Hypertensive patients commonly use home. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors Steven A. Yarows and Robert D.

Size: px
Start display at page:

Download "Hypertensive patients commonly use home. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors Steven A. Yarows and Robert D."

Transcription

1 AJH 2000;13: Measurement Variation Among 12 Electronic Home Blood Pressure Monitors Steven A. Yarows and Robert D. Brook Although there are AAMI and BHS standards for accuracy of electronic home electronic blood pressure monitors (HBPM), patient composition differences and differences in manufacturer s algorithm for calculation of the systolic and diastolic measurement may result in measurement differences between monitors. The aim of this study was the measurement of differences among HBPM. Paired comparisons were performed between the Omron 712c electronic home monitor and each of 12 other HBPM (Sunbeam 7654, Sunbeam 7623, Omron 711, Omron 432c, A&D-UA767, Lumiscope 1085M, Omron 725CIC, Assure A30, Lumiscope 1083N, Omron 815, Omron 605, and Assure BD- W20), in addition to comparison to the auscultatory method by trained observers. Measurements were made in normotensive subjects in an ambulatory setting. The main outcome measures were systolic and diastolic blood pressure measurements. All of the HBPM, except for the Sunbeam 7654 and the Assure A30/ BD-W20 (wrist) models, demonstrated small differences of <4/4 mm Hg for systolic/diastolic measurements with pulse measurement differences of <3 beats/min. These differences were less than the differences previously reported for office BP auscultation of 6/5 10 mm Hg for systolic/diastolic measurements. The Omron 712c, passing previous AAMI and BHS standards, measured the systolic reading within 2 mm Hg of auscultatory mercury or aneroid measurement and under-measured the diastolic by 6 9 mm Hg. Differences in the patient composition studied could account for the difference. The wrist and finger manometers performed clinically similar to the Omron 712c, except for the Assure BD-W20, which overmeasured the diastolic by 7 mm Hg. It is concluded that the small differences among the various HBPM, which are less than those in clinical office auscultation, should encourage greater use of electronic manometers in the office and at home. Am J Hypertens 2000;13: American Journal of Hypertension, Ltd. KEY WORDS: Blood pressure, self-care, home blood pressure devices, validation, standards. Hypertensive patients commonly use home blood pressure monitors, and electronic oscillometric monitors are increasingly being used in hospitals and outpatient settings to measure blood pressure (BP), although there is usually little physician input into the purchase. 1 3 Digital electronic monitors are also being used in hypertensive outcome-based studies. 4 Home blood pressure monitor (HBPM) types include aneroid and electronic models. Most modern electronic HBPM Received April 12, Accepted July 19, From the Division of Hypertension, Internal Medicine, University of Michigan Health System, Ann Arbor, Michigan. An educational grant from Bristol-Myers-Squibb was used for the purchase of the electronic devices. Address correspondence to Steven A. Yarows, MD, FACP, Chelsea Internal Med. Consultants PC, 128 Van Buren Street, Chelsea, MI 48118; syarows@umich.edu 2000 by the American Journal of Hypertension, Ltd /00/$20.00 Published by Elsevier Science, Inc. PII S (99)00182-X

2 AJH MARCH 2000 VOL. 13, NO. 3, PART 1 MEASUREMENT VARIATION AMONG HOME BP MONITORS 277 use the oscillometric method to determine BP at the upper arm, wrist, or finger. Some have disparaged the use of wrist and finger monitors as having uncertain accuracy and marked postural variability. 5 7 Each of the manufacturers of HBPM uses a unique proprietary algorithm to calculate systolic and diastolic BP. 8 There are many distributors of electronic blood pressure devices, which results in many different model names, although many distributors use similar manufacturer models. The Association for the Advancement of Medical Instrumentation (AAMI) and the British Hypertensive Society Working Party on Blood Pressure Measurement (BHS) standards have been established to assure accuracy of the HBPM The AAMI and BHS standards both require the monitors to be tested on 85 different subjects with defined characteristics. Manufacturers must comply with the AAMI standards to distribute and sell HBPM in the United States, however, the Food and Drug Administration (FDA) does not actually review the data from the AAMI test, nor are these results usually published in peer-reviewed journals. Thus, physicians have only to assume that the proper tests have been performed. Because each monitor that is studied using AAMI/ BHS standards has its own subject composition, different monitors could use different types of subjects, which could potentially result in measurement differences among different HBPM. We wanted to determine the magnitude of these differences, as this has not been previously studied. METHODS The HBPM were purchased in retail stores for the study, with the exception of the Omron 725CIC (Omron Healthcare, Inc., Vernon Hills, IL) and the Omron 605 (wrist), which were donated by pharmaceutical companies (Novartis and Bristol-Myer Squibb, respectively) for patient use. One author (R.B.) previously owned the aneroid sphygmomanometer. All BP measurements were performed on two authors (S.Y. and R.B.). Two different Omron 712C monitors were used in the study. A paired comparison of the Omron 712C upper arm manometer to another HBPM, mercury sphygmomanometer, or aneroid sphygmomanometer was performed. Each paired comparison was performed only by one subject and was independently performed of the other comparisons. Each monitoring was performed in the seated position with the back supported, always using the left arm. A total of 85 readings were performed on each HBPM monitor and, as each monitor was always compared to either the Omron 712c or a mercury/aneroid manometer, each paired set of readings had 170 BP measurements. Ten readings were first obtained at a single setting with the Omron 712C, and the next 10 readings were obtained with the comparison monitor. The beginning test monitor sequence was alternated to prevent any measurement bias due to the order of readings. The measurements were not separated by any specific interval and often were obtained immediately after the last reading. The final measurement set of each monitor contained five readings to complete the 85 measurements. The aneroid or mercury sphygmomanometer readings were performed by each of the authors on the other author. The aneroid readings were also performed by a licensed practical nurse (LPN) that has practiced clinical internal medicine for 10 years. The wrist and finger manometers were tested with the wrist placed on a surface, which resulted in the top of the wrist horizontal to 5 inches below the superior border of the left clavicle (heart level). The measurements were entered into Microsoft Excel 97 (Microsoft Corporation, Redmond, WA). Standard deviation, 95% confidence intervals, two-sample equal variance two-tailed Student t test, mean average, and differences among the simulation settings and the machines were calculated using these functions in Excel. All of the Omron and A&D devices were calibrated for static measurements (80, 110, 120, and 150 mm Hg) of accuracy using the BP Pump (Bio-Tek Instruments, Winooski, VT). The machine was last calibrated May 6, 1998, and was to be due for recalibration in May Nether the wrist nor finger HBPM could be calibrated by any known method. The other HBPM were calibrated with the BP Pump using dynamic measurements of 120/80 and 110/70 mm Hg through the previously described technique. 12 The aneroid manometer was tested with the BP pump at 150, 120, 110, and 80 mm Hg and the difference was 1 mmhg. RESULTS Ninety-Five Percent Confidence Intervals of the Paired Measurements The 12 devices tested were all sold for home BP measurement. The devices were compared to the Omron 712c, the calibrated aneroid manometer, and/or to a mercury manometer. A lack of any statistically significant difference was seen in the comparisons of the Omron 712c and other Omron devices for both systolic and diastolic measurement, except for the systolic comparison of the Omron 725CIC and diastolic comparison of the Omron 432c, Omron 725CIC, and Omron 815 (finger). The Sunbeam 7654, Lumiscope 1085M, Assure A30, Lumiscope 1083N, and Assure W20 demonstrated significant differences in both systolic and diastolic measurements; however, the Sunbeam 7623 and A&DUA767 did not have significant differences in either measurement. The pulse measurements were concordant for all devices except the Omron 712c versus the Lumiscope 1083N (finger) and the Omron 605 (wrist).

3 278 YAROWS AND BROOK AJH MARCH 2000 VOL. 13, NO. 3, PART 1 FIGURE 1. The systolic, diastolic, and pulse differences of the electronic device minus Omron712c (BPM). All are corrected for a single Omron712c device by adding or subtracting the measured average difference of the other Omron712c device. Correction for the Two Omron 712c Machines There were two different Omron 712c machines used for the different paired measurements. Author R.B. had 1530 BP measurements performed and author S.Y. had 2040 measurements performed for a total of 3570 BP measurements in this study. Each Omron 712c machine was compared another other using the same technique of 85 readings for each machine. The first Omron 712c machine read the systolic higher by 1.4 mm Hg and the diastolic higher by 1.5 mm Hg than the second Omron 712c machine (P.05 for both). The second Omron 712c measured the pulse 0.4 beats/ min higher than the first Omron 712c (P.4). Although the differences between the two Omron 712c machines were not clinically significant, the measurements were adjusted to correct for the differences between machines. If the second Omron 712c machine was used, adding 1.4 and 1.5 mm Hg to the respective systolic and diastolic readings corrected the differences in the systolic and diastolic measurements to compare to the first Omron 712c measurements. Similarly, 0.4 beats/min was subtracted from the second Omron 712c readings to compare to the first Omron 712c machine. Differences Between HBPM The adjusted differences between the devices are seen in Figure 1. All of the HBPM had a small diffrence between the HBPM and Omron 712c of 4 mm Hg for the systolic measurements, excluding Sunbeam 7654 (6 mm Hg difference). Self-reading the aneroid systolic measurements demonstrated a larger difference of 7 to 12 mm Hg compared to the automatic Omron 712c. There were small differences ( 1.3 mm Hg) between the Omron 712c and the aneroid or mercury manometers if another trained observer took the systolic measurements. All of the HBPM demonstrated small differences when compared to the Omron 712c in diastolic measurement of 4 mm Hg with the exception of the Assure BD-W10 wrist (over-measured by 7 mm Hg), and Assure BD A30 (over-measured by 7 mm Hg). When the Omron 712c device was compared to the aneroid or mercury manometer, the Omron 712c was

4 AJH MARCH 2000 VOL. 13, NO. 3, PART 1 MEASUREMENT VARIATION AMONG HOME BP MONITORS 279 TABLE 1. PREVIOUSLY PUBLISHED METHODS FOR COMPARISON OF DIFFERENT HBPM Author, Ref. No. of Subjects BP/Subject No. of Models No. of BP/Models Van Egmond et al O Brien-Pre-Home models (3 monitors each) 6 (2 monitors for 3 models) O Brien-Device Validation * 225 Bortolotto 30/model Imai-Simultaneous /model models (3 monitors each) 90 Imai-Inter-device Evans Yarows/Brook and * Four of the initial seven models passed the pre-home and after-use variability tests. BP, blood pressure; HBPM, home blood pressure monitors. found consistently to under-measure the diastolic pressures. Self-measurement of the diastolic pressure using an aneroid manometer showed the Omron 712c to under-measure the diastolic by 2 to 11 mm Hg. Measurement of the diastolic by the other author using a mercury manometer demonstrated an undermeasurement of 6 to 9 mm Hg with a similar 9 mm Hg under-measurement by the LPN using an aneroid manometer. All devices showed a clinically insignificant difference of 3 beats/min of the pulse measurement compared to the Omron 712c. DISCUSSION What is the best technique to compare measurement differences between HBPM? Most articles that determine accuracy of HBPM test only a single device against a standard device There are limited comparisons using multiple HBPM (Table 1) Different methods were used for the comparisons in these studies with a variety of subject numbers (13 to 120 per model), number of models compared (2 to 30), BP per subjects (3 to 100), and number of BP per models (6 to 100). As there is no established gold standard, we chose to use two normotensive subjects, so as to limit human variation, using 85 readings per model to minimize the measurement errors. This study was not meant to validate the accuracy of the HBPM tested but, rather, to compare the monitors. The development of AAMI and BHS standards has improved the accuracy of the monitors. The HBPM that were manufactured before AAMI or BHS standards usually failed these tests. 18,22 Most manufacturers except for Omron Healthcare, Inc, do not publish the monitor testing in peer-reviewed journals. The following monitors have passed the AAMI and the BHS standards: Omron 711, Omron 722c, Omron 713c, Omron 403c, Terumo ES-H51, Assure A30, Assure W20, A&D UA-767, and A&D UA-767PC The Omron 706 failed the BHS standards by only receiving a B grade by BHS standards for systolic and a C grade for diastolic measurements. Lumiscope and Sunbeam did not make AAMI or BHS reports available to us. Our main finding was that there were only small differences by all the HBPM (Figure 1), which would unlikely be clinically important. The second important finding is a constant under-measurement of the diastolic measurement by all of the HBPM when compared to measurements by trained personnel using the auscultatory technique. The third clinically relevant finding was that, contrary to previous reports, the wrist and finger manometers tested produced similarly precise BP readings comparable to upper arm HBPM when used correctly. The Omron 712c usually measured the systolic slightly lower than the other devices, whereas the diastolic was both under- and over-measured compared to the other monitors. Generally, there were smaller differences between different Omron models and the Omron 712c than other manufacturers. The paired pulse measurements were similar, with the greatest difference ( 3 beats/min) measured between the Omron 712c and the Lumiscope 1083N (finger). Thus, generally the sensors are equivalent at sensing the pulse, however, the different algorithms used by the different manufacturers probably account for the small differences in the measurement of systolic and diastolic readings. The largest difference between the Omron 712c and the electronic device was over-measurement of the systolic by 6 mm Hg (Sunbeam 7654), diastolic by 7 mm Hg (Assure-BD-W20 wrist), and pulse by 3 beats/ min (Lumiscope 1083N finger). This measurement error is similar to the human error measured in an actual clinical, non research setting. The measurement error in a Brazil University Hospital was 6 mm Hg for

5 280 YAROWS AND BROOK AJH MARCH 2000 VOL. 13, NO. 3, PART 1 systolic and 10 mm Hg for diastolic, even after taking a test on proper BP measurement technique. 30 A trained registered nurse was compared to usual healthcare providers at a family practice center, with the majority of the healthcare providers being medical assistants and licensed practical nurses, along with a few registered nurses. 31 A difference of 6 mm Hg systolic and 5 mm Hg diastolic was measured. Potentially correctable errors that could be improved by training were estimated to be only 2 mm Hg for systolic and 1 mm Hg for diastolic. Although the Assure A30 measured the greatest difference compared to the Omron 712c by over-measuring the systolic by 4 mm Hg and diastolic by 7 mm Hg, this difference was comparable to the human error in measurement. The human error is variable with each BP measurement. If the same electronic device is used, most of the difference would be due to BP variation rather than machine error, as the difference in measurement that we observed would be constant because of the algorithm defining systolic and diastolic BP and the low variability of the machines. 12 The Omron 712c very accurately measured the systolic pressure within 2 mm Hg of the measurement by trained observers using an aneroid or mercury manometer. Manually inflating a cuff was previously reported to increase the systolic BP by 12 to 13 mm Hg because of the required muscular activity. 32 Self-reading the systolic measurements by the aneroid method demonstrated a7to12mmhgdifference compared to the automatic Omron 712c; however the two manually inflated electronic machines (Sunbeam 7623 and Omron 432c) were easy to inflate and only demonstrated a 2 mm Hg difference compared to the fully automated Omron 712c. The Omron 712c under-measured the diastolic reading by 6 to 9 mm Hg when compared to an aneroid or mercury manometer measured by a trained observer. The Omron 711 and the Omron 725CIC, when compared to a mercury manometer, under-measured the diastolic by 7 mm Hg (P.01) and 6 mm Hg (P.05), respectively. 33 The Omron R3 wrist manometer was also found to under-measure diastolic compared to mercury measurement; however, there was no significant difference when compared to intra-arterial measurements. 34 This would suggest that the diastolic measurement algorithm with Omron monitors better reflects diastolic intra-arterial than auscultatory pressure. The Omron HEM705CP, which uses the same protocol as the Omron 712c, achieved a B systolic grade and an A diastolic grade using the British Hypertensive Society (BHS) protocol. 35,36 The Omron HEM705CP passed the AAMI accuracy criteria with the systolic under-measured by 2 7 mm Hg and the diastolic under-measured by 1 6 mm Hg. The difference of this study with our results is unclear; however, this study used two normotensive subjects, and the AAMI test was performed in mostly normotensive subjects, along with additional hypertensive and hypotensive patients with a variety of arm circumferences. Additionally, the AAMI population was European versus American in this study. Measuring BP at the wrist or finger requires the measurement point to be placed at the heart level to avoid systemic error from hydrostatic pressure. 5 Raising the wrist to a vertically up position lowered the BP by 13/18 mm Hg compared to heart level, whereas lowering the wrist to a vertically down position artificially raised the BP by 18/10 mm Hg. We carefully measured the wrist and finger BP at heart level and found that the readings were clinically similar to the systolic measurement of the Omron 712c. The wrist and finger measurements were also similar to the diastolic measurement of the Omron 712c, except that the Assure BD-W20 (wrist) over-measured the diastolic by 7 mm Hg. The Omron HEM-601 (wrist) was compared by hourly measurements to the boso Oscillomat (upper arm) and SpaceLabs ambulatory BP machines and the results were similar to our study. 37 No significant differences were recorded between the daily mean systolic (P.07) or diastolic (P.97) BP determined by ambulatory measurements. The Ormon HEM-601 also passed an AAMI test. 38 The following finger devices did not pass AAMI or BHS standards: Omron HEM-812F, Marshal F-88, EW276H (National Ltd), and Omron HEM804F Increased systolic pressure in a geriatric population increased the difference between the finger and upper arm measurement technique, although this may not be clinically important. 41 A limitation of this study is that BP was only measured in two subjects that were both normotensive. Comparison of different electronic monitors using hypertensive and hypotensive patients should be performed. Nonetheless, accuracy of measurement is most important between 130 and 150 mm Hg systolic and between 80 and 100 mm Hg diastolic, as clinical decisions are determined based on these measurements at these levels. It is reassuring that each machine had small differences between each other. The small differences between the machines were equal to or less than the difference that is seen in the usual clinical measurement by auscultation. The small differences between wrist/finger and upper arm machines may encourage increased patient usage, if patients can be given instruction about the correct method of usage. If the same electronic manometer brand is used, the instrument error will be less than measurement with auscultation. Although the diastolic measurement for the

6 AJH MARCH 2000 VOL. 13, NO. 3, PART 1 MEASUREMENT VARIATION AMONG HOME BP MONITORS 281 HBPM was lower than the mercury measurement in our study, the recommended upper limit of normal for home BP measurement is also lower (135/85 mm Hg) and this should correct for this difference. 43 Additionally, home BP readings over a 2-week period were similar to office readings when the same model of HBPM was used for both measurements. 4 Electronic manometers should be encouraged for wider clinical usage both in the office and at home. ACKNOWLEDGMENTS We are grateful to Judy Williams, LPN, for her assistance in the measurement of the aneroid blood pressures. We are also grateful for the careful review by Alan Weder, MD. REFERENCES 1. Jeck T, Edmonds D, Mengden T, Schubert M, Vetter W: [Performing self-measurement of blood pressure: a patient survey] Umgang mit der Blutdruck-selbstmessung: eine Patientenumfrage: Schweiz Rundsch Med Prax 1991;80: Personal communication, Eric Vennemeyer, Frost & Sullivan, Krecke HJ, Fleischmann C, Bokmann M: [Distribution and acceptance of self measurement of blood pressure in the Hamburg area] Verbreitung und Akzeptanz der Blutdruckselbstmessung im Grossraum Hamburg. Schweiz Rundsch Med Prax 1989;78: Hansson L, Zanchetti A, Carruthers SG, Dahlof B, Elmfeldt D, Julius S, Menard J, Rahn KH, Wedel H, Westerling S: Effects of intensive blood-pressure lowering and low-dose aspirin in patients with hypertension: principal results of the Hypertension Optimal Treatment (HOT). Lancet 1998;351: Pickering TG: Technologic advances in home blood pressure monitoring. Blood Press Monit 1998;3(suppl 1):S3 S6. 6. Nesselroad JM, Flacco VA, Phillips DM, Kruse J: Accuracy of automated finger blood pressure devices. Fam Med Mar 1996;28: Latman NS, Latman A: Evaluation of instruments for noninvasive blood pressure monitoring of the wrist. Biomed Instrum Technol 1997;31: Yong P, Geddes LA: A surrogate arm for evaluating the accuracy of instruments for indirect measurement of blood pressure. Biomed Instrum Technol 1990;24: Electronic or automated sphygmomanometers. Developed by Association for the Advancement of Medical Instrumentation. Approved October 16, American National Standards Institute, Inc.; 330 Washington Boulevard, Suite 400 Arlington, Virginia : O Brien E, Petrie J, Littler W, et al: The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990;8: O Brien E, Petrie J, Littler W, et al: Short report: an outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993;11: Yarows SA, Amerena JV: Accuracy of 10 models of home blood pressure monitoring using an oscillometric simulator. What is 120/80? Blood Pressure Monitor : James GD, Yee LS, Cates EM, Schlussel YR, Pecker MS, Pickering TG: A validation study of the Instromedix Baro-Graf Qd home blood pressure monitor. Am J Hypertens 1990;3: Foster C, McKinlay S, Cruickshank JM, Coats AJS: Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Hum Hypertens 1994;8: Mufunda J, Sparks B, Chifamba J, Dakwa C, Matenga JA, Adams JM, Sparks HV: Comparison of the Omron HEM-713C automated blood pressure monitor with a standard ausculatory method using a mercury manometer. Cent Afr J Med 1996;42: Walma EP, van Dooren C, van der Does E, Prins A, Mulder P, Hoes AW: Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. J Hum Hypertens 1995;9: Van Egmond J, Lenders JWM, Weernink E, Thien T: Accuracy and reproducibility of 30 devices for selfmeasurement of arterial blood pressure. Am J Hypertension 1993;6: O Brien E, Mee F, Atkins N, O Malley K: Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure. J Hypertens 1990;8: Evans CE, Haynes RB, Goldsmith CH, Hewson SA: Home blood pressure-measuring devices: a comparative study of accuracy. J Hypertens 1989;7: Imai Y, Abe K, Sasaki S, Minami N, Munakata M, Sakuma H, Hashimoto J, Sekino H, Imai K, Yoshinaga K: Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: comparison between cuff-oscillometric and microphone methods. J Hypertens 1989;7: Borotolotto LA, Henry O, Hanon O, Sikias P, Mourad JJ, Girerd X: Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722c and HEM-735c. Blood Press Monit 1999;4: Harrison DW, Kelly PL. Home health-care: accuracy, calibration, exhaust, and failure rate comparisons of digital blood pressure monitors. Med Instrum 1987;21: Personal communication, Accuracy of the Omron HEM-711 portable monitor for home measurement of blood pressure. Omron Healthcare Inc, Osamu Shirasaki, September 2, Cordoba R, Fuertes MI, Alvarez A, Molina I, Solans R, Melero I: [The evaluation of a self-measurement arterial pressure monitor: the OMRON-HM 722C]. [Spanish] Atencion Primaria 1997;20: Mufunda J, Sparks B, Chifamba J, Dakwa C, Matenga JA, Adams JM, Sparks HV: Comparison of the Omron HEM-713C automated blood pressure monitor with a

7 282 YAROWS AND BROOK AJH MARCH 2000 VOL. 13, NO. 3, PART 1 standard ausculatory method using a mercury manometer. Cent Afr J Med 1996;42: Walma EP, van Dooren C, van der Does E, Prins A, Mulder P, Hoes AW: Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. J Hum Hypertens 1995;9: Kwek K, Chan Y, Tan KH, Yeo G: Validation of an oscillometric electronic sphygmomanometer in an obstetric population. Am J Hypertens 1998;11: Kerestan SP: Enhanced hypertensive patient health care: performance of the B-D Assure blood pressure monitors. Personal communication, July Personal communication, A&D, Jerry Wang. 30. Rocha JC, Rocha AT, Magossi AMG, Leao RW, Palu MJF, Moreira DC: Evaluation of the technique for taking blood pressure by health care workers in a university hospital (Abstract). Am J Hypertens 1988;11:66A. 31. Kay LE: Accuracy of blood pressure measurement in the family practice center. JABFP 1998;11: Veerman DP, van Montfrans GA, Wieling W: Effects of cuff inflation on self-recorded blood pressure. Lancet 1990;335: Yarows SA: Home blood pressure monitoring in primary care. Blood Press Monit 1998;3(suppl 1):S11 S Watson S, Wenzel RR, di Matteo C, Meier B, Luscher TF: Accuracy of a new wrist cuff oscillometric blood pressure device. Am J Hypertens 1998;11: Personal communication, Omron Healthcare, Osamu Shirasaki, Director of Technology, September 2, O Brien E, Mee F, Atkins N, Thomas M: Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Philips HP5332, and Nissei DS-175. Blood Press Monitor 1996;1: Eckert S, Gleichmann S, Gleichmann U: Blood pressure self-measurement in upper arm and in wrist for treatment of arterial hypertension. Zeitschrift für Kardiologie. 1996;85: Izumi K: A comparison between blood pressure measurements taken with an Omron digital wrist blood pressure monitor (Model HEM-601) and those taken by the asculatory method. Data on file at Omron Healthcare, Inc. 39. Wambach G, Jacob R, Treeger A, Stimpel M: [How reliable is oscillometric blood pressure determination at the finger using the Omron device?] [German] Schweizerische Rundschau für Medizin Praxis 1992;81: Veerman DP, Lenders JW, Thien T, van Montfrans GA: LAM 100/Marshall F-88: accuracy and precision of a new device for discontinuous finger blood pressure measurement. J Hum Hypertens 1993;7: Teshima M, Kuwajima I, Inukai M, Suzuki Y, Matsusita S, Kuramoto K: [Clinical evaluation of finger blood pressure measurement devices for home-use] [Japanese] Nippon Ronen Igakkai Zasshi Jpn J Geriatr 1993;30: Iyriboz Y: Oscillometric finger blood pressure versus brachial auscultative blood pressure recording. J Fam Pract 1990;31: Thijs L, Staessen JA, Celis H, de Gaudemaris R, Imai Y, Julius S, Falard R: Reference values for self-recorded blood pressure. A meta-analysis of summary data. Arch Intern Med 1998;158:

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 16 Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-71-E)

More information

There is convincing evidence in clinical studies

There is convincing evidence in clinical studies AJH 1998;11:1413 1417 Reliability of Reporting Self-Measured Blood Pressure Values by Hypertensive Patients Thomas Mengden, Rosa Maria Hernandez Medina, Belen Beltran, Elena Alvarez, Karin Kraft, and Hans

More information

Home blood pressure measurement with oscillometric upper-arm devices

Home blood pressure measurement with oscillometric upper-arm devices REVIEW Home blood pressure measurement with oscillometric upper-arm devices R.L. Braam, Th. Thien *# Department of Internal Medicine, University Medical Centre St Radboud, PO Box 9101, 6500 HB Nijmegen,

More information

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

Hypertension is estimated to affect 43 to 56 million adults or 24% to 31% of the US

Hypertension is estimated to affect 43 to 56 million adults or 24% to 31% of the US REVIEW ARTICLE Home Blood Pressure Monitoring Steven A. Yarows, MD; Stevo Julius, MD, ScD; Thomas G. Pickering, MD, DPhil Hypertension is estimated to affect 43 to 56 million adults or 24% to 31% of the

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

The increasing awareness of hypertension as a serious

The increasing awareness of hypertension as a serious AJH 2002; 15:787 792 Evaluation of the Performance of a Wrist Blood Pressure Measuring Device With a Position Sensor Compared to Ambulatory 24-Hour Blood Pressure Measurements Sakir Uen, Burkard Weisser,

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Original article http://dx.doi.org/10.3345/kjp.2011.54.11.463 Korean J Pediatr 2011;54(11):463-469 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

In clinical practice, blood pressure (BP) treatment

In clinical practice, blood pressure (BP) treatment Original Paper Does Hidden Undercuffing Occur Among Obese Patients? Effect of Arm Sizes and Other Predictors of the Difference Between Wrist and Upper Arm Blood Pressures Hardik Doshi, MD; Alan B. Weder,

More information

Arm position and blood pressure: a risk factor for hypertension?

Arm position and blood pressure: a risk factor for hypertension? (3) 17, 389 395 & 3 Nature Publishing Group All rights reserved 09-9240/03 $25.00 www.nature.com/jhh ORIGINAL ARTICLE Arm position and blood pressure: a risk factor for hypertension? A Mourad, S Carney,

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009 42 Original article Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP, the Omron HBP T and the Pic Indolor Professional Roland

More information

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? O R I G I N A L A R T I C L E Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? I. Aksoy, J. Deinum, J.W.M. Lenders, Th. Thien *# Department of

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice.

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice. AJH 26; 19:81 89 BP Measurement Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice Tony C. Y. Pang and Mark A. Brown Background: To determine the extent of discrepancies between

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License Zheng DC, Liu CY, Amoore J, Mieke S, Murray A. Need for Re-validation of Automated Blood Pressure Devices for use in Unstable Conditions. In: 215 Computing in Cardiology Conference (CinC). 215, Nice, France:

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 223 Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol Paolo Verdecchia, Fabio Angeli, Roberto Gattobigio,

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

Data Collection Worksheet

Data Collection Worksheet Data Collection Worksheet 1. Has a doctor or nurse ever said that you have: High blood pressure or hypertension? [ ] No [ ] Yes [ ] Not Sure 1.a. If Yes, then at what age were you first told this? Age

More information

Importance Of Taking Manual Blood Pressure In Both Arms

Importance Of Taking Manual Blood Pressure In Both Arms Importance Of Taking Manual Blood Pressure In Both Arms Its importance to health is now well-recognised, in that high blood pressure can lead to Have a look at our slide show on how to take a blood pressure

More information

Blood Pressure Monitors Nissei

Blood Pressure Monitors Nissei We have made it easy for you to find a PDF Ebooks without any digging. And by having access to our ebooks online or by storing it on your computer, you have convenient answers with blood pressure monitors

More information

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon (2009) 23, 794 800 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Unreliable oscillometric blood pressure measurement: prevalence, repeatability

More information

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator MSc Thesis in Clinical Science and Technology Sara Rose Newell Department of Health Science

More information

Accuracy and reliability of wrist-cuff devices for selfmeasurement

Accuracy and reliability of wrist-cuff devices for selfmeasurement Original article 629 Accuracy and reliability of wrist-cuff devices for selfmeasurement of blood pressure Masahiro Kikuya a, Kenichi Chonan a,yutaka Imai a, Eiji Goto b and Masao Ishii c, on behalf of

More information

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER Naser KA. Teaching Hospital Peradeniya, Peradeniya, Sri Lanka Zawahir S

More information

greater than 90% of patients having blood pressure measured consistently according to the protocol.

greater than 90% of patients having blood pressure measured consistently according to the protocol. Clinical methods and pathophysiology 1 Reliability and validity of blood pressure measurement in the Secondary Prevention of Small Subcortical Strokes study Pablo E. Pérgola a, Carole L. White b, John

More information

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04) SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: MEASURING BLOOD PRESSURE - MANUAL Nursing DATE: REVIEWED: PAGES: 2/80 7/17 1 of 5 RESPONSIBILITY: RN, LPN, Patient Care Technician Multi-skilled

More information

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION*

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Progress in Clinical Medicine 1 ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Keishi ABE** Asian Med. J. 44(2): 83 90, 2001 Abstract: J-MUBA was a large-scale clinical

More information

Citation Acta medica Nagasakiensia. 1997, 42

Citation Acta medica Nagasakiensia. 1997, 42 NAOSITE: Nagasaki University's Ac Title Age and Gender Differences in White Author(s) Li, Zhang Ting Citation Acta medica Nagasakiensia. 1997, 42 Issue Date 1997-12-20 URL http://hdl.handle.net/10069/16086

More information

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home.

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home. Welch Allyn ABPM 6100S Comprehensive blood pressure monitoring that brings patient comfort home. Features and Benefits of the Welch Allyn ABPM 6100S Pictorial application instructions, artery markers and

More information

Home monitoring of blood glucose

Home monitoring of blood glucose Home Monitoring of Glucose and Blood Pressure JAMES R. TAYLOR, PharmD, CDE, and KENDALL M. CAMPBELL, MD, University of Florida, Gainesville, Florida Home monitoring of blood glucose and blood pressure

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT Blood Pressure Measurement Can Isik Electrical Engineering and Computer Science Department Syracuse University Syracuse, NY 13244 cisik@syr.edu Abstract Blood pressure is one of the fundamental vital signs,

More information

Accurate blood pressure (BP) measurement is essential

Accurate blood pressure (BP) measurement is essential AJH 2006; 19:468 474 BP Measurement A Comparison of Home Measurement and Ambulatory Monitoring of Blood Pressure in the Adjustment of Antihypertensive Treatment Teemu J. Niiranen, Ilkka M. Kantola, Risto

More information

Automated Blood Pressure Monitor

Automated Blood Pressure Monitor Automated Blood Pressure Monitor For Cardiac Stress and Exercise Testing SunTech Tango M2 Our cardiac stress blood pressure (BP) monitoring products give you the confidence to know that patient movement,

More information

Accuracy of Digital Arm and Wrist Manometers: Clinical Implications for the Dental Hygienist. Abstract

Accuracy of Digital Arm and Wrist Manometers: Clinical Implications for the Dental Hygienist. Abstract Accuracy of Digital Arm and Wrist Manometers: Clinical Implications for the Dental Hygienist Danielle Furgeson, RDH, EFDA, MS; Nancy Mickels-Foster, RDH, EFDA, EdM Introduction Thorough patient assessment

More information

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA Case 1 What should be your BP goal for an elderly (> 75 yrs of

More information

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Anthony J. Viera, MD, MPH, FAHA Department of Family Medicine Hypertension Research Program UNC School of Medicine Objectives Review limitations

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

Use and Interpretation of Home Blood Pressure Monitoring

Use and Interpretation of Home Blood Pressure Monitoring Use and Interpretation of Home Blood Pressure Monitoring BLOOD PRESSURE in year 2015 PARAMETERS TODAY Gianfranco Parati University of Milano-Bicocca & Ospedale S.Luca, IRCCS, Istituto Auxologico Italiano

More information

White coat and masked hypertension

White coat and masked hypertension White coat and masked hypertension Conflict of interest Support from Spacelabs, Microlife. Honoraria from Novartis, Elpen, Boeringer-Ingelheim, CANA, Lilly, MSD, Sanofi, Menarini, Ciezi, Astra-Zeneca.

More information

Blood Pressure Estimation Using Photoplethysmography (PPG)

Blood Pressure Estimation Using Photoplethysmography (PPG) Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4

More information

Blood Pressure Measurement Does it matter where you do it or how often?

Blood Pressure Measurement Does it matter where you do it or how often? Blood Pressure Measurement Does it matter where you do it or how often? James (Jim) M. Wright Coordinating Editor Cochrane Hypertension Group Conflict of Interest First half evidence based, no conflicts.

More information

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room. Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room Abstract Ahmad Mirdamadi (1), Mostafa Etebari (2) Original Article BACKGROUND:

More information

How accurate are your sphygmomanometers?

How accurate are your sphygmomanometers? Article introduction A&D Company, Limited 3-23-14, Higashi-Ikebukuro Toshima-ku, Tokyo 170-0013 JAPAN TEL: 81-3-5391-6132 FAX: 81-3-5391-1566 URL: http://www.aandd.jp February 14, 2017 How accurate are

More information

An Automated System for Blood Pressure Determination During Exercise

An Automated System for Blood Pressure Determination During Exercise An Automated System for Blood Pressure Determination During Exercise SEPHEN P. GLASSER, M.D., AND MAYNARD R. RAMSEY III, M.D., PH.D. with the technical assistance ofblanche Price SUMMARY Automation of

More information

The magnitude and duration of ambulatory blood pressure reduction following acute exercise

The magnitude and duration of ambulatory blood pressure reduction following acute exercise Journal of Human Hypertension (1999) 13, 361 366 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE The magnitude and duration of ambulatory

More information

doi: /bmj

doi: /bmj doi: 10.1136/bmj.322.7285.531 Blood pressure measuring devices: s of the European Society of Hypertension Eoin O Brien, Bernard Waeber, Gianfranco Parati, Jan Staessen, Martin G Myers on behalf of the

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials. AJH 1995; 8:311-315 BRIEF COMMUNICATIONS Lack of Placebo Effect on Ambulatory Blood Pressure Giuseppe Mancia, Stefano Omboni, Gianfranco Parati, Antonella Ravogli, Alessandra Villani, and Alberto Zanchetti

More information

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008 Analytical methods and statistical analyses 285 PA.NET International Quality Certification Protocol for blood pressure monitors Stefano Omboni a,b, Carlo Costantini a,b,c, Claudio Pini b,d, Roberto Bulegato

More information

Disclosures. Funding: CIHR, HSF, AIHS, UHF. Research: Novo Nordisk, CVRx, Valencia. Consulting: Valeant

Disclosures. Funding: CIHR, HSF, AIHS, UHF. Research: Novo Nordisk, CVRx, Valencia. Consulting: Valeant Hypertension Update Raj Padwal Associate Professor General Internal Medicine and Clinical Pharmacology Director, Hypertension Clinic University of Alberta Disclosures Funding: CIHR, HSF, AIHS, UHF Research:

More information

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood SPECIAL ARTICLE Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Eoin O Brien a,thomas

More information

CLINICAL SCIENCE. Angela M. G. Pierin a, Edna C. Ignez a, Wilson Jacob Filho b, Alfonso Júlio Guedes Barbato b, Décio Mion Jr. b

CLINICAL SCIENCE. Angela M. G. Pierin a, Edna C. Ignez a, Wilson Jacob Filho b, Alfonso Júlio Guedes Barbato b, Décio Mion Jr. b CLINICS 2008;63(1):43-50 CLINICAL SCIENCE BLOOD PRESSURE MEASUREMENTS TAKEN BY PATIENTS ARE SIMILAR TO HOME AND AMBULATORY BLOOD PRESSURE MEASUREMENTS Angela M. G. Pierin a, Edna C. Ignez a, Wilson Jacob

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

Indirect ambulatory blood pressure monitoring devices

Indirect ambulatory blood pressure monitoring devices AJH 1992;5:880-886 Comparison of Ambulatory and Clinic Blood Pressure and Heart Rate in Older Persons With Isolated Systolic Hypertension Gale H. Rutan, Robert H. McDonald, and Lewis H. Kuller We compared

More information

Validation Study Result Form

Validation Study Result Form Validation Study Result Form Please complete Section 1 to Section 3 of this form and return it to dabl Educational Trust with copies of the validation plots. The requirements for each of these sections

More information

INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS

INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS INCREASED PULSE PRESSURE AND SYSTOLIC x HEART RATE DOUBLE PRODUCT AND CARDIOVASCULAR AUTONOMIC NEUROPATHY IN TYPE 2 DIABETIC PATIENTS A.J. Scheen, J.C. Philips, M. Marchand Division of Diabetes, Nutrition

More information

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the A comparison of blood pressure measurement over a sleeved arm versus a bare arm Grace Ma MD, Norman Sabin MD, Martin Dawes MBBS MD @ See related article page 591 DOI:10.1503/cmaj.070975 Abstract Background:

More information

Special articles from the ESH working group on blood pressure monitoring 23

Special articles from the ESH working group on blood pressure monitoring 23 Special articles from the ESH working group on blood pressure monitoring 23 European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices

More information

Blood Pressure Measurement in SPRINT

Blood Pressure Measurement in SPRINT Blood Pressure Measurement in SPRINT Karen C. Johnson, MD, MPH, FAHA Vice Chair, SPRINT Steering Committee University of Tennessee Health Science Center, Department of Preventive Medicine For the SPRINT

More information

Results of the recently published Systolic Blood

Results of the recently published Systolic Blood Eπίκαιρο Άρθρο Blood Pressure goals of hypertension therapy in 2017: Which patient and which goal? Vasilios Papademetriou, MD, PhD, FACC, FAHA, FASH ABSTRACT The goal of hypertension treatment in 2017

More information

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

PDF hosted at the Radboud Repository of the Radboud University Nijmegen PDF hosted at the Radboud Repository of the Radboud University Nijmegen The following full text is a publisher's version. For additional information about this publication click this link. http://hdl.handle.net/2066/53625

More information

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods Vol. 2, Issue 1, pages 31-36 ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY By Alejandro de la Sierra, MD Luis M. Ruilope, MD Hypertension Units, Hospital Clinico, Barcelona & Hospital 12 de Octubre,

More information

Self-measurement of blood pressure

Self-measurement of blood pressure Bulletin of the World Health Organisation, 66 (2): 155-159 (1988) Self-measurement of blood pressure WORLD HYPERTENSION LEAGUE' Although experience is still limited and more research is needed, the World

More information

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 64 CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 4.1 GENERAL This chapter presents the methodologies that are usually adopted for the measurement of blood pressure, heart rate and pulse

More information

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right Anthony J. Viera, MD, MPH, FAHA Professor and Chair Objectives Review limitations of office BP in making a correct diagnosis of hypertension

More information

SIGNAL PROCESSING and KNOWLEDGE ACQUISITION in OBJECTIVE BLOOD PRESSURE MEASUREMENT with CONVENTIONAL CLINICAL SPHYGMOMANOMETRY

SIGNAL PROCESSING and KNOWLEDGE ACQUISITION in OBJECTIVE BLOOD PRESSURE MEASUREMENT with CONVENTIONAL CLINICAL SPHYGMOMANOMETRY SIGNAL PROCESSING and KNOWLEDGE ACQUISITION in OBJECTIVE BLOOD PRESSURE MEASUREMENT with CONVENTIONAL CLINICAL SPHYGMOMANOMETRY Ventzas, D. 1 - TEI Larisa Angelopoulos A. Biomedical Eng. Cardiologist,

More information

Managing Blood Pressure: It Takes a Team

Managing Blood Pressure: It Takes a Team Managing Blood Pressure: It Takes a Team Module Two How to Take Blood Pressure October 2011 MODULE TWO: How to Take Blood Pressure Module Two: HOW TO TAKE BLOOD PRESSURE Objectives Demonstrate proper technique

More information

High Blood Pressure Blood Pressure Monitoring at Home -- F...

High Blood Pressure Blood Pressure Monitoring at Home -- F... Return to Web version High Blood Pressure Blood Pressure Monitoring at Home Why should I measure my blood pressure at home and keep a record of it? Measuring your blood pressure at home and keeping a record

More information

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

!#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 !"#$ Blood Pressure Monitor Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 Introduction

More information

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License Pan F, Zheng DC, He PY, Murray A. Does the Position or Contact Pressure of the Stethoscope Make Any Difference to Clinical Blood Pressure Measurements: An Observational Study. Medicine 214, 93(29). Copyright:

More information

BP210 BP250. Automatic Blood Pressure Monitor. Medical Diagnostic Device

BP210 BP250.  Automatic Blood Pressure Monitor. Medical Diagnostic Device Medical Diagnostic Device BP210 BP250 Automatic Blood Pressure Monitor Our best-selling easy-to-use ACCUNIQ Sphygmomanometer ensures fast and accurate results www.accuniq.com ACCUNIQ Brand Introduction

More information

Assessment of techniques for measurement of blood pressure in infants and children

Assessment of techniques for measurement of blood pressure in infants and children Archives of Disease in Childhood, 1973, 48, 932. Assessment of techniques for measurement of blood pressure in infants and children A. M. ELSEED, E. A. SHINEBOURNE, and M. C. JOSEPH From the Paediatric

More information

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines TITLE: Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines DATE: 25 September 2012 CONTEXT AND POLICY ISSUES A mercury manual sphygmomanometer

More information

THE ACCURATE ASSESSMENT

THE ACCURATE ASSESSMENT ORIGINAL INVESTIGATION Consistency of Blood Pressure Differences Between the Left and Right s Kazuo Eguchi, MD; Mona Yacoub, MD; Juhee Jhalani; William Gerin, PhD; Joseph E. Schwartz, PhD; Thomas G. Pickering,

More information

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL Douglas G. Kelling Jr., MD Carmella Gismondi-Eagan, MD, FACP George C. Monroe, III, MD Revised, April 8, 2012 The information contained in this protocol

More information

BP Hour Ambulatory Blood Pressure Monitor. Medical Diagnostic Device

BP Hour Ambulatory Blood Pressure Monitor. Medical Diagnostic Device Medical Diagnostic Device BP800 24Hour Ambulatory Blood Pressure Monitor Continuous accurate readings over a 24hour period to monitor patients during their normal daily routines www.accuniq.com Brand Introduction

More information

Blood Pressure Protocol Training

Blood Pressure Protocol Training Blood Pressure Protocol Training May 31, 2018 Pat Spier, RN-BC, PCMH-CCE Barb Rice, RN-BC BCBS of North Dakota Fargo, ND 2013 Template and icons provided by The Advisory Board Company. Hypertension Education

More information

Ambulatory blood pressure monitoring has shown that in

Ambulatory blood pressure monitoring has shown that in Difference Between Clinic and Daytime Blood Pressure Is Not a Measure of the White Coat Effect Gianfranco Parati, Luisa Ulian, Cinzia Santucciu, Stefano Omboni, Giuseppe Mancia Abstract The purpose of

More information

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015 Original article 1 Clinical practice of ambulatory versus home blood pressure monitoring in hypertensive patients Jorge A. Paolasso, Florencia Crespo, Viviana Arias, Eduardo A. Moreyra, Ariel Volmaro,

More information

The introduction of oscillometric automated blood pressure

The introduction of oscillometric automated blood pressure Atrial Fibrillation Impact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring Nikolaos Pagonas,* Sven Schmidt,* Jörg Eysel, Friederike Compton, Clemens Hoffmann, Felix Seibert,

More information

THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM

THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM Cader FA, Das A, Kahhar MA Dhaka Medical College Hospital March 2015 Dr Aaysha Cader MBBS,

More information

Alternatives to the mercury sphygmomanometer

Alternatives to the mercury sphygmomanometer Original Article Alternatives to the mercury sphygmomanometer Susan Buchanan a, *, Peter Orris b and Joshua Karliner c a Environmental and Occupational Health Sciences Division, University of Illinois

More information