Cerebrospinal fluid shunting is required when endogenous. Flow characteristics of cerebrospinal fluid shunt tubing. Laboratory investigation

Size: px
Start display at page:

Download "Cerebrospinal fluid shunting is required when endogenous. Flow characteristics of cerebrospinal fluid shunt tubing. Laboratory investigation"

Transcription

1 J Neurosurg Pediatrics 9: , 9: , 2012 Flow characteristics of cerebrospinal fluid shunt tubing Laboratory investigation Joseph T. Cheatle, M.D., Alexis N. Bowder, B.S., B.A., Sandeep K. Agrawal, Ph.D., Michael D. Sather, M.D., and Leslie C. Hellbusch, M.D. Division of Neurosurgery, Department of Surgery, University of Nebraska Medical Center, Omaha, Nebraska Object. Cerebrospinal fluid shunt systems malfunction for a multitude of reasons, including malpostitioning, obstruction of the ventricular or distal catheter, obstruction of the shunt valve, and catheter disruptions or disconnections. The goal of this study was to examine the hydrodynamic resistance and flow in new and explanted catheters and also in catheters with 1 or 2 straight connectors. Methods. Explanted catheters of multiple lengths, 2-piece catheters, 3-piece catheters, and new catheters were attached to a proximal and distal manometer. A flask with artificial CSF attached to the proximal end provided flow. The flow was allowed to stabilize over 1 hour; then the change in pressure between the proximal and distal end of the catheter was measured. Results. The resistance to flow was calculated for new, never-implanted catheters and compared with the resistance of explanted distal shunt catheters. The resistance of the new catheters was examined after the addition of 1 and 2 straight connectors. Explanted catheters exhibited a slight increase in the resistance to flow of artificial CSF compared with new catheters. Two-piece and 3-piece catheters had a significant increase in resistance to flow compared with new catheters. For all catheters, resistance to flow increased as length increased (new, p = 0.01; explanted, p = 0.009; 1 connector, p = 0.01; 2 connectors, p = 0.03). In this paper, effective diameter is defined as the available crosssectional area of catheter contacted by the artificial CSF. For new and explanted catheters, a decrease in the effective diameter of the catheter was associated with an increase in the resistance to flow of artificial CSF (new, p = ; explanted, p = ). However, after the addition of 1 or 2 connectors, an inverse trend was observed: resistance to flow increased with effective diameter. Conclusions. There appears to be some increase in resistance of CSF shunt catheters as they age, altering flow dynamics. In addition, the use of straight connectors within a CSF shunt system increases the resistance to flow of artificial CSF within the shunt system. The increase in resistance appears to be related to the duration of implantation and the length of the catheter and inversely related to the diameter of the catheter. This increase in resistance may be related to sterile shunt malfunction. The addition of straight connectors is associated with a significant increase in resistance in comparison with catheters without connectors (p = 0.005). ( Key Words cerebrospinal fluid shunt shunt malfunction cerebrospinal fluid dynamics hydrocephalus Cerebrospinal fluid shunting is required when endogenous pathways of circulation and reabsorption of CSF are disrupted. Cerebrospinal fluid shunt system failure is a significant cause of morbidity in patients with hydrocephalus. Shunt system malfunctions can occur due to catheter breakage or disconnection, obstruction of either the ventricular or distal catheter, obstruction of the shunt valve, or malpositioning. As biomechanical devices, CSF shunt systems are inherently prone to failure and infection. In the 1st year after implantation alone, the average failure rate of a CSF shunt is 30%, and reports show that up to 50% of CSF shunt systems fail over time. 6,16 The flow of CSF within a shunt system depends on intraventricular pressure, the diameter and length of the catheter, and the resistance of the shunt system, including the internal resistance in the shunt tubing and the distal pressure at the site of CSF drainage. Distal catheter malfunction is well documented in the literature. Infection, bowel or bladder perforation, pseudocyst formation, volvulus, abdominal wall or diaphragm penetration, umbilical or inguinal hernia formation, scrotal extrusion, and unexplained failure are all described. Late shunt malfunction secondary to smoldering infections, embolization of choroid plexus and leptomeninges, or cellular invasion may contribute to late-term shunt malfunction and changes in resistance to CSF flow. 10 This article contains some figures that are displayed in color on line but in black and white in the print edition. 191

2 J. T. Cheatle et al. Sterile shunt malfunction is a known cause of shunt failure. Previous studies outline reaction to silicone within the body via a delayed hypersensitivity reaction. 2,3,11 Upon electron microscopy examination, shunts with sterile malfunction have been found to have increased proteinaceous and cellular debris. 10 This reaction may change the resistance to flow of CSF and result in less than optimal shunt function or outright failure. Calcification of the external surface of catheters and microstructure alteration of silicone catheters has been demonstrated in in vitro samples of explanted silicone catheters. 5 The resistance of different shunt valves alone and in series with their catheters has been well documented in the literature. 7 Valves have low intrinsic hydrodynamic resistance. A valve in series with a catheter increases the resistance in the system by 100% 200%, 10 but little is known about how the addition of straight connectors to the distal catheter can change the resistance to flow of the CSF shunt system. To date, no formal studies have been performed on explanted shunt catheters to determine if the resistance changes over the duration of implantation. The distal catheter is responsible for a significant portion of the total resistance, and a significant increase in resistance over the life of a shunt can alter the function of the CSF shunt system. In this study we compare the hydrodynamic resistance properties of explanted distal shunt catheters to identical new distal shunt catheters. In addition, we examine the effects of straight connectors on CSF shunt systems in vitro. Methods New Tubing and Straight Connectors New CSF shunt tubing and straight connectors were donated by Codman. All of the tubing was suitable for implantation without having any defects or being past the expiration date. The shunt tubing from Codman had an inner diameter of 1.0 mm and outer diameter of 2.2 mm and was originally 120 cm in length. The Codman straight connectors with beveled tips were 11.1 mm in length and had an inner diameter of 1.0 mm and outer diameter of 1.9 mm. Used Tubing Under the approval of the institutional review board, distal CSF shunt tubing was obtained from patients in the course of noninfected CSF shunt revision surgery. The tubing was used to study the flow and resistance of used tubing. The manufacturer and model number of the retrieved shunt tubing were identified by visual inspection and confirmed by reviewing operative reports. Subsequent chart review allowed us to determine the length of time the shunt tubing was implanted, as well as the original diagnosis necessitating placement of a CSF shunt system. If an infection was known or suspected at the time of surgery, the shunt tubing was excluded from the study. Retrieved shunt specimens were tested while this study was active between 2004 and Artificial CSF Artificial CSF was made to closely resemble natural CSF: ml aliquots of distilled water were mixed with 126 mm sodium chloride, 3 mm potassium chloride, and 1.25 mm monosodium phosphate until dissolved; 2 mm magnesium chloride and 2 mm of calcium chloride were subsequently added and allowed to dissolve; and 26 mm sodium bicarbonate and 10 mm of dextrose were added last and allowed to dissolve. The artificial CSF was allowed to equilibrate to the temperature of the water bath to closely resemble in vivo conditions. Measurement of CSF Shunt Tubing Resistance A water bath system as pictured in Fig. 1 was maintained at 37 ± 2 C. Manometers were placed at each end of the water bath with a constant height ± 2 mm. A flask with a redundant catheter containing 350 ml of artificial CSF was connected to the proximal CSF drainage catheter and placed in the water bath. The flask and catheter were constant in all experiments. The catheter was of adequate length to ensure that the artificial CSF was equilibrated to the temperature of the water bath. The proximal end of the catheter was attached to the proximal manometer. The distal end of the catheter was then attached to the distal manometer. All air was purged from the system. The pump was set to provide flow of 15 ml per hour. After 1 hour the water column heights of the proximal and distal manometer were read and recorded, as was the total volume of artificial CSF remaining in the beaker. To examine the effects of straight connectors in vitro, the new distal catheter tubing was cut to lengths of 18, 35, 52, 85.7, 93, 96, and 98 cm. These specific lengths were chosen to match lengths of explanted catheters. For each length the resistance was measured using the above procedure for a control catheter (with no connector), a catheter with 1 connector (2-piece catheter), and a catheter with 2 connectors (3-piece catheter). Resistance was measured 5 times for condition (control, 1 connector, and 2 connectors) with catheters that were 18, 35, 52, 85.7, 93, 96, and 98 cm in length. Therefore, for each length 15 trials were performed. Determination of Resistance to Flow The following variables were recorded during the measurement of resistance: starting volume of artificial CSF, ending volume of artificial CSF, total volume of artificial CSF used, height of the water column in proximal Fig. 1. Schematic of shunt flow system within a water bath maintained at 37.0 ± 2.0 C. acsf = artificial CSF. 192

3 Flow characteristics of CSF shunt tubing manometer, and height of water column in distal manometer. The Poiseuille law states that the volume flow of an incompressible fluid through a circular tube is equal to p/8 times the pressure differences between the ends of the tube, times the fourth power of the tube s radius divided by the product of the tube s length and the dynamic viscosity of the fluid. 9 Using the data obtained and the Poiseuille law, shown in Equation 1, the resistance to flow was calculated for each trial. Equation 1. Using the Poiseuille equation assuming a constant velocity, the resistance to flow for CSF (a Newtonian fluid) through the CSF shunt system was calculated, where r = radius of the catheter, l = length of distal catheter, and h = viscosity of the fluid. A predetermined viscosity of CSF at 0.79 mpa sec 4 was used for the viscosity in each resistance to flow calculation. Q = (P proximal P distal ) π r 4 / 8lη Eq. 1 Effective diameter was calculated by solving the Poiseuille equation for effective diameter, and the resulting equation (Equation 2) was used to calculate the effective diameter for each trial. Equation 2. The following equation was derived from the Poiseuille equation and used to calculate effective diameter (D) in each catheter, where l = length of tube, h = viscosity of the fluid in mpa sec, and Q = resistance to flow of CSF in Pa sec: D = 2 ( η 8l/Q/π) 1/4 Eq. 2 Finally, for each trial the Reynolds number was calculated (Equation 3), and the flow through the shunt system was characterized as either laminar or turbulent. Averages for resistance to flow, effective diameter, and Reynolds numbers and their corresponding standard deviations were calculated for each condition. Equation 3. The following equation was used to calculate the Reynolds number for each trial, where h = viscosity (in centistokes), Q = resistance to flow of CSF (in gallons/minute), and D = effective diameter (in millimeters) 9 : R e = (3160 Q)/(η D) Eq. 3 Results New Versus Explanted Shunts Seven different lengths of shunt tubing were examined in the laboratory (Table 1). The mean resistances of these catheters are illustrated in Fig. 2. The increasing length of catheters is associated with an arithmetic increase in the resistance (p = 0.028) seen in Fig. 2 lower. Another factor regarding resistance is the diameter of the tubing, as seen in Fig. 2 upper (p = 0.028). Here the resistance of the catheter is inversely related to the diameter of the shunt tubing. Table 2 summarizes the data obtained in explanted catheters from 23 patients (age range 1 35 years) who underwent ventriculoperitoneal shunt revisions without any signs of infection. The resistance to flow seen in Fig. TABLE 1: Reynolds number values obtained in new catheters* Shunt Length (cm) Reynolds Number controls (no connectors) New New New New New New New piece catheters (1 connector) New New New New New New New piece catheters (2 connectors) New New New New New New New * For all measurements involving new catheters, the tubing used was Codman Bactiseal (Ref ; SN ). 3, upper, demonstrates that length is much more important in the explanted catheters than in new catheters (p = ), suggesting the internal diameter becomes stenotic with age. The change in resistance with an increase in internal effective diameter is seen in Fig. 3, lower (p = 0.009). Comparison of new shunt tubing and older shunt tubing showed that new tubing has a slightly lower resistance than older tubing (Fig. 4). New Shunts With the Addition of Straight Connectors Seven different lengths of shunt tubing with 1 or 2 connectors were examined in the laboratory. The mean resistances to flow for each length and condition are shown in Fig. 5. The mean resistance to flow increased with length (1 connector, p = 0.01; 2 connectors, p = 0.03). More importantly, the mean resistance to flow increased significantly after the addition of 1 connector or 2 connectors to the shunt system (Fig. 5). A t-test revealed that there was also a significant difference in resistance between 2- and 3-piece catheters (p = 0.005). After the calculations of mean effective diameter were completed for each condition and at each length, the mean resistance to flow for the mean effective diameter was examined. In the control condition, increased effective diameter was associated with decreased resistance to 193

4 J. T. Cheatle et al. Fig. 2. Line graphs demonstrating how average resistance per trial (average of 5 trials) is affected by the effective diameter and length of new catheters. Upper: The average resistance to flow plotted as a function of effective diameter. In new catheters as length increases and effective diameter decreases, resistance to flow increases. The average resistance to flow measured in Pa sec/m 3 is displayed on the y-axis where E+10 denotes scientific notation (1.00E+10 = ). Lower: The average resistance to flow plotted as a function of length. flow as shown in Fig. 3. In contrast, when effective diameter increased in catheters containing either 1 or 2 straight connectors, resistance to flow increased as well (Fig. 6). To characterize the flow through shunts, Reynolds numbers were used to determine whether flow was laminar or turbulent. Laminar flow indicates smooth, constant flow of a fluid through a system and correlates to Reynolds number values less than 2,300. The calculations of Reynolds numbers were completed for each condition at each length, and the results are displayed in Fig. 7 and Table 1. For all 3 conditions, the Reynolds number indicated laminar flow with values that were well below the cutoff for turbulent flow. In addition, the Reynolds numbers appeared to decrease as the length of the catheters increased. Discussion Cerebrospinal fluid shunts are the mainstay of hydrocephalus treatment. However, dysfunction of CSF shunt systems continues to be a common problem. 1 As with all mechanical implanted devices, shunts are prone to failure, infection, and infiltration by host tissue. Little is known about the changes in resistance associated with aging of shunts or the effect of such changes on shunt function, although it appears that shunt tubing predictably becomes more resistive to flow, possibly contributing to shunt failure. In this study we examined new shunt tubing Fig. 3. Line graphs demonstrating how average resistance for the 23 explanted catheters is affected by the effective diameter and length of each catheter. Upper: The average resistance to flow plotted as a function of length. Lower: The average resistance to flow plotted as a function of effective diameter. As seen with the new catheters in Fig. 2, an increase in the length of explanted catheters corresponds to an increase in resistance to flow. In addition, smaller effective diameters lead to an increase in resistance to flow in explanted catheters. and compared this to explanted tubing removed because of CSF shunt malfunction. Explanted shunt tubing has a slight increase in resistance to flow compared with new tubing. This gradual increase in resistance is of unknown physiological significance, but it may contribute to sterile shunt malfunction. It is possible that during shunt replacements the shunt tubing may be stretched or damaged upon removal. Such damage could possibly have altered the resistance to flow in the explanted shunts that we used in our study, although we very carefully examined the entire visible surface of the shunts before testing. Nevertheless, in situ studies of resistance to flow would not be practical because it would be difficult to obtain multiple trials for each patient and would potentially increase the risk of shunt revision surgery. For incompressible flow within a viscid fluid, the ki- Fig. 4. Bar graph comparing the change in average resistance to flow between new and explanted catheters. 194

5 Flow characteristics of CSF shunt tubing TABLE 2: Resistance and effective diameter in explanted catheters* Shunt Mean Resistance (Pa sec/m 3 ) SD Mean Effective Diameter (mm) SD Explanted E E E 02 Explanted E E E 02 Explanted E E E 02 Explanted E E E 02 Explanted E E E 01 Explanted E E E 02 Explanted E E E 02 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 Explanted E E E 02 Explanted E E E 02 Explanted E E E 02 Explanted E E E 01 Explanted E E E 01 Explanted E E E 01 * Values are presented in scientific E notation: 1.00E+10 = ; 1.00E 01 = Fig. 5. Bar graph depicting the average resistance to flow for the control, 2-piece, and 3-piece catheters at 7 different lengths. None, one, and two refer to the number of connectors. *Significant increase in average resistance to flow (p = 0.05). netic and potential energy of a fluid must remain constant. Therefore, when a fluid moves from a region of low pressure to a region of higher pressure, the speed at which it flows must decrease. Cerebrospinal fluid is a Newtonian fluid, and its viscosity is nearly that of water ranging from 0.7 to 1 mpa sec at 37 C. The viscosity of our artificial CSF was found to be 0.79 mpa sec. 4,12 In nonimplanted CSF shunts, CSF flow is laminar. The Reynolds number for each of the explanted catheters remains low, well below the threshold of turbulent or even transitional flow, and therefore the characteristic nature of the flow remains laminar. The change in the resistance to flow, although significant, does not cause turbulent flow. This suggests that a relatively uniform stenosis along the length of the tube occurs rather than a stenotic area with a relatively spared diameter in other areas, as occurs in blood vessels in vivo. 13 The standard shunt system consists of tubing with fixed resistance and a 1-way differential pressure valve. 15,16 The flow of fluid through this system is reliant on both the differential pressure across the shunt and the resistance of the shunt itself. This resistance can be reproducibly set by the manufacturers of the shunt tubing. However, within a shunt system there exists a series of resistors: the opening of the ventricular catheter (Rc), the valve (Rv), any shunt accessories (Ra), and finally, the distal tubing (Rt). Therefore the total resistance of a shunt system can be represented in the following equation: 15 Rs = Rc + Rv + Ra + Rt. In some shunts the values of Rv and Ra are designed to change with position. 14 The changes in resistance to flow after the addition of various shunt valves has been well documented. A valve in series with a catheter increases the resistance in the system by 100% 200%. 10 In 195

6 J. T. Cheatle et al. Fig. 7. Bar graph depicting the Reynolds number for control, 2-, and 3-piece catheters at 7 different lengths of catheter. As expected all trials had Reynolds numbers indicating laminar flow with values well below the threshold for turbulent flow of 2,300. *Significant difference in Reynolds numbers (p = 0.05). Fig. 6. Line graphs showing how effective diameter affects the average resistance to flow in 2- and 3-piece catheters. Upper: The average resistance for 2-piece catheters plotted as a function of effective diameter. Lower: The average resistance to flow for 3-piece catheters plotted as a function of effective diameter. In contrast to the trends displayed by the control or new catheters in Fig. 3, as internal diameter increased resistance to flow also increased. This suggests that the connectors themselves create an additional source of resistance within a CSF shunt system. addition, a study examining the effect of the use of Luer connector devices in arterial and ventricular catheters showed that these devices exerted an increased resistance to flow within the catheter system. 8 Therefore, our findings that the resistance to flow within CSF shunt systems increases with the addition of 1 or 2 straight connectors is consistent with previous research. If the resistance of the connectors (Rsc 1 and Rsc 2 ) were added to the above equation, one would expect the resistance to flow to be significantly different between shunts of similar length containing either 1 or 2 connectors. This assumption is supported by our findings: Rc + Rv + Ra + Rt + Rsc 1 Rc + Rv + Ra + Rt + Rsc 1 + Rsc 2. Connectors are added to a shunt system for 2 reasons: to repair a break and to lengthen the shunt as the patient grows. 15 Not only must shunts sometimes be lengthened using connectors, but shunt valves must also be adjusted for each patient based on activity level, resistance to flow within their system, and their production of CSF. An increased resistance to flow may contribute to shunt malfunction by causing a slower flow through the entire system. A higher resistance to flow could also result in mildly increased intracranial pressure in shunt-treated patients. Knowing that the addition of connectors to a shunt system results in a significant increase in resistance to flow within a CSF shunt system will enable physicians to consider adjusting the CSF shunt valves accordingly. It should be noted that all of the new shunts and straight connectors used in this experiment were from Codman. The different trends between the control conditions and connector conditions in respect to the effect of diameter on resistance to flow support the conclusion that the connectors create additional resistance within the shunt system. As diameter increases within a viscous fluid system, the resistance to flow naturally decreases because there is more area through which the fluid can flow. 9 This trend was exhibited in the control group. When connectors were added to the CSF shunt system we saw an inverse trend. As effective diameter increased, resistance to flow increased as well. This increase in resistance is due to the presence of a connector within the shunt system. Resistance to flow increased as the viscous CSF fluid was forced through the opening of the straight connectors within the CSF shunt systems. As expected, the Reynolds numbers for all runs were well under the thresholds of turbulent or even transitional flow, and therefore the characteristic nature of the flow remained laminar. Therefore, the change in resistance to flow with the addition of connectors, although significant, does not cause turbulent flow. The changes in the resistance to flow observed in this study represent a preliminary look at the dynamics of a CSF shunt system. In addition, it might be worthwhile to include histological examination using electron microscopy of the explanted shunt tubing to see if proteinaceous debris or other components are contributing to stenosis over time. Conclusions An increase in resistance within a CSF shunt system appears to be directly related to the length of the catheter and inversely related to the effective diameter of the catheter. This increase in resistance may be related to sterile shunt malfunction. The addition of straight connectors shows a significant increase in resistance in comparison with catheters without connectors (p = 0.005). There is also a significant difference between resistances to flow in catheters containing 1 connector versus the resistance to flow in catheters containing 2 connectors (p = 0.005). 196

7 Flow characteristics of CSF shunt tubing Disclosure New shunt tubing and connectors for this study were donated by Codman. Author contributions to the study and manuscript preparation include the following. Conception and design: Bowder, Agrawal, Sather, Hellbusch. Acquisition of data: Bowder, Agrawal, Satire. Analysis and interpretation of data: all authors. Drafting the article: Bowder, Cheatle, Agrawal, Hellbusch. Critically revising the article: all authors. Reviewed submitted version of manuscript: all authors. Approved the final version of the manuscript on behalf of all authors: Bowder. Statistical analysis: Bowder, Cheatle. Administrative/ technical/material support: Agrawal, Hellbusch. Study supervision: Agrawal, Hellbusch. References 1. Aschoff A, Kremer P, Hashemi B, Kunze S: The scientific history of hydrocephalus and its treatment. Neurosurg Rev 22:67 95, Baldwin CM Jr, Kaplan EN: Silicone-induced human adjuvant disease? Ann Plast Surg 10: , Bass SJ, Gastwirth CM, Green R, Knights EM, Weinstock RE: Phagocytosis of silastic material following silastic great toe implant. J Foot Surg 17:70 72, Bloomfield IG, Johnston IH, Bilston LE: Effects of proteins, blood cells and glucose on the viscosity of cerebrospinal fluid. Pediatr Neurosurg 28: , Boch AL, Hermelin E, Sainte-Rose C, Sgouros S: Mechanical dysfunction of ventriculoperitoneal shunts caused by calcification of the silicone rubber catheter. J Neurosurg 88: , Borgbjerg BM, Gjerris F, Albeck MJ, Hauerberg J, Børgesen SE: Frequency and causes of shunt revisions in different cerebrospinal fluid shunt types. Acta Neurochir (Wien) 136: , Czosnyka Z, Czosnyka M, Richards H, Pickard JD: Hydrodynamic properties of hydrocephalus shunts. Acta Neurochir Suppl 71: , Eloot S, De Vos JY, Hombrouckx R, Verdonck P: How much is catheter flow influenced by the use of closed luer lock access devices? Nephrol Dial Transplant 22: , Fox RW, Pritchard PJ, McDonald AT: Introduction to Fluid Mechanics, ed 7. Hoboken JN: John Wiley and Sons, 2005, p Gower DJ, Lewis JC, Kelly DL Jr: Sterile shunt malfunction. A scanning electron microscopic perspective. J Neurosurg 61: , Heggers JP, Kossovsky N, Parsons RW, Robson MC, Pelley RP, Raine TJ: Biocompatibility of silicone implants. Ann Plast Surg 11:38 45, Howden L, Giddings D, Power H, Aroussi A, Vloeberghs M, Garnett M, et al: Three-dimensional cerebrospinal flow within the human ventricular system. Comput Methods Biomech Biomed Engin 11: , Ku DN: Blood flow in arteries. Annu Rev Fluid Mech 29: , Magram G: Cerebrospinal fluid shunt management: Part 1. Neurologist 2: , Magram G, Liakos AM: Cerebrospinal fluid flow through an implanted shunt. Neurol Res 22:43 50, Malm J, Lundkvist B, Eklund A, Koskinen LO, Kristensen B: CSF outflow resistance as predictor of shunt function. A longterm study. Acta Neurol Scand 110: , Sendelbeck L: Recipe for preparation of artificial CSF. Special Delivery 8:4, 1987 Manuscript submitted June 29, Accepted November 7, Please include this information when citing this paper: DOI: / PEDS Address correspondence to: Alexis Bowder, B.S., B.A., Department of Surgery, Division of Neurosurgery, Nebraska Medical Center, Omaha, Nebraska alexis.bowder@unmc. edu. 197

THESIS. Christian L. Blough. Undergraduate Program in Biomedical Engineering. The Ohio State University. Thesis Committee:

THESIS. Christian L. Blough. Undergraduate Program in Biomedical Engineering. The Ohio State University. Thesis Committee: Measuring Cerebrospinal Fluid Pressure With a Fluid Manometer THESIS Presented in Partial Fulfillment of the Requirements for the Degree Bachelors of Science in the Undergraduate School of The Ohio State

More information

Clinical Study The Value of Programmable Shunt Valves for the Management of Subdural Collections in Patients with Hydrocephalus

Clinical Study The Value of Programmable Shunt Valves for the Management of Subdural Collections in Patients with Hydrocephalus The Scientific World Journal Volume 2013, Article ID 461896, 4 pages http://dx.doi.org/10.1155/2013/461896 Clinical Study The Value of Programmable Shunt Valves for the Management of Subdural Collections

More information

Physics of the Cardiovascular System

Physics of the Cardiovascular System Dentistry College Medical Physics Physics of the Cardiovascular System The cells of the body act like individual engines. In order for them to function they must have: - 1. Fuel from our food to supply

More information

Principlesof shunt testing in-vivo Zofia Czosnyka, Matthew Garnett, Eva Nabbanja, Marek Czosnyka

Principlesof shunt testing in-vivo Zofia Czosnyka, Matthew Garnett, Eva Nabbanja, Marek Czosnyka Principlesof shunt testing in-vivo Zofia Czosnyka, Matthew Garnett, Eva Nabbanja, Marek Czosnyka Neurosurgery Unit, Addenbrooke s Hospital, Cambridge, UK Shunt helps to control hydrocephalus not to cure

More information

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm I. Husain, C. Langdon and J. Schwark Department of Mathematics Luther College University of Regina Regina, Saskatchewan

More information

Aesculap Neurosurgery Hydrocephalus Shunt Accessories

Aesculap Neurosurgery Hydrocephalus Shunt Accessories Aesculap Neurosurgery Hydrocephalus Shunt Accessories MIETHKE Accessories for MIETHKE shunt systems MIETHKE Accessories CATHETERS All catheters are made of radiopaque silicone d i = 1.2 mm d o = 2.5 mm

More information

Original Article. Emergency Department Evaluation of Ventricular Shunt Malfunction. Is the Shunt Series Really Necessary? Raymond Pitetti, MD, MPH

Original Article. Emergency Department Evaluation of Ventricular Shunt Malfunction. Is the Shunt Series Really Necessary? Raymond Pitetti, MD, MPH Original Article Emergency Department Evaluation of Ventricular Shunt Malfunction Is the Shunt Series Really Necessary? Raymond Pitetti, MD, MPH Objective: The malfunction of a ventricular shunt is one

More information

Cardioplegia Circuit Products { CARDIOPLEGIA ADAPTERS}

Cardioplegia Circuit Products { CARDIOPLEGIA ADAPTERS} Cardioplegia Circuit Products { CARDIOPLEGIA ADAPTERS} Cardioplegia adapters are designed to permit customization of the cardioplegia circuit. Medtronic adapters can be used to connect multiple cardioplegia

More information

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 Presented at the COMSOL Conference 2010 India ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 1 DEPARTMENT OF MECHANICAL ENGINEERING & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING

More information

CVS Hemodynamics. Faisal I. Mohammed, MD,PhD.

CVS Hemodynamics. Faisal I. Mohammed, MD,PhD. CVS Hemodynamics Faisal I. Mohammed, MD,PhD. Objectives point out the physical characteristics of the circulation: distribution of blood volume total cross sectional area velocity blood pressure List the

More information

1 Non-invasive measurement of arterial pressure

1 Non-invasive measurement of arterial pressure Non-invasive measurement of arterial pressure I. Background A. Circulatory systems Human circulation takes place in a closed system that consists of two subsystems, pulmonary circulation and systemic circulation,

More information

Researcher 2018;10(3)

Researcher 2018;10(3) Comparative study between frontal and posterior parietal ventriculo-peritoneal shunts in management of hydrocephalus Prof. Dr. Abdelhalim Moasa, Dr. Mohamed Hasan and Mohamed Abdelrazek Department of Neurosurgery,

More information

Perioperative Management Of Extra-Ventricular Drains (EVD)

Perioperative Management Of Extra-Ventricular Drains (EVD) Perioperative Management Of Extra-Ventricular Drains (EVD) Dr. Vijay Tarnal MBBS, FRCA Clinical Assistant Professor Division of Neuroanesthesiology Division of Head & Neck Anesthesiology Michigan Medicine

More information

Radiological evaluation of ventriculoperitoneal shunt systems

Radiological evaluation of ventriculoperitoneal shunt systems Radiological evaluation of ventriculoperitoneal shunt systems Poster No.: C-702 Congress: ECR 2009 Type: Educational Exhibit Topic: Neuro Authors: T. Rodt, C. von Falck, C. Tschan, M. Diensthuber, J. Zajaczek,

More information

Anesthetic Management of Laparoscopic Surgery for a Patient with

Anesthetic Management of Laparoscopic Surgery for a Patient with Anesthetic Management of Laparoscopic Surgery for a Patient with a Ventriculoperitoneal shunt Abstract With the advances in the management of hydrocephalus, patients with ventriculoperitoneal shunt are

More information

S ome hydrocephalic patients with extracranial shunts

S ome hydrocephalic patients with extracranial shunts PAPER Quantitative analysis of continuous intracranial pressure recordings in symptomatic patients with extracranial shunts P K Eide... Competing interests: none declared... Correspondence to: Dr Per Kristian

More information

BACTISEAL Endoscopic Ventricular Catheter (REF & )

BACTISEAL Endoscopic Ventricular Catheter (REF & ) BACTISEAL Endoscopic Ventricular Catheter (REF 82-3087 & 82-3088) LCN 206966-001/E 2011 2017 DePuy Synthes. All rights reserved. Revised 07/17 206966-001-E.indd 1 ENGLISH IMPORTANT INFORMATION Please Read

More information

The cardiovascular system is composed of a pump the heart and blood

The cardiovascular system is composed of a pump the heart and blood 5 E X E R C I S E Cardiovascular Dynamics O B J E C T I V E S 1. To understand the relationships among blood flow, pressure gradient, and resistance 2. To define resistance and describe the main factors

More information

Infusion studies in clinical practice. Kristian Aquilina Consultant paediatric neurosurgeon Great Ormond Street Hospital London

Infusion studies in clinical practice. Kristian Aquilina Consultant paediatric neurosurgeon Great Ormond Street Hospital London Infusion studies in clinical practice Kristian Aquilina Consultant paediatric neurosurgeon Great Ormond Street Hospital London 10 th September 2018 infusion study + hydrocephalus 216 publications Clinical

More information

Ventriculoperitoneal Shunt with Communicating Peritoneal & Subcutaneous Pseudocysts Formation

Ventriculoperitoneal Shunt with Communicating Peritoneal & Subcutaneous Pseudocysts Formation International Journal of Health Sciences, Qassim University, Vol. 8, No. 1 (January-March 2014) Case Report Ventriculoperitoneal Shunt with Communicating Peritoneal & Subcutaneous Pseudocysts Formation

More information

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

The resistance to CSF outflow in hydrocephalus what it is and what it isn t.

The resistance to CSF outflow in hydrocephalus what it is and what it isn t. The resistance to CSF outflow in hydrocephalus what it is and what it isn t. Davson et al 1970, The mechanism of drainage of CSF. Brain 93:665-8 1989, Copenhagen, Alfred Benzon Foundation CSf outflow is

More information

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM H. GIRIJA BAI 1 and K.B. NAIDU 2 Department of Mathematics, Sathyabama University, Chennai-600 119, Tamil Nadu, India 1 girijanameprakash@gmail.com

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft BURASKORN NUNTADILOK 1, BENCHAWAN WIWATANAPATAPHEE 1 MEECHOKE CHUEDOUNG 1, THANONGCHAI SIRIAPISITH

More information

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER Page 1 of 5 ASPIRATION CATHETER Carefully read all instructions prior to use, observe all warnings and precautions noted throughout these instructions. Failure to do so may result in complications. STERILE.

More information

3. Urinary Catheters. Indications. Methods of Bladder Catheterization. Hashim Hashim

3. Urinary Catheters. Indications. Methods of Bladder Catheterization. Hashim Hashim 3. Urinary Catheters Hashim Hashim Indications Urinary catheters are used to drain urine from the bladder. The main indications are: A. Diagnostic Measure post-void residual in the absence of ultrasound

More information

UC Davis Dermatology Online Journal

UC Davis Dermatology Online Journal UC Davis Dermatology Online Journal Title Scalp necrosis overlying a ventriculoperitoneal shunt: a case report and literature review Permalink https://escholarship.org/uc/item/2rs544f9 Journal Dermatology

More information

Shunt malfunction and Slight edema surrounding the ventricles: Ten case series

Shunt malfunction and Slight edema surrounding the ventricles: Ten case series ISSN: 47-5 Volume Number 8 (August-04) pp. 4-45 www.ijcrar.com Shunt malfunction and Slight edema surrounding the ventricles: Ten case series Firooz Salehpoor, Arastoo Pezeshki, Amirhossein Haghir *, Aidin

More information

Evaluation of Shunt Malfunction Using Shunt Site Reservoir

Evaluation of Shunt Malfunction Using Shunt Site Reservoir Original Paper Pediatr Neurosurg 2000;32:180 186 Received: January 27, 1999 Accepted: April 27, 2000 Evaluation of Shunt Malfunction Using Shunt Site Reservoir S. Sood A.I. Canady Steven D. Ham Section

More information

Branching Blood Vessels

Branching Blood Vessels Jonathan Borne Union Springs Academy Union Springs, NY Summer 2000 Research Host: Mary D. S. Frame, Ph.D. University of Rochester School of Medicine and Dentistry Page 1 of 8 Suggestions for Teachers Purpose:

More information

A Two-layered Model for the Analysis of Arterial Rheology

A Two-layered Model for the Analysis of Arterial Rheology IJCSIT International Journal of Computer Science and Information Technology, Vol., o., June, pp. 37- A Two-layered Model for the Analysis of Arterial Rheology Sapna Singh Department of Mathematics, Harcourt

More information

Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel

Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel RESEARCH ARTICLE OPEN ACCESS Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel S.R. Verma Department of Mathematics, D.A-V. (P.G.) College, Kanpur-208001, India Abstract

More information

Two computerized methods used to analyze intracranial pressure B waves: comparison with traditional visual interpretation

Two computerized methods used to analyze intracranial pressure B waves: comparison with traditional visual interpretation J Neurosurg 94:392 396, 2001 Two computerized methods used to analyze intracranial pressure B waves: comparison with traditional visual interpretation ANDERS EKLUND, MSC.E.P., AINA ÅGREN-WILSSON, M.D.,

More information

ISO INTERNATIONAL STANDARD. Sterile single-use intravascular catheter introducers

ISO INTERNATIONAL STANDARD. Sterile single-use intravascular catheter introducers INTERNATIONAL STANDARD ISO 11070 First edition 1998-05-01 Sterile single-use intravascular catheter introducers Introducteurs de cathéters intravasculaires stériles, non réutilisables A Reference number

More information

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases Proceedings of the 215 2nd International Symposium on Physics and Technology of Sensors, 8-1th March, 215, Pune, India Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

More information

CHAPTER 4 Basic Physiological Principles

CHAPTER 4 Basic Physiological Principles 4-1 CHAPTER 4 Basic Physiological Principles Now that we have a working anatomical knowledge of the heart and circulatory system, we will next develop a functional and quantitative knowledge of the cardiovascular

More information

A STUDY ON VASCULAR RECONSTRUCTION BY FLOW VISUALIZATION*

A STUDY ON VASCULAR RECONSTRUCTION BY FLOW VISUALIZATION* Nagoya J. med. Sci. 36: 91-100, 197-1 A STUDY ON VASCULAR RECONSTRUCTION BY FLOW VISUALIZATION* MASARU ESAKI 1st Departmetlt of Surgery, Nagoya University School of Medicine (Director: Prof. Yotaro lyomasa)

More information

BME. Lesson Title. Grade Level. Stent Design 6-8

BME. Lesson Title. Grade Level. Stent Design 6-8 BME Lesson Title Stent Design Grade Level 6-8 Authors Dr. Brian Davis, Dr. Carin Helfer Purpose This activity aims at using everyday materials to design and develop stents to unclog blood vessels. Materials

More information

Treatment Approaches for Abdominal Migration of Peritoneal Catheter of Ventriculoperitoneal Shunt

Treatment Approaches for Abdominal Migration of Peritoneal Catheter of Ventriculoperitoneal Shunt Treatment Approaches for Abdominal Migration of Peritoneal Catheter of Ventriculoperitoneal Shunt Metin KAPLAN 1 fieyhmuz Kerem ÖZEL 2 Osman DÖNMEZ 3 Ahmet KAZEZ 4 ABSTRACT Migration of peritoneal catheter

More information

Mathematical and Computational study of blood flow through diseased artery

Mathematical and Computational study of blood flow through diseased artery Mathematical and Computational study of blood flow through diseased artery Abstract This paper presents the study of blood flow through a tapered stenosed artery. The fluid (blood) medium is assumed to

More information

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER Page 1 of 5 NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER STERILE. SINGLE USE ONLY. Sterilized with ethylene oxide gas. Non pyrogenic. Do not resterilize. Do not use opened or damaged packages.

More information

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure #6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure Objectives: Observe slide of artery and vein cross-section Auscultate heart sounds using a stethoscope Measure

More information

Hydrocephalus 1/16/2015. Hydrocephalus. Functions of Cerebrospinal fluid (CSF) Flow of CSF

Hydrocephalus 1/16/2015. Hydrocephalus. Functions of Cerebrospinal fluid (CSF) Flow of CSF Hydrocephalus Hydrocephalus Ruth Arms, MSN, CNS-BC, SCRN Hydrocephalus is the buildup of fluid in the cavities (ventricles) deep within the brain. The excess fluid increases the size of the ventricles

More information

ER REBOA Catheter. Instructions for Use

ER REBOA Catheter. Instructions for Use ER REBOA Catheter Instructions for Use Prytime Medical Devices, Inc. 229 N. Main Street Boerne, TX 78006, USA feedback@prytimemedical.com www.prytimemedical.com US 1 210 340 0116 U.S. and Foreign Patents

More information

What is the mechanism of the audible carotid bruit? How does one calculate the velocity of blood flow?

What is the mechanism of the audible carotid bruit? How does one calculate the velocity of blood flow? CASE 8 A 65-year-old man with a history of hypertension and coronary artery disease presents to the emergency center with complaints of left-sided facial numbness and weakness. His blood pressure is normal,

More information

SURGICAL TECHNIQUE GUIDE

SURGICAL TECHNIQUE GUIDE DANGER indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING indicates a potentially hazardous situation which, if not avoided, could result

More information

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

Over the last few decades, endoscopic third ventriculostomy

Over the last few decades, endoscopic third ventriculostomy clinical article J Neurosurg Pediatr 17:734 738, 2016 Long-term follow-up of endoscopic third ventriculostomy performed in the pediatric population Matthew G. Stovell, MBBS, 1 Rasheed Zakaria, MA, BMBCh,

More information

Hydrocephalus is extremely common in the developing

Hydrocephalus is extremely common in the developing J Neurosurg Pediatrics 13:140 144, 2014 AANS, 2014 Effectiveness of the Bactiseal Universal Shunt for reducing shunt infection in a sub-saharan African context: a retrospective cohort study in 160 Ugandan

More information

3908 ½ River Ave. Newport Beach, California contact[at]dollecommunications[dot]com. Hydrocephalus Consult Report Published On-Line

3908 ½ River Ave. Newport Beach, California contact[at]dollecommunications[dot]com. Hydrocephalus Consult Report Published On-Line STEPHEN M. DOLLE Hydrocephalus Monitoring & Consults 3908 ½ River Ave. Newport Beach, California 92663 contact[at]dollecommunications[dot]com Hydrocephalus Consult Report Published On-Line March 4, 2014

More information

MANITOBA RENAL PROGRAM

MANITOBA RENAL PROGRAM MANITOBA RENAL PROGRAM SUBJECT Use of Closed Needleless Access Device with Hemodialysis Central Venous Catheters (CVC) SECTION CODE 30.20.04 30.20 Vascular Access AUTHORIZATION Professional Advisory Committee,

More information

The effect of cerebrospinal fluid on spinal cord deformation in an in vitro burst fracture model

The effect of cerebrospinal fluid on spinal cord deformation in an in vitro burst fracture model The effect of cerebrospinal fluid on spinal cord deformation in an in vitro burst fracture model C. F. Jones 1,2, S. G. Reed 2, P. A. Cripton 2 and R. M. Hall 1 1 School of Mechanical Engineering, University

More information

UNDERSTANDING HYDROCEPHALUS

UNDERSTANDING HYDROCEPHALUS UNDERSTANDING HYDROCEPHALUS Offering solutions for patients. 1 UNDERSTANDING HYDROCEPHALUS Dear Reader: Your doctor has either recommended the CODMAN HAKIM Precision Fixed Pressure Valve, the CODMAN HAKIM

More information

CVS Hemodynamics. Change in blood pressure:

CVS Hemodynamics. Change in blood pressure: CVS Hemodynamics -The distribution of blood inside the circulation: The major part of blood volume is found in the venous system 60% (2/3), that s why veins are called the capacitance vessels. -Arteries

More information

Balloon in Balloon (BIB )

Balloon in Balloon (BIB ) in (BIB ) Dilatation Catheter INSTRUCTIONS FOR USE CAUTION: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN. Read all instructions prior to use. Distributed by: B. Braun

More information

WHILE it is generally agreed that elevation

WHILE it is generally agreed that elevation The Derivation of Coronary Sinus Flow During Elevation of Right Ventricular Pressure By HERMAN M. GELLER, B.S., M.D., MARTIN BRANDFONBRENEU, M.D., AND CARL J. WIGGERS, M.D., The derivation of coronary

More information

Lumbar infusion test in normal pressure hydrocephalus

Lumbar infusion test in normal pressure hydrocephalus Acta Neurol Scand 2005: 111: 379 384 DOI: 10.1111/j.1600-0404.2005.00417.x Copyright Ó Blackwell Munksgaard 2005 ACTA NEUROLOGICA SCANDINAVICA Lumbar infusion test in normal pressure hydrocephalus Kahlon

More information

PTA Balloon Dilatation Catheter

PTA Balloon Dilatation Catheter PTA Balloon Dilatation Catheter March 2016 1/6(3214-0) Catalogue number Serial number Lot number Use by Do not reuse Do not resterilize Do not use if package is damaged Consult instructions for use Keep

More information

A telescopic ventriculoatrial shunt that elongates with growth

A telescopic ventriculoatrial shunt that elongates with growth A telescopic ventriculoatrial shunt that elongates with growth Technical note BURTON L. WISE, M.D. Department of Surgery (Neurosurgery) and Neurological Institute, Mount Zion Hospital and Medical Center,

More information

Clinical Study Endoscopic Third Ventriculostomy in Previously Shunted Children

Clinical Study Endoscopic Third Ventriculostomy in Previously Shunted Children Minimally Invasive Surgery Volume 2013, Article ID 584567, 4 pages http://dx.doi.org/10.1155/2013/584567 Clinical Study Endoscopic Third Ventriculostomy in Previously Shunted Children Eva Brichtova, 1

More information

Part. 2. Urodynamics. Urodynamics. 1 Urethro-cystomanometry Cystomanometry Urethral pressure Abdominal pressure 87

Part. 2. Urodynamics. Urodynamics. 1 Urethro-cystomanometry Cystomanometry Urethral pressure Abdominal pressure 87 80 Catalogue 2007 Part. 2 Urodynamics Urodynamics 1 Urethro-cystomanometry 82 2 Cystomanometry 84 3 Urethral pressure 85 4 Abdominal pressure 87 Catalogue 2007 81 Urethro-cystomanometry Soft Neoplex. Proximal

More information

BMI Medical. Ventricular Catheter T

BMI Medical. Ventricular Catheter T Shunting System Ventricular Catheter 01101 01101T Barium-impregnated silicone catheter provides resistance to kinking and compression. Stainless steel stylet allows catheter to be directed during catheter

More information

Case report. Open Access. Abstract

Case report. Open Access. Abstract Open Access Case report Spontaneous bowel perforation complicating ventriculoperitoneal shunt: a case report Theodosios Birbilis 1, Petros Zezos 2 *, Nikolaos Liratzopoulos 3, Anastasia Oikonomou 4, Michael

More information

NEUROSURGEON VS. HOSPITALIST Pediatric Hospital Medicine meeting Nashville, TN July 21, 2017*±

NEUROSURGEON VS. HOSPITALIST Pediatric Hospital Medicine meeting Nashville, TN July 21, 2017*± NEUROSURGEON VS. HOSPITALIST Pediatric Hospital Medicine meeting Nashville, TN July 21, 2017*± *no pediatricians were harmed in the making of this presentation ±nonetheless, please do not try this at home

More information

Advancing Lives and the Delivery of Health Care. The High-Flow Port Designed for Apheresis

Advancing Lives and the Delivery of Health Care. The High-Flow Port Designed for Apheresis Advancing Lives and the Delivery of Health Care The High-Flow Port Designed for Apheresis Optimized for Long Device Life Bench Tested up to 1,000 Accesses 1 Bard is proud to introduce the first and only

More information

A new method for CSF shunt patency assessment

A new method for CSF shunt patency assessment A new method for CSF shunt patency assessment Marek Swoboda, Matias G. Hochman, Jenna S Fritz, Mark E. Mattiucci, Frederick J Fritz NeuroDx Development LLC CSF shunts employed to treat hydrocephalus patients

More information

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study Available at http://pvamu.edu/aam Appl. Appl. Math. ISSN: 1932-9466 Vol. 12, Issue 1 (June 2017), pp. 291-304 Applications and Applied Mathematics: An International Journal (AAM) Hematocrit Level on Blood

More information

Often, traumatically injured patients need

Often, traumatically injured patients need Comparison of Flow Dynamics of Peripherally and Centrally Inserted Intravenous Catheters Using a Rapid Infusion System (ThermaCor 1200) Eric A. Wrenn, CRNA Racquel Wohlers, CRNA Mandrill Montgomery, CRNA

More information

Assisted Reproduction

Assisted Reproduction Assisted Reproduction Embryo Transfer, Follicle Aspiration, Insemination w w w. s o m a t e x. c o m Assisted Reproduction Top Made in Germany Quality SOMATEX presents a range of products attuned to the

More information

Part 1: General requirements

Part 1: General requirements Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 10555-1 Second edition 2013-06-15 Corrected version 2014-01-15 Intravascular catheters Sterile and single-use catheters Part 1: General requirements

More information

intravenous tubing for the rapid

intravenous tubing for the rapid Archives of Emergency Medicine, 1987, 4, 173-177 Evaluation of 3-mm diameter intravenous tubing for the rapid infusion of fluids G. D. CROSS Department of Anaesthesia, Odstock Hospital, Salisbury, Wiltshire,

More information

PTA Balloon Dilatation Catheter

PTA Balloon Dilatation Catheter PTA Balloon Dilatation Catheter 1/6(3211-0) February 2016 TIS-488-06272017 Catalogue number Serial number Lot number Use by Do not reuse Do not resterilize Do not use if package is damaged Consult instructions

More information

Pediatric hydrocephalus, affects 125,000 children in. Nonprogrammable and programmable cerebrospinal fluid shunt valves: a 5-year study

Pediatric hydrocephalus, affects 125,000 children in. Nonprogrammable and programmable cerebrospinal fluid shunt valves: a 5-year study See the corresponding editorial in this issue, p 461. J Neurosurg Pediatrics 9:000 000, 9:462 467, 2012 Nonprogrammable and programmable cerebrospinal fluid shunt valves: a 5-year study Clinical article

More information

Physiology - 8 Hemodynamics - 1 M.jafar 24/3/2016 Turquoise Team

Physiology - 8 Hemodynamics - 1 M.jafar 24/3/2016 Turquoise Team 21 Physiology - 8 Hemodynamics - 1 M.jafar 24/3/2016 Turquoise Team Hemodynamics Today we will take about hemodynamics which is the study of the movement of blood and of the forces concerned. Now how the

More information

Surgical Support System for Cerebral Aneurysm Coil Embolization

Surgical Support System for Cerebral Aneurysm Coil Embolization Technical Paper Surgical Support System for Cerebral Aneurysm Coil Embolization Yoshitaka NAGANO Yukihiro NISHIO Noriaki MATSUBARA Shigeru MIYACHI Hideo FUJIMOTO Endovascular coil embolization is becoming

More information

VARIATIONS IN THE FLOW OF CEREBROSPINAL FLUID THROUGH SPINAL NEEDLES

VARIATIONS IN THE FLOW OF CEREBROSPINAL FLUID THROUGH SPINAL NEEDLES Br. J. Anaesth. (1987), 59, 1465-1471 VARIATIONS IN THE FLOW OF CEREBROSPINAL FLUID THROUGH SPINAL NEEDLES S. P. GERRISH AND J. E. PEACOCK The use of intrathecal drugs to produce either analgesia or anaesthesia

More information

Collin County Community College

Collin County Community College Collin County Community College BIOL. 2402 Anatomy & Physiology WEEK 6 Blood Vessels 1 Anatomy of Blood Vessels Walls of blood vessels contain 3 distinct layers : Tunica intima innermost layer includes

More information

Structure. Arteries. 21_01d 4/18/12. The Cardiovascular System: Blood Vessels and Hemodynamics. Dr Badri Paudel GMC

Structure. Arteries. 21_01d 4/18/12. The Cardiovascular System: Blood Vessels and Hemodynamics. Dr Badri Paudel GMC Goal of the Cardiovascular System: deliver blood to all parts of the body The Cardiovascular System: Blood Vessels and Hemodynamics Dr Badri Paudel GMC Does so by using different types of tubing, attached

More information

Peel-Apart Percutaneous Introducer Kits for

Peel-Apart Percutaneous Introducer Kits for Bard Access Systems Peel-Apart Percutaneous Introducer Kits for Table of Contents Contents Page Bard Implanted Ports Hickman*, Leonard*, Broviac*, Tenckhoff*, and Groshong* Catheters Introduction....................................

More information

Mathematical Model for the Rupture of Cerebral Saccular Aneurysms through Threedimensional Stress Distribution in the Aneurysm Wall

Mathematical Model for the Rupture of Cerebral Saccular Aneurysms through Threedimensional Stress Distribution in the Aneurysm Wall Mathematical Model for the Rupture of Cerebral Saccular Aneurysms through Threedimensional Stress Distribution in the Aneurysm Wall Hans R. Chaudhry Department of Biomedical Engineering, New Jersey Institute

More information

AXS Catalyst Distal Access Catheter

AXS Catalyst Distal Access Catheter AXS Catalyst Distal Access Catheter Directions for Use 2 AXS Catalyst Distal Access Catheter ONLY Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician. warning Contents

More information

Z-MED II BALLOON AORTIC AND PULMONIC VALVULOPLASTY CATHETER. Instructions for Use

Z-MED II BALLOON AORTIC AND PULMONIC VALVULOPLASTY CATHETER. Instructions for Use Z-MED II BALLOON AORTIC AND PULMONIC VALVULOPLASTY CATHETER Instructions for Use CAUTION: Federal (USA) Law restricts this device to sale by or on the order of a physician. Manufactured for: B. Braun Interventional

More information

longitudinal sinus. A decrease in blood flow was observed when the pressure

longitudinal sinus. A decrease in blood flow was observed when the pressure 362 J. Physiol. (I942) IOI, 362-368 6I2.I44:6I2.824 THE EFFECT OF VARIATIONS IN THE SU.BARACHNOID PRESSURE ON THE VENOUS PRESSURE IN THE SUPERIOR LONGITUDINAL SINUS AND IN THE TORCULAR OF THE DOG BY T.

More information

Postprint.

Postprint. http://www.diva-portal.org Postprint This is the accepted version of a paper published in Journal of Neurology, Neurosurgery and Psychiatry. This paper has been peer-reviewed but does not include the final

More information

Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL

Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL Table of Contents Bryan-DUmon Series II rigid bronchoscope 1. 2. 3. 4. 5. Diagram and Overview Universal Barrel Bronchial and

More information

PHYSIOEX 3.0 EXERCISE 33B: CARDIOVASCULAR DYNAMICS

PHYSIOEX 3.0 EXERCISE 33B: CARDIOVASCULAR DYNAMICS PHYSIOEX 3.0 EXERCISE 33B: CARDIOVASCULAR DYNAMICS Objectives 1. To define the following: blood flow; viscosity; peripheral resistance; systole; diastole; end diastolic volume; end systolic volume; stroke

More information

CP STENT. Large Diameter, Balloon Expandable Stent

CP STENT. Large Diameter, Balloon Expandable Stent CP STENT Large, Expandable CP STENT OPTIONS 12mm to Expansion 26mm to Expansion CP Matrix (number of zigs) 1.6 2.2 2.8 3.4 3.9 4.5 5 5.5 6 12 14 15 16 18 20 22 24 26 28 30 CP Details: CP is composed of

More information

Microtubule Teardrop Patterns

Microtubule Teardrop Patterns Supporting Information Microtubule Teardrop Patterns Kosuke Okeyoshi 1, Ryuzo Kawamura 1, Ryo Yoshida 2, and Yoshihito Osada 1 * 1 RIKEN Advanced Science Institute, 2-1 Hirosawa, Wako-shi, Saitama 351-0198,

More information

Top Vent Study: Optimizing Urine Collection System Flow to Improve Patient Outcomes

Top Vent Study: Optimizing Urine Collection System Flow to Improve Patient Outcomes Top Vent Study: Optimizing Urine Collection System Flow to Improve Patient Outcomes Stephen Tully, R&D Project Engineer, Tyco Healthcare Group LP, Kendall Division, Mansfield, MA Top Vent Study: Optimizing

More information

Ann Marie Flannery, M.D., 1 Ann-Christine Duhaime, M.D., 2 1

Ann Marie Flannery, M.D., 1 Ann-Christine Duhaime, M.D., 2 1 J Neurosurg Pediatrics (Suppl) 14:24 29, 2014 AANS, 2014 Pediatric hydrocephalus: systematic literature review and evidence-based guidelines. Part 3: Endoscopic computer-assisted electromagnetic navigation

More information

The High-Flow Port Designed & Indicated for Apheresis

The High-Flow Port Designed & Indicated for Apheresis The High-Flow Port Designed & Indicated for Apheresis Advancing Lives and the Delivery of Health Care Bard is proud to introduce the first and only high-flow, power-injectable port designed and indicated

More information

Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno. Biophysics of cardiovascular system

Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno. Biophysics of cardiovascular system Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno Biophysics of cardiovascular system 1 Lecture outline Mechanical properties of blood vessels Reynolds

More information

Blood Flow Simulation toward Actual Application at Hospital

Blood Flow Simulation toward Actual Application at Hospital THE 5 TH ASIAN COMPUTAITIONAL FLUID DYNAMICS BUSAN, KOREA, OCTOBER 27 ~ OCTOBER 30, 2003 Blood Flow Simulation toward Actual Application at Hospital Abstract R. Himeno 1 1. Advanced Center for Computing

More information

INSTRUCTIONS FOR USE FOR:

INSTRUCTIONS FOR USE FOR: INSTRUCTIONS FOR USE FOR: en English INSTRUCTIONS FOR USE GORE ENFORM PREPERITONEAL BIOMATERIAL Carefully read all instructions prior to use. Observe all instructions, warnings, and precautions noted throughout.

More information

HISTORY. Question: What category of heart disease is suggested by this history? CHIEF COMPLAINT: Heart murmur present since early infancy.

HISTORY. Question: What category of heart disease is suggested by this history? CHIEF COMPLAINT: Heart murmur present since early infancy. HISTORY 18-year-old man. CHIEF COMPLAINT: Heart murmur present since early infancy. PRESENT ILLNESS: Although normal at birth, a heart murmur was heard at the six week check-up and has persisted since

More information

Novatech Products for Interventional Pulmonology

Novatech Products for Interventional Pulmonology Novatech Products for Novatech and Boston Medical Products Bringing you the finest products for Novatech is a manufacturer of top-quality medical products used successfully worldwide in the growing specialty

More information

Adult Cannulae. product catalog

Adult Cannulae. product catalog Adult Cannulae product catalog cardiac cardiac surgery solutions Table of Contents Arterial Cannulae 4 Curved Tip Aortic Arch Cannulae 4 Straight Tip Aortic Arch Cannulae 11 Flexible Aortic Arch Cannulae

More information

Occult Cerebrospinal Fluid Fistula between Ventricle and Extra-Ventricular Position of the Ventriculoperitoneal Shunt Tip

Occult Cerebrospinal Fluid Fistula between Ventricle and Extra-Ventricular Position of the Ventriculoperitoneal Shunt Tip 197 Occult Cerebrospinal Fluid Fistula between Ventricle and Extra-Ventricular Position of the Ventriculoperitoneal Shunt Tip Ching-Yi Lee 1, Chieh-Tsai Wu 1, Kuang-Lin Lin 2, Hsun-Hui Hsu 3 Abstract-

More information

CONFIGURE YOUR OWN CUSTOM-MADE SUPERIOR FOLLICLE ASPIRATION SET IN 7 STEPS

CONFIGURE YOUR OWN CUSTOM-MADE SUPERIOR FOLLICLE ASPIRATION SET IN 7 STEPS CONFIGURE YOUR OWN CUSTOM-MADE SUPERIOR FOLLICLE ASPIRATION SET IN 7 STEPS COPY OR DOWNLOAD AT WWW.GYNETICS.BE, COMPLETE AND RETURN FOR INQUIRY STEP 1 CHOOSE YOUR FLUSH NEEDLE TICK BOX Needle Lumen Gauge

More information

Gastric / EUS Metal Stents

Gastric / EUS Metal Stents Gastric / EUS Metal Stents In a fast paced and maturing market, Diagmed Healthcare s Hanarostent has managed to continue to innovate and add unique and clinically superior features to its already premium

More information