1 ASE ACR NEMA Digital Imaging and Communications in Medicine (DICOM) Supplement 45: Ultrasound Protocol Supplement Prepared by: DICOM Standards Committee, Working Group 12 Ultrasound 1300 N. 17 th Street Rosslyn, Virginia USA Version: 0.7

2 ii 8 June, 1999 ii

3 Supplement 45: Ultrasound Protocol Page i Foreword...iii Scope and Field of Application... iv A.1.4 Overview of the Composite IOD Module Content...2 A.6.4 US Image IOD Module Table...4 A.7.4 US Multi-Frame Image IOD Module Table...6 C.8.5.x US Series Module...7 C.8.5.y US Protocol Module REGISTRY OF DICOM DATA ELEMENTS...15

4 ii DOCUMENT HISTORY Document Version Date Content June, 1998 New draft based on February meeting output June, 1998 Added data elements to Part 6. Added note in C.8.5.y US Protocol Module June, 1998 Synchronized with Supplement 36. Added tables with preliminary picklists for microglossary inclusion November, 1998 Added supplement number (45). Added DICOMDIR support for Part 11. Made new modules optional. Defined relationship of ultrasound protocols to DICOM Real-World Model December, 1998 Clean up of Forward, Scope and Field of Application and page numbering December, 1998 Clarify relationship of ultrasound protocols to DICOM Real-World Model June, 1999 Updated with changes from 3/29/99 teleconference. Added informative SDM codes from ASE. OPEN ISSUES Finish defining picklists and context ID s. Get element numbers for new attributes. ii

5 Supplement 45: Ultrasound Protocol Page iii Foreword This draft Supplement to the DICOM Standard was developed according to NEMA Procedures. This Supplement to the Standard is developed in liaison with other standardization organizations including CEN TC251 in Europe, MEDIS-DC and JIRA in Japan, with review also by other organizations including IEEE, HL7 and ANSI in the USA. The DICOM Standard is structured as a multi-part document using the guidelines established in the following document: - ISO/IEC Directives, 1989 Part 3 : Drafting and Presentation of International Standards. This document is a Supplement to the DICOM Standard. It is an extension to PS 3.3, 3.6 and 3.11 of the published DICOM Standard which consists of the following parts: PS Introduction and Overview PS Conformance PS Information Object Definitions PS Service Class Specifications PS Data Structures and Encoding PS Data Dictionary PS Message Exchange PS Network Communication Support for Message Exchange PS Point-to-Point Communication Support for Message Exchange PS 3.10 PS 3.11 PS 3.12 PS 3.13 PS 3.14 Media Storage and File Format for Data Interchange Media Storage Application Profiles Media Formats and Physical Media for Data Interchange Print Management Point-to-Point Communication Support Gray-scale Display Function Standard These parts are related but independent documents.

6 iv Scope and Field of Application This Supplement to the DICOM Standard specifies changes to the DICOM Image Information Object Definition (IOD) for Ultrasound (US) and Ultrasound Multi-frame (USMF) Images. It specifies the new modules necessary to identify images acquired as part of an ultrasound acquisition protocol. Since this document proposes changes to existing Parts of DICOM the reader should have a working understanding of the Standard. This proposed Supplement includes a number of Addenda to existing Parts of DICOM: 1. PS 3.3 Addenda (Extension to Annex A and Annex C) 2. PS 3.6 Addenda (Extension to Section 6) 3. PS 3.11 Addenda (Extension to Annex C) iv

7 Supplement 45: Ultrasound Protocol Page 1 Changes to: NEMA Standards Publication PS Digital Imaging and Communications in Medicine (DICOM) Part 3: Information Object Definitions

8 2 3.X Ultrasound The following definitions are commonly used in the specification of the Ultrasound IOD s. 3.X.1 Acquisition Protocol An Acquisition Protocol is a Procedure Type that defines the particular ultrasound images that will be acquired. Note: These can be well known or locale specific. 3.X.2 Staged Protocol A Staged Protocol is an Acquisition Protocol that is repeated at more than one specific instance of time. Note: These instances commonly occur due to simple events. The introduction of contrast media, drugs or stress to the patient are typical events. Item: Add to PS 3.3 Table A.1-1 A.1.4 Overview of the Composite IOD Module Content Table A.1-1 provides an overview of the Modules used throughout the Composite IODs. This table is for informative purposes only. It is based on the IOD definitions found in the remaining sections of Annex A which are normative. Table A.1-1 COMPOSITE INFORMATION OBJECT MODULES OVERVIEW IODs Modules US USmf Patient M M Patient Summary General Study M M Patient study U U Study Content General Series CR Series NM Series M M US Series U U Frame Of Reference U S Frame of Ref. General Equipment SC Equipment U C M U C M

9 Supplement 45: Ultrasound Protocol Page 3 General Image Image Plane M* M Image Pixel M* M NM Image Pixel Pallet Color L o oku p Ta b le Contrast/ Bolus Cine Multi-frame C C C* C M M NM Multiframe Frame Pointers Mask Display Shutter Device Therapy CR Image CT Image MR Image NM Image NM Isotope NM Detector NM TOMO Acquisition NM Multi- Gated Acquisition NM Phase NM Re co n str u ct io n US Region Calibration U* U US Image M* M US Protocol U U SC Image X-Ray Image

10 4 X-Ray Acquisition X-Ray Collimator X-Ray Table XRF Positioner XRF Tomo Acquisition XA Positioner Bi-Plane Sequence Bi-Plane Image Overlay Identification Overlay Plane U* Multi-frame Overlay Bi-Plane Overlay Curve Identification M* M* Curve M* M* Audio U U Modality LUT VOI LUT U* U LUT Identification SOP Common M* M* Item: Add to PS 3.3 Table A for clarity, those modules that are changed from the existing standard or new are shown in Bold and Underlined A.6.4 US Image IOD Module Table Table A US IMAGE IOD MODULES IE Module Reference Usage

11 Supplement 45: Ultrasound Protocol Page 5 Patient Patient C M Study General Study C M Patient Study C U Series General Series C M Frame of Reference US Series C.8.5.x U Suggested if image is part of an acquisition protocol Frame of Reference C U US Frame of Reference C C - Required if images are spatially related Equipment General Equipment C M Image General Image C M (See A.6.4.1) Image Pixel C M Contrast/bolus C C - Required if contrast media was used in this image Palette Color Lookup table C.7.9 C - Required if Photometic Interpretation (0028,0004) has a value of PALETTE COLOR US Region Calibration C U US Image C M US Staged Protocol C.8.5.y U Suggested if image is part of a staged protocol Overplay Plane C.9.2 U VOI LUT C.11.2 U SOP Common C.12.1 M Curve Curve Identification C.10.1 M (See A.6.4.1) Curve C.10.2 M Audio C.10.3 U SOP Common C.12.1 M A Mutually Exclusive IEs The Image and Curve IEs are mutually exclusive. Each SOP Instance using this IOD shall contain exactly one of these IEs. A.6.5 US Image IOD Relationship to the DICOM Real-World Model Ultrasound images that make use of the US Series & Protocol Modules have a more specific relationship to the DICOM Real-World Model than non-protocol images. An ultrasound acquisition protocol is considered one Requested Procedure and has one Procedure Plan. The Procedure Plan will have one or more Action Items. An ultrasound acquisition protocol will typically consist

12 6 of one Scheduled Procedure Step containing one or more Action Items. Assuming the acquisition protocol was successfully completed, this will result in one Modality Performed Procedure Step. Therefore, all images that are part of this particular instance of an ultrasound acquisition protocol will have the same Procedure Code Sequence (0032,1064). It is recommended that all the images that are part of this particular instance of an ultrasound acquisition protocol have the same Performed Procedure Step ID. In an environment with information system connectivity, the Procedure Code Sequence can be used to identify all the images that were acquired as part of that protocol. If no information system connectivity is available, then Performed Procedure Step ID may be used to identify all the images that were acquired as part of that protocol. It is recommended that each stage should map to a separate series. Item: Add to PS 3.3 Table A for clarity, those modules that are changed from the existing standard or new are shown in Bold and Underlined A.7.4 US Multi-Frame Image IOD Module Table Table A US MULTI-FRAME IMAGE IOD MODULES IE Module Reference Usage Patient Patient C M Study General Study C M Patient Study C U Series General Series C M Frame of Reference US Series C.8.5.x U Suggested if image is part of an acquisition protocol Frame of Reference C U US Frame of Reference C C - Required if images are spatially related Equipment General Equipment C M Image General Image C M (See A.7.4.1) Image Pixel C M Contrast/bolus C C - Required if contrast media was used in this image. Cine C M Multi-frame C M Palette Color Lookup Table C.7.9 C - Required if Photometric Interpretation (0028,0004) has a value of PALETTE COLOR

13 Supplement 45: Ultrasound Protocol Page 7 US Region Calibration C U US Image C M US Staged Protocol C.8.5.y U Suggested if image is part of a staged protocol VOI LUT C.11.2 U SOP Common C.12.1 M Curve Curve Identification C.10.1 M (see A.7.4.1) Curve C.10.2 M Audio C.10.3 U SOP Common C.12.1 M A Mutually Exclusive IEs The Image and Curve IEs are mutually exclusive. Each SOP Instance using this IOD shall contain exactly one of these IEs. A.7.5 US Multi-Frame Image IOD Relationship to the DICOM Real-World Model Ultrasound images that make use of the US Series & Protocol Modules have a more specific relationship to the DICOM Real-World Model than non-protocol images. An ultrasound acquisition protocol is considered one Requested Procedure and has one Procedure Plan. The Procedure Plan will have one or more Action Items. An ultrasound acquisition protocol will typically consist of one Scheduled Procedure Step containing one or more Action Items. Assuming the acquisition protocol was successfully completed, this will result in one Modality Performed Procedure Step. Therefore, all images that are part of this particular instance of an ultrasound acquisition protocol will have the same Procedure Code Sequence (0032,1064). It is recommended that all the images that are part of this particular instance of an ultrasound acquisition protocol have the same Performed Procedure Step ID. In an environment with information system connectivity, the Procedure Code Sequence can be used to identify all the images that were acquired as part of that protocol. If no information system connectivity is available, then Performed Procedure Step ID may be used to identify all the images that were acquired as part of that protocol. It is recommended that each stage should map to a separate series. Item: Add to PS 3.3 New Sections C.8.5.x and C.8.5.y C.8.5.x US Series Module Table C.8-x contains IOD attributes that describe an ultrasound series.

14 8 Table C.8-x US SERIES MODULE ATTRIBUTES Attribute Name Tag Type Attribute Description Protocol Name (0018,1030) 1C User-defined description of the conditions under which the Series was performed. Required if Protocol Type Sequence (0018,xxxx) has no value. Protocol Type Sequence (0018,xxxx) 2 Sequence of one or more Items that identifies the protocol used to acquire this image. See section C.8.5.x.1.y >Include Code Sequence Macro Table Baseline Context ID is???. >Protocol Type Modifier Sequence (0018,xxxx) 3 Sequence of one or more Items that modifies the primary protocol type used to acquire this image. See section C.8.5.x.1.y >>Include Code Sequence Macro Table Baseline Context ID is???. Performed Procedure Step ID (0040,0253) 1C Identification of that Procedure that has been carried out within this step. Required if image was acquired in an acquisition protocol. C.8.5.x.1 US Series Attributes Descriptions C.8.5.x.1.y Protocol Type Sequence These Data Elements are intended to replace Protocol Name (0018,1030) Data Element. Protocol Type Sequence (0018,xxxx) describes the acquisition protocol used to acquire this image. The Coding Scheme Designator (0008,0102) shall be SNM3. The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID XXX. Note: The existence of this attribute means that this image is part of an acquisition protocol. Note: The following will not be included in the standard. Code Meaning (0008,0104) Transesophageal echocardiography Transthoracic echocardiography Epicardial echocardiography Intravascular echocardiography Intracardiac echocardiography Limited M-mode only echocardiography Limited Doppler only echocardiography Exercise stress echocardiography Maximal stress echocardiography Code Value (0008,0100) P5-B3002 P5-B3003 P5-B3004 P5-B3005 P5-B3006 P5-B301F P5-B303F P5-B3050 P5-B3051

15 Supplement 45: Ultrasound Protocol Page 9 Submaximal stress echocardiography Treadmill exercise stress echocardiography Bruce treadmill stress echocardiography Modified Bruce treadmill stress echocardiography Naughton treadmill stress echocardiography Bicycle exercise stress echocardiography Echocardiography with administered drug stress Dobutamine stress echocardiography High dose dobutamine stress echocardiography Low dose dobutamine stress echocardiography Arbutamine stress echocardiography Dipyridamole stress echocardiography Cardiac pacing echocardiography Adult echocardiography Pediatric echocardiography Intraoperative echocardiography Upright echocardiography Supine echocardiography Contrast echocardiography Contrast left ventricular opacification echocardiography Contrast perfusion echocardiography Contrast Doppler enhancement echocardiography 2D complete echocardiography Limited 2D only echocardiography Fetal echocardiography P5-B3052 P5-B3053 P5-B3054 P5-B3055 P5-B3056 P5-B3058 P5-B3060 P5-B3061 P5-B3062 P5-B3063 P5-B3065 P5-B3066 P5-B3070 P5-B3081 P5-B3082 P5-B3083 P5-B3084 P5-B3085 P5-B3090 P5-B3091 P5-B3092 P5-B3093 P5-B3191 P5-B3192 P5-B8215 Protocol Type Modifier Sequence (0018,xxxx) modifies the primary acquisition protocol stage when the image was acquired. The Coding Scheme Designator (0008,0102) shall be SNM3. The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID XXX. C.8.5.y US Staged Protocol Module Table C.8-y contains IOD attributes that describe an ultrasound staged protocol. Table C.8-y US STAGED PROTOCOL MODULE ATTRIBUTES Attribute Name Tag Type Attribute Description Number of Stages (0008,2124) 2 Number of Stages in this protocol. Number of Views in Stage (0008,212A) 2 Number of views in this Stage.

16 10 Stage Name (0008,2120) 1C A Stage is a particular time slice of a protocol in which a set of images are collected. The names can be free form text. Recommended text for Stress Echo stage names are: PRE-EXERCISE, POST-EXERCISE, PEAK-EXERCISE, RECOVERY, BASELINE, LOW DOSE, PEAK DOSE Required if Stage Type Sequence (0008,xxxx) has no value. Stage Type Sequence (0008,xxxx) 2 Sequence of one or more items that identifies the acquisition protocol stage when the image was acquired. See Section 8.5.y.1.1 >Include Code Sequence Macro Table Baseline Context ID is???. >Stage Type Modifier Sequence (0008,xxxx) 3 Sequence of one or more Items that modifies the primary stage type in this image. See Section 8.5.y.1.1 >>Include Code Sequence Macro Table Baseline Context ID is???. Stage Number (0008,2122) 1 A number that identifies the Stage. Stage Number starts at one. See Section C.8.5.y.1.4 View Number (0008,2128) 2 A number that identifies the View. Can be empty if View Number is unknown or if known, View Number starts at one. See Section C.8.5.y.1.5 Transducer Position Sequence (0008,2240) 2 Sequence of one or more Items that identifies the transducer position used in this image. See section C.8.5.y.1.2. >Include Code Sequence Macro Table Baseline Context ID is 4. > Transducer Position Modifier Sequence (0008,2242) 3 Sequence of one or more Items that modifies the primary transducer position of interest in this image. See Section C.8.5.y.1.2. >>Include Code Sequence Macro Table Baseline Context ID is 5. Transducer Orientation Sequence (0008,2244) 2 Sequence of one or more Items that identifies the Transducer Orientation used in this image. See section C.8.5.y.1.3.

17 Supplement 45: Ultrasound Protocol Page 11 >Include Code Sequence Macro Table Baseline Context ID is 6. > Transducer Orientation Modifier Sequence (0008,2246) 3 Sequence of one or more Items that modifies the primary Transducer Orientation of interest in this image. See Section C.8.5.y.1.3 >>Include Code Sequence Macro Table Baseline Context ID is 7. Note: Images that are part of a staged acquisition protocol should be identified by using Stage Type Sequence (0008,xxxx), Transducer Position Sequence (0008, 2240) and Transducer Orientation Sequence (0008,2244). Stage Number (0008,2122), View Number (0008,2128), Number of Stages (0008,2124), Number of Views in Stage (0008,212A) and Stage Name (0008, 2120) are maintained for backwards compatibility but may produce ambiguous results when used to identify a staged acquisition protocol. C.8.5.y.1 C.8.5.y.1.1 US Staged Protocol Attribute Descriptions Stage Type Sequence These Data Elements are intended to replace Stage Name (0008,2120) Data Element. Stage Type Sequence (0008,xxxx) describes the acquisition protocol stage when the image was acquired. The Coding Scheme Designator (0008,0102) shall be SNM3. The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID XXX Note: The existence of this attribute means that this image is part of an staged protocol. Note: The following will not be included in the standard. The following terms must be defined and a new Context ID created for them: These codes have been defined by SNOMED and are available. There is some debate that SNOMED requires a licensing fee, so these terms need to be defined in such a way that is agreeable to Working Group 6. Code Meaning (0008,0104) Code Value (0008,0100) Image acquisition procedure P Image acquisition after administration of contrast agent P Image acquisition during cardiac pacing P Image acquisition at user-defined cardiac pacing rate P Image acquisition during hand grip maneuver P Image acquisition during Valsalva P Image acquisition during postural maneuver P Pre-procedure image acquisition P Preoperative image acquisition P Intra-procedure image acquisition P Intra-operative image acquisition P Post-procedure image acquisition P Post-operative image acquisition P Image acquisition following first cardiopulmonary bypass P

18 12 Image acquisition following second cardiopulmonary P bypass Image acquisition following third cardiopulmonary bypass P Image acquisition during stress procedure P Image acquisition at baseline P Pre-stress image acquisition P Mid-stress image acquisition P Peak-stress image acquisition P Image acquisition during recovery P Image acquisition after drug administration P Image acquisition at user-defined dobutamine dose P Image acquisition at low-dose dobutamine P Image acquisition at mid-dose dobutamine P Image acquisition at peak dose dobutamine P Image acquisition at dobutamine 5 mcg/kg/min P Image acquisition at dobutamine 10 mcg/kg/min P Image acquisition at dobutamine 20 mcg/kg/min P Image acquisition at dobutamine 30 mcg/kg/min P Image acquisition at dobutamine 40 mcg/kg/min P Image acquisition at dobutamine 50 mcg/kg/min P Image at dobutamine 40 mcg/kg/min plus atropine P5-0131A Image acquisition at dobutamine 50 mcg/kg/min plus P5-0131B atropine Image acquisition at peak Arbutamine dose P Image acquisition at peak dipyridamole P Image acquisition after nitroglycerin P Image acquisition after amyl nitrite P Image acquisition after adenosine P Stage Type Modifier Sequence (0008,xxxx) modifies the primary acquisition protocol stage when the image was acquired. The Coding Scheme Designator (0008,0102) shall be SNM3. The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID XXX. C.8.5.y.1.2 Transducer Position Sequence Transducer Position Sequence (0008,2240) identifies the transducer position used in this image. The Coding Scheme Designator (0008,0102) shall be SNM3. The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID 4. Transducer Position Modifier Sequence (0008,2242) modifies the transducer position used in this image. The Coding Scheme Designator (0008,0102) shall be SNM3. The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID 5.

19 C.8.5.y.1.3 Transducer Orientation Sequence Supplement 45: Ultrasound Protocol Page 13 Transducer Orientation Sequence (0008,2244) identifies the transducer orientation used in this image. The Coding Scheme Designator (0008,0102) shall be SNM3. The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID 6. Transducer Orientation Modifier Sequence (0008,2246) modifies the transducer orientation used in this image. The Coding Scheme Designator (0008,0102) shall be SNM3. The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID 7. C.8.5.y.1.4 Stages The Number of Stages (0008,2124) shall be the total number of stages performed in this instance of the protocol. Stage Number (0008,2122) shall be the number that indicates the default order in which the stages are to be presented. Stage Number shall start from 1. C.8.5.y.1.5 Views The View Number (0008,2128) shall be used identify the same view across stages and defines a default order for presentation within and across stages. If a particular view appears in one stage but not in a subsequent stage, then that View Number will not be used during that subsequent stage. View Number shall start from 1. The Number of Views in Stage (0008,212A) shall be used to indicate how many unique View Numbers exist for this stage. The addition of modification of the value for this attribute shall not by itself force a SOP Instance UID change.

20 14 Changes to: NEMA Standards Publication PS Digital Imaging and Communications in Medicine (DICOM) Part 6: Data Dictionary

21 Supplement 45: Ultrasound Protocol Page REGISTRY OF DICOM DATA ELEMENTS Add the following Data Elements to PS 3.6, Section 6. Tag Name VR VM (0008,xxxx) Stage Type Sequence SQ 1 (0008,xxxx) Stage Type Modifier Sequence SQ 1 (0018,xxxx) Protocol Type Sequence SQ 1 (0018,xxxx) Protocol Type Modifier Sequence SQ 1

22 16 Changes to: NEMA Standards Publication PS Digital Imaging and Communications in Medicine (DICOM) Part 11: Media Storage Application Profiles

23 Supplement 45: Ultrasound Protocol Page 17 2 Item: Add to PS 3.11 Section C for clarity, that text which is removed from the existing standard is shown as Strikethrough and that text which is changed or new is shown in Underlined C Additional Keys File Set Creators and Updaters are only required to generate mandatory elements specified in PS Table C.3-4 specifies the additional associated keys that shall be supported. Refer to the Basic Directory IOD in PS 3.3. At each directory record level any additional data elements can be added as keys, but is not required by File Set Readers to be able to use them as keys. Table C.3-4 STD-US-XX-XX-XXXX ADDITIONAL DICOMDIR KEYS Key Attribute Tag Directory Record Type Protocol Type Sequence Type Notes (0018,xxxx) SERIES 2C Sequence of one or more Items that identifies the protocol used to acquire this image. Required if US Series module is included. >Include Code Sequence Macro Table Baseline Context ID is???. >Protocol Type Modifier Sequence See Part 3.3 Section C.8.5.x.1.y (0018,xxxx) SERIES 3 Sequence of one or more Items that modifies the primary protocol type used to acquire this image. >>Include Code Sequence Macro Table Baseline Context ID is???. Performed Procedure Step ID See Part 3.3 Section C.8.5.x.1.y (0040,0253) SERIES 1C Identification of that Procedure that has been carried out within this step. Required if series was part of a performed procedure step. Number of Stages (0008,2124) IMAGE 2C Number of Stages in this protocol. Required if Stage Type Sequence (0008,xxxx) is present. Number of Views in Stage See Part 3.3 Section C.8.5.y.1.4 (0008,212A) IMAGE 2C Number of unique views in this Stage. Required if Stage Type Sequence (0008,xxxx) is present. See Part 3.3 Section C.8.5.y.1.5

24 18 Stage Type Sequence (0008,xxxx) IMAGE 2C Sequence of one or more items that identifies the acquisition protocol stage when the image was acquired. Required US Protocol module is present. >Include Code Sequence Macro Table Baseline Context ID is???. >Stage Type Modifier Sequence See Part 3.3 Section 8.5.y.1.1 (0008,xxxx) IMAGE 3 Sequence of one or more Items that modifies the primary stage type in this image. >>Include Code Sequence Macro Table Baseline Context ID is???. See Part 3.3 Section 8.5.y.1.1 Stage Number (0008,2122) IMAGE 1C A number that identifies the Stage. Stage Number starts at one. Required if Stage Type Sequence (0008,xxxx) is present. See Part 3.3 Section C.8.5.y.1.4 View Number (0008,2128) IMAGE 1C A number that identifies the View. View Number starts at one. Required if Stage Type Sequence (0008,xxxx) is present. Transducer Position Sequence See Part 3.3 Section C.8.5.y.1.5 (0008,2240) IMAGE 2C Sequence of one or more Items that identifies the transducer position used in this image. Required if Stage Type Sequence (0008,xxxx) is present. >Include Code Sequence Macro Table Baseline Context ID is 4. > Transducer Position Modifier Sequence See Part 3.3 Section C.8.5.y.1.2. (0008,2242) IMAGE 3 Sequence of one or more Items that modifies the primary transducer position of interest in this image. >>Include Code Sequence Macro Table Baseline Context ID is 5. Transducer Orientation Sequence See Part 3.3 Section C.8.5.y.1.2. (0008,2244) IMAGE 2C Sequence of one or more Items that identifies the Transducer Orientation used in this image. Required if Stage Type Sequence (0008,xxxx) is present. >Include Code Sequence Macro Table Baseline Context ID is 6. See Part 3.3 Section C.8.5.y.1.3.

25 > Transducer Orientation Modifier Sequence Supplement 45: Ultrasound Protocol Page 19 (0008,2246) IMAGE 3 Sequence of one or more Items that modifies the primary Transducer Orientation of interest in this image. >>Include Code Sequence Macro Table Baseline Context ID is 7. See Part 3.3 Section C.8.5.y.1.3 2

New Information Object Definitions for DICONDE

New Information Object Definitions for DICONDE 4th European-American Workshop on Reliability of NDE - Th.5.B.4 New Information Object Definitions for DICONDE M. JOBST, GE Sensing & Inspection Technologies, Hürth, Germany; J. HANSEN, GE Sensing & Inspection

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 50: Mammography Computer-Aided Detection SR SOP Class

Digital Imaging and Communications in Medicine (DICOM) Supplement 50: Mammography Computer-Aided Detection SR SOP Class Digital Imaging and Communications in Medicine (DICOM) Supplement 50: Mammography Computer-Aided Detection SR SOP Class Status: Letter Ballot Text February 2, 2001 DICOM Standards Committee 1300 N. 17

More information

Health informatics Digital imaging and communication in medicine (DICOM) including workflow and data management

Health informatics Digital imaging and communication in medicine (DICOM) including workflow and data management INTERNATIONAL STANDARD ISO 12052 Second edition 2017-08 Health informatics Digital imaging and communication in medicine (DICOM) including workflow and data management Informatique de santé Imagerie numérique

More information

DICOM Conformance Statement BK Medical Ultrasound System, 1202, 2202 and 2300

DICOM Conformance Statement BK Medical Ultrasound System, 1202, 2202 and 2300 DICOM Conformance Statement BK Medical Ultrasound System, 1202, 2202 and 2300 Copyright BK Medical, Denmark. 2014. All rights reserved. Unauthorized use prohibited. Network Services Networking SOP Classes

More information

RS85 Ultrasound System DICOM Conformance Statement

RS85 Ultrasound System DICOM Conformance Statement RS85 Ultrasound System DICOM Conformance Statement Revision 1.0 System Version 1.0 Created: October 20, 2017 Last Updated: October 20, 2017 RS85 DICOM Conformance SAMSUNG MEDISON 0 COVER PAGE Company Name:

More information

DICOM Correction Proposal

DICOM Correction Proposal DICOM Correction Proposal STATUS Final Text Date of Last Update 2015/03/18 Person Assigned Andrew Casertano, DICOM WG 22 Submitter Name Harry Solomon Submission Date 2003/03/19 Correction Number CP-375

More information

UGEO HM70A Ultrasound System DICOM Conformance Statement

UGEO HM70A Ultrasound System DICOM Conformance Statement UGEO HM70A Ultrasound System DICOM Conformance Statement Revision 1.0 System Version 1.0 Created: Dec 06, 2013 Last Updated: Dec 06, 2013 UGEO HM70A DICOM Conformance SAMSUNG MEDISON 0 COVER PAGE Company

More information

20 CP Transverse positioning of pre-clinical research small animal subjects

20 CP Transverse positioning of pre-clinical research small animal subjects 20 CP-1473 - Transverse positioning of pre-clinical research small animal subjects Page 1 1 Status Final Text 2 Date of Last Update 2015/11/10 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com

More information

DICOM. Conformance Statement EPIQ 7 Release 1.1.x.x EPIQ 5 Release 1.0.x.x Rev C

DICOM. Conformance Statement EPIQ 7 Release 1.1.x.x EPIQ 5 Release 1.0.x.x Rev C Koninklijke Philips Electronics N.V. 2013 All rights are reserved. DICOM Conformance Statement EPIQ 7 Release 1.1.x.x EPIQ 5 Release 1.0.x.x 000217000000068 Rev C 2013-10-14 0.1 Revision History Document

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement Document Version 2.0 Revision Summary Date Revision Changes 30 th December 2016 2.0 Update to OrthoView 7.0 Specification 20 October 2005 1.0 Release of OrthoView 3.1 WK3 build

More information

19 CP Add teaching file export related codes

19 CP Add teaching file export related codes 19 CP-1680 - Add teaching file export related codes Page 1 1 Status Final Text 2 Date of Last Update 2017/09/14 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5 Submitter Name David Clunie

More information

DICOM Conformance Statement (Rev. 2) E-CUBE 15 v.3.2.x - v.3.x.x

DICOM Conformance Statement (Rev. 2) E-CUBE 15 v.3.2.x - v.3.x.x DICOM Conformance Statement (Rev. 2) E-CUBE 15 v.3.2.x - v.3.x.x Revision History Revision Date of Issue Author Description 0 2014-04-17 Hyun-deok Choi Initial release 1 2014-10-07 Hyun-deok Choi Update

More information

19 CP Add teaching file export related codes

19 CP Add teaching file export related codes 19 CP-1680 - Add teaching file export related codes Page 1 1 Status Letter Ballot 2 Date of Last Update 2017/06/21 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5 Submitter Name David Clunie

More information

RS80A Ultrasound System DICOM Conformance Statement

RS80A Ultrasound System DICOM Conformance Statement RS80A Ultrasound System DICOM Conformance Statement Revision 1.1 System Version 1.00.03 Created: June 02, 2014 Last Updated: August 22, 2014 RS80A DICOM Conformance SAMSUNG MEDISON 0 COVER PAGE Company

More information

DICOM Conformance Statement for. Hitachi ARIETTA Precision

DICOM Conformance Statement for. Hitachi ARIETTA Precision DICOM Conformance Statement for Hitachi ARIETTA Precision Company Name : Hitachi, Ltd. Product Name : Diagnostic Ultrasound System Hitachi ARIETTA Precision Version : Version 1. 1 Date : 2016. 04. 01 Internal

More information

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL TUS-AI900 V1.5 MODEL TUS-AI800 V1.5 MODEL TUS-AI700 V1.5

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL TUS-AI900 V1.5 MODEL TUS-AI800 V1.5 MODEL TUS-AI700 V1.5 DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL TUS-AI900 V1.5 MODEL TUS-AI800 V1.5 MODEL TUS-AI700 V1.5 TOSHIBA MEDICAL SYSTEMS CORPORATION 2016 ALL RIGHTS RESERVED Trademarks Aplio

More information

DICOM. Conformance Statement CX x Rev B Koninklijke Philips Electronics N.V All rights are reserved.

DICOM. Conformance Statement CX x Rev B Koninklijke Philips Electronics N.V All rights are reserved. DICOM Conformance Statement CX30 2.0.x 000292000000004 Rev B 2012-11-21 Koninklijke Philips Electronics N.V. 2012 All rights are reserved. 0.1 REVISION HISTORY Document Version A August 27, 2012 B November

More information

DICOM. Conformance Statement. CX x Rev B. Koninklijke Philips Electronics N.V All rights are reserved.

DICOM. Conformance Statement. CX x Rev B. Koninklijke Philips Electronics N.V All rights are reserved. DICOM Conformance Statement CX50 3.0.x 000250000000016 2012-10-15 Rev B Koninklijke Philips Electronics N.V. 2012 All rights are reserved. 0.1 REVISION HISTORY Document Version A August 27, 2012 B October

More information

DICOM. Conformance Statement. ie33 Software Version x

DICOM. Conformance Statement. ie33 Software Version x DICOM Conformance Statement ie33 Software Version 1.0.0. 0.1 REVISION HISTORY Document Version Date of Issue Author Description A October 29, 2004 M. Leif Initial Release B November 17, 2004 M. Leif Changed

More information

Provläsningsexemplar / Preview TECHNICAL REPORT. Medical electrical equipment Guidelines for implementation of DICOM in radiotherapy

Provläsningsexemplar / Preview TECHNICAL REPORT. Medical electrical equipment Guidelines for implementation of DICOM in radiotherapy TECHNICAL REPORT IEC TR 62266 First edition 2002-03 Medical electrical equipment Guidelines for implementation of DICOM in radiotherapy Reference number IEC/TR 62266:2002(E) Publication numbering As from

More information

DICOM. Conformance Statement. iu22 Release 5.0.X.X ie33 Release 5.0.X.X Rev F,

DICOM. Conformance Statement. iu22 Release 5.0.X.X ie33 Release 5.0.X.X Rev F, DICOM Conformance Statement iu22 Release 5.0.X.X ie33 Release 5.0.X.X 453561239620009 Rev F, 2008-08-01 Koninklijke Philips Electronics N.V. 2008 All rights are reserved. 0.1 REVISION HISTORY Document

More information

Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version 0.1 7/27/2011

Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version 0.1 7/27/2011 Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version 0.1 7/27/2011 I. Purpose Drawing from the profile development of the QIBA-fMRI Technical Committee,

More information

Click here for Link to References: CMS Website HOPPS CY 2018 Final Rule. CMS Website HOPPS CY2018 Final Rule Updated November 2017.

Click here for Link to References: CMS Website HOPPS CY 2018 Final Rule. CMS Website HOPPS CY2018 Final Rule Updated November 2017. Final Compared to 3Q 2017 Rates Medicare Hospital Outpatient Prospective Payment System HOPPS () Nuclear Cardiology Procedures, Radiopharmaceuticals, and Drugs Click here for Link to References: CMS Website

More information

DICOM Conformance Statement for. Hitachi F31

DICOM Conformance Statement for. Hitachi F31 DICOM Conformance Statement for Hitachi F31 Company Name : Hitachi, Ltd. Product Name : Diagnostic Ultrasound System Hitachi F31 Version : Version 1. 1 Date : 2016. 04. 01 Internal document number : SFT

More information

Cardiovascular nuclear imaging employs non-invasive techniques to assess alterations in coronary artery flow, and ventricular function.

Cardiovascular nuclear imaging employs non-invasive techniques to assess alterations in coronary artery flow, and ventricular function. National Imaging Associates, Inc. Clinical guidelines CARDIOVASCULAR NUCLEAR MEDICINE -MYOCARDIAL PERFUSION IMAGING -MUGA CPT4 Codes: Refer to pages 6-9 LCD ID Number: L33960 J 15 = KY, OH Responsible

More information

Transducers. ACUSON SC2000 Ultrasound System. Precision at the Speed of Life Release 5.0

Transducers. ACUSON SC2000 Ultrasound System. Precision at the Speed of Life Release 5.0 Transducers ACUSON SC2000 Ultrasound System Precision at the Speed of Life Release 5.0 Table of Contents Z6Ms Transducer 3 4Z1c Transducer 3 4V1c Transducer 4 8V3 Transducer 4 10V4 Transducer 5 V5Ms Transducer

More information

MyLabSix MyLabSeven MyLabAlpha MyLabGamma Ultrasound Scanners. DICOM Conformance Statement. Document Version 7.2

MyLabSix MyLabSeven MyLabAlpha MyLabGamma Ultrasound Scanners. DICOM Conformance Statement. Document Version 7.2 MyLabSix MyLabSeven MyLabAlpha MyLabGamma Ultrasound Scanners DICOM Conformance Statement Document Version 7.2 Date: APR 8, 2015 1 CONFORMANCE STATEMENT OVERVIEW MyLab is a family of Ultrasound scanners

More information

TW Hamilton, EP Ficaro, TA Mitchell, JN Kritzman, JR Corbett University of Michigan Health System, Ann Arbor, MI

TW Hamilton, EP Ficaro, TA Mitchell, JN Kritzman, JR Corbett University of Michigan Health System, Ann Arbor, MI Accuracy and Variability of of 3D-MSPECT for Estimating the Left Ventricular Ejection Fraction as as a Function of of Gating Frames and Reconstruction Filters TW Hamilton, EP Ficaro, TA Mitchell, JN Kritzman,

More information

MyLab Ultrasound Scanners. DICOM Conformance Statement. Document Version 6.3

MyLab Ultrasound Scanners. DICOM Conformance Statement. Document Version 6.3 MyLab Ultrasound Scanners DICOM Conformance Statement Document Version 6.3 Date: MAY 21, 2010 1 CONFORMANCE STATEMENT OVERVIEW MyLab is a family of Ultrasound scanners made by Esaote; their software is

More information

Cardiovascular nuclear imaging employs non-invasive techniques to assess alterations in coronary artery flow, and ventricular function.

Cardiovascular nuclear imaging employs non-invasive techniques to assess alterations in coronary artery flow, and ventricular function. National Imaging Associates, Inc. Clinical guidelines CARDIOVASCULAR NUCLEAR MEDICINE -MYOCARDIAL PERFUSION IMAGING -MUGA Original Date: October 2015 Page 1 of 9 FOR CMS (MEDICARE) MEMBERS ONLY CPT4 Codes:

More information

Transducers. ACUSON SC2000 Ultrasound System. Release 4.0. Answers for life.

Transducers. ACUSON SC2000 Ultrasound System. Release 4.0. Answers for life. Transducers ACUSON SC2000 Ultrasound System Release 4.0 Answers for life. ACUSON SC2000 Ultrasound System TABLE OF CONTENTS Z6Ms Transducer 3 4Z1c Transducer 3 4V1c Transducer 4 8V3 Transducer 4 10V4 Transducer

More information

DICOM. Conformance Statement. HD x Rev C Koninklijke Philips Electronics N.V All rights are reserved.

DICOM. Conformance Statement. HD x Rev C Koninklijke Philips Electronics N.V All rights are reserved. DICOM Conformance Statement HD15 1.5. 000258000000002 Rev C 2011-06-29 Koninklijke Philips Electronics N.V. 2011 All rights are reserved. 0.1 REVISION HISTORY Document Version Date of Issue Author Description

More information

3D-stress echocardiography Bernard Cosyns, MD, PhD

3D-stress echocardiography Bernard Cosyns, MD, PhD 3D-stress echocardiography Bernard Cosyns, MD, PhD No Disclosure The Pro-Technology bias Sicari et al. Cardiovascular Ultrasound 2006, 4:11 Overview 2D stress echocardiography: main limitations 3D echocardiography:

More information

INTRODUCTION TO DICOM FOR THE PRACTICING VETERINARIAN

INTRODUCTION TO DICOM FOR THE PRACTICING VETERINARIAN INTRODUCTION TO DICOM FOR THE PRACTICING VETERINARIAN MATTHEW A. WRIGHT, DENNIS BALLANCE, IAN D. ROBERTSON, BRIAN POTEET Digital Imaging and Communication in Medicine (DICOM) is a communication protocol

More information

Alignment and Calibration of the APEX Instrument. March 30, Chemical Crystallography Laboratory

Alignment and Calibration of the APEX Instrument. March 30, Chemical Crystallography Laboratory Alignment and Calibration of the APEX Instrument Page 1 of 8 Alignment and Calibration of the APEX Instrument March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Alignment and

More information

DICOM Correction Item

DICOM Correction Item Correction Number CP-759 DICOM Correction Item Log Summary: Type of Modification Modification Rationale for Correction Name of Standard PS 3.16 The templates for Cardiovascular Patient History, originally

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement HI VISION Ascendus Special Notes to Operators and Maintenance Managers Before using this system, be sure to thoroughly read this DICOM Conformance Statement and make yourself

More information

Pediatric Echo Structured Reporting An Experience Sharing

Pediatric Echo Structured Reporting An Experience Sharing THE DICOM 2013 INTERNATIONAL CONFERENCE & SEMINAR March 14-16 Bangalore, India Bindhu M Pediatric Echo Structured Reporting An Experience Sharing Philips Healthcare Technical Specialist Bangalore, India

More information

IHE Cardiology Technical Framework Supplement. Cardiac Imaging Report Content (CIRC) Trial Implementation

IHE Cardiology Technical Framework Supplement. Cardiac Imaging Report Content (CIRC) Trial Implementation Integrating the Healthcare Enterprise 5 IHE Cardiology Technical Framework Supplement 10 Cardiac Imaging Report Content (CIRC) 15 Trial Implementation Date: July 1, 2011 20 Editor: IHE Cardiology Technical

More information

TECHNICAL REPORT. Medical electrical equipment Guidelines for implementation of DICOM in radiotherapy

TECHNICAL REPORT. Medical electrical equipment Guidelines for implementation of DICOM in radiotherapy TECHNICAL REPORT IEC TR 62266 First edition 2002-03 Medical electrical equipment Guidelines for implementation of DICOM in radiotherapy IEC 2002 Copyright - all rights reserved No part of this publication

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-449 Log Summary: Type of Modification Correct Value Name of Standard PS 3.16-2004 Rationale for Correction: Clinical applications use arterial and venous branches

More information

General Cardiovascular Magnetic Resonance Imaging

General Cardiovascular Magnetic Resonance Imaging 2 General Cardiovascular Magnetic Resonance Imaging 19 Peter G. Danias, Cardiovascular MRI: 150 Multiple-Choice Questions and Answers Humana Press 2008 20 Cardiovascular MRI: 150 Multiple-Choice Questions

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 17510 First edition 2015-08-01 Medical devices Sleep apnoea breathing therapy Masks and application accessories Dispositifs médicaux Thérapie respiratoire de l apnée du sommeil

More information

Hitachi Aloka Medical manufactured one of the world s first diagnostic ultrasound platforms, and today this imaging modality has become the first

Hitachi Aloka Medical manufactured one of the world s first diagnostic ultrasound platforms, and today this imaging modality has become the first Hitachi Aloka Medical manufactured one of the world s first diagnostic ultrasound platforms, and today this imaging modality has become the first choice examination for many disorders. If the subtlest

More information

Anthem Blue Cross and Blue Shield Virginia Advanced Imaging Procedures Requiring Precertification Revised 02/13/2013

Anthem Blue Cross and Blue Shield Virginia Advanced Imaging Procedures Requiring Precertification Revised 02/13/2013 Anthem Blue Cross and Blue Shield Virginia Advanced Imaging Procedures Requiring Precertification Revised 02/13/2013 Modality and CT Head CTA Head: Cerebrovascular MRI Head MRA Head: Cerebrovascular Functional

More information

Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version 0.2 8/10/2011 I. Purpose

Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version 0.2 8/10/2011 I. Purpose Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version 0.2 8/10/2011 I. Purpose Drawing from the profile development of the QIBA-fMRI Technical Committee,

More information

DICOM CONFORMANCE STATEMENT FOR DC-8/DC-8 PRO/DC-8 CV/ DC-8 EXP/DC-8S DIAGNOSTIC ULTRASOUND SYSTEM

DICOM CONFORMANCE STATEMENT FOR DC-8/DC-8 PRO/DC-8 CV/ DC-8 EXP/DC-8S DIAGNOSTIC ULTRASOUND SYSTEM DICOM CONFORMANCE STATEMENT FOR DC-8/DC-8 PRO/DC-8 CV/ DC-8 EXP/DC-8S DIAGNOSTIC ULTRASOUND SYSTEM Version V4.0 2014-09-02 Mindray 2011-2014 All rights reserved SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement HI VISION Ascendus Special Notes to Operators and Maintenance Managers Before using this system, be sure to thoroughly read this DICOM Conformance Statement and make yourself

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 26782 First edition 2009-07-15 Anaesthetic and respiratory equipment Spirometers intended for the measurement of time forced expired volumes in humans Matériel d'anesthésie et

More information

Tc-99m Sestamibi/Tetrofosmin Stress-Rest Myocardial Perfusion Scintigraphy

Tc-99m Sestamibi/Tetrofosmin Stress-Rest Myocardial Perfusion Scintigraphy APPROVED BY: Director of Radiology Page 1 of 6 Tc-99m Sestamibi/Tetrofosmin Stress-Rest Myocardial Primary Indications: Evaluation of myocardial perfusion and viability in patients with known or suspected

More information

CPT Code Details

CPT Code Details CPT Code 93925 Details Code Descriptor Duplex scan of lower extremity arteries or arterial bypass grafts; complete bilateral study Lay Term The provider performs a duplex ultrasound scan of the lower extremity

More information

Automatic quality control processing for detection of element/channel dropout in ultrasound transducers

Automatic quality control processing for detection of element/channel dropout in ultrasound transducers Automatic quality control processing for detection of element/channel dropout in ultrasound transducers Paul L. Carson, Ph.D. University of Michigan, Department of Radiology, Ann Arbor, MI, USA Acknowledgements

More information

Stress echo workshop STRESSORS

Stress echo workshop STRESSORS Stress echo workshop STRESSORS Adham Ahmed, MD Lecturer of Cardiology, Ain Shams Indications of Stress Echo CAD Diagnosis Prognosticat ion 1 Physiologic Basis 1930s: Tennant and Wiggers Relationship between

More information

IEC 87: Ultrasonics Overview update July 2002

IEC 87: Ultrasonics Overview update July 2002 - 1 - CCAUV/02-07 IEC 87: Ultrasonics Overview update July 2002 Roy C Preston Secretary IEC TC 87: Ultrasonics British Electrotechnical Committee, British Standards Institution, 389 Chiswick High Road,

More information

evicore cardiology procedures and services requiring prior authorization

evicore cardiology procedures and services requiring prior authorization evicore cardiology procedures and services requiring prior authorization Moda Health Commercial Group and Individual Members* *Check EBT to verify member enrollment in evicore program Radiology Advanced

More information

Prior Authorization for Non-emergency Cardiac Imaging Procedures

Prior Authorization for Non-emergency Cardiac Imaging Procedures Attention: All Providers Prior Authorization for Non-emergency Cardiac Imaging Procedures The N.C. Medicaid Program is considering implementation of a prior authorization (PA) program for non-emergency

More information

ACUSON Sequoia C512 Echocardiography System. Transducers.

ACUSON Sequoia C512 Echocardiography System. Transducers. ACUSON Sequoia C512 Echocardiography System Transducers www.siemens.com/medical 1 10V4c Transducer 10 4 MHz ACUSON Sequoia 512 ultrasound system Neonatal and pediatric echocardiography MICROSON high resolution

More information

ISO INTERNATIONAL STANDARD. Sterile hypodermic syringes for single use Part 4: Syringes with re-use prevention feature

ISO INTERNATIONAL STANDARD. Sterile hypodermic syringes for single use Part 4: Syringes with re-use prevention feature INTERNATIONAL STANDARD ISO 7886-4 First edition 2006-10-01 Sterile hypodermic syringes for single use Part 4: Syringes with re-use prevention feature Seringues hypodermiques stériles, non réutilisables

More information

Automated Volumetric Cardiac Ultrasound Analysis

Automated Volumetric Cardiac Ultrasound Analysis Whitepaper Automated Volumetric Cardiac Ultrasound Analysis ACUSON SC2000 Volume Imaging Ultrasound System Bogdan Georgescu, Ph.D. Siemens Corporate Research Princeton, New Jersey USA Answers for life.

More information

Digital mammography imaging from Carestream Health solutions for great workflow, productivity, and patient care.

Digital mammography imaging from Carestream Health solutions for great workflow, productivity, and patient care. Digital Mammography Imaging on KODAK CR Systems Digital mammography imaging from Carestream Health solutions for great workflow, productivity, and patient care. Commercial distribution of the CR Mammography

More information

Underwater Acoustic Measurements in Megahertz Frequency Range.

Underwater Acoustic Measurements in Megahertz Frequency Range. Underwater Acoustic Measurements in Megahertz Frequency Range. Current State and Prospects of Development in Russia Alexander M. Enyakov,, Many medical applications of underwater acoustic measurements

More information

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 122 Annex C Edition 1999 (E) Equipment for speech audiometry Annex C Test report format Appareils pour l audiométrie vocale Annexe C Format du rapport d essai OIML R

More information

HEARTWORKS ADVANCING EDUCATION IN ULTRASOUND. Simulation in Cardiac Anatomy, Transesophageal & Transthoracic Echocardiography and Lung Ultrasound

HEARTWORKS ADVANCING EDUCATION IN ULTRASOUND. Simulation in Cardiac Anatomy, Transesophageal & Transthoracic Echocardiography and Lung Ultrasound ADVANCING EDUCATION IN ULTRASOUND Simulation in Cardiac Anatomy, Transesophageal & Transthoracic Echocardiography and Lung Ultrasound Learn the theory HeartWorks is the creation of leading cardiac anesthesiologists

More information

ACUSON Antares Ultrasound System. Transducers.

ACUSON Antares Ultrasound System. Transducers. ACUSON Antares Ultrasound System Transducers www.siemens.com/healthcare 1 CH6-2 Transducer 6 2 MHz Advanced SieClear spatial compounding, Cadence contrast pulse sequencing technology, Axius direct ultrasound

More information

Policy #: 222 Latest Review Date: March 2009

Policy #: 222 Latest Review Date: March 2009 Name of Policy: MRI Phase-Contrast Flow Measurement Policy #: 222 Latest Review Date: March 2009 Category: Radiology Policy Grade: Active Policy but no longer scheduled for regular literature reviews and

More information

Cigna - Prior Authorization Procedure List Cardiology

Cigna - Prior Authorization Procedure List Cardiology Cigna - Prior Authorization Procedure List Cardiology Category CPT Code CPT Code Description 33206 Insertion of new or replacement of permanent pacemaker with transvenous electrode(s); atrial 33207 Insertion

More information

GE Healthcare. Logiq P5 Premium

GE Healthcare. Logiq P5 Premium GE Healthcare Logiq P5 Premium LOGIQ P5 PREMIUM The GE Logiq P5 premium is an economical shared service ultrasound machine that is unique in it s price range by offering deep support for all applications

More information

ISO INTERNATIONAL STANDARD. Non-active surgical implants Joint replacement implants Particular requirements

ISO INTERNATIONAL STANDARD. Non-active surgical implants Joint replacement implants Particular requirements INTERNATIONAL STANDARD ISO 21534 Second edition 2007-10-01 Non-active surgical implants Joint replacement implants Particular requirements Implants chirurgicaux non actifs Implants de remplacement d'articulation

More information

Joint ADA and Health Level 7 Collaboration On Electronic Dental Attachments

Joint ADA and Health Level 7 Collaboration On Electronic Dental Attachments Joint ADA and Health Level 7 Collaboration On Electronic Dental Attachments Jean Narcisi, Director, Department of Dental Informatics Patrick Cannady, Manager, Department of Dental Informatics ADA Practice

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 169: Simplified Adult Echocardiography Report

Digital Imaging and Communications in Medicine (DICOM) Supplement 169: Simplified Adult Echocardiography Report 5 Digital Imaging and Communications in Medicine (DICOM) Supplement 169: Simplified Adult Echocardiography Report 10 15 20 DICOM Standards Committee 1300 N. 17th Street, Suite 900 Rosslyn, Virginia 22209

More information

Echocardiography A powerful module of MediConnect

Echocardiography A powerful module of MediConnect Echocardiography A powerful module of MediConnect WHERE IT ALL COMES TOGETHER MEDICONNECT AN OVERVIEW Integrating the latest IT technologies with medical expertise within an existing clinical infrastructure

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:

UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Electrophysiology Implant Code Classification Table The

More information

AudioConsole. User Guide. Doc. No EN/01 Part No EN

AudioConsole. User Guide. Doc. No EN/01 Part No EN AudioConsole Doc. No. 7-50-2180-EN/01 Part No. 7-50-21800-EN Copyright notice [2003], 2018 Inmedico A/S. All rights reserved. Oscilla is aregistered trademark of Inmedico A/S in the U.S.A. and/or other

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 15197 First edition 2003-05-01 In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus Systèmes d'essais

More information

Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version /13/2011 I. Purpose

Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version /13/2011 I. Purpose Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version 0.6 10/13/2011 I. Purpose Drawing from the profile development of the QIBA-fMRI Technical Committee,

More information

Final October 24, 2001

Final October 24, 2001 American Society of Echocardiography and Society of Cardiovascular Anesthesiologists Task Force Guidelines for Training in Perioperative Echocardiography Introduction and Overview When expertly utilized,

More information

STANDARDIZED PROCEDURE CARDIAC STRESS TESTING-DOBUTAMINE INFUSION (Adult)

STANDARDIZED PROCEDURE CARDIAC STRESS TESTING-DOBUTAMINE INFUSION (Adult) I. Definition: This test is performed to evaluate for cardiac ischemia, arrhythmias, and/or response to exercise. II. Background Information A. Setting: The setting (inpatient vs outpatient) and population

More information

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging Provläsningsexemplar / Preview TECHNICAL REPORT ISO/TR 16379 First edition 2014-03-01 Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR

More information

ISO/TR TECHNICAL REPORT. Non-destructive testing Guidelines for NDT personnel training organizations

ISO/TR TECHNICAL REPORT. Non-destructive testing Guidelines for NDT personnel training organizations TECHNICAL REPORT ISO/TR 25108 First edition 2006-07-01 Non-destructive testing Guidelines for NDT personnel training organizations Essais non destructifs Lignes directrices pour les organismes de formation

More information

Course Learning Objectives Sunday, February 17 Friday, February 22

Course Learning Objectives Sunday, February 17 Friday, February 22 Course Learning Objectives Sunday, February 17 Friday, February 22 1. Define the physical principles of ultrasound technology and its applications for two and threedimensional (2D, 3D) imaging and use

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-37 2001 AMENDMENT 1 2004-08 Amendment 1 Medical electrical equipment Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring

More information

GE Healthcare. Transducer Guide. VIVID Series

GE Healthcare. Transducer Guide. VIVID Series GE Healthcare Transducer Guide VIVID Series VIVID E9 Series Introducing Vivid TM E9 D-Series transducers. You now have the power to acquire extraordinary images on every patient, with transducer technology

More information

Avaya B179 Conference Telephone Voluntary Product Accessibility Template (VPAT)

Avaya B179 Conference Telephone Voluntary Product Accessibility Template (VPAT) Avaya B179 Conference Telephone Voluntary Product Accessibility Template (VPAT) The Avaya B179 Conference Telephone is a sophisticated speakerphone, intended for use by groups of ten or more individuals

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Equipment for speech audiometry Appareils pour l audiométrie vocale OIML R 122 Edition 1996 (E) CONTENTS Foreword... 3

More information

2019 Qualified Clinical Data Registry (QCDR) Performance Measures

2019 Qualified Clinical Data Registry (QCDR) Performance Measures 2019 Qualified Clinical Data Registry (QCDR) Performance Measures Description: This document contains the 18 performance measures approved by CMS for inclusion in the 2019 Qualified Clinical Data Registry

More information

CEA Technical Report. Closed Captioning in IP-delivered Video Programming CEA-TR-3

CEA Technical Report. Closed Captioning in IP-delivered Video Programming CEA-TR-3 CEA Technical Report Closed Captioning in IP-delivered Video Programming CEA-TR-3 February 2013 NOTICE Consumer Electronics Association (CEA ) Standards, Bulletins and other technical publications are

More information

TECHNICAL SPECIFICATION

TECHNICAL SPECIFICATION TECHNICAL SPECIFICATION IEC TS 62462 First edition 2007-05 Ultrasonics Output test Guide for the maintenance of ultrasound physiotherapy systems Commission Electrotechnique Internationale International

More information

DreamMapper 2.4. Mobile App User Guide

DreamMapper 2.4. Mobile App User Guide DreamMapper 2.4 Mobile App User Guide June 8, 2016 Table of Contents Sleep Screen... 1 How to navigate... 2 Changing Data Categories... 5 See Additional Details for the Day and Category... 6 Other Menu

More information

ACUSON S2000 Ultrasound System. Transducers Release 1.0.

ACUSON S2000 Ultrasound System. Transducers Release 1.0. ACUSON S2000 Ultrasound System Transducers Release 1.0 www.siemens.com/healthcare 1 7CF2 Transducer 7 2.5 MHz Reduced housing size at the grip zones for improved ergonomics Advanced materials for high

More information

ACUSON P500. Ultrasound Anytime, Anywhere. ACUSON P500. siemens.com/acusonp500. siemens.com/acusonp500 1

ACUSON P500. Ultrasound Anytime, Anywhere. ACUSON P500. siemens.com/acusonp500. siemens.com/acusonp500 1 Ultrasound Anytime, Anywhere. ACUSON P500 siemens.com/acusonp500 siemens.com/acusonp500 1 Enabling ultrasound imaging anytime, anywhere. 2 siemens.com/acusonp500 Ultrasound Anytime, Anywhere Siemens Healthineers

More information

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 12749-4 First edition 2015-08-15 Nuclear energy, nuclear technologies, and radiological protection Vocabulary Part 4: Dosimetry for radiation processing

More information

Progress in CT. Jim Green Senior Vice President & General Manager Philips CT

Progress in CT. Jim Green Senior Vice President & General Manager Philips CT Progress in CT Jim Green Senior Vice President & General Manager Philips CT 2 3 Agenda What is CT New clinical applications CT market Why we win 4 What is CT? Rotate an X-ray source and detection system

More information

Radiation Dose Structured Report Current and Future. Donald Peck, PhD Henry Ford Health System Detroit, Michigan USA

Radiation Dose Structured Report Current and Future. Donald Peck, PhD Henry Ford Health System Detroit, Michigan USA Radiation Dose Structured Report Current and Future Donad Peck, PhD Henry Ford Heath System Detroit, Michigan USA 1 Radiation Dose SR Origina deveoped in 2005 for Projection Radiography Added 3 new tempate

More information

Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version /310/2011 I. Purpose

Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version /310/2011 I. Purpose Incorporation of Imaging-Based Functional Assessment Procedures into the DICOM Standard Draft version 0.54 10/310/2011 I. Purpose Drawing from the profile development of the QIBA-fMRI Technical Committee,

More information

RADIOLOGIC TECHNOLOGY (526)

RADIOLOGIC TECHNOLOGY (526) RADIOLOGIC TECHNOLOGY (526) 526-133 DMS General Procedures 2 Radiologic Technology (526) 1 526-130 Introduction to Diagnostic Medical Sonography This course introduces the student to the history of ultrasound

More information

GEX Recommended Procedure Eff. Date: 09/21/10 Rev.: D Pg. 1 of 7

GEX Recommended Procedure Eff. Date: 09/21/10 Rev.: D Pg. 1 of 7 GEX Recommended Procedure Eff. Date: 09/21/10 Rev.: D Pg. 1 of 7 NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100

More information

Xcelera version comparison of features and functionality

Xcelera version comparison of features and functionality Xcelera version comparison of features and functionality Xcelera is a robust multimodality cardiology image management, analysis and reporting solution that provides patient centric access to cardiology

More information