Vascular and Interventional Radiology Original Research

Size: px
Start display at page:

Download "Vascular and Interventional Radiology Original Research"

Transcription

1 Vascular and Interventional Radiology Original Research Apfaltrer et al. High-Pitch Versus Standard-Pitch CTA of the Aorta Vascular and Interventional Radiology Original Research Paul Apfaltrer 1,2 E. Lexworth Hanna 1 U. Joseph Schoepf 1,3 J. Reid Spears 1 Stefan O. Schoenberg 2 Christian Fink 2 Rozemarijn Vliegenthart 1,4 Apfaltrer P, Hanna EL, Schoepf UJ, et al. Keywords: aorta, CT angiography, high-pitch CT angiography, radiation dose DOI: /AJR Received January 22, 2012; accepted after revision March 26, U. J. Schoepf is a consultant to and receives research support from Bayer-Schering, Bracco, GE Healthcare, Medrad, and Siemens Healthcare. 1 Department of Radiology and Radiological Science, Medical University of South Carolina, Ashley River Tower, 25 Courtenay Dr, MSC 226, Charleston, SC Address correspondence to U. J. Schoepf (schoepf@musc.edu). 2 Institute of Clinical Radiology and Nuclear Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany. 3 Department of Medicine, Division of Cardiology, Medical University of South Carolina, Charleston, SC. 4 Department of Radiology, Center for Medical Imaging, North East Netherlands, University Medical Center Groningen/University of Groningen, Groningen, The Netherlands. CME This article is available for CME credit. AJR 2012; 199: X/12/ American Roentgen Ray Society Radiation Dose and Image Quality at High-Pitch CT Angiography of the Aorta: Intraindividual and Interindividual Comparisons With Conventional CT Angiography OBJECTIVE. The objective of our study was to evaluate radiation dose and quantitative image quality parameters at high-pitch CT angiography (CTA) of the aorta compared with conventional CTA. MATERIALS AND METHODS. We studied the examinations of 110 patients (65 men and 45 women; mean age ± SD, 64 ± 15 years) who had undergone CTA of the entire aorta on a second-generation dual-source CT system; 50 examinations were performed in highpitch mode. The mean arterial attenuation, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figure of merit (FOM) were calculated for the high-pitch CTA and conventional CTA groups. Radiation exposures were compared. RESULTS. All studies were considered of diagnostic quality. At high-pitch CTA, the mean tube voltage and tube current exposure time product were 118 ± 7 kv (SD) and 197 ± 78 mas compared with 120 ± 1 kv and 258 ± 78 mas, respectively, at conventional CTA (p < 0.05). The mean volume CT dose index, dose-length product, and effective dose were 8.1 ± 2.4 mgy, ± mgy cm, and 9.6 ± 3.0 msv at high-pitch CTA and 18.3 ± 7.7 mgy, ± mgy cm, and 19.8 ± 8.2 msv at conventional CTA (p < 0.001). Attenuation was similar for both protocols, whereas significantly less contrast medium was injected for high-pitch CTA than for standard-pitch CTA (87.3 ± 16 ml vs 97.9 ± 16 ml, respectively; p < 0.01). The SNR and CNR were significantly lower in the high-pitch CTA examinations (p < 0.01), whereas the FOM was nonsignificantly higher. Twenty patients underwent both high-pitch CTA and conventional CTA, with a 45% reduction in radiation dose (p < 0.001). CONCLUSION. High-pitch CTA of the aorta yields 45 50% reduction of radiation exposure as well as contrast medium savings with maintained vessel attenuation. T he technical advances of MDCT technology have enabled acquisition of image data about the entire torso within seconds. The clinical success of CT angiography (CTA) is based on precise synchronization of image acquisition with optimal vascular contrast medium attenuation, high spatial resolution, and fast image acquisition times [1, 2]. More detectors, shorter CT rotation times, and narrower section profiles have resulted in progressively shorter image acquisition times, improved volume coverage, and substantial improvements in image quality. Thus, MDCT studies can be performed 3 20 times faster than studies performed using the fastest single-detector CT systems [3, 4]. Speed is essential to limit motion artifacts by reducing the time required for the patient to hold his or her breath. The recently introduced second-generation dual-source CT (DSCT) system is equipped with a high-pitch data acquisition mode with pitch values of up to 3.4 [5]. With a highpitch dual-helical CTA technique, datasets of the thorax and abdomen can be acquired in less than 2 seconds [6] with low radiation dose and gapless coverage of the entire patient volume. However, imaging patients at high pitch values has raised questions about the subsequent increase in CT image noise [7] and the potential to outrun the contrast medium bolus, resulting in insufficient and therefore nondiagnostic vascular attenuation. In prior studies, investigators have reported the feasibility of high-pitch acquisitions of the thorax [6, 8] and abdomen [9]. To date, a direct comparison of thoracoabdominal CTA performed with a high-speed, high-pitch technique and CTA performed with conventional 1402 AJR:199, December 2012

2 High-Pitch Versus Standard-Pitch CTA of the Aorta techniques is limited to an initial report that did not correct for differences in patient-related factors [10]. We aimed to investigate with interindividual and intraindividual comparisons the quantitative image quality, imaging parameters, and potential radiation dose reduction of high-pitch CTA of the aorta compared with conventional CTA. Materials and Methods Subject Population In a retrospective search of the institution s radiology information system (RIS), 110 patients (65 men, 45 women; mean age ± SD, 64 ± 15 years) who had undergone thoracoabdominal CTA using second-generation DSCT between September 2009 (installation date) and March 2011 were identified. During the study period, 50 patients underwent high-pitch CTA with an identical imaging protocol. The remaining 60 patients underwent standard-pitch CTA, which was performed using an identical standardized imaging protocol. For patients who had undergone high-pitch CTA, the RIS was searched for prior CTA studies performed using a standard-pitch technique for intraindividual comparisons. If a patient had undergone more than one standard-pitch CTA examination, the study performed closest to the date of the highpitch CTA study was selected for comparison. In total, 20 of the 50 patients had previously (January 2008 March 2011) undergone standard-pitch CTA. The indications for thoracoabdominal CTA included suspected aortic syndrome (n = 12), postoperative follow-up after thoracoabdominal vascular surgery or endovascular aneurysm repair (n = 35), and follow-up of aneurysm (n = 37) or dissection (n = 20). Age, height, weight, and body mass index (BMI) were recorded for all patients. Because of the retrospective nature of the study protocol, the institutional ethics review board waived the need for informed consent. CT Procedures All studies were acquired using a second-generation DSCT system (Somatom Definition Flash, Siemens Healthcare). This system uses two tubedetector pairs with a mechanical offset of nearly 90 ; one detector covers a 50-cm FOV and the second detector covers a smaller, 33-cm, FOV. The following image acquisition parameters were used: two physical mm channel detector arrays, effective slice collimation of mm by using a z-flying focal spot technique, and 0.28-second gantry rotation time. The pitch was 3.4 for the high-pitch CTA examinations and 0.8 for the standard-pitch CTA examinations. Selection of the high-pitch versus the standard-pitch image acquisition protocol was at the discretion of the attending radiologist of the day. The acquired range covered the area above the aortic arch to the common iliac arteries. All patients had a cannula placed in the antecubital fossa. A dual-syringe injection system (Stellant D, Medrad) was used to administer 350 mg I/mL of iohexol (Omnipaque, GE Healthcare). Personalized contrast medium injection protocol software (P3T version 1.0, Medrad), customizing a triphasic injection protocol to each patient and procedure, was used for all 110 subjects. A test bolus was used to determine the circulation time of the contrast material. The test bolus was a 20-mL bolus of contrast medium followed by a 40-mL chaser bolus of saline solution administered at 5 ml/s. The personalized contrast medium injection protocol software uses patient weight, study duration, and con- Fig. 1 Examples of aortoiliac CT attenuation measurements in 55-year-old woman with suspected aortic syndrome. A, CT angiography images show exact locations of measurements in ascending aorta (white arrow), thoracic descending aorta at level of pulmonary trunk (blue arrow) and at level of diaphragm (yellow arrow), abdominal aorta at level of renal arteries (black arrow), and right iliac artery (red arrow). B, Three-dimensional volume-rendered image reconstruction shows excellent image quality at all described levels. AJR:199, December

3 Apfaltrer et al. trast medium concentration and accounts for the attenuation peak and time-to-peak of a timing bolus acquisition with an ROI placed in the ascending aorta. Studies were reconstructed at a section thickness of 1.5 mm with a 1-mm overlap using a vascular image reconstruction algorithm. The total amount of iodine injected and the iodine delivery rate were calculated using the contrast medium volume, contrast concentration, and injection rate. Radiation Dose Volume CT dose index (CTDI vol ) values were indicated in the dose report of the CT system provided for each CT study, and individual radiation dose was estimated using the dose-length product (DLP) given by the CT system. The effective radiation dose delivered at thoracoabdominal CTA was calculated with a method proposed by the European Working Group for Guidelines on Quality Criteria for CT [11]. Because a combination of chest, abdominal, and pelvic acquisitions was performed for thoracoabdominal CTA, the mean of these region-specific conversion coefficients (k = msv / mgy cm) was used as previously described [6, 12]. Qualitative and Quantitative Evaluations of Arterial Contrast Medium Attenuation A radiologist with 5 years of experience in CTA assessed the image quality for evaluating the aortoiliac system as diagnostic or nondiagnostic. The quantitative image analysis was performed on reconstructed transverse sections. Aortoiliac CT attenuation was calculated in Hounsfield units by measuring a circular region of interest (ROI) in the TABLE 1: Characteristics of Patients in the High-Pitch CT Angiography (CTA) and Standard-Pitch CTA Groups Characteristic High-Pitch CTA Group (n = 50) Standard-Pitch CTA Group (n = 60) p Sex, no. (%) of patients Male 29 (58) 36 (60) Female 21 (42) 24 (40) Age (y) a 61.1 ± ± 11.8 < 0.05 Height (cm) a ± ± 10.1 NS Weight (kg) a 81.0 ± ± 28.0 NS Body mass index (kg/m 2 ) a 26.9 ± ± 6.5 NS Indication for CTA, no. (%) of patients Suspected aortic syndrome 3 (6) 9 (15) Known aneurysm 14 (28) 23 (38) Known dissection 13 (26) 7 (12) Follow-up imaging after aortic repair or grafting 18 (36) 17 (28) Stenosis or atherosclerosis 1 (2) 3 (5) Other 1 (2) 1 (2) Note NS = not significant. a Mean ± SD. center of the vessel at the following seven different levels within the arterial system (Fig. 1): ascending aorta, aortic arch, thoracic descending aorta at the level of pulmonary trunk, thoracic descending aorta at the diaphragm, abdominal aorta at the level of the renal arteries, abdominal aorta above the bifurcation, and left and right common iliac arteries (both measurements averaged to one measurement, IlA). Each ROI was drawn as large as the vessel lumen diameter allowed while carefully avoiding atherosclerotic plaques of the vessel wall as previously described [13] (Fig. 1). Attempts were made to select ROI areas of at least 40 mm 2 for the aorta and at least 20 mm 2 for the iliac arteries. Mean arterial attenuation was calculated as the average of the mean attenuation values of the ROIs. At each of the described levels, an ROI was also placed in an adjacent muscle (Fig. 1D). Image noise was defined as the SD of the CT attenuation value of the adjacent muscle ROI [14]. Signal-to-noise ratio (SNR) was calculated as the mean attenuation of the artery divided by the image noise per level. Furthermore, the contrast-to-noise ratio (CNR) was calculated as the mean attenuation of the artery minus the mean attenuation of the muscle divided by image noise (mean HU artery mean HU muscle / image noise). SNR and CNR were calculated similar to previously published studies [15, 16] as follows: and SNR = HU vessel / noise CNR = (HU vessel HU muscle ) / noise. To account for differences in the radiation exposure of the two protocols, the CNR of each artery was normalized by effective dose (ED) using the figure of merit (FOM) as previously described [14]: FOM = CNR 2 / ED. Statistical Analysis Acquisition parameters, contrast medium parameters, quantitative image quality parameters, and radiation dose estimations were compared for high-pitch CTA and standard-pitch CTA. Continuous variables were expressed as means ± SD and categoric variables, as frequencies or percentages. Statistical significance of differences in continuous variables was assessed by an independent-samples Student t test for interindividual comparison and by a paired-samples Student t test for intraindividual comparison. For categoric variables, the chi-square test was used. The influence of BMI on image noise was evaluated by linear regression analysis. Statistical analyses were performed using commercially available software (SPSS, release 17, SPSS). All p values were two-sided and a p value of < 0.05 was considered statistically significant. Results Table 1 shows the characteristics of the study population. The majority of the study population was male (58%). Patients who had undergone high-pitch CTA were significantly younger than those who had undergone standard-pitch CTA (mean age ± SD, 61.1 ± 15.3 vs 67.1 ± 11.8 years, p < 0.05), but their BMI was nonsignificantly lower (26.9 ± 4.5 vs 29.1 ± 6.5 kg/m 2, respectively; p = not significant). The CT indications for both groups were very similar (p = not significant). The NS NS 1404 AJR:199, December 2012

4 High-Pitch Versus Standard-Pitch CTA of the Aorta TABLE 2: Scanning Parameters for High-Pitch CT Angiography (CTA) and Standard-Pitch CTA Scanning Parameter High-Pitch CTA Group (n = 50) Standard-Pitch CTA Group (n = 60) p Tube voltage setting, no. (%) of patients NS 80 kv 1 (2) 100 kv 2 (4) 120 kv 47 (94) 60 (100) mas ± 80.7 a ± 76.3 a < CTDI vol (mgy) 8.1 ± 2.4 a 18.3 ± 7.7 a < Scan length (cm) 69.0 ± 6.6 a 63.6 ± 5.8 a < DLP (mgy cm) ± a ± a < Effective dose (msv) 9.6 ± 3.0 a 19.8 ± 8.2 a < Contrast volume (ml) 87.3 ± 15.7 a 97.9 ± 15.6 a < 0.01 Iodine delivery rate (g/s) 1.67 ± 0.22 a 1.68 ± 0.19 a NS Iodine amount (g) 31.4 ± 6.2 a 34.3 ± 5.9 a < 0.05 Note NS = not significant, CTDI vol = CT dose index, DLP = dose-length product. a Mean ± SD. Fig. 2 Interindividual comparison: 67-yearold woman (A) who underwent CT angiography (CTA) for evaluation of aortic aneurysm and 58-yearold woman (B) who underwent CTA for evaluation of aortic dissection. A and B, Both conventional (A) and high-pitch (B) CTA images show excellent image quality. However, patient undergoing conventional CTA technique (A) received greater volume of contrast medium volume and radiation dose compared with patient undergoing high-pitch technique (B) (100 vs 85 ml, respectively; 202 vs 154 mas). tube current exposure time product and contrast medium volume were significantly lower in the high-pitch CTA cohort and the coverage length was somewhat longer (Table 2). The iodine delivery rate was similar for both cohorts. The CTDI vol, DLP, and ED were all significantly lower in the high-pitch CTA cohort than in the standard-pitch cohort. The ED in the high-pitch CTA group was 9.6 ± 3.0 msv (mean ± SD) versus 19.8 ± 8.2 msv in the standard-pitch CTA group (p < 0.001). All studies were considered of diagnostic quality for their respective indication. An example illustrating the comparative image quality of studies obtained in the high-pitch mode and those obtained in the standard-pitch mode is shown in Figure 2. The minimum arterial attenuation differed by level and ranged from 168 HU (for Ila) to 229 HU (for the thoracic descending aorta at the level of the pulmonary trunk) in high-pitch CTA studies and from 89 HU (for the abdominal aorta above the bifurcation) to 134 HU (for the aortic arch) in standardpitch CTA studies. The mean arterial attenuation at high-pitch CTA was, overall, similar to that of standard-pitch CTA (354 ± 95 vs 332 ± 61 HU, respectively; p > 0.05), although there was a trend toward higher attenuation of highpitch CTA studies (not significant). The noise level was significantly higher in the high-pitch CTA group (23.2 ± 6.6 vs 16.7 ± 2.9 HU; p < 0.001). The SNR and CNR were significantly lower at high-pitch CTA than standard-pitch CTA (16.4 ± 5.6 vs 20.7 ± 5.5 [p < 0.001] and 14.4 ± 5.0 vs 17.8 ± 5.2 [p < 0.01], respectively). However, the FOM, a measure of image quality independent of radiation dose, tended to be higher for high-pitch CTA studies than for the standard-pitch CTA studies (32.1 ± 50.3 vs 21.4 ± 15.2), although this difference did not reach statistical significance. The arterial attenuation, noise, SNR, CNR, and FOM are provided by level in Table 3. Of the patients who had undergone highpitch CTA, 20 had also undergone at least one standard-pitch CTA examination using DSCT (Table 4). The mean interval between the two studies was 353 ± 257 days (range, days). The use of the high-pitch acquisition technique reduced ED by 45 50% (p < 0.001) within the same patient. Although on average 20 ml less iodinated contrast medium and a significantly lower amount of iodine were injected during high-pitch CTA (30.0 ± 6.1 vs 36.4 ± 4.2 g; p < 0.01), differences in arterial attenuation were nonsignificant. However, noise levels were higher for high-pitch CTA than for standard-pitch CTA. An intraindividual comparison of a patient who underwent both high-pitch and standard-pitch CTA examinations is shown in Figure 3. Discussion The purpose of this study was to compare image quality and radiation dose estimates AJR:199, December

5 Apfaltrer et al. TABLE 3: Quantitative Image Quality Parameters of the High-Pitch CT Angiography (CTA) and Standard-Pitch CTA Groups, by Level Level Arterial Attenuation (HU) Noise (HU) SNR CNR FOM Ascending aorta High-pitch CTA 349 ± ± ± ± ± 33.6 Standard-pitch CTA 337 ± ± 3.4 a 20.8 ± 5.6 a 17.9 ± 5.1 a 20.5 ± 14.9 Aortic arch High-pitch CTA 366 ± ± ± ± ± Standard-pitch CTA 333 ± 60 b 14.8 ± 3.0 a 23.7 ± 7.3 a 20.4 ± 6.7 c 27.7 ± 23.1 Thoracic descending aorta at the level of the pulmonary trunk High-pitch CTA 351 ± ± ± ± ± 39.0 Standard-pitch CTA 332 ± ± 3.5 a 20.7 ± 6.1 a 17.8 ± 5.5 c 20.8 ± 15.2 Thoracic descending aorta at diaphragm High-pitch CTA 348 ± ± ± ± ± 70.6 Standard-pitch CTA 326 ± ± 3.7 a 21.4 ± 6.7 c 18.6 ± 6.1 b 23.2 ± 19.5 Abdominal aorta at the level of the renal arteries High-pitch CTA 352 ± ± ± ± ± 31.1 Standard-pitch CTA 329 ± ± 3.9 a 19.8 ± 6.0 a 16.9 ± 5.8 c 20.1 ± 21.3 Abdominal aorta above the bifurcation High-pitch CTA 356 ± ± ± ± ± 35.7 Standard-pitch CTA 333 ± ± 4.5 a 19.3 ± 6.4 c 16.2 ± 6.2 c 18.2 ± 13.6 Left and right common iliac arteries High-pitch CTA 358 ± ± ± ± ± 42.7 Standard-pitch CTA 334 ± ± 4.7 a 19.7 ± 6.7 c 16.6 ± 6.4 b 19.4 ± 16.1 Note Values are means ± SD. SNR = signal-to-noise ratio, CNR = contrast-to-noise ratio, FOM = figure of merit. a p < b p < c p < 0.01 of high-pitch CTA performed on a DSCT system of the entire aorta with standard-pitch CTA. Our results show that in comparison with the conventional technique, high-pitch thoracoabdominal CTA yields a radiation dose reduction of 45 50%, at maintained diagnostic quality. Also, a considerably lower amount of contrast medium was needed for the same arterial attenuation. Second-generation DSCT systems, which have double the number of detectors (and, hence, double the volume coverage) of firstgeneration DSCT systems, offer high-pitch or Flash mode acquisition at pitch values of up to 3.4 [5]. Increasing the pitch decreases the time necessary to cover the entire patient volume. This high-pitch mode has primarily been investigated for coronary CTA. In addition to limiting radiation exposure to a single cardiac cycle, the high-pitch mode eliminates overlapping volume coverage of sequential transverse sections, enabling coronary CTA examinations at EDs of less than 1 msv in suitable patients [17, 18]. However, the principle can be readily applied to thoracoabdominal CTA studies, where data are still limited. The potential application of this technique for the evaluation of the thoracoabdominal aorta is attractive because of the potential radiation dose reduction that can be achieved in patients with aortic disease who often need to undergo repeated CTA at regular intervals. In the current investigation all imaging studies were considered of diagnostic quality for their indication even though the noise level with high-pitch CTA was found to be higher. The maintained diagnostic quality of high-pitch CTA despite lower radiation dose observed here is in agreement with recent investigations evaluating high-pitch acquisition of the thorax [6] and abdomen [9] with diagnostic image quality in 83% and 100% of cases, respectively. Recently, Beeres et al. [10] showed significantly better subjective image quality using the high-pitch acquisition technique compared with conventional CTA. The investigators focused especially on the ability of CTA in imaging the aortic annulus, coronary arteries, and ascending aorta without motion artifacts. A higher aortic attenuation value was achieved with lower contrast medium injection volumes in the high-pitch CTA group than with the standard-pitch CTA group. However, the iodine delivery rates in grams per second were almost identical. One potential explanation for the differences in aortic attenuation favoring the high-pitch mode could be that the patients undergoing highpitch CTA had a lower body weight. However, a difference in body weight was not found in our study. The BMI of the patients who had undergone high-pitch CTA was only nonsignificantly lower than that of patients who had undergone standard-pitch CTA (26.9 ± 4.5 vs 29.1 ± 6.5 kg/m 2, respectively). Additionally, we specifically investigated a group of patients who had undergone both high-pitch and standard-pitch CTA to account for patient-related differences that could potentially (partly) explain differences in CT protocol, amount of 1406 AJR:199, December 2012

6 High-Pitch Versus Standard-Pitch CTA of the Aorta TABLE 4: Results of 20 Patients Who Underwent Both High-Pitch and Standard-Pitch CT Angiography (CTA) Parameter High-Pitch CTA Standard-Pitch CTA p Indication, no. (%) of patients < 0.05 Suspected aortic syndrome 1 (5) Known aneurysm 4 (20) 5 (25) Known dissection 8 (40) 7 (35) Follow-up imaging after aortic repair or grafting 8 (40) 7 (35) 120 kv 20 (100) 20 (100) NS Tube current exposure time product (mas) ± 61.7 a ± 56.3 a < 0.01 CTDI (mgy) 8.4 ± 1.6 a 16.5 ± 4.1 a < Scan length (cm) 69.0 ± 5.1 a 63.6 ± 6.5 a < 0.01 DLP (mgy cm) ± a ± a < Effective dose (msv) 10.0 ± 2.0 a 18.1 ± 5.1 a < Contrast volume (ml) 83.6 ± 15.3 a ± 12.5 a < 0.01 Iodine amount (g) 30.0 ± 6.1 a 36.4 ± 4.2 a < 0.01 Averaged arterial attenuation (HU) b 338 ± 57 a 316 ± 73 a NS Averaged image noise (HU) b 21.9 ± 5.7 a 15.8 ± 3.7 a < Averaged SNR b 16.4 ± 4.9 a 21.3 ± 6.8 a < 0.01 Averaged CNR b 14.2 ± 4.3 a 18.1 ± 6.5 a < 0.05 Averaged FOM b 24.2 ± 15.5 a 23.7 ± 16.2 a NS Note NS = not significant, CTDI = CT dose index, DLP = dose-length product, SNR = signal-to-noise ratio, CNR = contrast-to-noise ratio, FOM = figure of merit. a Mean ± SD. b Averaged over the seven measurement levels. Fig. 3 Intraindividual comparison of initial CT angiography (CTA) and repeated CTA for followup of ascending aortic aneurysm in 41-year-old woman. A and B, Both conventional standardpitch (A) and high-pitch (B) CTA images show excellent image quality. However, for standardpitch CTA, amount of contrast medium and radiation dose were higher than for highpitch CTA (130 vs 97 ml of contrast medium and 245 vs 152 mas, respectively). iodinated contrast material, and CT attenuation. The results for that group were in agreement with the findings for the overall cohort. Thus, we do not believe that differences in patient weight explain the difference in aortic CT attenuation in our study. Rather, one might speculate that shorter acquisition times combined with appropriate contrast medium bolus timing enable capturing the vascular system during a phase of higher, more concentrated intraluminal contrast medium attenuation, despite lower contrast medium volumes, than with slower image acquisition and possible greater dilution of intravascular contrast medium. In patients undergoing abdominal, pelvic, or abdominopelvic imaging, Sahani et al. [19] reported similar image quality using high-pitch or low-pitch technique. Factors affected by the pitch include effective section thickness, image noise, beam-hardening artifacts, image acquisition speed, and radiation dose [19, 20]. Lowering the pitch reduces the slice-broadening effect, minimizes helical artifacts, and lowers image noise [21, 22]. These reported results correlate with our results because we also found significantly higher SNR and CNR in the conventional low-pitch CTA group (p < 0.01). However, lowering the pitch also increases image acquisition time and radiation dose [19]. To account for differences in the radiation exposure of the two protocols, the CNR of each arterial segment and the overall CNR were normalized by the ED using the FOM as previously described [14]. Although the CNR with highpitch CTA was significantly lower than that with standard-pitch CTA, the FOM was higher with high-pitch CTA at all arterial levels but these differences did not reach statistical significance. Thus, one might argue that high-pitch AJR:199, December

7 Apfaltrer et al. CTA has the tendency to yield higher image quality per unit of radiation exposure. A slight increase in image noise may be acceptable as long as diagnostic information is not compromised, in keeping with the as low as reasonably achievable (ALARA) principle. The high-pitch acquisition technique resulted in significant reductions of CTDI vol, DLP, and ED compared with the conventional CTA technique (p < 0.001); these results are in line with recently published data on the high-pitch acquisition technique [10]. This was true although no significant difference in mean tube voltage and mean BMI of both investigated groups was found (both p > 0.05). Despite these similar patient cohorts, the mean tube current exposure time product for the high-pitch CTA studies was significantly lower compared with that of the conventional CTA studies. However, we think that the effect of this reduction in tube current time product does not solely explain the radiation dose reduction of up to 45 50% achieved with the high-pitch technique; rather, we believe that an interaction of various attributes of this new acquisition technique is responsible for the dose reduction. Effective reduction of patient exposure of ionized radiation is of particular importance in patients with aortic disease who tend to undergo regular follow-up CT studies for disease surveillance and often incur high cumulative radiation doses. In our patient cohort, 20 individuals had undergone at least one high-pitch CTA and one conventional CTA examination within a mean of about 1 year for surveillance of aneurysms, dissections, or postaortic repairs. By using the high-pitch acquisition techniques, radiation doses could be reduced by up to 45% (p < 0.001) within the same patient while differences in arterial attenuation were nonsignificant. Because the potential detrimental effects of radiation are assumed to be linearly related to both dose and frequency of exposure [23, 24], low radiation techniques such as the high-pitch CTA method investigated here should be performed whenever feasible especially in patients who require frequent follow-up. In addition, high-pitch CTA allows an average contrast medium savings of ml, which may be relevant in patients with decreased renal function [25, 26] and patients with cardiovascular compromise who are at risk for volume overload. Equal CT attenuation values for high-pitch CTA and standard-pitch CTA could potentially have been obtained by further lowering the amount of iodinated contrast medium injected. However, because the noise level is higher in high-pitch CTA due to the reduced radiation dose, suboptimal diagnostic quality could potentially have occurred. Thus, in the absence of further prospective investigations in this field, we do not advocate lowering the amount of iodinated contrast medium even further for high-pitch CTA. In a previous study of abdominopelvic CT, investigators described image quality for highpitch acquisitions in patients weighing more than 90 kg as unacceptable [19]. Nevertheless, in our study, a quarter of the patients who underwent high-pitch CTA had a BMI of greater than 30 kg/m 2 and nearly a third weighed more than 90 kg. This divergence in observations is likely related to the use of more mature current technology in our study. If, however, these obese patients had been excluded from our analysis, the difference in radiation dose would have been even more pronounced. Our study is limited by its retrospective nature. Also, we decided not to perform a subjective image quality evaluation but chose to rely on objective parameters of image quality. However, a radiologist considered all 110 studies of diagnostic quality for their indication. Finally, in this study we did not investigate the effect of more recent iterative image reconstruction techniques, which may improve the image noise characteristics as well as subjective and objective image quality. In conclusion, the use of the high-pitch technique for thoracoabdominal CTA substantially reduces radiation dose and contrast medium volume while maintaining arterial contrast medium attenuation. Given the recent focus on the importance to image wisely, we have since adjusted our clinical practice and we advocate the preferential use of this high-pitch acquisition technique especially in patients who undergo frequent follow-up CT studies for aortic disease surveillance. References 1. Holalkere NS, Matthes K, Kalva SP, et al. 64-Slice multidetector row CT angiography of the abdomen: comparison of low versus high concentration iodinated contrast media in a porcine model. Br J Radiol 2011; 84: Hiatt MD, Fleischmann D, Hellinger JC, Rubin GD. Angiographic imaging of the lower extremities with multidetector CT. Radiol Clin North Am 2005; 43: , ix 3. Napoli A, Fleischmann D, Chan FP, et al. Computed tomography angiography: state-of-the-art imaging using multidetector-row technology. J Comput Assist Tomogr 2004; 28(suppl 1):S32 S45 4. Prokop M. Multislice CT angiography. Eur J Radiol 2000; 36: Achenbach S, Marwan M, Schepis T, et al. Highpitch spiral acquisition: a new scan mode for coronary CT angiography. J Cardiovasc Comput Tomogr 2009; 3: Goetti R, Baumuller S, Feuchtner G, et al. Highpitch dual-source CT angiography of the thoracic and abdominal aorta: is simultaneous coronary artery assessment possible? AJR 2010; 194: Tacelli N, Remy-Jardin M, Flohr T, et al. Dualsource chest CT angiography with high temporal resolution and high pitch modes: evaluation of image quality in 140 patients. Eur Radiol 2010; 20: Karlo C, Leschka S, Goetti RP, et al. High-pitch dual-source CT angiography of the aortic valve aortic root complex without ECG-synchronization. Eur Radiol 2011; 21: Hardie AD, Mayes N, Boulter DJ. Use of highpitch dual-source computed tomography of the abdomen and pelvis to markedly reduce scan time: clinical feasibility study. J Comput Assist Tomogr 2011; 35: Beeres M, Schell B, Mastragelopoulos A, et al. High-pitch dual-source CT angiography of the whole aorta without ECG synchronisation: initial experience. Eur Radiol 2012; 22: Menzel HG, O Sullivan D, Beck P, Bartlett D. European measurements of aircraft crew exposure to cosmic radiation. Health Phys 2000; 79: Macari M, Chandarana H, Schmidt B, et al. Abdominal aortic aneurysm: can the arterial phase at CT evaluation after endovascular repair be eliminated to reduce radiation dose? Radiology 2006; 241: Schindera ST, Graca P, Patak MA, et al. Thoracoabdominal-aortoiliac multidetector-row CT angiography at 80 and 100 kvp: assessment of image quality and radiation dose. Invest Radiol 2009; 44: Utsunomiya D, Oda S, Funama Y, et al. Comparison of standard- and low-tube voltage MDCT angiography in patients with peripheral arterial disease. Eur Radiol 2010; 20: Pontana F, Duhamel A, Pagniez J, et al. Chest computed tomography using iterative reconstruction vs filtered back projection. Part 2. Image quality of low-dose CT examinations in 80 patients. Eur Radiol 2011; 21: Szucs-Farkas Z, Strautz T, Patak MA, et al. Is body weight the most appropriate criterion to select patients eligible for low-dose pulmonary CT angiography? Analysis of objective and subjective image quality at 80 kvp in 100 patients. Eur Radiol 2009; 19: Deak PD, Langner O, Lell M, et al. Effects of adaptive section collimation on patient radiation 1408 AJR:199, December 2012

8 High-Pitch Versus Standard-Pitch CTA of the Aorta dose in multisection spiral CT. Radiology 2009; 252: Fink C, Krissak R, Henzler T, et al. Radiation dose at coronary CT angiography: second-generation dual-source CT versus single-source 64- MDCT and first-generation dual-source CT. AJR 2011; 196:1075; [web]w550 W Sahani D, Saini S, D Souza RV, et al. Comparison between low (3:1) and high (6:1) pitch for routine abdominal/pelvic imaging with multislice computed tomography. J Comput Assist Tomogr 2003; 27: Fuchs T, Kachelriess M, Kalender WA. Technical advances in multi-slice spiral CT. Eur J Radiol 2000; 36: Hidajat N, Mäurer J, Schröder RJ, Wolf M, Vogl T, Felix R. Radiation exposure in spiral computed tomography: dose distribution and dose reduction. Invest Radiol 1999; 34: Hu H, Fox SH. The effect of helical pitch and beam collimation on the lesion contrast and slice profile in helical CT imaging. Med Phys 1996; 23: Preston DL, Ron E, Tokuoka S, et al. Solid cancer incidence in atomic bomb survivors: Radiat Res 2007; 168: Fazel R, Krumholz HM, Wang Y, et al. Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med 2009; 361: Katzberg RW, Haller C. Contrast-induced nephrotoxicity: clinical landscape. Kidney Int Suppl 2006; 100:S3 S7 26. Solomon R, Dumouchel W. Contrast media and nephropathy: findings from systematic analysis and Food and Drug Administration reports of adverse effects. Invest Radiol 2006; 41: FOR YOUR INFORMATION This article is available for CME credit. Log onto click on AJR (in the blue Publications box); click on the article name; add the article to the cart; proceed through the checkout process. AJR:199, December

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax

Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax Application of CARE kv and SAFIRE in Contrast-Enhanced CT Examination on Thorax Poster No.: C-0013 Congress: ECR 2013 Type: Authors: Scientific Exhibit Z. Dejian 1, X. Zhuodong 2, J. Hui 2, L. Xiao 3 ;

More information

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination

Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out (tro) examination Int J Clin Exp Med 2015;8(11):21898-21905 www.ijcem.com /ISSN:1940-5901/IJCEM0015005 Original Article Application of flash dual-source CT at low radiation dose and low contrast medium dose in triple-rule-out

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T. The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume

More information

Original Article. Yohei Marukawa a,b*, Shuhei Sato b, Takashi Tanaka b, Akihiro Tada b, Yuichiro Kanie c, and Susumu Kanazawa b

Original Article. Yohei Marukawa a,b*, Shuhei Sato b, Takashi Tanaka b, Akihiro Tada b, Yuichiro Kanie c, and Susumu Kanazawa b Original Article http ://escholarship.lib.okayama-u.ac.jp/amo/ Evaluating Low-kV Dual-source CT Angiography by High-pitch Spiral Acquisition and Iterative Reconstruction in Pediatric Congenital Heart Disease

More information

Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit detector dual-source computed tomography

Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit detector dual-source computed tomography Acta Radiol OnlineFirst, published on June 11, 2014 as doi:10.1177/0284185114535391 Original Article Image quality and radiation dose of lower extremity CTangiography at 70 kvp on an integrated circuit

More information

CT angiography techniques. Boot camp

CT angiography techniques. Boot camp CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

Update on Acute Aortic Syndrome

Update on Acute Aortic Syndrome SUNDAY Update on Acute Aortic Syndrome Diana Litmanovich, MD Learning objectives To be familiar with the definition, natural history, and imaging findings of acute aortic syndrome, including: I. Aortic

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation

More information

Summary. Rong WangABCE, Xiang-Jiu XuADG, Gang HuangAG, Xing ZhouEF, Wen-Wen ZhangA, Ya-Qiong MaF, Xiao-na ZuoA. Background

Summary. Rong WangABCE, Xiang-Jiu XuADG, Gang HuangAG, Xing ZhouEF, Wen-Wen ZhangA, Ya-Qiong MaF, Xiao-na ZuoA. Background Signature: Pol J Radiol, 2017; 82: 114-119 DOI: 10.12659/PJR.899876 ORIGINAL ARTICLE Received: 2016.06.03 Accepted: 2016.07.21 Published: 2017.02.28 Authors Contribution: A Study Design B Data Collection

More information

Cardiac Computed Tomography

Cardiac Computed Tomography Cardiac Computed Tomography Authored and approved by Koen Nieman Stephan Achenbach Francesca Pugliese Bernard Cosyns Patrizio Lancellotti Anastasia Kitsiou Contents CARDIAC COMPUTED TOMOGRAPHY Page 1.

More information

Dual Energy CT of the Heart: Perfusion and Beyond

Dual Energy CT of the Heart: Perfusion and Beyond Dual Energy CT of the Heart: Perfusion and Beyond U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant

More information

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3

Simon Nepveu 1, Irina Boldeanu 1, Yves Provost 1, Jean Chalaoui 1, Louis-Mathieu Stevens 2,3, Nicolas Noiseux 2,3, Carl Chartrand-Lefebvre 1,3 Coronary Artery Bypass Graft Imaging with CT Angiography and Iterative Reconstruction: Quantitave Evaluation of Radiation Dose Reduction and Image Quality Simon Nepveu 1, Irina Boldeanu 1, Yves Provost

More information

Cardiopulmonary Imaging Original Research

Cardiopulmonary Imaging Original Research Cardiopulmonary Imaging Original Research Bischoff et al. Versus Helical Mode for Coronary CTA Cardiopulmonary Imaging Original Research Bernhard Bischoff 1 Franziska Hein 1 Tanja Meyer 1 Markus Krebs

More information

Cardiac Imaging Tests

Cardiac Imaging Tests Cardiac Imaging Tests http://www.medpagetoday.com/upload/2010/11/15/23347.jpg Standard imaging tests include echocardiography, chest x-ray, CT, MRI, and various radionuclide techniques. Standard CT and

More information

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre Pushing the limits of cardiac CT Steven Dymarkowski Radiology / Medical Imaging Research Centre 5 X 2013 Introduction Rapid technological advances and new clinical applications in cardiovascular imaging

More information

CT Perfusion. U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging

CT Perfusion. U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging CT Perfusion U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant for / research support from Bayer Bracco

More information

Cardiac CT Lowering the Dose Dramatically

Cardiac CT Lowering the Dose Dramatically Cardiac CT Lowering the Dose Dramatically U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant for / research

More information

Dual Energy CT of Pulmonary Embolism

Dual Energy CT of Pulmonary Embolism Dual Energy CT of Pulmonary Embolism U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures Consultant for / research

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

Dual Energy CT Aortography: Can We Reduce Iodine Dose??

Dual Energy CT Aortography: Can We Reduce Iodine Dose?? Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

Ultralow Dose Chest CT with MBIR

Ultralow Dose Chest CT with MBIR Ultralow Dose Chest CT with MBIR Ella A. Kazerooni, M.D. Professor & Director Cardiothoracic Radiology Associate Chair for Clinical Affairs University of Michigan Disclosures Consultant: GE Healthcare

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015

EXPERIMENTAL AND THERAPEUTIC MEDICINE 9: , 2015 2384 Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients GUIRU HE, XIAOPEI LIU, YAN LIU, WEI

More information

Radiation Dose Reduction Strategies in Coronary CT Angiography

Radiation Dose Reduction Strategies in Coronary CT Angiography Radiation Dose Reduction Strategies in Coronary CT Angiography Noor Diyana Osman, PhD noordiyana@usm.my Contents: Introduction Radiation dosimetry in CT Radiation risk associated with coronary CT angiography

More information

MEDICAL UNIVERSITY of SOUTH CAROLINA

MEDICAL UNIVERSITY of SOUTH CAROLINA U. Joseph Schoepf, MD Prof. (h.c.), FAHA, FSCBT-MR, FNASCI, FSCCT Professor of Radiology, Medicine, and Pediatrics Director, Division of Cardiovascular Imaging MEDICAL UNIVERSITY of SOUTH CAROLINA Disclosures

More information

Computed Tomography of the Coronary Arteries

Computed Tomography of the Coronary Arteries Cardiology Update DAVOS 2011 Computed Tomography of the Coronary Arteries Anders Persson M.D., Ph.D Director, Assoc. Professor Center for Medical Image Science and Visualization Linköping University SWEDEN

More information

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Cardiac CT - Coronary Calcium Basics Workshop II (Basic) Cardiac CT - Coronary Calcium Basics Workshop II (Basic) J. Jeffrey Carr, MD, MSCE Dept. of Radiology & Public Health Sciences Wake Forest University School of Medicine Winston-Salem, NC USA No significant

More information

Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich

Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse CH-0 Zurich www.zora.uzh.ch Year: 01 Image quality and radiation dose comparison of prospectively triggered low-dose

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing Cademartiri et al. Heart Rhythm Irregularities on MDCT Angiography Cardiac Imaging Original Research A C M E D E N T U R I C A L I M A G I N G AJR 2006; 186:634 638 0361 803X/06/1863 634 American Roentgen

More information

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Poster No.: C-0863 Congress: ECR 2015 Type: Scientific Exhibit Authors: A. Nakamoto,

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Minimizing CT Radiation Dose CT examinations improve health care and are an essential part

More information

CT Versus MR for the Runoff

CT Versus MR for the Runoff CT Versus MR for the Runoff Robert R. Edelman, M.D. Dept. of Radiology NorthShore University HealthSystem Feinberg School of Medicine, Northwestern University Magnetic Resonance Computed Tomography Radio

More information

Purpose. Methods and Materials

Purpose. Methods and Materials Comparison of iterative and filtered back-projection image reconstruction techniques: evaluation of heavily calcified vessels with coronary CT angiography Poster No.: C-1644 Congress: ECR 2011 Type: Scientific

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction.

A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction. Biomedical Research 2017; 28 (17): 7683-7687 ISSN 0970-938X www.biomedres.info A study on CT pulmonary angiography at low kv and low-concentration contrast medium using iterative reconstruction. Yanhe

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose?

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose? Journal of Physics: Conference Series PAPER OPEN ACCESS Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose? To cite this article: M Smarda et

More information

The radiation dose in retrospective

The radiation dose in retrospective The radiation dose in retrospective gated tdcoronary computed td tomography (CCT) Saeed AL Ahmari, Ghormallah AL Zahrani, Sumiah AL Helali, Samir AL Dulikan, Abdullah Bafagih, HibaKhashojji Prince Sultan

More information

Introduction ORIGINAL ARTICLE

Introduction ORIGINAL ARTICLE Pediatr Radiol (2015) 45:1472 1479 DOI 10.1007/s00247-015-3372-2 ORIGINAL ARTICLE Image quality at low tube voltage (70 kv) and sinogram-affirmed iterative reconstruction for computed tomography in infants

More information

An Introduction to Dual Energy Computed Tomography

An Introduction to Dual Energy Computed Tomography An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the

More information

CT Myocardial Perfusion: Is there Added Value to Coronary CT?

CT Myocardial Perfusion: Is there Added Value to Coronary CT? CT Myocardial Perfusion: Is there Added Value to Coronary CT? U. Joseph Schoepf, MD, FAHA, FSCBT MR, FSCCT Professor of Radiology, Medicine, and Pediatrics Director of Cardiovascular Imaging Disclosures

More information

Low-dose CT coronary angiography in the step-andshoot mode: diagnostic performance

Low-dose CT coronary angiography in the step-andshoot mode: diagnostic performance 1 Institute of Diagnostic Radiology, University Hospital Zurich, Zurich, Switzerland; 2 Clinic for Cardiovascular Surgery, University Hospital Zurich, Zurich, Switzerland; 3 Cardiovascular Centre, University

More information

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School OUTLINE 1 st Presentation: Dose optimization strategies

More information

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center

Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Correlation of Cardiac CTA to Conventional Cardiac Angiography in Diagnosing Coronary Artery Stenosis in a Community Based Center Mathieu Sabbagh, R3 Michigan State University Radiology Garden City Hospital

More information

Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography.

Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography. ISPUB.COM The Internet Journal of Radiology Volume 11 Number 2 Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography. T Niemann, G Bongartz Citation T Niemann, G Bongartz. Detectability

More information

Optimizing radiation dose by varying age at pediatric temporal bone CT

Optimizing radiation dose by varying age at pediatric temporal bone CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Optimizing radiation dose by varying age at pediatric temporal bone CT Daichi Noto, 1 Yoshinori Funama, 2a Mika Kitajima, 3 Daisuke

More information

Cardiac CT Techniques in Neonates (and infants)

Cardiac CT Techniques in Neonates (and infants) Cardiac CT Techniques in Neonates (and infants) Siddharth P. Jadhav, MD Director, Body CT and MRI Edward B. Singleton Department of Pediatric Radiology Texas Children s Hospital Disclosures None Objectives

More information

Low-dose CT coronary angiography: Role of adaptive statistical iterative reconstruction (ASIR)

Low-dose CT coronary angiography: Role of adaptive statistical iterative reconstruction (ASIR) Low-dose CT coronary angiography: Role of adaptive statistical iterative reconstruction (ASIR) Poster No.: 456 Congress: ESCR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit M. A. Glazkova, I. Arkhipova,

More information

Dose levels at coronary CT angiography a comparison of Dual Energy-, Dual Source- and 16-slice CT

Dose levels at coronary CT angiography a comparison of Dual Energy-, Dual Source- and 16-slice CT Eur Radiol (2011) 21:530 537 DOI 10.1007/s00330-010-1954-9 CARDIAC Dose levels at coronary CT angiography a comparison of Dual Energy-, Dual Source- and 16-slice CT J. Matthias Kerl & Ralf W. Bauer & Tobias

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

Exploration the Method of Low Dose Coronary Artery Imaging with Dual-Source CT

Exploration the Method of Low Dose Coronary Artery Imaging with Dual-Source CT Journal of Biosciences and Medicines, 2013, 1, 6-10 doi:10.4236/jbm.2013.11002 Published Online February 2013 (http://www.scirp.org/journal/jbm/) Exploration the Method of Low Dose Coronary Artery Imaging

More information

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design 2 Cardiovascular Computed Tomography: Current and Future Scanning System Design Wm. Guy Weigold Introduction The heart can be visualized in gross form on any standard chest computed tomography (CT), but

More information

Keywords Coronary CT angiography Low dose Iterative reconstruction

Keywords Coronary CT angiography Low dose Iterative reconstruction European Heart Journal Cardiovascular Imaging (2014) 15, 1238 1245 doi:10.1093/ehjci/jeu113 Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction

More information

Radiation Dose Optimization in Cardiac CT: A Technical Review. CHENG Wai Kwong. 17 May Contents

Radiation Dose Optimization in Cardiac CT: A Technical Review. CHENG Wai Kwong. 17 May Contents Radiation Dose Optimization in Cardiac CT: A Technical Review CHENG Wai Kwong 17 May 2014 Contents 1. Introduction and background 2. Cardiac CT synchronization techniques 3. 4. Conclusion 1 Introduction

More information

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial Poster No.: C-0595 Congress: ECR 2013 Type: Authors:

More information

Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography

Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography Review Article Coronary CT angiography with prospective ECG-triggering: an effective alternative to invasive coronary angiography Zhonghua Sun Discipline of Medical Imaging, Department of Imaging and Applied

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA

With the ongoing evolution of ever faster and more. Coronary CTA. Image Acquisition and Interpretation SYMPOSIA SYMPOSIA Image Acquisition and Interpretation Josef Matthias Kerl, MD,*w Lars K. Hofmann, MD,z Christian Thilo, MD,*y Thomas J. Vogl, MD,w Philip Costello, MD,* and U. Joseph Schoepf, MD*y Abstract: Computed

More information

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography

Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography Journal of the American College of Cardiology Vol. 46, No. 3, 2005 2005 by the American College of Cardiology Foundation ISSN 0735-1097/05/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2005.05.056

More information

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications

SYMPOSIA. Coronary CTA. Indications, Patient Selection, and Clinical Implications SYMPOSIA Indications, Patient Selection, and Clinical Implications Christian Thilo, MD,* Mark Auler, MD,* Peter Zwerner, MD,w Philip Costello, MD,* and U. Joseph Schoepf, MD* Abstract: Recent technical

More information

Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience

Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience Low-dose prospective ECG-triggering dual-source CT angiography in infants and children with complex congenital heart disease: first experience Ximing Wang, M.D., Zhaoping Cheng, M.D., Dawei Wu, M.D., Lebin

More information

Vascular CT Protocols

Vascular CT Protocols Vascular CT Protocols V 1D: Chest and abdominal CT angiogram (aortic dissection protocol) V 1T: Chest CT angiogram (aortic trauma protocol) V 2: Abdominal and pelvis CT angiogram (aortic aneurysm protocol)

More information

PERIPHERAL CTA. Richard L. Hallett, MD

PERIPHERAL CTA. Richard L. Hallett, MD RC 812B Lakeside E351 1 December 2017 0830 1000 Richard L. Hallett, MD Chief, Cardiovascular Imaging Northwest Radiology Network Indianapolis, IN Adjunct Assistant Professor Radiology Cardiovascular Imaging

More information

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT

Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Impact of SSF on diagnostic performance of coronary CT angiography within one heart beat in patients with high heart rate using a 256-row detector CT Junfu Liang 1,2, Hui Wang 1, Lei Xu 1, Li Dong 1, Zhanming

More information

Gender differences in CT calcium scoring: A phantom study

Gender differences in CT calcium scoring: A phantom study Gender differences in CT calcium scoring: A phantom study Nicholas Petrick, Qin Li, Benjamin Berman, Marios A Gavrielides, Rongping Zeng, Berkman Sahiner CDRH/OSEL/DIDSR U.S. Food and Drug Administration

More information

Performance of dual-source CT with high pitch spiral mode for coronary stent patency compared with invasive coronary angiography

Performance of dual-source CT with high pitch spiral mode for coronary stent patency compared with invasive coronary angiography Journal of Geriatric Cardiology (2016) 13: 817 823 2016 JGC All rights reserved; www.jgc301.com Research Article Open Access Performance of dual-source CT with high pitch spiral mode for coronary stent

More information

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16.

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16. NIH Public Access Author Manuscript Published in final edited form as: Acad Radiol. 2008 January ; 15(1): 40 48. doi:10.1016/j.acra.2007.07.023. Analysis of Heart Rate and Heart Rate Variation During Cardiac

More information

(Non-EKG Gated) CTA Thoracic Aorta = CTA Chest

(Non-EKG Gated) CTA Thoracic Aorta = CTA Chest (Non-EKG Gated) CTA Thoracic Aorta = CTA Chest Reviewed By: Dan Verdini, MD, Rachael Edwards, MD Last Reviewed: January 2019 Contact: (866) 761-4200, Option 1 In accordance with the ALARA principle, TRA

More information

Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source coronary CT angiography

Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source coronary CT angiography www.nature.com/scientificreports Received: 1 March 2017 Accepted: 27 June 2017 Published online: 31 July 2017 OPEN Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Imaging Strategies for Endovascular Cardiovascular Procedures and Percutaneous Aortic Valves. Roy K Greenberg, MD

Imaging Strategies for Endovascular Cardiovascular Procedures and Percutaneous Aortic Valves. Roy K Greenberg, MD Imaging Strategies for Endovascular Cardiovascular Procedures and Percutaneous Aortic Valves Roy K Greenberg, MD Disclosure Research support Cook Inc, Boston Scientific, W.L.Gore, Cordis, Vascutek, Terarecon

More information

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation Poster No.: C-2014 Congress: ECR 2014 Type: Educational Exhibit Authors: M. A. Ottesen; Oslo/NO Keywords: Cardiac, Arteries / Aorta, CT, CT-Angiography,

More information

Missed Pulmonary Embolism on Abdominal CT

Missed Pulmonary Embolism on Abdominal CT Cardiopulmonary Imaging Original Research Lim et al. Pulmonary Emboli Missed on CT of the Abdomen Cardiopulmonary Imaging Original Research Kun Young Lim 1 Seth J. Kligerman 2 Cheng Ting Lin 2 Charles

More information

Radiation dose of cardiac CT what is the evidence?

Radiation dose of cardiac CT what is the evidence? Eur Radiol (2009) 19: 1311 1315 DOI 10.1007/s00330-009-1312-y CARDIAC Hatem Alkadhi Radiation dose of cardiac CT what is the evidence? Received: 10 December 2008 Revised: 2 January 2009 Accepted: 12 January

More information

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling Investigations and research Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling J. Wang Z. Ying V. Yao L. Ciancibello S. Premraj S. Pohlman

More information

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Int J Cardiovasc Imaging (2008) 24:547 556 DOI 10.1007/s10554-007-9282-0 ORIGINAL PAPER Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study Jaap M. Groen Æ Marcel J.

More information

Academy of Medical Sciences, Guangdong General Hospital, Guangzhou, P.R. China. of multidetector computed tomography

Academy of Medical Sciences, Guangdong General Hospital, Guangzhou, P.R. China. of multidetector computed tomography Original Research Hellenic J Cardiol 2014; 55: 393-401 Potentially Optimal Body Size to Adjust Tube Current for Individualized Radiation Dose Control in Retrospective ECG-Triggered 256-Slice CT Coronary

More information

Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT

Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT Eur Radiol (2017) 27:3904 3912 DOI.07/s00330-017-472-9 CARDIAC Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT Gert Jan Pelgrim 1 & Robbert W.

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Angio-CT: heart and coronary arteries

Angio-CT: heart and coronary arteries European Journal of Radiology 45 (2003) S32/S36 www.elsevier.com/locate/ejrad Angio-CT: heart and coronary arteries Andreas F. Kopp * Tübingen University Hospital, Tübingen, Germany Received 22 November

More information

Lung Perfusion Analysis New Pathways in Lung Imaging. Case Study Brochure PLA 309 Hospital

Lung Perfusion Analysis New Pathways in Lung Imaging. Case Study Brochure PLA 309 Hospital Lung Perfusion Analysis New Pathways in Lung Imaging Case Study Brochure PLA 309 Hospital http://www.toshibamedicalsystems.com Toshiba Medical Systems Corporation 2012 all rights reserved. Design and specifications

More information

Computed tomography (CT) and magnetic resonance

Computed tomography (CT) and magnetic resonance Trends in Emergency ortic Imaging Determining optimal imaging techniques for each patient. Y PERRY CHOI, MD, ND HMID MOJIIN, MD Computed tomography (CT) and magnetic resonance imaging (MRI) play vital

More information

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could

More information

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Poster No.: B-0742 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Paper

More information

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography

Radiation Dose Reduction and Coronary Assessability of Prospective Electrocardiogram-Gated Computed Tomography Coronary Angiography Journal of the American College of Cardiology Vol. 52, No. 18, 2008 2008 by the American College of Cardiology Foundation ISSN 0735-1097/08/$34.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.07.048

More information

CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings

CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings CT angiography of pulmonary arteries to detect pulmonary embolism with low kv settings Poster No.: C-3289 Congress: ECR 2010 Type: Scientific Exhibit Topic: Chest - Your latest results Authors: M. K. Gill,

More information