HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

Size: px
Start display at page:

Download "HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER"

Transcription

1 HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER Naser KA. Teaching Hospital Peradeniya, Peradeniya, Sri Lanka Zawahir S School of Pharmacy, Management and Science University Redhwan Ahmed Al-Naggar Faculty of Medicine, Universiti Teknologi MARA ABSTRACT Electronic blood pressure devices commonly used in home monitory and clinical setup. Our objective was to determine the accuracy of different types of digital (electronic) blood pressure and pulse measuring devices. Methods: Two hundred patients were studied at a Diabetes clinic in Sri Lanka. Five digital devices were tested comparing with that of mercury sphygmomanometer (40 patients for each digital device). Blood pressure and pulse rate were recorded in each patient with mercury sphygmomanometer and one of the study digital sphygmomanometers. Precision of measurement for mercury sphygmomanometer was 2mm Hg and for digital device was 1mm Hg. Data were descriptively analyzed by using Kappa measurement of agreement method and paired t-test. Results: Compared to sphygmomanometer readings, Systolic Blood Pressure (SBP) readings were higher in all digital devices and the median pulse rate readings were also showing the same results. Kappa measure of agreement between digital devices and mercury device readings were calculated. In device A both SBP and (Diastolic Blood Pressure)DBP showed Kappa<0.3 but findings are not significant. The pulse rate showed Kappa >0.3 with a significant fair agreement with manual counting. In device B pulse rate showed very good agreement (Kappa-0.8, P<0.01), while Blood Pressure showed poor agreement between devices (p=0.09). Device C and E showed considerable agreement between the devices (P<0.001). The device D showed moderate to good agreement between the devices (P<0.001) except for SBP. A paired-samples t-test was conducted to evaluate the mean differences between each of digital measure of blood pressure and pulse rates with a mercury sphygmomanometer. Significant mean differences were found in BP and pulse between electronic devices and manual readings in all five groups (device A to device E) except pulse rate on device B (p=0.05), SBP on device D (p=0.45) and DBP in device E (p=0.89). The mean BP and pulse rates were higher in digital devices except diastolic BP in device D. Discussion: Blood pressure is an important determinant of cardiovascular outcome in diabetes. In order to ensure that these new devices provide accurate data, they should be calibrated and validated periodically. Calibration ensures that measurements start from zero on all occasions. Clinicians should consider device accuracy when selecting the method for BP measurement in situations where readings are likely to influence medical management, such as when BP readings will be used to adjust cardiac medications for inpatients or in the out patients clinics. Routine use of electronic BP devices in hospitalized patients, when accuracy of the BP measurement is often critical for appropriate disease management, should be reconsidered and any abnormal values have to be reconfirmed with a mercury sphygmomanometer.

2 1. INTRODUCTION There is no visible way one can understand changes in human blood pressure, and naturally, checking regularly is the only means to keep track of it. The blood pressure results can predict the risks for body and health, thereby enabling physician to suggest appropriate medications, food and work schedules for the patients. Hypertension is an important worldwide public health problem and is a major risk factor for cardiovascular disease, the leading cause of death and disability worldwide [1, 2]. Screening for hypertension is the most commonly performed screening test in medical practice and is important not least because individuals with a normal blood pressure (BP) at the age of 50 still run a 90% life-time risk of developing hypertension [3]. Twentytwo percent of persons who have hypertension are unaware that they have it [3]. In addition, one-fourth of the population 20 years and older is estimated to be pre-hypertensive, which is defined as a systolic blood pressure of mm Hg, and/or a diastolic blood pressure of mm Hg. Pre-hypertension contributes to about 15% of blood pressure related deaths from coronary artery disease [4]. Hypertension and prehypertension in children and adolescents 8 to 17 years old have increased since 1999, and both are frequently undiagnosed in children 3 to 18 years of age [5, 6]. Childhood blood pressure is a strong predictor of High blood pressure levels in adulthood, and hypertension and prehypertension are a significant health issue in the young, especially with the prevalence of childhood obesity [7, 8]. Therefore it is very important for both young and older individuals with hypertension and prehypertension to be identified and treated [8-10]. Furthermore, trial evidence indicates that even small differences in systolic BP of 2 4mmHg are clinically important, thus accurate measurement is vital [11, 12]. Different blood pressure monitors provide different readings, errors in measurement occur through three main sources: device errors, user errors and patient errors [13]. Device errors are potentially the most fundamental of these three sources, and, in particular, accurate measurement is underpinned by the requirement for an instrument that measures pressure accurately in patients across the range of pressures. To date, the gold standard instrument for BP measurement is a mercury to gauge the blood pressure with a hand pump attached to the arm of the patient (mercury sphygmomanometer) [4, 14-16], but in recent years this has been superseded in many clinical settings by automated electronic devices. Medical engineers, on the hunt for more convenience, took the principles of digital monitors and adapted them to make a device for the arm and wrist. But, increased usage of these digital devices by ordinary patients will fulfill the requirement of BP measures accurately? Therefore, it is important to understand the accuracy of different digital BP devices used by the public. 1.1 Objective Our objective was to determine the accuracy of different types of digital (electronic) BP and pulse measuring devices. 2. METHODOLOGY Data were prospectively collected for three consecutive months by four trained nurses from five different groups of consented patients (group A to group E ); they were presented to private general clinic at Royal Care Hospital Akurana, Sri Lanka. For each group, 40 subjects were selected by simple random sampling method. The age range was years. Every patient was asked to randomly select one rolled paper from randomization box to allocate the digital device to be used (the randomization box was filled with 200 rolled papers and each of five devices (A to E) name was written in forty rolled paper) In addition to demographical data from each group of participants; resting position BP and pulse rate were measured three times (to strengthen the accuracy of the blood pressure measurements) using one of digital wrist or upper arm digital BP apparatus, the average was calculated [4, 17, 18]. And the same procedure was repeated with mercury sphygmomanometer for BP (as a standard one) and pulse rates were measured manually from every respondent. Body weight, height, hip circumferences and chronic co-morbid illnesses also were recorded from each of the patients.

3 Data were descriptively analyzed and the comparisons between each digital apparatus and mercury sphygmomanometer were done by Kappa measurement of agreement method and paired t-test, using SPSS 18. The level of significant was set as p < RESULTS There were total of 200 patients studied in five groups and each group consisted of 40 patients. The median age in each group was laid between 37 and 48 year old and the most of the participants were females. Group D and E participants were obese compare to other groups and the highest median Body Mass Index (BMI) (27; IQR 24-29) was reported in group D. The median hip circumference was also greater than the normal on group D patients (96.5 cm; IQR ). As compare to sphygmomanometer readings, the median systolic and diastolic blood pressure readings were higher in all digital devices except group D. The median pulse rate readings were also showing higher in digital devices than manual readings (table 1). As shown in table 2, in group A and B, both systolic BP and diastolic BP, and group D systolic BP shown Kappa (KA) <0.14 with poor agreements between digital device and sphygmomanometer readings. However, these agreements were not statistically significant (p>0.05). The diastolic BP of devices C, D and E were showing significant fair, good and very good agreement with sphygmomanometer readings (KA=0.3, p=0.007; KA=0.64, p<0.001, and KA=1, p<0.001) respectively. The systolic BP of devices C and E were showing significant moderate agreement with sphygmomanometer. The pulse readings from digital devices of group B, C, D and E were showing significant moderate to good agreement with manual pulse readings (KA between 0.42 and 0.8 and P<0.05), whereas group A digital pulse reading showed significant fair agreement with manual reading (KA=0.31 and P=0.022). A paired-samples t-test was conducted to determine the mean differences between each of digital measure of blood pressure and pulse rates with manual methods, there were significant mean differences on BP and pulse rates between electronic devices and manual readings in almost all five groups (group A to group E ) except pulse rate on group B (p=0.053), group D (p=0.45) and diastolic BP in group E (p=0.89). The mean BP and pulse rates were higher in all digital devices except diastolic BP in group D (table 3). 4. DISCUSSION This study revealed that there were statistically significant difference on mean BP and pulse readings between each of the electronic devices and manual readings.the findings of this study are similar to prior studies of earlier models of the automated BP device in which statistically significant systolic differences were noted between the electronic and manual readings for BP determination [19-23]. All these studies findings indicate that despite technologic improvements in the automated device over the years, significant differences still exist in BP readings obtained with the two different methods. Different from the majority of method-comparison studies of electronic and manual devices which used one or two individuals for data collection. This study had BP and pulse rate data collected by three different clinical research assistants. This situation more closely approximates usual practice on inpatient nursing care units, where BP readings are obtained commonly by multiple providers during a 24-hour shift. The manufacturer s operating instructions highlight the concern that cardiac rhythm irregularity may influence the accuracy of the device, few clinical users of the device restrict its use to patients with regular cardiac rhythms. Given that if the electronic BP measurement is done commonly in patients with irregular rhythms, it is especially important to determine if this practice is leading to even greater differences in BP between the two methods. Further studies are needed to confirm these results in a variety of clinical situations, including patients with irregular cardiac rhythms, and patients with systolic BP less than 90 mmhg and greater than 170 mmhg. The widespread use of and reliance on automated devices should be brought into question for inpatient clinical use. Many nursing and medical interventions for hospitalized patients and outpatients clinics are influenced by BP values obtained by nursing staff (for example, administration of antihypertensive and vasoactive

4 medications; monitoring for changes in BP after surgery or procedures). Using a less accurate BP method may lead to inappropriate alterations in medication or medical care which could negatively impact patient outcomes. 5. CONCLUSION Different types of devices for BP and pulse measurement are used commonly in the out patients clinics and hospitals. We found a statistically significant differences in BP (systolic and diastolic) measurements taken with the electronic devices and manual methods. Similarly, significant differences in pulse rates were also seen with electronic devices and manual methods. Clinicians should consider device accuracy when selecting the method for BP measurement in situations where readings are likely to influence medical management, such as when BP readings will be used to adjust cardiac medications for inpatients or in the out patients clinics. Routine use of electronic BP devices in hospitalized patients, when accuracy of the BP measurement is often critical for appropriate disease management, should be reconsidered and any abnormal values have to be reconfirmed with a mercury sphygmomanometer. Acknowledgment Special thanks to Dr. Mahindaratne KMNK, Farzana MMF, Nursing staff of the Royal Care Hospital for their voluble contribution in data collection and the authors acknowledge all the respondents for their voluntary participation on this study. Corresponding Author: Mohamed Shukry Zawahir, School of Pharmacy, Management and Science University (MSU), Section 13, 40100, Shah Alam, Selangor, Malaysia. Telephone: Fax: shukry2010@gmail.com

5 References: Bern, L., et al., (2007). Differences in blood pressure values obtained with automated and manual methods in medical inpatients. Medsurg Nurs,. 16(6), Blacher, J., et al. (2000). Pulse pressure not mean pressure determines cardiovascular risk in older hypertensive patients. Arch Intern Med., 160(8), Braam, R.L. and T. Thien. (2005). Is the accuracy of blood pressure measuring devices underestimated at increasing blood pressure levels? Blood Press Monit,.,10(5), Chang, J.J., D. Rabinowitz, and S. Shea. (2003).Sources of variability in blood pressure measurement using the Dinamap PRO 100 automated oscillometric device. Am J Epidemiol, 158(12), Chen, X., et al., Impacts of measurement protocols on blood pressure tracking from childhood into adulthood: a metaregression analysis. Hypertension, (2008). 51(3): p Chobanian, A.V., et al.(2003). The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure: the JNC 7 report. JAMA, 289(19), Frese, E.M., A. Fick, and H.S. Sadowsky. (2011). Blood pressure measurement guidelines for physical therapists. Cardiopulm Phys Ther J., 22(2), Gillman, M.W. and N.R. (1995). Cook, Blood pressure measurement in childhood epidemiological studies. Circulation, 92(4), Julius, S., et al., (1992). White coat hypertension: a follow-up. Clin Exp Hypertens A., 14(1-2), McAlister, F.A. and S.E. Straus, (2001). Evidence based treatment of hypertension. Measurement of blood pressure: an evidence based review. BMJ, 322(7291), O'Brien, E., et al. (2003 ). European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. J Hypertens, 21(5), Ostchega, Y., et al. (2003). National Health and Nutrition Examination Survey : effect of observer training and protocol standardization on reducing blood pressure measurement error. J Clin Epidemiol., 56(8), Padwal, R., S.E. Straus, and F.A. McAlister. (2001). Evidence based management of hypertension. Cardiovascular risk factors and their effects on the decision to treat hypertension: evidence based review. BMJ, 322(7292), Padwal, R.S., et al., (2008). The 2008 Canadian Hypertension Education Program recommendations for the management of hypertension: Part 1 - blood pressure measurement, diagnosis and assessment of risk. Can J Cardiol, 24(6), Pater, C. (2005). Beyond the Evidence of the New Hypertension Guidelines. Blood pressure measurement - is it good enough for accurate diagnosis of hypertension? Time might be in, for a paradigm shift (I). Curr Control Trials Cardiovasc Med,. 6(1), 6.

6 Pickering, T.G., et al. (2005). Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Circulation, 111(5), Pickering, T.G. (2003). What will replace the mercury sphygmomanometer? Blood Press Monit, 8(1), Shahriari, M., et al., (2003)Measurement of arm blood pressure using different oscillometry manometers compared to auscultatory readings. Blood Press, 12(3), The fourth report on the diagnosis, evaluation, and treatment of high blood pressure in children and adolescents. (2004). Pediatrics. 114(2 Suppl 4th Report), Urbina, E.M., et al., (1995). Effect of body size, ponderosity, and blood pressure on left ventricular growth in children and young adults in the Bogalusa Heart Study. Circulation, 91(9), Van Ittersum, F.J., et al. (1998). Determinants of the limits of agreement between the sphygmomanometer and the SpaceLabs device for blood pressure measurement in health volunteers and insulin-dependent diabetic patients. J Hypertens, 16(8), Wingfield, D., G.K. Freeman, and C.J. (2002). Bulpitt, Selective recording in blood pressure readings may increase subsequent mortality. QJM, 95(9), Wolf-Maier, K., et al. (2003). Hypertension prevalence and blood pressure levels in 6 European countries, Canada, and the United States. JAMA, 289(18),

7 Table 1: Demographic characteristics and baseline median BP and Pulse rate from five different types of electronic BP apparatus and Sphygmomanometer Group-A Group-B Group-C Group-D Group-E Age; median (IQR) 46 ( ) 37 (27-52) 40 (28-53) 39 (30-49) 48 (38-59) Sex (Female); (N) % (25) 72% (23) 65% (26) 65% (26) 57% (29) 62% BMI (KgM -2 ); median (IQR) 23 (20-27) 23.5 ( ) 23.5 (21-26) 27 (24-29) 25 (22-28) Hip circumference (cm) median (IQR) 87 (71-94) 87 (77-97) 83 (71-93) 96.5 ( ) 94 ( ) Average systolic BP (mmhg) (electronic); median(iqr) 162 ( ) 149 ( ) 124 ( ) 119 ( ) 124 ( ) Average systolic BP (mmhg); median(iqr)(sphygmomanometer) 120 (98-130) 113 ( ) 120 ( ) 120 ( ) 113 ( ) Average diastolic BP (mmhg); median (IQR) (electronic) 99.5 (80-105) 87.5 ( ) 81 ( ) 72 ( ) 72.5 ( ) Average diastolic BP (mmhg); median (IQR) (manual) 77.0 ( ) 73 (70-80) 78 (64-80) 80 ( ) 71 ( ) Average pulse rate (beats/minute); median (IQR) (electronic) 80 (74-90) 82 (74-92) 84 (76-90) 85 (76-96) 85 (78-95) Average pulse rate (beats/minute); median (IQR) (manual) 77 (71-83) 80 (74-84) 80 (73-85) 81 (72-88) 79 (71-85)

8 Table 2: Measure of agreement Kappa of BP and Pulse rate between digital devices and mercury sphygmomanometer (N=40 in each group) Group A Pulse rate Kappa Agreement p- value Strength of agreement Fair Group B Pulse Rate <0.001 Good Group C Pulse Rate <0.001 Moderate Fair Good Group D Pulse Rate < Good Moderate Group E Pulse Rate < Moderate Very good Moderate Kappa; <0.2 poor, fair agree, moderate, good, Very good agree

9 Table 3: Mean differences of BP and Pulse rate between digital devices and mercury Mean (SD) Electronic Manual t-value (df) P value Group- A (35) (21.04) (39) <0.001 group-a (16.45) (13.97) 7.3 (39) <0.001 Pulse rate Group-A (14.52) (12.10) 2.97 (39) Group-B 151 (32.55) (18.60) 7,68 (39) <0.001 Group-B (19.76) (11.50) 5.14 (39) <0.001 Pulse rate Group-B (13.37) (9.74) 1.94 (39) Group-C (23.70) (20.71) 3.58 (39) Group-C (14.87) (12.64) 4.86 (39) <0.001 Pulse rate group-c (12.58) (10.97) 2.90 (39) Group-D (24.87) (19.31) 0.76 (39) Group-D (15.58) (12.67) (39) Pulse Electronic-D (12.15) (11.38) 3.58 (39) Group-E (16.60) (18.33) 5.58 (39) <0.001 Electronic-E (10.66) (9.55) (39) Pulse Electronic-E (14.35) (11.97) 3.68 (39) Mean differences from paired t-test, significant level 0.05 (N=40 in each group)

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room. Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room Abstract Ahmad Mirdamadi (1), Mostafa Etebari (2) Original Article BACKGROUND:

More information

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Original article http://dx.doi.org/10.3345/kjp.2011.54.11.463 Korean J Pediatr 2011;54(11):463-469 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

Reference Values and Simplified Methods for Interpretation of Blood Pressure in Children and Adolescents

Reference Values and Simplified Methods for Interpretation of Blood Pressure in Children and Adolescents Reference Values and Simplified Methods for Interpretation of Blood Pressure in Children and Adolescents Simonetti G.D. Istituto Pediatrico della Svizzera Italiana Ente Ospedaliero Cantonale e Università

More information

4/4/17 HYPERTENSION TARGETS: WHAT DO WE DO NOW? SET THE STAGE BP IN CLINICAL TRIALS?

4/4/17 HYPERTENSION TARGETS: WHAT DO WE DO NOW? SET THE STAGE BP IN CLINICAL TRIALS? HYPERTENSION TARGETS: WHAT DO WE DO NOW? MICHAEL LEFEVRE, MD, MSPH PROFESSOR AND VICE CHAIR DEPARTMENT OF FAMILY AND COMMUNITY MEDICINE UNIVERSITY OF MISSOURI 4/4/17 DISCLOSURE: MEMBER OF THE JNC 8 PANEL

More information

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM Faculty Disclosure I have no financial interest to disclose No off-label use of medications will be discussed FIFTH ANNUAL SYMPOSIUM Recognize changes between

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the A comparison of blood pressure measurement over a sleeved arm versus a bare arm Grace Ma MD, Norman Sabin MD, Martin Dawes MBBS MD @ See related article page 591 DOI:10.1503/cmaj.070975 Abstract Background:

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

PREVALENCE OF HYPERTENSION AND ITS ASSOCIATED RISK FACTORS AMONG SCHOOL AGE CHILDREN

PREVALENCE OF HYPERTENSION AND ITS ASSOCIATED RISK FACTORS AMONG SCHOOL AGE CHILDREN PREVALENCE OF HYPERTENSION AND ITS ASSOCIATED RISK FACTORS AMONG SCHOOL AGE CHILDREN DR.AZAD ABDUL JABAR HALEEM AL-MEZORI- Dohuk Medical school/ University of Duhok-Kurdistan- Iraq Prof. Dr. AbbasAl-Rabaty-HawlerMedical

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon (2009) 23, 794 800 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Unreliable oscillometric blood pressure measurement: prevalence, repeatability

More information

White coat and masked hypertension

White coat and masked hypertension White coat and masked hypertension Conflict of interest Support from Spacelabs, Microlife. Honoraria from Novartis, Elpen, Boeringer-Ingelheim, CANA, Lilly, MSD, Sanofi, Menarini, Ciezi, Astra-Zeneca.

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines TITLE: Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines DATE: 25 September 2012 CONTEXT AND POLICY ISSUES A mercury manual sphygmomanometer

More information

Prevalence of Hypertension in Semi-Urban area of Nepal

Prevalence of Hypertension in Semi-Urban area of Nepal ORIGINAL ARTICLE Prevalence of Hypertension in Semi-Urban area of Nepal Koju R*, Manandhar K*, Gurung R*, Pant P*, Bedi TRS* *Department of Internal Medicine, Dhulikhel Hospital KUH ABSTRACT Hypertension

More information

ARIC Manuscript Proposal # PC Reviewed: 2/10/09 Status: A Priority: 2 SC Reviewed: Status: Priority:

ARIC Manuscript Proposal # PC Reviewed: 2/10/09 Status: A Priority: 2 SC Reviewed: Status: Priority: ARIC Manuscript Proposal # 1475 PC Reviewed: 2/10/09 Status: A Priority: 2 SC Reviewed: Status: Priority: 1.a. Full Title: Hypertension, left ventricular hypertrophy, and risk of incident hospitalized

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

In clinical practice, blood pressure (BP) treatment

In clinical practice, blood pressure (BP) treatment Original Paper Does Hidden Undercuffing Occur Among Obese Patients? Effect of Arm Sizes and Other Predictors of the Difference Between Wrist and Upper Arm Blood Pressures Hardik Doshi, MD; Alan B. Weder,

More information

THE ACCURATE ASSESSMENT

THE ACCURATE ASSESSMENT ORIGINAL INVESTIGATION Consistency of Blood Pressure Differences Between the Left and Right s Kazuo Eguchi, MD; Mona Yacoub, MD; Juhee Jhalani; William Gerin, PhD; Joseph E. Schwartz, PhD; Thomas G. Pickering,

More information

Prevalence, awareness, treatment and control of hypertension in North America, North Africa and Asia

Prevalence, awareness, treatment and control of hypertension in North America, North Africa and Asia (2004) 18, 545 551 & 2004 Nature Publishing Group All rights reserved 0950-9240/04 $30.00 www.nature.com/jhh REVIEW ARTICLE Prevalence, awareness, treatment and control of hypertension in North America,

More information

THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM

THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM Cader FA, Das A, Kahhar MA Dhaka Medical College Hospital March 2015 Dr Aaysha Cader MBBS,

More information

Protocol. Automated Ambulatory Blood Pressure Monitoring for the Diagnosis of Hypertension in Patients with Elevated Office Blood Pressure

Protocol. Automated Ambulatory Blood Pressure Monitoring for the Diagnosis of Hypertension in Patients with Elevated Office Blood Pressure Automated Ambulatory Blood Monitoring for the Diagnosis of Hypertension in Patients with Elevated Office Blood (10102) Medical Benefit Effective Date: 07/01/14 Next Review Date: 03/15 Preauthorization

More information

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Anthony J. Viera, MD, MPH, FAHA Department of Family Medicine Hypertension Research Program UNC School of Medicine Objectives Review limitations

More information

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods Vol. 2, Issue 1, pages 31-36 ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY By Alejandro de la Sierra, MD Luis M. Ruilope, MD Hypertension Units, Hospital Clinico, Barcelona & Hospital 12 de Octubre,

More information

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION*

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Progress in Clinical Medicine 1 ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Keishi ABE** Asian Med. J. 44(2): 83 90, 2001 Abstract: J-MUBA was a large-scale clinical

More information

A pre experimental study to assess the effect of administration of pomegranate juice among prehypertensive men and women aged between 25-45

A pre experimental study to assess the effect of administration of pomegranate juice among prehypertensive men and women aged between 25-45 2015; 1(11): 597-601 ISSN Print: 2394-7500 ISSN Online: 2394-5869 Impact Factor: 5.2 IJAR 2015; 1(11): 597-601 www.allresearchjournal.com Received: 04-08-2015 Accepted: 06-09-2015 Anoop Masih Sandhu Assistant

More information

CARDIOVASCULAR RISK FACTORS & TARGET ORGAN DAMAGE IN GREEK HYPERTENSIVES

CARDIOVASCULAR RISK FACTORS & TARGET ORGAN DAMAGE IN GREEK HYPERTENSIVES CARDIOVASCULAR RISK FACTORS & TARGET ORGAN DAMAGE IN GREEK HYPERTENSIVES C. Liakos, 1 G. Vyssoulis, 1 E. Karpanou, 2 S-M. Kyvelou, 1 V. Tzamou, 1 A. Michaelides, 1 A. Triantafyllou, 1 P. Spanos, 1 C. Stefanadis

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

Echocardiographic definition of left ventricular hypertrophy in the hypertensive: which method of indexation of left ventricular mass?

Echocardiographic definition of left ventricular hypertrophy in the hypertensive: which method of indexation of left ventricular mass? Journal of Human Hypertension (1999) 13, 505 509 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE Echocardiographic definition of left ventricular

More information

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License Zheng DC, Liu CY, Amoore J, Mieke S, Murray A. Need for Re-validation of Automated Blood Pressure Devices for use in Unstable Conditions. In: 215 Computing in Cardiology Conference (CinC). 215, Nice, France:

More information

Director of the Israeli Institute for Quality in Medicine Israeli Medical Association July 1st, 2016

Director of the Israeli Institute for Quality in Medicine Israeli Medical Association July 1st, 2016 The differential effect of Atherosclerosis on end organ damage in adult and elderly patients with CVRF: New Algorithm for Hypertension Diagnosis and Treatment R. Zimlichman, FAHA, FASH, FESC, FESH Chief

More information

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice.

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice. AJH 26; 19:81 89 BP Measurement Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice Tony C. Y. Pang and Mark A. Brown Background: To determine the extent of discrepancies between

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Automatic oscillometric blood pressure devices are frequently

Automatic oscillometric blood pressure devices are frequently Arterial Stiffness as Underlying Mechanism of Disagreement Between an Oscillometric Blood Pressure Monitor and a Sphygmomanometer Nicole M. van Popele, Willem Jan W. Bos, Nicole A.M. de Beer, Deirdre A.M.

More information

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA)

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) THE NEW ARMENIAN MEDICAL JOURNAL Vol.5 (2011), Nо 2, p.29-34 DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA) Zelveian P.H. 1, 2, Podosyan G.A. 2 1 Institute

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI.

DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI. DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI. Franco Cipollini, Carlo Porta, Enrica Arcangeli, Carla Breschi, & Giuseppe Seghieri Azienda USL 3, Ambulatorio

More information

greater than 90% of patients having blood pressure measured consistently according to the protocol.

greater than 90% of patients having blood pressure measured consistently according to the protocol. Clinical methods and pathophysiology 1 Reliability and validity of blood pressure measurement in the Secondary Prevention of Small Subcortical Strokes study Pablo E. Pérgola a, Carole L. White b, John

More information

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04) SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: MEASURING BLOOD PRESSURE - MANUAL Nursing DATE: REVIEWED: PAGES: 2/80 7/17 1 of 5 RESPONSIBILITY: RN, LPN, Patient Care Technician Multi-skilled

More information

Importance of Ambulatory Blood Pressure Monitoring in Adolescents

Importance of Ambulatory Blood Pressure Monitoring in Adolescents Importance of Ambulatory Blood Pressure Monitoring in Adolescents Josep Redon, MD, PhD, FAHA Internal Medicine Hospital Clinico Universitario de Valencia University of Valencia CIBERObn Instituto de Salud

More information

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License Pan F, Zheng DC, He PY, Murray A. Does the Position or Contact Pressure of the Stethoscope Make Any Difference to Clinical Blood Pressure Measurements: An Observational Study. Medicine 214, 93(29). Copyright:

More information

Egyptian Hypertension Guidelines

Egyptian Hypertension Guidelines Egyptian Hypertension Guidelines 2014 Egyptian Hypertension Guidelines Dalia R. ElRemissy, MD Lecturer of Cardiovascular Medicine Cairo University Why Egyptian Guidelines? Guidelines developed for rich

More information

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009 42 Original article Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP, the Omron HBP T and the Pic Indolor Professional Roland

More information

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice (2005) 19, 801 807 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Prognostic significance of blood pressure measured in the office, at home and

More information

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? O R I G I N A L A R T I C L E Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? I. Aksoy, J. Deinum, J.W.M. Lenders, Th. Thien *# Department of

More information

Prevalence, awareness, treatment and control of hypertension in employees of factories of Northern Greece: the Naoussa study

Prevalence, awareness, treatment and control of hypertension in employees of factories of Northern Greece: the Naoussa study (2004) 18, 623 629 & 2004 Nature Publishing Group All rights reserved 0950-9240/04 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Prevalence, awareness, treatment and control of hypertension in employees of

More information

Risk Assessment of developing type 2 diabetes mellitus in patient on antihypertensive medication

Risk Assessment of developing type 2 diabetes mellitus in patient on antihypertensive medication 41 Research Article Risk Assessment of developing type 2 diabetes mellitus in patient on antihypertensive medication Amarjeet Singh*, Sudeep bhardwaj, Ashutosh aggarwal Department of Pharmacology, Seth

More information

There is convincing evidence in clinical studies

There is convincing evidence in clinical studies AJH 1998;11:1413 1417 Reliability of Reporting Self-Measured Blood Pressure Values by Hypertensive Patients Thomas Mengden, Rosa Maria Hernandez Medina, Belen Beltran, Elena Alvarez, Karin Kraft, and Hans

More information

Dementia Retinal arteriolar narrowing and HT. What comes first?

Dementia Retinal arteriolar narrowing and HT. What comes first? Dementia 2016 Retinal arteriolar narrowing and HT What comes first? Jan A. Staessen, MD, PhD University of Leuven / Maastricht University jan.staessen@med.kuleuven.be / ja.staessen@maastrichtuniversity.nl

More information

NIH Public Access Author Manuscript JAMA Intern Med. Author manuscript; available in PMC 2015 August 01.

NIH Public Access Author Manuscript JAMA Intern Med. Author manuscript; available in PMC 2015 August 01. NIH Public Access Author Manuscript Published in final edited form as: JAMA Intern Med. 2014 August ; 174(8): 1397 1400. doi:10.1001/jamainternmed.2014.2492. Prevalence and Characteristics of Systolic

More information

Hypertension with Comorbidities Treatment of Metabolic Risk Factors in Children and Adolescents

Hypertension with Comorbidities Treatment of Metabolic Risk Factors in Children and Adolescents Hypertension with Comorbidities Treatment of Metabolic Risk Factors in Children and Adolescents Stella Stabouli Ass. Professor Pediatrics 1 st Department of Pediatrics Hippocratio Hospital Evaluation of

More information

Prevalence, awareness of hypertension in rural areas of Kurnool

Prevalence, awareness of hypertension in rural areas of Kurnool Original article: Prevalence, awareness of hypertension in rural areas of Kurnool Dr. Sudhakar Babu*, Dr.Aruna MS** *Associate Professor, Dept of Community Medicine, Vishwa Bharathi Medical College Kurnool,

More information

Dr Doris M. W Kinuthia

Dr Doris M. W Kinuthia Dr Doris M. W Kinuthia Objectives Normal blood pressures in children Measurement of blood pressure in children Aetiology of Hypertension in children Evaluation of children with hypertension Treatment of

More information

Please do not hesitate to contact our dedicated service unit for more details regarding this program at

Please do not hesitate to contact our dedicated service unit for more details regarding this program at The Blue Cross and Blue Shield Federal Employee Program (FEP) and the American Medical Association (AMA) have come together in a collaborative effort to provide physicians with resources designed to improve

More information

Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements

Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements Original Article J Clin Med Res. 2018;10(4):309-313 Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements Kannayiram Alagiakrishnan a, d, Ruojin

More information

Prevalence of Pre Hypertension Among the Women Aged Years in Coastal and Non Coastal Areas

Prevalence of Pre Hypertension Among the Women Aged Years in Coastal and Non Coastal Areas 2016 IJSRST Volume 2 Issue 5 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Prevalence of Pre Hypertension Among the Women Aged 20-60 Years in Coastal and Non Coastal

More information

How well do office and exercise blood pressures predict sustained hypertension? A Dundee Step Test Study

How well do office and exercise blood pressures predict sustained hypertension? A Dundee Step Test Study (2000) 14, 429 433 2000 Macmillan Publishers Ltd All rights reserved 0950-9240/00 $15.00 www.nature.com/jhh ORIGINAL ARTICLE How well do office and exercise blood pressures predict sustained hypertension?

More information

University of Padova, Padua, Italy, and HARVEST Study Group, Italy

University of Padova, Padua, Italy, and HARVEST Study Group, Italy University of Padova, Padua, Italy, and HARVEST Study Group, Italy ISOLATED SYSTOLIC HYPERTENSION IN THE YOUNG DOES NOT IMPLY AN INCREASED RISK OF FUTURE HYPERTENSION NEEDING TREATMENT Mos L, Saladini

More information

Slide notes: References:

Slide notes: References: 1 2 3 Cut-off values for the definition of hypertension are systolic blood pressure (SBP) 135 and/or diastolic blood pressure (DBP) 85 mmhg for home blood pressure monitoring (HBPM) and daytime ambulatory

More information

Stopping oral contraceptives: an effective blood pressure-lowering intervention in women with hypertension

Stopping oral contraceptives: an effective blood pressure-lowering intervention in women with hypertension (2005) 19, 451 455 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Stopping oral contraceptives: an effective blood pressure-lowering intervention

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

Validation Study Result Form

Validation Study Result Form Validation Study Result Form Please complete Section 1 to Section 3 of this form and return it to dabl Educational Trust with copies of the validation plots. The requirements for each of these sections

More information

Adolescent Hypertension Roles of obesity and hyperuricemia. Daniel Landau, MD Pediatrics, Soroka University Medical Center

Adolescent Hypertension Roles of obesity and hyperuricemia. Daniel Landau, MD Pediatrics, Soroka University Medical Center Adolescent Hypertension Roles of obesity and hyperuricemia Daniel Landau, MD Pediatrics, Soroka University Medical Center Blood Pressure Tables BP standards based on sex, age, and height provide a precise

More information

PREVALENCE OF METABOLİC SYNDROME İN CHİLDREN AND ADOLESCENTS

PREVALENCE OF METABOLİC SYNDROME İN CHİLDREN AND ADOLESCENTS PREVALENCE OF METABOLİC SYNDROME İN CHİLDREN AND ADOLESCENTS Mehmet Emre Atabek,MD,PhD Necmettin Erbakan University Faculty of Medicine, Department of Pediatrics, Division of Pediatric Endocrinology and

More information

doi: /fampra/cmw053

doi: /fampra/cmw053 doi: 10.1093/fampra/cmw053 Manuscript type Original research Title Comparison of blood pressure measurements on the bare arm, over a sleeve, and over a rolled-up sleeve in the elderly Short title Blood

More information

MEDICAL POLICY SUBJECT: AUTOMATED AMBULATORY BLOOD PRESSURE MONITORING

MEDICAL POLICY SUBJECT: AUTOMATED AMBULATORY BLOOD PRESSURE MONITORING MEDICAL POLICY SUBJECT: AUTOMATED AMBULATORY 02/19/09 PAGE: 1 OF: 5 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product, including

More information

Comparison of arbitrary definitions of circadian time periods with those determined by wrist actigraphy in analysis of ABPM data

Comparison of arbitrary definitions of circadian time periods with those determined by wrist actigraphy in analysis of ABPM data Journal of Human Hypertension (1999) 13, 449 453 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE Comparison of arbitrary definitions of

More information

Normal blood pressure values as perceived by normotensive and hypertensive subjects

Normal blood pressure values as perceived by normotensive and hypertensive subjects ORIGINAL ARTICLE Normal blood pressure values as perceived by normotensive and hypertensive subjects (2003) 17, 87 91 & 2003 Nature Publishing Group All rights reserved 0950-9240/03 $25.00 www.nature.com/jhh

More information

OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension. Ontario Health Technology Advisory Committee

OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension. Ontario Health Technology Advisory Committee OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension Ontario Health Technology Advisory Committee May 2012 Background Hypertension in Canada Hypertension occurs when

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

Depok-Indonesia STEPS Survey 2003

Depok-Indonesia STEPS Survey 2003 The STEPS survey of chronic disease risk factors in Indonesia/Depok was carried out from February 2003 to March 2003. Indonesia/Depok carried out Step 1, Step 2 and Step 3. Socio demographic and behavioural

More information

Assessment of techniques for measurement of blood pressure in infants and children

Assessment of techniques for measurement of blood pressure in infants and children Archives of Disease in Childhood, 1973, 48, 932. Assessment of techniques for measurement of blood pressure in infants and children A. M. ELSEED, E. A. SHINEBOURNE, and M. C. JOSEPH From the Paediatric

More information

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation (2009), 1 5 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Hypertension Diagnosis and Treatment

Hypertension Diagnosis and Treatment Health Care Guideline Hypertension Diagnosis and Treatment The November 2014, Fifteenth Edition of ICSI s Hypertension Diagnosis and Treatment Health Care Guideline incorporates a revision of our previous

More information

A CROSS SECTIONAL STUDY OF RELATIONSHIP OF OBESITY INDICES WITH BLOOD PRESSURE AND BLOOD GLUCOSE LEVEL IN YOUNG ADULT MEDICAL STUDENTS

A CROSS SECTIONAL STUDY OF RELATIONSHIP OF OBESITY INDICES WITH BLOOD PRESSURE AND BLOOD GLUCOSE LEVEL IN YOUNG ADULT MEDICAL STUDENTS Original Article A CROSS SECTIONAL STUDY OF RELATIONSHIP OF OBESITY INDICES WITH BLOOD PRESSURE AND BLOOD GLUCOSE LEVEL IN YOUNG ADULT MEDICAL STUDENTS Renu Lohitashwa, Parwati Patil ABSTRACT Overweight

More information

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL Douglas G. Kelling Jr., MD Carmella Gismondi-Eagan, MD, FACP George C. Monroe, III, MD Revised, April 8, 2012 The information contained in this protocol

More information

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right Anthony J. Viera, MD, MPH, FAHA Professor and Chair Objectives Review limitations of office BP in making a correct diagnosis of hypertension

More information

Awareness of Hypertension, Risk Factors and Complications among Attendants of a Primary Health Care Center In Jeddah, Saudi Arabia

Awareness of Hypertension, Risk Factors and Complications among Attendants of a Primary Health Care Center In Jeddah, Saudi Arabia IOSR Journal of Nursing and Health Science (IOSR-JNHS e-issn: 2320 1959.p- ISSN: 2320 1940 Volume 6, Issue 1 Ver. VIII (Jan. - Feb. 2017), PP 16-21 www.iosrjournals.org Awareness of Hypertension, Risk

More information

Hypertension: What s new since JNC 7. Harold M. Szerlip, MD, FACP, FCCP, FASN, FNKF

Hypertension: What s new since JNC 7. Harold M. Szerlip, MD, FACP, FCCP, FASN, FNKF Hypertension: What s new since JNC 7 Harold M. Szerlip, MD, FACP, FCCP, FASN, FNKF Disclosures Spectral Diagnostics Site investigator Eli Lilly Site investigator ACP IM ITE writing committee NBME Step

More information

The magnitude and duration of ambulatory blood pressure reduction following acute exercise

The magnitude and duration of ambulatory blood pressure reduction following acute exercise Journal of Human Hypertension (1999) 13, 361 366 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE The magnitude and duration of ambulatory

More information

Association between arterial stiffness and cardiovascular risk factors in a pediatric population

Association between arterial stiffness and cardiovascular risk factors in a pediatric population + Association between arterial stiffness and cardiovascular risk factors in a pediatric population Maria Perticone Department of Experimental and Clinical Medicine University Magna Graecia of Catanzaro

More information

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure?

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure? International Journal of Nursing December 2018, Vol. 5,. 2, pp. 94-99 ISSN 2373-7662 (Print) 2373-7670 (Online) Copyright The Author(s). All Rights Reserved. Published by American Research Institute for

More information

Manual or Automated Sphygmomanometer? A Historical Cohort to Quantify Measurement Bias in Blood Pressure Recording

Manual or Automated Sphygmomanometer? A Historical Cohort to Quantify Measurement Bias in Blood Pressure Recording ORIGINAL PAPER Manual or Automated Sphygmomanometer? A Historical Cohort to Quantify Measurement Bias in Blood Pressure Recording Arash A. Nargesi, MD, MPH; 1,2 Zaniar Ghazizadeh, MD; 1 Mehrdad Larry,

More information

T. Suithichaiyakul Cardiomed Chula

T. Suithichaiyakul Cardiomed Chula T. Suithichaiyakul Cardiomed Chula The cardiovascular (CV) continuum: role of risk factors Endothelial Dysfunction Atherosclerosis and left ventricular hypertrophy Myocardial infarction & stroke Endothelial

More information

Prognostic significance of blood pressure measured on rising

Prognostic significance of blood pressure measured on rising (2001) 15, 413 417 2001 Nature Publishing Group All rights reserved 0950-9240/01 $15.00 www.nature.com/jhh ORIGINAL ARTICLE Prognostic significance of blood pressure measured on rising P Gosse, C Cipriano,

More information

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine Techniques of Vital Signs John Gazewood, MD, MSPH Department of Family Medicine Objectives Define normal in several ways. Describe correct technique of measuring vital signs. Describe correct technique

More information

The Hypertension Clinic is a part of the Internal Medicine

The Hypertension Clinic is a part of the Internal Medicine Original Article Hypertension Registry at the Bangkok Hospital Medical Center: The First 7 Months Experience OBJECTIVE: The Hypertension Registry at the Bangkok Hospital Medical Center was established

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information