1 Continuous Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio Xiao-Rong Ding, Student Member, IEEE, Yuan-Ting Zhang, Fellow, IEEE, Jing Liu, Wen-Xuan Dai, Hon Ki Tsang*, Senior Member, IEEE Abstract Pulse transit time (PTT) has attracted much interest for cuffless blood pressure (BP) measurement. However, its limited accuracy is one of the main problems preventing its widespread acceptance. Arterial BP oscillates mainly at high frequency (HF) because of respiratory activity, and at low frequency (LF) because of vasomotor tone. Prior studies suggested that PTT can track BP variation in HF range, but was inadequate to follow the LF variation, which is probably the main reason for its unsatisfactory accuracy. This paper presents a new indicator, the photoplethysmogram intensity ratio () which can be affected by changes in the arterial diameter and thus trace the LF variation of BP. Spectral analysis of BP, PTT, and respiratory signal confirmed that PTT was related to BP in HF at the respiratory frequency, while was associated with BP in LF range. We therefore develop a novel BP estimation algorithm by using both PTT and. The proposed algorithm was validated on 7 healthy subjects with continuous Finapres BP as reference. The results showed that the mean ± standard deviation (SD) for the estimated systolic, diastolic and mean BP with the proposed method against reference were -.37±5.1 mmhg, -.8±4.6 mmhg, -.18±4.13 mmhg, and mean absolute difference (MAD) were 4.9 mmhg, 3.18 mmhg, 3.18 mmhg, respectively. Furthermore, the proposed method outperformed the two most cited PTT algorithms for about mmhg in SD and MAD. These results demonstrated that the proposed BP model using and PTT can estimate continuous BP with improved accuracy. Index Terms Arterial diameter change, cuffless blood pressure, photoplethysmogram intensity ratio, pulse transit time, respiration, vasomotion This work was supported in part by the Guangdong Innovation Research Team Fund for Low-cost Healthcare Technologies in China, the External Cooperation Program of the Chinese Academy of Sciences (Grant GJHZ11). Asterisk indicates corresponding author. Xiao-Rong Ding, Yuan-Ting Zhang, Jing Liu, and Wen-Xuan Dai is with the Joint Research Centre for Biomedical Engineering, Department of Electronic Engineering, The Chinese University of Hong Kong, Hong Kong SAR, China ( xrding@ee.cuhk.edu.hk; ytzhangapple@icloud.com; jingliu@ee.cuhk.edu.hk; wxdai@ee.cuhk.edu.hk) *Hon Ki Tsang is with the Department of Electronic Engineering, The Chinese University of Hong Kong, Hong Kong SAR, China ( hktsang@ee.cuhk.edu.hk). Copyright (c) 15 IEEE. Personal use of this material is permitted. However, permission to use this material for any other purposes must be obtained from the IEEE by sending an to pubs-permissions@ieee.org. B I. INTRODUCTION LOOD pressure (BP) is an important hemodynamic parameter varying between systolic BP (SBP) to diastolic BP (DBP) in each heartbeat. High BP, also known as hypertension, is one of the major modifiable risk factors leading to the development of cardiovascular diseases (CVDs) the number one killer in the world. Hypertension is highly prevalent but poorly controlled because of the low awareness and treatment rate [1], which enhances development of CVDs and results in significant burdens on individuals and society. BP variability (BPV) has been reported to have prognostic value for hypertension [], and thus continuous BP measurement is crucial for early prevention, detection, evaluation and treatment of hypertension and related CVDs. Conventional 4-hour ambulatory BP monitoring can facilitate monitoring of BPV through measuring BP at regular intervals with auscultatory or oscillometric approaches. However, it has limitations including the discontinuous nature and the discomfort caused by the repeated cuff inflations. Compared with cuff-based BP techniques, pulse transit time (PTT) method has received much attention over the recent decades because of its capability to track BP change, as well as its advantages as a noninvasive, continuous and most importantly cuffless tool for BP measurement [3-5]. PTT is the time taken by the arterial pulse propagating from the heart to a peripheral site, and can be calculated as the time interval between the R wave peak of electrocardiogram (ECG) and a characteristic point of photoplethysmogram (PPG). The fundamental principle of PTT-based method is based upon the pulse wave velocity (PWV) recording through the Moens-Korteweg (M-K) equation: Eh PWV (1) d which relates PWV with the elastic modulus of vessel wall E, blood density ρ and arterial dimension properties such as vessel thickness h and arterial diameter d. PWV is inversely related with PTT, i.e., PWV=K/PTT, where K is the distance between heart and certain peripheral site; and E can be exponentially correlated BP through the following equation [6]:

2 E E e () P where E is the elastic modulus at zero pressure; γ is a coefficient depending on particular vessel, and is BP. Therefore, PTT can be translated into BP with an initial calibration under the assumption that h/d keeps constant. PTT-based BP estimation has been extensively studied ever since [7-], when Chen et al [7] estimated SBP using pulse arrival time with intermittent calibration, and showed that the estimated SBP was highly correlated with reference SBP (r=.97±.). In 5, Poon et al [9] established a model with PTT to estimate BP with initial calibration, and achieved an accuracy of.6±9.8 mmhg and.9±5.6 mmhg for SBP and DBP, respectively. Recently, Wibmer et al [19] investigated the relationship of PTT and SBP through regression analysis and found a nonlinear approach was better than linear one. Although PTT has been considered a promising surrogate of BP and could become the most widely used technique for noninvasive continuous BP monitoring in the future [4, 5], there are still several problems to be solved before its widespread application. Frist, some PTT-BP models could only provide one BP parameter, e.g., exclusively SBP [7, 8, 1, ], DBP [11], or mean BP (MBP) [18], but SBP, DBP, and MBP all have clinical significance. Second, a calibration procedure is required to map PTT to BP. However, re-calibration at intermittent intervals is often necessary for accurate estimation, potentially owing to the inadequacy of PTT to track BP variation over a long period. Last and most importantly, the accuracy of PTT-based BP estimation is unsatisfactory. The possible reasons are the influences of the vascular or vasomotor tone and the pre-ejection period (PEP) [3]. Regarding PEP issue, impedance cardiogram (ICG), phonocardiogram (PCG) [], ballistocardiogram (BCG) or two peripheral PPG have been adopted to eliminate the effect of PEP. Nonetheless, several studies indicated that PTT with PEP included actually performs better for BP estimation than that with PEP removed, which demonstrates the positive effect of PEP on BP estimation [8,, 3]. For the vasomotor tone, previous research has examined its influence on BP-PTT relationship [4], and central PPG instead of peripheral PPG was suggested to alleviate such effect [18]. However, few studies have attempted to take this factor into account in the PTT-BP estimation model to improve the accuracy. BP is dynamic and its rhythmic oscillations can be identified with the appearance in its spectrum as individual peaks, which reflect: (1) the oscillations with a frequency typically between.-.35 Hz, a frequency similar to that of normal respiratory activity, defined as high frequency (HF); () oscillations with a frequency of approximately Hz, suggesting the sympathetic modulation of vasomotor tone, defined as low frequency (LF) [5-7]. According to our prior research work [8, 9], PTT could track BP in HF range, but was inadequate to follow LF variations in BP. This is probably the most important reason for the inaccuracy of estimated BP with only the PTT and the requirement of intermittent calibration to maintain accuracy. However, to date, few study has addressed the LF component in cuffless BP estimation. Here we propose a new indicator, the PPG intensity ratio () [3], that can reflect changes in the arterial diameter and thereby the arterial vasomotion, and thus allow tracking of BP in the LF range. Furthermore, we develop a novel BP estimation algorithm which employs both the and the PTT to improve the estimation accuracy. II. METHODOLOGY A. PTT and Fig. 1 illustrates the diagram of PTT and calculation, where PTT is determined as the time interval between the R wave peak of ECG and the peak of first derivative of PPG in the same cardiac cycle, and is the ratio of PPG peak intensity I H to PPG valley intensity I L of one cardiac cycle. Our earlier study shows that can theoretically reflect the arterial diameter change d during one cardiac cycle from systole to diastole, and is exponentially linked with d through the following expression [3]: d e (3) where α is considered to be a constant related to the optical absorption coefficients in the light path. Fig. 1. Diagram of pulse transit time (PTT) and photoplethysmogram (PPG) intensity ratio () calculation, where I H indicates PPG peak intensity, I L the valley intensity, and 1st dppg is the first derivative of PPG. From a physiological perspective, BP is mainly affected by four factors: arterial compliance, cardiac output, peripheral resistance, and blood volume [31]. Arterial compliance can be evaluated by PTT, since PTT is an index of arterial stiffness [3]. Also, cardiac output can be related to PTT through heart rate. With regard to peripheral resistance and blood volume, one of the primary sources is the arterial diameter change which can be assessed by as described above. Accordingly, PTT and can capture BP variations indirectly, and be used for BP estimation.

3 B. Model-based BP Estimation with PTT and Arterial BP is a hemodynamic parameter that fluctuates on a beat-to-beat basis as a result of the dynamic interplay involving vasomotion, arterial mechanisms and neural regulation [33]. Beat-to-beat BP fluctuations are usually attributed to two rhythmic events: respiration and vasomotion [34-36]. The respiratory rhythm is a HF spectral component that occurs in BP variability, and also considered to be a marker of vagal modulation, whereas the slow oscillation, corresponding to the vasomotor waves, is a LF component that is present in BPV, and is also a marker of sympathetic modulation. Physiological study of BPV showed that both the slow variability and fast variability can be observed in SBP, while the DBP exhibits only on the slow variability [37]. Since SBP is the summation of pulse pressure (PP) and DBP, it is hypothesized that the HF component is mainly dominant in PP. Furthermore, PTT and have been investigated to reflect HF and LF component in BPV [3], respectively. It is therefore speculated that PP and DBP can be derived from PTT and, respectively, and consequently the SBP can be estimated accurately. 1) PP Estimation with PTT Based on M-K Equation In the M-K equation (1), the elastic modulus E is given by [38]: P E R O 1 RO Ri (4) RO Ri where R O is the external radius, R O is the external radius change in response of the pressure change P, and P is the PP in the artery; Ri is the internal radius, and σ is the Poisson s ratio. PWV is reversely proportional to PTT: 1 PWV (5) PTT Since the artery radius change is quite small compared with elastic modulus change, it is assumed to be constant. According to (1) and (4)-(5), P has a relationship with PTT as follows: Fig.. The two-element Windkessel [39]. In a pure Windkessel, the DBP can be theoretically expressed in terms of RC using the following equation: DBP t / RC P e (8) where P is the end-systolic aortic pressure. Since C is constant in a relatively short period, DBP mainly varies with R. Noting that the major regulator of peripheral vascular resistance is the vessel diameter, R will mainly rely on the arterial diameter change d. As mentioned above in (3), d is related with as follows: 1 d (9) e And d is inversely related to R. Thus DBP depends on the reciprocal of : 1 DBP (1) Therefore, DBP can be derived with calibrated DBP and. 3) SBP Estimation with PP and DBP DBP DBP (11) SBP is the sum of PP and DBP. Thereupon, beat-to-beat SBP can be estimated with the addition of (7) and (11): PTT PP SBP DBP (1) PTT 1 PP (6) PTT With initial calibrated PP and PTT, PP can be derived in terms of measured PTT: PTT PP PP (7) PTT ) DBP Estimation with based on Windkessel Model Two-element Windkessel model, originally proposed by Frank, consists of peripheral resistance R and arterial compliance C [39], as shown in Fig.. C. Experiment To validate the proposed BP estimation using both PTT and as given by (11) and (1), an experiment was conducted on 7 healthy adults (14 males) with mean age of 5.6±.1 years (range 1-9 years), who were nonsmokers with no history of cardiovascular disease. Reference BP was measured by Finapres (Finapres Medical System), a noninvasive continuous BP measurement system, with the finger cuff on the left thumb, and brachial cuff on the left upper arm. ECG and PPG were acquired with one-lead ECG electrodes placed on left and right arms, and PPG sensor on left index finger, respectively. Synchronous respiratory activity continuously monitored by recording the chest movement with respiratory monitoring belt (Vernier Software & Technology). All tests were performed with subjects in the seated position, and the signals were

4 recorded at the sampling rate of 1 Hz for five minutes. All the subjects gave their informed consent prior to the experiments, in accordance with the guidelines of the Institutional Research Ethics Board. D. Signal Processing and Data Analysis In order to verify the capability of PTT to track the HF component of SBP as well as respiratory activity, and to reflect LF fluctuations of BP, the power spectrum analysis of SBP, DBP, PP, respiratory signal, PTT and were conducted in -.5 Hz frequency range based on Lomb-Scargle periodogram method [4]. Difference mean and standard deviation (SD), as well as mean absolute difference (MAD) between estimated BP with the proposed method and reference BP were used as the evaluation metrics. The agreement between reference BP and estimated BP with the proposed method were analyzed according to the Bland-Altman approach [41], with the agreement limits defined by mean ±1.96 SD. In addition, the proposed method was compared with two most cited PTT-based algorithms [7, 9] for cuffless BP estimation. Statistical significance was estimated using Student s t-test. P<.5 is regarded as statistically significant. Furthermore, time series of PP, PTT and respiratory signal with their corresponding spectrums are presented in Fig. 4. As can be observed from Fig. 4 (a)-(c), they have a similar variation pattern with almost the same frequency components at.-.3 Hz. Moreover, from the PSD curves, the respiration-synchronous variation is seen as peaks at the respiratory frequency of PP spectrum, as well as in PTT. This indicates that the HF component of PP can be reflected by PTT, which might be caused by the respiratory activity, as discussed in the next section. Example recordings of DBP and and their power spectra are shown in Fig. 5. The variation of DBP is slower compared with PP, with its frequency range concentrated at approximately.1 Hz. It can be seen from Fig. 5 (a) and (c) that the amplitude of DBP is inversely related with that of on the whole, and they have the similar spectral components. This is in line with preliminary study about which demonstrated could potentially evaluate the LF modulation of BP. But it is worth noting that there is also minor HF component appeared in spectrum. III. RESULTS A. Spectral Analysis of BP, Respiratory Signal, PTT and As can be seen in Fig. 3 (a), slow variability and fast variability can be observed in a typical continuous BP signal. Correspondingly, the variations of SBP, DBP and PP are shown in Fig. 3 (b). Obviously, SBP contained both slow variation and fast variation, whereas the DBP only showed slow variability, with PP presented the fast variability. Power spectral density (PSD) of SBP, DBP, and PP as illustrated in Fig. (c)-(e) can further describe this, where SBP showed LF variation centered at around.1 Hz, with HF variation dominated between.-.3 Hz; and the LF spectrum components were more pronounced in DBP, while PP mainly contained the HF variation. This is consistent with previous study about BP variation. Fig. 4. Time series and corresponding PSD of PP (a-b), PTT (c-d) and respiratory signal (e-f) of a representative subject. Fig. 3. Continuous BP signal (a); beat-to-beat SBP, DBP and PP (b) of a representative subject, with corresponding PSD of SBP (c), DBP (d) and PP (e). Fig. 5. Typical time series and corresponding PSD of DBP (a-b) and (c-d) of a representative subject.

5 B. Performance of Proposed BP Model with PTT and The estimation results of proposed BP model were compared with the continuous Finapres BP as the reference with regard to correlation and agreement. The correlation and the Bland-Altman plot of the SBP, DBP and MBP estimation for our proposed method versus Finapres BP are given in Fig. 6. The Person s correlation coefficient between overall estimated SBP, DBP, and MBP and that of Finapres is.91,.88, and.89, respectively. For the Bland-Altman plot, the x-axis of the plots presents the average of the estimation with the proposed algorithm and the Finapres, while the y-axis shows the difference between the two methods. The bias (mean) and the limits of agreement (bias±1.96 SD) are illustrated in red solid line and black dash dot lines, respectively. It is observed that the majority of the points lie within the limit of agreement, indicating that the estimated BP with the proposed method are in close agreement with those made by Finapres. The bias for SBP, DBP, and MBP estimates are -.37, -.8 and -.18 mmhg, respectively. Fig. 6. Correlation and Bland-Altman plots of SBP (a-b), DBP (c-d), and MBP (e-f) with the reference of Finapres BP. C. Comparison of Proposed Model against Current PTT Algorithms To further evaluate the efficiency of the proposed method, we compared the proposed method with two of the most cited PTT algorithms in the cuffless BP estimation area [7, 9]. The first PTT algorithm [7] estimates SBP in terms of relative PTT change through the following equation: SBP SBP PTT PTT PTT (13) And the second PTT algorithm [9] can obtain SBP and DBP based on the following equation set: PTT DBP MBP ln PTT 1 SBP DBP 3 PTT PTT (14a) PTT SBP DBP SBP DBP (14b) PTT Fig. 7 shows a representative example from the subjects of the beat-to-beat Finapres SBP, estimated SBP by proposed

6 algorithm with PTT and, compared to those by PTT algorithms, where SBP_PTT (1) represents the first PTT algorithm, and SBP_PTT () the second PTT algorithm. We can see that estimated SBP with proposed method tracked better with Finapres SBP compared with those of PTT methods. For this dataset with 66 beat-to-beat SBP, the beat by beat variation can be observed, and the average value measured by Finapres was ±4.87 mmhg, and the estimated values by proposed method, first and second PTT methods were ±4.8 mmhg, ±3.65 mmhg, and 1.76±7.73 mmhg, respectively. The corresponding errors were.59±3.1 mmhg, -1.83±4.6 mmhg and 5.56±5.9 mmhg, respectively, which further reveals that the SBP estimated with our method shows better correlation with Finapres SBP than the control PTT algorithms. Since the first PTT method only contains the algorithm for SBP estimation, DBP and MBP estimation were only compared with those of the second PTT method. It can be seen from Fig. 8 that DBP varied slower than SBP, where the average Finapres value was 73.5±.98 mmhg. The estimated DBP with proposed method and PTT method were 71.98±.3 mmhg and 79.38±4. mmhg, respectively, with the corresponding errors -1.7±.11 mmhg and 6.13±5.3 mmhg. Obviously, the estimation with the PTT method overestimated DBP not only in terms of the average level but the variability. Fig. 8. Estimated beat-to-beat DBP with proposed method (green) and PTT algorithm (dash blue) with the reference of Finapres DBP (red) (a) and the estimated errors (b) of one representative subject. Fig. 7. Estimated beat-to-beat SBP with proposed method (green) and PTT algorithms (solid blue and dash blue) with the reference of Finapres SBP (red) (a) and the estimated errors (b) of one representative subject. The variation of MBP is also slow which is similar to that of DBP, with average value of 87.3±3. mmhg, as depicted in Fig. 9. The estimated MBP with proposed method and the PTT method are 85.78±.3 mmhg and 93.18±5.6 mmhg, with corresponding estimation error of -1.45±.7 mmhg and 5.94±5.37 mmhg, respectively, suggesting that the estimation with the proposed method correlates better with Finapres MBP than that of the PTT method. Fig. 9. Estimated beat-to-beat MBP with proposed method (green) and PTT algorithm (dash blue) with the reference of Finapres MBP (red) (a) and the estimated errors (b) of one representative subject. Furthermore, the overall performance was analyzed in terms of difference mean, SD and MAD. Difference mean is a measure of the bias of BP estimates, while difference SD is a measure of error variability. MAD is a measure of overall accuracy in estimating BP. The smaller MAD, the better overall performance. Table I summarizes the values of mean, SD and MAD for our proposed BP estimation method and the PTT methods tested on 7 subjects, including 1713 heart beats. It is observed that the difference mean of our proposed method in estimating SBP, DBP, and MBP is within -.37, -.8, and -.18 mmhg of the Finapres, respectively; the difference SD of the proposed method in estimating SBP, DBP and MBP is

7 within 5.1, 4.6, and 4.13 mmhg of the Finapres, respectively; and the MAD of the proposed method in estimating SBP, DBP and MBP is within 4.9, 3.18, and 3.18 mmhg of the Finapres, respectively. Comparing with the PTT methods, it is evident that our proposed method in estimating SBP, DBP and MAP achieves smaller difference mean, SD, as well as MAD, and the difference is significant, indicating a better accuracy. TABLE I ACCURACY, PRECISION, AND AGREEMENT BETWEEN FINAPRES BP AND ESTIMATED BP WITH PTT, AND THAT WITH PTT AND PTT PTT Proposed Method (1) Method () Method [7] [9] SBP Mean DBP N/A (mmhg) MBP N/A SD (mmhg) MAD (mmhg) SBP * DBP N/A MBP N/A SBP * DBP N/A MBP N/A *Statistically significant at the level.5 compared with PTT method (1); statistically significant at the level.5 compared with PTT method (). IV. DISCUSSION In the present study, we proposed a new indicator, the that can reflect the arterial diameter change, to trace the LF variation of BP, and established a novel BP model with the combination of and PTT. We found that the beat-to-beat BP contained both the HF and LF variation components, where the HF component was dominant in PP and similar to that of PTT as well as respiratory signal, while the LF range was primarily in DBP and was also coupled with the. The PTT and were therefore adopted to estimated PP and DBP, respectively, and SBP can be obtained accordingly. The preliminary results on healthy subjects demonstrated the feasibility of using both PTT and to enhance the accuracy of PTT-based BP estimation. A. Effect of Respiration and Vasomotor Tone on BP It has been long recognized that arterial BP fluctuates on a beat-to-beat basis, and the application of spectral techniques to continuous BP has revealed the presence of spontaneous fluctuations including the oscillations at the HF range similar to respiratory frequency, and vasomotion waves in the LF range slower than the respiratory frequency [35, 36]. Power spectral analysis of SBP in this study was consistent with this, i.e., SBP spectrum contained both the HF peak at around.5 Hz, which was coupled with the respiratory frequency, and the LF peaks focused at approximately.1 Hz. Previous investigations demonstrated that the HF component oscillation in BP is related to the respiratory activity [4]. The underlying mechanisms of the fluctuation of BP with respiration is probably due to the intrathoracic pressure change with breathing which has a mechanical effect on venous return, pulmonary vascular and aortic pressure, thereby leading to the cyclic variation in BP. Several researchers have found that the modulation of this HF component for SBP spectrum was linearly related to the respiratory sinus arrhythmia (RSA), indicating that respiratory fluctuations in BP was attributable to RSA [43]. Moreover, Drinnan et al [44] quantified the relation between heart rate and PTT through paced respiration, and found that there was a strong relationship between PTT and heart rate interval. Johansson et al [45] confirmed that PTT varied in pace with respiration and detected the respiration rate reliably from PTT. The result of present study about the coupled frequency between PTT and respiratory signal further verified this. Besides, PTT has been used as a measure of respiratory effort [46], because PTT is inversely proportional to BP, and BP falls with inspiration with the rise in PTT. This can also be observed from time series of PP, PTT and respiratory signal as depicted in Fig. 4. With these considerations, we thus reasoned that PTT could reflect the effect of respiration activity on BP, and can be used to estimate the HF component of BP variation, particularly PP. On the other hand, the LF component of BP is associated with vasomotion waves that result from an oscillation of the sympathetic vasomotor tone. Except for the respiratory synchronous oscillations in BP, the vasomotor tone is also an essential determinant of BP. A number of investigators evidenced that the oscillations of vasomotor tone is caused by local changes in smooth muscle constriction and dilation through the modulation of the sympathetic nervous activity [35, 36]. Though the underlying mechanism has remained elusive, the fluctuations in vasomotor tone are considered to relate to the local adjustment of peripheral resistance to regulate the blood flow thus to meet the local metabolic demand. The adjustment of the peripheral resistance is mainly determined by the arterial diameter change, which is the result of the variations in the tension exerted by the smooth muscle in the vessel walls [47]. In other words, the arterial diameter change might be the primary factor that affects the vasomotor tone. is related to the arterial diameter change, and thus is hypothesized to reflect the vasomotor tone and further the LF oscillations in BP. The spectrum of and BP in this study could explain the hypothesis, which is also consistent with earlier reported results. Our previous study analyzed the relationship between and BP under the influence of autonomic nervous activities, such as deep breathing, Valsalva maneuver, and sustained handgrip, and showed that increase level of BP was associated with a shift of spectral power toward the LF component, suggesting the capability of to evaluate the modulation of sympathetic nervous activity on BP [3]. In addition, Nitzan et al [48] used a similar indicator, the relative amplitude variability of PPG signal, to evaluate autonomic nervous system, and claimed that the variation of PPG baseline and amplitude were mainly concentrated in the LF range, implying the influence of the sympathetic nervous activity. B. PTT Methods for BP Estimation PTT has been extensively studied for continuous BP

8 estimation without a cuff. Despite of its theoretical feasibility, the PTT-based BP measurement method hasn t been widely applied clinically because of several existed problems; particularly the accuracy issue remains the overarching challenge facing researchers. Prior studies have reported that PTT mainly presents the HF component variation of BP, but is inadequate to follow the LF variation [8, 9], which we hypothesize to be the utmost reason leading to the unsatisfactory accuracy of most current PTT-based BP measurement approaches. Continuous BP could be accurately tracked if the noninvasive physiological parameters used for cuffless BP estimation can not only follow the HF respiratory influence on BP, but also its LF oscillation due to the sympathetic modulation of vasomotor tone. McCarthy et al [17] examined two popular PTT-BP algorithms, which are also used in this study as contrast methods, and found that neither these two algorithms could provide reliable BP estimations over a long period. The reason for Chen s method with acceptable accuracy with intermittent calibration is probably due to the combination of LF variation of cuff-based BP with HF variation of PTT. Without the intermittent calibration, the estimation accuracy of SBP was reported to deteriorate from.64±1.55 mmhg to -3.4±9. mmhg within 1 min. As for Poon s method, the accuracy remained relatively stable, from 1.79±1.5 mmhg to 1.4±9.74 mmhg within 1 min. The larger SD after the initial calibration was potentially attributed to the HF variation of PTT for DBP estimation, as can be seen in Fig. 9 (a). However, DBP varies mainly in LF range, as shown in Fig. 5 (b). As a result, the accuracy of SBP estimation was decreased, because SBP was derived from DBP and PP. Besides, most of PTT-based approaches estimated BP with PTT through linear or nonlinear regression method, mostly for SBP estimation, rather than theoretical model based on physiological significance. And DBP has been reported to correlate less with PTT than SBP, which is probably a consequence of the smaller range of variations in DBP as found in those studies [1, 1, 16, 19]. Investigation by Liu et al [4] about the relationship between PTT and BP during exercise and recovery revealed the influence of the vascular smooth muscle tone on BP-PTT relationship, suggesting that the vascular tone should be considered into PTT-based BP estimation. Nevertheless, few studies have attempted to take account this factor into the PTT-based BP model to enhance the accuracy. The strengths of this study, compared with previous studies of PTT-based cuffless BP methods, are the introduction of a novel indicator to evaluate the LF variation of BP, and its combination with PTT to estimate BP to achieve more accurate estimations. With the LF and HF of BP evaluated by and PTT, continuous BP is supposed to be predicted accurately without frequent calibration. Moreover, despite of SBP, estimations of DBP and MBP have also been achieved with acceptable accuracy. C. Limitations There are some limitations in this study. One concern is that we validated our proposed method with Finapres monitor as the reference, while the gold standard for continuous BP measurement should be the invasive intra-arterial method. Second, the beat-to-beat estimation with the proposed method and PTT methods were only within a short period of time. The estimation with long-term period should be further validated. Third, though the effects of breathing pattern on the respiratory component of SBP spectrum have already been analyzed in previous study [49], the pattern on PTT should be further investigated. Furthermore, the utilization of to assess the vasomotor tone on BP should be further validated with more experiments. Finally, the proposed method was tested on healthy subjects, where the sample size is not large enough and subjects with CVDs were not included. V. CONCLUSION AND FUTURE WORK In summary, we have presented a new model for BP estimation using PTT and, and validated the model experimentally. We obtained more accurate results for the cuffless BP estimation by using both PTT and the new indicator, the. By means of one-point calibration, the beat-to-beat SBP, DBP and MBP can be calculated. We found that can indicate the LF variation of BP, whereas PTT reflects the HF fluctuations of BP. In addition, a novel BP model was established with the combination of and PTT, which outperformed the compared PTT-based methods in terms of accuracy on 7 subjects. Most notably, this is the first study to our knowledge to consider the use of to estimate the influence of vasomotor tone on BP. Our results provide evidence for BP variations in both the LF and the HF range where PTT can mainly reflect HF component, and indicates that LF component of BP should be considered to improve the estimation accuracy. Although the pilot study offers a potential method for estimating cuffless BP with better accuracy, it should be further validated with larger sample following with corresponding standard requirement, for example, the IEEE 178 standard for wearable cuffless BP measuring devices. With better accuracy, we expect this method to provide insight for cuffless BP estimation technique. Moreover, through continuous monitoring BP in an unobtrusive way with an acceptable accuracy permits better prevention and management of hypertension, thus reducing the global burden generated by CVDs. ACKNOWLEDGMENT The authors would like to thank the student volunteers from Dept. of Electronic Engineering, The Chinese University of Hong Kong for their participation in the experiment, and Dr. W. Karlen from Mobile Health Systems Laboratory, ETH Zurich for kindly helping to revise this manuscript. REFERENCES [1] C. K. Chow, et al., "Prevalence, awareness, treatment, and control of hypertension in rural and urban communities in high-, middle-, and low-income countries," JAMA, vol. 31, pp , 13. [] G. Parati, et al., "Assessment and management of blood-pressure variability," Nat. Rev. Cardiol., vol. 1, pp , 13. [3] D. Buxi, et al., "A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time," Physiol. Meas., vol. 36, pp. R1-R6, 15.

9 [4] Y. L. Zheng, et al., "Unobtrusive sensing and wearable devices for health informatics," IEEE Trans. Biomed. Eng., vol. 61, pp , 14. [5] L. Peter, et al., "A review of methods for non-invasive and continuous blood pressure monitoring: Pulse transit time method is promising?," IRBM, vol. 35, pp. 71-8, 14. [6] D. J. Hughes, et al., "Measurements of Young's modulus of elasticity of the canine aorta with ultrasound," Ultrason. Imaging., vol. 1, pp , [7] M. W. Chen, et al., "Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration," Med. Biol. Eng. Comput., vol. 38, pp ,. [8] J. Muehlsteff, et al., "Cuffless estimation of systolic blood pressure for short effort bicycle tests: the prominent role of the pre-ejection period," in Proc. 7th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., Shanghai, China, 5, pp [9] C. C. Y. Poon and Y. T. Zhang, "Cuff-less and noninvasive measurements of arterial blood pressure by pulse transit time," in Proc. 7th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., Shanghai, China, 6, pp [1] J. S. Kim, et al., "Effect of confounding factors on blood pressure estimation using pulse arrival time," Physiol. Meas., vol. 9, p. 615, 8. [11] Y. Chen, et al., "Continuous and noninvasive blood pressure measurement: a novel modeling methodology of the relationship between blood pressure and pulse wave velocity," Ann. Biomed. Eng., vol. 37, pp. -33, 9. [1] M. Y. M. Wong, et al., "An evaluation of the cuffless blood pressure estimation based on pulse transit time technique: a half year study on normotensive subjects," Cardiovasc. Eng., vol. 9, pp. 3-38, 9. [13] M. Masè, et al., "Feasibility of cuff-free measurement of systolic and diastolic arterial blood pressure," J. Electrocardiol., vol. 44, pp. 1-7, 11. [14] H. Gesche, et al., "Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method," Eur. J. Appl. Physiol., vol. 11, pp , 1. [15] M. Forouzanfar, et al., "Coefficient-free blood pressure estimation based on pulse transit time-cuff pressure dependence," IEEE Trans. Biomed. Eng., vol. 6, pp , 13. [16] A. Hennig and A. Patzak, "Continuous blood pressure measurement using pulse transit time," Somnologie, vol. 17, pp , 13. [17] B. McCarthy, et al., "An examination of calibration intervals required for accurately tracking blood pressure using pulse transit time algorithms," J. Hum. Hypertens., vol. 7, pp , 13. [18] J. Sola, et al., "Noninvasive and nonocclusive blood pressure estimation via a chest sensor," IEEE Trans. Biomed. Eng., vol. 6, pp , 13. [19] T. Wibmer, et al., "Pulse transit time and blood pressure during cardiopulmonary exercise tests," Physiol. Res., vol. 63, pp , 14. [] V. Chandrasekaran, et al., "Cuffless differential blood pressure estimation using smart phones," IEEE Trans. Biomed. Eng., vol. 6, pp , Apr 13. [1] J. H. Shin, et al., "Non-constrained monitoring of systolic blood pressure on a weighing scale," Physiol. Meas., vol. 3, p. 679, 9. [] M. Y. M. Wong, et al., "The effects of pre-ejection period on post-exercise systolic blood pressure estimation using the pulse arrival time technique," Eur. J. Appl. Physiol., vol. 111, pp , 11. [3] C. Douniama, et al., "Blood pressure tracking capabilities of pulse transit times in different arterial segments: a clinical evaluation," in Comput. Cardiol., 9, pp [4] Q. Liu, et al., "Attenuation of systolic blood pressure and pulse transit time hysteresis during exercise and recovery in cardiovascular patients," IEEE Trans. Biomed. Eng., vol. 61, pp , 13. [5] A. Malliani, et al., "Physiology and Clinical Implications of Variability of Cardiovascular Parameters with Focus on Heart Hate and Blood-Pressure," Am. J. Cardiol., vol. 73, pp. C3-C9, Apr [6] A. Malliani, et al., "Cardiovascular Neural Regulation Explored in the Frequency-Domain," Circ., vol. 84, pp , Aug [7] G. Parati, et al., "Spectral analysis of blood pressure and heart rate variability in evaluating cardiovascular regulation a critical appraisal," Hypertension, vol. 5, pp , Jun [8] H. T. Ma and Y. T. Zhang, "Spectral analysis of pulse transit time variability and its coherence with other cardiovascular variabilities," in Proc. 8th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., New York, USA, 6, pp [9] Q. Liu, et al., "Time-frequency analysis of variabilities of heart rate, systolic blood pressure and pulse transit time before and after exercise using the recursive autoregressive model," Biomed. Signal. Proces., vol. 6, pp , Oct 11. [3] X. R. Ding and Y. T. Zhang, "Photoplethysmogram intensity ratio: a potential indicator for improving the accuracy of PTT-based cuffless blood pressure estimation," To be published in Proc. 37th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., Milan, Italy, 15. [31] Factors that affect blood pressure. Available: ascular/fact_aff_blood_pressure.pdf [3] Y. L. Zhang, et al., "Radial pulse transit time is an index of arterial stiffness," Hypertens. Res., vol. 34, pp , 11. [33] G. Mancia, "Short-and long-term blood pressure variability present and future," Hypertension, vol. 6, pp , 1. [34] J. M. Karemaker and J. Strackee, "Hemodynamic fluctuations and baroreflex sensitivity in humans: a beat-to-beat model," Am. J. Physiol. Heart. Circ. Physiol., vol. 53, pp. H68-H689, [35] A. Malliani, et al., "Cardiovascular neural regulation explored in the frequency domain," Circ., vol. 84, pp , [36] M. Pagani, et al., "Low and high frequency components of blood pressure variability," Ann. N. Y. Acad. Sci., vol. 783, pp. 1-3, [37] R. W. de Boer, "Beat-to-beat blood-pressure fluctuations and heart-rate variability in man: physiological relationships, analysis techniques and a simple model," Ph.D. Dissertation, Department of Physiology, University of Amsterdam, Amsterdam, Netherland, [38] D. H. Bergel, "The dynamic elastic properties of the arterial wall," J. Physiol., vol. 156, pp , [39] N. Westerhof, et al., "The arterial windkessel," Med. Biol. Eng. Comput., vol. 47, pp , 9. [4] T. Ruf, "The Lomb-Scargle periodogram in biological rhythm research: analysis of incomplete and unequally spaced time-series," Biol. Rhythm. Res., vol. 3, pp , [41] J. M. Bland and D. Altman, "Statistical methods for assessing agreement between two methods of clinical measurement," Lancet, vol. 37, pp , [4] A. C. Dornhorst, et al., "Respiratory variations in blood pressure," Circ., vol. 6, pp , 195. [43] J. A. Taylor and D. L. Eckberg, "Fundamental relations between short-term RR interval and arterial pressure oscillations in humans," Circ., vol. 93, pp , Apr [44] M. J. Drinnan, et al., "Relation between heart rate and pulse transit time during paced respiration," Physiol. Meas., vol., pp , Aug 1. [45] A. Johansson, et al., "Pulse wave transit time for monitoring respiration rate," Med. Biol. Eng. Comput., vol. 44, pp , 6. [46] O. Contal, et al., "Pulse transit time as a measure of respiratory effort under noninvasive ventilation," Eur. Respir. J., vol. 41, pp , Feb 13. [47] L. H. Peterson, "Regulation of blood vessels," Circ., vol. 1, pp , 196. [48] M. Nitzan, et al., "The variability of the photoplethysmographic signal-a potential method for the evaluation of the autonomic nervous system," Physiol. Meas., vol. 19, p. 93, [49] D. Laude, et al., "Spectral-Analysis of Systolic Blood-Pressure and Heart-Rate Oscillations Related to Respiration," Clin. Exp. Pharmacol. Physiol., vol., pp , May 1995.

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062

More information

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS R. Mukkamala, R. G. Mark, R. J. Cohen Haard-MIT Division of Health Sciences and Technology, Cambridge, MA, USA Abstract We

More information

Blood Pressure Determination Using Analysis of Biosignals

Blood Pressure Determination Using Analysis of Biosignals Blood Pressure Determination Using Analysis of Biosignals RADIM ČÍŽ 1, MILAN CHMELAŘ 2 1 Department of Telecommunications Brno University of Technology Purkyňova 118, 612 00 Brno CZECH REPUBLIC cizr@feec.vutbr.cz,

More information

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 64 CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 4.1 GENERAL This chapter presents the methodologies that are usually adopted for the measurement of blood pressure, heart rate and pulse

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (  The story so far. Medical Electronics Dr Neil Townsend Michaelmas Term 2001 (wwwrobotsoxacuk/~neil/teaching/lectures/med_elec) Blood Pressure has been measured (in some way) for centuries However, clinical interpretation

More information

Which method is better to measure arterial stiffness; augmentation index, pulse wave velocity, carotid distensibility? 전북의대내과 김원호

Which method is better to measure arterial stiffness; augmentation index, pulse wave velocity, carotid distensibility? 전북의대내과 김원호 Which method is better to measure arterial stiffness; augmentation index, pulse wave velocity, carotid distensibility? 전북의대내과 김원호 Arterial stiffness Arterial stiffness is inversely related to arterial

More information

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND D. Špulák 1, R. Čmejla 1, V. Fabián 2 Czech Technical University in Prague, Faculty of Electrical Engineering, 1 Department of Circuit Theory 2

More information

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Soo-young Ye, Gi-Ryon Kim, Dong-Keun Jung, Seong-wan Baik, and Gye-rok Jeon Abstract In this paper, algorithm estimating the blood

More information

Heart Abnormality Detection Technique using PPG Signal

Heart Abnormality Detection Technique using PPG Signal Heart Abnormality Detection Technique using PPG Signal L.F. Umadi, S.N.A.M. Azam and K.A. Sidek Department of Electrical and Computer Engineering, Faculty of Engineering, International Islamic University

More information

Blood Pressure Estimation Using Photoplethysmography (PPG)

Blood Pressure Estimation Using Photoplethysmography (PPG) Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4

More information

Estimating Mental Stress Using a Wearable Cardio-Respiratory Sensor

Estimating Mental Stress Using a Wearable Cardio-Respiratory Sensor Estimating Mental Stress Using a Wearable Cardio-Respiratory Sensor Jongyoon Choi and Ricardo Gutierrez-Osuna Department of Computer Science and Engineering Texas A&M University, College Station, TX {goonyong,

More information

Effect of Postural Changes on Baroreflex Sensitivity: A study on the EUROBAVAR data set

Effect of Postural Changes on Baroreflex Sensitivity: A study on the EUROBAVAR data set Effect of Postural Changes on Baroreflex Sensitivity: A study on the EUROBAVAR data set Younhee Choi University of Rhode Island Department of ECE Kingston, RI 02881 USA choi.younhee@gmail.com Seok-Bum

More information

Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement

Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement DOI 10.1007/s11517-014-1150-1 Original Article Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement Dingchang Zheng Luigi Yuri Di Marco Alan

More information

A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD

A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD A REVIEW ON NON-INVASIVE HYPERTENSION MONITORING SYSTEM BY USING PHOTOPLETHYSMOGRAPHY METHOD Zahari Taha, Lum Shirley * and Mohd Azraai Mohd Razman Innovative Manufacturing, Mechatronics and Sports (imams)

More information

EVALUATION OF NONINVASIVE PULSE TRANSIT TIME METHODOLOGIES FOR DIAGNOSIS OF HYPERTENSION DANIEL JOHN BADGER. A thesis submitted to the

EVALUATION OF NONINVASIVE PULSE TRANSIT TIME METHODOLOGIES FOR DIAGNOSIS OF HYPERTENSION DANIEL JOHN BADGER. A thesis submitted to the EVALUATION OF NONINVASIVE PULSE TRANSIT TIME METHODOLOGIES FOR DIAGNOSIS OF HYPERTENSION by DANIEL JOHN BADGER A thesis submitted to the Graduate School-New Brunswick Rutgers, The State University of New

More information

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS V.S. Murthy 1, Sripad Ramamoorthy 1, Narayanan Srinivasan 2, Sriram Rajagopal 2, M. Mukunda Rao 3 1 Department of Physics, Indian Institute

More information

Systolic Blood Pressure Estimation Using Video Processing

Systolic Blood Pressure Estimation Using Video Processing 1 Systolic Blood Pressure Estimation Using Video Processing Dao Lam, Gung-Yu Pan, Patrick Hsu, Albert Xu and Chung-Kuan Cheng Department of Computer Science and Engineering, University of California, San

More information

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device

Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable Device 2016 3 rd International Conference on Engineering Technology and Application (ICETA 2016) ISBN: 978-1-60595-383-0 Preprocessing PPG and ECG Signals to Estimate Blood Pressure Based on Noninvasive Wearable

More information

Results of geometric ballistocardiography

Results of geometric ballistocardiography Results of geometric ballistocardiography D. Jezbera, J. Kříž, J. Rajsner, F. Studnička, P. Šeba Abstract Results of application of differential geometry on ballistocardiographic signals is presented.

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Glabella. Continuously Sensing Blood Pressure Behavior using an Unobtrusive Wearable Device. Christian Holz Edward Wang Microsoft Research

Glabella. Continuously Sensing Blood Pressure Behavior using an Unobtrusive Wearable Device. Christian Holz Edward Wang Microsoft Research Glabella Continuously Sensing Blood Pressure Behavior using an Unobtrusive Wearable Device Christian Holz Edward Wang Microsoft Research Glabella continuously records the pulse at 3 locations, pulse transit

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

(received 23 September 2004; accepted 18 October 2004)

(received 23 September 2004; accepted 18 October 2004) ARCHIVES OF ACOUSTICS 29, 4, 597 606 (2004) NON-INVASIVE ULTRASONIC EXAMINATION OF THE LOCAL PULSE WAVE VELOCITY IN THE COMMON CAROTID ARTERY T. POWAŁOWSKI, Z. TRAWIŃSKI Institute of Fundamental Technological

More information

Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference

Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference ORIGINAL RESEARCH Physiological Reports ISSN 2051-817X Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference

More information

ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE

ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE 84TH ASMA ANNUAL SCIENTIFIC MEETING CHICAGO - MAY 12-16, 213 ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE PROGRAM ID NUMBER 445 MANEN O, DUSSAULT C, SAUVET

More information

Cardiovascular Diseases Detecting via Pulse Analysis

Cardiovascular Diseases Detecting via Pulse Analysis Engineering, 2013, 5, 176-180 http://dx.doi.org/10.4236/eng.2013.510b038 Published Online October 2013 (http://www.scirp.org/journal/eng) Cardiovascular Diseases Detecting via Pulse Analysis Jingjing Xia,

More information

Heart rate variability and respiratory sinus arrhythmia assessment of affective states by bivariate autoregressive spectral analysis

Heart rate variability and respiratory sinus arrhythmia assessment of affective states by bivariate autoregressive spectral analysis Heart rate variability and respiratory sinus arrhythmia assessment of affective states by bivariate autoregressive spectral analysis The MIT Faculty has made this article openly available. Please share

More information

COMPARISON OF LINEAR AND NON-LINEAR ANALYSIS OF HEART RATE VARIABILITY IN SEDATED CARDIAC SURGERY PATIENTS

COMPARISON OF LINEAR AND NON-LINEAR ANALYSIS OF HEART RATE VARIABILITY IN SEDATED CARDIAC SURGERY PATIENTS COMPARISON OF LINEAR AND NON-LINEAR ANALYSIS OF HEART RATE VARIABILITY IN SEDATED CARDIAC SURGERY PATIENTS I. Korhonen 1, L.T. Mainardi 2, H.Yppärilä 3, T. Musialowicz 4 1 VTT Information Technology, Tampere,

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

Special Lecture 10/28/2012

Special Lecture 10/28/2012 Special Lecture 10/28/2012 HYPERTENSION Dr. HN Mayrovitz Special Lecture 10/28/2012 Arterial Blood Pressure (ABP) - Definitions ABP Review Indirect Oscillographic Method Resistance (R), Compliance (C)

More information

NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION. D. Barschdorff*, M. Erig*, E. Trowitzsch**

NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION. D. Barschdorff*, M. Erig*, E. Trowitzsch** NONINVASIVE CONTINUOUS BLOOD PRESSURE DETERMINATION D. Barschdorff*, M. Erig*, E. Trowitzsch** *Institute of Electrical Measurement University of Paderborn, D3398 Paderborn, Germany **Sleep Laboratories

More information

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Testing the Accuracy of ECG Captured by through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Data Analysis a) R-wave Comparison: The mean and standard deviation of R-wave amplitudes

More information

Impedance Cardiography (ICG) Method, Technology and Validity

Impedance Cardiography (ICG) Method, Technology and Validity Method, Technology and Validity Hemodynamic Basics Cardiovascular System Cardiac Output (CO) Mean arterial pressure (MAP) Variable resistance (SVR) Aortic valve Left ventricle Elastic arteries / Aorta

More information

Note: At the end of the instructions, you will find a table which must be filled in to complete the exercise.

Note: At the end of the instructions, you will find a table which must be filled in to complete the exercise. Autonomic Nervous System Theoretical foundations and instructions for conducting practical exercises carried out during the course List of practical exercises 1. Deep (controlled) breath test 2. Cold pressor

More information

Experimental Study of Heart Rate Variability by Wrist-Pulse Wave

Experimental Study of Heart Rate Variability by Wrist-Pulse Wave Experimental Study of Heart Rate Variability by Wrist-Pulse Wave Ming Luo a, Liangguang Peng b School of Electronics Engineering, Chongqing University of Posts and Telecommunications, Abstract Chongqing

More information

INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS

INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS A.J. Scheen, J.C. Philips, M. Marchand Division of Diabetes, Nutrition

More information

Estimation of Pulse Arrival Time Using Impedance Plethysmogram from Body Composition Scales

Estimation of Pulse Arrival Time Using Impedance Plethysmogram from Body Composition Scales Estimation of Pulse Arrival Time Using Impedance Plethysmogram from Body Composition Scales Birutė Paliakaitė 1, Saulius Daukantas 2, Andrius Sakalauskas 2, Vaidotas Marozas 1,2 1 Department of Electronics

More information

Chronobiology in Dysautonomia and Cerebrovascular Disease

Chronobiology in Dysautonomia and Cerebrovascular Disease Progress in Clinical Medicine Chronobiology in Dysautonomia and Cerebrovascular Disease JMAJ 44(4): 171 176, 2001 Tsutomu KAMO* and Yoichi TAKAHASHI** *Assistant Professor and **Associate Professor, Department

More information

Heart Rate Variability Analysis Using the Lomb-Scargle Periodogram Simulated ECG Analysis

Heart Rate Variability Analysis Using the Lomb-Scargle Periodogram Simulated ECG Analysis Page 1 of 7 Heart Rate Variability Analysis Using the Lomb-Scargle Periodogram Simulated ECG Analysis In a preceding analysis, our focus was on the use of signal processing methods detect power spectral

More information

Complexity of cardiovascular control in amyotrophic lateral sclerosis patients is related to disease duration

Complexity of cardiovascular control in amyotrophic lateral sclerosis patients is related to disease duration Complexity of cardiovascular control in amyotrophic lateral sclerosis patients is related to disease duration 1,2, Laura Dalla Vecchia 1, Kalliopi Marinou 1, Gabriele Mora 1, Alberto Porta 3,4 ¹ IRCCS

More information

TODAY S TOPIC Blood Pressure & Pulse Wave Measurement Combined in One Procedure Re-classification of Risk Patients

TODAY S TOPIC Blood Pressure & Pulse Wave Measurement Combined in One Procedure Re-classification of Risk Patients CARDIOVASCULAR TECHNOLOGY AND INDICATION SERVICE TODAY S TOPIC Blood Pressure & Pulse Wave Measurement Combined in One Procedure Re-classification of Risk Patients SERIES Hypertension Management in the

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

Author's copy - produced at Lamar University

Author's copy - produced at Lamar University BLOOD PRESSURE ESTIMATION USING PHOTOPLETHYSMOGRAPHY WITH TELEMEDICINE APPLICATION (Centered, double spaced if more than one line, all capitals) (2 inches from title to A Thesis ) A Dissertation Presented

More information

Estimation of arterial pulse wave velocity with a new improved Tissue Doppler method

Estimation of arterial pulse wave velocity with a new improved Tissue Doppler method Estimation of arterial pulse wave velocity with a new improved Tissue Doppler method M. Persson, A. Eriksson, H. W. Persson and K. Lindström Department of Electrical Measurements, Lund University, Sweden

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

(1) Universidad de Zaragoza, Spain. (2) Politecnico di Milano, Italy

(1) Universidad de Zaragoza, Spain. (2) Politecnico di Milano, Italy Time-frequency phase differences and phase locking to characterize dynamic interactions between cardiovascular signals M. Orini 1,2, R. Bailón 1, L.T. Mainardi 2 and P. Laguna 1 (1) Universidad de Zaragoza,

More information

Development of Non-Restraint Blood Pressure Measurement Method Utilizing PWV

Development of Non-Restraint Blood Pressure Measurement Method Utilizing PWV Development of Non-Restraint Blood Pressure Measurement Method Utilizing PWV Yasunari Asakura School of Integrative and Global Majors, University of Tsukuba, Tsukuba, Japan Email: asakura@ golem.kz.tsukuba.ac.jp

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model Gye-rok Jeon, Jae-hee Jung, In-cheol Kim, Ah-young Jeon, Sang-hwa Yoon, Jung-man Son, Jae-hyung Kim, Soo-young Ye,

More information

Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model

Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model Simulation of the Blood Pressure Estimation using the Artery Compliance Model and Pulsation Waveform Model Ahyoung Jeon*, Sooyoung Ye**, Gyerok Jeon* *Department of Biomedical-engineering Pusan National

More information

CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE

CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE VASILE MANOLIU Electrical Engineering Faculty, POLITEHNICA University of Bucharest, Splaiul

More information

Figure removed due to copyright restrictions.

Figure removed due to copyright restrictions. Harvard-MIT Division of Health Sciences and Technology HST.071: Human Reproductive Biology Course Director: Professor Henry Klapholz IN SUMMARY HST 071 An Example of a Fetal Heart Rate Tracing Figure removed

More information

Research Article A Blood Pressure Monitoring Method for Stroke Management

Research Article A Blood Pressure Monitoring Method for Stroke Management Hindawi Publishing Corporation BioMed Research International Volume 214, Article ID 571623, 7 pages http://dx.doi.org/1.11/214/571623 Research Article A Blood Pressure Monitoring Method for Stroke Management

More information

Blood Pressure (BP) measurement becomes a vital parameter now days due to continuous increase in chronic heart

Blood Pressure (BP) measurement becomes a vital parameter now days due to continuous increase in chronic heart ISSN: 0975-766X CODEN: IJPTFI Available Online through Research Article www.ijptonline.com FAST AND ACCURATE METHOD FOR CONTINUOUS BP MEASUREMENT Harkirat Kaur*, Dr. Sunil Agrawal, Mr. Sarvjit Singh Dept.

More information

A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks

A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks A New Approach for Blood Pressure Monitoring based on ECG and PPG Signals by using Artificial Neural Networks Younessi Heravi Mohammad Amin Department of Basic Sciences, School of Medicine, North Khorasan

More information

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection Proceedings of the 2012 International Conference on Industrial Engineering and Operations Management Istanbul, Turkey, July 3 6, 2012 Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter

More information

An Armband Wearable Device for Overnight and Cuff-less Blood Pressure Measurement

An Armband Wearable Device for Overnight and Cuff-less Blood Pressure Measurement TBME-1457-213.R2 1 Abstract 24-hour blood pressure (BP) has significant prognostic value for cardiovascular risk screening, but the present BP devices are mainly cuff-based, which are unsuitable for long-term

More information

Carotid and radial pulse feature analysis with EMFi sensor

Carotid and radial pulse feature analysis with EMFi sensor Carotid and radial pulse feature analysis with EMFi sensor Jarmo Alametsä 1, Ari Palomäki 2, Jari Viik 1 1 Tampere University of Technology, Tampere, Finland, 2 Kanta Häme Central Hospital, Hämeenlinna,

More information

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE No conflict of interest u Technology, based on Pulse Wave Transit Time u Limits of the technology u Precision

More information

Arterial Pressure in CKD5 - ESRD Population Gérard M. London

Arterial Pressure in CKD5 - ESRD Population Gérard M. London Arterial Pressure in CKD5 - ESRD Population Gérard M. London INSERM U970 Paris 150 SBP & DBP by Age, Ethnicity &Gender (US Population Age 18 Years, NHANES III) 150 SBP (mm Hg) 130 110 80 Non-Hispanic Black

More information

Two-point Method for Arterial Local Pulse Wave Velocity Measurement by Means of Ultrasonic RF Signal Processing

Two-point Method for Arterial Local Pulse Wave Velocity Measurement by Means of Ultrasonic RF Signal Processing ARCHIVES OF ACOUSTICS Arch. Acoust., 35, 1, 3 11 (2010) DOI: 10.2478/v10168-010-0001-9 Two-point Method for Arterial Local Pulse Wave Velocity Measurement by Means of Ultrasonic RF Signal Processing Zbigniew

More information

Time-domain cross-correlation baroreflex sensitivity: performance on the EUROBAVAR data set

Time-domain cross-correlation baroreflex sensitivity: performance on the EUROBAVAR data set Appendix II 95 Appendix II Time-domain cross-correlation baroreflex sensitivity: performance on the EUROBAVAR data set B.E. Westerhof, J. Gisolf, W.J. Stok, K.H. Wesseling and J.M. Karemaker, J Hypertens

More information

Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique

Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique www.jbpe.org Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique Original 1 Department of Biomedical Engineering, Amirkabir University of technology, Tehran, Iran Abbaspour

More information

HRV ventricular response during atrial fibrillation. Valentina Corino

HRV ventricular response during atrial fibrillation. Valentina Corino HRV ventricular response during atrial fibrillation Outline AF clinical background Methods: 1. Time domain parameters 2. Spectral analysis Applications: 1. Evaluation of Exercise and Flecainide Effects

More information

Medical University of Gdansk, Gdansk, Poland

Medical University of Gdansk, Gdansk, Poland Increased inspiratory resistance alters the cardiac contribution to the dynamic relationship between blood pressure and pial artery pulsation oscillations in healthy subjects Pawel J. Winklewski 1, Jacek

More information

Department of Electronics and Communication Engineering, MNM Jain Engineering College, Chennai, India. Abstract

Department of Electronics and Communication Engineering, MNM Jain Engineering College, Chennai, India. Abstract Biomedical Research 2017; 28 (2): 689-694 ISSN 0970-938X www.biomedres.info Statistical analysis of pulse rate variability quantified through second derivative photoplethysmogram (SDPPG) and its compatibility

More information

Clinical application of Arterial stiffness. pulse wave analysis pulse wave velocity

Clinical application of Arterial stiffness. pulse wave analysis pulse wave velocity Clinical application of Arterial stiffness pulse wave analysis pulse wave velocity Arterial system 1. Large arteries: elastic arteries Aorta, carotid, iliac, Buffering reserve: store blood during systole

More information

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1982 Characterization of the Oscillometric Method for Measuring Indirect

More information

Sepsis Wave II Webinar Series. Sepsis Reassessment

Sepsis Wave II Webinar Series. Sepsis Reassessment Sepsis Wave II Webinar Series Sepsis Reassessment Presenters Nova Panebianco, MD Todd Slesinger, MD Fluid Reassessment in Sepsis Todd L. Slesinger, MD, FACEP, FCCM, FCCP, FAAEM Residency Program Director

More information

Long-term Sleep Monitoring System and Long-term Sleep Parameters using Unconstrained Method

Long-term Sleep Monitoring System and Long-term Sleep Parameters using Unconstrained Method Long-term Sleep Monitoring System and Long-term Sleep Parameters using Unconstrained Method Jaehyuk Shin, Youngjoon Chee, and Kwangsuk Park, Member, IEEE Abstract Sleep is a most important part of a human

More information

Measurement and Analysis of Radial Artery Blood Velocity in Young Normotensive Subjects

Measurement and Analysis of Radial Artery Blood Velocity in Young Normotensive Subjects Informatica Medica Slovenica 2003; 8(1) 15 Research Paper Measurement and Analysis of Radial Artery Blood in Young Normotensive Subjects Damjan Oseli, Iztok Lebar Bajec, Matjaž Klemenc, Nikolaj Zimic Abstract.

More information

Association of Arg16Gly Polymorphism of the Beta-2 Adrenergic receptor with Baroreflex Sensitivity and Indices of Autonomic Cardiovascular Modulation

Association of Arg16Gly Polymorphism of the Beta-2 Adrenergic receptor with Baroreflex Sensitivity and Indices of Autonomic Cardiovascular Modulation Association of Arg16Gly Polymorphism of the Beta-2 Adrenergic receptor with Baroreflex Sensitivity and Indices of Autonomic Cardiovascular Modulation Ochoa JE 1,2, Correa M 1,5, Gallo J 3,5, McEwen J 1,3,

More information

Design of the HRV Analysis System Based on AD8232

Design of the HRV Analysis System Based on AD8232 207 3rd International Symposium on Mechatronics and Industrial Informatics (ISMII 207) ISB: 978--60595-50-8 Design of the HRV Analysis System Based on AD8232 Xiaoqiang Ji,a, Chunyu ing,b, Chunhua Zhao

More information

Slide notes: References:

Slide notes: References: 1 2 3 Cut-off values for the definition of hypertension are systolic blood pressure (SBP) 135 and/or diastolic blood pressure (DBP) 85 mmhg for home blood pressure monitoring (HBPM) and daytime ambulatory

More information

Special Lecture 11/08/2013. Hypertension Dr. HN Mayrovitz

Special Lecture 11/08/2013. Hypertension Dr. HN Mayrovitz Special Lecture 11/08/2013 Hypertension Dr. HN Mayrovitz Arterial Blood Pressure (ABP) Major Factors Summarized Sympathetic Hormones Arteriole MAP ~ Q x TPR + f (V / C) SV x HR Renal SBP Hypertension =

More information

Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms

Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms J. Jilek*, M. Stork** *Carditech, Culver City, California, USA **Department of Applied Electronics and Telecommunications University

More information

UvA-DARE (Digital Academic Repository)

UvA-DARE (Digital Academic Repository) UvA-DARE (Digital Academic Repository) Beat-to-beat blood-pressure fluctuations and heart-rate variability in man: physiological relationships, analysis techniques and a simple model de Boer, R.W. Link

More information

Chapter 9, Part 2. Cardiocirculatory Adjustments to Exercise

Chapter 9, Part 2. Cardiocirculatory Adjustments to Exercise Chapter 9, Part 2 Cardiocirculatory Adjustments to Exercise Electrical Activity of the Heart Contraction of the heart depends on electrical stimulation of the myocardium Impulse is initiated in the right

More information

Figure 1 Uncontrolled state where ZPFV is increased by 500 ml.

Figure 1 Uncontrolled state where ZPFV is increased by 500 ml. Question 1 Part I Description The initial venous zero pressure filling volume was 2500 ml. The heart rate was fixed at 72 beats/min. The systemic arterial pressure (SAP) was 114/75 mmhg (average = 94 mmhg).

More information

Impedance Cardiography (ICG) Application of ICG for Hypertension Management

Impedance Cardiography (ICG) Application of ICG for Hypertension Management Application of ICG for Hypertension Management 1mA @ 100 khz Impedance Cardiography (ICG) Non-invasive Beat-to-beat Hemodynamic Monitoring Diastole Systole Aortic valve is closed No blood flow in the aorta

More information

Coronary artery disease (CAD) risk factors

Coronary artery disease (CAD) risk factors Background Coronary artery disease (CAD) risk factors CAD Risk factors Hypertension Insulin resistance /diabetes Dyslipidemia Smoking /Obesity Male gender/ Old age Atherosclerosis Arterial stiffness precedes

More information

Disclosures. Afterload on the PV loop. RV Afterload THE PULMONARY VASCULATURE AND ASSESSMENT OF THE RIGHT VENTRICLE

Disclosures. Afterload on the PV loop. RV Afterload THE PULMONARY VASCULATURE AND ASSESSMENT OF THE RIGHT VENTRICLE THE PULMONARY VASCULATURE AND ASSESSMENT OF THE RIGHT VENTRICLE Ryan J. Tedford, MD Heart Failure, Mechanical Circulatory Support, and Cardiac Transplantation Division of Cardiology, Department of Medicine

More information

BARORECEPTOR-SENSITIVE FLUCTUATIONS OF HEART RATE AND PUPIL DIAMETER

BARORECEPTOR-SENSITIVE FLUCTUATIONS OF HEART RATE AND PUPIL DIAMETER BARORECEPTOR-SENSITIVE FLUCTUATIONS OF HEART RATE AND PUPIL DIAMETER G. Calcagnini, P.Giovannelli, F. Censi, P. Bartolini, V. Barbaro Biomedical Engineering Laboratory, Istituto Superiore di Sanità, Roma,

More information

Biomedical Instrumentation D. The Photoplethysmogram

Biomedical Instrumentation D. The Photoplethysmogram Biomedical Instrumentation D. The Photoplethysmogram Dr Gari Clifford Based on slides by Prof. Lionel Tarassenko The need for real-time oxygen saturation monitoring Respiratory failure & pulmonary disease

More information

The Value of a BP Determination Method Using a Novel Non-Invasive BP Device against the Invasive Catheter Measurement

The Value of a BP Determination Method Using a Novel Non-Invasive BP Device against the Invasive Catheter Measurement The Value of a BP Determination Method Using a Novel Non-Invasive BP Device against the Invasive Catheter Measurement Jinsong Xu 1., Yanqing Wu 1., Hai Su 1 *, Weitong Hu 1, Juxiang Li 1, Wenying Wang

More information

Diverse Techniques to Detect Arterial Stiffness

Diverse Techniques to Detect Arterial Stiffness Diverse Techniques to Detect Arterial Stiffness 백상홍가톨릭대학교강남성모병원순환기내과 혈관연구회창립심포지움 2005, 3, 3 Small Arteries Arterial Remodelling Thickness Internal diameter Wall / Lumen Large Arteries Cross sectional

More information

LabVIEW based Non-Invasive prototype device for cardiac diagnosis using Nadi Shastra

LabVIEW based Non-Invasive prototype device for cardiac diagnosis using Nadi Shastra Original Article LabVIEW based Non-Invasive prototype device for cardiac diagnosis using Nadi Shastra Parth Shah 1, A.V. Leio 1, Siddaram Mahajan* 1, Shaik Jameel 2 1 Department of Biomedical Engineering,

More information

SYNCHRONIZED MEASUREMENTS OF MAXIMUM BLOOD FLOW VELOCITIES IN CAROTID, BRACHIAL AND FEMORAL ARTERIES, AND ECG IN HUMAN POSTURE CHANGES

SYNCHRONIZED MEASUREMENTS OF MAXIMUM BLOOD FLOW VELOCITIES IN CAROTID, BRACHIAL AND FEMORAL ARTERIES, AND ECG IN HUMAN POSTURE CHANGES SYNCHRONIZED MEASUREMENTS OF MAXIMUM BLOOD FLOW VELOCITIES IN CAROTID, BRACHIAL AND FEMORAL ARTERIES, AND ECG IN HUMAN POSTURE CHANGES Y.Hirao*, **, T.Kuroda*, D.Zhang*, Y.Kinouchi*, H.Yamaguchi*** and

More information

PHYSIOEX 3.0 EXERCISE 33B: CARDIOVASCULAR DYNAMICS

PHYSIOEX 3.0 EXERCISE 33B: CARDIOVASCULAR DYNAMICS PHYSIOEX 3.0 EXERCISE 33B: CARDIOVASCULAR DYNAMICS Objectives 1. To define the following: blood flow; viscosity; peripheral resistance; systole; diastole; end diastolic volume; end systolic volume; stroke

More information

Study and Design of a Shannon-Energy-Envelope based Phonocardiogram Peak Spacing Analysis for Estimating Arrhythmic Heart-Beat

Study and Design of a Shannon-Energy-Envelope based Phonocardiogram Peak Spacing Analysis for Estimating Arrhythmic Heart-Beat International Journal of Scientific and Research Publications, Volume 4, Issue 9, September 2014 1 Study and Design of a Shannon-Energy-Envelope based Phonocardiogram Peak Spacing Analysis for Estimating

More information

Cardiovascular system

Cardiovascular system Cardiovascular system L-4 Blood pressure & special circulation Dr Than Kyaw 27 February 2012 Blood Pressure (BP) Pressure generation and flow Blood is under pressure within its closed system. Pressure

More information

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes Dual channel photoplethysmography studies of cardio-vascular response to the body position changes R. Erts, I. Kukulis, J. Spigulis, L. Kere University of Latvia, Raina Blvd. 9, Riga, LV-, Latvia Renars.erts@lu.lv

More information

EFFECTS OF MECHANICAL PUMPING ON THE ARTERIAL PULSE WAVE VELOCITY : PERIPHERAL ARTERY AND MICRO-VESSELS

EFFECTS OF MECHANICAL PUMPING ON THE ARTERIAL PULSE WAVE VELOCITY : PERIPHERAL ARTERY AND MICRO-VESSELS EFFECTS OF MECHANICAL PUMPING ON THE ARTERIAL PULSE WAVE VELOCITY : PERIPHERAL ARTERY AND MICRO-VESSELS Shu-Mei Wu*, Yio-Wha Shau**, Bor-Shyh Lin*, Fok-Ching Chong* *Institute of Electrical Engineering,

More information

Insulin resistance influences 24h heart rate and blood pressure variabilities and cardiovascular autonomic modulation in normotensive healthy adults

Insulin resistance influences 24h heart rate and blood pressure variabilities and cardiovascular autonomic modulation in normotensive healthy adults Insulin resistance influences 24h heart rate and blood pressure variabilities and cardiovascular autonomic modulation in normotensive healthy adults Ochoa JE 1, Correa M 2, Valencia AM 2, Gallo J 2, McEwen

More information

Reinvestigation of the relationship between the amplitude of the first heart sound to cardiac dynamics

Reinvestigation of the relationship between the amplitude of the first heart sound to cardiac dynamics ORIGINAL RESEARCH Physiological Reports ISSN 251-817X Reinvestigation of the relationship between the amplitude of the first heart sound to cardiac dynamics Hong Tang 1, Chengjie Ruan 1, Tianshuang Qiu

More information

Theoretical and practical questions in the evaluation of arterial function Miklós Illyés MD. Ph.D.

Theoretical and practical questions in the evaluation of arterial function Miklós Illyés MD. Ph.D. Theoretical and practical questions in the evaluation of arterial function Miklós Illyés MD. Ph.D. TensioMed Arterial Stiffness Centre, Budapest Heart Institute, Faculty of Medicine, University of Pécs

More information

A PARADIGM FOR QUANTIFYING VENTRICULAR CONTRACTION USA

A PARADIGM FOR QUANTIFYING VENTRICULAR CONTRACTION USA CELLULAR & MOLECULAR BIOLOGY LETTERS 331 A PARADIGM FOR QUANTIFYING VENTRICULAR CONTRACTION JOSEPH L. PALLADINO 1 and ABRAHAM NOORDERGRAAF 2 1 Department of Engineering, Trinity College, Hartford, CT,

More information

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah CARDIOVASCULAR MONITORING Prof. Yasser Mostafa Kadah Introduction Cardiovascular monitoring covers monitoring of heart and circulatory functions It makes it possible to commence interventions quickly in

More information