, Gill RR, et al. Vascular enhancement and image quality of MDCT pulmonary angiography in 400 cases: comparison of standard and low kilovoltage settings. AJR 2009; 192: Nakayama Y, Awai K, Funama Y, et al. Abdominal 842 AJR:197, October 2011

6 Vascular Enhancement and Image Quality of CTV CT with low tube voltage: preliminary observa- C. Compression sonography in patients with inde- pelvis be imaged? Radiology 2008; 246: tions about radiation dose, contrast enhancement, terminate or low-probability lung scans: lack of 25. Begemann PG, Bonacker M, Kemper J, et al. image quality, and noise. Radiology 2005; 237: usefulness in the absence of both symptoms of Evaluation of the deep venous system in patients deep-vein thrombosis and thromboembolic risk with suspected pulmonary embolism with multi- 14. Kubo T, Lin PJ, Stiller W, et al. Radiation dose factors. AJR 1996; 166: detector CT: a prospective study in comparison to reduction in chest CT: a review. AJR 2008; 190: 20. Hunsaker AR, Zou KH, Poh AC, et al. Routine Doppler sonography. J Comput Assist Tomogr Boone JM, Geraghty EM, Seibert JA, Wootton- Gorges SL. Dose reduction in pediatric CT: a rational approach. Radiology 2003; 228: Perrier A, Bounameaux H. Accuracy or outcome in suspected pulmonary embolism. N Engl J Med 2006; 354: Perrier A, Roy PM, Sanchez O, et al. Multidetectorrow computed tomography in suspected pulmonary embolism. N Engl J Med 2005; 352: Righini M, Le Gal G, Aujesky D, et al. Diagnosis of pulmonary embolism by multidetector CT alone or combined with venous ultrasonography of the leg: a randomised non-inferiority trial. Lancet 2008; 371: Rosen MP, Sheiman RG, Weintraub J, McArdle pelvic and lower extremity CT venography in patients undergoing pulmonary CT angiography. AJR 2008; 190: Dodd JD. Evidence-based practice in radiology: steps 3 and 4 appraise and apply diagnostic radiology literature. Radiology 2007; 242: Zierler BK. Ultrasonography and diagnosis of venous thromboembolism. Circulation 2004; 109: I9 I Goodman LR, Stein PD, Matta F, et al. CT venography and compression sonography are diagnostically equivalent: data from PIOPED II. AJR 2007; 189: Kalva SP, Jagannathan JP, Hahn PF, Wicky ST. Venous thromboembolism: indirect CT venography during CT pulmonary angiography should the 2003; 27: Garg K, Kemp JL, Russ PD, Baron AE. Thromboembolic disease: variability of interobserver agreement in the interpretation of CT venography with CT pulmonary angiography. AJR 2001; 176: Cham MD, Yankelevitz DF, Shaham D, et al. Deep venous thrombosis: detection by using indirect CT venography The Pulmonary Angiography-Indirect CT Venography Cooperative Group. Radiology 2000; 216: Baldt MM, Zontsich T, Stumpflen A, et al. Deep venous thrombosis of the lower extremity: efficacy of spiral CT venography compared with conventional venography in diagnosis. Radiology 1996; 200: AJR:197, October

Detection of deep venous thrombosis on CT venography: Comparison of dose of contrast material

Detection of deep venous thrombosis on CT venography: Comparison of dose of contrast material Detection of deep venous thrombosis on CT venography: Comparison of dose of contrast material Poster No.: C-1140 Congress: ECR 2010 Type: Scientific Exhibit Topic: Contrast Media Authors: T. Ichikawa,

More information

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Radiation Exposure in Pregnancy John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Illustrative Clinical Scenario 32 year old female 34 weeks pregnant with recent onset shortness of breath and central chest pain

More information

Pulmonary Embolism. Thoracic radiologist Helena Lauri

Pulmonary Embolism. Thoracic radiologist Helena Lauri Pulmonary Embolism Thoracic radiologist Helena Lauri 8.5.2017 Statistics 1-2 out of 1000 adults annually are diagnosed with deep vein thrombosis (DVT) and/or pulmonary embolism (PE) About half of patients

More information

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T. The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume

More information

Eun-Suk Cho, MD, Jae-Joon Chung, MD, Sungjun Kim, MD, Joo Hee Kim, MD, Jeong-Sik Yu, MD, Choon-Sik Yoon, MD

Eun-Suk Cho, MD, Jae-Joon Chung, MD, Sungjun Kim, MD, Joo Hee Kim, MD, Jeong-Sik Yu, MD, Choon-Sik Yoon, MD Original Article Cardiovascular Imaging http://dx.doi.org/1.3348/kjr.213.14.2.183 pissn 1229-6929 eissn 25-833 Korean J Radiol 213;14(2):183-193 CT Venography for Deep Vein Thrombosis Using a Low Tube

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

ACR Appropriateness Criteria Suspected Lower Extremity Deep Vein Thrombosis EVIDENCE TABLE

ACR Appropriateness Criteria Suspected Lower Extremity Deep Vein Thrombosis EVIDENCE TABLE . Fowkes FJ, Price JF, Fowkes FG. Incidence of diagnosed deep vein thrombosis in the general population: systematic review. Eur J Vasc Endovasc Surg 003; 5():-5.. Hamper UM, DeJong MR, Scoutt LM. Ultrasound

More information

PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT OF VASCULAR OCCLUSION EXTENT AND LOCALIZATION OF EMBOLI 1. BACKGROUND

PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT OF VASCULAR OCCLUSION EXTENT AND LOCALIZATION OF EMBOLI 1. BACKGROUND JOURNAL OF MEDICAL INFORMATICS & TECHNOLOGIES Vol. 11/2007, ISSN 1642-6037 Damian PTAK * pulmonary embolism, AngioCT, postprocessing techniques, Mastora score PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT

More information

Missed Pulmonary Embolism on Abdominal CT

Missed Pulmonary Embolism on Abdominal CT Cardiopulmonary Imaging Original Research Lim et al. Pulmonary Emboli Missed on CT of the Abdomen Cardiopulmonary Imaging Original Research Kun Young Lim 1 Seth J. Kligerman 2 Cheng Ting Lin 2 Charles

More information

CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings

CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings Poster No.: C-3289 Congress: ECR 2010 Type: Scientific Exhibit Topic: Chest - Your latest results Authors: M. K. Gill,

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

Dual Energy CT Aortography: Can We Reduce Iodine Dose??

Dual Energy CT Aortography: Can We Reduce Iodine Dose?? Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle   holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/19768 holds various files of this Leiden University dissertation. Author: Langevelde, Kirsten van Title: Are pulmonary embolism and deep-vein thrombosis

More information

CT angiography techniques. Boot camp

CT angiography techniques. Boot camp CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary

More information

Deep Vein Thrombosis: Can a Second Sonographic Examination Be Avoided?

Deep Vein Thrombosis: Can a Second Sonographic Examination Be Avoided? Alfonsa Friera 1 Nuria R. Giménez 2 Paloma Caballero 1 Pilar S. Moliní 2 Carmen Suárez 2 Received August 15, 2001; accepted after revision October 16, 2001. 1 Radiology Department, Hospital de la Princesa,

More information

Proper Diagnosis of Venous Thromboembolism (VTE)

Proper Diagnosis of Venous Thromboembolism (VTE) Proper Diagnosis of Venous Thromboembolism (VTE) Whal Lee, M.D. Seoul National University Hospital Department of Radiology 2 nd EFORT Asia Symposium, 3 rd November 2010, Taipei DVT - Risk Factors Previous

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

Qualitative evaluation of pulmonary CT angiography findings in pregnant and postpartum women with suspected pulmonary thromboembolism

Qualitative evaluation of pulmonary CT angiography findings in pregnant and postpartum women with suspected pulmonary thromboembolism Original Article Qualitative evaluation of pulmonary CT angiography findings in pregnant and postpartum women with suspected pulmonary thromboembolism Maryam Moradi, Leyla Jalali Monfared Department of

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

Optimizing radiation dose by varying age at pediatric temporal bone CT

Optimizing radiation dose by varying age at pediatric temporal bone CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Optimizing radiation dose by varying age at pediatric temporal bone CT Daichi Noto, 1 Yoshinori Funama, 2a Mika Kitajima, 3 Daisuke

More information

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation Petter Bugge Askeland Project thesis at the Faculty of Medicine UNIVERSITETET

More information

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Introduction. Methods and Materials

Introduction. Methods and Materials Radiology Insights A Clinical Evaluation of an Automated Software Program (Certegra P3T PA) for Patient Specific Contrast Injection During Chest CTA to Exclude Pulmonary Embolism Christopher R Deible MD,

More information

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling Investigations and research Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling J. Wang Z. Ying V. Yao L. Ciancibello S. Premraj S. Pohlman

More information

Pulmonary embolism: Strategies to optimize pulmonary MDCT angiography studies

Pulmonary embolism: Strategies to optimize pulmonary MDCT angiography studies Pulmonary embolism: Strategies to optimize pulmonary MDCT angiography studies Poster No.: C-0152 Congress: ECR 2013 Type: Educational Exhibit Authors: G. Viteri, A. Garcia-lallana, I. Simon Yarza, A. Villanueva

More information

FADAK PUBLICATIONS WWW.FADAKPUBLICATIONS.COM Persian Journal of Medical Sciences PJMS.FADAKPUBLICATIONS.COM Comparison of CT Pulmonary Angiography Quality at Low Tube Voltage (80 Kvp) With Standard Conditions

More information

Clinical Guide - Suspected PE (Reviewed 2006)

Clinical Guide - Suspected PE (Reviewed 2006) Clinical Guide - Suspected (Reviewed 2006) Principal Developer: B. Geerts Secondary Developers: C. Demers, C. Kearon Background Investigation of patients with suspected pulmonary emboli () remains problematic

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

Volumetric Perfusion CT Using Prototype 256 Detector Row CT Scanner: Preliminary Study with Healthy Porcine Model V

Volumetric Perfusion CT Using Prototype 256 Detector Row CT Scanner: Preliminary Study with Healthy Porcine Model V AJNR Am J Neuroradiol 26:2536 2541, November/December 2005 Technical Note Volumetric Perfusion CT Using Prototype 256 Detector Row CT Scanner: Preliminary Study with Healthy Porcine Model V Shinichiro

More information

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Original Article Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Shun-Yu Gao, Xiao-Peng Zhang, Yong Cui, Ying-Shi

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

BrightSpeed Elite CT with ASiR: Comparing Dose & Image Quality Rule Out Pulmonary Embolism on Initial & Follow-Up Exam

BrightSpeed Elite CT with ASiR: Comparing Dose & Image Quality Rule Out Pulmonary Embolism on Initial & Follow-Up Exam GE Healthcare BrightSpeed Elite CT with ASiR: Comparing Dose & Image Quality Rule Out Pulmonary Embolism on Initial & Follow-Up Exam Michael Swack, MD Diagnostic Radiologist Irvington Radiologists, PC

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Cardiac CT Techniques in Neonates (and infants)

Cardiac CT Techniques in Neonates (and infants) Cardiac CT Techniques in Neonates (and infants) Siddharth P. Jadhav, MD Director, Body CT and MRI Edward B. Singleton Department of Pediatric Radiology Texas Children s Hospital Disclosures None Objectives

More information

Evaluation of Deep Vein Thrombosis With Reduced Radiation and Contrast Material Dose at Computed Tomography Venography

Evaluation of Deep Vein Thrombosis With Reduced Radiation and Contrast Material Dose at Computed Tomography Venography Circulation Journal Official Journal of the Japanese Circulation Society http://www.j-circ.or.jp ORIGINAL ARTICLE Imaging Evaluation of Deep Vein Thrombosis With Reduced Radiation and Contrast Material

More information

Epidermiology Early pulmonary embolism

Epidermiology Early pulmonary embolism Epidermiology Early pulmonary embolism Sitang Nirattisaikul Faculty of Medicine, Prince of Songkla University 3 rd most common cause of cardiovascular death in the United States, following ischemic heart

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Vascular and Interventional Radiology Original Research

Vascular and Interventional Radiology Original Research Vascular and Interventional Radiology Original Research Apfaltrer et al. High-Pitch Versus Standard-Pitch CTA of the Aorta Vascular and Interventional Radiology Original Research Paul Apfaltrer 1,2 E.

More information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School OUTLINE 1 st Presentation: Dose optimization strategies

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/40114 holds various files of this Leiden University dissertation Author: Exter, Paul L. den Title: Diagnosis, management and prognosis of symptomatic and

More information

High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous?

High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous? High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous? Poster No.: C-1753 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit B. Y. Lee, H. R. KIM, J. I. Jung,

More information

Bottom up cardiac CT for CABG assessment to resolve breathing artefact

Bottom up cardiac CT for CABG assessment to resolve breathing artefact Bottom up cardiac CT for CABG assessment to resolve breathing artefact Poster No.: C-0589 Congress: ECR 2010 Type: Educational Exhibit Topic: Cardiac Authors: P. Glass, P. Donnelly, P. Hanley, D. Higginson,

More information

A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction.

A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction. Biomedical Research 2017; 28 (17): 7683-7687 ISSN 0970-938X www.biomedres.info A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction. Yanhe

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Minimizing CT Radiation Dose CT examinations improve health care and are an essential part

More information

A low probability interpretation of a ventilation/

A low probability interpretation of a ventilation/ Very Low Probability Interpretation of V/Q Lung Scans in Combination with Low Probability Objective Clinical Assessment Reliably Excludes Pulmonary Embolism: Data from PIOPED II Alexander Gottschalk 1,

More information

The Location and Size of Pulmonary Embolism in Antineoplastic Chemotherapy Patients 1

The Location and Size of Pulmonary Embolism in Antineoplastic Chemotherapy Patients 1 The Location and Size of Pulmonary Embolism in Antineoplastic Chemotherapy Patients 1 Yun Joo Park, M.D., Woocheol Kwon, M.D., Won-Yeon Lee, M.D. 2, Sang Baek Koh, M.D. 3, Seong Ah Kim, M.D., Myung Soon

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Clinical Image Gallery Next Generation Volume 1

Clinical Image Gallery Next Generation Volume 1 Clinical Image Gallery Next Generation Volume 1 Dr. Russell Bull Royal Bournemouth Hospital, Bournemouth, United Kingdom After long experience with the first generation, a next generation Aquilion ONE

More information

Acute abdominal venous thromboses- the hyperdense noncontrast CT sign

Acute abdominal venous thromboses- the hyperdense noncontrast CT sign Acute abdominal venous thromboses- the hyperdense noncontrast CT sign Poster No.: C-1095 Congress: ECR 2011 Type: Educational Exhibit Authors: M. Goldstein, K. Jhaveri; Toronto, ON/CA Keywords: Abdomen,

More information

Imaging of acute pulmonary thromboembolism*

Imaging of acute pulmonary thromboembolism* Silva, Isabela et al. Imaging of acute pulmonary thromboembolism Imaging of acute pulmonary thromboembolism* C. ISABELA S. SILVA, NESTOR L. MÜLLER The diagnosis of acute pulmonary thromboembolism is based

More information

New Criteria for Ventilation-Perfusion Lung Scan Interpretation: A Basis for Optimal Interaction with Helical CT Angiography 1

New Criteria for Ventilation-Perfusion Lung Scan Interpretation: A Basis for Optimal Interaction with Helical CT Angiography 1 1206 July-August 2000 RG Volume 20 Number 4 New Criteria for Ventilation-Perfusion Lung Scan Interpretation: A Basis for Optimal Interaction with Helical CT Angiography 1 Alexander Gottschalk, MD Introduction

More information

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference?

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference? Pediatric Imaging Original Research Reid et al. Parameter Selection for Pediatric Abdominal CT Downloaded from www.ajronline.org by 46.3.3.24 on 2/3/18 from IP address 46.3.3.24. Copyright ARRS. For personal

More information

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial Poster No.: C-0595 Congress: ECR 2013 Type: Authors:

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

CTA Chest Pulmonary Embolism & Routine CT Abdomen + Pelvis W CTA Chest W (arterial) & CT Abdomen + Pelvis W (venous)

CTA Chest Pulmonary Embolism & Routine CT Abdomen + Pelvis W CTA Chest W (arterial) & CT Abdomen + Pelvis W (venous) CTA Chest Pulmonary Embolism & Routine CT Abdomen + Pelvis W CTA Chest W (arterial) & CT Abdomen + Pelvis W (venous) Reviewed By: Anna Ellermeier, MD; Brett Mollard, MD Last Reviewed: August 2018 Contact:

More information

FOR CMS (MEDICARE) MEMBERS ONLY NATIONAL COVERAGE DETERMINATION (NCD) FOR COMPUTED TOMOGRAPHY:

FOR CMS (MEDICARE) MEMBERS ONLY NATIONAL COVERAGE DETERMINATION (NCD) FOR COMPUTED TOMOGRAPHY: National Imaging Associates, Inc. Clinical guidelines CHEST CTA Original Date: September 1997 Page 1 of 5 CPT Codes: 71275 Last Review Date: August 2014 NCD 220.1 Last Effective Date: March 2008 Guideline

More information

Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit detector dual-source computed tomography

Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit detector dual-source computed tomography Acta Radiol OnlineFirst, published on June 11, 2014 as doi:10.1177/0284185114535391 Original Article Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit

More information

Kevin M. Duwe 1 Maria Shiau Nancy E. Budorick John H. M. Austin Yahya M. Berkmen

Kevin M. Duwe 1 Maria Shiau Nancy E. Budorick John H. M. Austin Yahya M. Berkmen 2000 RRS Executive Council ward I Kevin M. Duwe 1 Maria Shiau Nancy E. udorick John H. M. ustin Yahya M. erkmen Received February 24, 2000; accepted after revision June 5, 2000. 1 ll authors: Department

More information

Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography.

Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography. ISPUB.COM The Internet Journal of Radiology Volume 11 Number 2 Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography. T Niemann, G Bongartz Citation T Niemann, G Bongartz. Detectability

More information

Cardiovascular Imaging in Pregnancy. Diana Litmanovich, MD

Cardiovascular Imaging in Pregnancy. Diana Litmanovich, MD Cardiovascular Imaging in Pregnancy Diana Litmanovich, MD Cardiovascular disorders 4% of all pregnancies PE DVT Aortic dissection Peripartum cardiomyopathy Acute coronary syndrome Hypertension Pneumonia

More information

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination Int J Clin Exp Med 2015;8(11):21898-21905 www.ijcem.com /ISSN:1940-5901/IJCEM0015005 Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out

More information

General Imaging. Imaging modalities. Incremental CT. Multislice CT Multislice CT [ MDCT ]

General Imaging. Imaging modalities. Incremental CT. Multislice CT Multislice CT [ MDCT ] General Imaging Imaging modalities Conventional X-rays Ultrasonography [ US ] Computed tomography [ CT ] Radionuclide imaging Magnetic resonance imaging [ MRI ] Angiography conventional, CT,MRI Interventional

More information

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Poster No.: C-0863 Congress: ECR 2015 Type: Scientific Exhibit Authors: A. Nakamoto,

More information

Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT

Organ-Based Dose Current Modulation and Thyroid Shields: Techniques of Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Hoang et al. Radiation Dose Reduction for Neck CT Medical Physics and Informatics Original Research Jenny K. Hoang 1,2 Terry T. Yoshizumi 1 Kingshuk Roy

More information

Anterior Spinal Artery and Artery of Adamkiewicz Detected by Using Multi-Detector Row CT

Anterior Spinal Artery and Artery of Adamkiewicz Detected by Using Multi-Detector Row CT AJNR Am J Neuroradiol 24:13 17, January 2003 Anterior Spinal Artery and Artery of Adamkiewicz Detected by Using Multi-Detector Row CT Kohsuke Kudo, Satoshi Terae, Takeshi Asano, Masaki Oka, Kenshi Kaneko,

More information

Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT

Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT YAO ET AL ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 6, 252 257 p.p., 2003 Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/40114 holds various files of this Leiden University dissertation Author: Exter, Paul L. den Title: Diagnosis, management and prognosis of symptomatic and

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760 Can Weight-Adapted IV Contrast Media Protocols Reduce Iodine Dose and Still Produce a Diagnostic Contrast Enhancement Level in Abdomino-Pelvic CT Scans? Poster No.: C-1760 Congress: ECR 2015 Type: Authors:

More information

Citation for published version (APA): Douma, R. A. (2010). Pulmonary embolism: advances in diagnosis and prognosis

Citation for published version (APA): Douma, R. A. (2010). Pulmonary embolism: advances in diagnosis and prognosis UvA-DARE (Digital Academic Repository) Pulmonary embolism: advances in diagnosis and prognosis Douma, R.A. Link to publication Citation for published version (APA): Douma, R. A. (2010). Pulmonary embolism:

More information

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck

Comparative Analysis of Radiation Dose and Image Quality Between Thyroid Shielding and Unshielding During CT Examination of the Neck Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Young Hen Lee 1 Eun-tae Park 1 Pyong Kon Cho 1 Hyung Suk Seo 1 Bo-Kyung Je 1 Sang-il Suh 2 Kyung-Sook

More information

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could

More information

Lung Perfusion Analysis New Pathways in Lung Imaging. Case Study Brochure PLA 309 Hospital

Lung Perfusion Analysis New Pathways in Lung Imaging. Case Study Brochure PLA 309 Hospital Lung Perfusion Analysis New Pathways in Lung Imaging Case Study Brochure PLA 309 Hospital http://www.toshibamedicalsystems.com Toshiba Medical Systems Corporation 2012 all rights reserved. Design and specifications

More information

Spectral CT imaging as a new quantitative tool? Assessment of perfusion defects of pulmonary parenchyma in patients with lung cancer

Spectral CT imaging as a new quantitative tool? Assessment of perfusion defects of pulmonary parenchyma in patients with lung cancer Original Article Spectral CT imaging as a new quantitative tool? Assessment of perfusion defects of pulmonary parenchyma in patients with lung cancer Ying-Shi Sun, Xiao-Yan Zhang, Yong Cui, Lei Tang, Xiao-Ting

More information

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation Poster No.: C-2014 Congress: ECR 2014 Type: Educational Exhibit Authors: M. A. Ottesen; Oslo/NO Keywords: Cardiac, Arteries / Aorta, CT, CT-Angiography,

More information

Imaging of DVT: A Multi-modality Overview

Imaging of DVT: A Multi-modality Overview Imaging of DVT: A Multi-modality Overview Douglas S. Katz, MD Imaging of DVT: A Multi-modality Overview STR 2010 Annual Meeting Workshop Douglas S. Katz, M.D., F.A.C.R. Vice Chair for Clinical Research

More information

Computed tomography findings in 10 cases of iliac vein compression (May Thurner) syndrome

Computed tomography findings in 10 cases of iliac vein compression (May Thurner) syndrome European Journal of Radiology 55 (2005) 421 425 Computed tomography findings in 10 cases of iliac vein compression (May Thurner) syndrome Levent Oguzkurt a,, Fahri Tercan a, M. Ali Pourbagher a, Osman

More information

C h a p t e r T h r e e

C h a p t e r T h r e e Dose Reduction and Image Quality: Reducing kvp and Increasing mas Settings for CT Angiography of the Circle of Willis C h a p t e r T h r e e A Waaijer, MS van Leeuwen, BK Velthuis, CJG Bakker, GAP de

More information

P ulmonary embolism (PE) is a common disease estimated to

P ulmonary embolism (PE) is a common disease estimated to 53 PULMONARY EMBOLISM Prospective evaluation of unsuspected pulmonary embolism on contrast enhanced multidetector CT (MDCT) scanning Gillian Ritchie, Simon McGurk, Catriona McCreath, Catriona Graham, John

More information

CT Pulmonary Angiography 2010

CT Pulmonary Angiography 2010 2010 Annual Course CT Pulmonary Angiography 2010 Vassilios Raptopoulos, MD Beth Israel Deaconess Medical Center Harvard Medical School CTPA 2010 Utilization Technique DVT Radiation In Pregnancy Large Small

More information

Provider Led Entity. CDI Quality Institute PLE Chest / Pulmonary Embolus AUC 07/31/2018

Provider Led Entity. CDI Quality Institute PLE Chest / Pulmonary Embolus AUC 07/31/2018 Provider Led Entity CDI Quality Institute PLE Chest / Pulmonary Embolus AUC 07/31/2018 Appropriateness of advanced imaging procedures* in patients with suspected or known pulmonary embolus and the following

More information

Usefulness of Clinical Pre-test Scores for a Correct Diagnostic Pathway in Patients with Suspected Pulmonary Embolism in Emergency Room

Usefulness of Clinical Pre-test Scores for a Correct Diagnostic Pathway in Patients with Suspected Pulmonary Embolism in Emergency Room Send Orders for Reprints to reprints@benthamscience.net The Open Emergency Medicine Journal, 2013, 5, (Suppl 1: M-4) 19-24 19 Open Access Usefulness of Clinical Pre-test Scores for a Correct Diagnostic

More information

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Poster No.: B-0742 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Paper

More information

Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT

Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT The Harvard community has made this article openly available. Please share how this access benefits you. Your story

More information

CT SCAN PROTOCOL. Shoulder

CT SCAN PROTOCOL. Shoulder CT SCAN PROTOCOL Shoulder Purpose and Summary CT images made with this protocol are used to provide the orthopedic surgeon with a detailed 3D anatomical reconstruction of the patient s scapula and proximal

More information

Les Outils Cliniques de Demain en Scanner Cardiaque. Cardiaque Status en ECR 2018 From Diagnosis to Prognosis

Les Outils Cliniques de Demain en Scanner Cardiaque. Cardiaque Status en ECR 2018 From Diagnosis to Prognosis ECR 2018 From Diagnosis to Prognosis ECR 2018 From Diagnosis to Prognosis Thursday, March 1, 2018/08:30-10:00/Room N Les Outils Cliniques de Demain en Scanner Cardiaque Cardiaque Status en 2018 Rodrigo

More information

CT Versus MR for the Runoff

CT Versus MR for the Runoff CT Versus MR for the Runoff Robert R. Edelman, M.D. Dept. of Radiology NorthShore University HealthSystem Feinberg School of Medicine, Northwestern University Magnetic Resonance Computed Tomography Radio

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Mismetti P, Laporte S, Pellerin O, Ennezat P-V, Couturaud F, Elias A, et al. Effect of a retrievable inferior vena cava filter plus anticoagulation vs anticoagulation alone

More information

Imaging in Pulmonary Embolism. Gamal Rabie Agmy, MD,FCCP Professor of Chest Diseases, Assiut university

Imaging in Pulmonary Embolism. Gamal Rabie Agmy, MD,FCCP Professor of Chest Diseases, Assiut university Imaging in Pulmonary Embolism Gamal Rabie Agmy, MD,FCCP Professor of Chest Diseases, Assiut university Background Information Pulmonary embolism is a life-threatening condition that occurs when a clot

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

Pulmonary Nodules: Contrast- Enhanced Volumetric Variation at Different CT Scan Delays

Pulmonary Nodules: Contrast- Enhanced Volumetric Variation at Different CT Scan Delays Cardiopulmonary Imaging Original Research Rampinelli et al. Contrast-Enhanced CT of Pulmonary Nodules Cardiopulmonary Imaging Original Research Cristiano Rampinelli 1 Sara Raimondi 2 Mauro Padrenostro

More information